Q-TECH™ PROGRAMMER
A COMPONENT OF THE S-ICD® SYSTEM
USER’S MANUAL
3200
© Copyright 2013 Boston Scientific or its affiliates.
All rights reserved.
Limited Software License and Equipment Use.
S-ICD®, SQ-RX®, and Q-TRAK® are all registered trademarks of Cameron Health, Inc.
Q-GUIDE™ and Q-TECH™ are all trademarks of Cameron Health, Inc.
Manuals or other written documentation may not be copied or distributed without Cameron Health, Inc.
authorization.
Cameron Health, Inc.
4100 Hamline Avenue North
St. Paul, MN 55112-5798 USA
1.800.CARDIAC (227.3422)
+1.651.582.4000
www.cameronhealth.com
Guidant Europe NV/SA
Boston Scientific
Green Square,
Lambroekstraat 5D
1831 Diegem, Belgium
1.800.CARDIAC (227.3422)
+1.651.582.4000
www.cameronhealth.com
TABLE OF CONTENTS
general description
Description��������������������������������������������������������������������������������������������������������������������������������1
Indications for Use��������������������������������������������������������������������������������������������������������������������1
Contraindications���������������������������������������������������������������������������������������������������������������������1
Programmer Warnings and Cautions��������������������������������������������������������������������������������������1
General��������������������������������������������������������������������������������������������������������������������������������������������������������������������1
Electrostatic Discharge (ESD)��������������������������������������������������������������������������������������������������������������������������3
S-ICD System Warnings and Cautions�������������������������������������������������������������������������������������3
General��������������������������������������������������������������������������������������������������������������������������������������������������������������������3
Use of Other Medical Therapies/Diagnostic Procedures����������������������������������������������������������������������4
Electromagnetic Interference (EMI) Outside the Hospital Environment����������������������������������������5
Potential Adverse Events����������������������������������������������������������������������������������������������������������������������������������5
Operation
Setup of the Q-TECH Programmer�������������������������������������������������������������������������������������������7
Packaging���������������������������������������������������������������������������������������������������������������������������������������������������������������7
Charging the Programmer�������������������������������������������������������������������������������������������������������������������������������7
Using the Q-TECH Programmer�����������������������������������������������������������������������������������������������8
Turning the Programmer On���������������������������������������������������������������������������������������������������������������������������8
Placing the Programmer in Suspend Mode�����������������������������������������������������������������������������������������������9
Turning the Programmer Off����������������������������������������������������������������������������������������������������������������������� 10
Using the Programmer Touch Screen������������������������������������������������������������������������������������������������������� 10
Using the Wand������������������������������������������������������������������������������������������������������������������������������������������������� 10
Navigation���������������������������������������������������������������������������������������������������������������������������������������������������������� 12
Screen Header��������������������������������������������������������������������������������������������������������������������������������������������������� 12
Navigation Bar��������������������������������������������������������������������������������������������������������������������������������������������������� 12
Restarting the Programmer�������������������������������������������������������������������������������������������������������������������������� 14
Configuring the Q-TECH Programmer���������������������������������������������������������������������������������� 15
Configuring Programmer Settings������������������������������������������������������������������������������������������������������������� 15
Date and Time Format������������������������������������������������������������������������������������������������������������������������������������ 17
Time Zone����������������������������������������������������������������������������������������������������������������������������������������������������������� 17
Language Preference�������������������������������������������������������������������������������������������������������������������������������������� 18
Printer Selection����������������������������������������������������������������������������������������������������������������������������������������������� 18
Programmer Software Version�������������������������������������������������������������������������������������������������������������������� 21
OPERATION (continued)
Modes of Operation for the Q-TECH Programmer��������������������������������������������������������������� 21
Online Behavior������������������������������������������������������������������������������������������������������������������������������������������������ 21
Offline Behavior������������������������������������������������������������������������������������������������������������������������������������������������ 22
Modes of Operation for the SQ-RX Device��������������������������������������������������������������������������� 23
Shelf Mode���������������������������������������������������������������������������������������������������������������������������������������������������������� 23
Therapy On Mode�������������������������������������������������������������������������������������������������������������������������������������������� 23
Therapy Off Mode�������������������������������������������������������������������������������������������������������������������������������������������� 23
Connecting and Disconnecting from the SQ-RX Device����������������������������������������������������� 24
Scanning For SQ-RX Devices������������������������������������������������������������������������������������������������������������������������ 24
Connecting to the SQ-RX Device���������������������������������������������������������������������������������������������������������������� 26
Connecting to an SQ-RX Device in Shelf Mode:������������������������������������������������������������������������������������ 26
Ending a Patient Session������������������������������������������������������������������������������������������������������������������������������� 27
Programming the SQ-RX Device at Implant������������������������������������������������������������������������ 29
Entering Electrode Information������������������������������������������������������������������������������������������������������������������ 29
Creating the Patient Chart���������������������������������������������������������������������������������������������������������������������������� 31
Automatic Setup����������������������������������������������������������������������������������������������������������������������������������������������� 33
Programming Therapy Parameters������������������������������������������������������������������������������������������������������������ 36
Defibrillation Testing��������������������������������������������������������������������������������������������������������������������������������������� 39
Performing an SQ-RX Follow-up������������������������������������������������������������������������������������������� 42
Sensing Configuration and Automatic Setup���������������������������������������������������������������������������������������� 42
Viewing SQ-RX Device Status���������������������������������������������������������������������������������������������������������������������� 42
Viewing Stored Episodes������������������������������������������������������������������������������������������������������������������������������� 44
Printing Reports from the Programmer������������������������������������������������������������������������������� 45
Printing Reports������������������������������������������������������������������������������������������������������������������������������������������������ 45
Summary Report���������������������������������������������������������������������������������������������������������������������������������������������� 46
Captured S-ECG Report���������������������������������������������������������������������������������������������������������������������������������� 47
Episodes Report������������������������������������������������������������������������������������������������������������������������������������������������ 48
S-ECG Features����������������������������������������������������������������������������������������������������������������������� 49
S-ECG Rhythm Strip Markers������������������������������������������������������������������������������������������������������������������������ 49
S-ECG Scale Settings��������������������������������������������������������������������������������������������������������������������������������������� 50
Capture and View S-ECG Strips�������������������������������������������������������������������������������������������������������������������� 51
Utilities Menu������������������������������������������������������������������������������������������������������������������������� 52
Acquire Reference S-ECG������������������������������������������������������������������������������������������������������������������������������� 53
Beeper Control�������������������������������������������������������������������������������������������������������������������������������������������������� 53
Manual Setup���������������������������������������������������������������������������������������������������������������������������������������������������� 55
Smart Charge����������������������������������������������������������������������������������������������������������������������������������������������������� 57
Additional Programmer Functions��������������������������������������������������������������������������������������� 59
Rescue Shock����������������������������������������������������������������������������������������������������������������������������������������������������� 59
Manual shock����������������������������������������������������������������������������������������������������������������������������������������������������� 61
S-ICD System Magnet Model 4520������������������������������������������������������������������������������������������������������������ 61
maintenance
Charging the Programmer���������������������������������������������������������������������������������������������������������������������������� 62
Cleaning the Programmer���������������������������������������������������������������������������������������������������������������������������� 62
Service������������������������������������������������������������������������������������������������������������������������������������������������������������������ 63
Maintenance Check����������������������������������������������������������������������������������������������������������������������������������������� 63
Safety Measurements������������������������������������������������������������������������������������������������������������������������������������� 63
Programmer End of Life �������������������������������������������������������������������������������������������������������������������������������� 63
troubleshooting
Inability to Print������������������������������������������������������������������������������������������������������������������������������������������������ 64
No Printer Available����������������������������������������������������������������������������������������������������������������������������������������� 64
Touch Screen Inactive while Connected to AC Power������������������������������������������������������������������������ 64
Loss of Communication with Printer��������������������������������������������������������������������������������������������������������� 65
Inability to Communicate With the SQ-RX Device������������������������������������������������������������������������������� 65
Compliance statements
Federal Communications Commission (FCC) US���������������������������������������������������������������������������������� 66
1999/5/EC (R&TTE Directive)������������������������������������������������������������������������������������������������������������������������ 66
EMI/RFI����������������������������������������������������������������������������������������������������������������������������������������������������������������� 66
Declarations tables
Declaration Electromagnetic Emission���������������������������������������������������������������������������������������������������� 67
Declaration Electromagnetic Immunity Part 1�������������������������������������������������������������������������������������� 67
Declaration Electromagnetic Immunity Part 2�������������������������������������������������������������������������������������� 68
Recommended Separation Distances������������������������������������������������������������������������������������������������������ 69
EMI/RFI Information���������������������������������������������������������������������������������������������������������������������������������������� 69
Specifications
Product Guidelines������������������������������������������������������������������������������������������������������������������������������������������ 70
Specifications����������������������������������������������������������������������������������������������������������������������������������������������������� 70
Nominal Specifications���������������������������������������������������������������������������������������������������������������������������������� 71
Definition of package label symbols
Packaging Symbols: Q-TECH Programmer���������������������������������������������������������������������������������������������� 72
Packaging Symbols: Q-TECH Programmer Wand��������������������������������������������������������������������������������� 73
Warranty
Warranty�������������������������������������������������������������������������������������������������������������������������������������������������������������� 74
Q-TECH™ PROGRAMMER: general description
Description
The Q-TECH Programmer (the “programmer”), Model 3200, is a component of the Cameron Health/
Boston Scientific S-ICD System, which is prescribed for patients when cardiac arrhythmia management
is warranted. The other components of the S-ICD System include the SQ-RX Pulse Generator (the “SQRX device”), Q-TRAK Subcutaneous Electrode (“electrode”) and the Q-GUIDE Subcutaneous Electrode
Insertion Tool (“EIT”). The programmer communicates wirelessly with the SQ-RX device to enable
adjustment of programmable settings and data collection. The S-ICD System is designed for ease of use
and simplicity of patient management. The S-ICD System has a number of automatic functions designed
to reduce the amount of time required for implantation, initial programming and patient follow-up.
Indications for Use
The S-ICD System is intended to provide defibrillation therapy for the treatment of life-threatening
ventricular tachyarrhythmias in patients who do not have symptomatic bradycardia, incessant
ventricular tachycardia, or spontaneous, frequently recurring ventricular tachycardia that is reliably
terminated with anti-tachycardia pacing.
Contraindications
Unipolar pacemakers are contraindicated for use with the S-ICD System.
Programmer Warnings and Cautions
General
•
External defibrillation equipment should be available for immediate use during the
implantation procedure and follow-up.
•
Use the programmer only with the external power supply packaged with the programmer.
Using other accessories may cause damage to the programmer.
•
The US power cord is for connection to 110 VAC supply mains. All other power cords are for
connection to 230 VAC supply mains. Outside North America, use the supplied power cord that
exactly matches your AC electrical outlet.
•
To avoid risk of electric shock, the programmer must only be connected to a supply mains with
protective earth.
•
Mains isolation is achieved by disconnecting the AC power cord. Do not position the programmer
or the external power supply in a manner that would make it difficult to disconnect that cord.
•
Never use a damaged external power supply or a damaged programmer.
1
Q-TECH™ PROGRAMMER: general description
2
•
The programmer is not waterproof or explosion-proof and it cannot be sterilized. Do not use
it in the presence of flammable gas mixtures, including anesthetic mixtures containing air,
oxygen, or nitrous oxide.
•
Confirm that the programmer is in communication with the intended implanted SQ-RX Pulse
Generator.
•
The programmer is intended for use by healthcare professionals only. The programmer
may cause radio interference or may disrupt the operation of nearby equipment. It may be
necessary to take mitigation measures, such as reorienting or relocating the programmer or
shielding the location.
•
By design, the programmer emits radio frequencies (403.5 MHz and 2.5 GHz). This may interfere
with nearby equipment. Closely monitor medical electrical equipment in the vicinity when
using the programmer.
•
The presence of other equipment in the same frequency band (403.5 MHz for the SQ-RX Pulse
Generator and 2.5 GHz for the printer) may interfere with communication. If communication
problems occur, refer to the Troubleshooting section of this manual.
•
If the programmer is used near other medical or office equipment, all equipment should be
observed to verify normal operation. Other equipment may interfere with the programmer,
even if that other equipment complies with the International Special Committee on Radio
Interference (CISPR) emission requirements.
•
Use of non-Cameron Health/Boston Scientific-approved accessories may result in increased
emissions or decreased functionality of the programmer and may cause unintended
operational behavior of the programmer.
•
No modification of the programmer is allowed.
•
Use only the Cameron Health/Boston Scientific S-ICD programmer and appropriate software to
communicate with and program the SQ-RX Pulse Generator.
•
Do not discard the programmer in a fire, incinerate, or subject it to temperatures that exceed
100° C (212° F). This could cause the programmer to explode.
•
Do not disassemble or alter any parts of the programmer.
•
Do not immerse the programmer in liquid of any kind. If the programmer does get wet, contact
customer service for information about returning the programmer to Cameron Health/Boston
Scientific. Do not attempt to dry the programmer in an oven, microwave, or dryer because this
poses a risk of overheating or explosion.
Q-TECH™ PROGRAMMER: general description
•
Do not subject the programmer to temperatures outside of the -10° C to 55° C (14° F to 131° F)
storage range. Exposure to high temperatures may cause the programmer to overheat or ignite,
and may possibly reduce its performance and service life.
•
Mishandling (such as dropping or crushing) could damage the programmer. If you suspect
damage to the programmer, contact your Cameron Health/Boston Scientific representative or the
customer service department for instructions and return packaging.
•
The display on the programmer is made of glass or acrylic and could break if the programmer is
dropped or if it receives significant impact. Do not use if screen is broken or cracked as this could
cause injury .
•
Contact your Cameron Health/Boston Scientific representative or the customer service
department for assistance or a replacement part.
Electrostatic Discharge (ESD)
The programmer may be affected by ESD. If ESD occurs and the programmer’s functionality is
affected, attempt to reset the programmer or contact your Cameron Health/Boston Scientific
representative or the customer service department for instructions.
•
Do not touch or connect the telemetry wand to the programmer unless ESD precautionary
procedures are used.
S-ICD System Warnings and Cautions
Before using the S-ICD System, read and follow all warnings and cautions provided in this manual. Refer
to the SQ-RX Pulse Generator, Q-TRAK Electrode or Q-GUIDE Electrode Insertion Tool User’s Manuals for
information on implanting and explanting the S-ICD System.
The S-ICD System contains sterile products for single use only. Do not resterilize. Handle the
components of the S-ICD System with care at all times and maintain proper sterile technique.
General
•
External defibrillation equipment should be available for immediate use during the
implantation procedure and follow-up.
•
Placing a magnet over the SQ-RX Pulse Generator suspends arrhythmia detection and therapy
response. Removing the magnet resumes arrhythmia detection and therapy response.
•
Battery depletion will eventually cause the SQ-RX Pulse Generator to stop functioning.
Defibrillation and excessive numbers of charging cycles shorten the SQ-RX Pulse Generator’s
battery longevity.
3
Q-TECH™ PROGRAMMER: general description
•
The S-ICD System has not been evaluated for pediatric use.
•
The S-ICD System does not provide long-term bradycardia pacing, Cardiac Resynchronization
Therapy (CRT) or Anti-Tachycardia Pacing (ATP).
Use of Other Medical Therapies/Diagnostic Procedures
4
•
External defibrillation or cardioversion may damage the S-ICD System. Avoid placing the
defibrillation paddles directly over the SQ-RX device or electrode.
•
Cardio Pulmonary Resuscitation (CPR) may temporarily interfere with sensing and may cause
delay of therapy.
•
Do not expose a patient with an implanted S-ICD System to diathermy. The interaction of
diathermy therapy with an implanted SQ-RX Pulse Generator can damage the SQ-RX Pulse
Generator and cause patient injury.
•
Do not expose the patient to MRI scanning. MRI scanning can damage the SQ-RX Pulse
Generator and cause patient injury.
•
Electrical interference or “noise” from sources such as electrosurgical and monitoring
equipment can interfere with the communication between the programmer and SQ-RX Pulse
Generator or cause inappropriate therapy. If interference occurs, move and reposition the
programmer or wand away from the source of the interference.
•
Ionizing radiation therapy, such as radioactive cobalt, linear accelerators, and betatrons
may adversely affect the S-ICD System operation. Therapeutic ionizing radiation cannot be
immediately detected; however, it can damage the electronic components of the SQ-RX Pulse
Generator. To minimize the risks of ionizing radiation:
»»
Shield the SQ-RX Pulse Generator with a radiation-resistant material, regardless of the
distance between the SQ-RX Pulse Generator and the radiation beam.
»»
Do not project the radiation port directly at the SQ-RX Pulse Generator.
»»
Evaluate the S-ICD System operation after each radiation treatment.
•
Lithotripsy and other therapeutic forms of ultrasound can damage the SQ-RX Pulse Generator.
If required, avoid direct flow of the pulse waves near the site of the implanted SQ-RX Pulse
Generator.
•
Use caution during ablation procedures. Program the S-ICD System to Therapy Off. Keep the
current path (electrode tip to ground) as far away as possible from the implanted SQ-RX Pulse
Generator and electrode.
•
The programmer is not suitable for use in the presence of a flammable anesthetic mixture with
air or with oxygen or nitrous oxide.
Q-TECH™ PROGRAMMER: general description
Electromagnetic Interference (EMI) Outside the Hospital Environment
Exposure to EMI or static magnetic fields may suspend tachycardia detection and possibly cause
temporary inhibition of therapy delivery. EMI may also trigger delivery of a shock in the absence
of a tachycardia. Automatic sensing and detection of tachycardias will resume when the patient
moves away from the EMI or static magnetic field source.
To minimize the risk, advise patients to avoid sources of EMI or static magnetic fields having
strengths of >10 gauss or 1 mTesla.
•
•
Sources of EMI include, but are not limited to:
»»
High-voltage power lines
»»
Arc welding equipment
»»
Electrical smelting furnaces
»»
Large radio-frequency transmitters (such as radar)
»»
Alternators on running engines in automobiles
»»
Communications equipment (such as high-power radio transmitters)
Sources of strong static magnetic fields may include the following:
»»
Industrial transformers and motors
»»
Large stereo speakers
»»
Magnetic wands, such as those used for airport security
Patients should seek medical guidance from their physician before entering an area where a posted
sign prohibits patients with an implantable cardioverter defibrillator or pacemaker.
Potential Adverse Events
Potential adverse events related to implantation of the S-ICD System may include, but are not
limited to, the following:
•
Acceleration/induction of atrial or
ventricular arrhythmia
•
Bleeding
•
Adverse reaction to induction testing
•
Conductor fracture
•
Allergic/adverse reaction to system or
medication
•
Cyst formation
•
Death
•
Delayed therapy delivery
5
Q-TECH™ PROGRAMMER: general description
•
Discomfort or prolonged healing of
incision
•
Infection
•
Electrode deformation and/or breakage
•
Keloid formation
•
Electrode insulation failure
•
Migration or dislodgement
•
Erosion/extrusion
•
Muscle stimulation
•
Failure to deliver therapy
•
Nerve damage
•
Fever
•
Pneumothorax
•
Hematoma
•
Post-shock/post-pace discomfort
•
Hemothorax
•
Premature battery depletion
•
Improper electrode connection to the
pulse generator
•
Random component failures
•
Stroke
•
Inability to communicate with the pulse
generator
•
Subcutaneous emphysema
•
•
Inability to defibrillate or pace
Surgical revision or replacement of the
system
•
Inappropriate post-shock pacing
•
Syncope
•
Inappropriate shock delivery
•
Tissue redness, irritation, numbness or
necrosis
If any adverse events occur, invasive corrective action and/or S-ICD System modification or removal
may be required.
Patients who receive an S-ICD System may also develop psychological disorders that include, but
are not limited to, the following:
6
•
Depression
•
Fear of shocks
•
Phantom shocks
Q-TECH™ PROGRAMMER: Operation
Setup of the Q-TECH Programmer
Packaging
Programmer components include:
•
Model 3200 Programmer with pre-loaded software
•
Model 3203 Telemetry Wand
•
External power supply and AC power cord
Visually inspect the packaging to ensure the contents are complete. Do not use if there is evidence
of damage.
In case of damage return the product to Cameron Health/Boston Scientific. Contact your local
representative or the Cameron Health/Boston Scientific customer service department for
instructions and return packaging.
Charging the Programmer
The programmer is primarily intended to be operated while connected to the AC-powered external
power supply, but may also be operated on battery power provided that the internal battery is
adequately charged. The programmer is recharged whenever it is connected to the AC-powered
external power supply. When not in use, it is recommended that the programmer remain connected
to the external power supply in order to maintain an adequate battery charge.
Typical charge time for a fully discharged battery is 5 hours. However, more time may be required if
the programmer is in use while being recharged.
The Battery Status indicator located on the upper right corner of the screen displays the status of
the main battery power when the unit is in use:
•
All four bars are illuminated (green) - The battery is 100% charged
•
Three bars are illuminated (green) - The battery is 75% charged
•
Two bars are illuminated (yellow) - The battery is 50% charged
•
One bar is illuminated (red) - The battery is 25% charged
7
Q-TECH™ PROGRAMMER: Operation
The programmer displays one of the following alert screens as battery power gets progressively lower.
•
Programmer Battery Low
•
Programmer Battery Critical
•
Out Of Power
To charge the programmer:
1.
Connect the external power supply cable to the programmer (Figure 1).
External power supply port
Figure 1: Connecting the external power supply
2.
Plug the external power supply cord into an AC power outlet.
Using the Q-TECH Programmer
Turning the Programmer On
The programmer power button is located in the recess above and behind the left corner of the
screen. Press and hold the button until the display screen is active (Figure 2).
8
Q-TECH™ PROGRAMMER: Operation
Power On/Off
Figure 2: Turning the programmer On/Off
Note: If the programmer cannot be turned on while it is connected to AC power via
the external power supply, first unplug the external power supply cord from
the programmer. Press and hold the programmer power button until the
display screen is active (Figure 2). AC power via the external power supply can
then be reconnected.
Placing the Programmer in Suspend Mode
The programmer has a Suspend Mode which is activated automatically to conserve power. The
display will be blank when this mode is in effect.
The programmer enters Suspend Mode whenever:
•
The power button (Figure 2) is momentarily pressed and released
•
The programmer is not connected to the external power supply, it is not in active
communication with an SQ-RX pulse generator, and no user activity has occurred for 15
minutes
Momentarily pressing the power button will resume normal operation.
9
Q-TECH™ PROGRAMMER: Operation
Turning the Programmer Off
There are two ways to turn the programmer off:
1.
Press and hold the power button (Figure 2) until the System shutdown menu
appears. Select Power off from the popup and confirm by pressing OK.
2.
From the on-screen Main Menu, press the Power Off button and select OK at the
confirmation prompt.
Using the Programmer Touch Screen
The programmer is equipped with an LCD touch screen. The screen can be adjusted to the desired
viewing angle by using the kick-stand located on the back of the programmer. All interaction with
the programmer is conducted using the fingers to touch the appropriate areas on the screen.
Scroll on-screen lists by sliding a finger up and down the list (Figure 3). An on-screen keyboard is
presented whenever text entry is required.
Figure 3: Scrolling on-screen lists
Using the Wand
Model 3203 wand (“the wand”) makes it possible for this programmer to communicate with the
SQ-RX device.
Note: The wand is a non-sterile device. DO NOT STERILIZE the wand or the
programmer. The wand must be contained in a sterile barrier before use in
the sterile field.
10
Q-TECH™ PROGRAMMER: Operation
To connect the wand to the programmer, slide the wand cable connector over the communication
connector port located on the rear edge of the programmer (Figure 4).
To disconnect the wand, grasp the wand cable connector and gently pull it straight off the
communication connector port.
Note: Do not pull or yank on the cable to disconnect the wand from the
programmer. Such action could cause hidden damage to the cable. A
damaged cable might reduce wireless communication capabilities and
require a replacement wand.
Figure 4: Connecting the wand
To obtain optimal telemetry, the wand should be placed directly over the implanted SQ-RX device.
Although the programmer may receive S-ECG telemetry at greater distances, programming should
be performed with the wand placed over the SQ-RX device.
When telemetry loss occurs, the display screen will turn yellow and a message will appear with the
text “Communication Loss” to alert the user. Reposition the wand to establish communication. The
programmer will display a screen informing the user if the SQ-RX device is found and programming
can continue.
Note: If communication cannot be reestablished, the session should be ended and
restarted by scanning for the SQ-RX device.
11
Q-TECH™ PROGRAMMER: Operation
Navigation
The programmer’s graphic user interface (GUI) facilitates management and control of the S-ICD
System. The Navigation Bar and on-screen icons at the top of the screen allow the user to navigate
programming software screens. In addition, a continuous subcutaneous electrocardiogram (S-ECG)
is displayed along the bottom of the screen during Online (active) communication with the SQ-RX
device.
Screen Header
When the programmer is Offline (inactive communication), the screen header displays the Battery
Status Indicator.
When viewing Offline Stored Sessions, the screen header displays:
•
Patient name
•
Therapy On/Off
•
Battery status indicator
When the programmer is online (active communication), the screen header displays:
•
Therapy On/Off
•
Patient name
•
Patient heart rate
•
Programmer Battery and Telemetry status indicator
•
Screen title
•
Rescue shock icon
Navigation Bar
The Navigation Bar is the primary method for navigating the Online programmer screens. The bar
is located along the top edge of the programmer screen and chosen screens appear with their
selection icon highlighted.
Table 1 displays a list of the programmer icons and their corresponding descriptions.
12
Q-TECH™ PROGRAMMER: Operation
Table 1: Icon descriptions
Icon
Description
User Application

Main Menu Button
Allows user to return to the main menu.

Auto Setup Button
Allows user to access the auto setup menu.

Device Settings Button
Allows user to access the SQ-RX device settings screen.

Device Status Button (open folder and closed folder)
Allows user to access the SQ-RX device status screen. User can view number of shocks
delivered since the last update as well as the SQ-RX device battery life.

Patient View Button
Allows user to access the patient chart screen.
Captured and Stored Episodes S-ECG Button
Allows user to access captured S-ECG and stored episode screens.

Induction Test Button
Allows user to access induction screen.

Manual Shock Button
Allows user to access the manual shock screen.

Battery & Telemetry Meter
Left side of the meter allows user to view the programmer’s battery status.
The right side of the meter allows viewing of telemetry signal.

Capture S-ECG
Allows user to capture a live S-ECG.

S-ECG Display Settings
Allows user to modify the zoom and sweep speed on the live S-ECG.

Heart Rate Icon
Allows user to view current heart rate.

Rescue Button
Allows user to administer a shock from any programmer screen.
Option Selection Slider Switch
Allows user to select one of two options, e.g. A or B

13
Q-TECH™ PROGRAMMER: Operation
Restarting the Programmer
The programmer’s operating system is self-monitoring and is generally able to sense many system
error conditions and automatically initiate a restart sequence in response. Follow the on-screen
instructions to complete the programmer-initiated restart sequence.
The programmer may need to be manually restarted if:
•
You cannot exit a screen
•
The operating system stops responding
A manual restart is accomplished by pressing and holding the power button until the system
shutdown menu appears on the screen. Select Restart from the popup and confirm
by pressing OK.
If the programmer does not respond to a restart process, contact your Cameron Health/Boston
Scientific/Boston Scientific representative or the customer service department for assistance.
14
Q-TECH™ PROGRAMMER: Operation
Configuring the Q-TECH Programmer
Configuring Programmer Settings
The programmer should be configured before communication with an SQ-RX device is attempted.
This includes setting the date and time format, time zone, language and printer. Once these settings
are configured during the initial setup process they become the default parameters and will not
normally need to be changed with each session.
To configure the programmer settings:
1.
Select the Programmer Settings button on the Startup screen (Figure 5) to
display the Programmer Settings screen (Figure 6).
Figure 5: Startup screen
15
Q-TECH™ PROGRAMMER: Operation
Figure 6: Programmer Settings screen
2.
16
Select the corresponding line to access each setting. The settings that can be
configured include:
•
Date and time format
•
Time zone
•
Language
•
Printer
Q-TECH™ PROGRAMMER: Operation
Date and Time Format
To set the date and time format:
1.
Select Set Date and Time Format on the Programmer Settings screen (Figure 6).
The Date and Time Settings screen appears (Figure 7).
2.
Select the desired date format.
3.
Select the Save button to save the changes and return to the Programmer Settings screen,
or select Cancel to return to the Programmer Settings screen without saving the changes.
Figure 7: Date and Time Settings screen
Time Zone
To set the time zone in which the programmer will be used:
1.
Select Set Time Zone on the Programmer Settings screen. The time zone selection screen appears (Figure 8).
2.
Select the country/time zone button for the zone in which the programmer will
be used. A checkmark will appear in the selected button. The SQ-RX device line
frequency of 50 Hz or 60 Hz is automatically programmed based on the country/time
zone selected.
17
Q-TECH™ PROGRAMMER: Operation
Figure 8: Time Zone selection screen (scrollable list)
3.
Select the Save button to save the changes and return to the Programmer Settings
screen, or select Cancel to return to the Programmer Settings screen without saving
the changes.
Language Preference
To set the language preference:
1.
Select Set Language on the Programmer Settings screen. The Language Settings
screen appears.
2.
Select the Save button to save the changes or select Cancel to return to the
Programmer Settings screen without saving the changes. If the language is changed
the programmer will automatically restart and return to the Startup screen.
Printer Selection
Only Cameron Health/Boston Scientific-approved printers can be used with the programmer. To
select the printer to be used with the programmer:
1.
18
Ensure the printer is on and, as applicable, the wireless function is enabled or the
wireless adapter is in the printer’s USB port.
Q-TECH™ PROGRAMMER: Operation
2.
Select Printer Setup on the Programmer Settings screen. A previously configured
printer becomes the default printer and will be displayed at this time. If a default
printer has not already been selected and configured, the programmer will scan the
area to locate wireless printers. A Scan Progress Bar will appear informing the user
that the programmer is currently scanning for printers.
3.
Select the printer of choice (Figure 9). If no printers are found, a window will appear
stating that there are no printers. Select the Scan Again button or the Cancel button
to return to the Programmer Settings screen. If a printer was selected during another
session, the Printer Setup screen is displayed.
Figure 9: Printer Selection screen
4.
Once the scan is complete, the Printer Selection screen appears (Figure 9).
Note: Refer to “Troubleshooting” section for printer problems.
5.
Select the desired printer name and enter the name using the on-screen keyboard
(up to 15 characters). A unique printer identifier should appear with the printer
selection (Figure 10).
19
Q-TECH™ PROGRAMMER: Operation
Figure 10: Use the on-screen keyboard to enter a name for the selected printer
6.
Select the Save button to save the changes and return to the Programmer Settings
screen, or select Cancel to return to the Programmer Settings screen without saving
the changes. A confirmation screen will appear when the printer setup is completed
(Figure 11).
Figure 11: Printer Setup confirmation screen
20
Q-TECH™ PROGRAMMER: Operation
Programmer Software Version
To view the programmer’s software version:
1.
Select About Programmer on the Programmer Settings screen. The Programmer
Software Version information screen appears (Figure 12).
Figure 12: Programmer Software Version information screen
2.
This screen displays the current version of the programmer software. Select the
Continue button to return to the Programmer Settings screen.
Note: The patient printed report contains the programmer software version.
Modes of Operation for the Q-TECH Programmer
Online Behavior
The programmer’s interface varies according to whether the programmer is Online (actively
communicating) or Offline (not communicating) with a selected SQ-RX device.
An Online session begins when the programmer establishes a telemetry link with a specific SQ-RX
device. A yellow alert screen is displayed if the telemetry signal is lost between the programmer and
the SQ-RX device for more than five seconds during active communication. This may occur if the
21
Q-TECH™ PROGRAMMER: Operation
wand is moved out of the telemetry communications range or if noise or interfering objects inhibit
communication. Programming commands, including Rescue Shocks, will not be available until
telemetry is reestablished.
Telemetry reconnection may occur automatically within one minute when the SQ-RX device and
wand are within telemetry range.
Note: Whenever the programmer is in active communication with an SQ-RX
device, charging of the pulse generator in preparation for delivering a shock
(whether commanded or in response to a detected arrhythmia) is indicated
by an audible notification. The notification continues until the shock is
delivered or aborted.
Offline Behavior
When the programmer is not actively communicating with an SQ-RX device it is Offline. During
Offline sessions, programmer settings can be accessed, and stored patient sessions can be viewed
and/or printed.
Stored Patient Sessions
During a patient follow-up visit, the programmer will retrieve data from the SQ-RX device’s
memory. The programmer can store up to 15 patient sessions. When the 16th session occurs,
the programmer will automatically replace the oldest stored session with the new data. A stored
session includes the following information:
•
Captured S-ECG Reports
•
Episode History (including any downloaded episodes)
•
Patient Data
•
Programmed Device Settings
To view stored patient sessions:
22
1.
From the Startup screen, select Stored Patient Sessions.
2.
Select the desired patient session.
Q-TECH™ PROGRAMMER: Operation
Modes of Operation for the SQ-RX Device
The SQ-RX device has three modes of operation:
•
Shelf
•
Therapy On
•
Therapy Off
Shelf Mode
The Shelf mode is a low power consumption state intended for storage only. When communication
is initiated between the SQ-RX device and the programmer, a full-energy capacitor reformation is
automatically performed and the SQ-RX device is prepared for set-up. Once the SQ-RX device is
taken out of Shelf mode, it cannot be reprogrammed back into Shelf mode.
Therapy On Mode
The Therapy On mode is the primary operating mode of the SQ-RX device, allowing automatic
detection of, and response to, ventricular tachyarrhythmias.
Note: The SQ-RX device must be programmed out of Shelf mode before being
programmed to Therapy On.
Therapy Off Mode
The Therapy Off mode disables automatic therapy delivery and enables manual control of shock
delivery. Programmable parameters may be viewed and adjusted via the programmer. Also, the
subcutaneous electrogram (S-ECG) may be displayed or printed.
The SQ-RX device automatically defaults to Therapy Off mode when the SQ-RX device is taken out
of Shelf mode.
Note: Manual and rescue shock therapy are available once the initial Setup process
is complete.
23
Q-TECH™ PROGRAMMER: Operation
Connecting and Disconnecting from the SQ-RX Device
This section provides the information necessary for selecting, connecting to, and disconnecting from
the SQ-RX device.
Scanning For SQ-RX Devices
1.
Select the Scan For Devices button on the Startup Screen (Figure 13). The Device
List screen appears after the Scan Progress Bar displayed during the scanning
process. Select the Cancel button at any time to end the scanning process.
Figure 13: Startup screen
24
Q-TECH™ PROGRAMMER: Operation
2.
When the scanning process is complete, a list of all SQ-RX devices detected (up to 16)
will be displayed on the Device List (Figure 14). The SQ-RX devices that are in Shelf
mode will be displayed as “Not Implanted.” Any SQ-RX devices that were previously
taken out of Shelf mode are displayed either as “Implanted” or with the stored patient
name.
Figure 14: Device List screen (scrollable list)
3.
If the desired SQ-RX device is not listed, select the Scan Again button to re-initiate the
scanning process. Select the Cancel button to return to the Startup screen.
Note: Refer to the Inability to Communicate with the SQ-RX Device section for
further assistance.
25
Q-TECH™ PROGRAMMER: Operation
Connecting to the SQ-RX Device
From the Device List screen, select the desired SQ-RX device to initiate the communication session.
Note: Regardless of how many SQ-RX devices are located by a scan, the user must
select an SQ-RX device from the list to begin active communication.
Connecting to an SQ-RX Device in Shelf Mode:
1.
When the SQ-RX device selection is made, the programmer connects to the
selected SQ-RX device. A window will appear indicating connection is in process.
2.
Once communication is established with the SQ-RX device, the Device Identification
screen appears (Figure 15).
Figure 15: Device Identification screen
Note:
3.
26
The Device Identification screen is visible only while connecting to an SQ-RX device in Shelf
mode.
The SQ-RX device model and serial numbers are automatically acquired and
displayed during the initial scanning process. Select Continue to remove the SQ-RX
device from Shelf mode and prepare for implantation, or select Cancel to return to
the Device List screen.
Q-TECH™ PROGRAMMER: Operation
Ending a Patient Session
To end an Online patient session and return the programmer to its Offline operation mode:
1.
Select the Main Menu button on the Navigation Bar. The Main Menu screen appears.
2.
Select the End Session button (Figure 16).
Figure 16: Main Menu screen
3.
4.
A warning message will be displayed to the user if any of the following conditions
exists:
•
Automatic Therapy is programmed to Off
•
Reference S-ECG has not been acquired
•
Automatic Setup or Optimization has not been completed. This message is
typically displayed following the implant procedure as Setup Optimization was not
performed (Figure 17).
Select the Continue button to end the patient session and return to the Startup Menu
screen, or select Cancel to remain online and return to the Main Menu screen.
27
Q-TECH™ PROGRAMMER: Operation
Figure 17: Session incomplete message
Note: Once the Continue button is selected, the session is stored and
communication is terminated.
Note: In order to confirm that Automatic Therapy is programmed On upon
disconnection, always use the End Session process and review all displayed
warning messages.
28
Q-TECH™ PROGRAMMER: Operation
Programming the SQ-RX Device at Implant
This section provides the information necessary for programming the SQ-RX device during an implant
setting.
Entering Electrode Information
The programmer maintains information on the implanted electrode. To record this information for a
patient’s new or replacement electrode:
1.
Select the Main Menu button.
2.
Select the Implant button.
3.
Select the Automatic Setup button in the Navigation Bar.
4.
Select Set Electrode ID button (Figure 18).
Figure 18: Select the Set Electrode ID button to enter electrode information
Note: ECG and heart rate information is not present on the Automatic Setup and
Electrode ID setup screens until the electrode has been connected to the SQRX device.
5.
Enter the electrode model and serial number (Figure 19).
29
Q-TECH™ PROGRAMMER: Operation
Figure 19: Entering electrode information
6.
Select the Program button to save the information. A confirmation screen will appear
during communication with the SQ-RX device (Figure 20). Select Cancel to cancel
information storage and return to the Automatic Setup screen.
Figure 20: Confirmation screen showing communication in progress
30
Q-TECH™ PROGRAMMER: Operation
Creating the Patient Chart
This chart contains reference information for the patient. To set up the patient chart:
1.
Select the Main Menu button on the Navigation Bar.
2.
Select the Implant button (Figure 21).
Figure 21: Select the Implant button to create a patient chart
3.
Select the Patient View icon to access the Patient View screen.
4.
The SQ-RX device model and serial numbers appear on the first line of the chart.
The electrode model and serial numbers appear on the second line of the chart. The
implant date appears on the third line of the chart (Figure 22). Using the on-screen
keyboard, enter the following patient information:
•
Patient Name:
up to 25 characters
•
Doctor Name:
up to 25 characters
•
Doctor Info:
up to 25 characters
•
Notes:
up to 100 characters
31
Q-TECH™ PROGRAMMER: Operation
Figure 22: Patient Chart screen
Note: The Notes field will automatically wrap the text with the presence of a space
between any characters within the first line.
5.
Select the Save button to update the SQ-RX device with the patient information.
Note: Failure to save the new patient information will result in loss of the entered
data.
32
Q-TECH™ PROGRAMMER: Operation
Automatic Setup
Before the SQ-RX device can be activated, it must go through an initial Automatic Setup Process at
the time of the implant.
The Automatic Setup Process is initiated as follows:
1.
Select the Main Menu button.
2.
Select the Implant button.
3.
Select the Auto Setup button on the Navigation Bar. The Automatic Setup screen
appears (Figure 23).
4.
Select Continue if the patient’s heart rate is less than 130 bpm. For rates greater than
130 bpm, select the Cancel button and refer to the instructions for Manual Setup.
Figure 23: Automatic Setup screen
5.
Once started the Automatic Setup will:
•
Perform the shock electrode integrity check to measure electrode impedance
(Figure 24). Normal sub-threshold impedance range is < 400 Ohms.
•
Select the best sensing configuration. The sense electrode configuration appears on
the printed report and can be viewed via the Manual Setup process.
•
Select the appropriate gain selection. The selected sense gain appears on the
Printed Report and can be viewed via the Manual Setup process.
33
Q-TECH™ PROGRAMMER: Operation
Figure 24: Measuring electrode impedance
The progress of the Automatic Setup process is shown in the status bar. When each
function is complete, the arrow next to the function moves to a down position.
6.
The Automatic Setup Sensing Optimization screen will be displayed. The programmer
will display a message requesting that the patient sit up; however, this process can be
omitted during the implantation procedure by selecting the Skip button (Figure 25).
Figure 25: Automatic Setup screen
34
Q-TECH™ PROGRAMMER: Operation
7.
Select the Continue button to finish the Automatic Setup process. A confirmation
screen will appear when Automatic Setup is complete (Figure 26).
Figure 26: Confirmation of successful Automatic Setup
8.
Following the optional optimization process, the Acquire Reference S-ECG screen is
displayed (Figure 27). Select the Continue button to acquire a reference S-ECG.
Figure 27: Acquire Reference S-ECG screen
35
Q-TECH™ PROGRAMMER: Operation
9.
Once the Reference S-ECG acquisition process begins, a status screen appears. The
process may take up to one minute, during which the patient should remain still.
During this process, a template of the patient’s baseline QRS complex is stored in the
SQ-RX device. Select Cancel at any time to end Reference S-ECG acquisition. When
acquisition is complete, select the Continue button.
Programming Therapy Parameters
Once Automatic Setup has been completed, the SQ-RX device therapy parameters may be selected.
Note: Data from clinical testing of the S-ICD System demonstrated a significant
reduction in inappropriate therapy with the activation of the Conditional
Shock Zone prior to hospital discharge.
To set the therapy parameters:
1.
Select the Main Menu button on the Navigation Bar.
2.
Select the Implant button.
3.
Select the Device Settings button on the Navigation Bar to display the Device
Settings screen (Figure 28).
Figure 28: Device Settings screen
36
Q-TECH™ PROGRAMMER: Operation
4.
Select the On/Off Therapy switch to set the desired therapy mode (Figure 29).
Figure 29: Setting the On/Off Therapy switch
5.
Select and drag the Conditional Shock Zone (yellow) and Shock Zone (red) slider bars
to set the desired zone configuration.
•
•
6.
7.
The Shock Zone is programmable between 170 and 250 bpm in steps of 10 bpm.
The Conditional Shock Zone is programmable between 170 and 240 bpm in steps
of 10 bpm. Enhanced detection criteria are automatically enabled when the Conditional Shock Zone is programmed.
• When programming both the Shock Zone and Conditional Shock Zone, maintain
at least a 10 bpm difference between the two zones. If the Conditional Shock Zone
slider (yellow) is dragged over the Shock Zone slider (red), the two sliders will merge
to create a single Shock Zone.
If post-shock pacing is desired, slide the Post Shock Pacing switch to the On position.
(Post-shock bradycardia pacing occurs at a non-programmable rate of 50 bpm for up
to 30 seconds. Pacing is inhibited if the intrinsic rate is greater than 50 bpm.)
Select the Program button to apply the changes and program the SQ-RX device.
A screen will appear to confirm that the SQ-RX device settings were programmed
(Figure 30). If the SQ-RX device does not accept the SQ-RX device programming, a
Program Device alert screen will appear. Select the Try Again button to return to the
Device Settings screen.
37
Q-TECH™ PROGRAMMER: Operation
Figure 30: Confirmation of programming
8.
Once programming is confirmed, select the Continue button to proceed to the next
operation.
Note: If any of the SQ-RX device settings are changed from the Device Settings
screen and not subsequently programmed, the Pending Program Changes
screen will appear. You may select Cancel to return to the Device Settings
screen or select Continue to cancel all SQ-RX device setting changes.
38
Q-TECH™ PROGRAMMER: Operation
Defibrillation Testing
Once the SQ-RX device is implanted and Automatic Therapy is programmed On, defibrillation
testing may be conducted.
Note: Defibrillation testing is recommended at implant to confirm the ability of the
S-ICD System to sense and convert VF.
Note: Episode data associated with rescue shocks, manual shocks, and induction
testing is not stored in the SQ-RX device.
To induce VF and test the S-ICD System:
1.
Select the Programmer Main Menu button on the Navigation Bar to access the Main
Menu.
2.
Select the Patient Test button to setup the induction test (Figure 31).
Figure 31: Select the Patient Test Button to access the Induction Test screen.
39
Q-TECH™ PROGRAMMER: Operation
3.
Select either standard (STD) or reverse (REV) polarity (Figure 32).
4.
Select and drag the red marker to set the desired shock energy for the first delivered
shock. The shock energy may be programmed from 10 to 80 J. A 15 J safety margin is
recommended for defibrillation testing.
Figure 32: Setting the desired first shock energy for defibrillation testing
5.
Select the Continue button to display the next Induction Test screen.
Note: Ensure that noise markers (“N”) are not present on the S-ECG prior to
induction. The presence of noise markers may delay detection and therapy
delivery.
40
Q-TECH™ PROGRAMMER: Operation
6.
Select and hold the Hold To Induce button for the desired duration (Figure 33).
Figure 33: Induction Test screen
The following functions occur during the test:
•
The S-ICD System induces ventricular fibrillation using 200 mA alternating current
(AC) at 50 Hz. Induction continues until the Hold To Induce button is released (up to
a maximum of 10 seconds per attempt).
•
Arrhythmia detection and the Live S-ECG are suspended during AC induction. Once
the Induction button is released, the programmer displays the patient’s rhythm.
•
Upon detection and confirmation of an induced arrhythmia, the S-ICD System
automatically delivers a shock at the programmed energy output and polarity.
•
If the shock fails to convert the arrhythmia, re-detection occurs and subsequent
shocks are delivered at the SQ-RX device’s maximum energy output (80 J).
Note: Following the release of the Hold To Induce button, evaluate the sensing
markers during the induced rhythm. The S-ICD System uses a lengthened
rhythm detection period. Consistent tachy “T” markers indicate that
tachyarrhythmia detection is occurring, and that capacitor charging is
imminent. If a high degree of amplitude variation is noted during the
arrhythmia, a slight delay may be expected prior to capacitor charging or
shock delivery.
41
Q-TECH™ PROGRAMMER: Operation
Note: The SQ-RX device can deliver a maximum of five shocks per episode. At any
time, an 80 J rescue shock can be delivered by pressing the Rescue Shock
button.
7.
At any time prior to therapy delivery, the programmed energy may be aborted by
selecting the red Abort button.
8.
Select the Exit button to exit the induction process and return to the Main Menu
screen.
Performing an SQ-RX Follow-up
Sensing Configuration and Automatic Setup
It is not necessary to perform Automatic Setup at each follow-up. If Automatic Setup is performed,
resulting in a vector change, sensing should be reevaluated. After the setup process is complete,
evaluate the streaming S-ECG during a pectoral exercise. Sensing performance during high rate
exercises can also be performed. Acceptable sensing will yield “S” markers synchronous to all QRS
complexes. If other markers are noted, use the Manual Setup process to evaluate other sensing
configurations.
Note: If Manual Setup was previously used to override a sensing configuration,
careful consideration should be taken when selecting Automatic Setup.
If an update to the reference S-ECG is desired due to a change in the patient’s resting ECG, follow
the Acquire Reference S-ECG instructions.
Viewing SQ-RX Device Status
Once communication is established, the programmer displays the Device Status screen with
information regarding the current episodes and battery status of the SQ-RX device.
42
Q-TECH™ PROGRAMMER: Operation
To view this information:
1.
Select the Main Menu button.
2.
Select the Follow Up button.
3.
Select the Device Status button on the Navigation Bar.
4.
The Device Status screen will appear showing an overview of all SQ-RX device activity
since the last communication session (Figure 34).
Figure 34: Device Status screen
The Device Status overview reports:
•
Date of the last follow-up session
•
Total number of shocks delivered since the last follow-up session
•
Total number of treated episodes since the last follow-up session
•
Total number of untreated episodes since the last follow-up session
•
Remaining SQ-RX device battery life
43
Q-TECH™ PROGRAMMER: Operation
Viewing Stored Episodes
The SQ-RX device stores up to 25 treated and 20 untreated tachycardia episodes, which can be
viewed during a patient’s follow-up session. When the maximum number of episodes is exceeded,
the most recent episode replaces the oldest stored episode; however, the first treated episode will
not be overwritten.
Note: If a spontaneous episode occurs during a patient’s follow-up session while
the SQ-RX device is wirelessly communicating with the programmer, the
episode will not be stored.
To view stored episodes:
1.
Select the Main Menu button.
2.
Select the Follow Up button.
3.
Select the Captured and Stored Episodes S-ECG icon from the Navigation Bar.
4.
Select the Episodes option to access the episodes pick list.
5.
Select an episode (treated or untreated) from the list (Figure 35). The selected episode
will be downloaded from the SQ-RX device and displayed.
Figure 35: Stored Episodes screen (scrollable list)
44
Q-TECH™ PROGRAMMER: Operation
6.
Each selected Episode screen also displays the programmed parameters and the
stored S-ECG data at the time of episode declaration.
7.
Select the Continue button to return to the Episodes pick list screen.
The following details are available for each episode:
Treated Episodes
Up to 128 seconds of S-ECG data is stored for each Treated Episode:
•
Pre-episode S-ECG: Up to 44 seconds
•
First shock: Up to 24 seconds of pre-shock S-ECG and up to 12 seconds of post-shock
S-ECG
•
Subsequent shocks: 6 seconds of pre-shock and 6 seconds post-shock S-ECG
Untreated Episodes
An Untreated Episode is defined as any high-rate episode that spontaneously terminates
during the charging process, before a shock is delivered.
Up to 128 seconds of S-ECG data is stored for each Untreated Episode:
•
Pre-episode S-ECG: 44 seconds of pre-episode S-ECG
•
Episode S-ECG: Up to 84 seconds of tachycardia S-ECG data
Printing Reports from the Programmer
Printing Reports
Patient reports can be printed before or after a patient session is ended. It is recommended that
a final report be printed immediately following the implant procedure. There are three patient
reports:
•
Summary Report
•
Captured S-ECG Report
•
Episode Reports
To print patient reports from either an Online or Offline session:
1.
Select the Main Menu button to display the Main Menu screen.
2.
Select the Print Reports button.
45
Q-TECH™ PROGRAMMER: Operation
3.
Select the desired report type. A checkmark will appear next to the selected report.
Report types are described below.
4.
Select the Print button to print the selected reports.
5.
Select the Cancel button to return to the previously accessed screen.
Summary Report
The Summary Report includes the following information:
•
Patient Name
•
Date of Current Follow-Up
•
Date of Last Follow-Up
•
Printed Report Date
•
SQ-RX Device Model/Serial Number
•
Electrode Model/Serial Number
•
Implant Date
•
Therapy Parameters
•
Programmed Gain Settings and Sensing Configuration
•
Initial Shock Polarity Configuration
•
Episode Summary: Since Last Follow-Up and Since Initial Implant
•
Battery Status
•
Electrode Impedance Measurement
•
SQ-RX Device Integrity Check, if applicable
•
Programmer Software Version
•
SQ-RX Device Software Version
To print a summary report, select the Summary Report button on the Print Reports screen.
46
Q-TECH™ PROGRAMMER: Operation
Captured S-ECG Report
To print a Captured S-ECG report:
1.
Select the Captured S-ECG Report button from the Print Reports pick list screen.
A list of the captured S-ECG strips is displayed. Select the button next to the Captured S-ECG(s) to be printed. A checkmark appears next to the selected Captured
S-ECG (Figure 36).
Figure 36: Selecting Captured S-ECG Reports for printing
2.
Select Continue to return to the Print Reports pick list screen.
3.
Select the Print button to print the selected report.
4.
Select the Cancel button to return to the previously accessed screen.
47
Q-TECH™ PROGRAMMER: Operation
Episodes Report
To print an Episode Report:
1.
Select the Episodes Report button on the Print Reports pick list screen. The Episode
List screen appears showing a list of the stored episodes (Figure 37). Select the button
next to the episode(s) to be printed. A checkmark appears next to the selected
episode(s).
Figure 37: Stored Episode List screen (scrollable).
2.
Select Continue to return to the Print Reports pick list screen.
3.
Select the Print button to print the selected report.
4.
Select the Cancel button to return to the previously accessed screen.
Note: In order to be available for printing, episodes must be individually
downloaded from the Stored Episodes Screen (Figure 35).
48
Q-TECH™ PROGRAMMER: Operation
S-ECG Features
The programmer provides the capability to view, adjust and capture the streaming S-ECG from the
SQ-RX device.
S-ECG Rhythm Strip Markers
The system provides annotations to identify specific events on the S-ECG. These markers are shown
in the S-ECG Markers on Programmer Display Screens and Printed Reports table (Table 2)and an
example is shown in the Sample ECG Markers illustration (Figure 38).
Table 2: S-ECG Markers on Programmer Display Screens and Printed Reports
Description
Marker
Charging*
C
Sensed Beat
S
Noisy Beat
N
Paced Beat
P
Tachy Detection
T
Discard Beat
•
Shock
Return to NSR*
* Marker present on printed report but not on programmer display screen.
Figure 38: Sample ECG markers
49
Q-TECH™ PROGRAMMER: Operation
S-ECG Scale Settings
To adjust the real-time S-ECG amplitude and display speed scale settings:
1.
Select the S-ECG Display Settings icon located to the right of the Live S-ECG window. The S-ECG Settings screen is displayed.
2.
Select and drag the Gain or Sweep Speed Scale bars as desired (Figure 39). The S-ECG
scale will change according to the selected setting. The gain setting controls the
visual gain. The programmer defaults to Full Range for SQ-RX devices with a gain
setting of x1 and to 1/2 Range for SQ-RX devices with a gain setting of x2. The Sweep
Speed slider controls the display speed of the scrolling Live S-ECG. The nominal
sweep speed setting is 25 mm/sec.
Figure 39: Adjusting Gain and Sweep Speed
Note: Amplitude settings and display speed adjustments on scrolling real-time
S-ECG and Captured S-ECGs affect the display screen settings only and have
no impact on the SQ-RX device settings for sensing.
50
Q-TECH™ PROGRAMMER: Operation
Capture and View S-ECG Strips
The programmer can display, capture and store real-time S-ECG rhythm strips. The programmer
saves a maximum of fifteen, 12-second, Captured S-ECG recordings comprised of:
•
8.5 seconds before activation of the Capture S-ECG button
•
3.5 seconds after activation of the Capture S-ECG button
If an additional recording is required, then the oldest previous recording is replaced with the new
recording.
To capture a new S-ECG rhythm strip:
1.
Select the Capture S-ECG button located to the left of the Live S-ECG window
(Figure 40). The S-ECG will scroll across the display screen. Calipers appear below the
Captured S-ECG rhythm strip (Figure 41). Each 12-second recording is date and time
stamped according to the programmer’s date and time setting.
Figure 40: Select the Capture S-ECG button
Figure 41: Captured S-ECG rhythm strip
51
Q-TECH™ PROGRAMMER: Operation
2.
Select and move the calipers across the S-ECG strip to measure intervals as desired.
3.
Select the Continue button to return to the previously accessed screen.
To view a previously Captured S-ECG:
1.
Select the Main Menu button.
2.
Select the Follow Up button.
3.
Select the Captured and Stored Episode S-ECG button. The Captured S-ECG pick list
screen appears.
4.
Select one Captured S-ECG from the list. The Captured S-ECG Details screen appears.
5.
Select and drag the calipers to view details.
6.
Select the Continue button to return to the Captured S-ECG list screen.
Utilities Menu
The programmer Utilities menu provides access to additional SQ-RX device features. These include
Acquire Reference S-ECG, Beeper Control, Manual Setup and Smart Charge.
To access the Utilities menu during an Online session:
1.
Select the Main Menu button to display the Main Menu screen.
2.
Select the Utilities button. The Utilities screen appears (Figure 42).
Figure 42: Utilities screen
52
Q-TECH™ PROGRAMMER: Operation
Acquire Reference S-ECG
To acquire a manual Reference S-ECG:
1.
Select the Acquire Reference S-ECG button on the Utilities menu to access the
Acquire Reference S-ECG screen.
2.
Select Continue to acquire a Reference S-ECG. The programmer will begin acquiring
the Reference S-ECG. A message will appear requesting that the patient remain still
(Figure 43). The reference S-ECG QRS template is recorded and stored in the SQ-RX
device.
3.
Select the Continue button to complete the process and return to the Utilities screen.
The Cancel button can be used at any time to end S-ECG acquisition and return to the
Utilities screen.
Figure 43: The patient should remain still while the Reference S-ECG is acquired
Beeper Control
The SQ-RX device has an internal alert system that emits an audible beeping tone when an alert
condition occurs. Alert conditions may include Elective Replacement Indicator (ERI), End of Life
(EOL), electrode impedance out of range, prolonged charge times, failed Device Integrity Check and
irregular battery depletion. Beeper Control allows activation or suppression of the beeper.
53
Q-TECH™ PROGRAMMER: Operation
Once an alert condition is triggered, audible tones beep for 16 seconds every nine hours until the
alert condition is resolved. Beeper Control is automatically enabled when the SQ-RX device is taken
out of Shelf mode.
To reset the Beeper Control:
Note: The Beeper Control display screen is enabled only when an alert condition
occurs. If an alert condition is activated, a notification screen will appear
upon connection.
1.
Select the Main Menu button to display the Main Menu.
2.
Select the Utilities button to display the Utilities Menu.
3.
Select the Beeper Control button.
4.
Select the Reset Beeper button to suspend audible beeping tones triggered by an
alert condition (Figure 44). If the alert condition is not corrected, the audible beeping
tones will be reactivated during the next automatic S-ICD System self-check.
Figure 44: Resetting the Beeper
5.
The disable Beeper feature is enabled when ERI is reached or EOL occurs. If desired,
select disable Beeper when ERI or EOL occurs. Beeper tones are permanently
disabled.
Note: If beeper is disabled alert conditions will not trigger audible tones.
54
Q-TECH™ PROGRAMMER: Operation
Manual Setup
Manual Setup enables the user to perform the electrode integrity test and select the electrode
sensing configuration and gain setting in the SQ-RX device.
1.
From the Utilities Menu, select the Manual Setup button. The Impedance Test
screen appears (Figure 45).
2.
Select the Test button to perform the electrode integrity test.
Figure 45: Impedance test screen
3.
Select Continue to proceed or select Test to retry. The sensing configuration can be
evaluated by temporarily activating any of the vectors or gain settings. The Live S-ECG
and markers will reflect the temporary selections for evaluation purposes.
4.
There are three available sensing vectors that can be manually selected:
•
Primary: Sensing from the proximal electrode ring on the subcutaneous electrode
to the surface of the active SQ-RX device
•
Secondary: Sensing from the distal sensing electrode ring on the subcutaneous
electrode to the surface of the active SQ-RX device
•
Alternate: Sensing from the distal sensing electrode ring to the proximal sensing
electrode ring on the subcutaneous electrode
55
Q-TECH™ PROGRAMMER: Operation
The gain setting adjusts the sensed S-ECG signal sensitivity. It may be manually selected via the
Select Gain slider switch (Figure 46).
Figure 46: Gain adjustment
56
•
1x Gain (± 4 mV):
1x Gain should be selected when the signal amplitude is large
enough to cause clipping when the 2x gain is selected.
•
2x Gain (± 2 mV):
2x Gain should be selected when the signal amplitude is small
enough to allow use of a more sensitive setting without causing
clipping of the captured signal. The 2x gain selection amplifies
the signal twice as much as the 1x gain selection.
Q-TECH™ PROGRAMMER: Operation
To program the manually selected sense configuration:
1.
Select the Program button to save the sense vector and gain settings.
2.
The Acquire Reference S-ECG process is automatically enabled during the Manual
Setup Process. Select the Continue button to acquire a reference S-ECG. A
confirmation screen will appear when the captured reference S-ECG is acquired
(Figure 47).
3.
Select the Continue button.
Figure 47: Captured Reference S-ECG confirmation
Smart Charge
Through the Smart Charge feature, the SQ-RX device charge initiation sequence adapts to the
occurrence of non-sustained ventricular arrhythmia episodes by delaying capacitor charging. This
conserves battery longevity and may prevent unnecessary shocks for non-sustained arrhythmias.
Smart Charge is enabled automatically when an untreated ventricular arrhythmia episode is
recorded. To reset the Smart Charge feature:
1.
Select the Main Menu button.
2.
Select the Utilities button on the Main Menu to display the Utilities Menu.
3.
Select the Smart Charge button. The Reset Smart Charge screen appears (Figure 48).
57
Q-TECH™ PROGRAMMER: Operation
Figure 48: Reset Smart Charge screen
58
4.
Select the Reset button to reset the Smart Charge to zero or press Cancel to return to
the Utilities Menu without resetting the Smart Charge.
5.
A confirmation window will appear with the message: “Smart Charge successfully
reset.”
6.
Press the Continue button to return to the Utilities screen.
Q-TECH™ PROGRAMMER: Operation
Additional Programmer Functions
Rescue Shock
The Rescue Shock button is available on the programmer display when the SQ-RX device Setup
Process is complete and an SQ-RX device is actively communicating with the programmer. During
active communication, a maximum (80 J) rescue shock can be delivered upon programmer
command.
To deliver a rescue shock:
1.
Select the red Rescue Button at the top of the programmer screen. The Rescue
Shock screen appears (Figure 49).
Figure 49: Rescue Shock screen
2.
Select the Shock button to begin charging the SQ-RX device for a Rescue shock. A
red background screen with the word “Charging” will appear (Figure 50). Selecting
the Abort button will prevent delivery of a rescue shock and will return to the Device
Settings screen.
59
Q-TECH™ PROGRAMMER: Operation
Figure 50: Abort Rescue Shock
3.
A confirmation screen will appear with notification that the shock was delivered
successfully along with the corresponding shock impedance (Figure 51).
Figure 51: Confirmation of Rescue Shock delivery
60
Q-TECH™ PROGRAMMER: maintenance
If for any reason the shock could not be delivered, a red background screen will appear with a
message stating “The Shock could not be delivered.”
Note: In the event telemetry is lost, SQ-RX device commands, including Rescue
Shocks, will not be available until telemetry is reestablished.
Manual shock
Manual Shock allows the user to deliver a synchronized shock during a sinus rhythm, an atrial
rhythm or a ventricular rhythm. The shock energy level is user-configured in the 10 to 80 joule range
(Figure 52). Manual shock may also be utilized at a low energy to assess system impedance/integrity
either at implant or as warranted by patient condition. A manual shock may be administered with
the S-ICD detection zone(s) either ON or OFF.
Figure 52: Manual shock
S-ICD System Magnet Model 4520
Model 4520 Magnet (the “magnet”) is a nonsterile accessory used to inhibit the delivery of therapy
from the SQ-RX device. Apply the magnet flat against the skin directly over the implanted SQ-RX
device for a minimum of one (1) second to suspend arrhythmia detection. Removal of the magnet
will return the SQ-RX device to normal operation. If the magnet is applied during an episode, the
episode will not be stored in the SQ-RX device memory.
61
Q-TECH™ PROGRAMMER: maintenance
Other behaviors of magnet application:
•
Inhibit shock therapy delivery
•
Terminate post-shock pacing therapy
•
Prohibit arrhythmia induction testing
•
Activate the SQ-RX device’s beeper with each detected QRS complex for 60 seconds
Note: A programmer commanded Rescue Shock can override the use of the magnet
as long as the magnet was in place prior to the initiation of the programming
command. If the magnet is applied after the initial command, the Rescue
Shock will be terminated.
Note: Magnet application does not affect wireless communication between the SQRX device and the programmer.
Charging the Programmer
When not in use, it is recommended that the programmer remain connected to the external power
supply. This will ensure that the internal battery is adequately charged.
Cleaning the Programmer
Keep the programmer away from dust and dirt. Do not use harsh chemicals, cleaning solvents, or
strong detergents to clean the programmer and wand.
To clean the programmer and wand as needed:
62
1.
Turn the Programmer Off.
2.
Gently wipe the programmer screen with a soft, clean, dry cloth.
3.
Clean the programmer plastic case and the wand by wiping them with an isopropyl
alcohol-moistened cloth.
4.
Dry the programmer immediately to remove residue.
Q-TECH™ PROGRAMMER: maintenance
Service
There are no user-accessible or user-serviceable parts or components in the programmer. If any
service, repair, or replacement of internal compontents is needed, the programmer must be
returned to Cameron Health/Boston Scientific. Contact your Cameron Health/Boston Scientific
Representative or the customer service department for instructions and return packaging.
When requesting service, please provide information concerning the nature of the failure and the
manner in which the equipment was used when the failure occurred. The model number and serial
number should also be provided. Before returning any products to the factory, call Cameron Health/
Boston Scientific customer service or your local representative for a Return Authorization number.
Maintenance Check
Prior to each use, you should perform a visual inspection and verify the following:
•
Mechanical and functional integrity of the programmer, cables, and accessories.
•
Legibility and adherence of the programmer labels.
•
Startup screen appears a few seconds after you turn on the programmer. (The normal powerup process verifies that the programmer has passed its internal checks and is ready for use.)
Safety Measurements
National regulations may require that the user, manufacturer, or manufacturer representative
periodically perform and document safety tests of the programmer. If such testing is required in
your country, follow the testing interval and extent of testing as regulated in your country. If you
do not know the national regulations in your country, please contact your Cameron Health/Boston
Scientific representative. If IEC/EN 62353 is a required standard in your country, but no specific
testing or interval is specified, it is recommended that you perform these safety tests using the
direct method as specified in IEC/EN 62353 at an interval of every 24 months. Test values are shown
in the Nominal Specifications table (Table 10).
Programmer End of Life
The programmer and accessories are designed to provide years of service under typical use. To
dispose of, return, or exchange a programmer, contact your Cameron Health/Boston Scientific
representative or the customer service department for instructions and return packaging. Do not
dispose of the programmer in the trash or at electronics recycling facilities.
63
Q-TECH™ PROGRAMMER: troubleshooting
troubleshooting
This section presents potential programmer issues and the possible solutions. Of note, restarting
the programmer can often resolve many of the issues listed below. The programmer can be
restarted by pressing and holding the power button until the system shutdown menu appears and
then choosing the “Restart” option.
Contact your Cameron Health/Boston Scientific representative or the customer service department
for additional assistance.
Inability to Print
If unable to print, follow the steps below:
1.
Ensure that the printer is turned On and that it contains paper and a sufficient ink
supply.
2.
Check printer feed for paper jam.
3.
Ensure, as applicable, that the wireless function is enabled on the printer or that the Cameron
Health/Boston Scientific approved wireless adapter is fully inserted into the left side of the
printer.
No Printer Available
The No Printer Available screen will appear if a printer was not set up. Select the Try Again button or
refer to the Printer Selection section for instructions.
Touch Screen Inactive while Connected to AC Power
If the touch screen does not function while the programmer is connected AC power via the external
power supply, disconnect and reconnect the external power supply and restart the programmer.
64
Q-TECH™ PROGRAMMER: troubleshooting
Loss of Communication with Printer
When communication between the programmer and the printer fails, a Printing Error screen will
appear with a message stating “Error while printing reports. Press “Continue” to try printing any
remaining reports, or “Cancel” to cancel the current print job.”
If this occurs:
1.
Select the Try Again button to reconnect to the printer.
2.
Verify that the selected printer has, as applicable, the wireless function enabled or a
Cameron Health-approved wireless adapter inserted in the left side of the printer.
3.
Move the programmer closer to the printer.
4.
Move any devices and the associated cables that may be interfering with the RF
communication.
Inability to Communicate With the SQ-RX Device
If the programmer is unable to communicate with the SQ-RX device, follow the steps below:
1.
Attempt to reposition the wand.
2.
Select Scan For Devices from the Main Menu or select Scan Again from the Device List
screen to locate the desired SQ-RX device.
3.
Move any equipment and associated cables that may be interfering with RF
communication.
4.
If available, attempt to communicate using a different S-ICD System programmer
and/or wand.
5.
Apply a pulse generator magnet to the SQ-RX device to elicit beeper tones. Remove
the magnet and re-attempt communication.
65
Q-TECH™ PROGRAMMER: Compliance statements
Compliance statements
Federal Communications Commission (FCC) US
This transmitter is authorized by rule under the Medical Device Radiocommunication Service (in part
95 of the FCC Rules) and must not cause harmful interference to stations operating in the 400.150 406.00 MHz band in the Meteorological Aids (i.e., transmitters and receivers used to communicate
weather data), the Meteorological Satellite, or the Earth Exploration Satellite Services and must accept
interference that may be caused by such stations, including interference that may cause undesired
operation. This transmitter shall be used only in accordance with the FCC Rules governing the Medical
Device Radiocommunication Service. Analog and digital voice communications are prohibited.
Although this transmitter has been approved by the Federal Communications Commission, there is no
guarantee that it will not receive interference or that any particular transmission from this transmitter
will be free from interference.
FCC ID SDYCHI2020
1999/5/EC (R&TTE Directive)
The S-ICD System contains radio equipment in the frequency range 402 MHz to 405 MHz for ultra low
power active medical implants. Cameron Health hereby declares that this transmitter is in compliance
with the essential requirements and other relevant provisions of Directive 1999/5/EC. To obtain a fulltext Declaration of Conformity please contact Cameron Health/Boston Scientific using the information
given on the inside front cover or back page of this manual.
EMI/RFI
This equipment has been tested and found to comply with the applicable limits for medical devices, IEC
60601-1-2:2007 or Active Implantable Medical Device Directive 90/385/EEC.
Although this testing shows the device to provide reasonable protection against harmful interference
in a typical medical installation, there is no guarantee that interference will not occur in a particular
installation. If the device does cause harmful interference the user is encouraged to try and correct the
interference by the following measures:
• Reorient or relocate the device
• Increase the separation between the devices
• Connect the equipment to an outlet on a different circuit
• Contact your Cameron Health/Boston Scientific representative or customer service
66
Q-TECH™ PROGRAMMER: Declarations tables
Table 3: Declaration Electromagnetic Emission
The Q-TECH Model 3200 programmer is intended for use in the electromagnetic environment specified below. The customer or the user of the Q-TECH
Model 3200 should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment Guidance
RF Emissions CISPR 11
Group 1
The Q-TECH Model 3200 programmer uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF Emissions CISPR 11
Class A
Harmonic Emissions IEC 61000-3-2
Class A
The Q-TECH Model 3200 programmer is suitable for use in all establishments
other than domestic and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Voltage Fluctuations/Flicker Emissions IEC 61000-3-3
Complies
Table 4: Declaration Electromagnetic Immunity Part 1
The Q-TECH Model 3200 programmer is intended for use in the electromagnetic environment specified below. The customer or the user of the Q-TECH Model 3200
should assure that it is used in such an environment.
Immunity Test
IEC 60601
Compliance Level
Electromagnetic Environment
Test Level
Guidance
Electrostatic Discharge
± 6 kV contact
± 6 kV contact
Floors should be wood, concrete, or
(ESD)
± 8 kV air
± 8 kV air
ceramic tile. If floors are covered
IEC 61000-4-2
with synthetic material, then the
relative humidity should be at least
30%.
Electrical Fast
± 2 kV for power supply lines
± 2 kV for power supply lines
Mains power quality should be that
Transient/Burst
± 1 kV for input/output lines
± 1 kV for input/output lines
of a typical commercial or hospital
environment.
IEC 61000-4-4
Surge
± 1 kV line(s) to line(s)
± 1 kV line(s) to line(s)
Mains power quality should be that of a
IEC 61000-4-5
± 2 kV line(s) to earth
± 2 kV line(s) to earth
typical commercial or hospital environment.
<5% UT (>95% dip in UT) for 0.5 cycles
Voltage Dips, Short
<5% UT (>95% dip in UT) for 0.5 cycles
Mains power quality should be that of a
interruptions, and voltage 40% UT (60% dip in UT) for 5 cycles
40% UT (60% dip in UT) for 5 cycles
typical commercial or hospital environment.
variations on power
70% UT 30% dip in UT) or 25 cycles
70% UT (30% dip in UT) for 25 cycles
If the user of the Q-TECH Model 3200
supply input lines
<5% UT (>95% dip in UT) for 5 sec
<5% UT (>95% dip in UT) for 5 sec
programmer requires continued operation
IEC 61000-4-11
during power mains interruptions, it is
recommended that the Q-TECH Model
3200 programmer be powered from an
uninterruptible power supply or a battery.
Power Frequency (50/60
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
Hz) Magnetic Fields
IEC 61000-4-8
location in a typical commercial or hospital
environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
67
Q-TECH™ PROGRAMMER: Declarations tables
Table 5: Declaration Electromagnetic Immunity Part 2
The Q-TECH Model 3200 programmer is intended for use in the electromagnetic environment specified below. The customer or the user of the Q-TECH Model 3200
should assure that it is used in such an environment.
Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601 Test
Level
3 Vrms
150 kHz to
80 MHz
Compliance
Level
3V
3 V/m
80 Mhz to
2.5 GHz
3 V/m
Electromagnetic Environment Guidance
Portable and mobile RF communications equipment should be used no closer
to any part of the Q-TECH Model 3200 programmer, including cables, than the
recommended separation distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended Separation Distance
[ V ]√ P
d = [3,5]√ P
E
d = [ 7 ]√ P
E
d = 3,5
150 KHz to 80 MHz
1
80 MHz to 800 MHz
1
800 MHz to 2.5 GHz
1
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic
site survey,a should be less than the compliance level in each frequency range.b
Interference may occur in the vicinity of equipment marked with the following
symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
b
68
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the Q-TECH Model 3200 is used exceeds the applicable RF compliance level above, the Q-TECH Model 3200 should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting
or relocating the Q-TECH Model 3200.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m
Q-TECH™ PROGRAMMER: Declarations tables
Table 6: Recommended Separation Distances
Recommended separation distances between portable and mobile RF communications equipment and the Q-TECH
Model 3200 programmer
The Q-TECH Model 3200 programmer is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the
user of the Q-TECH Model 3200 programmer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the Q-TECH Model 3200 programmer as recommended below, according to the maximum output power of the
communications equipment.
Separation distance according to frequency of transmitter
m
Rated maximum output power
80 MHz to 800 MHz
150 KHz to 80 MHz
800 MHz to 2.5 GHz
of transmitter
[ V ]√ P
d = 3,5
W
1
[ E ]√ P
d = 3,5
1
d=
[ E7 ]√ P
1
0.01
0.117
0.117
0.233
0.1
0.369
0.369
0.738
1
1.17
1.17
2.34
10
3.69
3.69
7.38
100
11.7
11.7
23.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Table 7: EMI/RFI Information
Wireless Connection
From Programmer
Frequency
Bandwidth
FCC Designation
R&TTE Class
Programmer Wireless Adapter
and Device
403.5 MHz
< 300 KHz
CFR 47, Part 951,
MICS
Class 1
Subclass 47
Printer Adapter
2.400 - 2.4835 GHz
< 1000 KHz Maximum
CFR 47, Part 15,
Class B
Class 2
Subclass 21
69
Q-TECH™ PROGRAMMER: Specifications
Table 8: Product Guidelines
Component
Requirement
DC Power
Battery pack type
4000 mAh 3.7 volt lithium-ion battery pack
Charge time
Approximately 5 hours
Power Supply
Input
100 – 240 VAC, 50 - 60 Hz, 0.5A
Output
5.5 VDC, 3.64A
Power: 20 W
Manufacturer/Model
Elpac Power Systems MWA020005A
Environment
Operating
Storage
Temperature
15° C to +38° C
(+59° F to +100° F)
-10° C to +55° C
(+14° F to +131° F)
Relative humidity
5% to 93% maximum at 40° C,
non-condensing
5% to 93% maximum at 40° C,
non-condensing
Atmospheric pressure
50 kPa to 106 kPa
(7.252 psi to 15.374 psi)
50 kPa to 106 kPa
(7.252 psi to 15.374 psi)
Table 9: Specifications
Component
Specification
Dimensions
Width x Depth x Height
24.0 cm x 12.7 cm x 2.6 cm
9.4 in x 5.0 in x 1.0 in
Weight
.6 kg, 1.3 lbs
Standard Screen Display
WVGA, 1024 x 600 pixels, 16M TFT
70
Q-TECH™ PROGRAMMER: Specifications
Table 10: Nominal Specifications
Characteristic
Nominal
Electrical Safety Testing – IEC 60601-1:2005 / ANSI/AAMI ES60601-1:2005 allowed values
Earth resistance
100 mΩ
Earth leakage current
5 mA Normal Condition (NC)
10 mA Single Fault Condition (SFC)
Patient leakage current
100 μA Normal Condition (NC)
500 μA Single Fault Condition (SFC) (mains on applied parts)
Electrical Safety Testing – IEC 62353:2008 allowed values
Earthbond testing (Groundbond testing)
300 mΩ including power cable not exceeding 3 meters
Equipment leakage — direct method
500 μA
Patient leakage current — direct method
(Wand, BF)
<= 5000 μA
Safety Features
Defibrillator protection
to 5000 V, 400 J
71
Q-TECH™ PROGRAMMER: Definition of package label symbols
Table 11: Packaging Symbols: Q-TECH Programmer
Symbol
72
Specification
Symbol
Specification
Type BF Applied Part Symbol – Indicates
connection of Type BF applied part
Manufacturer – Name and address of the
manufacturer
European Community Representative
– Authorized representative in the EU
community
Storage Temperature – Product stored with
temperature limitations
Part Number– Component number
Radio – Non-ionizing electromagnetic
radiation
Do Not Dispose – Do not dispose of in trash
Follow instructions for use
Serial Number – Serial number of the
programmer
Risk Of Shock – Indicates the risk of electric
shock
Conformité Européenne – Product fully
complies with European Directive AIMD
90/385/EEC
Fragile: Handle with Care – Transport and
store with care
Keep Dry – Ship and store in a dry place
Date of Manufacture – Date programmer
was manufactured
Electrostatic Discharge – Identifies
a connector not tested for immunity to
electrostatic discharge
Reference Number – Catalog number
Power plug – Storage location of power cord
Door, open – Open the flap
Q-TECH™ PROGRAMMER: Definition of package label symbols
Table 12: Packaging Symbols: Q-TECH Programmer Wand
Symbol
Specification
Lot number – Batch code
Symbol
Specification
Non-Sterile – Used to indicate the
product is not sterile
73
Q-TECH™ PROGRAMMER: Warranty
Warranty
Cameron Health/Boston Scientific, Inc. offers a 2-year limited warranty on the Q-TECH Programmer
to the user or using facility. Please contact your Cameron Health/Boston Scientific office or local
representative for details. The warranty on the Q-TECH Programmer does not extend to any product that
has been tampered with, altered, or repaired by any person other than an employee of Cameron Health/
Boston Scientific organization.
74
Cameron Health, Inc.
4100 Hamline Avenue North
St. Paul, MN 55112-5798 USA
1.800.CARDIAC (227.3422)
+1.651.582.4000
www.cameronhealth.com
Guidant Europe NV/SA
Boston Scientific
Green Square,
Lambroekstraat 5D
1831 Diegem, Belgium
1.800.CARDIAC (227.3422)
+1.651.582.4000
www.cameronhealth.com
PN 104749-004 en 2013-09