- No category
advertisement
http://www.espec.co.jp/english
Head Office
3-5-6, Tenjinbashi, Kita-ku, Osaka 530-8550, Japan
Tel : 81-6-6358-4741 Fax : 81-6-6358-5500
ESPEC NORTH AMERICA, INC.
Tel : 1-616-896-6100 Fax : 1-616-896-6150
ESPEC EUROPE GmbH
Tel : 49-89-1893-9630 Fax : 49-89-1893-96379
ESPEC ENVIRONMENTAL EQUIPMENT (SHANGHAI) CO., LTD.
Head Office
Tel : 86-21-51036677 Fax : 86-21-63372237
BEIJING Branch
Tel : 86-10-64627025 Fax : 86-10-64627036
TIANJIN Branch
Tel : 86-22-26210366 Fax : 86-22-26282186
GUANGZHOU Branch
Tel : 86-20-83317826 Fax : 86-20-83317825
SHENZHEN Branch
Tel : 86-755-83674422 Fax : 86-755-83674228
SUZHOU Branch
Tel : 86-512-68028890 Fax : 86-512-68028860
ESPEC TEST TECHNOLOGY (SHANGHAI) CO., LTD.
Tel : 86-21-68798008 Fax : 86-21-68798088
ESPEC SOUTH EAST ASIA SDN.BHD.
Tel : 60-3-8945-1377 Fax : 60-3-8945-1287
TS7E26C02 (The contents of this catalog is as of July, 2014.)
Quality Management System Assessed and Registered
ESPEC CORP. has been assessed by and registered in the Quality Management System based on the International Standard ISO
9001:2008 (JIS Q 9001:2008) through the
Japanese Standards Association (JSA).
*
Registration : ESPEC CORP.
(Overseas subsidiaries not included)
Environmental Management System Assessed and Registered
ESPEC CORP.
● Specifications are subject to change without notice due to design improvements.
●
Corporate names and trade names mentioned in this catalog are trademarks or registered trademarks.
Temperature Chamber Series
CAT.NO.E94190-Q1407
1
Ideal for numerous applications ranging from high-temperature tests to drying and heat processing.
The "Perfect Oven" epitomizes the features and performance of the ideal oven.
It is a versatile product, conducting high-temperature tests, but also drying and heat treatment for production lines, with unrivaled reliability and performance.
The 56 models offered by ESPEC precisely answers the various needs of our customers.
2
MODEL VARIATION
PV H
+ 200℃/+ 300℃
(+ 392
°
F/+ 572
°
F)
Temperature Chamber
(
Vertical type
)
PH H
+ 200℃/+ 300℃
(+ 392
°
F/+ 572
°
F)
Temperature Chamber
(
Horizontal type
)
STPH
+ 500℃
(+ 932
°
F)
Ultra-High Temperature Chamber
SSPH
+700℃
(+1292
°
F)
Ultra-High Temperature Chamber
OVEN SERIES FOR VARIOUS APPLICATIONS
ST H
)
DESK-TOP TYPE
HIGH-TEMP CHAMBER
+
200
℃/+
300
℃
15
L/
28
L/
39
L
VAC
VACUUM OVEN
+
200
℃
933
×
10 2 to 1
×
10 2 Pa
3
LC
CONVECTION OVEN
+
200
℃/+
250
℃
90
L/
165
L/
360
L
LCV
VACUUM OVEN
+
200
℃
0 to
−
101 kPa
(
Gauge
)
SPH H
+ 200℃/+ 300℃
(+ 392
°
F/+ 572
°
F)
Temperature Chamber with Explosion Vent
IPH H
+ 200℃/+ 300℃
(+ 392
°
F/+ 572
°
F)
Anaerobic Temperature Chamber
GPH H
+ 200℃/+ 300℃
(+ 392
°
F/+ 572
°
F)
Temperature Chamber with Rotating Specimen Rack
PV H C
CLEAN OVEN
+
200
℃/+
300
℃/+
350
℃
178
L/
380
L/
678
L
Class 5
(
H
)
LARGE VOLUME
TEMPERATURE CHAMBER
+
200
℃/
+
300
℃
2250
L/
4050
L
4
5
Control operation
Two types of program instrumentation to suit different applications.
Standard Instrumentation and M-Instrumentation.
●
Constant operation mode
●
Alarm
User-friendly Standard
Instrumentation
St a nd a rd I nst r u ment at ion feat u res programmed operation with operational settings such as constant mode and automatic start/stop. Suitable for heat treatment, drying, and similar productionline applications.
M-Instrumentation features
programs with up to 20 steps
Suitable for a range of applications from temperature-characteristics testing to heat treatment and drying. Programmed operation now allows storing ten patterns, each up to twenty steps. Provides a wide range of functions, including temperature ramp settings and a maximum of 999 repeat cycles.
Easy setup with on-screen
display
Employs interactive settings for ease of use. Text can be displayed and entered in Japanese or English alphanumeric characters.
Four optional functions
Four optional functions, namely, air flow adjuster, automatic damper, integrating hour meter, and calendar timer can be included in the instrumentation. These functions can be set by using main panel instrumentation keys.
Interface (Option)
Interface for device communication can be selected between RS-485, GPIB and
RS-232C.
Control operation
Examples of Programmed Operation
(M-Instrumentation)
Temp.
③
Step program*
①
②
④
⑤
⑥
⑦…⑳
Time
Temp.
③ ⑦…⑳
Ramp program*
①
② ④
⑤
⑥
Time
Temp.
Automatic start*
①
②
③
⑥
⑦…⑳
Time
1.HOLD
Temp.
2.CONST
Temp.
3.OFF
Temp.
End of program
( selection
)
END END END
Time
Temperature at the final step is maintained.
Time
Operation at preset constant temperature.
Time
Oven is turned OFF
*
The number of repetitions of a program can be preset between 1 and 999.
Stepwise damper setting is possible using an optional automatic damper.
Guarantee soak function, whereby the timer is used to maintain a preset temperature
for a preset length of time, can also be performed.
Temperature Indicator-controller
Instrumentation Standard Instrumentation M-Instrumentation
Operation mode
Constant operation, programmed operation and remote operation through communication interface
Setting and indication ranges
Temperature: 0 to
+
210
℃
(
+
32 to
+
410°F)
0 to
+
310
℃
(
+
32 to
+
590°F)
0 to
+
510
℃
(
+
32 to
+
950°F)
0 to
+
710
℃
(
+
32 to
+
1310°F)
Time: 0 to 9999 hours 59 minutes
Setting resolution
Temperature: 1℃
Time: 1 minute
Programming function
One-pattern, two-steps program entry is possible.
10-patterns, 20-steps program entry is possible.
Ramp setting
:
Step or ramp temperature changes possible.
OFF mode
:
The oven can be turned off during programmed operation.
Automatic start
:
Timed start-up is possible by setting
the first step to 0℃
(i.e. oven OFF).
Automatic stop
:
Timed termination is possible by setting the oven to turn OFF upon completion of a program.
End mode
:
The operating status upon completion of a program can be set to
HOLD, CONST or OFF.
Repetition
:
Up to 999 times.
Auxiliary functions
Input burnout detection
Upper and lower temperature limit alarm
Upper deviation limit temperature alarm
Buzzer alarm
Automatic overheat protection
Trouble indication
Alarm indication
Self-diagnostic
Guarantee soak
Power failure recovery selection
Power failure protection
Quick timer
Quick operation
6
PV(H)
+ 200℃/+ 300℃
TEMPERATURE CHAMBER (Vertical type)
7
Test area
*Cable port and N
2
gas injector are options.
A space-saving upright chamber
Components are incorporated into the top portion of the vertical chamber, reducing installation space by
20~60% (comparison with conventional model).
Increases productivity on the production line, and saves laboratory space.
Seamless door interior structure
Door back is a single molded structure preventing heat losses from loose joints.
Large processing capacity
Since the f loor and shelves of the chamber have b e e n g r e a t l y r e i n fo r c e d , a l a r g e a m o u n t of specimens can be loaded and processed at the same time. The sliding shelves ensure easy handling of the specimens.
Excellent heating performance
Heating performance is greatly enhanced so that the chamber temperature remains constant even if the ventilation damper is opened. ( at
+20℃ ambient temperature)
SPECIFICATIONS
Model
System
Temperature range
*2
Temperature fluctuation
*2
Temperature uniformity
*2
Temperature heat-up time
Exterior material
Interior material
Insulation material
Heater
PV
−
212 PV
−
222 PV
−
232 PV
−
332 PVH
−
212 PVH
−
222 PVH
−
232 PVH
−
332
Forced hot-air circulation / ventilation system
Ambient temp.
+
20
℃
(
+
68°F) to
+
200
℃
(
+
392°F) Ambient temp.
+
20
℃
(
+
68°F) to
+
300
℃
(
+
572°F)
±
±
1.0
0.2
℃
℃
at
at
+
+
100
100 (
+
℃
(
+
212°F),
212°F),
Ambient temp. to
±
+
2.0
200
℃
+
at
℃ within 40 min.
(
200
+
+
℃
(
200
+
℃
(
392°F)
392°F)
+
392°F)
±
0.2
℃
at
+
100
℃
(
+
212°F),
+
200
℃
(
+
392°F),
±
0.3
℃
at
+ 300℃ (+
572°F)
±
1.0
℃
at
+
100
℃
(
+
212°F),
±
2.0
℃
at
+
200
℃
(
+
392°F),
±
3.0
℃
at
+ 300℃ (+
572°F)
Ambient temp. to
+
300
℃
(
+
572°F) within 60 min.
Cold rolled rust-proof steel plate, Melamine resin coating
Stainless steel plate
Glass wool
Sheathed heater
Air circulator
Damper
Fittings
Inside dimensions
W
×
H
×
Dmm (in)
Outside dimensions
*3
W
×
H
×
Dmm (in)
Capacity (L)
Stainless steel sirocco fan
Circulation
/
Ventilation (manual switching
)
600
×
600
×
600
(23.6
×
23.6
×
23.6)
770
(30.3
×
×
1200
47.2
×
×
925
36.4)
Power cable (approx 2m from chamber), Specimen power supply control terminals
(relay contact is opened/stop during malfunction. Voltage capacity 250V AC 3A)
600
(23.6
770
(30.3
×
×
×
×
900
35.4
1500
59
×
×
×
×
600
23.6)
925
36.4)
600
(23.6
770
(30.3
×
×
×
×
1200
47.2
1800
70.9
×
×
×
×
600
23.6)
925
36.4)
800
(31.5
1030
(40.6
×
×
×
×
1200
47.2
1800
70.8
×
×
×
×
800
31.5)
1145
45.1)
600
(23.6
770
(30.3
×
×
×
×
600
23.6
1200
47.2
×
×
×
×
600
23.6)
925
36.4)
600
(23.6
770
(30.3
×
×
×
×
900
35.4
1500
59
×
×
×
×
600
23.6)
925
36.4)
600
×
1200
×
600
(23.6
×
47.2
×
23.6)
800
×
1200
×
800
(31.5
×
47.2
×
31.5)
770
×
1800
×
925
(30.3
×
70.9
×
36.4)
1030
×
1800
×
1145
(40.6
×
70.8
×
45.1)
216 324 432 768 216 324 432 768
Weight (kg)
Allowable ambient conditions ts Power supply
(Voltage fluctuation:
±
10
%
of rated value)
165 190 210 325 165 190 210 325
200
/
220
/
230
/
1
φ
50
/
60Hz
Temperature: 0 to
+
40
℃
(
+
32 to
+
104°F) Humidity: to 75
% rh
240V AC
200
/
220V AC
3
φ
3W 50
/
60Hz
─
380V AC
3
φ
4W 50Hz
230
1
200
φ
/
/
220
/
240V AC
50
/
60Hz
200
/
220V AC 3
─
φ
3W 50
3
φ
/
60Hz
380V AC
4W 50Hz
Max. power consumption
(kVA)
4.0
4.8
5.8
6.8
*1 Values assume circulatory operation with no specimens at an ambient temperature of
+
23
℃ ±
5.
*2 Conforms to Japan Testing Machinery standard K05:2000.
*3 Excluding protrusions.
4.0
5.8
6.2
8.8
Shelf pitch, quantity and load resistance
Model
Shelf pitch
Shelves
Shelf load resistance
*1 *2
Chamber total load resistance
*1
PV(H)-212 11
25kg PV(H)-222
PV(H)-232
50mm
PV(H)-332 80mm
*1 Including shelf weight
*2 Equally distributed load
17
23
14 45kg
200kg
SAFETY DEVICES
●
Leakage breaker
●
Electrical compartment door switch
●
Door switch
●
Thermal fuse
●
Temperature switch for air circulator
●
Heater wiring breaker
●
Upper and lower temperature limit alarm
(built inside temperature controller)
●
Overheat protector
●
Cartridge fuse
●
Specimen power supply control terminal
ACCESSORIES
●
Shelf (stainless steel wire)
(stainless steel plate for type 332)
●
Shelf bracket (stainless steel)
●
Cartridge fuse
●
User's manual
2
2 sets (4)
2
1 set
8
PH(H)
+ 200℃/+ 300℃
TEMPERATURE CHAMBER (Horizontal type)
9
Test area
Energy Saver Duct
PH-201
Damper open 50%
SET 200
℃
SET 100
℃
0 0.5
1
Energy save approx.
19
%
1.5
2
Standard
Energy Saver Duct
Energy save approx.
23
%
2.5
3
( kWh
)
*N
2
gas injector unit is option.
High performance chamber
A t e mp e r at u re -i nd icat ion cont rol le r w it h a n a d v a n c e d PI D o p e r a t io n , a n d a n o r ig i n a l l y d e v e l o p e d c h a m b e r c o n f i g u r a t i o n p r o v i d e u n m a t c h e d ove n p e r for m a nce. Te mp e r at u r e uniformity, temperature constancy, temperature heat-up rate, and temperature recovery time are performed with the upmost reliability.
Safety measures
Tr iple s a fe t y m e ch a n i s m s a r e e m ploye d fo r excessive overheating.
Wide model selection
We provide a total of 16 ovens with combination of temperature range, capacity, and instrumentation.
Energy Saver Duct (Option)
Energy saving approximately 20% by heat recycling f r o m e x h a u s t t h r o u g h t h e d u c t t o m a i n t a i n temperature while damper opens.
Standard
Energy Saver Duct
Standard
Energy Saver Duct
Intake air
Exhaust air
SPECIFICATIONS
Model
System
Temperature range
*2
Temperature fluctuation
*2
Temperature uniformity
*2
Temperature heat-up time
PH
−
102 PH
−
202 PH
−
302 PH
−
402 PHH
−
102 PHH
−
202 PHH
−
302 PHH
−
402
Forced hot-air circulation / ventilation system
Ambient temp.
+
20
℃
(
+
68°F) to
+
200
℃
(
+
392°F) Ambient temp.
+
20
℃
(
+
68°F) to
+
300
℃
(
+
572°F)
±
±
±
±
0.1
0.2
0.5
1.5
℃
℃
℃
℃
at
at
at
at
+
+
+
+
100
200
100
200
℃
℃
℃
℃
(
(
(
(
+
+
+
+
212°F)
392°F)
212°F)
392°F)
Ambient temp. to
+
±
±
±
±
0.2
0.4
1.0
2.0
200
℃
℃
℃
℃
℃
at
at
at
at
(
+
+
+
+
+
100
200
100
200
℃
℃
℃
℃
392°F)
(
(
(
(
+
+
+
+
212°F)
392°F)
212°F)
392°F)
±
0.1
℃
at
+
100
℃
(
+
212°F)
±
0.2
℃
at
+
200
℃
(
+
392°F)
+
300
℃
(
+
572°F)
±
0.5
℃
at
+
100
℃
(
+
212°F)
±
1.5
℃
at
+
200
℃
(
+
392°F)
±
2.5
℃
at
+
300
℃
(
+
572°F)
±
±
±
±
±
±
0.2
0.4
0.6
1.0
2.0
3.0
℃
℃
℃
℃
℃
℃
at
at
at
at
at
at
+
+
+
+
+
+
100
200
300
100
200
300
℃
℃
℃
℃
℃
℃
Ambient temp. to
+
300
℃
(
+
572°F)
(
(
(
(
(
(
+
+
+
+
+
+
212°F)
392°F)
572°F)
212°F)
392°F)
572°F) within 40 min.
within 60 min.
within 60 min.
within 70 min.
Exterior material
Interior material
Insulation material
Heater
Cold rolled rust-proof steel plate, Melamine resin coating
Stainless steel plate
Glass wool
Iron chrome strip wire heater
Air circulator
Damper
Fittings
Inside dimensions
W
W
×
×
H
H
×
×
Dmm (in)
Dmm (in)
Outside dimensions
*3
Stainless steel propeller fan
Circulation
/
Ventilation (manual switching)
450
×
450
×
450
(17.7
×
17.7
×
17.7)
1040
(41
×
×
820
32.3
×
×
635
25)
Power cable (approx 2m from chamber), Specimen power supply control terminals
(relay contact is opened/stop during malfunction. Voltage capacity 250V AC 3A)
600
(23.6
1190
(46.9
×
×
×
×
600
23.6
970
38.2
×
×
×
×
600
23.6)
785
30.9)
800
(31.5
1500
(59.1
×
×
×
×
800
31.5
1210
47.6
×
×
×
×
800
31.5)
1065
41.9)
1000
(39.4
1730
(68.1
×
×
×
×
1000
39.4
1480
58.3
×
×
×
×
1000
39.4)
1275
50.2)
450
×
450
×
450
(17.7
×
17.7
×
17.7)
600
×
600
×
600
(23.6
×
23.6
×
23.6)
1040
×
820
×
635
(41
×
32.3
×
25)
1190
×
970
×
785
(46.9
×
38.2
×
30.9)
800
(31.5
1500
(59.1
×
×
×
×
800
31.5
1210
47.6
×
×
×
×
800
31.5)
1065
41.9)
1000
×
1000
×
1000
(39.4
×
39.4
×
39.4)
1730
×
1480
×
1275
(68.1
×
58.3
×
50.2)
91 216 512 1000 91 216 512 1000 Capacity (L)
Weight (kg)
Allowable ambient conditions ts Power supply
(Voltage fluctuation:
±
10
%
of rated value)
95 130
Temperature: 0 to
200
/
220
/
230
/
240V AC
1
φ
50
/
60Hz
240
3
200
φ
/
220V AC
3W 50
380V AC 3
+
φ
40
/
430
℃
(
+
60Hz,
32 to
4W 50Hz
+
200
/
95
220
1
φ
/
230
50
/
/
130
104°F) Humidity: to 75
240V AC
60Hz
% rh
240
3
200
φ
/
430
220V AC
3W 50
380V AC 3
φ
/
60Hz,
4W 50Hz
Max. power consumption
(kVA)
2.0
2.7
5.0
6.5
*1 Values assume circulatory operation with no specimens at an ambient temperature of
+
23
℃ ±
5.
*2 Conforms to Japan Testing Machinery standard K05:2000.
*3 Excluding protrusions.
2.7
3.8
6.5
9.5
Shelf pitch, quantity and load resistance
Model
Shelf pitch
Shelves
Shelf load resistance
*1 *2
Chamber total load resistance
*1
SAFETY DEVICES
PH(H)-102
PH(H)-202
50mm
PH(H)-302 80mm
PH(H)-402 140mm
*1 Including shelf weight
*2 Equally distributed load
8
11
9
6
20kg
40kg
50kg
60kg
100kg
●
Leakage breaker
●
Electrical compartment door switch
●
Door switch (type 402 only)
●
Thermal fuse
●
Temperature switch for air circulator (except type 402)
●
Air circulator overload relay (type 402 only)
●
Heater wiring breaker
●
Reverse-prevention relay
●
Upper and lower temperature limit alarm
ACCESSORIES
(built inside temperature controller)
●
Overheat protector
●
Shelf (stainless steel wire for type102
●
Shelf bracket (stainless steel)
・
202) 2
2 sets (4)
●
●
Cartridge fuse
(stainless steel punched plate for type 302
・
402) 2
Specimen power supply control terminal
●
Cartridge fuse
●
User's manual
2
1 set
10
STPH
+ 500℃
ULTRA-HIGH TEMPERATURE CHAMBER
Temperature control to
+
500
℃
Effective temperature range of (ambient temp.
+) 20℃ to +500℃ . The chamber can be used for a variety of applications, including tests of viability under hightemperatures and temperature resistance.
Door equipped with a single-action
lever
The door can be firmly locked by an easy-to-use single-action lever. It prevents accidents from unlocked doors.
SPECIFICATIONS
11
Model
System
Temp. range
*2
Temp. fluctuation
*2
Temp.
uniformity
Temp.
Interior
Insulation
Heater
*2 heat-up time
Air circulator
STPH-102 STPH-202
Forced hot-air circulation / ventilation system
Ambient temp.
+
20
℃
(
+
68°F) to
+
500
℃
(
+
932°F)
±
0.5
℃
±
0.8
℃
at
+
100
℃
(
+
212°F)
±
1.8
℃
at
+
200
℃
(
+
392°F)
±
2.8
℃
at
+
300
℃
(
+
572°F)
±
3.8
℃
at
+
400
℃
(
+
752°F)
±
4.8
℃
at
+
500
℃
(
+
932°F)
Ambient temp. to
+
500
℃
(
+
932°F) within 60min.
Stainless steel plate
Glass wool, MG wool
Iron chrome strip wire heater
Stainless steel propeller fan
Damper
Fittings
Inside dimensions
W
×
H
×
Dmm (in)
Outside dimensions
W
×
H
×
Dmm (in)
*3
Circulation/ Ventilation (manual switching)
Power cable (approx 2m from chamber),
Specimen power supply control terminals
Electrical compartment cooling fan
450
×
450
×
450
(17.7
×
17.7
×
17.7)
1190
×
1110
×
795
(46.9
×
43.7
×
31.3)
600
(23.6
1340
(52.8
×
×
×
×
600
23.6
49.6
×
×
1260
×
600
23.6)
×
945
37.2)
Capacity (L)
Weight (kg)
Allowable ambient conditions
Power supply
(
±
10
%
of rated value)
Max. power consumption
91
190
Temp.: 0 to
6.5 kVA
+
40
℃
Humid.: to 75
% rh
380V AC 3
(
200 / 220V AC 3
φ
+
φ
32 to
216
250
+
104°F)
50/60Hz,
4W 50Hz
8.3 kVA
*1 Values assume circulatory operation with no specimens at an ambient temperature of
+
23
℃ ±
5.
*2 Conforms to Japan Testing Machinery standard K05:2000.
*3 Excluding protrusions.
ACCESSORIES
●
Shelf (stainless steel wire)
●
Shelf bracket (stainless steel)
●
Cartridge fuse
●
User's manual
SAFETY DEVICES
●
Leakage breaker
●
Electrical compartment door switch
●
Thermal fuse
●
Temperature switch for air circulator
●
Electrical compartment thermal switch
●
Heater wiring breaker
●
Upper and lower temperature limit alarm
(built inside temperature controller)
●
Overheat protector
●
Cartridge fuse
●
Specimen power supply control terminal
2
2 sets (4)
2
1 set
SSPH
Saving-energy insulated structure
Ceramic f iber and aluminium foil are used as insulation materials. It increases effective insulation and prevents heat loss, thus saving energy.
A Double seal gasket configuration
A gasket made of stainless steel fiber and a leaf spring are used to form a double seal between the door and the chamber. Prevents heat radiation on door.
Door equipped with a single-action lever
The door can be firmly locked by an easy-to-use single-action lever.
+ 700℃
ULTRA-HIGH TEMPERATURE CHAMBER
ACCESSORIES
●
Shelf (stainless steel wire)
●
Shelf bracket (stainless steel)
●
Cartridge fuse
●
User's manual
SAFETY DEVICES
●
Leakage breaker
●
Electrical compartment door switch
●
Thermal fuse
●
Temperature switch for air circulator
●
Air circulator rotation detector
●
Electrical compartment thermal switch
●
Heater wiring breaker
●
Upper and lower temperature limit alarm
(built inside temperature controller)
●
Overheat protector
●
Cartridge fuse
●
Specimen power supply control terminal
2
2 sets (4)
2
1 set
SPECIFICATIONS
Model
System
Temp. range
*2
Temp.
fluctuation
Temp.
uniformity
Temp.
*2
*2 heat-up time
SSPH-102 SSPH-202
Forced hot-air circulation / ventilation system
+
100 to
+
700
℃
(
+
212 to
+
1292°F)
±
0.5
℃
at
+
100 to
+
500
℃
(
+
212 to
+
932°F)
±
0.8
℃
at
+
501 to
+
700
℃
(
+
933 to
+
1292°F)
±
0.8
℃
at
+
100
℃
(
+
212°F)
±
2.8
℃
at
+
300
℃
(
+
572°F)
±
4.8
℃
at
+
500
℃
(
+
932°F)
±
7.0
℃
at
+
700
℃
(
+
1292°F)
Ambient temp. to
+
700
℃
(
+
1292°F) within 120min.
within 160min.
Stainless steel plate
Glass wool, Ceramic fiber
Iron chrome strip wire heater
Stainless steel propeller fan
Interior
Insulation
Heater
Air circulator
Damper
Fittings
Inside dimensions
W
×
H
×
Dmm (in)
Outside dimensions
W
×
H
×
Dmm (in)
*3
Capacity (L)
Circulation/ Ventilation (manual switching)
Power cable (approx 2m from chamber),
Specimen power supply control terminals
Electrical compartment cooling fan
450
×
450
×
450
(17.7
×
17.7
×
17.7)
1190
×
1110
×
795
(46.9
×
43.7
×
31.3)
91
600
(23.6
1340
(52.8
×
×
×
×
600
23.6
49.6
216
×
×
1260
×
600
23.6)
×
945
37.2)
Weight (kg)
Allowable ambient conditions
Power supply
(
±
10
%
of rated value)
250 330
Temp.: 0 to
+
40
℃
(
+
32 to
+
104°F)
Humid.: to 75
% rh
200
/
220V AC 3
φ
50
/
60Hz,
380V AC 3
φ
4W 50Hz
Max. power consumption
8.3 kVA 9.5 kVA
*1 Values assume circulatory operation with no specimens at an ambient temperature of
+
23
℃ ±
5.
*2 Conforms to Japan Testing Machinery standard K05:2000.
*3 Excluding protrusions.
12
SPH(H)
+ 200℃/+ 300℃
TEMPERATURE CHAMBER WITH EXPLOSION VENT
Temperature chamber
with Explosion Vent
This temperature chamber is suitable for drying and heat-treatment of f lammable synthetic resins or volatile solvents. It is equipped with an explosion vent which releases explosion and a safety door to ensure security.
Door equipped
with a single-action lever
The door can be securely locked by an easy-to-use single-action lever. Even if the operator accidentally turns on the power when door is unlocked, the door lock detection switch prevents heater fan from starting. Besides, in three minutes, the alarm buzzer sounds to call for warning.
●
Release explosion safely
Aluminium plate
Insulation material
Flame
Specimen
Explosion
In case an explosion occurs inside the test chamber, as shown in the above image, insulation material is bent and blown upward together with the aluminium plate to the metal screen at the top of the chamber.
This way the explosion is safely channeled and released through the top metal screen. For SPH(H)-402, explosion is released through the top metal screen by bending insulation material on the rear wall.
■
Release explosion safely
13
In the event that an explosion occure inside the test chamber, as shown in the below figures, insulation material is bent and blown upward together with the aluminium plate to the metal screen at the top of the chamber.
In this way the explosion is safely channeled and released through the top metal screen.
Aluminium plate
Insulation material
Flame
WARNING
1) The following flammables or materials containing them can be subjected to drying (heat treatment) with this chamber. However, to avoid explosion, ventilate the chamber well and use the chamber below the explosive limit.
注意
Inflammables:
●
Ignitable Substances
1. Ethyl ether, gasoline, acetaldehyde, propylene oxide, c a r b o n d i s u l f i d e , c a r b o n d i o x i d e a n d o t h e r substances with an ignition point of below
−
30
℃ .
2. Normal hexane, ethylene oxide, acetone, benzene, methyl ethyl ketone and other substances with an ignition point above
−
30
℃
and below 0
℃
.
3 . M e t h a n o l , e t h a n o l , x y l e n e , p e n t y l a c e t a t e amylacetate and other substances with an ignition point above 0
℃ and below
+
30
℃
.
4. Kerosene, light oil, turpentine oil, isopentyl alcohol
(also called isoamyl alcohol), acetic acid and other substances with an ignition point above
+
30
℃ and below
+
65
℃
.
●
Combustible Gases
Hydrogen, acetylene, ethylene, methance, ethane, propane, butane, and other combustible substances that are in a gaseous state at a temperature of
+
15
℃ and at a pressure of 1 atmosphere.
2) Temperature chamber with explosion vent is fitted with a comprehensive range of devices to ensure safety. In addition to the regular inspection, these must be carefully inspected before reusing after an explosion.
3) This equipment is designed to prevent any damage to people or equipment in the vicinity for explosion pressures not exceeding 29.4kPa. If the explosion pressure exceeds 9.8kPa, reuse of the equipment itself may not be possible.
4) Please refer to the instruction manual before using the chamber to ensure safe operation.
Specimen
Explosion
SPECIFICATIONS
Model
System
Temperature range
*2
Temperature fluctuation
*2
Temperature uniformity
*2
Temperature heat-up time
SPH
−
102 SPH
−
202 SPH
−
302 SPH
−
402 SPHH
−
102 SPHH
−
202 SPHH
−
302 SPHH
−
402
Forced hot-air circulation / ventilation system
Ambient temp.
+
20
℃
(
+
68°F) to
+
200
℃
(
+
392°F) Ambient temp.
+
20
℃
(
+
68°F) to
+
300
℃
(
+
572°F)
±
±
±
±
0.1℃ at +
0.2℃ at +
0.5℃ at +
1.5℃ at +
100
200
100
200
℃
℃
℃
℃
(
(
(
(
+
+
+
+
212°F)
392°F)
212°F)
392°F)
Ambient temp. to
+
±
±
±
±
0.2
0.4
1.0
2.0
200
℃
℃
℃
℃
℃
at
at
at
at
(
+
+
+
+
+
100
200
100
200
℃
℃
℃
℃
392°F)
(
(
(
(
+
+
+
+
212°F)
392°F)
212°F)
392°F)
±
±
±
±
±
±
0.1
0.2
0.2
0.5
1.5
2.5
℃
℃
℃
℃
℃
℃
at
at
at
at
at
at
+
+
+
+
+
+
100
200
300
100
200
300
℃
℃
℃
℃
℃
℃
(
(
(
(
(
(
+
+
+
+
+
+
212°F)
392°F)
572°F)
212°F)
392°F)
572°F)
± 0.2℃ at +
100
℃
(
+
212°F)
± 0.4℃ at +
200
℃
(
+
392°F)
± 0.6℃ at +
300
℃
(
+
572°F)
±
1.0℃ at +
100
℃
(
+
212°F)
± 2.0℃ at +
200
℃
(
+
392°F)
±
3.0℃ at +
300
℃
(
+
572°F)
Ambient temp. to
+
300
℃
(
+
572°F) within 40 min.
within 60 min.
within 60 min.
within 70 min.
Exterior material
Interior material
Insulation material
Explosion vent
Heater
Cold rolled rust-proof steel plate, Melamine resin coating
Stainless steel plate
Glass wool
Safety vent to release inside pressure on explosion,
Explosion exhaust duct, Protective wire mesh, Insulation, Outer plate
Stainless steel, Sheated heater with fin
Stainless steel propeller fan
Circulation
/
Ventilation (manual switching)
Air circulator
Damper
Fittings
Inside dimensions
W
×
H
×
Dmm (in)
Outside dimensions
*3
W
×
H
×
Dmm (in)
Capacity (L)
450
×
450
×
450
(17.7
×
17.7
×
17.7)
1040
(41
×
×
1260
49.6
×
×
635
25)
Power cable (approx 2m from chamber), Specimen power supply control terminals
(relay contact is opened/stop during malfunction. Voltage capacity 250V AC 3A).
600
(23.6
1190
(46.9
×
×
×
×
600
23.6
×
1370
53.9
×
×
×
600
23.6)
785
30.9)
800
(31.5
1500
(59.1
×
×
×
×
800
31.5
1715
68.1
×
×
×
×
800
31.5)
1065
41.9)
1000
(39.4
1730
(68.1
×
×
×
×
1000
39.4
1800
70.9
×
×
×
×
1000
39.4)
1775
69.9)
450
×
450
×
450
(17.7
×
17.7
×
17.7)
600
×
600
×
600
(23.6
×
23.6
×
23.6)
1040
×
1260
×
635
(41
×
49.6
×
25)
1190
×
1370
×
785
(46.9
×
53.9
×
30.9)
800
(31.5
1500
(59.1
×
×
×
×
800
31.5
1715
68.1
×
×
×
×
800
31.5)
1065
41.9)
1000
×
1000
×
1000
(39.4
×
39.4
×
39.4)
1730
×
1800
×
1775
(68.1
×
70.9
×
69.9)
91 216 512 1000 91 216 512 1000
Weight (kg)
Allowable ambient conditions ts Power supply
(Voltage fluctuation:
±
10
%
of rated value)
95 130 270 500 95 130 270 500
200
/
220
/
230
/
Temperature: 0 to
+
40
℃
(
+
32 to
+
104°F) Humidity: to 75
% rh
240V AC
1
φ
50
/
60Hz
200
/
220V AC
3
φ
3W 50
/
60Hz,
380V AC 3
φ
4W 50Hz
200
/
220
1
φ
/
230
50
/
/
240V AC
60Hz
200
/
220V AC
3
φ
3W 50
/
60Hz,
380V AC 3
φ
4W 50Hz
Max. power consumption
(kVA)
2.0
2.7
5.0
6.5
*1 Values assume circulatory operation with no specimens at an ambient temperature of
+
23
℃ ±
5.
*2 Conforms to Japan Testing Machinery standard K05:2000.
*3 Excluding protrusions.
2.7
3.8
6.5
9.5
ACCESSORIES SAFETY DEVICES
●
Shelf (stainless steel wire for type 102
・
202) 2
(stainless steel punched plate for type 302
・
402) 2
●
●
Leakage breaker
Electrical compartment door switch
●
Shelf bracket (stainless steel) 2 sets (4)
●
Chamber door lock detection switch
●
Cartridge fuse
●
Protective wire mesh
2
1
●
●
Explosion detection limit switch
Thermal fuse
(stainless steel mesh with soft aluminium foil)
●
Temperature switch for air circulator (except type 402)
●
Insulation (glass wool)
●
Outer plate (thin soft aluminium panel)
●
Stand bracket and hexagon socket head cap screw
(for type 102
・
202)
3
1
4 each
●
●
●
Air circulator overload relay (for type 402 only)
Heater wiring breaker
Reverse-prevention relay (for type 402 only)
●
Hexagon socket screw key (for type 102
・
202) 1
●
User's manual 1 set
●
Upper and lower temperature limit alarm
(built inside temperature controller)
●
Overheat protector
●
Cartridge fuse
●
Specimen power supply control terminal
14
IPH(H)
+ 200℃/+ 300℃
ANAEROBIC TEMPERATURE CHAMBER
Low oxygen level testing
Equipped with a non-oxidizing gas intake structure which fills the chamber with non-oxidizing gas such as CO
2
or N
2
for heat treatment or temperature cha r a cter ist ics t est i ng requ i r i ng low oxygen concentration atmosphere.
Hermetically sealed configuration
The chamber is hermetically sealed to decrease oxygen inside the chamber. The inner stainless steel plate is seamless welded with argon gas.
O
2
concentration indicator controller (optional)
An optional O
2
concentration indicator controller equipped with an oxygen sensor is available. It allows precise regulation of the O
2
level throughout the range 0.5 to 21% (using N
2
).
SPECIFICATIONS
Model
System
Temp. range
Temp. fluctuation
Temp. uniformity
*2
*2
*2
IPH
−
202 IPHH
−
202
±
±
±
±
0.1
0.2
0.5
1.5
℃
Forced hot-air circulation system
Ambient temp.
+
20
℃
(
+
68°F) to
+
200
℃
(
+
392°F)
℃
℃
℃
at
at
at
at
+
+
+
+
100
200
100
200
℃
℃
℃
℃
(
Ambient temp. to
(
+
(
(
(
+
+
+
+
+
212°F)
392°F)
212°F)
392°F)
200
℃
392°F) within 40min.
Ambient temp. to
+
300
℃
+
(
20
+
℃
(
+
572°F)
68°F)
±
0.1
℃
at
+
100
℃
(
+
212°F)
±
0.2
℃
at
+
200
℃
(
+
392°F)
±
0.2
℃
at
+
300
℃
(
+
572°F)
±
0.1
℃
at
+
100
℃
(
+
212°F)
±
0.2
℃
at
+
200
℃
(
+
392°F)
±
0.2
℃
at
+
300
℃
(
+
572°F)
Ambient temp. to
+
300
℃
(
+
572°F) within 60min.
CO
2
, N
2
gas (ordinary temperature, dry gas)
Temp. heat-up time
Fluid
Fluid pressure
Flow rate
Chamber O
2
level
Allowed max. pressure: 2.0MPa (Gauge) (primary side of valve)
Secondary side is adjusted with the valve to 0.05MPa (Gauge)
Max. flow rate: 20 L /
min. (0.05MPa (Gauge), 20
℃
)
0.5
% ( lowest
)
Chamber injector pressure
29Pa (Gauge) and over (at max flow rate)
Valve
Pressure gauge
Flow meter
Scale range
1/4" brass needle valve
φ
75mm embedded type class 2.5
Scale range: 0
〜
0.1MPa (Gauge)
Floating type (provided with needle valve for flow rate control)
0 to 30L
/
min. N
2
gas
Safety valve
Gas inlet
Trip pressure: 2.0kPa (Gauge)
1
/
4" ring joint
Fittings
Power cable (approx 2m from chamber),
Specimen power supply control terminal
Inside dimensions (in) W600mm
×H
600mm
×D
600mm (23.6
×
23.6
×
23.6)
Outside dimensions (in)
*3
W1190mm
×H
970mm
×D
785mm (46.9
×
38.2
×
30.9)
Capacity (L) 216
15
Weight (kg) 130
Allowable ambient conditions Temp.: 0 to
+
40
℃
(
+
32 to
+
104°F) Humid.: to 75
% rh
Power supply
(
±
10
%
of rated value)
200
/
220
/
230
/
240V AC 1
φ
50
/
60Hz
Max. power consumption 2.7 kVA 3.8 kVA
*1 Values assume circulatory operation with no specimens at an ambient temperature of
+
23
℃ ±
5.
*2 Conforms to Japan Testing Machinery standard K05:2000.
*3 Excluding protrusions.
ACCESSORIES
●
Shelf (stainless steel wire)
●
Shelf bracket (stainless steel)
●
Cartridge fuse
●
User's manual
SAFETY DEVICES
●
Leakage breaker
●
Electrical compartment door switch
●
Thermal fuse
●
Temperature switch for air circulator
●
Heater wiring breaker
●
Upper and lower temperature limit alarm
(built inside temperature controller)
●
Overheat protector
●
Cartridge fuse
●
Specimen power supply control terminal
2
2 sets (4)
2
1 set
GPH(H)
+ 200℃/+ 300℃
TEMPERATURE CHAMBER WITH ROTATING SPECIMEN RACK
Suitable for heat deterioration test
Based on the PH Temperature Chambers, these models incorporate a detachable rotating specimen rack and is especially designed for heat deterioration testing of rubbers and plastics including polyesters and vinyls.
Equipped with a rotating specimen rack
The rack drive unit is installed inside, enhancing function and lending them a simple appearance.
By removing the rack, this equipment may also be operated as an ordinary temperature chamber.
SPECIFICATIONS
Model
System
Temp.range
*2
Temp. fluctuation
Temp.uniformity
*2
*2
Temp.heat-up time
GPH−102 GPH−202 GPHH−102 GPHH−202
Forced hot-air circulation / ventilation system
Ambient temp.
+
20
℃
(
+
68°F) to
+
200
℃
(
+
392°F)
±
±
±
±
0.1
0.2
0.5
1.5
℃
℃
℃
℃
at
at
at
at
+
+
+
+
100
200
100
200
℃
℃
℃
℃
(
(
(
(
+
+
+
+
212°F)
392°F)
212°F)
392°F)
Ambient temp. to
+
300
℃
+
(
20
+
℃
(
+
572°F)
68°F)
±
0.1
℃
at
+
100
℃
(
+
212°F)
±
0.2
℃
at
+
200
℃
(
+
392°F)
±
0.2
℃
at
+
300
℃
(
+
572°F)
±
0.5
℃
at
+
100
℃
(
+
212°F)
±
1.5
℃
at
+
200
℃
(
+
392°F)
±
2.5
℃
at
+
300
℃
(
+
572°F)
Ambient temp. to
+
200
℃
(
+
392°F) within 40min.
Ambient temp. to
+
300
℃
(
+
572°F) within 60min.
1 2 1 2
320mm (12.6in.)
Number of racks
Outside diameter
Available numbers of specimens/weight
56pcs per rack (up to 0.7kg)
Specimen clip
RPM of specimen rack
Motor
W
×
Hmm (in)
Construction
Chamber lamp
50pcs per rack
5rpm
/
50Hz, 6rpm
/
60Hz
1φ 15W
190
×
340(7.48
×
13.39) ————
Heat-resistant reinforced glass
3-plate sets
5.5W
incandescent lamp
————
————
Fittings
Power cable (approx 2m from chamber),
Specimen power supply control terminals
(relay contact is opened during malfunction.
Voltage capacity 250V AC 3A)
Inside dimensions
W
×
H
×
Dmm (in)
Outside dimensions
*3
W
×
H
×
Dmm (in)
450
×
450
×
450
(17.7
×
17.7
×
17.7)
600
×
600
×
600
(23.6
×
23.6
×
23.6)
450
×
450
×
450
(17.7
×
17.7
×
17.7)
600
×
600
×
600
(23.6
×
23.6
×
23.6)
1040
(91
×
×
820
32.3
×
×
635
25
)
1190
(46.9
×
×
970
38.2
×
×
785
30.9)
1040
(91
×
×
820
32.3
×
×
635
25)
1190
(46.9
×
×
970
38.2
×
×
785
30.9)
Capacity (L) 91 216 91 216
Weight (kg) 95 130 95 130
Allowable ambient conditions Temp.: 0 to
+
40
℃
(
+
32 to
+
104°F) Humid.: to 75
% rh ts Power supply
(
±
10
%
of rated value)
200
/
220
/
230
50
/
/
240V AC 1
60Hz
φ
Max. power consumption 2.0 kVA 2.7 kVA 3.8 kVA
*1 Values assume circulatory operation with no specimens at an ambient temperature of
+
23
℃ ±
5.
*2 Conforms to Japan Testing Machinery standard K05:2000.
*3 Excluding protrusions.
Test area
ACCESSORIES
●
Shelf (stainless steel wire)
●
Shelf bracket (stainless steel)
2
2 sets (4)
●
Cartridge fuse 2
●
Specimen clip type102 50 type202 100
●
Shaft insulation filters
●
User's manual
1 set
1 set
SAFETY DEVICES
●
Leakage breaker
●
Electrical compartment door switch
●
Thermal fuse
●
Temperature switch for air circulator
●
Heater wiring breaker
●
Upper and lower temperature limit alarm
(built inside temperature controller)
●
Overheat protector
●
Cartridge fuse
●
Specimen power supply control terminal
16
17
OPTIONS
Model
PV
PVH
PH
PHH
STPH SSPH
SPH
SPHH
IPH
IPHH
GPH
GPHH
212 222 232 332 102 202 302 402 102 202 102 202 102 202 302 402 202 102 202
Option
Time run-out output
Calendar timer
Integrating hour meter
● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●
● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●
● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●
Temperature recorder terminal
● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●
Paperless recorder/
Temperature recorder
● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●
Recorder wiring
Automatic damper
Exhaust port flange
● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
—
—
●
●
●
●
Exhaust duct
Energy Saver Duct
● ● ● ● ● ● ● — ● ● ● ● ● ● ● — — ● ●
— — — — ● ● — — — — — — ● ● — — — ● ●
Nitrogen gas injector
● ● ● ● ● ● ● ● ● ● ● ● — — ● ● — — —
Inert specification
350℃ Specification
— — — — — — — — ● ● — — — — — — — — —
— — — — —
●
—
●
—
●
—
●
— — — — — —
●
—
●
— — — —
O
2
concentration indicator-controller
— — — — — — — — — — — — — — — — ● — —
Air flow adjuster ● ● ● ● ● ● ● ● — — — — — — — — ● ● ●
Fin heater
— — — — ● ● ● ● — — — — ※ ※ ※ ※ ● ● ●
Shelf and shelf bracket
18-8 Cr-Ni stainless steel wire
18-8 Cr-Ni punched stainless steel shelf
●
—
●
—
●
—
—
●
●
●
●
●
—
●
—
●
●
—
●
—
●
—
●
—
●
●
●
●
—
●
—
●
●
●
●
●
●
●
Mesh shelf
Heavyduty shelf
Vertical type
● ● ● ● — — — — — — — — — — — — — — —
● ● ● ● — — — — — — — — — — — — — — —
Horizontal type (25kg)
— — — — ● ● — — — — — — ● ● — — ● ● ●
Horizontal type (60kg)
— — — — — ● ● — — — — — — ● ● — ● — ●
Cable port
Cable port rubber plug
Casters
Viewing window
Chamber lamp
Anchoring fixtures
Floor reinforcement
● ● ● ● ● ● ● ● ● ● — — — — — — — ● ●
●
—
●
—
●
—
●
—
●
—
●
—
●
—
●
— —
— — — — — — — —
●
—
●
—
— ● ● ● ● ● — — ● ● ● ● — — — — ● ● ●
●
—
—
●
—
—
●
—
—
●
—
—
●
●
—
—
●
●
—
—
●
●
—
—
●
●
— —
— — — — — — — — — —
— —
— — — — — — — — — —
● ● ● ● ● ● ● — — — — — ● ● ● — ● ● ●
— — — — — ● ● ● — — — — — ● ● ● ● — ●
Stand
Vertical type
Horizontal type
Angle type stand
Casters for stand
● ● — — — — — — — — — — — — — — — — —
— — — — ● ● ● — ● ● ● ● ● ● ● — ● ● ●
— — — — — — — ● — — — — — — — ● — — —
— — — — ● ● — — — — — — — — — — ● ● ●
Stacking brackets
— — — — ● ● — — — — — — — — — — ● — —
L-type-stand and stacking brackets
— — — — ● ● — — — — — — — — — — ● — —
External alarm terminal
● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●
Emergency stop pushbutton
Color specification
Interface
Power cable
※
Standard specification
● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●
● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●
● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●
● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ● ●
OPTIONS
Time run-out output
This option enables turning the power to the specimen ON or OFF with contact signal output when the time is up by using the timer function on the controller.
Power supply rating: 250VAC 1A
Actuation: Contact close when program time overflows
Where located: Right side of chamber
Temperature recorder terminal
Outputs chamber temperature through thermocouple type K (JIS C 1602)
(Thermocouple type N for STPH, SSPH)
Where located:
Rear of electrical compartment
Temperature recorder
Temp. range: 0 to
+200℃
0 to
+300℃
0 to
+600℃
0 to
+800℃
Recording system:
Pen recorder (1 pen) or multi-point recorder (6 dots)
* If performing simultaneous installation of a recorder and an N
2
gas injector, they must be handled individually. Otherwise, installation may be limited depending on what other options are chosen.
Calendar timer
Automatically starts and stops chamber operation.
Setting range:
Sunday to Saturday
(Possible to set multiple days)
0:00 to 23:59
(Setting resolution 1 minute)
Margin of error per month:
±1 minute
Paperless recorder
Records temperature of each section such as the temperature inside the chamber.
Temp. range: 0
〜+200℃
0
〜+300℃
0
〜+600℃
0
〜+800℃
Number of inputs: Temperture 1
(5 more channels can be turned ON)
Data saving cycle: 5 sec
External recording media:
CF memory card port
(Includes a 256MB CFcard)
Language support: ENG, JPN
Temperature recorder wiring
Preparation of a power cable, temperature sensor and a grounding wire for additional installation in the future.
Automatic damper
A u t o m a t i c a l l y o p e n s o r c l o s e s synchronously with program operation for ventilation and faster cooling of chamber temperature.
Integrating hour meter
Displays cumulative chamber operation time.
Available with or without reset feature.
* Operating time is not accumulated when operation is stopped due to malfunction or for other reasons.
Measuring time: 999,999 hr
●
Please refer to chart on p.17 for applicable models.
18
Energy Saver Duct
PH-201
Damper open 50%
SET 200
℃
SET 100
℃
0 0.5
1
Energy save approx.
19
%
1.5
2
Standard
Energy Saver Duct
Energy save approx.
23
%
2.5
3
( kWh
)
OPTIONS
Exhaust port flange
Flange connects an exhaust duct to the chamber to exhaust hot air from the chamber.
(for oven with damper.)
Material: Cold rolled steel plate with chromate conversion coatings
Stainless steel sheet
STPH-102, 202
SSPH-102, 202
Dimensions: External diameter 87mm
Location: Chamber rear side
* When connecting to exhaust duct, the length of duct must be less than 4m.
Energy Saver Duct
Discharges exhaust air towards to ceiling with heat exchanger for energy
Dimensions of exhaust port flange:
External diameter 87mm
* The height of chamber will be 30mm higher than standard unit by the additional lower flame.
Intake air
Exhaust air
350
℃
specification
Ad apted to prov ide a ma x i mu m temperature of 350
℃.
* PHH only.
O
2
concentration indicator-controller
Controls oxygen concentration inside the oven.
O
2
concentration range:
0.5 to 21% oxygen concentration (v/v)
Gas: N
2
gas
(ordinary temperature dry gas)
* IPH(H) only.
Exhaust duct
Discharges hot air towards the ceiling.
(for oven with damper.)
Dimensions: External diameter 87mm
Location: Chamber rear side
* Exhaust port f lange is provided at end of exhaust duct.
Nitrogen gas injector
Used for reducing specimen oxidation.
Fluid pressure: Max. allowable pressure
2.0MPa (Gauge) on primary side of valve
0.05MPa(Gauge)on secondary side using valve.
Max flow rate: 30 L min.
Flow meter: Float type flow meter
* If performing simultaneous installation of a recorder and an N
2
gas introducing unit, they must be handled individually. Otherwise, installation may be limited depending on what other options are chosen.
Air flow adjuster
Allows low air velocity in chamber
PV(H) 0.3 to 2.3m/s
PH(H)-102/202
GPH(H)-102/202 0.2 to 2.3m/s
IPH(H)-202
PH(H)-302
PH(H)-402
0.3 to 2.3m/s
0.3 to 2.6m/s
Average wind velocity across chamber central longitudinal section. Represents the typical mean value for each chamber.
19
●
Please refer to chart on p.17 for applicable models.
Inert specification
Used to minimize the oxidation of specimens.
* STPH only.
* Standard dampers are not fitted.
Fin heater
Used when anti-corrosive is required.
Stainless steel sheathed heater with fins.
OPTIONS
Shelf and shelf bracket
Equivalent to standard accessory.
PH(H)-102/202, SPH(H)-102/202,
GPH(H), and IPH(H) include stainless steel punched plate that differs from the standard shelf provided.
Heavy-duty shelf
Used to hold heavy specimen exceeding the load capacity of the standard shelf.
Additional cable port
A through hole provided on the wall of chamber.
Material: Stainless steel plate
Inside diameter: 25, 50, 100mm
(
φ50mm for STPH-102・202)
* The cable port may not be able to be used at the same time as the optional exhaust duct. (Except when used with PV(H))
* If several cable ports are installed, the surface temperature may rise or the chamber may not be able to meet standard performance.
Stainless steel wire
Stainless steel punched plate
Mesh shelf
For testing small specimens.
Material: 18-8 Cr-Ni stainless steel
φ0.8, 5 mesh
* To use, place this shelf on a standard shelf.
<
Vertical type
>
Material: 18-8 Cr-Ni stainless steel wire
Shelf support load resistance: Max 200kg
Model
PV(H)-212
-222
-232
Shelf load resistance*
PV(H)-332
*Uniformly distributed load
45kg
90kg
<
Horizontal type
>
For 25kg
Material: 18-8 Cr-Ni stainless steel wire
Shelf support load resistance: Max 50kg
* Equipped with 2 sets of shelf and shelf bracket.
For 60kg
Material: 18-8 Cr-Ni punched stainless steel
Shelf support load resistance: Max 200kg
* Standard shelves not provided.
<
Possible installation points
>
Model Top Rear Left side right side
× × ○ ○
PV(H)
PH(H)-
102
・
202
・
302
PH(H)-402
GPH(H)
STPH(H)
○
×
×
×
○
○
○
○
○
○
○
×
×
×
×
×
Cable port rubber plug
Prevents airleakage from the cable port.
Inside diameter: 25, 50, 100mm
* This rubber plug cannot be used when operating the chamber at
+200℃ or higher.
Model
PV(H)-212
-222
-232
PV(H)-332
Size
W550
×
D600
×
H35 mm
W740
×
D740
×
H38 mm
*Uniformly distributed load.
Shelf load resistance*
10kg
15kg
Casters
Installed for mobility.
・Adjustable type (Height 92mm)
4 casters
4 leveling feet
・ Non-adjustable type (Height 85mm)
2 casters with stoppers
2 fixed wheels
●
Please refer to chart on p.17 for applicable models.
20
OPTIONS
Viewing window
Used for observation of the specimens inside the chamber.
Dimensions: W190
×L340 mm
Stand
Exterior: Cold rolled and rust-proof steel plate with melamine baked finish
<
Vertical type
>
Type Top Model
MV-23
MV-23C
300mm
321mm
PV(H)-212
MV-26 600mm
PV(H)-212
MV-26C 621mm
*Type C: Casters and leveling feet
*with door
・
222
Angle type stand
Added to the chamber's original stand, this stand makes it easier to load and unload the specimen to the lower part of the test chamber.
Exterior: Equal-angle steel
Melamine baked finish
Model Type
L
M
H
Height
150mm
300mm
450mm
PH(H)-402
SPH(H)-402
Chamber lamp
Required when the door is fitted with viewing windows.
Location (incandescent light bulb):
PH-102, 202-Test area ceiling
PH-302, 402-Test area rear wall
Anchoring fixtures
Used to bolt the chamber to the floor.
Floor reinforcement
Used when testing load is larger than standard maximum load capacity.
* This option should be ordered together with the chamber.
Model
Floor load resistance*
Standard load resistance*
PH(H)-202
SaPH(H)-202
GPH(H)-202
IPH(H)-202
Up to 200kg
PH(H)-302
SPH(H)-302
PH(H)-402
SPH(H)-402
Up to 300kg
* Equally distributed load
50kg
60kg
100kg
MV-23C
<
Horizontal type
>
Type
L-1
L-2
L-3
M-1
M-2
Height
140mm
Model
PH(H)-102, GPH(H)-102
PH(H)-202, GPH(H)-202
IPH(H)-202
200mm PH(H)-302, SPH(H)-302
365mm PH(H)-102, GPH(H)-102
PH(H)-202, GPH(H)-202
IPH(H)-202
M-3
MS-1
MS-2
H-1(D)
400mm
505mm
PH(H)-302, SPH(H)-302
STPH-102, SSPH-102
STPH-202, SSPH-202
PH(H)-102, SPH(H)-102,
GPH(H)-102
H-2(D) 540mm
PH(H)-202, SPH(H)-202,
GPH(H)-202, IPH(H)-202
H-3(D) 585mm PH(H)-302, SPH(H)-302
*Type(D): with door
Casters for stand
Attached to the optional stand.
・Height adjustable (Height 92mm)
Free-turning wheel 4
Leveling feet
4
<
Horizontal type
>
L-1, L-2, M-1, M-2, H-1(D), H-2(D),
MS-1, MS-2
Stacking brackets
When stacking two chambers, this plate couples the top and bottom chambers securely.
*Only the L model optional stand can be used when chambers are stacked.
L-type-stand and stacking brackets
An L-type stand is fitted to the optional stacking brackets.
21
●
Please refer to chart on p.17 for applicable models.
■
Some photographs listed in this catalog contain Japanese display.
From the side, L-2,
M-2 (casters are optional) and H-2
OPTIONS
External alarm terminal
Used as a contact that relays an alarm to a remote point when one of the safety devices trips.
Output point: 1
Power supply: 250V AC 1A
Actuation: Signal generated when troubles occurs (contact closed)
Where located: Right side of chamber
Color specification
Chamber can be painted to any desired color.
Does not apply to:
・Door handle and handle cover
・Specimen power supply control
terminal frame
・Instrumentation frame
・Operation panel
・Damper operation panel
(including knob)
・Hinge cover
・Breaker cover
*Submit a color sample when specifying a color.
Emergency stop pushbutton
Stops the chamber immediately.
Interface
Communications ports to connect the chamber to a PC.
・RS-485
・GPIB
・RS-232C
Communication cables
・ RS-485
5m/ 10m/ 30m
・ RS-232C 1.5m/ 3m/ 6m
・ GPIB
2m/ 4m
Power cable
・5m
・10m
Safety precautions
● Do not use specimens which are explosive or inflammable, or which contain such substances. To do so could be hazardous, as this may lead to fire or explosion.
●
Periodical cleaning of the chamber and exhaust duct is required for it may cause combustion and fire when vapor of specimen is built up. Furthermore, an interior argon welding can be applied to the insulation layer of the chamber to minimize vapor penetration which may cause fire (except
IPH(H)). For more information, please contact us.
● Be sure to read the operation manual before operation.
22
Safety precautions
●
Do not use specimens which are explosive or inflammable, or which contain such substances.
To do so could be hazardous, as this may lead to fire or explosion.
● Periodical cleaning of the chamber and exhaust duct is required for it may cause combustion and fire when vapor of specimen is built up. Furthermore, an interior argon welding can be applied to the insulation layer of the chamber to minimize vapor penetration which may cause fire (except IPH(H)).
For more information, please contact us.
● Be sure to read the operation manual before operation.
DANGER
Do not use specimens which are explosive or inflammable, or which contain such substances.
To do so could be hazardous, as this may lead to fire or explosion.
CAUTION
Periodical cleaning of the chamber and exhaust duct is required for it may cause combustion and fire when vapor of specimen is built up.
Furthermore, an interior argon welding can be applied to the insulation layer of the chamber to minimize vapor penetration which may cause fire (except IPH(H)). For more information, please contact us.
Be sure to read the user’s manual before operation.
http://www.espec.co.jp/english
Head Office
3-5-6, Tenjinbashi, Kita-ku, Osaka 530-8550, Japan
Tel : 81-6-6358-4741 Fax : 81-6-6358-5500
ESPEC NORTH AMERICA, INC.
Tel : 1-616-896-6100 Fax : 1-616-896-6150
ESPEC EUROPE GmbH
Tel : 49-89-1893-9630 Fax : 49-89-1893-96379
ESPEC ENVIRONMENTAL EQUIPMENT (SHANGHAI) CO., LTD.
Head Office
Tel : 86-21-51036677 Fax : 86-21-63372237
BEIJING Branch
Tel : 86-10-64627025 Fax : 86-10-64627036
TIANJIN Branch
Tel : 86-22-26210366 Fax : 86-22-26282186
GUANGZHOU Branch
Tel : 86-20-83317826 Fax : 86-20-83317825
SHENZHEN Branch
Tel : 86-755-83674422 Fax : 86-755-83674228
SUZHOU Branch
Tel : 86-512-68028890 Fax : 86-512-68028860
ESPEC TEST TECHNOLOGY (SHANGHAI) CO., LTD.
Tel : 86-21-68798008 Fax : 86-21-68798088
ESPEC SOUTH EAST ASIA SDN.BHD.
Tel : 60-3-8945-1377 Fax : 60-3-8945-1287
TS7E26C02 (The contents of this catalog is as of July, 2014.)
Quality Management System Assessed and Registered
ESPEC CORP. has been assessed by and registered in the Quality Management System based on the International Standard ISO
9001:2008 (JIS Q 9001:2008) through the
Japanese Standards Association (JSA).
*
Registration : ESPEC CORP.
(Overseas subsidiaries not included)
Environmental Management System Assessed and Registered
ESPEC CORP.
● Specifications are subject to change without notice due to design improvements.
●
Corporate names and trade names mentioned in this catalog are trademarks or registered trademarks.
Temperature Chamber Series
CAT.NO.E94190-Q1407