Temperature Chamber Series

Add to my manuals
24 Pages

advertisement

Temperature Chamber Series | Manualzz

http://www.espec.co.jp/english

Head Office

3-5-6, Tenjinbashi, Kita-ku, Osaka 530-8550, Japan

Tel : 81-6-6358-4741 Fax : 81-6-6358-5500

ESPEC NORTH AMERICA, INC.

Tel : 1-616-896-6100 Fax : 1-616-896-6150

ESPEC EUROPE GmbH

Tel : 49-89-1893-9630 Fax : 49-89-1893-96379

ESPEC ENVIRONMENTAL EQUIPMENT (SHANGHAI) CO., LTD.

Head Office

Tel : 86-21-51036677 Fax : 86-21-63372237

BEIJING Branch

Tel : 86-10-64627025 Fax : 86-10-64627036

TIANJIN Branch

Tel : 86-22-26210366 Fax : 86-22-26282186

GUANGZHOU Branch

Tel : 86-20-83317826 Fax : 86-20-83317825

SHENZHEN Branch

Tel : 86-755-83674422 Fax : 86-755-83674228

SUZHOU Branch

Tel : 86-512-68028890 Fax : 86-512-68028860

ESPEC TEST TECHNOLOGY (SHANGHAI) CO., LTD.

Tel : 86-21-68798008 Fax : 86-21-68798088

ESPEC SOUTH EAST ASIA SDN.BHD.

Tel : 60-3-8945-1377 Fax : 60-3-8945-1287

TS7E26C02 (The contents of this catalog is as of July, 2014.)

Quality Management System Assessed and Registered

ESPEC CORP. has been assessed by and registered in the Quality Management System based on the International Standard ISO

9001:2008 (JIS Q 9001:2008) through the

Japanese Standards Association (JSA).

Registration : ESPEC CORP.

(Overseas subsidiaries not included)

Environmental Management System Assessed and Registered

ESPEC CORP.

● Specifications are subject to change without notice due to design improvements.

Corporate names and trade names mentioned in this catalog are trademarks or registered trademarks.

Temperature Chamber Series

CAT.NO.E94190-Q1407

1

Ideal for numerous applications ranging from high-temperature tests to drying and heat processing.

The "Perfect Oven" epitomizes the features and performance of the ideal oven.

It is a versatile product, conducting high-temperature tests, but also drying and heat treatment for production lines, with unrivaled reliability and performance.

The 56 models offered by ESPEC precisely answers the various needs of our customers.

2

MODEL VARIATION

PV H

+ 200℃/+ 300℃

(+ 392

°

F/+ 572

°

F)

Temperature Chamber

Vertical type

PH H

+ 200℃/+ 300℃

(+ 392

°

F/+ 572

°

F)

Temperature Chamber

Horizontal type

STPH

+ 500℃

(+ 932

°

F)

Ultra-High Temperature Chamber

SSPH

+700℃

(+1292

°

F)

Ultra-High Temperature Chamber

OVEN SERIES FOR VARIOUS APPLICATIONS

ST H

DESK-TOP TYPE

HIGH-TEMP CHAMBER

200

℃/+

300

15

L/

28

L/

39

L

VAC

VACUUM OVEN

200

933

×

10 2 to 1

×

10 2 Pa

3

LC

CONVECTION OVEN

200

℃/+

250

90

L/

165

L/

360

L

LCV

VACUUM OVEN

200

0 to

101 kPa

Gauge

SPH H

+ 200℃/+ 300℃

(+ 392

°

F/+ 572

°

F)

Temperature Chamber with Explosion Vent

IPH H

+ 200℃/+ 300℃

(+ 392

°

F/+ 572

°

F)

Anaerobic Temperature Chamber

GPH H

+ 200℃/+ 300℃

(+ 392

°

F/+ 572

°

F)

Temperature Chamber with Rotating Specimen Rack

PV H C

CLEAN OVEN

200

℃/+

300

℃/+

350

178

L/

380

L/

678

L

Class 5

H

LARGE VOLUME

TEMPERATURE CHAMBER

200

℃/

300

2250

L/

4050

L

4

5

Control operation

Two types of program instrumentation to suit different applications.

Standard Instrumentation and M-Instrumentation.

Constant operation mode

Alarm

User-friendly Standard

Instrumentation

St a nd a rd I nst r u ment at ion feat u res programmed operation with operational settings such as constant mode and automatic start/stop. Suitable for heat treatment, drying, and similar productionline applications.

M-Instrumentation features

programs with up to 20 steps

Suitable for a range of applications from temperature-characteristics testing to heat treatment and drying. Programmed operation now allows storing ten patterns, each up to twenty steps. Provides a wide range of functions, including temperature ramp settings and a maximum of 999 repeat cycles.

Easy setup with on-screen

display

Employs interactive settings for ease of use. Text can be displayed and entered in Japanese or English alphanumeric characters.

Four optional functions

Four optional functions, namely, air flow adjuster, automatic damper, integrating hour meter, and calendar timer can be included in the instrumentation. These functions can be set by using main panel instrumentation keys.

Interface (Option)

Interface for device communication can be selected between RS-485, GPIB and

RS-232C.

Control operation

Examples of Programmed Operation

(M-Instrumentation)

Temp.

Step program*

⑦…⑳

Time

Temp.

③ ⑦…⑳

Ramp program*

② ④

Time

Temp.

Automatic start*

⑦…⑳

Time

1.HOLD

Temp.

2.CONST

Temp.

3.OFF

Temp.

End of program

( selection

)

END END END

Time

Temperature at the final step is maintained.

Time

Operation at preset constant temperature.

Time

Oven is turned OFF

*

The number of repetitions of a program can be preset between 1 and 999.

Stepwise damper setting is possible using an optional automatic damper.

Guarantee soak function, whereby the timer is used to maintain a preset temperature

for a preset length of time, can also be performed.

Temperature Indicator-controller

Instrumentation Standard Instrumentation M-Instrumentation

Operation mode

Constant operation, programmed operation and remote operation through communication interface

Setting and indication ranges

Temperature: 0 to

210

(

32 to

410°F)

0 to

310

(

32 to

590°F)

0 to

510

(

32 to

950°F)

0 to

710

(

32 to

1310°F)

Time: 0 to 9999 hours 59 minutes

Setting resolution

Temperature: 1℃ 

Time: 1 minute

Programming function

One-pattern, two-steps program entry is possible.

10-patterns, 20-steps program entry is possible.

Ramp setting

Step or ramp temperature changes possible.

OFF mode

The oven can be turned off during programmed operation.

Automatic start

Timed start-up is possible by setting

the first step to 0℃

(i.e. oven OFF).

Automatic stop

Timed termination is possible by setting the oven to turn OFF upon completion of a program.

End mode

The operating status upon completion of a program can be set to

HOLD, CONST or OFF.

Repetition

Up to 999 times.

Auxiliary functions

Input burnout detection

Upper and lower temperature limit alarm

Upper deviation limit temperature alarm

Buzzer alarm

Automatic overheat protection

Trouble indication

Alarm indication

Self-diagnostic

Guarantee soak

Power failure recovery selection

Power failure protection

Quick timer

Quick operation

6

PV(H)

+ 200℃/+ 300℃

TEMPERATURE CHAMBER (Vertical type)

7

Test area

*Cable port and N

2

gas injector are options.

A space-saving upright chamber

Components are incorporated into the top portion of the vertical chamber, reducing installation space by

20~60% (comparison with conventional model).

Increases productivity on the production line, and saves laboratory space.

Seamless door interior structure

Door back is a single molded structure preventing heat losses from loose joints.

Large processing capacity

Since the f loor and shelves of the chamber have b e e n g r e a t l y r e i n fo r c e d , a l a r g e a m o u n t of specimens can be loaded and processed at the same time. The sliding shelves ensure easy handling of the specimens.

Excellent heating performance

Heating performance is greatly enhanced so that the chamber temperature remains constant even if the ventilation damper is opened. ( at

+20℃ ambient temperature)

SPECIFICATIONS

Model

System

Temperature range

*2

Temperature fluctuation

*2

Temperature uniformity

*2

Temperature heat-up time

Exterior material

Interior material

Insulation material

Heater

PV

212 PV

222 PV

232 PV

332 PVH

212 PVH

222 PVH

232 PVH

332

Forced hot-air circulation / ventilation system

Ambient temp.

20

(

68°F) to

200

(

392°F) Ambient temp.

20

(

68°F) to

300

(

572°F)

±

±

1.0

0.2

at

at

100

100 (

(

212°F),

212°F),

Ambient temp. to

±

2.0

200

at

℃ within 40 min.

(

200

(

200

(

392°F)

392°F)

392°F)

±

0.2

at

100

(

212°F),

200

(

392°F),

±

0.3

at

+ 300℃ (+

572°F)

±

1.0

at

100

(

212°F),

±

2.0

at

200

(

392°F),

±

3.0

at

+ 300℃ (+

572°F)

Ambient temp. to

300

(

572°F) within 60 min.

Cold rolled rust-proof steel plate, Melamine resin coating

Stainless steel plate

Glass wool

Sheathed heater

Air circulator

Damper

Fittings

Inside dimensions

 

W

×

H

×

Dmm (in)

Outside dimensions

*3

W

×

H

×

Dmm (in)

Capacity (L)

Stainless steel sirocco fan

Circulation

/

Ventilation (manual switching

600

×

600

×

600

(23.6

×

23.6

×

23.6)

770

(30.3

×

×

1200

47.2

×

×

925

36.4)

Power cable (approx 2m from chamber), Specimen power supply control terminals

(relay contact is opened/stop during malfunction. Voltage capacity 250V AC 3A)

600

(23.6

770

(30.3

×

×

×

×

900

35.4

1500

59

×

×

×

×

600

23.6)

925

36.4)

600

(23.6

770

(30.3

×

×

×

×

1200

47.2

1800

70.9

×

×

×

×

600

23.6)

925

36.4)

800

(31.5

1030

(40.6

×

×

×

×

1200

47.2

1800

70.8

×

×

×

×

800

31.5)

1145

45.1)

600

(23.6

770

(30.3

×

×

×

×

600

23.6

1200

47.2

×

×

×

×

600

23.6)

925

36.4)

600

(23.6

770

(30.3

×

×

×

×

900

35.4

1500

59

×

×

×

×

600

23.6)

925

36.4)

600

×

1200

×

600

(23.6

×

47.2

×

23.6)

800

×

1200

×

800

(31.5

×

47.2

×

31.5)

770

×

1800

×

925

(30.3

×

70.9

×

36.4)

1030

×

1800

×

1145

(40.6

×

70.8

×

45.1)

216 324 432 768 216 324 432 768

Weight (kg)

Allowable ambient conditions ts Power supply

(Voltage fluctuation:

±

10

of rated value)

165 190 210 325 165 190 210 325

200

/

220

/

230

/

1

φ

50

/

60Hz

Temperature: 0 to

40

(

32 to

104°F) Humidity: to 75

% rh

240V AC

200

/

220V AC

3

φ

3W 50

/

60Hz

380V AC

3

φ

4W 50Hz

230

1

200

φ

/

/

220

/

240V AC

50

/

60Hz

200

/

220V AC 3

φ

3W 50

3

φ

/

60Hz

380V AC

4W 50Hz

Max. power consumption

(kVA)

4.0

4.8

5.8

6.8

*1 Values assume circulatory operation with no specimens at an ambient temperature of

23

℃ ±

5.

*2 Conforms to Japan Testing Machinery standard K05:2000.

*3 Excluding protrusions.

4.0

5.8

6.2

8.8

Shelf pitch, quantity and load resistance

Model

Shelf pitch

Shelves

Shelf load resistance

*1 *2

Chamber total load resistance

*1

PV(H)-212 11

25kg PV(H)-222

PV(H)-232

50mm

PV(H)-332 80mm

*1 Including shelf weight

*2 Equally distributed load

17

23

14 45kg

200kg

SAFETY DEVICES

Leakage breaker

Electrical compartment door switch

Door switch

Thermal fuse

Temperature switch for air circulator

Heater wiring breaker

Upper and lower temperature limit alarm

(built inside temperature controller)

Overheat protector

Cartridge fuse

Specimen power supply control terminal

ACCESSORIES

Shelf (stainless steel wire)

(stainless steel plate for type 332)

Shelf bracket (stainless steel)

Cartridge fuse

User's manual

2

2 sets (4)

2

1 set

8

PH(H)

+ 200℃/+ 300℃

TEMPERATURE CHAMBER (Horizontal type)

9

Test area

Energy Saver Duct

PH-201

Damper open 50%

SET 200

SET 100

0 0.5

1

Energy save approx.

19

%

1.5

2

Standard

Energy Saver Duct

Energy save approx.

23

%

2.5

3

( kWh

*N

2

gas injector unit is option.

High performance chamber

A t e mp e r at u re -i nd icat ion cont rol le r w it h a n a d v a n c e d PI D o p e r a t io n , a n d a n o r ig i n a l l y d e v e l o p e d c h a m b e r c o n f i g u r a t i o n p r o v i d e u n m a t c h e d ove n p e r for m a nce. Te mp e r at u r e uniformity, temperature constancy, temperature heat-up rate, and temperature recovery time are performed with the upmost reliability.

Safety measures

Tr iple s a fe t y m e ch a n i s m s a r e e m ploye d fo r excessive overheating.

Wide model selection

We provide a total of 16 ovens with combination of temperature range, capacity, and instrumentation.

Energy Saver Duct (Option)

Energy saving approximately 20% by heat recycling f r o m e x h a u s t t h r o u g h t h e d u c t t o m a i n t a i n temperature while damper opens.

Standard

Energy Saver Duct

Standard

Energy Saver Duct

Intake air

Exhaust air

SPECIFICATIONS

Model

System

Temperature range

*2

Temperature fluctuation

*2

Temperature uniformity

*2

Temperature heat-up time

PH

102 PH

202 PH

302 PH

402 PHH

102 PHH

202 PHH

302 PHH

402

Forced hot-air circulation / ventilation system

Ambient temp.

20

(

68°F) to

200

(

392°F) Ambient temp.

20

(

68°F) to

300

(

572°F)

±

±

±

±

0.1

0.2

0.5

1.5

at

at

at

at

100

200

100

200

(

(

(

(

212°F)

392°F)

212°F)

392°F)

Ambient temp. to

±

±

±

±

0.2

0.4

1.0

2.0

200

at

at

at

at

(

100

200

100

200

392°F)

(

(

(

(

212°F)

392°F)

212°F)

392°F)

±

0.1

at

100

(

212°F)

±

0.2

at

200

(

392°F)

300

(

572°F)

±

0.5

at

100

(

212°F)

±

1.5

at

200

(

392°F)

±

2.5

at

300

(

572°F)

±

±

±

±

±

±

0.2

0.4

0.6

1.0

2.0

3.0

at

at

at

at

at

at

100

200

300

100

200

300

Ambient temp. to

300

(

572°F)

(

(

(

(

(

(

212°F)

392°F)

572°F)

212°F)

392°F)

572°F) within 40 min.

within 60 min.

within 60 min.

within 70 min.

Exterior material

Interior material

Insulation material

Heater

Cold rolled rust-proof steel plate, Melamine resin coating

Stainless steel plate

Glass wool

Iron chrome strip wire heater

Air circulator

Damper

Fittings

Inside dimensions

W

W

×

×

H

H

×

×

Dmm (in)

Dmm (in)

 

Outside dimensions

*3

Stainless steel propeller fan

Circulation

/

Ventilation (manual switching)

450

×

450

×

450

(17.7

×

17.7

×

17.7)

1040

(41

×

×

820

32.3

×

×

635

25)

Power cable (approx 2m from chamber), Specimen power supply control terminals

(relay contact is opened/stop during malfunction. Voltage capacity 250V AC 3A)

600

(23.6

1190

(46.9

×

×

×

×

600

23.6

970

38.2

×

×

×

×

600

23.6)

785

30.9)

800

(31.5

1500

(59.1

×

×

×

×

800

31.5

1210

47.6

×

×

×

×

800

31.5)

1065

41.9)

1000

(39.4

1730

(68.1

×

×

×

×

1000

39.4

1480

58.3

×

×

×

×

1000

39.4)

1275

50.2)

450

×

450

×

450

(17.7

×

17.7

×

17.7)

600

×

600

×

600

(23.6

×

23.6

×

23.6)

1040

×

820

×

635

(41

×

32.3

×

25)

1190

×

970

×

785

(46.9

×

38.2

×

30.9)

800

(31.5

1500

(59.1

×

×

×

×

800

31.5

1210

47.6

×

×

×

×

800

31.5)

1065

41.9)

1000

×

1000

×

1000

(39.4

×

39.4

×

39.4)

1730

×

1480

×

1275

(68.1

×

58.3

×

50.2)

91 216 512 1000 91 216 512 1000 Capacity (L)

Weight (kg)

Allowable ambient conditions ts Power supply

(Voltage fluctuation:

±

10

of rated value)

95 130

Temperature: 0 to

200

/

220

/

230

/

240V AC

1

φ

50

/

60Hz

240

3

200

φ

/

220V AC

3W 50

380V AC 3

φ

40

/

430

(

60Hz,

32 to

4W 50Hz

200

/

95

220

1

φ

/

230

50

/

/

130

104°F) Humidity: to 75

240V AC

60Hz

% rh

240

3

200

φ

/

430

220V AC

3W 50

380V AC 3

φ

/

60Hz,

4W 50Hz

Max. power consumption

(kVA)

2.0

2.7

5.0

6.5

*1 Values assume circulatory operation with no specimens at an ambient temperature of

23

℃ ±

5.

*2 Conforms to Japan Testing Machinery standard K05:2000.

*3 Excluding protrusions.

2.7

3.8

6.5

9.5

Shelf pitch, quantity and load resistance

Model

Shelf pitch

Shelves

Shelf load resistance

*1 *2

Chamber total load resistance

*1

SAFETY DEVICES

PH(H)-102

PH(H)-202

50mm

PH(H)-302 80mm

PH(H)-402 140mm

*1 Including shelf weight

*2 Equally distributed load

8

11

9

6

20kg

40kg

50kg

60kg

100kg

Leakage breaker

Electrical compartment door switch

Door switch (type 402 only)

Thermal fuse

Temperature switch for air circulator (except type 402)

Air circulator overload relay (type 402 only)

Heater wiring breaker

Reverse-prevention relay

Upper and lower temperature limit alarm

ACCESSORIES

(built inside temperature controller)

Overheat protector

Shelf (stainless steel wire for type102

Shelf bracket (stainless steel)

202) 2

2 sets (4)

Cartridge fuse

(stainless steel punched plate for type 302

402) 2

Specimen power supply control terminal

Cartridge fuse

User's manual

2

1 set

10

STPH

+ 500℃

ULTRA-HIGH TEMPERATURE CHAMBER

Temperature control to

500

Effective temperature range of (ambient temp.

+) 20℃ to +500℃ . The chamber can be used for a variety of applications, including tests of viability under hightemperatures and temperature resistance.

Door equipped with a single-action

lever

The door can be firmly locked by an easy-to-use single-action lever. It prevents accidents from unlocked doors.

SPECIFICATIONS

11

Model

System

Temp. range

*2

Temp. fluctuation

*2

Temp.

uniformity

Temp.

Interior

Insulation

Heater

*2 heat-up time

Air circulator

STPH-102 STPH-202

Forced hot-air circulation / ventilation system

Ambient temp.

20

(

68°F) to

500

(

932°F)

±

0.5

±

0.8

at

100

(

212°F)

±

1.8

at

200

(

392°F)

±

2.8

at

300

(

572°F)

±

3.8

at

400

(

752°F)

±

4.8

at

500

(

932°F)

Ambient temp. to

500

(

932°F) within 60min.

Stainless steel plate

Glass wool, MG wool

Iron chrome strip wire heater

Stainless steel propeller fan

Damper

Fittings

Inside dimensions

W

×

H

×

Dmm (in)

Outside dimensions

W

×

H

×

Dmm (in)

*3

Circulation/ Ventilation (manual switching)

Power cable (approx 2m from chamber),

Specimen power supply control terminals

Electrical compartment cooling fan

450

×

450

×

450

(17.7

×

17.7

×

17.7)

1190

×

1110

×

795

(46.9

×

43.7

×

31.3)

600

(23.6

1340

(52.8

×

×

×

×

600

23.6

49.6

×

×

1260

×

600

23.6)

×

945

37.2)

Capacity (L)

Weight (kg)

Allowable ambient conditions

Power supply

(

±

10

of rated value)

Max. power consumption

91

190

Temp.: 0 to

6.5 kVA

40

Humid.: to 75

% rh

380V AC 3

(

200 / 220V AC 3

φ

φ

32 to

216

250

104°F)

50/60Hz,

4W 50Hz

8.3 kVA

*1 Values assume circulatory operation with no specimens at an ambient temperature of

23

℃ ±

5.

*2 Conforms to Japan Testing Machinery standard K05:2000.

*3 Excluding protrusions.

ACCESSORIES

Shelf (stainless steel wire)

Shelf bracket (stainless steel)

Cartridge fuse

User's manual

SAFETY DEVICES

Leakage breaker

Electrical compartment door switch

Thermal fuse

Temperature switch for air circulator

Electrical compartment thermal switch

Heater wiring breaker

Upper and lower temperature limit alarm

(built inside temperature controller)

Overheat protector

Cartridge fuse

Specimen power supply control terminal

2

2 sets (4)

2

1 set

SSPH

Saving-energy insulated structure

Ceramic f iber and aluminium foil are used as insulation materials. It increases effective insulation and prevents heat loss, thus saving energy.

A Double seal gasket configuration

A gasket made of stainless steel fiber and a leaf spring are used to form a double seal between the door and the chamber. Prevents heat radiation on door.

Door equipped with a single-action lever

The door can be firmly locked by an easy-to-use single-action lever.

+ 700℃

ULTRA-HIGH TEMPERATURE CHAMBER

ACCESSORIES

Shelf (stainless steel wire)

Shelf bracket (stainless steel)

Cartridge fuse

User's manual

SAFETY DEVICES

Leakage breaker

Electrical compartment door switch

Thermal fuse

Temperature switch for air circulator

Air circulator rotation detector

Electrical compartment thermal switch

Heater wiring breaker

Upper and lower temperature limit alarm

(built inside temperature controller)

Overheat protector

Cartridge fuse

Specimen power supply control terminal

2

2 sets (4)

2

1 set

SPECIFICATIONS

Model

System

Temp. range

*2

Temp.

fluctuation

Temp.

uniformity

Temp.

*2

*2 heat-up time

SSPH-102 SSPH-202

Forced hot-air circulation / ventilation system

100 to

700

(

212 to

1292°F)

±

0.5

at

100 to

500

(

212 to

932°F)

±

0.8

at

501 to

700

(

933 to

1292°F)

±

0.8

at

100

(

212°F)

±

2.8

at

300

(

572°F)

±

4.8

at

500

(

932°F)

±

7.0

at

700

(

1292°F)

Ambient temp. to

700

(

1292°F) within 120min.

within 160min.

Stainless steel plate

Glass wool, Ceramic fiber

Iron chrome strip wire heater

Stainless steel propeller fan

Interior

Insulation

Heater

Air circulator

Damper

Fittings

Inside dimensions

W

×

H

×

Dmm (in)

Outside dimensions

W

×

H

×

Dmm (in)

*3

Capacity (L)

Circulation/ Ventilation (manual switching)

Power cable (approx 2m from chamber),

Specimen power supply control terminals

Electrical compartment cooling fan

450

×

450

×

450

(17.7

×

17.7

×

17.7)

1190

×

1110

×

795

(46.9

×

43.7

×

31.3)

91

600

(23.6

1340

(52.8

×

×

×

×

600

23.6

49.6

216

×

×

1260

×

600

23.6)

×

945

37.2)

Weight (kg)

Allowable ambient conditions

Power supply

(

±

10

of rated value)

250 330

Temp.: 0 to

40

(

32 to

104°F)

Humid.: to 75

% rh

200

/

220V AC 3

φ

50

/

60Hz,

380V AC 3

φ

4W 50Hz

Max. power consumption

8.3 kVA 9.5 kVA

*1 Values assume circulatory operation with no specimens at an ambient temperature of

23

℃ ±

5.

*2 Conforms to Japan Testing Machinery standard K05:2000.

*3 Excluding protrusions.

12

SPH(H)

+ 200℃/+ 300℃

TEMPERATURE CHAMBER WITH EXPLOSION VENT

Temperature chamber

with Explosion Vent

This temperature chamber is suitable for drying and heat-treatment of f lammable synthetic resins or volatile solvents. It is equipped with an explosion vent which releases explosion and a safety door to ensure security.

Door equipped

with a single-action lever

The door can be securely locked by an easy-to-use single-action lever. Even if the operator accidentally turns on the power when door is unlocked, the door lock detection switch prevents heater fan from starting. Besides, in three minutes, the alarm buzzer sounds to call for warning.

 

Release explosion safely

Aluminium plate

Insulation material

Flame

Specimen

Explosion

In case an explosion occurs inside the test chamber, as shown in the above image, insulation material is bent and blown upward together with the aluminium plate to the metal screen at the top of the chamber.

This way the explosion is safely channeled and released through the top metal screen. For SPH(H)-402, explosion is released through the top metal screen by bending insulation material on the rear wall.

Release explosion safely

13

In the event that an explosion occure inside the test chamber, as shown in the below figures, insulation material is bent and blown upward together with the aluminium plate to the metal screen at the top of the chamber.

In this way the explosion is safely channeled and released through the top metal screen.

Aluminium plate

Insulation material

Flame

   

WARNING

1) The following flammables or materials containing them can be subjected to drying (heat treatment) with this chamber. However, to avoid explosion, ventilate the chamber well and use the chamber below the explosive limit.

注意

Inflammables:

Ignitable Substances

1. Ethyl ether, gasoline, acetaldehyde, propylene oxide, c a r b o n d i s u l f i d e , c a r b o n d i o x i d e a n d o t h e r substances with an ignition point of below

30

℃ .

2. Normal hexane, ethylene oxide, acetone, benzene, methyl ethyl ketone and other substances with an ignition point above

30

and below 0

.

3 . M e t h a n o l , e t h a n o l , x y l e n e , p e n t y l a c e t a t e amylacetate and other substances with an ignition point above 0

℃ and below

30

.

4. Kerosene, light oil, turpentine oil, isopentyl alcohol

(also called isoamyl alcohol), acetic acid and other substances with an ignition point above

30

℃ and below

65

.

Combustible Gases

Hydrogen, acetylene, ethylene, methance, ethane, propane, butane, and other combustible substances that are in a gaseous state at a temperature of

15

℃ and at a pressure of 1 atmosphere.

2) Temperature chamber with explosion vent is fitted with a comprehensive range of devices to ensure safety. In addition to the regular inspection, these must be carefully inspected before reusing after an explosion.

3) This equipment is designed to prevent any damage to people or equipment in the vicinity for explosion pressures not exceeding 29.4kPa. If the explosion pressure exceeds 9.8kPa, reuse of the equipment itself may not be possible.

4) Please refer to the instruction manual before using the chamber to ensure safe operation.

Specimen

Explosion

SPECIFICATIONS

Model

System

Temperature range

*2

Temperature fluctuation

*2

Temperature uniformity

*2

Temperature heat-up time

SPH

102 SPH

202 SPH

302 SPH

402 SPHH

102 SPHH

202 SPHH

302 SPHH

402

Forced hot-air circulation / ventilation system

Ambient temp.

20

(

68°F) to

200

(

392°F) Ambient temp.

20

(

68°F) to

300

(

572°F)

±

±

±

±

0.1℃ at +

0.2℃ at +

0.5℃ at +

1.5℃ at +

100

200

100

200

(

(

(

(

212°F)

392°F)

212°F)

392°F)

Ambient temp. to

±

±

±

±

0.2

0.4

1.0

2.0

200

at

at

at

at

(

100

200

100

200

392°F)

(

(

(

(

212°F)

392°F)

212°F)

392°F)

±

±

±

±

±

±

0.1

0.2

0.2

0.5

1.5

2.5

at

at

at

at

at

at

100

200

300

100

200

300

(

(

(

(

(

(

212°F)

392°F)

572°F)

212°F)

392°F)

572°F)

± 0.2℃ at +

100

(

212°F)

± 0.4℃ at +

200

(

392°F)

± 0.6℃ at +

300

(

572°F)

±

1.0℃ at +

100

(

212°F)

± 2.0℃ at +

200

(

392°F)

±

3.0℃ at +

300

(

572°F)

Ambient temp. to

300

(

572°F) within 40 min.

within 60 min.

within 60 min.

within 70 min.

Exterior material

Interior material

Insulation material

Explosion vent

Heater

Cold rolled rust-proof steel plate, Melamine resin coating

Stainless steel plate

Glass wool

Safety vent to release inside pressure on explosion,

Explosion exhaust duct, Protective wire mesh, Insulation, Outer plate

Stainless steel, Sheated heater with fin

Stainless steel propeller fan

Circulation

/

Ventilation (manual switching)

Air circulator

Damper

Fittings

Inside dimensions

 

W

×

H

×

Dmm (in)

Outside dimensions

*3

W

×

H

×

Dmm (in)

Capacity (L)

450

×

450

×

450

(17.7

×

17.7

×

17.7)

1040

(41

×

×

1260

49.6

×

×

635

25)

Power cable (approx 2m from chamber), Specimen power supply control terminals

(relay contact is opened/stop during malfunction. Voltage capacity 250V AC 3A).

600

(23.6

1190

(46.9

×

×

×

×

600

23.6

×

1370

53.9

×

×

×

600

23.6)

785

30.9)

800

(31.5

1500

(59.1

×

×

×

×

800

31.5

1715

68.1

×

×

×

×

800

31.5)

1065

41.9)

1000

(39.4

1730

(68.1

×

×

×

×

1000

39.4

1800

70.9

×

×

×

×

1000

39.4)

1775

69.9)

450

×

450

×

450

(17.7

×

17.7

×

17.7)

600

×

600

×

600

(23.6

×

23.6

×

23.6)

1040

×

1260

×

635

(41

×

49.6

×

25)

1190

×

1370

×

785

(46.9

×

53.9

×

30.9)

800

(31.5

1500

(59.1

×

×

×

×

800

31.5

1715

68.1

×

×

×

×

800

31.5)

1065

41.9)

1000

×

1000

×

1000

(39.4

×

39.4

×

39.4)

1730

×

1800

×

1775

(68.1

×

70.9

×

69.9)

91 216 512 1000 91 216 512 1000

Weight (kg)

Allowable ambient conditions ts Power supply

(Voltage fluctuation:

±

10

of rated value)

95 130 270 500 95 130 270 500

200

/

220

/

230

/

Temperature: 0 to

40

(

32 to

104°F) Humidity: to 75

% rh

240V AC

1

φ

50

/

60Hz

200

/

220V AC

3

φ

3W 50

/

60Hz,

380V AC 3

φ

4W 50Hz

200

/

220

1

φ

/

230

50

/

/

240V AC

60Hz

200

/

220V AC

3

φ

3W 50

/

60Hz,

380V AC 3

φ

4W 50Hz

Max. power consumption

(kVA)

2.0

2.7

5.0

6.5

*1 Values assume circulatory operation with no specimens at an ambient temperature of

23

℃ ±

5.

*2 Conforms to Japan Testing Machinery standard K05:2000.

*3 Excluding protrusions.

2.7

3.8

6.5

9.5

ACCESSORIES SAFETY DEVICES

Shelf (stainless steel wire for type 102

202) 2

(stainless steel punched plate for type 302

402) 2

Leakage breaker

Electrical compartment door switch

Shelf bracket (stainless steel) 2 sets (4)

Chamber door lock detection switch

Cartridge fuse

Protective wire mesh

2

1

Explosion detection limit switch

Thermal fuse

(stainless steel mesh with soft aluminium foil)

Temperature switch for air circulator (except type 402)

Insulation (glass wool)

Outer plate (thin soft aluminium panel)

Stand bracket and hexagon socket head cap screw

(for type 102

202)

3

1

4 each

Air circulator overload relay (for type 402 only)

Heater wiring breaker

Reverse-prevention relay (for type 402 only)

Hexagon socket screw key (for type 102

202) 1

User's manual 1 set

Upper and lower temperature limit alarm

(built inside temperature controller)

Overheat protector

Cartridge fuse

Specimen power supply control terminal

14

IPH(H)

+ 200℃/+ 300℃

ANAEROBIC TEMPERATURE CHAMBER

Low oxygen level testing

Equipped with a non-oxidizing gas intake structure which fills the chamber with non-oxidizing gas such as CO

2

or N

2

for heat treatment or temperature cha r a cter ist ics t est i ng requ i r i ng low oxygen concentration atmosphere.

Hermetically sealed configuration

The chamber is hermetically sealed to decrease oxygen inside the chamber. The inner stainless steel plate is seamless welded with argon gas.

O

2

concentration indicator controller (optional)

An optional O

2

concentration indicator controller equipped with an oxygen sensor is available. It allows precise regulation of the O

2

level throughout the range 0.5 to 21% (using N

2

).

SPECIFICATIONS

Model

System

Temp. range

Temp. fluctuation

Temp. uniformity

*2

*2

*2

IPH

202 IPHH

202

±

±

±

±

0.1

0.2

0.5

1.5

Forced hot-air circulation system

Ambient temp.

20

(

68°F) to

200

(

392°F)

at

at

at

at

100

200

100

200

(

Ambient temp. to

(

(

(

(

212°F)

392°F)

212°F)

392°F)

200

392°F) within 40min.

Ambient temp. to

300

(

20

(

572°F)

68°F)

±

0.1

at

100

(

212°F)

±

0.2

at

200

(

392°F)

±

0.2

at

300

(

572°F)

±

0.1

at

100

(

212°F)

±

0.2

at

200

(

392°F)

±

0.2

at

300

(

572°F)

Ambient temp. to

300

(

572°F) within 60min.

CO

2

, N

2

gas (ordinary temperature, dry gas)

Temp. heat-up time

Fluid

Fluid pressure

Flow rate

Chamber O

2

level

Allowed max. pressure: 2.0MPa (Gauge) (primary side of valve)

Secondary side is adjusted with the valve to 0.05MPa (Gauge)

Max. flow rate: 20 L /

min. (0.05MPa (Gauge), 20

)

0.5

% ( lowest

)

Chamber injector pressure

29Pa (Gauge) and over (at max flow rate)

Valve

Pressure gauge

Flow meter

Scale range

1/4" brass needle valve

φ

75mm embedded type class 2.5

Scale range: 0

0.1MPa (Gauge)

Floating type (provided with needle valve for flow rate control)

0 to 30L

/

min. N

2

gas

Safety valve

Gas inlet

Trip pressure: 2.0kPa (Gauge)

1

/

4" ring joint

Fittings

Power cable (approx 2m from chamber),

Specimen power supply control terminal

Inside dimensions (in) W600mm

×H

600mm

×D

600mm (23.6

×

23.6

×

23.6)

Outside dimensions (in)

*3

W1190mm

×H

970mm

×D

785mm (46.9

×

38.2

×

30.9)

Capacity (L) 216

15

Weight (kg) 130

Allowable ambient conditions Temp.: 0 to

40

(

32 to

104°F) Humid.: to 75

% rh

Power supply

(

±

10

of rated value)

200

/

220

/

230

/

240V AC 1

φ

50

/

60Hz

Max. power consumption 2.7 kVA 3.8 kVA

*1 Values assume circulatory operation with no specimens at an ambient temperature of

23

℃ ±

5.

*2 Conforms to Japan Testing Machinery standard K05:2000.

*3 Excluding protrusions.

ACCESSORIES

Shelf (stainless steel wire)

Shelf bracket (stainless steel)

Cartridge fuse

User's manual

SAFETY DEVICES

Leakage breaker

Electrical compartment door switch

Thermal fuse

Temperature switch for air circulator

Heater wiring breaker

Upper and lower temperature limit alarm

(built inside temperature controller)

Overheat protector

Cartridge fuse

Specimen power supply control terminal

2

2 sets (4)

2

1 set

GPH(H)

+ 200℃/+ 300℃

TEMPERATURE CHAMBER WITH ROTATING SPECIMEN RACK

Suitable for heat deterioration test

Based on the PH Temperature Chambers, these models incorporate a detachable rotating specimen rack and is especially designed for heat deterioration testing of rubbers and plastics including polyesters and vinyls.

Equipped with a rotating specimen rack

The rack drive unit is installed inside, enhancing function and lending them a simple appearance.

By removing the rack, this equipment may also be operated as an ordinary temperature chamber.

SPECIFICATIONS

Model

System

Temp.range

*2

Temp. fluctuation

Temp.uniformity

*2

*2

Temp.heat-up time

GPH−102 GPH−202 GPHH−102 GPHH−202

Forced hot-air circulation / ventilation system

Ambient temp.

20

(

68°F) to

200

(

392°F)

±

±

±

±

0.1

0.2

0.5

1.5

at

at

at

at

100

200

100

200

(

(

(

(

212°F)

392°F)

212°F)

392°F)

Ambient temp. to

300

(

20

(

572°F)

68°F)

±

0.1

at

100

(

212°F)

±

0.2

at

200

(

392°F)

±

0.2

at

300

(

572°F)

±

0.5

at

100

(

212°F)

±

1.5

at

200

(

392°F)

±

2.5

at

300

(

572°F)

Ambient temp. to

200

(

392°F) within 40min.

Ambient temp. to

300

(

572°F) within 60min.

1 2 1 2

320mm (12.6in.)

Number of racks

Outside diameter

Available numbers of specimens/weight

56pcs per rack (up to 0.7kg)

Specimen clip

RPM of specimen rack

Motor

W

×

Hmm (in)

Construction

Chamber lamp

50pcs per rack

5rpm

/

50Hz, 6rpm

/

60Hz

1φ 15W

190

×

340(7.48

×

13.39) ————

Heat-resistant reinforced glass

3-plate sets

5.5W

incandescent lamp

————

————

Fittings

Power cable (approx 2m from chamber),

Specimen power supply control terminals

(relay contact is opened during malfunction.

Voltage capacity 250V AC 3A)

Inside dimensions

 

W

×

H

×

Dmm (in)

Outside dimensions

*3

W

×

H

×

Dmm (in)

 

450

×

450

×

450

(17.7

×

17.7

×

17.7)

600

×

600

×

600

(23.6

×

23.6

×

23.6)

450

×

450

×

450

(17.7

×

17.7

×

17.7)

600

×

600

×

600

(23.6

×

23.6

×

23.6)

1040

(91

×

×

820

32.3

×

×

635

25

)

1190

(46.9

×

×

970

38.2

×

×

785

30.9)

1040

(91

×

×

820

32.3

×

×

635

25)

1190

(46.9

×

×

970

38.2

×

×

785

30.9)

Capacity (L) 91 216 91 216

Weight (kg) 95 130 95 130

Allowable ambient conditions Temp.: 0 to

40

(

32 to

104°F) Humid.: to 75

% rh ts Power supply

(

±

10

of rated value)

200

/

220

/

230

50

/

/

240V AC 1

60Hz

φ

Max. power consumption 2.0 kVA 2.7 kVA 3.8 kVA

*1 Values assume circulatory operation with no specimens at an ambient temperature of

23

℃ ±

5.

*2 Conforms to Japan Testing Machinery standard K05:2000.

*3 Excluding protrusions.

Test area

ACCESSORIES

Shelf (stainless steel wire)

Shelf bracket (stainless steel)

2

2 sets (4)

Cartridge fuse 2

Specimen clip type102 50 type202 100

Shaft insulation filters

User's manual

1 set

1 set

SAFETY DEVICES

Leakage breaker

Electrical compartment door switch

Thermal fuse

Temperature switch for air circulator

Heater wiring breaker

Upper and lower temperature limit alarm

(built inside temperature controller)

Overheat protector

Cartridge fuse

Specimen power supply control terminal

16

17

OPTIONS

Model

PV

PVH

PH

PHH

STPH SSPH

SPH

SPHH

IPH

IPHH

GPH

GPHH

212 222 232 332 102 202 302 402 102 202 102 202 102 202 302 402 202 102 202

Option

Time run-out output

Calendar timer

Integrating hour meter

● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●

● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●

● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●

Temperature recorder terminal

● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●

Paperless recorder/

Temperature recorder

● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●

Recorder wiring

Automatic damper

Exhaust port flange

● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●

Exhaust duct

Energy Saver Duct

● ● ● ● ● ● ● — ● ● ● ● ● ● ● — — ● ●

— — — — ● ● — — — — — — ● ● — — — ● ●

Nitrogen gas injector

● ● ● ● ● ● ● ● ● ● ● ● — — ● ● — — —

Inert specification

350℃ Specification

— — — — — — — — ● ● — — — — — — — — —

— — — — —

— — — — — —

— — — —

O

2

concentration indicator-controller

— — — — — — — — — — — — — — — — ● — —

Air flow adjuster ● ● ● ● ● ● ● ● — — — — — — — — ● ● ●

Fin heater

— — — — ● ● ● ● — — — — ※ ※ ※ ※ ● ● ●

Shelf and shelf bracket

18-8 Cr-Ni stainless steel wire

18-8 Cr-Ni punched stainless steel shelf

Mesh shelf

Heavyduty shelf

Vertical type

● ● ● ● — — — — — — — — — — — — — — —

● ● ● ● — — — — — — — — — — — — — — —

Horizontal type (25kg)

— — — — ● ● — — — — — — ● ● — — ● ● ●

Horizontal type (60kg)

— — — — — ● ● — — — — — — ● ● — ● — ●

Cable port

Cable port rubber plug

Casters

Viewing window

Chamber lamp

Anchoring fixtures

Floor reinforcement

● ● ● ● ● ● ● ● ● ● — — — — — — — ● ●

— —

— — — — — — — —

— ● ● ● ● ● — — ● ● ● ● — — — — ● ● ●

— —

— — — — — — — — — —

— —

— — — — — — — — — —

● ● ● ● ● ● ● — — — — — ● ● ● — ● ● ●

— — — — — ● ● ● — — — — — ● ● ● ● — ●

Stand

Vertical type

Horizontal type

Angle type stand

Casters for stand

● ● — — — — — — — — — — — — — — — — —

— — — — ● ● ● — ● ● ● ● ● ● ● — ● ● ●

— — — — — — — ● — — — — — — — ● — — —

— — — — ● ● — — — — — — — — — — ● ● ●

Stacking brackets

— — — — ● ● — — — — — — — — — — ● — —

L-type-stand and stacking brackets

— — — — ● ● — — — — — — — — — — ● — —

External alarm terminal

● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●

Emergency stop pushbutton

Color specification

Interface

Power cable

Standard specification

● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●

● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●

● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●

● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●

OPTIONS

Time run-out output

This option enables turning the power to the specimen ON or OFF with contact signal output when the time is up by using the timer function on the controller.

Power supply rating: 250VAC 1A

Actuation: Contact close when program time overflows

Where located: Right side of chamber

Temperature recorder terminal

Outputs chamber temperature through thermocouple type K (JIS C 1602)

(Thermocouple type N for STPH, SSPH)

Where located:

Rear of electrical compartment

Temperature recorder

Temp. range: 0 to

+200℃

0 to

+300℃

0 to

+600℃

0 to

+800℃

Recording system:

Pen recorder (1 pen) or multi-point recorder (6 dots)

* If performing simultaneous installation of a recorder and an N

2

gas injector, they must be handled individually. Otherwise, installation may be limited depending on what other options are chosen.

Calendar timer

Automatically starts and stops chamber operation.

Setting range:

Sunday to Saturday

(Possible to set multiple days)

0:00 to 23:59

(Setting resolution 1 minute)

Margin of error per month:

±1 minute

Paperless recorder

Records temperature of each section such as the temperature inside the chamber.

Temp. range: 0

〜+200℃

0

〜+300℃

0

〜+600℃

0

〜+800℃

Number of inputs: Temperture 1

(5 more channels can be turned ON)

Data saving cycle: 5 sec

External recording media:

CF memory card port

(Includes a 256MB CFcard)

Language support: ENG, JPN

Temperature recorder wiring

Preparation of a power cable, temperature sensor and a grounding wire for additional installation in the future.

Automatic damper

A u t o m a t i c a l l y o p e n s o r c l o s e s synchronously with program operation for ventilation and faster cooling of chamber temperature.

Integrating hour meter

Displays cumulative chamber operation time.

Available with or without reset feature.

* Operating time is not accumulated when operation is stopped due to malfunction or for other reasons.

Measuring time: 999,999 hr

Please refer to chart on p.17 for applicable models.

18

Energy Saver Duct

PH-201

Damper open 50%

SET 200

SET 100

0 0.5

1

Energy save approx.

19

%

1.5

2

Standard

Energy Saver Duct

Energy save approx.

23

%

2.5

3

( kWh

OPTIONS

Exhaust port flange

Flange connects an exhaust duct to the chamber to exhaust hot air from the chamber.

(for oven with damper.)

Material: Cold rolled steel plate with chromate conversion coatings

Stainless steel sheet

STPH-102, 202

SSPH-102, 202

Dimensions: External diameter 87mm

Location: Chamber rear side

* When connecting to exhaust duct, the length of duct must be less than 4m.

Energy Saver Duct

Discharges exhaust air towards to ceiling with heat exchanger for energy

Dimensions of exhaust port flange:

External diameter 87mm

* The height of chamber will be 30mm higher than standard unit by the additional lower flame.

Intake air

Exhaust air

350

specification

Ad apted to prov ide a ma x i mu m temperature of 350

℃.

* PHH only.

O

2

concentration indicator-controller

Controls oxygen concentration inside the oven.

O

2

concentration range:

0.5 to 21% oxygen concentration (v/v)

Gas: N

2

gas

(ordinary temperature dry gas)

* IPH(H) only.

Exhaust duct

Discharges hot air towards the ceiling.

(for oven with damper.)

Dimensions: External diameter 87mm

Location: Chamber rear side

* Exhaust port f lange is provided at end of exhaust duct.

Nitrogen gas injector

Used for reducing specimen oxidation.

Fluid pressure: Max. allowable pressure

2.0MPa (Gauge) on primary side of valve

0.05MPa(Gauge)on secondary side using valve.

Max flow rate: 30 L min.

Flow meter: Float type flow meter

* If performing simultaneous installation of a recorder and an N

2

gas introducing unit, they must be handled individually. Otherwise, installation may be limited depending on what other options are chosen.

Air flow adjuster

Allows low air velocity in chamber

PV(H) 0.3 to 2.3m/s

PH(H)-102/202

GPH(H)-102/202 0.2 to 2.3m/s

IPH(H)-202

PH(H)-302

PH(H)-402

0.3 to 2.3m/s

0.3 to 2.6m/s

Average wind velocity across chamber central longitudinal section. Represents the typical mean value for each chamber.

19

Please refer to chart on p.17 for applicable models.

Inert specification

Used to minimize the oxidation of specimens.

* STPH only.

* Standard dampers are not fitted.

Fin heater

Used when anti-corrosive is required.

Stainless steel sheathed heater with fins.

OPTIONS

Shelf and shelf bracket

Equivalent to standard accessory.

PH(H)-102/202, SPH(H)-102/202,

GPH(H), and IPH(H) include stainless steel punched plate that differs from the standard shelf provided.

Heavy-duty shelf

Used to hold heavy specimen exceeding the load capacity of the standard shelf.

Additional cable port

A through hole provided on the wall of chamber.

Material: Stainless steel plate

Inside diameter: 25, 50, 100mm

(

φ50mm for STPH-102・202)

* The cable port may not be able to be used at the same time as the optional exhaust duct. (Except when used with PV(H))

* If several cable ports are installed, the surface temperature may rise or the chamber may not be able to meet standard performance.

Stainless steel wire

Stainless steel punched plate

Mesh shelf

For testing small specimens.

Material: 18-8 Cr-Ni stainless steel

φ0.8, 5 mesh

* To use, place this shelf on a standard shelf.

Vertical type

Material: 18-8 Cr-Ni stainless steel wire

Shelf support load resistance: Max 200kg

Model

PV(H)-212

-222

-232

Shelf load resistance*

PV(H)-332

*Uniformly distributed load

45kg

90kg

Horizontal type

For 25kg

Material: 18-8 Cr-Ni stainless steel wire

Shelf support load resistance: Max 50kg

* Equipped with 2 sets of shelf and shelf bracket.

For 60kg

Material: 18-8 Cr-Ni punched stainless steel

Shelf support load resistance: Max 200kg

* Standard shelves not provided.

Possible installation points

Model Top Rear Left side right side

× × ○ ○

PV(H)

PH(H)-

102

202

302

PH(H)-402

GPH(H)

STPH(H)

×

×

×

×

×

×

×

×

Cable port rubber plug

Prevents airleakage from the cable port.

Inside diameter: 25, 50, 100mm

* This rubber plug cannot be used when operating the chamber at

+200℃ or higher.

Model

PV(H)-212

-222

-232

PV(H)-332

Size

W550

×

D600

×

H35 mm

W740

×

D740

×

H38 mm

*Uniformly distributed load.

Shelf load resistance*

10kg

15kg

Casters

Installed for mobility.

・Adjustable type (Height 92mm)

4 casters

4 leveling feet

・ Non-adjustable type (Height 85mm)

2 casters with stoppers

2 fixed wheels

Please refer to chart on p.17 for applicable models.

20

OPTIONS

Viewing window

Used for observation of the specimens inside the chamber.

Dimensions: W190

×L340 mm

Stand

Exterior: Cold rolled and rust-proof steel plate with melamine baked finish

Vertical type

Type Top Model

MV-23

MV-23C

300mm

321mm

PV(H)-212

MV-26 600mm

PV(H)-212

MV-26C 621mm

*Type C: Casters and leveling feet

*with door

222

Angle type stand

Added to the chamber's original stand, this stand makes it easier to load and unload the specimen to the lower part of the test chamber.

Exterior: Equal-angle steel

Melamine baked finish

Model Type

L

M

H

Height

150mm

300mm

450mm

PH(H)-402

SPH(H)-402

Chamber lamp

Required when the door is fitted with viewing windows.

Location (incandescent light bulb):

PH-102, 202-Test area ceiling

PH-302, 402-Test area rear wall

Anchoring fixtures

Used to bolt the chamber to the floor.

Floor reinforcement

Used when testing load is larger than standard maximum load capacity.

* This option should be ordered together with the chamber.

Model

Floor load resistance*

Standard load resistance*

PH(H)-202

SaPH(H)-202

GPH(H)-202

IPH(H)-202

Up to 200kg

PH(H)-302

SPH(H)-302

PH(H)-402

SPH(H)-402

Up to 300kg

* Equally distributed load

50kg

60kg

100kg

MV-23C

Horizontal type

Type

L-1

L-2

L-3

M-1

M-2

Height

140mm

Model

PH(H)-102, GPH(H)-102

PH(H)-202, GPH(H)-202

IPH(H)-202

200mm PH(H)-302, SPH(H)-302

365mm PH(H)-102, GPH(H)-102

PH(H)-202, GPH(H)-202

IPH(H)-202

M-3

MS-1

MS-2

H-1(D)

400mm

505mm

PH(H)-302, SPH(H)-302

STPH-102, SSPH-102

STPH-202, SSPH-202

PH(H)-102, SPH(H)-102,

GPH(H)-102

H-2(D) 540mm

PH(H)-202, SPH(H)-202,

GPH(H)-202, IPH(H)-202

H-3(D) 585mm PH(H)-302, SPH(H)-302

*Type(D): with door

Casters for stand

Attached to the optional stand.

・Height adjustable (Height 92mm)

 Free-turning wheel 4

 Leveling feet

4

Horizontal type

 L-1, L-2, M-1, M-2, H-1(D), H-2(D),

 MS-1, MS-2

Stacking brackets

When stacking two chambers, this plate couples the top and bottom chambers securely.

*Only the L model optional stand can be used when chambers are stacked.

L-type-stand and stacking brackets

An L-type stand is fitted to the optional stacking brackets.

21

Please refer to chart on p.17 for applicable models.

Some photographs listed in this catalog contain Japanese display.

From the side, L-2,

M-2 (casters are optional) and H-2

OPTIONS

External alarm terminal

Used as a contact that relays an alarm to a remote point when one of the safety devices trips.

Output point: 1

Power supply: 250V AC 1A

Actuation: Signal generated when troubles occurs (contact closed)

Where located: Right side of chamber

Color specification

Chamber can be painted to any desired color.

Does not apply to:

・Door handle and handle cover

・Specimen power supply control

terminal frame

・Instrumentation frame

・Operation panel

・Damper operation panel

(including knob)

・Hinge cover

・Breaker cover

*Submit a color sample when specifying a color.

Emergency stop pushbutton

Stops the chamber immediately.

Interface

Communications ports to connect the chamber to a PC.

・RS-485

・GPIB

・RS-232C

Communication cables

・ RS-485

5m/ 10m/ 30m

・ RS-232C 1.5m/ 3m/ 6m

・ GPIB

2m/ 4m

Power cable

・5m

・10m

Safety precautions

● Do not use specimens which are explosive or inflammable, or which contain such substances. To do so could be hazardous, as this may lead to fire or explosion.

Periodical cleaning of the chamber and exhaust duct is required for it may cause combustion and fire when vapor of specimen is built up. Furthermore, an interior argon welding can be applied to the insulation layer of the chamber to minimize vapor penetration which may cause fire (except

IPH(H)). For more information, please contact us.

● Be sure to read the operation manual before operation.

22

Safety precautions

Do not use specimens which are explosive or inflammable, or which contain such substances.

 

To do so could be hazardous, as this may lead to fire or explosion.

● Periodical cleaning of the chamber and exhaust duct is required for it may cause combustion and fire when vapor of specimen is built up. Furthermore, an interior argon welding can be applied to the insulation layer of the chamber to minimize vapor penetration which may cause fire (except IPH(H)).

For more information, please contact us.

● Be sure to read the operation manual before operation.

DANGER

Do not use specimens which are explosive or inflammable, or which contain such substances.

To do so could be hazardous, as this may lead to fire or explosion.

CAUTION

Periodical cleaning of the chamber and exhaust duct is required for it may cause combustion and fire when vapor of specimen is built up.

Furthermore, an interior argon welding can be applied to the insulation layer of the chamber to minimize vapor penetration which may cause fire (except IPH(H)). For more information, please contact us.

Be sure to read the user’s manual before operation.

http://www.espec.co.jp/english

Head Office

3-5-6, Tenjinbashi, Kita-ku, Osaka 530-8550, Japan

Tel : 81-6-6358-4741 Fax : 81-6-6358-5500

ESPEC NORTH AMERICA, INC.

Tel : 1-616-896-6100 Fax : 1-616-896-6150

ESPEC EUROPE GmbH

Tel : 49-89-1893-9630 Fax : 49-89-1893-96379

ESPEC ENVIRONMENTAL EQUIPMENT (SHANGHAI) CO., LTD.

Head Office

Tel : 86-21-51036677 Fax : 86-21-63372237

BEIJING Branch

Tel : 86-10-64627025 Fax : 86-10-64627036

TIANJIN Branch

Tel : 86-22-26210366 Fax : 86-22-26282186

GUANGZHOU Branch

Tel : 86-20-83317826 Fax : 86-20-83317825

SHENZHEN Branch

Tel : 86-755-83674422 Fax : 86-755-83674228

SUZHOU Branch

Tel : 86-512-68028890 Fax : 86-512-68028860

ESPEC TEST TECHNOLOGY (SHANGHAI) CO., LTD.

Tel : 86-21-68798008 Fax : 86-21-68798088

ESPEC SOUTH EAST ASIA SDN.BHD.

Tel : 60-3-8945-1377 Fax : 60-3-8945-1287

TS7E26C02 (The contents of this catalog is as of July, 2014.)

Quality Management System Assessed and Registered

ESPEC CORP. has been assessed by and registered in the Quality Management System based on the International Standard ISO

9001:2008 (JIS Q 9001:2008) through the

Japanese Standards Association (JSA).

Registration : ESPEC CORP.

(Overseas subsidiaries not included)

Environmental Management System Assessed and Registered

ESPEC CORP.

● Specifications are subject to change without notice due to design improvements.

Corporate names and trade names mentioned in this catalog are trademarks or registered trademarks.

Temperature Chamber Series

CAT.NO.E94190-Q1407

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement