Pro30
U SE R MA N UA L
Eng lish
For Sales & Service Contact
2650 E. 40th Ave. • Denver, CO 80205
Phone 303-320-4764 • Fax 303-322-7242
1-800-833-7958
www.geotechenv.com
C o ntents
Warranty............................................................................................................... i
Introduction........................................................................................................ 1
Getting Started.................................................................................................... 1
Initial Inspection.................................................................................. 1
Battery Installation.............................................................................. 1
Key Pad................................................................................................. 2
Connecting the Probe/Cable Assembly to the Instrument.............. 3
Run Screen............................................................................................ 4
Backlight............................................................................................... 5
Powering Off........................................................................................ 5
Navigation............................................................................................. 5
First Power On..................................................................................... 5
System Setup Menu............................................................................................. 6
Audio . ................................................................................................. 7
Contrast................................................................................................ 7
Conductivity Units (Cond. Units)...................................................... 8
Conductivity Auto Stable (Cond. Auto Stable)................................. 9
Temperature Units............................................................................... 9
Specific Conductance Reference Temperature
(SPC Ref. Temp.)................................................................................ 10
Specific Conductance Temperature Coefficient (SPC %/°C)........ 10
TDS Constant..................................................................................... 10
Language............................................................................................. 11
Item #606082
Rev A
Drawing # A606082
July 2011
©2011 YSI Incorporated.
The YSI logo is a registered trademarks of YSI Incorporated.
Auto Shutoff....................................................................................... 11
Cell Constant...................................................................................... 11
Resetting the System Setup Menu to Factory Default.................... 12
Calibration ........................................................................................................ 12
Temperature....................................................................................... 12
Conductivity Calibration.................................................................. 12
Taking Measurements....................................................................................... 15
Saving and Viewing Data................................................................................. 15
Saving Data......................................................................................... 16
Viewing and Erasing Saved Data - Data Mode............................... 16
Care, Maintenance and Storage....................................................................... 19
General Maintenance........................................................................ 19
Sensor Maintenance........................................................................... 19
Sensor Storage.................................................................................... 20
Warranty
The YSI Professional 30 instrument (Pro30) is warranted for three (3) years from
date of purchase by the end user against defects in materials and workmanship,
exclusive of batteries and any damage caused by defective batteries. Pro30 cable/
probe assemblies are warranted for two (2) years from date of purchase by the end
user against defects in material and workmanship. Pro30 instruments & cables
are warranted for 90 days from date of purchase by the end user against defects in
material and workmanship when purchased by rental agencies for rental purposes.
Within the warranty period, YSI will repair or replace, at its sole discretion, free of
charge, any product that YSI determines to be covered by this warranty.
Declaration of Conformity............................................................................... 24
To exercise this warranty, call your local YSI representative, or contact YSI
Customer Service in Yellow Springs, Ohio at +1 937 767-7241, 800-897-4151 or visit
www.YSI.com for a Product Return Form. Send the product and proof of purchase,
transportation prepaid, to the Authorized Service Center selected by YSI. Repair
or replacement will be made and the product returned, transportation prepaid.
Repaired or replaced products are warranted for the balance of the original warranty
period, or at least 90 days from date of repair or replacement.
Recycling............................................................................................................ 25
LIMITATION OF WARRANTY
Troubleshooting................................................................................................ 20
Specifications..................................................................................................... 22
Accessories / Part Numbers............................................................................. 23
Battery Disposal................................................................................. 25
Contact Information......................................................................................... 26
Ordering and Technical Support...................................................... 26
Service Information........................................................................... 26
This Warranty does not apply to any YSI product damage or failure caused by:
1.
Failure to install, operate or use the product in accordance with YSI’s written
instructions;
2. Abuse or misuse of the product;
3. Failure to maintain the product in accordance with YSI’s written instructions
or standard industry procedure;
4. Any improper repairs to the product;
5. Use by you of defective or improper components or parts in servicing or
repairing the product;
6. Modification of the product in any way not expressly authorized by YSI.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR
IMPLIED, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS
FOR A PARTICULAR PURPOSE. YSI’S LIABILITY UNDER THIS WARRANTY
IS LIMITED TO REPAIR OR REPLACEMENT OF THE PRODUCT, AND THIS
SHALL BE YOUR SOLE AND EXCLUSIVE REMEDY FOR ANY DEFECTIVE
PRODUCT COVERED BY THIS WARRANTY. IN NO EVENT SHALL YSI BE
LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL
DAMAGES RESULTING FROM ANY DEFECTIVE PRODUCT COVERED BY
THIS WARRANTY.
i
I ntr o ducti o n
T his page left intenti o nally blank
Thank you for purchasing the YSI Pro30, an instrument from the YSI Professional
Series product family. The Pro30 measures conductivity and temperature in
water. The Pro30 features an impact resistant and waterproof (IP-67) case, a
rugged MS-8 (military-spec) cable connector, backlit display, user-selectable
sensor options, 50 data set memory, internal barometer and a rubber over-mold
case.
The Pro30 provides valuable instructions and prompts near the bottom of the
display that will guide you through operation and use. However, reading the
entire manual is recommended for a better understanding of the instrument’s
features.
i
The Pro30 cannot communicate to a PC via a Pro Series communications
saddle. Connecting the Pro30 to a communication saddle may cause
erratic instrument behavior.
G etting S tarted
Initial Inspection
Carefully unpack the instrument and accessories and inspect for damage.
Compare received parts with items on the packing list. If any parts or materials
are damaged or missing, contact YSI Customer Service at 800-897-4151 (+1 937
767-7241) or the authorized YSI distributor from whom the instrument was
purchased.
Battery Installation
The instrument requires 2 alkaline C-cell batteries. Under normal conditions,
the average battery life is 425 hours at room temperature without using the back
light. A battery symbol
will blink in the lower, left corner of the display
to indicate low batteries when approximately 1 hour of battery life remains.
To install or replace the batteries:
1. Turn the instrument off and flip over to view the battery cover on the
back.
2. Unscrew the four captive battery cover screws.
3. Remove the battery cover and remove the old batteries if necessary.
4. Install the new batteries, ensuring correct polarity alignment (figure
1).
ii
1
5. Place the battery cover on the back of the instrument and tighten the
four screws. Do not over-tighten.
Figure 1. Pro30 with battery cover removed. Notice
battery symbols indicating polarities.
The waterproof instrument case is sealed at the factory and is not to
be opened, except by authorized service technicians. Do not attempt to
separate the two halves of the instrument case as this may damage the
instrument, break the waterproof seal, and will void the warranty.
i
Key Pad
2
3
1
4
5
6
Number
Key
Description
1
Calibrate
Press and hold for 3 seconds to calibrate.
Opens Calibrate menu from the Run screen.
2
Up Arrow
Use to navigate through menus, to navigate
through box options along the bottom of the
Run screen and to increase numerical inputs.
3
Power and Backlight
Press once to turn instrument on. Press a
second time to turn backlight on. Press a
third time to turn backlight off. Press and
hold for 3 seconds to turn instrument off.
4
Menu
Use to enter the System Setup
menu from the Run screen.
5
Enter
Press to confirm entries and selections.
6
Down Arrow
Use to navigate through menus, to navigate
through box options at the bottom of the Run
screen and to decrease numerical inputs.
Connecting the Probe/Cable
Assembly to the Instrument
The conductivity and temperature sensors are integral to the cable assembly;
therefore, they cannot be removed from the cable.
To connect the cable, align the keys on the cable connector to the slots on the
instrument connector. Push together firmly and then twist the outer ring until it
locks into place (figure 3). This connection is water-proof.
Figure 2, Keypad
2
3
Backlight
Once the instrument is powered on, pressing the power/backlight key
will turn on the display backlight. The backlight will remain on until the key is
pressed again or after two minutes of not pressing any key on the keypad.
Powering Off
To turn the instrument off, press and hold the power/backlight key
three seconds.
for
Navigation
Figure 3, Note the keyed connector.
The up
and down
of the Pro30.
arrow keys allow you to navigate through the functions
Navigating the Run Screen
Run Screen
Press the power/backlight key
to turn the instrument on. The instrument
will run through a self test and briefly display a splash screen with system
information before displaying the main Run screen (figure 4). The first time the
Pro30 is turned on, it will prompt you to select a language; see the First Power
On section of this manual for more information.
When in the Run screen, the up
and down arrow keys will move the
highlighted box along the bottom options. Once a box is highlighted, press enter
to access the highlighted option.
Description of Run screen box functions from left to right:
Option
Description
Highlight and press enter to save displayed data to
memory.
SAVE
Highlight and press enter to view and/or erase saved
data.
DATA
Navigating the System Setup Menu
When in the System Setup menu, the up and down arrow keys will move the
highlighted bar up and down the system setup options. See the System Setup
menu section of this manual for more information about these options.
First Power On
Figure 4, Run screen.
The instrument will step through an initial language configuration when
powered on for the first time. Use the up or down arrow keys to highlight the
4
5
appropriate language then press enter to confirm (figure 5). If an incorrect
language is selected, it may be changed in the System Setup menu.
Select Language:




English
Français
Español
Deutsch
Use  to select
Language
Press  to confirm
Figure 5, Select language.
After selecting a language, the Run screen will be displayed. The next time the
instrument is powered up, the Run screen will display immediately after the
splash screen.
S ystem S etup M enu
Press the menu
key to access the System Setup menu. The System Setup
menu contains multiple screens that are notated as ‘pages’. The current page is
indicated near the bottom of the display (figure 6).
Figure 6, page 1 of System Setup menu. Audio is enabled.
Audio
Audio can be enabled or disabled by using the up or down arrow keys to highlight
Audio and pressing enter. When enabled, there will be an ‘X’ in the box next to
Audio.
Use the up and down arrow keys to scroll through menu options and menu
pages.
When Audio is enabled, the Pro30 will beep twice to indicate stability when
Auto Stable is enabled. The instrument will also beep when a key is pressed.
When Audio is disabled, the Pro30 will not beep.
Exiting the System Setup Menu
Contrast
To exit the System Setup menu, press the down arrow key until the ESC - Exit
box is highlighted, then press enter to return to the Run screen.
To adjust the display Contrast, use the up or down arrow keys to highlight
Contrast, then press enter. Next, use the up or down arrow keys to adjust the
contrast. The up arrow key will darken the contrast and the down arrow key will
lighten the contrast. After adjusting the contrast, press enter to save and exit the
Contrast adjustment option.
Emergency Contrast Adjustment
If necessary, there is an alternate method of adjusting the contrast. To adjust the
contrast, press and hold the menu key, then press the up arrow key to darken the
contrast or press the down arrow key to lighten the contrast.
6
7
Conductivity Units (Cond. Units)
Highlight Cond. Units (Conductivity Units) and press enter to open a submenu
that allows you to select the conductivity units to be displayed on the Run screen.
Highlight a unit and press enter to enable or disable it. An enabled conductivity
unit will have an ‘X’ in the box next to it. Highlight the ESC-Exit box along
the bottom of the display and press enter to save any changes and to close the
conductivity units submenu.
There are seven options for displaying conductivity. Only four units can be
enabled at the same time:
•
•
•
•
•
•
•
COND-mS/cm displays conductivity in milliSiemens per centimeter.
COND-uS/cm displays conductivity in microSiemens per centimeter.
SPC-mS/cm displays Specific Conductance in milliSiemens per
centimeter. Specific Conductance is temperature compensated
conductivity.
SPC-uS/cm displays Specific Conductance in microSiemens per
centimeter. Specific Conductance is temperature compensated
conductivity.
Sal ppt displays salinity in parts per thousand. The salinity reading is
calculated from the instrument’s conductivity and temperature values
using algorithms found in Standard Methods for the Examination of
Water and Wastewater.
TDS g/L displays Total Dissolved Solids in grams per liter. TDS is
calculated from conductivity and temperature using a user-selectable
TDS constant.
TDS mg/L displays Total Dissolved Solids in milligrams per liter. TDS
is calculated from conductivity and temperature using a user-selectable
TDS constant.
Note: 1 milliSiemen = 1,000 microSiemens.
each measurement to calculate a specific conductance value compensated to a
user selected reference temperature, see below. Additionally, the user can select
the temperature coefficient from 0% to 4%.
Using the Pro30’s default reference temperature and temperature coefficient (25
°C and 1.91%), the calculation is carried out as follows:
Specific Conductance (25°C) = Conductivity of sample
1 + 0.0191 * (T - 25)
T = Temperature of the sample in °C
Conductivity Auto Stable (Cond. Auto Stable)
Auto Stable utilizes preset values to indicate when a reading is stable. The preset
values are adjustable in the System Setup menu. The user can input a % change
in readings (0.0 to 1.9) over ‘x’ amount of time in seconds (3-19).
Highlight Cond. Auto Stable, then press enter to open the submenu.
Use the up or down arrow keys to highlight the % change or seconds (secs) input
field, then press enter to make the highlighted field adjustable. Use the up or
down arrow keys to adjust the selected value, then press enter to confirm changes.
Once you have confirmed any changes, highlight the ESC-Exit box along the
bottom of the display and press enter to close the Auto Stable submenu.
To disable Auto Stable, set the % Change input to 0.0.
When Auto Stable is enabled, an AS symbol will display next to the reading
on the Run screen and blink during stabilization. When the dissolved oxygen
and/or conductivity reading stabilizes based on the Auto Stable settings, the
AS symbol will display steadily and the instrument will beep twice if Audio
is turned on.
Specific Conductance
The conductivity of a sample is highly dependent on temperature, varying as
much as 3% for each change of one degree Celsius (temperature coefficient =
3%/°C). In addition, the temperature coefficient itself varies with the nature of
the ionic species present in the sample. Therefore, it is useful to compensate for
this temperature dependence in order to quickly compare conductivity readings
taken at different temperatures.
The Pro30 can display non-temperature compensated conductivity as well as
temperature compensated Specific Conductance. If Specific Conductance is
selected, the Pro30 uses the temperature and conductivity values associated with
8
Temperature Units
Highlight Temperature Units and press enter to open a submenu that allows
you to change the temperature units displayed on the Run screen. Highlight
the desired unit (Celsius or Fahrenheit) and press enter to enable. The enabled
temperature unit will have an ‘X’ in the box next to it. Only one unit may be
enabled at a time. Highlight the ESC-Exit box and press enter to save any changes
and to close the Temperature Units submenu.
9
Specific Conductance Reference
Temperature (SPC Ref. Temp.)
SPC Ref. Temp. (Specific Conductance Reference Temperature) is the reference
temperature used to calculate Specific Conductance. The reference temperature
range is 15 and 25 °C. The default value is 25 °C.
To change the reference temperature, highlight SPC Ref. Temp. and press enter
to open the submenu. With the reference temperature highlighted, press enter
to make the field adjustable. Next, use the up or down arrow key to increase
or decrease the value. Press enter to save the new reference temperature. Next,
highlight the ESC-Exit box and press enter to close the submenu.
Specific Conductance Temperature
Coefficient (SPC %/°C)
SPC %/°C (Specific Conductance Temperature Coefficient) is the temperature
coefficient used to calculate Specific Conductance. The coefficient range is 0.00
to 4.00. The default value is 1.91% which is based on KCl standards.
To change the temperature coefficient, highlight SPC %/°C and press enter to
open the submenu. With the temperature coefficient highlighted, press enter
to make the field adjustable. Next, use the up or down arrow key to increase or
decrease the value. Press enter to save the new coefficient. Next, highlight the
ESC-Exit box and press enter to close the submenu.
TDS Constant
TDS Constant is a multiplier used to calculate an estimated TDS (Total Dissolved
Solids) value from conductivity. The multiplier is used to convert Specific
Conductance in mS/cm to TDS in g/L. The Pro30’s default value is 0.65. This
multiplier is highly dependent on the nature of the ionic species present in the
water sample. To be assured of moderate accuracy for the conversion, you must
determine a multiplier for the water at your sampling site. Use the following
procedure to determine the multiplier for a specific sample:
1. Determine the specific conductance of a water sample from the site;
2. Filter a sample of water from the site;
3. Completely evaporate the water from a carefully measured volume of
the filtered sample to yield a dry solid;
4. Accurately weigh the remaining solid;
5. Divide the weight of the solid (in grams) by the volume of water used
(in liters) to yield the TDS value in g/L for this site;
6. Divide the TDS value in g/L by the specific conductance of the water in
mS/cm to yield the conversion multiplier. Be certain to use the correct
units.
10
If the nature of the ionic species at the site changes between sampling studies, the
TDS values will be in error. TDS cannot be calculated accurately from specific
conductance unless the make-up of the chemical species in the water remains
constant.
To change the TDS Constant in the Pro30, highlight TDS Constant and press
enter to open the submenu. With the TDS Constant highlighted, press enter
to make the field adjustable. Next, use the up or down arrow key to increase or
decrease the value. The input range is 0.30 to 1.00. Press enter to save the new
TDS Constant. Next, highlight the ESC-Exit box and press enter to close the
submenu.
Language
Highlight Language and press enter to open a submenu that allows you to
change the language. Highlight the desired language (English, Spanish, German,
or French) and press enter to enable. The enabled language will have an ‘X’ in the
box next to it. Highlight ESC-Exit box and press enter to save any changes and
to close the Language submenu.
The text in the boxes along the bottom of the Run screen will always be displayed
in English regardless of the language enabled in the System Setup menu.
Auto Shutoff
Auto Shutoff allows you to set the instrument to turn off automatically after a
period of time. Use the up or down arrow keys to highlight Auto Shutoff, then
press enter to open the submenu. Press enter while the minute field is highlighted
to make it adjustable. Next, use the up or down arrow keys to adjust the shut
off time from 0 to 60 minutes. Press enter to save the new shutoff time. Next,
highlight the ESC-Exit box and press enter to close the submenu.
To disable Auto Shutoff, set the Time in Minutes to 0 (zero).
Cell Constant
The Cell Constant displays the cell constant of the conductivity cell. The cell
constant is calculated and updated each time a conductivity calibration is
performed. The cell constant range is 4.0 to 6.0. Resetting the System Menu
resets the cell constant to 5.0.
11
Resetting the System Setup Menu
to Factory Default
To reset the Pro30 settings to factory default, press the down arrow key while in
the System Setup menu until the Reset - box is highlighted, then press enter.
The instrument will ask you to confirm the reset. Highlight Yes and press enter
to continue with the reset or highlight No and press enter to cancel the reset. A
Factory Reset will not affect data saved in the instrument’s memory.
The following will be set in the Pro30 after performing a reset:
Parameter
Reset Defaults
Audio
On
Contrast
Conductivity Units
Set to mid range
cond uS/cm, spc mS/cm, spc uS/cm
and sal ppt
Conductivity Auto Stable
Off (0.0 % Change and 10 seconds)
SPC Reference Temperature
25°C
SPC Temperature Coefficient
1.91%/°C
TDS Constant
0.65
Temperature Units
°C
Language
English
Auto Shutoff
30 minutes
Conductivity Cell Constant
Cell constant reset to 5.0*
*It is recommended to perform a Conductivity calibration after performing a
reset.
i
It is not necessary to calibrate conductivity, specific conductance
and salinity. Calibrating one of these parameters will simultaneously
calibrate the others. YSI recommends calibrating specific conductance
for greatest ease.
Calibrating
Specific
Conductivity
(Sp.)
or
Note: When calibrating Specific Conductance, the Pro30 uses the factory default
values for the Specific Conductance Reference Temperature and the Specific
Conductance Temperature Coefficient regardless of what is configured in the
System Setup Menu. The default value for the Reference Temperature is 25°C
and the default value for the Temperature Coefficient is 1.91%/°C. It is important
to note that the Temperature Coefficient of a calibration solution is dependent
on the contents of the solution. Therefore, YSI recommends using a traceable
calibration solution made of KCl (potassium chloride) when calibrating Specific
Conductance since these solutions typically have a Temperature Coefficient of
1.91%/°C. Additionally, be sure to enter the value of the solution as it is listed for
25°C when calibrating Specific Conductance.
1. Fill a clean container (i.e. plastic cup or glass beaker) with fresh, traceable
conductivity calibration solution and place the sensor into the solution.
The solution must cover the holes of the conductivity sensor that are
closest to the cable (figure 7). Ensure the entire conductivity sensor is
submerged in the solution or the instrument will read approximately
half the expected value. Gently move the probe up and down to remove
any air bubbles from the conductivity sensor.
C alibrati o n
Temperature
All Pro30 cables have built-in temperature sensors. Temperature calibration is
not required nor is it available.
Conductivity Calibration
Ensure the conductivity sensor is clean and dry before performing a conductivity,
specific conductance or salinity calibration.
12
Conductance
13
Calibrating in Salinity
Figure 7, solution above two holes near cable.
2. Turn the instrument on and allow the conductivity and temperature
readings to stabilize. Press and hold the Cal key for 3 seconds. Highlight
Conductivity and press enter. Next, highlight the desired calibration
method, Sp. Conductance or Conductivity, and press enter.
3. Highlight the units you wish to calibrate, either uS/cm or mS/cm, and
press enter. 1 mS = 1,000 uS. Next, use the up or down arrow key to
adjust the value on the display to match the value of the conductivity
calibration solution. If calibrating conductivity, it is necessary to look
up the value of the solution at the current temperature and enter that
value into the Pro30. Most conductivity solutions are labeled with a
value at 25°C. If calibrating specific conductance, enter the value listed
for 25°C. Depressing either the up or down arrow key for 5 seconds will
move the changing digit one place to the left. The Pro30 will remember
the entered calibration value and display it the next time a conductivity
calibration is performed.
4. Press enter to complete the calibration. Or, press Cal to cancel the
calibration and return to the Run screen.
5. ‘Calibration Successful’ will display for a few seconds to indicate a
successful calibration and then the instrument will return to the Run
screen.
6. If the calibration is unsuccessful, an error message will display on the
screen. Press the Cal key to exit the calibration error message and
return to the Run screen. See the Troubleshooting guide for possible
solutions.
14
1. Fill a clean container (i.e. plastic cup or glass beaker) with fresh, traceable
salinity calibration solution and place the sensor into the solution.
The solution must cover the holes of the conductivity sensor that are
closest to the cable (figure 7). Ensure the entire conductivity sensor is
submerged in the solution or the instrument will read approximately
half the expected value. Gently move the probe up and down to remove
any air bubbles from the conductivity sensor.
2. Turn the instrument on and allow the conductivity and temperature
readings to stabilize. Press and hold the Cal key for 3 seconds. Highlight
Conductivity and press enter. Next, highlight Salinity and press enter.
3. Use the up or down arrow key to adjust the value on the display to
match the value of the salinity solution. Depressing either the up or
down arrow key for 5 seconds will move the changing digit one place
to the left. The Pro30 will remember the entered calibration value and
display it the next time a salinity calibration is performed.
4. Press enter to complete the calibration. Or, press Cal to cancel the
calibration and return to the Run screen.
5. ‘Calibration Successful’ will display for a few seconds to indicate a
successful calibration and then the instrument will return to the Run
screen.
6. If the calibration is unsuccessful, an error message will display on the
screen. Press the Cal key to exit the calibration error message and
return to the Run screen. See the Troubleshooting guide for possible
solutions.
T aking M easurements
Before taking measurements, be sure the instrument has been calibrated to ensure
the most accurate readings. Place the probe in the sample to be measured and
give the probe a quick shake to release any air bubbles. Be sure the conductivity
sensor is completely submerged in the sample. The two holes near the cable
should be covered by the sample for accurate conductivity readings (figure 7).
Allow the temperature readings to stabilize.
S aving and V iewing D ata
The Pro30 can store 50 data sets in non-volatile memory for later viewing. A
data set includes the values currently on the display, i.e. temperature, dissolved
oxygen and two conductivity parameters. Each data point is referenced with a
data set number, 01 through 50.
15
Saving Data
i
The Pro30 can not communicate to a PC via a Pro Series communications
saddle. Connecting the Pro30 to a communication saddle may cause
erratic instrument behavior.
From the Run screen, use the up or down arrow keys to highlight the Save box
and press enter to save the current readings. The instrument will indicate the
data set is saved and display the saved data set’s number (figure 8).
Figure 9, Data mode.
Viewing Data
Once in Data mode, use the up and down arrow keys to view saved data sets in
sequential order or press enter to access the bottom functions. After accessing
the bottom functions, highlight the Data box and press enter to regain access
to viewing data. The data set displayed is indicated by the data set number, 01
through 50.
Figure 8, data set saved.
The instrument will display ‘Memory Full’ if all 50 data sets have been saved and
you attempt to save another data set.
Viewing and Erasing Saved Data - Data Mode
Erasing Data
While viewing saved data, press the enter key to access the function boxes at the
bottom of the display. Next, use the up or down arrow keys to highlight Erase,
then press enter. The instrument will give you the option to erase one data set
or all data sets (figure 10).
Data mode allows you to view and erase saved data. From the Run screen, use
the up or down arrow keys to highlight Data and press enter to access Data
mode. Note that the function boxes at the bottom of the display are different in
Data mode (figure 9).
16
17
C are , M aintenance and
S t o rage
This section describes the proper procedures for care, maintenance and storage
of the instrument. The goal is to maximize their lifetime and minimize downtime associated with improper instrument usage.
General Maintenance
General Maintenance - Gasket
The instrument utilizes a gasket as a seal to prevent water from entering the
battery compartment. Following the recommended procedures will help keep
the instrument functioning properly.
Use the up or down arrow key to select Erase Data Set, Erase All Sets or the ESCExit function box, then press enter to confirm.
If the gasket and sealing surfaces are not maintained properly, it is possible that
water can enter the battery compartment. If water enters this area, it can severely
damage the battery terminals causing loss of battery power and corrosion to the
battery terminals. Therefore, when the battery compartment lid is removed, the
gasket that provides the seal should be carefully inspected for contamination (i.e.
debris, grit, etc.) and cleaned with water and mild detergent if necessary.
Select ESC-Exit and press enter to exit Erase mode without erasing any data.
Sensor Maintenance
Select Erase Data Set and press enter to erase the data set that was displayed
before entering Erase mode. For example, if data set 12 was displayed before
entering erase mode, and Erase Data Set is selected, Data Set 12 will be erased
from memory and the data sets AFTER that number will move up to keep them
sequential. For example, if there are15 records and number 12 is erased then 13
becomes 12, 14 becomes 13, and 15 becomes 14. The instrument will return to
Data mode after erasing one data set.
Sensor Maintenance - Temperature
Figure 10, Erase data mode.
Select Erase All Data Sets and press enter to clear the Pro30 memory and return
to Data mode.
Exiting Data Mode
While in Data mode, press enter to access the bottom functions. Next, highlight
the ESC-Exit box and press enter to return to the Run screen.
18
You must keep the temperature sensor free of build up. Other than that, no
additional maintenance is required. A toothbrush can be used to scrub the
temperature sensor if needed.
Sensor Maintenance - Conductivity
The openings that allow sample access to the conductivity electrodes should be
cleaned regularly. The small cleaning brush included in the Maintenance Kit is
intended for this purpose. Dip the brush in clean water and insert it into each
hole 10 to 12 times. In the event that deposits have formed on the electrodes, it
may be necessary to use a mild detergent (laboratory grade soap or bathroom
foaming tile cleaner) with the brush. Rinse thoroughly with clean water, then
check the response and accuracy of the conductivity cell with a calibration
solution.
19
Sensor Storage
Symptom
SHORT AND Long TERM STORAGE
For both short and long term storage, the conductivity sensor should be stored
clean and dry.
Conductivity readings
are inaccurate.
1. Ensure the conductivity sensor is
clean. Follow the cleaning procedures
in the Care, Maintenance and
Storage section of this manual.
2. Verify the sample is above the two
holes near the cable, see figure 8.
3. Verify calibration.
4. Verify temperature readings
are accurate.
5. Verify the correct units are setup in the
System Setup menu, i.e. uS vs mS and
Conductivity vs. Specific Conductance.
6. Contact YSI Tech Support.
Conductivity values display
Over or Undr on Run screen.
1. Ensure the conductivity sensor is
clean. Follow the cleaning procedures
in the Care, Maintenance and
Storage section of this manual.
2. Verify the sample is above the two
holes near the cable, see figure 8
3. Verify calibration.
4. Verify temperature readings
are accurate.
5. Sample conductivity is outside
the measurement range of the
instrument, i.e. 0-200 mS.
6. Contact YSI Tech Support.
Remove the batteries from the instrument when storing it for long periods of
time (>30 days).
Long Term Storage Temperature: -5 to 70°C (23 to 158°F)
T r o ublesh o o ting
Symptom
Possible Solution
Instrument will not turn on,
a battery symbol appears,
or “Critical Shutdown”
displays on the screen.
1. Low battery voltage, replace batteries.
2. Batteries installed incorrectly,
check battery polarity.
3. Return system for service.
Temperature values display
Over or Undr on Run screen.
1. Sample temperature is less than -5° C or
more than +55°C. Increase or decrease
the sample temperature to bring within
the allowable range.
2. Contact YSI Tech Support.
Instrument will not calibrate the 1. Ensure the conductivity sensor is
Conductivity sensor; instrument
clean. Follow the cleaning procedures
displays “Calibration Over”,
in the Care, Maintenance and
“Calibration Under”, or “Unstable
Storage section of this manual.
Reading” during calibration.
2. Verify the calibration solution is above
the two holes near the cable, see figure 8.
3. Verify the calibration solution
is not expired or contaminated.
Try a new bottle of solution.
4. Ensure you are entering in the correct
value for the solution according to the
measurement units. 1 mS = 1,000 uS.
5. Allow sufficient stabilization time
for conductivity and temperature
AND wait at least 3 seconds before
confirming a calibration.
6. Contact YSI Tech Support.
20
Possible Solution
21
S pecificati o ns
A ccess o ries / P art N umbers
These specifications represent typical performance and are subject to change
without notice. For the latest product specification information, please visit
YSI’s website at www.ysi.com or contact YSI Tech Support.
Parameter
Range
60530-1, -4, -10, -20, or -30
1, 4, 10, 20, 30-meter cable assembly*
603077
Flow cell
603056
Flow cell mounting spike
603075
Carrying case, soft-sided
603074
Carrying case, hard-sided
603069
Belt clip
063517
Ultra clamp for instrument
063507
Tripod for instrument
603062
Cable management kit, included with all cables
longer then 1 meter.
605978
Cable weight, 4.9 oz, stackable
± 1.0% of the reading
or ± 0.1 ppt, whichever
is greater.
603070
Shoulder strap
060907
Conductivity Calibration Solution, 1,000 μS/
cm. 1 box of 8 pints.
Dependent on accuracy of
temperature, conductivity
and TDS Constant.
060911
Conductivity Calibration Solution, 10,000 μS/
cm. 1 box of 8 pints.
060660
Conductivity Calibration Solution, 50,000 μS/
cm. 1 box of 8 pints.
065274
Conductivity Calibration Solution, 100,000 μS/
cm. 1 box of 8 pints.
Resolution
Accuracy
-5 to 55°C
0.1°C
± 0.2°C
Conductivity
0-500 uS/cm
0-5 mS/cm
0-50 mS/cm
0-200 mS/
cm (auto
ranging)
0.0001 to
0.1 mS/cm;
0.1 to 0 uS/
cm (range
dependent)
Instrument only: ± 0.5%
of the reading or 1 uS/
cm, whichever is greater.
Instrument with 1
or 4 meter cables:
± 1.0% of the reading or 1
uS/cm, whichever is greater.
Instrument with 10, 20,
or 30 meter cables:
± 2.0% of the reading or 1
uS/cm, whichever is greater.
Total
Dissolved
Solids (TDS)
0 to 70 ppt
0.1 ppt
0 to 100 g/L.
0.0001 to 0.1
TDS Constant g/L (range
range: 0.3 to
dependent)
1.00 (0.65
default)
Description
Pro30 Instrument
Temperature
Salinity
Part Number
6050030
*All cables include a temperature and conductivity sensor.
22
23
D eclarati o n o f C o nf o rmity
The undersigned hereby declares on behalf of the named manufacturer under our
sole responsibility that the listed product conforms to the requirements for the listed
European Council Directive(s) and carries the CE mark accordingly.
Manufacturer:
YSI Incorporated
1725 Brannum Lane
Yellow Springs, OH 45387
USA
Product Name:
Pro30 Water Quality Instrument
Model Numbers
Instrument/Accessory:
Pro30 (6050030)
Probe/Cable Assemblies:
60530-1, -4, -10, -20, and -30
R ecycling
YSI is committed to reducing the environmental footprint in the course of doing
business. Even though materials reduction is the ultimate goal, we know there
must be a concerted effort to responsibly deal with materials after they’ve served
a long, productive life-cycle. YSI’s recycling program ensures that old equipment
is processed in an environmentally friendly way, reducing the amount of
materials going to landfills.
•
•
•
Conforms to the following:
Directives:
IEC 61326-1:2005
RoHS 2002/95/EC
WEEE 2002/96/EC
IP-67 Protection per ANSI/IEC 60529-2004
Harmonized Standards:
•
Supplementary
Information:
All performance met the operation criteria as follows:
1. ESD, IEC 61000-4-2:2001, Performance Criterion B
2. Radiated Immunity, IEC 61000-4-3, Performance
Criterion A
3. Electrical Fast Transient (EFT), IEC
61000-4-4:2004, +Corr. 1:2006 + Corr.
2:2007, Performance Criterion B
4. Radio Frequency, Continuous Conducted
Immunity, IEC61000-4-6, Performance Criterion A
5. Radiated Emissions, EN 613261:2006 (IEC61326-1:2005) Class B
Authorized EU
Representative
YSI Hydrodata Ltd
Unit 2 Focal Point, Lacerta Court, Works Road
Letchworth, Hertfordshire, SG6 1FJ UK
EN61326-1:2006 (IEC 613261:2005) Basic Immunity
Signed: Lisa M. Abel
Title: Director of Quality
24
Printed Circuit Boards are sent to facilities that process and reclaim as much
material for recycling as possible.
Plastics enter a material recycling process and are not incinerated or sent
to landfills.
Batteries are removed and sent to battery recyclers for dedicated metals.
When the time comes for you to recycle, follow the easy steps outlined at
www.ysi.com.
Battery Disposal
The Pro30 is powered by alkaline batteries which the user must remove and
dispose of when the batteries no longer power the instrument. Disposal
requirements vary by country and region, and users are expected to understand
and follow the battery disposal requirements for their specific locale.
Date: 27 June 2011
25
C o ntact I nf o rmati o n
Ordering and Technical Support
Telephone: 800 897 4151 (USA)
+1 937 767 7241 (Globally)
Monday through Friday, 8:00 AM to 5:00 ET
Fax: +1 937 767 9353 (orders)
+1 937 767 1058 (technical support)
Email: Mail: environmental@ysi.com
YSI Incorporated
1725 Brannum Lane
Yellow Springs, OH 45387
USA
Internet: www.ysi.com
When placing an order please have the following available:
1.) YSI account number (if available)
2.) Name and phone number
3.) Purchase Order or Credit Card number
4.) Model Number or brief description
5.) Billing and shipping addresses
6.) Quantity
Service Information
YSI has authorized service centers throughout the United States and
Internationally. For the nearest service center information, please visit www.ysi.
com and click ‘Support’ or contact YSI Technical Support directly at 800-8974151 (+1 937-767-7241).
When returning a product for service, include the Product Return form with
cleaning certification. The form must be completely filled out for a YSI Service
Center to accept the instrument for service. The form may be downloaded from
www.ysi.com by clicking on the ‘Support”.
Item # 606082
Rev A
Drawing # A606082
July 2011
©2011 YSI Incorporated.
26