TD 783 / TDS 890
Do not sterilise
TD 783 1600161-001
TD 783 1600162-001
TDS 890 1600163-001
1/8“ (3.17 mm)
2.5 bar
2V
4V f ig. 1
Typ 3 / ISO 1797
Code 4-6 / ISO 6360/1 min. 11 mm
ø 1.6 mm max. 25 mm fig. 2 fig. 6
4-hole REF 1600243-001
3-hole REF 1600242-001
REF 1302401-010
Instruction
REF 2100038-ENG/12.10
English
Devices entirely made in Switzerland by
Bien-Air Dental SA.
Type
Straight laboratory turbine with or without spray.
Intended use
Product intended for professional use only. Use in for laboratory work. Any use other than that for which this device is intended is prohibited and may prove dangerous.
Technical data
Coupling
2- or 4 hole pressure gauge and direct fig. 1 .
Recommanded rotation speed
Up to 300,000 rpm.
Weight
60 g (2,11 oz).
Air consumption
35 normliter/min on average.
Air spray rate
2 normliter/min to 200 kPa (29 psi).
Water spray rate
100 ml/min to 200 kPa (29 psi).
Information
The technical specifications, illustrations and dimensions contained in these instructions are given only as a guide. They may not be the subject of any claim. The manufacturer reserves the right to make technical improvements to its equipment, without amending these instructions. For all additional information, please contact Bien-Air Dental SA at the address indicated on the back cover.
Using
Air pressure output
Do not use the turbine at above the required air pressure
For use without unit:
Adjust the required air pressure to 2.5 bar (36.28
psi) with the assistance of a Bien-Air Dental inspection pressure gauge - 3-hole or 4-hole version placed between hose and connection, fig. 1 .
For use with the S001 unit:
Please refer to the indications of the unit’s manometer. The required air pressure is 3 bar (44 psi).
Important
The device must not be started without a bur inserted into the chuck.
Bur
Diameter of shaft 1.60 mm (0.06 in), type 3 in accordance with ISO 1797; max. length 25 mm (0.98
in), code 4-6 in accordance with ISO 6360-1 fig. 2 .
Inserting the tool
Place the bur in the turbine’s chuck fig. 3 . Then insert the turbine head in the bur securing tool, press the bur against the metal disc in the bur securing tool and press firmly until you hear a «clic» fig. 4 .
Extracting the tool
Hold the central part of the straight turbine’s handle firmly in one hand and press the rear part of the handle forward. Release and remove the bur fig. 5 .
Seals replacement
2V: Remove the seal according to fig. 6
4V: Remove the seal according to fig. 7
Removing seal, place it on the tubes and remount the attachment fig. 1
Maintenance
Non-sterilisable
Only use maintenance products and components from Bien-Air Dental. The use of other products and components can void the guarantee.
Servicing
Never disassemble the device. For any modification and repair, we recommend that you contact your regular supplier or Bien-Air Dental directly.
Bien-Air Dental asks the user to have its dynamic instruments checked or inspected at least once a year.
Cleaning
Use Spraynet cleaner fig. 8 . Spray the inside and outside of the equipment to remove the residues and carefully clean the surfaces.
Never immerse in disinfectant solutions.
Not suitable for ultrasonic bath.
Lubrication
Lubrication at least twice a day .
Only use Lubrimed medical grease and greaser.
• First-time use: Remove the cap, screw the knurled back-housing until the grease appears at the end of the rod. Wipe off surplus grease and replace cap after use.
• Remove the bur and introduce the rod of the greaser to the bottom. The dose corresponds to one half-turn of the knurled backhousing. Use the guidemarks fig. 9 .
• Replace a bur, turn on the air and run the turbine for about 15 seconds.
fig. 3
TD 783
TDS 890 fig. 4 fig. 5 fig. 9
REF 1000005-001
Important
After cleaning and disinfecting the instrument and before using it, operate at moderate speed, with a bur in the chuck, for 10 to 15 seconds to distribute, and remove, the excess lubricant.
Cleaning of spray tubes fig. 10
Wire cleaner.
Transport and storage conditions
Temperature between -40°C (-40°F) and 70°C
(158°F), relative humidity between 10% and 100%, atmospheric pressure 50 kPa to 106 kPa (7.3 to 15.3
psi).
Other precautions for use
The device must be used by a qualified person in accordance with the current legal provisions concerning industrial safety, health and accident pre-
1/2
TD 783 / TD 783
TDS 890 4-hole
2-hole
STS Trimmer
Pressure of the pressure gauge in bar
–
4,8
S001
Pressure of the pressure gauge in bar
3
–
Without module
(on the unit) in bar
2,5
2,5 fig. 7 fig. 8 fig. 10
REF 1302403-010
Lubrimed
REF 1600037-006
1 sec.
vention measures, and these working instructions.
In accordance with these requirements, the operators:
• must only use operating devices that are in perfect working order; in the event of irregular functioning, excessive vibration, abnormal heating or other signs indicating malfunction of the device, the work must be stopped immediately; in this case, contact a repair centre that is approved by Bien-Air Dental;
• must ensure that the device is used only for the purpose for which it is intended, must protect themselves, their patients and third parties from any danger, and must avoid contamination through the use of the product.
Rest the device on a suitable support in order to eliminate the risk of injury to you yourself and to third parties.
Recommendations
It is essential to use dry, purified compressed air in order to ensure the long working life of the device. Maintain the quality of the air and the water by regular maintenance of the compressor and the filtration systems.
The use of unfiltered hard water will lead to early blockage of the tubes, connectors and spray cones.
After a period of prolonged use, the device must be cleaned and serviced according to these instructions.
REF 2100038-ENG/12.10
1/2
Adhere to the Instructions for Use, in accordance with the instructions of the tool manufacturer. Never use a tool with an incorrect rod, as there is a risk of this becoming detached during the use and causing injury to you yourself and to third parties.
The device is not intended for medical treatment; any use other than that for which this product is intended is not allowed and may proved to be dangerous. The device meets all the current European legal requirements (according to the European
Directive 93/42/EEC relating to medical devices).
Guarantee
Terms of guarantee
Bien-Air Dental grants the user a guarantee covering all functional defects, material or production faults.
The device is covered by this guarantee for
12 months from the date of invoicing.
Bien-Air Dental offers a 24-month guarantee for the glass-bar light conductors.
In the event of justified claim, Bien-Air Dental or its authorised representative will fulfil the company’s obligations under this guarantee by repairing or replacing the product free of charge. Any other claims, of whatever nature, in particular in the form of a claim for damages and interest, are excluded.
Bien-Air Dental shall not be held responsible for damage or injury and the consequences thereof, resulting from:
• excessive wear and tear
• improper use
• non-observance of the instructions for installation, operation and maintenance
• unusual chemical, electrical or electrolytic influences
• poor connections, whether of the air, water or electricity supply.
The guarantee does not cover flexible “fibre optic” type conductors, or any parts made of synthetic materials.
The guarantee shall become null and void if the damage and its consequences are due to improper manipulation of the product, or modifications to the product carried out by persons not authorised by Bien-Air Dental.
Claims under the terms of the guarantee will be considered only on presentation, together with the product, of the invoice or the consignment note, on which the date of purchase, the product reference and the Serial No. should be clearly indicated.
REF Legend
1600161-001 TD 783 turbine for hose with internal Ø 1/8" (3,17mm), with Friction Grip chuck (FG)
1600162-001 TD 783 turbine with 2-hole connection, Friction Grip (FG)
1600163-001 TDS 890 turbine with spray, 4-hole connection,
Friction Grip (FG)
1000005-001 Friction Grip bur-locking tool for TD 783 and TDS 890
1302401-010 Seal 3V
1302403-010 Seal 4V
1600242-001 Gauge for 3-hole attachment
1600243-001 Gauge for 4-hole attachment
1600037-006 Lubrimed medical grease cartridges, box of 6 cartridges
1000003-001 Lubrimed greaser
1600036-006 Spraynet, cleaning spray 500 ml, box of 6 cans
List of Bien-Air Dental SA registered trade mark products ®:
Aquilon ®
Bora ®
Boralina ®
ChiroPro ®
Diamline
Eolia
Gyro
®
®
Gyrolina
®
®
Isolite
Lubrifluid
Lubrimed
MX2 ®
®
®
®
MX ®
PowerCare
Prestilina
Spraynet ®
®
®
In these instructions, “Device” corresponds to the product described in the heading “Type”. For example, turbine, contra-angle, handpiece, micromotor, tube, electronics, connectors, station etc.
Manufacturer.
Wear rubber gloves.
Attention.
CE Marking.
Move in the direction indicated.
Move fully to the stop, in the direction indicated.
This product may be covered by one or more of the following patents:
EP Europe : 745358 / 688539 / 948294 / 1145688 / 1563800 / 1563801 / 1675523 / 1753360 DE Germany : 29616023.7
DK Denmark : 9600315 FR France : 2722972 CH Switzerland : 693922 CN China: 100528099 / 100522100 / 100522099 /
100553584 JP Japan : 3892485 / 4298933 / 7000419 US United-States : 5453008 / 6033220 / 6319003 / 7214060 / 7448870
RU Russia : 2361540 / 2361541 / 2372046
REF 1600161-001 TUR TD783 / REF 1600162-001 TUR TD783 2H / REF 1600163-001 TUR TDS890 4H
Set supplied
TD 783
REF 1600161-001
Optional accessories
10x
REF 1000001-010
REF 1000005-001
10x 10x
REF 1302401-010 REF 1302403-010
4-hole
REF 1600243-001
REF 2100038-ENG/12.10
TD 783
REF 1600162-001
3-hole
REF 1600242-001
TDS 890
REF 1600163-001
REF 1000003-001
6x
REF 1600037-006 REF 1600036-006
Bien-Air Dental SA
Länggasse 60
Case postale
2500 Bienne 6, Switzerland
Tel. +41 (0)32 344 64 64
Fax +41 (0)32 344 64 91 [email protected]
Bien-Air Deutschland GmbH
Jechtinger Strasse 11
79111 Freiburg, Deutschland
Tel. +49 (0)761 45 57 40
Fax +49 (0)761 47 47 28 [email protected]
Bien-Air España SA
Entença, 169 Bajos
08029 Barcelona, España
Tel. +34 934 25 30 40
Fax +34 934 23 98 60 [email protected]
Bien-Air USA, Inc.
Medical Technologies
5 Corporate Park
Suite 160
Irvine, CA 92606 USA
Phone 1-800-433-BIEN
Phone 949-477-6050
Fax 949-477-6051 [email protected]
Bien-Air France Sàrl
55-57, avenue Jean Lolive
93508 Pantin Cedex, France
Tel. +33 (0)1 41 83 60 70
Fax +33 (0)1 48 96 07 40 [email protected]
Bien-Air Italia s.r.l.
Via Vaina 3
20122 Milano, Italia
Tel. +39 (02) 58 32 12 51/52/54
Fax +39 (02) 58 32 12 53 [email protected]
Bien-Air UK Limited
63, The Street Capel,
Surrey RH5 5JZ, England
Tel. +44 (0)1306 711 303
+44 (0)1306 712 505
Fax +44 (0)1306 711 444 [email protected]
Bien-Air Asia Ltd.
Nishi-Ikebukuro
Daiichi-Seimei Bldg. 10F
2-40-12 Ikebukuro, Toshimaku
Tokyo, 171-0014, Japan
Tel. +81 (3) 5954-7661
Fax +81 (3) 5954-7660 [email protected]
Beijing Bien-Air
Medical Instrument
Technology Service Co. Ltd.
Room 907, The Exchange Beijing,
No 118 Jian Guo Lu Yi,
Chao Yang District,
Beijing 100022, China
Tel. +86 10 6567 0651
Fax +86 10 6567 8047 [email protected]
2/2
![](http://s2.manualzz.com/store/data/067036515_1-ec17742ebb10489b5fb97332aff124f0-210x147.png)
Advertisement