DataView 3 User manual

Add to my manuals
229 Pages

advertisement

DataView 3 User manual | Manualzz

ATS400

ETL DataView 3

User Manual

Contents

Contents

1 Introduction

2 Administration

2.1

System setup

2.1.1

2.1.2

2.1.3

2.1.4

2.1.5

2.1.5.1

2.1.5.1.1

2.1.5.1.2

2.1.5.2

2.1.5.2.1

2.1.5.2.2

2.1.5.2.3

2.1.5.2.4

2.1.5.2.5

2.2

Windows users

Installing ETL DataView 3

Installing the PDF creator

Setting up the PDF creator

Adding to a network

Variants X6 or X8

Create network drive

Setup a network printer

Variants X4 oder X5

Network setup

Checking the network connection

Saving the settings

Checking the share

Configuration

Configuration

2.2.8.1

2.2.8.2

2.2.8.3

2.2.8.4

2.2.8.5

2.2.9

2.2.9.1

2.2.9.2

2.2.9.3

2.2.9.4

2.2.9.5

2.2.10

2.2.11

2.2.12

2.2.13

2.2.1

2.2.2

2.2.3

2.2.4

2.2.5

2.2.6

2.2.6.1

2.2.6.1.1

2.2.6.1.2

2.2.6.1.3

2.2.6.1.4

2.2.6.1.5

2.2.6.1.6

2.2.6.1.7

2.2.7

2.2.8

Setting up language

Naming the workstation

Signal tones

Safety cabinet

Startmode

Test plan selection

Examples

Article number

Article groups

Article- and serial number

Article groups and serial number

Combined entrance

Combined entrence with article groups

Variable sequence in one field

Dummy test

User administration

Add an new user

Enter password

Delete user

Assign rights

Automatic logout

File storage

Result files

Test plans

Network

Temporary files

Tool

Serial interface

Push button

Change voltage

Dummy test plan

User Manual ATS400 Version 3.14.41 - 1

6

7

7

37

37

41

43

45

51

53

21

22

22

31

15

15

7

8

78

79

80

80

82

83

83

73

74

74

77

72

72

72

72

67

68

69

71

63

64

65

66

58

60

61

62

54

55

55

57

2

Contents

2.2.14

3

3.1

3.2

3.3

3.3.1

3.3.1.1

3.3.1.2

3.3.1.3

3.3.1.4

3.3.1.5

3.3.1.6

3.3.1.7

3.3.2

3.3.3

3.3.4

3.3.5

3.3.6

3.3.7

3.3.7.1

3.3.7.2

3.3.7.3

3.3.7.4

3.3.8

3.3.9

3.3.10

3.3.11

3.3.12

3.4

3.4.1

3.4.2

3.4.3

3.4.4

3.4.5

4

4.1

4.1.1

4.2

4.2.1

4.2.2

4.3

Printserver configuration

Test plan editing

Administering test plans

Changing test plans

Test types

Common dialogs

Ramp

Start conditions

Matrix

Jump

Text

Multitest

Create log

Protective Earth test

High voltage AC

High voltage DC

Insulation test

Function test

Leakage current test

Testing modules

Supply options

Polarity for protective conducter

Polarity for touch current

Batch

Data input

Resistance

Matrix

Continuity test

Test plan settings

Text

Identification

Report options

Info

Options

Inspector

Test individual

Protective earth test

Test plan

Manual test plan selection

Automatic test plan selection

Safety cabinet

5 Report creation

5.1

Creating templates

5.1.1

5.1.1.1

5.1.1.2

5.1.1.2.1

Key words

Key words of the results data

Key words of the test step parameters

Key words of the Protective earth test type

User Manual ATS400 Version 3.14.41 - 1

146

146

146

146

149

151

87

90

90

92

93

103

104

105

107

108

110

115

117

98

99

100

102

94

94

95

97

119

121

124

125

126

127

129

130

131

132

133

137

138

139

139

140

141

142

143

144

3

Contents

5.1.1.2.2

5.1.1.2.3

5.1.1.2.4

5.1.1.2.5

5.1.1.2.6

5.1.1.2.7

5.1.1.2.8

5.1.1.2.9

5.1.1.3

5.1.1.4

5.1.1.5

5.1.1.6

5.1.2

5.1.2.1

5.1.3

5.1.3.1

Key words of the High Voltage AC test type

Key words of the High voltage DC test type

Key words of the Insulation test type

Key words of the FCT-Current test type

Key words of the Leakage current test type

Key words of the Sight check test type

Key words of the Data input test type

Key words of the Batch run test type

5.1.1.2.10

Key words of the User-Interface test type

5.1.1.2.11

Key words of the Resistance test type

5.1.1.2.12

Key words of the Continuity test type

5.1.1.2.13

Key words of the PT100 test type

5.1.1.2.14

Key words of the Lightcontrol test type

5.1.1.2.15

Key words of the Dummy load test type

5.1.1.2.16

Key words of the Light Filament current test type

5.1.1.2.17

Key words of the External programm test type

Key words of the test step measurement values

Key words of the multiple test values test step

Key words of the test step advices

Key words of the test step process control

HTML templates

Creation with Microsoft Word 2012

CSV templates

Creation with Microsoft Excel 2010

6 Reference

6.1

6.2

6.2.1

6.3

Patterns

Protective earth test

Leading of the cable

High voltage test

6.3.1

6.3.1.1

6.3.1.2

6.3.1.3

6.3.1.4

6.4

Evaluation of the measurement

Error messages for high voltage AC

Error messages for high voltage DC

Test for required voltage

Current error at upper limit

Results files

6.4.1

6.4.2

6.4.3

6.4.3.1

6.4.3.1.1

6.4.3.1.2

6.4.3.1.3

6.4.3.1.4

Header

Identification

Teststeps

TestStep

Protective Earth

High voltage AC

High voltage DC

Insulation

6.4.3.1.5

6.4.3.1.6

6.4.3.1.7

6.4.3.1.8

FCT-Current

Leakage current

Sight check

Data input

6.4.3.1.9

Batch run

6.4.3.1.10

User-Interface

6.4.3.1.11

Resistance

6.4.3.1.12

Continuity test

User Manual ATS400 Version 3.14.41 - 1

158

159

159

160

160

160

161

162

152

153

154

155

157

158

158

158

162

163

163

166

167

168

172

173

176

176

177

178

180

199

202

204

205

207

208

210

212

182

183

183

183

188

190

193

196

180

180

180

181

181

181

4

Contents

6.4.3.1.13

PT 100

6.4.3.1.14

Light control

6.4.3.1.15

Dummy load

6.4.3.1.16

Light filament current

6.4.3.1.17

External program

6.4.4

6.4.5

Result

PlanOptionen

6.4.5.1

6.4.5.1.1

ReportOptionen

ReportOption

6.4.6

6.5

6.6

6.7

6.7.1

6.7.2

Validation

Plan files

Printserver configuration

Matrix

RelaisMatrix.cfg

ExtendedRelais.conf

225

225

225

226

226

227

214

216

217

219

221

223

224

224

224

User Manual ATS400 Version 3.14.41 - 1 5

Introduction

1 Introduction

This manual describes the functions of

ETL DataView 3

.

This manual is aimed at various roles, which use

ETL DataView 3

. This manual differentiates between the roles of

administration ,

test plan editing

,

inspector and

report creation

. Furthermore, the manual contains a

reference

about the tests, for the files used and created by

ETL DataView 3

.

Note

This manual has the status in progress. All details are correct to the best of knowledge, but may be incomplete. The information in this manual supplements and replaces the existing documentation.

All text in

Courier New blue

is either text as shown in the user interface or keywords in files.

Note

This manual is aimed to be used with an ATS400. When using devices of

Series 36 not all chapters are applicapable. There will be no note on this fact.

User Manual ATS400 Version 3.14.41 - 1 6

Administration

2 Administration

This part of the manual is aimed at system administrators.

It describes information that is necessary to set up and operate

ETL DataView 3

.

It describes activities that a system administrator must carry out if he wants to install

ETL DataView 3

on his own PC.

All instructions and screenshots refer to the operating system

Windows 7 SP1 32 bit

English.

Section you will only need if you are installing

ETL DataView 3

on your own computer:

Installing ETL DataView 3

Installing the PDF-Creator

Setting up the PDF Creator

You must also consider all other sections when changing the system environment, e.g.

want to integrate the

ATS400

in a network environment.

2.1

System setup

In this chapter the work is described which must be done by a

Windows

system administrator.

2.1.1

Windows users

On the variants X4 and X5 of the

ATS400

no users can be created on system level since Windows CE does not have users.

On the variants X6 and X8 of the

ATS400

one user with the name

ATS400

is created as adminstrator. This user has no password and is logged on automatically.

You can make the

ATS400

part of a domain and allow to log in with domain accounts.

You need additional users only when making the

ATS400

part of a network and the logged in user will use windows explorer or other applications using the network.

ETL DataView 3

will not make use of Windows users and does not support single sign on.

Important

If reports will be created and you will use multiple printers, e. g. for local printing and creating a PDF-file, the logged in user must have local adminstrative rights.

For each user logged into Windows the configuration of the PDF-Creator must be applied again, see

here

User Manual ATS400 Version 3.14.41 - 1 7

Administration

In chapter

Create network drive is described how to use a

ATS400

without adding to a domain to use test plans and store results in a network.

2.1.2

Installing ETL DataView 3

For the here describes installation of

ETL DataView 3

into the programs folder the following conditions must be met:

You must have local adminstration rights.

You must have

ETL DataView 3

on an data storage.

You must have aknowledgement of using

Windows Explorer

.

On a

Windows 7 64Bit

system the programs folder is named

Program Files (X86)

instead of

Program Files

.

Dependend on the settings of the User Account Control additional dialogs may be opened. These are not displayed in the following instruction. Additionally some buttons may be overlayed by the administrator symbol.

Create a new folder

ETL

in the programs folder of the PC.

User Manual ATS400 Version 3.14.41 - 1 8

Administration

Open using the contex menu the dialog

Properties

and step to the property page

Security

. Click on the button

Edit...

.

Click in the following dialog on the button

Add...

.

User Manual ATS400 Version 3.14.41 - 1 9

Administration

Enter

Everyone

and close the dialog with the button

OK

.

Activate for

Everyone

the checkbox

Full control

in column

Allow

.

Close both dialogs with the button

OK

.

User Manual ATS400 Version 3.14.41 - 1 10

Administration

Navigate to the folder

ETL

and create a new folder

DataView 3

.

Copy the files for

ETL DataView 3

from your data storage to

DataView 3

.

User Manual ATS400 Version 3.14.41 - 1 11

Administration

This folder should now contain the files like in the following screenshot.

User Manual ATS400 Version 3.14.41 - 1 12

Administration

Create a shortcut for

ETL DataView 3

.

Rename the shortcut into

ETL DataView 3

.

Allow

Windows Explorer

to show hidden files and folders. Open under

Organize

the

Folder and Searchoptions

. Step to property page

View

.

User Manual ATS400 Version 3.14.41 - 1 13

Administration

Activate the radio button

Show hidden files, folders and drives

. Close the dialog with the button

OK

.

To offer a start from the desktop copy the shortcut to the folder

C:

->

Users

->

Public

->

Public Desktop

.

To enable a start when a user logs in copy the shortcut to the folder

C:

->

ProgramData

->

Microsoft

->

Windows

->

Start Menu

->

Programs

->

Startup

.

Important

Before using

ETL DataView 3

you must configure the

serial port .

User Manual ATS400 Version 3.14.41 - 1 14

Administration

2.1.3

Installing the PDF creator

ETL Prüftechnik

uses the PDF creator in version 1.6.2 to create PDF documents.

The PDF creator is already installed on the ATS400 variants X6 and X8. For the

ATS400 variants X4 and X5, it is not possible to install the PDF creator.

You only need to carry out the activities in this section if you are installing

ETL DataView 3

on your own PC and want to use the creation of PDF files as the report option.

Download the PDF creator from the site ftp://ETL-FTP:[email protected]/

Tools/PDFCreator-1_6_2_setup.exe

.

Copy the downloaded setup program to the computer to be used. To this end, we recommend using the

C:\Setups

folder. If the device is connected to a company network, please disconnect the network connection by removing the cable. This means that you will not need to reply to additional dialogues, and no further unnecessary components will be installed.

Start the setup by right clicking on the file and selecting

Run as an administrator

.

Then follow the standard setup. Part of the dialogues are not displayed if the computer is not connected to a network.

2.1.4

Setting up the PDF creator

To automatically save the PDF files, the PDF creator needs to be set up accordingly. A corresponding printer profile is created.

Important

For every user logged into

Windows

, this setup will partly have to be made separately.

User Manual ATS400 Version 3.14.41 - 1 15

Administration

Start the setup by right clicking on the desktop on the PDF creator icon. Select

Run as an administrator

from the menu.

Select

Printer

->

Options

from the menu.

User Manual ATS400 Version 3.14.41 - 1 16

Administration

Create a new profile by clicking on the button with the

plus symbol

.

A window

Profile

opens, enter

ETL

and confirm with

Ok

.

User Manual ATS400 Version 3.14.41 - 1 17

Administration

In the list box, in addition to

Standard

,

ETL

is now also included as a profile. Select

ETL

to change the settings.

For the

ETL DataView 3

to automatically create PDF files, the

Auto-save

tab needs to be saved. The file name and the directory must be stated precisely in doing so.

User Manual ATS400 Version 3.14.41 - 1 18

Administration

Activate the

Use 'Auto-save'

checkbox.

Enter the

etl_dataview

value in the

Filename

field.

Activate the

Use this directory for auto-save

checkbox. Select the

C:\Program

Files\ETL\DataView 3\Printing

folder in the relevant field.

End the dialogue with the

Save

button. The window will close.

User Manual ATS400 Version 3.14.41 - 1 19

Administration

Select from the

Printer

->

Printers

menu.

In the

New printer name

field, enter

ETLPrinter

and confirm the

Add printer

button. If the

Add printer

button cannot be selected, the setup program was not started with administrator rights.

User Manual ATS400 Version 3.14.41 - 1 20

Administration

Change the profile of the ETLPrinter to

ETL

.

End the dialogue with the

Save

button.

2.1.5

Adding to a network

ETL DataView 3

is able to store test plans and result files in a mapped network drive.

You can use one share to store all files or two separate shares for test plans and result files. You can use different hosts for test plans and result files. Preparing you need the following details:

When using a fixed IP address:

- IP address of the DNS server, z. B. 10.2.1.50

- IP address of the locale system, z. B. 10.2.1.186

- Subnetmask of the network, z. B. 255.255.255.0

In all cases:

- Name of the host, e. g. etldats

- Name of the share for the test plans, e. g. etl_nobackup

- Name of the share for the result files, e. g. etl_nobackup

- Name of the domain, e. g. etl.local

- Name of the user with sufficent rights for the share

- Password of the user with sufficent rights for the share

To do the work you need a USB-keyboard an propably a mouse. Be aware that on the variants X4 and X5 you can use only the left USB connector of the system. To

User Manual ATS400 Version 3.14.41 - 1 21

Administration

connect more than one device you also need an USB-hub.

With the variants X4 or X5 the

integration is done in

ETL DataView 3

.

With the variants X6 or X8 the integration can be done in

ETL DataView 3

. It is also possilble to do the integration using a

network drive .

2.1.5.1 Variants X6 or X8

2.1.5.1.1 Create network drive

It is not neccessary to integrate the

ATS400

into a domain to load test plans or store result files on to a network drive.

You must know the domain name, the share name, the user and the password.

If you want to use any drive letter you map it in

Windows Explorer

.

Select in the folder view the entry

Computer

and choose for the menu bar

Map network drive...

. Enter into the following dialog your connection an user data.

Activate the checkboxes like displayed.

User Manual ATS400 Version 3.14.41 - 1 22

Administration

Alternativelly you can map the drive using a batch file in the startup folder. This integration must be done by your system administrator.

You cannot setup this configuration in

ETL DataView 3

. You must apply this settings manually. In the following example it is assusmed that you will store the test plans on drive N:\ and the result files on drive M:\.

Check in

ETL DataView 3

that under

Settings

->

File storage

->

Results

->

Storage

the radio button

Local

is activated.

User Manual ATS400 Version 3.14.41 - 1 23

Administration

Navigate in

Windows Explorer

to the folder

C:\Program Files\ETL\DataView 3

.

Open the file

Settings.conf

with the Windows Notepad.

Open the file with a double click and select in the opening dialog

Select a program from a list of installed programs

and close it with the button

OK

.

In the next dialog choose

Notepad

and close the dialog with the button

OK

.

User Manual ATS400 Version 3.14.41 - 1 24

Administration

Windows Notepad will open. Find the line with the entry

strPath_Plans

.

User Manual ATS400 Version 3.14.41 - 1 25

Administration

Change the text

Plans\

into

N:\

. Be carefull not to forget the ending character

\

.

User Manual ATS400 Version 3.14.41 - 1 26

Administration

Find the line with the entry

strPath_Results

.

User Manual ATS400 Version 3.14.41 - 1 27

Administration

Change the text

Results\

into

M:\

. Be carefull not to forget the ending character

\

.

Save the file and close Notepad.

User Manual ATS400 Version 3.14.41 - 1 28

Administration

In

ETL DataView 3

the entries are now visible under

Settings

->

File storage

.

User Manual ATS400 Version 3.14.41 - 1 29

Administration

User Manual ATS400 Version 3.14.41 - 1 30

Administration

2.1.5.1.2 Setup a network printer

If you want to use a network printer you must setup the printer in Windows.

Open

Devices and Printers

from the start menu. Chose from the menu bar

Add a printer

. In the opening dialog choose

Add a network, wireless or Bluetooth printer

.

User Manual ATS400 Version 3.14.41 - 1 31

Administration

Now the avalable network printers are searched and will be displyed in a list box.

Choose the printer from the list box.

User Manual ATS400 Version 3.14.41 - 1 32

Administration

You will be asked to install a driver. Choose the corresponding driver.

User Manual ATS400 Version 3.14.41 - 1 33

Administration

After the installation of the printer driver you can enter a name for the printer.

User Manual ATS400 Version 3.14.41 - 1 34

Administration

Do not share the printer.

User Manual ATS400 Version 3.14.41 - 1 35

Administration

After that dialog the printer will be diplayed in the list of the local printers.

User Manual ATS400 Version 3.14.41 - 1 36

Administration

2.1.5.2 Variants X4 oder X5

2.1.5.2.1 Network setup

After powering on the system end

ETL DataView 3

using the buttons

Workstation

,

Exit Dataview

and close the opening dialog with

Yes

.

Chosse from start menu

Settings

->

Network and Dial-up Connections

.

User Manual ATS400 Version 3.14.41 - 1 37

Administration

Then chosse the network adapter and open it with a double click.

User Manual ATS400 Version 3.14.41 - 1 38

Administration

In the opening dialog enter the desired settings. On the property page

IP Address

you choose using a DHCP-server or a fix IP address.

User Manual ATS400 Version 3.14.41 - 1 39

Administration

If you are using a fix IP address you must supply the address of the name server on the property page

Name Servers

.

Close the dialog using button

OK

.

User Manual ATS400 Version 3.14.41 - 1 40

Administration

2.1.5.2.2 Checking the network connection

Next you will check if the host can be reached over the network. For that the

ATS400

must be connected with the network. You need to open a

Command Prompt

window.

User Manual ATS400 Version 3.14.41 - 1 41

Administration

In the opening windows enter ping <hostname>. Replace <hostname> with the name of the host you will connect, in the example etldats. If there is a connection the answer is like displyed below.

There are two errors which can occur.

First ping cannot resolve the name of the host. In this case you get a message that the host could not be reached.

Second there is no connection. In this case data pakets are lost. In both cases contact your network admninistrator.

Close the window entering Exit.

User Manual ATS400 Version 3.14.41 - 1 42

Administration

2.1.5.2.3 Saving the settings

The settings are not persistant and must be saved. Open the

Run…

dialog.

User Manual ATS400 Version 3.14.41 - 1 43

Administration

In

Run

dialog enter ndcucfg and press the Enter key.

User Manual ATS400 Version 3.14.41 - 1 44

Administration

A new window will open. Enter reg save into the window and press the Enter key.

Close the window entering quit.

2.1.5.2.4 Checking the share

The share can now be checked. You must open

Command Prompt

again as described above.

Enter into the window net use Plans \\<Hostname>\<share name>. Replace

<Hostname> with the name of the host, in the example etldats, and <Share name> with the share name for test plans in the example etl_nobackup.

User Manual ATS400 Version 3.14.41 - 1 45

Administration

User Manual ATS400 Version 3.14.41 - 1 46

Administration

The dialog with the account data for logon will be opened. Enter user name, password and domain.

User Manual ATS400 Version 3.14.41 - 1 47

Administration

Next you will be asked if you want the data to be stored. If you store the data you will never be asked again for the logon data.

User Manual ATS400 Version 3.14.41 - 1 48

Administration

The successfully mapping wil be displayed.

If you get the message

Command failed: status 53

the share can not be used.

Contact in this case your network administrator.

User Manual ATS400 Version 3.14.41 - 1 49

Administration

Delete the mapping with net use Plans /d.

Close the window with exit.

User Manual ATS400 Version 3.14.41 - 1 50

Administration

2.1.5.2.5 Configuration

So

ETL DataView 3

can connect with the host it needs the corresponding information.

Enter these in

Settings

->

File storage

in the property pages

Network

->

Shared folder

->

Result

rsp.

Test plan

.

After entering the data you can establish a connection clicking on button

Connect

. If the connection is successfull the text changes from

Not connected

to

Connected

.

User Manual ATS400 Version 3.14.41 - 1 51

Administration

You can check a succefull connection by opening

Windows Explorer

.

User Manual ATS400 Version 3.14.41 - 1 52

Administration

In

Windows Explorer

click on the icon

Network

. The mappe dfolders will be displayed.

2.2

Configuration

This chapter covers work to be done for configuring

ETL DataView 3

.

User Manual ATS400 Version 3.14.41 - 1 53

Administration

2.2.1

Setting up language

Open the dialog choosing

Settings

->

Language

.

Select in the list the desired language. Click on button

OK

to apply the language. This setting will be stored in file

Settings.conf

. The language will be active immediately for most of the dialogs. The dialog will be closed and the window

Settings

will be displayed again.

With a click on the button

Menu

the change will be lost. The dialog will be closed and the window

Mainmenu

will be displayed again.

User Manual ATS400 Version 3.14.41 - 1 54

Administration

2.2.2

Naming the workstation

Open the dialog choosing

Settings

->

Workstation

->

Base settings

.

Click into the field right of

Name

. A keyboard window will be opened. Enter the name of the workstation. Click on one of the buttons

Back

or

Menu

. The settings will be stored in the file

Settings.conf

.

When

creating reports you can access the field with the keyword

WORKSTATION

.

2.2.3

Signal tones

Additional to the signal tones generated for the overall result also signal tones can be given for each test step. The duration of the tone is according to the setting for the overall result.

User Manual ATS400 Version 3.14.41 - 1 55

Administration

Open the dialog choosing

Settings

->

Workstation

->

Base settings

.

These setting is done by the 4 top checkboxes. Click on one of the buttons

Back

or

Menu

. The settings will be stored in the file

Settings.conf

.

Selection

Buzzer Individual test IO

Buzzer Individual test NIO

Buzzer Multi testIO

Buzzer Multi test NIO

Behaviour

Is the result of a test step IO a signal tone is generated. This is independent from the fact of the test step is part of a multi test step or not.

Is the result of a test step NIO a signal tone is generated. This is independent from the fact of the test step is part of a multi test step or not.

Is the result of a test step IO a signal tone is generated. This is done only when the test step is part of a multi test step.

Is the result of a test step NIO a signal tone is generated. This is done only when the test step is part of a multi test step.

User Manual ATS400 Version 3.14.41 - 1 56

Administration

2.2.4

Safety cabinet

When using a safety cabinet you can choose several settings. To use this settings it is neccessary to use a safety cabinet support the signals

IN OP

and

Locking

on the

ETL-Interfaces

of the

ATS400

.

This setting applies to the workstation and is valid for all test plans.

Open the dialog choosing

Settings

->

Workstation

->

Base settings

.

Choose a value from the list

Safety cabinet

. This will change the behaviour of the signal

Out 6

on the

ETL-Interface

. Click on one of the buttons

Back

or

Menu

. The settings will be stored in the file

Settings.conf

.

Selection

No cabinet

Locked during plan

Behaviour

The signal has the meaning

READY FOR

OPERATION

and behaves as described in the base device manual in chapter

ETL-

Interface for Selection and

Operation Panels

.

The signal has the meaning

Locking

. The safety cabinet will not open as long as a

User Manual ATS400 Version 3.14.41 - 1 57

Administration

Selection

Unlock on pass

Behaviour

test is running. After the test the safety cabinet can be opened independent from the overall result of the test.

The signal has the meaning

Locking

. The safety cabinet can not be opened as long as the overall result is not passed. Will the safety cabinet stay locked a message is displayed. You can unlock the safety cabinet with a button on the screen or the signal

Button Stop

on the

ETL-

Interface

.

2.2.5

Startmode

You can define which window will be displayed after starting

ETL DataView 3

.

Open the dialog choosing

Settings

->

Worksation

->

Startmode

.

Dependend from the selection in

Startup mode

additional controls may be displayed.

These are labeld as

Individual Test

or as

Plan

.

User Manual ATS400 Version 3.14.41 - 1 58

Administration

Selection

Main menu

Select plan by ID

Start in batch mode

Select plan manually

Individual test

Menu individual test

Plan

Behaviour

The main menu will be displayed.

The window

Select ID

will be displayed.

The window

Test batch

will be displayed.

The window

Select manual

will be displayed.

Additionally the selection

Individual

Test

will be displayed. You can select which test type will be started. You can only select enabled test types. The test will be started immediatelly after the start.

The window

Test individual

will be displayed.

Additionally the selection

Plan

will be displayed. You can select one of the test plans. This test plan will be running immediatelly after the start. Will the test plan be renamed or deleted you get a message and the main menu will be displayed.

Important

Having

user administration active all user should have the right to use the

window configured as start up window.

User Manual ATS400 Version 3.14.41 - 1 59

Administration

2.2.6

Test plan selection

You can determine how to enter data during automatic test plan selection.

Open the dialog choosing

Settings

->

Workstation

->

Program select

.

Dependend from the selection in

Select plan

are the controls operational.

With the drop down box

Select plan

you can choose from which source the selection will be taken.

Choosing

ETL-Interface

the test plan will be selected by applying a number at the

ETL-Interface and active the acquisition. Details are described in the base device manual.

Choosing

Keyboard

a window will be displayed. Into this window a ID will be entered which is used to select the test plan.

Choosing

File Plan.ID

a file with the name

Plan.ID

must be created in the plans folder. This file contains a line with the ID which is used to select a test plan. Optional in a second line a serial number and in a third line a article description can be given.

The contents of the file will be periodically checked.

Using this selection the checkbox

Delete Plan.ID

is operational. Is this check box active the file will be deleted after it has been processed.

User Manual ATS400 Version 3.14.41 - 1 60

Administration

The fields

Typ-Pattern

,

ID-Pattern

,

Serial-Pattern

and

Article-Pattern

as well as the corresponding check boxes are operational if

Keyboard

is selected in

Select plan

. With the other selections the controls are disabled.

In the field

Typ-Pattern

is the pattern which is used to test the entered data using automatic test plan selection. This pattern must be meet to have a valid entered data. The entered data will be in the result file as tag

TYPE

.

The check box

Combibox

indicates that the field

Typ-Pattern

contains two parts.

When

Combibox

is active the fields

ID-Pattern

and

Serial-Pattern

are getting operational, the check boxes

Enable

will both be activated and disabled.

In the field

ID-Pattern

is the pattern which will be used to extract the type from the entered data. The extracted value will be in the result file as tag

USEDID

.

The corresponding check box

Enabled

indicates that the entrance contains a type.

In the field

Serial-Pattern

is the pattern which will be used to extract a serial number. The extracted value will be in the result file as tag

SERIALNUMBER

.

The corresponding check box

Enabled

indicates that the entrance contains a serial number.

In the field

Article-Pattern

is the pattern which will be used to extract additional article data. The extracted value will be in the result file as tag

ARTICLE

.

The corresponding check box

Enabled

indicates that the entrance contains additonal article data.

Details how to enter pattern are in topic

Patterns .

2.2.6.1 Examples

The following examples shows possible settings at hand of the number scheme used by

ETL

. This scheme consist of a article number with 6 ciphers followed by a blank and the serial number. The serial number consists of 4 ciphers for the production month a blank and 4 ciphers for a number. This leads to following pattern:

000000\ 0000\ 0000

An example number is 205589 0515 1000. The corresponding aricle description is UGP-

5025.

Other examples will show additional possibilities.

User Manual ATS400 Version 3.14.41 - 1 61

Administration

2.2.6.1.1 Article number

Should for automatic test plan selection only the article number be used the pattern has to be given in field

Typ-Pattern

. The check boxes remain inactive and the other fields remain empty.

User Manual ATS400 Version 3.14.41 - 1 62

Administration

2.2.6.1.2 Article groups

Should for automatic test plan selection only the article number be used the pattern has to be given in field

Typ-Pattern

. The check boxes remain inactive and the other fields remain empty. The pattern must be entered that the characters for the group are included in paranthesis. In the test plan only the article group ist entered for the identification.

User Manual ATS400 Version 3.14.41 - 1 63

Administration

2.2.6.1.3 Article- and serial number

Should for automatic test plan selection the article number and the serial number in sepearate fields be entered both check boxes for

ID-Pattern

and

Serial-Pattern

must be activated. In the fields

Typ-Pattern

and

Serial-Pattern

the patterns are entered. There is no need to enter something in

ID-Pattern

because this pattern is identical with the pattern in

Typ-Pattern

. The check box

Combibox

remains inactive.

User Manual ATS400 Version 3.14.41 - 1 64

Administration

2.2.6.1.4 Article groups and serial number

Should for automatic test plan selection the article number and the serial number in sepearate fields be entered both check boxes for

ID-Pattern

and

Serial-Pattern

must be activated. In the fields

Typ-Pattern

and

Serial-Pattern

the patterns are entered. The pattern must be entered that the characters for the group are included in paranthesis. In the test plan only the article group ist entered for the identification.

There is no need to enter something in

ID-Pattern

because this pattern is identical with the pattern in

Typ-Pattern

. The check box

Combibox

remains inactive.

User Manual ATS400 Version 3.14.41 - 1 65

Administration

2.2.6.1.5 Combined entrance

Should for automatic test plan selection the article number and the serial number in a common field be entered the check box

Combibox

has to be activated and in all fields an entry has to be made for the patterns. The pattern in field

Typ-Pattern

must be entered that the characters for the article number are include in paranthesis. The sequence of the fields is fixed first the article number and then the serial number. The setting look like shown below.

User Manual ATS400 Version 3.14.41 - 1 66

Administration

2.2.6.1.6 Combined entrence with article groups

Should for automatic test plan selection the article number and the serial number in a common field be entered the check box

Combibox

has to be activated and in all fields an entry has to be made for the patterns. The pattern in field

Typ-Pattern

must be entered that the characters for the group are included in paranthesis. The sequence of the fields is fixed first the article number and then the serial number. The setting looks like shown below.

User Manual ATS400 Version 3.14.41 - 1 67

Administration

2.2.6.1.7 Variable sequence in one field

Is the sequence of the fields undetermined so the fields must be separated by a character. The pattern for the type and the serial number must be different.

Otherwise the fields cannot be separated. In this case the setting looks like shown below.

User Manual ATS400 Version 3.14.41 - 1 68

Administration

2.2.7

Dummy test

These settings must be done if you want to check the

ATS400

and the test station for functionality.

Open the Dialog using

Settings

->

Workstation

->

Dummy

.

Parameter

Program Start

User Change

Count

Interval

Time

Description

The dummy test will be done when the program starts.

The dummy test will be done after a user logs in.

The dummy test will be done after the given number of tests. When opening a test plan or changing it the counter will be reset to 0.

Time interval between two dummy tests. Checking for the end of the interval will be done when waiting for the start condition of the test step or when waiting for the identification of the unit under test. The intervall starts after a successfull dummy test.

Up to 5 points in time can be defined when a dummy test will be performed. Checking for the time will be done when

User Manual ATS400 Version 3.14.41 - 1 69

Administration

Parameter

Automatical dummy

Description

waiting for the start condition of the test step or when waiting for the identification of the unit under test. The times will be repeated for each day.

This checkbox must be set when an automatic dummy will be used.

Any combination can be configured despite of such a setting makes sense.

Important

The name of the result file and the report files are according to the same rules as with normal test plans.

Will in the file name not choose to use DateTime only the last dummy test will be recorded. Will the dummy test be executed after a certain amount of tests the result file and the report of the last test will be overwritten.

Will by creating the folder name a preset keyword or a keyword from the test type Data input be used invalid folder names may arise when they should be stored in subfolders.

Will

Program Start

be used together with

Time

the following rule applies:

The dummy test of the next point in time in the future will not be executed.

Example:

The points in time 8:00 and 16:00 are configured. The successfull dummy test when starting the program was at 7:55. Therefore the dummy test at 8:00 will not be executed. The next dummy test will be at 16:00.

Will

Program Start

be used together with

User Change

then after the start of the program and loging in of the user the dummy test will be executed only once.

Following settings are recommended if no other rules apply:

Scenario

Laboratory use or single shift

Device will be powerd off.

Two shifts without user administration

Device will be powerd off

Multiple shifts with user administration

Setting

Select

Programm Start

.

Select begin.

Select

Program Start

User Change

.

and two times when the shifts

User Manual ATS400 Version 3.14.41 - 1 70

Administration

Scenario

Different Windows user

Setting

Select

Programm Start

.

2.2.8

User administration

With the user administration you can limit the possibilities of the users in

ETL DataView 3

.

Open the dialog choosing

Settings

->

User administration

.

The check box

User administration

determins if you have user administration active. Is the check box active user administration is also active.

The check box

Enable anonymous user

determins if a user must log in. Is this check box active this user will be logged in automatically.

ETL DataView 3

has two user as default which cannot be deleted.

First this is ADMIN, who holds all rights as default. This user is provided to get access to the system without help from

ETL Prüftechnik

in the case the password has been forgotten. This user cannot be deleted. The factory default password for ADMIN is

ETL. This can be changed but it is recommended to keep it.

User Manual ATS400 Version 3.14.41 - 1 71

Administration

Second this is Anonymous which allows to log in without displaying a login dialog. The factory default is that this user has no rights.

Important

As factory default the anonymous user has no rights. He cannot run a test, make settings or exit

ETL DataView 3

. This should be taken into consideration when choosing the

startup mode

.

2.2.8.1 Add an new user

Click on the button

New

to create an new user.

A new windows is opened to enter the user name. leaving with

OK

the user will be created and he will be added to the list.

The user name and the password are case sensitive.

2.2.8.2 Enter password

Select the user in the list and click on the button

Password

to add or change the password.

The user name and the password are case sensitive.

The password is encrypted an cannot be restored.

2.2.8.3 Delete user

Select the user in the list. Click on the button

Delete

. A confirmation dialog is displayed. If you confirm the user will be deleted.

2.2.8.4 Assign rights

Select the user in the list. Activate or unactivate the check boxes on the right side to allow or deny the desired rights.

Important

Be carefull to give all users the right to use the window defined in

Startup mode .

User Manual ATS400 Version 3.14.41 - 1 72

Administration

2.2.8.5 Automatic logout

Open the dialog with the button

Autologout

.

In this dialog the rules are configured a user will be logged out automatically.

Selection

Time

Menu

Plan select

Start condition

Behaviour

Determines the time interval the user is logged off.

The time intervall is active when the user is in a menu window. All menus and windows reached via button

Settings

are not considered.

The time intervalis active when the user is in a window for test plan selection.

The time intervall is active when waiting to read the start conditions of the first step in a running test plan.

User Manual ATS400 Version 3.14.41 - 1 73

Administration

2.2.9

File storage

In this menu you can setup additonal to the storage places for

result files

and

test plans also the

network settings . Furthermore

temporary options for storage of result

files experiencing problems with the storage media and

release storage place on the

internal storage media.

Open the dialog choosing

Settings

->

File storage

.

2.2.9.1 Result files

On the property page

Result

you can define in which cases result files will be stored and on which media.

In the upper part of the property page

Storage

you can select in which cases result files will be stored. There are three possibilities and you can choose one or multiple of them.

With

Test IO

result files will be stored in the case the overall result is passed.

With

Test NIO

result files will be stored in the case the overall result is failed.

User Manual ATS400 Version 3.14.41 - 1 74

Administration

Wiht

Not tested (Error)

result files will be stored in the case the test was aborted by the user or by a problem arising from the system, e. g. overtemperature.

In the lower part of the property page

Storage

you can select the media to store the files on. Selecting

Local

the media of the system will be used. With variants X4 and

X5 this is the buildt in SD-Card and with variants X6 and X8 this is the internal hard disk.

On the property

Path

you can define the naming of a folder for the result files and the report files.

On the property page

Path

you can make additional settings for the folder where result files and report files are stored. If you select

_NONE

in all fields no subfolder is created.

You can select from the keywords

PlanName

,

PlanID

,

Workstation

,

Type

,

UsedID

,

SerialNumber

and

Article

.

PlanID

is the identifikation of the test plan and

PlanName

the name of the test plan.

Workstation

is the

name of the workstation

.

Type

,

UsedID

,

SerialNumber

and

Article

result from the automatic plan selection.

The checkbox

Create Subdirs

determins how the fields are used. Is the checkbox inactive one subfolder will be created where its name is concatenated from the

User Manual ATS400 Version 3.14.41 - 1 75

Administration

diferrent fields. Is the checkbox active for each field a new subfolder is created in a hierarchical manner.

The combined path is displayed in the text field at the bottom.

On the Property page

Filename

you can define the name of the result file and report file.

The above dispayed setting is factory default. The combined file name is displayed in the text field at the bottom.

User Manual ATS400 Version 3.14.41 - 1 76

Administration

2.2.9.2 Test plans

You can select the storage of the test plans.

On the property page

Test Plan

you can select the storage media. You can select from one of the medias

Local

,

USB

or

LAN

.

Is the checkbox

Save after Execution

active the test plan will also be stored after executing a test. This will preserve data entered by the tester as default when using the test plan again.

User Manual ATS400 Version 3.14.41 - 1 77

Administration

2.2.9.3 Network

On the property page

Network

you must make the settings when chossing

LAN

for result files or test plans. The settings have to be made in similar way for result files and test plans.

Refer to the chapters in

Adding to a network on how to use this dialog.

User Manual ATS400 Version 3.14.41 - 1 78

Administration

2.2.9.4 Temporary files

On the property page

Temp

you define when temporary stored result files and report files will be moved to their original destination folder.

This setting is only valid whe the storage media is set to

USB

or

LAN

. In this case there is the possibility that the storage media is not available when storing result files or report files. In this case the files will be stored locally. These files will be moved in the case the storage media is available again. Since there is the possibility that a larger amount of data must be moved and you cannot test during that activity you can define when and how much files will be moved. Factory default is

If destination available

.

With

If destination available

the files will be moved after the next test.

With

On exit testplan

the files will be moved the test plan will be exited.

With

On startup only (All files)

all files will be moved when

ETL DataView 3

is starting.

The radio button below

Files to move

are only valid when

If destination available

or

On exit testplan

is selected.

You can define that all files or a limited number of files should be moved.

User Manual ATS400 Version 3.14.41 - 1 79

Administration

2.2.9.5 Tool

On the property page

Tool

you can permanently delete files. There will be no confirmation dialog.

With the button

Delete Results

the localy stored result files and report files will be deleted.

With the button

Delete Temp Results

the temporary stored result files and report files will be deleted.

With the button

Delete Individual Tests

the settings for the single step test will be reset to factory default.

2.2.10 Serial interface

For

ETL Dataview 3

can connect to the device the serial interface must be configured. You must do this only when installing

ETL Dataview 3

on your own system.

User Manual ATS400 Version 3.14.41 - 1 80

Administration

Important

With the variants

ATS400

X8 and X6 the serial interface is set to COM2 as factory default. With the variants

ATS400

X5 and X4 the serial interface cannot be configured.

Open the dialog choosing

Settings

->

Workstation

->

COM

.

All available serial interfaces on the system will be displayed. Choose those interface the

ATS400

is connected with.

After clicking on the button

Connect

the serial interface will be opened and tried to connect to the device. Is a connectin established the LED

Remote

at the front of the

ATS400

is on.

Close the dialog with the button

Back

and exit

ETL DataView 3

.

User Manual ATS400 Version 3.14.41 - 1 81

Administration

2.2.11 Push button

Using a second front together with variant X2 it is nesseccary to deactivate the push button on the front.

Open the dialog choosing

Settings

->

Workstation

->

Base settings

.

Is the checkbox

Wheel

active the push button can be used. This setting is stored in the file

Settings.conf

.

User Manual ATS400 Version 3.14.41 - 1 82

Administration

2.2.12 Change voltage

There is the possibility to change the voltage of a high voltage test during the test using the push button. This possibility must be activated.

Open the dialog choosing

Settings

->

Workstation

->

Base settings

.

Is the checkbox

Change voltage

active you can use the push button to change the voltage during a test. This setting is stored in the file

Settings.conf

.

2.2.13 Dummy test plan

The dummy test plan is created by ETL Prüftechnik and will be delivered with the dummy. You can use only one dummy test plan. The dummy test plan is stored in the subfolder

DummyPlan

of the

ETL DataView 3

folder. There must only one file with the extention

.plan

in this folder.

The dummy test plan will be used as a normal test plan. A result file will be created according to the settings in

Settings

->

File Storage

.

As factory default a dummy test plan doesn't create a report file.

The execution of the dummy test plan is done manually or according to the settings in

Dummy test .

User Manual ATS400 Version 3.14.41 - 1 83

Administration

Caution

The following procedure is not covered by any rights within

ETL DataView 3

.

The work described here can be done by any user having the right to change test plans.

Important

The name of the result file and the report files are according to the same rules as with normal test plans.

Will in the file name not choose to use DateTime only the last dummy test will be recorded. Will the dummy test be executed after a certain amount of tests the result file and the report of the last test will be overwritten.

Will by creating the folder name a preset keyword or a keyword from the test type Data input be used invalid folder names may arise when they should be stored in subfolders.

Exit a running

ETL DataView 3

.

Copy the dummy test plan for Subfolder

DummyPlan

into the folder where your test plans are located.Be carefull not to overwrite an existing test plan.

User Manual ATS400 Version 3.14.41 - 1 84

Administration

Navigate in Windows Explorer to the folder

C:\Programm files\ETL\DataView 3

.

Open the file

Settings.conf

with the Windows Editor.

Open the file with a double click and select in the opening dialog

Select a program from a list of installed programs

and close it with the button

OK

.

In the next dialog choose

Notepad

and close the dialog with the button

OK

.

User Manual ATS400 Version 3.14.41 - 1 85

Administration

Windows Notepad will open. Find the line with the entry

Edit_Invalid_Enabled

.

User Manual ATS400 Version 3.14.41 - 1 86

Administration

Change the value from

false

into

true

.

Save the file and close the editor.

Start

ETL DataView 3

again.

Change the dummy test plan by adding the desired report options. Exit

ETL DataView 3

again.

Change in the subfolder

DummyPlan

the extension of the file from

.plan

into

.plan$

.

Move the dummy test plan from the folder with your test plans into the subfolder

DummyPlan

.

Change the value in

Edit_Invalid_Enabled

in the file

Settings.conf

from

true

to

false

.

2.2.14 Printserver configuration

For configuring the print server the

configuration file must be present. You can

configure that the files from the report creation will be stored in a different location as the result files, the behaviour how to deal with unknown keywords in the report und the timeout for waiting on PDF-reports or printing.

User Manual ATS400 Version 3.14.41 - 1 87

Administration

Important

In the case the configured folder is on an external storage media, e. g. a memory stick or network, this media must be present before the first report will be created. The tester will see a runtime message and the file will not be stored.

This settings must be made manually and are not supported by

ETL DataView 3

.

Exit

ETL DataView 3

. The configuration will only be read during startup.

Navigate in Windows Explorer to the folder

C:\Programm Files\ETL\DataView 3

.

Open the file

Printserver.conf

with Notepad.

Open the file with a double click and select in the opening dialog

Select a program from a list of installed programs

and close it with the button

OK

.

User Manual ATS400 Version 3.14.41 - 1 88

Administration

In the next dialog choose

Notepad

and close the dialog with the button

OK

.

For configuring an alternative folder you must change the entry for the report type.

The file contains for each report type an empty element. You must create an opening and closing element.

Add a folder name or change an existing folder name. In this example for the report type

PDF

to folder

N:\

.

To change the timeout for the report file you must change one of the existing numbers. Be aware that the time is in milliseconds.

With the element

Debug

the output can be changed if an unknown keyword is found.

The element

PdfTimeout

is a time within the PDF-file must be created. Is the file created before this time is elasped the next report can be created.

The element

PrintTimeout

is a wait time. The complete time will be waited before the next report can be created.

Save the file and close Notepad. Start

ETL DataView 3

again.

User Manual ATS400 Version 3.14.41 - 1 89

Test plan editing

3 Test plan editing

This part of the manual is aimed at the persons who create and edit the test plans.

It contains all information that is necessary to create the test plans in accordance with the requirements for the test and the existing test types.

All illustrations and options refer to the Windows variant that you receive with the

ATS400

variants X2, X6 and X8. The Windows CE variants with the

ATS400

variants

X4 and X5 do not contain all options.

3.1

Administering test plans

Open the dialog choosing

Test plan

->

Create / Edit

.

On the left side a list of the currently available test plans is displayed.

In the midth upper part the test steps in the selected test plan are displayed. Is no test plan selected the list is empty.

In the midth lower part the data for the test step is displayed. Is no test plan or no test step selected the list is empty.

User Manual ATS400 Version 3.14.41 - 1 90

Test plan editing

Button

Change plan

New Plan

Copy plan

Delete plan

Plan information

Test

Back

Action

The window for

Changing test plans will

be displayed and the selected test plan can be changed.

A new plan will be created. Details for this you will find in the quick start guide.

The selected test plan will be copied. You must enter a new name.

After a confirmation dialog the test plan will be deleted permanently if you confirm this.

The window for

Plan options will be

opened.

In this mode no changes can be made.

The window

Test plan

will be opened.

The window will be closed.

User Manual ATS400 Version 3.14.41 - 1 91

Test plan editing

3.2

Changing test plans

Open the dialog choosing

Test plan

->

Create / Edit

->

Change plan

.

On the left side a list with the currently present test steps is displayed. The test steps can be expanded or collapsed.

Button

Edit

Add

Delete

Copy

Action

This button is available in the case a test step is selected.

The window for configuring the test step will be displayed.

A new test step will be created. Details can be found in the quick start guide.

This button is available in the case a test step is selected.

After a confirmation dialog the test step will be deleted permanently.

This button is available in the case a test step is selected.

The test step will be copyed to the

User Manual ATS400 Version 3.14.41 - 1 92

Test plan editing

Button

3.3

Cut

Paste

Up down

Expand

Collapse

Test

Close

Plan Options

Action

internal clip board.

This button is available in the case a test step is selected.

After a confirmation dialog the test step will be copied to the internal clip board and then will be deleted permanently.

This button is available when a test step is present in the internal clip board.

This test step will be inserted below the seleted test step.

Jump

targets of existing test steps will be adapted automatically. The

Jump targets of the

inserted test step will be set to default values.

This button is available in the case a test step is selected.

The selected test step will be moved one position upwards.

Jump targets will be

adapted automatically.

This button is available in the case a test step is selected.

The selected test step will be moved one position downwards.

Jump

targets will be adapted automatically.

All test steps of the test plan will be expanded. The text changes to

Collapse

.

All test steps of the test plan will be collapsed. The text changes to

Expand

.

The window

Test

will be opened.

The window will be closed.

The window for

Plan options will be

opened.

In this mode changes can be done.

Test types

For each test type exists specific parameters and common parameters. The specific parameters are unique for each test type and the common parameters are the same for all test types. Not all common parmeters are present for every test type and may not be present when configuring a single test step.

Common parameters are:

Ramp

Start conditions

Matrix

User Manual ATS400 Version 3.14.41 - 1 93

Test plan editing

Jump

Text

Multitest

Create log

3.3.1

Common dialogs

3.3.1.1 Ramp

The ramp configuration describes the ramp parameters.

Factroy default is that no ramp is used.

Ramp

: Turns the ramp on and off. The other parameters can be changed when ramp is turned on.

Start

: Voltage with that the ramp will start.

Up

: Time in seconds from the start voltage up to the test voltage.

Test

: Test time in seconds. This is the time the test voltage is applied to the unit under test. This is the same value a parameter

t

in the configuration dialog.

Down

: Time in seconds from the test voltage to the end voltage.

User Manual ATS400 Version 3.14.41 - 1 94

Test plan editing

3.3.1.2 Start conditions

The start conditions describe under which cirumstances the test will start.

Each signal can have one of four states.

Active

: The signal must be set to provide a start.

Not Active

: The signal must not be set to provide a start.

Ignore

: The signal will be ignored.

Monitored

: The signal must be set. It will be monitored that the signal will be set in the case all other signals with state

Active

are alredy set. Will the signal not be set within the configured time the test will be evaluated as failed and will be aborted. Are multiple signals with state

Monitored

the first signal reaching the time will fail and abort the test.

To select the state

Monitored

at least one other signal must have state

Active

.

Is there only one signal with state

Active

and no other signal has state

Monitored

, the state of that signal cannot be changed.

It is under the responsibility of the test plan editor to choose a combination that makes sense for his scenario.

The signal

Start-Button

allows the start with the

Starttaste

at the front or the signal

Button Start

on the ETL Interface or the

Start

button on the screen.

User Manual ATS400 Version 3.14.41 - 1 95

Test plan editing

The signal

PE-Testprobe

allows the start with the button at the PE-Testprobe.

The signal

Contact Monitoring

allows the start with contacting, e. g. with test pistols.

The singal

Safety Circuit HV

allows the start with closing the safety circuit. For the test type

HV-AC

this value is always

Active

and cannot be changed. For the test types

HV-DC

and

ISO

it depends from the settings in the

ATS400

. If the value can be changed. Factory default is the value

Active

and cannot be changed.

Statechange

requires that at least one of the signals with state

Active

or

Monitored

must change its state from not set to set to provide a start. When manually contacting the unit under test it ensures that the new test point has been contated.

This checkbox is not visible on single test and the first test step in a test plan and is always set.

Delay

will delay the start of the test after the start conditions are meet.

Repeating

allows to execute test steps only once, each time or nth time to execute.

Is the value 0 the test step will only be executed once. Is the value 1 it will be executed each time. For all other values the test step will be executed the first time and then be omited for n-1 times. This parameter will not be displayed for single test.

Buttons

is a combobox. You have four selections. This combobox is not visible for single test and set to

None

.

None

dosn't show any buttons.

Skip

allows the tester to omit the current test step. It will be evaluated as passed for the overall result.

Start

allows to start the test step.

Start and skip

shows the skip and the start button.

User Manual ATS400 Version 3.14.41 - 1 96

Test plan editing

3.3.1.3 Matrix

With this property page the settings of a matrix will be configured.

Shown is the dialog in the case a

configuration file is present.

Parameter

Off

Matrix ETL-

Interface

ETL-IO: Out 7

(Pin9)

ETL-IO: Out 8

(Pin10)

Matrix CAN

Description

Both outputs of the ETL-Interface are off and all relais of a matrix are in default position.

The ETL-Ingterface will be used. This setting is only possible if in

Settings

->

I/O-Interface

the check box

Disable SC, Con

is active.

This check box determines the state of the output Out7 of the ETL-Interface.

This check box determines the state of the output Out8 of the ETL-Interface.

This checkbox can be selected in the case a cofiguration file for a matrix is existent. If the checkbox is active the matrix can be configured.

On the sub property page

Relaismatrix

are controls to configure the matrix.

User Manual ATS400 Version 3.14.41 - 1 97

Test plan editing

Parameter

Bank

Reset

HV1

HV2

Description

With the buttons it can be switched between the banks of the different outputs. If there is a non default setting in a bank a yellow field will be displayed near the button.

Sets als outputs to the default setting.

If a button in this line is set the wire HV1 is switched to the coresponding output.

In this line the number of the outputs of the active bank are displayed.

If a button in this line is set the wire HV2 is switched to the coresponding output.

In the example above the output 1 is connected to wire HV1 and output 2 is connected to wire HV2.

3.3.1.4 Jump

On this property page jumps are configured.

User Manual ATS400 Version 3.14.41 - 1 98

Test plan editing

On the property page

Jump

conditions can be configured how the test plan will select the next test step. For the test steps can provide thre different results you have one property page for each different result. The settings are the same for each result.

On property page

IO.

you define the target if a test has ended with passed.

On property page

NIO

you define the target if a test has ended with failed.

On property page

Cancel

you define the target ist a test has been aborted.

You can select

Next

and the test plan will proceed with the next step.

You can select

End

and the test plan will end.

You can select

Jump

and you can select a test step in the drop down box. The stet plan will proceed with this test step.

3.3.1.5 Text

On this property page user advices are configured.

On the propety page

Text

an text or picture can been shown before, during or on an erroneous test.

User Manual ATS400 Version 3.14.41 - 1 99

Test plan editing

The text can be entered by a click on the white edit box and activated by setting the checkbox

Show

. The hight of the font can be changed in the drop down box.

Will an advice been shown

Before

it will be visible until the start conditions are meet.

Will an advice been shown

During

a test it will be visible until the test step ends.

Will an advice been shown on

Error

it will be visible until the dialog will be confirmed.

On

Cont. On Error

you can configure when the advice on

Error

will be confirmed.

You can use the same conditions as for start conditions.

On variant

ATS400 X4

there isn't the possibility to use pictures. The corresponding controls aren't visible.

3.3.1.6 Multitest

On this property page the multi test parameters are configured.

To summarize multiple test points into one result, e. g. the number of test points is different for different units under test, you can use

Multitest

.

User Manual ATS400 Version 3.14.41 - 1 100

Test plan editing

Parameter

Enable

Result

Show

Size

Text

Image

Search

Description

Enables multi test behaviour.

Configures how the results of the test are summarized.

Worstcase

: The worst measurement will be used.

Mode

Count

Configures the end of a multi test.

# of Measurements

: You must have the numer of tests with result passed as given in

Count

.

Button Pass

: The multi test ends when the pass button is pressed. It dosn't matter if this is coming from the front of the

ATS400

, the button on the screen or from the ETL-

Interface.

Configures how much tests must be done when used in mode

# of Measurements

. This parameter is only be visible when in

Mode

the value

# of Measurements

is selected.

Pass button visible

Configures to show the button

Pass

on the screen. This parameter is only be visible when in

Mode

the value

Button

Pass

is selected.

Configures that a user advice will be shown between two tests of a multi test.

Configures the hight of the font of the text of the user advice.

Text of the user advice.

The text will be shown as user advice.

A picture will be shown as user advice.

Shows the file select dialog to select the picture. This button is available when

Image

is selected.

On variant

ATS400 X4

there isn't the possibility to use pictures. The corresponding controls aren't visible.

User Manual ATS400 Version 3.14.41 - 1 101

Test plan editing

3.3.1.7 Create log

On this property page the logging is configured.

To log measurements during a test logging can be activated.

Parameter

Enable

Time interval

Description

Activates the logging.

Configures how often measurements are stored.

For each test step a file will be created. The rule for the base name is according to the result file. The number of the test step is added to the base name.

The file is a test file in csv-format and can be imported to other applications. The exact format depends on the test type.

User Manual ATS400 Version 3.14.41 - 1 102

Test plan editing

3.3.2

Protective Earth test

The protective earth test will be configured using the following dialog.

I

U

Parameter t

R max

Description

Test time in seconds. The test time is the time the current is flowing through the unit under test.

Test current in A. This current is flowing through the unit under test for the test time.

Open circuit voltage in Volt. This voltage is applied to the unit under test or the contacts in the case no current is flowing through the unit under test.

Maximum allowed resistance for evaluation.

User Manual ATS400 Version 3.14.41 - 1 103

Test plan editing

3.3.3

High voltage AC

The high voltage AC test will be configured using the following dialog.

Parameter t

U

I min

I max f

Result

Description

Test time in seconds. The test time is the time the testing voltage is applied to the unit under test. The times for powering on and off and the ramp times are not part of the test time.

Testing voltage in kilovolt. This voltage is applied to the unit under test for the test time.

Minimum test current. This current must be flowing through the unit under test during the test time. If the current fallse below this limit the test is evaluated as not ok.

Maximum test current. The current must not exceed this limit during the test time. If the current exceeds this limit the test will be evaluated as not ok.

Frequency in Hertz. Frequency of the testing voltage.

You can select how the evaluation is done. Selecting

Test

User Manual ATS400 Version 3.14.41 - 1 104

Test plan editing

Parameter

Sparc detection

Description

Time

the test ends after the test time. Selecting

Start-

Button

the evaluation ends with pressing the start button again.

This parameter will only be visible when the

ATS400

is supporting this function.

You can select if spark detection is disabled resp. how sensitive it is. Spark detection evaluates the break down of the measured voltage regarding to size and gradient.

Selecting

Off

spark detection is disabled. Selecting

Coarse

a large spark will be detected, selecting

Normal

the sensitivity is medium and selecting

Fine

small sparks will be detected.

Additional informations about the behaviour of the high voltage test, the error messages and the evaluation you will find in the

Reference

.

3.3.4

High voltage DC

The high voltage DC test will be configured using the following dialog.

User Manual ATS400 Version 3.14.41 - 1 105

Test plan editing

Parameter t

U

I min

I max

Result

U Discharge

Sparc detection

Description

Test time in seconds. The test time is the time the testing voltage is applied to the unit under test. The times for powering on and off and the ramp times are not part of the test time.

Testing voltage in Volt. This voltage is applied to the unit under test for the test time.

Minimum test current. This current must be flowing through the unit under test during the test time. If the current fallse below this limit the test is evaluated as not ok.

Maximum test current. The current must not exceed this limit during the test time. If the current exceeds this limit the test will be evaluated as not ok.

You can select how the evaluation is done. Selecting

Test

Time

the test ends after the test time. Selecting

Start-

Button

the evaluation ends with pressing the start button again.

Discharge voltage is the voltage which must be fallen short of after the test before the test step ends.

This parameter will only be visible when the

ATS400

is supporting this function.

You can select if spark detection is disabled resp. how sensitive it is. Spark detection evaluates the break down of the measured voltage regarding to size and gradient.

Selecting

Off

spark detection is disabled. Selecting

Coarse

a large spark will be detected, selecting

Normal

the sensitivity is medium and selecting

Fine

small sparks will be detected.

Additional informations about the behaviour of the high voltage test, the error messages and the evaluation you will find in the

Reference

.

User Manual ATS400 Version 3.14.41 - 1 106

Test plan editing

3.3.5

Insulation test

The insulation test will be configured using the following dialog.

Parameter t

U

R min

Result

Description

Test time in seconds. The test time is the time the testing voltage is applied to the unit under test. The times for powering on and off and the ramp times are not part of the test time.

Testing voltage in Volt. This voltage is applied to the unit under test for the test time.

Minimung allowed resistance for the evaluation.

You can select how the evaluation is done. Selecting

Test

Time

the test ends after the test time. Selecting

Start-

Button

the evaluation ends with pressing the start button again.

User Manual ATS400 Version 3.14.41 - 1 107

Test plan editing

3.3.6

Function test

The function test will be configured using the following dialog.

f t

Parameter

U

I min

I max

Description

Test voltage in Volt. This voltage is applied to the unit under test. This parameter will be ignored if the source is external.

Frequency of the test voltage. This parameter will be ignored if the source is a direct current source or external.

Test time in seconds. The test time is the time the evaluation is done against the limits. This time starts after the conditions for evaluation start is meet. This parameter will be ignored when

Button Pass/Fail

is selected.

Minimum current. Is the current below this limit the test is evaluated as failed. This parameter will be shown when

Current

is selected as measuring channel. This parameter will be ignored when

Button Pass/Fail

is selected.

Maximum current. Is the current above this limit the test is

User Manual ATS400 Version 3.14.41 - 1 108

Test plan editing

Parameter

U min

U max

Result

Graphic

Source

Voltage

Start

Description

evaluated as failed. This parameter will be shown when

Current

is selected as measuring channel. This parameter will be ignored when

Button Pass/Fail

is selected.

Minimum voltage. Is the voltage below this limit the test is evaluated as failed. This parameter will be shown when

Current

is not selected as measuring channel. This parameter will be ignored when

Button Pass/Fail

is selected.

Maximum voltage. Is the voltage above this limit the test is evaluated as failed. This parameter will be shown when

Current

is not selected as measuring channel. This parameter will be ignored when

Button Pass/Fail

is selected.

Here are two entries possible. In the the upper drop down box the method can been choosen. You can select between

Measuring canal

und

Button Pass/Fail

. In the lower drop down box the channel for evaluation is selected when

Measuring canal

is selected. Is

Button Pass/Fail selected

as method no evaluation is done and the parameters

I min

,

I max

,

U min

,

U max

and the paremeters for

Start

will be ignored.

You can select if a graphic will be diplayed and if it will be stored. Selecting

None

no graphic will be displayed and stored. Selecting

Show only

the graphic will only be displayed.

Selecting Show and Save

the graphic will be shown and saved.

Source for suppling the unit under test. The possible sources are dependent from your device. Is an external source selected you can configure if the source is a direct current source with the checkbox

Source is DC

.

Drop down box how the supplyment of the unit under test is to handle. Selecting

Turn off after test

the voltage will be turn off when test has finished. Selecting

Leave on after test

the voltage will remain on when the test has finished. Selecting

Turn off

dosn't do a test. Only the voltage will be turned off. Selecting

Turn off if test N.I.O.

the voltage will be turned off when the test is evaluated as failed. Selecting

Turn on

dosn't do a test. Only the voltage will be turned on.

Setting when the evaluation against the limits will start.

This parameter and the dependent parameters will be ignored when

Result

ist set to

Button Pass/Fail

. With the setting

After Delay

the evaluation begins after the time in

Delay

has passed. With the seting

> min

the evaluation begins when the measurement is above the limit in

I min

rsp.

U min

. With the setting

Delay and > min

the

User Manual ATS400 Version 3.14.41 - 1 109

Test plan editing

Parameter Description

time in

Delay

must have passed and the measurement is above the limit in

I min

rsp.

U min

. With the setting

Gradient

the gradinet of the measurements must have fallen below the value in

Gradient

. With the setting

< max

the evaluation begins when the mesurement has fallen below the limit in

I max

rsp.

U max

. With the setting

Delay und < max

the time in

Delay

must ahve passed and the measurement is fallen below the limit in

I max

rsp.

U max

. With the setting

> min

,

Gradient

and

< max

the evaluation must start before expiring the time in

Timeout

.

Otherwise the test is evaluated as failed.

3.3.7

Leakage current test

The leakage current test will be configured using the following dialog.

These parameters are relevant for checking.

Parameter

Delay

Description

This parameter is not used.

User Manual ATS400 Version 3.14.41 - 1 110

Test plan editing

Parameter

t

I max

I min

Result

Description

Time used for checking against the limits. If you are using a polarity with automatic change the time is valid for each polarity.

Maximum allow leakage current.

Minimum leakage current which must be reached.

Type of the channel. There are the following selesctions:

I AC rms

: Effective value of the AC-fraction

I DC

: DC-fraction of the leakage current

I min

: Lowest value of the current leakage current

I max

: Highest value of the current leakage current

I rms

: Effective value of the DC- and AC-fraction

These parameters are relevant for suppling the unit under test.

Parameter

U f triple phase sisngle phase

Source

Voltage

Description

Voltage the device under test will be supplied.

Frequency of the supply voltage for the device under test.

The unit under test is supplyed with triple phase. This radio button is active only on a test rig configured for such units under test.

Teh unit under test is supplyed with a single phase.

There are several sources available. The sources are explained in detail in chapter

Supply options .

Source for the supply of the device under test.

Mains voltage

: The supply is taken from the mains voltage of the ATS400.

AC 0..270V

: The supply is taken from the power converter of the ATS400.

External

: The supply is taken from an external source.

Defines how after a test the supply for the device under test is handeled.

Turn off after Test

: After the test the supply will be turned off.

This setting is fixed and cannot be changed.

These parameters are relevant for the test setup.

Parameter

Testing Modul

Description

There are several testing moduls available. the testing modules are explained in detail in chapter

Testing modules

.

IEC 60990 Pic 3

: Unweighted touch current

IEC 60990 Pic 4

: Touch current weighted for reaction

IEC 60990 Pic 5

: Touch current weighted for to let loose

User Manual ATS400 Version 3.14.41 - 1 111

Test plan editing

Parameter

Testing method

Polarity

Description

IEC 60601 Basic

: Unweighted leakage current

IEC 60601 Pic 12

: Weighted leakage current

Selects which testing method will be used.

Protective conducter current

: The current in the protective conducter will be meassured. This method is named earth leakage current in EN 60601-1:2013-12.

Touch current

: The current over the housing will be measured. For this measurement the PE-probe must be used. This method is named contact current in EN 60601-

1:2013-12.

Determines the connection of the unit unde rtest and the test setup. The contacting depends form the testing method. The contacting for the protective conducter current are explained in chapter

Polarity for protective conducter , for the touch current in chapter

Polarity for touch current .

In some selections the word „Auto“ is used. With this setting the change of the polarity is done automatically within one test step.

Using testing method

Protective conductor current

following connections are available:

Auto

n/S1

Closed

L1->PE

p/S5

Normal/Switched n/S1

Closed

L2-> PE

p/S5

Normal n/S1

Closed

Auto with SFC

p/S5

Switched n/S1

Open p/S5

Normal/Switched

User Manual ATS400 Version 3.14.41 - 1 112

Test plan editing

Parameter Description

L1->PE with SFC

n/S1

Open p/S5

Normal

L2-> PE with SFC

n/S1

Open n/S1

Closed

L1->PE

p/S5

Switched

Using testing method

touch current

following connections are available:

Auto

p/S5

Normal/Switched e/S7

Closed n/S1

Closed

L2-> PE

p/S5

Normal e/S7

Closed e/S7

Closed n/S1

Closed

Auto with SFC

p/S5

Switched n/S1

Open p/S5

Normal/Switched

L1->PE with SFC

n/S1

Open p/S5

Normal e/S7

Closed e/S7

Closed

User Manual ATS400 Version 3.14.41 - 1 113

Test plan editing

Parameter Description

L2-> PE with SFC

n/S1

Open p/S5

Switched

SFC PE Open Auto

n/S1

Closed

SFC PE Open

p/S5

Normal/Switched n/S1

Closed p/S5

Normal

SFC PE Open reverse

n/S1

Closed p/S5

Switched e/S7

Closed e/S7

Open e/S7

Open e/S7

Open

User Manual ATS400 Version 3.14.41 - 1 114

Test plan editing

3.3.7.1 Testing modules

IEC60990 Pic 3

Measuring circuit for unweighted touch current (corresponds EN61010-1 Pic A.3)

This measuring model emulates the impedance of the human body. This allows sto measure the current flowing through a human body touching the unit under test in a similar way.

IEC60990 Pic 4

Measuring circuit for touch current weighted for reaction (corresponds EN61010 Pic

A1)

Reaction in the human body results from the current flowing within the body. The measuring circuit according to picture 4 emulates the impedance of the body and

User Manual ATS400 Version 3.14.41 - 1 115

Test plan editing

weight the reaction depended from the current causing an unwanted reaction.

IEC60990 Pic 5

Measuring circuit for touch current weighted for let loose

The inability to let loose the unit under test results from a current flow within the human body (e.g. through muscles). the influence of the frequency for let losse is different from the influence for reaction. this is significant for frequencies above 1 kHz.

IEC 60601 Basic

Measuring circuit

This measuring circuit corresponds to the requirement of DIN EN 60601-1:2013-12 chapter 8.7.3 part e.

User Manual ATS400 Version 3.14.41 - 1 116

Test plan editing

IEC 60601 Pic 12

Measuring circuit with weighting

This measuring circuit corresponds to the requirement of DIN EN 60601-1:2013-12 chapter 8.7.3 part a - d.

3.3.7.2 Supply options

Mains voltage

The unit under test is supplied from the mains voltage of the ATS400.

There is no isolation from the mains voltage.

User Manual ATS400 Version 3.14.41 - 1 117

Test plan editing

AC 0..270V

The unit under test is supplied by the power converter in the ATS400.

The isolation is within the ATS400.

Extern

The unit under test is supplied over an external input.

An insulation can be done by an external transfomer.

User Manual ATS400 Version 3.14.41 - 1 118

Test plan editing

3.3.7.3 Polarity for protective conducter

The following polarities are available with the measuring method

protective conducter current

.

L1->PE

resp. 1. phase

Auto

L2->PE

resp. 2. phase

Auto

User Manual ATS400 Version 3.14.41 - 1 119

Test plan editing

L1->PE with SFC

resp. 1. phase

Auto with SFC

L2->PE with SFC

reap. 2. phase

Auto with SFC

User Manual ATS400 Version 3.14.41 - 1 120

Test plan editing

3.3.7.4 Polarity for touch current

The following polarities are available with the measuring method

touch current

.

L1->PE

resp. 1. phase

Auto

L2->PE

resp. 2. phase

Auto

User Manual ATS400 Version 3.14.41 - 1 121

Test plan editing

L1->PE with Auto

resp. 1. phase

Auto with Auto

L2->PE with Auto

resp. 2. phase

Auto with Auto

User Manual ATS400 Version 3.14.41 - 1 122

Test plan editing

SFC PE Open

resp. 1. phase

SFC PE Open Auto

SFC PE Open reverse

resp. 2. Phase

SFC PE Open Auto

User Manual ATS400 Version 3.14.41 - 1 123

Test plan editing

3.3.8

Batch

The batch data will be configured using the following dialog.

With the test step

Batch

you can enter data for a charge.

The entered data will be stored in the result file and report file and can be printed locally. This test type can be configured individually. In the top most entry field a preset user can be entered or by activating the checkbox

from Username

the currently logged in user will be used. Also the entry field

Test equipment

can be preset or the name of the worksation can be used. In the both middle fields a type and ordern number can be preset. To generate a sequential serial number you can preset the first serial number. The serial number can be automatical increased on passed and/or failed tests. In this case the serial numer must not contain any alphabetic characters. The start conditions can be set as usual.

Using the additional checkboxes the fields can provided with

Patterns

. This allows the input to be checked for plausibility.

User Manual ATS400 Version 3.14.41 - 1 124

Test plan editing

3.3.9

Data input

The data input will be configured using the following dialog.

You can configure up to 5 fields to enter any data. Is the corresponding checkbox

Pattern

active the preset will be used as a

Pattern . This allows the input to be

checked for plausibility.

User Manual ATS400 Version 3.14.41 - 1 125

Test plan editing

3.3.10 Resistance

The resistance test will be configured using the following dialog.

Depending on the configuration and firmware of the

ATS 400

not all parameters will be displayed.

Parameter t

R min

R max

R offset

Timeout

Measuring range

Description

Test time, the evaluation will be done when it is elapsed.

Lower limit for the evaluation.

Upper limit for the evaluation.

Offset to compensate the resistance of the measurement cables. This value will be subtracted from the measurement value before evaluation.

Time within that valid measurements must be present.

Selection for the measurement range.

0: Automatic

User Manual ATS400 Version 3.14.41 - 1 126

Test plan editing

Parameter Description

Other values depends from the configuration of the

ATS 400

.

3.3.11 Matrix

The matrix test step will be configured using the following dialog.

This step can only be chossen in the case a

configuration file for a matrix is present.

Parameter

Off

Description

Both outputs of the ETL-Interface are off and all relais of a matrix are in default position.

This checkbox is always disabled.

Matrix ETL-

Interface

ETL-IO: Out 7

(Pin9)

ETL-IO: Out 8

(Pin10)

This checkbox is always disabled.

This checkbox is always disabled.

User Manual ATS400 Version 3.14.41 - 1 127

Test plan editing

Parameter

Matrix CAN t

Description

This checkbox can be selected in the case a cofiguration file for a matrix is existent. If the checkbox is active the matrix can be configured.

This time determines how long the step will be executed.

This can be used to make external measurements.

On the sub property page

Relaismatrix

are controls to configure the matrix.

Parameter

Bank

Reset

HV1

HV2

Description

With the buttons it can be switched between the banks of the different outputs. If there is a non default setting in a bank a yellow field will be displayed near the button.

Sets als outputs to the default setting.

If a button in this line is set the wire HV1 is switched to the coresponding output.

In this line the number of the outputs of the active bank are displayed.

If a button in this line is set the wire HV2 is switched to the coresponding output.

In the example above the output 1 is connected to wire HV1 and output 2 is connected to wire HV2.

User Manual ATS400 Version 3.14.41 - 1 128

Test plan editing

3.3.12 Continuity test

The continuity test will be configured using the following dialog.

Parameter

Interpretation

Limit

Test time

Test voltage

Max. test current

Description

Closed

: The test will be evaluated as passed if a resistance lower than

Limit

will be measured.

Open

: The test will be evaluated as passed if a resistance higher than

Limit

will be measured.

Limit for the evaluation.

Time after that the measurement will be evaluated.

Maximum test voltage used for the measurement.

Maximum current during the test.

User Manual ATS400 Version 3.14.41 - 1 129

Test plan editing

3.4

Test plan settings

This section discusses the settings that are valid for the entire test plan.

The associated window can be opened using the

Plan Options

button in the

Create / Edit

window.

The window with the

Planoptions

opens.

User Manual ATS400 Version 3.14.41 - 1 130

Test plan editing

3.4.1

Text

You can enter a free text here.

User Manual ATS400 Version 3.14.41 - 1 131

Test plan editing

3.4.2

Identification

Here you can configure identifications which are used for automatic plan selection.

With the button

New

you can enter an new idetification.

With the button

Delete

the selected identification will be deleted.

The identifications will be used for automatic plan selection.

User Manual ATS400 Version 3.14.41 - 1 132

Test plan editing

3.4.3

Report options

The report options provide the possibility to convert the results data into another format. For part of the options,

templates need to be created.

When calling up for the first time, no report options are yet active.

User Manual ATS400 Version 3.14.41 - 1 133

Test plan editing

Click on the

Change

button. The dialogue changes, which means that the report selected can be edited.

Activate one or more of the

IO

,

NIO

or

Error

checkboxes so that the report is created for the corresponding overall result.

User Manual ATS400 Version 3.14.41 - 1 134

Test plan editing

Select the desired report type from the dropdown box.

For the

Printer

,

Save as HTML

and

Save as PDF

options, you need to select an

HTML template.

With the

Save as XML

option, you can select a style sheet.

With the

Save as CSV

option you need to select a CSV template.

With the

Print on Zebra Lableprinter

option, you need to select a ZPL file.

With the

Printer

and

Print on Zebra Lableprinter

options, you can additionally select a printer. If you do not select a printer, the current standard printer is used to print.

User Manual ATS400 Version 3.14.41 - 1 135

Test plan editing

When clicking on the Template or Printer fields, the corresponding standard dialogue of

Windows opens.

With the

Ok

button, you can assume the data from the report selected. With the

Cancel

button, the changes are dismissed.

With the

Save as HTML

,

Save as PDF

,

Save as XML

und

Save as CSV

options a new file is created. The place of saving and the file name are guided by the settings in

Settings

->

File storage

->

Result

.

User Manual ATS400 Version 3.14.41 - 1 136

Test plan editing

3.4.4

Info

Here it will displayed when and on which workstation and which user the test plan was created or changed. If there was no

workstation name configured rsp. the

user administration was not active the according fields are empty.

User Manual ATS400 Version 3.14.41 - 1 137

Test plan editing

3.4.5

Options

This option will not be used any more. This option has expanded and is now part of the

Start conditions .

User Manual ATS400 Version 3.14.41 - 1 138

Inspector

4 Inspector

This part of the manual is aimed for those persons using the program for testing.

This part described the general procedures for using. Regarding to the different situations this part cannot describe the concrete situation at a test station.

4.1

Test individual

Open the dialog choosing

Test individual

.

There is a button for each test type. The buttons for the available test types on your

ATS400

are operational.

The window for each test type is the same. In the following this is described with the protective earth test as an example.

User Manual ATS400 Version 3.14.41 - 1 139

Inspector

4.1.1

Protective earth test

You will do a protective earth test in this window.

Before the first test no values will be displayed. During and after the test the values will be displayed.

A white background means that no test has been done.

A yellow background means that a test is currently executing.

A green background means that the last test was evaluated as passed.

A red background means that the last test was evaluated as failed.

Button

Menu

Edit

Back

Action

This button is operational when no test is executed.

The main menu will be displayed.

This button is operational when no test is executed.

The

configuration dialog will be displayed.

The window will be closed. During a test the button changes to

Cancel

.

User Manual ATS400 Version 3.14.41 - 1 140

Inspector

Button

Cancel

4.2

Test plan

Open the dialog choosing

Test plan

.

Action

The currently running test will be aborted.

The test will be evaluated as failed. The button changes to

Back

.

Button

Menu

Select manual

Select ID

Test batch

User Manual ATS400

Action

The main menu will be displayed.

The window for

manual test plan selection

will be opened.

The window for

automatic test plan selection will be opened. Using this

selection the test plan will be closed after the test of one unit under test.

The window for

automatic test plan selection will be opened. Using this

Version 3.14.41 - 1 141

Inspector

Button

Create / Edit

Dummytest

Main menu

Action

selection the test plan will be closed by the tester.

This button is not operational when the

workstation is setup to use

ETL-

Interface

or

File Plan.ID

for automatic test plan selection.

The window for

test plan administering

will be opened.

This button is operational when a

dummy test plan is present. A dummy test will be

done.

The main menu will be displayed.

4.2.1

Manual test plan selection

Open the dialog choosing

Test plan

->

Select manual

.

On the left side the list of the existing test plans are displayed.

In the midth upper part the test steps in the selected test plan are displayed. Is no

User Manual ATS400 Version 3.14.41 - 1 142

Inspector

test plan selected the list is empty.

In the midth lower part the data for the test step is displayed. Is no test plan or no test step selected the list is empty.

Button

Menu

Test

Plan information

Back

Action

The main menu will be displayed.

The window

Test plan

will be opened.

The window for

Plan options will be

opened.

In this mode you cannot do any changes.

The window will be closed.

4.2.2

Automatic test plan selection

Open the dialog choosing

Test plan

->

Select ID

or

Test plan

->

Test batch

.

The number of the displayed fields depends on how the automatic test plan selection is

configured . In this example the maximum number of fields is displayed and is

acccording to the example

Article number and serial number with additionally activated

User Manual ATS400 Version 3.14.41 - 1 143

Inspector

checkbox for the article description.

Button

Menu

Back

Action

The main menu will be displayed.

The window will be closed.

4.3

Safety cabinet

The safety cabinets are built this way that they are locked when they are not supplied by the

ATS400

. After powering on the

ATS400

they are unlocked and can be opened.

Depending how the

ATS400

is setup in the

administration

ETL DataView 3

behaves different together with the safety cabinet.

The setting

Locked during plan

prevents an influence during executing a test plan.

The safety cabinet can be opened after the test plan has completed.

The setting

Unlock on pass

also contains a process control. In the case a test will be evaluated as failed the safety cabinet will be unlocked after a second action for unlocking. Dependend from the configuration of the saftey cabinet this can be done with the button

Cancel

and always with the button

Unlock

in

ETL DataView 3

.

User Manual ATS400 Version 3.14.41 - 1 144

Inspector

When the safety cabinet is unlocked the message

Open safety cabinet

will be displayed until the safety cabinet has been opened.

User Manual ATS400 Version 3.14.41 - 1 145

Report creation

5 Report creation

This part of the manual is aimed at the persons who create and edit the report templates.

It contains information that is necessary to create report templates that are required for the individual report options.

5.1

Creating templates

For the use of the

Printer

,

Save as HTML

and

Save as PDF

report options, you need to create an

HTML template

.

When using the

Save as XML

report option, you can create a style sheet.

To use the

Save As CSV

report option, you need to create a

CSV template

.

To use the

Print on Zebra Lableprinter

option, you need to create a ZPL file as a template.

All templates have in common that the data can be accessed via

key words .

5.1.1

Key words

You can access the data of the results files via key words. The key words always consist of prefix

TAG_

and the postfix

_TAG

.They are not displayed in the following tables. Upper and lower case is considered for the key words. Assume the key words in the way as they are written in the following tables.

Key words that apply for the entire test plan have the form

TAG_<key word>_TAG

.

Key words that apply for a test view have the form

TAG_<step>_<key word>_TAG

. In doing so, it is defined in

<step>

what test step is accessed. If

<step>

has the value

##

, every test step of the test plan is accessed. In this case, the key word must exist in a cell of a table. If

<step>

has the value

#n

, while

n

is a number counted from

1, precisely this step is accessed. In this case, the key word can also be used outside a table.

Key words that refer to the multiple test data of a test step have the form

TAG_<step>_Multitest_<Multitest>_<key word>_TAG

. In doing so, it is defined in

<Multitest>

what multiple test is accessed. If

<Multitest>

has the value

##

, every multiple test is accessed. In this case, the key word must exist in a cell of a table. If

<Multitest>

has the value

#n

, while

n

is a number counted from 1, precisely this multiple test is accessed.

5.1.1.1 Key words of the results data

This section describes all key words that apply for the total test plan and are addressed with the form

TAG_<key word>_TAG

.

Key words that are part of every results file:

Key word

DataView_Version

Explanation

Version of

ETL DataView 3

with which this file

User Manual ATS400 Version 3.14.41 - 1 146

Report creation

Key word

PlanName

PlanPath

PlanDescription

Explanation

was created.

Name of the test plan, is filled with the file name when being created.

Relative path to the application where the test plan was saved.

Description of the test plan. This value can be entered by the user in

Settings

->

Text

.

PlanCreatedByUser

User who created the test plan first. If user administration is not active, the field is empty.

PlanCreatedByWorkstation

Test station where the test plan was created.

The name of the test station can be stated in

Settings

->

Workstation

->

Base settings

.

PlanCreatedOnDate

PlanEditedByUser

PlanEditedByWorkstation

Date of creating the test plan. The date is in the format in accordance with the country settings valid at this moment in time.

User who modified the test plan last. If user administration is not active, the field is empty.

Test station where the test plan was modified last. The name of the test station can be stated in

Settings

->

Workstation

->

Base settings

.

PlanEditedOnDate

PlanIdentificationEnabled

The

<ResultData>\<Identification>

block exists.

PlanBatchEnabled

Date of modifying the test plan. The date is in the format in accordance with the country settings valid at this moment in time.

PlanTestSteps

Intended for future purposes.

Number of test steps in the test plan.

PlanSelectMode

PlanSelectPatternID

PlanSelectPatternTyp

PlanSelectPatternSerie

Selection method for plan selection via identification

Template to identify the PlanID

Template to identify the type if it is entered with the ID.

Template to identify the serial number if it is entered with the ID.

PlanSelectPatternMask

DATE

Bit template:

1: Combined field (2+4)

2: Entry type

4: Entry series

8: Entry articelnumber

Date of the test in the local date format.

User Manual ATS400 Version 3.14.41 - 1 147

Report creation

Key word

TIME

TESTER

PLAN_NAME

WORKSTATION

USEDID

RESULT

FileSavedByUser

FileSavedByWorkstation

FileSavedOnDate

Count_Tests

Count_IO

Count_NIO

Count_ERROR

PlanCycleTime

PlanStandbyTime

IDs

ID_n

Explanation

Time of the test in the local time format.

User logged in.

Name of the test plan used.

Name of the test station.

ID used for the automatic test plan selection.

Total result of the plan.

Passed

-> The test item has passed the test.

Failed

-> the test item has not passed the test or the test was cancelled.

User who saved the test plan or the result file.

Test station where the test plan or the result file was saved. The name of the test station can be stated in

Settings

->

Workstation

->

Base settings

.

Date of saving the test plan or the result file.

The date is in the format in accordance with the country settings valid at this moment in time.

Number of tests that were carried out since starting the test plan.

Number of IO results of the test since starting the test plan.

Number of NIO results of the test since starting the test plan.

Number of errors of the test since starting the test plan.

Duration for carrying out the test. The value is always 0.

Duration between the end of the prior test and the start of the current test.

Number of IDs.

Identification for the test plan. The tags are numbered from 0 to the number of IDs - 1, i.e.

have the values ID_0, ID_1, etc.

Key words that are part of the

Batch run

test step:

Key word

Benutzer

Pruefplatz

Explanation

User field, can be adopted from the login.

Test station field, can be adopted from

User Manual ATS400 Version 3.14.41 - 1 148

Report creation

Key word

Typ

Auftragsnummer

Seriennummer

Explanation

the workstation settings.

Type, can already be entered during the automatic test plan selection.

Order number

Serial number, can already be entered during the automatic test plan selection. If this value is numerical, it can be automatically increased for IO or

NIO.

Key words that are entered via the automatic test plan selection:

Key word

TYPE

USEDID

SERIALNUMBER

ARTICLE

Explanation

Type of the test item, is only filled in if the combibox is active.

ID used to select the test plan, is only filled in when the

ID-Pattern

field is active.

Serial number of the test item, is only filled in if the

Serial-Pattern

field is active.

Article name, is only filled in if the

Article-Pattern

field is active.

5.1.1.2 Key words of the test step parameters

The following key words are common for all test steps. These key words are addressed with the form

TAG_<step>_<key word>_TAG

.

Key words for the parameters:

Key word

UNIT

UNIT2

MIN

MAX

StartCondition

Explanation

Unit of the measurement value

Unit of the test size

Minimum value of the measurement value.

Maximum value of the measurement value.

Start conditions of the test step. The start conditions are saved in a bitcoded manner and can only be analysed together with the

StartConditionMask

field.

User Manual ATS400 Version 3.14.41 - 1 149

Report creation

Key word

StartConditionMask t_delay

StateChangeBits

StartStateChange

StartStateJump

Explanation

Start conditions of the test step. The start conditions are saved in a bitcoded manner and can only be analysed together with the

StartCondition

field.

Start delay between fulfilling the start condition and the start of the test.

Always 0, never used.

Indicates whether the switching condition needs to be maintained when testing the start conditions.

0 = no waiting for switch condition

1 = Wait for switching condition

Indicates whether the

Jump

button is displayed.

Repeating

StartButton

PassButton

StartandJump

Indicates how often the test step is carried out.

Indicates whether the

Start

button is displayed.

Indicates whether the

Pass

button is displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

The

Batch run

,

Data Input

,

Dummy load

,

External program

,

FCT-Current

,

High

Voltage AC

,

High voltage DC

,

Insulation

,

Leakage current

,

Matrix

,

Light

Filament current

,

Protective earth

,

PT100

,

Resistance a

nd

User-Interface

test types use the following parameters in addition:

Key word

Polung

PolungCAN_00

bis

Multitest_Enable

PolungCAN_15

Explanation

Contacting setting of an external relay matrix.

Setting of the relay matrix in accordance with the choice in the program. The display is bit coded.

Multitest is active, if the value does not equal 0.

Multitest_Auswertung

Multitest_Endekriterium

Method of analysing the multitest.

0 = worst measurement value.

Method of ending the multitest.

0 = number of measurements

User Manual ATS400 Version 3.14.41 - 1 150

Report creation

Key word

Multitest_Endekriterium_Anzahl

Explanation

1 = pass key

Number of multitest inspections.

The

Lightcontrol

,

Data Input

,

External program

,

FCT-Current

,

High Voltage

AC

,

High voltage DC

,

Insulation

,

Leakage current

,

Matrix

,

Light Filament current

,

Protective earth

,

PT100

,

Resistance a

nd

User-Interface

test types use the following parameters in addition:

Key word

ErrorStartCondition

ErrorStartConditionMask

ErrorStateChangeBits

ErrorWSBDelay

ErrorWSBTimeValid

Explanation

Condition for switching forward in the event of an error.

Mask for the condition for switching further in the event of an error.

Always 0, never used.

Delay of switching further in the event of an error.

Delay time active.

5.1.1.2.1 Key words of the Protective earth test type

The

Protective earth

test type additionally uses the following key words:

Key word

U

I

R_max t f

EN60204

DC_Enable

Explanation

Idle voltage

Test current

Upper threshold value of the resistance.

Test time

Frequency of the voltage

Indication whether the test is performed in accordance with EN 60204.

Always 0.

Indicates whether the test is carried out with direct current voltage.

0: Alternating current

1: Direct current

User Manual ATS400 Version 3.14.41 - 1 151

Report creation

5.1.1.2.2 Key words of the High Voltage AC test type

The

High voltage AC

test type additionally uses the following key words:

Key word

U

I_min

I_max t

Ramp

U_start t_up t_down f

Quelle f_Source

U_Source

Management

Auswertung

CreateLog

Timeinterval

Explanation

Test voltage

Lower threshold

Upper threshold

Test duration

Indicates whether the ramp is active.

Start voltage

Increase time of the ramp. This parameter is only valid if the ramp is active.

Drop time of the ramp. This parameter is only valid if the ramp is active.

Frequency of the voltage

Selected current source.

This parameter is only used in the event of a hot high voltage.

Frequency of the source.

This parameter is only used in the event of a hot high voltage.

Voltage of the source.

This parameter is only used in the event of a hot high voltage.

Indicates how the supply of the test item is to be carried out.

0: Power off after test

1: Keep power on after test

2: Only power off

3: Power off only on error

4: Only power on

Type of the analysis of the test.

0 = end the test time

1 = end with start signal

Indicates whether a log file will be created.

0 = No log file will be created

1 = A log file be created

Time interval between two samples which will be written to the log file. The value is in seconds.

This value will only be used in case the value in

CreateLog

is 1.

User Manual ATS400 Version 3.14.41 - 1 152

Report creation

5.1.1.2.3 Key words of the High voltage DC test type

The

High voltage DC

test type additionally uses the following key words:

Key word

U

I_min

I_max t

Ramp

U_start t_up t_down

U_discharge

Quelle f_Source

U_Source

Management

Auswertung

CreateLog

Timeinterval

Explanation

Test voltage

Lower threshold

Upper threshold

Test duration

Indicates whether the ramp is active.

Start voltage

Increase time of the ramp. This parameter is only valid if the ramp is active.

Drop time of the ramp. This parameter is only valid if the ramp is active.

Discharge voltage

Selected current source.

This parameter is only used in the event of a hot high voltage.

Frequency of the source.

This parameter is only used in the event of a hot high voltage.

Voltage of the source.

This parameter is only used in the event of a hot high voltage.

Indicates how the supply of the test item is to be carried out.

0: Power off after test

1: Keep power on after test

2: Only power off

3: Power off only on error

4: Only power on

Type of the analysis of the test.

0 = end the test time

1 = end with start signal

Indicates whether a log file will be created.

0 = No log file will be created

1 = A log file be created

Time interval between two samples which will be written to the log file. The

User Manual ATS400 Version 3.14.41 - 1 153

Report creation

Key word

CheckCurrentInRamp

Explanation

value is in seconds.

This value will only be used in case the value in

CreateLog

is 1.

During a HVDC7 test the current will be checked against the limit Imax also during executing a ramp.

5.1.1.2.4 Key words of the Insulation test type

The

Insulation

test type additionally uses the following key words:

Key word

U

R_min t

Ramp

U_start t_up t_down

U_discharge

Quelle f_Source

U_Source

Management

Auswertung

Explanation

Test voltage

Lower threshold of the insulation resistance.

Test duration

Indicates whether the ramp is active.

Start voltage

Increase time of the ramp. This parameter is only valid if the ramp is active.

Drop time of the ramp. This parameter is only valid if the ramp is active.

Discharge voltage

Selected current source.

This parameter is only used in the event of a hot high voltage.

Frequency of the source.

This parameter is only used in the event of a hot high voltage.

Voltage of the source.

This parameter is only used in the event of a hot high voltage.

Indicates how the supply of the test item is to be carried out.

0: Power off after test

1: Keep power on after test

2: Only power off

3: Power off only on error

4: Only power on

Type of the analysis of the test.

0 = end the test time

1 = end with start signal

User Manual ATS400 Version 3.14.41 - 1 154

Report creation

Key word

CreateLog

Timeinterval

Explanation

Indicates whether a log file will be created.

0 = No log file will be created

1 = A log file be created

Time interval between two samples which will be written to the log file. The value is in seconds.

This value will only be used in case the value in

CreateLog

is 1.

5.1.1.2.5 Key words of the FCT-Current test type

The

FCT-Current

test type additionally uses the following key words:

Key word

U_Source

U

I_min

I_max t f t_start t_timeout

Gradient

Management

Explanation

Source of the test supply.

Voltage of the test item supply

Lower threshold value during the analysis. Not valid if the analysis is set to

Pass/Fail

. The unit and thresholds depend on the channel.

Upper threshold value during the analysis. Not valid if the analysis is set to

Pass/Fail

. The unit and thresholds depend on the channel.

Test time

Frequency of the voltage

Delay time for the start scenario when starting the test until starting the analysis. This value is not valid for all start scenarios.

Timeout for the start of the analysis.

The value is not used for all start scenarios.

Gradient for the start of the analysis.

This value is not used for all start scenarios.

Indicates how the supply of the test item is to be carried out.

0: Power off after test

1: Keep power on after test

2: Only power off

3: Power off only on error

4: Only power on

User Manual ATS400 Version 3.14.41 - 1 155

Report creation

Key word

Scenario

Auswertung

Kanal

SourceIsDC

Graphic

CreateLog

Timeinterval

UseAnalogConversion

ConversionOffset

ConversionGradient

PhysicalUnit

Explanation

Type of the start of the analysis

Type of the analysis

0 = measurement

1 =

Pass/Fail

button

Measurement channel

0 = current

1 = analogue channel1

2 = analogue channel2

3 = analogue channel3

4 = analogue channel4

5 = voltage measurement

External source provides direct current voltage

Type of the graphics settings

0 = no graphics

1 = only display graphic

2 = display and save graphic

Indicates whether a log file will be created.

0 = No log file will be created

1 = A log file be created

Time interval between two samples which will be written to the log file. The value is in seconds.

This value will only be used in case the value in

CreateLog

is 1.

This check box inidcates whether the analog input will be converted to customer specific units.

0 = No conversion

1 = Do conversion

Value of the physical value when the voltage has valaue 0.

This field is only valid when

UseAnalogConversion

has the value 1.

Slope for the conversion of the voltage into the physical value.

This field is only valid when

UseAnalogConversion

has the value 1.

Physical unit of the customer specific conversion.

This field is only valid when

UseAnalogConversion

has the value 1.

User Manual ATS400 Version 3.14.41 - 1 156

Report creation

5.1.1.2.6 Key words of the Leakage current test type

The

Leakage current

test type additionally uses the following key words:

Key word

Messmodell

Messmethode

Management

U_Source

U

I_max t f t_start

Auswertung

Polaritaet

SelftestMode

I_min

Explanation

Measurement model used.

Measurement method used.

0: Protective conductor current

1: Housing discharge current

Indicates how the supply of the test item is to be carried out.

0: Power off after test

1: Keep power on after test

2: Only power off

3: Power off only on error

4: Only power on

Source of the test supply.

Voltage of the test item supply

Upper threshold for the discharge current.

Test time

Frequency of the voltage

Will not be used.

Measurement channel used for the test:

0: I AC rms

1: I DC

2: I min

3: I max

4: I rms

Polarity used of the supply of the test item.

For single-phase test item:

0: Automotive

1: L1-> PE

2: L2-> PE

3: Mode B

4: Automatic with first error

5: L1 -> PE with first error

6: L2 -> PE with first error

With three-phase test item:

0: Clockwise rotation

1: Counterclockwise rotation

Always 0

Lower threshold for the discharge

User Manual ATS400 Version 3.14.41 - 1 157

Report creation

Key word

Phase

CreateLog

Timeinterval

Explanation

current.

Number of phases of the test item.

Indicates whether a log file will be created.

0 = No log file will be created

1 = A log file be created

Time interval between two samples which will be written to the log file. The value is in seconds.

This value will only be used in case the value in

CreateLog

is 1.

5.1.1.2.7 Key words of the Sight check test type

The

Sight check

test type additionally uses the following key words:

Key word

Abfrage

Explanation

Mode for termination.

5.1.1.2.8 Key words of the Data input test type

The

Data input

test type additionally uses the following key words:

Key word

EingabeMasken

Explanation

Bit samples containing screen fields instead of requirements.

5.1.1.2.9 Key words of the Batch run test type

The

Batch run

test type additionally uses the following key words:

Key word

SettingFlags

EingabeMasken

Explanation

Settings that indicate that certain fields are configured at other locations. This field is bit coded.

Bit samples containing screen fields instead of requirements.

5.1.1.2.10 Key words of the User-Interface test type

The

User-Interface

test type additionally uses the following key words:

Key word

In

Explanation

Entry bits

User Manual ATS400 Version 3.14.41 - 1 158

Report creation

Key word

InMask

Out

OutMask

Duration

Timeout

TimeoutEnable

NumInterface

Explanation

Mask for entry bits.

Output bits

Mask for output bits.

Duration of the output pulse.

Timeout when waiting for the status of the entry bits.

This parameter is only valid if the

TimeoutEnable

parameter is set.

States whether timeout is used.

0: Timeout is not used

1: Timeout is used

Number of the user IO interface used.

0: User-IO Interface of the internal IO-

CPU

1: User-IO Interface of the external IO-

CPU

5.1.1.2.11 Key words of the Resistance test type

The

Resistance

test type additionally uses the following key words:

Key word t

R_min

R_max

R_offset t_timeout

Explanation

Test time

Lower threshold value of the resistance.

Upper threshold value of the resistance.

Resistance of the measurement structure.

Maximum time after starting until the first measurement value is recorded.

5.1.1.2.12 Key words of the Continuity test type

The

Continuity

test type additionally uses the following key words:

Key word

TestVoltage

TestTime

MaxTestCurrent

Limit

Explanation

Maximum test voltage used for the measurement.

Time after that the measurement will be evaluated.

Maximum current during the test.

Limit for the evaluation.

User Manual ATS400 Version 3.14.41 - 1 159

Report creation

Interpretation

TypeOfSource

Evaluation of the measurement:

1: Pass on continuity

2: Pass on discontinuity

Type of the used source:

0: Alternating current

1: direct current

5.1.1.2.13 Key words of the PT100 test type

The

PT100

test type additionally uses the following key words:

Key word

T_min

T_max

R_offset

R0

Explanation

Lower threshold for the temperature.

Upper threshold for the temperature.

Resistance of the measurement structure.

Basic resistance of the measurement sensor.

5.1.1.2.14 Key words of the Lightcontrol test type

The

Lightcontrol

test type additionally uses the following key words:

Key word

Leuchte_ART

Leuchte_BEFEHL

Leuchte_DIM

Leuchte_DALIC

Leuchte_DALIV

Leuchte_DIMT

Explanation

Type of the light

Command to the light

Dim value

DALIC

DALIV

DIMT

5.1.1.2.15 Key words of the Dummy load test type

The

Dummy load

test type additionally uses the following key words:

Key word

Management

U_Source

Explanation

Indicates how the supply of the test item is to be carried out.

0: Power off after test

1: Keep power on after test

2: Only power off

3: Power off only on error

4: Only power on

Source of the test supply.

User Manual ATS400 Version 3.14.41 - 1 160

Report creation

Key word

U f

Channels

RGas

RCoil

Filament_enable

Filament_t

Filament_delay

Filament_I

Ignition_enable

Ignition_t

Ignition_delay

Ignition_U

Fct_enable

Fct_t

Fct_delay

Fct_Imin

Fct_Imax t_timeout

Filament_Imax

Explanation

Voltage of the test item supply

Frequency of the voltage

Number of channels

Gas resistance

Coil resistance

Active flame

Active flame

Flame delay

Current of the flame

Ignition available

Ignition time

Ignition delay

Ignition voltage

Function available

Duration of the function

Delay of the analysis

Lower threshold value of the current

Upper threshold value of the current

Always 0

Max. current of the flame

5.1.1.2.16 Key words of the Light Filament current test type

The

Light Filament current

test type additionally uses the following key words:

Key word

Management

Filament_enable

Filament_t

Filament_delay

Filament_I

Filament_Imax

Explanation

Indicates how to proceed after the test with supplying the test item.

0: Deactivate

1: Activate

Active flame

Active flame

Flame delay

Current of the flame

Max. current of the flame

User Manual ATS400 Version 3.14.41 - 1 161

Report creation

5.1.1.2.17 Key words of the External programm test type

The

External program

test type additionally uses the following key words:

Key word

SettingFlags

Explanation

Settings

5.1.1.3 Key words of the test step measurement values

Every test step has the following fields:

Key word

COUNTIO

COUNTNIO

COUNTERR

CountTests

StepCycleTime

Result dblResultValue1 dblResultValue2 strResultValue1 strResultValue2 strResultTestTime

ERROR

Explanation

Number of times this step was successfully carried out.

Number of times not carried out successfully.

Number of cancelled tests.

Total number of tests carried out.

Duration of the test.

Result of the test step:

UnTested

-> This test step was not carried out during this test.

Active

-> Test step is active, cannot occur in the file.

IO

-> This test step was rated as IO.

NIO

-> This test step was rated as NIO.

ERROR

-> This test step was cancelled with an error.

Jumped

-> This test step was skipped.

Numeric measurement value in SI unit.

Numeric value of the test size in SI units.

Formatted measurement value with a unit.

Formatted test value with a unit.

Formatted duration of the test in seconds.

Indication of the error when cancelling the measurement.

None

-> No error

Timeout_StartMeasurement

-> Time error when starting the measurement

Timeout_SetPassFail

-> not used

Timeout_MeasureTimeOvershoot

-> measurement time exceeded

User Manual ATS400 Version 3.14.41 - 1 162

Report creation

Key word

ERRORinfo

TestingUser

SightCheckInputText

Explanation

Invalid_TestState

-> invalid test status

Invalid_TestResult

-> invalid test result

Invalid_PVSteuerLT

-> invalid control word

Invalid_PVStatusPruefung

-> invalid status of the test

Invalid_TestParameter

-> invalid test parameter

Cancel

-> cancelled

Expanded error number.

User loged in during the test step.

User entry during the visual inspection.

5.1.1.4 Key words of the multiple test values test step

Every multiple test has the following fields:

Key word dblTestResult strTestResult

TestPoint

MultitestUser

Explanation

Result value of the multiple test in the

SI unit. This field contains only a figure value.

Result value as text.

User input for the multitest.

User loged in during the multitest.

5.1.1.5 Key words of the test step advices

Every test step has the following fields:

Key word

TextCount

Text_0

TextEnable_0

TextSize_0

Picture_0

Explanation

Always 3

Text to be output with indication beforehand.

Activation of the output with indication beforehand.

Font size with indication beforehand.

File name of the image with indication beforehand.

If the file can be reached via a relative path from the folder of the plan files,

User Manual ATS400 Version 3.14.41 - 1 163

Report creation

Key word

Outputtype_0

OutputSize_0

Text_1

TextEnable_1

TextSize_1

Picture_1

Outputtype_1

User Manual ATS400

Explanation

the relative path is entered, otherwise the absolute path is entered.

Image or text display with indication beforehand.

Values used:

Text

displays the text only.

Bild

displays the image only.

TextBild

displays the text and the image.

TextWithInputField

displays the text and an input field.

Small or large display with indication beforehand

Values used:

TextKlein

displays the small window for the text.

TextGroß

displays the large window for the text.

BildKlein

displays the small window with the image.

BildGroß

displays the large window with the image.

TextMitEingabeFeldKlein

displays the small window for the text and the input field.

TextMitEingabeFeldGroß

displays the large window for the text and the input field.

Nichts

displays no field.

Text to be output with indication during.

Activation of the output with indication during.

Font size with indication during.

File name of the image with indication during.

If the file can be reached via a relative path from the folder of the plan files, the relative path is entered, otherwise the absolute path is entered.

Image or text display with indication during

Values used:

Text

displays the text only.

Bild

displays the image only.

TextBild

displays the text and the image.

Version 3.14.41 - 1 164

Report creation

Key word

OutputSize_1

Text_2

TextEnable_2

TextSize_2

Picture_2

Outputtype_2

OutputSize_2

User Manual ATS400

Explanation

TextWithInputField

displays the text and an input field.

Small or large display with indication during.

Values used:

TextKlein

displays the small window for the text.

TextGroß

displays the large window for the text.

BildKlein

displays the small window with the image.

BildGroß

displays the large window with the image.

TextMitEingabeFeldKlein

displays the small window for the text and the input field.

TextMitEingabeFeldGroß

displays the large window for the text and the input field.

Nichts

displays no field.

Text to be issued in the event of error.

Activation of the output in the event of an error.

Font size in the event of an error.

File name of the image in the event of an error.

If the file can be reached via a relative path from the folder of the plan files, the relative path is entered, otherwise the absolute path is entered.

Image or text display in the event of an error.

Values used:

Text

displays the text only.

Bild

displays the image only.

TextBild

displays the text and the image.

TextWithInputField

displays the text and an input field.

Small or large display in the event of an error.

Values used:

TextKlein

displays the small window for the text.

TextGroß

displays the large window for the text.

Version 3.14.41 - 1 165

Report creation

Key word Explanation

BildKlein

displays the small window with the image.

BildGroß

displays the large window with the image.

TextMitEingabeFeldKlein

displays the small window for the text and the input field.

TextMitEingabeFeldGroß

displays the large window for the text and the input field.

Nichts

displays no field.

5.1.1.6 Key words of the test step process control

Every test step has the following fields:

Key word

Test

TestName

TestNameAbbr

Execute

InvertResult

View

Jump_IO_Mode

Jump_IO_Jump

Jump_IO_Repeat

Jump_NIO_Mode

Explanation

Non-localised name of the test step.

Localised name of the test step.

Short name of the test step.

Indicates whether the test step is to be carried out. The value is always

true

. The value is reserved for future purposes.

The result valuation IO and NIO is replaced. This field is only used for dummy tests.

Indicates whether the test step is to be displayed. The value is reserved for future purposes.

Indicates how the jump is to be carried out in the event of an IO event of the test step.

0 -> Next step

1 -> Go to the end

2 -> Go to a jump destination

Indicates the jump destination in the event of an IO result of the test step.

Indicates how often the test step is to be repeated in the event of an IO result of the test step. This value is always 0.

Indicates how the jump is to be carried out in the event of an NIO event of the test step.

User Manual ATS400 Version 3.14.41 - 1 166

Report creation

Key word

Jump_NIO_Jump

Jump_NIO_Repeat

Jump_Error_Mode

Jump_Error_Jump

Jump_Error_Repeat

Jump_Cancel_Mode

Jump_Cancel_Jump

Jump_Cancel_Repeat

TestStep

Explanation

0 -> Next step

1 -> Go to the end

2 -> Go to a jump destination

Indicates the jump destination in the event of an NIO result of the test step.

Indicates how often the test step is to be repeated in the event of an NIO result of the test step.

Indicates how the jump is to be carried out in the event of an error of the test step.

0 -> Next step

1 -> Go to the end

2 -> Go to a jump destination

This value is always 1.

Indicates the jump destination in the event of an error of the test step.

Indicates how often the test step is to be repeated in the event of an error of the test step. This value is always 0.

Indicates how the jump is to be carried out in the event of a cancellation of the test step.

0 -> Next step

1 -> Go to the end

2 -> Go to a jump destination

Indicates the jump destination in the event of a cancellation of the test step.

Indicates how often the test step is to be repeated in the event of a cancellation of the test step. This value is always 0.

Sequential number of the step.

5.1.2

HTML templates

HTML templates can be created and edited with any program that is able to save in the file type. Depending on the program, you must observe corresponding framework conditions.

Problematic in this context is that some programmes break down the key words into parts when editing them and insert formatting instructions. In this case, the key word is no longer identified.

User Manual ATS400 Version 3.14.41 - 1 167

Report creation

Use the HTML template; for the

Printer

or

Save as PDF

report option, the settings of the local Internet Explorer and of the printer are also required. The header and footer can be configured in the

Internet Explorer

in the

Page Setup

dialogue. The settings for the side orientation and paper size are made in the printer. The side edges must be set in the registration database with the

RegEdit

program from

Microsoft

Windows

.

Creation with Microsoft Word 2010

Creation with Open Office Writer

5.1.2.1 Creation with Microsoft Word 2012

Create a new empty document or use a template already created by you.

Save this document with

Save as

as a

Website, filtered (*.htm;*.html)

oder

website (*.htm;*.html)

. Note that you must edit the extension suggested from

htm

to

html

after setting all options.

User Manual ATS400 Version 3.14.41 - 1 168

Report creation

Under

Tools

open the dialogue

Web options...

.

In the

Browser

tab, deactivate the

Save new websites as websites in a file

option.

User Manual ATS400 Version 3.14.41 - 1 169

Report creation

In the

Files

tab, activate the

Save help files in a folder

option.

In the

Coding

tab, select from the

Save document as

selection the

Unicode (UTF8)

entry.

If you have selected

Website, filtered (*.htm;*.html)

you are given the

User Manual ATS400 Version 3.14.41 - 1 170

Report creation

following warning. Click on the

Yes

button.

Create your report.

If you save the report, you are given a file and a folder. The help files can be found in the folder. If you need to copy or move the template, you always also need to copy or move the folder with it.

User Manual ATS400 Version 3.14.41 - 1 171

Report creation

5.1.3

CSV templates

CSV is a text-based table format. You can edit it with every text editor or with most table calculations. If you do not use Microsoft Excel, it must be ensured that ISO-

8859-15 is used as the character coding. If you create CSV files with the

ATS400 X4

or

ATS400 X5

device variants, it must be a 7-bit ASCII, which means that the illustration is compatible. In this case, no umlauts may be used.

Creation with Microsoft Excel 2010

Creation with Open Office Calc

User Manual ATS400 Version 3.14.41 - 1 172

Report creation

5.1.3.1 Creation with Microsoft Excel 2010

Create a new empty document or use a template already created by you.

Delete all worksheets except for the first one.

Save this document using

Save as

as a

CSV (separator-separated) (*.csv)

.

User Manual ATS400 Version 3.14.41 - 1 173

Report creation

Create your report.

When saving, information is displayed to you. Click on the

Ok

button to save.

User Manual ATS400 Version 3.14.41 - 1 174

Report creation

User Manual ATS400 Version 3.14.41 - 1 175

Reference

6 Reference

This area describes all files used by

ETL DataView 3

that are managed or created within

ETL DataView 3

.

Results files

Plan files

Printserverkonfiguration

Matrixkonfiguration

6.1

Patterns

At various places in the configuration patterns can be used. This patterns are used similar to regular expressions .

A pattern consists of characters which have a special meaning. The following patterns are defined:

c t r a

C Description h a e r

? one character, e. g. A

*a string, e. g. Text

Aa letter (a-z, A-Z)

0a cipher (0-9)

Wan alphanumerical letter (a-z, A-Z, 0-9, _)

\ exact this character, e. g. \W at tis place a W is expected

( Extract the enclosed characters. Is used in the automatic test plan selection. It is

) not allowed to nest parantheses into each other.

Examples:

t e r n

P Description a t

0 It is expected to enter 6 ciphers, e. g. 202201

0

User Manual ATS400 Version 3.14.41 - 1 176

Reference t e r n

P Description a t

0

0

0

0

0

0

0

\

0

0

0 It is expected to enter 4 ciphers, a space and 4 ciphers, e. g. 0112 1000

0

0

0

0

0

\

0

0

0

0

\

0

0

0

0

0

0 It is expected to enter 6 ciphers, a space, 4 ciphers, a space and 4 ciphers, e. g.

0

202201 0112 1000

0

0

A It is expected to enter 3 letters and two ciphers , e. g. Typ25

A

A

6.2

Protective earth test

The PE-measurement is realized int 4-wire technic. This means that the measurement cables have different wires for the current (source) and the voltage measurement

(sense). These are connected together directly at the measurement point. Therefore the resistance of the measurement cable does not influence the result. This is true only for the effective resistance of the measurement cable but not for the inductivity and the resulting apperent resistance. Especially with small test resistance and

User Manual ATS400 Version 3.14.41 - 1 177

Reference

unfavourable installation could this lead to measure an increased resistance.

The resistance of the unit under test is normally a effective resistance. Errors resulting from the inductivity of the measurement cables will decrease with higer resistance of the unit under test. An apperent resistance of 10 mΩ for the measurement cable and

10 mΩ for the unit under test results in a measurement of 14 mΩ. With a unit under test of 50 mΩ the result will be 51 mΩ and with 100 mΩ it will be 100,5 mΩ.

In the case the inductivity of the measurement cable is disturbant the inductivity should be kept small. To achive this keep the aerea between the measurement cables small.

Additonally you must take care of a sufficent contact pressure to keep the contact resistance small.

6.2.1

Leading of the cable

Some examples for leading of the wires.

User Manual ATS400 Version 3.14.41 - 1 178

Reference

Unfavourable leading of the measurement cable. A large aerea is spanned between the both measurement cables and therefore the apperant resistance will increase.

Also unfavourable leading of the measurement cables. The aerea between the measurement cables is smaller but the apperant resistance will increase due to the loop.

Favourable leading of the measurement wires. The aerea between the measurement cables is small keeping the apperant resistane small.

User Manual ATS400 Version 3.14.41 - 1 179

Reference

6.3

High voltage test

For the

ATS400

three groups of high voltage test modules are offered. These will be named in the following with there short names. The short name HVDC will be used if the text is valid for the HVDC7 and HVDC3 module.

Short name

HVAC

HVDC7

HVDC3

Module

Test with alternate current.

Test with direct current. These modules have a testing voltage of 5 KV or more.

Test with direct current. These modules have a testing voltage up to 3600 Volt and are current limited.

Informations for configuration of the high voltage test you will find for

HVAC-Test , the

HVDC7-Test and the HVDC3-Test in the part

test plan editing

.

6.3.1

Evaluation of the measurement

6.3.1.1 Error messages for high voltage AC

Value

19

20

21

22

0

16

17

18

23

24

25

Description

No error

Upper current limit exceeded.

Break through deteced with peak current.

Testing voltage not reached.

Lower current limit not reached.

Break through in ramp. Limit for spark detection exceeded.

Upper current during ramp exceeded.

Break through during static phase. Limit for spark detection exceeded.

Upper current limit during static phase exceeded.

Break through in ramp. Gradient for spark detection exceeded.

Break through during static phase. Gradient for spark detection exceeded.

6.3.1.2 Error messages for high voltage DC

Value

0

32

33

Description

No error

Upper current limit exceeded.

Break through deteced with peak current.

User Manual ATS400 Version 3.14.41 - 1 180

Reference

Value

34

35

36

37

38

39

40

41

42

Description

Lower current limit not reached.

Break through.

Testing voltage not reached.

Break through in ramp. Limit for spark detection exceeded.

Upper current during ramp exceeded.

Break through during static phase. Limit for spark detection exceeded.

Upper current limit during static phase exceeded.

Break through in ramp. Gradient for spark detection exceeded.

Break through during static phase. Gradient for spark detection exceeded.

6.3.1.3 Test for required voltage

After powering on resp. at the end of the ramp when the current voltage does not change any more it is checked that the current voltage has reached more than 95% of the voltage in parameter

Test voltage

.

6.3.1.4 Current error at upper limit

This error will be detected when the measured current exceeds the current in parameter

Maximum Current

.

This evaluation will be done in the following cases:

Phase

Rising ramp

Static

Falling ramp

HVAC

Yes

Yes

Yes

HVDC

No

Yes

No

6.4

Results files

The results files have the extension

.result

. The place of saving and the file name result from the details in

Settings

->

File storage

->

Result

. This file has the format of an XML file with UTF-8 coding with BOM.

<ResultData>

is the enclosing tag.

<ResultData>/<Header>

contains the header data.

<ResultData>/<Identification>

contains the identifications.

<ResultData>/<TestSteps>

contains the list of test steps.

<ResultData>/<TestSteps>/<TestStep Index="n">

contains a test step. The

number in

Index

is incremented for every test step.

<ResultData>/<Result>

contains the overall result as well as the data of the

User Manual ATS400 Version 3.14.41 - 1 181

Reference

automatic test plan selection.

<ResultData>/PlanOptionen>

contains the options of the test plan.

<ResultData>/PlanOptionen>/<ReportOptionen>

contains the list of reports to be

created.

<ResultData>/PlanOptionen>/<ReportOptionen>/<ReportOption>

contains the

report settings for a report.

<ResultData>/<Validation>

contains validation data.

6.4.1

Header

The data in this block can be found under

<ResultData>/<Header>

.

Tag

DataView_Version

TestPlan_Version

PlanName

PlanPath

PlanDescription

Explanation

Version of

ETL DataView 3

with which this file was created.

Version of the format of the file.

Name of the test plan, is filled with the file name when being created.

Relative path to the application where the test plan was saved.

Description of the test plan. This value can be entered by the user in

Settings

->

Text

.

PlanCreatedByUse

r User who created the test plan first. If user administration is not active, the field is empty.

PlanCreatedByWorkstation

Test station where the test plan was created. The name of the test station can be stated in

Settings

-

>

Workstation

->

Base settings

.

PlanCreatedOnDate

PlanEditedByUser

PlanEditedByWorkstation

PlanEditedOnDate

FileSavedByUser

FileSavedByWorkstation

Date of creating the test plan. The date is in the format in accordance with the country settings valid at this moment in time.

User who modified the test plan last. If user administration is not active, the field is empty.

Test station where the test plan was modified last.

The name of the test station can be stated in

Settings

->

Workstation

->

Base settings

.

Date of modifying the test plan. The date is in the format in accordance with the country settings valid at this moment in time.

User who saved the test plan or the result file.

Test station where the test plan or the result file was saved. The name of the test station can be stated in

Settings

->

Workstation

->

Base settings

.

FileSavedOnDate

Date of saving the test plan or the result file. The date is in the format in accordance with the country

User Manual ATS400 Version 3.14.41 - 1 182

Reference

Tag Explanation

settings valid at this moment in time.

PlanIdentificationEnabled

The

<ResultData>\<Identification>

block exists.

PlanBatchEnabled

PlanTestSteps

Intended for future purposes.

Number of test steps in the test plan.

6.4.2

Identification

The data in this block can be found under

<ResultData>/<Identification>

.

Tag

IDs

ID_n

Explanation

Number of IDs.

Identification for the test plan. The tags are numbered from 0 to the number of IDs - 1, i.e. have the values

ID_0, ID_1, etc.

6.4.3

Teststeps

The data in this block can be found under

<ResultData>/<TestSteps>

.

This block contains a list of the individual test steps. Every test step is a

block of its

own.

6.4.3.1 TestStep

The data in this block can be found under

<ResultData>/<TestSteps>/<TestStep

Index="n">

.

Attribute

Index

Explanation

Index of the test step, counting starts from 0

Tag

Test

TestName

Execute

InvertResult

View

Parameter

Explanation

ID of the test type

Non-localised name of the test step.

Indicates whether the test step is to be carried out.

The value is always

true

. The value is reserved for future purposes.

The result valuation IO and NIO is replaced. This field is only used for dummy tests.

Indicates whether the test step is to be displayed. The value is reserved for future purposes.

Number of parameters of this test step. The number of parameters depends on the test step.

User Manual ATS400 Version 3.14.41 - 1 183

Reference

Tag

Parameter_n

Jump

Jump_IO_Mode

Jump_IO_Jump

Jump_IO_Repeat

Jump_NIO_Mode

Jump_NIO_Jump

Jump_NIO_Repeat

Jump_Error_Mode

Jump_Error_Jump

Jump_Error_Repeat

Jump_Cancel_Mode

Jump_Cancel_Jump

Jump_Cancel_Repeat

Text

Explanation

Parameter of the test step. These parameters are explained individually for every test type.

This fields always has the value

Jump

.

Indicates how the jump is to be carried out in the event of an IO event of the test step.

0 -> Next step

1 -> Go to the end

2 -> Go to a jump destination

Indicates the jump destination in the event of an IO result of the test step.

Indicates how often the test step is to be repeated in the event of an IO result of the test step. This value is always 0.

Indicates how the jump is to be carried out in the event of an NIO event of the test step.

0 -> Next step

1 -> Go to the end

2 -> Go to a jump destination

Indicates the jump destination in the event of an NIO result of the test step.

Indicates how often the test step is to be repeated in the event of an NIO result of the test step.

Indicates how the jump is to be carried out in the event of an error of the test step.

0 -> Next step

1 -> Go to the end

2 -> Go to a jump destination

This value is always 1.

Indicates the jump destination in the event of an error of the test step.

Indicates how often the test step is to be repeated in the event of an error of the test step. This value is always 0.

Indicates how the jump is to be carried out in the event of a cancellation of the test step.

0 -> Next step

1 -> Go to the end

2 -> Go to a jump destination

Indicates the jump destination in the event of a cancellation of the test step.

Indicates how often the test step is to be repeated in the event of a cancellation of the test step. This value is always 0.

Immer 3.

User Manual ATS400 Version 3.14.41 - 1 184

Reference

Tag

Text_0

TextEnable_0

TextSize_0

Picture_0

Outputtype_0

OutputSize_0

Text_1

TextEnable_1

TextSize_1

Picture_1

Outputtype_1

OutputSize_1

Explanation

Text to be output with indication beforehand.

Activation of the output with indication beforehand.

Font size with indication beforehand.

File name of the image with indication beforehand.

If the file can be reached via a relative path from the folder of the plan files, the relative path is entered, otherwise the absolute path is entered.

Image or text display with indication beforehand.

Values used:

Text

displays the text only.

Bild

displays the image only.

TextBild

displays the text and the image.

TextWithInputField

displays the text and an input field.

Small or large display with indication beforehand

Values used:

TextKlein

displays the small window for the text.

TextGroß

displays the large window for the text.

BildKlein

displays the small window with the image.

BildGroß

displays the large window with the image.

TextMitEingabeFeldKlein

displays the small window for the text and the input field.

TextMitEingabeFeldGroß

displays the large window for the text and the input field.

Nichts

displays no field.

Text to be output with indication during.

Activation of the output with indication during.

Font size with indication during.

File name of the image with indication during.

If the file can be reached via a relative path from the folder of the plan files, the relative path is entered, otherwise the absolute path is entered.

Image or text display with indication during

Values used:

Text

displays the text only.

Bild

displays the image only.

TextBild

displays the text and the image.

TextWithInputField

displays the text and an input field.

Small or large display with indication during.

Values used:

TextKlein

displays the small window for the text.

TextGroß

displays the large window for the text.

BildKlein

displays the small window with the image.

User Manual ATS400 Version 3.14.41 - 1 185

Reference

Tag

Text_2

TextEnable_2

TextSize_2

Picture_2

Outputtype_2

OutputSize_2

Tags

TagName_n

TagData_n

Template

Show

Size

Explanation

BildGroß

displays the large window with the image.

TextMitEingabeFeldKlein

displays the small window for the text and the input field.

TextMitEingabeFeldGroß

displays the large window for the text and the input field.

Nichts

displays no field.

Text to be issued in the event of error.

Activation of the output in the event of an error.

Font size in the event of an error.

File name of the image in the event of an error.

If the file can be reached via a relative path from the folder of the plan files, the relative path is entered, otherwise the absolute path is entered.

Image or text display in the event of an error.

Values used:

Text

displays the text only.

Bild

displays the image only.

TextBild

displays the text and the image.

TextWithInputField

displays the text and an input field.

Small or large display in the event of an error.

Values used:

TextKlein

displays the small window for the text.

TextGroß

displays the large window for the text.

BildKlein

displays the small window with the image.

BildGroß

displays the large window with the image.

TextMitEingabeFeldKlein

displays the small window for the text and the input field.

TextMitEingabeFeldGroß

displays the large window for the text and the input field.

Nichts

displays no field.

Number of the field names of the test step. This value is

0, except for the batch and data input test steps.

Name of the field. The

n

postfix corresponds to the number of the field counted from 0. This entry only exists if

Tags

is not equal to 0.

Pre-occupation or sample of the field. The

n

postfix corresponds to the number of the field counted from 0.

This entry only exists if

Tags

is not equal to 0.

Outdated, is no longer used.

Indicates whether a user notice is displayed before the multiple test.

Font size of the text for a multiple test.

User Manual ATS400 Version 3.14.41 - 1 186

Reference

Tag

Text

Display

PicturePath

StepCycleTime

Count_Tests

Count_IO

Count_NIO

Count_ERROR

Result

ResultValue_1

ResultValue_2

TestTime

Error

Explanation

Text to be issued in the event of a multiple test.

Image or text display in the event of a multiple test.

Values used:

Text

displays the text only.

Bild

displays the image only.

TextBild

displays the text and the image.

TextWithInputField

displays the text and an input field.

File name of the image in the event of a multiple test.

If the file can be reached via a relative path from the folder of the plan files, the relative path is entered, otherwise the absolute path is entered.

Duration of the test.

Total number of tests carried out.

Number of times this step was successfully carried out.

Number of times not carried out successfully.

Number of cancelled tests.

Result of the test step:

UnTested

-> This test step was not carried out during this test.

Active

-> Test step is active, cannot occur in the file.

IO

-> This test step was rated as IO.

NIO

-> This test step was rated as NIO.

ERROR

-> This test step was cancelled with an error.

Jumped

-> This test step was skipped.

Numeric measurement value in SI unit.

Numeric value of the test size in SI units.

Formatted duration of the test in seconds.

Indication of the error when cancelling the measurement.

None

-> No error

Timeout_StartMeasurement

-> Time error when starting the measurement

Timeout_SetPassFail

-> not used

Timeout_MeasureTimeOvershoot

-> measurement time exceeded

Invalid_TestState

-> invalid test status

Invalid_TestResult

-> invalid test result

Invalid_PVSteuerLT

-> invalid control word

Invalid_PVStatusPruefung

-> invalid status of the test

Invalid_TestParameter

-> invalid test parameter

Cancel

-> cancelled

User Manual ATS400 Version 3.14.41 - 1 187

Reference

Tag Explanation

ErrorInfo

TestPoint_n

TestingUser

SightCheckInputText

MultitestUser_n

Expanded error number.

User input for the multitest.

The

n

post fix is counted upwards from 0.

User loged in during the test step.

User entry during the visual inspection.

User loged in during the multitest.

The

n

post fix is counted upwards from 0.

6.4.3.1.1 Protective Earth

Parameters of the

Protective Earth

test step.

Tag

Parameter_0

Parameter_1

Parameter_2

Parameter_3

Parameter_4

Parameter_5

Parameter_6

Parameter_7

Parameter_8

Parameter_9

Parameter_10

Parameter_11

bis

Parameter_26

Parameter

StartCondition

U

I

StartConditionMas k

R_max t t_delay f t_start

EN60204

Polung

PolungCAN_00

PolungCan_15

bis

Explanation

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartConditionMask

field.

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartCondition

field.

Idle voltage

Test current

Upper threshold value of the resistance.

Test time

Start delay between fulfilling the start condition and the start of the test.

Frequency of the voltage

Delay time for the start scenario when starting the test until starting the analysis. This value is not valid for all start scenarios.

Indication whether the test is performed in accordance with EN 60204.

Always 0.

Contacting setting of an external relay matrix.

Setting of the relay matrix in accordance with the choice in the program. The

User Manual ATS400 Version 3.14.41 - 1 188

Reference

Tag

Parameter_27

Parameter_28

Parameter_29

Parameter_30

Parameter_31

Parameter_32

Parameter_33

Parameter_34

Parameter_35

Parameter_36

Parameter_37

Parameter_38

Parameter_39

Parameter_40

Parameter_41

Parameter_42

Parameter_43

Parameter Explanation

StateChangeBits

display is bit coded.

Always 0, never used.

Multitest_Enable

Multitest is active, if the value does not equal 0.

Multitest_Auswert ung

Multitest_Endekri terium

Method of analysing the multitest.

0 = worst measurement value.

Method of ending the multitest.

0 = number of measurements

1 = pass key

Number of multitest inspections.

Multitest_Endekri terium_Anzahl

ErrorStartConditi on

ErrorStartConditi onMask

Condition for switching forward in the event of an error.

Mask for the condition for switching further in the event of an error.

Always 0, never used.

ErrorStateChangeB its

ErrorWSBDelay

Delay of switching further in the event of an error.

ErrorWSBTimeValid

Delay time active.

StartStateChange

Indicates whether the switching condition needs to be maintained when testing the start conditions.

0 = no waiting for switch condition

1 = Wait for switching condition

DC_Enable

Indicates whether the test is carried out with direct current voltage.

0: Alternating current

1: Direct current

StartStateJump

Indicates whether the

Jump

button is displayed.

Repeating

StartButton

PassButton

StartandJump

Indicates how often the test step is carried out.

Indicates whether the

Start

button is displayed.

Indicates whether the

Pass

button is displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

User Manual ATS400 Version 3.14.41 - 1 189

Reference

Tag

Parameter_44

Parameter_45

Parameter_46

Parameter_47

Parameter_48

Parameter

StartTime_Delay

Explanation

Wait time for the signal

Start-Button

, in case the start condition is

monitored

.

StartPETime_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition is

monitored

.

StartKUTime_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition is

monitored

.

StartSHKHVTime_De lay

Wait time for the signal

Safety Circuit

HV

, in case the start condition is

monitored

.

PassButtonVisible

Indicates whether the button

Pass

will be shown during a Multitest.

6.4.3.1.2 High voltage AC

Parameters of the

High voltage AC

test step.

Tag

Parameter_0

Parameter_1

Parameter_2

Parameter_3

Parameter_4

Parameter_5

Parameter_6

Parameter_7

Parameter_8

Parameter_9

Parameter_10

Parameter

StartCondition

StartConditionMas k

U

I_min

I_max t t_delay

Ramp

U_start t_up t_down

Explanation

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartConditionMask

field.

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartCondition

field.

Test voltage

Lower threshold

Upper threshold

Test duration

Start delay between fulfilling the start condition and the start of the test.

Indicates whether the ramp is active.

Start voltage

Increase time of the ramp. This parameter is only valid if the ramp is active.

Drop time of the ramp. This parameter is only valid if the ramp is active.

User Manual ATS400 Version 3.14.41 - 1 190

Reference

Tag

Parameter_11

Parameter_12

Parameter_13

Parameter_14

bis

Parameter_29

Parameter_30

Parameter_31

Parameter Explanation f t_start

Polung

PolungCAN_00

PolungCan_15

bis

StateChangeBits

Frequency of the voltage

Delay time for the start scenario when starting the test until starting the analysis. This value is not valid for all start scenarios.

Contacting setting of an external relay matrix.

Setting of the relay matrix in accordance with the choice in the program. The display is bit coded.

Always 0, never used.

Multitest_Enable

Multitest is active, if the value does not equal 0.

Parameter_32

Parameter_33

Parameter_34

Parameter_35

Parameter_36

Parameter_37

Parameter_38

Parameter_39

Parameter_40

Parameter_41

Parameter_42

Parameter_43

Multitest_Auswert ung

Multitest_Endekri terium

Method of analysing the multitest.

0 = worst measurement value.

Method of ending the multitest.

0 = number of measurements

1 = pass key

Number of multitest inspections.

Multitest_Endekri terium_Anzahl

Offset

ErrorStartConditi on

ErrorStartConditi onMask

Always 0, will not be used.

Condition for switching forward in the event of an error.

Mask for the condition for switching further in the event of an error.

Always 0, never used.

ErrorStateChangeB its

ErrorWSBDelay

Delay of switching further in the event of an error.

ErrorWSBTimeValid

Delay time active.

StartStateChange

Indicates whether the switching condition needs to be maintained when testing the start conditions.

0 = no waiting for switch condition

1 = Wait for switching condition

Quelle

Selected current source.

This parameter is only used in the event of a hot high voltage.

f_Source

Frequency of the source.

This parameter is only used in the event of a hot high voltage.

User Manual ATS400 Version 3.14.41 - 1 191

Reference

Tag

Parameter_44

Parameter_45

Parameter_46

Parameter_47

Parameter_48

Parameter_49

Parameter_50

Parameter_51

Parameter_52

Parameter_53

Parameter_54

Parameter_55

Parameter_56

Parameter_57

Parameter_58

Parameter

U_Source

StartStateJump

Explanation

Voltage of the source.

This parameter is only used in the event of a hot high voltage.

Indicates whether the

Jump

button is displayed.

Management

Auswertung

Repeating

StartButton

PassButton

StartandJump

StartandJump

CreateLog

Timeinterval

Indicates how the supply of the test item is to be carried out.

0: Power off after test

1: Keep power on after test

2: Only power off

3: Power off only on error

4: Only power on

Type of the analysis of the test.

0 = end the test time

1 = end with start signal

Indicates how often the test step is carried out.

Indicates whether the

Start

button is displayed.

Indicates whether the

Pass

button is displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

Indicates whether a log file will be created.

0 = No log file will be created

1 = A log file be created

Time interval between two samples which will be written to the log file. The value is in seconds.

This value will only be used in case the value in

CreateLog

is 1.

StartTime_Delay

Wait time for the signal

Start-Button

, in case the start condition is

monitored

.

StartPETime_Delay

Wait time for the signal

PE-Testprobe

, in case the start condition is

monitored

.

StartKUTime_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition is

monitored

.

StartSHKHVTime_De

Wait time for the signal

Safety Circuit

User Manual ATS400 Version 3.14.41 - 1 192

Reference

Tag

Parameter_59

Parameter_60

Parameter_61

Parameter lay

Explanation

HV

, in case the start condition is

monitored

.

PassButtonVisible

Indicates whether the button

Pass

will be shown during a Multitest.

SparkDetection

ContinousRamp

This parameter is only valid on deivices with HVdc7/8 module and a HMP supporting spark detection.

Setting of the spark detection.

0 = Off

1 = Coarse

2 = Normal

3 = Fine

Indicates that the rampe will be continued after the test step.

This parameter is only valid for the variant for KWP protocol and high voltage tester of Series 36.

0: Ramp will not be continued

1: Ramp will be continued

6.4.3.1.3 High voltage DC

Parameters of the

High voltage DC

test step.

Tag

Parameter_0

Parameter_1

Parameter_2

Parameter_3

Parameter_4

Parameter_5

Parameter_6

Parameter_7

Parameter_8

Parameter_9

Parameter

StartCondition

StartConditionMas k

U

I_min

I_max t t_delay

Ramp

U_start t_up

Explanation

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartConditionMask

field.

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartCondition

field.

Test voltage

Lower threshold

Upper threshold

Test duration

Start delay between fulfilling the start condition and the start of the test.

Indicates whether the ramp is active.

Start voltage

Increase time of the ramp. This

User Manual ATS400 Version 3.14.41 - 1 193

Reference

Tag

Parameter_10

Parameter_11

Parameter_12

Parameter_13

Parameter_14

bis

Parameter_29

Parameter_30

Parameter_31

Parameter Explanation t_down t_start

parameter is only valid if the ramp is active.

Drop time of the ramp. This parameter is only valid if the ramp is active.

Delay time for the start scenario when starting the test until starting the analysis. This value is not valid for all start scenarios.

U_discharge

Polung

PolungCAN_00

PolungCan_15

bis

StateChangeBits

Discharge voltage

Contacting setting of an external relay matrix.

Setting of the relay matrix in accordance with the choice in the program. The display is bit coded.

Always 0, never used.

Multitest_Enable

Multitest is active, if the value does not equal 0.

Parameter_32

Parameter_33

Parameter_34

Parameter_35

Parameter_36

Parameter_37

Parameter_38

Parameter_39

Parameter_40

Parameter_41

Multitest_Auswert ung

Multitest_Endekri terium

Method of analysing the multitest.

0 = worst measurement value.

Method of ending the multitest.

0 = number of measurements

1 = pass key

Number of multitest inspections.

Multitest_Endekri terium_Anzahl

ErrorStartConditi on

ErrorStartConditi onMask

Condition for switching forward in the event of an error.

Mask for the condition for switching further in the event of an error.

Always 0, never used.

ErrorStateChangeB its

ErrorWSBDelay

Delay of switching further in the event of an error.

ErrorWSBTimeValid

Delay time active.

StartStateChange

Indicates whether the switching condition needs to be maintained when testing the start conditions.

0 = no waiting for switch condition

1 = Wait for switching condition

Quelle

Selected current source.

This parameter is only used in the event of a hot high voltage.

User Manual ATS400 Version 3.14.41 - 1 194

Reference

Tag

Parameter_42

Parameter_43

Parameter_44

Parameter_45

Parameter_46

Parameter_47

Parameter_48

Parameter_49

Parameter_50

Parameter_51

Parameter_52

Parameter_53

Parameter_54

Parameter_55

Parameter f_Source

U_Source

StartStateJump

Explanation

Frequency of the source.

This parameter is only used in the event of a hot high voltage.

Voltage of the source.

This parameter is only used in the event of a hot high voltage.

Indicates whether the

Jump

button is displayed.

Management

Auswertung

Repeating

StartButton

PassButton

StartandJump

CreateLog

Timeinterval

Indicates how the supply of the test item is to be carried out.

0: Power off after test

1: Keep power on after test

2: Only power off

3: Power off only on error

4: Only power on

Type of the analysis of the test.

0 = end the test time

1 = end with start signal

Indicates how often the test step is carried out.

Indicates whether the

Start

button is displayed.

Indicates whether the

Pass

button is displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

Indicates whether a log file will be created.

0 = No log file will be created

1 = A log file be created

Time interval between two samples which will be written to the log file. The value is in seconds.

This value will only be used in case the value in

CreateLog

is 1.

StartTime_Delay

Wait time for the signal

Start-Button

, in case the start condition is

monitored

.

StartPETime_Delay

Wait time for the signal

PE-Testprobe

, in case the start condition is

monitored

.

StartKUTime_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition is

monitored

.

User Manual ATS400 Version 3.14.41 - 1 195

Reference

Tag

Parameter_56

Parameter_57

Parameter_58

Parameter_59

Parameter

StartSHKHVTime_De lay

Explanation

Wait time for the signal

Safety Circuit

HV

, in case the start condition is

monitored

.

PassButtonVisible

Indicates whether the button

Pass

will be shown during a Multitest.

SparkDetection

ContinousRamp

This parameter is only valid on deivices with HVdc7/8 module and a HMP supporting spark detection.

Setting of the spark detection.

0 = Off

1 = Coarse

2 = Normal

3 = Fine

Indicates that the rampe will be continued after the test step.

This parameter is only valid for the variant for KWP protocol and high voltage tester of Series 36.

0: Ramp will not be continued

1: Ramp will be continued

6.4.3.1.4 Insulation

Parameters of the

Insulation

test step.

Tag

Parameter_0

Parameter_1

Parameter_2

Parameter_3

Parameter_4

Parameter_5

Parameter_6

Parameter_7

Parameter_8

Parameter

StartCondition

StartConditionMas k

U

R_min t t_delay

Ramp

U_start t_up

Explanation

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartConditionMask

field.

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartCondition

field.

Test voltage

Lower threshold of the insulation resistance.

Test duration

Start delay between fulfilling the start condition and the start of the test.

Indicates whether the ramp is active.

Start voltage

Increase time of the ramp. This

User Manual ATS400 Version 3.14.41 - 1 196

Reference

Tag

Parameter_9

Parameter_10

Parameter_11

Parameter_12

Parameter_13

bis

Parameter_28

Parameter_29

Parameter_30

Parameter Explanation t_down t_start

parameter is only valid if the ramp is active.

Drop time of the ramp. This parameter is only valid if the ramp is active.

Delay time for the start scenario when starting the test until starting the analysis. This value is not valid for all start scenarios.

U_discharge

Polung

PolungCAN_00

PolungCan_15

bis

StateChangeBits

Discharge voltage

Contacting setting of an external relay matrix.

Setting of the relay matrix in accordance with the choice in the program. The display is bit coded.

Always 0, never used.

Multitest_Enable

Multitest is active, if the value does not equal 0.

Parameter_31

Parameter_32

Parameter_33

Parameter_34

Parameter_35

Parameter_36

Parameter_37

Parameter_38

Parameter_39

Parameter_40

Multitest_Auswert ung

Multitest_Endekri terium

Method of analysing the multitest.

0 = worst measurement value.

Method of ending the multitest.

0 = number of measurements

1 = pass key

Number of multitest inspections.

Multitest_Endekri terium_Anzahl

ErrorStartConditi on

ErrorStartConditi onMask

Condition for switching forward in the event of an error.

Mask for the condition for switching further in the event of an error.

Always 0, never used.

ErrorStateChangeB its

ErrorWSBDelay

Delay of switching further in the event of an error.

ErrorWSBTimeValid

Delay time active.

StartStateChange

Indicates whether the switching condition needs to be maintained when testing the start conditions.

0 = no waiting for switch condition

1 = Wait for switching condition

Quelle

Selected current source.

This parameter is only used in the event of a hot high voltage.

User Manual ATS400 Version 3.14.41 - 1 197

Reference

Tag

Parameter_41

Parameter_42

Parameter_43

Parameter_44

Parameter_45

Parameter_46

Parameter_47

Parameter_48

Parameter_49

Parameter_50

Parameter_51

Parameter_52

Parameter_53

Parameter_54

Parameter f_Source

U_Source

StartStateJump

Explanation

Frequency of the source.

This parameter is only used in the event of a hot high voltage.

Voltage of the source.

This parameter is only used in the event of a hot high voltage.

Indicates whether the

Jump

button is displayed.

Management

Auswertung

Repeating

StartButton

PassButton

StartandJump

CreateLog

Timeinterval

Indicates how the supply of the test item is to be carried out.

0: Power off after test

1: Keep power on after test

2: Only power off

3: Power off only on error

4: Only power on

Type of the analysis of the test.

0 = end the test time

1 = end with start signal

Indicates how often the test step is carried out.

Indicates whether the

Start

button is displayed.

Indicates whether the

Pass

button is displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

Indicates whether a log file will be created.

0 = No log file will be created

1 = A log file be created

Time interval between two samples which will be written to the log file. The value is in seconds.

This value will only be used in case the value in

CreateLog

is 1.

StartTime_Delay

Wait time for the signal

Start-Button

, in case the start condition is

monitored

.

StartPETime_Delay

Wait time for the signal

PE-Testprobe

, in case the start condition is

monitored

.

StartKUTime_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition is

monitored

.

User Manual ATS400 Version 3.14.41 - 1 198

Reference

Tag

Parameter_55

Parameter_56

Parameter

StartSHKHVTime_De lay

Explanation

Wait time for the signal

Safety Circuit

HV

, in case the start condition is

monitored

.

PassButtonVisible

Indicates whether the button

Pass

will be shown during a Multitest.

6.4.3.1.5 FCT-Current

Parameters of the

FCT-Current

test step.

Tag

Parameter_0

Parameter_1

Parameter_2

Parameter_3

Parameter_4

Parameter_5

Parameter_6

Parameter_7

Parameter_8

Parameter_9

Parameter_10

Parameter_11

Parameter

StartCondition

StartConditionMas k

U_Source

U

I_min

I_max t t_delay f t_start t_timeout

Gradient

Explanation

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartConditionMask

field.

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartCondition

field.

Source of the test supply.

Voltage of the test item supply

Lower threshold value during the analysis. Not valid if the analysis is set to

Pass/Fail

. The unit and thresholds depend on the channel.

Upper threshold value during the analysis. Not valid if the analysis is set to

Pass/Fail

. The unit and thresholds depend on the channel.

Test time

Start delay between fulfilling the start condition and the start of the test.

Frequency of the voltage

Delay time for the start scenario when starting the test until starting the analysis. This value is not valid for all start scenarios.

Timeout for the start of the analysis.

The value is not used for all start scenarios.

Gradient for the start of the analysis.

This value is not used for all start

User Manual ATS400 Version 3.14.41 - 1 199

Reference

Tag

Parameter_12

Parameter_13

Parameter_14

Parameter_15

bis

Parameter_30

Parameter_31

Parameter_32

Parameter_33

Parameter

Management

Scenario

Polung

PolungCAN_00

PolungCan_15

Auswertung

bis

Explanation

scenarios.

Indicates how the supply of the test item is to be carried out.

0: Power off after test

1: Keep power on after test

2: Only power off

3: Power off only on error

4: Only power on

Type of the start of the analysis

Contacting setting of an external relay matrix.

Setting of the relay matrix in accordance with the choice in the program. The display is bit coded.

Type of the analysis

0 = measurement

1 =

Pass/Fail

button

StateChangeBits

Always 0, never used.

Multitest_Enable

Multitest is active, if the value does not equal 0.

Parameter_34

Parameter_35

Parameter_36

Parameter_37

Parameter_38

Parameter_39

Parameter_40

Parameter_41

Parameter_42

Multitest_Auswert ung

Multitest_Endekri terium

Method of analysing the multitest.

0 = worst measurement value.

Method of ending the multitest.

0 = number of measurements

1 = pass key

Number of multitest inspections.

Multitest_Endekri terium_Anzahl

ErrorStartConditi on

ErrorStartConditi onMask

Condition for switching forward in the event of an error.

Mask for the condition for switching further in the event of an error.

Always 0, never used.

ErrorStateChangeB its

ErrorWSBDelay

Delay of switching further in the event of an error.

ErrorWSBTimeValid

Delay time active.

Kanal

Measurement channel

0 = current

1 = analogue channel1

2 = analogue channel2

3 = analogue channel3

4 = analogue channel4

User Manual ATS400 Version 3.14.41 - 1 200

Reference

Tag

Parameter_43

Parameter_44

Parameter_45

Parameter_46

Parameter_47

Parameter_48

Parameter_49

Parameter_50

Parameter_51

Parameter_52

Parameter_53

Parameter_54

Parameter_55

Parameter_56

Parameter Explanation

SourceIsDC

5 = voltage measurement

External source provides direct current voltage

StartStateChange

Indicates whether the switching condition needs to be maintained when testing the start conditions.

0 = no waiting for switch condition

1 = Wait for switching condition

StartStateJump

Indicates whether the

Jump

button is displayed.

Graphic

Repeating

StartButton

PassButton

StartandJump

CreateLog

Timeinterval

Type of the graphics settings

0 = no graphics

1 = only display graphic

2 = display and save graphic

Indicates how often the test step is carried out.

Indicates whether the

Start

button is displayed.

Indicates whether the

Pass

button is displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

Indicates whether a log file will be created.

0 = No log file will be created

1 = A log file be created

Time interval between two samples which will be written to the log file. The value is in seconds.

This value will only be used in case the value in

CreateLog

is 1.

StartTime_Delay

Wait time for the signal

Start-Button

, in case the start condition is

monitored

.

StartPETime_Delay

Wait time for the signal

PE-Testprobe

, in case the start condition is

monitored

.

StartKUTime_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition is

monitored

.

StartSHKHVTime_De lay

Wait time for the signal

Safety Circuit

HV

, in case the start condition is

monitored

.

User Manual ATS400 Version 3.14.41 - 1 201

Reference

6.4.3.1.6 Leakage current

Parameters of the

Leakage current

test step.

Tag

Parameter_0

Parameter_1

Parameter_2

Parameter_3

Parameter_4

Parameter_5

Parameter_6

Parameter_7

Parameter_8

Parameter_9

Parameter_10

Parameter_11

Parameter_12

Parameter_13

Parameter

StartCondition

StartConditionMas k

Messmodell

Messmethode

Management

U_Source

U

I_max t t_delay f t_start

Auswertung

Polaritaet

Explanation

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartConditionMask

field.

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartCondition

field.

Measurement model used.

Measurement method used.

0: Protective conductor current

1: Housing discharge current

Indicates how the supply of the test item is to be carried out.

0: Power off after test

1: Keep power on after test

2: Only power off

3: Power off only on error

4: Only power on

Source of the test supply.

Voltage of the test item supply

Upper threshold for the discharge current.

Test time

Start delay between fulfilling the start condition and the start of the test.

Frequency of the voltage

Will not be used.

Measurement channel used for the test:

0: I AC rms

1: I DC

2: I min

3: I max

4: I rms

Polarity used of the supply of the test item.

For single-phase test item:

0: Automotive

User Manual ATS400 Version 3.14.41 - 1 202

Reference

Tag

Parameter_14

Parameter_15

Parameter_16

bis

Parameter_31

Parameter_32

Parameter_33

Parameter Explanation

1: L1-> PE

2: L2-> PE

3: Mode B

4: Automatic with first error

5: L1 -> PE with first error

6: L2 -> PE with first error

SelftestMode

Polung

With three-phase test item:

0: Clockwise rotation

1: Counterclockwise rotation

Always 0.

Contacting setting of an external relay matrix.

PolungCAN_00

PolungCan_15

bis

StateChangeBits

Setting of the relay matrix in accordance with the choice in the program. The display is bit coded.

Always 0, never used.

Multitest_Enable

Multitest is active, if the value does not equal 0.

Parameter_34

Parameter_35

Parameter_36

Parameter_37

Parameter_38

Parameter_39

Parameter_40

Parameter_41

Parameter_42

Parameter_43

Multitest_Auswert ung

Multitest_Endekri terium

Method of analysing the multitest.

0 = worst measurement value.

Method of ending the multitest.

0 = number of measurements

1 = pass key

Number of multitest inspections.

Multitest_Endekri terium_Anzahl

I_Min

ErrorStartConditi on

ErrorStartConditi onMask

Lower threshold for the discharge current.

Condition for switching forward in the event of an error.

Mask for the condition for switching further in the event of an error.

Always 0, never used.

ErrorStateChangeB its

ErrorWSBDelay

Delay of switching further in the event of an error.

ErrorWSBTimeValid

Delay time active.

StartStateChange

Indicates whether the switching condition needs to be maintained when testing the start conditions.

0 = no waiting for switch condition

1 = Wait for switching condition

User Manual ATS400 Version 3.14.41 - 1 203

Reference

Tag

Parameter_44

Parameter_45

Parameter_46

Parameter_47

Parameter_48

Parameter_49

Parameter_50

Parameter_51

Parameter_52

Parameter_53

Parameter_54

Parameter_55

Parameter

Phase

StartStateJump

Explanation

Number of phases of the test item.

Indicates whether the

Jump

button is displayed.

Repeating

StartButton

PassButton

StartandJump

CreateLog

Timeinterval

Indicates how often the test step is carried out.

Indicates whether the

Start

button is displayed.

Indicates whether the

Pass

button is displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

Indicates whether a log file will be created.

0 = No log file will be created

1 = A log file be created

Time interval between two samples which will be written to the log file. The value is in seconds.

This value will only be used in case the value in

CreateLog

is 1.

StartTime_Delay

Wait time for the signal

Start-Button

, in case the start condition is

monitored

.

StartPETime_Delay

Wait time for the signal

PE-Testprobe

, in case the start condition is

monitored

.

StartKUTime_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition is

monitored

.

StartSHKHVTime_De lay

Wait time for the signal

Safety Circuit

HV

, in case the start condition is

monitored

.

6.4.3.1.7 Sight check

Parameters of the

Sight check

test step.

Tag

Parameter_0

Parameter_1

Parameter Explanation

StartCondition

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartConditionMask

field.

StartConditionMas

Start conditions of the test step. The

User Manual ATS400 Version 3.14.41 - 1 204

Reference

Tag

Parameter_2

Parameter_3

Parameter_4

Parameter_5

Parameter_6

Parameter_7

Parameter_8

Parameter_9

Parameter_10

Parameter_11

Parameter_12

Parameter_13

Parameter k

Explanation

start conditions are saved in a bit-coded manner and can only be analysed together with the

StartCondition

field.

t_delay

StateChangeBits

Abfrage

Start delay between fulfilling the start condition and the start of the test.

Always 0, never used.

Mode for termination.

StartStateChange

Indicates whether the switching condition needs to be maintained when testing the start conditions.

0 = no waiting for switch condition

1 = Wait for switching condition

StartStateJump

Indicates whether the

Jump

button is displayed.

Repeating

StartButton

StartandJump

StartTime_Delay

StartSHKHVTime_De lay

Indicates how often the test step is carried out.

Indicates whether the

Start

button is displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

Wait time for the signal

Start-Button

, in case the start condition is

monitored

.

StartPETime_Delay

Wait time for the signal

PE-Testprobe

, in case the start condition is

monitored

.

StartKUTime_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition is

monitored

.

Wait time for the signal

Safety Circuit

HV

, in case the start condition is

monitored

.

6.4.3.1.8 Data input

Parameters of the

Data Input

test step.

Tag

Parameter_0

Parameter

StartCondition

Explanation

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartConditionMask

field.

User Manual ATS400 Version 3.14.41 - 1 205

Reference

Tag

Parameter_1

Parameter_2

Parameter_3

bis

Parameter_17

Parameter_18

Parameter_19

Parameter_20

Parameter_21

Parameter_22

Parameter_23

Parameter_24

Parameter_25

Parameter_26

Parameter_27

Parameter_28

Parameter_29

Parameter_30

Parameter_31

Parameter_32

Parameter

StartConditionMas k

Explanation

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartCondition

field.

Polung

PolungCAN_00

PolungCan_15

bis

StateChangeBits

Contacting setting of an external relay matrix.

Setting of the relay matrix in accordance with the choice in the program. The display is bit coded.

Always 0, never used.

Multitest_Enable

Multitest is active, if the value does not equal 0.

Multitest_Auswert ung

Multitest_Endekri terium

Method of analysing the multitest.

0 = worst measurement value.

Method of ending the multitest.

0 = number of measurements

1 = pass key

Number of multitest inspections.

Multitest_Endekri terium_Anzahl t_delay

Eingabmasken

StartStateJump

Start delay between fulfilling the start condition and the start of the test.

Bit samples containing screen fields instead of requirements.

Indicates whether the

Jump

button is displayed.

Repeating

StartButton

PassButton

StartandJump

StartTime_Delay

Indicates how often the test step is carried out.

Indicates whether the

Start

button is displayed.

Indicates whether the

Pass

button is displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

Wait time for the signal

Start-Button

, in case the start condition is

monitored

.

StartPETime_Delay

Wait time for the signal

PE-Testprobe

, in case the start condition is

monitored

.

StartTimeKU_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition

User Manual ATS400 Version 3.14.41 - 1 206

Reference

Tag

Parameter_33

Parameter

StartTimeSHKHV_De lay

Explanation

is

monitored

.

Wait time for the signal

Safety Circuit

HV

, in case the start condition is

monitored

.

6.4.3.1.9 Batch run

Parameters of the

Batch run

test step.

Tag

Parameter_0

Parameter_1

Parameter_2

Parameter_3

bis

Parameter_18

Parameter_19

Parameter_20

Parameter_21

Parameter Explanation

StartCondition

StartConditionMas k

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartConditionMask

field.

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartCondition

field.

Polung

PolungCAN_00

PolungCan_15

bis

StateChangeBits

Contacting setting of an external relay matrix.

Setting of the relay matrix in accordance with the choice in the program. The display is bit coded.

Always 0, never used.

SettingFlags

Settings that indicate that certain fields are configured at other locations. This field is bit coded.

Multitest_Enable

Multitest is active, if the value does not equal 0.

Parameter_22

Parameter_23

Parameter_24

Parameter_23

Parameter_26

Multitest_Auswert ung

Multitest_Endekri terium

Method of analysing the multitest.

0 = worst measurement value.

Method of ending the multitest.

0 = number of measurements

1 = pass key

Number of multitest inspections.

Multitest_Endekri terium_Anzahl t_delay

Eingabmasken

Start delay between fulfilling the start condition and the start of the test.

Bit samples containing screen fields instead of requirements.

User Manual ATS400 Version 3.14.41 - 1 207

Reference

Tag

Parameter_27

Parameter_28

Parameter_29

Parameter_30

Parameter_31

Parameter_32

Parameter_33

Parameter_34

Parameter_35

Parameter

StartStateJump

Explanation

Indicates whether the

Jump

button is displayed.

Repeating

StartButton

PassButton

StartandJump

StartTime_Delay

StartSHKHVTime_De lay

Indicates how often the test step is carried out.

Indicates whether the

Start

button is displayed.

Indicates whether the

Pass

button is displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

Wait time for the signal

Start-Button

, in case the start condition is

monitored

.

StartPETime_Delay

Wait time for the signal

PE-Testprobe

, in case the start condition is

monitored

.

StartKUTime_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition is

monitored

.

Wait time for the signal

Safety Circuit

HV

, in case the start condition is

monitored

.

6.4.3.1.10 User-Interface

Parameters of the

User-Interface

test step.

Tag

Parameter_0

Parameter_1

Parameter_2

Parameter_3

Parameter_4

Parameter_5

Parameter_6

Parameter

StartCondition

StartConditionMas k

In

InMask

Out

OutMask

Duration

Explanation

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartConditionMask

field.

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartCondition

field.

Entry bits

Mask for entry bits.

Output bits

Mask for output bits.

Duration of the output pulse.

User Manual ATS400 Version 3.14.41 - 1 208

Reference

Tag

Parameter_7

Parameter_8

Parameter_9

Parameter_10

Parameter_11

bis

Parameter_26

Parameter_27

Parameter_28

Parameter

Timeout

Explanation

Timeout when waiting for the status of the entry bits.

This parameter is only valid if the

TimeoutEnable

parameter is set.

TimeoutEnable t_delay

Polung

States whether timeout is used.

0: Timeout is not used

1: Timeout is used

Start delay between fulfilling the start condition and the start of the test.

Contacting setting of an external relay matrix.

PolungCAN_00

PolungCan_15

bis

StateChangeBits

Setting of the relay matrix in accordance with the choice in the program. The display is bit coded.

Always 0, never used.

Multitest_Enable

Multitest is active, if the value does not equal 0.

Parameter_29

Parameter_30

Parameter_31

Parameter_32

Parameter_33

Parameter_34

Parameter_35

Parameter_36

Parameter_37

Parameter_38

Multitest_Auswert ung

Multitest_Endekri terium

Method of analysing the multitest.

0 = worst measurement value.

Method of ending the multitest.

0 = number of measurements

1 = pass key

Number of multitest inspections.

Multitest_Endekri terium_Anzahl

ErrorStartConditi on

ErrorStartConditi onMask

Condition for switching forward in the event of an error.

Mask for the condition for switching further in the event of an error.

Always 0, never used.

ErrorStateChangeB its

ErrorWSBDelay

Delay of switching further in the event of an error.

ErrorWSBTimeValid

Delay time active.

NumInterface

Number of the user IO interface used.

0: User-IO Interface of the internal IO-

CPU

1: User-IO Interface of the external IO-

CPU

StartStateChange

Indicates whether the switching condition needs to be maintained when testing the start conditions.

0 = no waiting for switch condition

User Manual ATS400 Version 3.14.41 - 1 209

Reference

Tag

Parameter_39

Parameter_40

Parameter_41

Parameter_42

Parameter_43

Parameter_44

Parameter_45

Parameter_46

Parameter_47

Parameter

StartStateJump

Explanation

1 = Wait for switching condition

Indicates whether the

Jump

button is displayed.

Repeating

StartButton

PassButton

StartandJump

StartTime_Delay

StartSHKHVTime_De lay

Indicates how often the test step is carried out.

Indicates whether the

Start

button is displayed.

Indicates whether the

Pass

button is displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

Wait time for the signal

Start-Button

, in case the start condition is

monitored

.

StartPETme_Delay

Wait time for the signal

PE-Testprobe

, in case the start condition is

monitored

.

StartKUTime_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition is

monitored

.

Wait time for the signal

Safety Circuit

HV

, in case the start condition is

monitored

.

6.4.3.1.11 Resistance

Parameters of the

Resistance

test step.

Tag

Parameter_0

Parameter_1

Parameter_2

Parameter_3

Parameter

StartCondition

StartConditionMas k t t_start

Explanation

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartConditionMask

field.

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartCondition

field.

Test time

Delay time for the start scenario when starting the test until starting the analysis. This value is not valid for all start scenarios.

User Manual ATS400 Version 3.14.41 - 1 210

Reference

Tag

Parameter_4

Parameter_5

Parameter_6

Parameter_7

Parameter_8

bis

Parameter_23

Parameter_24

Parameter_25

Parameter_26

Parameter_27

Parameter_28

Parameter_29

Parameter_30

Parameter_31

Parameter_32

Parameter_33

Parameter_34

Parameter_35

Parameter_36

Parameter_37

Parameter Explanation t_delay

R_min

R_max

Start delay between fulfilling the start condition and the start of the test.

Lower threshold value of the resistance.

Upper threshold value of the resistance.

Polung

PolungCAN_00

PolungCan_15

bis

StateChangeBits

Contacting setting of an external relay matrix.

Setting of the relay matrix in accordance with the choice in the program. The display is bit coded.

Always 0, never used.

Multitest_Enable

Multitest is active, if the value does not equal 0.

Multitest_Auswert ung

Multitest_Endekri terium

Method of analysing the multitest.

0 = worst measurement value.

Method of ending the multitest.

0 = number of measurements

1 = pass key

Number of multitest inspections.

Multitest_Endekri terium_Anzahl

ErrorStartConditi on

ErrorStartConditi onMask

Condition for switching forward in the event of an error.

Mask for the condition for switching further in the event of an error.

Always 0, never used.

ErrorStateChangeB its

ErrorWSBDelay

Delay of switching further in the event of an error.

ErrorWSBTimeValid

Delay time active.

R_offset

Resistance of the measurement structure.

StartStateChange

Indicates whether the switching condition needs to be maintained when testing the start conditions.

0 = no waiting for switch condition

1 = Wait for switching condition

StartStateJump

Indicates whether the

Jump

button is displayed.

Repeating

Indicates how often the test step is carried out.

User Manual ATS400 Version 3.14.41 - 1 211

Reference

Tag

Parameter_38

Parameter_39

Parameter_40

Parameter_41

Parameter_42

Parameter_43

Parameter_44

Parameter_45

Parameter

StartButton

PassButton t_timeout

StartandJump

StartTime_Delay

Explanation

Indicates whether the

Start

button is displayed.

Indicates whether the

Pass

button is displayed.

Maximum time after starting until the first measurement value is recorded.

Indicates whether the

Start

and

Jump

buttons are displayed.

Wait time for the signal

Start-Button

, in case the start condition is

monitored

.

StartPETime_Delay

Wait time for the signal

PE-Testprobe

, in case the start condition is

monitored

.

StartKUTime_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition is

monitored

.

StartSHKHVTime_De lay

Wait time for the signal

Safety Circuit

HV

, in case the start condition is

monitored

.

6.4.3.1.12 Continuity test

Parameter of the

Continuity

test step.

Tag

Parameter_0

Parameter_1

Parameter_2

Parameter_3

Parameter_4

Parameter_5

Parameter_6

Parameter_7

Parameter

StartCondition

StartConditionMas k

TestVoltage

TestTime

MaxTestCurrent t_delay

Limit

Interpretation

Explanation

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartConditionMask

field.

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartCondition

field.

Maximum test voltage used for the measurement.

Time after that the measurement will be evaluated.

Maximum current during the test.

Start delay between fulfilling the start condition and the start of the test.

Limit for the evaluation.

Evaluation of the measurement:

User Manual ATS400 Version 3.14.41 - 1 212

Reference

Tag

Parameter_8

Parameter_9

Parameter_10

bis

Parameter_25

Parameter_26

Parameter_27

Parameter Explanation

TypeOfSource

Polung

PolungCAN_00

PolungCan_15

bis

StateChangeBits

1: Pass on continuity

2: Pass on discontinuity

Type of the used source:

0: Alternating current

1: direct current

Contacting setting of an external relay matrix.

Setting of the relay matrix in accordance with the choice in the program. The display is bit coded.

Always 0, never used.

Multitest_Enable

Multitest is active, if the value does not equal 0.

Parameter_28

Parameter_29

Parameter_30

Parameter_31

Parameter_32

Parameter_33

Parameter_34

Parameter_35

Parameter_36

Parameter_37

Parameter_38

Parameter_39

Multitest_Auswert ung

Multitest_Endekri terium

Method of analysing the multitest.

0 = worst measurement value.

Method of ending the multitest.

0 = number of measurements

1 = pass key

Number of multitest inspections.

Multitest_Endekri terium_Anzahl

ErrorStartConditi on

ErrorStartConditi onMask

Condition for switching forward in the event of an error.

Mask for the condition for switching further in the event of an error.

Always 0, never used.

ErrorStateChangeB its

ErrorWSBDelay

Delay of switching further in the event of an error.

ErrorWSBTimeValid

Delay time active.

StartStateChange

Indicates whether the switching condition needs to be maintained when testing the start conditions.

0 = no waiting for switch condition

1 = Wait for switching condition

StartStateJump

Indicates whether the

Jump

button is displayed.

Repeating

StartButton

Indicates how often the test step is carried out.

Indicates whether the

Start

button is displayed.

User Manual ATS400 Version 3.14.41 - 1 213

Reference

Tag

Parameter_40

Parameter_41

Parameter_42

Parameter_43

Parameter_44

Parameter_45

Parameter

PassButton

StartandJump

StartTime_Delay

Explanation

Indicates whether the

Pass

button is displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

Wait time for the signal

Start-Button

, in case the start condition is

monitored

.

StartPETime_Delay

Wait time for the signal

PE-Testprobe

, in case the start condition is

monitored

.

StartKUTime_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition is

monitored

.

StartSHKHVTime_De lay

Wait time for the signal

Safety Circuit

HV

, in case the start condition is

monitored

.

6.4.3.1.13 PT 100

Parameters of the

PT 100

test step.

Tag

Parameter_0

Parameter_1

Parameter_2

Parameter_3

Parameter_4

Parameter_5

Parameter_6

Parameter_7

Parameter_8

bis

Parameter_23

Parameter

StartCondition

StartConditionMas k t t_start t_delay

T_min

T_max

Polung

PolungCAN_00

PolungCan_15

bis

Explanation

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartConditionMask

field.

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartCondition

field.

Test time

Delay time for the start scenario when starting the test until starting the analysis. This value is not valid for all start scenarios.

Start delay between fulfilling the start condition and the start of the test.

Lower threshold for the temperature.

Upper threshold for the temperature.

Contacting setting of an external relay matrix.

Setting of the relay matrix in accordance with the choice in the program. The

User Manual ATS400 Version 3.14.41 - 1 214

Reference

Tag

Parameter_24

Parameter_25

Parameter_26

Parameter_27

Parameter_28

Parameter_29

Parameter_30

Parameter_31

Parameter_32

Parameter_33

Parameter_34

Parameter_35

Parameter_36

Parameter_37

Parameter_38

Parameter_39

Parameter_40

Parameter_41

Parameter Explanation

StateChangeBits

display is bit coded.

Always 0, never used.

Multitest_Enable

Multitest is active, if the value does not equal 0.

Multitest_Auswert ung

Multitest_Endekri terium

Method of analysing the multitest.

0 = worst measurement value.

Method of ending the multitest.

0 = number of measurements

1 = pass key

Number of multitest inspections.

Multitest_Endekri terium_Anzahl

ErrorStartConditi on

ErrorStartConditi onMask

Condition for switching forward in the event of an error.

Mask for the condition for switching further in the event of an error.

Always 0, never used.

ErrorStateChangeB its

ErrorWSBDelay

Delay of switching further in the event of an error.

ErrorWSBTimeValid

Delay time active.

R_offset

Resistance of the measurement structure.

StartStateChange

Indicates whether the switching condition needs to be maintained when testing the start conditions.

0 = no waiting for switch condition

1 = Wait for switching condition

StartStateJump

Indicates whether the

Jump

button is displayed.

R0

Repeating

StartButton

PassButton

StartandJump

Basic resistance of the measurement sensor.

Indicates how often the test step is carried out.

Indicates whether the

Start

button is displayed.

Indicates whether the

Pass

button is displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

User Manual ATS400 Version 3.14.41 - 1 215

Reference

Tag

Parameter_42

Parameter_43

Parameter_44

Parameter_45

Parameter

StartTime_Delay

Explanation

Wait time for the signal

Start-Button

, in case the start condition is

monitored

.

StartPETime_Delay

Wait time for the signal

PE-Testprobe

, in case the start condition is

monitored

.

StartKUTime_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition is

monitored

.

StartSHKHVTime_De lay

Wait time for the signal

Safety Circuit

HV

, in case the start condition is

monitored

.

6.4.3.1.14 Light control

Parameters of the

Light control

test step.

Tag

Parameter_0

Parameter_1

Parameter_2

Parameter_3

Parameter_4

Parameter_5

Parameter_6

Parameter_7

Parameter_8

Parameter_9

Parameter_10

Parameter_11

Parameter_12

Parameter_13

Parameter

StartCondition

StartConditionMas k t_delay

StateChangeBits

Leuchte_ART

Leuchte_Befehl

Leuchte_DIM

Leuchte_DALIC

Leuchte_DALIV

Leuchte_DIMT

ErrorStartConditi on

ErrorStartConditi onMask

ErrorStateChangeB its

ErrorWSBDelay

Explanation

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartConditionMask

field.

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartCondition

field.

Start delay between fulfilling the start condition and the start of the test.

Always 0, never used.

Type of the light

Command to the light

Dim value

DALIC

DALIV

DIMT

Condition for switching forward in the event of an error.

Mask for the condition for switching further in the event of an error.

Always 0, never used.

Delay of switching further in the event of

User Manual ATS400 Version 3.14.41 - 1 216

Reference

Tag

Parameter_14

Parameter_15

Parameter_16

Parameter_17

Parameter_18

Parameter_19

Parameter_20

Parameter_21

Parameter_22

Parameter_23

Parameter Explanation

an error.

ErrorWSBTimeValid

Delay time active.

StartStateChange

Indicates whether the switching condition needs to be maintained when testing the start conditions.

0 = no waiting for switch condition

1 = Wait for switching condition

StartStateJump

Indicates whether the

Jump

button is displayed.

Repeating

StartButton

StartandJump

StartTime_Delay

StartSHKHVTime_De lay

Indicates how often the test step is carried out.

Indicates whether the

Start

button is displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

Wait time for the signal

Start-Button

, in case the start condition is

monitored

.

StartPETime_Delay

Wait time for the signal

PE-Testprobe

, in case the start condition is

monitored

.

StartKUTime_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition is

monitored

.

Wait time for the signal

Safety Circuit

HV

, in case the start condition is

monitored

.

6.4.3.1.15 Dummy load

Parameters of the

Dummy load

test step.

Tag

Parameter_0

Parameter_1

Parameter_2

Parameter

StartCondition

StartConditionMas k

Management

Explanation

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartConditionMask

field.

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartCondition

field.

Indicates how the supply of the test item is to be carried out.

User Manual ATS400 Version 3.14.41 - 1 217

Reference

Parameter_3

Parameter_4

Parameter_5

Parameter_6

Parameter_7

Parameter_8

Parameter_9

Parameter_10

Parameter_11

Parameter_12

Parameter_13

Parameter_14

Parameter_15

Parameter_16

Parameter_17

Parameter_18

Parameter_19

Parameter_20

Parameter_21

Parameter_22

Parameter_23

Tag

Parameter_24

bis

Parameter_39

Parameter_40

Parameter_41

Parameter Explanation

U_Source

U f

Channels

RGas

RCoil

Filament_enable

Filament_t

Filament_delay

Filament_I

Ignition_enable

Ignition_t

Igniton_delay

Ignition_U

0: Power off after test

1: Keep power on after test

2: Only power off

3: Power off only on error

4: Only power on

Source of the test supply.

Voltage of the test item supply

Frequency of the voltage

Number of channels

Gas resistance

Coil resistance

Active flame

Active flame

Flame delay

Current of the flame

Ignition available

Ignition time

Ignition delay

Ignition voltage

Fct_Enable

Fct_t

Fct_delay

Fct_Imin

Function available

Duration of the function

Delay of the analysis

Lower threshold value of the current

Fct_Imax t_timeout

Polung

PolungCAN_00

PolungCan_15

bis

StateChangeBits

Upper threshold value of the current

Always 0

Contacting setting of an external relay matrix.

Setting of the relay matrix in accordance with the choice in the program. The display is bit coded.

Always 0, never used.

Multitest_Enable

Multitest is active, if the value does not equal 0.

Parameter_42

Parameter_43

Multitest_Auswert ung

Multitest_Endekri terium

Method of analysing the multitest.

0 = worst measurement value.

Method of ending the multitest.

0 = number of measurements

User Manual ATS400 Version 3.14.41 - 1 218

Reference

Tag

Parameter_44

Parameter_45

Parameter_46

Parameter_47

Parameter_48

Parameter_49

Parameter_50

Parameter_51

Parameter_52

Parameter_53

Parameter_54

Parameter_55

Parameter_56

Parameter_57

Parameter_58

Parameter_59

Parameter Explanation

1 = pass key

Number of multitest inspections.

Multitest_Endekri terium_Anzahl

ErrorStartConditi on

ErrorStartConditi onMask

Condition for switching forward in the event of an error.

Mask for the condition for switching further in the event of an error.

Always 0, never used.

ErrorStateChangeB its

ErrorWSBDelay

StartStateJump

Delay of switching further in the event of an error.

ErrorWSBTimeValid

Delay time active.

Filament_Imax

Max. current of the flame

Indicates whether the

Jump

button is displayed.

Repeating

StartButton

PassButton

StartandJump

StartTime_Delay

StartSHKHVTime_De lay

Indicates how often the test step is carried out.

Indicates whether the

Start

button is displayed.

Indicates whether the

Pass

button is displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

Wait time for the signal

Start-Button

, in case the start condition is

monitored

.

StartPETime_Delay

Wait time for the signal

PE-Testprobe

, in case the start condition is

monitored

.

StartKUTime_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition is

monitored

.

Wait time for the signal

Safety Circuit

HV

, in case the start condition is

monitored

.

6.4.3.1.16 Light filament current

Parameters of the

Light filament current

test step.

Tag

Parameter_0

Parameter

StartCondition

Explanation

Start conditions of the test step. The

User Manual ATS400 Version 3.14.41 - 1 219

Reference

Tag

Parameter_1

Parameter_2

Parameter_3

Parameter_4

Parameter_5

Parameter_6

Parameter_7

Parameter_8

bis

Parameter_23

Parameter_24

Parameter_25

Parameter_26

Parameter_27

Parameter_28

Parameter_29

Parameter_30

Parameter_31

Parameter_32

Parameter_33

Parameter

StartConditionMas k

Explanation

start conditions are saved in a bit-coded manner and can only be analysed together with the

StartConditionMask

field.

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartCondition

field.

Management

Filament_enable

Filament_t

Indicates how to proceed after the test with supplying the test item.

0: Deactivate

1: Activate

Active flame

Active flame

Filament_delay

Filament_I

Polung

PolungCAN_00

PolungCan_15

bis

StateChangeBits

Flame delay

Current of the flame

Contacting setting of an external relay matrix.

Setting of the relay matrix in accordance with the choice in the program. The display is bit coded.

Always 0, never used.

Multitest_Enable

Multitest is active, if the value does not equal 0.

Multitest_Auswert ung

Multitest_Endekri terium

Method of analysing the multitest.

0 = worst measurement value.

Method of ending the multitest.

0 = number of measurements

1 = pass key

Number of multitest inspections.

Multitest_Endekri terium_Anzahl

ErrorStartConditi on

ErrorStartConditi onMask

Condition for switching forward in the event of an error.

Mask for the condition for switching further in the event of an error.

Always 0, never used.

ErrorStateChangeB its

ErrorWSBDelay

Delay of switching further in the event of an error.

ErrorWSBTimeValid

Delay time active.

User Manual ATS400 Version 3.14.41 - 1 220

Reference

Tag

Parameter_34

Parameter_35

Parameter_36

Parameter_37

Parameter_38

Parameter_39

Parameter_40

Parameter_41

Parameter_42

Parameter_43

Parameter_44

Parameter Explanation

Filament_Imax

Max. current of the flame

StartStateChange

Indicates whether the switching condition needs to be maintained when testing the start conditions.

0 = no waiting for switch condition

1 = Wait for switching condition

StartStateJump

Indicates whether the

Jump

button is displayed.

Repeating

StartButton

PassButton

StartandJump

StartTime_Delay

StartSHKHVTime_De lay

Indicates how often the test step is carried out.

Indicates whether the

Start

button is displayed.

Indicates whether the

Pass

button is displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

Wait time for the signal

Start-Button

, in case the start condition is

monitored

.

StartPETime_Delay

Wait time for the signal

PE-Testprobe

, in case the start condition is

monitored

.

StartKUTime_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition is

monitored

.

Wait time for the signal

Safety Circuit

HV

, in case the start condition is

monitored

.

6.4.3.1.17 External program

Parameters of the

External program

test step.

Tag

Parameter_0

Parameter_1

Parameter_3

Parameter

StartCondition

StartConditionMas k

Polung

Explanation

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartConditionMask

field.

Start conditions of the test step. The start conditions are saved in a bit-coded manner and can only be analysed together with the

StartCondition

field.

Contacting setting of an external relay

User Manual ATS400 Version 3.14.41 - 1 221

Reference

Tag

Parameter_4

bis

Parameter_19

Parameter_20

Parameter_21

Parameter_22

Parameter_23

Parameter_24

Parameter_25

Parameter_26

Parameter_27

Parameter_28

Parameter_29

Parameter_30

Parameter_31

Parameter_32

Parameter_33

Parameter_34

Parameter_35

Parameter_36

Parameter_37

Parameter Explanation

PolungCAN_00

PolungCan_15

bis

StateChangeBits

matrix.

Setting of the relay matrix in accordance with the choice in the program. The display is bit coded.

Always 0, never used.

Multitest_Enable

Multitest is active, if the value does not equal 0.

Multitest_Auswert ung

Multitest_Endekri terium

Method of analysing the multitest.

0 = worst measurement value.

Method of ending the multitest.

0 = number of measurements

1 = pass key

Number of multitest inspections.

Multitest_Endekri terium_Anzahl

SettingsFlags t_delay

ErrorStartConditi on

ErrorStartConditi onMask

Settings

Start delay between fulfilling the start condition and the start of the test.

Condition for switching forward in the event of an error.

Mask for the condition for switching further in the event of an error.

Always 0, never used.

ErrorStateChangeB its

ErrorWSBDelay

Delay of switching further in the event of an error.

ErrorWSBTimeValid

Delay time active.

StartStateJump

Indicates whether the

Jump

button is displayed.

Repeating

StartButton

PassButton

StartandJump

StartTime_Delay

Indicates how often the test step is carried out.

Indicates whether the

Start

button is displayed.

Indicates whether the

Pass

button is displayed.

Indicates whether the

Start

and

Jump

buttons are displayed.

Wait time for the signal

Start-Button

, in case the start condition is

monitored

.

User Manual ATS400 Version 3.14.41 - 1 222

Reference

Tag

Parameter_38

Parameter_39

Parameter_40

Parameter Explanation

StartPETime_Delay

Wait time for the signal

PE-Testprobe

, in case the start condition is

monitored

.

StartKUTime_Delay

Wait time for the signal

Contact

Monitoring

, in case the start condition is

monitored

.

StartSHKHVTime_De lay

Wait time for the signal

Safety Circuit

HV

, in case the start condition is

monitored

.

6.4.4

Result

The data in this block can be found under

<ResultData>/<Result>

.

Tag

Count_Tests

Count_IO

Count_NIO

Count_ERROR

PlanCycleTime

PlanStandbyTime

Result

Type

UsedID

SerialNumber

Article

Explanation

Number of tests that were carried out since starting the test plan.

Number of IO results of the test since starting the test plan.

Number of NIO results of the test since starting the test plan.

Number of errors of the test since starting the test plan.

Duration for carrying out the test. The value is always

0.

Duration between the end of the prior test and the start of the current test.

Total result of the plan.

Passed

-> The test item has passed the test.

Failed

-> the test item has not passed the test or the test was cancelled.

Type of the test item, is only filled in if the combibox is active.

ID used to select the test plan, is only filled in when the

ID-Pattern

field is active.

Serial number of the test item, is only filled in if the

Serial-Pattern

field is active.

Article name, is only filled in if the

Article-Pattern

field is active.

User Manual ATS400 Version 3.14.41 - 1 223

Reference

6.4.5

PlanOptionen

The data in this block can be found under

<ResultData>/<PlanOptionen>

.

This block contains the subblock for the

ReportOptionen and subsequent entries.

Tag

Start_Timeout

Explanation

If the vlaue is not 0 it will be checked that all other start conditions are meet after the start condition start signal is set.

6.4.5.1 ReportOptionen

The data in this block can be found under

<ResultData>/<PlanOptionen>/

<ReportOptionen>

.

This block contains a list of the individual report options. Every report option is a

block

of its own.

6.4.5.1.1 ReportOption

The data in this block can be found under

<ResultData>/<PlanOptionen>/

<ReportOptionen>/<ReportOption>

.

This block contains the subblock for the

output data

and subsequent entries.

Tag

PrintFlags

Explanation

Indicates for what overall result the report is to be created. The indication is bit-coded.

6.4.5.1.1.1 SpoolData

The data in this block can be found under

<ResultData>/<PlanOptionen>/

<ReportOptionen>/<ReportOption>/<SpoolData>

.

Tag

Template

Output

ConversionType

Explanation

File name of the template.

If the file can be reached via a relative path from the folder of the plan files, the relative path is entered, otherwise the absolute path is entered.

Indication where the report is to be sent. This indication is only filled in if the report is to be issued on a printer.

Indicates what report is to be created.

printer

-> printout on a printer using an HTML template

html

-> creation of an HTML file using an HTML template

User Manual ATS400 Version 3.14.41 - 1 224

Reference

Tag Explanation pdf

-> creation of a PDF file using an HTML template

xml

-> creation of an XML file

csv

-> creation of a CSV file using a CSV template

zebra

-> printout on a zebra printer using a ZPL template

6.4.6

Validation

The data in this block can be found under

<ResultData>/<Validation>

.

Tag

Checksum

Explanation

Not used, always 42.

6.5

6.6

Plan files

The plan files have the extension

.plan

. The place of saving and the file name result from the details in

Settings

->

File storage

->

Test plan

. This file has the format of an XML file with UTF-8 coding with BOM.

The plan files are largely identical to the result files. They differ by the encompassing

PlanData

tag instead of

ResultData

. Furthermore, the result fields in the individual test steps and the

Result

block do not apply.

Printserver configuration

This file has the name

PrintServer.conf

and resides in the working folder. This file must not be present and will not be created automatically.

In this file are settings which are used only by the printserver. These settings are for the storage of the report files, handling missing keywords and timeouts. It is possible to configure an alternative folder for each report type.

The file is in XML-format and the main tag is

Settings

.

Tag

HTML

PDF

XML

CSV

Debug

Explanation

Alternative folder for the report type

html

.

Alternative folder for the report type

pdf

.

Alternative folder for the report type

xml

.

Alternative folder for the report type

csv

.

false

: Will a keyword not be recognized no text be generated.

true

: Will a keyword not be recognized the keyword in parantheseis will be generated.

Default setting is

false

.

User Manual ATS400 Version 3.14.41 - 1 225

Reference

Tag

PdfTimeout

PrintTimeout

Explanation

Time within the PDF-document must be generated.

The value is in milliseconds.

Default is 6 seconds.

time within the printing on the printer must be completed. The value is in milliseconds.

Default is 10 seconds.

6.7

Matrix

6.7.1

RelaisMatrix.cfg

This file will be used for all relais matrix types und describes which relais will be switched for the different outputs and wires. For the types 3 and 4 also the file for the

additional relais is neccessary.

Description: Relais Matrix, 8Channel, 1xRelais_Platine, 1xLT_CPU

MatrixTyp: 1

MatrixNum: 1

Channels: 2

-----------------------------

Channel 1

A: 1, 2,

B: 1,

0:

-----------------------------

Channel 2

A: 3, 4,

B: 3,

0:

-----------------------------

Tag

Description

MatrixTyp

MatrixNum

Channels

Description

Comment which will not be proccessed.

Determines the different modules.

1

: Matrix with max. 24 relais for each board

2

: Matrix with max. 8 relais for each board

3

: Matrix with max. 16 relais for each board

4

: Matrix with max. 32 relais for each board

Number of boards in the matrix.

Number of channels in the matrix.

Between the header and the first channel resp. between the channels is a separator line. After the last channel the seperator line must be present.

Each channel consists of 4 Lines. There must be configured as much lines as set in

Channels

.

User Manual ATS400 Version 3.14.41 - 1 226

Reference

Tag

A:

B:

0:

Description

Comment which will not be proccessed.

These relais will be switched in the case wire HV1 will connected to the apropiate output.

These relais will be switched in the case wire HV2 will connected to the apropiate output.

These relais will be switched in the case none of the wires will be connected to an output.

6.7.2

ExtendedRelais.conf

This file is neccessary when a matrix of type

MatrixTyp

3 or 4 will be used.

This file is in XML-Format. For each test type as much lines must be present as boards are used according to the value

MatrixNum

in the file

RelaisMatrix.cfg

. The values in

the tag

unsignedInt

have a range from 0 ... 3.

1

2

Value

0

3

Description

No additional relais switched.

Additional relais K1 switched.

Additional relais K2 switched.

Both additional relais switched.

User Manual ATS400 Version 3.14.41 - 1 227

Reference

<?xml version="1.0" encoding="utf-8"?>

<

ExtendedRelais

xmlns:xsi ="

http://www.w3.org/2001/XMLSchema-instance

" xmlns:xsd ="

http://www.w3.org/2001/XMLSchema

">

<!--

Für jede Prüfart müssen so viele Zeilen vorhanden sein, wie Platinen verbaut sind = MatrixNum aus RelaisMatrix.cfg

0 = Kein Relais geschaltet, 1 = K1 geschaltet, 2 = K2 geschaltet, 3 = beide geschaltet !

-->

<

HV_AC

>

<

unsignedInt

>

0

</

unsignedInt

>

</

HV_AC

>

<

HV_DC

>

<

unsignedInt

>

0

</

unsignedInt

>

</

HV_DC

>

<

ISO

>

<

unsignedInt

>

0

</

unsignedInt

>

</

ISO

>

<

PE

>

<

unsignedInt

>

0

</

unsignedInt

>

</

PE

>

<

FCT_Current

>

<

unsignedInt

>

0

</

unsignedInt

>

</

FCT_Current

>

<

Resistance

>

<

unsignedInt

>

0

</

unsignedInt

>

</

Resistance

>

<

LeakageCurrent

>

<

unsignedInt

>

0

</

unsignedInt

>

</

LeakageCurrent

>

<

DummyLoad

>

<

unsignedInt

>

0

</

unsignedInt

>

</

DummyLoad

>

<

FCT_Extern

>

<

unsignedInt

>

0

</

unsignedInt

>

</

FCT_Extern

>

<

Notlicht

>

<

unsignedInt

>

0

</

unsignedInt

>

</

Notlicht

>

<

PT100

>

<

unsignedInt

>

0

</

unsignedInt

>

</

PT100

>

<

Matrix

>

<

unsignedInt

>

0

</

unsignedInt

>

</

Matrix

>

</

ExtendedRelais

>

User Manual ATS400 Version 3.14.41 - 1 228

Lembergstr. 23

70825 Korntal

Phone: +49 711 83 99 39-0

Fax: +49 711 83 99 39-9

Internet: www.etl-prueftechnik.de

E-Mail: [email protected]