HumaClot Pro

Specifications & more
HumaClot Pro
>> System Specifications
>> Scope of Supply
>> Obligatory Items
>> Optional Items
System OverviewCat.No.15800
Analyzer type
Fully automated coagulation analyzer
Class
Bench top / Desktop
Throughput
Up to 100 PT / h
Up to 58 tests /h with mixed parameters (PT, aPTT, FIB)
Reagent systemClosed1 (validated HUMAN application settings)2
# of parameters
50 freely definable applications3
Consumable system
Closed for cuvette rings
Operation mode
Batch mode (sample oriented, test selective)
Random-access (STAT)
Languages softwareMultilingual4
Languages manualEnglish
Operation software
1
HumaClot Pro GUI
Open at launch, but system will be closed mandatorily in the future.
Hemostat Prothrombin-SI, Prothrombin Liquid, aPTT, TT & FIB preinstalled. D-Dimer will follow after launch
3
After closing the system for non-Hemostat reagents, the number of freely definable channels will be reduced
4
At launch: English and German, additional languages (RU, FR, SP, PO) will follow in due course
2
HumaClot Pro
Fully Automated Coagulation Instrument
Tests
Measure typePhoto-optical
Analysis modes
Parameters
Coagulation, Chromogenic, Immunologic
Prothrombin Time (PT)
Activated Partial Thromboplastin Time (aPTT)
Thrombin Time (TT)
Fibrinogen (FIB)
D-Dimer (DD)
Protein C (PC)*
Antithrombin (AT3)*
Factors (FII – FXII)*
Venom Time (VT)* or Batroxobin Time (BT)*
Heparin (HEP)*
Activated PC Resistance (APCR)*
Lupus Anticoagulant (Screen, Confirm)*
Protein S (PS)*
von Willebrand Factor (VWF)*
Ecarin Chromogenic Assay Thrombin (ECAT)*
Hirudin (ECAH) *
Plasminogen (PLG) *
α 2-Antiplasmin (A2AP) *
PT/FIB*
Prothrombinase induced Clotting Time (PiCT)*
Thrombin Dynamics Test (TDT)*
Thrombin Activatable Fibrinolysis Inhibitor (TAFI)*
Calibration
Up to 7 Point, ISI , normal PT
Self check
On start-up
Result display
Clotting - Sec, % of Normal, INR
Kinetic – Sec, mE, Unit
Samples / Controls /
Calibrators
Supported sample vessels
Primary tubes: D11-13 x 65-100 mm
Secondary tubes: Human 4 ml, Hitachi cup,
Sample positions
25 (22 + 3 STAT)
STAT capability
3 positions
Supp. calibration/control vessels
Human 4 ml, Hitachi cup
Calibration/control positions8
Sample identification
Sample typePlasma
Auto-repeat
Manual internal barcode reader
Yes (double measure time)
Auto-calibration / -dilutionYes
Calibration points
*
Reagent currently not available from HUMAN
Up to 7
Reagent
Reagent positions15
Mixing
Yes, 3 positions
3 x 15 ml ø 31.2 mm
Supp. reagent vessels
12 x 5 ml stirred ø 22.5 mm
Wash & cleaner positions
2
Supp. wash / cleaner vessels
2 x 15 ml ø 31.2 mm
Internal barcode reader
Sample ID
Barcode Reader Types
Code 39, codabar min. 6 characters,
2 of 5 interleave min 6 characters
UPCA -> EAN13
UPAC -> UPAC
Reaction unit / Incubation
Reaction system Cuvette ring
Capacity
32 cuvettes / ring
Mixing during measurementYes
Incubation temperature 37.4°C
Accuracy
± 0,4°C
Uniformity
± 0,1°C
Min. reaction volume (test vol.)
150 µl
Max. reaction volume (test vol.)
300 µl
Liquid Pipetting
Liquid transportation
X, Y, Z- pipetting arm
Liquid level detection/ level sense Yes for reagents & samples (capacity)
Needle crash detection
Yes, XYZ Arm
Needle type
Steel needle
Needle heatingYes
Min. pipetting volume
2 µl
Max. pipetting volume
250 µl
Pipetting resolution
1 µl
HumaClot Pro
Fully Automated Coagulation Instrument
WashingCapacity
2.5 l DI water tank / 2.5 l waste tank
Level sensor
DI water empty / Waste full
DI water filterYes
Probe wash stationYes
Water consumption min.
Wash profiles5
Reading
Optical systemPhotometric
Number of wavelengths3
Wavelengths < 6 ml
405 nm, 570 nm, 740 nm
Light sourceLED
Optical range
0 - 2 abs.
Optical sensitivity
0.001 abs.
Number of channels2
Reference channels2
Data Processing Interfaces 1 x RS232C (Mini DIN), 1 x RS232C (9 pin), 1x Chipcard , 1xLAN, 2x USB,
1x USB Device, 1x SD Card HC (Internal)
Supp. printer External USB PCL3 e.g
HP Deskjet D5560 / HP Officejet 6000 / HP Deskjet 6940
HP Deskjet 6930 / HP Officejet Pro 8500 / HP Officejet Pro 8100
Compatible USB storage
FAT, up to 32GB
LIS/LIMS
Yes, by LAN bidirectional (FTP auto-send, auto-request)
User levels
3 (anyone, expert, service)
Real-time reaction curve displayYes
Calibration curve displayYes
User interface
Integrated colour touch screen 17 x 12.5 cm, resolution 800 x 600
Result memory Up to 10.000 results
Result flags
OK, none clotting, deviation error, temperature, mixer sample,
mixer reagent photometer, liquid, QC
Number of applications50
Quality control module Yes, with Levey Jennings
QC rules Westgard rules (1 _3S, 2_2S, R_4S, 4_1S, 8x, 1_2S)
Max. # of QC 8
# of QC‘s / parameter3
Update parameter by USB
Export integrated service softwareYes
Yes export
GeneralDimension(WxDxH) Instrument without any components: 83 x 46 x 61 cm
Space requirements routine use: 100 x 60 x 80 cm
Packaging:93 x 71 x 60 cm
Weight:Gross: 45 kg, Net: 38 kg
Electrical requirements
100-240 VAC ± 10 %, 50/60 hz, IEC-60320C13 connector
Power consumption:150 W
Fuse: 5A250 V time lag (T5A L250V) IEC 60127-2II
Protection class
IP20 according IEC60529
2 = < 12.5 mm, finger or similar object
0 = Not protected against h
armful ingress of water
Environmental conditions
Operation : Temperature 15…30°C
Humidity max. 10-85 % non condensing
Altitude max. 4000 m
Temperature 2...50°C
Scope of Supply
HumaClot Pro
Packlist
User Manual
1
15800/1
Power Cable Length (1.8 m)
1
.................
Distilled Water Container (incl. screw lid)1
.................
Waste Water Container (incl. screw lid)1
.................
Distilled Water Sensor
1
.................
Waste Water Sensor
1
.................
Filter for Distilled Water Sensor
Syringe 250 µl (Built- in)
Cuvette Rings
10
.................
Sample Cups 4 ml
11
.................
Reagent Vial 5 ml Plastic
5
.................
Bottle Washing Solution
1
.................
Bottle Cleaner
1
.................
Reducing Rings for Reagent Vials 5 ml3
.................
USB Stick Teflon Coated Reagent Stirrer (3 * 13 mm)3
.................
Primary Fuses (250 V/5 A) Transport:
Humidity max. 10-85 % non condensing
Noise level: ≤ 56 db
Unit/Size
Cat.No.
1 15800
12
1
.................
.................
1 .................
2
.................
HumaClot Pro
Fully Automated Coagulation Instrument
Obligatory Items
Service Kits
Cat.No.
Need to be ordered with each instrument:
Maintenance Kit (Sufficient for twelve months)
Need to be ordered with the first instrument:
Alignment Tool Kit 15800/501
Starter Spare Part Kit
15800/502
15800/500
Optional
Consumables
Cuvette Rings for 1920 Tests
Unit/Size
6 Boxes with 10 cuvette rings 32 cuvettes incl. mixers and chip card
1
Cat.No.
15800/10
1
15800/20
HumaClot Pro Washing Solution 5 x 15m
To prevent carry-over. Instrument consumption approx. 10 Kits/Year depending on assay combination
Sample Cups 4 ml 2 x 250 pcs.
1
15800/25
1
15800/30
For sample, calibrator and controls
Cleaner 5 x 15 ml
For end of week cleaning of the instrument. Instrument consumption approx. 1 Kits/Year.
Reagent Vials 5 ml with Caps 50 pcs.
1
15800/40
Necessary for Hemostat FIB Buffer and Hemostat aPTT CaCl
Teflon Stirrer Bar 10 pcs.
1
15800/50
For reagent mixing e.G PT-SI
Reagents Unit/Size
Hemostat aPTT-EL 6 x 4 ml / 240 Tests
Cat.No.
1
33002
Hemostat D-Dimer 40 Tests
1
36002
Hemostat Fibrinogen 100 Tests
1
32002
Hemostat Thrombin Time 60 Tests 1
34002
Hemostat Thromboplastin-SI 6 x 2 ml / 60 Tests
1
31002
Hemostat Thromboplastin-SI 6 x 10 ml / 300 Tests
1
31003
Hemostat Control Plasma Normal 6 x 1 ml1
35001
Hemostat Control Plasma Abnormal 6 x 1 ml1
35002
Hemostat Calibrator – 4x1ml (lyophilized)
Accessories
USB Printer
1
35500
To generate a calibration curve for the determination of prothrombin time
Unit/SizeCat.No.
115800/80
15800/2015-03 © 2015 Human
Legal statement
While Human is making every effort to include accurate and up-to-date information, we make
no representations or warranties, express or implied, as to the accuracy or completeness of the
information provided in this document and disclaim any liability for the use of it.