MULTILINE NEXT DC
Item No.: 4225502
Last updated: 15.07.2019 / Rev.: 2
Copyright: © 2019 GREINER GmbH
All copyrights to this instruction manual remain with GREINER GmbH. Reprinting - even in extracts - is not permitted. The transfer of this manual to third parties requires the written permission of GREINER GmbH. Any use or modification outside the narrow limits of copyright law is not permitted without the consent of GREINER GmbH.
This applies in particular to duplication, translation, microfilming and storage as well as processing in electronic systems.
2 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
Table of contents
Table of contents
Note on pictorial representations
Safety instructions for commissioning
Attaching the leg section drive
Pushing the couch off the transport pallet
Disposal of the packaging material
Requirements on the place of operation
Check before commissioning and reassignment
Safety instructions for operation
Unlocking and locking the hand control
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC
3 / 72
Table of contents
Castors with central locking and directional control
Sitting down and standing up of the patients
Infusion rod holder and infusion rod
Neck roll / Relax pillow in premium material
Arm supports in premium material
Legrest upholstery protection cover
Push handle with paper roll holder
Return delivery / repair order
4 / 72 Instruction manual - MULTILINE NEXT DC
© 2019 GREINER GmbH
1 Introduction
1 Introduction
With this instruction manual (hereinafter also referred to as the "manual") we would like to provide the user and operator with useful information for safe and proper operation of the MULTILINE NEXT DC couch (hereinafter also referred to as the "couch").
This manual also describes functions or features that may not be included in your couch.
Observance of the instructions helps to
avoid risks.
reduce downtime. reduce ongoing operating costs.
increase the reliability and service life of the couch.
We reserve the right to make technical changes without notice within the scope of further development of the couch described in this manual.
The original of this manual was written in German. All other languages are translations of this manual.
1.1
Validity
This manual applies only to the couch and accessories supplied by GREINER GmbH (also abbreviated to
GREINER).
1.2
Manufacturer
GREINER GmbH
Wettestraße 1
74385 Pleidelsheim / Germany
Phone: +49-7144-8112-0
Fax: +49-7144-8112-99 [email protected] www.greiner-gmbh.de
1.3
Note on pictorial representations
The photos / illustrations are general illustrations and may differ from the actual conditions or from the components actually used.
1.4
Warranty and liability
Our General Terms and Conditions (GTC) apply in the currently valid version at www.greiner-gmbh.de.
In the event of a complaint (damage, defects or other reasons for complaint), your authorised specialist distributor is the appropriate contact person. Please provide the data of the type plate (and possibly a photo) (see
The more accurate and better the data, the better and more targeted the remedy can be provided.
NOTE This couch is not approved for the North American market, especially the United States of America. The distribution and use of the couch in these markets, also through third parties, is prohibited by the manufacturer.
1.5
Service manual
A service manual can be provided on request. This supports the service technicians of the authorised specialist distributors or other qualified personnel commissioned by the operator in the maintenance or repair of the couch.
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 5 / 72
1 Introduction
1.6
Type plate
The type plate of the couch is located (depending on the version) either at the rear of the plastic cover of the lower part or on the outside of the bearing (not shown).
It contains the following information (see also 9.4 Technical data, if necessary).
Symbol Meaning Symbol Meaning
Manufacturer Product name
Cushion Cushion colour or type Serial number
Customer
V in:
I in max.
IPX4
CE marking
Follow the instructions
Customer number or name
Input voltage/frequency
Input current
IP rating
Indoor application
Thermal fuse element
No domestic waste
U out:
Op:
Warning
Date of manufacture
Safe working load
Alternating current
Output voltage
Duty cycle
Protection class I
Applied Part: Type B
Isolating transformer
1
Fig. 1 Type plate - plastic cover
Further type plates can be found on all electrical components.
1.
Type plate (example)
Type plates are documents that may not be altered or removed.
Damaged or lost type plates must be replaced true to the original.
6 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
2 Safety
2 Safety
The main objective of the safety instructions is to prevent injury to persons.
In addition, observation of the safety instructions helps to avoid material damage.
As the operator you have purchased a MULTILINE NEXT DC couch from GREINER GmbH. As a result, you are also responsible for its proper and intended operation.
2.1
Signal words
In order to point out dangers, prohibitions and important information, the following signal words and symbols are used in this manual. These must be read and strictly observed.
DANGER This signal word indicates an imminent danger resulting in serious injuries, and in some cases
death.
WARNING This signal word indicates a potential imminent danger, which can result in serious injuries, and in some cases death.
CAUTION This signal word indicates a potential imminent danger, which can result in light to serious
injuries.
NOTE This signal word indicates a potential imminent danger, which can result in damage to prop-
erty and the environment.
TIP Indicates tips for use or other particularly important information when handling the couch.
2.2
Device safety
Greiner couches were built in accordance with the applicable national and international standards and regulations according to the current state of the art and tested by an independent body.
This couch meets the requirements for safety and functionality. It bears the CE mark, which documents compliance with the basic requirements for medical devices.
The couch described in this instruction was designed and manufactured in compliance with the EC Directive 93/42
EEC (MDD - Medical Device Directive) and meets its requirements. According to the Medical Devices Act (MPG), it is an active medical device of risk class I.
The validity of the declaration of conformity expires if changes are made to the couch by the customer or third parties, e.g. modifications of any kind, use of external accessories, changes to the software, removal of warning and information signs (no claim to completeness).
Due to the chosen robust, interference-proof design of the electrical adjustment devices, the couch is largely tolerant to EMC interference from other electrical equipment operated in the vicinity.
Despite comprehensive EMC testing, it is possible that other devices may be disturbed by the couch.
In such cases, either switch off the couch completely by disconnecting the mains plug or disconnect it from the mains, or, if necessary, keep a larger distance from the disturbed device and change the orientation to each other. If possible, do not use the same outlet.
This applies in particular to the simultaneous use of highly sensitive measuring instruments. Their measurement results may be influenced under certain circumstances.
The couch can be equipped with accessories on delivery (see 5 Accessories).
If necessary, observe the instruction manuals for the accessories. For further information, please contact your authorised specialist distributor.
All prohibition, warning and message symbols or instructions on the couch must be observed. The symbols and notes must always be legible and complete. Damaged or lost symbols or notes must be replaced true to the original.
The couch is mainly made of steel tubes or steel profiles. The surfaces are either powder-coated or coated with zinc or chrome.
All surfaces are harmless against skin contact.
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 7 / 72
2 Safety
2.2.1
Safety inspection
Legal foundations
Operators of electrically driven couches for medical use are, in accordance with
EC Medical Device Directive 93/42 EEC (MDD) and the resulting national laws/regulations (e.g. in Germany)
German Medical Device Operator Ordinance (MPBetreibV) § 4 (repair)
DGUV Information 203-071 (periodic inspections of mobile electrical work equipment).
DGUV Regulation 3 Accident Prevention Regulation (electrical systems and equipment). obliged to maintain the safe condition of medical devices throughout the entire period of use. This also includes regular professional maintenance and regular safety checks.
In other countries, outside Germany or the EU, the applicable national regulations must be observed.
GREINER GmbH generally recommends an annual maintenance according to the maintenance table as well as a safety inspection with simultaneous electrical inspection according to IEC 62353 (see service manual).
2.3
Personal safety
To avoid errors and to ensure trouble-free operation of the couch, the safety instructions in this instruction manual must have been read and understood completely and must always be made available to the user at all times
(in printed or electronic form).
Basic instruction for the user can be given by GREINER GmbH or its authorised specialist distributor at the request of the operator.
Patients must be instructed in the use of the couch before activating the hand control.
In other countries, outside Germany or the EU, the applicable national regulations and laws must be observed. In
Germany, the provisions of the Medical Devices Act (MPG), the Medical Devices Operator Ordinance (MPBetreibV) and the relevant statutory regulations must be observed.
Notes to the operator
This couch meets all requirements of the EC directive 93/42 EEC for medical devices. It is classified, designed and manufactured as an active class I medical device in accordance with the Medical Devices Act (MPG).
Observe your obligations as operator according to the Medical Device Operator Ordinance
(MPBetreibV) to ensure a permanently safe operation of this medical device without endangering patients, users and third parties.
Notes to the user
In accordance with MPBetreibV, the user must ensure that the couch is functional and in proper condition before using it in medical areas and must observe the Instruction manual.
The same applies to any accessories that have been attached to the couch.
Notes to the patients
With each new assignment, the patient must be instructed in the important functions of the couch.
Instruction is usually given by the user.
8 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
2 Safety
2.4
Intended use
The MULTILINE NEXT DC couch from GREINER GmbH is a medical couch for use by patients in health facilities such as hospitals, clinics, medical practices and dialysis centres. It is intended exclusively for use inside buildings and under normal environmental conditions.
The MULTILINE NEXT DC couch was specially developed for use in dialysis treatment, e.g. in a dialysis centre.
The safe working load of the couch is 220 kg.
The couch is used mainly as a place for sitting or lying while receiving a dialysis treatment.
The couch allows the convenient and comfortable positioning of the patient during dialysis therapy.
The user is supported in his or her work with regard to ergonomics and quality.
As a treatment couch for sitting and lying treatments or examinations.
The couch can be used for patient transport under certain conditions within buildings.
The intended users are trained and instructed medical personnel of the relevant department
(e.g.: nurses, carers, doctors, assistants, etc.).
Installation and commissioning by technical personnel of the operator or authorised specialist distributors, if necessary.
2.4.1
Reasonably foreseeable misuse
Reasonably foreseeable misuse can lead to hazards.
Some examples of "reasonably foreseeable misuse" are given below.
(no claim to completeness)
Use in the vicinity of high-frequency surgical units, MRI units or defibrillators.
Use in an environment where flammable or explosive gases or vapours (e.g. anaesthetics) are to be expected.
Use as a work chair.
Use in the operating theatre.
Use as operating chair or operating table.
Use as a spare hospital bed or spare bed.
Overloading of the couch above the specified safe working load.
Operation of the couch by patients without prior instruction.
Improper operation of electrical functions and uncontrolled positioning.
Outdoor use.
Use in wet areas, e.g. bathing establishments.
Use as a climbing aid (ladder).
Use as a children's toy, e.g. gymnastics or sports equipment.
Use as a vehicle or for the transport of goods.
Try to move the couch although the castors are in the braked position.
Pull on cables to move the couch.
Disconnect electrical connectors by pulling on the cable.
Cleaning in washing facility.
Cleaning with a spray lance of a high-pressure cleaner or water jet.
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 9 / 72
3 Commissioning
3 Commissioning
The chapter commissioning describes the preparation of the couch for use including a final check of the functional
safety (see 3.7 Check before commissioning and reassignment).
TIP The use of specially trained and instructed service personnel during commissioning is neither necessary nor intended.
3.1
Safety instructions for commissioning
Before commissioning the couch, the user must be instructed in the handling of the couch using the instruction manual, having previously read it in detail. In addition, the potential dangers that may occur despite proper operation of the couch must be pointed out in detail.
Operator's obligation - instruction
The operator must observe the respective valid national specifications and legal regulations.
The operator must ensure that the users are instructed before using the couch.
The instruction can be given either by an authorised person of an authorised specialist distributor or directly by the manufacturer.
TIP If the delegated person of the operator, e.g. the person responsible for medical devices, has been instructed by an authorised person of an authorised specialist distributor or by the manufacturer, we would like to point out that we authorise the person responsible for medical devices of the operator within the scope of the purpose for the instruction of the couch on the basis of training, experience and knowledge.
The operator must ensure that the instruction manual has been read and understood completely by the users. The operator must disclose the place where the instruction manual is stored and make it accessible at all times.
The operator must ensure that representative personnel also receive adequate instruction in the operation of the couch.
The operator must ensure that the couch is operated exclusively by trained users.
The operator must inform the users that the patients may only operate the couch after their instruction, if necessary only under supervision.
TIP Electrical testing before commissioning.
A further electrical inspection of the couch before the first commissioning is not necessary, as the couch has been tested for electrical safety in accordance with EN 60601 (test report in instruction manual) and has left our factory in perfect condition.
In other countries, outside Germany or the EU, the applicable national regulations and laws must be observed. Additional commissioning specifications of the operator must always be observed. In individual cases, a new electrical inspection must be carried out and documented in accordance with IEC 62353 (see service manual).
DANGER Danger of suffocation from packaging material.
Packaging material is not a toy and must be kept away from babies and small children.
In particular, do not pull the plastic bags or sacks over your head or crawl into them.
3.2
Installation requirements
The couch is only approved for use in buildings and under normal ambient conditions, or for use in dry rooms
The following environmental conditions must be observed on site.
Operating temperature [C]: + 40
Air humidity [rH] (non-condensing):
Air pressure [hPa]:
Operation at estimated altitude [m]:
+ 5
20 % - 90 %
700 - 1060
≤ 2000
10 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
3 Commissioning
3.3
Scope of delivery
Configuration options
There are basically 3 different versions of the MULTILINE NEXT DC couches. They differ in the number of electrical adjustment options.
In addition, the type of brake system installed for the castors determines the principle variant.
The MULTILINE NEXT DC couch therefore has 8 configuration options.
The couch can be equipped with accessories (see 5 Accessories).
Actuators Castors
MULTI-
LINE NEXT DC
4223000
4223100
4224000
4224100
4224200
4225000
4225100
4225200
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 11 / 72
3 Commissioning
3.4
Packaging
The MULTILINE NEXT DC couch is packed and shipped as shown below.
TIP Carry out a visual check of the outer packaging immediately after delivery and before unpacking.
Make a written record of any major obvious damage directly on the carrier's delivery note [A] [B].
Later complaints are facilitated by the immediate creation of meaningful photos, if possible in the presence of the delivery driver.
A
B
Fig. 2 Shipment packaging
3.5
Preparation
The preparation section describes and shows how to unpack the couch and make it ready for operation using the example of a fully equipped MULTILINE NEXT DC with 5 adjustment drives and central locking of the castors
(4225100) [A] [B].
NOTE The procedure and sequence of the simple work steps must be strictly observed, otherwise personal injury may occur.
The couch or parts of the couch will be severely damaged in any case if the following activities are not carried out correctly.
A B
Fig. 3 Shipping position
12 / 72
Fig. 4 Operating position
Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
3 Commissioning
3.5.1
Removing the packaging
Personnel requirements
1 User
Required tools and aids
Scissors
Way of proceeding
A
1
B
2
Fig. 5 Removing the tensioning straps
1.
Use scissors to cut through the tensioning straps (1) and then remove them [A] [B].
C D
4
3
Fig. 6 Removing the packaging material
2.
Remove the cardboard lid (2), the outer packaging (3) and the protective film (4) upwards [B] [C] [D].
NOTE Do not cut the packaging with a knife or long blade.
The upholstery material or other parts of the couch could be damaged.
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 13 / 72
3 Commissioning
3.5.2
Preparation for assembly
Personnel requirements
1 User
Required tools and aids
Scissors
Way of proceeding
A
1
D
B
2
4
Fig. 7 Removing the seat upholstery, the manual and accessories
1.
Remove the seat upholstery (1) and the manual (2) [A] [B].
2.
3.
Manual with electrical test (3) (test report) [C].
Remove the infusion rod (4) if necessary [D]
E
G
C 3
F
7
5
6
8
Fig. 8 Removing the cable ties / removing the wing screws
4.
Cut through the cable tie (5) for securing the leg section drive (6) during transportation [F].
5.
Cut through the cable tie (7) on the wing screws (8) for fastening the seat upholstery. Remove the wing screws [F]. Do not lose the wing screws, as both are needed later again [G].
14 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
3 Commissioning
3.5.3
Releasing the transport lock
Personnel requirements
1 User
Required tools and aids
None
Way of proceeding
A
C
B
3
2
1
Fig. 9 Pulling out the axle of the transport lock
1.
Remove the circlip (1) on the axle (2) of the transport lock [B] [C]. Do not lose the circlip and the axle, as both are needed later again.
2.
Release the axle when pulling it out. To do this, move the leg section (3) slightly and hold it firmly
[A].
CAUTION Risk of trapping and crushing.
Use your hand to guide the leg section when lowering. Hold the leg section while lowering and do not let it fall freely.
D
4
Fig. 10 Folding down the cross bar
3.
Fold down the cross bar (4) vertically [D].
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 15 / 72
3 Commissioning
3.5.4
Attaching the leg section drive
Personnel requirements
1 User
Required tools and aids
None
Way of proceeding
A
C
B
1
Fig. 11 Pulling out the axle on the leg section drive
1.
Remove the circlip (1) on the axle (2) of the leg section drive (3) [A] [B] [C].
D
E
4
3
2
Fig. 12 Attaching the leg section drive
2.
Join leg section (4) and leg section drive [D].
3.
Insert the axle and reinstall the circlip.
The circlips on the axles must re-engage correctly [E].
CAUTION Risk of trapping and crushing.
The axle may slip out and the leg section may then drop unexpectedly.
16 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
3 Commissioning
3.5.5
Electrical commissioning
Personnel requirements
1 User
Required tools and aids
None
Operating statuses
When the couch is connected to the mains and the hand control is disabled, the yellow operation control LED (1) lights up.
When the couch is connected to the mains and the hand control is enabled with the magnetic key, the green operation control LED (2) lights up to indicate that the couch is ready for operation.
Way of proceeding
A
1
B
2
4
5
3
Fig. 13 Electrical commissioning
1.
Insert the mains plug into a socket (without Fig.).
Unlocking the hand control
2.
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 17 / 72
3 Commissioning
3.5.6
Attaching the footrest
Personnel requirements
1 User
Required tools and aids
None
Prearrangement
TIP Bring the couch into the highest sitting position and the backrest into a flat position. Swivel the
seat all the way up and the leg section all the way down (see 4.3 Hand control).
Way of proceeding
A
Fig. 14 Couch position for attaching the footrest
1.
Bring the couch into the shown position [A].
B
3
D
5
4
C
1
2
Fig. 15 Aligning the cross bar and the rocker lever
2.
Position the cross bar (1) and the rocker lever (2) so that both can be connected to the axle [C].
Move the footrest (3) manually until the correct position is reached [B].
NOTE Ensure that the cables cannot be squeezed or jammed.
3.
Insert the axle (4) and reinstall the circlip (5).
The circlips on the axles must re-engage correctly [D].
CAUTION Risk of trapping and crushing.
The axle may slip out and the footrest may then drop unexpectedly.
18 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
3
E
Commissioning
2
1
Fig. 16 Checking the position after assembly
4.
Check the correct assembly of the cross bar (1) and the rocker lever (2).
After assembly, both components must be arranged as shown [E].
F
Fig. 17 Checking the laying of the cables
5.
Checking the laying of the cables.
The cables must not be trapped or crushed [F].
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 19 / 72
3 Commissioning
3.5.7
Attaching the seat upholstery
Personnel requirements
1 User
Required tools and aids
None
Prearrangement
TIP Bring the couch into the highest sitting position and the backrest into a flat position. Swivel the
seat all the way up and the leg section all the way down (see 4.3 Hand control).
Swivel the armrests upwards and bring them into the widest position (see 4.9 Armrests).
Way of proceeding
A
Fig. 18 Couch position for attaching the seat upholstery
1.
Bring the couch into the shown position [A].
B
1
3
2
Fig. 19 Align the seat upholstery centrically and fasten it
2.
Place seat (1) on backrest upholstery (2) and align centrically with the backrest upholstery [B].
3.
Align the Velcro connection (3) between the seat and backrest upholstery and attach [B].
20 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
3
C
Commissioning
4
5
Fig. 20 Align the mounting bolts
4.
Align the mounting bolts (4) with the recesses (5) [C].
D
Fig. 21 Hook in the seat upholstery
5.
Slide the seat upholstery backwards to hook the mounting bolts into the recesses. Simultaneously press down the seat upholstery in the rear part [D].
6.
Ensure correct fitting of the mounting bolts by lifting the seat upholstery [D].
E
Fig. 22 Check the Velcro connection
7.
Check that the Velcro connection is centred and wrinkle-free [E].
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 21 / 72
3
F
Commissioning
6
Fig. 23 Tighten the wing screws
8.
Screw in and tighten the wing screws (6) [F].
G
H
Fig. 24 Check the seat upholstery assembly
9.
Check correct assembly of the seat upholstery [G].
10.
Bring the couch into the basic position (see 4.3.2 Basic position) [H].
11.
Swivel armrests down and bring them into the narrowest position (see 4.9 Armrests) [H].
12.
Lock the hand control and hang it on the brackets at the front end of the arm supports (see
4.3.1 Unlocking and locking the hand control) [H].
22 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
3 Commissioning
3.5.8
Pushing the couch off the transport pallet
Personnel requirements
1 User
Required tools and aids
Cutter-knife
Way of proceeding
A
B
2
1
Fig. 25 Attaching the mains cable
1.
Remove the mains plug (1) from the socket and attach it to the mains cable holder (2) [A] [B].
C
Fig. 26 Pushing the couch off the transport pallet
2.
Cut and fold down the corners of the cardboard [C].
3.
4.
Release the brake (see 4.12 Braking). Carefully push the couch forward off the transport pallet.
For this purpose, use aids, e.g. cardboard packaging, to make a ramp.
Carry out a functional check (see 3.7 Check before commissioning and reassignment).
3.5.9
Disposal of the packaging material
The packaging material must be separated according to substance groups and disposed of properly in accordance with national regulations. If you have any queries, please contact your operator, your local municipalities or
waste disposal companies (see 8.4 Disposal).
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 23 / 72
3 Commissioning
3.6
Requirements on the place of operation
For the operation of the couch, an appropriate mains supply and, if applicable, an equipotential bonding connection is required at the place of use.
If this is available and the building's wiring permits it, the couch must always be connected to the equipotential
bonding system (see 4.7 Equipotential bonding pin).
Battery
TIP
Charge the battery for at least 24 hours before first use (see 4.5 Battery mode).
The battery is then available for emergency operation without restriction.
3.7
Check before commissioning and reassignment
After the commissioning work has been completed, and before any occupation by a new patient, the couch must
be cleaned and disinfected (see 6 Cleaning and disinfection).
CAUTION Risk of infection from contaminated couch.
The couch must be cleaned and then disinfected before the first use.
This also applies before each use (occupation by a new patient).
A functional check of the couch must be carried out immediately after cleaning and disinfection and before use.
WARNING Risk of injury for patient, user and third parties.
The couch must be checked for functional safety after first commissioning and every occupation by a new patient.
It must be ensured that the couch can be used as intended without endangering the patient, the user or third parties.
If the couch is put into operation for the first time or before any occupation by a new patient, the same functional
General check of the overall condition of the couch for soiling, condition and damage, completeness and legibility of stickers, symbols and instructions, clean if necessary in accordance with the manufacturer's specifications or those of the operator and repair if necessary.
Visually inspect the mains cable and plug, hand control, as well as all other electrical cables and connectors for breakage, proper installation (e.g. strain relief), clamping or friction points etc. and repair if necessary.
Functional check of the electrical adjustment devices and the control system. All electric actuators must be operated once up to their end positions. Pay particular attention to changed operating behaviour, unusual noises, speed, smooth running, odours and increased temperature and repair if necessary.
Carry out a functional check according to the installed brake system for the castors (locking, freewheel) and repair if necessary.
Any defects or damage found must be remedied immediately.
The couch must not be used before the defects or damage have been remedied.
WARNING Risk of injury due to defective couch.
The couch must not be operated in a defective or faulty condition in which the couch could endanger patients, users or third parties.
A defective or faulty couch must be clearly marked "DEFECT". The marking must be done in such a way that the defective condition is clearly identifiable for everyone.
Inform the operator for intensive inspection during repair.
24 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
4 Operation
4 Operation
The chapter operation describes the functions of the couch and its performance features.
Prerequisites for operation
The couch may only be operated and used in accordance with its intended use, the generally recognised rules of technology and the national occupational safety and accident prevention regulations valid at the place of use (see
In Germany, the provisions of the Medical Devices Act (MPG), the Medical Devices Operator Ordinance
(MPBetreibV) and the relevant statutory regulations must be observed.
To ensure safe operation, the following points must be strictly observed.
Before using the couch, the user must familiarise him- or herself with the contents of this manual and observe the safety instructions for the individual points of danger.
The couch must be cleaned first and then disinfected (see 6 Cleaning and disinfection) before com-
missioning and before any further use (occupation by a new patient). A functional check must then
be carried out afterwards (see 3.7 Check before commissioning and reassignment).
The couch does not have an EMERGENCY STOP command device. Access to the mains plug must therefore be guaranteed at all times in order to allow the deactivation of the couch in an emergency situation by disconnecting the mains plug.
Avoid mechanical stress on the mains cable or plug. Do not crush or trap the mains cable anywhere. Carry out an immediate visual check after each mechanical load on the mains cable or plug.
Tighten all handwheels, clamping levers and clamping screws etc., including those on accessories, before use.
Due to the widely differing stresses, operating conditions and frequencies at the place of use, the setting by the manufacturer of a central specification for the frequency of safety inspections does not make sense. In accordance with the Medical Device Operator Ordinance (MPBetreibV), the operator must specify the test period for the safety inspection on the basis of a risk assessment (see
4.1
Safety instructions for operation
When adjusting the couch, ensure that there are no limbs of the patient, user or other persons, especially children playing between the lower and upper sections, as well as between the back or armrest(s) or the leg section and the floor, which could be trapped and injured. Pets should generally be kept away from the couch and no other objects should be stored under the couch. Ensure that the couch does not get caught on walls or furniture.
DANGER Risk of injury due to unintentional movements of the couch.
The couch is not intended for unsupervised use by children or handicapped persons. In such situations, always ensure that the electrical adjustment functions are disabled as long as there is no trained user on site.
Keep the magnetic key separate from the hand control at all times.
(see 4.3.1 Unlocking and locking the hand control)
The couch may only be occupied by one person at a time.
The safe working load must not be exceeded and must be evenly distributed on the lying surface.
Do not sit or rest on the backrest upholstery, the legrest upholstery, the arm support or the footrest.
CAUTION Risk of falling or injury if lying position is set too high.
In order to avoid or reduce injuries caused by falling, we recommend (except when carrying out care measures on the patient) moving the couch to the lowest position. This generally applies
when moving the couch (see 4.11 Driving) and in particular when transporting and sitting down
When operating other devices on or near the couch, that are equipped with cables, air hoses or similar, make sure that these lines cannot get trapped in the moving parts of the couch or be damaged in any other way.
WARNING Risk of injury due to the use of unsuitable accessories.
Only original accessories from GREINER GmbH may be used.
The attachment of accessories from other manufacturers (e.g. fixing belts etc.) lies within the responsibility and duty of care of the operator.
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 25 / 72
4 Operation
4.2
Structure
The structure of the couch is shown here using the example of a fully equipped MULTILINE NEXT DC with 5 adjustment drives and central locking of all 4 castors (4225100).
7 8
9 10
6
11
5
4
3
12
2
1
17
20
19 18
Fig. 27 Structure of the couch
1.
Footrest
2.
Legrest upholstery
3.
Hand control
4.
Seat upholstery
5.
Backrest upholstery
6.
Relax pillow (accessory)
7.
LED reading lamp (accessory)
8.
Mains cable (not shown)
9.
Push handle with paper roll holder (accessory)
10.
Infusion rod (accessory)
26 / 72
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
Integral foam arm support
Armrest
Equipotential bonding pin (not shown)
Standard rails (accessory)
Infusion rod holder (accessory)
Foot switch - both sides (accessory)
Central locking - both sides
Castor
Footrest protection cover (accessory)
Legrest upholstery protection cover (accessory)
Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
13
14
15
16
4 Operation
4.3
Hand control
Depending on the electrical version, there are three different hand controls. The hand controls enable the user or patient to make continuous adjustments using electric drives.
Patients must be instructed in the functions of the hand control in any case.
When adjusting the couch, the patient's forearms must be in the normal position on the arm supports.
General information on operating the hand control.
The drives run as long as the buttons are pressed.
Each adjustment stops if more than one button is pressed simultaneously.
Except for the button combinations for Relax and Bed position (see 4.3.3 Relax position and
4.3.4 Bed position (flat position)).
Always hang the hand control on the brackets at the front end of the arm supports. In this way, the patient cannot accidentally sit or lie down on the hand control and thus trigger unwanted movements of the couch.
The spiral cable provides plenty of freedom for movement.
Operation
Backrest
Seat inclination
Leg section up (3) - down (4) up (5) - down (6) up (7) - down (8)
Seat height
Footrest
Basic position up (9) - down (10) retract (11) - extend (12)
(13)
Shock position
Relax
(14)
(5+6)
Bed (7+8)
Both the yellow and red buttons (13) + (14) for basic position and shock position (Trendelenburg position) and
tion, 4.3.5 Shock position, 4.3.3 Relax position and 4.3.4 Bed position (flat position)).
1 2 1 2 1 2
3
5
7
4
6
8
3
5
7
4
6
8
3
5
7
4
6
8
9 10 9
11
10
12
13 14
Fig. 28 Hand control (3 drives)
1.
Yellow operation control LED
(hand control locked)
2.
Green operation control LED
(hand control unlocked)
3.
Backrest - up
4.
Backrest - down
5.
Seat inclination - up
6.
Seat inclination - down
13
Fig. 29 Hand control (4 drives)
14 13
7.
Leg section - up
8.
Leg section - down
9.
Seat height - up
10.
Seat height - down
11.
Footrest - retract
12.
Footrest - extend
13.
Basic position
14.
Shock position (Trendelenburg position)
Fig. 30 Hand control (5 drives)
14
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 27 / 72
4 Operation
4.3.1
Unlocking and locking the hand control
The unlocking and locking of the hand control is shown here using the example of a MULTILINE NEXT DC with
5 adjustment drives (only 4225000/4225100).
Locking the hand control prevents the couch from moving accidentally or unintentionally.
TIP After disconnecting the mains plug and then plugging it back in, the couch remains in its previous operating status. This means, for example, that the hand control was locked before disconnecting the mains supply, so the hand control remains locked even after the mains supply has been restored.
The control unit remembers the current operating status, even without an existing battery for
emergency operation (see 4.5 Battery mode).
If the battery is available and the mains supply is disconnected, the control unit changes to standby after 10 minutes. The corresponding operating control LED on the hand control turns off.
A couch in stand-by mode cannot be activated by the magnetic key. Press any button on the hand control to activate.
Now the couch can be unlocked or locked.
Operating statuses
When the couch is connected to the mains and the hand control is disabled, the yellow operation control LED (1) lights up.
When the couch is connected to the mains and the hand control is enabled with the magnetic key, the green operation control LED (2) lights up to indicate that the couch is ready for operation.
2
1
4
5
3
Fig. 31 Unlocking and locking the hand control
Unlocking the hand control
1.
Yellow operation control LED (hand control locked)
2.
Green operating control LED (hand control unlocked)
3.
Hand control
4.
Magnetic key
5.
Key symbol
To unlock the hand control (3), swipe the magnetic key (4) over the key symbol (5).
Locking the hand control
To lock the hand control, swipe the magnetic key over the key symbol again.
TIP When the hand control is locked, any foot switch for moving to the shock position (Trendelenburg
position) is also locked (see 4.6 Foot switch).
28 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
4 Operation
4.3.2
Basic position
If patients want to sit down on or get up from the couch, as well as before driving, the couch must be brought into the basic position by pressing the yellow button (13).
Pressing the button for basic position requires increased attention from the user, as different actuators are activated simultaneously. All lying surfaces (backrest, seat surface and leg section) move simultaneously to the basic position at maximum speed. Depending on the model, the seat height is also lowered completely.
The backrest is fully upright.
The seat surface is placed horizontally.
The leg section is swivelled all the way down.
The seat height may be lowered completely if required (not 4223000/4223100).
TIP The electric actuator of the footrest (only 4225000/4225100) remains in the existing position.
13
Fig. 32 Button for basic position
13.
Basic position
13 13
© 2019 GREINER GmbH
Fig. 33 Basic position
Instruction manual - MULTILINE NEXT DC 29 / 72
4 Operation
4.3.3
Relax position
Pressing the two seat inclination adjustment buttons (15) simultaneously brings the couch in a comfortable reclining position for the patient.
Pressing the buttons for the relax position requires increased attention from the user, as different actuators are activated simultaneously. All lying surfaces (backrest, seat surface and leg section) move simultaneously to the fixed position at maximum speed.
The backrest lowers slightly.
The seat surface is slightly raised.
The leg section swivels upwards by a predetermined amount.
Careless handling can endanger the patient and third parties or lead to damage to objects or the couch.
TIP The seat height (not 4223000/4223100) and the electric adjustment drive of the footrest
(only 4225000/4225100), if present, remain in the existing position.
The position is permanently stored in the control unit and cannot be changed.
15 15 15
Fig. 34 Button combination for relax position
15.
Relax position: seat inclination - up+down
30 / 72
Fig. 35 Relax position
Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
4 Operation
4.3.4
Bed position (flat position)
Pressing the two leg section adjustment buttons (16) simultaneously brings the couch into the bed position (flat position).
Pressing the buttons for the bed position requires increased attention from the user, as different actuators are activated simultaneously. All lying surfaces (backrest, seat surface and leg section) move simultaneously to the fixed position at maximum speed.
The backrest is brought into a flat position.
The seat surface is placed horizontally.
The leg section is swivelled all the way up.
Careless handling can endanger the patient and third parties or lead to damage to objects or the couch.
TIP The seat height (not 4223000/4223100) and the electric adjustment drive of the footrest
(only 4225000/4225100), if present, remain in the existing position.
The position is permanently stored in the control unit and cannot be changed.
16 16 16
Fig. 36 Button combination for bed position (flat position)
16.
Bed position (flat position): leg section - up+down
Fig. 37 Bed position (flat position)
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 31 / 72
4 Operation
4.3.5
Shock position
WARNING Risk of injury due to patient slipping off the couch.
This function is intended for medical emergencies only and not for daily use.
The patient position known as the shock position (Trendelenburg position), is also referred to by others as the low head position.
Pressing the button (14) for shock position or the foot switch, if present, requires maximum attention from the user, as different actuators are activated simultaneously. All lying surfaces (backrest, seat surface and leg section) move simultaneously, at maximum speed, in shock position.
The backrest lowers to the head-down position.
The seat surface and the leg section are swivelled all the way up.
When pressing the button or the foot switch (Trendelenburg position), if present, make sure that
the patient does not slip off the couch (see 4.6 Foot switch).
Do not leave patients unattended while in the shock position (Trendelenburg position). The patient may lose consciousness.
Careless handling can endanger the patient and third parties or lead to damage to objects or the couch.
TIP The seat height (not 4223000/4223100) and the electric adjustment drive of the footrest
(only 4225000/4225100), if present, remain in the existing position.
14
Fig. 38 Button for shock position (Trendelenburg position)
14.
Shock position (Trendelenburg position)
14 14
32 / 72
Fig. 39 Shock position
Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
4 Operation
4.4
Duty cycle
The maximum duty cycle of the electrical couch functions is indicated on the type plate and in the technical data
(see 1.6 Type plate and 9.4 Technical data).
NOTE 10 % 2 min. / 18 min. means that each electrical adjustment may be operated continuously for a maximum of 2 minutes and then a pause of 18 minutes is required (overheating protection).
TIP The couch is equipped with a self-resetting thermal protection device, which prevents an overload of the control unit or the adjustment drives. The couch is automatically switched off in the event of overloading.
If the maximum duty cycle of two minutes is exceeded several times or for an extended period of time, the electrical adjustment functions are temporarily unavailable.
The couch has been adjusted again once it has cooled down sufficiently.
4.5
Battery mode
The couch can also be equipped with a battery (lead gel) for shock position (Trendelenburg position). This means that in the event of a mains failure or if the couch is disconnected from the mains, the patient can be quickly brought into the shock position independently of the mains supply.
The battery is not intended as a replacement for mains operation.
Notes on handling the battery
TIP Charge the battery for at least 24 hours before first use.
The battery is then available for emergency operation without restriction.
It is not possible to overcharge the battery.
The couch can be adjusted during the charging process.
NOTE Proper handling increases the service life of the battery.
Avoid deep discharge of the battery.
Deeply discharged batteries can be damaged to such an extent that premature replacement is necessary.
In order to guarantee the electrical functionality at all times or the longest battery life, the couch should be connected to the mains at all times.
Fully charged batteries should not be stored for longer than 3 months without recharging. The battery must be fully charged or replaced with a fully charged battery every 3 months at the latest, otherwise it will be damaged due to self-discharge.
The battery (accessory) in the control unit is a wearing part and must be replaced after 4 years at the latest, or even earlier, depending on the intensity of use.
Frequent and fast discharges reduce the service life of the batteries.
The control unit distinguishes between three charging levels:
Charge level high: > 20 V - Normal
Charge level medium: 18-20 V - Alarm
Charge level low: < 18 V - Critical
If the charge level falls to the medium voltage (18-20 V), a permanent alarm sounds as soon as any key of the hand control is pressed. The battery should be charged immediately to ensure the best possible service life.
When the charge level reaches the critical level (< 18 V), the control unit immediately deactivates all movements.
If you still try to move the couch by pressing a button on the hand control, the battery can be deeply discharged.
There is a danger that the initiated movement cannot be finished or that the processor in the control unit can no longer detect the position of the drives. The longest life span is achieved when the battery is fully charged.
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 33 / 72
4 Operation
4.6
Foot switch
The couch can also be equipped with a foot switch (1).
Pressing the orange foot switch brings the couch directly into the shock position (Trendelenburg position) (see
TIP When the hand control is locked, the foot switch for moving to the shock position (Trendelenburg
position) is also locked (see 4.3.1 Unlocking and locking the hand control).
1
Fig. 40 Foot switch
1.
Foot switch
4.7
Equipotential bonding pin
The equipotential bonding cable connects the connection pin (1) on the couch to the connection on the wall strip.
To do this, place the plug completely onto the corresponding connection pin against the noticeable resistance. To remove, simply disconnect the plug.
1
34 / 72
Fig. 41 Equipotential bonding pin
Instruction manual - MULTILINE NEXT DC
1.
Connection pin
© 2019 GREINER GmbH
4 Operation
4.8
LED reading lamp
The couch can also be equipped with an LED reading lamp.
Switching on and off
Press the button (1) on the lamp head (2) to switch it on and off.
CAUTION Danger of glare / eye injuries from LED reading lamp.
Never look directly into the LED reading lamp and do not use the LED reading lamp to illuminate the eye area. The bright light can cause glare or eye injuries.
2
1
TIP
1.
On/off button
2.
Lamp head
Fig. 42 LED reading lamp
The LED reading lamp switches on automatically when the mains plug is inserted.
NOTE The LED reading lamp cannot be switched on in purely battery operation - deep discharge protection.
NOTE Despite comprehensive EMC testing, it is possible that the LED reading lamp may cause interference to other devices.
In such cases, either switch off the LED reading lamp completely or, if necessary, maintain a greater distance from the disturbed device, or change the orientation of the devices to each other.
If possible, do not use the same mains outlet.
TIP Before cleaning and disinfecting, switch off the LED reading lamp and then disconnect the mains
plug - allow to cool sufficiently (see 6 Cleaning and disinfection).
Use commercially available plastic cleaners to clean the electrical components.
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 35 / 72
4 Operation
4.9
Armrests
Folding up and lowering
Guide the armrest (1) by hand until it reaches the end position (when folding up and lowering). Always hold the armrest when lowering and do not let it fall freely.
TIP The left and right armrests and their arm supports can be adjusted independently of each other.
1
1.
Armrest
Fig. 43 Armrests - folding up and lowering
Adjusting the width
The distance between the armrests (1) can be continuously adjusted. To do this, lift the whole armrest slightly.
Now insert or pull out the armrest. Always hold the armrest when lowering and do not let it fall freely.
1
36 / 72
Fig. 44 Armrests - adjusting the width
Instruction manual - MULTILINE NEXT DC
1.
Armrest
© 2019 GREINER GmbH
4 Operation
Adjusting the inclination
The inclination of the arm support (2) can be adjusted in 3 positions. To do this, disengage the locking pin (3) by pulling. Slightly raise the arm support to relieve the load on the locking pin. Now incline the arm support downwards and then lock it again with the locking pin. Always hold the arm support when inclining and do not let it fall freely.
2
3
2.
Arm support
3.
Locking pin (inclination)
Fig. 45 Arm supports - adjusting the inclination
Adjusting the height
The height of the armrest is continuously adjustable. To raise, turn the handwheel (4) clockwise and to lower, unscrew it counterclockwise.
NOTE Do not force the handwheel over the noticeable end position when in the highest position. When the lowest position is reached, there is no longer any noticeable resistance on the handwheel.
4
Fig. 46 Armrests - adjusting the height
4.
Handwheel (height)
Swivelling
The arm support (2) can be swivelled inwards or outwards independently of the armrest. To do this, loosen the handwheel (5) and swivel the arm support inwards or outwards, then retighten the handwheel.
2
Fig. 47 Arm supports - swivelling
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC
2.
Arm support
5.
Handwheel (swivelling)
5
37 / 72
4 Operation
4.10
Footrest
Removing and attaching the footrest
Release the two handwheels (2) and then pull out the footrest upwards.
When reinstalling, make sure that the connecting bolts (3) are fully engaged in the counterpart. Now tighten the two handwheels firmly again.
1 3
2
Fig. 48 Removing and attaching the footrest
1.
Footrest
2.
Handwheel
3.
Connecting bolt
TIP In the version with 5 adjustment drives (only 4225000/4225100), the footrest is continuously electrically adjustable.
The footrest can be adjusted manually (4223000/4223100 or 4224000/4224100) in both the 3 and
4 actuator versions.
Manual footrest
The position of the footrest (1) is continuously adjustable. Press down the clamping lever (4) to release. Push in or pull out the footrest. When moving with the other hand, lift the footrest slightly to relieve the load on the guides. To fix them in place, pull the clamping lever upwards.
NOTE Do not use force to tighten the clamping lever beyond the noticeable end position.
1
4
38 / 72
Fig. 49 Manual footrest
Instruction manual - MULTILINE NEXT DC
1.
Footrest
4.
Clamping lever
© 2019 GREINER GmbH
4 Operation
4.11
Driving
The couch may only be moved over a solid surface.
Before moving the couch, it must be brought into the basic position by pressing the yellow button (13) on the
hand control (see 4.3.2 Basic position).
Retract the footrest completely. Either by pressing the button (11) on the hand control (see
4.3 Hand control) or by pushing in manually (see 4.10 Footrest).
Always hang the hand control on the brackets at the front end of the arm supports. The cable of the hand control must not touch or run over the floor, as it is not resistant to being run over.
If necessary, the cable of the hand control must be wound up or hung up.
Running over or crushing the hand control cable will result in damage which can lead to malfunctions.
Disconnect the mains plug and hang it onto the mains cable holder (on the backrest).
WARNING Risk of electric shock / fire and malfunction.
Make sure that the mains cable cannot be overstretched, run over or otherwise damaged. The mains cable must be attached to the mains cable holder and must not touch or run over the floor, as it is not resistant to being run over.
If necessary, the mains cable must be wound up or hung up.
Running over or crushing the mains cable will result in damage which can lead to electrical hazards and malfunctions.
The armrests must be pushed inwards in the narrowest position. Furthermore, the armrests must
be folded down into the normal position and aligned parallel to the couch (see 4.9 Armrests).
Do not use any existing infusion rod etc. as a push handle.
Additional attached or suspended objects must not cause any danger to the user or third parties.
Possible damage (e.g. to door frames etc.) must be avoided.
Do not drive faster than walking speed and slow down when turning corners and curves.
Drive carefully over obstacles on the ground, e.g. thresholds, and do not brake abruptly.
The use of stairs is strictly prohibited.
When driving on slopes or inclines, e.g. ramps, a second user must additionally secure the couch at the front due to its weight.
The couch is equipped with four lockable castors.
All castors must be braked at all times the couch is parked (see 4.12 Braking).
Fig. 50 Driving
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 39 / 72
4 Operation
4.12
Braking
Depending on the version, 3 different braking systems can be installed.
The operation of the braking systems is different.
4.12.1
Castors with central locking
Braking
To brake the couch, step on the red side of one of the two pedals (1).
When braked, both pedals are horizontal.
All castors (2) are locked simultaneously.
Make sure that the couch cannot move on its own.
2
1
Driving
Fig. 51 Locking the castors with central locking
To release the brakes, step on the green side of one of the two pedals (1).
In the driving position, the green side of both pedals points downwards.
All castors (2) are released simultaneously.
1.
Pedal
2.
Castor
2
1
40 / 72
Fig. 52 Releasing the castors with central locking
Instruction manual - MULTILINE NEXT DC
1.
Pedal
2.
Castor
© 2019 GREINER GmbH
4 Operation
4.12.2
Castors with central locking and directional control
In order to allow the couch to move smoothly, the right front castor (marked yellow) can be locked in the direction of travel. The electrical conductivity of the right front castor additionally prevents possible electrostatic charging of the couch.
Braking
To brake the couch, step down the red side of one of the two pedals (1) completely.
When braked, the red side of both pedals points downwards.
All castors (2+3) are locked simultaneously.
Make sure that the couch cannot move on its own.
1
3
2
Manoeuvring
Fig. 53 Locking the castors with central locking and directional control
1.
Pedal
2.
Castor
3.
Castor (directional)
To release the brakes and bring the couch into the manoeuvring position, step on the green side of one of the two pedals (1).
In the manoeuvring position, both pedals are horizontal.
All castors (2+3) are released simultaneously.
TIP In the manoeuvring position, all castors are completely free to move. This means that the couch can be manoeuvred in the tightest of spaces.
1
3
2
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC
1.
Pedal
2.
Castor
3.
Castor (directional)
Fig. 54 Manoeuvring the couch with central locking and directional control
41 / 72
4 Operation
Driving
To release the brakes and bring the couch into the driving position, step down the green side of one of the two pedals (1) completely.
In the driving position, the green side of both pedals points downwards.
All castors (2+3) are released simultaneously.
The right front castor (marked yellow) (3) can now be engaged in the direction of travel.
1
3
2
TIP
1.
Pedal
2.
Castor
3.
Castor (directional)
Fig. 55 Driving the couch with central locking and directional control - directional castor set correctly
The right front castor can engage in two positions offset by 180°. A castor that is locked in the forward position starts to wobble when pushed, which makes you feel unpleasant and unsafe to drive.
To lock the right front castor correctly in the driving position, always push the couch forward a short distance first. Ensure that the directional castor is locked in the retracted position.
3
42 / 72
2 1
1.
Pedal
2.
Castor
3.
Castor (directional)
Fig. 56 Driving the couch with central locking and directional control - directional castor set incorrectly
Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
4 Operation
4.12.3
Single-braked castors
Braking
Step on the locking pedals (1) on all 4 castors (2).
All 4 castors must be braked at all times.
CAUTION Risk of falling or accident due to insufficiently braked couch.
If not all 4 castors are braked, the couch is insufficiently fixed. This applies in particular to slopes or inclines such as ramps.
Make sure that the couch cannot move on its own.
2
1
Fig. 57 Locking the single-braked castors
1.
Locking pedal
2.
Castor
Driving
To release the brakes, push all 4 locking levers at the upper end with the tip of your foot.
2
1
Fig. 58 Releasing the single-braked castors
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC
1.
Locking pedal
2.
Castor
43 / 72
4 Operation
4.13
Sitting down and standing up of the patients
If patients want to sit down on or stand up from the couch, it must first be brought into the basic position. There-
fore press the yellow button (13) on the hand control (see 4.3.2 Basic position).
Locking the couch by activating the corresponding brake system.
When the patient sits down or stands up, all castors of the couch must be braked (see 4.12 Braking).
CAUTION Risk of falling or accident due to insufficiently braked couch.
If the couch is not being moved, the castors must be locked at all times, as the couch may be used by patients as a support when sitting down or standing up. Rolling the unbraked couch away can lead to serious falls. After activating the central locking or the 4 single braked castors, check whether the couch is actually fixed, i.e. the castors are sufficiently braked.
The couch may be in an unbraked position before each new use and must therefore be checked immediately for properly locked castors.
Adjust the footrest as required or desired - this may depend on the size of the patient. Either by
pressing the button (11) on the hand control (see 4.3 Hand control) or by pushing in manually (see
Provide support as needed.
CAUTION Risk of falling or accident due to moving arm supports.
When patients get in or out from the front, they must not lean on the arm supports as they may suddenly swing sideways.
4.14
Fig. 59 Sitting down and standing up of the patients
Patient transportation
When transporting patients, proceed in principle in the same way as when moving the unoccupied couch (see
4.11 Driving). In addition, the following requirements must be observed.
TIP The castors are essentially used only to move the couch when it is unoccupied.
The couch can be used for patient transportation within buildings under certain conditions.
The patient may only be transported in a seated position.
The forearms of the patient lie in normal position on the arm supports.
The patient's feet are on the footrest.
Retract the footrest as fully as possible or adjust as required or desired - this may depend on the
by pushing in manually (see 4.10 Footrest).
When driving on slopes or inclines, e.g. ramps, a second user must additionally secure the couch at the front due to its weight and for the safety of the patient.
44 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
4 Operation
4.15
Infusion rod holder and infusion rod
The couch can also be equipped with an infusion rod holder and an infusion rod.
Swivel the infusion rod holder
The infusion rod holder (1) or the infusion rod (2) can be simply swivelled from left to right or in the opposite direction. The infusion rod holder sensitively snaps into the relevant end position.
CAUTION Risk of trapping due to infusion rod.
Never leave the infusion rod holder in an intermediate position. When lowering the back, there is a risk of trapping between the infusion rod and the back cover.
The infusion rod will definitely be damaged. The infusion rod holder may also be bent and will have to be repaired.
Intermediate positions represent a danger and must therefore be avoided at all costs.
This also applies to couches that appear to be standing around unused. Patients and third parties have no way of detecting unsafe intermediate positions of the infusion rod holder and of assessing the resulting danger.
Inserting and adjusting the infusion rod
After inserting the infusion rod, tighten the handwheel (3) on the infusion rod holder. Make sure that the infusion rod has also been pushed into the infusion rod holder as far as it will go. The height of the infusion rod can be continuously adjusted. To do this, loosen the clamping screw (4) on the infusion rod. Hold the upper pull-out part of the infusion rod while doing this. Adjust the height as desired or required. Retighten the clamping screw.
NOTE
Do not use the infusion rod as a push handle - risk of damage (see 4.11 Driving).
4
3
2
© 2019 GREINER GmbH
1
Fig. 60 Infusion rod holder and infusion rod
Instruction manual - MULTILINE NEXT DC
1.
Infusion rod holder
2.
Infusion rod
3.
Handwheel
4.
Clamping screw
45 / 72
5 Accessories
5 Accessories
Every
MULTILINE NEXT DC
couch can be equipped with accessories.
TIP Depending on the type of accessories, retrofitting (after initial delivery) can range from very simple to quite complex.
5.1
Premium material
The premium upholstery cover material (PREMIUM 422) allows even greater customisation of the couch. Different colour combinations create highlights and give expression to creativity.
TIP
White P800 Lightgrey P200 Anthracite P210
The innovative "premium" line sets new standards with its super soft, yet extremely hard-wearing surface.
The "premium." upholstery cover material is currently available in 12 different colours.
Grass P400 Mocca P300 Chocolate P310
Citron P500 Carmin P600 Fuchsia P610
Aqua P700 Navy P710
Fig. 61 Premium material
5.2
Neck roll
Black P100
The couch can be ordered with a neck roll (4222600) to support the cervical spine.
TIP
Of course, a neck roll, in the appropriate cover material and the desired colour, can be re-ordered at any time.
Fig. 62 Neck roll
46 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
5 Accessories
5.3
Relax pillow
The couch can be ordered with a relax pillow (4222610) for soft positioning of the head.
TIP
Of course, a relax pillow, in the appropriate cover material and the desired colour, can be re-ordered at any time.
Fig. 63 Relax pillow
5.4
Neck roll / Relax pillow in premium material
TIP The neck roll and the relax pillow feel more comfortable and valuable in direct skin contact.
They are also available in combination with compact cover material on the large surfaces for backrest, seat and legrest upholstery.
5.5
Arm supports in premium material
Not for integral foam arm supports.
TIP
The padded arm supports feel more comfortable and valuable in direct skin contact.
They are also available in combination with compact cover material on the large surfaces for backrest, seat and legrest upholstery.
Fig. 64 Arm supports in premium material
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 47 / 72
5 Accessories
5.6
LED reading lamp
For handling the LED reading lamp (4222100) (see 4.8 LED reading lamp).
TIP
The LED reading lamp enables glare-free illumination of the patient's direct field of vision.
The position need only be adjusted once and the ergonomic gooseneck holds it safely and reliably.
No other lighting equipment is required.
No tripping hazards due to additional mains cables etc. - safety issue.
Fig. 65 LED reading lamp
5.7
Legrest upholstery protection cover
The legrest upholstery cover (4221100) protects against soiling and damage. The robust, transparent design of the plastic film ensures a long service life.
TIP
The protection cover is attached to the back of the legrest upholstery using Velcro fasteners. This allows the protection cover to be attached and removed quickly and user-friendly - easy handling during cleaning.
Fig. 66 Legrest upholstery protection cover
48 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
5 Accessories
5.8
Footrest protection cover
The protection cover (4221200) covers the footrest completely and protects it from soiling and damage. The robust, transparent design of the plastic film ensures a long service life.
TIP
The protection cover is simply pulled over the footrest. This allows the protection cover to be attached and removed quickly and user-friendly - easy handling during cleaning.
Fig. 67 Footrest protection cover
5.9
Battery
The additional battery built into the control unit for shock position (Trendelenburg position) enables the user to bring the patient reliably and quickly into shock position, even if there is no mains connection.
TIP
The battery for the shock position
(Trendelenburg position) gives the user the secure feeling of having control over the situation and being able to act at all times, even in the event of unforeseeable events - safety aspect.
Fig. 68 Battery
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 49 / 72
5 Accessories
5.10
Push handle with paper roll holder
The ergonomically shaped push handle (4221400) enables the user to manoeuvre the couch safely and precisely.
TIP
The push handle also serves as a paper roll holder (roll width 55 cm).
This makes it very easy to replace the paper cover when changing patients - saving time.
Fig. 69 Push handle with paper roll holder
5.11
Padded arm supports
The couch is supplied with integral foam arm supports as standard. However, padded arm supports (4221520)
can also be ordered instead of integral foam arm supports (see 4.9 Armrests).
TIP
Of course, padded arm supports, in the appropriate cover material and the desired colour, can be reordered at any time.
Fig. 70 Padded arm supports
50 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
5 Accessories
5.12
Infusion rod holder
For handling the infusion rod holder (4221800) (see 4.15 Infusion rod holder and infusion rod).
Fig. 71 Infusion rod holder
5.13
Infusion rod
For handling the infusion rod (4221900) (see 4.15 Infusion rod holder and infusion rod).
Fig. 72 Infusion rod
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 51 / 72
5 Accessories
5.14
Standard rails
The pair of standard rails on the seat (4222200) enables the quick and safe attachment of accessories from other manufacturers. The highly stable and extra-long design provides safety and a wide variety of positioning options.
NOTE Observe the safety aspects when moving the electric actuators.
Take particular care to check for possible trapping points, both on the patient, first and foremost, and with regard to the couch itself.
Tearing off or trapping cables and hoses of the accessories attached to the standard rail can lead
to considerable risks for patients, users and third parties (see 4.3 Hand control, 4.11 Driving and
TIP
When manoeuvring the couch, pay attention to excess width, protruding components, additional mains cables for accessories, etc.
Fig. 73 Standard rails
5.15
Seat with pressure relief
The combination of pressure-relieving springs in the seat upholstery (4222400) with "premium." cover material
(PREMIUM 422) provides unrivalled comfort, even during treatment lasting several hours.
TIP
1
5
The seat with pressure-relieving springs is only available with premium material (PREMIUM 422)
2
3
4
Fig. 74 Seat upholstery with pressure-relieving springs
1.
"premium." cover material
2.
Wadding
3.
Cut foam
6
4.
Felt cover
5.
Single spring element
6.
Ventilation opening in the wood
52 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
5 Accessories
5.16
Foot switch
For handling the foot switch (4222500) (see 4.6 Foot switch).
Pressing the orange foot switch brings the couch directly into the shock position (Trendelenburg position) (see
TIP
Using the foot switch means that the user has both hands free right from the start for important, possibly life-saving, measures on the patient - saving time.
In an emergency, every second counts.
Fig. 75 Foot switch
5.17
Mains cable
The couch is supplied with an EU mains cable as standard. However, a mains cable for Switzerland (CH -
4222700) can also be ordered instead of an EU mains cable.
NOTE Never replace the mains cable yourself. The service of your authorised specialist distributor will be pleased to take care of this.
Work on live components should be carried out by qualified experts - safety aspect.
See "4.3.2 Replacing the mains cable" in the service manual in any case.
TIP
The range of possible variants is constantly being expanded.
Just ask.
1
2
Fig. 76 Mains cable CH
1.
Mains cable CH
2.
Mains plug CH
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 53 / 72
6 Cleaning and disinfection
6 Cleaning and disinfection
This chapter describes all regular cleaning and disinfection activities.
To ensure that the couch is always functional and usable, it should be cleaned and disinfected as soon as possible after each use so that it can be used again immediately. Improper cleaning or disinfection may cause hazards.
WARNING Risk of electric shock / fire and malfunction.
The couch must always be disconnected from the mains supply by unplugging the mains plug before performing any maintenance work.
Before recommissioning, ensure that no residual moisture remains on the electrical contacts by drying or blowing off the mains plug.
None of the electrical components may have external damage.
Failure to observe these instructions may result in the penetration of cleaning agents or disinfectants, resulting in malfunctions or damage.
DANGER Risk of electric shock / fire as well as functional failure due to penetration of liquids.
If liquids are suspected to have penetrated electrical components, immediately unplug the mains plug or do not reconnect it.
Take the couch out of operation immediately and clearly mark it with "DEFECT".
Inform the operator for intensive inspection during repair.
NOTE The information contained in the instructions of the manufacturer of the relevant detergent or disinfectant must be strictly observed.
The dilution ratio recommended by the manufacturers in the respective instructions must be observed. The specified concentrations must not be exceeded or undershot.
Failure to observe these instructions may result in material damage or personal injury.
TIP Before the extensive use of cleaning agents and disinfectants, we always recommend testing the surface compatibility on a non-visible area.
As a rule, the solutions should be freshly prepared.
TIP Do not use a brush and high pressure to clean or disinfect type plates, stickers, symbols and notes.
Readability must always be guaranteed and compatibility tested over a small area if necessary.
Cleaning and disinfecting agents must not contain any corrosive or corrosive components.
They must not contain substances that alter the surface structure or the adhesive properties of the materials.
The lubricants used on the couch must not be attacked.
The pH value of the water must not exceed or fall below 6-8.
The total water hardness of 0.9 mmol/l (corresponding to 5°dH) must not be exceeded.
TIP The information in the VAH list (German Association for Applied Hygiene), as well as the information provided by us, do not exempt the user from carrying out her or his own tests and trials, as the conditions (e.g. water hardness) can vary greatly from place to place. A legally binding assurance of certain characteristics cannot be derived from this.
NOTE The use of unsuitable cleaning agents and disinfectants, the wrong mixing ratio and insufficient care of the couch can cause damage to the surface coating for which GREINER GmbH is not liable.
TIP Ensure that no liquid residues remain on the metallic parts of the couch after cleaning and disinfection (avoid droplet formation on edges). Otherwise corrosion cannot be excluded in these areas in the long term.
TIP To avoid degreasing of the piston rods, all electric actuators should be moved to the smallest stroke position before cleaning or disinfection.
Dry-running piston rods should be greased thinly with industrial Vaseline.
54 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
6 Cleaning and disinfection
6.1
Cleaning
Cleaning and subsequent disinfection must be carried out at regular intervals.
The following times are usually to be observed.
If required.
After each patient change.
In accordance with the guidelines of the operator's hygiene plan.
Depending on the degree of soiling, we recommend cleaning the couch with a damp cloth (microfiber cloth or similar). Oil, grease, sweat, urine and blood must be removed immediately. Use a warm, mild soapy water. A soft hand brush can also be used for stubborn dirt or stains. Do not use excessive amounts of water to clean the couch.
NOTE The upholstery materials must be protected from the effects of intense heat and cleaned regularly.
Take care that the upholstery are not soaked.
Do not use organic solvents (e.g. petrol, benzene, turpentine, toluene, xylene or acetone etc.), chlorides, polishes, chemical cleaning agents or wax polishes to clean the upholstery. Do not use agents containing oil, grease or alcohol.
The upholstery materials are not resistant to dry cleaning.
Under no circumstances should scouring sponges or cleaning agents containing abrasive particles, abrasives or other blunting substances be used to clean the steel parts.
For cleaning the plastic parts also use a warm, mild soapy water or, if necessary, commercially available plastic cleaner.
Use commercially available plastic cleaners to clean the electrical components.
NOTE The couch may not be cleaned in washing facilities, by means of spray lances of a high-pressure cleaner or water jet.
6.2
Disinfection
For wipe and spray disinfection, the disinfectants listed in the VAH list (German Association for Applied Hygiene) can be used in their intended concentration.
The dilution ratio recommended by the manufacturers in the respective instructions must be observed.
NOTE The concentrations indicated in the VAH list should not be exceeded or undershot. Under no circumstances may the user add cleaning agents such as soap or detergents to a disinfectant at his own discretion.
Do not allow the disinfectant to penetrate into the upholstery beyond the prescribed contact time.
Disinfectants are aggressive and can change the surface.
DANGER Fire and explosion hazard due to wiping / spraying disinfection containing alcohol.
There is a fire and explosion hazard when wiping or spraying disinfectants containing alcohol are used over large areas.
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 55 / 72
7 Troubleshooting
7 Troubleshooting
The chapter troubleshooting contains information about possible malfunctions that can be corrected by the user.
TIP The use of specially trained and instructed service personnel is neither necessary nor intended for troubleshooting.
Safety Instructions for troubleshooting
DANGER Fatal risk of electric shock.
Under no circumstances may the user attempt to rectify faults in electrical components him or herself.
General troubleshooting tips
TIP Before each troubleshooting check whether the couch is connected to the mains (mains plug in a live socket).
Voltage test for everyone.
At the appropriate mains outlet, perform a simple test with another functioning electrical device.
Device works when switched on - mains voltage present.
Device does not function when switched on - no mains voltage present.
When the couch is connected to the mains and the hand control is enabled with the magnetic key, the green op-
56 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
7 Troubleshooting
7.1
Initialisation of actuators
backrest, seat inclination or leg section, the control unit does not know the current position of the corresponding actuator.
In this case, you must move the affected actuator completely to the end position indicated below. Do not press any other keys until the process is complete.
Carry out the initialization as follows.
Raise the backrest completely (3).
TIP
Swivel the seat all the way down (6).
Swivel the leg section all the way down (8).
The signal tone stops and the current position of the corresponding actuator is reinitialized. The actuator is ready for operation again.
First bring the couch completely into the shock position (Trendelenburg position) (14) (see
4.3.5 Shock position). Immediately afterwards, move the couch completely to the basic position
(13) (see 4.3.2 Basic position) - safely initialises any faulty actuator(s).
This procedure is only recommended for an unoccupied couch.
3 3 3
6 6 6
13
8
14 13
8
14 13
8
14
Fig. 77 Initialisation of actuators
3.
Backrest - up
6.
Seat inclination - down
8.
Leg section - down
13.
Basic position
14.
Shock position (Trendelenburg position)
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 57 / 72
7 Troubleshooting
7.2
Control unit RESET
Fluctuations in the mains voltage or similar can lead to the control unit of the couch no longer reacting to the ac-
In this case, the control unit must be "RESET".
Carry out the control unit RESET as follows.
Press and hold the two top buttons (3) and (4) simultaneously.
After a short time, a pulsating signal tone sounds.
Continue to press and hold the two buttons until the sound stops.
A "RESET" has been carried out on the control unit and the couch is ready for operation again.
TIP In some cases, the actuators may have to be reinitialised after a "RESET" of the control unit (see
7.1 Initialisation of actuators).
3 3 3
4 4 4
Fig. 78 Control unit RESET
15.
Backrest - up
16.
Backrest - down
58 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
7 Troubleshooting
7.3
Troubleshooting table
The following table provides assistance in remedying malfunctions.
Electrical malfunction
Problem
Couch completely without electrical function.
The green power indicator
LED on the control unit does not light up.
(may only be carried out by service personnel, as the backrest cover must be removed)
Hand control without function.
Possible causes
Mains cable not plugged in.
Socket without mains voltage.
Mains plug not plugged in properly.
Defective hand control, mains cable or control unit.
Plug in the mains cable. The green or yellow operating control LED on the hand control must light up.
Check socket and fuse box.
Check the plug connection of the mains plug.
Inform the operator for intensive inspection during repair.
If necessary, replace the hand control, mains cable or control unit.
Socket without mains voltage.
Mains cable or plug defective.
Defective control unit.
Hand control locked.
Defective hand control.
Solution
Use an intact mains socket.
Inform the operator for intensive inspection during repair.
If necessary, replace the mains cable or control unit.
Unlock the hand control.
(see 4.3.1 Unlocking and locking the hand control)
Inform the operator for intensive inspection during repair.
If necessary, replace the hand control.
Foot switch without function.
Despite a proper mains supply, operation is not possible.
Hand control locked. (the corresponding function of the foot switch is also locked)
Defective foot switch.
Unlock the hand control.
(see 4.3.1 Unlocking and locking the hand control)
Inform the operator for intensive inspection during repair.
If necessary, replace foot switch.
Control unit has switched off due to overheating.
Defective control unit.
Max. duty cycle:
Op: 2 / 18 min note
Inform the operator for intensive inspection during repair.
If necessary, replace the control unit.
Actuator only starts briefly, then stops.
An acoustic signal sounds when the actuators for the backrest, seat inclination or leg section are activated.
Actuator overloaded.
The control unit does not have the current position of the corresponding actuator.
Remove overload.
Retest.
Recalibrate the actuator(s).
(see 7.1 Initialisation of actuators)
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 59 / 72
7 Troubleshooting
Problem
The control unit does not respond and a pulsing signal tone sounds when any button on the hand control is pressed.
Battery operation not possible.
Battery operation with alarm tone only.
Electrical malfunction
Possible causes
Fluctuations in the mains voltage etc. lead to a total failure of the unit.
Battery discharged.
No battery was ever, or is not currently, installed.
Battery capacity exhausted or medium charge level reached.
Solution
Carry out a "RESET" of the control unit.
Connect the couch to the mains for at least 24 hours.
Inform the operator for intensive inspection during repair.
If necessary, install a battery.
Connect to the mains as soon as possible to charge the battery.
Battery operation with sufficient charge is only possible for a short time.
Potential equalisation does not function properly.
The LED reading lamp cannot be switched on.
(the connection cable may only be checked by service personnel, as the backrest cover must be removed)
Battery life limit reached.
Equipotential bonding pin damaged or not present.
Inform the operator for intensive inspection during repair.
If necessary, replace the battery.
Inform the operator for intensive inspection during repair.
If necessary, replace the equipotential bonding pin.
Defective LED reading lamp.
Connection cable of the LED reading lamp plugged out or defective.
Inform the operator for intensive inspection during repair.
If necessary, plug in the connection cable or replace the LED reading lamp.
Mechanical malfunction
Possible causes Solution Problem
Central locking of the 4 castors does not function properly.
Single-braked castor(s) do(s) not function(s) properly.
Defective castor(s).
Brake linkage bent or defective.
Defective castor(s).
Inform the operator for intensive inspection during repair.
If necessary, replace the castor(s).
Inform the operator for intensive inspection during repair.
If necessary, replace the castor(s).
60 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
8 Decommissioning
8 Decommissioning
8.1
Service life
Provided that the intended use, safety inspections and necessary maintenance measures are observed, the service life of the MULTILINE NEXT DC couch is approx. 7 to 10 years, depending on stress, operating conditions and frequency.
8.2
Storage and transportation
NOTE
The MULTILINE NEXT DC couch should only be moved (see 4.3.2 Basic position and 4.11 Driving)
and carefully stored (see 4.12 Braking) in basic position.
TIP
The couch must be cleaned and disinfected before storage (see 6 Cleaning and disinfection).
Store the couch in a dry, well-ventilated room, protected from dirt.
Furthermore, the couch should preferably be protected from direct sunlight or UV radiation, if necessary by a light-proof foil.
The following environmental conditions must be observed at the storage location or during transportation.
Storage or transportation temperature [C]: + 50
Air humidity [rH] (non-condensing):
Air pressure [hPa]:
Storage at an estimated altitude [m]:
- 10
20 % - 90 %
700 - 1060
≤ 2000
8.2.1
Battery
NOTE Fully charged batteries should not be stored for longer than 3 months without recharging. The battery must be fully charged every 3 months at the latest, otherwise it will be damaged due to
self-discharge (see 4.5 Battery mode).
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 61 / 72
8 Decommissioning
8.3
Check before recommissioning
CAUTION Risk of infection from contaminated couch.
After storage, the couch must be cleaned and disinfected before being put back into operation.
A functional check of the couch must be carried out immediately after cleaning and disinfection and before use
(occupation by a new patient).
WARNING Risk of injury for patient, user and third parties.
The couch must be checked for functional safety once recommissioning is complete.
It must be ensured that the couch can be used as intended without endangering the patient, the user or third parties.
If the couch is recommissioned after storage, the same functional check must be carried out as for the check be-
fore commissioning (see 3.7 Check before commissioning and reassignment).
General check of the overall condition of the couch for soiling, condition and damage, completeness and legibility of stickers, symbols and instructions, clean if necessary in accordance with the manufacturer's specifications or those of the operator and repair if necessary.
Visually inspect the mains cable and plug, hand control, as well as all other electrical cables and connectors for breakage, proper installation (e.g. strain relief), clamping or friction points etc. and repair if necessary.
Functional check of the electrical adjustment devices and the control system. All electric actuators must be operated once up to their end positions. Pay particular attention to changed operating behaviour, unusual noises, speed, smooth running, odours and increased temperature and repair if necessary.
Carry out a functional check according to the installed brake system for the castors (locking, freewheel) and repair if necessary.
Any defects or damage found must be remedied immediately.
The couch must not be used before the defects or damage have been remedied.
WARNING Risk of injury due to defective couch.
The couch must not be operated in a defective or faulty condition in which the couch could endanger patients, users or third parties.
A defective or faulty couch must be clearly marked "DEFECT". The marking must be done in such a way that the defective condition is clearly identifiable for everyone.
Inform the operator for intensive inspection during repair.
62 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
8 Decommissioning
8.4
Disposal
During the construction of the couch, care was taken to ensure that, wherever possible, no composites were used. This design concept allows a high degree of recycling once the couch has reached the end of its service life.
Country-specific disposal regulations must be observed!
General disposal instructions
CAUTION Risk of infection from contaminated couch.
The operator must ensure that all components to be disposed of are not infectious or contaminated.
In the event of scrapping, dispose of metal and plastic parts separately and properly.
If you have any queries, please contact your local municipality, local waste disposal company or our service (see
Disposal of electrical parts
Labelling of electrical and electronic equipment in accordance with Directive
2002/96/EC (WEEE) and the German Electrical and Electronic Equipment Act (ElektroG)
The symbol on the product indicates that it must not be disposed of as normal household waste.
This couch is classified as a commercially used electrical appliance in accordance with WEEE Directive 2002/96/EC (WEEE) and the German Electrical and Electronic Equipment Act (ElektroG).
The electrical components used are free of prohibited harmful substances in accordance with the RoHS Directive.
Replaced electrical components (manual switches, actuators, control units, etc.) must be treated like electrical waste in accordance with the WEEE Directive and disposed of properly.
GREINER GmbH, or its authorised specialist distributors, shall take back the delivered couch after termination of use at the customer's expense and dispose of the goods in accordance with statutory regulations. The customer undertakes to make the device available to
The device is identified by its GREINER serial number and type plate. The customer is responsible for proving that the product was produced by GREINER. For further information, please refer to our General Terms and Conditions at www.greiner-gmbh.de.
Disposal of batteries (if available)
Batteries that are no longer usable must be disposed of properly in accordance with the Battery Ordinance and should never be disposed of with household waste.
Pb
If you have any queries on waste disposal, please contact your local municipality, local waste
disposal company or our service (see 1.2 Manufacturer).
Disposal of gas springs/hydraulic units (if available)
Gas springs and hydraulic units are mainly made of metal and can be recycled after the oil contained within them has been drained.
Gas springs must be depressurised in accordance with the manufacturer's instructions before disposal or draining of the oil. These details are available on request from the corresponding gas spring manufacturer (see type plate or manufacturer label). For the correct procedure for the disposal (recycling) of Stabilus gas springs, see www.stabilus.com.
CAUTION Danger of injury to the eyes or face, risk of injury due to trapping
Due to the high internal pressure, suitable measures for personal protection must be taken before degassing (eye and face protection, covering the degassing opening, etc.).
When the gas springs are removed, there is a risk of trapping when the release mechanism is activated by sudden movement of the piston rod.
Environmental protection
Before disposal, the oil must be drained and disposed of properly.
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 63 / 72
9 Appendix
9 Appendix
9.1
Explanation of symbols
The following symbols are used in this manual, on the couch and its components or parts.
Symbol Meaning Symbol Meaning
TÜV approval mark QM system test mark
Follow the instructions Safety information
Environmental protection RoHS label
No domestic waste Electrical scrap
Manufacturer Date of manufacture
Product name Serial number
Warning Applied Part: Type B
Safe working load Indoor application
Protection class I Alternating current
Thermal fuse element Isolating transformer
9.2
Ordering spare parts
If required, a spare parts list can be provided on request (see 1.2 Manufacturer).
NOTE To ensure functional reliability and to maintain warranty claims, only original spare parts from
GREINER GmbH may be used.
9.3
Return delivery / repair order
Returns of medical devices, whether used or unused, may have come into contact with infectious agents or hazardous substances and are therefore generally regarded as "contaminated returns". This applies until the sender can prove that they are not infectious or contaminated.
Medical devices must be sent in their packaging only. Ensure that a "Declaration of Hygiene Status" is attached to the outside of the packaging. The accompanying documents must show that the medical device has been properly disinfected.
Medical devices must be disinfected before repair. The operator or client is responsible for execution and credible proof.
For further information, please refer to our General Terms and Conditions at www.greiner-gmbh.de.
CAUTION Risk of infection from contaminated couch.
For all returns or repair orders, the sender or operator must ensure that the corresponding medical devices are neither infectious nor contaminated.
64 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
9 Appendix
9.4
Technical data
Dimensions and weights
Fig. 79 Dimensions
Width B:
Length in basic position L1:
Length in bed position L2:
Lying surface width:
Length of lying surface:
Entrance height H1:
Height adjustment H2:
Backrest inclination:
Seat inclination:
Inclination of leg section:
Castor:
Empty weight of couch depending on model:
Safe working load:
Electrical data
Protection class I:
Application Part Type B (special protection against electric shock):
Protection type:
Control unit [CB16]
Bus version:
Number of channels depending on version:
Mains connection [U in]:
Current consumption [I in]:
Total current [I out]:
Current per channel [I out]:
Op = duty cycle 1) :
Thermal fuse element:
Isolating transformer:
Actuators
Rated voltage [U in]:
Current consumption [I in] depending on version:
Battery (shock position)
Battery voltage (battery set - series connection):
Battery capacity:
Version / type:
1) Duty cycle 10 % max. 2 minutes continuous use followed by 18 minutes not in use.
Failure to do so may result in a malfunction.
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC
80 cm
110 cm
192 cm
60 cm
220 cm
54 cm
21 cm (54 - 75 cm)
-5° - 78°
0° - 22°
-25° - 0°
10 cm
105 - 110 kg
220 kg
IPX4
OpenBus™
3 - 5
230 V AC, 50/60 Hz max. 2 A max. 8 A
5.5 - 8 A
10 % max. 2 min / 18 min
24 V DC
3.8 - 7 A
24 V DC
1.3 Ah battery set / lead gel
65 / 72
9 Appendix
9.5
Proof of instruction
Form for proof of instruction
MULTILINE NEXT DC
Fig. 80 Proof of instruction MULTILINE NEXT DC
Type of instruction
Initial instruction
Repeated instruction
Instruction by
Manufacturer
Authorised person of the authorised specialist distributor
Authorised representative of the operator
Other ______________________________
Instruction for
User
Authorised representative of the operator
Authorised person of the authorised specialist distributor
Service personnel of the operator
Service personnel of the authorised specialist distributor
Device data (see 1.6 Type plate)
Manufacturer GREINER GmbH
Wettestraße 1
74385 Pleidelsheim / Germany
66 / 72
Product name
Instruction manual - MULTILINE NEXT DC
MULTILINE NEXT DC
© 2019 GREINER GmbH
9 Appendix
Form for proof of instruction
MULTILINE NEXT DC
The following points, which are particularly relevant, must be clearly emphasized during the instruction.
The maximum safe working load is 220 kg.
Do not sit or rest on backrest upholstery, legrest upholstery, arm support or footrest - danger of tipping over. This must be expressly pointed out to patients.
Immediately after parking (with or without a patient) the couch must always be fixed in place - risk of accident due to sudden rolling away when sitting down or standing up and when leaning against it.
Always lock the hand control when the couch is unattended - risk of injury by jamming limbs. This is particularly relevant in corridors and unoccupied rooms - playing children.
Patients may sit down or stand up only when the couch is in the basic position - risk of accident.
When transporting patients or driving, follow the procedure in accordance with the instruction manual.
Secure the mains and hand control cables against being run over - risk of electric shock / damage.
When making electrical adjustments, always pay attention to the immediate danger area - danger of jamming.
A sufficient distance to furniture, walls, window ledges etc. must be maintained - risk of damage.
Only activate the hand control for patients after detailed instruction - critically assess the suitability of the patient.
Never unlock the hand control for children or handicapped persons without surveillance.
Do not leave patients in shock position (Trendelenburg position) or head down position without surveillance - risk of loss of consciousness and/or slipping out of the couch.
The mains plug must always be accessible - EMERGENCY STOP.
Use potential equalization - danger of deviating or even wrong measured values when using electronic medical devices on patients despite electromagnetic compatibility (EMC).
Not for use in potentially explosive areas - danger of explosion. The control unit and the drives of the couch are not explosion-proof. The battery may release gases which, together with other gases, may form explosive mixtures.
Do not clean in washing facilities or with a high pressure cleaner - risk of electric shock / damage. Electrical components are splash-proof only.
Maintenance periods or deadlines for safety inspections by the operator must be observed.
Never continue to use a defective couch. The couch must be clearly marked as "DEFECT". Inform the operator of any need for repairs.
Instructor
The instructor confirms with his or her signature that the corresponding participants have been instructed by him or her in the use of the medical device in accordance with the instruction manual.
Last name First name
Company / Department
Date / Place: Stamp / Signature:
Participant
The participant confirms with his or her signature that the content of the instruction has been understood.
Last name First name
Company / Department Signature
Last name
Company / Department
Last name
Company / Department
First name
Signature
First name
Signature
Last name
Company / Department
© 2019 GREINER GmbH
First name
Signature
Instruction manual - MULTILINE NEXT DC 67 / 72
9 Appendix
9.6
Glossary
The following groups of persons and terms are defined in this glossary.
Operator
Under the German Medical Device Operator Ordinance (also abbreviated as MPBetreibV), the operator of a medical device is any natural or legal person who is responsible for the operation of the health facility in which the medical device is operated or used by its employees.
An operator (e.g. hospital, dialysis centre, hospital, hospital management, medical practice, company doctor etc.) can be the legal owner of the MULTILINE NEXT DC couch, as well as have actual control over it. The operator is responsible for the safe operation of this medical device.
User
Under the MPBetreibV, the user is the person who uses a medical device on the patient.
Users (e.g. medical specialists, nurses, doctors, assistants, nursing staff, caregivers etc.) are persons who are authorised by their training, experience or instruction to operate the couch independently or to carry out their work with it, or who have been instructed in the handling of the couch. Furthermore, they can recognise and avoid possible dangers and assess the clinical condition of the patient.
Patient
A patient is a person who is ill, in need of care or handicapped and who is placed on this couch.
Service personnel
Service personnel (e.g. service technicians, specially trained employees of a service support point, customer service employees of GREINER GmbH etc.) are persons who are authorised to carry out maintenance or repair work beyond that performed by the user. They are entitled to carry out the necessary maintenance or repair work on their own responsibility on the basis of their training, experience or through further instruction. They can identify and avoid any hazards that may occur during maintenance or repair work.
Healthcare facility
Under the MPBetreibV a health facility is any facility, body or institution, including rehabilitation and care facilities, in which medical devices are professionally operated or used by medical personnel, persons in the nursing profession or other authorised persons.
68 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
9
9.7
Appendix
Notes
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 69 / 72
9 Appendix
70 / 72 Instruction manual - MULTILINE NEXT DC © 2019 GREINER GmbH
9 Appendix
© 2019 GREINER GmbH Instruction manual - MULTILINE NEXT DC 71 / 72