Precision Medical PM60E Easy Vac Aspirator User manual
Below you will find brief information for Easy Vac Aspirator PM60E. The Easy Vac Aspirator PM60E is a medical device designed to provide a continuous vacuum source for aspirating fluids during medical suctioning procedures. It features a user-friendly design with an adjustable vacuum control knob, inline hydrophobic filter, and a standard 800cc suction canister with float shutoff. The device is equipped with a thermal limit switch for safety and comes with a 5-year warranty.
Advertisement
Advertisement
U S E R M A N U A L
Model No. PM60E
CAUTION
SAVE THESE INSTRUCTIONS
Federal (USA) law restricts this device to sale by or on the order of a physician.
300 Held Drive Tel: (+001) 610-262-6090
Northampton, PA 18067 USA Fax: (+001) 610-262-6080 www.precisionmedical.com
RECEIVING / INSPECTION
Remove the Precision Medical, Inc. Easy Vac Aspirator from packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider.
INTENdEd USE
To supply a continuous vacuum source adequate within the stated operating vacuum range, to aspirate fluids in medical suctioning procedures by trained healthcare professionals.
REAd ALL INSTRUCTIONS BEFORE USING
This manual instructs a Professional to install and operate the Easy Vac
Aspirator. This is provided for your safety and to prevent damage to the
Easy Vac Aspirator. If you do not understand this manual, DO NOT USE the Easy Vac Aspirator and contact your Provider.
ExPLANATION OF ABBREVIATIONS
inHg mmHg
Inches of Mercury
Millimeters of Mercury
SAFETY INFORMATION - WARNINGS ANd CAUTIONS
dANGER
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
Used without the safety alert symbol indicates a potentially hazardous situation which, if not avoided, may result in property damage.
CONSULT ACCOMPANYING DOCUMENTS
Type B equipment
1
dANGER
To reduce the risk of Electrocution:
• Always unplug the Easy Vac Aspirator immediately after using.
• dO NOT use while bathing.
• dO NOT place or store Aspirator where it can fall or be pulled into a tub or sink.
• dO NOT place or drop into water or other liquid.
• dO NOT reach for a Aspirator that has fallen into water. Unplug immediately.
• dO NOT attempt to repair Aspirator. Any attempt could result in electrical shock.
• dO NOT use near any type of flame or flammable/explosive substances.
• dO NOT allow Aspirator to exceed Operating or Storage Temperature specifications.
• Caution must be exercised when using this Aspirator where oxygen is being administered.
• The Easy Vac Aspirator is not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen or nitrous oxide.
WARNING
To reduce risk of burns, electrocution, fire, or injury to persons:
• The Easy Vac Aspirator should never be left unattended when running.
• Close supervision is necessary when the Aspirator is used by, on, or near children or invalids.
• Use the Easy Vac Aspirator only for its "Intended Use" as described in this manual.
• dO NOT use attachments not recommended by the Precision Medical, Inc.
• dO NOT use when sleepy or drowsy.
• dO NOT use if Easy Vac Aspirator has a damaged cord or plug.
• dO NOT use if Easy Vac Aspirator has been dropped or damaged.
• dO NOT cover air vents. Restriction of air flow will cause Aspirator to overheat, causing the thermal reset to shut down the Aspirator.
• dO NOT drop or insert any object into any opening or hole.
• Connect this Aspirator to a properly grounded outlet only.
See “GROUNDING INSTRUCTIONS”.
• Keep the cord away from heated surfaces.
2
CAUTION
Ground reliability can only be achieved when the power cord is connected to an equivalent receptacle marked "Hospital Grade".
SPECIFICATIONS
Operating Vacuum Range: 0-26 inHg (0-660 mmHg)
dimensions:
Length:
Width:
13.25 in
8.00 in
(33.7 cm)
(20.3 cm)
Weight:
Shipping Weight:
Electrical:
Height:
12.6 lbs
9.75 in
(5.7 kg)
14.5 lbs (6.6 kg)
(24.8 cm)
207 - 253 VAC, 50 Hz, 2.4 Amps
Operating Temperature Range: 50°F to 104°F (10°C to 40°C)
Storage Temperature Range: -4°F to 104°F (-20°C to 40°C)
Max 95% Noncondensing
Filter: Inline, Hydrophobic
Safety Systems: Automatic Resetting Thermal Limit Switch
Equipment Classification:
Class I
Type B
IPX0
- with respect to protection from electrical shock
- degree of protection against electrical shock
- degree of protection against ingress of liquids
Mode of Operation - continuous
The Easy Vac Aspirator meets the EMC requirements of EN 55011:2007
.
Specifications are subject to change without prior notice.
3
COMPONENT dESCRIPTION
CAUTION
If for any reason any label becomes illegible or lost, contact
Precision Medical, Inc.
Vacuum
Gauge
Patient
Tubing
Connection
Control
Knob
On/Off
Switch
Suction Feet
Standard 800cc
Suction Canister with Float Shutoff
Inline
Hydrophobic
Filter
Power Cord
Power Cord
Keeper
4
GROUNdING INSTRUCTIONS
The Easy Vac Aspirator should be grounded. In the event of an electrical short circuit, grounding reduces the risk of electrical shock by providing an escape wire for the electrical current. This Aspirator is equipped with a cord having a grounding wire with a grounding plug. The plug must be plugged into an outlet that is properly installed and grounded.
dANGER
Improper use of the grounding plug can result in a risk of electrical shock.
WARNING
Ground reliability can only be achieved when the power cord is connected to an equivalent receptacle marked "Hospital Grade".
OPERATING INSTRUCTIONS
CAUTION
Inspect the Easy Vac Aspirator for visual damage, DO NOT USE if damaged.
1. Place Easy Vac Aspirator near a properly grounded electrical receptacle on a table, floor or other horizontal, smooth, hard surface.
Allow a minimum of 4 in. (10 cm) clearance on all sides to ensure adequate air flow for cooling of Aspirator.
2. Press down on Aspirator until Suction Feet are secure.
3. Be sure the On/Off Switch is in the “OFF” position.
4. Be sure the Inline Hydrophobic Filter is properly connected to the vacuum port on the Suction Canister lid.
• The Inline Hydrophobic Filter prevents liquid from entering the vacuum pump.
5
CAUTION
dO NOT run the Easy Vac Aspirator without Inline
Hydrophobic Filter. This will void the warranty.
5. Plug the Aspirator into a properly grounded receptacle.
6. Turn the On/Off Switch to the “ON” position.
7. Block the patient connection on the Suction Canister and adjust the Vacuum Control Knob to the desired vacuum.
To INCREASE vacuum - Turn knob CLOCKWISE
To DECREASE vacuum - Turn knob COUNTERCLOCKWISE
• Vacuum should be set as prescribed by Provider.
8. Connect one end of the Patient Suction Tubing to the port on the
Canister marked Patient Tubing.
9. Connect the other end of the Patient Suction Tubing to the suction catheter and begin procedure as instructed by Provider.
10. When procedure is completed, press the On/Off Switch to the
“OFF” position.
• DO NOT overfill the canister
11. Empty the Suction Canister as follows:
• Remove the canister from its bracket and empty as instructed by your Provider.
CAUTION
Canister is for single patient use only.
12. Replace Suction Canister and reconnect all suctioning tubing.
6
CLEANING / MAINTENANCE
WARNING
Turn the Easy Vac Aspirator off and disconnect from any external power source before cleaning.
Inline Hydrophobic Filter:
This filter cannot be cleaned. If it becomes contaminated or clogged, it must be replaced.
To replace Inline Hydrophobic Filter:
1. Disconnect filter assembly from Suction Canister and remove Suction
Canister from the Aspirator.
2. Remove filter assembly from the Aspirator.
3. Dispose of filter.
4. Install new filter (Part #1690) and reconnect tubing and elbow.
Tubing is single patient use only. DO NOT clean.
Cleaning for Exterior Components:
1. Disconnect the Aspirator from power source and other connections before cleaning.
2. Use a cloth or sponge dampened with soap and water.
3. Wipe dry with a clean cloth or paper towel.
4. Store the Aspirator in a clean area free from grease, oil, and other sources of contamination.
RETURNS
Returned products require a Returned Goods Authorization (RGA) number, contact Precision Medical, Inc. All returns must be packaged in sealed containers to prevent damage. Precision Medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy available on the Internet, www.precisionmedical.com.
Manuals are available on our website; www.precisionmedical.com
7
dISPOSAL INSTRUCTIONS
Dispose of Bio-waste products as described by your Healthcare
Professional or Healthcare Facility protocol.
Please Recycle
WARNING
If Aspirator is used for treatment of infectious disease, consult your Physician or Healthcare Professional for recommended procedures for proper disposal.
REPLACEMENT PARTS description
Patient Suction Tubing
Filter, Hydrophobic
Filter Assembly (with tube & elbow)
Elbow Canister Outlet
800cc Canister, (Case-10 ea.)
Part #
1955
1690
502012
1616
502519-10
8
TROUBLESHOOTING
If the Easy Vac Aspirator fails to function, consult the Troubleshooting
Guide below. If the problem cannot be corrected, consult your Provider.
Problem
Aspirator will not operate
Probable Cause
1. On/Off switch in
“OFF” position
2. Power Cord not connected properly
3. Thermal limit exceeded
Remedy
1. Push On/Off switch to “ON” position
2. Check plug at outlet
NO vacuum on gauge / NO vacuum at Patient
Connection Tubing
NO vacuum on gauge / vacuum at Patient
Connection Tubing
Vacuum on gauge
/ NO vacuum at
Patient Suction
Tubing
Unable to adjust vacuum
1. Hydrophobic Filter not connected properly
2. Suction Canister not connected or assembled properly
3. Defective Suction
Canister
4. Defective Regulator
1. Defective vacuum gauge
1. Clogged
Hydrophobic filter
2. Canister overfilled
3. Shut Aspirator off, wait 15 minutes; be sure vents are clear
1. Confirm and tighten connections
2. Confirm and tighten
3. Replace Suction
Canister
4. Contact your
Provider
1. Contact your
Provider
1. Replace
Hydrophobic filter
2. Empty canister
1. Defective regulator 1. Contact your
Provider
9
LIMITEd WARRANTY
ANd
LIMITATION OF LIABILITY
Precision Medical, Inc. warrants that EASY VAC Aspirator
(the Product) will be free of defects in workmanship and/or material for the following period:
Five (5) years from shipment
Should any failure to conform to this warranty appear within the applicable period,
Precision Medical, Inc. shall, upon written notification thereof and substantiation that the goods have been stored, installed, maintained and operated in accordance with Precision Medical, Inc.’s instructions and standard industry practice, and that no modifications, substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.
The representatives of Precision Medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale. Thus, this writing is a final, complete and exclusive statement of the terms of that contract.
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF
MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER
WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.
Precision Medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property. Correction of non-conformities as provided above shall constitute fulfillment of all liabilities of Precision Medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision Medical, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or specifications without notice.
Precision Medical, Inc. reserves the right to correct clerical or typographical errors without penalty.
Precision Medical, Inc.
300 Held Drive
Northampton, PA 18067 USA
Toll Free Tel: 800-272-7285
Toll Free Fax: 800-353-1240
Tel:(+001) 610-262-6090
Fax:(+001) 610-262-6080
ISO 13485 Certified
506958 Rev0 2/25/11 P rinted in
USA