Advertisement
Advertisement
Puritan Bennett™
980 Series Ventilator
Operator's Manual
0123
COVIDIEN, COVIDIEN with logo and Covidien logo are trademarks of Covidien. Third-party trademarks (“TM*”) belong to their respective owners.
The following list includes trademarks or registered trademarks of a Covidien entity in the United States and/or in other countries.
Argyle™, DAR™, F&P™, IE Sync™, PAV+™, Puritan Bennett™, Shiley™
Copyright Information
The information contained in this manual is the sole property of Covidien and may not be duplicated without permission. This manual may be revised or replaced by Covidien at any time and without notice. Ensure this manual is the most current applicable version. If in doubt, contact
Covidien’s technical support department or visit the Puritan Bennett™ product manual web page at:
While the information set forth herein is believed to be accurate, it is not a substitute for the exercise of professional judgment.
The ventilator should be operated and serviced only by trained professionals. Covidien’s sole responsibility with respect to the ventilator and software, and its use, is as stated in the limited warranty provided.
Nothing in this document shall limit or restrict in any way Covidien’s right to revise or otherwise change or modify the equipment (including its software) described herein, without notice. In the absence of an express, written agreement to the contrary, Covidien has no obligation to furnish any such revisions, changes, or modifications to the owner or user of the equipment (including its software) described herein.
Covidien is a Medtronic company. The names Covidien and Medtronic are used interchangeably throughout this manual.
Contents
3
4
5
6
Figures
7
8
9
Tables
10
11
Monitored Exhaled Tidal Volume (V
12
1
1 Introduction
1.1 Overview
This manual contains information for operating the Puritan Bennett™ 980 (PB980) Series Ventilators. Before operating the ventilator system, thoroughly read this manual.
To obtain an additional copy of this manual, contact Medtronic Customer Service or your local representative.
1.1.1 Related Documents
Medtronic makes available all appropriate information relevant to use and service of the ventilator. For further assistance, contact your local Medtronic representative.
The Puritan Bennett™ 980 Series Ventilator Operator’s Manual – Provides basic information on operating the ventilator and troubleshooting errors or malfunctions. Before using the ventilator, thoroughly read this manual.
The Puritan Bennett™ 980 Series Ventilator Service Manual – Provides information to Medtronic-trained service technicians for use when testing, troubleshooting, repairing, and upgrading the ventilator.
This chapter contains the following:
• Symbol definitions
• Safety Information, including Warnings, Cautions, and Notes
• Technical assistance information
• How to access on-screen Help
• How to access warranty information
• Serial number interpretation
• Information regarding electromagnetic susceptibility
1.2 Global Symbol Definitions
describes the symbols shown on the ventilator shipping cartons. Other symbols appearing on various labels are shown in
.
Table 1.
Shipping Carton Symbols and Descriptions
Symbol Description
CE Mark 0123: signifies compliance with Medical Device Directive 93/42/EEC
Serial number
Part number
Authorized representative
Manufacturer
1
This side up
Fragile
13
Table 1.
Shipping Carton Symbols and Descriptions (continued)
Symbol Description
Humidity limitations: 10% to 95% relative humidity, non-condensing (operation and storage)
Temperature limitations: 10°C to 40°C (50°F to 104°F) (operation) –20°C to 70°C (–4°F to
158°F) (storage)
Atmospheric pressure limitations: 70 kPa to 106 kPa (10.2 psi to 15.4 psi)
Keep dry
CSA certification mark that signifies the product has been evaluated to the applicable
ANSI/Underwriters Laboratories Inc. (UL) and CSA standards for use in the US and Canada.
This device is for sale by or on the order of a physician.
Refer to instruction manual.
1.3 Safety Information
Please pay close attention to warnings and their associated consequences, which could result in adverse events.
Adverse Events (Residual Risks):
Adverse events associated with the use of the PB980 ventilator in order of least to most clinically severe are: anxiety, delay of treatment, cross patient exposure to body fluids, oxygen saturation high, oxygen toxicity, bronchospasm, tissue damage/tissue trauma, hyperventilation, hypoventilation, thermal injury, electric shock, lung overinflation, infection, oxygen saturation low, hypoxia, barotrauma, respiratory failure, extubate (unplanned), cardiac arrest, asphyxia. Any serious incident that occurs in relation to the device (or its accessories where applicable) should be reported to Medtronic and to the national competent authority. The most frequent events are delay of treatment due to transfer to another ventilator (without harm), anxiety, and oxygen saturation low (recoverable without harm). The remaining harms are infrequent or rare.
1.3.1 Safety Term Definitions
This section contains safety information for users, who should always exercise appropriate caution while using the ventilator.
Table 2.
Safety Term Definitions
Term
WARNING
Caution
Note
Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse events) to the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the prod‐ uct.
Note
Notes provide additional guidelines or information.
14
1
1.3.2 Warnings Regarding Fire Hazards
Warning: Explosion hazard — Do not use in the presence of flammable gases. An oxygen-rich environment accelerates combustibility.
Warning: To avoid a fire hazard, keep all components of the system away from all sources of ignition (such as matches, lighted cigarettes, flammable medical gases, and heaters). Oxygen-rich environments accelerate combustibility.
Warning: In case of fire or a burning smell, immediately take the following actions if it is safe to do so: disconnect the patient from the ventilator and disconnect the ventilator from the oxygen supply, facility power, and all batteries. Provide alternate method of ventilatory support to the patient, if required.
Warning: Replacement of ventilator batteries by inadequately trained personnel could result in an unacceptable risk, such as excessive temperatures, fire, or explosion.
Warning: To minimize fire hazard, inspect and clean or replace, as necessary, any damaged ventilator parts that come into contact with oxygen.
Warning: To prevent electrostatic discharge (ESD) and potential fire hazard, do not use antistatic or electrically conductive hoses or tubing in or near the ventilator breathing system.
1.3.3 General Warnings
Warning: To ensure proper operation and avoid the possibility of physical injury, only qualified medical personnel should attempt to set up the ventilator and administer treatment with the ventilator.
Warning: In case of ventilator failure, the lack of immediate access to appropriate alternative means of ventilation can result in patient death. An alternative source of ventilation, such as a self-inflating, manually-powered resuscitator (as specified in ISO 10651-4 with mask) should always be available when using the ventilator.
Warning: Patients on mechanical ventilation should be monitored by clinicians for proper patient ventilation.
Warning: The ventilator system is not intended to be a comprehensive monitoring device and does not activate alarms for all types of conditions. For a detailed understanding of ventilator operations, be sure to thoroughly read this manual before attempting to use the ventilator system.
Warning: To prevent patient injury, do not use the ventilator if it has a known malfunction. Never attempt to override serious malfunctions. Replace the ventilator and have the faulty unit repaired by trained service personnel.
Warning: To prevent patient injury, do not make unauthorized modifications to the ventilator.
Warning: To prevent injury and avoid interfering with ventilator operation, do not insert tools or any other objects into any ventilator openings.
Warning: The audio alarm volume level is adjustable. The operator should set the volume at a level that allows the operator to distinguish the audio alarm above background noise levels. See
Section 3.8.2.6, Alarm Volume, page 79
for instructions on alarm volume adjustment.
Warning: Do not silence, disable, or decrease the volume of the ventilator’s audible alarm if patient safety could be compromised.
Warning: If increased pressures are observed during ventilation, it may indicate a problem with the ventilator.
Check for blocked airway, circuit occlusion, and run SST.
Warning: The LCD panel contains toxic chemicals. Do not touch broken LCD panels. Physical contact with a broken LCD panel can result in transmission or ingestion of toxic substances.
Warning: If the graphical user interface (GUI) display (LCD panel) is blank, experiences interference and cannot be read or is unresponsive, check the patient and then verify via patient observation and the status display that ventilation is continuing as set. See instructions at
Section 2.10.2, GUI Touch Screen Unresponsive or Blank:
15
1
to an alternative source of ventilation in accordance with your institution’s protocol.
Warning: The scalar waveforms are not intended to represent a patient physiological parameter nor a qualifiable characteristic of gas (air or O
2
) delivered to, or removed from, the human body.
Warning: The Puritan Bennett™ 980 Series Ventilator contains phthalates. When used as indicated, very limited exposure to trace amounts of phthalates may occur. There is no clear clinical evidence that this degree of exposure increases clinical risk. However, in order to minimize risk of phthalate exposure in children and nursing or pregnant women, this product should only be used as directed.
Warning:
Even though the 980 Series Ventilator meets the standards listed in Section 1.1
battery of the device is considered to be Dangerous Goods (DG) Class 9 - Miscellaneous, when transported in commerce. The 980 Series Ventilator and the associated lithium-ion battery are subject to strict transport conditions under the Dangerous Goods Regulation for air transport (IATA: International Air Transport Association),
International Maritime Dangerous Goods code for sea and the European Agreement concerning the International
Carriage of Dangerous Goods by Road (ADR) for Europe. Private individuals who transport the device are excluded from these regulations although for air transport some requirements may apply.
1.3.4 Warnings Regarding Environment of Use
Warning: Do not position the ventilator next to anything that blocks or restricts the gas inlet or cooling air circulation openings, gas exhaust port, fan intake, or alarm speaker, as this may:
• limit the air circulation around the ventilator, potentially causing overheating
• limit the ventilator’s ability to exhaust patient exhaled gas leading to potential harm
• limit the clinician’s ability to hear ventilator alarms
Warning: To avoid injury, do not position the ventilator in a way that makes it difficult to disconnect the patient.
Warning: To ensure proper operation, do not position the ventilator in a way that makes it difficult to access the
AC power cord.
Warning: Do not use the ventilator in a hyperbaric chamber. It has not been validated for use in this environment.
Warning: Do not use the ventilator in the presence of strong magnetic fields. Doing so could cause a ventilator malfunction.
Warning: Do not use the ventilator during radiotherapy (i.e. cancer treatment using ionizing radiation), as doing so could cause a ventilator malfunction.
Warning: To avoid the risk of ventilator malfunction, operate the ventilator in an environment that meets
specifications. See Table 62, page 245
.
Warning: Do not use the ventilator as an EMS transport ventilator. It has not been approved or validated for this use.
1.3.5 Warnings Before Using Equipment
Warning: Before activating any part of the ventilator, be sure to check the equipment for proper operation and,
if appropriate, run SST as described in this manual. See Section 3.9.1.2, SST Test Sequence, page 83
.
Warning: Check for leaks in the ventilator breathing system by running SST prior to ventilating a patient.
Warning: Lock the ventilator’s casters during use to avoid the possibility of extubation due to inadvertent ventilator movement.
Warning: The ventilator accuracies listed in the Ventilator Settings, Alarm Settings, and Patient Data tables in
Chapter 11 are applicable under specified operating conditions. See
. If the ventilator is operated outside specified ranges, the ventilator may supply incorrect information and the accuracies listed in the
16
1 aforementioned tables do not apply. A hospital Biomedical Technician must verify the ventilator is operated in the environmental conditions specified.
1.3.6 Warnings Regarding Electrical Power
Warning: To avoid the risk of electrical shock:
• Use only Covidien-branded batteries, adapters, and cables.
• Do not use batteries, adapters or cables with visible signs of damage.
• Do not touch internal components.
1.3.7 Warnings Regarding Ventilator Settings
Warning: The ventilator offers a variety of breath delivery options. Throughout the patient’s treatment, the clinician should carefully select the ventilation mode and settings to use for that patient, based on clinical judgment, the condition and needs of the patient, and the benefits, limitations and characteristics of the breath delivery options. As the patient’s condition changes over time, periodically assess the chosen modes and settings to determine whether or not those are best for the patient’s current needs.
Warning: Avoid nuisance alarms by applying appropriate alarm settings.
Warning: To prevent inappropriate ventilation, select the correct Tube Type (ET or Tracheostomy) and tube inner diameter (ID) for the patient’s ventilatory needs. Inappropriate ventilatory support leading to over-or under-ventilation could result if an ET tube or trach tube setting larger or smaller than the actual value is entered.
Warning: Setting expiratory volume alarms to OFF increases the risk of not detecting a low returned volume.
Warning: Setting any alarm limits to OFF or extreme high or low values can cause the associated alarm not to activate during ventilation, which reduces its efficacy for monitoring the patient and alerting the clinician to situations that may require intervention.
1.3.8 Warnings Regarding Hoses, Tubing, and Accessories
Warning: Do not use heat and moisture exchangers (HMEs) and heated humidifiers together. This may result in the HME absorbing water and becoming obstructed, resulting in high airway pressures.
Warning: To prevent electrostatic discharge (ESD) and potential fire hazard, do not use antistatic or electrically conductive hoses or tubing in or near the ventilator breathing system.
Warning: Adding accessories to or removing accessories from the ventilator breathing system (VBS) can change the pressure gradient and affect ventilator performance. Ensure that any changes to the ventilator circuit configurations do not exceed the specified values for circuit compliance and for inspiratory or expiratory limb total
resistance. See Table 58, page 242
. If adding accessories to or removing accessories from the VBS, always run SST to establish circuit compliance and resistance prior to ventilating the patient.
Warning: Use of a nebulizer or humidifier can lead to an increase in the resistance of inspiratory and expiratory filters. Monitor the filters frequently for increased resistance or blockage.
Warning: During transport, the use of breathing tubing without the appropriate cuffed connectors may result in the circuit becoming detached from the ventilator.
Warning: The added gas from an external pneumatic nebulizer can adversely affect spirometry, delivered O lead to an increase in expiratory filter resistance.
2
%, delivered tidal volumes, and breath triggering. Additionally, aerosolized particulates in the ventilator circuit can
Warning: Carefully route patient tubing and cabling to reduce the possibility of patient entanglement or strangulation.
Warning: Always use filters designed for use with the Puritan Bennett™ 980 Series Ventilator. Do not use filters designed for use with other ventilators. See
for relevant filter part numbers.
1
17
Warning: To avoid liquid entering the ventilator, empty the expiratory condensate vial before fluid reaches the maximum fill line.
Warning: Accessory equipment connected to the analog and digital interfaces must be certified according to IEC
60601-1. Furthermore, all configurations shall comply with the system standard IEC 60601-1-1. Any person who connects additional equipment to the signal input part or signal output part of the ventilator system configures a medical system, and is therefore responsible for ensuring the system complies with the requirements of the system standard IEC 60601-1-1. If in doubt, consult Medtronic Technical Services at 1 800 255 6774 or your local representative.
1.3.9 Warnings Regarding Gas Sources
Warning: Do not use nitric oxide, helium or mixtures containing helium as high-pressure inlet gas sources with the ventilator. They have not been validated for use as primary gas sources and could result in compromised ventilation and monitoring.
Warning: To avoid the risk of ventilator malfunction, do not use the ventilator with anesthetic gases.
Warning: For proper ventilator operation, use only clean, dry, medical grade gases when ventilating a patient.
Warning: Use of only one gas source could lead to loss of ventilation or hypoxemia if that one gas source fails and is not available. Therefore, always connect at least two gas sources to the ventilator to ensure a constant gas supply is available to the patient in case one of the gas sources fails. The ventilator has two connections for gas sources: air inlet and oxygen inlet.
Warning: Use of the ventilator in altitudes higher or barometric pressures lower than those specified could compromise ventilator operation. See
for a complete list of environmental specifications.
Warning: The ventilator should be connected to a gas pipeline system compliant to ISO 7396-1:2007 because:
• Installation of the ventilator on a non-ISO 7396-1:2007 compliant gas pipeline system may exceed the pipeline design flow capacity.
• The ventilator is a high-flow device and can interfere with the operation of other equipment using the same gas source if the gas pipeline system is not compliant to ISO 7396-1:2007.
1.3.10 Warnings Regarding Infection Control
Warning: Patients receiving mechanical ventilation may experience increased vulnerability to the risk of infection. Dirty or contaminated equipment is a potential source of infection. It is recognized that cleaning, sterilization, sanitation, and disinfection practices vary widely among health care institutions. Always follow your hospital infection control guidelines for handling infectious material. Follow the instructions in this manual and your institution’s protocol for cleaning and sterilizing the ventilator and its components. Use all cleaning solutions and products with caution. Follow manufacturer’s instructions for individual cleaning solutions. See
Warning: To prevent infection and contamination, always ensure inspiratory and expiratory bacteria filters are installed before ventilating the patient.
Warning: Never attempt to reuse single-use components or accessories. Doing so increases risk of cross-contamination and reprocessing of single-use components or accessories may compromise functionality leading to possible loss of ventilation.
1.3.11 Warnings Regarding Ventilator Maintenance
Warning: To ensure proper operation and avoid the possibility of physical injury, this ventilator should only be serviced by qualified technicians who have received appropriate Medtronic-provided training for the maintenance of this ventilator.
Warning: Follow preventive maintenance according to specified intervals. See
and
18
1
1.3.12 Cautions
Caution: To prevent possible equipment damage, ensure the casters are locked to prevent inadvertent movement of the ventilator during routine maintenance, or when the ventilator is on an incline.
Caution: Do not use sharp objects to make selections on the display or keyboard.
Caution: To ensure optimal performance, keep the GUI touch screen and keyboard clean and free from foreign
substances. See Table 39, page 169 .
Caution: To avoid moisture entering the ventilator and possibly causing a malfunction, Covidien recommends using a wall air water trap when using piped medical air from a facility-based air compressor.
Caution: Use only the cleaning agents specified. See
for approved cleaning agents.
Caution:
Clean compressor inlet filter according to the interval listed in Chapter 7 . See Table 38, page 167
.
Caution: Do not block cooling vents.
Caution: Ensure proper connection and engagement of expiratory and inspiratory filters.
Caution: Follow instructions for proper GUI and BDU (breath delivery unit) mounting as described in the
Puritan Bennett™ 980 Series Ventilator Installation Instructions .
Caution: Follow proper battery installation instructions as described in this manual.
Caution: When transferring the ventilator from storage conditions, allow its temperature to stabilize at ambient conditions prior to use.
Caution: Remove extended and primary batteries from ventilator prior to transporting in a vehicle. Failure to do so could result in damage to the ventilator.
1.3.13 Notes
Note: When using Non-Invasive Ventilation (NIV), the patient’s actual exhaled volume may differ from the exhaled volume reported by the ventilator due to leaks around the noninvasive interface.
Note: When utilizing a closed-suction catheter system, the suctioning procedure can be executed using existing mode, breath type, and settings. To reduce potential for hypoxemia during the procedure, elevated delivered oxygen can be enabled using the Elevate O
2
control. See
1.4 Obtaining Technical Assistance
1.4.1 Technical Services
For technical information and assistance, to order parts, or to order an operator’s manual or service manual, contact Medtronic Technical Services at 800 255 6774 or a local Medtronic representative.
If unable to correct a problem while using the ventilator, contact Medtronic Technical Services at 800 255 6774 or a local Medtronic representative. The Puritan Bennett™ 980 Series Ventilator Service manual , used by qualified, factory-trained service personnel, provides additional troubleshooting information.
When calling Medtronic Technical Services, or a local Medtronic representative, have the BDU and GUI serial numbers available, as well as the firmware version number of the ventilator system.
The ventilator’s configuration is available by touching the wrench icon on the GUI screen.
1
Have this information available whenever requesting technical assistance.
Covidien Australia’s service center contact information is:
Covidien Australia
19
52A Huntingwood Drive
Huntingwood
NSW Australia 2148
Tel: 1800 336 693
Fax: 1800 929 630
For online technical support, visit the SolvIT SM Center Knowledge Base at www.covidien.com.
The SolvIT Center provides answers to frequently asked questions about the ventilator system and other Puritan Bennett products 24 hours a day, 7 days a week.
1.4.2 On-screen Help
The ventilator is equipped with an on-screen help system that enables users to select an item on the screen and display a description of that item. Follow the procedure to access and use on-screen help:
1.4.2.1 Accessing On-screen Help Topics
Help topics on the ventilator are called tooltips. If a tooltip is available, a glowing blue outline appears around the item in question.
To access tooltips
1. Touch the item in question for a period of at least 0.5 s, or drag the help icon (the question mark icon appearing at the lower right of the GUI screen) to the item in question.
A tooltip appears with a short description of the item. Most screen items have tooltips associated with them, providing the operator with access to a multitude of help topics.
2. Touch “more” on the dialog to display an expanded description.
3. Touch “close” to close the dialog, or let it fade away after 5 s
Notes:
• Dragging the help icon causes the tooltip to display in its unexpanded state.
• Dragging the help icon and pausing causes a tooltip to display. Continue dragging to another item to dismiss the last tooltip and display another tooltip.
1.4.2.2 Other Resources
Additional resources for information about the ventilator can be found in the Puritan Bennett™ 980 Series
Ventilator Service Manual and appendices in this manual for BiLevel 2.0, Leak Sync, PAV+™, NeoMode 2.0, and
Proximal Flow options.
1.5 Warranty Information
To obtain warranty information for a covered product, contact Medtronic Technical Services at 1 800 255 6774 or call a local Medtronic representative.
1.6 Manufacture Date
The graphical user interface (GUI) and breath delivery unit (BDU) each possess a specific year of manufacture applicable only for that assembly. These dates are contained in the serial numbers for each assembly or option. Serial numbers for the 980 Ventilator final units consist of ten digits, in the following format:
20
1
35ZYYXXXXX where
• 35 signifies the unit was manufactured in Galway, Ireland
• Z represents the product code (B= breath delivery unit, G= GUI, C = compressor, P= Proximal Flow monitoring option). The product codes shown here are typically the most common. There may be other product codes shown in the serial number depending upon the particular option(s) purchased.
• YY is a two-digit year code that changes with each year
• XXXXX is a sequential number that resets at the beginning of each new year
Serial numbers are located on labels on the back panels of the GUI and BDU, and in various locations on product options.
1.7 Manufacturer
Covidien llc, 15 Hampshire Street, Mansfield, MA 02048 USA.
1
Covidien Ireland Limited, IDA Business &Technology Park, Tullamore, Ireland.
1.8 Electromagnetic Compatibility
The ventilator system complies with the requirements of IEC 60601-1-2:2007, IEC 60601-1-2: 2014 (EMC Collateral
Standard), and AIM Standard 7351731 Rev 2.00.2017. Certain transmitting devices (cellular phones, two-way radios, cordless phones, paging transmitters, RFID devices, etc.) emit radio frequencies that could interrupt ventilator operation if operated in a range too close to the ventilator. Practitioners should be aware of possible radio frequency interference if portable devices are operated in close proximity to the ventilator.
The Puritan Bennett™ 980 Series Ventilator requires special precautions to be taken regarding electromagnetic compatibility (EMC) and must be installed and put into service according to the EMC information provided in
21
22
2
2 Product Overview
2.1 Overview
This chapter contains introductory information for the Puritan Bennett™ 980 Series Ventilator.
Notes:
• Items shown in bold-italic font are contained as entries in the glossary.
• Items shown in bold font are physical hardware features (e.g., to patient port , exhaust port ).
• Alarms are shown in ALL CAPITAL letters.
Communication between the ventilator’s graphical user interface (GUI) and the breath delivery unit ( BDU ) occurs continuously via independent central processing units (CPUs) .
See
Figure 12, page 52 and their associated reference designators when reading the
following paragraphs.
Gas delivery starts with the ventilator connected to wall (or bottled) air and oxygen. Gas travels to the mix module where gas pressures are regulated by their respective proportional solenoid valves (PSOLs) . The PSOLs meter the gases according to the ventilator settings entered, then the gases flow through individual air and oxygen flow sensors into the mix manifold and accumulator for mixing. The individual gas pressures are continuously monitored both before and after they are mixed in the mix manifold and accumulator assemblies. The mixed gas then flows to the inspiratory pneumatic system where it flows through the breath delivery flow sensor and then the inspiratory PSOL for delivery to the patient.
Before the gas reaches the patient, it passes through an internal inspiratory bacteria filter, then through an external inspiratory bacteria filter attached to the ventilator’s gas outlet (the to patient port) where the breathing circuit is attached. When the gas returns from the patient, it flows through the expiratory limb of the breathing circuit, to the from patient port on the expiratory bacteria filter (which includes a condensate vial) before flowing through the exhalation flow sensor and exhalation valve (EV) . A gas exhaust port allows exhaled gas to exit the ventilator and flow to the room.
The ventilator recognizes the patient’s breathing effort using pressure triggering ( P
TRIG
) or flow triggering ( ⩒
TRIG
During pressure triggering, as the patient inhales, the airway pressure decreases and the inspiratory pressure transducer (PI) monitors this pressure decrease. When the pressure drops to at least the value of the pressure sensitivity ( P
SENS
)
) setting, the ventilator delivers a breath. During flow triggering, the difference between inspiratory and expiratory flows is monitored. As the patient inhales, the exhalation flow sensor measures less flow, while the delivery flow sensor measurement remains constant. When the difference between the two measurements is at least the value of the operator-set flow sensitivity ( ⩒
.
the clinician can increase the
SENS
SENS
) , the ventilator delivers a breath. If the patient is not inhaling, any difference between delivered flow and expiratory flow is due to flow sensor inaccuracy or leaks in the ventilator breathing circuit. To compensate for leaks, which can cause autotriggering,
setting or enable Leak Sync, if available.
Note: Leak Sync is a software function that is enabled by the clinician. Details on its operation are provided in
A backup pressure triggering threshold of 2 cmH
2
O is also in effect. This provides enough pressure sensitivity to avoid autotriggering, but will still allow the ventilator to trigger with acceptable patient effort.
The exhalation valve controls positive end expiratory pressure (PEEP) using feedback from the expiratory pressure transducer (PE) . the valve controller also cycles the ventilator into the exhalation phase if the PE measurement equals or exceeds the operator-set high circuit pressure limit. The PE measurement also controls when the safety valve (SV) opens. If PE measures 110 cmH
2
O or more in the ventilator breathing circuit, the safety valve opens, allowing the patient to breathe room air through the valve.
2
23
2.2 Ventilator Description
The ventilator system is available in three models. All ventilators provide continuous ventilation to patients requiring respiratory support.
Puritan Bennett™ 980 Pediatric–Adult Ventilator – The Pediatric–Adult model ventilates pediatric or adult patients with predicted body weights from 7.0 kg to 150 kg, and with tidal volumes from 25 mL to 2500 mL.
Puritan Bennett™ 980 Neonatal Ventilator – The Neonatal model ventilates neonatal patients with predicted body weights from 0.3 kg to 7.0 kg, and with tidal volumes for mandatory volume-controlled breaths from 2 mL to 315 mL.
Puritan Bennett™ 980 Universal Ventilator – The Universal model ventilates neonatal, pediatric, and adult patients with predicted body weights from 0.3 kg to 150 kg, and with tidal volumes for mandatory volume-controlled breaths from 2 mL to 2500 mL.
To ventilate neonatal patients on the Pediatric–Adult or Universal models, the NeoMode 2.0 software option is required. For details regarding the NeoMode 2.0 software option, see
The ventilator should have a service life of approximately 10 years, provided the preventive maintenance schedule stated in the Puritan Bennett™ 980 Series Ventilator Service Manual is followed.
The ventilator’s IEC 60601-1/EN 60601-1 classification is:
• Protection class I
• Type BF
• Mobile
• Internally powered
• IP 21 equipment
• Continuous operation
• Not suitable for use with flammable medical gases (not AP or APG)
See
for a description of the meaning of the IP classification.
The ventilator system uses a graphical user interface (GUI) and breath delivery unit (BDU) for entering patient settings and delivering breaths to the patient. The GUI contains electronics capable of transferring the clinician’s input (by touching the screen) to the BDU where pneumatic and electronic systems generate the breathing parameters.
2.3 Indications For Use/Intended Purpose
The Puritan Bennett™ 980 Ventilator System is designed for use on patient population sizes from neonatal (NICU) through adult who require respiratory support or mechanical ventilation and weigh a minimum of 0.3 kg (0.66 lb).
It is suitable for service in hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilatory support using medical oxygen and compressed medical air from either an internal air compressor or external air sources to deliver oxygen concentrations of 21% to 100%. Ventilatory support can be delivered invasively or non-invasively, to patients who require the following types of ventilator support:
• Positive Pressure Ventilation, delivered invasively (via endotracheal tube or trach tube) or non-invasively (via mask or nasal prongs)
• Assist/ Control, SIMV, or Spontaneous modes of ventilation
Note: Intended typical usage may be defined to include the following for the ventilator system:
Hospital Use – Typically covers areas such as operating rooms, special procedure areas, intensive and critical care areas within the hospital.
Intra-hospital transport – Includes transport of a patient within the hospital or hospital-type facility. All external hospital transportation (i.e., ambulance or aircraft) is excluded.
24
2
2.4 Contraindications
Do not operate the ventilator in a magnetic resonance imaging (MRI) environment.
2.5 Components List
Note: The ventilator system has no components made of natural rubber latex.
Note: The components in the gas pathway that can become contaminated with bodily fluids or expired gases during both normal and single fault conditions are:
• External inspiratory filter
• Internal inspiratory filter
• Exhalation filter and condensate vial
• Exhalation valve assembly
The typical ventilator system ships with the following packing list. Depending upon the ventilator system purchased, your list may vary.
Table 3.
Typical Packing List
Quantity
1
1
Graphical user interface
Breath delivery unit
1
1
1
1
1
2
1
1
1
1
1
Item
Inspiratory filter
Exhalation filter
Condensate vial
Gas hoses (air and oxygen)
Standard caster base
Power cord
Operator’s manual CD
Puritan Bennett™ 980 Series Ventilator Installation Instructions
Flex arm
Drain bag
Gold standard circuit (for running EST)
2
25
2.6 Product Views
2.6.1 GUI Front View
Figure 1.
GUI Front View
1 Display brightness key
2 Display lock key
3 Alarm volume key
4 Manual Inspiration key
5 Rotary encoder (knob)
6 Inspiratory pause key
7 Expiratory pause key
8 Alarm reset key
9 Audio paused key
10 Omni-directional LED
26
2.6.2 GUI Rear View
Figure 2.
GUI Rear View
2
See
for symbols found on the GUI or BDU. The “Do Not Push” symbol found on the GUI, only, is shown in this table.
2
27
2.6.3 BDU Front View
Figure 3.
BDU Front View
1 Condensate vial
2 Exhalation filter
3 Exhalation filter latch
4 AC power indicator
5 Power switch
6 Status display
7 Internal inspiratory filter
8 Option connector panel door
Table 4.
BDU Front Label Symbols and Descriptions
Symbol
To patient port
Description
From patient port
Exhalation filter latch locked (down) and unlocked (up)
28
2.6.4 BDU Rear View
Figure 4.
BDU Rear View
2
1 Standard base
2 Air inlet
3 Oxygen inlet
4 Labels indicating installed software options
5 Service mode button
6 Remote alarm port
7 Cylinder mount (optional)
, item 4 and
Figure 5.
Installed Software Options
lists the symbols and descriptions found on BDU or base labels.
29
2
Table 5.
BDU Rear Label or Panel Symbols and Descriptions
Symbol Description
This device is for sale by or on the order of a physician.
User must consult instructions for use. Symbol is also found on “Do not obstruct” labels on both left and right sides of the ventilator, and on label indication supply gas connec‐ tions.
Keep away from fire or flame. Oxygen rich environments accelerate combustibility.
Atmospheric pressure limitations—The operational atmospheric pressure range 70 kPa to 106 kPa (10.2 psi to 15.4 psi).
Humidity limitations—The operational humidity limit range 10% to 95%.
Temperature limitations—The operational temperature limit range 50°F to 104°F (10°C to
40°C).
Type BF applied part.
IEC Ingress protection classification—Protected against ingress of fingers or similar objects and protected from condensation.
Explosive hazard. Do not use in the presence of flammable gases.
Authorized to bear the CSA certification mark signifying the product has been evaluated to the applicable ANSI/Underwriters Laboratories Inc. (UL) and CSA standards for use in the US and Canada.
The ventilator contains components manufactured with phthalates.
CB1
CB2
Potential equalization point (ground) (on AC panel).
BDU circuit breaker (on AC panel).
Compressor circuit breaker (on AC panel).
USB port (at rear of ventilator).
HDMI port (at rear of ventilator).
S
T
T
E
C
I
E
R
V
S
E
H
D
Service port (at rear of ventilator).
Service mode button (at rear of ventilator).
Remote alarm port (at rear of ventilator).
Ethernet connector (at rear of ventilator).
Serial port (at rear of ventilator).
30
2
Table 6.
Common Symbols found on GUI or BDU Labels
Symbol Description
CE Mark—Signifies compliance with Medical Device Directive 93/42/EEC.
Do Not Push—Do not push on the GUI.
Manufacturer—Name of the ventilator manufacturer.
Authorized representative.
Serial number.
Manufacture date—The manufacture date is contained in the serial number. See
Sec‐ tion 1.6, Manufacture Date, page 20 for details regarding interpretation of the serial num‐
ber.
WEEE – Proper waste disposal. Follow local governing ordinances regarding disposing of waste labeled with the WEEE symbol.
2.6.5 Ventilator Side Views
Figure 6.
Ventilator Right Side View
2
31
Figure 7.
Ventilator Left Side View
2.7 Mounting Configurations
The ventilator system can be mounted as a free-standing unit standing at the patient’s bedside; the BDU with the
GUI is mounted on a base with casters and includes a handle for ease of movement.
2.8 Battery Backup
The ventilator system uses a battery to provide backup power in case AC power is lost. When operating on battery power, the status display shows the “On Battery Power” image. See
is available to lengthen the amount of time the ventilator can operate on battery power. See
.
2.9 Graphical User Interface
There are two displays on the ventilator — the primary display (GUI) and the status display.
2.9.1 Primary Display
The GUI incorporates a 15inch display that rotates 170° about a vertical axis in either direction. The GUI can also be tilted up to 45° from vertical.
The clinician enters ventilation parameters via the GUI’s touch screen, also known as the ventilator’s primary display. The GUI’s keys activate other ventilator functions including screen brightness, display lock, alarm volume, manual inspiration, inspiratory pause, expiratory pause, alarm reset, and audio paused.
32
2
The GUI displays the following information depending on the state of the ventilator:
• Ventilator, apnea, and alarm settings
• Patient data
• Waveforms
• Current alarm banners
2.10 GUI Controls and Indicators
2.10.1 Control Keys
The GUI bezel has eight off-screen control keys as shown in
.
Table 7.
GUI Control Keys
Key symbol Description
Brightness control key — Adjusts the GUI screen brightness. Press the key and turn the knob to adjust the brightness.
Display lock key — Actuates a lock to prevent inadvertent settings changes to the ven‐ tilator (including the knob function) while the display is locked. The display lock is useful when cleaning the touch screen. Press the key again to unlock the display.
Also use the display lock key to reset the GUI touch screen as described in Section 2.10.2
.
Alarm volume key — Adjusts the alarm volume. The alarm volume cannot be turned OFF.
Manual inspiration key — In A/C, SIMV, and SPONT modes, delivers one manual breath to the patient in accordance with the current mandatory breath parameters. In BiLevel mode, transitions from low pressure (P
L
) to high pressure (P
H
) (or vice versa). To avoid breath stacking, a manual inspiration is not delivered during inspiration or during the
restricted phase of exhalation. See Section 10.7.5, Manual Inspiration, page 205 for infor‐
mation on the restricted phase of exhalation.
The Manual inspiration key can be used to deliver mandatory breaths to the patient or to run an inspiratory pause maneuver in SPONT mode. The manual inspiration key cannot be used to run an expiratory pause maneuver in SPONT mode.
Inspiratory pause key — Initiates an inspiratory pause which closes the inspiratory and exhalation valves and extends the inspiratory phase of a mandatory breath for the pur‐ poses of measuring end inspiratory pressure (P
I END
(P
PL
), static compliance (C
STAT
) for calculation of plateau pressure
), and static resistance (R
STAT
).
Expiratory pause key — Initiates an expiratory pause which extends the expiratory phase of the current breath in order to measure total PEEP (PEEP
TOT
).
Alarm reset key — Clears active alarms or resets high-priority alarms and cancels an active audio paused condition. An alarm reset is recorded in the alarm log if there is an active alarm. DEVICE ALERT alarms cannot be reset.
Audio paused key — Pauses alarms for 2 minutes. Cancel the audio paused function by touching the on-screen Cancel button.
2.10.2 GUI Touch Screen Unresponsive or Blank: Recommended Responses
Rarely the graphical user interface (GUI) display (LCD panel) may become blank, cannot be read or becomes unresponsive. Audible and visual alarm functions are maintained and may be activated, and ventilation should continue as set.
2
33
In these conditions, first assess the patient, including an examination of the patient and a review of the monitored parameters, to confirm ventilation is in progress and to confirm the patient’s status. Examine the status display on top of the breath delivery unit (BDU) to confirm ventilator function.
If, in your clinical judgement, the patient’s status is satisfactory, and ventilation is continuing as set, follow the steps below for each specific GUI condition.
Note: If, in your clinical judgement, the patient’s status is unsatisfactory, consider patient transfer to another ventilator or provide alternate source of ventilation in accordance with your institution’s protocols.
The ventilator should be replaced as soon as possible and repaired by qualified service personnel.
Specific GUI Conditions (Unresponsive or Blank) and Recommended Responses:
1. Unresponsive screen with display still visible and waveforms active: If you observe an unresponsive GUI, inaccurate GUI responses, or unintended GUI responses, reset the touch screen to restore proper touch screen functionality.
To reset the touch screen
1. Touch the display lock key on the GUI bezel to lock the screen.
The locked padlock icon appears on the screen and the display lock key illuminates.
2. Touch the display lock key again. Doing so displays a progress bar below the locked padlock icon, after which time the locked icon will “unlock”, indicating a successful GUI touch screen reset.
Alternatively, ensure that a patient is not connected to the ventilator and power cycle the ventilator.
Note: Do not touch the screen during the unlock period.
Note: The manual GUI touch screen reset described in this section is different than in the automatic
30-second transient reset of the GUI described in
.
2. Momentary Blank GUI Screen / GUI Transient Reset: Rarely, the GUI resets spontaneously to ensure proper function. If this occurs, the status display shows a count-down timer until completion of the reset (see
, and illustration below). Assess the patient as described above and wait for the GUI reset to finish (approximately 30 seconds).
3. Blank GUI Screen / GUI Failure: An error message will be displayed on the status display (See
illustration below). If this occurs, confirm that ventilation continues as set, provide alternate ventilatory support or transfer the patient to another ventilator in accordance with your institution’s protocols as soon as it is appropriate to do so. Service the ventilator prior to returning it for use on patients.
34
2 If the GUI condition is not remedied, consider patient transfer to another ventilator or provide alternate source of ventilation in accordance with your institution’s protocols. Service the ventilator prior to returning it for use on patients.
2.10.3 Visual Indicators
shows the GUI’s visual indicators. See
Figure 33, page 92 for area names.
The audio paused function has two visual indicators—the audio paused key on the GUI bezel glows yellow during an audio pause interval, and a visual countdown timer appears, showing the amount of time the audio paused interval has remaining.
Table 8.
GUI Visual Indicators
Symbol Description
Ventilator setup (vent setup) button. Located at the lower left corner of the GUI.
Touch this button to open the ventilator setup screen.
2
Manual Event
Adult patient circuit indicator. Indicates adult circuit type tested during SST, and in use. Appears above the Vent Setup button.
Pediatric patient circuit indicator. Indicates pediatric circuit type tested during SST, and in use. Appears above the Vent Setup button.
Neonatal patient circuit indicator. Indicates neonatal circuit type tested during SST, and in use. Appears above the vent setup button.
Home icon. A constant access icon. See Figure 33, page 92
. Touch this icon to dismiss all open dialogs on the GUI screen. The display resumes showing the ven‐ tilator waveforms.
Touching this text causes the manual event screen to appear, where a variety of events can be recorded for viewing in the Trending layout. See the Trending addendum for more information about events.
Alarms icon. A constant access icon. See Figure 33, page 92
. Touch this icon to display the alarm settings screen, which allows alarm limits to be changed.
Logs icon. A constant access icon. See
Figure 33, page 92 . Touch this icon to display
the logs screen, which contains tabs for Alarms, Settings, Patient Data, Diagnostics,
EST/SST status, General Event, and Service logs.
35
Table 8.
GUI Visual Indicators (continued)
Symbol Description
Elevate O
2
control. A constant access icon. See
Figure 33, page 92 . Touch this icon
to temporarily increase the delivered oxygen concentration by the percent dis‐ played in the Elevate O
2
dialog. The ventilator will deliver the elevated O2 concen‐ tration for 2 minutes and then return the O
2
concentration to the set O
2
% value.
For the 2-minute period, the Elevate O value between 22% and 100% O interval. The Elevate O
2
2
2
concentration can be set to result in any
delivery. Touching “Extend” restarts the 2-minute
function can be terminated prior to completion of the
2-minute interval by touching Stop. Any time the Elevate O entry is made to the patient data log.
2
control is activated, an
Screen capture icon. A constant access icon. See
. Touch this icon to capture the image displayed on the GUI screen. See to read the complete pro‐ cedure for capturing screen images.
Help icon. A constant access icon. See Figure 33, page 92
. Drag this icon to the item in question and release. A tooltip appears, describing the item’s function.
Unread items icon. When this icon appears overlaid on another icon or tab (the logs icon, for example) it indicates there are unread items at this location.
Configure icon. A constant access icon. See Figure 33, page 92
. Touch this icon to display the configure screen. From this screen, perform all the SST tests or a single
SST test. If performing a single test, all SST tests must subsequently be performed and passed in order to ventilate a patient.
Pause icon. Located above the constant access icons. Touch this icon to pause the waveform graph.
Waveform layout icon. Located above the constant access icons area. Touch this icon to open the waveform layout dialog.
Grid lines icon. Located above the constant access icons area. Touch this icon to turn waveform grid lines on or off.
Maximize waveform icon. Located at the upper right portion of each waveform.
Touch this icon to enlarge the waveform to its maximum size.
Restore waveform icon. Restores waveform to its original size. Located at the upper right of the maximized waveform.
Pushpin icon— pinned state. When in the pinned state, prevents a dialog from closing (under certain conditions). Located in the upper right corner of the GUI on
the vent setup screen. See Section 4.4.2.2
Pushpin icon—unpinned state. When the unpinned icon is touched, the pinned state becomes active. Located in the upper right corner of the GUI on the vent setup
Low priority alarm icon (appears on alarm banner).
Medium priority alarm icon (appears on alarm banner).
36
2
Table 8.
GUI Visual Indicators (continued)
Symbol Description
High priority alarm icon (appears on alarm banner).
Audio paused symbol. The audio paused symbol appears together with a 2-minute countdown timer in the constant access area when the audio paused key is pressed.
2.10.4 On-screen Symbols and Abbreviations
Touch an on-screen symbol briefly (0.5 s) to display a tooltip on the GUI screen. The tooltip contains a definition of the symbol and other descriptive text, available with either short or long descriptions. The short description expands to show more information by touching more on the tooltip dialog or collapses by touching less. The tooltip closes by touching close or fades in 5 seconds if left alone. Expanding the tooltip dialog prevents the tooltip from timing out. Touching outside the tooltip causes the dialog to close.
summarizes the ventilator’s symbols and abbreviations.
Table 9.
Symbols and Abbreviations
Symbol or Abbreviation
T
A
D
SENS
C
DYN
R
DYN
EEF
P
I END
P
CIRC
LEAK
LEAK
Y
V
⩒
TE MAND
⩒
E TOT
E SPONT
V
TE SPONT
V
TE
E
SENS
T
E
Definition
Apnea interval
Disconnect sensitivity
Dynamic compliance
Dynamic resistance
End expiratory flow
End inspiratory pressure
Monitored total circuit pressure
Exhalation leak
Exhalation leak at PEEP (Leak Sync enabled) as measured by the proximal flow sensor
Exhaled mandatory tidal volume
Exhaled minute volume
Exhaled spontaneous minute volume
Exhaled spontaneous tidal volume
Exhaled tidal volume
Expiratory sensitivity
Expiratory time
Flow pattern (ramp)
⩒
⩒
⩒
CIRC
SENS
TRIG
⩒
P
Y
H
Flow pattern (square)
Monitored total inspiratory and expiratory flow
Flow sensitivity
Flow triggering
Monitored inspiratory and expiratory flow measured at the prox‐ imal airway
High pressure setting (in BiLevel)
2
37
O
2
%
P
0.1
C
PAV
E
PAV
% Supp
R
PAV
R
TOT
WOB
TOT
T
H
T
H
:T
L
I:E
C
20
/C
V
LEAK
T
I
3Tau
I
P
I
V
TI
V
TL
P
DRIVE
PEEP
I
PEEP
I PAV
P
L
T
L
P
MEAN
NIF
Table 9.
Symbols and Abbreviations (continued)
Symbol or Abbreviation
P
Y
P
PEAK
PEF
⩒
MAX
PSF
PEEP
%Leak
P
PL
T
PL
P
COMP
P
SENS
P
SUPP
P
TRIG
V
TIY
Definition
Monitored circuit pressure throughout the breath cycle measured at the proximal airway
High pressure time (in BiLevel)
High pressure time to low pressure time ratio (in BiLevel)
Inspiratory time to expiratory time (I:E)
Inspiration compliance ratio
Inspiratory leak
Inspiratory time
Inspiratory time constant
Inspiratory pressure
Inspired tidal volume
Inspired tidal volume (when Leak Sync is enabled)
Driving Pressure
Intrinsic PEEP (auto PEEP)
PAV-based intrinsic PEEP
Low pressure setting (in BiLevel)
Low pressure time (in BiLevel)
Mean circuit pressure
Negative inspiratory force
Oxygen percentage
Airway occlusion pressure at 100 ms
PAV-based lung compliance
PAV-based lung elastance
Percent support setting for tube compensation and PAV+™
PAV-based patient resistance
PAV-based total airway resistance
PAV-based work of breathing of patient and ventilator during inspiration
Peak circuit pressure
Peak expiratory flow
Peak inspiratory flow
Peak spontaneous flow
Set or monitored positive end expiratory pressure
Percent leak
Plateau pressure
Plateau time
Compensation pressure
Pressure sensitivity
Pressure support level
Pressure triggering
Proximal inspired tidal volume
38
2
Table 9.
Symbols and Abbreviations (continued)
Symbol or Abbreviation
V
TEY
V
TI MANDY
V
TI SPONTY
V
TLY f
Definition
Proximal exhaled tidal volume
Proximal mandatory inspired tidal volume
Proximal spontaneous inspired tidal volume
Proximal inspired tidal volume with Leak Sync enabled
Respiratory rate or apnea respiratory rate
Rise time percent f/V
T
T
I SPONT
T
I
/T
TOT
C
STAT
R
STAT
V
T
V
T CIRC
V
T Y
PEEP f
TOT
TOT
VC
VS
Spontaneous rapid/shallow breathing index
Spontaneous inspiratory time
Spontaneous inspiratory time ratio
Static compliance
Static resistance
Tidal volume
Monitored total inspiratory and expiratory volumes
Monitored inspiratory and expiratory patient volumes measured throughout the breath cycle at the proximal airway
Total PEEP
Total respiratory rate (monitored)
Vital capacity
Volume support
2.10.5 Audible Indicators
A tone sounds when a button on the GUI is touched, and also when settings are accepted. Audible indicators include pitched tones, beeps, and key clicks. Key clicks sound whenever a key on the GUI is pressed. Various tones annunciate patient alarms.
Note: Pressing the audio paused key mutes alarms for the 2-minute audio paused period.
Caregivers may choose to mute alarms by pressing the audio paused key.
A 2-minute countdown timer appears on the GUI during the audio paused interval. Cancel the audio paused function by touching Cancel.
Table 10.
GUI Audible Indicator Functions
Function
Low priority alarm tone
Medium priority alarm tone
High priority alarm tone
Soft bound tone
Hard bound tone (invalid entry)
39
Description
A series of two tones. Sounds when a low priority alarm occurs.
A repeating series of three tones. Sounds when a medium priority alarm occurs.
A repeating series of five tones. Sounds when a high priority alarm occurs.
One tone. Sounds when a soft bound is reached when making changes to ventilator settings. A soft bound is a selected value that exceeds or goes below its limit and requires acknowledgment to continue.
The invalid entry sound occurs when a hard bound is reached when making changes to ventilator settings. A
2
Table 10.
GUI Audible Indicator Functions (continued)
Function Description hard bound defines the upper or lower limit of the set‐ ting, where the setting cannot be adjusted higher or lower.
ventilator functions. See
2.11 Breath Delivery Unit
The breath delivery unit contains the hardware and software to enable the ventilator to provide patient support.
2.11.1 BDU Controls and Indicators
2.11.1.1 BDU Controls
• On/Off switch—Lift the switch cover and turn the ventilator on or off.
Figure 8.
Ventilator Power Switch and AC Indicator
1 AC power indicator
2 On/Off switch
• Service mode button—Press and release this button when the Covidien splash screen appears on the status display after powering on the ventilator to enter Service mode.
40
Figure 9.
Service Mode Button (TEST)
2
1 Service mode button
Note: The Covidien splash screen shows the Covidien logo and appears momentarily as a banner on the status display.
2.11.1.2 BDU AC Indicator
The status display and the AC power indicator are the only visual indicators on the BDU. The AC indicator illuminates green whenever the ventilator is connected to AC power. All other visual indicators on the ventilator
are on the GUI. See Section 2.11.1.4, Typical Status Display Indicators and Messages, page 42
for a description of the status display indicators and symbols. The next section summarizes the information appearing on the status display.
2.11.1.3 Status Display
the following information according to the state of the ventilator:
During normal ventilation the status display shows:
• Current power source (AC or DC)
• Safe state status: safety valve open (SVO) or vent inop
• Presence of primary and extended batteries and their charging status
• Relative available battery charge level
• Circuit pressure graph displaying pressure units, ⤒ P
PEAK
alarm setting and current P
PEAK
and PEEP values
• Connection of air and oxygen
41
2
• Ventilator operational hours
• Visual indication of current alarm volume setting
Note: The status display provides a redundant check of ventilator operation. If the GUI stops operating for any reason, ventilation continues as set.
Figure 10.
Sample Status Display During Normal Ventilation
1 Primary and extended battery status (presence or absence).
2 Alarm volume setting
3 Gas connection status
4 Power status
5 Measured peak circuit pressure (updated at the end of the current breath)
6 P
PEAK
alarm setting
7 Measured inspiratory pressure (changes as pressure changes)
8 Selected pressure units
9 Measured PEEP
During Service mode the status display shows:
• Ventilator serial number
• Ventilator operational time
• EST and SST history
• Power on self test (POST) status
• Hours until next preventive maintenance is due
• Gas pressure at the manifold inlets
See
Table 11 for status display possibilities.
2.11.1.4 Typical Status Display Indicators and Messages
Note: Status display images are shown without the optional DC compressor installed.
Table 11 lists indicators and messages that appear on the status display:
42
2
Table 11.
Status Display Indicators and Descriptions
Status Display Indicator or Message Meaning
Splash screen. Appears when the ventilator’s power switch is turned on. When this image appears, press and release the TEST button at the back of the ventila‐ tor to enter Service mode.
POST failure. This image appears if a POST error occurs at ventilator startup, along with the error code (in this case a missing primary battery).
2
Failure of the exhalation flow sensor assembly (EVQ) during power on self test.
Confirm proper installation of the exhalation flow sen‐ sor assembly and power cycle the ventilator. See
for proper EVQ installation instructions.
Failure of the EVQ during power on self test. Reinstall or replace the EVQ and run flow sensor calibration. See
for proper EVQ installation instructions.
43
Table 11.
Status Display Indicators and Descriptions (continued)
Status Display Indicator or Message Meaning
Prior to patient connection. The status display appears as shown when the patient has not been connected to the ventilator. Note the absence of P
PEAK ues.
and PEEP val‐
Stand-By state. The status display appears as shown when the ventilator is in stand-By state.
Battery charged. The ventilator’s primary battery (in the right-most receptacle) is shown fully charged, repre‐ sented by a + symbol and green color.
44
Battery charging. Identifies that the ventilator’s primary battery is charging. This icon is animated; orange bars scroll upward towards a “+” sign indicating the battery is charging. Green bars show the relative remaining battery capacity. If an extended battery is installed, the image shows a similar representation in the extended battery location (left-most receptacle).
2
Table 11.
Status Display Indicators and Descriptions (continued)
Status Display Indicator or Message Meaning
Battery icon. Denotes the ventilator is operating on battery power when this image appears on any status display indicator. Alerts the operator there is insuffi‐ cient AC power to operate the ventilator. The indicator is replaced by the “on AC power” indicator when ade‐ quate AC power is restored.
On battery power. Alerts the operator there is insuffi‐ cient AC power to operate the ventilator. Ventilator is operating on battery power with greater than 10 minutes of capacity remaining. Note the appearance of the battery icon.
2
Low battery. Identifies that the ventilator’s primary bat‐ tery (right-most receptacle) is discharging and there are 10 minutes or less of battery capacity remaining. A percentage indicator shows the remaining battery capacity. If an extended battery is installed, the image would show a similar representation in the extended battery location (left-most receptacle).
Critically low battery. Identifies that the ventilator’s pri‐ mary battery has less than 5 minutes of battery capacity remaining. A percentage indicator shows the remaining battery capacity. If an extended battery is installed, the image would show a similar representa‐ tion in the extended battery location.
45
Table 11.
Status Display Indicators and Descriptions (continued)
Status Display Indicator or Message Meaning
Power failure. Alerts the user that the ventilator’s bat‐ tery is depleted or depletion is imminent. Replace pri‐ mary or extended battery with a fully charged battery or connect ventilator to AC power.
Battery Inoperative. This image appears on the status display when a battery fault renders the battery inop‐ erative.
Battery not installed. This image appears when there is no primary battery installed, and renders the ventilator inoperative.
46
GUI Transient Reset. Indicates there is a transient loss of communication between the BDU and the GUI. It occurs in the ventilator by design to maintain full GUI display functionality. During the GUI transient reset, ventilation continues as currently set, audible and vis‐ ual alarms are not annunciated, and the status display shows a count-down timer until the completion of the
GUI transient reset. The count-down lasts for approxi‐ mately 30 s.
2
Table 11.
Status Display Indicators and Descriptions (continued)
Status Display Indicator or Message Meaning
GUI Failure. Indicates a loss of communication between the BDU and the GUI that cannot be recov‐ ered by the ventilator system. During the GUI failure, ventilation continues as currently set, audible and vis‐ ual alarms are annunciated, and the status display shows “Display Failed”. Replace the ventilator as soon as it is appropriate to do so. Service the ventilator prior to returning it for use on patients.
Recommended actions for GUI failure condition:
• Verify the patient’s respiratory and physiological stability.
• Confirm that the patient is receiving ventilatory support by observing expansion and contraction of the patient’s chest.
• Assess patient status by reviewing other monitor‐ ing indicators (for example, oxygen saturation, heart rate, blood pressure, etc.).
• Transfer the patient to an alternate source of ven‐ tilation consistent with your institution’s protocol.
Ventilator inoperative (vent inop). Indicates the venti‐ lator is no longer capable of ventilating a patient and requires service. The alarm reset key cannot be used to restore function to the ventilator during a ventilator inoperative condition. Provide alternate means of ven‐ tilation immediately. Note the display of the safety valve open indicator.
2
Safety valve open (SVO) indicator. During SVO, the patient can breathe room air through the safety valve, to the extent the patient is able to breathe unaided. See
Section 4.11.6, Safety Valve Open (SVO), page 115 .
47
Table 11.
Status Display Indicators and Descriptions (continued)
Status Display Indicator or Message Meaning
Backup ventilation (BUV) indicator. Indicates the ventilator has entered the backup ventilation state. See
Section 10.16.4, Background Diagnostic System, page 237
for a description of BUV.
AC power indicator. When this image appears on any status display indicator, indicates the ventilator is oper‐ ating on AC power.
Status display appearance when ventilator is breathing in Normal mode. Note the appearance of the AC power icon.
Air available indicator. When this image appears on any status display indicator, indicates the ventilator is con‐ nected to a pressurized air source.
O
2
available indicator. Indicates ventilator is connected to a pressurized O
2
source.
2.11.1.5 BDU Audible Indicators
The continuous tone alarm is the only audible indicator in the BDU, and is described in
48
2
Table 12.
BDU Audible Indicator Functions
Indicator
Continuous tone alarm (Immediate priority)
Description
A continuous tone annunciated when there is a ventilator inoperative (vent inop) con‐ dition. This alarm lasts for a minimum of 2 minutes.
2.11.2 Connectors
The ventilator incorporates the following connectors:
Ventilator outlet port (to patient) – A coaxial 15 mm (ID) / 22 mm (OD) conical connection to which the external inspiratory bacteria filter attaches.
Exhalation port (from patient) – The expiratory limb of the patient circuit attaches to the inlet of the exhalation bacteria filter. This port is compatible with a standard 22 mm (OD) conical connection.
Proximal flow sensor – A keyed pneumatic connector for the proximal flow sensor is provided with a locking feature to prevent inadvertent disconnection. The proximal flow sensor measures flow and pressure at the patient wye. The proximal flow sensor is an optional sensor. Details on operation are provided in the appendix in this manual. See
Standard interface connectors – USB, HDMI, and Ethernet connectors are provided. The USB connector allows screens to be captured on an external USB storage device and allows communication with an external patient monitor via serial over USB protocol, and the HDMI connector allow the GUI image to be displayed on an external video display device. The Ethernet connector is used by service personnel to upload new software and options.
See Section 5.5.1, Port Use, page 129 for more information. See
Section 5.4.3, Comm Port Configuration, page 119
for information on serial over USB data transfer when configuring Comm ports for external devices.
2.12 Additional Equipment
An optional DC compressor is available to provide compressed air in the event the wall or bottled air supply is lost or is unavailable. The compressor receives DC power from its own power supply if AC power is present. If there is no AC power available, the compressor is powered by its internal battery. The compressor interface printed circuit board assembly (PCBA) communicates with the BDU CPU PCBA. See the Compressor Operator’s Manual
Addendum for details regarding compressor operation.
Warning: Use of the compressor in altitudes higher or barometric pressures lower than those specified could compromise ventilator/compressor operation. See
2.13 Special Features
A Proximal Flow option is available. The proximal flow sensor is used to measure low flows and pressures associated with neonatal ventilation. If the ventilator is configured with this option, See
information.
2.14 Color Definitions
provides a legend to interpret gas colors in the pneumatic diagrams shown in Figure 11 and
.
Table 13.
Color Legend
Color or Symbol Description
High-pressure oxygen (NFPA 99 designation)
High-pressure air (NFPA 99 designation)
2
49
Table 13.
Color Legend (continued)
Color or Symbol
Mixed gases, including air
Atmosphere
Vacuum
Water
Description
2.15 Pneumatic Diagrams
Note: Both the compressor and the Proximal Flow option are hardware options.
Figure 12 illustrate the ventilator’s pneumatics with and without the Proximal Flow option. The
Proximal Flow option is only for use with neonatal patients.
Figure 11.
Pneumatic Diagram (Compressor Shown)
50
2
1 Pressure switch, mix accumulator (PS1)
2 Solenoid valve, options supply (SOL2)
3 Pressure sensor, mix accumulator (P
MX
)
4 Accumulator, mix (ACC
M
)
5 Tube, mix (T
M
)
6 Proportional solenoid valve, patient gas delivery (PSOL
D
)
7 Solenoid valve, BUV (SOL 3)
8 Safety valve (SV)
9 Pressure sensor, safety valve (P
SV
)
10 Solenoid valve, inspiratory pressure sensor autozero (SOL4)
11 Pressure sensor, inspiratory (P
I
)
12 Pressure sensor, barometric (PA)
13 Vial, exhalation condensate (ECV)
14 Filter, exhalation (F4)
15 Flow sensor assembly, exhalation valve
16 Exhalation valve (EV)
17 Filter, exhalation pressure line (F5)
18 Solenoid valve, exhalation pressure autozero (SOL 5)
19 Pressure sensor, exhalation (PE)
20 Humidifier
21 Filter, external bacteria (F
D2
)
22 Filter, internal bacteria (F
D1
)
23 Check valve, patient gas delivery (CV
D
)
24 Sensor, oxygen (OS)
25 Restrictor, breath delivery bypass (R2)
26 Flow sensor, patient gas delivery (FS
D
)
27 Accumulator, compressor (ACC
C
)
28 Relief valve, compressor accumulator (RV
CA
)
29 Solenoid valve, compressor unload (SOL7)
30 Motor compressor (MC)
31 Heat exchanger, compressor (HE)
32 Filter, compressor air (F7)
33 Dryer, compressor
34 Filter, muffler (F6)
35 Check valve, compressor accumulator (CV
CA
)
36 Pressure sensor, compressor accumulator (PC)
37 Check valve, oxygen (CVO
2
)
38 Check valve, air (CV
Air
)
39 Proportional solenoid valve, oxygen (PSOLO
2
)
40 Flow sensor, air (FS
Air
)
41 Proportional solenoid valve, air (PSOL
Air
)
42 Pressure sensor, air gas inlet (P
Air
)
43 Restrictor, wall air bleed outlet (R1)
44 Check valve, compressor air inlet (CV
CAir
)
45 Filter bowl assembly, air (WT2)
46 Filter element, air (F2)
47 Check valve, wall air inlet (CV
WAir
)
48 Filter, oxygen impact (F1)
49 Filter, element, oxygen (F3)
50 Pressure sensor, oxygen gas inlet (PO
2
)
51 Flow sensor, oxygen (FSO
2
)
52 Restrictor, prox flow (R4)
53 Relief valve, mix accumulator (RV
MA
)
54 Solenoid valve, mix accumulator purge (SOL1)
2
51
Figure 12.
Pneumatic Diagram — Compressor and Prox Flow Systems
1 Restrictor, prox flow (R4)
2 Solenoid valve, prox flow (SOL 6)
3 Module, proximal flow system
4 Pressure sensor, prox flow accumulator (P
PROX
)
5 Humidifier
6 Wye, patient circuit
7 Sensor, proximal flow
8 Filter, neonatal exhalation
9 Condensate vial, neonatal expiratory
Items enclosed by a dotted line represent components internal to the ventilator.
52
3
3 Installation
3.1 Overview
This chapter contains information for the installation and set up of the Puritan Bennett™ 980 Series Ventilator.
Before operating the ventilator system, thoroughly read this operator’s manual.
Topics include:
• Safety reminders
• Ventilator setup
• Battery information
• Ventilator operating modes
• Preparing the ventilator for use
• Tests to perform prior to ventilating a patient
3.2 Safety Reminders
Warning: Explosion hazard—Do not use in the presence of flammable gases. An oxygen-rich environment accelerates combustibility.
Warning: To ensure proper operation and avoid the possibility of physical injury, only qualified medical personnel should attempt to set up the ventilator and administer treatment with the ventilator.
Warning: To prevent electrostatic discharge (ESD) and potential fire hazard, do not use antistatic or electrically conductive hoses or tubing in or near the ventilator breathing system.
Warning: Use only gas supply hoses approved by Covidien. Other hoses may be restrictive and may cause improper ventilator operation.
Warning: To avoid possible injury, lock the ventilator’s casters prior to installing or removing ventilator components.
Caution: To ensure optimum performance, Covidien recommends preventive maintenance be performed by
factory-trained biomedical engineers per the schedule specified. See Table 43, page 181 .
3.3 Product Assembly
3.3.1 How to Assemble Ventilator Components
Ventilator setup should have already been completed by factory-trained service personnel including successfully passing EST. This manual does not include ventilator assembly instructions.
3.3.2 Product Power Sources
3.3.2.1 Using AC Power
The ventilator is normally AC-powered. See
Section 3.5.1, Connecting the Ventilator to AC Power, page 55 to connect
the ventilator to AC power.
3.3.2.2 Using Battery Power
Warning: Use only Covidien-branded batteries. Using other manufacturer’s brands could result in the batteries operating the ventilator for less than the specified amount of time or could cause a fire hazard.
Warning: One primary battery must be installed at all times in the BDU’s primary battery slot for proper ventilator operation. The ventilator will not complete the startup process without the primary battery installed. See
for identification of battery slots.
3
53
The ventilator’s primary battery must be installed by qualified service personnel (as it is shipped separately) before patient use. The ventilator will not complete power on self test (POST) if the battery is not present, and ventilation is prohibited. Ensure the battery is fully charged before placing the ventilator into service.
The ventilator employs a battery backup system if AC power becomes unavailable or drops below approximately
90 volts. A new, fully charged battery provides at least 1 hour of power to the ventilator assuming ambient temperature of 20°C (68°F) to 25°C (77°F), PBW =70 kg, and at factory default ventilator settings.
The battery backup systems for the ventilator and compressor contain one primary battery each. Backup power is supplied to the ventilator in the event of an AC power loss.
One extended battery slot is available for the ventilator and the compressor. If both primary and extended ventilator and compressor batteries are present, these batteries can power the ventilator and compressor for 2 hours (1 hour for the primary battery and 1 hour for the extended battery) under the environmental conditions described previously. When using battery power, the ventilator and compressor operate from their extended batteries, if present, first and then switch to the primary batteries. The ventilator and compressor primary and extended batteries are charged whenever the ventilator is plugged into AC power (the ventilator does not have to be powered up). If the ventilator or compressor is operating on battery power, the status display shows which battery is in use and its charge level, and the remaining time the battery will operate before charging is required again.
3.3.2.3 Battery Charging
Batteries requiring charging are charged whenever the ventilator is connected to AC power, whether operating or not.
The ventilator and compressor charge their primary batteries first, then their extended batteries. The time required to charge a single battery (either primary or extended) is approximately 6 hours at room temperature whether the ventilator is turned off (but connected to AC power) or operating, but charging time can vary based on temperature or depletion state of the battery. The status display provides the batteries’ capacities.
The compressor’s battery charging system (if a compressor is present) operates independently from the ventilator’s charging system and batteries are charged in parallel.
If a battery fault occurs, the fault is annunciated, charging of the faulty battery discontinues, but charging of any other non-faulty battery continues. A faulty battery will cause annunciation of the error and battery power will not be available for the ventilator.
The ventilator status display indicates the charge level of the installed batteries, the presence of one or more battery faults, and which battery is being charged.
The ventilator operates no differently when its batteries are charging than it does when the batteries are fully charged.
The ventilator continues operating as set when the ventilator switches from AC power to battery power and illuminates an indicator on the status display alerting the operator that the ventilator is now operating on battery power and AC POWER LOSS alarm annunciates. A medium priority alarm annunciates when the remaining run-time for the ventilator drops to 10 minutes and a high priority alarm annunciates when the remaining time drops to 5 minutes.
3.4 Product Placement
,
or if using a pendant-mounted configuration, Figure 14, page 55
.
Move the ventilator using the handle encircling the BDU and roll the ventilator to the desired location.
54
Figure 13.
Example of Freestanding Ventilator Placement
3
Figure 14.
Example of a Pendant-mounted Ventilator
VEN_10308_B
3
3.5 Product Connectivity
3.5.1 Connecting the Ventilator to AC Power
Note: Power outlet access and power cord position— Ensure that the power outlet used for the ventilator is easily accessible; disconnection from the outlet is the only way to completely remove power from the ventilator.
To connect the power cord to AC power
1. Plug the ventilator into a properly grounded power outlet rated for at least 15 A.
55
2. Verify the connection by checking the AC indicator below the power switch on the front of the BDU. See
for the power switch and AC indicator locations.
To connect the power cord to the ventilator
1. Remove the power cord retainer and connect the female end of the power cord to the ventilator’s power cord receptacle. See
.
2. Set hex nuts and spacers, if present, aside for reassembly.
3. Replace the power cord retainer, using the hex nuts and spacers removed during disassembly. See
.
Note: Your ventilator may have either combination of power cord and retainer.
Use the power cord hook located at the back of the ventilator for power cord storage.
Warning: For proper ventilator operation, and to avoid the risk of electric shock, connect the ventilator to a grounded, hospital grade, AC electrical outlet.
Figure 15.
Power Cord Retainer on BDU (without spacers)
1 1/4 in. hex nuts
2 Power cord retainer
56
3 AC power cord
Figure 16.
Power Cord Retainer on BDU (with spacers)
3
1 1/4 in. hex nuts
2 Power cord retainer
3 AC power cord
4 Spacers
3.5.2 Connecting the Gas Supplies
and O
2
supply pressure ranges must be between 35 psig and 87 psig (241.3 kPa and 599.8 kPa) and the average flow requirement for both gases is 60 L/min at 279.9 kPa (40.61 psi). The transient will not exceed 200 L/min for
≥3 s.
Warning: Due to excessive restriction of the Air Liquide™, SIS, and Dräger™ hose assemblies, reduced ventilator performance levels may result when oxygen or air supply pressures < 345 kPa (50 psi) are employed.
Gas cross flow from one high pressure input port of one type of gas to another high pressure input port of a different gas will not exceed 100 mL/h under normal or single fault conditions. If, during a single fault condition, cross flow exceeds 100 mL/h, an audible alarm annunciates.
Warning: Use of only one gas source could lead to loss of ventilation or hypoxemia if that one gas source fails and is not available. Therefore, always connect at least two gas sources to the ventilator to ensure a constant gas supply is available to the patient in case one of the gas sources fails. The ventilator has two connections for gas sources:
To connect the gas sources
1. Connect the oxygen hose to the oxygen inlet fitting (item 1) as shown. Ensure use of a medical grade oxygen source.
2. Connect the air hose to the air inlet fitting (item 2). See Figure 17, page 58
.
57
3
Figure 17.
Connecting the Ventilator to the Gas Supplies
1 O
2
gas connection
2 Air gas connection
Warning: To prevent a potential fire hazard and possible damage to the ventilator, ensure the connections to the gas supplies are clean and unlubricated, and there is no water in the supply gas. If water is suspected, use an external wall air water trap to prevent damage to the ventilator or its components.
The ventilator system can be purchased with the following gas inlet fittings for both air and O
2 female, NIST, Air Liquide, SIS, and Dräger.
: BOC, DISS, DISS
See
Table 44, page 191 for part numbers of gas hoses.
3.5.3 Filter Installation
and contamination, both inspiratory and exhalation filters must be used with the ventilator.
Warning: In order to reduce the risk of infection, always use the ventilator with inspiratory and exhalation bacteria filters.
Warning: Do not attempt to use inspiratory or exhalation filters designed for use with ventilators other than the
Puritan Bennett 980 Series Ventilator. See Table 44, page 191 for relevant part numbers.
Warning: Refer to the filter’s instructions for use (IFU) for details such as cleaning and sterilization requirements, filtration efficiency, proper filter usage, and maximum filter resistance, particularly when using aerosolized medications.
Warning: Refer to the exhalation filter IFU for information on reusable filter cleaning and sterilization and filter efficiency.
Warning: Do not reuse disposable inspiratory or exhalation filters, and dispose according to your institution’s policy for discarding contaminated waste.
58
3
Caution: Ensure both inspiratory and exhalation filters are properly attached to the ventilator.
To install the inspiratory filter
1. Attach the inspiratory filter to the to patient port.
2. Ensure the direction of flow arrow is pointing outward, toward the patient circuit’s inspiratory limb.
Note: Refer to the inspiratory filter IFU for information on proper use and handling of the filter.
Note: Refer to the exhalation filter IFU for information on proper use and handling of the filter and for emptying the condensate vial for adult and pediatric patients. See
Chapter 15, Appendix NeoMode 2.0, page 297 for
information on emptying the condensate vial when using neonatal exhalation filters.
The condensate vial must be assembled to the reusable exhalation filter prior to installing the assembly to the ventilator.
To assemble the adult/pediatric reusable exhalation filter and condensate vial
1. Seat the filter to the condensate vial, ensuring alignment of the condensate vial’s seal with the mating edge of the exhalation filter.
2. Twist the condensate vial in a counterclockwise direction until the stops on the vial and exhalation filter meet.
Warning: Do not operate the exhalation filter latch during patient ventilation. Opening the latch during ventilation will result in a patient disconnect condition and corresponding alarm.
To install the adult/pediatric exhalation filter
1. If necessary, remove expiratory limb of patient circuit from exhalation filter.
2. Raise the exhalation filter latch to unlock (item 6).
3
This raises the exhalation valve assembly and allows the filter door to swing away from the ventilator. See
3. Open the exhalation filter door.
4. Remove the existing filter.
5. Insert the new filter by sliding the filter along the tracks in the door. Ensure the from patient port aligns with the cutout in the door and points away from the ventilator.
6. Close the exhalation filter door.
7. Lower the exhalation filter latch to secure the filter.
Note: To prevent EVQ misalignment and potential damage, ensure there is an exhalation filter in place any time the ventilator is transported from one location to another.
59
Figure 18.
Adult and Pediatric Filter Installation
1 Condensate drain port
2 Condensate vial
3 Exhalation filter
4 Condensate vial gasket
5 Condensate drain port cap
6 Exhalation filter latch
7 Exhalation filter door
To install the neonatal exhalation filter adapter door
1. If necessary, remove the expiratory limb of the patient circuit from the exhalation filter.
2. Lift the exhalation filter latch. See
3. Remove the existing exhalation filter door by lifting it off of the pivot pins.
4. Fit the neonatal adapter door onto the pivot pins.
Figure 19.
Installing the Neonatal Filter
3
4
2
1
1 Neonatal exhalation filter
2 Neonatal adapter door
VEN_11162_A
60
3 Exhalation filter latch
4 Filter door pivot pin
3
To install the neonatal exhalation filter assembly
1. With the door still open, push the neonatal filter assembly straight up into the adapter.
2. Close the door.
3. Lower the exhalation filter latch.
4. Reattach the expiratory limb of the patient circuit to the filter.
To use the drain bag
1. Remove the drain port cap from the exhalation filter condensate vial drain port.
2. Attach the drain bag tube to the condensate vial’s drain port.
3. Hang the drain bag on the holder located on the ventilator’s accessory rail, as shown. See
. See Table 44, page 191 for the part number of the drain bag holder.
Figure 20.
Drain Bag
3
VEN_11209_A
3.5.4 Flex Arm
Use the flex arm to support the patient circuit between the patient and the ventilator. See
which illustrates flex arm installation into the sockets provided.
61
Figure 21.
Flex Arm Installation
VEN_10330_A
To attach or remove the flex arm
1. Locate the threaded inserts in the ventilator’s handle.
2. Fasten the flex arm into one of the inserts.
3. Hang the patient circuit using the circuit management supports included with the flex arm.
4. Remove the flex arm by first removing the patient circuit, then unfastening the flex arm from the threaded fastener in the handle.
3.5.5 Humidifier
Use the humidifier to add heat and moisture to the inhaled gas. Connect the humidifier to a hospital grade electrical outlet. Choose the humidifier (type and volume appropriate for the patient). The humidifier may be
mounted with the humidifier bracket as shown. See Figure 22, page 63
for the part number of the humidifier bracket.
Warning: Selection of the incorrect humidifier type or volume during SST or during patient ventilation can affect the accuracy of delivered volume to the patient by allowing the ventilator to incorrectly calculate the compliance correction factor used during breath delivery. This can be a problem, as the additional volume required for circuit compressibility compensation could be incorrectly calculated, resulting in over- or under-delivery of desired volume.
Warning: To ensure proper compliance and resistance calculations, perform SST with the humidifier and all accessories used for patient ventilation installed in the ventilator breathing system.
Warning: Follow the humidifier manufacturer’s IFU when using a humidifier with patient ventilation.
62
Caution: Follow the humidifier manufacturer’s IFU for proper humidifier operation.
To install the humidifier bracket
• Attach humidifier bracket to the ventilator’s accessory rail by placing the bracket behind the railing and fastening the bracket clamp to the bracket with four 5/32 inch hex screws, capturing the railing between the bracket and the clamp. Ensure the humidifier mounting slots are facing outward from the ventilator.
Figure 22.
Bracket Installation on Rail
3
To install the humidifier
1. Slide the rear of the humidifier into the corresponding slot on the humidifier bracket, until it is fully seated.
See
Figure 23, page 64 Some humidifiers slide into the narrow slot in the humidifier bracket, and some
humidifiers use the wide slot.
3
63
Figure 23.
Humidifier Installation to Ventilator
2. Fill the humidification system with water to the desired volume.
3. Install the chamber to the humidifier, connect the patient circuit, then run SST.
4. Plug the humidifier into a grounded, hospital grade electrical outlet.
5. Turn the humidifier on.
Note: Complete instructions for the humidifier bracket and humidifier installation are given in the
Puritan Bennett™ 980 Series Ventilator Humidifier Bracket Installation Instructions , which includes humidifier bracket part numbers and descriptions.
3.5.6 Connecting the Patient Circuit
See
to connect the adult, pediatric, and neonatal circuits.
Warning: Use patient circuits of the lowest compliance possible with the ventilator system to ensure optimal compliance compensation and to avoid reaching the compliance compensation limit. See
types corresponding with predicted body weight (PBW).
Table 14.
Patient Types and PBW Values
Circuit Type
Neonatal
Pediatric
Adult
PBW in kg (lb)
0.3 kg to 7.0 kg (0.66 lb to 15 lb)
7.0 kg to 24 kg (15 lb to 53 lb)
25 kg to 150 kg (55 lb to 331 lb)
Note: Refer to the patient circuit’s IFU for information on proper use and handling and care and maintenance of the circuit.
64
3
A list of breathing system components and accessories is provided. See
. Use only Covidien components and accessories in the patient circuit.
Follow your institution’s protocol for safe disposal of the patient circuit.
Follow the patient circuit’s IFU for cleaning and disinfection information for reusable circuits.
Orient the patient circuit by hanging the patient circuit on the circuit management supports provided with the flex arm.
Figure 24.
Connecting the Adult or Pediatric Patient Circuit
8 9
7
6
5
3
1 Humidifier
2 Inspiratory limb
3 Circuit wye
4 Expiratory limb
5 Condensate vial
4 3 2 1
6 From Patient port
7 Exhalation filter
8 To patient port
9 Inspiratory filter
VEN_11168_A
65
Figure 25.
Connecting the Neonatal Patient Circuit
1 Humidifier
2 Patient circuit inspiratory limb
3 Circuit wye
4 Patient circuit expiratory limb
5 Condensate vial
6 From patient port
7 Neonatal exhalation filter (installed in adapter door)
8 To patient port
9 Inspiratory filter
Warning: Do not attempt to sterilize single-patient use circuits.
3.6 Batteries
3.6.1 Batteries
Warning: Use only Covidien-branded batteries. Using other manufacturer’s brands or remanufactured batteries could result in the batteries operating the ventilator for less than the specified amount of time or could cause a fire hazard.
Warning: To reduce the risk of infection due to cross-contamination, using a damp cloth, disinfect the batteries with one of the solutions listed before installation and whenever transferring to or from another ventilator. During use, clean external surfaces of batteries as necessary. See
Table 39, page 169 . Do not spray disinfectant directly onto
the battery or its connector.
Warning: Even though the Puritan Bennett 980 Ventilator meets the standards listed in
lithium-ion battery of the device is considered to be Dangerous Goods (DG) Class 9 - Miscellaneous, when transported in commerce. As such, the Puritan Bennett 980 Ventilator and the associated lithium-ion battery are subject to strict transport conditions under the Dangerous Goods Regulation for air transport (IATA: International
Air Transport Association), International Maritime Dangerous Goods code for sea and the European Agreement
66
concerning the International Carriage of Dangerous Goods by Road (ADR) for Europe. Private individuals who transport the device are excluded from these regulations although for air transport some requirements may apply.
Warning: To avoid the risk of fire, explosion, electric shock, or burns, do not short circuit, puncture, crush, heat above 60°C, incinerate, disassemble the battery, or immerse the battery in water.
Caution:
Ensure the batteries are oriented properly. See Figure 27, page 68
.
Figure 26.
Ventilator Battery
3
1 Battery connector
3.6.1.1 Primary Batteries
The ventilator’s primary battery is located in the rearward battery receptacle on the right side of the BDU. The compressor’s primary battery is located in the rearward battery receptacle in the compressor base. See
. The primary battery may be “hot swapped,” that is it can be replaced while the ventilator is operating.
To install or replace the primary battery in the BDU or compressor
1. Check the charge level by pressing the charge level button on the battery and verifying the charge level
LEDs illuminate. See
. for the location of the charge level button. Five green LED segments illuminate, indicating ≥90% battery capacity. From bottom to top, the first LED indicates ≥10% capacity, the second LED indicates ≥25% capacity, the third LED indicates ≥50% capacity, and the fourth LED indicates
≥75% capacity. An illuminated red LED at the top of the battery indicates a battery fault. If no LEDs illuminate it means there is <10% battery capacity remaining.
2. If the charge level is sufficient, orient the battery as shown, face the front of the ventilator and locate the battery compartments on the right side of the appropriate module. See
. The receptacle
67
3
towards the rear of the ventilator houses the primary battery while the receptacle towards the front of the ventilator houses the extended battery.
3. The primary battery is fastened in place with a thumbscrew (item 3). Loosen the thumbscrew approximately four to five turns to allow battery installation.
4. Insert the battery and push into its receptacle all the way until it clicks, indicating it is latched. The battery will only fit into the slot one way.
Figure 27.
Proper Battery Orientation
1 Charge status LEDs
2 Charge level button
5. Tighten the thumbscrew to secure the battery and prevent the primary battery from being removed.
Note: Remove the primary battery by reversing the steps. After loosening the thumbscrew, slide the battery ejector to the left to eject the battery.
68
Figure 28.
Battery Compartment Locations
1 2
3
3
4
5
3
6 VEN_10278_B
1 BDU extended battery receptacle and ejector
2 BDU primary battery receptacle and ejector
3 BDU and compressor primary battery thumbscrews
4 Compressor primary battery receptacle and ejector
5 Compressor extended battery receptacle and ejector
6 BDU primary battery (positioned for installation)
Note: Remove either primary battery by sliding the battery ejector to the left. The battery ejects itself from its receptacle.
3.6.1.2 Extended batteries
The extended battery receptacle is located forward of the primary battery. Like the primary battery, the extended battery may be hot swapped.
To install or remove an extended battery in either the BDU or compressor
1. Properly orient the battery as shown. See Figure 27, page 68
.
2. Push the battery into the forward receptacle of the appropriate module of the ventilator all the way until it clicks, indicating the battery is latched. See
Note: Remove the battery by sliding the battery ejector to the left. The battery ejects itself from its receptacle.
There is no thumbscrew for extended batteries.
Note:
in ventilator.
3.6.2 Battery Testing
To test the batteries
69
3
With the battery installed in the ventilator, push the battery charge level button located on the battery. A series of LEDs illuminates, indicating the charge level of the battery. When the bottom LED is illuminated, there is ≥10% of full battery capacity. The next LED illuminates when there is ≥25% capacity. the third lamp illuminates when there is ≥50% capacity available. The fourth LED illuminates when there is ≥75% capacity, and when the top LED
is illuminated, it represents ≥90% capacity. See Figure 27, page 68
to view the battery test button and LEDs.
3.6.3 Battery Life
Battery life for both primary and extended batteries is approximately 3 years from the date of manufacture. Actual battery life depends on the history of use and ambient conditions. As the batteries age with use, the time the ventilator will operate on battery power from a fully charged battery will decrease. Replace the battery every 3 years or sooner if battery operation time is insufficient for your usage or if indicated in the GUI prompt area. The user is informed on the GUI prompt area that the battery is nearing the end of service life. The prompt occurs a minimum of 90 days ahead of end of life for nominal use conditions. The battery requiring replacement is identified in the System Communication Log.
Determine which battery is nearing the end of its service life by accessing the ventilator logs.
1. Touch the Logs icon in the constant access area of the GUI.
2. Touch Diagnostics.
3. Touch System Comm Log. For each impacted battery, there will be an entry like the example shown in
Table 15 , including the location of the battery and its serial number.
Table 15.
Example of Battery End of Service Life Log
Time/Date
HH:MM:SS
DD-Month-YYYY
Test/Event
Ventilator Primary battery is nearing end of service life
HH:MM:SS
DD-Month-YYYY
HH:MM:SS
DD-Month-YYYY
HH:MM:SS
DD-Month-YYYY
Ventilator Extended bat‐ tery is nearing end of serv‐ ice life
Compressor Primary bat‐ tery is nearing end of serv‐ ice life
Compressor Extended battery is nearing end of service life
Code
E00041
E00042
E00043
E00044
Notes
SN XXxxXXXXXX
SN XXxxXXXXXX
SN XXxxXXXXXX
SN XXxxXXXXXX
3.6.4 Battery Manufacturing Date
The date of manufacture of the battery is contained in the serial number. The serial number is listed on the side of the battery near the connection point. The serial number consists of 10 alphanumeric characters and provides the following information: 2 character year, 2 character week, 2 character text character, and 4 character sequence.
70
Figure 29.
Battery manufacturing date
3
1 Year
2 Week number (XX of 52)
3 Text
4 Build sequence number
3.6.5 Battery Disposal
The battery is considered electronic waste and must be disposed of according to local regulations.
Follow local governing ordinances and recycling plans regarding disposal or recycling of the battery.
3.6.6 Li-ion Battery Pack Handling, Storage and Replacement
3.6.6.1 Handling, Storage and Maintenance
Lithium-ion (Li-ion) battery packs (P/N 10086042) handling, storage and maintenance is recommended as follows:
• Battery packs should be installed in the ventilator or compressor (if the option is installed) while connected to AC Power to maintain charge, whether the ventilator is operating, in standby, or off.
• Battery packs not installed in a ventilator should have the charge level checked prior to use. This can be done
by pressing the charge level button. See Section 3.6
and
. It is recommended that battery packs displaying a charge level equivalent to one LED or fewer be fully charged prior to patient use.
• Battery pack life will be reduced if the battery pack is repeatedly allowed to fully deplete.
• Battery packs that display a red fault LED (noted in Section 3.6.6.2
Note: For optimal battery life, it is recommended to store batteries below 25°C and charged to 2 LEDs or fewer.
3.6.6.2 Replacement and Disposal
Li-ion battery pack (P/N 10086042) replacement and disposal:
In
, users are reminded to replace the battery pack every 3 years from date of manufacture and dispose of the battery packs according to local regulations. The date of manufacture is reflected within the serial number
sequence found on the battery pack. See Figure 31
for guidance on how to determine this date or follow the use-by date indicated on the battery pack labeling.
If a battery fault occurs, it will be indicated by an illuminated red fault LED or on the ventilator status display. If this occurs within 3 years from date of manufacture, contact our service experience group
([email protected]) to arrange for return and replacement of the battery pack.
3
71
Figure 30.
Battery Charge Level
1 White battery in use LED
2 Red battery fault indicator
Figure 31.
Battery Date of Manufacture
3 Green charge status LEDs
4 Charge level button
1 Year
2 Weak number (XX of 52)
3 Text
4 Build sequence number
3.7 Ventilator Operating Modes
3.7.1 Normal Mode
Normal mode is the default mode used for patient ventilation. The ventilator enters Normal mode after it has been turned on and POST completes, the ventilator is set up, and breath delivery parameters have been entered. If the clinician chooses, s/he can select Quick Start which uses default values or institutionally configured breath delivery settings after PBW has been entered. Entry into Normal mode is not allowed if a primary battery is not
72
3 detected in the ventilator BDU, a major POST fault occurs, or there is an uncorrected major system fault, or uncorrected short self test (SST) or extended self test (EST) failures or non-overridden alerts.
During Normal mode, the omni-directional LED on the top of the GUI appears green in color, in a steadily lit state.
for details regarding alarm priority. If another alarm occurs concurrently with an existing alarm, the LED displays the color corresponding to the highest priority level. If the alarm deescalates, the latched area (located on either side of the alarm LED indicator) of the alarm LED displays the color of the highest priority alarm while the center of the
LED displays the color of the current alarm’s priority. For more information on specific alarms, touch the logs icon in the constant access icons area of the GUI.
3.7.2 Quick Start
Quick Start is an extension of Normal mode, where institutionally configured default settings are applied after the patient’s PBW or gender and height are entered and Quick Start is touched to begin ventilation.
3.7.3 Stand-By State
Stand-By state can be used when the clinician needs to disconnect the patient for any reason (prior to a suction procedure, for example). The ventilator enters Stand-By state if a request is made by the clinician, a patient is disconnected within a fixed time period determined by the ventilator software, and the clinician confirms the patient has been disconnected intentionally. If a patient becomes disconnected from the patient circuit after the time period elapses, an alarm sounds and the patient-disconnect sequence is initiated. In Stand-By state, gas output is reduced to 10 L/min to limit gas consumption and to allow for detection of patient reconnection
Stand-By state is available in all ventilation modes except during inspiratory and expiratory BUV, occlusion status cycling (OSC) , safety valve open (SVO), or ventilator inoperative (vent inop) conditions.
Note: Do not block patient circuit wye while in Stand-By state. If the wye is blocked, the ventilator detects a patient connection and will attempt to resume normal ventilation.
To enter Stand-By state
1. Touch the Menu tab on the left side of the GUI. The menu appears.
2. Touch Stand-By. A Stand-By state pending dialog appears instructing the clinician to disconnect the patient circuit. A timer starts which allows 30 s to disconnect the patient.
3. Disconnect the patient circuit and confirm the disconnection by touching Confirm. A timer starts, which allows 30 s for confirmation of disconnect.
To exit Stand-By state
Reconnect the patient circuit. The ventilator resumes ventilation at the settings in use before the disconnection.
The following ventilator setting becomes active during Stand-By state:
• Base flow is set to 10 L/min
During Stand-By state:
• The exhalation valve is open.
• Current ventilator settings are retained in memory.
• Flow sensors are monitored to detect patient reconnection.
• Patient-related alarms are temporarily suppressed.
• Ventilator settings can be changed, if desired, and will be applied upon patient reconnection.
• The ventilator displays an indicator that it is in Stand-By state, and a timer indicating the elapsed time the ventilator has been in Stand-By state.
• Ventilator background checks continue to be made.
3
73
The ventilator automatically exits Stand-By state when patient reconnection is detected, the clinician completes patient setup (if ventilation was mistakenly started before setup was complete), or the ventilator power is cycled.
Prior to entering Stand-By state, the ventilator measures pressure and flow in the patient circuit to determine if a patient is attached. If a patient is detected, the ventilator continues ventilation as set prior to the request, alerts the operator that Stand-By state is pending, and requests the patient be disconnected. A countdown timer appears alerting the operator of the time remaining to disconnect the patient. After the patient is disconnected, the ventilator requests confirmation of the disconnection. When the ventilator enters Stand-By state, a message appears on the GUI, any active alarms are muted and reset and the associated alarm reset entries are logged in the
Alarm Event Log. Alarm detection is suspended, and breath delivery is suspended while a bias flow is maintained for patient detection. During Stand-By state, the ventilator displays the elapsed time the patient has been without ventilation. Since the ventilator maintains a bias flow for patient detection, it resumes ventilation at the previous settings when the patient is reconnected. There is no need to touch Exit Stand-By. Reconnecting the patient returns the ventilator to normal operation. During Stand-By state, patient data values are not displayed and the
LED located at the top of the GUI cycles between yellow and green. Entry into and exit from Stand-By state is recorded in the General Event log.
3.7.4 Service Mode
Warning: Before entering Service mode, ensure that a patient is not connected to the ventilator. Ventilatory support is not available in Service mode.
Service mode is used for extended self test (EST), ventilator calibration, configuration, software upgrades, option installation (all of which must be performed by Covidien factory-trained service personnel), and for making adjustments to institutional settings. All information stored in the individual logs is available in Service mode.
Service mode logs include:
• System Diagnostic
• System Comm.
• EST/SST Diagnostic
• Settings
• Alarms
• General Event
• Service
• Patient Data
See the Puritan Bennett™ 980 Series Ventilator Service Manual for details about Service mode logs.
A patient must not be attached to the ventilator when entering Service mode. Specific actions must be performed to enter this mode, prior to POST completion.
To access Service mode
1. Remove the ventilator from patient usage.
2. Turn the ventilator’s power switch on.
3. Press and release the Service mode button (TEST) at the back of the ventilator, when the Covidien splash screen appears on the status display after powering on the ventilator. See
for an image of the splash screen. The ventilator prompts to confirm no patient is attached.
74
Figure 32.
Service Mode Button (TEST)
1 E
S
T
3
D
I
I
E
S
E
R
V
VEN_11246_D
1 Service Mode button
4. Wait to enter Service mode.
5. Confirm that a patient is not connected to the ventilator by touching the corresponding button. The message SERVICE MODE VENTILATION SUPPORT IS NOT AVAILABLE appears on the graphical user interface.
6. Perform the required service.
7. Turn off the ventilator to exit Service mode.
Note: The Covidien splash screen shows the Covidien logo and appears momentarily as a banner on the status display.
See the Puritan Bennett™ 980 Series Ventilator Service Manual for information on the keys that are disabled during EST.
for a list of institutionally and operator-configurable items.
3.8 Product Configuration
Warning: If the ventilator fleet in your institution uses multiple institutionally configured presets and defaults, there can be risks of inappropriate alarm settings.
The ventilator is shipped configured with factory defaults for new patient parameters can be configured to suit institutional preferences. The operator may configure any desired parameter as long as this option has not been locked out and rendered unavailable. When configuring the ventilator, it displays the parameters associated with the operator’s last configuration. The following table lists the factory-configured settings, the institutionally configurable settings, and the operator-configurable settings.
75
3
Table 16.
Ventilator Configuration
Feature Factory Config‐ ured
X Vital patient data banner
Large font patient data panel
Waveform layout
Display bright‐ ness (Light set‐ tings)
Alarm volume
Elevate O
2 trol
con‐
Date/time for‐ mat
Default mL/kg ratio
New patient startup defaults
(including PBW, ventilation type, mode, manda‐ tory type, trigger type, O
2 vate O
2
)
%, ele‐
Opacity
X
X
X
X
X
X
X
X
X
Institutionally
Configurable
X
X
X
X
X
X
X
X
X
X
Operator Con‐ figurable
X
X
X
X
X
X
X
Configured by
Circuit Type
X
X
X
User Lockable
X
X
X
X
Can’t be changed in Normal mode
X X
3.8.1 Preparing the Ventilator for Use
Caution: Do not lean on the GUI or use it to move the ventilator. Doing so could break the GUI, its locking mechanism, or tip the ventilator over.
Prior to ventilating a patient, configure the GUI so it is capable of displaying all the desired parameters, information, and patient data. This eliminates the necessity for taking the patient off the ventilator, as configuration of many of the items requires the unit to be in Service mode.
To perform institutional configuration
Service Mode, page 74 for instructions on entering Service mode.
2. Touch Configuration at the top of the screen in Service mode. A list of buttons appears allowing configuration of the corresponding parameters.
3. See the next sections for specific instructions on institutional configuration of each parameter.
To return to factory default configuration
1. Enter Service mode, and confirm that no patient is attached by touching Confirm. See
Mode, page 74 for instructions on entering Service mode.
2. Touch Configuration at the top of the screen in Service mode. A list of buttons appears allowing configuration of the corresponding parameters.
76
3
3. Select the desired modified setting from the menu options on the left side.
4. Touch Default.
3.8.2 Configuring the GUI
The display can be configured in various ways. See Table 16, page 76
for the parameters which are factory configured, institutionally configurable and operator-configurable. Once the factory or institutionally configurable items have been configured, they remain the default values. Factory configured values cannot be
memory as default settings. If changes are made to operator-configurable parameters, they remain in memory during a ventilator power cycle as long as the same patient is set up when returned to ventilation. If a new patient is set up, the factory configured values or institutionally configured values (if the parameter has been configured) are used. No alarm settings are institutionally configurable, which prevents changes to factory default alarm settings. However, the default mL/kg ratio is institutionally configurable, which can affect the default alarm setting values. Always review the alarm defaults prior to beginning ventilation, and set appropriately.
3.8.2.1 Date and Time Format
The date and time may be configured to the institution’s preference. The time can be specified as 12-hour or
24-hour time in HH:MM:SS format with 1-hour and 1-minute resolutions, respectively. The date formats are:
• DD-MMM-YYYY where DD is a two-digit day format, MMM is a three-letter abbreviation for the month, and
YYYY is a four-digit representation of the year
• MM-DD-YYYY where MM is a two digit month format, DD is a two-digit day format
• YYYY is a four-digit representation of the year
The settable date corresponds to the number of days in the set month and accounts for leap years.
To institutionally configure the ventilator’s date and time settings
1. Perform steps 1 and 2 of the
Section 3.8.1, Preparing the Ventilator for Use, page 76 .
2. Touch Date and Time.
3. Touch the button corresponding to 12-hour or 24-hour time.
4. Touch Hour and turn the knob to enter the correct hour.
5. Repeat for the minutes, and am or pm.
6. Touch the button corresponding to the date format desired (DD-MMM-YYYY or MM-DD-YYYY).
7. Touch Accept to confirm the date and time.
8. If done configuring parameters, exit Service mode.
3.8.2.2 Pressure Units
The ventilator’s pressure units can be configured for hPa or cmH
2
O.
To institutionally configure pressure units
1. Perform steps 1 and 2 of the
Section 3.8.1, Preparing the Ventilator for Use, page 76 .
2. Touch the vent setup button.
3
3. Touch the button corresponding to the desired pressure units.
4. If done configuring parameters, exit Service mode by touching Exit.
3.8.2.3 Screen Brightness and Keyboard Backlight (Light Settings)
To institutionally configure screen brightness and keyboard backlight
77
1. Perform steps 1 and 2 of the
Section 3.8.1, Preparing the Ventilator for Use, page 76 .
2. Touch Light Settings. Sliders appear to adjust the screen brightness and keyboard backlight.
3. Move the sliders to increase or decrease the brightness and backlight levels. Alternatively, turn the knob to increase or decrease the brightness and backlight levels.
4. Touch Accept to apply the changes, or Cancel to revert to original settings.
5. If done configuring parameters, exit Service mode.
To adjust display brightness
1. Press the display brightness key.
2. Slide the brightness slider or turn the knob to adjust the brightness level.
3. Dismiss the slider by touching anywhere on the GUI screen or allow to time out in 5 s.
Note: Keyboard backlight (bezel key brightness) can only be adjusted in Service mode.
3.8.2.4 New Patient Setup Defaults
To institutionally configure new patient default settings
1. Perform steps 1 and 2 of the
Section 3.8.1, Preparing the Ventilator for Use, page 76 .
2. Touch the button corresponding to adult, pediatric, or neonatal new patient defaults.
3. Touch the Vent Type, Mode, Mandatory Type, and Trigger Type buttons corresponding to the desired parameters.
4. Configure the default PBW and mL/kg ratio, Elevate O
2 knob.
and O
2
% by touching its button and turning the
5. Repeat for each patient type by selecting the corresponding button.
6. Touch Accept or Accept ALL when the default configuration is complete.
7. If done configuring parameters, exit Service mode.
3.8.2.5 Elevate O
2
Note: The Elevate O
2
control adds a percentage of O in any value between 22% and 100% O
2
delivery.
2
to the breathing mixture for 2 minutes. The additional percentage is shown on the icon in the constant access icon area. The Elevate O
2
concentration can be set to result
To adjust the amount of elevated O
2
delivered for 2 minutes
1. In the vent setup dialog in Normal mode, touch the Elevate O
2
GUI screen.
icon in the constant access icons area of the
The icon glows and a dialog appears with a countdown timer, Elev O
2 changes, and Extend, Stop, and Close buttons.
button highlighted and ready for
2. Turn the knob to increase or decrease the amount of oxygen by the amount shown on the button. The allowable range can result in any value between 22% and 100% O
2
delivery.
3. Touch Extend to extend the 2-minute interval. Touching Extend restarts the 2-minute countdown timer.
4. Touch Stop to stop additional oxygen from being delivered and dismiss the countdown timer.
The Elevate O
2
function follows these rules:
78
3
• If apnea ventilation occurs during the 2-minute interval, the apnea % O
2 configured amount.
delivery also increases by the
• During LOSS OF AIR SUPPLY or LOSS OF O
2
SUPPLY alarm conditions, the Elevate O progress, and is temporarily disabled until the alarm condition no longer exists.
2
function is canceled if in
• During Safety PCV, the Elevate O
2
control has no effect. During circuit disconnect and stand-By states (when the ventilator is turned on but not ventilating) the Elevate O concentration, not the set oxygen concentration.
2
function affects the currently delivered oxygen
3.8.2.6 Alarm Volume
Warning: The audio alarm volume level is adjustable. The operator should set the volume at a level that allows the operator to distinguish the audio alarm above background noise levels.
To institutionally configure the alarm volume
1. Perform steps 1 and 2 of the
Section 3.8.1, Preparing the Ventilator for Use, page 76 .
2. Touch Alarm Volume Defaults. A screen appears allowing configuration of the alarm volume by circuit type.
3. Slide the alarm slider for each circuit type (adult, pediatric, or neonatal) or turn the knob to configure the alarm volume. The volume settings range from 1 (minimum) to 10 (maximum).
4. If done configuring the alarm volume, exit Service mode.
To adjust alarm volume
1. Set the alarm volume by pressing the alarm volume key, then sliding the alarm volume slider or turning the knob.
3
The alarm values range from 1 (minimum) to 10 (maximum).
2. Dismiss the slider by touching anywhere on the GUI screen or allow to time out in 5 seconds.
Note: A sample alarm tone sounds for verification at each volume level change. If necessary, readjust the alarm volume by moving the alarm volume slider to increase or decrease the volume.
Note: The alarm volume reverts to the institutionally configured default alarm volume or factory default if the ventilator’s power is cycled.
3.8.2.7 Vital Patient Data
Patient data are displayed in the Vital Patient Data banner. The operator can configure the banner for displaying
the desired patient data. See Figure 33, page 92
. A total of 14 values may be configured at one time, with eight values visible, and six more visible by scrolling the values using the left- and right- pointing arrows in the patient data area.
Two pages of additional patient data may be viewed by touching or swiping down on the patient data tab at the top of the GUI. Choose the respective buttons to view page one or page two. Additional patient data values may not be changed.
To institutionally configure patient data displayed on the GUI
1. Perform steps 1 and 2 of the
Section 3.8.1, Preparing the Ventilator for Use, page 76 .
2. Touch Patient Data Defaults. Five layout preset buttons appear along with a list of parameters and descriptions.
3. Touch a preset button and individually select a parameter from the scrollable list to appear in that preset’s vital patient data banner. Use the right- and left- pointing arrows to configure default values for all available parameters. Additionally, touch the padlock icon above each patient data parameter on the data banner to allow (unlocked) or restrict (locked) operator configurablity of that parameter during normal ventilation.
79
4. When done configuring the selected preset, touch Accept and select another preset to configure, if desired.
5. Touch Defaults to return configuration to factory settings.
6. If done configuring parameters, exit Service mode by touching Exit.
To configure the patient data displayed on the GUI
1. Double-tap a patient data parameter at the top of the GUI screen. A menu of buttons appears identified with patient data parameters. The parameter at the location touched will be replaced with the new parameter of choice. To view more parameters, touch the left- or right- pointing arrows to reveal more parameters.
2. Touch the button corresponding to the replacement parameter. The existing parameter is replaced with the new parameter.
3. Repeat steps 1 and 2 for as many parameters as desired.
3.8.2.8 Displaying Patient Data With a Larger Font
To improve visibility of patient data, a screen is available that appears with a larger font. Up to 14 data values may be displayed, which include:
• Institutional default patient data values (if configured)
• Remaining user selected patient data values (up to 14, including waveforms and loops)
To institutionally configure the large font patient data defaults
1. Perform steps 1 and 2 of the
Section 3.8.1, Preparing the Ventilator for Use, page 76 .
2. Touch Large Font Patient Data Defaults. Five layout presets appear along with a list of parameters and descriptions.
3. Touch a preset button and individually select a parameter for each of the desired patient data values.
4. Choose the desired scalar and loop waveforms for the large font patient data display. Waveform thumbnails only appear in the three right-most cells of the large font data panel.
5. Touch any of the padlock icons along the right-most edge of the selected layout to prevent operator configurability of the selected row.
6. Touch Accept or Accept ALL when finished.
7. If factory defaults are desired for a preset, touch Defaults.
8. If done configuring parameters, exit Service mode by touching Exit.
To display the large font patient data panel
1. Swipe the vital patient data banner tab downward or touch the vital patient data tab. The additional patient data panel appears.
2. Swipe the additional patient data banner’s tab downward or touch the additional patient data banner’s tab.
Patient data appear in a larger font.
3. Swipe the large font patient data panel tab upward or touch the tab to return to the banner to its normal font size.
The large font patient data parameters are configured in the same way as described in the patient data configuration section above.
3.8.2.9 Waveforms
Green waveforms denote a mandatory inspiration, yellow waveforms denote exhalation, and orange waveforms denote a spontaneous inspiration.
The GUI can be configured to display up to three waveforms and two loops simultaneously in the waveform area.
See
Allowable loops include pressure vs. volume and flow vs. volume. The waveforms display 60 seconds of
80
3 information and can be shown in a redrawing format, or paused with the ability to enable a cursor to trace the waveform by turning the knob.
The ventilator-generated waveforms provide immediate and dynamic qualitative information to the clinician about the subtleties of ventilation in real time. In many cases the shape and character of the drawn graphics for volume, flow, and pressure can provide advanced and early warning to the clinician of potential problems such as air leaks, air-trapping, breath asynchrony, over-distension, and flow mismatching.
The scalar waveforms are not intended to represent a patient physiological parameter nor a qualifiable characteristic of gas (air or O
2
) delivered to, or removed from, the human body.
V
T CIRC
Symbol
⩒
P
CIRC
CIRC
Definition
The V
T CIRC
waveform is reflective of the volume going into and out of the breathing circuit throughout the breath cycle. The volume values expressed by the V not compensated for circuit compliance or BTPS.
The scale representing measured volume (mL) can be set from a minimum range of –1 mL to 2 mL, to a maximum range of –2000 mL to 6000 mL.
T CIRC
waveform are
The ⩒
CIRC
waveform displays the total inspiratory and expiratory flow throughout the breath cycle measured by the ventilator’s internal (inspiratory and exhalation) flow sen‐ sors. The flow values expressed by the ⩒
CIRC
waveform are not compensated for circuit compliance or BTPS.
The scale representing measured flow (L/min) can be set from a minimum range of –2 L/min to 2 L/min, to a maximum range of –200 L/min to 200 L/min.
The P
CIRC
waveform displays the estimated pressure at the wye of the breathing circuit throughout the breath cycle by utilizing the ventilator’s internal (inspiratory and expira‐ tory) pressure sensors.
The scale representing measured pressure cmH range of –2 (cmH
2
O or hPa) to 10 (cmH or hPa) to 120 (cmH
2
O or hPa).
2
2
O or hPa) can be set from a minimum
O or hPa) to a maximum range of –20 (cmH
2
O
To institutionally configure waveforms and loops
1. Perform steps 1 and 2 of the
Section 3.8.1, Preparing the Ventilator for Use, page 76 .
2. Touch Graph Defaults. Five layout presets appear along with a list of parameters and descriptions.
3. Touch a layout preset button. The parameter button outline glows, signifying that it can be changed. If more than one parameter can be changed, touch that parameter to make its outline glow.
4. Select the parameter from the list whose waveform is desired to appear on the waveforms screen.
5. Configure each of the graphic display layouts as described above.
6. Touch the padlock icon above each graphic layout to prevent operator configuration of the selected layout.
7. If factory defaults are desired for a preset, touch Defaults.
8. If done configuring parameters, exit Service mode by touching Exit.
To configure waveforms and loops
1. Touch Waveform Layout, located below the displayed waveforms or the vent setup screen.
3
The icon glows and a menu of various waveform layouts appears.
2. Touch the desired waveform icon to display. The selected waveform appears on the GUI screen and the dialog closes.
To change the axis scaling
81
1. Touch the desired waveform axis.
2. Turn the knob to change the value. For each axis, turn the knob to the right to decrease the values, and turn to the left to increase the values.
To pause waveforms
1. Touch the pause icon, located below the waveforms area.
The icon glows yellow and allows the breath to complete. A cursor appears and travels along the waveform while turning the knob, displaying the x- and y-axis values.
2. Touch the pause icon again to reactivate the waveform.
See
Section 5.4.1, GUI Screen Capture, page 117
for information on storing waveforms.
3.8.2.10 Opacity
To institutionally configure screen opacity
1. Perform steps 1 and 2 of the
Section 3.8.1, Preparing the Ventilator for Use, page 76 .
2. Touch the opacity icon.
3. Turn the knob to increase or decrease the opacity.
4. Touch the padlock icon at the right side of the screen to allow or prevent operator adjustment of the screen opacity.
5. Touch Accept to close the dialog.
To adjust the screen opacity
1. Touch the opacity icon.
The icon glows when the opacity can be changed.
2. Turn the knob to increase or decrease the opacity.
Note: The opacity icon can be found on the vent setup screen and on any of the respiratory mechanics maneuvers screens.
3.9 Installation Testing
Fully charge the batteries before placing the ventilator into clinical use. See
Section 3.3.2.3, Battery Charging, page 54
status LEDs and for the location of the battery test switch and status LEDs.
Prior to connecting a patient to the ventilator for the first time, a qualified service technician must have calibrated the ventilator’s exhalation valve, flow sensors, and atmospheric pressure transducer and performed and successfully passed EST. See the Puritan Bennett™ 980 Ventilator Service Manual for instructions.
In addition, the clinician must also perform SST.
3.9.1 SST (Short Self Test)
Warning: Always disconnect the patient from the ventilator prior to running SST or EST. If SST or EST is performed while a patient is connected, patient injury may occur.
82
3
Warning: Check for circuit occlusion and run SST if increased pressures are observed during ventilation.
Warning: When changing any accessories in the patient circuit or changing the patient circuit itself, run SST to check for leaks and to ensure the correct circuit compliance and resistance values are used in ventilator calculations.
Note: When extending ventilator circuits for neonatal patients, the resulting VBS compliance may trigger a
COMPLIANCE LIMITED V support (PS).
T
alarm such that the VC+ or VS software will not continue to update the pressure target during breath delivery. In this case the user may change the breath type to pressure control (PC) or pressure
When a patient is not attached to the ventilator, run SST to check the patient circuit for:
• Gas leaks
• Circuit compliance and resistance calculations
SST must be run under any of the following conditions:
• Prior to ventilating a new patient
• When replacing the patient circuit and exhalation filter
• When connecting a different patient circuit to the ventilator
• When changing the patient circuit type
• When installing a new or sterilized exhalation filter
• When changing the humidification device type
• When adding accessories to or removing accessories from the breathing system (such as a humidifier or water trap)
No external test equipment is required, and SST requires minimal operator participation.
Humidification type and volume can be adjusted after running SST, however the ventilator makes assumptions when calculating resistance and compliance if these changes are made without rerunning SST. For optimal breath delivery, run SST after changing humidification type or humidifier volume.
SST results are recorded in the SST results log, viewable in Service mode and in Normal mode using the configuration (wrench) icon.
3.9.1.1 Required Equipment
• Proposed patient circuit for patient ventilation
• Accessories (water traps, etc.)
• Exhalation filter and condensate vial
• Humidifier, if applicable
• A number 1 stopper to block the patient airway at the patient wye
• Two gas sources (air and oxygen) connected to the ventilator) at a pressure between 241.3 kPa and 599.8 kPa
(35 psi and 87 psi)
3.9.1.2 SST Test Sequence
To run SST
1. Ensure that a patient is not connected to the ventilator.
2. So that the ventilator does not detect a patient connection, ensure that the breathing circuit wye is not attached to a test lung or covered in any way that would cause an increase in pressure at the wye.
3. Turn the ventilator on using the power switch located at the front of the BDU, below the status display. The ventilator runs POST when the power switch is turned on. Ensure the ventilator is operating on full AC power. Otherwise, SST test failures may result.
83
3
4. Wait at least 15 minutes to allow the ventilator to warm up and stabilize to ensure accurate results.
5. At the ventilator startup screen, touch SST or the configure icon (wrench) displayed in the lower right area of the GUI.
The SST history log appears along with Patient Setup, Run Leak Test, and Run All SST buttons.
6. Connect the patient circuit, filters, condensate vial, and all accessories to be used in patient ventilation.
Ensure that the patient wye is not blocked.
7. Touch Run All SST to perform all SST tests or touch Run Leak Test to perform the SST Leak Test of the ventilator breathing circuit.
8. Touch Accept to continue or Cancel to go back to the previous screen.
9. After accepting, touch the Circuit Type button corresponding to the patient circuit type used to perform
SST and to ventilate the patient (adult, pediatric, or neonatal).
10. Touch the Humidification Type button corresponding to the humidification type used for patient ventilation. If no humidifier is used, touch HME. If a humidifier is used, touch Humidification Volume and turn the knob to enter the volume. See
Table 18 for adult and pediatric patients or Table 19
for neonatal patients to determine the correct volume to enter.
11. Touch Accept to start SST.
12. Follow the prompts. Certain SST tests require operator intervention, and will pause indefinitely for a response.
Table 20 for a summary of the SST test sequence and results.
13. After each test, the ventilator displays the results. If a particular test fails, the test result appears on the screen and a choice to repeat the test or perform the next test is given. When all of the SST tests are complete, the
SST status screen displays the individual test results.
14. To proceed to patient set up, (if SST did not detect Alert or Failed) touch Exit SST, then touch Accept or cycle the ventilator’s power.
Table 17 lists the tests performed during SST.
Table 17.
SST tests
Test step
SST Flow Sensor Cross Check Test
SST Exhalation Valve Performance
SST Circuit Pressure Test
SST Leak Test
SST Exhalation Filter Test
SST circuit Resistance Test
SST circuit Compliance Test
SST Prox (if Proximal Flow option is installed)
Function
Tests O
2
and air flow sensors
Calibrates the exhalation valve and creates a table for use during calculations
Exercises delivery PSOL.
Checks inspiratory and expiratory autozero solenoids.
Cross-checks inspiratory and expiratory pressure trans‐ ducers at various pressures.
Tests ventilator breathing system for leaks
Checks for exhalation filter occlusion and exhalation compartment occlusion.
Checks for inspiratory and expiratory limb occlusions, and calculates and stores the inspiratory and expiratory limb resistance parameters.
Calculates the attached patient circuit compliance.
Verifies functionality of the proximal flow subsystem
Note: For adult and pediatric patients, enter the humidifier volume shown in
Table 18 . For neonatal patients, follow
the neonatal specific instructions during SST and enter the chamber volume shown in
84
3
Table 18.
Humidifier Volumes For Adult and Pediatric Patients
Manufacturer
Fisher & Paykel
Fisher & Paykel
Fisher & Paykel
Fisher & Paykel
Fisher & Paykel
Model
MR225
MR290
MR250
MR210
MR370
Description
Ped, disposable, Manual
Feed
Ped/Adult disposable,
Autofeed
Adult, disposable, Manual feed
Adult, Disposable, Manual
Feed
Adult, Reusable, Manual
Feed
Humidifier volume set‐ ting for SST (mL)
300
380
480
480
725
Table 19.
Humidifier Volumes—Neonatal Patients
Manufacturer
Fisher & Paykel MR290
Model Description
Neo/adult, disposable, autofill humidification chamber
SST humidifier volume setting (mL)
550
a If the following neonatal patient circuits are used with Fisher & Paykel MR290 humidification chamber, enter 500 mL as the humidifier volume: DAR neonatal patient circuit with single heated wire (DAR 307S9910)–for incubator use; DAR neonatal patient circuit with single heated wire (DAR 307/8682)–not for incubator use.
3.9.1.3 SST Results
SST reports results for each individual test. Three status indicators identify the SST results and actions to take for each.
Pass – The individual SST test has met its requirements.
Alert – Alerts occur when the ventilator detects one or more non-critical faults.
Fail – The individual SST test did not meet its requirements.
Table 20.
Individual SST Results
Pass
Alert
Test status
Failed
Meaning
Individual SST test passed
The test result is not ideal, but is not critical.
If SST is in progress, it halts further testing and prompts for decision.
The ventilator has detected a critical prob‐ lem and SST cannot complete until the ven‐ tilator passes the failed test.
Response
No need to do anything, unless prompted by the ventilator.
When the system prompts, touch one of these buttons:
• Repeat Test
• Exit SST
Eliminate leaks in the ventilator breathing system and rerun SST. Otherwise, service the ventilator and rerun SST.
a WARNING —Completing SST with an Alert status for an individual test produces an Override SST button. Overriding an
Alert in SST may result in ventilator performance outside of the stated specification for accuracy. Choose to override the
Alert status and authorize ventilation only when absolutely certain this cannot create a patient hazard or add to risks arising from other hazards. To override the alert, touch Override SST, then touch Accept.
85
3
3.9.1.4 SST Outcomes
When SST completes all of the tests, analyze the results.
Table 21.
Overall SST Outcomes
Final Outcome
PASS
OVERRIDDEN
FAIL
Meaning
All SST tests passed.
The ventilator detected one or more faults.
Choose to override the Alert status and authorize ventilation only when absolutely certain this cannot create a patient hazard or add to risks arising from other hazards.
One or more critical faults were detected.
The ventilator enters the SVO state and can‐ not be used for normal ventilation until SST passes.
Response
Touch Patient Setup to set up the patient for ventilation.
Check the patient circuit to determine the problem or restart SST with a different patient circuit.
Check the patient circuit to determine the problem or restart SST with a different patient circuit.
If touching Override SST, observe the following warning:
Warning: Overriding an Alert in SST may result in ventilator performance outside of the stated specification for accuracy. Choose to override the Alert status and authorize ventilation only when absolutely certain this cannot create a patient hazard or add to risks arising from other hazards.
A single circuit-leak test can be run without changing the SST outcome.
If a complete SST is interrupted and ventilation was allowed before starting SST, normal ventilation is allowed if all of the following conditions are met:
• SST did not detect any failures or alerts before the interruption
• No other errors that would prevent ventilation occurred
• There were no changes to the circuit type at the start of the interrupted SST
During SST, the ventilator displays the current SST status, including the test currently in progress, results of completed tests. Test data are available in Service mode where applicable or are displayed on the screen. The ventilator logs SST results, and that information is available following a power failure. The audio paused and alarm reset keys are disabled during SST, as well as the manual inspiration, inspiratory pause, and expiratory pause keys.
3.9.2 EST (Extended Self Test)
The ventilator’s extended self test (EST) function is designed to verify the ventilator’s operational subsystem integrity.
All required software support to perform EST is resident on the ventilator. EST requires approximately 10 minutes to complete.
Note: SST is not part of the EST test suite. To determine patient circuit resistance and compliance, run SST.
3.9.2.1 EST Prerequisites
Follow all identified guidelines when performing EST. Inspect all equipment required for any self test to ensure it is not damaged in any way.
1. Collect all required equipment prior to performing any self test of the ventilator. Successful self test is not possible without the use of the listed equipment.
2. Disconnect the ventilator from the patient.
3. Fully charge the primary ventilator battery.
86
3
4. Connect the ventilator to AC power using the hospital-grade power cord until completion of any self test.
5. Ensure the ventilator is powered down.
6. Ensure both air and oxygen sources register pressure between 241 kPa to 599 kPa (35 psi and 87 psi).
To perform extended self test (EST) or to access additional service functions, the ventilator must be in Service
mode. See Section 3.7.4, Service Mode, page 74 .
Note: While in the Service mode, normal ventilation is not allowed.
Warning: Always disconnect the ventilator from the patient before running EST. Running EST while the ventilator is connected to the patient can injure the patient.
Warning: A fault identified during this test indicates the ventilator or an associated component is defective.
Rectify the fault and perform any required repairs prior to releasing the ventilator for patient use, unless it can be determined with certainty that the defect cannot create a hazard for the patient, or add to the risks which may arise from other hazards.
Perform EST during any of the listed conditions:
• Prior to initial installation and first time usage of the ventilator
• Every 12 months
• Before any preventive maintenance
• Following ventilator service or repair
• As part of the ventilator’s routine performance verification
During EST, the ventilator displays the current EST status, including the test currently in progress, results of completed tests, and measured data (where applicable). The ventilator logs EST results, and that information is available following a power failure. The ventilator disables several offscreen keys located on the bezel of the GUI during EST:
• Audio paused
• Alarm reset
• Manual inspiration
• Inspiratory pause
• Expiratory pause
Run tests either as a group or as single tests for troubleshooting purposes.
3.9.2.2 Equipment for EST
• Covidien gold standard test circuit
• Number 1 stopper
• Air and oxygen sources, both at 241 kPa to 599 kPa (35 psi to 87 psi)
• Adult-sized exhalation filter
Note: Attempts to run EST with a neonatal filter can cause some EST tests to fail.
Note: If using Air Liquide™, Dräger™, and SIS air or oxygen hose assemblies, certain EST tests may fail when using supply pressures less than 345 kPa (50 psi) based on excessive hose restriction.
3.9.3 EST Test Sequence
Note: If the ventilator has not reached normal operating temperature from recent usage, allow it to warm up for at least 15 minutes in Service mode prior to running EST to ensure accurate testing.
To perform EST
3
87
1. Review and perform all self test prerequisites. See
Section 3.9.2.1, EST Prerequisites, page 86
.
2. Collect the appropriate equipment. Section 3.9.2.2, Equipment for EST, page 87 .
3. Access Service mode. See Section 3.7.4, Service Mode, page 74 .
4. Verify that all calibration tests under the Calibration tab have passed.
5. Touch the Self Test tab from the horizontal banner at the top of the monitoring screen.
6. Touch the EST tab from the menu options on the left side.
7. Touch Run All to run all tests in sequence or select the desired individual test.
8. Choose one of the available options: touch Accept to continue; touch Cancel to go back to the previous screen; or touch Stop to cancel EST.
9. Follow the prompt to remove the inspiratory filter and connect the gold standard circuit.
10. Touch Accept.
11. Follow prompts to complete EST. The EST tests require operator intervention, and will pause indefinitely for a response. See
Section 3.9.3, EST Test Sequence, page 87
.
12. At the Disconnect O
2
prompt, disconnect the high pressure oxygen source.
13. At the Disconnect air prompt, disconnect the high pressure air source.
14. At the Connect air and O
2
prompt, connect both high pressure air and oxygen sources.
15. Touch Run All or select the desired individual test. After each test, the ventilator displays the results.
16. If a particular test fails, either repeat the test or perform the next test.
17. When all of the EST tests complete, review test results by touching each individual test listed on the left side of the GUI.
18. At the top of the GUI, touch Exit.
19. Touch Confirm at the prompt to return to normal ventilation mode. The ventilator reruns POST and then displays the ventilator startup screen.
Table 22.
EST Tests
EST Test Step
Zero Offset
Function Required User
Interaction
Follow prompts
Leak Test
Mix Leak
Mix PSOL
Mix Accumulator
Circuit Pressure
Tests inspiratory and expiratory pressure transducers and flow sensors at ambient pressure.
Determines ability of system to hold pressure.
Verifies integrity of the mix system.
Verifies mix PSOL function.
Verifies mix accumulator pressure sensor and overpressure switch function.
• Checks inspiratory and expiratory autozero solenoids.
• Cross-checks safety valve, inspiratory and expiratory pressure transducers at various pressures.
• Verifies the autozero solenoid’s function.
Verifies all flow sensors and PSOLs at specified flow volumes.
Follow prompts
Follow prompts
None
None
None
None Flow Sensor Cross
Check Test
Delivery PSOL
Exhalation Valve (EV)
Loopback
Exhalation Valve (EV)
Pressure Accuracy
Verifies delivery PSOL current function.
Verifies exhalation valve current and loopback current are within range.
Verifies current versus pressure values in flash memory correspond with actual installed exhalation valve.
None
None
None
88
3
Table 22.
EST Tests (continued)
EST Test Step Function
Exhalation Valve (EV)
Performance
Exhalation Valve (EV)
Velocity Transducer
Safety System
Verifies the exhalation valve operates within specifications of the last exhalation valve calibration.
Verifies the velocity transducer is sending a signal and the control circuit recognizes it. It does not verify the quality of the signal.
Tests safety valve operation.
Backup Ventilation Verifies backup ventilation systems: mix, inspiratory, and exhala‐ tion.
Communication Verifies GUI communication ports function, both serial and ether‐ net.
Internal Storage
LCD Backlight
Status Display
Verifies internal storage device function.
Verifies GUI LCD backlight intensity function.
• Verifies status display function.
• Verifies LCD function.
GUI Alarms
BD Alarms
• Communicates with BD CPU.
• Communicates with compressor, if installed.
Tests GUI alarm indicators, cycling through each alarm status indi‐ cation.
Verifies BD audible alarm is functional. Also verifies power fail capacitor can operate LOSS OF POWER alarm.
Rotary Knob Test Verifies knob rotation function.
Offscreen Key Test Verifies GUI bezel key function.
Ventilatory Battery Tests ventilator battery and power distribution.
Run only if compressor installed
Compressor Battery Tests compressor battery function, as well as compressor power system and fan function.
Compressor Tests overall compressor operation: pressure transducer, fan, motor, and pressure relief valve.
Compressor Leak Checks compressor system for leaks.
Compressor Per‐ formance
Tests compressor operational performance under load.
Required User
Interaction
None
None
None
None
None
None
None
None
Follow prompts
Follow prompts
Follow prompts
Follow prompts
Follow prompts
Follow prompts
Follow prompts
None
None
3.9.4 EST Test Results
Table 23.
EST Test Step Results
Pass
Test Status
ALERT
Failed
Meaning
Individual EST test passed.
The test result is not ideal, but is not critical.
If EST is in progress, it halts further testing and prompts for decision.
EST not successfully passed.
Response
No need to do anything unless prompted by the ventilator
When the system prompts, select:
Repeat Test
Next Test
Stop, then touch Accept.
Select:
Repeat Test,
89
3
Table 23.
EST Test Step Results (continued)
Test Status Meaning Response
Next Test
Stop, then touch Accept.
Run all EST tests.
NEVER RUN Test still requires successful PASS.
a WARNING —Completing EST with an ALERT status for an individual test produces an Override EST button. Choose to override the ALERT status and authorize ventilation only when absolutely certain this cannot create a patient hazard or add to risks arising from other hazards. To override the alert, touch Override EST, then touch Accept.
When EST completes all of the tests, analyze the results.
Table 24.
Overall EST Outcomes
Final Outcome
PASSED
OVERRIDDEN
FAIL
Meaning
All EST tests passed.
ALERT status overridden by user.
One or more critical faults were detec‐ ted. The ventilator enters the SVO state and cannot be used for normal ventilation until SST passes. Service is required.
Response
EST successfully completed. Select other Service mode functions or pre‐ pare for SST tests prior to returning the ventilator for patient usage.
Repair the ventilator and rerun EST.
Repair the ventilator and rerun EST.
Touching Override EST results in the following warning:
Warning: Choose to override the ALERT status and authorize ventilation only when absolutely certain this cannot create a patient hazard or add to risks arising from other hazards.
3.10 Operation Verification
Before ventilating a patient, you must perform SST and alarms tests with passing results . See
SST Test Sequence, page 83 . See
Section 6.5.5, Alarm Testing, page 138 , as well.
90
4
4 Operation
4.1 Overview
This chapter describes Puritan Bennett™ 980 Series Ventilator operation and includes the following sections:
• Setting up the ventilator
• How to use the ventilator
• How to use the ventilator’s graphical user interface (GUI)
• How to set or change main, alarm, or apnea settings
• How to test alarms
• How to calibrate, enable, or disable the O
2
sensor
• How to perform inspiratory and expiratory pause maneuvers.
• How to use Non-Invasive Ventilation (NIV)
4.2 Ventilator Function
Air and oxygen from wall sources, cylinders, or the optional compressor enter the ventilator and flow through individual oxygen and air flow sensors. The gases are then mixed in the mix module’s accumulator. A pressure-relief valve in the mix module’s accumulator prevents over-pressurization. The mix module also contains an oxygen sensor which monitors the air-oxygen mixture according to the operator-set O
2
% setting.
After the gas mixes, it flows to the inspiratory pneumatic system, where the breath delivery flow sensor measures the gas flow and controls a PSOL valve for proper breath delivery tidal volumes and pressures. The inspiratory pneumatic system contains a safety valve to avoid over-pressure conditions before flowing through bacteria filters to the patient through the inspiratory limb of the patient circuit. Upon exhalation, gas flows out the patient circuit expiratory limb, through the expiratory bacteria filter, through the exhalation valve, which includes the exhalation flow sensor, and through the exhalation port.
4.3 Ventilator Setup
Warning: To avoid interrupted ventilator operation or possible damage to the ventilator, always use the ventilator on a level surface in its proper orientation.
To set up the ventilator
1. Connect the ventilator to the electrical and gas supplies. See Figure 15, page 56
, and
2. Connect the patient circuit to the ventilator. See the figures on pages Figure 24 and
adult/pediatric or neonatal patient circuits, respectively.
3. Turn the ventilator on using the power switch. See Figure 8, page 40 .
4. Before ventilating a patient, run SST to calculate the compliance and resistance with all items included in the patient circuit. See
Section 3.9.1.2, SST Test Sequence, page 83 .
4.4 User Interface Management
The user interface is structured with a GUI and a status display. The GUI provides access to ventilator controls and patient data. The status display is a small LCD panel which acts as a back up to the GUI in the event of a GUI failure.
See
Section 2.11.1.3, Status Display, page 41
for more information about the status display.
The status display is not interactive.
During normal ventilator operation, the following information appears on the status display:
4
91
• Current power state (AC or DC)
• Batteries installed / charge status (BDU and compressor, if present)
• Visual indication of audible alarm volume
• Circuit pressure graph displaying P
PEAK
, PEEP, and pressure-related alarm settings
4.4.1 Using the GUI
The GUI is used to interact with the ventilator while it is ventilating a patient or in any of its operating modes.
Caution: Do not lean on the GUI or use it to move the ventilator. Doing so could break the GUI, its locking mechanism, or tip the ventilator over.
The GUI is divided into several areas.
Figure 33.
Areas of the GUI
1.
Prompt area — Located beneath the waveforms. Any prompts or messages related to soft or hard bounds display here. A soft bound is a selected value that exceeds its recommended limit and requires acknowledgment to continue. Hard bounds have minimum and maximum limits beyond which values cannot be selected, but if the desired value is equal to a settings hard bound, then it is allowable.
2.
Menu tab — Located on the left side of the GUI screen. Swiping the tab to the right and touching Setup causes the Vent, Apnea, and Alarm tabs to appear. Touching those tabs opens screens so that changes to ventilator settings, apnea settings and alarm settings can be made.
3.
Waveform area — Located in the center of the GUI screen. Shows various breath waveforms. See
Section 3.8.2.9, Waveforms, page 80
for information on how to configure graphics.
4.
Breath Phase Indicator — During normal ventilation, the GUI displays a breath indicator in the upper left corner which shows the type of breath [Assist (A), Control (C), or Spontaneous (S)] currently being delivered to the patient, and whether it is in the inspiratory or expiratory phase. The breath indicator is updated at the beginning of every inspiration, and persists until the next breath type update. During inspiration, assist (A) and control (C) breath indicators glow green and spontaneous (S) breath indicators glow orange, each appearing in inverse video where the indicator appears black surrounded by the colored glow. See
92
4
. During the expiratory phase the breath indicators appear as solid colors (green during assist or control breaths and orange during spontaneous breaths).
5.
Vital Patient data banner — Located across the top of the GUI screen. The patient data banner displays monitored patient data and can be configured to show desired patient data. See
Section 3.8.2.7, Vital Patient
Data, page 79 for information on configuring patient data for display.
6.
Alarm banners — Located on the right side of the GUI screen. Indicates to the operator which alarms are active, and shown in a color corresponding to priority (high is red and flashing, medium is yellow and flashing, low is yellow and steady).
7.
Constant access icons — Located at the lower right of the GUI screen. This area allows access to home
(house), configure (wrench), logs (clipboard), elevate oxygen percentage (O
2
), and help (question mark) icon. These icons are always visible regardless of the function selected on the GUI.
8.
Constant access area — Consists of the Current Settings area and the Constant access icons. This area allows access to any of the patient setup variables shown in these areas. Touching an icon causes the particular menu for that variable to appear.
9.
Current settings area — Located at the lower center of the GUI screen. The ventilator’s current active settings display here. Touching any of the current settings buttons causes a dialog to appear, allowing changes using the knob.
10.
Vent Setup Button — Located at the lower left of the GUI screen. Touching this button allows access to the ventilator setup screen.
See
Section 2.11.1.3, Status Display, page 41
for information about displayed items during Service mode.
4.4.2 Adjusting GUI Viewing Properties
4.4.2.1 Screen Opacity
The opacity control enables the operator to adjust the opacity of the displayed information between
50% and 100%. At 50%, the displayed image is semi-transparent, and at 100%, the displayed image is opaque. The opacity value remains as set if power is cycled. See
Section 3.8.2.10, Opacity, page 82 for
instructions on adjusting this feature.
4.4.2.2 Pushpin Feature
The pushpin feature prevents a dialog from closing under certain conditions. Like the opacity control, the pushpin appears on the settings screen after a new patient starts ventilation.
Figure 34.
Pushpin Icon – unpinned state
4
Figure 35.
Pushpin Icon – pinned state
To use the pushpin
1. When a dialog is open, for example, if Accept or Accept ALL buttons are available, touch the unpinned pushpin icon to pin the dialog and hold it open.
2. Touch Close to close the dialog.
93
4.4.2.3 Display Brightness
Display brightness can be controlled manually. This feature is institutionally configurable.
See Section 3.8.2.3, Screen Brightness and Keyboard Backlight (Light Settings), page 77
. The brightness range is from
1% to 100% with 1% resolution. The default value is 80%.
To manually adjust display brightness
1. Press the display brightness key.
2. Slide the brightness slider to the right to increase the brightness level or to the left to decrease the brightness level. Alternatively, turn the knob to increase or decrease the brightness level. The control disappears from the screen in approximately 5 seconds.
4.4.2.4 Display Lock
The primary display provides a display lock key to prevent inadvertent changes to settings. When active, the display lock disables the touch screen, knob, and off-screen keys and illuminates an LED on the display bezel. An image of the display lock icon appears transparently over anything displayed on the GUI, should the operator attempt to use the GUI. Any new alarm condition disables the display lock and enables normal use of the GUI.
To lock and unlock the display
1. Press the display lock key on the GUI.
The keyboard LED illuminates and a transparent icon appears on the screen, indicating display lock. The icon shortly disappears, but if the operator tries to activate any of the touch screen controls, the icon reappears.
2. To unlock the display, press the display lock key again. The display lock LED turns off.
4.4.3 Using Gestures When Operating the GUI
The GUI incorporates a gesture-based interface where features can be actuated with the fingers using different motions. The following table explains gestures used with the GUI.
Table 25.
Gestures and Their Meanings
Gesture Description
Swipe Quickly brush the screen surface with the fingertip.
Used for
Opening or closing dialogs or panels that slide in and out from the screen sides or top, moving waveform data, expanding or collapsing tooltips, scrolling lists, or alarm banners, maximiz‐ ing or minimizing waveforms.
How to Use
Swipe toward the center of the screen to open dialogs or panels. Swipe toward the side of the screen (or upward if viewing the additional patient data or large font patient data panels) to close.
To move a paused waveform, swipe in the desired direction.
Swipe upward anywhere on a waveform to maximize it, and swipe downward on the maximized waveform to minimize it.
Swipe a tooltip upward to expand to a long description and downward to collapse to a short description. A downward swipe any‐ where in the patient data area opens the additional patient data panel, and another
94
4
Table 25.
Gestures and Their Meanings (continued)
Gesture
Doubletap
Drag
Touch and hold
Drag and drop
Description
Rapidly touch the screen surface twice with one finger.
Move the fingertip over the screen sur‐ face without losing contact.
Used for
Maximizing or minimizing the viewable area of a dialog, con‐ trol, or waveform, expanding or collapsing tooltips
Changing x- and y- axis scales, repositioning waveforms, mov‐ ing the waveform cursor, mov‐ ing scrollbars, scrolling lists.
Scrolling speed varies depend‐ ing upon how far outside the list boundary the finger is posi‐ tioned.
How to Use swipe on the additional patient data tab displays the large font patient data panel.
Double-tapping maximizes the viewable waveform area or shows the long descrip‐ tion of a tooltip. Double-tapping again min‐ imizes the viewable waveform area or shows the short description of a tooltip. If the control is configurable, double tapping produces the configuration pop-up menu.
Touch the axis and drag to the right to increase the waveform x-axis scale, and to the left to decrease. Touch the axis and drag upward to increase the y-axis scale and downward to decrease. To reposition wave‐ forms, touch and drag the graph to the new position. To move the waveform cursor, touch the cursor and drag it right or left. The graph responds similarly. Scroll a list by dragging the scrollbar right or left or up or down. The list scrolls according to the direc‐ tion of the finger movement. An automatic scrolling feature starts if the finger is drag‐ ged from the inside of a list to outside its boundary. The farther outside the boun‐ dary the finger is dragged, the faster the list scrolls.
N/A Touch an item and hold for at least 0.5
seconds.
Touch and drag an item to another loca‐ tion and lift finger to drop.
Displaying a tooltip dialog on whatever item is touched. The tooltip appears to glow indicat‐ ing the touch and hold action.
Dragging help icon to describe an onscreen item.
Drag the help icon, located at the lower right of the GUI screen, to the item in ques‐ tion and drop. If a blue glow appears, a tool‐ tip is available and appears with informa‐ tion about that item (for example, a control or symbol).
4.5 Ventilator Operation
Warning: Prior to patient ventilation, select the proper tube type and tube ID.
Caution: Do not set containers filled with liquids on the ventilator, as spilling may occur.
After turning on the ventilator it will display a “splash screen,” and run power on self test (POST). After the splash screen appears, the ventilator gives a choice to ventilate the same patient or a new patient, or run SST.
Ventilation parameters are entered via the graphical user interface (GUI) using the following general steps:
4
95
1. Touch the setting displayed on the GUI.
2. Turn the knob to the right to increase or to the left to decrease the value.
3. Touch Accept to apply the setting or Accept ALL to apply several settings at once.
Note: Quick Start allows for rapid setup and initiation of mechanical ventilation. Review Quick Start parameters and ensure they are consistent with institutional practice before using this feature.
To use Quick Start
1. Touch New Patient.
2. Touch the highlighted PBW button or Gender/Height.
3. Turn the knob to adjust the patient’s PBW or gender and height (if gender is selected, the height selection becomes available).
4. Touch Quick Start.
5. Connect the circuit wye adapter to the patient’s airway or interface connection. The patient is ventilated with the institutionally configured Quick Start defaults according to the PBW or gender/height entered, and circuit type used during SST. There is no prompt to review the settings and the waveforms display appears.
Note: Connecting the circuit wye adapter to the patient’s airway or interface connection prior to making the ventilation settings causes the ventilator to begin ventilation using Safety Pressure Control Ventilation (Safety
PCV) and annunciate a PROCEDURE ERROR alarm. As soon as the ventilator receives confirmation of its settings
(by touching Accept or Accept ALL ), it transitions out of safety PCV, resets the alarm, and delivers the chosen settings. See
for a listing of these settings.
To resume ventilating the same patient
1. Touch Same Patient on the GUI screen. The previous ventilator settings are displayed on the screen for review prior to applying the settings to the patient.
2. If the settings are acceptable, touch Accept to confirm. To change any settings, touch the setting, turn the knob clockwise to increase the value of the setting or counter-clockwise to decrease the value of the setting, and touch Accept to confirm. To make several settings changes at once, make the desired changes, then touch Accept ALL to confirm. The appearance of the settings changes from white, non-italic font showing the current setting to yellow italics (noting the pending setting). After the settings are accepted, the appearance changes back to white non-italic font.
3. Connect the circuit to the patient’s airway to initiate ventilation.
To ventilate a new patient
1. Touch New Patient on the GUI screen. The New Patient settings screen appears to enter the ventilation
control parameters. See Table 63, page 246
for default ventilator parameter settings.
96
Figure 36.
New Patient Settings
4
2. Enter the patient’s PBW or gender and height (if gender is selected, the height selection becomes available).
3. If the default ventilator settings are appropriate for the patient, touch Start to confirm the settings, otherwise, touch a ventilator setting and turn the knob to adjust the parameter. Continue this process for all parameters needing adjustment.
4. Touch Accept or Accept ALL to confirm the changes.
5. Connect the circuit to the patient’s airway to start ventilation.
4.5.1 Ventilator Settings
Warning: The ventilator offers a variety of breath delivery options. Throughout the patient’s treatment, the clinician should carefully select the ventilation mode and settings to use for that patient based on clinical judgment, the condition and needs of the patient, and the benefits, limitations and characteristics of the breath delivery options. As the patient’s condition changes over time, periodically assess the chosen modes and settings to determine whether or not those are best for the patient’s current needs.
The following ventilator settings appear at the new patient setup screen:
• Predicted Body Weight (PBW) — Adjust the patient’s PBW, or select the patient’s gender and height. See
Section 4.6, Predicted Body Weight (PBW) Calculation, page 104 .
• Ventilation type — Determines the type of ventilation to be delivered [Invasive or Non-Invasive (NIV)].
– Invasive — Conventional ventilation using endotracheal (ET) or tracheostomy (trach) tubes.
– Non-Invasive (NIV ) — Ventilation using non-vented full face masks, nasal masks, infant nasal prongs, or
uncuffed ET tubes. See Section 4.7, Non-Invasive Ventilation (NIV), page 105 .
• Mode — Specify the breathing mode (A/C (assist/control), SIMV (synchronous intermittent mandatory ventilation), SPONT (spontaneous ventilation), BiLevel, or CPAP.
• Mandatory type — Select PC (pressure control), VC (volume control), or VC+ (volume control plus).
• Spontaneous type — If SIMV or BiLevel was selected as the Mode, specify PS (pressure support) or TC (tube compensation. If SPONT was selected as the Mode, specify PS (pressure support), TC (tube compensation), or
VS (volume support) or PAV+™ (Proportional Assist™ ventilation).
Note: VS, PAV+™, and TC are only available during Invasive Ventilation.
97
4
• Trigger type — Select pressure- triggering (P
TRIG
) or flow-triggering ( ⩒
TRIG
). Pressure-triggering is not available when ventilation type is NIV. If ventilating a neonatal patient, only flow triggering is available.
Other ways to access the vent setup screen:
• Touch the vent setup button at the bottom left of the GUI display
• Swipe the menu tab on the left side of the GUI and touch Setup
Figure 37.
Open Menu Tab
1 Setup button
98
Figure 38.
New Patient Setup Screen
4
To enter settings into the ventilator
1. Select Vent type, Mode, Mandatory Type, Spontaneous Type and Trigger Type by touching the corresponding button.
2. Touch the ventilator setting button needing changes.
3. Adjust the setting value.
4. Continue in this manner until all changes are made, then touch Accept or Accept ALL .
5. Touch START. Ventilation does not begin until the breathing circuit is connected to the patient’s airway. After
ventilation begins, waveforms begin plotting on the displayed waveforms axes. See Section 3.8.2.9,
for information on setting up the graphics display.
If changes to any settings are required, return to the Vent Setup screen as described above.
Note: A yellow triangle icon appears on tabs and buttons displayed on the GUI containing unread or unviewed items. When the item containing the icon is touched, the icon disappears.
Note: To make any settings changes after completing patient setup, touch the Vent tab on the left side of the
Setup dialog and make settings changes as described above. The current setting appears in white font and changes to yellow italics to note the new value is pending.Touch Accept or Accept ALL to confirm a single change or a batch of changes. Once the settings are accepted, their appearance changes to white font.
Note: Selecting Quick Start , Accept , Accept ALL or Start from the Setup dialog implements all settings in ALL four setup tabs (Vent Setup, Apnea, Alarms, and More Settings) and dismisses the setup dialog.
4.5.1.1 Tube Compensation
Tube compensation is a spontaneous type selected during ventilator setup. It allows the ventilator to deliver
To enable TC
4
99
1. Touch the Vent tab on the GUI screen. See
2. Touch SPONT for the mode selection.
3. Touch TC for Spontaneous type.
settings.
5. Ensure to select the tube type (either endotracheal or tracheostomy) and set the tube ID to correspond to patient settings.
6. After making the changes, touch Accept to apply the new settings, or Cancel to cancel all changes and dismiss the dialog.
4.5.1.2 Adjust Tube Type, Tube ID, and Humidification
Warning: To prevent inappropriate ventilation with TC, select the correct tube type ET or Tracheostomy) and tube inner diameter (ID) for the patient’s ventilatory needs. Inappropriate ventilatory support leading to over-or under-ventilation could result if an ET tube or trach tube setting larger or smaller than the actual value is entered.
To select new settings for the tube, follow these steps
1. Touch Vent Setup on the GUI screen to display the ventilator setup screen.
2. Touch Tube Type or Tube ID for the value to be changed.
3. Turn the knob to change the setting.
4. Make other tube settings, as necessary.
5. Touch Accept or Accept ALL to apply the new settings, or Cancel to cancel all changes and dismiss the dialog.
Note: The tube type and tube ID indicators flash if TC is a new selection, indicating the need for entry of the correct tube type and tube ID.
To select new settings for the humidifier, follow these steps
1. From the Ventilator setup screen, touch the More Settings tab. A dialog appears containing selections for humidifier type and volume.
Note: A Humidification Volume button appears below the selection only if Non-Heated Expiratory Tube or
Heated Expiratory Tube is selected as the humidifier type.
2. Turn the knob to enter a value equal to the dry volume of the humidifier chamber being used.
3. Touch Accept or Accept ALL to apply the new settings, or Cancel to cancel all changes and dismiss the dialog.
Table 26 lists the allowable ventilator settings according to patient type and ventilation type.
Table 26.
Allowable Ventilator Settings
Patient type
Ventilation type
Mode
Invasive
Adult
NIV
Mandatory type
A/C, SIMV,
SPONT, BiLe‐ vel
A/C, SIMV,
SPONT
PC, VC, VC+ PC, VC vel
Pediatric
Invasive
A/C, SIMV,
SPONT, BiLe‐
NIV
A/C, SIMV,
SPONT
PC, VC, VC+ PC, VC vel
Neonatal
Invasive
A/C, SIMV,
SPONT, BiLe‐
NIV
A/C, SIMV,
SPONT, CPAP
PC, VC, VC+ PC, VC
100
Table 26.
Allowable Ventilator Settings (continued)
Spontane‐ ous type
PS, TC, VS,
PAV+™
(≥25 kg)
PS
Trigger type ⩒
TRIG
, P-Trig ⩒
TRIG
PS, TC,
(≥7.0 kg), VS
⩒
TRIG
PS
, P-Trig ⩒
TRIG
PS, VS
⩒
TRIG
PS
⩒
TRIG
Note: To use neonatal ventilator settings, the NeoMode 2.0 software option must be installed on the ventilator, or a Puritan Bennett™ 980 Neonatal Ventilator must be in use.
4.5.2 Apnea Settings
After making the necessary changes to the ventilator settings touch the Apnea tab on the left side of the setup dialog. Although changing the apnea settings is not required, confirm the default settings are appropriate for the patient. Apnea ventilation allows pressure control or volume control breath types. Parameters in pressure-controlled apnea breaths include f, P f, V
T
, ⩒
MAX
, flow pattern, O
2
%, and T
A
.
I
, T
I
, O
2
%, and T
A
. Volume-controlled apnea breath parameters are
Note: If Quick Start is chosen, and the apnea tab on the vent setup screen shows a yellow triangle, indicating the apnea settings have not been reviewed.
Figure 39.
Apnea Setup Screen
4
To set apnea parameters
1. Select the desired apnea breath type (PC or VC).
2. Enter the desired apnea settings in the same manner as for the ventilator settings.
3. Touch Accept or Accept ALL to confirm apnea settings.
During apnea pressure ventilation, apnea rise time% is fixed at 50%, and the constant parameter during a respiratory rate change is T .
I
101
4
4.5.3 Alarm Settings
After accepting the apnea settings, the display returns once more to show the waveforms. Return to the vent setup dialog and touch the Alarms tab on the left side of the GUI screen. The alarms screen appears with the default alarm settings. See
. Review and adjust the alarm settings appropriately for the patient.
Note: If Quick Start is chosen, the alarms tab on the dialog shows a yellow triangle, indicating the alarm settings have not been reviewed.
Note: See
Table 64, page 252 for new patient default alarm values. These
defaults cannot be changed. The clinician can adjust alarm settings by following the procedure described. The alarm settings are retained in memory when the ventilator’s power is cycled, and current settings revert to new patient defaults when a new patient is selected.
Figure 40.
Alarms Settings Screen
To adjust the alarm settings
1. Touch each alarm setting slider of the alarm(s) to change. Alarm settings are available for P
PEAK
V
TE MAND
, V
TE SPONT
, and V
TI
parameters.
, f
TOT
, ⩒
E TOT
,
2. Turn the knob clockwise to increase the value, or counterclockwise to decrease the value.
3. Continue until all desired alarms are set as necessary.
4. Touch Accept ALL to confirm the alarm settings.
Note: There is an additional alarm setting for TC, PAV+™, VS, and VC+ breath types: high inspired tidal volume
( ⤒ V
TI
). This alarm condition occurs when the inspired tidal volume is larger than the setting value. A ↑ V
TI
alarm will also cause breath delivery to transition to the exhalation phase to avoid delivery of excessive inspiratory volumes.
Warning: Prior to initiating ventilation and whenever ventilator settings are changed, ensure the alarm settings are appropriate for the patient.
Warning: Setting any alarm limits to OFF or extreme high or low values, can cause the associated alarm not to activate during ventilation, which reduces its efficacy for monitoring the patient and alerting the clinician to situations that may require intervention.
102
See
Section 3.8.2.6, Alarm Volume, page 79
to ensure alarm volume is adjusted properly.
Note: Do not block the patient wye while the ventilator is waiting for a patient connection. Otherwise the blockage could imitate a patient connection.
Note: A sample alarm tone sounds for verification at each volume level change. Readjust the alarm volume by moving the alarm volume slider to increase or decrease the alarm volume.
4.5.4 Alarm Screen During Operation
During ventilator operation, the alarm screen appears with indicators to let the operator know the current patient data value for each parameter (item 1), the parameter alarm settings (items 2 and 3), recent range of patient data values for the last 200 breaths (item 4). If an alarm occurs, the slider and corresponding limit button show a color
matching the alarm’s priority. See Figure 41, page 103
.
Figure 41.
Alarm Screen During Operation
4
1 Pointers show current value of patient data corresponding to the alarm parameter
2 High alarm setting (in this case ⤒ V
TE SPONT
)
3 Low alarm setting (in this case ⤓ V
TE SPONT
)
4 Range of patient data values for the particular parameter during the last 200 breaths
4.5.5 Making Ventilator Settings Changes
If, during ventilation, settings changes are necessary that don’t involve changes to PBW, Mode, Breath types, or
Trigger types, the current settings area located at the lower portion of the GUI screen can be used. See
Figure 41, page 103 for the location of the current settings area.
To change a ventilator setting using the “current settings” area
1. In the current settings area, touch the parameter whose value needs to be changed. A dialog appears containing buttons for all ventilator settings, with the selected setting highlighted.
2. Touch and turn the knob for any other settings that need to be changed.
3. Touch Accept or Accept ALL .
To change a setting using the vent setup button
1. Touch the vent setup button.
4
103
2. Change the settings as described previously.
3. Touch Accept or Accept ALL to confirm the changes
The ventilator settings and the alarm settings chosen remain in memory after a power cycle, as long as the same patient is chosen when the ventilator is set up again. If a new patient is being ventilated, the ventilator and alarm settings revert to their default values. If all power is lost (both AC and battery), the ventilator and alarm settings in effect prior to the power loss are automatically restored if the power loss duration is 5 minutes or less. If the power loss lasts longer than 5 minutes, ventilation resumes in Safety PCV. Ventilator and alarm settings must be reset for
the patient being ventilated. See Table 53, page 236 for a list of these settings.
To use the previous setup button
1. To return to the previous setup, touch the vent setup button then touch Previous Setup on the GUI screen.
The ventilator restores the main control and breath settings previously used, as well as the alarm and apnea settings, and prompts a review by highlighting the previous values in yellow. The ventilator, alarm, and apnea settings tab text is also shown in yellow and the tabs show a yellow triangle, indicating there are previous settings that have not been reviewed.
2. If the settings are acceptable, touch Accept or Accept ALL .
The Previous Setup button disappears when the previous settings are confirmed and reappears when ventilating with new settings.
4.5.6 Constant Timing Variable During Rate Changes
A breath timing graph appears at the bottom of the setup screen which illustrates the relationship between inspiratory time, expiratory time, I:E ratio, respiratory rate, and the effects on breath timing due to flow pattern, tidal volume, and ⩒
MAX
during mandatory PC, VC, BiLevel, or VC+ breaths. With BiLevel, PC and VC+ breaths, three padlock icons are located underneath the breath timing graph allowing the operator to select, from left to right, T
I
I:E ratio, or T
E
as the constant variable during rate changes (or T
H
, T
H
:T
L
ratio, or T
L
in BiLevel). If the ventilation mode is SPONT, the padlock icons do not appear, and the breath timing graph only displays T
I
for a manual inspiration.
If the mandatory type is VC, the icons do not appear, but the breath timing graph displays T
I
, I:E ratio, and T
E
.
To choose a constant timing variable for rate changes
1. Touch a padlock icon corresponding to the parameter to make constant during rate changes (this changes the padlock’s appearance from unlocked to locked). The locked parameter glows in the settings area.
2. Turn the knob to adjust the parameter’s value.
3. Touch Accept .
4.6 Predicted Body Weight (PBW) Calculation
Many default ventilator and alarm settings are based on patient PBW. Either through the entry of height and gender or directly via setting PBW, the PBW range spans at least 3.5 kg (7.7 lb) through at least the 155 kg (342 lb) male and the 150 kg (331 lb) female. Understanding how the ventilator operates at the very low end of the range of PBW requires awareness that an entry or prediction for PBW drives the value of a delivered volume, which has a lower limit of 2.0 mL (if using the NeoMode 2.0 option). Data for adult male and female PBW as a function of height were calculated by applying the equations presented on www.ards.net.
Assume the ventilator (via direct height or PBW entry) registered a PBW of 0.3 kg. If a delivered volume of 4 mL/kg
(PBW) was specified, the required volume would equal only 1.2 mL, which is less than the ventilator minimum of
2.0 mL. At a desired 4 mL/kg, the infants’ PBW would need to be at least 0.5 kg or the desired volume must be reset to greater than 4 mL/kg (PBW). Once the PBW of the premature infant approaches 1.0 kg (2.2 lb), this restriction disappears.
104
4
After entering PBW, review and change all settings as needed.
The correlation function PBW = height was derived from the sources referenced. For subjects whose body weight/height data define the range of PBWs that include the 20- to 23-week gestational-age neonates and the young male and female adolescent adults at the foot of the ARDS tables, their PBW values were taken as the 50th percentile numbers in the Fenton tables and the CDC and NCHS charts and tables, respectively. Note that the
Fenton tables provided the exclusive information for premature and infant data between 20 weeks and 50 weeks
of fetal and gestational growth.
Note: Any repeated values noted in the tables are the result of decimal rounding.
4.7 Non-Invasive Ventilation (NIV)
Warning: Use only non-vented patient interfaces with NIV. Leaks associated with vented interfaces could result in the ventilator’s inability to compensate for those leaks, even if Leak Sync is employed.
Warning: Full face masks used for Non-Invasive Ventilation should provide visibility of the patient’s nose and mouth to reduce the risk of emesis aspiration.
Warning: When using NIV, the patient’s exhaled tidal volume (V patient data V
TE
) could differ from the ventilator’s monitored
TE
reading due to leaks around the mask. When NIV is selected, Leak Sync is automatically enabled.
See
Non-Invasive Ventilation (NIV) is used when the clinician determines a mask or other non-invasive patient interface rather than an endotracheal tube would result in the desired patient outcome.
4.7.1 NIV Intended Use
NIV is intended for use by neonatal, pediatric, and adult patients possessing adequate neural-ventilatory coupling and stable, sustainable, respiratory drive.
4.7.2 NIV Breathing Interfaces
The PB980 ventilator can be used with non-vented NIV interfaces that comply with ISO 5356-1:
Full face Mask — ResMed Mirage™* non-vented full-face mask
Nasal Mask — ResMed Ultra Mirage™* non-vented mask
Infant Nasal Prongs — Argyle ™*
Medin ™*
nasal prongs, Hudson RCI ™* infant nasal prongs, Fisher & Paykel ™* nasal prongs,
nasal prongs, and Ram cannula
Uncuffed neonatal ET tube — Shiley™ Uncuffed tracheal tube, Murphy (3.0 mm)
4.7.3 NIV Setup
NIV can be initiated from either the new patient setup screen during Vent startup or while the patient is being
ventilated invasively. See Table 27
to use NIV patient setup information.
4
1 Fenton TR, BMC Pediatrics 2003, 3:13. http://www.biomedcentral.com/1471–2431/3/13.
2
3
Hamill, PV V. 1977 NCHS growth curves for children birth to 18 years for the United States: National Center for Health Stat
(Vital and Health Statistics: Series 11, Data from the National Health Survey; no. 165) (DHEW publication; (PHS) 78–1650).
1977.
Kuczmarski RJ, Ogden CL, Guo SS, et al. 2000 CDC growth charts for the United States: Methods and development.
National Center for Health Statistics. Vital Health Stat 11(246). 2002.
105
Table 27.
Setting Up a Patient for NIV
To set up a new patient
1. Turn the ventilator on.
2. Select New Patient.
3. Enter patient’s PBW or gender and height.
4. Touch NIV ventilation type.
5. Select mode.
6. Select mandatory type.
7. Complete ventilator settings, including apnea and alarm settings.
To set up a patient currently being ventilated
1. Touch or swipe the menu tab on the left side of the
GUI.
2. Touch Vent Setup.
3. Perform steps 4 through 7 as if setting the ventilator up for a new patient.
4. Review the settings, including apnea and alarm set‐ tings and change if necessary.
To function as intended, D patient.
SENS
needs to be set between the LEAK data value displayed while the patient is connected, and the LEAK data value displayed when the interface is open to ambient and not connected to the
To set D
SENS
with NIV interfaces when Leak Sync is enabled
1. After adjusting the patient settings, start ventilation.
2. Ensure that Leak Sync is enabled and set D
SENS
to the highest setting.
3. Select LEAK (L/min) to be displayed in the patient data banner.
4. With the NIV interface connected to the breathing circuit and open to ambient, use the patient data value called LEAK to quantify the leak in L/min.
5. Set the D
SENS
(in L/min) below the leak rate (in L/min) to ensure that the disconnect alarm is violated during a disconnect. Note that this will cause a CIRCUIT DISCONNECT alarm.
6. Connect the patient interface to the patient and ensure that the CIRCUIT DISCONNECT alarm is resolved.
7. D
SENS
needs to be set higher than the Leak data value displayed while the patient is connected and lower than the Leak data value displayed when the interface is open to ambient and not connected to the patient.
8. Periodically assess the leak rate, especially with PEEP changes, and adjust the D is connected to the patient.
SENS
setting as needed to ensure the presence of an alarm during disconnect and the absence of nuisance alarms while the interface
9. Always use alternative methods of monitoring during NIV.
4.7.4 Conversion from Invasive to NIV Ventilation Type
Note: Before switching to non-invasive ventilation from invasive ventilation when the EtCO2 sensor was used, ensure the EtCO2 option is disabled.
Warning: For proper ventilation when changing the ventilation type on the same patient, review the automatic settings changes described. Adjust appropriately based on the relevant tables.
Some ventilator settings available during Invasive Ventilation are not available during NIV. See
automatic settings changes when changing ventilation type from Invasive to NIV.
Table 28.
Invasive to NIV on Same Patient
Current invasive setting
Breath mode: BiLevel
Breath mode: SIMV or SPONT
Mandatory type: VC+
Spontaneous type: any type except PS
New NIV setting
Breath Mode: A/C
High T
I SPONT
( ⤒ T
I SPONT
) limit setting available
Mandatory type:
Neonatal: PC
Adult and pediatric: VC
Spontaneous type: PS
106
Table 28.
Invasive to NIV on Same Patient (continued)
Current invasive setting
Trigger type: pressure
Alarm settings:
⤓ V
D
TE SPONT
SENS
⤓ P
PEAK
(if applicable), ⤓⩒
E TOT
, ⤓ V
TE MAND
,
INSPIRATION TOO LONG (not user-settable)
New NIV setting
Trigger type: flow
(Flow triggering is the only allowable trigger type dur‐ ing NIV)
Alarm settings: ⤓ P
PEAK
, ⤓⩒
E TOT
, ⤓ V
TE MAND
, ⤓ V
TE SPONT default to NIV new patient values. See
Table 63, page 246 . INSPIRATION TOO LONG alarm not available.
D
SENS
setting defaults to OFF if Leak Sync is disabled.
Note: In any delivered spontaneous breath, either Invasive or NIV, if pressure support is set to 0 cmH
2 always a target inspiratory pressure of 1.5 cmH
2
O applied.
O, there is
When in NIV, the vent setup button’s appearance changes, letting the operator know the ventilation type is NIV.
Figure 42.
Vent Setup Button “NIV” Indicating NIV ventilation type
4
4.7.5 Conversion from NIV to Invasive Ventilation Type
Table 29 shows automatic settings changes made when changing ventilation type from NIV to Invasive.
Table 29.
NIV to Invasive on Same Patient
Current NIV setting
Ventilator settings: ⤒
TI SPONT
Alarm settings:
⤓ P
PEAK
, ⤓⩒
E TOT
, ⤓ V
TE MAND
, ⤓ V
TE SPONT
D
SENS
N/A
New invasive setting
Alarm settings: Default to new patient values depend‐ ent upon selected Invasive ventilator settings. See
. INSPIRATION TOO LONG alarm becomes available.
D
SENS
setting defaults to Invasive new patient value.
See
107
4
4.7.6 High Spontaneous Inspiratory Time Limit Setting
NIV includes a setting in SIMV or SPONT modes for High Spontaneous Inspiratory Time limit ( ⤒ T
I SPONT
). When a patient’s inspiratory time reaches or exceeds the set limit, the ventilator transitions from inspiration to exhalation, and the ↑ T
I SPONT
symbol appears at the lower left on the GUI screen, indicating the ventilator has truncated the
). The ( ⤒ T
I SPONT
) setting does not restrict changes to PBW; if the PBW is decreased, ( ⤒ T
I
SPONT
) may decrease automatically to remain within its allowable limits.
Figure 43.
⤒ T
I SPONT
Indicator
Warning: No audible alarm sounds in conjunction with the visual ⤒ T
I SPONT appear in any alarm log or alarm message.
indicator, nor does the indicator
It is possible the target inspiratory pressure may not be reached if the ⤒ T system leaks are so large as to cause the ventilator to truncate the breath at the maximum allowable ⤒ T setting.
I SPONT
setting is not long enough, or if
I SPONT
Note: To reduce the potential for not reaching the target pressure, minimize the leaks in the system and increase the rise time% and decrease the E
SENS
setting, if appropriate.
4.7.7 NIV Apnea Setup
Set the patient’s apnea parameters as described. See
Section 4.5.2, Apnea Settings, page 101
. NIV does not change the way apnea parameters are set.
4.7.8 NIV Alarm Settings
The system initially sets most alarm settings based on the patient’s PBW. Review all alarm settings, and change as necessary, but startup does not require confirmation of the settings. Alarm settings are made in exactly the same way in NIV as for Invasive Ventilation.
108
Figure 44.
Default NIV Alarm Settings
4
Touch the Alarms tab at any time during ventilation to show the current limits and the monitored patient value
color changes based on alarm priority See
for colors and meanings of alarms and their priorities.
Note: The upper and lower limits of an alarm cannot conflict with each other.
Note: The upper limits for the spontaneous exhaled tidal volume and mandatory exhaled tidal volume alarms are always the same value. Changing the upper limit of one alarm automatically changes the upper limit of the other.
4.8 Manual Inspiration
A manual inspiration is an operator-initiated mandatory (OIM) inspiration. When the operator presses the manual inspiration key while the ventilator is in a mode that includes mandatory breaths (includ‐ ing mixed modes BiLevel and SIMV), the ventilator delivers the manual inspiration using the currently set mandatory breath parameters. A manual inspiration performed while the ventilator is in the SPONT mode uses the currently set apnea breath parameters. A volume-based manual inspiration is com‐ pliance-compensated. Pressing the manual inspiration key while in BiLevel mode will transition from
T
H
to T
L
or T
L
to T
H
depending on when in the breath cycle the key was pressed.
4.9 Respiratory Mechanics Maneuvers
To access respiratory mechanics maneuvers
1. Touch or swipe the Menu tab on the left side if the GUI.
2. Touch RM.
4
109
Figure 45.
RM in Menu Tab
3. Touch the particular tab for the desired maneuver.
Figure 46.
Respiratory Maneuver Tabs
4. Follow the prompts on the GUI screen.
5. Accept or reject the maneuver results. If the result is accepted, its value is saved.
4.9.1 Inspiratory Pause Maneuver
An inspiratory pause maneuver closes the inspiration and exhalation valves and extends the inspiratory phase of a single, mandatory breath for the purpose of measuring end inspiratory circuit pressure in order to calculate inspiratory Plateau Pressure (P
PL
), lung Static Compliance (C
STAT
), and Static Resistance (R
STAT
) of the respiratory system. Pressures on either side of the artificial airway are allowed to equilibrate, which determines the pressure
110
4 during a no-flow state. A request for an inspiratory pause is ignored during apnea ventilation, safety PCV, OSC, BUV, and in Stand-By state. Inspiratory pauses are allowed in A/C, SIMV, BiLevel and SPONT modes. If an inspiratory pause maneuver has already occurred during the breath, a second inspiratory pause maneuver is not allowed.
Inspiratory pauses can be classified as automatic or manual . The automatic inspiratory pause lasts at least 0.5
second but no longer than 3 seconds. A manual inspiratory pause starts by pressing and holding inspiratory pause key. The pause lasts for the duration of the key-press (up to 7 seconds).
To perform an automatic inspiratory pause
1. Press and release the inspiratory pause key on the GUI bezel or touch and release START if performing an inspiratory pause from the GUI screen as shown above. The ventilator performs the inspiratory pause maneuver and displays P
PL
, C
STAT
, and R
STAT
along with the date and time.
2. Touch the Accept or Reject button to save or dismiss results. If the Accept button is touched, the results are displayed.
Cancel an automatic inspiratory pause maneuver by touching Cancel on the GUI screen.
To perform a manual inspiratory pause
1. Press and hold the inspiratory pause key on the GUI bezel or touch and hold START on the GUI screen if performing an inspiratory pause from the GUI screen as shown above. The ventilator prompts that the maneuver has started, and to release to end the maneuver. The ventilator performs the inspiratory pause maneuver and displays P
PL
, C
STAT
, and R
STAT
, along with the date and time.
2. Touch Accept or Reject to save or dismiss results. If Accept is touched, the results are displayed.
Cancel a manual inspiratory pause maneuver by releasing the inspiratory pause key.
4.9.2 Expiratory Pause Maneuver
An expiratory pause extends the expiratory phase of the current breath for the purpose of measuring end expiratory lung pressure (PEEP
TOT
) or total PEEP. It has no effect on the inspiratory phase of a breath, and only one expiratory pause per breath is allowed. For I:E ratio calculation purposes, the expiratory pause maneuver is considered part of the exhalation phase.
During an expiratory pause, both inspiratory and exhalation valves are closed, allowing the pressures on both sides of the artificial airway to equilibrate. This allows intrinsic PEEP (PEEP
I
) to be calculated. PEEP
I
is PEEP
TOT
minus the set PEEP level. An expiratory pause can be either automatically or manually administered, and is executed at the next mandatory breath in A/C, SIMV, or BiLevel modes. In SIMV, the breath cycle in which the pause becomes active (when the next scheduled ventilator initiated mandatory (VIM) breath occurs) will be extended by the amount of time the pause is active. For A/C and SIMV, the expiratory pause maneuver is scheduled for the next end-of-exhalation prior to a mandatory breath. In BiLevel the expiratory pause maneuver is scheduled for the next end-of-exhalation prior to a transition from P
L
to P
H
. During the expiratory pause maneuver, PEEP
I
and PEEP
TOT equilibration time values are displayed and regularly updated because stabilization of one of these values can indicate the pause can be ended. During the expiratory pause, the Apnea Interval T
A
is extended by the amount of time the pause is active. Expiratory pause maneuver requests are ignored if the ventilator is in apnea ventilation, safety PCV, OSC, BUV, and Stand-By state. Additionally, SEVERE OCCLUSION alarms are suspended during expiratory pause maneuvers. If flow triggering is active, backup pressure sensitivity (P
SENS
) detects patient breathing effort.
Maximum duration for a manual expiratory pause is 15 seconds and 3 seconds for an automatic expiratory pause.
During a manual or automatic expiratory pause, PEEP
I
and PEEP
TOT
appear on the GUI with the next VIM to allow the clinician to view when these values are stabilized, indicating the maneuver can be ended.
To perform an automatic expiratory pause
4
111
• Press and release the expiratory pause key on the GUI or touch and release START if performing the expiratory pause from the GUI screen. The ventilator performs the expiratory pause maneuver and displays a circuit pressure graph, PEEP
TOT
, and PEEP
I
, along with the date and time.
To perform a manual expiratory pause
1. Press and hold the expiratory pause key on the GUI bezel or touch and hold the START button if performing the expiratory pause from the GUI screen. The ventilator prompts that the maneuver has started, and to release the button to end the maneuver. The ventilator performs the expiratory pause maneuver and displays a circuit pressure graph, PEEP
TOT
, and PEEP
I
, along with the date and time.
2. Accept or reject the pause results.
To cancel an expiratory pause maneuver, touch Cancel on the GUI screen.
4.9.3 Other Respiratory Maneuvers
To perform other respiratory maneuvers, touch the corresponding tab on the desired maneuver, and follow the prompts on the GUI screen.
4.10 Oxygen Sensor Function
The ventilator’s oxygen sensor monitors O
2
%. This cell is mounted in the mix module in the BDU and monitors the percentage of oxygen in the mixed gas delivered to the breathing circuit (it may not reflect the actual oxygen concentration in the gas the patient inspires).
See the Puritan Bennett™ 980 Series Ventilator Service Manual for instructions on replacing the O
2
sensor.
New patient default O
2
% settings are as follows:
• O
2
sensor enabled
• Neonatal: 40% O
2
• Pediatric and adult: 100% O
2
Note: The oxygen sensor can possess three states: Enabled, Disabled, and Calibrate. The oxygen sensor is enabled at ventilator startup regardless if New Patient or Same Patient setup is selected.
To enable, or disable the O
2
sensor
1. Touch the vent setup button.
2. Touch the More Settings tab. The more settings screen appears.
112
Figure 47.
More Settings Screen with O
2
Sensor Enabled
4
3. Touch the button corresponding to the desired O
2
sensor function (Enable or Disable).
4. Touch Accept .
4.10.1 Oxygen Sensor Life
The O
2
% setting can range from room air (21% O
2
) up to a maximum of 100% oxygen. The sensor reacts with oxygen to produce a voltage proportional to the partial pressure of the mixed gas. Since ambient atmosphere contains approximately 21% oxygen, the sensor constantly reacts with oxygen and always produces a voltage.
The useful life of the cell can also be shortened by exposure to elevated temperatures and pressures. During normal use in the ICU, the oxygen sensor lasts for approximately 1 year — the interval for routine preventive maintenance.
Because the oxygen sensor constantly reacts with oxygen, it requires periodic calibration to prevent inaccurate
O
2
% alarm annunciation. Once a calibrated oxygen sensor and the ventilator reach a steady-state operating temperature, the monitored O
2
% will be within three percentage points of the actual value for at least 24 hours.
To ensure the oxygen sensor remains calibrated, recalibrate the oxygen sensor at least once every 24 hours.
Typically, the clinician uses an O
2
analyzer in conjunction with the information given by the ventilator. If a NO O
SUPPLY alarm occurs, compare the O
2
analyzer reading with the ventilator’s O
2
reading for troubleshooting purposes. The ventilator automatically switches to 100% air delivery.
2
4.10.2 Oxygen Sensor Calibration
The oxygen sensor should be calibrated every 24 hours and before use. The calibration function provides a single-point O
2
sensor calibration.
To calibrate the O
2
sensor
1. Touch the vent setup button.
4
2. Touch the More Settings tab.
113
3. Touch Calibrate for the O
2
sensor. The calibration procedure results in 100% O
2
being delivered through the breathing circuit for the 2-minute calibration period. See
4.10.3 Oxygen Sensor Calibration Testing
To test the O
2
sensor calibration
1. Connect the ventilator’s oxygen hose to a known 100% O
2 cylinder).
source (for example, a medical-grade oxygen
2. Calibrate the oxygen sensor as described above.
3. Connect the ventilator oxygen hose to another known 100% O
2 medical-grade oxygen cylinder).
source (for example, a second
4. Set O
2
% to each of the following values, and allow 1 minute after each for the monitored value to stabilize:
21%, 40%, 90%
5. Watch the GUI screen to ensure the value for O
2 of selecting each setting.
(delivered O
2
%) is within 3% of each setting within 1 minute
4.11 Ventilator Protection Strategies
The ventilator incorporates a number of strategies to support patient safety. These include power on self test
(POST), SST and a strategy called Ventilation Assurance which provides alternate means of ventilation in the case of certain serious faults in the breath delivery system. The descriptions detail the system response to potential failures.
4.11.1 Power on Self Test (POST)
The first strategy is to detect potential problems before the ventilator is placed on a patient. POST checks the integrity of the ventilator’s electronics and prevents ventilation if a critical fault is found. See the Puritan Bennett™
980 Series Ventilator Service Manual for a complete description of POST). POST may detect major or minor system faults which manifest themselves as DEVICE ALERTS. See
Section 6.5.13, DEVICE ALERT Alarm, page 157
for more information.
4.11.2 Technical Fault
modified settings, or enter the vent inop state. A technical fault cannot be cleared by pressing the alarm reset key.
It can only be cleared by correcting the fault that caused it or if alarm reset criteria have been met.
4.11.3 SST
In addition to characterizing the ventilator breathing circuit, SST performs basic checks on the ventilator’s pneumatic system including the breath delivery PSOL, the Flow Sensors and the Exhalation Valve. Faults detected during SST must be corrected before ventilation can be started.
4.11.4 Procedure Error
A procedure error occurs when the ventilator senses a patient connection before ventilator setup is complete. The ventilator provides ventilatory support using default safety pressure controlled ventilation (safety PCV) settings.
See
114
4
4.11.5 Ventilation Assurance
During ventilation, the ventilator performs frequent background checks of its breath delivery subsystem (See
Assurance provides for continued ventilatory support using one of three backup ventilation (BUV) strategies, bypassing the fault to maintain the highest degree of ventilation that can be safely delivered (See
Background Diagnostic System, page 237
for a full description of the backup ventilation strategies).
Note: Do not confuse BUV with safety PCV, which occurs when a patient is connected before ventilator setup is complete, or with apnea ventilation, which occurs in response to patient apnea.
4.11.6 Safety Valve Open (SVO)
In the event of a serious fault occurring that cannot be safely bypassed, the ventilator, as a last resort, reverts to a safe state. In Safe State, the ventilator opens the safety valve and the exhalation valve, allowing the patient to breathe room air (if able to do so), provided the patient circuit is not occluded. During SVO, the patient (if connected) can breathe room air through the safety valve after it releases pressure in the patient circuit. The patient exhales through the exhalation valve with minimal resistance and the exhalation valve also acts like a check valve, limiting gas from being drawn in through the expiratory filter or expiratory limb of the circuit. SVO conditions are logged into the event and alarm logs as are the events leading to the SVO condition. If the condition causing SVO clears, the ventilator clears the SVO state. Patient data do not display on the GUI, but graphics are still plotted. During SVO, the ventilator ignores circuit occlusions and disconnects. If the condition causing SVO can only be corrected by servicing the ventilator, the SVO alarm cannot be reset by pressing the alarm reset key.
4.11.7 Ventilator Inoperative (Vent Inop)
Vent Inop occurs when the ventilator detects a catastrophic error which prevents all other safety states from operating. Vent Inop limits pressure to the patient as the ventilator enters the SVO state, disables (closes) the inspiratory valves (PSOLs), and purges the gas mixing system accumulator. The safety valve is opened and a Vent
Inop indicator illuminates and a high priority alarm annunciates from the primary alarm, and the secondary alarm
(continuous tone) is activated. The ventilator can only exit the vent inop state by power cycling and successfully passing EST. The Vent Inop alarm cannot be reset with the alarm reset key. All detection and annunciation of patient data alarm conditions is suspended.
4.12 Ventilator Shutdown
When the ventilator power switch is turned off, the ventilator executes an orderly shutdown routine, saving patient data before removing power. If the ventilator detects a patient connected when the power switch is turned off, a high priority alarm is annunciated and a banner on the display requires the operator to confirm that a power down was requested. Only after the operator confirms will the ventilator execute the shutdown command.
All logs are retained in the ventilator’s memory upon ventilator shutdown. When the logs reach the maximum
information on ventilator logs.
4
115
116
5
5 Product Data Output
5.1 Overview
This chapter describes the features of the Puritan Bennett™ 980 Series Ventilator designed to provide output to the clinician. This includes language, methods of displaying and transferring data, types of displayed data, and types of external device ports. Connectivity to an external patient monitoring system is also included.
5.2 Language
The language used on the ventilator is configured at the factory.
5.3 Data Display
Displayed data are updated in real-time. The practitioner can display up to 60 seconds of waveform data and pause and capture up to two loops using the screen capture function. The operator can pause the displays and when the displays are paused, a cursor appears with the relevant numeric values for the intersecting points of the cursor and waveform or loop. The scalar waveform contains a single value, but loops contain both x- and y-axis data. The operator can move the cursor along the waveform or loop using the knob, and read the corresponding
5.4 Data Transfer
Data from the ventilator can be accessed via USB or RS-232 connectors. The following data are available for downloading via connection to a remote device or flash drive:
• Waveform images (screen capture function): USB port
• Waveform data: RS-232 port, USB port with USB to serial conversion capability (per Comm port configuration)
• Results from DCI commands: RS-232 port, USB port with USB to serial conversion capability (per Comm port configuration)
5.4.1 GUI Screen Capture
Caution: The USB interface should be used for saving screen captures and interfacing with an external patient monitor. It should not provide power to other types of devices containing a USB interface.
Caution: Only compatible USB devices should be used, otherwise GUI performance may be impacted.
A 128 MB flash drive storage device formatted in the 32-bit file format is required for downloading images from the USB ports. The USB device listed in
only compatible USB device currently available for use on the ventilator. To order a compatible USB device, contact Medtronic Technical Services at 800 255 6774 or a local
Medtronic representative.
To capture GUI screens
1. Navigate to the desired screen from which you wish to capture an image (for example, the waveforms screen). There is no need to pause the waveform before performing the screen capture.
2. Touch the screen capture icon in the constant access icons area of the GUI screen.
5
If desired, navigate to another screen and repeat steps 1 and 2 for up to 10 images. If another image is captured, increasing the queue to 11 images, the newest image overwrites the oldest image so there are always only 10 images available.
117
Note: If the camera icon appears dim, it means that the screen capture function is currently processing images, and is unavailable. When processing is finished, the camera icon is no longer dim and the screen capture function is available.
To transfer captured images to a USB storage device
1. Swipe the Menu on the left side of the GUI. See
2. Touch Screen Capture. A list of screen captures appears, identified by time and date. A slider also appears indicating more images than shown are present.
3. Insert a passive USB storage device (flash drive) into one of the USB ports at the rear of the ventilator. The
.
If more than one USB storage device is installed in the ventilator, touch the button of the destination USB device where the image will be copied. If an incompatible device is inserted, the port becomes disabled until the device is removed, and removal is confirmed by touching the confirm button. The message shown
Figure 48.
Incompatible USB Device Message
Note: Removal of the external USB storage device while screenshot files are being written to it may result in incomplete file transfer and unusable files.
4. In the list of images, touch the image name.
5. Touch Copy. The image is stored on the destination USB storage device.
6. Alternatively, touch Select All, and all images in the list are stored on the USB device and which can then be viewed and printed from a personal computer.
Note: The file format of screen captures is .PNG.
5.4.2 Communication Setup
To specify the communication configuration for the ventilator
1. Touch the Configure icon in the constant access icons area of the GUI.
118
A menu appears with several tabs.
2. Touch the Comm Setup tab. The Comm Setup screen appears allowing three ports to be configured. These ports can be designated as DCI, DCI 2.0, DCI 3.0 Philips, Spacelabs, or Waveforms.
Figure 49.
Comm Setup Screen
5
Note: Waveforms can be selected on any port, but only on one port at a time.
5.4.3 Comm Port Configuration
Configuring the Comm port allows the ventilator to communicate with devices listed in the Comm Setup screen, or to capture waveform data (in ASCII format) from the ventilator.
To configure Comm ports
1. Touch COM1, COM2, or COM3.
2. Turn the knob indicating the desired device configuration.
3. Select the desired baud rate. If waveforms was selected, the baud rate automatically becomes configured to 38400.
4. Select 7 or 8 data bits.
5. Select parity of even, odd, or none if data bits = 8.
Connect the device to the previously configured port. See
Figure 50, page 129 for a description and the locations
of the Comm ports.
Note: When a USB port is configured as a Comm port, it is necessary to use a USB-to-serial adapter cable. This adapter must be based on the chipset manufactured by Prolific. For further information, contact your Medtronic representative.
Selecting waveforms when configuring a Comm port allows the ventilator to continuously transmit pressure, flow, and sequence numbers in ASCII format from the selected serial port, at a baud rate of 38400 bits/s, and the operator- selected stop bits, and parity. A sample of pressure and flow readings is taken every 20 ms. This sample of readings is transmitted on the selected serial port at the end of each breath at breath rates of 10/min and higher.
For longer duration breaths, at least the first 8 seconds of the breath is transmitted. The format of the data is as follows: the beginning of inspiration is indicated by “BS, S:nnn,<LF>“ where ’BS’ identifies the Breath Start, ‘S:nnn’ is a sequence number incremented at every breath, and <LF> is a line feed character. The fff, and ppp fields show
119
5
the breath flow and pressure data. The end of exhalation is indicated by: “BE<LF>“ where ‘BE’ indicates Breath End, and <LF> is a line feed character.
5.4.4 Serial Commands
The ventilator system offers commands that allow communication to and from the ventilator using a Comm port.
Commands to the ventilator from a remote device include:
• RSET: see
Section 5.4.5, RSET Command, page 120 .
• SNDA: see Section 5.4.6, SNDA Command, page 120 .
• SNDF: see Section 5.4.7, SNDF Command, page 123 .
Note: The ventilator responds only if it receives a carriage return <CR> after the command string.
5.4.5 RSET Command
The RSET command clears data from the ventilator receive buffer. The ventilator does not send a response to the host system. Enter the RSET command exactly as shown:
RSET<CR>
5.4.6 SNDA Command
The SNDA command instructs the ventilator to send information on ventilator settings and monitored patient data to the host system. Enter the SNDA command exactly as shown:
SNDA<CR>
When the ventilator receives the command SNDA<CR>, it responds with the code MISCA, followed by ventilator settings and monitored patient data information.
The MISCA response follows this format:
MISCA
1
706
2
97
3
<STX>
4
FIELD 5, … FIELD 101,
5
<ETX>
6
<CR>
7
1 Response code to SNDA command
2 Number of bytes between <STX> and <CR>
3 Number of data fields between <STX> and
<ETX>
4 Start of transmission (02 hex)
5 Data field, left-justified and padded with spaces
6 End of transmission (03 hex)
7 Terminating carriage return
Fields not available are marked as “Not used.” Underscores represent one or more spaces that pad each character string.
Table 30 lists MISCA responses to SNDA commands.
Table 30.
MISCA Response
Component
MISCA
706
97
<STX>
Field 5
Field 6
Description
Response to SNDA command (5 characters)
The number of bytes between <STX> and <CR> (3 characters)
The number of fields between <STX> and <ETX> (2 characters)
Start of transmission character (02 hex)
Ventilator time (HH:MM_) (6 characters)
Ventilator ID to allow external hosts to uniquely identify each Puritan Bennett™ 980 Ven‐ tilator (18 characters)
120
5
Table 30.
MISCA Response (continued)
Field 22
Field 23
Field 24
Field 25
Field 26
Field 27
Field 28-29
Field 30
Field 31-33
Field 34
Field 35
Field 36
Field 37
Field 38
Field 39
Field 40
Field 41
Component
Field 7
Field 8
Field 9
Field 10
Field 11
Field 12
Field 13
Field 14
Field 15
Field 16
Field 17-20
Field 21
Field 42
Field 43-44
Field 45
Field 46
Field 47
Field 48
Field 49-50
Field 51
Field 52
Field 53
Description
Room number (6 characters)
Date (MMM_DD_YYYY_) (12 characters)
Mode (CMV___, SIMV__, CPAP__ or BILEVL) (CMV = A/C) setting (6 characters)
Respiratory rate setting in breaths per minute (6 characters)
Tidal volume setting in liters (6 characters)
Peak flow setting in liters per minute (6 characters)
O
2
% setting (6 characters)
Pressure sensitivity setting in cmH
2
O (6 characters)
PEEP or P
L
(in BiLevel) setting in cmH
2
O (6 characters)
Plateau time in seconds (6 characters)
Not used (6 characters)
Apnea interval in seconds (6 characters)
Apnea tidal volume setting in liters (6 characters)
Apnea respiratory rate setting in breaths per minute (6 characters)
Apnea peak flow setting in liters per minute (6 characters)
Apnea O
2
% setting (6 characters)
Pressure support setting in cmH
2
O (6 characters)
Flow pattern setting (square or ramp__) (6 characters)
Not used (6 characters)
Elevate O
2
state (on____ or off___) (6 characters)
Not used (6 characters)
Total respiratory rate in breaths per minute (6 characters)
Exhaled tidal volume in liters (6 characters)
Exhaled minute volume in liters (6 characters)
Spontaneous minute volume in liters (6 characters)
Maximum circuit pressure in cmH
2
O (6 characters)
Mean airway pressure in cmH
2
O (6 characters)
End inspiratory pressure in cmH
2
O (6 characters)
Expiratory component of monitored value of I:E ratio, assuming inspiratory component of 1 (6 characters)
High circuit pressure limit in cmH
2
O (6 characters)
Not used (6 characters)
Low exhaled tidal volume limit in liters (6 characters)
Low exhaled minute volume limit in liters (6 characters)
High respiratory rate limit in breaths per minute (6 characters)
High circuit pressure alarm status (NORMAL, ALARM_, or RESET_) (6 characters)
Not used (6 characters)
Low exhaled tidal volume (mandatory or spontaneous) alarm status (NORMAL, ALARM_, or RESET_) (6 characters)
Low exhaled minute volume alarm status (NORMAL, ALARM_, or RESET_) (6 characters)
High respiratory rate alarm status (NORMAL, ALARM_, or RESET_) (6 characters)
121
5
Field 86
Field 87
Field 88
Field 89
Field 90
Field 91
Field 92
Field 93
Field 67
Field 68
Field 69
Field 70
Field 71
Field 72-83
Field 84
Field 85
Field 94
Field 95
Field 96
Component
Field 54
Field 55
Field 56
Field 57
Field 58-59
Field 60
Field 61
Field 62
Field 63
Field 64
Field 65
Field 66
Table 30.
MISCA Response (continued)
Field 97
Field 98
Field 99
Field 100
Field 101
<ETX>
<CR>
Description
No O
2
supply alarm status (NORMAL, ALARM_, or RESET_) (6 characters)
No air supply alarm status (NORMAL, ALARM_, or RESET_) (6 characters)
Not used (6 characters)
Apnea alarm status (NORMAL, ALARM_, or RESET_) (6 characters)
Not used (6 characters)
Ventilator time (HH:MM_) (6 characters)
Not used (6 characters)
Date (MMM_DD_YYYY_) (12 characters)
Static compliance (C
STAT
) from inspiratory pause maneuver in mL/cmH
2
O (6 characters)
Static resistance (R
STAT
) from inspiratory pause maneuver in cmH
2
O/L/s (6 characters)
Dynamic compliance (C
DYN
) in mL/cmH
2
O (6 characters)
Dynamic resistance (R
DYN
) in cmH
2
O/L/s (6 characters)
Negative inspiratory force (NIF) in cmH
2
O (6 characters)
Vital capacity (VC) in L (6 characters)
Peak spontaneous flow (PSF) in L/min (6 characters)
Ventilator-set base flow in liters per minute (6 characters)
Flow sensitivity setting in L/min (6 characters)
Not used (6 characters)
End inspiratory pressure in cmH
2
O (6 characters)
Inspiratory pressure or P
H
setting in cmH
2
O (6 characters)
Inspiratory time or T
H
setting in seconds (6 characters)
Apnea interval setting in seconds (6 characters)
Apnea inspiratory pressure setting in cmH
2
O (6 characters)
Apnea respiratory rate setting in breaths per minute (6 characters)
Apnea inspiratory time setting in seconds (6 characters)
Apnea O
2
% setting (6 characters)
Apnea high circuit pressure limit in cmH
2
O (6 characters)
Audio paused state (ON____ or OFF___) (6 characters)
Apnea alarm status (NORMAL, ALARM_ or RESET_) (6 characters)
Severe Occlusion/Disconnect alarm status (NORMAL, ALARM_ or RESET_) (6 characters)
Inspiratory component of I:E ratio or High component of H:L (BiLevel) setting (6 charac‐ ters)
Expiratory component of I:E ratio setting or Low component of H:L (BiLevel) (6 characters)
Inspiratory component of apnea I:E ratio setting (6 characters)
Expiratory component of apnea I:E ratio setting (6 characters)
Constant during rate setting change for pressure control mandatory breaths (I-time or
I/E___ or______) (6 characters) (where ______ represents T
E
or PCV not active)
Monitored value of I:E ratio (6 characters)
End of transmission character (03 hex)
Terminating carriage return
122
5
5.4.7 SNDF Command
SNDF is a command sent from an external host device to the ventilator system instructing it to transmit all ventilator settings data, monitored patient data, and alarm settings and occurrences. Enter the SNDF command exactly as shown:
SNDF<CR>
When the ventilator receives the command SNDF<CR>, it responds with the code MISCF, followed by ventilator settings, monitored patient data, and alarm information
The MISCF response follows this format:
MISCF
1
1225*
2
169
3
<STX>
4
FIELD 5, … FIELD
173,
5
<ETX>
6
<CR>
7
1 Response code to SNDF command 5 Data field, left-justified and padded with spaces
6 End of transmission (03 hex) 2 Number of bytes between <STX> and
<CR>
3 Number of data fields between <STX> and
<ETX>
4 Start of transmission (02 hex)
7 Terminating carriage return
* 1229 if Philips is selected for serial port in communication setup
Table 31 lists MISCF responses to SNDF commands.
Note: Non-applicable fields will either contain zero or be blank.
Note: Ensure your external devices are compatible with the latest DCI software to prevent incompatibilities as data fields may have been modified.
Table 31.
MISCF Response
Component
MISCF
1225*
169
<STX>
Field 5
Field 6
Field 7
Field 8
Field 9
Field 10
Field 11
Field 12
Field 13
Field 14
Field 15
Response to SNDF command (5 characters)
Number of bytes between <STX> and <CR> (4 characters) *1229 if Phillips is selected for the
Comm port in Communication Setup
Description
Number of fields between <STX> and <ETX> (3 characters)
Start of transmission character (02 hex)
Ventilator time (HH:MM_) (6 characters)
Ventilator ID to allow external hosts to uniquely identify each Puritan Bennett™ 980 Ventilator
(18 characters)
Date (MMM_DD_YYYY_) (12 characters)
Ventilation Type (NIV______ or INVASIVE_ or HFO
2
T____) (9 characters)
Mode (A/C___, SIMV__, SPONT_ or CPAP__) (6 characters)
Mandatory Type (PC____, VC____, VC+___) (6 characters)
Spontaneous Type (PS____, TC____, VS____, PA____) (6 characters)
Trigger Type setting ( ⩒ -Trig, P-Trig, IE
SYNC
) (6 characters)
Respiratory rate setting in breaths/min (6 characters)
Tidal volume (V
T
Peak flow ( ⩒
MAX
) setting in L (6 characters)
) setting in L/min (6 characters)
123
5
Field 28
Field 29
Field 30
Field 31
Field 32
Field 33
Field 34
Field 35
Field 36
Component
Field 16
Field 17
Field 18
Field 19
Field 20
Field 21
Field 22
Field 23
Field 24
Field 25
Field 26
Field 27
Field 37
Field 38
Field 39
Field 40
Field 41
Field 42
Field 43
Field 44
Field 45
Field 46
Field 47
Field 48
Field 49
Field 50
Field 51
Field 52
Field 53
Field 54
Field 55
Field 56
Table 31.
MISCF Response (continued)
Description
O
2
% setting (6 characters)
Pressure sensitivity setting in cmH
2
O (6 characters)
PEEP/CPAP setting in cmH
2
O (6 characters)
Plateau setting in seconds (6 characters)
Apnea interval setting in seconds (6 characters)
Apnea tidal volume setting in L (6 characters)
Apnea respiratory rate setting in breaths/min (6 characters)
Apnea peak flow setting in L/min (6 characters)
Apnea O
2
% setting (6 characters)
PCV apnea inspiratory pressure setting in cmH
2
O (6 characters)
PCV Apnea Inspiratory Time setting in seconds (6 characters)
Apnea flow pattern setting (square or ramp) (6 characters)
Apnea mandatory type setting (PC or VC) (6 characters)
Inspiratory component of Apnea I:E ratio (if apnea mandatory type is PC) (6 characters)
Expiratory component of Apnea I:E ratio (if apnea mandatory type is PC) (6 characters)
Support pressure setting (cmH
2
O)
Flow pattern setting (square or ramp) (6 characters)
Elevate O
2
state (ON or OFF) (6 characters)
High inspiratory pressure alarm setting ( ⤒ P
PEAK
) in cmH
2
O (6 characters)
Low inspiratory pressure alarm setting ( ⤓ P
PEAK
) in cmH
2
O or OFF (6 characters)
High exhaled minute volume ( ⤒ V
E TOT
) alarm setting in L/min or OFF (6 characters)
Low exhaled minute volume ( ⤓ V
E TOT
) alarm setting in L/min or OFF (6 characters)
High exhaled mandatory tidal volume ( ⤒ V
TE MAND
) alarm setting in mL or OFF (6 characters)
Low exhaled mandatory tidal volume ( ⤓ V
TE MAND
) alarm setting in mL or OFF (6 characters)
High exhaled spontaneous tidal volume ( ⤒ V
TE SPONT
) alarm setting in mL or OFF (6 characters)
Low exhaled spontaneous tidal volume ( ⤓ V
TE SPONT
) alarm setting in mL or OFF (6 characters)
High respiratory rate ( ⤒ f
TOT
) alarm setting in breaths/min or OFF (6 characters)
High inspired tidal volume ( ⤒ V
TI
) alarm setting in mL (6 characters)
Base flow setting in L/min (6 characters)
Flow sensitivity ( ⩒
SENS
) setting in L/min (6 characters)
PCV inspiratory pressure (P
I
) setting in cmH
2
O (6 characters)
PCV inspiratory time (T
I
) setting in seconds (6 characters)
Inspiratory component of I:E ratio setting or High component of H:L ratio setting (6 characters)
Expiratory component of I:E ratio setting or Low component of H:L ratio setting (6 characters)
Constant during rate change setting (I-time, I/E, or E-time) (6 characters)
Tube ID setting in mm (6 characters)
Tube Type setting (ET or TRACH) (6 characters)
Humidification type setting (Non-heated exp tube, Heated exp tube, or HME) (18 characters)
Humidifier volume setting in L (6 characters)
O
2 sensor setting (Enabled or Disabled) (9 characters)
Disconnect sensitivity (D
SENS
) setting in %, L/min or OFF (6 characters)
124
5
Component
Field 57
Field 58
Field 59
Field 60
Field 61
Field 62
Field 63
Field 64
Field 65
Field 66
Field 67
Field 68
Field 69
Field 70
Field 71
Field 72
Field 73
Field 74
Field 75
Field 79
Field 80
Field 81
Field 82
Field 83
Field 84
Field 85
Field 86
Field 87
Field 88
Field 89
Field 90
Field 91
Field 92
Field 93
Field 94
Field 95
Table 31.
MISCF Response (continued)
Field 76
Field 77
Field 78
Description
Rise time% setting (6 characters)
PAV+™ percent support setting (6 characters)
Expiratory sensitivity (E
SENS
) setting in % or L/min for PAV+™ breath type (6 characters)
PBW setting in kg (6 characters)
Target support volume (V
T SUPP
) setting in L (6 characters)
High pressure (P
H
) setting (in BiLevel) in cmH
2
O (6 characters)
Low pressure (P
L
) setting (in BiLevel) in cmH
2
O (6 characters)
High pressure time (T
H
) setting (in BiLevel) in seconds (6 characters)
High spontaneous inspiratory time limit ( ⤒ T
I SPONT
) setting in seconds (6 characters)
Circuit type setting (ADULT or PEDIATRIC) (9 characters)
Low pressure time (T
L
) setting (in BiLevel) in seconds (6 characters)
Expiratory time (T
E
) setting in seconds (6 characters)
End inspiratory pressure (P
I END
) in cmH
2
O (6 characters)
Respiratory rate (f
TOT
) in breaths/min (6 characters)
Exhaled tidal volume (V
TE
) in L (6 characters)
Patient exhaled minute volume ( ⩒
E TOT
) in L/min (6 characters)
Peak airway pressure (P
PEAK
) in cmH
2
O (6 characters)
Mean airway pressure (P
MEAN
) in cmH
2
O (6 characters)
Expiratory component of monitored value of I:E ratio, assuming inspiratory component of 1 (6 characters)
I:E ratio (6 characters)
Delivered O
2
% (6 characters)
Monitored inspired tidal volume (V
TI
) in L # (6 characters)
# VTL (L), If Leak Sync ON.
Intrinsic PEEP (PEEP
I
) in cmH
2
O (6 characters)
Estimated total resistance (R
TOT
) in cmH
2
O/L/s (6 characters)
Estimated patient resistance (R
PAV
) in cmH
2
O/L/s (6 characters)
Estimated patient elastance (E
PAV
) in cmH
2
O/L (6 characters)
Estimated patient compliance (C
PAV
) in mL/cmH
2
O (6 characters)
Not used
Rapid shallow breathing index (f/V
T
) (6 characters)
Spontaneous percent inspiratory time (T
I
/T
TOT
) (6 characters)
Monitored Positive end expiratory pressure (PEEP) in cmH
2
O (6 characters)
Spontaneous inspiratory time (T
I SPONT
) in seconds (6 characters)
Exhaled spontaneous minute volume ( ⩒
E SPONT
) in L/min (6 characters)
Intrinsic PEEP (PEEPI) from expiratory pause maneuver in cmH
2
O (6 characters)
Monitored total PEEP (PEEP
TOT
) from expiratory pause maneuver in cmH
2
O (6 characters)
Static compliance (C
STAT
) from inspiratory pause maneuver in mL/cmH
2
O (6 characters)
Static resistance (R
STAT
) from inspiratory pause maneuver in cmH
2
O/L/s (6 characters)
Plateau pressure (P
PL
) from inspiratory pause maneuver in cmH
2
O (6 characters)
High spontaneous inspiratory time (ALERT_ or blank) (6 characters)
125
5
Table 31.
MISCF Response (continued)
Component
Field 96
Field 97
Field 98
Field 99
Field 100
Field 101
Field 102
Field 103
Field 104
Field 105
Field 106
Field 107
Field 108
Field 109
Field 110
Field 111
Field 112
Field 113
Field 114
Field 115
Field 116
Field 117
Field 118
Field 119
Field 120
Field 121
Field 122
Field 123
Field 124
Field 125
Field 126
Field 127
Field 128
Field 129
Field 130
Field 131
Field 132
Field 133
Field 134
Field 135
Dynamic compliance (C
DYN
) in mL/cmH
Dynamic resistance (R
DYN
) in cmH
2
O/L/s (6 characters)
Peak spontaneous flow (PSF) in L/min (6 characters)
Peak expiratory flow (PEF) in L/min (6 characters)
End expiratory flow (EEF) in L/min (6 characters)
Proximal Flow Sensor state (ON or OFF) (6 characters)
Vital capacity (VC) in L (6 characters)
Description
2
O (6 characters)
Negative inspiratory force (NIF) in cmH
2
O (6 characters)
P
0.1
pressure change in cmH
2
O (6 characters)
Audio paused (ON or OFF) (6 characters)
Apnea ventilation alarm* (6 characters)
High exhaled minute volume alarm* ( ↑ ⩒
E TOT
) (6 characters)
High exhaled tidal volume alarm* ( ↑ V
TE
) (6 characters)
High O
2
% alarm* (6 characters)
High inspiratory pressure alarm* ( ↑ P
PEAK
) (6 characters)
High ventilator pressure alarm* ( ↑ P
VENT
) (6 characters)
High respiratory rate alarm* ( ↑ f
TOT
) (6 characters)
AC power loss alarm* (6 characters)
Inoperative battery alarm* (6 characters)
Low battery alarm* (6 characters)
Inadvertent Power Off alarm* (6 characters)
Low exhaled mandatory tidal volume alarm* ( ↓ V
TE MAND
) (6 characters)
Low exhaled minute volume alarm* ( ↓ ⩒
E TOT
) (6 characters)
Low exhaled spontaneous tidal volume ( ↓ V
TE SPONT
) alarm* (6 characters)
Low O
2
% alarm* (6 characters)
Low air supply pressure alarm* (6 characters)
Low O
2
supply pressure alarm* (6 characters)
Compressor inoperative alarm* (6 characters)
Disconnect alarm* (6 characters)
Severe occlusion alarm* (6 characters)
Inspiration too long alarm* (6 characters)
Procedure error alarm* (6 characters)
Compliance limited tidal volume (V
T
) alarm* (6 characters)
I
High inspired spontaneous tidal volume* ( ↓ T
I SPONT
) alarm (6 characters)
Reserved
High compensation limit ( ↑ P
COMP
) alarm* (6 characters)
PAV+™ startup too long alarm* (6 characters)
PAV+™ R and C not assessed alarm* (6 characters)
Volume not delivered (VC+) alarm* (6 characters)
Volume not delivered (VS) alarm* (6 characters)
126
5
Field 165
Field 166
Field 167
Field 168
Field 169
Field 170
Field 171
Field 172
Field 173
<ETX>
Field 137
Field 138
Field 139
Field 140
Field 141
Field 142
Field 143
Field 144
Field 145
Field 146
Field 147
Field 148
Field 149
Field 150
Field 151
Field 152
Field 153
Field 154
Field 155
Field 156
Field 157
Field 158
Field 159
Field 160
Field 161
Field 162
Field 163
Field 164
Table 31.
MISCF Response (continued)
Component
Field 136 Low inspiratory pressure ( ↓ P
PEAK
) alarm* (6 characters)
Technical malfunction A5* (6 characters)
Description
Technical malfunction A10* (6 characters)
Technical malfunction A15* (6 characters)
Technical malfunction A20* (6 characters)
Technical malfunction A25* (6 characters)
Technical malfunction A30* (6 characters)
Technical malfunction A35* (6 characters)
Technical malfunction A40* (6 characters)
Technical malfunction A45* (6 characters)
Technical malfunction A50* (6 characters)
Technical malfunction A55* (6 characters)
Technical malfunction A60* (6 characters)
Technical malfunction A65* (6 characters)
Technical malfunction A70* (6 characters)
Technical malfunction A75* (6 characters)
Technical malfunction A80* (6 characters)
High ETCO
2
Alarm* (6 characters)
Spontaneous tidal volume (V
TE SPONT
) in liters (6 characters)
Total work of breathing (WOB
TOT
) in Joules/L (6 characters)
Leak Sync state (9 characters) (ON, or OFF)
%LEAK (6 characters)
LEAK in L/min (6 characters)
VLEAK in mL (6 characters)
Prox Inop alarm* (ALARM or NORMAL)
ETCO
2
(mmHg) when COM port is set to DCI 2.0 or DCI 3.0 (6 characters) Otherwise Blank
Inspiratory Compliance ratio (C20/C) (6 characters)
Three times Inspiratory Time Constant (3Tau
I
) in seconds (6 characters)
#
Delivered mL/kg Volume (VTI/PBW) # .
VTL/PBW (mL/kg), If Leak Sync ON. (6 characters)
Monitored Driving Pressure (P
DRIVE
) in cmH
2
O (6 characters)
Monitored Positive End Expiratory Pressure at Patient Interface (PEEP
IF
) in cmH
2
O (6 characters)
Monitored End Inspiratory Pressure at Patient Interface (P
Monitored Constant Flow ( ⩒
CONST
) in L/min (6 characters)
I END IF
) in cmH
2
O (6 characters)
IE Sync Trigger Sensitivity (I
SYNC
) setting (6 characters)
IE Sync Cycle Sensitivity (E
SYNC
) setting (6 characters)
Constant flow ( ⩒
CONST
) setting in L/min (6 characters)
Ventilator State (6 characters) f
Nebulizer State (6 characters)
End of transmission character (03 hex)
5
127
Table 31.
MISCF Response (continued)
Component
<CR> Terminating carriage return
Description
* Possible responses are: NORMAL, LOW, MEDIUM, HIGH, or RESET.
a Fields 146 to 149 are blank for DCI 3.0
b Nebulizer Inoperative Alarm*, when COM port is set to DCI 3.0
c e
CO2 Monitor Inoperative Alarm* when COM port is set to DCI 3.0
d Low EtCO2 Alarm*, when COM port is set to DCI 3.0
Fields 162 to 173 are configured for DCI 3.0, otherwise Blank.
f Possible Ventilator State responses are: STNDBY (Stand-By Mode), SAFPCV (Safety PCV), BREATH (Normal Breathing Mode),
DISCON (Circuit Disconnect), OCCLUD (Occlusion), SVOPEN (Safety Valve Open), BUV (Backup Ventilation), MIX-BUV (Mixer
Backup Ventilation), HFO2T, APNEA.
g Possible Nebulizer State responses are: ON, OFF, SUSPND (Suspended).
5.5 Communication Ports
Warning: The ventilator does not confirm the receipt of the signal to distributed remote devices, including those with alarms. Therefore, ensure alarm volume is appropriately set to avoid missed alarms.
Warning: To avoid possible injury, only connect devices that comply with IEC 60601-1 standard to any of the ports at the rear of the ventilator, with the exception of passive memory storage devices (“flash drives”) and serial-to-USB adapter cables. If a serial-to-USB adapter cable is used, it must be connected to an IEC 60601-1-compliant device.
Warning: To avoid possible injury, do not connect a device that is attached to the patient to any of the non-clinical ports listed in
Figure 50 when the ventilator is ventilating a patient.
128
Figure 50.
Port Locations
3
2
1
S
T
T
E
E
R
V
I
H
D
M
I
4
5
6
7
5
1 RS-232 Port (COM 1)
2 Ethernet Port (non-clinical use)
3 Nurse call Port (remote alarm port)
4 USB Port (USB 1) (COM 2) (non-clinical use)
VEN_11244_D
5 USB Port (USB 2) (COM 3) (non-clinical use)
6 HDMI Port (non-clinical use)
7 Service Port (non-clinical use)
5.5.1 Port Use
See
Section 5.4, Data Transfer, page 117
for data transfer details.
5.5.1.1 RS-232 Port
To use the RS-232 port
1. Obtain a cable with a male DB-9 connector to connect to the RS-232 port on the ventilator.
2. Make the appropriate connection to a monitoring device. A gender changer, null modem cable, or socket saver may be required. Consult with the institution’s Information Technology professional as required.
3. Ensure to specify the baud rate, parity, and data bits in the ventilator communication setup to correctly match the parameters of the monitoring device.
4. A monitor designed to use this port is required for obtaining data from the ventilator. Set up the monitoring device to receive ventilator data. These data can include waveform data.
5. Program the remote device to send the appropriate RS-232 commands as described in the next section.
See
Table 31, page 123 for MISCA and MISCF responses to SNDA and SNDF commands.
5
129
5.5.1.2 Ethernet Port
The Ethernet port is used by service personnel for accessing various logs and updating ventilator software.
5.5.1.3 Nurse Call Port
A remote alarm or nurse call interface is available on the ventilator system which can be used to remotely annunciate the alarm status of the ventilator. Medium and high priority alarms are remotely annunciated. The
nurse call connector is located at the back of the ventilator, as shown. See Figure 50, page 129 .
See the remote alarm manufacturer’s IFU for information regarding proper nurse call connection.
5.5.1.4 USB Ports
The USB ports are used for screen captures, or receiving serial data when a USB port has been configured as a serial port. This is also known as transferring data via a serial over USB protocol. See
Section 5.4.2, Communication Setup, page 118
for comm setup configuration. Screen captures require an external USB memory storage device (“flash drive”) for screen captures. Instructions for using this port for screen captures are given. See
5.5.1.5 HDMI Port
An external display can be used via connection with the HDMI port.
To use the HDMI port with an external display
1. Connect one end of an HDMI cable to the HDMI port at the back of the ventilator (item 6, above).
2. Connect the other end of the cable to the external display. An HDMI to DVI adapter may be used.
3. Turn the device on. The appearance of the GUI now displays on the external display device.
5.5.1.6 Service Port
The service port is used by service personnel only.
5.6 Retrieving Stored Data
Ventilator data are stored in various logs, accessible using the logs icon.
Some logs may be accessed during normal ventilation, and some are only available to Medtronic personnel when the ventilator is in Service mode. See
Section 8.5, Ventilator Logs, page 185
for more information on data stored in various logs.
5.7 Display Configurability
view which parameters are configurable and by whom.
See
Section 3.8.1, Preparing the Ventilator for Use, page 76
for information on configuring each display item.
5.8 Printing Data or Screen Captures
The ventilator cannot be connected directly to a printer.
for instructions on using the screen-capture feature.
5.9 Connectivity to External Systems
The ventilator is compatible with the Philips Medical IntelliVue MP50 and Spacelabs Ultraview patient monitoring systems.
130
Note: Not all patient monitors are compatible with the Puritan Bennett™ 980 Series Ventilator.
5
131
5
132
6
6 Performance
6.1 Overview
This chapter contains detailed information about Puritan Bennett™ 980 Series Ventilator performance including:
• Ventilator settings
• Alarm interpretation and alarm testing
• A detailed description of selected alarms
• Monitored patient data
6.2 System Options
Various software functions and options are available for the ventilator. Details for each of these functions and options are described in the appendices included in this manual. Information regarding the compressor hardware option is included in the included Compressor Operator’s Manual Addendum.
6.3 Environmental Considerations
Warning: Use of the ventilator/compressor in altitudes higher or barometric pressures lower than those specified
for a complete list of environmental specifications.
6.4 Ventilator Settings
Default ventilator settings are based on the circuit type selected during SST. A neonatal, pediatric or adult patient circuit can be used, and all accessories needed to ventilate the patient should be attached when SST is performed.
6.4.1 Ventilation Type
The clinician enters the ventilation type, specifying how the patient will be ventilated; invasively or non-invasively
(NIV). The ventilation type optimizes the alarm limits for NIV patients, and disables some settings for NIV.
6.4.2 Mode
Available ventilation modes are mandatory (A/C) or spontaneous (SPONT) modes, as well as two “mixed” modes:
SIMV and BiLevel.
A/C (Assist-Control) – A/C mode guarantees delivery of a minimum number of mandatory breaths based on the frequency (f ) set by the clinician. Breaths in A/C can be patient-initiated (PIM) or ventilator-initiated (VIM).
SPONT (Spontaneous) – SPONT mode delivers only spontaneous breaths which are all patient-initiated.
SIMV (Synchronized Intermittent Mandatory Ventilation) – SIMV is a mixed mode allowing both mandatory and spontaneous breaths. SIMV guarantees at least one mandatory breath per set breath cycle, which is either patient-initiated or ventilator-initiated. The mandatory type of an SIMV breath can be PC, VC
BiLevel – BiLevel is also a mixed mode which overlays the patient’s spontaneous breaths onto the breath structure for PC mandatory breaths. Two levels of pressure, P
L
and P
H
are employed. The breath cycle interval for both SIMV and BiLevel modes is 60/f where f is the respiratory rate set by the operator.
CPAP – CPAP is available only when circuit type is neonatal and ventilation type is NIV. CPAP mode allows spontaneous breathing with a desired PEEP level. In order to limit inadvertent alarms associated with the absence of returned volumes in nasal CPAP breathing, CPAP does not make available exhaled minute volume and exhaled tidal volume alarm settings.
6
133
6.4.3 Breath Type
Mandatory breath types for A/C and SIMV modes include volume controlled (VC), pressure controlled (PC), or volume control plus (VC+) breath types, also called mandatory type.
VC (Volume Control) – The ventilator delivers an operator-set tidal volume.
PC (Pressure Control) – The ventilator delivers an operator-set pressure.
VC+ (Volume Control Plus) – Volume control plus (a mandatory, pressure controlled breath type that does not restrict flow during the inspiratory phase, and automatically adjusts the inspiratory pressure target from breath to
for more information on VC+.
Mandatory inspirations are triggered in the following ways:
Pressure Trigger (P
TRIG
) – Changes in circuit pressure cause the ventilator to deliver a breath. These pressure changes relate to the pressure sensitivity (P
SENS
) set by the operator. If the patient makes an effort to inspire, the airway pressure drops. If the pressure drops by at least the value of P
SENS
, the ventilator delivers a breath.
Flow Trigger ( ⩒
TRIG
) — Changes in flow in the circuit cause the ventilator to deliver a breath. The breath delivery and exhalation flow sensors measure gas flow in the ventilator breathing system. As the patient inspires, the delivered flow remains constant, and the exhalation flow sensor measures decreased flow. When the difference between the two flow measurements is at least the operator-set value for flow sensitivity ( ⩒ delivers a breath.
SENS
), the ventilator
Time Trigger – The ventilator delivers a breath after a specific amount of time elapses.
Operator Trigger (OIM) – The operator presses the manual inspiration key. An operator initiated mandatory breath is also called an OIM breath. During an OIM breath, the breath delivered is based on the current settings for a mandatory breath.
Spontaneous breathing modes such as SIMV, BiLevel, and SPONT include the following breath types (called spontaneous types):
PS (Pressure Support) – The ventilator delivers an operator-set positive pressure above PEEP (or P
L
in BiLevel) during a spontaneous breath. If SIMV is selected as the mode, PS is automatically selected for spontaneous type.
VS (Volume Support) – The ventilator delivers an operator-set positive pressure above PEEP during a spontaneous breath and automatically adjusts the pressure level from breath to breath to consistently deliver the set tidal volume.
TC (Tube Compensation) – Additional positive pressure delivered to the patient during spontaneous breaths to overcome resistance of the artificial airway.
PAV+™ (Proportional Assist™ Ventilation) – A software function that allows the ventilator to reduce the work of breathing (WOB) by assisting the patient’s inspiration by an operator-set amount proportional to the pressure generated by the patient. See
Chapter 14 for more information on PAV+™.
The inspiratory trigger methods for spontaneous breaths are:
Pressure Trigger (P
TRIG
) – Same as described for mandatory inspiration triggers.
Flow Trigger ( ⩒
TRIG
) – Same as described for mandatory inspiration triggers.
Operator Trigger (OIM) – Since the operator can only initiate a mandatory breath by pressing the manual inspiration key, spontaneous mode allows OIMs, but the breath delivered is based on the current settings for an apnea breath.
See
134
6
6.5 Alarms
Warning: The ventilator system is not intended to be a comprehensive monitoring device and does not activate alarms for all types of conditions. For a detailed understanding of ventilator operations, be sure to thoroughly read this manual before attempting to use the ventilator system.
Warning: Setting any alarm limits to OFF or extreme high or low values can cause the associated alarm not to activate during ventilation, which reduces its efficacy for monitoring the patient and alerting the clinician to situations that may require intervention.
This manual uses the following conventions when discussing alarms:
A description or name of an alarm without specifying the alarm setting is denoted with an upward or downward pointing arrow ( ↑ or ⤒ ) preceding the specific alarm name. An alarm setting is denoted as an upward or downward pointing arrow with an additional horizontal limit symbol ( ↓ or ⤓ ) preceding the specific alarm. Some alarm conditions actually limit breath delivery such as ↑ P
PEAK
and ↑ V
TI
by truncating inspiration and transitioning to the exhalation phase. These alarm conditions are denoted as alarm limits . See
6.5.1 Alarm Messages
Alarms are visually annunciated using an indicator on the top of the GUI for an operator to view from 360 degrees around the ventilator with a radius of four meters (13.1 feet) centered on the visual alarm indicator. If an alarm occurs, this indicator flashes at a frequency and color matching the alarm priority. The alarms also appear as colored banners on the right side of the GUI screen, which can be viewed by the operator from a distance of four meters and at any point within the base of a cone subtended by an angle of 30° to the axis normal to the center of the plane of the monitoring display or visual indication. If an alarm occurs, this indicator appears in the color matching the alarm priority (yellow for low (!) and medium (!!) priority; red for high (!!!) priority). Specific alarm detail
(shown in the alarm message area) is designed for the operator to read from up to one meter from the screen. For
technical alarm and non-technical alarm details, see the respective tables: Table 35
and
An alarm is defined as a primary alarm if it is the initial alarm. A dependent alarm arises as a result of conditions that led to the primary alarm. This is also referred to as an augmentation. An augmentation strategy is built into the ventilator software to handle occurrences where the initial cause of the alarm has the potential to precipitate one or more additional alarms. When an alarm occurs, any subsequent alarm related to the cause of this initial alarm augments the initial alarm instead of appearing on the GUI as a new alarm. The initial alarm’s displayed analysis message is updated with the related alarm’s information, and the Alarm Log Event column shows the initial alarm as Augmented .
A primary alarm consists of a base message , analysis message , and a remedy message . The base message describes the primary alarm. The analysis message describes the likely cause of the alarm and may include alarm augmentations. The remedy message provides information on what to do to correct the alarm condition.
Alarm banners, when dragged leftward from the right side of the GUI, display messages for the indicated active alarms.
shows the alarm message format.
6
135
Figure 51.
Alarm Message Format
1 Base message
2 Analysis message
3 Remedy message
A latched alarm is one whose visual alarm indicator remains illuminated even if the alarm condition has autoreset.
Latched alarm indicators are located on the sides of the omni-directional LED. A latched alarm can be manually reset by pressing the alarm reset key. If no alarms are active, the highest priority latched alarm appears on the omni-directional LED on the GUI. A lockable alarm is one that does not terminate an active audio paused function (it does not sound an audible alert during an active audio paused function), while a non-lockable alarm cancels the audio paused period and sounds an audible alert. All patient data alarms and the CIRCUIT
DISCONNECT alarm are lockable alarms.
Note: When a new lockable alarm occurs, the alarm will not start to sound audibly if the previous lockable alarm was muted.
The following rules define how alarm messages are displayed:
• Primary alarms precede any dependent alarms.
• The system adds dependent alarms to the analysis messages of each active primary alarm with which they are associated. If a dependent alarm resets, the system removes it from the analysis message of the primary alarm.
• The priority level of a primary alarm is equal to or greater than the priority level of any of its active dependent alarms.
• An alarm cannot be a dependent alarm of any alarm that occurs subsequently.
• If a primary alarm resets, any active dependent alarms become primary unless they are also dependent alarms of another active primary alarm. This is due to different reset criteria for primary and dependent alarms.
• The system applies the new alarm limit to alarm calculations from the moment a change to an alarm limit is accepted.
• The priority level of a dependent alarm is based solely on its detection conditions (not the priority of any associated alarms.
• When an alarm causes the ventilator to enter OSC, or safety valve open (SVO), the patient data display
(including waveforms) is blanked. The elapsed time without ventilatory support (that is, since OSC, or SVO
136
6 began) appears on the GUI screen. If the alarm causing OSC, or SVO is autoreset, the ventilator resets all patient data alarm detection algorithms.
Table 32.
Alarm Descriptions and Symbols
High compensation pressure
Alarm description
High delivered oxygen percentage
High exhaled minute volume
High exhaled minute volume setting
High exhaled tidal volume
High exhaled tidal volume setting
High inspired tidal volume limit
High internal ventilator pressure
High respiratory rate
High respiratory rate setting
High spontaneous inspiratory time
High spontaneous inspiratory time limit
High circuit pressure
High circuit pressure limit
Low circuit pressure
Low circuit pressure setting
Low exhaled mandatory tidal volume
Low exhaled mandatory tidal volume setting
Low exhaled minute volume
Low exhaled minute volume setting
Low exhaled spontaneous tidal volume
Low exhaled spontaneous tidal volume setting
Low delivered oxygen percentage
Symbol
↑ T
I SPONT
⤒ T
I SPONT
↑ P
PEAK
⤒ P
PEAK
↓ P
PEAK
⤓ P
PEAK
↓ V
TE MAND
⤓ V
TE MAND
↓ ⩒
E TOT
⤓⩒
E TOT
↓ V
TE SPONT
⤓ V
TE SPONT
↓ O
2
%
↑ P
COMP
↑ O
2
%
↑ ⩒
E TOT
⤒⩒
E TOT
↑ V
TE
⤒ V
⤒ V
TE
TI
↑ P
VENT
↑ f
TOT
⤒ f
TOT
6.5.2 Alarm Reset Key
The alarm reset function can be used for any non-technical alarm.
See Section 6.5.8, Alarm Handling, page 142
for an explanation of technical vs. non-technical alarms . Alarm reset reinitializes the algorithm the ventilator uses to initially detect the alarm except for A/C POWER LOSS, COMPRESSOR
INOPERATIVE, LOW BATTERY, NO AIR SUPPLY, NO O
2
SUPPLY, PROCEDURE ERROR alarms and active battery alarms. If the cause of the alarm still exists after the alarm reset key is pressed, the alarm becomes active again. The ventilator logs all actuations of the alarm reset key.
6.5.3 Audio Paused Key
Warning: Do not pause, disable, or decrease the volume of the ventilator’s audible alarm if patient safety could be compromised.
The audio paused feature temporarily mutes the audible portion of an alarm for 2 minutes. After the
2-minute period, if the alarm condition still exists, the alarm sounds again. Pressing the audio paused key again restarts the 2-minute interval during which an alarm is muted. An LED within the key illuminates and a count-down timer appears on the GUI next to an audio paused indicator symbol,
137
6
indicating an active audio paused function. The audio paused feature does not allow the audible alarm to be turned off; the audible portion of the alarm is temporarily muted for 2 minutes. The GUI’s omni-directional LED flashes during an active alarm state, and during an audio paused period and its appearance changes with the priority if the alarm escalates. Pressing the alarm reset key cancels an audio paused condition. If the condition that caused the alarm still exists, the alarm activates again.
6.5.4 Alarm Volume Key
An alarm volume key is available for setting the desired alarm volume. The alarm volume is auto‐ matically set to the factory default setting of 10 (maximum) or to the institutional default setting based on circuit type if it has been so configured. When setting the alarm volume, a sample tone is generated, allowing the practitioner to decide the appropriate alarm volume for the surrounding ambient con‐ ditions. If a high priority alarm occurs, the alarm volume increases one increment from its current volume level if it is not acknowledged within 30 s. If a high priority alarm is not acknowledged within
60 s, the audible alarm volume escalates to its maximum volume.
See
Section 3.8.2.6, Alarm Volume, page 79
for instructions on adjusting the alarm volume.
Warning: The audio alarm volume level is adjustable. The operator should set the volume at a level that allows the operator to distinguish the audio alarm above background noise levels. See
Section 3.8.2.6, Alarm Volume, page 79
.
6.5.5 Alarm Testing
Testing the alarms requires oxygen and air sources and stable AC power. To verify the alarm system the operator needs to be standing in front of the ventilator. Test the alarms at least every 12 months, using the procedures described.
Required Equipment
• Test lung (P/N 4-000612-00)
• Adult patient circuit
If the alarm does not annunciate as indicated, verify the ventilator settings and repeat the test. The alarm tests check the operation of the following alarms:
• CIRCUIT DISCONNECT
• LOW EXHALED MANDATORY TIDAL VOLUME ( ↓ V
TE MAND
)
• LOW EXHALED TOTAL MINUTE VOLUME ( ↓ ⩒
E TOT
)
• HIGH CIRCUIT PRESSURE ( ↑ P
PEAK
)
• SEVERE OCCLUSION
• AC POWER LOSS
• APNEA
• LOW EXHALED SPONTANEOUS TIDAL VOLUME ( ↓ V
TE SPONT
)
• NO O
2
SUPPLY
• LOW DELIVERED O
2
% ( ↓ O
2
%)
• HIGH DELIVERED O
2
% ( ↑ O
2
%)
Ventilator setup for alarms tests
1. Disconnect the patient circuit from the ventilator and turn the ventilator off for at least 5 minutes.
2. Turn the ventilator or on. The ventilator runs POST.
3. On the GUI, select NEW PATIENT.
4. Set up new patient using the following settings.
138
6
PBW : 70 kg
Ventilation type : Invasive
Mode : A/C
Mandatory type : VC
Trigger type : ⩒
TRIG
5. Set the following new patient settings f : 6.0 1/min
V
T
: 500 mL
V
MAX
T
PL
: 30 L/min
: 0 s
Flow pattern : square
⩒
SENS
: 3 L/min
O
2
%: 21%
PEEP : 5 cmH
2
O
6. Set the following apnea settings
T
A
: 10 s f : 6.0 1/min
O
2
%: 21%
V
T
: 500 mL
7. Set the following alarm settings
⤒ P
PEAK
: 70 cmH
2
O f
TOT
: OFF
⤓⩒
⤒⩒
⤓ V
⤒ V
⤓ V
⤒ V
E TOT
E TOT
: 1 L/min
: 3.5 L/min
TE MAND
: 300 mL
TE MAND
: OFF
TE SPONT
TE SPONT
: OFF
: OFF
8. Set the graphics display to a volume-time plot (for use in the APNEA alarm test).
9. Connect an Adult patient circuit to the ventilator and attach a test lung to the patient wye.
Note: To ensure proper test results, do not touch the test lung or patient circuit during the CIRCUIT DISCONNECT alarm test.
CIRCUIT DISCONNECT alarm test
1. Allow the ventilator to deliver at least four breaths. During the inspiratory phase of a breath, disconnect the inspiratory filter from the to patient port. The ventilator annunciates a CIRCUIT DISCONNECT alarm after the inspiratory filter is disconnected.
2. Connect the inspiratory filter to the to patient port to autoreset the alarm.
LOW EXHALED MANDATORY TIDAL VOLUME ( ↓ V
TE MAND
) alarm test
Set V
T
to 225 mL. The ventilator annunciates a LOW EXHALED MANDATORY TIDAL VOLUME ( ↓ V the third consecutive breath after Accept is touched.
TE MAND
) alarm on
LOW EXHALED TOTAL MINUTE VOLUME ( ↓ ⩒
E TOT
) alarm test
6
139
Set ⤓⩒
TOT
E TOT
alarm limit to 3.45 L/min. The ventilator annunciates a LOW EXHALED TOTAL MINUTE VOLUME ( ↓ ⩒
) alarm on the next breath after Accept is touched.
E
HIGH CIRCUIT PRESSURE ( ↑ P
PEAK
) alarm test
1. Make the following patient and alarm settings changes.
a. V
T
: 500 mL b.
⩒
MAX
: 30 L/min c.
⤒ P
PEAK
: 20 cmH
2
O
2. After one breath, the ventilator annunciates a HIGH CIRCUIT PRESSURE ( ↑ P
PEAK sound, check the patient circuit for leaks.
) alarm. If the alarm does not
SEVERE OCCLUSION alarm test
1. Make the following alarm settings change:
⤒ P
PEAK
: 50 cmH
2
O
2. Press the alarm reset key to reset all alarms.
3. Adjust D
SENS
to the ⩒
MAX
setting.
4. Disconnect the ventilator breathing circuit from the from patient port and block the gas flow.
5. While maintaining the occlusion, ensure the safety valve open indicator appears on the status display, the
GUI shows the elapsed time without normal ventilation support, and the test lung inflates and deflates rapidly with small pulses as the ventilator delivers trial pressure-based breaths.
6. Press the alarm reset key to reset all the alarms.
AC POWER LOSS alarm test
1. Allow the ventilator to deliver at least four breaths, then disconnect the power cord from AC facility power.
If any battery is charged, the GUI annunciates an AC POWER LOSS alarm. If less than 10 minutes of battery backup are available, the GUI annunciates a LOW BATTERY alarm. If no battery power is available, the BDU annunciates a LOSS OF POWER alarm.
2. Connect the power cord to AC facility power. The AC POWER LOSS or LOW BATTERY alarm autoresets.
APNEA alarm test
1. Make the following alarm settings changes: a.
⤒ P
PEAK
: 70 cmH
2
O b.
Mode : SPONT c.
Spontaneous type : PS
2. The GUI annunciates an APNEA alarm within 10 s after touching Accept .
3. Squeeze the test lung twice to simulate two subsequent patient-initiated breaths. The APNEA alarm autoresets.
4. Let the ventilator return to apnea ventilation.
Note: To avoid triggering a breath during the apnea interval, do not touch the test lung or patient circuit.
Note: For the apnea alarm test, the exhaled tidal volume (V
TE
) displayed in the patient data area must be greater than half the delivered volume shown on the volume-time plot in the graphics display in order for apnea to
LOW EXHALED SPONTANEOUS TIDAL VOLUME alarm test
140
6
1. Make the following patient and alarm settings changes a.
Trigger type : P
TRIG b. P
SENS
: 4 cmH
2
O c.
⤓ V
TE SPONT
: 2500 mL
2. Press the alarm reset key to reset the apnea alarm.
3. Slowly squeeze the test lung to simulate spontaneous breaths. The ventilator annunciates a LOW EXHALED
SPONTANEOUS TIDAL VOLUME ( ↓ V inspiration.
TE SPONT
) alarm at the start of the fourth consecutive spontaneous
4. Make the following patient settings changes: a.
Mode : A/C b.
⤓ V
TE SPONT
: OFF
5. Press the alarm reset key to reset the ⤓ V
TE SPONT
alarm.
NO O
2
SUPPLY alarm test
1. Disconnect the oxygen inlet supply. The ventilator annunciates a NO O
2
SUPPLY alarm within one breath.
2. Connect the oxygen inlet supply. The NO O
2 reconnected.
SUPPLY alarm autoresets within two breaths after oxygen is
LOW DELIVERED O
2
% and HIGH DELIVERED O
2
% alarms tests
1. Make the following patient and alarm settings changes:
P
SENS
: 2 cmH
2
O
O
2
%: 100%
2. Make the following apnea settings changes:
T
A
: 60 s
3. Attach the ventilator’s oxygen gas hose to a known air supply (for example, a medical grade air cylinder) or a wall air outlet.
4. Attach the ventilator’s air gas hose to a known medical oxygen supply.
5. Observe the GUI screen. The delivered O a medium priority ↓ O
2
2
% display should decrease, and the ventilator should annunciate
% alarm within 60 s and a high priority ↓ O
2
% alarm within 2 minutes.
6. Set the O
2
% to 21%.
7. Observe the GUI screen. The delivered O a medium priority ↑ O
2
2
% display should increase, and the ventilator should annunciate
% alarm within 60 s and a high priority ↑ O
2
% alarm within 2 minutes.
8. Remove the air gas hose from the oxygen supply and reconnect the hose to a known medical air supply.
9. Remove the oxygen gas hose from the air supply and reconnect the hose to a known oxygen supply.
10. Press the alarm reset key to clear all alarms.
Warning: Before returning the ventilator to service, review all settings and set appropriately for the patient to be ventilated.
6.5.6 Viewing Alarms
When an alarm occurs, the omni-directional LED at the top of the GUI flashes in a color corresponding to the alarm
. When the alarm banner appears, it displays its base message. Touching the individual alarm causes an expanded explanation to appear, containing analysis and remedy messages, and may contain a link to the alarm log or the alarms settings screen. Touch the link to display requested information. The omni-directional LED remains steadily lit and may appear multicolored, meaning that multiple alarms with varying priority levels have occurred. During an event that causes multiple alarms, the ventilator simultaneously displays the two highest priority active alarms.
141
6
6.5.7 Alarm Delay
6.5.7.1 Determination of an Alarm Condition
The delay time from the moment the alarm condition first occurs until the alarm is annunciated is imperceptible.
6.5.7.2 Delay to/from a Distributed Alarm System
For alarm conditions relayed via the serial port, the overall delay is dependent upon the polling rate of the external device. The delay from the time the serial port is polled by the external device, until the alarm message leaves the serial port does not exceed 3 seconds. An example of an external device is a patient monitor.
6.5.8 Alarm Handling
Current alarm settings are saved in the ventilator’s non-volatile memory (NVRAM) . If the alarm settings are changed by another clinician, those settings become applicable. For example, there are no operator-selectable default alarm settings.
The ventilator system’s alarm handling strategy is intended to
• Detect and call attention to legitimate causes for caregiver concern as quickly as possible, while minimizing nuisance alarms.
• Identify the potential cause and suggest corrective action for certain types of alarms. However, the clinician must make the final decision regarding any clinical action.
• Make it easy to discern an alarm’s priority level.
• Allow quick and easy alarm setup.
Ventilator alarms are categorized as high priority, medium priority, or low priority, and are classified as technical or non-technical.
The ventilator is equipped with two alarms — the primary alarm secondary alarms. The primary alarm annunciates high, medium, and low priority alarms when they occur. The secondary alarm (also named “immediate” priority in
alarm is powered by a capacitor and lasts for at least 120 seconds.
lists alarm priority levels and their visual, audible, and autoreset characteristics. An alarm autoresets when the condition causing the alarm no longer exists.
Table 33.
Alarm Prioritization
Priority Level
Immediate
High: immediate atten‐ tion required to ensure patient safety.
Visual indicator
Specific to alarm condi‐ tion or component failure.
Flashing red LED located on the top of the GUI, red alarm banner on GUI screen, red bar next to alarm setting icon on
Alarms screen.
Audible indicator
Continuous tone alarm sounding for at least 120 s.
High-priority audible alarm (a sequence of five tones that repeats twice, pauses, then repeats again).
Autoreset characteris‐ tics
N/A
Visual alarm does not auto reset. Visual alarm indica‐ tors remain steadily illumi‐ nated following an autor‐ eset. The alarm reset key must be pressed to extin‐ guish visual indicator.
Medium: prompt atten‐ tion necessary.
Flashing yellow LED loca‐ ted on the top of the GUI, yellow alarm banner on
GUI screen, and yellow bar
Medium-priority audible alarm (a repeating sequence of three tones).
LED indicator turns off and autoreset is entered into the alarm log.
142
6
Table 33.
Alarm Prioritization (continued)
Priority Level Visual indicator next to alarm setting icon on Alarms screen.
Audible indicator Autoreset characteris‐ tics
Low: a change in the patient-ventilator system has occurred.
Steadily illuminated yel‐ low LED located on the top of the GUI, yellow alarm banner on GUI screen, and yellow bar next to alarm setting icon on Alarms screen.
Low-priority audible alarm
(two tone, non-repeating).
LED indicator turns off and autoreset is entered into the alarm log.
Normal: normal ventilator operation
Immediate
Steadily illuminated green
LED located on the top of
GUI, no alarm banner, and white values next to alarm setting icon on Alarms screen.
Status display shows the
GUI has failed.
None
The secondary alarm annunciates a repeating sequence of single tones, since the primary alarm
(part of the GUI) has failed.
None
None
A technical alarm is one that is caused by a violation of any of the ventilator’s self monitoring conditions, such as failure of POST or a fault detected by the ventilator’s background diagnostic system. This includes faults detected by the ventilator’s background diagnostic system. Technical alarms cannot be reset by pressing the
alarm reset key. (See Section 10.16.4, Background Diagnostic System, page 237
. Technical alarms fall into eight categories, shown in
Table 34.
Technical Alarm Categories
Category
1
Name
Vent-Inop
Priority
High
2
3
4
5
6
7
8
Exh BUV
Insp BUV
Mix BUV
SVO
Caution
Warning
Notification
High
High
High
High
High
Medium
Low
System Response
Ventilator goes to safe state. See Section 4.11, Ven‐ tilator Protection Strategies, page 114
.
Backup ventilation
Backup ventilation
Backup ventilation
Ventilator goes to safe state. See Section 4.11, Ven‐ tilator Protection Strategies, page 114
.
Ventilation continues as set
Ventilation continues as set
Ventilation continues as set (not displayed on alarm banner)
See
Table 35 for a list of ventilator technical alarms, their meaning, and what to do if they occur.
6
143
See
ventilator alarms.
Table 35.
Technical Alarms
Alarm message
O
2
SENSOR
DEVICE ALERT
O
2 failed.
Meaning
sensor is out of calibration or has
Various. Technical alarm category is described. See
.
More information for the particular technical alarm can be found in the
System diagnostic log, a link to which is provided on the expanded alarm banner.
What to do
Re-calibrate or replace O
2
sensor.
Follow remedy message displayed on GUI.
A non-technical alarm is an alarm caused due to a fault in the patient-ventilator interaction or a fault in the electrical or gas supplies that the practitioner may be able to alleviate.
Table 36.
Non-technical Alarm Summary
Base message
AC POWER
LOSS
AC POWER
LOSS
Priority
Low
Low
Analysis message
Operating on vent main battery.
Operating on vent main and compressor battery.
APNEA
(patient data alarm)
Medium Apnea ventilation. Breath
High interval > apnea interval.
Extended apnea duration or multiple apnea events.
Remedy message
N/A
N/A
Check patient & set‐ tings.
Comments
Ventilator automatically switches to battery power. Power switch on.
AC power not available.
Battery operating indica‐ tor on status display turns on. Resets when AC power is restored.
The set apnea interval has elapsed without the ventilator, patient, or operator triggering a breath. Resets after patient initiates a third consecutive breath. Pos‐ sible dependent alarm:
↓ ⩒
E TOT
144
6
Table 36.
Non-technical Alarm Summary (continued)
Base message
CIRCUIT DIS‐
CONNECT
COMPLIANCE
LIMITED V
T
(alarm is not adjustable)
(patient data alarm)
COMPRESSOR
INOPERATIVE
↑ P
PEAK
(patient data alarm)
Priority
High
High
Low
Low
Analysis message
No ventilation.
No ventilation.
Compliance compensation limit reached.
No compressor air.
Low Last breath ≥ set limit.
Medium Last 3 breaths ≥ set limit.
High Last 4 or more breaths ≥ set limit.
Remedy message
Check patient
Reconnect circuit.
Check patient.
Reconnect circuit.
Inspired volume may be < set. Check patient and circuit type.
Comments
Ventilator has recovered from unintended power loss lasting more than 5 minutes, detects circuit disconnect. The GUI screen displays elapsed time without ventilator support. Resets when ventilator senses recon‐ nect ion.
Ventilator detects circuit disconnect and switches to Stand-By state; the GUI screen displays elapsed time without ventilator support. Resets when ventilator senses recon‐ nection.
Compliance volume required to compensate delivery of a VC, VC+ or VS breath exceeds the max‐ imum allowed for three consecutive breaths.
Replace compressor No compressor ready indicator on status dis‐ play.
Check patient, cir‐ cuit & ET tube.
Measured airway pres‐ sure ≥ set limit. Ventilator truncates current breath unless already in exhala‐ tion. Possible dependent alarms: ↓ V
E
TOT
, ↑ f
TE MAND
, ↓ ⩒
TOT
. Corrective action: Check patient.
Check for leaks, tube type/ID setting. Consider reducing % Supp setting or increasing ⤒ P
PEAK
.
6
145
Table 36.
Non-technical Alarm Summary (continued)
Base message
↑ P
COMP
(patient data alarm)
↓ P
PEAK
(patient data alarm)
Priority
Low
Medium Last 3 spont breaths ≥ set
P
PEAK
limit –5 cmH
2
O.
High Last 4 or more spont breaths ≥ set P limit–5 cmH
2
O.
PEAK
Low
Analysis message
Last spont breath ≥ set
P
PEAK
limit –5 cmH
2
O.
Last 2 breaths, pressure
≤ set limit.
Medium Last 4 breaths, pressure
≤ set limit.
High Last 10 or more breaths, pressure ≤ set limit.
Remedy message
In TC or PAV+™:
Check for leaks, tube type, and tube ID setting.
Check for leaks.
Comments
Pressure of spontaneous breaths ≥ set limit. Possi‐ ble dependent alarms:
↓ V
TE SPONT
, ↓ ⩒
E TOT
, ↑ f
↓ V
TE SPONT
Corrective
TOT
, action: Check for leaks.
Check for the correct tube type. Check that the tube inside diameter cor‐ responds to the patient
PBW.
Check the ⤒ P
PEAK
setting.
Peak inspiratory pressure
≤ alarm setting. (Availa‐ ble only when Manda‐ tory Type is VC+* or when ventilation type is NIV.
Target pressure = the low limit: PEEP +3 cmH
2
O.
Ventilator cannot deliver target volume. Possible dependent alarms: ↑ f
TOT
.
Corrective action: check patient and settings.
* Because the VC+ pressure control algorithm does not allow the target inspiratory pressure to fall below
PEEP +3 cmH
2
O, attempting to set the ⤓ P
PEAK
alarm limit at or below this level will turn the alarm off.
↑ O
2
% (patient data alarm)
Medium Measured O
2
% > set for
≥ 30 s but < 2 min.
High Measured O
≥ 2 min.
2
% > set for
Check patient, gas sources, O
2
& ventilator.
analyzer
The O
2
% measured dur‐ ing any phase of a breath cycle is 7% (12% during the first hour of opera‐ tion) or more above the
O
2
% setting for at least 30 seconds. (These percen‐ tages increase by 5% for 4 minutes following a decrease in the O ting).
2
% set‐
↓ O
2
% (patient data alarm)
High Measured O
2
% < set O
2
%. Check patient, gas sources, O
2
& ventilator.
analyzer
The O
2
% measured dur‐ ing any phase of a breath cycle is 7% (12% during the first hour of opera‐ tion) or more below the
O
2
% setting for at least 30 seconds, or below 18%.
(These percentages increase by 5% for 4 minutes following an
146
6
Table 36.
Non-technical Alarm Summary (continued)
Base message
↑ V
TE
(patient data alarm)
↑
(patient data alarm)
↑ f
⩒
E TOT
TOT
(patient data alarm)
Priority
Low
Medium
High
Low
High
Analysis message
⩒
E TOT
Medium ⩒
E TOT
≥ set limit for ≤ 30 s. Check patient and
≥ set limit for > 30 s.
settings.
⩒
E TOT
≥ set limit for > 120 s.
Remedy message
Last 2 breaths ≥ set limit.
Check settings,
Last 4 breath s≥ set limit.
Last 10 or more breaths
≥ set limit.
changes in patient’s
R&C.
Low f
TOT
Medium f
TOT
≥ set limit for ≤ 30 s.
Check patient & set‐
≥ set limit for > 30 s.
tings.
High f
TOT
≥ set limit for > 120 s.
Comments increase in the O
2 ting).
% set‐
Exhaled tidal volume
≥ set limit. Alarm upda‐ ted whenever exhaled tidal volume is recalcula‐ ted. Possible dependent alarm: ↑ ⩒
E TOT
Expiratory minute vol‐ ume ≥ set limit. Alarm updated whenever an exhaled minute volume is recalculated. Possible dependent alarm: ↑ V
TE
.
Total respiratory rate
≥ set limit. Alarm upda‐ ted at the beginning of each inspiration. Reset when measured respira‐ tory rate falls below the alarm limit. Possible dependent alarms: ↓ V
MAND
, ↓ V
TE SPONT
, ↓ ⩒
TE
E TOT
.
147
6
Table 36.
Non-technical Alarm Summary (continued)
Base message
↑ P
VENT
(patient data alarm)
INOPERATIVE
BATTERY
INOPERATIVE
BATTERY
INOPERATIVE
BATTERY
Priority
Low
Medium
High
Low
Low
Low
Analysis message
1 breath ≥ limit.
2 breaths ≥ limit.
3 or more breaths ≥ limit.
Remedy message
Check patient, cir‐ cuit & ET tube.
Inadequate charge or non‐ functional ventilator bat‐ tery system.
Inadequate charge or non‐ functional compressor bat‐ tery system.
Inadequate charge or non‐ functional ventilator bat‐ tery system.
Inadequate charge or non‐ functional compressor bat‐ tery system.
Service/replace ven‐ tilator battery.
Service/replace compressor battery.
Service/replace ven‐ tilator battery. Serv‐ ice/replace com‐ pressor battery.
Comments
Inspiratory pressure >
110 cmH
2
O and manda‐ tory type is VC or sponta‐ neous type is TC or
PAV+™. Ventilator trun‐ cates current breath unless already in exhala‐ tion. Possible dependent alarms:
↑ f
TOT
.
↓ V
TE MAND
Corrective action:
, ↓ ⩒
E TOT
1. Check patient for agitation. Agitated breathing, com‐ bined with high %
Supp setting in
PAV+™ can cause over-assistance.
Consider reducing
% Supp setting.
2. Provide alternate ventilation.
Remove ventilator from use and con‐ tact Service.
Battery installed but not functioning or charging for ≥6 hours. Resets when battery is functional.
148
6
Table 36.
Non-technical Alarm Summary (continued)
Base message
INOPERATIVE
BATTERY
INOPERATIVE
BATTERY
INOPERATIVE
BATTERY
INSPIRATION
TOO LONG
(patient data alarm)
Priority Analysis message
Medium Inadequate charge or non‐ functional ventilator bat‐ tery system.
Medium Inadequate charge or non‐ functional compressor bat‐ tery system.
Medium Inadequate charge or non‐ functional ventilator bat‐ tery system.
Inadequate charge or non‐ functional compressor bat‐ tery system.
Low Last 2 spont breaths = PBW based T
I
limit.
Medium Last 4 spont breaths = PBW based T
I
limit.
High Last 10 or more spont breaths = PBW based T limit.
I
PAV STARTUP
TOO LONG
(patient data alarm) (occurs only if PAV+™ is in use)
Low PAV startup not complete for ≥ 45 s.
Medium PAV startup not complete for ≥ 90 s.
High PAV startup not complete for ≥ 120 s.
Remedy message
Service/replace ven‐ tilator battery.
Service/replace compressor battery.
Service/replace ven‐ tilator battery. Serv‐ ice/replace com‐ pressor battery.
Check patient.
Check for leaks.
Check for leaks, shal‐ low breathing, & set‐ tings for ↑ V
↑ P
PEAK
.
TI
and
Comments
Defective battery detec‐ ted. Corrective action-
Replace battery as soon as possible.
Inspiratory time for spon‐ taneous breath
≥ PBW-based limit. Venti‐ lator transitions to exha‐ lation. Resets when T
I
falls below PBW-based limit.
Active only when ventila‐ tion type is Invasive.
Unable to assess patient’s resistance and compli‐ ance during PAV startup.
Possible dependent alarms ↓ V
TE SPONT
, ↓ ⩒
E
TOT
, ↑ f
TOT
. Corrective action: check patient.
(Patient’s inspiratory times may be too short to evaluate resistance and compliance.) Check that selected humidification type and empty humidi‐ fier volume are correct.
6
149
Table 36.
Non-technical Alarm Summary (continued)
Base message
PAV R & C NOT
ASSESSED
(patient data alarm) (occurs only if PAV+™ is in use)
LOSS OF
POWER
LOW BATTERY Medium
LOW BATTERY Medium
LOW BATTERY Medium
LOW BATTERY
LOW BATTERY
LOW BATTERY
Priority
Low
Medium
Immediate
High
High
High
R and/or C over 15 minutes old.
R and/or C over 30 minutes old.
N/A
Analysis message
Vent main battery opera‐ tional time <10 minutes.
Compressor battery opera‐ tional time <10 minutes.
Vent main battery opera‐ tional time <10 minutes and compressor battery operational time <10 minutes.
Vent main battery opera‐ tional time <5 minutes.
Compressor battery opera‐ tional time <5 minutes.
Vent main battery opera‐ tional time <5 minutes and
Remedy message
Check for leaks, shal‐ low breathing, & set‐ tings for tube ID, ↑ V and ↑ P
PEAK
.
TI
N/A
Comments
Unable to assess resist‐ ance or compliance dur‐ ing PAV+™ steady-state.
Startup was successful, but later assessments were unsuccessful. Cor‐ rective action: check patient. (Patient’s inspira‐ tory times may be too short to evaluate resist‐ ance and compliance).
Check that selected humidification type and empty humidifier vol‐ ume are correct.
The ventilator power switch is ON and there is insufficient power from
AC and the battery. There may not be a visual indi‐ cator for this alarm, but an independent audio alarm sounds for at least
120 seconds. Alarm annunciation can be reset by turning power switch to OFF position.
Resets when battery has
≥10 minutes of opera‐ tional time remaining.
Replace or allow recharge vent main battery.
Replace or allow recharge compres‐ sor battery.
Replace or allow recharge vent main battery and com‐ pressor battery.
Replace or allow recharge vent main battery.
Replace or allow recharge compres‐ sor battery.
Replace or allow recharge vent main
Resets when main bat‐ tery or compressor bat‐ tery has ≥10 minutes of operational time remain‐ ing or when AC power is restored.
Resets when battery has
≥5 minutes of opera‐ tional time remaining or when AC power is restored.
Resets when battery has
≥5 minutes of opera‐ tional time remaining or
150
6
Table 36.
Non-technical Alarm Summary (continued)
Base message
↓
(patient data alarm)
↓ V
TE SPONT
(patient data alarm)
↑ data alarm)
↓
V
V
V
TE MAND
TI
(patient
E TOT
(patient data alarm)
Priority Analysis message compressor battery opera‐ tional time <5 minutes.
Low Last 2 mand breaths ≤ set limit.
Medium Last 4 mand breaths ≤ set limit.
High Last 10 or more mand breaths ≤ set limit.
Low Last 4 spont breaths ≤ set limit.
Medium Last 7 spont breaths ≤ set limit.
High Last 10 or more spont breaths ≤ set limit.
Low
High
Last spont breath ≥ set limit.
Medium Last 3 spont breaths ≥ set limit.
Last 4 or more spont breaths ≥ set limit.
Remedy message battery and com‐ pressor battery.
Check for leaks, changes in patient’s
R & C.
Check patient & set‐ tings.
In TC, VS, or
PAV+™:
• Check patient and settings.
Low ⩒
E TOT
Medium ⩒
E TOT
High
≤ set limit for ≤30 s. Check patient & set‐
≤ set limit for >30 s.
tings.
⩒
E TOT
≤ set limit for >120 s.
Comments when AC power is restored.
Exhaled mandatory tidal volume.≤ set limit. Alarm updated whenever exhaled mandatory tidal volume is recalculated.
Possible dependent alarms: ↓ ⩒
E TOT
, ↑ f
TOT
.
Exhaled spontaneous tidal volume ≤ set limit.
Alarm updated when‐ ever exhaled spontane‐ ous tidal volume is recal‐ culated. Possible dependent alarms: ↓ ⩒
TOT
↑ f
TOT
.
E
Delivered inspiratory vol‐ ume ≥ inspiratory limit.
Ventilator transitions to exhalation. Possible dependent alarms: ↓ V
SPONT
, ↓ ⩒
E TOT
, ↑ f
TE
TOT
Corrective action: check for leaks. Check for the correct tube type.
TI
or V
TI
set‐ Check the V ting. In PAV+™, check for patient agitation, which can cause miscalculation of R
PAV
and C
PAV
Check ⤒ V
TI
.
. Consider reducing % Supp setting.
Total minute volume
≤ set limit. Alarm upda‐ ted whenever exhaled minute volume is recal‐ culated. Possible dependent alarms ↓ V
TE
MAND
, ↓ V
TE SPONT
, ↑ f
TOT
6
151
Table 36.
Non-technical Alarm Summary (continued)
Base message
VOLUME NOT
DELIVERED
(not adjusta‐ ble) (patient data alarm)
NO AIR SUPPLY
NO AIR SUPPLY
NO O
2
SUPPLY
PROCEDURE
ERROR
SEVERE
OCCLUSION
INADVERTENT
POWER OFF
PROX INOPER‐
ATIVE
Low
Priority
High
Low
Low
High
Low
High
High
High
Analysis message
Last 2 spont (or mand) breaths pressure > max allowable level.
Medium Last 10 or more spont (or mand) breaths, pressure > max allowable level.
Remedy message
Check patient and setting for ↑ P
PEAK
Comments
Insp target pressure >
(P
PEAK
–PEEP –3 cmH
2
O), when spontaneous type is VS or mandatory type is
VC+ Ventilator cannot deliver target volume.
Possible dependent alarms: for VC+ breaths:
↓ V
TE MAND
↓ ⩒
E TOT
↑ f
TOT
.
For VS breaths: ↓ V
TE SPONT
,
↓ ⩒
E TOT
, ↑ f settings.
TOT
Corrective action: check patient and
Ventilator delivers 100%
O
2
. Air supply pressure
≤17 psig. Resets if air sup‐ ply pressure ≥35 psig is connected.
Compressor inoperative.
Ventilation continues as set.
Only O
2
available.
Compressor inoperative.
Ventilation continues as set, except y O
2
% = 100.
Ventilation continues as set.
Only air available.
Check air source.
Check patient air source.
Check O
2
source.
Ventilation continues as set, except O
2
% = 21.
Patient connected before setup complete.
Little/no ventilation.
Ventilator switched OFF with patient connected to breathing circuit.
Data from proximal flow sensor is not being used.
Check patient & O source.
Provide alternate ventilation.
2
Complete setup pro‐ cess.
Check patient. Pro‐ vide alternate venti‐ lation. Clear occlu‐ sions; drain circuit.
Return power switch to on position and disconnect patient before turning power off.
Check proximal flow sensor connections and tubes for occlu‐ sions or leaks.
Operator-set O
2
21%. Resets if O
2 connected.
% equals
supply
Ventilator delivers 21%
O
2
instead of set O
2
%.
Resets if oxygen supply connected.
Ventilator begins safety ventilation. Resets when ventilator startup proce‐ dure is complete.
Ventilator enters occlu‐ sion status cycling (OSC).
Patient data displays are blanked and GUI screen displays elapsed time without ventilator sup‐ port.
User must acknowledge turning the power off by touching Power Off on the GUI.
Data for real time wave‐ forms and monitored vol‐ umes are obtained from internal sensors.
152
6
Table 37.
Non-Technical Alarms and Suggested Responses
Alarm message
AC POWER LOSS
Meaning
The ventilator or compressor is run‐ ning on battery power.
What to do
Monitor the battery charge level to ensure there is enough power remaining to operate the ventila‐ tor/compressor.
Check patient and settings.
APNEA (patient data alarm)
CIRCUIT DISCONNECT
Compliance limited V alarm)
T
(patient data
COMPRESSOR INOPERATIVE
The time between patient breaths exceeds the set apnea interval.
The patient circuit has become dis‐ connected or there is a large leak in the patient circuit.
Compliance volume required to compensate delivery of a VC, VC+, or
VS breath exceeds the maximum allowed for three consecutive breaths.
Air pressure not detected in the compressor’s accumulator. Status display indicates the compressor is inoperative
Reconnect the patient circuit, or eliminate the leak.
Check patient and circuit type.
Inspired volume may be less than set.
Service or replace compressor.
↑ P
PEAK
(patient data alarm) The measured airway pressure is
≥ set limit. Reduced tidal volume likely.
• Check the patient.
• Check the patient circuit.
↓ P
PEAK
(patient data alarm) The peak inspiratory pressure in the patient circuit ≤ alarm setting. This alarm is only available when NIV is the selected ventilation type or when VC+ is the selected Mandatory type during Invasive Ventilation.*
• Check the endotracheal tube.
Check for leaks in the patient circuit and VBS.
* Because the VC+ pressure control algorithm does not allow the target inspiratory pressure to fall below
PEEP +3 cmH
2
O, attempting to set the ⤓ P
PEAK
alarm setting at or below this level will turn the alarm off.
↑ O
2
% (patient data alarm) The O
2
% measured during any phase of a breath cycle is 7% (12% during the first hour of operation) or more above the O
2
% parameter for at least 30 seconds. The percentage window increases by 5% for 4 minutes after increasing the set O value.
2
%
Check the patient, the air and oxy‐ gen supplies, the oxygen analyzer, and the ventilator.
6
153
Table 37.
Non-Technical Alarms and Suggested Responses (continued)
Alarm message
↓ O
2
% (patient data alarm)
↑
↑
↑ f
↑
V
⩒
P
TE
E TOT
TOT
(patient data alarm)
(patient data alarm)
VENT
(patient data alarm)
(patient data alarm)
INOPERATIVE BATTERY
INSPIRATION TOO LONG (patient data alarm)
LOSS OF POWER
LOW BATTERY
Meaning
The O
2
% measured during any phase of a breath cycle is 7% (12% during the first hour of operation) or more below the O
2
% parameter for at least 30 seconds. The percentage window increases by 5% for 4 minutes after increasing the set O value.
2
%
Exhaled tidal volume ≥ alarm setting for the last two breaths.
Minute volume ≥ alarm setting.
The breath rate from all breaths is
≥ alarm setting.
The inspiratory pressure transducer has measured a pressure
>110 cmH
2
O in VC, TC, or PAV+™. The ventilator transitions to exhalation. A reduced tidal volume is likely.
Alarm indicates unable to charge battery or the battery system is non‐ functional
The PBW-based inspiratory time for the last two spontaneous breath exceeds the ventilator-set limit.
Active only when ventilation type is
Invasive.
The ventilator power switch is ON, but there is insufficient power from the mains AC and the battery. There may not be a visual indicator for this alarm, but an independent audio alarm (immediate priority) sounds for at least 120 seconds.
Medium priority alarm indicating
<10 minutes of battery power remaining to operate the ventilator
What to do
• Check the patient, the air and oxygen supplies, the oxygen analyzer, and the ventilator.
• Calibrate the oxygen sensor.
See Section 4.10.2, Oxygen Sen‐ sor Calibration, page 113 for
details regarding calibrating the oxygen sensor.
• Use an external O
2 disable the O
2
monitor and
sensor.
• Check patient settings.
• Check for changes in the patient’s resistance and com‐ pliance.
Check patient settings.
Check the patient and the ventila‐ tor settings.
• Check the patient, the patient circuit (including filters), and the endotracheal tube. Ensure the ET tube ID is the correct size. Check the ventilator flow and volume settings.
• Rerun SST.
• Obtain and alternate ventila‐ tion source.
• Remove the ventilator from clinical use and obtain service.
Replace inoperative battery.
• Check the patient.
• Check the patient circuit for leaks.
• Check rise time% and E
SENS tings.
set‐
• Check the integrity of the AC power and battery connec‐ tions. Obtain alternative venti‐ lation, if necessary.
• Install and extended battery.
• Turn the power switch off to reset the alarm.
Recharge the battery, by plugging the ventilator into AC power or
154
6
Table 37.
Non-Technical Alarms and Suggested Responses (continued)
↓
↓
↓
V
V
⩒
TE MAND
TE SPONT
E TOT
NO O
↑ P
2
SUPPLY
COMP
Alarm message
(patient data alarm)
(patient data alarm)
(patient data alarm)
NO AIR SUPPLY
PROCEDURE ERROR
SEVERE OCCLUSION
Meaning or compressor. High priority alarm indicating <5 minutes of battery power remain to operate the venti‐ lator or compressor.
The patient’s exhaled mandatory tidal volume is ≤ alarm setting for the last two mandatory breaths.
What to do replace the battery, or install an extended battery.
• Check the patient.
• Check for leaks in the patient circuit.
The patient’s exhaled spontaneous tidal volume is ≤ alarm setting for the last two spontaneous breaths.
The minute volume for all breaths is
≤ alarm setting.
• Check for changes in the patient’s resistance or compli‐ ance.
• Check the patient.
• Check the ventilator settings.
• Check the patient.
• Check the ventilator settings.
The air supply pressure is less than the minimum pressure required for correct ventilator operation. The ventilator delivers 100% O
2
if availa‐ ble. If an oxygen supply is not avail‐ able, the safety valve opens. The ven‐ tilator displays the elapsed time without ventilator support. This alarm cannot be set or disabled.
The patient circuit is severely occlu‐ ded. The ventilator enters occlusion status cycling. The elapsed time without ventilatory support appears.
• Check the patient.
• Check the air and oxygen sour‐ ces.
• Obtain alternative ventilation, if necessary.
The oxygen supply pressure is less than the minimum pressure required for correct ventilator oper‐ ation. The ventilator delivers 100% air if available. If an air supply is not avail‐ able, the safety valve opens. The ven‐ tilator displays the elapsed time without ventilatory support. This alarm cannot be set or disabled.
The patient is attached before venti‐ lator startup is complete. Safety ven‐ tilation is active.
• Check the patient.
• Check the air and oxygen sour‐ ces.
• Obtain alternative ventilation, if necessary.
Target pressure ≥ ⤒ P
PEAK
–5 cmH
2
O In TC:
Check for leaks and check tube type and tube ID settings.
In PAV+™:
Limit target pressure to
⤒ P
PEAK
–5 cmH
2
O.
• Provide alternate ventilation, if necessary.
• Complete ventilator startup procedure.
• Check the patient.
• Obtain alternative ventilation, if necessary.
155
6
Table 37.
Non-Technical Alarms and Suggested Responses (continued)
↑ V
TI
Alarm message
(patient data alarm)
VOLUME NOT DELIVERED (patient data alarm)
PAV STARTUP TOO LONG (occurs only if PAV+™ is in use)
PAV R & C NOT ASSESSED (occurs only if PAV+™ is in use)
PROX INOPERATIVE
Meaning
Delivered inspiratory volume ≥ high inspiratory volume limit.
Insp target pressure >
(P
PEAK
–PEEP –3 cmH
2
O), when spon‐ taneous type is VS or when manda‐ tory type is VC+.
Unable to assess resistance or com‐ pliance during PAV+™ startup.
Unable to assess resistance or com‐ pliance during PAV+™ steady-state.
A malfunction occurred with the proximal flow sensor or the pneu‐ matic lines are occluded.
What to do
• Check the patient circuit for bulk liquid, crimps, blocked fil‐ ter.
• If problem persists, remove ventilator from use and obtain service.
• Ventilator transitions to exha‐ lation
• Check for leaks and check tube type and tube ID setting.
• Check patient and ventilator settings.
• Check for leaks, tube type, tube
ID, % Supp setting, and for patient agitation.
Check patient and ↑ P
PEAK
setting.
Check for leaks, shallow breathing, and settings for ↑ V
TI
and ↑ P
PEAK
.
Check for leaks, shallow breathing, and settings for tube ID, ↑ V
TI
and
↑ P
PEAK
.
Replace the proximal flow sensor or purge its pneumatic lines. Does not affect data from the ventilator’s delivery or exhalation flow sensors.
The next sections provide detailed descriptions of selected alarms.
6.5.9 AC POWER LOSS Alarm
The AC POWER LOSS alarm indicates the ventilator power switch is on and the ventilator is being powered by the battery and an alternate power source may soon be required to sustain normal ventilator operation. The ventilator annunciates a medium-priority LOW BATTERY alarm when the ventilator has less than 10 minutes of battery power remaining. The ventilator annunciates a high-priority LOW BATTERY alarm when less than 5 minutes of battery power are estimated available.
The compressor is a DC device, in which AC power is converted to DC power, and it has its own primary and extended batteries (if the extended battery was purchased). If AC power is lost, there is no conversion to DC power for the compressor as in normal operation, but the compressor supplies air, providing the charge level of its batteries is sufficient.
6.5.10 APNEA Alarm
The APNEA alarm indicates neither the ventilator nor the patient has triggered a breath for the operator-selected apnea interval (T
A
). T
A
is measured from the start of an inspiration to the start of the next inspiration and is based on the ventilator’s inspiratory detection criteria. T
A
can only be set via the apnea ventilation settings.
156
6
The APNEA alarm autoresets after the patient initiates two successive breaths, and is intended to establish the patient’s inspiratory drive is reliable enough to resume normal ventilation. To ensure the breaths are patient-initiated (and not due to autotriggering), exhaled volumes must be at least half the V
T returning to normal ventilation if there is a disconnect).
(this avoids
6.5.11 CIRCUIT DISCONNECT Alarm
The CIRCUIT DISCONNECT alarm indicates the patient circuit is disconnected at the ventilator or the patient side of the patient wye, or a large leak is present. The methods by which circuit disconnects are detected vary depending on breath type. Time, pressure, flow, delivered volume, exhaled volume, and the D discussion of the CIRCUIT DISCONNECT detection methods.
SENS
setting may be
The CIRCUIT DISCONNECT alarm sensitivity is adjusted via the D
SENS
setting. During a CIRCUIT DISCONNECT condition, the ventilator enters an idle state and delivers a base flow of oxygen to detect a reconnection.
When the ventilator determines the patient circuit is reconnected, the CIRCUIT DISCONNECT alarm autoresets and normal ventilation resumes without having to manually reset the alarm (for example, following suctioning).
A disconnected patient circuit interrupts gas delivery and patient monitoring. Notification of a patient circuit disconnect is crucial, particularly when the patient cannot breathe spontaneously. The ventilator does not enter apnea ventilation when a disconnect is detected to avoid changing modes during a routine suctioning procedure.
Note: When utilizing a closed-suction catheter system, the suctioning procedure can be executed using existing mode, breath type and settings. To reduce potential for hypoxemia during the procedure, elevate the oxygen concentration using the Elevate O
2
control. See Section 3.8.2.5, Elevate O
.
6.5.12 LOSS OF POWER Alarm
This alarm alerts the operator that there is insufficient battery power and no AC power to support ventilator or compressor operation. The alarm annunciates as long as the ventilator’s power switch is in the ON position, and lasts for at least 120 seconds.
6.5.13 DEVICE ALERT Alarm
A DEVICE ALERT alarm indicates a background test or power on self test (POST) has failed. Depending on which test failed, the ventilator either declares an alarm and continues to ventilate according to current settings, or ventilates with modified settings, or enters the ventilator inoperative state. The DEVICE ALERT alarm relies on the
Background Diagnostic System, page 237
.
6.5.14 HIGH CIRCUIT PRESSURE Alarm
The ↑ P
PEAK
⤒ P
PEAK
alarm indicates the currently measured airway pressure is equal to or greater than the set limit. The
limit is active during all breath types and phases to provide redundant patient protection (for example, to detect air flow restrictions downstream of the pressure-sensing device). The ventilation modes. The ⤒ P
PEAK and adult patients. The ⤒ P
PEAK
↑ P
PEAK
limit is not active during a SEVERE OCCLUSION alarm.
limit is active in all normal
alarm new patient default values are separately configurable for neonatal, pediatric,
The ↑ P
PEAK
alarm truncates inspiration and transitions the ventilator into the exhalation phase and the limit cannot be set less than
• PEEP +7 cmH
2
O or
• PEEP + P
I
+2 cmH
2
O, or
• PEEP + P
SUPP
+2 cmH
2
O, nor can it be set less than or equal to ⤓ P
PEAK
.
157
6
The ⤒ P
PEAK
limit cannot be disabled. The ventilator phases in changes to the ⤒ P
PEAK
limit immediately to allow prompt notification of a high circuit pressure condition.
The minimum ⤒ P
PEAK
limit (7 cmH
2
O) corresponds to the lowest peak pressures not due to autotriggering anticipated during a mandatory breath. The maximum ⤒ P
PEAK
limit (100 cmH maximum pressure required to inflate very low-compliance lungs.
2
O) was selected because it is the
The ventilator allows circuit pressure to rise according to a computed triggering profile for the initial phase of PC and PS breaths without activating the ⤒ P
PEAK
alarm. This triggering profile helps avoid nuisance alarms due to possible transient pressure overshoot in the airway when aggressive values of rise time% are selected. A brief pressure overshoot measured in the patient circuit is unlikely to be present at the carina.
6.5.15 HIGH DELIVERED O
2
% Alarm
The ↑ O
2
% alarm indicates the measured O
30 seconds to eliminate transient O
2
2
% is at or above the error percentage above the O
% delivery variation nuisance alarms. The ↑ O
2 in ventilator gas delivery or oxygen monitor. The ventilator declares a ventilator automatically sets the ↑ O
2
↑ O
2
2
% setting for at least
% alarm detects malfunctions
% alarm after 30 seconds. Although the
% alarm limits, the oxygen sensor can be disabled. (The error percentage is
12% above setting for the first hour of ventilator operation, 7% above the setting after the first hour of operation, and an additional 5% above the setting for the first 4 minutes following a decrease in the setting).
The ventilator automatically adjusts the ↑ O
2
% alarm limit when O
2 occlusion, circuit disconnect, or a NO AIR/O
2 measured oxygen percentage at 1-second intervals.
% changes due to 100% O
SUPPLY alarm. The ventilator checks the ↑ O
2
2
, apnea ventilation,
% alarm limit against the
6.5.16 HIGH EXHALED MINUTE VOLUME Alarm
The ↑ ⩒
E TOT
alarm indicates the measured exhaled total minute volume for spontaneous and mandatory breaths is equal to or greater than the alarm setting. The ↑ ⩒
E TOT
alarm is effective immediately upon changing the setting, to ensure prompt notification of prolonged high tidal volumes.
The ↑ ⩒
E TOT resistance. The ↑ ⩒ hypocarbia.
alarm can be used to detect a change in a patient’s breathing pattern, or a change in compliance or
E TOT
alarm can also detect too-large tidal volumes, which could lead to hyperventilation and
6.5.17 HIGH EXHALED TIDAL VOLUME Alarm
The ↑ V
TE
alarm indicates the measured exhaled tidal volume for spontaneous and mandatory breaths is equal to or greater than the set ↑ V
TE
alarm. The ↑ V
TE
alarm is updated whenever a new measured value is available.
The ↑ V
TE
alarm can detect increased exhaled tidal volume (due to greater compliance and lower resistance) and prevent hyperventilation during pressure control ventilation or pressure support. Turn the ↑ V nuisance alarms. (Hyperventilation due to increased compliance is not a concern during volume-based ventilation, because the tidal volume is fixed by the clinician’s choice and the ventilator’s compliance-compensation algorithm).
TE
alarm OFF to avoid
6.5.18 HIGH INSPIRED TIDAL VOLUME Alarm
The high inspired tidal volume alarm indicates the patient’s inspired volume exceeds the set limit. When this condition occurs, the breath terminates and the alarm sounds. The ventilator displays monitored inspired tidal volume values in the patient data area on the GUI screen. When ventilation type is NIV, there is no high inspired tidal volume alarm or setting available, but the monitored inspired tidal volume (V data area on the GUI screen.
TI
) may appear in the patient
158
6
6.5.19 HIGH RESPIRATORY RATE Alarm
The ↑ f
TOT
alarm indicates the measured breath rate is greater than or equal to the ⤒ f
TOT
alarm setting. The ↑ f
TOT alarm is updated whenever a new total measured respiratory rate is available.
The ↑ f
TOT
alarm can detect tachypnea, which could indicate the tidal volume is too low or the patient’s work of breathing has increased. The ventilator phases in changes to the ⤒ f notification of a high respiratory rate condition.
TOT
limit immediately to ensure prompt
6.5.20 INSPIRATION TOO LONG Alarm
The INSPIRATION TOO LONG alarm, active only when ventilation type is Invasive, indicates the inspiratory time of a spontaneous breath exceeds the following time limit:
(1.99 +0.02 × PBW) seconds (adult and pediatric circuits)
(1.00 +0.10 × PBW) seconds (neonatal circuits) where PBW is the current setting for predicted body weight in kg.
When the ventilator declares an INSPIRATION TOO LONG alarm, the ventilator terminates inspiration and transitions to exhalation. The INSPIRATION TOO LONG alarm applies only to spontaneous breaths and cannot be set or disabled.
Because leaks (in the patient circuit, around the endotracheal tube cuff, or through chest tubes) and patient-ventilator mismatch can affect accurate exhalation detection, the INSPIRATION TOO LONG alarm can act as a backup method of safely terminating inspiration. If the INSPIRATION TOO LONG alarm occurs frequently, check for leaks and ensure E
SENS
and rise time% are properly set.
6.5.21 LOW CIRCUIT PRESSURE Alarm
Warning: Because the VC+ pressure control algorithm does not allow the target inspiratory pressure to fall below
PEEP +3 cmH
2
O, attempting to set the ⤓ P
PEAK
alarm limit at or below this level will turn the alarm off.
The ↓ P
PEAK
alarm indicates the measured maximum airway pressure during the current breath is less than or equal to the set alarm level during a non-invasive inspiration or during a VC+ inspiration.
The ↓ P
PEAK
alarm is active for mandatory and spontaneous breaths, and is present only when ventilation type is
NIV or Mandatory Type is VC+. During VC+, the ↓ P turned OFF during NIV. The limit.
⤓ P
PEAK
PEAK
alarm can be turned OFF. The ↓ P
PEAK
alarm can always be
alarm limit cannot be set to a value greater than or equal to the ⤒ P
PEAK
alarm
In VC+, whenever PEEP is changed, ↓ P
≥16 cmH
2
O, or PEEP +3.5 cmH
2
PEAK
is set automatically to its new patient value, PEEP +4 cmH
O when PEEP <16 cmH
2
O.
2
O when PEEP
There are no alarms dependent upon ↓ P
PEAK
, and the ↓ P
PEAK
alarm does not depend on other alarms.
6.5.22 LOW DELIVERED O
2
% Alarm
The ↓ O
2
% alarm indicates the measured O below the O
2
2
% during any phase of a breath is at or below the error percentage
% setting, or less than or equal to 18%, for at least 30 seconds. Although the ventilator automatically sets the ↓ O
2
% alarm, replace (if necessary) or disable the oxygen sensor to avoid nuisance alarms. The error percentage is 12% below setting for the first hour of ventilator operation following a reset, 7% below setting after the first hour of operation, and an additional 5% below setting for the first 4 minutes following an increase in the setting.
The ventilator automatically adjusts the ↓ O
2 disconnect, or a NO O
2
The ventilator checks the ↓ O
2
% alarm limit when O
/AIR SUPPLY alarm. The ↓ O
2
% changes due to apnea ventilation, circuit
2
% alarm is disabled during a safety valve open (SVO) condition.
% alarm against the measured oxygen percentage at 1-second intervals.
6
159
The ↓ O
2
% alarm can detect malfunctions in ventilator gas delivery or the oxygen monitor, and can ensure the patient is adequately oxygenated. The ventilator declares a alarms from transient O
2
% delivery variations. The O
2
↓ O
2
% alarm after 30 seconds to eliminate nuisance
% measured by the oxygen sensor is shown in the patient data area. See
Section 3.8.2.7, Vital Patient Data, page 79
to include O
2
% if it is not displayed.
6.5.23 LOW EXHALED MANDATORY TIDAL VOLUME Alarm
The alarm indicates the measured exhaled mandatory tidal volume is less than or equal to the ↓ V
TE MAND setting. The ↓ V
TE MAND
alarm
alarm updates when a new measured value of exhaled mandatory tidal volume is available.
The ↓ V
TE MAND
alarm can detect an obstruction, a leak during volume ventilation, or a change in compliance or resistance during pressure-based ventilation (that is, when the same pressure is achieved but tidal volume decreases). There are separate alarms for mandatory and spontaneous exhaled tidal volumes for use during SIMV,
SPONT, and BiLevel. The ventilator phases in a change to the ↓ V notification of a low exhaled tidal volume condition.
TE MAND
alarm immediately to ensure prompt
6.5.24 LOW EXHALED SPONTANEOUS TIDAL VOLUME Alarm
The ⤓
SPONT
V
TE SPONT available.
alarm indicates the measured exhaled spontaneous tidal volume is less than or equal to the ↓ V
alarm setting. The alarm updates when a new measured value of exhaled spontaneous tidal volume is
TE
The ↓ V
TE SPONT
alarm can detect a leak in the patient circuit or a change in the patient’s respiratory drive during a single breath. The ↓ V
TE SPONT
alarm is based on the current breath rather than on an average to detect changes as quickly as possible. There are separate alarms for mandatory and spontaneous exhaled tidal volumes for use during SIMV and BiLevel. The ventilator phases in a change to the ⤓ V prompt notification of a low exhaled tidal volume condition.
TE SPONT
alarm limit immediately to ensure
6.5.25 LOW EXHALED TOTAL MINUTE VOLUME Alarm
The ↓ ⩒
E TOT
alarm indicates the measured minute volume (for mandatory and spontaneous breaths) is less than or equal to the volume.
↓ ⩒
E TOT
alarm setting. The ↓ ⩒
E TOT
alarm updates with each new calculation for exhaled minute
The ↓ ⩒
E TOT
alarm can detect a leak or obstruction in the patient circuit, a change in compliance or resistance, or a change in the patient’s breathing pattern. The ↓ ⩒
E TOT
alarm can also detect too-small tidal volumes, which could lead to hypoventilation and hypoxia (oxygen desaturation).
The ventilator phases in changes to the ↓ ⩒
E TOT prolonged low tidal volumes.
alarm limit immediately to ensure prompt notification of
6.5.26 PROCEDURE ERROR Alarm
The ventilator declares a PROCEDURE ERROR alarm if it is powered up (either by turning on the power switch or if power is regained following a power loss of at least 5 minutes) and the ventilator detects a patient attached before Ventilator Startup is complete. Until confirmation of the ventilator settings, the ventilator annunciates a
high-priority alarm and enters Safety PCV. See Table 53, page 236
.
The PROCEDURE ERROR alarm requires confirmation of ventilator settings after restoration of ventilator power, in case a new patient is attached to the ventilator. Safety PCV is an emergency mode of ventilation providing ventilation according to displayed settings until settings confirmation, and is not intended for long-term patient ventilation.
6.5.27 SEVERE OCCLUSION Alarm
A severe occlusion alarm occurs when gas flow in the ventilator breathing system is severely restricted. The ventilator enters occlusion status cycling (OSC) where the ventilator periodically attempts to deliver a
160
6 pressure-based breath while monitoring inspiration and exhalation breath phases for a severe occlusion. If an occlusion is not detected, the ventilator considers the occlusion condition reset, clears the occlusion alarm, and continues ventilation with the settings in use before the occlusion occurred. The ventilator indicates an occlusion was detected.
6.6 Monitored Patient Data
Monitored patient data appear in the Patient Data Banner at the top of the GUI screen above the waveforms
display. See Figure 33, page 92
. Where applicable, factory defaults are indicated.
See
Section 3.8.2.7, Vital Patient Data, page 79
to change the displayed patient data parameters or the order in which they are displayed.
If any patient data values are displayed continuously blinking, it means their values are shown clipped to what has been defined as their absolute limits. If the values are displayed in parentheses “()”, it means they are clipped to their variable limits. Variable limits are based on various patient and ventilator settings. Each of these data points should be viewed as suspect.
Dashes (--) are displayed if the patient data value is not applicable based on mode/breath type combinations.
Note: A blinking patient data value means that the displayed value is greater-than or less-than either of its absolute limits and has been “clipped” to its limit. A data value that appears in parentheses means it has questionable accuracy. If no value is displayed, then the ventilator is in a state where the value cannot be measured.
The following sections contain descriptions of all patient data parameters shown in the patient data displays.
Note: All displayed patient volume data represent lung volumes expressed under BTPS conditions.
6.6.1 Total Exhaled Minute Volume
Total exhaled minute volume ( ⩒
E TOT
) is the BTPS and compliance compensated sum of exhaled gas volumes from both mandatory and spontaneous breaths for the previous 1-minute interval. A factory default parameter.
6.6.2 Exhaled Spontaneous Minute Volume
Exhaled spontaneous minute volume ( ⩒
E SPONT
) is the BTPS- and compliance-compensated sum of exhaled spontaneous volumes for the previous minute. A factory default parameter.
6.6.3 Exhaled Tidal Volume
Exhaled tidal volume (V
TE breath. Displayed V
TE default parameter.
) is the volume of the patient’s exhaled gas for the previous mandatory or spontaneous
is both compliance-and BTPS compensated, and updates at the next inspiration. A factory
6.6.4 Proximal Exhaled Minute Volume
Proximal exhaled minute volume ( ⩒
E TOTY
) is the BTPS- and compliance-compensated sum of exhaled spontaneous volumes for the previous minute.
6.6.5 Proximal Exhaled Tidal Volume
Proximal exhaled tidal volume (V
TEY
) is the exhaled tidal volume for the previous breath measured by the proximal flow sensor (for neonatal patients, only). V
TEY
is updated at the beginning of the next inspiration.
6.6.6 Exhaled Spontaneous Tidal Volume
Exhaled spontaneous tidal volume (V
TE SPONT
) is the exhaled volume of the last spontaneous breath, updated at the beginning of the next inspiration following a spontaneous breath.
6
161
6.6.7 Exhaled Mandatory Tidal Volume
Exhaled mandatory tidal volume (V
TE MAND
) is the exhaled volume of the last mandatory breath, updated at the beginning of the next inspiration following a mandatory breath. If the mode is SPONT and the ventilator has not delivered mandatory breaths in a time period of greater than 2 minutes (for example via a manual inspiration), the
V
TE MAND
patient data indicator becomes hidden.The indicator reappears when the value updates. A factory default parameter.
6.6.8 Exhaled mL/kg Volume
The patient’s exhaled volume displayed in mL/kg PBW.
6.6.9 Inspired Tidal Volume
Inspired tidal volume (V
TI
) is the BTPS- and compliance-compensated volume of inspired gas for all pressure-based or NIV breaths, updated at the beginning of the following expiratory phase. V data are available. A factory default parameter.
TI
is displayed when
6.6.10 Proximal Inspired Tidal Volume
Proximal inspired tidal volume (V
TIY
) is the inspired tidal volume for a mandatory or spontaneous breath measured by the proximal flow sensor (for neonatal patients, only). V expiratory phase and is displayed when data are available.
TIY
is updated at the beginning of the following
6.6.11 Delivered mL/kg Volume
The delivered gas volume in mL/kg PBW.
6.6.12 I:E Ratio
The ratio of inspiratory time to expiratory time for the previous breath, regardless of breath type. Updated at the beginning of the next inspiration. When I:E ratio is ≥1:1, it is displayed as XX:1. Otherwise it is displayed as 1:XX. A factory default parameter.
Note: Due to limitations in setting the I:E ratio in PC ventilation, the monitored data display may not exactly match the I:E ratio setting.
6.6.13 Mean Circuit Pressure
Mean circuit pressure (P
MEAN
) is the average circuit pressure for a complete breath period, including both inspiratory and expiratory phases whether mandatory or spontaneous. The displayed value can be either positive or negative. A factory default parameter.
6.6.14 Peak Circuit Pressure
Peak circuit pressure (P
PEAK
) is the maximum circuit pressure at the patient wye during the previous breath, including both inspiratory and expiratory phases. A factory default parameter.
6.6.15 End Inspiratory Pressure
End inspiratory pressure (P
I END default parameter.
) is the pressure at the end of the inspiratory phase of the current breath. A factory
6.6.16 End Expiratory Pressure
End expiratory pressure (PEEP) is the pressure at the end of the expiratory phase of the previous breath, updated at the beginning of the next inspiration. During an expiratory pause, the displayed value includes any active lung
PEEP. A factory default parameter.
162
6
6.6.17 Intrinsic PEEP
Intrinsic PEEP (PEEP
I
) is an estimate of the pressure above the PEEP level at the end of an exhalation. PEEP
I
is determined during an expiratory pause maneuver.
6.6.18 PAV-based Intrinsic PEEP
PAV-based intrinsic PEEP (PEEP
I PAV breath.
) is an estimate of intrinsic PEEP, updated at the end of a spontaneous PAV+™
6.6.19 Total PEEP
Total PEEP (PEEP
TOT
) is the estimated pressure at the circuit wye during the expiratory pause maneuver.
6.6.20 Plateau Pressure
Plateau pressure (P
PL
) is the pressure measured and displayed during an inspiratory pause maneuver.
6.6.21 Driving Pressure (P
DRIVE
)
While ventilating with PAV+™, Driving Pressure (P
PEEP
TOT be calculated by dividing V
T
by C
PAV
DRIVE
and represents the amount of pressure required to overcome the elastic recoil of the lungs. P
(V
T
/C
PAV
).
) is the difference between Plateau Pressure and estimated
DRIVE
can also
PAV+™ software automatically performs a 300 end-inspiratory hold in a random pattern every four to ten PAV+™ breaths. Each time the end-inspiratory hold is conducted this pressure change is measured from the end expiratory lung pressure at the beginning of the breath to the pressure at the end of the plateau maneuver.
6.6.22 Total Respiratory Rate
Total respiratory rate (f
TOT
) is the total number of mandatory and spontaneous breaths per minute delivered to the patient. A factory default parameter.
6.6.23 PAV-based Lung Compliance
For a PAV+™ breath, PAV-based lung compliance (C each calculation. C
PAV
PAV
) is the change in pulmonary volume for an applied change in patient airway pressure, measured under zero-flow conditions and updated upon successful completion of
is displayed on the waveform screen.
6.6.24 PAV-based Patient Resistance
For a PAV+™ breath, PAV-based patient resistance (R in patient lung flow and updated upon successful completion of each calculation. R waveform screen.
PAV
) is the change in pulmonary pressure for an applied change
PAV
is displayed on the
6.6.25 PAV-based Lung Elastance
For a PAV+™ breath, PAV-based lung elastance (E
PAV completion of each calculation.
) is the inverse of C
PAV
and is updated upon successful
6.6.26 Spontaneous Rapid Shallow Breathing Index
Spontaneous rapid shallow breathing index ( f /V
T indicate the inverse. A factory default parameter.
) is an indication of the patient’s ability to breathe spontaneously.
High values generally mean the patient is breathing rapidly, but with low tidal volumes. Low values generally
6
163
6.6.27 Spontaneous Inspiratory Time Ratio
In SPONT mode, spontaneous inspiratory time ratio (T default parameter.
I
/T
TOT
) is the percentage of a spontaneous breath consumed by the inspiratory phase. Updated at the successful completion of a spontaneous breath. A factory
6.6.28 Spontaneous Inspiratory Time
Spontaneous inspiratory time (T
I SPONT
) is the duration of the inspiratory phase of a spontaneous breath and updated at the end of each spontaneous breath. T
I SPONT
is only calculated when the breathing mode allows spontaneous breaths and the breaths are patient-initiated. A factory default parameter.
6.6.29 PAV-based Total Airway Resistance
For a PAV+™ breath, PAV-based total airway resistance (R
TOT
) is the change in pulmonary pressure for an applied change in total airway flow and updated upon the successful completion of each calculation. If the R appears in parentheses as described at the beginning of this section, the R
TOT
PAV
value
value also appears in parentheses.
6.6.30 Static Compliance and Static Resistance
Static compliance (C
STAT
) is an estimate of the elasticity of the patient’s lungs, expressed in mL/cmH
2
O. It is computed during a mandatory breath.
Static resistance (R
STAT
) is the total inspiratory resistance across the artificial airway and respiratory system, displayed at the start of the next inspiration after the inspiratory pause maneuver. It is an estimate of how restrictive the patient’s airway is, based on the pressure drop at a given flow, expressed in cmH computed during a VC mandatory breath with a square flow waveform.
2
O/L/s. R
STAT
is
C
STAT
is calculated using this equation:
Figure 52.
Static Compliance (C
STAT
)
1 C
STAT
Static compliance
2 V pt
Total expiratory volume (patient and breathing circuit)
3 C ckt
Compliance of the breathing circuit during the pause maneuver (derived from SST)
4 P ckt
The pressure in the patient circuit measured at the end of the 100 ms interval defining the pause-mechanics plateau
5 PEEP The pressure in the patient circuit measured at the end of expiration
R
STAT
is calculated using this equation after C
STAT
is computed and assuming a VC breath type with a square waveform:
Figure 53.
Static Resistance (R
STAT
)
1 R
STAT
Static resistance
2 C ckt
Compliance of the breathing circuit during the pause maneuver (derived from SST)
3 ⩒ pt
Flow into the patient during the last 100 ms of the waveform
4 C
STAT
Static compliance
5 P
PL
Mean pressure in the patient circuit over the 100 ms interval defining the pause-mechanics plateau
6 P
PEAK
Peak circuit pressure
164
6
During the pause, the most recently selected graphics are displayed and frozen, to determine when inspiratory pressure stabilizes. C and take this format:
STAT
and R
STAT
are displayed at the start of the next inspiration following the inspiratory pause
C
STAT xxx or
R
STAT
yyy
Special formatting is applied if the software determines variables in the equations or the resulting C
STAT values are out of bounds:
or R
STAT
• Parentheses ( ) signify questionable C
STAT
or R
STAT
values, derived from questionable variables.
• Flashing C
STAT
or R
STAT
values are out of bounds.
• R
STAT
------ means resistance could not be computed, because the breath was not of a mandatory, VC type with square flow waveform.
6.6.31 Dynamic Compliance
Dynamic compliance (C
DYN
) is a dynamic estimate of static compliance for each mandatory breath delivered.
6.6.32 Dynamic Resistance
Dynamic resistance (R
DYN
) is a dynamic estimate of static resistance for each mandatory breath delivered.
6.6.33 C
20
/C
C
20
/C is the ratio of compliance of the last 20% of inspiration to the compliance of the entire inspiration.
6.6.34 End Expiratory Flow
End expiratory flow (EEF) is a measurement of the end expiratory flow for an applicable breath.
6.6.35 Peak Spontaneous Flow
Peak spontaneous flow (PSF) is a measurement of the maximum inspiratory spontaneous flow for an applicable spontaneous breath.
6.6.36 Displayed O
2
%
Displayed O
2
% is the percentage of oxygen in the gas delivered to the patient, measured at the ventilator’s outlet, upstream of the inspiratory filter. It is intended to provide a check against the set O not as a measurement of oxygen delivered to the patient. O
2
2
% for alarm determination, and
% data can be displayed as long as the O
2 is enabled. If the monitor is disabled, dashes (--) are displayed. If a device alert occurs related to the O
2 a blinking 0 is displayed.
monitor
monitor,
6.6.37 Inspiratory Time Constant (3Tau
I
)
The 3Tau
I
parameter is three times the product of the patient’s resistance and compliance, and is used to determine the adequacy of the set inspiratory time (pressure ventilation), or the inspiratory time determined by the flow pattern, tidal volume, (V
T
), peak flow ( ⩒
MAX
), and plateau time (T
PL
) settings (volume ventilation).
6
165
166
7
7 Preventive Maintenance
7.1 Overview
This chapter contains information on maintenance of the Puritan Bennett™ 980 Series Ventilator. It includes:
• How to perform routine preventive maintenance procedures, including frequency
• How to clean, disinfect, or sterilize the ventilator and main components
• How to store the ventilator for extended periods
• How to dispose of used parts
7.2 Ventilator Operational Time
The ventilator contains an hour meter that records the number of operational hours since the ventilator was manufactured. An additional timer tracks the number of hours since the last preventive maintenance activity was performed. Both the GUI and the status display show the number of hours before the next preventive maintenance is due.
7.3 Preventive Maintenance Intervals
Warning: To ensure proper ventilator operation, perform preventive maintenance intervals as specified in
and
.
Table 38.
Operator Preventive Maintenance Frequency
Part
Patient circuit: inspiratory and expir‐ atory limbs
Frequency
Several times a day or as required by the institution’s policy.
Condensate vial, water traps, and drain bag
Oxygen sensor calibration
Maintenance
• Check both limbs for water accumulation
• Empty and clean.
Check and empty as needed.
Inlet air filter bowl
Reusable inspiratory filter • Before every use
• After 15 days of continuous use in the inspiratory limb (replace)
• Yearly, or after 50 autoclave cycles (replace)
• Whenever excess resistance is suspected
From the ventilator setup screen, touch the More Settings tab. To calibrate the oxygen sensor, touch
Calibrate in the oxygen sensor area of the screen.
See Section 4.10.3, Oxygen Sensor
for information on testing the oxygen sensor calibration.
• Replace the bowl if it is cracked.
• If any sign of moisture is visible, remove the ventilator from use and contact service personnel.
• Inspect and replace if cracked, crazed, or damaged. Sterilize between patients and circuit changes, or according to the institution’s policy. Sterilize before non-destructive dis‐ posal, or dispose of filter
7
167
Table 38.
Operator Preventive Maintenance Frequency (continued)
Battery
Battery
Exhalation flow sensor assembly
(EVQ)
Part
Reusable exhalation filter
Disposable inspiratory filter
Disposable exhalation filter
Frequency
• Before every use
• After 15 days of continuous use in the exhalation limb (replace)
• Yearly or after 50 autoclave cycles (replace)
• Whenever excess resistance is suspected
After 15 days of continuous use (dis‐ card)
After 15 days of continuous use (dis‐ card)
When transferring battery to or from another ventilator
Every 3 years or as prompted
Per institutional guidelines, or if SST flow sensor cross check fails.
NOTE: the EVQ is removable and can be disinfected. DO NOT STER‐
ILIZE the EVQ.
Every 100 disinfection cycles. A dis‐ infection cycle is defined as one dis‐
infection event as described in Sec‐ tion 7.5.1, Exhalation Flow Sensor
Assembly (EVQ) Disinfection, page 172
.
Maintenance according to the institution’s policy.
• Run SST to check resistance of the inspiratory limb.
• Use care when changing inspir‐ atory filter to avoid filter dam‐ age and minimize the potential for introduction of particles.
• Inspect and replace if cracked, crazed, or damaged. Sterilize between patients and circuit changes, or according to the institution’s policy. Sterilize before non-destructive dis‐ posal, or dispose of filter according to the institution’s policy.
• Run SST to check resistance of the expiratory limb and filter.
• Use care when changing the exhalation filter to avoid filter damage and minimize the potential for introduction of particles.
Discard according to the institu‐ tion’s protocol.
Discard according to the institu‐ tion’s protocol.
Disinfect by wiping with a damp cloth using one of the solutions lis‐
ted. See Table 39, page 169 for
approved cleaning agents.
Replace
See Section 7.5, Component Cleaning and Disinfection, page 170 and
Sec‐ tion 7.5.1, Exhalation Flow Sensor
Assembly (EVQ) Disinfection, page 172
.
Replace. Discard used flow sensor according to the institutions.s pro‐ tocol. Run exhalation flow sensor calibration and SST.
168
7
Table 38.
Operator Preventive Maintenance Frequency (continued)
Part
Neonatal door/adapter
Compressor inlet air filter
Frequency
When gas pathway surfaces are visi‐ bly soiled or per institutional guide‐ lines.
When exterior surfaces of door are soiled.
Every 250 hours
Maintenance
Disinfect per
Wash in mild soapy water and rinse thoroughly. Let air dry.
Caution: Use specified cleaning, disinfection, and sterilization agents and procedures for the appropriate part as instructed.
7.4 Surface Cleaning of Exterior Surfaces
External surfaces of the GUI, BDU, and compressor base may become soiled and should be cleaned periodically.
To clean the GUI, BDU, or compressor base
1. Moisten a soft cloth with one of the disinfectants listed or use Sani Cloths (PDI, Inc.). See
.
2. Wipe the GUI, BDU, and compressor base, removing any dirt or foreign substances.
3. Dry all components thoroughly.
4. If necessary, vacuum any cooling vents on the GUI and BDU with an electrostatic discharge (ESD)-safe vacuum to remove any dust.
Table 39.
Surface Cleaning Agents
Part
Ventilator exterior (including touch screen and flex arm
Procedure
Wipe clean with a cloth dampened with one of the cleaning agents lis‐ ted or its equivalent. Use a damp cloth and water to rinse off chemical residue as necessary.
Mild dish washing detergent solu‐ tion
• Isopropyl alcohol (70% solu‐ tion)
• Bleach (10% solution)
• Window cleaning solution (iso‐ propyl alcohol and ammonia)
• Ammonia (15% solution)
• Hydrogen peroxide (3% solu‐ tion)
• Formula 409™* cleaner (Clorox
Company)
• CaviCide™* surface disinfec‐ tant (Metrex Research Corpora‐ tion)
• Control III™* germicide (Maril
Products, Inc.)
Comments and cautions
Do not allow liquid or sprays to pen‐ etrate the ventilator openings or cable connections.
Do not attempt to sterilize the ven‐ tilator by exposure to ethylene oxide (ETO) gas.
Do not use pressurized air to clean or dry the ventilator, including the
GUI cooling vents.
Do not submerge the ventilator or pour cleaning solutions over or into the ventilator.
7
169
Table 39.
Surface Cleaning Agents (continued)
Part
Ventilator cooling vents
Procedure
• Mr. Muscle™* Window & Glass
(SC Johnson)
• Sani Cloth™* (PDI, Inc.)
• [Propan-2-ol, Isopropanol, Iso‐
Vacuum the vents at the back of the
GUI and BDU to remove dust.
a Chemicals stated are the generic equivalents of Mr. Muscle™* Window & Glass.
N/A
Comments and cautions
7.5 Component Cleaning and Disinfection
Warning: To avoid microbial contamination and potential performance problems, do not clean, disinfect, or reuse single-patient use (SPU) or disposable components. Discard per local or institutional regulations.
Risks associated with reuse of single-patient use items include but are not limited to microbial cross-contamination, leaks, loss of part integrity, and increased pressure drop. When cleaning reusable components, do not use hard brushes or implements that could damage surfaces.
Table 40.
Component Cleaning Agents and Disinfection Procedures
EVQ
Part Cleaning agent and procedure
Before disinfection, presoak in
EMPower™* Dual Enzymatic Solu‐ tion (Metrex Inc.).
Perform high level disinfection using liquid chemical disinfectant using any of the following agents:
• Cidex™* (2.5%)
• Metricide 28™* 2.5%
• Cidex OPA™* (0.55%)
• Metricide™* OPA Plus (0.6%)
Neonatal door/adapter
• Sporox™* II (Sultan)
Follow the manufacturer’s instruc‐ tions
See
Section 7.5.1, Exhalation Flow
Sensor Assembly (EVQ) Disinfection, page 172 for specific instructions.
Before disinfection, presoak in
EMPower™* Dual Enzymatic Solu‐ tion (Metrex Inc.).
Perform high level disinfection using liquid chemical disinfectant using any of the following agents:
• Cidex™* (2.5%)
• Metricide 28™*
• Cidex™* OPA (0.55%)
• Metricide OPA™* Plus (0.6%)
Comments and cautions
Do not drop the EVQ or handle roughly during disinfection or storage
N/A
170
7
Table 40.
Component Cleaning Agents and Disinfection Procedures (continued)
Part
Reusable patient circuit tubing
Cleaning agent and procedure
Follow the manufacturer’s instruc‐ tions.
Disinfect per manufacturer’s instructions for use.
Comments and cautions
Disposable patient circuit tubing
Breathing circuit in-line water traps Disinfect per manufacturer’s
Breathing circuit components
Disposable drain bag and tubing
(single unit)
Inlet air filter bowl
Discard instructions for use.
Disinfect per manufacturer’s instructions for use.
Discard when filled to capacity or when changing patient circuit.
Wash the bowl with mild soap sol‐ ution, if needed.
• Inspect for nicks and cuts, and replace if damaged.
• Run SST to check for leaks when reinstalling the circuit or when installing a new circuit.
Discard per the institution’s proto‐ col.
• Inspect water traps for cracks and replace if damaged.
• Run SST to check for leaks when reinstalling the circuit or when installing new components.
• Inspect components for nicks and cuts, and replace if dam‐ aged.
• Run SST to check for leaks when reinstalling the circuit or when installing new components.
N/A
Battery
• Avoid exposing the inlet air fil‐ ter bowl to aromatic solvents, especially ketones.
• Replace if cracks or crazing are visible.
Do not immerse the battery or get the contacts wet.
Cooling fan filter
Other accessories
Wipe with a damp cloth using one of the cleaning agents listed. See
.
Clean every 250 hours or as neces‐ sary. Wash in mild soap solution, rinse, and air dry.
Follow manufacturer’s instructions for use.
N/A
N/A
To clean and disinfect parts
1. Wash parts in warm water using a mild soap solution.
2. Thoroughly rinse parts in clean, warm water (tap water is acceptable) and wipe dry.
cleaning and disinfection agents.
4. After the components are cleaned or disinfected, inspect them for cracks or other damage.
5. Dispose of damaged parts according to the institution’s policy.
Note: Steps 1 through 3 above do not apply to the EVQ. See
Section 7.5.1, Exhalation Flow Sensor Assembly
for disinfection instructions.
171
7
Whenever replacing or reinstalling a component, run SST before ventilating a patient.
7.5.1 Exhalation Flow Sensor Assembly (EVQ) Disinfection
Note: EVQ disinfection is not required on a routine basis but it should be disinfected if SST flow sensor cross check fails. See
for a list of appropriate disinfectants.
Note: Follow the institution’s infection control protocol for handling, storage, and disposal of potentially bio-contaminated waste.
Caution: To avoid damaging the hot film wire, do not insert fingers or objects into the center port when disinfecting the EVQ.
The EVQ contains the exhalation flow sensor electronics, exhalation valve diaphragm, exhalation filter seal, and pressure sensor filter. The exhalation flow sensor electronics consist of the hot film wire and the thermistor.
Because it is protected by the exhalation filter, it does not require or need replacement or disinfection on a regular basis. It is, however, removable and should be disinfected if the SST flow sensor cross check fails. Expected service life is 100 disinfection cycles.
Caution: To avoid damage to the exhalation flow sensor element:
• Do not touch the hot film wire or thermistor in the center port
• Do not vigorously agitate fluid through the center port while immersed.
• Do not forcefully blow compressed air or any fluid into the center cavity.
• Do not drop or handle roughly during disinfection or storage.
Warning: Damaging the flow sensor’s hot film wire or thermistor in the center port can cause the ventilator’s spirometry system to malfunction.
Figure 54.
EVQ
1 Top View
2 Bottom View
172
7
Figure 55.
EVQ Components
1 Hot film wire and thermistor
2 Diaphragm sealing surface
3 Electrical contacts
4 Filter grommet
7.5.1.1 Removal
Warning: Prior to cleaning and disinfection, remove and dispose of the disposable components of the EVQ.
To remove the EVQ
1. Lift up on the exhalation filter latch and open the exhalation filter door.
2. With thumb inserted into the plastic exhalation port and four fingers under the EVQ, pull it down until it snaps out. To avoid damaging the flow sensor element, do not insert fingers into the center port.
Figure 56.
EVQ Removal
7
To remove disposable components of the EVQ
1. Remove and discard the exhalation valve diaphragm, the exhalation filter seal, and the pressure sensor filter.
Lift the exhalation filter seal out of the exhalation flow sensor to remove it.
173
Figure 57.
Exhalation Valve Diaphragm Removal
Figure 58.
Exhalation Filter Seal Removal
174
Figure 59.
Pressure Sensor Filter Removal
7
2. Dispose of the removed items according to the institution’s protocol. Follow local governing ordinances regarding disposal of potentially bio-contaminated waste.
7.5.1.2 Disinfection
Warning: Do not steam autoclave the EVQ or sterilize with ethylene oxide gas. Either process could cause the ventilator’s spirometry system to malfunction when reinstalled in the ventilator.
Warning:
Covidien may damage the plastic enclosure or electronic sensor components, resulting in malfunction of the ventilator’s spirometry system.
Warning: Follow disinfectant manufacturer’s recommendations for personal protection (such as gloves, fume hood, etc.) to avoid potential injury.
down any bio-film that may be present. Follow manufacturer’s instructions regarding duration of soak process.
Caution: Do not use any type of brush to scrub the EVQ, as damage to the flow sensing element could occur.
2. Rinse in clean, deionized water.
3. Prepare the chemical disinfectant according to the manufacturer’s instructions or as noted in the institution’s protocol. See
for the proper disinfecting agents.
4. Immerse in the disinfectant solution, oriented as shown, and rotate to remove trapped air bubbles in its cavities. Keep immersed for the minimum time period specified by the manufacturer or as noted in the institution’s protocol.
175
7
Figure 60.
Immersion Method
5. At the end of the disinfecting immersion period, remove and drain all disinfectant. Ensure all cavities are completely drained.
7.5.1.3 Rinsing
Warning: Rinse according to manufacturer’s instructions. Avoid skin contact with disinfecting agents to prevent possible injury.
1. Rinse the EVQ using clean, deionized water in the same manner used for the disinfection step.
2. Drain and repeat rinsing three times with clean, deionized water.
3. After rinsing in deionized water, immerse in a clean isopropyl alcohol bath for approximately 15 seconds.
Slowly agitate and rotate to empty air pockets.
7.5.1.4 Drying
• Dry in a low temperature warm air cabinet designed for this purpose. Covidien recommends a convective drying oven for this process, with temperature not exceeding 60°C (140°F).
Caution: Exercise care in placement and handling in a dryer to prevent damage to the assembly’s flow sensor element.
7.5.1.5 Inspection
See
Figure 55, page 173 while inspecting the EVQ.
1. Inspect the plastic body, diaphragm sealing surface, filter grommet and the seal groove on the bottom side for any visible damage, degradation, or contamination.
2. Inspect electrical contacts for contaminating film or material. Wipe clean with a soft cloth if necessary.
3. Inspect the hot film wire and thermistor in the center port for damage and for contamination. DO NOT
ATTEMPT TO CLEAN EITHER OF THESE . If contamination exists, rinse again with deionized water. If rinsing is unsuccessful or hot film wire or thermistor is damaged, replace the EVQ.
176
7
7.5.2 EVQ Reassembly
Figure 61 shows the reprocessing kit:
Figure 61.
EVQ Reprocessing Kit
1 Diaphragm
2 Pressure sensor filter
3 Exhalation filter seal
1. After drying the EVQ, remove the pressure sensor filter from the reprocessing kit and install its large diameter into the filter grommet with a twisting motion until flush with the plastic valve body, as shown. The narrow end faces out.
Figure 62.
Installing the Pressure Sensor Filter
7
177
Figure 63.
Installing the Pressure Sensor Filter
2. Remove the exhalation filter seal from the kit and turn the assembly so its bottom is facing up.
3. Install the seal into the exhalation flow sensor as shown in Figure 64
. Ensure that the seal fits completely within the recess and sits flat.
Figure 64.
Installing the Exhalation Filter Seal
1 Exhalation filter seal
4. Remove the diaphragm from the kit and install it. See Figure 65 .
178
7
Figure 65.
Installing the Diaphragm
Figure 66.
Installing the Diaphragm
1 Diaphragm bead located in the EVQ’s groove
5. Carefully inspect component placement and the complete assembly.
7.5.3 EVQ Replacement
1. Replace the EVQ any time if cracked or damaged in use, or if a malfunction occurs during SST or EST.
2. Replace assembly if damage is noted to the hot film wire and thermistor in the center port.
3. Perform required calibrations. See
.
To install the EVQ into the ventilator
179
7
1. With the exhalation filter door open, insert the assembly directly under the exhalation valve and push
straight up until it snaps into place. See Figure 67, page 180
. To avoid damaging the hot film wire, do not insert fingers into any opening.
2. Install the exhalation filter by sliding it onto the tracks in the door, and orienting the filter’s from patient port through the hole in the door.
3. Close exhalation filter door and lower exhalation filter latch.
Figure 67.
Installing the EVQ
4. Calibrate the flow sensor.
7.5.4 Storage
1. Pre-test the EVQ before storage by installing it into the ventilator and running SST to test the integrity of the
breathing system. See Section 3.9.1.2, SST Test Sequence, page 83
.
2. After performing SST, remove the assembly and place it into a protective bag or similar covered container.
7.6 Component Sterilization
To sterilize parts
1. Sterilize per the component’s instructions-for use or the steam sterilization procedure described. See
.
2. After the components are sterilized, visually inspect them for cracks or other damage.
3. Dispose of damaged parts according to the institution’s policy.
Table 41.
Sterilization Parameters
Autoclave sterilization
Effective sterilization occurs by steam autoclaving at 132°C (170°F) for 15 minutes for gravity displacement cycles. Pre-vac sterilization of wrapped goods (132°C for 4 minutes) may also be used. Refer to pre-vac system manufacturer’s program parameters or follow the steam sterilizer manufacturer’s instructions.
1. Disassemble the component.
2. Clean the component, then steam autoclave*.
3. Wrap each component in muslin or equivalent paper for autoclaving.
4. Place the wrapped parts in the steam autoclave and sterilize.
5. Inspect the sterilized parts for damage, and discard if damaged.
180
7
Table 41.
Sterilization Parameters (continued)
Autoclave sterilization
6. Reassemble the component.
7. Install the component on the ventilator.
8. Run SST.
*If performing pre-vac sterilization, follow system manufacturer’s instructions for use (IFU).
Table 42.
Component Sterilization Procedures
Part
Reusable exhalation and inspiratory filters
Procedure
Steam autoclave per manufactur‐ er’s instructions-for-use
Exhalation filter condensate vial
Reusable drain bag tubing (short piece of tubing attached to drain bag) and clamp
Steam autoclave per manufactur‐ er’s instructions-for-use
Clean and autoclave the reusable tubing; clean the clamp. See
cleaning agents.
for approved
Comments/Cautions
• Do not chemically disinfect or expose to ETO gas.
• Check filter resistance using ventilator SST or other means before use.
• Follow the manufacturer’s rec‐ ommendations for reuse.
• Inspect the condensate vial for cracks after processing.
• Replace the condensate vial if damaged.
N/A
Whenever replacing or reinstalling a component, run SST before ventilating a patient.
7.7 Service Personnel Preventive Maintenance
Covidien recommends only qualified service personnel perform preventive maintenance activities summarized
. Complete details are described in the Puritan Bennett™ 980 Series Ventilator Service Manual .
At ventilator startup, and in Service mode, the GUI and status display indicate when there are 500 hours or less before preventive maintenance is due.
Table 43.
Service Preventive Maintenance Frequency
Frequency
Every 12 months
When ventilator location changes by 1000 feet of altitude
Entire ventilator
Part Maintenance
• Run Extended Self Test (EST).
Test alarm system.
• Run performance verification.
This includes running an elec‐ trical safety test and inspecting ventilator for mechanical dam‐ age and for label illegibility.
Primary and extended batteries Perform battery test (as part of EST and perform stand-alone battery test in Service mode).
Atmospheric pressure transducer Perform atmospheric pressure transducer calibration.
7
181
Table 43.
Service Preventive Maintenance Frequency (continued)
Frequency
Every 10 000 operational hours
Every year or as needed
Part
Internal inspiratory filter
Maintenance
Replace. Do not attempt to auto‐ clave or reuse.
BDU 10K hour kit, p/n 10097275 Install. See the Puritan Bennett™
980 Series Ventilator Service
Manual for information on tests required after installation of the 10K
PM Kit.
Compressor 10K hour kit, p/n
10097258
Oxygen sensor
Install. See the Puritan Bennett™
980 Series Ventilator Service
Manual for information on tests required after installation of the 10K
PM Kit.
• Replace the oxygen sensor as needed.
Every 7 years or sooner as needed Coin cell batteries on GUI and BD
CPU PCBAs
• Actual sensor life depends on operating environment. Oper‐ ation at higher temperature or
O
2
% levels will result in shorter sensor life.
Replace.
7.8 Safety Checks
Covidien factory-trained service personnel should perform extended self test (EST) on the ventilator after
servicing it at the intervals specified in Table 43
. See the Puritan Bennett™ 980 Series Ventilator Service Manual for details on performing EST.
7.9 Inspection and Calibration
Ventilator inspection and calibration should be performed by Covidien factory-trained service personnel at the intervals specified in
7.10 Documentation
Covidien factory-trained service personnel should manually enter the service date, time, and nature of repair/preventive maintenance performed into the log using a keyboard on the GUI.
To manually document a service or preventive maintenance activity
1. Enter Service mode.
2. Select the Logs tab.
3. Select the Service Log tab.
4. Select Add Entry, and using the buttons to the right of each line, complete the entry.
5. Touch Accept when complete.
7.11 Storage for Extended Periods
To store the ventilator
182
7
1. Clean the unit thoroughly.
2. Remove any batteries and accessories.
To return the ventilator to service
1. Replace batteries.
2. Recharge batteries prior to patient ventilation. If batteries are older than 3 years, use new batteries.
3. Perform EST and SST prior to patient ventilation.
7
183
184
8
8 Troubleshooting
8.1 Overview
This chapter contains information regarding ventilator logs on the Puritan Bennett™ 980 Series Ventilator.
Warning: To avoid a potential electrical shock, do not attempt to correct any electrical problem with the ventilator while it is connected to AC power.
8.2 Problem Categories
For the Puritan Bennett™ 980 Series Ventilator Operator’s Manual , troubleshooting is limited to responding to ventilator alarms and reviewing various ventilator logs. For detailed alarm information, including how to
service personnel who have attended the Medtronic training class for Puritan Bennett 980 Series Ventilators should consult the Puritan Bennett™ 980 Series Ventilator Service Manual for detailed repair information and ventilator diagnostic codes.
8.3 How to Obtain Ventilator Service
To obtain service for the ventilator, call Medtronic Customer Service at 1 800 255 6774 and follow the prompts.
8.4 Used Part Disposal
Follow local governing ordinances and recycling plans regarding disposal or recycling of device components.
Discard all damaged parts removed from the ventilator during the maintenance procedures according to your institution’s protocol. Sterilize contaminated parts before non-destructive disposal.
8.5 Ventilator Logs
The ventilator uses various logs to store event information for later retrieval when managing a patient’s treatment.
Some of the logs are accessible during ventilation and some logs are only available to Medtronic personnel when the ventilator is in Service mode. The Puritan Bennett™ 980 Series Ventilator
Service Manual gives more details regarding logs available to qualified service personnel.
When New Patient is selected during ventilator setup, patient data, ventilator settings, and alarm logs are cleared, but this information is available for service personnel review following New Patient selection when the ventilator is set up.
• Alarms Log — The alarm log records up to 1000 alarms that have occurred, whether they have been reset or autoreset, the priority level, and their analysis messages.
8
The alarm log is accessible during normal ventilation and in Service mode. A date- and time-stamped entry is made in the log whenever an alarm is detected, escalated, reset or autoreset. An entry is also made when an audio paused interval begins, ends, or is canceled. If one or more alarms have occurred since the last time the alarm log was viewed, a triangular icon appears on the GUI indicating there are unread items. The alarm log is stored in non-volatile memory (NVRAM) and may be re-displayed after the ventilator’s power is cycled.
If the ventilator enters BUV for any reason, this is also entered into the alarm log. The alarm log is cleared by setting the ventilator up for a new patient.
• Settings Log — The settings log records changes to ventilator settings for retrospective analysis of ventilator-patient management. The time and date, old and new settings. and alarm resets are recorded. A
185
maximum of 500 settings changes can be stored in the log. The settings log is cleared when the ventilator is set up for a new patient. The settings log is accessible in normal ventilation mode and Service mode.
• Patient Data Log — This log records every minute (up to 4320 patient data entries) consisting of date and time of the entry, patient data name, and the patient data value during ventilator operation. It is cleared when the ventilator is set up for a new patient. Three tabs are contained in the patient data log:
– Vital Patient data — The log contains the same information that the clinician has configured in the patient data banner at the top of the GUI. If the patient data parameters in the banner are changed, these changes are reflected the next time the patient data log is viewed.
– Additional Patient Data – 1 — This log corresponds to the patient data parameters set on page 1 of the additional patient data banner. A total of 15 parameters are stored here, consisting of date and time of the entry (recorded every minute), patient data name, and the patient data value during ventilator operation.
– Additional Patient Data – 2 — This log corresponds to the patient data parameters set on page 2of the additional patient data banner. A total of ten parameters are stored here, consisting of date and time of the entry (recorded every minute), patient data name, and the patient data value during ventilator operation.
• Diagnostic Log — The Diagnostic Log is accessible during normal ventilation and Service modes and contains tabs for the System Diagnostic Log (default), the System Communication Log, and the EST/SST
Diagnostic Log. The diagnostic log contains tabs for the following:
– System Diagnostic Log — The System Diagnostic Log contains the date and time when an event occurred, the type of event, the diagnostic code(s) associated with each fault or error that occurred, the type of error that occurred, and any notes. See the Puritan Bennett™ 980 Series Ventilator Service Manual
(10078090) for specific information contained in the System Diagnostic Log. The diagnostic log is not cleared when the ventilator is set up for a new patient.
– System Communication Log — This log contains information generated by the ventilator’s communication software. See the Puritan Bennett™ 980 Series Ventilator Service Manual (10078090) for specific information contained in the System Communication Log.
– EST/SST Diagnostic Log — The EST/SST diagnostic log displays the time, date, test/event, system code
(see the Puritan Bennett™ 980 Series Ventilator Service Manual), type, and notes.
• EST/SST status log — The EST and SST status log displays the time, date, test/event, test status (passed or failed).
• General Event log — The general event log contains ventilator-related information not found in any other logs. It includes date and time of compressor on and off, changes in alarm volume, when the ventilator entered and exited Stand-By, GUI key presses, respiratory mechanics maneuvers, O
2 connection, elevate O
2 is not cleared upon new patient setup.
calibration, patient
, and warning notifications. The General event log can display up to 256 entries and
• Service Log — The service log is accessible during normal ventilation and Service modes and contains the nature and type of the service, reference numbers specific to the service event (for example, sensor and actuator ID numbers), manual and automatic serial number input, and the time and date when the service event occurred. It is not cleared upon new patient setup.
To view ventilator logs
1. Touch the clipboard icon in the constant access icon area of the GUI.
The log screen appears with tabs for the various logs.
2. Touch the tab of the log desired.
3. View the information for each parameter desired.
186
Figure 68.
Log Screen
8
1 Individual logs tabs
2 Pages contained in the log being viewed
Ventilator logs can be saved by entering Service mode, and downloading them via the ethernet port. See the
Puritan Bennett™ 980 Series Ventilator Service Manual for instructions on downloading ventilator logs.
8.6 Diagnostic Codes
Refer to the diagnostic log for the codes generated during patient ventilation. For more information on the diagnostic codes, contact Medtronic Technical Support.
8
187
188
9
9 Accessories
9.1 Overview
The following commonly available accessories from the listed manufacturers can be used with the ventilator system:
Filters – DAR/Covidien, Puritan Bennett
Heated Humidification Systems – Fisher & Paykel
Patient Circuits – commonly available breathing circuits with standard ISO 15 mm/ 22 mm connection for neonatal, pediatric, and adult patients. Manufacturers include Fisher & Paykel, DAR, and Intersurgical
Masks – ResMed, Respironics, Fisher & Paykel
Patient Monitoring Systems – See
Section 5.9, Connectivity to External Systems, page 130
for information on which systems can be used with the ventilator
Nasal Interfaces – Fisher & Paykel, Argyle
Compressed air filter and water trap – Covidien
Warning: The Puritan Bennett™ 980 Series Ventilator contains phthalates. When used as indicated, very limited exposure to trace amounts of phthalates may occur. There is no clear clinical evidence that this degree of exposure increases clinical risk. However, in order to minimize risk of phthalate exposure in children and nursing or pregnant women, this product should only be used as directed.
9.2 General Accessory Information
The patient circuit support arm (flex arm) can be fastened to the ventilator handle on either the right or left side.
Flex arms used on the Puritan Bennett™ 840 Ventilator System can also be used on the Puritan Bennett™ 980
Ventilator System.
9
189
Figure 69.
Ventilator with Accessories
Figure 70.
Additional Accessories
See
and
for the parts listed in Table 44
.
Note: Occasionally, part numbers change. If in doubt about a part number, contact your local Medtronic representative.
190
9
Table 44.
Accessories and Options
Item num‐ ber
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Accessory or option description
Test lung
Drain bag tubing (package of 10)
Drain bag (package of 25)
Drain Bag tubing clamp, reusable (package of 5)
Pediatric–adult exhalation filter
Pediatric–adult exhalation filtration system (carton of 12)
980 FRU, exhalation flow sensor
Wall air water trap
Power cord, 10A, RA, ANZ
Air hose assembly; Australia, New Zealand
Oxygen hose assembly; Australia, New Zealand
Cylinder mount for compressed Air and O
2
gas
Flex arm assembly
FRU, caster base assembly
Rechargeable lithium ion battery
Humidifier bracket
Drain bag clip
Inspiratory bacteria filter, reusable (Re/Flex)
16
17
18
Inspiratory bacteria filter, disposable (carton of 50) (DAR)
Condensate vial, reusable
Condensate vial drain cap
Patient circuit, adult dual heated wire, disposable, for F&P
MR850 –(Medtronic /DAR) use with adapter 111/1149
Patient circuit, single heated wire, adult, disposable, for F&P
MR850–(Medtronic / DAR) use with adapter 111/1146
Ventilator breathing circuit, adult, dual heated system, disposable, Fisher &
Paykel—(Fisher & Paykel)
19
Assy, patient circuit, with single water trap, heated insp. limb, pediatric, dis‐ posable for F&P MR850–(Medtronic / DAR) use with adapter 111/1146
Assy, patient circuit, dual heated wire, pediatric, disposable, F&P MR850—
Assy, patient circuit, neonatal, single heated wire, disposable, incubator use, for F&P MR850—(Medtronic / DAR) use with adapter 111/1146
Assy, patient circuit, neonatal, single heated wire, disposable, not for incu‐ bator use, for F&P MR850–(Medtronic / DAR) use with adapter 111/1146
20
21
Ventilator breathing circuit, neonatal, dual heated system, disposable,
Fisher & Paykel—(Fisher & Paykel)
O-ring seal, condensate vial, reusable
Neonatal expiratory filtration system, disposable, with condensate vial
22 Proximal Flow monitoring sensor (disposable, 10/box)
Not shown Exhalation valve module reprocessing kit (6/ carton)
Part number
10005490
4-048493-00
4-048491-00
4-048492-00
10063033
10043551
10097468
10086051
GR106800
4-074712-00
4-074711-00
10086050
4-032006-00
980CASBAS
10086042
10086049
10087137
4-074600-00
351U5856
10063031
4-074613-00
304S14300
304S14402Z
RT280
306S8987
5505850
307S9910
307/8682
RT265
10085527
4-076900-00
10047078
10086048
191
9
Table 44.
Accessories and Options (continued)
Item num‐ ber
Accessory or option description
Not shown Gold standard test circuit, 21 inch (for performing EST)
Not shown End tidal CO
2
monitoring option
Not shown Integrated Nebulizer Upgrade Option
Not shown PB980 Exhaust Port Adapter
Hardware options
Not shown Proximal Flow monitoring option
Not shown 980, USB flash drive
Not shown NeoMode 2.0 Software Upgrade
Not shown IE Sync Trigger Upgrade Option
Software options
Not shown High Flow O2 Therapy Upgrade Option
Not shown NIV plus Software Upgrade Option
Part number
4-018506-00
10084332
980NEB
980EVQADAPT-30
10084331
PT00011076
PT00081810
980HFO2T
980IESYNC
980NIVPLUS a Reusable filtration system does not include condensate vial. Reusable condensate vial must be ordered separately.
b The part numbers listed reflect the breathing circuit manufacturer part numbers and are subject to change. Refer to the breathing circuit manufacturer for exact circuit details regarding ordering information.
192
10
10 Theory of Operations
10.1 Overview
This chapter provides specific details on breath delivery functions of the Puritan Bennett™ 980 Series Ventilator.
The chapter is organized as shown.
Section number
10.1
10.2
10.3
10.4
10.5
10.6
10.7
10.8
10.9
10.10
10.11
10.12
10.13
10.14
10.15
10.16
10.17
10.18
10.19
Title and Page
Section 10.2, Theoretical Principles, page 194
Section 10.3, Applicable Technology, page 194
Section 10.4, Inspiration — Detection and initiation, page 195
Section 10.5, Exhalation — Detection and Initiation, page 198
Section 10.6, Compliance and BTPS Compensation, page 199
Section 10.7, Mandatory Breath Delivery, page 202
Section 10.8, Spontaneous Breath Delivery, page 205
Section 10.9, A/C Mode, page 211
Section 10.10, SIMV Mode, page 213
Section 10.11, Spontaneous (SPONT) Mode, page 216
Section 10.12, Apnea Ventilation, page 217
Section 10.13, Detecting Occlusion and Disconnect, page 220
Section 10.14, Respiratory Mechanics, page 222
Section 6.4, Ventilator Settings, page 133
Section 10.16, Safety Net, page 235
Section 10.17, Power On Self Test (POST), page 238
Section 10.18, Short Self Test (SST), page 239
Section 10.19, Extended Self Test (EST), page 239
Warning: The ventilator offers a variety of breath delivery options. Throughout the patient’s treatment, the clinician should carefully select the ventilation mode and settings to use for that patient based on clinical judgment, the condition and needs of the patient, and the benefits, limitations and characteristics of the breath delivery options. As the patient’s condition changes over time, periodically assess the chosen modes and settings to determine whether or not those are best for the patient’s current needs.
The gas supplies to which the ventilator are connected must be capable of delivering 200 L/min flow with the proper supply pressure between 35 psig and 87 psig (241.8 kPa to 599.8 kPa). These supplies may be compressed air from an external source (wall or bottled) air or oxygen. (An optional compressor is available to be used as an external air source.)
Air and oxygen hoses connect directly to the rear of the breath delivery unit (BDU). The flow of each gas is metered by a Proportional Solenoid (PSOL) valve to achieve the desired mix in the Mix Module. The flow through each PSOL is monitored by separate flow sensors to ensure the accuracy of the mix. The mixed gases then flow to the
Inspiratory Module.
The blended gas in the Inspiratory Module is metered by the Breath Delivery PSOL and monitored by the Breath
Delivery Flow Sensor to ensure that the gas is delivered to the patient according to the settings specified by the operator. Delivered tidal volumes are corrected to standard respiratory conditions (BTPS) to ensure consistent interpretation by the clinician. The Inspiratory Module also incorporates the Safety Valve, which opens to vent excess pressure and allows the patient to breathe room air (if able to do so) in the event of a serous malfunction.
10
193
An optional compressor, capable of delivering flows of 140 L/min (BTPS) and minute volumes of up to 40 L/min
(BTPS), can be connected to the ventilator. Gas mixing occurs in the accumulator, protected by a relief valve. A one-way valve allows a maximum reverse flow into the gas supply system up to 100 mL/min under normal conditions.
Air and O
2
gases travel through proportional solenoid valves (PSOLs), flow sensors, and one-way valves, and are mixed in the mix module (according to the operator-set O
2
concentration), which also has a pressure-relief valve.
From here, the gas flows through another PSOL, to the inspiratory pneumatic system, where it passes by a safety valve, then through a one-way valve, an internal bacteria filter, an external bacteria filter, through the humidifier, if used, and then to the patient via the connected breathing circuit.
During exhalation, the gas flows through the expiratory limb of the breathing circuit, through a condensate vial, a bacteria filter, through the exhalation flow sensor, through the exhalation valve, and out the exhaust port . The exhalation valve actively controls PEEP while minimizing pressure overshoots and relieving excess pressures.
Pressure transducers in the inspiratory pneumatic system (PI) and exhalation compartment (PE) monitor pressures for accurately controlling breath delivery.
10.2 Theoretical Principles
This theory of operations is described mainly from a clinical standpoint, discussing how the ventilator responds to various patient inputs, but also including a general description of the ventilator’s components and how they work together to manage breath delivery.
10.3 Applicable Technology
The ventilator’s control is provided by Breath Delivery (BD) and Graphical User Interface (GUI) Central Processing
Units (CPUs). The BD CPU manages all breath delivery functions and provides background checks on the subsystems required for breath delivery. The GUI CPU controls the primary display, operator input devices, and the alarm system. The status display, a small, non-interactive LCD display located on the Breath Delivery Unit (BDU) is controlled by its own processor. See
Section 2.11.1.3, Status Display, page 41
for more information.
USB, Ethernet, and HDMI interfaces are provided on the ventilator. The USB interface supports items such as transferring data to an external monitor via a serial over USB protocol and saving screen captures to a memory
serial-over-USB data transfer. The Ethernet interface is used by qualified service personnel for accessing ventilator logs and performing software options installation, and the HDMI interface provides the ability to display the GUI screen on an external video display device.
Pressure and flow sensors in the inspiratory and expiratory modules manage breath delivery processes. Sensor signals are used as feedback to the breath delivery PSOL and exhalation valve controllers. Additional flow and pressure sensors are used in the mix module to control the breathing gas composition. In addition, gas temperature is measured for temperature compensation of flow readings. Atmospheric pressure is measured in the inspiratory module and used for BTPS compensation. The sensor signals are filtered using anti-aliasing filters and sampled with A/D converters. Additional low-pass filters precondition the signals, the signals are then used for controls and display purposes.
Closed-loop control is used to maintain consistent pressure and flow waveforms in the face of changing patient/system conditions. This is accomplished by using the output as a feedback signal that is compared to the operator-set input. The difference between the two is used to drive the system toward the desired output. For example, pressure-control modes use airway pressure as the feedback signal to control gas flow from the ventilator. See the following figure. This diagram shows a schematic drawing of a general feedback control system.
The input is a reference value (e.g., operator preset inspiratory pressure) that is compared to the actual output value (e.g., instantaneous value of airway pressure). The difference between those two values is the error signal.
The error signal is passed to the controller (e.g., the software control algorithm). The controller converts the error signal into a signal that can drive the actuator (e.g., the hardware drivers and valves) to cause a change in the manipulated variable (e.g., inspiratory flow).
194
Figure 71.
General Feedback Control System
10
Note: In the diagram, the “plant” is the patient and the connected breathing circuit.
10.4 Inspiration — Detection and initiation
When ventilator inspiration occurs, it is called triggering. Breaths are delivered to the patient based on ventilator settings the practitioner has entered and are determined by pressure, flow, or time measurements, or operator action. The ventilator uses the following methods to trigger an inspiration:
• Pressure triggering (P
TRIG
)
• Flow triggering ( ⩒
TRIG
)
• Time-triggered
• Operator-initiated
If the ventilator detects a drop in pressure at the circuit wye or when there is a decrease in base flow measured at the exhalation valve, the patient is said to trigger the breath. Mandatory breaths triggered by the patient are referred to as PIM or patient-initiated mandatory breaths.
All spontaneous breaths are patient-initiated, and are also triggered by a decrease in circuit pressure or measured base flow indicating the patient is initiating an inspiration.
Another term, autotriggering , is used to describe a condition where the ventilator triggers a breath in the absence of the patient’s breathing effort. Autotriggering can be caused by inappropriate ventilator sensitivity settings, water in the patient circuit, or gas leaks in the patient circuit.
10.4.1 Pressure Triggering
If pressure triggering (P
TRIG
) is selected, the ventilator transitions into inspiration when the pressure at the patient circuit wye drops below positive end expiratory pressure (PEEP) minus the operator-set sensitivity level (P
SENS
). See
. As the patient begins the inspiratory effort and breathes gas from the circuit (event 5, the A–B
), pressure decreases below PEEP. When the pressure drops below PEEP minus P
SENS
(event 6), the ventilator delivers a PIM breath. The pressure-decline time interval between events A and B determines how aggressive the patient’s inspiratory effort is. A short time interval signifies an aggressive breathing effort. The A–B interval is also affected by P
SENS
. A smaller P
SENS
setting means a shorter A–B time interval. (The minimum P setting is limited by autotriggering, and the triggering criteria include filtering algorithms that minimize the
SENS probability of autotriggering).
10
195
Figure 72.
Inspiration Using Pressure Sensitivity
1 Exhalation
2 Inspiration
3 Event A: (patient inspires)
4 Event B: patient-triggered inspiration begins
5 A–B interval
6 Operator-set pressure sensitivity
10.4.2 Flow Triggering
If flow triggering ( ⩒
TRIG
) is selected the BDU provides a constant gas flow through the ventilator breathing circuit
(called base flow) during exhalation. The base flow is 1.5 L/m greater than the value selected for flow sensitivity
( ⩒
SENS
). See
where the top graphic represents expiratory flow and the bottom graphic represents inspiratory flow.
The ventilator’s breath delivery flow sensor measures the base flow delivered to the circuit and the exhalation flow sensor measures the flow entering the exhalation valve. The ventilator monitors patient flow by measuring the difference between the inspiratory and exhaled flow measurements. If the patient is not inspiring, any difference in measured flows is due to leaks in the breathing system or flow sensor inaccuracy. The clinician can compensate for leaks in the breathing system by increasing ⩒
SENS
to a value equal to desired ⩒
SENS
+ leak flow.
As the patient begins the inspiratory effort and inspires from the base flow, less exhaled flow is measured, while the delivered flow remains constant. See
Figure 73 , (event A). As the patient continues to inspire, the difference
between the delivery and exhalation flow sensor measurements increases. The ventilator initiates an inspiration when the difference between the two flow measurements is greater than or equal to the operator-set flow
sensitivity value. See Figure 73
, (event B).
As with pressure triggering, the time delay between onset of the patient’s effort and actual gas delivery depends on:
• how quickly the exhaled flow declines (that is, the aggressiveness of the inspiratory effort). The more aggressive the inspiratory effort, the shorter the interval, and
• the flow sensitivity value. The smaller the value, the shorter the delay.
During flow triggering, a backup pressure sensitivity of 2 cmH
2 that the flow trigger fails.
O is present to detect a breath trigger in the event
196
Figure 73.
Inspiration Using Flow Sensitivity
10
1 Software-set base flow (L/min)
2 Start of patient effort
3 Event A: flow is decreasing
4 Event B: gas delivery begins
5 Operator-set flow sensitivity
6 1.5 L/min
7 Flow delivered to patient
10.4.3 Time Triggers
The ventilator measures the time interval for each breath and breath phase. If the ventilator is in Assist/Control
(A/C) mode (where the ventilator delivers breaths based on the breath rate setting), a VIM or ventilator initiated mandatory breath is delivered after the appropriate time interval. The duration of the breath in seconds ( T b
) is 60/f.
Figure 74.
Breath Activity During Time-triggered Inspiration
10
1 Breath activity (VIM)
2 Breath activity (PIM)
3 Time period (Tb) = (60/ f )
10.4.4 Operator-initiated Triggers
If the operator presses the manual inspiration key, an OIM (operator-initiated mandatory) breath is delivered.
The ventilator will not deliver an OIM under the following conditions:
• During an active inspiration, whether mandatory or spontaneous
• During the restricted phase of exhalation
• During circuit disconnect and occlusion status cycling (OSC) conditions
See
later in this chapter for information on the restricted phase of exhalation.
197
10.5 Exhalation — Detection and Initiation
When exhalation occurs, it is called cycling. Mandatory breaths can be volume-cycled or time-cycled by the ventilator or pressure cycled by the patient. Spontaneous breaths can be flow-cycled or pressure-cycled by the patient or time-cycled by the ventilator. A patient-cycled exhalation relies on measurements such as inspiratory flow rate or airway pressure. The ventilator uses the three methods described to detect exhalation:
• Airway pressure method (spontaneous breaths)
• Percent peak flow method (spontaneous breaths)
• Time-cycling method (mandatory breaths)
10.5.1 Airway Pressure Method
If expiratory sensitivity (E
SENS
) is set to a value too low for the patient-ventilator combination, a forceful expiratory effort could cause circuit pressure (P
PEAK
) to rise to its limit. The ventilator monitors circuit pressure throughout the inspiratory phase, and initiates an exhalation when the pressure equals the inspiratory pressure (P
I volume support (VS).
) target value + an incremental value. This transition to exhalation occurs during spontaneous pressure-based ventilation and in
Note: The allowable incremental value above the target pressure is 1.5 cmH
2
O once a portion of inspiration time
(Tn) has elapsed. Before Tn, the incremental value is higher to allow for transient pressure overshoots. For the first
200 ms of inspiration, the incremental pressure is 10% of the target pressure, up or 8 cmH
2
O, whichever is greater.
From 200 ms to Tn, the incremental pressure decreases in a linear fashion from the initial value to 1.5 cmH
2
O.
Figure 75.
Exhalation via the Airway Pressure Method
1 Pressure target
2 Pressure target +incremental value (n)
3 Start breath
4 200 ms
5 Tn
10.5.2 Percent Peak Flow Method
For spontaneous breath types including PS (pressure supported), TC (tube compensated), and VS (volume supported), the ventilator captures the value of the delivered peak inspiratory flow, then monitors the inspiratory flow decline until the value of current flow to peak flow (expressed as a percentage) is less than or equal to the set
E
SENS
value. The ventilator then cycles from inspiration into exhalation.
See
for an example of exhalation using the percent peak flow method.
198
Figure 76.
Typical Flow-cycled breath
10
1 Flow (y-axis)
2 Peak flow
3 E
SENS
= a % of peak flow
4 Inspiration
5 Exhalation
6 Time (x-axis)
Note: PAV+™ uses a flow-based cycling method, also called E
SENS
⩒
MAX
.
but it is expressed in L/min rather than in % of
10.5.3 Time-cycling Method
In pressure ventilation, the set inspiratory time (T ventilation, T
I
depends on the tidal volume (V
T
I
) defines the duration of the inspiratory phase. In volume
) setting, peak flow ( ⩒
MAX
), flow pattern, and plateau time (T
PL
). The ventilator cycles into exhalation when the set T
I
(pressure ventilation) or computed T
I
(volume ventilation) lapses.
10.5.4 Backup Methods
There are four backup methods for preventing excessive duration or pressure during inspiration.
Time limit – For adult and pediatric patients, the time limit method ends inspiration and begins exhalation when the duration of a spontaneous inspiration is greater than or equal to [1.99+0.02 × PBW (kg)] s.
High circuit pressure limit – During any type of inspiration, inspiration ends and exhalation begins when the monitored airway pressure (P
CIRC
) is greater than or equal to the set high circuit pressure limit.
High ventilator pressure limit – The ventilator transitions from inspiration to exhalation if the high ventilator pressure ( ⤒ P
VENT
) limit of 110 cmH
2
O is reached.
High inspired tidal volume limit – The high inspired tidal volume limit terminates inspiration and commences exhalation during VC+, VS, tube compensated (TC), or proportionally assisted (PAV+™) breaths if the delivered volume is greater than or equal to ⤒ V
TI
.
Note: The ventilator does not generate subatmospheric airway pressures during exhalation.
10.6 Compliance and BTPS Compensation
10.6.1 Compliance Compensation in Volume-based Breaths
Compliance compensation accounts for the gas volume not actually delivered to the patient during inspiration.
This gas is known as the compliance volume, VC. VC is the gas lost to pressurizing the breathing circuit and includes the volumes of the patient circuit, any accessories such as a humidifier and water traps, and internal ventilator gas passages.
10
199
Figure 77.
Square Flow Pattern
1 Flow (y-axis)
2 Actual ⩒
MAX
3 Set ⩒
MAX
Figure 78.
Descending Ramp Flow Pattern
4 Compliance volume (VC)
5 Set V
T
6 T
I
1 Flow (y-axis)
2 Actual ⩒
MAX
3 Set ⩒
MAX
4 Compliance volume (VC)
5 Set V
T
6 T
I
7 Minimum ⩒
MAX
In the ventilator, an iterative algorithm automatically computes the compliance volume. There is a maximum tubing-to-patient compliance ratio to reduce the potential for over-inflation due to an erroneous patient compliance estimation. The maximum ratio is determined by the selected patient circuit type and predicted body weight (PBW):
200
10
Factor Compliance volume factor
C pt ckt
Compliance of the patient circuit
The compliance volume is calculated as
C pt
Compliance of the patient
V
C
C pt ckt
Compliance volume
Compliance of the patient circuit
P
P wye
Pressure at the patient wye at the end of the current inspiration
Pressure at the end of the current exhalation
Without automated compliance compensation, practitioners would have to compute volume in the patient circuit, then increase the V
T
V
C increment to compensate for compliance volume provides only partial compensation, and requires extra effort and understanding by the practitioner. Additionally, P wye
and P can change with time.
to estimate the loss of
setting by that amount. Increasing the tidal volume by a single
An iterative algorithm in the ventilator automatically computes the compliance volume and compensates for it.
Compliance compensation does not change inspiratory time (T amplitude of the selected flow pattern). Keeping T
I
I
). It is achieved by increasing flow (increasing the
constant maintains the original I:E ratio.
There is a maximum compliance volume to reduce the potential for overinflation due to an erroneous compliance volume calculation. The maximum compliance volume is determined by the selected patient circuit type and predicted body weight (PBW), and is summarized by this equation:
V comp,max
= Factor × Tidal volume where:
V comp,max
= maximum compliance volume
Factor = linear interpolation of the values in the following table for adult, pediatric, and neonatal circuit types.
Factor is calculated as:
MIN (10, MAX (2.5, 1.0 + (2.0/0.3 × kg PBW)))
Table 45.
Compliance Volume Factors
Adult patient circuit type
PBW (kg) Factor
-
25
30
3.8
3.4
60
150
2.75
2.5
Pediatric patient circuit type
PBW (kg) Factor
≤10 5
11
12.5
3.5
2.9
15
24
2.7
2.58
Neonatal patient circuit type
PBW (kg) Factor
≤ 0.7
1
2
3
≥4.5
10
7.67
4.33
3.22
2.5
Note: Compliance compensation calculations are also in effect during exhalation to ensure spirometry accuracy.
If the patient’s compliance decreases beyond the limits of compliance compensation, the ventilator relies on the
⤒ P
PEAK
alarm setting to truncate the breath and switch to exhalation.
10
201
10.6.2 BTPS Compensation in Volume-based Breaths
Volumes and flows are BTPS compensated, that is, they are reported by the ventilator at existing barometric pressure, 37°C (98.6°F), and fully saturated with water vapor.
10.7 Mandatory Breath Delivery
Three mandatory breath types are offered in the ventilator — volume control (VC) which bases breath delivery on the delivered inspiratory tidal volume, pressure control (PC), which bases breath delivery on achieving and sustaining a pressure target for a set period of time, and volume control plus (VC+) which is a pressure-controlled breath based on a target tidal volume. VC+ can be used in situations where a patient’s lungs become more compliant due to treatment as it reduces the target pressure (lessening the forces on the alveoli) to achieve the target tidal volume.
Mandatory breaths are delivered by the ventilator, are either assisted (if patient initiated or PIM), or controlled (if ventilator initiated or VIM), or initiated by the operator (OIM). In A/C mode, the breath period ( T using the breath rate ( f ) according to the equation b
) is calculated
If, during T b
, patient effort is detected, a PIM breath is initiated and a new breath period starts. If no patient effort is detected before T b
lapses, the next breath delivered is a VIM, and a new breath period starts.
See
for details on the following VC+ settings:
• Expiratory time (T
E
• I:E ratio
)
• Inspiratory time (T
I
)
• Rise time%
• Target or tidal volume (V
T
)
VC and PC breath types require no initialization. A VC breath is based on meeting a delivered volume target and a PC breath is based on meeting a specific pressure target. VC+ breaths, however, go through a startup routine.
10.7.1 Volume Control (VC)
Volume Control is the control scheme that controls the flow with for the purpose of supplying a predetermined volume (set by the practitioner) to the patient. There are two basic flow wave forms to administer this volume: the
“square” that guarantees a constant flow during the inspiration time, or the “descending ramp” whose slope and initial value are determined to provide the required volume target. See
Figure 79 and Figure 80, page 203
. The inspiration time is determined indirectly by the characteristics of the selected flow wave.
202
Figure 79.
Ideal Waveform Using Square Flow Pattern
1 Pressure (cmH
2
O)
2 Flow (L/min)
3 Volume (mL)
4 Inspiratory phase
5 Expiratory phase
6 Constant flow
Figure 80.
Ideal Waveform Using Descending Ramp Flow Pattern
10
1 Pressure (cmH
2
O)
2 Flow (L/min)
3 Volume (mL)
4 Inspiratory phase
5 Expiratory phase
6 Descending ramp
10.7.2 Pressure Control (PC)
Pressure Control is the control scheme by which the pressure is controlled at the circuit wye to reach a constant level (set by the practitioner) during inspiration, and a PEEP level during exhalation. See
. This level is maintained for a time given by the set inspiration time, following followed by an exhalation regulated by the exhalation valve until the PEEP level is reached. As flow is not predetermined, the supplied volume varies depending on the patient’s pulmonary response.
203
10
Figure 81.
Ideal Waveform Using Pressure Control Ventilation
1 Pressure (cmH
2
O)
2 Flow (L/min)
3 Volume (mL)
4 Target pressure
5 PEEP
6 Inspiratory phase
7 Expiratory phase
10.7.3 VC+
VC+ breaths require initialization and must go through a startup routine.
10.7.3.1 VC+ Startup
Up to three test breaths will be delivered prior to ventilating the patient with VC+ breaths. The delivered VC+ test breaths will be volume control (VC) breaths (if Leak Sync is disabled) using the VC+ settings for V
T
and inspiratory time and a ramp flow pattern. Delivered peak flows are calculated based on the set T compliance and assuming an end-inspiratory pressure of 15 cmH types, and 10 cmH
2
O above PEEP for neonatal patient type.
2
I
(including 200 ms inspiratory plateau and assuming a descending ramp flow pattern). Tubing volume will be calculated based on SST tubing
O above PEEP for adult and pediatric patient
After each test breath, the measured delivered volume and the end-inspiratory pressure and end-expiratory pressure are used to estimate the patient’s lung compliance to determine the VC+ pressure target to achieve the set V
T
.
In VC+, if pressure and volume measurements from the test breaths are not valid, then a PC breath will be delivered with P
I
of 15 cmH
2
O for pediatric and adult patients or 10 cmH
% settings in VC+.
2
O for neonate patients using the T
I
and rise time
Note: To allow for optimal function of startup and operation of VC+ in the ventilator it is important not to block the tubing while the patient is undergoing suctioning or other treatment that requires disconnection from the ventilator. The ventilator has a disconnect detection algorithm that suspends ventilation while the patient is disconnected.
After VC+ Startup, the ventilator will make adjustments to the target pressure in order to deliver the set volume
(V
T
). To reach the desired volume promptly, the maximum allowed pressure adjustments for all patient circuit types are summarized in
204
10
Table 46.
Pressure Adjustment Limits According to Patient Body Weight (PBW)
Condition
Maximum pressure adjustment (up), cmH
2
O
Minimum pressure adjust‐ ment (down), cmH
2
O
+3
-6
+4.5
-6
7 ≤ PBW <
15 kg
+6
-6
15 ≤ PBW <
20 kg
+8
-8
20 ≤ PBW kg
See
Table 36, page 144 for details on the following VC+ alarms:
1. VOLUME NOT DELIVERED
2. HIGH INSPIRED TIDAL VOLUME ( ↑ V
TI
)
3. LOW CIRCUIT PRESSURE ( ↓ P
PEAK
)
4. COMPLIANCE LIMITED V
T
During VC+, inspiratory target pressure cannot be lower than PEEP +3 cmH
2
⤒ P
PEAK
–3 cmH
2
O.
O and cannot exceed
+10
-10
10.7.4 Rise Time%
If PC or VC+ is selected as the Mandatory type, adjust rise time% for optimum flow delivery into lungs. Patients with high impedance (low compliance, and high resistance) may benefit from a lower rise time% whereas patients with low impedance may better tolerate a more aggressive rise time setting. The rise time% setting specifies the speed at which the inspiratory pressure reaches 95% of the target pressure. The rise time setting applies to PS (including a setting of 0 cmH
2
O), PC, or VC+ breaths. To match the flow demand of an actively breathing patient, observe simultaneous pressure-time and flow-time curves, and adjust the rise time% to maintain a smooth rise of pressure to the target value. A rise time% setting reaching the target value well before the end of inspiration can cause the ventilator to supply excess flow to the patient. Whether this oversupply is clinically beneficial must be evaluated for each patient. Generally, the optimum rise time% for gently breathing patients is less than or equal to the default
(50%), while optimum rise time% for more aggressively breathing patients can be 50% or higher.
Warning: Under certain clinical circumstances (such as stiff lungs, or a small patient with a weak inspiratory drive), a rise time% setting above 50% could cause a transient pressure overshoot and premature transition to exhalation, or pressure oscillations during inspiration. Carefully evaluate the patient’s condition before setting the rise time
% above the default setting of 50%.
10.7.5 Manual Inspiration
When pressed, the manual inspiration key delivers one OIM breath to the patient, using set breath delivery parameters.
The ventilator will not allow a manual inspiration during the restricted phase of exhalation or when the ventilator is in the process of delivering a breath whether mandatory or spontaneous). All manual inspiration attempts are logged in the General Event log.
The restricted phase of exhalation is the time period during the exhalation phase where an inspiration trigger is not allowed. The restricted phase of exhalation is defined as the first 200 ms of exhalation or the time it takes for expiratory flow to drop to ≤ 50% of the peak expiratory flow, or the time it takes for the expiratory flow to drop to
≤0.5 L/min (whichever is longest). The restricted phase of exhalation will end after 5 s of exhalation have elapsed regardless of the measured expiratory flow rate.
10.8 Spontaneous Breath Delivery
The modes allowing spontaneous breaths are SIMV, SPONT, and BiLevel.
10
205
The spontaneous breath type setting determines which type of pressure-assist will be applied to the patient’s spontaneous breaths (PS, TC, VS, or PAV+™).
After selecting the spontaneous breath type, choose the level of pressure support (P
(V
T SUPP
) for VS or percent support for TC and PAV+™ and specify the rise time% and E
SENS
) for PS, Support volume
, where available. Changes to the spontaneous breath type setting phase in at the start the next inspiration.
SUPP
Note: In any delivered spontaneous breath, either Invasive or NIV, there is always a target inspiratory pressure of at least 1.5 cmH
2
O applied.
During spontaneous breathing, the patient’s respiratory control center rhythmically activates the inspiratory muscles. The support type setting allows selection of pressure-assist to supplement the patient’s pressure-generating capability.
Table 47.
Spontaneous Breath Delivery Characteristics
Characteristic
Inspiratory detection
Pressure or flow during inspiration
Spontaneous type = PS and P
<5 cmH
2
O
SUPP
1
2
Implementation
P
SENS
or ⩒
SENS
depending on the trigger type selected.
Pressure rises according to the selected rise time% and PBW setting, with target pressure equal to the effective pressure + PEEP:
P
SUPP
Effective pressure (cmH
2
O)
0 1.5
2.2
2.9
3
4
3.6
4.3
Pressure rises according to the selected rise time% and PBW setting, and target pressure equals P
SUPP
+ PEEP.
Pressure or flow during inspiration
Spontaneous type = PS and P
SUPP
≥5 cmH
2
O
Pressure or flow during inspiration
Spontaneous type = VS
Tube compensation (TC)
Pressure rises according to the selected rise time% and PBW setting, and target pressure equals the pressure determined during the test breath or pressure target determined from assessment of delivered volume from
the previous breath. For more information on VS, see Section 10.8.2, Volume
.
Tube compensation provides programmable, inspiratory pressure assis‐ tance during otherwise unsupported spontaneous breaths. This assists the patient in overcoming the flow resistance of the artificial airway. Pres‐ sure is programmed to help the patient overcome part or all of the resist‐ ance of the artificial airway. The ventilator continuously calculates the pressure differential and adjusts the compensation pressure accordingly.
For more information regarding TC, see Section 10.8.3, Tube Compensation, page 208 .
Inspiratory flow profile The inspiratory flow profile is determined by patient demand and the rise time% setting. As the rise time% setting is increased from minimum to maximum, the time to achieve the pressure target decreases. The maxi‐ mum available flow is up to 30 L/min for neonatal circuit types, 80 L/min for pediatric circuit types, and up to 200 L/min for adult circuit types with‐ out Leak Sync.
Exhalation valve during inspiration Adjusts to minimize pressure overshoot and maintain the target pressure.
Inspiratory valve during inspiration Adjust to maintain target pressure.
206
10
Table 47.
Spontaneous Breath Delivery Characteristics (continued)
Characteristic Implementation
Because the exhalation valve acts as a relief valve venting any excess flow, inspiratory flow can be delivered aggressively and allows reduced work of breathing.
Expiratory detection The end-inspiratory flow or airway pressure method, whichever detects exhalation first. Time backup and the ↑ P backup strategies.
PEAK
alarm are also available as
Pressure or flow during exhalation Pressure is controlled to PEEP.
For pressure triggering: set to deliver a bias flow of 1 L/min. For flow trig‐ gering: set to deliver base flow.
Inspiratory valves during exhalation For pressure triggering: set to deliver a bias flow of 1 L/min. For flow trig‐ gering: set to deliver base flow near the end of expiratory flow.
Exhalation valve during exhalation Adjusts to maintain the operator-selected value for PEEP.
10.8.1 Pressure Support (PS)
Pressure Support is a type of spontaneous breath, similar to PC, by which the pressure is controlled to reach a constant value, preset by the practitioner, once an inspiratory effort is detected. This target value is held until the detection of end of inspiration. Subsequently, the exhalation valve control initiates the exhalation, driving the pressure to the PEEP level.
10.8.2 Volume Support (VS)
Volume support is a pressure-supported spontaneous breath type available when SPONT is selected as the mode.
The target support volume (V
T SUPP
) is the target volume for pressure supported breaths.
See
Table 63, page 246 for details regarding the following VS settings:
• Expiratory sensitivity (E
SENS
)
• Rise time%
• Target support volume (V
T SUPP
)
10.8.2.1 Technical Description
Volume Support (VS) breaths are patient-triggered, pressure-supported spontaneous breaths. The VS algorithm varies the inspiratory pressure of each breath to deliver the operator-set target tidal volume (V
T SUPP
). If the delivered volume for a breath is above or below the set target volume, VS adjusts the target pressure for the next breath up or down, as necessary, to deliver more or less volume. As the patient’s condition improves allowing more patient control over spontaneous ventilation, the VS algorithm decreases the amount of inspiratory pressure necessary to deliver the target volume. Conversely, VS increases inspiratory pressure if the patient’s respiratory drive becomes compromised.
In the absence of leaks or changes in patient resistance or compliance, Volume Support achieves and maintains a steady, breath-to-breath tidal volume within five breaths of VS initiation or startup.
During VS, the inspiratory pressure target cannot be lower than PEEP +1.5 cmH
2
⤒ P
PEAK
–3 cmH
2
O.
O, and cannot exceed
10.8.2.2 VS Startup
Test breaths will be delivered prior to ventilating the patient with VS breaths.
The delivered VS test breaths will be pressure support breaths using a P adult patients or 10 cmH
2
SUPP
value of 15 cmH
2
O for neonate patients. The test breaths will use the E
SENS
O for pediatric and
and rise time% settings in VS.
10
207
After each test breath, the measured delivered volume and the end-inspiratory pressure and end-expiratory pressure will be used to estimate the patient’s lung compliance to determine the VS pressure target to achieve the set V
T
.
In VS, if pressure and volume measurements from test breaths are not valid, VS startup continues delivering test breaths until the pressure and volume measurements are valid.
Note: To allow for optimal function of startup and operation of VS in the ventilator it is important not to block the tubing while the patient is undergoing suctioning or other treatment that requires disconnection from the ventilator. The ventilator has a disconnect detection algorithm that suspends ventilation while the patient is disconnected.
After VS Startup, the ventilator makes adjustments to the target pressure in order to deliver the set volume (V
T
SUPP
). In order to reach the desired volume promptly, the maximum allowed pressure adjustments for all patient circuit types are summarized in
.
Table 48.
Pressure Adjustment Limits According to Patient Body Weight (PBW)
Condition
Maximum pressure adjustment (up), cmH
2
O
Minimum pressure adjust‐ ment (down), cmH
2
O
+3
-6
+4.5
-6
7 ≤ PBW <
15 kg
+6
-6
15 ≤ PBW <
20 kg
+8
-8
20 ≤ PBW kg
+10
-10
See
Table 36, page 144 for details on the following VS alarms:
• VOLUME NOT DELIVERED
• COMPLIANCE LIMITED V
T
• HIGH INSPIRED TIDAL VOLUME ( ↑ V
TI
)
10.8.2.3 Monitored Patient Data
See
during VS breaths.
10.8.3 Tube Compensation
Tube compensation (TC) is a pressure-supported spontaneous breath type available in SIMV, SPONT and BiLevel modes. When TC is enabled, the patient’s respiratory muscles are not required to work as hard to draw gases into the lungs as they would in the absence of the pressure assistance provided by the TC feature. This is particularly important for patients whose respiratory systems are already functioning poorly, and would have to exert even greater muscular effort to overcome the increased resistance to flow through the artificial airway.
Tube compensation provides programmable, inspiratory pressure assistance during otherwise unsupported spontaneous breaths. This assists the patient in overcoming the flow resistance of the artificial airway. Pressure is programmed to vary in accordance with the resistance to flow through the artificial airway. The ventilator continuously calculates the pressure differential and adjusts the compensation pressure accordingly.
Tube compensation also includes safety protection, safety checks, and logic checks which prevent the operator from entering certain incompatible settings, such as a large airway size paired with a small predicted body weight.
If the type of humidifier has been changed after running SST with TC, the volume can be adjusted at the same time to avoid a reduction in compensation compliance accuracy.
208
10
10.8.3.1 Technical Description
Tube compensation is a spontaneous mode enhancement which assists patients’ spontaneous breaths not already supported by specific pressure-based breath types (such as PS, VS, and PAV+™) by delivering positive pressure proportional to the flow-based, resistive pressure developed across the artificial airway. TC causes the sensation of breathing through an artificial airway to diminish because the TC algorithm instructs the ventilator to develop just the correct amount of forward pressure to offset (cancel) the back pressure developed across the artificial airway during the inspiratory phase. The degree of cancellation can be set by the clinician and is adjustable between 10% an 100% in increments of 5%.
Tube compensation can support all unsupported spontaneous breaths for patients with predicted body weights
≥7.0 kg (15.4 lb), and for endotracheal/tracheostomy tubes with an inside diameter (ID) of ≥4.5 mm. TC can be used within SPONT, BiLevel, or SIMV, all of which permit unsupported spontaneous breaths. With BiLevel selected,
TC supports spontaneous breaths at both pressure levels.
Tube compensation checks the flow rate every 5 ms, using an internal lookup table which contains the flow-to-pressure relationship of the selected artificial airway, and is used to calculate the amount of pressure needed to overcome all or part of the resistance of the artificial airway. Based the TC setting and the instantaneous flow measurement, the ventilator’s PSOL valves are continually adjusted, adjusting the circuit pressure to match the changing tube-pressure compensation requirements.
10.8.3.2 Tube Compensation Alarms
See
for details of the ↑ P
COMP
, ↑ P
VENT
, and ↑ V
TI
alarms associated with TC.
10.8.3.3 Monitored Patient Data
See
for details of the inspired tidal volume (V
TI with TC.
) monitored patient data parameter a associated
10.8.3.4 Tube Inside Diameter (ID)
The ventilator uses “soft bound” values for estimated tube inside diameter (ID) based on PBW. Soft bounds are ventilator settings that have reached their recommended high or low limits. When adjusting the tube size, if the inside diameter does not align with a valid predicted body weight, a Continue button appears. Setting the ventilator beyond these soft bounds requires the operator to acknowledge the prompt by touching Continue before continuing to adjust the tube size. The limit beyond which the tube ID cannot be adjusted is called a hard bound, and the ventilator emits an invalid entry tone when a hard bound is reached.
Warning: Greater than expected ventilatory support, leading to unknown harm, can result if the specified tube type or tube ID is smaller than the actual tube type or tube ID.
10.8.3.5 Ventilator Settings/Guidelines
The estimation of settings to use with TC is aided by an understanding of the ventilator settings, the data used for determination of the compensation values, and the specified performance or accuracy of the TC function.
The setting for ⤒ P
PEAK
must take the estimated tube compensation into consideration. The target pressure
(compensation) at the patient wye is derived from the knowledge of the approximate airway resistance of the ET or tracheostomy tube being used. The compensation pressure in cmH of PEEP for calculation and setting of ⤒ P
PEAK
.
2
O for available tube sizes and gas flows is shown. See
. The estimated compensation must be added to the value
10.8.3.6 Specified Performance
Performance using TC is specified to be ±(0.5 +10% of actual) joules/liter (residual work during inspiration at the
100% support (% Supp) level). Work is computed over the entire inspiratory interval. In terms of ventilation, resistive work is given by the following equation:
10
209
W
P
E END
Work [J/L]
End expiratory pressure
P k
TR
Tracheal pressure
Conversion constant (0.098)[J/cmH
2
O x L]
indicate pressures at steady-state flows for ET tubes and tracheostomy tubes, respectively, at 100% support at the wye for sizes between 4.5 mm and 10 mm.
Figure 82.
ET Tube Target Pressure vs. Flow
4.5 mm 5 mm 5.5 mm 6 mm
110
6.5 mm
100
1
40
30
20
60
50
10
0
0
90
80
70
20 40 60 80 100 120 140 160 180
7 mm
7.5 mm
8 mm
8.5 mm
9 mm
9.5 mm
10 mm
200
2
VEN_10249_A
1 Pressure (cmH
2
O)
2 Flow (L/min)
210
10
Figure 83.
Tracheostomy Tube Target Pressure vs. Flow
110
1
100
90
80
70
60
30
20
50
40
10
0
0 20 40 60 80 100
4.5 mm
120
5 mm
140 160
5.5 mm 6 mm
180 200
6.5 mm
7 mm
7.5 mm
8 mm
8.5 mm
9 mm
9.5 mm
10 mm
2 VEN_10250_A
1 Pressure (cmH
2
O)
2 Flow (L/min)
10.8.4 Proportional Assist Ventilation (PAV+™)
PAV+™ is another available type of spontaneous breath. For detailed description of the operating theory, See
10.9 A/C Mode
When the ventilator is in assist-control (A/C) mode, only mandatory breaths are delivered. These mandatory breaths can be PC, VC, or VC+ breaths. See
Section 10.7, Mandatory Breath Delivery, page 202 for a more detailed
explanation of VC+ breaths. As for any mandatory breath, the triggering methods can be P
TRIG
, ⩒
TRIG
, time-triggered, or operator initiated. If the ventilator senses the patient initiating the breath, a PIM or assist breath is delivered. Otherwise, VIM breaths (control breaths) are delivered based on the set respiratory rate. The length of the breath period is defined as:
10 where:
T b
= breath period (s) f = set respiratory rate (breaths per minute)
The inspiratory phase length is determined by the current breath delivery settings. At the end of the inspiratory phase, the ventilator enters the expiratory phase as determined by the following equation: where:
T
E
= length of the expiratory phase (s)
T
I
= length of inspiratory phase (s) including plateau time, T
PL
211
VIMs).
Figure 84.
No Patient Inspiratory Effort Detected
1 VIM
2 T b
to be delivered at a rate greater than or equal to the set respiratory rate.
Figure 85.
Patient Inspiratory Effort Detected
1 PIM
2 T b set
Figure 86 illustrates A/C breath delivery when there are both PIM and VIM breaths delivered.
Figure 86.
Combined VIM and PIM Breaths
1 VIM
2 PIM
3 T b set
If changes to the respiratory rate are made, they are phased in during exhalation only. The new breath period depends on the new respiratory rate, is based on the start of the current breath, and follows these rules:
• The current breath’s inspiratory time is not changed.
• A new inspiration is not delivered until at least 200 ms of exhalation have elapsed.
• The maximum time t until the first VIM for the new respiratory rate is delivered is 3.5 times the current inspiratory time or the length of the new breath period (whichever is longer), but t is no longer than the old breath period.
• If the patient generates a PIM after the ventilator recognizes the rate change and before time t , the new rate begins with the PIM.
212
10.9.1 Changing to A/C Mode
Switching to A/C mode from any other mode causes the ventilator to phase in a VIM and set the start time for the beginning of the next A/C breath period. Following this VIM, and before the next A/C period begins, the ventilator responds to the patient’s inspiratory efforts by delivering mandatory breaths.
The first A/C breath (VIM breath) is phased in while following these rules:
• The breath is not delivered during an inspiration.
• The breath is not delivered during the restricted phase of exhalation.
• The ventilator ensures the apnea interval elapses at least 5 s after the beginning of exhalation.
• Any other specially scheduled event (for example, a respiratory mechanics maneuver or any pause maneuver) is canceled and rescheduled at the next interval.
When the first VIM of the new A/C mode is delivered depends on the mode and breath type active when the mode change is requested.
10.10 SIMV Mode
Synchronous Intermittent Mandatory Ventilation (SIMV) mode is a mixed ventilation mode allowing both mandatory and spontaneous breaths using pressure- or flow-triggering. The mandatory breaths can be PC, VC, or VC+, and the spontaneous breaths are pressure-assisted with either PS or TC. SIMV guarantees one mandatory breath per SIMV breath period, which is either a PIM or VIM. OIM breaths are allowed in SIMV and are delivered at the setting selected for Mandatory Type. See
, which shows the two parts of the SIMV breath period.
Figure 87.
Mandatory and Spontaneous Intervals
10
1 T b
=SIMV breath period (includes T m
and T s
)
2 T m
=Mandatory interval (reserved for a PIM breath)
3 T s
=Spontaneous interval (VIM delivered if no PIM delivered during T m
)
The first part of the period is the mandatory interval ( T m
) which is reserved for a PIM. If a PIM is delivered, the T interval ends and the ventilator switches to the second part of the period, the spontaneous interval ( T s m
), which is reserved for spontaneous breathing for the remainder of the breath period. At the end of an SIMV breath period, the cycle repeats. If a PIM is not delivered during the mandatory interval, the ventilator delivers a VIM at the end of the mandatory interval, then switches to the spontaneous interval. The following figure shows an SIMV breath period where a PIM is delivered within the mandatory interval. Any subsequent trigger efforts during Its yield spontaneous breaths. As shown, T m
transitions to T s
when a PIM is delivered.
10
213
Figure 88.
PIM Delivered Within Mandatory Interval
1 PIM
2 T m
(T m
transitions to T s
when a PIM is delivered)
3 T s
(subsequent trigger efforts during T s
yield spontaneous breaths)
4 T b
The following figure shows an SIMV breath period where a PIM is not delivered within the mandatory interval.
Figure 89.
PIM Not Delivered Within Mandatory Interval
1 VIM
2 T m
(VIM delivered at end of T m
if no PIM delivered during T m
3 T s
4 T b
In SIMV, mandatory breaths are identical to those in A/C mode if the ventilator’s respiratory rate setting is greater than the patient’s natural respiratory rate. Spontaneous breaths are identical to those in SPONT mode if the ventilator setting for respiratory rate is significantly below the patient’s natural respiratory rate. Patient triggering must meet the requirements for pressure and flow sensitivity.
The procedure for setting the respiratory rate in SIMV is the same as in A/C mode. Once the respiratory rate f is set, the SIMV interval period T b
in seconds is:
T b
= 60/f
During the mandatory interval, if the patient triggers a breath according to the current setting for pressure or flow sensitivity, the ventilator delivers a PIM. Once a mandatory breath is triggered, T m
ends, T trigger efforts yield spontaneous breaths. During the spontaneous interval, the patient can take as many spontaneous breaths.as allowed. If no PIM or OIM is delivered by the end of the mandatory interval, the ventilator delivers a VIM and transitions to the spontaneous interval at the beginning of the VIM.
s
begins, and any further
The SIMV breathing algorithm delivers one mandatory breath each period interval, regardless of the patient’s ability to breathe spontaneously. Once a PIM or VIM is delivered, all successful patient efforts yield spontaneous breaths until the cycle interval ends. The ventilator delivers one mandatory breath during the mandatory interval, regardless of the number of successful patient efforts detected during the spontaneous interval. (An OIM
214
10 delivered during the mandatory interval satisfies the mandatory breath requirement, and causes T m to T s
.)
to transition
The maximum mandatory interval for any valid respiratory rate setting in SIMV is defined as the lesser of:
• 0.6 × the SIMV interval period ( T b
)
• 10 s.
There is no minimum value for T m
.
In SIMV, the interval from mandatory breath to mandatory breath can be as long as 1.6 × the SIMV period interval
(but no longer than the period interval +10 s). At high respiratory rates and too-large tidal volumes, breath stacking (the delivery of a second inspiration before the first exhalation is complete) is likely. In volume ventilation, breath stacking during inspiration and early exhalation leads to hyperinflation and increased airway and lung pressures, which can be detected by a high pressure limit alarm. In pressure control ventilation (with inspiratory pressure remaining constant), breath stacking leads to reduced tidal volumes, which can be detected by the low tidal volume and minute ventilation alarms.
In SIMV mode it is possible for the respiratory rate to drop temporarily below the f setting (unlike A/C mode, in which f
TOT
is always greater than or equal to the f setting). If the patient triggers a breath at the beginning of a breath period, then does not trigger another breath until the maximum mandatory interval for the following breath has elapsed, a monitored respiratory rate less than the respiratory rate setting can result.
If a spontaneous breath occurs toward the end of the spontaneous interval, inspiration or exhalation can still be in progress when the SIMV interval ends. No VIM, PIM, or OIM is allowed during the restricted phase of exhalation.
In the extreme, one or more expected mandatory breaths could be omitted. When the expiratory phase of the spontaneous breath ends, the ventilator reverts to its normal criteria for delivering mandatory breaths.
If an OIM is detected during the mandatory interval, the ventilator delivers the currently specified mandatory breath then closes delivered during T s
.
T m
and transitions to T s
. If an OIM is detected during the spontaneous interval, the ventilator delivers the currently specified mandatory breath, but the SIMV cycle timing does not restart if OIM breaths are
10.10.1 Changing to SIMV Mode
Switching the ventilator to SIMV from any other mode, causes the ventilator to phase in a VIM and set the start time for the next SIMV period. Following this VIM, but before the next SIMV period begins, the ventilator responds to successful patient inspiratory efforts by delivering spontaneous breaths. The first SIMV VIM breath is phased in according to the following rules:
• The VIM breath is not delivered during an inspiration or during the restricted phase of exhalation.
• If the current mode is A/C, the first SIMV VIM is delivered after the restricted phase of exhalation plus the shortest of the following intervals, referenced to the beginning of the last or current inspiration: 3.5 T
T
A
, or the length of the current breath period.
I
, current
• If the current mode is SPONT, and the current or last breath type was spontaneous or OIM, the first SIMV VIM is delivered after the restricted phase of exhalation plus the shortest of the following intervals, referenced to the beginning of the last or current inspiration: 3.5 × T
I
or current T
A
.
• If the current mode is BiLevel in the P
H reduced to P
L
state and the current breath is mandatory, the PEEP level will be
once the exhalation phase is detected.
• The time t until the first VIM of the new A/C mode is the lesser of:
– PEEP transition time + 2.5 × the duration of the active gas delivery phase, or
– the length of the apnea interval (T
A
), or
– the length of the current breath cycle
• If the current mode is BiLevel in the P
H
state and the current breath is spontaneous, the PEEP level will be reduced once the exhalation phase is detected.
10
215
• The time t until the first VIM of the new A/C mode is the lesser of:
– PEEP transition time + 2.5 × the duration of the spontaneous inspiration, or
– the start time of the spontaneous breath + the length of the apnea interval (T
A
).
• If the current mode is BiLevel in the P of the new A/C mode is the lesser of:
L
state and the current breath is mandatory, the time t until the first VIM
– PEEP transition time + 2.5 × the duration of the active gas delivery phase, or
– the length of the apnea interval (T
A
), or
– the length of the current breath cycle
• If the current mode is BiLevel in the P
L time has occurred during P
L
state and the current breath is spontaneous and the spontaneous start
, the time t until the first VIM of the new A/C mode is the lesser of:
– 3.5 × the duration of the spontaneous inspiration, or
– the length of the apnea interval (T
A
) or
– the length of the current breath cycle
• If the current mode is BiLevel in the P
L time has occurred during P
H
state and the current breath is spontaneous and the spontaneous start
, the time t until the first VIM of the new A/C mode is the lesser of:
– PEEP transition time + 2.5 × the duration of the spontaneous inspiration, or
– the start time of the spontaneous breath + the length of the apnea interval (T
A
).
If the command to change to SIMV occurs after the restricted phase of exhalation has ended, and before a next breath or the apnea interval has elapsed, the ventilator delivers the first SIMV VIM at the moment the command is recognized.
The point at which the new rate is phased in depends on the current phase of the SIMV interval and when the rate change command is accepted. If the rate change occurs during the mandatory interval, the maximum mandatory interval is that for the new or old rate, whichever is less. If the patient generates a successful inspiratory effort during the spontaneous interval, the ventilator responds by delivering a spontaneous breath.
Respiratory rate changes are phased in during exhalation only. The new SIMV interval is determined by the new respiratory rate and is referenced to the start of the current SIMV period interval, following these rules:
• Inspiratory time (T
I
) of current breath is neither truncated nor extended.
• The new inspiration is not delivered until 200 ms of exhalation have elapsed.
The time until the new SIMV interval begins is:
• whichever is greater: the new SIMV period interval or 3.5 × the last or current T
I
• not greater than the current SIMV period interval
10.11 Spontaneous (SPONT) Mode
In SPONT mode, the patient initiates inspiration according to the trigger type in effect, but OIM breaths are allowed which are delivered with the currently specified mandatory breath parameters. The following spontaneous breath types are available in SPONT mode:
• PS
• VS
• TC
• PAV+™
The inspiratory phase begins when the ventilator detects patient effort during the ventilator’s exhalation phase.
Breath delivery during the inspiratory phase is determined by the settings for pressure support, PEEP, rise time%, and expiratory sensitivity, unless the breath is an OIM breath.
216
10
If tube compensation (TC), or Proportional Assist Ventilation (PAV+™) is selected as the spontaneous type, breath delivery during the inspiratory phase is determined by the settings for% support (% Supp), expiratory sensitivity, tube ID, and tube type.
Note: Given the current ventilator settings, if PAV+™ would be an allowable spontaneous type (except that tube
ID <6 mm) then PAV+™ becomes selectable. If selected, tube ID is set to its New Patient default value based on the
PBW entered. An attention icon for tube ID appears.
If Volume Support (VS) is selected as the spontaneous type, breath delivery during the inspiratory phase is determined by rise time%, volume support level (V
T SUPP
), expiratory sensitivity, and PEEP.
Inspiratory pauses are only possible during OIM breaths, and expiratory pauses are not allowed during SPONT.
Expiratory trigger methods include:
• E
SENS
(% flow deceleration from peak inspiratory flow)
• PBW based time limit (T
I
too long)
• ↑ P
PEAK
• Inspiratory tidal volume limit (for VS only)
• Airway Pressure Cycling method
10.11.1 Changing to SPONT Mode
If the operator changes to SPONT mode during an A/C or SIMV inspiration (mandatory or spontaneous), the inspiration is completed, unaffected by the mode change. Because SPONT mode has no special breath timing requirements, the ventilator then enters the exhalation phase and waits for the detection of patient inspiratory effort, a manual inspiration, or apnea detection.
10.12 Apnea Ventilation
When a patient stops breathing or is no longer being ventilated, it is called apnea. When apnea is detected by the ventilator the ventilator alarms and delivers apnea ventilation according to the current apnea ventilation settings.
10.12.1 Apnea Detection
The ventilator declares apnea when no breath has been delivered by the time the operator-selected apnea interval elapses, plus a small increment of time (350 ms). This increment allows time for a patient who has begun to initiate a breath to trigger inspiration and prevent the ventilator from declaring apnea when the apnea interval is equal to the breath period.
The apnea timer resets whenever an inspiration begins, regardless of whether the inspiration is patient-triggered, ventilator-triggered, or operator-initiated. The ventilator then sets a new apnea interval beginning from the start of the current inspiration. To hold off apnea ventilation, another inspiration must be delivered before (the current apnea interval +350 ms) elapses. Apnea detection is suspended during a disconnect, occlusion, or safety valve open (SVO) state.
Apnea is not declared when the apnea interval setting equals or exceeds the breath period. For example, if the respiratory rate setting is 4/min, an apnea interval of 15 s or more means apnea cannot be detected. The ventilator bases apnea detection on inspiratory (not expiratory) flow, and allows detection of a disconnect or occlusion during apnea ventilation. Apnea detection is designed to accommodate interruptions to the typical breathing pattern due to other ventilator features that temporarily extend the inspiratory or expiratory intervals (rate changes, for example), but still detect a true apnea event.
Figure 90 shows an apnea breath where T
A
equals the breath period.
10
217
Figure 90.
Apnea Interval Equals Breath Period
1 T b0
2 T b1
3 PIM
4 T
A
(apnea interval)
Figure 91 shows an apnea breath with T
A
greater than the breath period.
Figure 91.
Apnea Interval Greater Than Breath Period
1 T b0
2 T b1
3 PIM
4 VIM
5 T
A
(apnea interval)
Figure 92 shows an apnea breath with T
A
less than the breath period.
Figure 92.
Apnea Interval Less Than Breath Period
1 T b0
2 T b1
3 PIM
4 Dashed line indicates a PIM to avoid apnea
5 Apnea VIM
218
6 Apnea interval
7 Apnea T b0
8 Apnea ventilation
9 T b
(T
A
<T b
)
10
10.12.2 Transition to Apnea Ventilation
When apnea is declared, the ventilator delivers apnea ventilation according to the current apnea ventilation settings and displays the apnea settings on the graphical user interface (GUI). Regardless of the apnea interval setting, apnea ventilation cannot begin until inspiration of the current breath is complete and the restricted phase of exhalation has elapsed.
10.12.3 Settings Changes During Apnea Ventilation
All apnea and non-apnea settings remain active on the GUI during apnea ventilation. Both non-apnea and apnea settings changes are phased in according to the applicable rules. If apnea ventilation is active, new settings are accepted but not implemented until non-apnea ventilation begins. Allowing key entries after apnea detection allows adjustment of the apnea interval at setup, regardless of whether apnea has been detected. During apnea ventilation, the manual inspiration key is active, but expiratory pause and inspiratory pause keys are not active. The increase O
2
control is active during apnea ventilation, because apnea detection is likely during suctioning.
The apnea respiratory rate must be ≥60/T
A
Additionally, apnea settings cannot result in an I:E ratio >1.00:1.
10.12.4 Resetting Apnea Ventilation
Apnea ventilation is intended as an auxiliary mode of ventilation when there is insufficient breath delivery to the patient over a specified period of time. Apnea ventilation can be reset to normal ventilation by the operator (by pressing the alarm reset key) or the patient (autoreset). It is also reset when a rate change is made that renders apnea ventilation inapplicable.
If the patient regains inspiratory control, the ventilator returns to the operator-selected mode of non-apnea ventilation. The ventilator determines whether the patient has regained respiratory control by monitoring triggered inspirations and exhaled volume. With exception of PAV+ and TC breath types, if the patient triggers two consecutive inspirations, and the exhaled volume is equal to or greater than 50% of the delivered volume
(including any compliance volume), the ventilator resets to non-apnea ventilation. Exhaled volume is monitored to avoid resetting due to autotriggering caused by large leaks in the patient circuit. During PAV+ and TC, on the first patient triggered breath, the ventilator returns to PAV+ or TC. PAV+ reenters its Startup phase to estimate the patient’s respiratory mechanics values. TC begins immediately at the set percent support.
10.12.5 Apnea Ventilation in SIMV
The following strategy is designed to allow SIMV to avoid triggering apnea ventilation if a VIM breath can be delivered instead:
• If the apnea interval (T
A
) elapses at any time during the mandatory interval, the ventilator delivers a VIM rather than beginning apnea ventilation.
• If T
A
elapses during the spontaneous interval, apnea ventilation begins.
shows an illustration of how SIMV is designed to deliver a VIM rather than trigger apnea ventilation, when possible.
10
219
Figure 93.
Apnea Ventilation in SIMV
1 T b
2 Last breath (PIM)
3 VIM
4 T m max
5 T
A
6 T m
(If T
A
elapses during T m
, ventilator delivers a VIM rather than beginning apnea ventilation)
7 T s
10.12.6 Phasing in New Apnea Intervals
How a new apnea interval is phased in depends on whether or not apnea ventilation is active. If apnea ventilation is active, the ventilator accepts and implements the new setting immediately. During normal ventilation (that is, apnea ventilation is not active), these rules apply:
• If the new apnea interval setting is shorter than the current (or temporarily extended) apnea interval, the new value is implemented at the next inspiration.
• If the new apnea interval setting is longer than the current (or temporarily extended) apnea interval, the old interval is extended to match the new interval immediately.
10.13 Detecting Occlusion and Disconnect
10.13.1 Occlusion
The ventilator detects severe patient circuit occlusions in order to protect the patient from excessive airway pressures, or from receiving little or no gas. Occlusions require immediate attention to remedy.
The ventilator detects a severe occlusion if:
• The inspiratory or expiratory limb of the breathing circuit is partially or completely occluded (condensate or secretions collected in a gravity-dependent loop, kinked or crimped tubing, etc.).
• The ventilator exhaust port is blocked or resistance through the port is too high.
The ventilator checks the patient circuit for occlusions during all modes of breathing (except Stand-By state and
Safety Valve Open) at delivery of every breath. Once the circuit check begins, the ventilator detects a severe occlusion of the patient circuit within 200 ms. The ventilator checks the exhaust port for occlusions during the expiratory phase of every breath (except during disconnect and safety valve open). Once the exhaust port check begins, the ventilator detects a severe occlusion within 100 ms following the first 200 ms of exhalation. All occlusion checking is disabled during pressure sensor autozeroing.
When an occlusion is detected, an alarm sounds, the ventilator enters the OSC (occlusion status cycling) state and displays a message indicating the length of time the patient has gone without ventilation (how long the ventilator
220
10 has been in OSC). This alarm has the capability to autoreset, since occlusions such as those due to patient activity
(for example, crimped, or kinked tubing) can correct themselves.
Once a severe occlusion is detected, the ventilator acts to minimize airway pressure. Because any severe occlusion places the patient at risk, the ventilator minimizes the risk while displaying the length of time the patient has been without ventilatory support. Severe occlusion is detected regardless of what mode or triggering strategy is in effect. When a severe occlusion is detected, the ventilator terminates normal ventilation, terminates any active audio paused interval, annunciates an occlusion alarm, and enters the safe state (exhalation and inspiratory valve deenergized and safety valve open) for 15 s or until inspiratory pressure drops to 5 cmH comes first.
2
O or less, whichever
During a severe occlusion, the ventilator enters OSC, in which it periodically attempts to deliver a pressure-based breath while monitoring the inspiration and expiratory phases for the existence of a severe occlusion. If the severe occlusion is corrected, the ventilator detects the corrected condition after two complete OSC breath periods during which no occlusion is detected. When the ventilator delivers an OSC breath, it closes the safety valve and waits 500 ms for the safety valve to close completely, delivers a breath with a target pressure of 15 cmH
2
O for 2000 ms, then cycles to exhalation. This breath is followed by a mandatory breath according to the current settings, but with PEEP =0 and O
2
% equal to 100% for adult/pediatric circuit types or 40% for neonatal circuits. During OSC (and only during OSC), the ⤒ P
PEAK
(high circuit pressure) alarm limit is disabled to ensure it does not interfere with the ability of the ventilator to detect a corrected occlusion. When the ventilator does not detect a severe occlusion, it resets the occlusion alarm, reestablishes PEEP, and reinstates breath delivery according to current settings.
Inspiratory and expiratory pause, and manual inspirations are suspended during a severe occlusion. Pause maneuvers are canceled by a severe occlusion. During a severe occlusion, ventilator settings changes are possible.
Severe occlusions are not detected when the ventilator is in the safety valve open (SVO) state.
A corrected occlusion is detected within 15 s.
10.13.2 Disconnect
A circuit disconnect condition is detected when the ventilator cannot ensure that a patient is receiving sufficient tidal volume (due to a large leak or disconnected patient circuit). This discussion applies when Leak Sync is disabled.
When a disconnect is detected, an alarm sounds, the ventilator indicates that a disconnect has been detected, and displays a message indicating the length of time the patient has gone without ventilation.
Patient data are not displayed during a circuit disconnect condition.
The ventilator monitors the expiratory pressure and flow, delivered volume, and exhaled volume to declare a disconnect using any of these methods:
• The ventilator detects a disconnect when the expiratory pressure transducer measures no circuit pressure and no exhaled flow during the first 200 ms of exhalation. The ventilator postpones declaring a disconnect for another 100 ms to allow an occlusion (if detected) to be declared first, because it is possible for an occlusion to match the disconnect detection criteria.
• Despite many possible variations of circuit disconnections or large leaks, it is possible for a patient to generate some exhaled flow and pressure. The ventilator then uses the disconnect sensitivity (D
SENS
), the percentage of delivered volume lost during the exhalation phase of the same breath to declare a disconnect) setting to detect a disconnect.
• If the disconnect occurs during a spontaneous breath, a disconnect is declared when the inspiration is terminated by maximum inspiratory time (or the ⤒ T
I SPONT
limit setting when ventilation type is non-invasive
[NIV]) and the ventilator detects inspiratory flow rising to the maximum allowable.
• If the disconnect occurs at the endotracheal tube, the exhaled volume will be much less than the delivered volume for the previous inspiration. The ventilator declares a disconnect if the exhaled volume is lower than the D
SENS
setting for three consecutive breaths. The D
SENS
setting helps avoid false detections due to leaks in
10
221
the circuit or the patient’s lungs, and the three-consecutive-breaths requirement helps avoid false detections due to a patient out-drawing the ventilator during volume control (VC) breaths.
• Flow less than a value determined using the D
SENS consecutive seconds during exhalation.
setting and pressure less than 0.5 cmH
2
O detected for 10
Warning: When ventilation type is NIV, and D
SENS leaks and some disconnect conditions.
setting is turned OFF, the system may not sound an alarm for
Once the ventilator detects a patient circuit disconnect, the ventilator declares a high-priority alarm and discontinues breath delivery, regardless of what mode (including apnea) was active when the disconnect was detected. If there is an active audio paused interval when the disconnect occurs, the audio paused interval is not canceled. The ventilator displays the length of time the patient has been without ventilatory support. During the disconnect, the exhalation valve closes, idle flow (10 L/min flow with Leak Sync disabled and 20 L/min with Leak
Sync enabled) begins, and breath triggering is disabled. A message appears identifying how long the patient has gone without ventilatory support.
The ventilator monitors both expiratory flow and circuit pressures to detect reconnection. The ventilator declares a reconnect if any of the following criteria are met for the applicable time interval:
• Exhaled idle flow within the reconnect threshold is detected.
• Inspiratory and expiratory pressures are both above or both below reconnect threshold levels.
• Inspiratory pressure rises to a reconnect level.
If the disconnect condition is corrected, the ventilator detects the corrected condition within 1 second.
Ventilator triggering, apnea detection, expiratory and inspiratory pause, manual inspirations, and programmed maneuvers or one-time events are suspended during a patient circuit disconnect condition. Spirometry is not monitored during a disconnect, and all alarms based on spirometry values are disabled. During a disconnect condition, ventilator settings changes are possible.
If the disconnect alarm is autoreset or manually reset, the ventilator reestablishes PEEP. Once PEEP is reestablished, the ventilator reinstates breath delivery according to settings in effect before the disconnect was detected.
10.13.3 Annunciating Occlusion and Disconnect Alarms
Occlusion and disconnection cannot be declared at the same time. Therefore, the ventilator annunciates only the first event to be declared.
Circuit disconnect detection is not active during OSC, SVO, or prior to patient connection.
10.14 Respiratory Mechanics
See Section 4.9, Respiratory Mechanics Maneuvers, page 109
for instructions on how to perform these maneuvers.
In addition to inspiratory pause and expiratory pause maneuvers, the ventilator can provide other respiratory maneuvers, including negative inspiratory force (NIF), occlusion pressure (P
0.1
) and vital capacity (VC), as well as automatic calculations of lung function and performance, such as dynamic compliance (C
DYN resistance (R
DYN
), peak expiratory flow (PEF), end expiratory flow (EEF), C
20
) and dynamic
/C, and peak spontaneous flow (PSF).
Respiratory maneuvers can be performed in all breathing modes (except as noted) but are not available during the following conditions:
• Apnea ventilation
• Safety PCV
• Occlusion status cycling (OSC)
• Non-invasive ventilation (NIV)
• When the circuit type is neonatal
• SVO
222
10
• Ventilator is in Stand-By state
• When any other respiratory maneuver has already taken place during the same breath
The GUI also displays any maneuver request, distinguishing between requests that are accepted or rejected, and any maneuver that has begun, ended, or has been canceled.
When a maneuver is selected, a GUI information panel is opened, displaying the maneuver name, user prompts and controls, and recent calculated results.
Any maneuver is canceled automatically upon declaration of any of the following alarms:
• ↑ P
PEAK
alarm
• ↑ P
VENT
alarm
• ↑ V
TI
The following Respiratory Mechanics maneuvers are not available in BiLevel ventilation:
• P
0.1
—Occlusion pressure
• NIF —Negative inspiratory force
• VC—Vital capacity
10.14.1 Inspiratory Pause
Note: Inspiratory pause and expiratory pause maneuvers can be performed directly by pressing the respective keys on the GUI or by swiping the Menu tab on the left side of the GUI. For more information on how to perform
.
An inspiratory pause maneuver extends the inspiratory phase of a single mandatory breath for the purpose of measuring end inspiratory circuit pressure which is used to calculate static compliance of the patient’s lungs and thorax (C
STAT
), static resistance of the respiratory system (R
STAT
), and inspiratory plateau pressure (P
PL
). To calculate these pressures, the inspiratory and exhalation valves are closed, allowing pressures on both sides of the artificial airway to equalize, revealing the actual lung inflation pressure during a no-flow condition. An inspiratory pause maneuver can be either automatically or manually administered, and is only available during the next mandatory breath in A/C, SIMV, BiLevel or SPONT modes. In BiLevel, an inspiratory pause maneuver is scheduled for the next inspiration prior to a transition from P
H
to P
L
. Only one inspiratory pause maneuver is allowed per breath. An inspiratory pause maneuver cannot occur during apnea ventilation, safety PCV, stand-by state, occlusion, and SVO.
An automatic inspiratory pause begins when the inspiratory pause key is pressed momentarily or the maneuver is started from the GUI screen. See
Section 4.9, Respiratory Mechanics Maneuvers, page 109
for more information on performing respiratory mechanics maneuvers from the Menu tab on the GUI rather than using the keys on the
GUI. The pause lasts at least 0.5 s but no longer than 3 s. A manual inspiratory pause starts by pressing and holding the inspiratory pause key. The pause lasts for the duration of the key-press (up to 7 s).
An active manual inspiratory pause maneuver is considered complete if any of the following occur:
• The inspiratory pause key is released and at least 2 s of inspiratory pause have elapsed or pressure stability conditions have been detected for not less than 0.5 s.
• Pause duration reaches 7 s.
A manual inspiratory pause maneuver request (if the maneuver is not yet active) will be canceled if any of events
1–10 occur. See
.
Table 49.
Inspiratory and Expiratory Pause Events
Event identifier
1
2
Event
There is a loss of communications with the GUI
High Ventilator pressure limit ( ⤒ P
VENT
) is reached
10
223
Table 49.
Inspiratory and Expiratory Pause Events (continued)
Event identifier
3
6
7
4
5
8
9
10
11
12
13
14
15
Event
High circuit pressure limit ( ⤒ P
PEAK
) is reached
A disconnect is detected
Occlusion is detected
Apnea is detected
72 seconds have elapsed without an inspiratory pause after one has been requested
INSPIRATION TOO LONG alarm is detected
High inspired tidal volume ( ↑ V
TI
) alarm is detected
High compensation pressure ( ↑ P
COMP
) alarm is detected
Cancel is touched if maneuver is initiated from the GUI screen.
Safety valve open (SVO) is detected
Patient trigger effort causes circuit pressure to go below sensitivity. The sensitivity level is the setting value for pressure trigger or the backup pressure value for flow trigger
BUV is entered
Expiratory pause key is pressed (inspiratory pause key if maneuver is an expiratory pause)
During a manual inspiratory pause, the maneuver is terminated if any of events 1, 3, 5, 6, 12, or 13 occur. See
An inspiratory pause maneuver is ignored if the ventilator is in Apnea ventilation, safety PCV, OSC, SVO, BUV, or
Stand-By state.
An active automatic inspiratory pause maneuver is terminated and exhalation begun if any of events –12, or 14
.
The active automatic inspiratory pause maneuver is considered complete if the pause duration reaches 3 seconds or pressure stability conditions have been detected for not less than 0.5 s.
An automatic inspiratory pause maneuver request (if the maneuver is not yet active) will be canceled if events
1–9, 11,12,14, or 15 occur. See Table 49
.
Other characteristics of inspiratory pause maneuvers include:
• During an inspiratory pause, the apnea interval (T
A
) is extended by the duration of the inspiratory pause.
• If the ventilator is in SIMV, the breath period during which the next scheduled VIM occurs will also be extended by the amount of time the inspiratory pause is active.
• All activations of the inspiratory pause control are logged in the Patient Data log.
• Severe occlusion detection is suspended.
• When calculating I:E ratio, inspiratory pause is considered part of the inspiratory phase.
• The expiratory time remains unchanged, and will result in a change in the I:E ratio for the breath that includes the inspiratory phase.
Once the inspiratory pause maneuver is completed the operator can review the quality of the maneuver waveform and accept or reject the maneuver data.
10.14.2 Expiratory Pause
An expiratory pause extends the exhalation phase of a single breath in order to measure end expiratory lung pressure (PEEP
TOT
) and allows intrinsic PEEP (PEEP
I
) to be calculated as PEEP
TOT
minus set PEEP. The pressures on
224
10 either side of the artificial airway are allowed to equalize by closing the inspiratory and exhalation valves.
Expiratory pause is available in A/C, SIMV, and BiLevel modes. For A/C and SIMV, the expiratory pause maneuver is scheduled for the next end-of-exhalation prior to a mandatory breath. In BiLevel, the expiratory pause occurs at the next end-of-exhalation prior to a transition from P
L
to P
H
. Only one expiratory pause per breath is allowed, and the expiratory pause request is rejected if an inspiratory pause has already taken place during the same breath.
A request for an expiratory pause maneuver is ignored in apnea ventilation, safety PCV, SPONT, OSC, BUV, and
Stand-By. See Section 4.9, Respiratory Mechanics Maneuvers, page 109
for more information on performing these maneuvers from the GUI screen rather than using the keys on the GUI.
Either manual or automatic expiratory pause maneuvers can occur. A momentary press of the expiratory pause key begins an automatic expiratory pause which lasts at least 0.5 s, but no longer than 3.0 s. A manual expiratory pause starts by pressing and holding the expiratory pause key and lasts for the duration of the key-press (up to
15 s).
An active manual
expiratory pause is terminated if any of events 1–12 occur. See Section 4.9, Respiratory Mechanics
.
An active manual expiratory pause is complete if the expiratory pause key is released and at least 3 s of expiratory pause have elapsed, pressure stability conditions have been detected for ≥0.5 s, or pause duration lasts 15 s.
An active automatic expiratory pause is terminated if any of events 1, 3, or 11–13 occur. See
An active automatic expiratory pause is complete if the pause duration reaches 3 s or pressure stability conditions have been detected for ≥0.5 s, or the pause duration lasts 15 s.
The automatic expiratory pause maneuver request (the maneuver is not yet active) is canceled if events 1–9, 11,
12, or 15 occur:
The automatic expiratory pause maneuver is terminated and inspiration begun if any of events 1, 3, or 11–13
.
Other characteristics of expiratory pause maneuvers include:
• During an active manual expiratory pause, severe occlusion detection is suspended.
• When calculating I:E ratio, the expiratory pause is considered part of the exhalation phase.
• During the expiratory pause, the inspiratory time remains unchanged, so the I:E ratio is changed for the breath that includes the expiratory pause.
• All activations of the expiratory pause control are logged in the Patient Data log.
Once the expiratory pause maneuver is completed the operator can review the quality of the maneuver waveform and accept or reject the maneuver data.
10.14.3 Negative Inspiratory Force (NIF) Maneuver
The Negative Inspiratory Force (NIF) maneuver is a coached maneuver where the patient is prompted to draw a maximum inspiration against an occluded airway (the inspiratory and exhalation valves are fully closed).
A NIF maneuver is canceled if:
• Disconnect is detected
• Occlusion is detected
• SVO is detected
• ↑ P
PEAK
alarm is declared
• ↑ P
VENT
alarm is declared
• ↑ V
TI
alarm is declared
• Communications with the GUI is lost
• The maneuver has been active for 30 s and an inspiration is not detected
225
10
• INSPIRATION TOO LONG alarm is declared
• A manual inspiration is requested
When a NIF maneuver is activated, a single pressure-time waveform grid is automatically displayed. During a NIF maneuver, the circuit pressure displays on the waveforms screen and is regularly updated, producing a real-time display.
When an active NIF maneuver ends successfully, the calculated NIF result appears on the waveforms screen and on the maneuver panel. The NIF value displayed represents the maximum negative pressure from PEEP.
When a NIF maneuver ends, a PEEP restoration breath is delivered to the patient, then normal breath delivery resumes.
10.14.4 P
0.1
Maneuver (Occlusion Pressure)
P
0.1
is the negative airway pressure (delta pressure change) generated during the first 100 ms of an occluded inspiration. It is an estimate of the neuromuscular drive to breathe.
When a P
0.1
maneuver ends successfully, the calculated airway pressure displays on the waveforms screen and on the maneuver panel. A P the maneuver.
0.1
maneuver is terminated if 7 s elapse and a trigger has not been detected to activate
A P
0.1
maneuver is canceled if:
• Disconnect is detected
• Occlusion is detected
• SVO is detected
• ↑ P
PEAK
alarm is declared
• ↑ P
VENT
alarm is declared
• ↑ V
TI
alarm is declared
• INSPIRATION TOO LONG alarm is declared
• Communications with the GUI is lost
• A manual inspiration is requested
10.14.5 Vital Capacity (VC) Maneuver
The Vital Capacity (VC) maneuver is a coached maneuver where the patient is prompted to draw a maximum inspiration (regardless of the current settings) and then to slowly and fully exhale.
When the vital capacity maneuver becomes active, the ventilator delivers a spontaneous inspiration in response to patient effort (with P
SUPP
=0, Rise time% =50, and E
SENS
=0), and then allows for a full exhalation effort.
• Disconnect is detected
• Occlusion is detected
• SVO is detected
• ↑ P
PEAK
alarm is declared
• ↑ P
VENT
alarm is declared
• ↑ V
TI
alarm is declared
• INSPIRATION TOO LONG alarm is declared
• Communications with the GUI is lost
• A manual inspiration is requested
• The maneuver as been active for 15 s and inspiration is not detected
• Cancel is touched.
226
10
When an active VC maneuver ends successfully, the calculated expiratory volume displays on the waveforms screen and on the maneuver panel and a PEEP restoration breath is delivered.
10.15 Ventilator Settings
10.15.1 Apnea Ventilation
Apnea ventilation is a backup mode and starts if the patient fails to breathe within the apnea interval (T the operator. T
A
A
) set by
defines the maximum allowable length of time between the start of inspiration and the start of the next inspiration. Available settings include mandatory type (PC or VC). For PC breaths the allowable settings are:
• Apnea interval (T
A
)
• Inspiratory pressure (P
I
)
• Inspiratory time (T
I
)
• Respiratory rate (f )
For VC breaths, the allowable settings are:
• Apnea interval (T
A
)
• Flow pattern
• O
2
%
• Peak inspiratory flow ( ⩒
MAX
)
• Respiratory rate (f )
• Tidal volume (V
T
)
During apnea ventilation with PC selected as the mandatory type, rise time% is fixed at 50%, and the constant parameter during a rate change is inspiratory time (T
I
).
If apnea is possible (that is, if (60/f ) > T ventilation O
2
A
) increasing the non-apnea O
2
% setting automatically changes apnea
% if it is not already set higher than the new non-apnea O
2
%. Apnea ventilation O
2
% does not automatically change by decreasing the non-apnea O
2
%. Whenever there is an automatic change to an apnea setting, a message appears on the GUI, and the apnea settings screen appears.
During apnea ventilation, changes to all non-apnea ventilation settings are allowed, but the new settings do not take effect until the ventilator resumes normal ventilation. Being able to change T
A
during apnea ventilation can avoid immediately re-entering apnea ventilation once normal ventilation resumes.
Because the minimum value for T
A
is 3 s, apnea ventilation cannot take place when non-apnea f is greater than or equal to 18/min. The ventilator does not enter apnea ventilation if T
A
is equal to the breath period interval. Set T to a value less than the expected or current breath period interval as a way of allowing the patient to initiate breaths while protecting the patient from the consequences of apnea.
A
10.15.2 Circuit Type and Predicted Body Weight (PBW)
Together, circuit type and PBW (displayed in lb or kg) provide the basis for new patient values and absolute limits on various ventilator settings such as tidal volume (V
T
) and peak flow ( ⩒
MAX
). Run SST in order to change the circuit
type. Table 50 gives the minimum, maximum, and new patient default values for V
T
based on circuit type.
Table 50.
Values for V
T
Based on Circuit Type
Circuit Type
Neonatal
New Patient Default
When mandatory type is
VC+, MAX {2 mL, (mL/kg
Ratio × PBW)} mL; When mandatory type is VC,
Minimum V
T
2 mL if NeoMode 2.0 soft‐ ware option is installed
Maximum V
315 mL
T
10
227
Table 50.
Values for V
T
Based on Circuit Type (continued)
Circuit Type New Patient Default
MAX {3 mL, (mL/kg Ratio ×
PBW)} mL
Pediatric
Adult mL/kg ratio × PBW mL mL/kg ratio × PBW mL
25 mL
75 mL
Minimum V
T
Maximum V
1590 mL
2500 mL
T
See
T
setting, for more information on V
T
calculations based on PBW and circuit type.
Table 51.
Peak Flow and Circuit Type (Leak Sync Disabled)
Circuit Type
Neonatal
Maximum peak flow ( ⩒
MAX
) setting
30 L/min
Pediatric
Adult
60 L/min
150 L/min
PBW determines constants for breath delivery algorithms, some user-settable alarms, the high spontaneous inspiratory time limit setting ( ⤒ T
I SPONT
) in NIV, and the non-settable INSPIRATION TOO LONG alarm.
10.15.3 Ventilation Type
There are two ventilation type choices: Invasive and NIV (Non-Invasive). Invasive Ventilation is conventional ventilation used with endotracheal or tracheostomy tubes. All installed software options, breathing modes, breath types, and trigger types are available during Invasive Ventilation.
for a list of interfaces that have been successfully tested with NIV.
NIV enables the ventilator to handle large system leaks associated with these interfaces by providing pressure-based disconnect alarms, minimizing false disconnect alarms, and replacing the INSPIRATION TOO
LONG alarm with a high spontaneous inspiratory time limit ( ⤒ T
I SPONT
) setting and visual indicator.
The following list shows the subset of invasive settings active during NIV:
Mode – A/C, SIMV, SPONT. (BiLevel is not available during NIV).
Mandatory Type – PC or VC. (VC+ is not available during NIV).
Spontaneous Type – PS (TC and VS are not available during NIV).
During NIV alarm setup, the clinician may set alarms to OFF and must determine if doing so is appropriate for the patient’s condition.
10.15.4 Mode and Breath Type
Specifying the mode defines the types and sequences of breaths allowed for both invasive and NIV ventilation types.
Table 52.
Modes and Breath Types
Mode
A/C
Mandatory Breath Type Spontaneous Breath Type
Invasive: VC, VC+ or PC NIV:
VC or PC
Not allowed
SIMV Invasive: PC, VC, or VC+
NIV: VC or PC
Pressure supported (PS) or
TC
Sequence
All mandatory (patient-, ventilator-, or operator-ini‐ tiated).
Each new breath begins with a mandatory interval, during which a patient effort yields a synchron‐
228
10
Table 52.
Modes and Breath Types (continued)
Mode
SPONT
BiLevel (invasive ventila‐ tion type only)
CPAP
Mandatory Breath Type Spontaneous Breath Type
Not allowed (PC or VC allowed only for manual inspirations).
PC
VC or PC (allowed only for
OIM breaths)
Sequence ized mandatory breath. If no patient effort is detec‐ ted during the mandatory interval, the ventilator delivers a mandatory breath. Subsequent patient efforts before the end of the breath yield spontaneous breaths.
All spontaneous (except for manual inspirations).
Invasive: pressure supported
(PS), tube compensated
(TC), volume supported (VS), proportionally assisted
(PAV+™) NIV: PS
PS, TC
N/A
Combines mandatory and spontaneous breathing modes. See
more information on BiLe‐ vel ventilation.
All spontaneous (except for manual Inspirations).
for more information on CPAP.
Breath types must be defined before settings can be specified. There are only two categories of breath type: mandatory and spontaneous. Mandatory breaths are volume controlled (VC) or pressure controlled (PC or VC+).
The ventilator currently offers spontaneous breaths that are pressure supported (PS) volume supported (VS), tube compensated (TC), or proportionally assisted (PAV+™). The following figure shows the modes and breath types available on the ventilator.
Figure 94.
Illustrated Modes and Breath Types
10
The mode setting defines the interaction between the ventilator and the patient.
• Assist/control (A/C) mode allows the ventilator to control ventilation within boundaries specified by the practitioner. All breaths are mandatory, and can be PC, VC, or VC+.
• Spontaneous (SPONT) mode allows the patient to control ventilation. The patient must be able to breathe independently, and exert the effort to trigger ventilator support.
• Synchronous Intermittent Mandatory Ventilation (SIMV) is a mixed mode that allows a combination of mandatory and spontaneous interactions. In SIMV, the breaths can be spontaneous or mandatory, mandatory breaths are synchronized with the patient’s inspiratory efforts, and breath delivery is determined by the f setting.
229
• BiLevel is a mixed mode that combines both mandatory and spontaneous breath types. Breaths are delivered in a manner similar to SIMV mode with PC selected, but providing two levels of pressure. The patient is free to initiate spontaneous breaths at either pressure level during BiLevel.
Changes to the mode are phased in at the start of inspiration. Mandatory and spontaneous breaths can be flowor pressure-triggered.
The ventilator automatically links the mandatory type setting to the mode setting. During A/C or SIMV modes, once the operator has specified volume or pressure, the ventilator displays the appropriate breath parameters.
Changes in the mandatory type are phased in at the start of inspiration.
10.15.5 Respiratory Rate (f)
The f setting determines the minimum number of mandatory breaths per minute for ventilator-initiated mandatory breaths in A/C, SIMV, and BiLevel modes.If the mode is A/C or SIMV and VC is the breath type, specifying
⩒
MAX
and flow pattern determines T
I
, T
E
, and I:E. In PC breaths, specifying T
I automatically determines the other timing variables. See
Section 10.15.13, Inspiratory Time, page 232 for an explanation of the interdependencies of f,
T
I
, T
E
and I:E. Changes to the f setting are phased in at the start of inspiration.
The ventilator does not accept a proposed f setting if it would cause the new T the T
I
I
or T
E
to be less than 0.2 second,
to be greater than 8 s, or I:E ratio greater than 4.00:1. (The ventilator also applies these restrictions to a proposed change to the apnea respiratory rate, except that apnea I:E cannot exceed 1.00:1. An exception to this rule occurs in BiLevel ventilation where the proposed f setting will allow the I:E ratio to be greater than 4.00:1 only until the minimum T
L
is reached).
10.15.6 Tidal Volume
The tidal volume (V
The delivered V
T
T
) setting determines the volume of gas delivered to the patient during a VC mandatory breath.
is compensated for BTPS and patient circuit compliance. Changes to the V in at the start of inspiration. The V
T
setting only affects the delivery of mandatory breaths.
T
setting are phased
When proposing a change to the V flow pattern, and T
PL or less than 0.2 s, or a T
E
T
setting, the ventilator compares the new value with the settings for f, ⩒
. If the proposed setting would result in an I:E ratio that exceeds 4.00:1 or a T less than 0.2 s, the ventilator disallows the change.
MAX
,
I
greater than 8 s
10.15.7 Peak Inspiratory Flow
The peak inspiratory flow ( ⩒
MAX
) setting determines the maximum rate of delivery of tidal volume to the patient during mandatory VC breaths, only. Changes to ⩒
MAX
are phased in at the start of inspiration. Mandatory breaths are compliance compensated, even at the maximum ⩒
MAX
setting. Circuit compliance compensation does not cause the ventilator to exceed the ventilator’s maximum flow capability.
When proposing a change to the ⩒ flow pattern, and T
T
I
MAX
setting, the ventilator compares the new value with the settings for V
PL
. It is impossible to set a new ⩒ greater than 8.0 s or less than 0.2 s, or a T
E
MAX
T
, f,
that would result in an I:E ratio that exceeds 4.00:1, or a
less than 0.2 s.
10.15.8 Plateau Time
The plateau time (T
PL
) setting determines the amount of time inspiration is held in the patient’s airway after inspiratory flow has ceased. T
PL
is available only during VC mandatory breaths (for A/C and SIMV mode, and operator-initiated mandatory breaths). T are phased in at the start of inspiration.
PL
is not available for PC mandatory breaths. Changes to the T
PL
setting
When proposing a change to the T
PL
setting, the ventilator computes the new I:E ratio and T
I settings for V
T
, f, ⩒
MAX
4.00:1, or a T
I
, and flow pattern. It is impossible to set a new T part of the inspiratory phase.
PL
, given the current
that would result in an I:E ratio that exceeds
greater than 8 s or less than 0.2 s, or a T
E
less than 0.2 s. For the I:E ratio calculation, T
PL
is considered
230
10
10.15.9 Flow Pattern
The flow pattern setting defines the gas flow pattern of volume-controlled (VC) mandatory breaths only. The selected values for V are held constant, T
I
T
and ⩒
MAX
apply to both the square or descending ramp flow patterns. If V
T
and ⩒
MAX
and
approximately halves when the flow pattern changes from descending ramp to square (and approximately doubles when flow pattern changes from square to descending ramp), and corresponding changes to the I:E ratio also occur. Changes in flow pattern are phased in at the start of inspiration.
The settings for flow pattern, V
T
, f, T
PL
, and ⩒
MAX
are interrelated. If any setting change would cause any of the following, the ventilator does not allow that change
• I:E ratio >4:1
• T
I
>8.0 s or T
I
<0.2 s
• T
E
<0.2 s
10.15.10 Flow Sensitivity
The flow sensitivity ( the patient circuit during the ventilator’s expiratory phase. Once a value for flow sensitivity is selected, the
Reductions to ⩒
SENS
⩒
SENS
) setting defines the rate of flow inspired by a patient that triggers the ventilator to deliver a mandatory or spontaneous breath. When ventilator delivers a base flow equal to ⩒
⩒
TRIG
is selected, a base flow of gas (1.5 L/min) travels through
SENS
+1.5 L/min (base flow is not user- selectable). When the patient inhales and their inspiratory flow exceeds the ⩒
SENS
setting, a trigger occurs and the ventilator delivers a breath.
are phased in immediately, while increases are phased in at the start of exhalation.
When ⩒
SENS
is active, it replaces pressure sensitivity (P
SENS
⩒
SENS controlled, and apnea ventilation). When ⩒
SENS
). The ⩒
SENS
is active, a backup P
setting has no effect on the P
SENS
setting of 2 cmH the patient’s inspiratory effort, even if the flow sensors do not detect flow.
2
SENS
can be active in any ventilation mode (including pressure supported, volume controlled, pressure
setting.
O is in effect to detect
Although the minimum ⩒
SENS
setting of 0.2 L/min (adult/pediatric circuit types) or 0.1 L/min (neonatal circuit type) can result in autotriggering, it can be appropriate for very weak patients. The maximum setting of 20 L/min
(adult/pediatric circuit types) or 10 L/min (neonatal circuit type) is intended to avoid autotriggering when there are significant leaks in the patient circuit.
10.15.11 Pressure Sensitivity
The pressure sensitivity (P
SENS
) setting selects the pressure drop below baseline (PEEP) required to begin a patient-initiated breath (either mandatory or spontaneous). Changes to P
P
SENS
setting has no effect on the ⩒
SENS
SENS
are phased in immediately. The
setting and is active only if the trigger type is P
TRIG
.
Lower P
SENS
settings provide greater patient comfort and require less patient effort to initiate a breath. However, fluctuations in system pressure can cause autotriggering at very low settings. The maximum P autotriggering under worst-case conditions if patient circuit leakage is within specified limits.
SENS
setting avoids
10.15.12 Inspiratory Pressure
The inspiratory pressure (PI) setting determines the pressure at which the ventilator delivers gas to the patient during a PC mandatory breath. The P
I
setting only affects the delivery of PC mandatory breaths. The selected PI is the pressure above PEEP. (For example, if PEEP is set to 5 cmH to the patient at 25 cmH
2
O). Changes to the P
I
2
O, and P
I
is 20 cmH
2
O, the ventilator delivers gas
setting are phased in at the start of inspiration.
The sum of PEEP + P
I
+2 cmH
2
O cannot exceed the high circuit pressure ( ⤒ P
PEAK
) limit. To increase this sum of pressures, first raise the ⤒ P
PEAK
5 cmH
2
limit before increasing the settings for PEEP or P
I
O and the maximum value is 90 cmH
2
O.
. The minimum value for P
I
is
10
231
10.15.13 Inspiratory Time
The inspiratory time (T mandatory breaths. The ventilator accepts a setting as long as the resulting I:E ratio and T
Changes to T
I
I
setting) determines the time during which an inspiration is delivered to the patient for PC
phase in at the start of inspiration. Directly setting T
I
E
settings are valid.
in VC mandatory breaths is not allowed.
The ventilator rejects settings that result in an I:E ratio greater than 4.00:1, a T
I a T
E
less than 0.2 s to ensure the patient has adequate time for exhalation.
greater than 8 s or less than 0.2 s, or
Setting f and T
I
automatically determines the value for I:E and T
E
.
This equation summarizes the relationship between T
I
, I:E, T
E
, and breath period time:
If the f setting remains constant, any one of the three variables (T
I
, I:E, or T
E
) can define the inspiratory and expiratory intervals. If the f setting is low (and additional spontaneous patient efforts are expected), T
I
can be a more useful variable to set than I:E. As the f setting increases (and the fewer patient-triggered breaths are expected), the I:E setting becomes more relevant. Regardless of which variable is chosen, a breath timing bar always shows the interrelationship between T
I
, I:E, T
E
and f.
10.15.14 Expiratory Time
The expiratory time (T
Changes to the T
E value for I:E ratio and T
I of f, T
I
, T
E
, and I:E.
E
) setting defines the duration of exhalation for PC and VC+ mandatory breaths, only.
setting are phased in at the start of exhalation. Setting f and T
E
automatically determines the
. See Section 10.15.13, Inspiratory Time, page 232
for an explanation of the interdependencies
10.15.15 I:E Ratio
The I:E ratio setting is available when I:E is selected as the constant during rate change. The I:E setting determines the ratio of inspiratory time to expiratory time for mandatory PC breaths. The ventilator accepts the specified range of direct I:E ratio settings as long as the resulting T
I
and T
E
settings are within the ranges established for mandatory breaths. Changes to the I:E ratio phase in at the start of inspiration. Directly setting the I:E ratio in VC
mandatory breaths is not allowed. See Section 10.15.13, Inspiratory Time, page 232
for an explanation of the interdependencies of f, T
I
, T
E
, and I:E.
Setting f and I:E automatically determine the values for T
I
and T
E
. The maximum I:E ratio setting of 4.00:1 is the maximum that allows adequate time for exhalation and is intended for inverse ratio pressure control ventilation.
10.15.16 High Pressure in BiLevel
The high pressure level (P
H
) setting is the pressure level entered by the operator for the inspiratory phase of the mandatory breath in BiLevel ventilation.
10.15.17 Low Pressure in BiLevel
The low pressure level (P
L
) setting is the pressure level entered by the operator for the expiratory phase of the mandatory breath in BiLevel ventilation.
10.15.18 High Time in BiLevel
The high time (T
H
) setting is the duration of time (in seconds) the ventilator maintains the set high pressure level in BiLevel ventilation.
232
10
10.15.19 Low Time in BiLevel
The low time (T
L
) setting is the duration of time (in seconds) the ventilator maintains the set low pressure level in
BiLevel ventilation.
10.15.20 T
H
:T
L
Ratio in BiLevel
The ratio of T
H
to T
L
in BiLevel ventilation, similar to I:E ratio when ventilating a patient without BiLevel.
10.15.21 PEEP
This setting defines the positive end-expiratory pressure (PEEP), also called baseline airway pressure. PEEP is the positive pressure maintained in the patient circuit during exhalation. Changes to the PEEP setting are phased in at the start of exhalation.
The sum of:
• PEEP +7 cmH
2
O
• PEEP + P
I
+2 cmH
2
O (if PC is active)
• PEEP + P
SUPP
+2 cmH
2
O (if PS is active) cannot exceed the ⤒ P
PEAK the settings for PEEP, P
I
limit. To increase the sum of pressures, first raise the ⤒ P
PEAK
, or P
SUPP
.
limit before increasing
If there is a loss of PEEP from occlusion, disconnect, Safety Valve Open, or loss of power conditions, PEEP is reestablished (when the condition is corrected) by the ventilator delivering a PEEP restoration breath. The PEEP restoration breath is a 1.5 cmH
2
O pressure-supported breath with exhalation sensitivity of 25%, and rise time% of
50%. A PEEP restoration breath is also delivered at the conclusion of vent startup. After PEEP is restored, the ventilator resumes breath delivery at the current settings.
Note: PEEP restoration breath parameters are not user adjustable.
10.15.22 Pressure Support
The pressure support (P
SUPP
) setting determines the level of positive pressure above PEEP applied to the patient’s airway during a spontaneous breath. P
SUPP
is only available in SIMV, SPONT, and BiLevel, in which spontaneous breaths are allowed. The PSUPP setting is maintained as long as the patient inspires, and patient demand determines the flow rate. Changes to the ⤒ P
SUPP
setting are phased in at the start of inspiration. The pressure support setting affects only spontaneous breaths.
The sum of PEEP + P
SUPP
⤒ P
PEAK
+2 cmH
2
O cannot exceed the ⤒ P
limit before increasing the settings for PEEP or P considered safe for the patient, a P maximum safe circuit pressure.
SUPP
PEAK
SUPP
limit. To increase the sum of pressures, first raise the
. Since the ⤒ P
setting that would cause a ↑ P
PEAK
PEAK
limit is the highest pressure
alarm requires reevaluating the
10.15.23 Volume Support
Volume support (V
T SUPP
Changes to the to the V
) is defined as the volume of gas delivered to the patient during spontaneous VS breaths.
T SUPP
setting are phased in at the start of inspiration.
10.15.24 % Supp in TC
In TC, the% Supp setting represents the amount of the imposed resistance of the artificial airway the TC breath type will eliminate by applying added pressure at the patient circuit wye. For example, if the % Supp setting is
100%, TC eliminates 100% of the extra work imposed the by the airway. At 50%, TC eliminates 50% of the added work from the airway. TC is also used with BiLevel, and is available during both P
H
and P
L
phases.
10
233
10.15.25 % Supp in PAV+™
In PAV+™, the % Supp setting represents the percentage of the total work of breathing provided (WOB) by the ventilator. Higher inspiratory demand yields greater support from the ventilator. The patient performs the remaining work. If the total WOB changes (resulting from a change to resistance or compliance) the percent support remains constant.
10.15.26 Rise Time%
The rise time% setting allows adjustment of the speed at which the inspiratory pressure reaches 95% of the target pressure. Rise time settings apply to PS (including a setting of 0 cmH
2
O), VS, PC, or VC+ breaths. The higher the value of rise time%, the more aggressive (and hence, the more rapid) the rise of inspiratory pressure to the target
(which equals PEEP + P
I
(or P
SUPP
)). The rise time% setting only appears when pressure-based breaths are available.
The range of rise time% is 1% to 100%. A setting of 50% takes approximately half the time to reach 95% of the target pressure as a setting of 1.
• For mandatory PC, VC+, or BiLevel breaths, a rise time setting of 1 produces a pressure trajectory reaching 95% of the inspiratory target pressure (PEEP + P
I
in 2 s or 2/3 of the T
I
, whichever is shortest).
• For spontaneous breaths (VS, or PS), a rise time setting of 1 produces a pressure trajectory reaching 95% of the inspiratory target (PEEP + P
SUPP
) in (0.4 × PBW-based T
I
TOO LONG × 2/3) s.
• When both PC and PS breaths are active, the slopes and thus the pressure trajectories can appear to be different. Changes to T
I
and P
I
cause PC pressure trajectories to change. Changes in rise time% are phased in at the start of inspiration.
• When P
SUPP
=0, the rise time% setting determines how quickly the ventilator drives circuit pressure to
PEEP +1.5 cmH
2
O.
10.15.27 Expiratory Sensitivity
The expiratory sensitivity (E
SENS
) setting defines the percentage of the measured peak inspiratory flow at which the ventilator cycles from inspiration to exhalation in all spontaneous breath types. When inspiratory flow falls to the level defined by E
SENS
Changes to E
SENS
, exhalation begins. E
SENS
is a primary setting and is accessible from the GUI screen.
are phased in at the next patient-initiated spontaneous inspiration.
E
SENS
complements rise time%. Rise time% should be adjusted first to match the patient’s inspiratory drive, and then the E
SENS
setting should cause ventilator exhalation to occur at a point most appropriate for the patient. The higher the E
SENS
setting, the shorter the inspiratory time. Generally, the most appropriate E
SENS
is compatible with the patient’s condition, neither extending nor shortening the patient’s intrinsic inspiratory phase.
E
SENS
in a PAV+™ breath is expressed in L/min instead of percent.
10.15.28 Disconnect Sensitivity
Leak Sync disabled: Disconnect sensitivity (D
SENS
) is defined as the percentage of returned volume lost due to a leak, above which the ventilator declares a CIRCUIT DISCONNECT alarm. When D
SENS
is set to its lowest value
(20%) it has the highest sensitivity for detecting a leak or disconnect. Conversely, when D
SENS
is set to its highest value (95%), the ventilator is least sensitive to declaring a leak or disconnect, because greater than 95% of the returned volume must be lost before the alarm annunciates. During NIV, the D
SENS
value is automatically set to OFF, which means that returned volume loss is not considered and the alarm will not sound.
Leak Sync enabled: Disconnect sensitivity (D a disconnect and vice versa.
SENS
) is defined as the leak at PEEP value in L/min above which the ventilator declares a CIRCUIT DISCONNECT alarm. The lowest setting is most sensitive to detecting and declaring
To set D
SENS
with NIV interfaces when Leak Sync is enabled
1. After adjusting the patient settings, start ventilation.
2. Ensure that Leak Sync is enabled.
234
10
3. With the NIV interface open to ambient (not connected to the patient), use the patient data leak value to quantify the leak in L/min.
4. Set the D
SENS
(in L/min) below the leak rate (in L/min).
5. Periodically assess the leak rate, especially with PEEP changes, and adjust the D
SENS
setting as needed.
6. Always use alternative methods of monitoring during NIV.
Note: If D
SENS
is set to OFF during NIV, the ventilator is still capable of declaring a CIRCUIT DISCONNECT alarm.
Note: D
SENS
cannot be turned OFF if Leak Sync is enabled.
Changes to D
SENS
are phased in at the start of inspiration.
10.15.29 High Spontaneous Inspiratory Time Limit
The high spontaneous inspiratory time limit setting ( ⤒ T
I SPONT
) is available only in SIMV or SPONT modes during
NIV, and provides a means for setting a maximum inspiratory time after which the ventilator automatically transitions to exhalation. The default ⤒ T
I SPONT
setting is based upon circuit type and PBW.
For pediatric/adult circuit types, the new patient default value is
(1.99 +(0.02 × PBW)) s
For neonatal circuit types, the new patient default value is
((1.00 +(0.10 × PBW) s
The ⤒ T
I SPONT
indicator appears on the primary display at the beginning of a ventilator-initiated exhalation and remains visible for as long as the ventilator truncates breaths in response to the ⤒ T indicator disappears when the patient’s inspiratory time returns to less than the start of inspiration.
⤒ T elapsed after the beginning of exhalation of the last truncated breath. Changes to
I SPONT
I SPONT
setting. The ⤒ T
I SPONT
setting, or after 15 s has
⤒ T
I SPONT
are phased in at the
10.15.30 Humidification Type
The humidification type setting sets the type of humidification system (heated expiratory tube, non-heated expiratory tube, or heat-moisture exchanger (HME) used on the ventilator and can be changed during normal ventilation or short self test (SST). Changes in humidification type phase in at the start of inspiration.
SST calibrates spirometry partly based on the humidification type. Changing the humidification type without rerunning SST can affect the accuracy of spirometry and delivery.
The accuracy of the exhalation flow sensor varies depending on the water vapor content of the expiratory gas, which depends on the type of humidification system in use. Because the temperature and humidity of gas entering the expiratory filter differ based on the humidification type being used, spirometry calculations also differ according to humidification type. For optimum accuracy, rerun SST to change the humidification type.
10.15.31 Humidifier Volume
The dry, compressible volume in mL of the humidification chamber for the humidification type entered during
SST. Only applies if a humidifier is used.
10.16 Safety Net
While the ventilator is designed to be as safe and as reliable as possible, Covidien recognizes the potential for problems to arise during mechanical ventilation, either due to user error, patient-ventilator interactions, or because of problems with the ventilator itself. Safety Net is a broad term that includes strategies for handling problems that arise in the “patient-ventilator“ system (patient problems) as well as strategies to minimize the impact of system faults on patient safety. In these scenarios, The ventilator is designed to alarm and to provide the highest level of ventilation support possible. in case of ventilator malfunction. If the ventilator is not capable of
10
235
ventilatory support, it opens the patient circuit and allows the patient to breathe from room air if able to do so (this emergency state is called safety valve open (SVO) . Safety mechanisms are designed to be verified periodically or to have redundancy. The ventilator is designed to ensure that a single-point failure does not cause a safety hazard or affect its ability to annunciate a high-priority audible alarm.
10.16.1 User Error
The ventilator is designed to prevent the operator from implementing settings that are clearly inappropriate for the patient’s predicted body weight (PBW). Each setting has either soft bounds (can be overridden) or hard bounds (no override allowed) that alert the operator to the fact that the settings may be inappropriate for the patient. In the event that the patient is connected without any parameters being specified, the ventilator enters
Safety PCV, a safe mode of ventilation regardless of the circuit type in use (neonatal, pediatric, or adult) or patient’s
PBW. Safety PCV is entered after POST, if a patient connection is made prior to settings confirmation. Safety PCV uses new patient default settings with exceptions shown in
Table 53.
Safety PCV Settings
Parameter
PBW f
TOT mode
Mandatory type
(total respiratory rate)
O
T
P
2
I
I
%
PEEP
Trigger type
P
SENS
⩒
SENS
↑ P
PEAK
↑ ⩒
E TOT alarm
↓ ⩒
E TOT alarm
↑ V
TE
alarm
↓ V
TE MAND
alarm
↓ V
TE SPONT
alarm
Circuit type
Humidification type
Humidifier volume
Safety PCV Value
Neonatal: 3 kg
Pediatric: 15 kg
Adult: 50 kg
A/C
PC
Neonatal: 25 1/min
Pediatric: 16 1/min
Adult: 16 1/min
Neonatal: 0.3 s Pediatric: 0.7 s Adult: 1 s
15 cmH
2
O
Neonatal: 40%
Pediatric: 100%
Adult: 100%
3 cmH
2
O
Neonatal: ⩒
TRIG
Pediatric: P
TRIG
Adult: P
TRIG
2 cmH
2
O
1.0 L/min
20 cmH
2
O
OFF
0.05 L/min
OFF
OFF
OFF
Last set value, or adult if none available
Set value, or non-heated exp tube if none available
Last set value, or 480 mL if none available
Note: In Safety PCV, expiratory pauses are not allowed.
236
10
10.16.2 Patient Related Problems
In case of patient problems, the ventilator remains fully operative and annunciates the appropriate alarm. The detection, response, and priority of each patient-related alarm is determined by the actual patient problem. See
for a comprehensive description of the patient alarm system.
10.16.3 System Related Problems
The ventilator is designed to prevent system faults. Its modular design allows the breath delivery unit (BDU) to operate independently of the graphical user interface (GUI) and several modules of the breath delivery subsystem have redundancy that, if certain faults occur, provides for ventilatory support using settings that do not depend on the suspect hardware. System faults include the following:
• Hardware faults (those that originate inside the ventilator and affect its performance)
• Soft faults (faults momentarily introduced into the ventilator that interfere with normal operation)
• Inadequate supply (AC power or external gas pressure)
• Patient circuit integrity (occluded or disconnected circuit)
10.16.4 Background Diagnostic System
The ventilator has an extensive system of continuous testing processes. If an error is detected in the background diagnostic system, the ventilator notifies the operator by posting an entry in the diagnostic log. If the ventilator experiences an anomaly which causes an unintended reset, the ventilator will recover from that reset and deliver a breath within 3 s without any operator intervention. After recovering from a reset, the ventilator uses the same settings that were in effect before the reset occurred.
The background test process compares monitored values of ventilator functions with expected values of ventilator sensors under normal conditions regardless of whether the ventilator is in Stand-By or is ventilating a patient. The ventilator will continue to ventilate the patient with the highest level of support possible, and may revert to one of the states described. See
Section 4.11, Ventilator Protection Strategies, page 114 .
Background tests include:
• Periodically initiated tests performed at intervals of a specific number of machine cycles. These tests check hardware components directly affecting breath delivery, safety mechanisms, and the GUI, and detect and correct corruption of control variable data.
• Boundary checks performed at every analog measurement. These checks verify measurement circuitry, including sensors.
Ventilation Assurance is a safety net feature invoked if the background diagnostics detect a problem with certain components in either the gas mix subsystem, the inspiratory subsystem, or the expiratory subsystem. Each subsystem has a backup ventilation strategy that allows ventilation to continue by bypassing the suspect components giving the operator time to replace the ventilator.
Mix BUV is invoked if the measured gas mix is significantly different from the set mix, if the accumulator pressure is out of range or if a fault is indicated in the mix PSOLs or flow sensors. During Mix BUV, the normal mix controller is bypassed and ventilation continues as set, except that the gas mix reverts to 100% oxygen or air, depending on where the fault indication was detected. Backup circuits then control the pressure in the accumulator to keep it in the proper range for the inspiratory module.
Inspiratory BUV is invoked if background diagnostics detect a problem in the inspiratory module (PSOL or flow
.
Table 54.
Inspiratory Backup Ventilation Settings
Backup ventilation parameter
PBW
Mode
Setting
Previously used setting during Vent Startup
A/C
237
10
Table 54.
Inspiratory Backup Ventilation Settings (continued) f
Backup ventilation parameter
Mandatory type
T
P
I
I
O
2
%
PEEP
Trigger type
Gas flow
Setting
PC
Neonatal: 25 1/min
Pediatric: 16 1/min
Adult: 16 1/min
Neonatal: 0.3 s
Pediatric: 0.7 s
Adult: 1 s
15 cmH
2
O
100% (21% if O
2
not available)
3 cmH
2
O
⩒
TRIG
; 2 L/min (adult/pediatric), 1.5 L/min
Controlled by pressure in the mix accumulator
During Inspiratory BUV, the delivery PSOL is disabled, but gas delivery is achieved via an inspiratory BUV solenoid valve, the gas flow being created by pressure in the mix accumulator.
Exhalation BUV is invoked if problems with the exhalation valve driver are detected. A backup analog circuit is enabled to control the exhalation valve though the more advanced control features (active exhalation valve control) are not functional.
Note: During Mix and Inspiratory BUV, gas supply to installed options is disabled.
Entry into BUV is logged in the alarm log and system diagnostic log, and the status display provides an indicator that the ventilator is in BUV and which subsystem is affected.
When in BUV, a high priority alarm is annunciated, and the GUI displays an alarm banner indicating BUV, blanks patient data, and a displays a pressure waveform.
If the ventilator cannot provide any degree of reliable ventilatory support and fault monitoring, then the ventilator alarms and enters the safety valve open (SVO) emergency state. During SVO, the ventilator deenergizes the safety, expiratory, and inspiratory valves, annunciates a high-priority alarm, and turns on the SVO indicator. During SVO, a patient can spontaneously inspire room air (if able to do so) and exhale. Check valves on the inspiratory and expiratory sides minimize rebreathing of exhaled gas during SVO. During SVO the ventilator:
• Displays the elapsed time without ventilatory support
• Does not display patient data (including waveforms)
• Does not detect patient circuit occlusion or disconnect conditions
Visible indicators on the ventilator’s GUI and status display illuminate when the ventilator is in the SVO state. Other safeguards built into the ventilator include a one-way valve (check valve) in the inspiratory pneumatic circuit allowing the patient to inhale through the safety valve (if able to do so) with limited resistance. This check valve also limits exhaled flow from entering the inspiratory limb to reduce the possibility of re-breathing exhaled CO gas.
2
10.17 Power On Self Test (POST)
Every time the ventilator is powered on or resets and at the beginning of short self test (SST) and extended self test (EST) it performs power on self test (POST). POST checks the integrity of the GUI and breath delivery subsystems and communication channels without operator intervention and takes approximately 12 s to complete.
If POST detects a major fault, qualified service personnel must correct the problem and successfully pass EST. See the Puritan Bennett™ 980 Series Ventilator Service Manual for more details on POST.
238
10
10.18 Short Self Test (SST)
SST is a short (about 5 minutes) and simple sequence of tests that verifies proper operation of breath delivery hardware (including pressure and flow sensors, checks the patient circuit (including tubing, humidification device, and filters) for leaks, and measures the circuit compliance and resistance. SST also checks the resistance of the expiratory filter. SST, in normal mode, can only be performed at start up, prior to initiation of ventilation.
Covidien recommends running SST every 15 days, between patients, and when changing the patient circuit or its configuration (including changing circuit type, adding or removing in-line water traps, or using a different type
or style of patient circuit). See Section 3.9.1.2, SST Test Sequence, page 83
. The ventilator does not allow access to SST if it senses a patient is connected.
10.19 Extended Self Test (EST)
EST verifies the integrity of the ventilator’s subsystems using operator participation. EST requires a “gold standard” test circuit and a stopper to block the patient wye. All test resources, including the software code to run EST, exist in the ventilator. EST testing, excluding tests of optional equipment such as the compressor and extended battery) takes about 10 minutes. If the compressor is used as the air source for EST and optional equipment is tested, then
EST takes approximately 15 minutes. See Section 3.9.2, EST (Extended Self Test), page 86 .
Warning: Do not enter Service mode with a patient attached to the ventilator. Serious injury could result.
10
239
240
11
11 Specifications
11.1 Overview
This chapter contains the following specifications for the Puritan Bennett™ 980 Series Ventilator:
• Physical
• Electrical
• Interface
• Environmental
• Performance (ranges, resolution, and accuracies for ventilator settings, alarm settings, and patient data)
• EMC compliance information
Warning: Due to excessive restriction of the Air Liquide™, SIS, and Dräger™ hose assemblies, reduced ventilator performance levels may result when oxygen or air supply pressures < 50 psi (345 kPa) are employed.
11.2 Measurement Uncertainty
Measurement uncertainties and the manner in which they are applied are listed in the following tables unless otherwise noted:
Table 55.
Performance Verification Equipment Uncertainty
Flow
Measured Parameter
Pressure
Oxygen Concentration
Temperature
Atmospheric Pressure
Offset
0.1001 SLPM
0.121594 cmH
2
O
0.0168 %O
2
0.886041°C
1.76 cmH
2
O
Gain
2.7642 % reading
0.195756 % reading
0.0973 % reading
0.128726 % reading
-
During breath delivery performance verification for flow and pressure based measurements, the equipment inaccuracy is subtracted from the acceptance specification as follows:
• Net Acceptance Gain = Requirement Specification Gain -–Measurement Uncertainty Gain
• Net Acceptance Offset = Requirement Specification Offset –Measurement Uncertainty Offset
• Acceptance Limit = ± [(Net Acceptance Offset) + (Net Acceptance Gain) × (Setting)]
• (Setting –Acceptance Limit) ≤ Measurement ≤ (Setting + Acceptance Limit)
For derived parameters, such as volume, compliance, etc., the individual sensor uncertainties are combined and applied as applicable to determine the acceptance limits.
11.3 Physical Characteristics
Table 56.
Physical Characteristics
Weight
Dimensions
Ventilator: 51.26 kg (113 lb) including BDU, GUI, standard base, and primary battery
BDU only: 31.3 kg (69 lb)
Ventilator and compressor: 171.2 kg (57 lb) including BDU, GUI, ventilator and compressor primary batteries, base assembly, and compressor
Compressor: 40.4 kg (89 lb) including base assembly
BDU only: 31.3 kg (69 lb)
Ventilator: 32 cm width by 30 cm depth by 111 cm height) (not including GUI screen)
241
11
Table 56.
Physical Characteristics (continued)
(12.5 in. width by 11.5 in. depth by 43.5 in. height (not including GUI screen)
Ventilator: 32 cm width by 30 cm depth by 148 cm height (including GUI screen)
(12.5 in. width by11.5 in. depth by 58 in. height) (including GUI screen) Stand‐ ard base: 58 cm width by 66 cm depth (22.5 in. width by 26 in. depth)
At a distance of 1 meter does not exceed 48 dBA at 5 L/min A-weighted sound pressure level, ventilator (average)
A-weighted sound pressure level, ventilator and compressor
At a distance of 1 meter does not exceed 54 dBA at 5 L/min
A-weighted sound power level, ventilator
A-weighted sound power level, ventilator and compressor
Connectors
Does not exceed 58 dBA below 500 mL/min
Does not exceed 63 dBA below 500 mL/min
Inspiratory and expiratory limb connectors are 22 mm OD conical fittings compliant with ISO 5356-1
Inspiratory and exhalation filters Refer to filter instructions for use for complete specifications
Pressure units (chosen by oper‐ ator)
Displayed weight units
Displayed length units
Hectopascal (hPa) or centimeters of water (cmH
2
O)
Kilograms (kg) or pounds (lb)
Centimeters (cm) or inches (in)
Table 57.
Pneumatic Specifications
Oxygen and air inlet supplies Pressure: 241 kPa to 600 kPa (35 psi to 87 psi) Flow: maximum of 200 L/min
Oxygen sensor life Up to 1 year. Operating life varies depending on oxygen usage and ambient temperature.
Gas mixing system Range of flow from the mixing system: Up to 150 L/min for adult patients.
Additional flow is available (peak flow to 200 L/min) for compliance com‐ pensation Up to 80 L/min for pediatric circuit type Up to 30 L/min for neonatal circuit type Leakage from one gas system to another: Meets IEC 80601-2-12 standard Operating pressure range: 35 psi to 87 psi (241 kPa to 600 kPa)
Table 58.
Technical Specifications
Maximum limited pressure (P
LIM max
Maximum working pressure (P
W max
)
)
Response time to change in FiO
2 to 90% O
2
setting from 21% O
(measured at the patient wye)
2
Measuring and display devices
A fixed pressure limit to the safety valve limits circuit pressure to <123 hPa (125 cmH
2
O) at the patient wye.
P
W max
is ensured by the high pressure limit ( ⤒ P
PEAK when P
I
is <100 cmH
2
O (98.07 hPa)
)
< 18 s for volumes >150 mL
< 19 s for volumes ≥30 mL but ≤150 mL
< 50 s for volumes ≥2 mL but <30 mL
Pressure measurements:
Type: solid state differential pressure transducer
Sensing position: inspiratory module; expiratory mod‐ ule
Mean circuit pressure (P
–20 cmH
2
O (–20 hPa) to 100 cmH
130 cmH
2
O (127 hPa)
Volume measurements:
MEAN
):
2
O (98 hPa)
Peak circuit pressure (P
PEAK
:–20 cmH
2
O (–20 hPa) to
242
11
Table 58.
Technical Specifications (continued)
Type: hot film anemometer
Sensing position: inspiratory module; expiratory mod‐ ule
Oxygen measurement:
Type: galvanic cell
Sensing position: inspiratory module
Up to 75 L/min Minute volume ( ⩒
E TOT
) capability, ventilator
Minute volume ( ⩒
E TOT
) capability, compressor Up to 40 L/min BTPS, including compliance compen‐ sation
Results of ventilator testing using circuits identified for use with the ventilator system
Internal Inspiratory filter bacterial/viral filtration effi‐ ciency
>99.999%
Internal Inspiratory filter particle filtration efficiency >99.97% retention of particles 0.3 µm nominal at
100 L/min flow
Internal Inspiratory filter resistance
Combined inspiratory limb resistance
0.2 cmH
2
O < resistance <2.2 cmH
2
0.2 cmH
2
O < resistance <1.7 cmH
2
O at 30 L/min flow
O at 15 L/min flow
0.2 cmH
2
O < resistance <5.5 cmH
2
0.2 cmH
2
O < resistance <1.7 cmH
2
O at 30 L L/min flow
O at 15 L/min flow
Exhalation filter particle filtration efficiency, reusable >99.97% retention of particles 0.3 mm nominal at
100 L/min flow
Exhalation filter bacterial/viral filtration efficiency, reus‐ able
>99.999%
Exhalation filter resistance (pediatric/adult, reusable and disposable)
Exhalation filter bacterial/viral filtration efficiency, dis‐ posable
< 2.0 cmH
2
<1.7 cmH
2
>99.999%
O at 30 L/min when new
O at 15 L/min when new
Exhalation filter particle filtration efficiency, disposable >99.97% retention of particles 0.3 mm nominal at
100 L/min flow
Exhalation filter resistance, disposable
Exhalation filter particle filtration efficiency
<2.5 cmH
2
O at 100 L/min when new
>99.97% retention of particles 0.3 mm nominal at
100 L/min flow
>99.999% Exhalation filter bacterial/viral filtration efficiency,
(neonatal, disposable)
Exhalation filter particle filtration efficiency (neonatal, disposable)
Exhalation filter resistance (neonatal, disposable)
>99.70% retention of particles 0.3 µm nominal at
30 L/min flow
Circuit compliance (acceptable ranges of VBS compli‐ ance for each patient type)
Inspiratory limb circuit resistance (acceptable ranges of
VBS inspiratory limb circuit resistance for each patient type)
Expiratory limb circuit resistance (acceptable ranges of
VBS Expiratory limb circuit resistance for each patient type)
<0.58 cmH
2
O at 2.5 L/min
Adult: 1.3 mL/cmH
2
O to 4.2 mL/cmH
Pediatric: 0.9 mL/cmH
2
2
O
O to 3.0 mL/cmH
2
O
Neonatal: 0.4 mL/cmH
2
O to 1.5 mL/cmH
2
O
Adult: (at 60 L/min): 1.15 cmH
2
O to 11.0 cmH
2
Pediatric: (at 30 L/min): 0.46 cmH
2
Neonatal: (at 10 L/min): 0.37 cmH
cmH
2
O
2
O
O to 4.5 cmH
2
O
O to 4.5 (6.0 for Prox)
Adult: (at 60 L/min): 1.15 cmH
2
O to 11.0 cmH
Pediatric: (at 30 L/min): 0.46 cmH
2
2
O
O to 4.5 cmH
2
O
11
243
Table 58.
Technical Specifications (continued)
Audio alarm volume (primary)
Measurement uncertainty: ± 3 dBA
Neonatal: (at 10 L/min): 0.37 cmH
2
cmH
2
O
O to 4.5 (6.0 for Prox)
Range: High priority alarm volume range (dBA): 58 (vol‐ ume setting 1) to 86 (volume setting 10)
Medium priority alarm volume range (dBA): 52 (volume setting 1) to 78 (volume setting 10)
Low priority alarm volume range (dBA): 50 (volume setting 1) to 76 (volume setting 10)
Measured 1 m from front, rear, and sides of ventilator
See
Section 6.5.4, Alarm Volume Key, page 138
for alarm volume behavior during an alarm condition.
Resolution: 1
Minimum 64 dBA measured 1 m from front, rear, and sides of ventilator.
Audio alarm volume (secondary)
Measurement uncertainty: ± 3 dBA
11.4 Electrical Specifications
Table 59.
Electrical Specifications
Electrical ratings, ventilator
Electrical ratings, ventilator and compressor
Mains overcurrent release
Earth leakage current
Touch current
Patient Leakage current
11.5 Interface Requirements
The pin-out for the RS-232 interface is as follows:
Table 60.
Interface Pin Designations
6
7
4
5
Pin
1
2
3
Signal
N/C
RxD
TxD
N/C
GND
N/C
RTS
100 V ~, 50–60 Hz, 2.25 A
120 V ~, 50–60 Hz 1.5 A
220–240 V ~, 50–60 Hz, 0.75 A
100V~, 50–60 Hz, 8.25 A
120V~, 50–60 Hz, 6.0 A
220–240V~, 50–60 Hz, 3.0 A
Ventilator: 4 A
Compressor: 6 A
Meets requirements of IEC 60601-1, type BF applied part
Meets requirements of IEC 60601-1, type BF applied part
Meets requirements of IEC 60601-1, type BF applied part
Not connected
Name
Receive data
Transmit data
Not connected
Ground
Not connected
Request to send
244
11
Table 60.
Interface Pin Designations (continued)
Pin
8
9
Signal
CTS
N/C
The pin-out for the nurse call interface is as follows:
Table 61.
Nurse Call Pin Designations
Pin
1
2
3
4
Normally closed (NC)
Relay common
Normally open (NO)
Not connected
Clear to send
Not connected
Name
Configuration
11.6 Environmental Specifications
The following table provides the environmental conditions appropriate for using the ventilator. Use the ventilator only in these specified conditions:
Table 62.
Environmental Specifications
Specification
Temperature
Atmospheric Pressure
Altitude
Relative Humidity
Operation
10°C to 40°C (50°F to 104°F) ventila‐ tor
10°C to 35°C (50°F to 95°F) internal battery charger
70 kPa to 106 kPa (10.15 psi to
(15.37 psi)
–411.5 m to 3048 m (–1350 ft to 10
000 ft)
10% to 95% non-condensing
Storage
–20°C to 70°C (–4°F to 158°F)
50 kPa to 106 kPa (7.25 psi to
15.37 psi)
6096 m max (20000 ft max)
10 to 95% non-condensing
Note: When using the compressor, reduced dryer performance may be expected if relative humidity exceeds 50% when temperature is 40°C.
When using the compressor, reduced dryer performance may be expected if temperature exceeds 32.8°C when relative humidity is 95%.
Note: The limits marked on the device label represent out-of-box storage conditions as follows:
• Temperature: (10°C to 40°C (50°F to 104°F)
• Pressure: 70 kPa to 106 kPa (10.15 psi to 15.37 psi)
• Relative Humidity: 10% to 95% non-condensing
11.7 Performance Specifications
11.7.1 Ranges and Resolutions
for ranges and resolutions for ventilator settings, Table 64, page 252 for alarm settings, and
for displayed patient data parameters.
11
245
Table 63.
Ventilator Settings Range and Resolution
Setting
Apnea ventilation
Description
A safety mode of ventilation that starts if the patient does not receive a breath for an elapsed time exceed‐ ing the apnea interval.
Apnea expiratory time (T
E
) For mandatory PC apnea breaths, the time interval between the end of inspiration and the beginning of the next inspiration.
Apnea I:E ratio In PC breath types, specifies the ratio of apnea inspiratory time to apnea expiratory time.
Range and resolution
See individual apnea settings.
Range: 0.20 s to 59.8 s
Resolution: 0.01 s
Apnea respiratory rate (f
A
) Sets the number of volume- or pres‐ sure-based breaths per minute for ventilator initiated mandatory (VIM) apnea breaths.
Apnea tidal volume (V
T
) Sets the volume of gas delivered to the patient’s lungs during a manda‐ tory, volume-controlled apnea breath. Apnea tidal volume is com‐ pensated for body temperature and pressure, saturated (BTPS) and the compliance of the patient circuit.
Range: I:E ≤ 1.00:1
Resolution:
0.01 for values > 1:10.0;
0.1 for values ≤ 1:10 and > 1:100;
1 for values ≤ 1:100
Range: square, descending ramp
(
Apnea flow pattern The flow shape of the delivered mandatory volume-based (VC) apnea breath.
Apnea inspiratory pressure
(P
I
)
The pressure above PEEP at which gas is delivered to the patient dur‐ ing mandatory PC apnea breaths.
Apnea inspiratory time (T
I
) Same as inspiratory time for nonap‐ nea ventilation
Apnea interval (T
Apnea O
Apnea peak inspiratory flow
⩒
MAX
)
2
%
A
)
Range: 5 cmH
2
O to 90-PEEP cmH
Resolution:1 cmH
2
O
The time after which the ventilator transitions to apnea ventilation
T
A
≥ 60/f
A
Determines the oxygen concentra‐ tion in a standard mixture of air and oxygen.
The maximum rate of tidal volume delivery during mandatory volumebased apnea breaths.
Range: 0.20 s to 8 s
Resolution:
0.01 s in PC or VC+, 0.02 s in VC
Range: 3 s to 60 s or OFF in CPAP
Resolution: 1 s
Range: 21% O
2
Resolution: 1%
to 100% O
2
2
O
Range: When mandatory type is VC:
NEONATAL: 1 L/min to 30 L/min
PEDIATRIC: 3.0 L/min to 60 L/min
ADULT: 3.0 L/min to 150 L/min
Resolution:
0.1 L/min for flows
< 20 L/min (BTPS);
1 L/min for flows ≥ 20 L/min (BTPS)
Range: 2.0 1/min to 40 1/min
Resolution:
0.1 1/min for 2.0 1/min to 9.9 1/min;
1 1/min for 10 1/min to 40 1/min
Range:
NEONATAL: 3 mL to 315 mL
PEDIATRIC/ADULT:
≥ 25 mL to 2500 mL
Resolution: 0.1 mL for values <20 mL;
0.5 mL for values ≥20 mL and <25 mL;
1 mL for values ≥25 mL and <100 mL; 5
246
11
Table 63.
Ventilator Settings Range and Resolution (continued)
Setting Description
Apnea constant during rate change
Specifies which of the three opera‐ tor-adjustable breath timing varia‐ bles remains constant when respi‐ ratory rate is changed during apnea ventilation.
Apnea mandatory type
Circuit type
The type of mandatory breath deliv‐ ered during apnea ventilation.
Specifies the circuit for which com‐ pliance and resistance values dur‐ ing SST have been calculated.
Constant during rate change Specifies which of the three opera‐ tor-adjustable breath timing varia‐ bles remains constant when respi‐ ratory rate is changed.
Disconnect sensitivity
(D
SENS
)
Leak Sync disabled: The percentage of returned volume lost, above which the ventilator declares a CIR‐
CUIT DISCONNECT alarm.
Leak Sync enabled: The leak at PEEP value in L/min, above which the ventilator declares a CIRCUIT DIS‐
CONNECT alarm.
Range and resolution
mL for values ≥100 mL and <400 mL;
10 mL for values ≥400 mL
Range: T
I
Range: PC, VC
Range: NEONATAL, PEDIATRIC, ADULT
Range: I:E ratio, T
I
, T
E
T
H
:T
L
ratio, T
H
,T
L
for PC or VC+ breaths;
in BiLevel
Expiratory sensitivity (E
Expiratory time (T
E
)
SENS
) The percentage of ⩒
MAX
that, when reached, causes the ventilator to cycle from inspiration to exhalation during spontaneous, pressurebased breaths.
For PC or VC+ breaths, the time interval between the end of inspi‐ ration and the beginning of the next inspiration. The end of the exhala‐ tion phase is considered to be when the flow rate at the patient wye
Range (Leak Sync disabled): 20% to 95% or
OFF
Range (Leak Sync enabled:
NEONATAL:
Invasive: 1 L/min to 15 L/min
NIV: 1 L/min to 30 L/min
PEDIATRIC: 1 L/min to 40 L/min
ADULT: 1 L/min to 65 L/min
Resolution (Leak Sync disabled): 1%
Resolution (Leak Sync enabled):
0.5 L/min for values <10 L/min;
1 L/min for values ≥10 L/min
Range: 1% to 80% when Spontaneous
Type is PS, or VS
1 L/min to 10 L/min when Spontaneous
Type is PAV+.
Resolution:1% when Spontaneous Type is
PS, TC, or VS; 1 L/min when Spontaneous
Type is PAV+™.
Note: Default value is not expected to need adjustment. Only adjust after becom‐ ing experienced with PAV+™ and only if it is suspected that the ventilator is not cycling at the patient’s end-of-inspiration.
Range: ≥ 0.20 s
Resolution: 0.01 s
11
247
Table 63.
Ventilator Settings Range and Resolution (continued)
Gender
Height
Setting
Flow pattern
Flow sensitivity ( ⩒
SENS
)
Description remains less than 0.5 L/min above the base flow.
The flow shape of the delivered mandatory or VC breath
For flow triggered breaths, deter‐ mines the volume of flow (below the base flow) required to begin a mandatory or spontaneous patient initiated breath.
The patient’s gender
The patient’s height
High spontaneous inspira‐ tory time limit (
Humidification type
Humidifier volume
Elevate O
I:E ratio
2
%
⤒ T
Inspiratory pressure (P
Inspiratory time (T
I SPONT
I
)
I
)
)
Active in NIV only, allows the oper‐ ator to select the maximum spon‐ taneous inspiratory time.
The type of humidification system used on the ventilator.
The empty fluid volume of the cur‐ rently installed humidifier.
The percentage of O to the current air/O minutes.
2
2
to be added
mixture for 2
In PC and VC+ breath types, speci‐ fies the ratio of inspiratory time to expiratory time.
The pressure above PEEP at which gas is delivered to the patient dur‐ ing mandatory PC breaths.
The time during which an inspira‐ tion is delivered to the patient dur‐ ing mandatory PC or VC+ breaths.
Range and resolution
Range: square, descending ramp
Range:
NEONATAL: 0.1 L/min to 10 L/min
PEDIATRIC/ADULT:
0.2 L/min to 20.0 L/min
Resolution: 0.1 L/min
Range: male or female
Range:
19.5 cm to 280 cm; 7.5 in to 110 in.
Resolution:
0.5 cm for heights < 35 cm;
1 cm for heights < 254 cm;
2 cm for heights ≥ 254 cm;
0.25 in for heights < 14 in; 0.5 in for heights
<100 in.
1 in for heights ≥ 100 in.
See Section 4.6, Predicted Body Weight (PBW)
.
Range:
NEONATAL: 0.2 s to 1.7 s
PEDIATRIC/ADULT: 0.4 s to 5 s
Resolution: 0.1 s
Range: HME, non-heated expiratory tube, heated expiratory tube
Range: 100 mL to 1000 mL
Resolution: 10 mL
Range: 1% to 100%
Resolution:1% between 1% and 10; 5% between 5% and 75%; jumps to 100% when increased above 75%
Range: 1:299 to 149:1
Resolution: 0.01 for values > 1:10; 0.1 for values ≤ 1:10.0 and > 1:100.0;
1 for values ≤ 1:100
Displayed as XX:1 when I:E ≥ 1; displayed as
1:XX when I:E < 1
Range:5 cmH
2
O to 90 cmH
2
Resolution:1 cmH
2
O
O
Range: 0.2 s to 8 s for mandatory PC and
VC+ breaths
T
PL
+ 0.2 s to 8 s in VC
Resolution: 0.01 s for PC or VC+ breaths;
0.02 s for VC breaths
248
11
Table 63.
Ventilator Settings Range and Resolution (continued)
Leak Sync (leak compensa‐ tion)
Setting
Mandatory type mL/kg ratio
Description
Compensates for leaks during INVA‐
SIVE or non-invasive (NIV) ventila‐ tion.
The type of mandatory breath deliv‐ ered in A/C, SPONT or SIMV modes.
SPONT mode allows mandatory type selection for operator initiated mandatory (OIM) breaths.
The default tidal volume/PBW ratio
(only adjustable in Service mode)
Mode
O
2
% (delivered)
The ventilation mode. The mode determines the allowable breath types:
A/C – (assist/control) – a mandatory mode allowing volume controlled
(VC), pressure controlled (PC), or
VC+ breath types.
SPONT – allows the patient to ini‐ tiate the breath. Applicable SPONT breath types are pressure support
(PS), volume support (VS), tube compensated (TC) or PAV+™.
SIMV – Synchronized Intermittent
Mandatory Ventilation – a mixed ventilatory mode providing manda‐ tory breaths and allowing a patient spontaneous breaths during the breath cycle.
BiLevel – a mixed ventilatory mode combining the attributes of both mandatory and spontaneous breaths incorporating two pressure levels, P
H
and P
L
.
Percentage of delivered oxygen in the gas mixture.
Peak inspiratory flow ( ⩒
MAX
) The maximum rate of tidal volume delivery during mandatory volumebased breaths.
PEEP Sets the positive end-expiratory pressure, defined as the pressure targeted in the patient circuit dur‐ ing exhalation.
Range and resolution
Range: enabled or disabled
Range: PC, VC, VC+
Range: 4.0 mL/kg to 10 mL/kg
Resolution: 0.5 mL/kg
Range: A/C, SPONT, SIMV, BiLevel but not available when ventilation type is NIV);
CPAP (only available when circuit type is neonatal and ventilation type is NIV)
Range: 21% to 100%
Resolution: 1%
Range: When mandatory type is VC:
NEONATAL: 1 L/min to 30 L/min
PEDIATRIC: 3.0 L/min to 60 L/min
ADULT: 3.0 L/min to 150 L/min
Resolution:
0.1 L/min for values < 20 L/min (BTPS);
1 L/min for values ≥ 20 L/min (BTPS)
Range: 0 cmH
2
O to 45 cmH
Resolution: 0.5 cmH
2
2
O
O from 0.0 cmH
19.5 cmH
2
45 cmH
2
O
O; 1 cmH
2
O from 20 cmH
2
2
O to
O to
11
249
Table 63.
Ventilator Settings Range and Resolution (continued)
P
P
H
L
Setting
Plateau time (T
PL
)
Predicted Body Weight
(PBW)
Description
The positive pressure during the insufflation phase in BiLevel venti‐ lation.
The positive pressure in the patient circuit during the expiratory phase of BiLevel ventilation.
The amount of time inspiration is held in the patient’s lungs after inspiratory flow ceases for volumebased mandatory breaths. Consid‐ ered part of inspiratory phase for I:E ratio calculations.
Indicates an approximation of the patient’s body weight based upon their gender and height (or length for neonatal patients). PBW deter‐ mines default limits and limits for breath delivery parameters.
Pressure sensitivity (P
SENS
) For pressure triggered breaths, determines the amount of pressure below PEEP required to begin a mandatory or spontaneous patient initiated breath.
Pressure support (P
SUPP
PS
) or The positive pressure above PEEP
(or P
L
in BiLevel) during a spontane‐ ous breath.
Respiratory rate ( f )
Rise time %
Sets the number of volume- or pres‐ sure-based breaths per minute for ventilator initiated mandatory (VIM) breaths in A/C, SIMV, and BiLevel modes.
Sets the speed at which inspiratory gas delivered to the patient reaches the pressure target in BiLevel, PC,
VC+ VS, or PS. Higher percentages of rise time produce inspiratory pres‐ sure trajectories with shorter time to the target value.
Spontaneous type
% Supp
The breath type for patient initiated spontaneous breaths in SIMV,
SPONT, and BiLevel modes.
In tube compensation, specifies the additional positive pressure desired
Range: 5 cmH
2
Range:
Range and resolution
NEONATAL:1.0 1/min to 150 1/min
PEDIATRIC/ADULT: 1.0 1/min to 100 1/ min
Resolution: 0.1 from 1.0 1/min to 9.9 1/ min;
1 1/min from 10 1/min to 150 1/ min
Range: 1% to 100%
Resolution: 1%
O to 90 cmH
2
Resolution: 1 cmH
2
O
O
Range: 0 cmH
2
O to 45 cmH
2
O
Resolution:
0.5 cmH
1 cmH
2
2
O from 0.0 to 19.5 cmH
O from 20 to 45 cmH
2
O
2
O;
Range: 0 s to 2 s
Resolution: 0.1 s
Range:
NEONATAL: 0.3 kg (0.66 lb) to 7.0 kg (15 lb) when the NeoMode 2.0 option is installed
PEDIATRIC: 7.0 kg (15 lb) to 24 kg (53 lb)
ADULT: ≥ 25 kg (55.12 lb) to 150 kg (330.69
lb)
Resolution: 0.01 kg for weights < 1 kg,
0.1 kg for weights ≥ 1 kg and < 10 kg, 1 kg for weights ≥ 10 kg
Range: 0.1 cmH
2
O to 20.0 cmH
2
Resolution: 0.1 cmH
2
O
O
Range: 0 cmH
Resolution: 5%
2
O to 70 cmH
Resolution: 1 cmH
2
O
Range: PS, TC, PAV+™, or VS
Range: 10% to 100%
2
O
250
11
Table 63.
Ventilator Settings Range and Resolution (continued)
% Supp
T
T
T
H
L
H
(time high)
(time low)
:T
L
ratio
Setting
Tidal volume (V
T
)
Volume support (V
VS
Trigger type
Tube ID
Tube type
Ventilation type
T SUPP
) or
(
Description to overcome resistance of the artifi‐ cial airway.
In PAV+™, specifies the percentage of total inspiratory work of breath‐ ing (WOB) performed by the venti‐ lator.
The duration of the insufflation phase during BiLevel ventilation.
The duration of the expiratory phase during BiLevel ventilation.
In BiLevel, specifies the ratio of insuf‐ flation time to expiratory time.
The volume of gas delivered to the patient during a mandatory vol‐ ume-based breath. V
T
compensates for body temperature and pressure, saturated (BTPS) and circuit compli‐ ance. Applicable for volume-based breaths.
The volume of gas delivered to the patient during spontaneous, vol‐ ume supported breaths.
Determines whether flow changes
⩒
TRIG
) or pressure changes (P trigger patient breaths
TRIG
)
The internal diameter of the artificial airway used to ventilate the patient.
Range and resolution
Range: 5% to 95%
Resolution: 5%
Range: 0.2 s to 30 s
Resolution: 0.01 s
Range: ≥ 0.20 s
Resolution: 0.01 s
Range: 1:299 to 4:1; in BiLevel T
H
:T
L
Resolution: 0.01 for < 10.00:1 and > 1:10.00;
0.1for [< 100.0:1 and ≥ 10.0:1] or [≤ 1:10.0
and > 1:100.0]; 1 for < 1:100.0 or ≥ 100:1
Range:
NEONATAL: 2 mL to 315 mL
PEDIATRIC: 25 mL to 1590 mL
ADULT: 75 mL to 2500 mL
Resolution:
0.1 mL for values <20 mL; 0.5 mL for values
≥20 mL and <25 mL; 1 mL for values
≥25 mL and <100 mL; 5 mL for values
≥100 mL and <400 mL; 10 mL for values
≥400 mL
Range:
NEONATAL: 2 mL to 310 mL
PEDIATRIC: 25 mL to 1590 mL
ADULT: 75 mL to 2500 mL
Resolution: 0.1 mL for values <20 mL;
0.5 mL for values ≥20 mL and <25 mL; 1 mL for values ≥25 mL and <100 mL; 5 mL for values ≥100 mL and <400 mL; 10 mL for values ≥400 mL
Range:
NEONATAL: ⩒
TRIG
PEDIATRIC/ADULT: ⩒
TRIG
or P
TRIG
Range: 4.5 mm to 10 mm when spontane‐ ous type is TC
Range: 6 mm to 10 mm when spontaneous type is PAV+™
Resolution: 0.5 mm
Range: Endotracheal (ET), tracheal (Trach) The type of artificial airway used to ventilate the patient.
Invasive or non-invasive (NIV) ven‐ tilation type based upon the type of breathing interface used. Invasive:
Range: Invasive, NIV
11
251
Table 63.
Ventilator Settings Range and Resolution (continued)
Setting Description
ET or Trach tubes. NIV: masks, infant nasal prongs, or uncuffed ET tubes.
Range and resolution
Table 64.
Alarm Settings Range and Resolution
Alarm volume
Setting
High exhaled minute volume alarm setting (
High exhaled tidal volume alarm setting ( limit ( ⤒ V
⤒⩒
⤒
TI
V
E TOT
TE
)
)
High inspired tidal volume alarm
The ↑ ⩒
E TOT
alarm indicates the measured total minute volume
≥ the set alarm limit.
The ↑ V
TE
alarm indicates that the measured exhaled tidal volume
≥ the set alarm limit for spontane‐ ous and mandatory breaths.
The ↑ V
TI
alarm indicates the deliv‐ ered volume of any breath ≥ the set alarm limit.
Description
Controls the volume of alarm annunciations
Apnea interval (T
A
) The apnea alarm condition indi‐ cates that neither the ventilator nor the patient has triggered a breath for the operator-selected apnea interval (T
A
). When the apnea alarm condition is true, the ventilator invokes mandatory ventilation as specified by the operator.
(
High circuit pressure setting
⤒ P
PEAK
)
The ↑ P
PEAK alarm level.
alarm indicates the patient’s airway pressure ≥ the set
Low circuit pressure setting ( ⤓ P
PEAK
) The ↓ P
PEAK
alarm indicates the measured airway pressure ≤ the set alarm limit during an NIV or VC+ inspiration.
Range and resolution
Range: 1 (minimum) to 10 (maxi‐ mum)
Resolution: 1
Range: 3 s to 60 s or OFF in CPAP
Resolution: 1
Range: 7 cmH
2 olution: 1 cmH
O to 100 cmH
2
2
O
O Res‐
Range:
NIV: OFF or ≥0.5 cmH
O
2
O to
<100 cmH
2
Resolution:
0.5 cmH
1 cmH
2
2
O for values <20.0 cmH
O for values ≥20 cmH
2
O
2
O;
Range: OFF and NEONATAL:
0.1 L/min to 10 L/min
PEDIATRIC:
0.1 L/min to 30 L/min
ADULT:
0.1 L/min to 100 L/min
Resolution:
0.005 L/min for values < 0.50 L/min;
0.05 L/min for values ≥ 0.5 L/min to
<5.0 L/min; 0.5 L/min for values
≥5.0 L/min
Range: OFF and NEONATAL: 5 mL to
500 mL
PEDIATRIC: 25 mL to 1500 mL
ADULT: 25 mL to 3000 mL
Resolution: 1 mL for values <100 mL;
5 mL for values ≥100 mL and
<400 mL; 10 mL for values ≥400 mL
Range: 6 mL to 6000 mL
Resolution: 1 mL for values <100 mL;
5 mL for values ≥100 mL to <400 mL;
10 mL for values ≥400 mL
252
11
Table 64.
Alarm Settings Range and Resolution (continued)
Setting
High respiratory rate alarm setting
( ⤒ f
TOT
)
High spontaneous inspiratory time limit ( ⤒ T
Low exhaled mandatory tidal vol‐ ume alarm setting (
Low exhaled minute volume alarm setting (
I SPONT
⤓⩒
)
E TOT
)
⤓ V
TE MAND
)
Low exhaled spontaneous tidal vol‐ ume alarm setting ( ⤓ V
TE SPONT
)
The ured breath rate ≥ the set alarm limit.
The ↑ T
I SPONT
indicator allows the operator to select the maximum spontaneous inspiratory time of an
NIV breath. No alarm is annunciated; only the symbol ⤒ T
I SPONT
appears on the screen near the NIV indicator when inspiration time exceeds the setting. If ⤒ T
I SPONT
is exceeded, the ventilator transitions from inspira‐ tion to exhalation.
The ↓ V
TE MAND
alarm indicates the measured mandatory tidal volume
≤ the set alarm limit.
The measured exhaled minute volume
≤ the set alarm limit for mandatory and spontaneous breaths.
The
↑
↓
↓ f
TOT
⩒
V
E TOT
Description
alarm indicates the meas‐
alarm indicates the
TE SPONT
alarm indicates the measured spontaneous tidal vol‐ ume ≤ the set alarm limit.
Range and resolution
Range: OFF or NEONATAL: 10 1/min to 170 1/ min
PEDIATRIC/ADULT: 10 1/min to 110
1/min
Resolution: 1 1/min
Range:
NEONATAL: 0.2 to ≤ the value of the
NIV inspiratory time limit trigger for the patient’s PBW and circuit type s
PEDIATRIC/ADULT: 0.4 s to ≤ the value of the NIV inspiratory time limit trigger for the patient’s PBW and circuit type s Resolution: 0.1 s
Range: OFF and
NEONATAL: 1 mL to 300 mL
PEDIATRIC: 1 mL to 1000 mL
ADULT: 1 mL to 2500 mL
Resolution: 1.0 mL for values
<100 mL; 5 mL for values ≥100 mL and <400 mL; 10 mL for values
≥400 mL
Range: OFF when ventilation type =
NIV and
NEONATAL: OFF, 0.01 L/min to
10 L/min
PEDIATRIC: 0.05 L/min to 30 L/min
ADULT: 0.05 L/min to 60 L/min
Resolution: 0.005 L/min for values
<0.50 L/min; 0.05 L/min for values
≥0.50 L/min and <5.0 L/min;
0.5 L/min for values >5.0 L/min
Range: OFF and
NEONATAL: 1 mL to 300 mL
PEDIATRIC: 1 to 1000 mL
ADULT: 1 to 2500 mL
Resolution: 1 mL for values <
100 mL;
5 mL from 100 mL to <400 mL;
10 mL for values ≥400 mL
Table 65.
Patient Data Range and Resolution
Data value
Breath phase
Description
The breath phase indicator displays the breath delivery phase (inspira‐
Range and resolution
Range: Control (C), Assist (A), Spon‐ taneous (S)
11
253
Table 65.
Patient Data Range and Resolution (continued)
Data value
Inspired tidal volume (V
TI
)
Description tion or exhalation) currently being delivered to the patient.
The volume inspired for a pressurebased breath.
Driving Pressure (P
DRIVE
Inspired tidal volume (V
Leak Sync
)
Dynamic compliance (C
Dynamic resistance (R
TL
DYN
DYN
End expiratory flow (EEF)
)
) during
End expiratory pressure (PEEP)
End inspiratory pressure (P
)
I END
)
Range and resolution
During PAV+™, P
DRIVE
is computed as the pressure difference between
Plateau Pressure and the estimated
PEEP
TOT
.
The volume inspired for each breath when Leak Sync is enabled.
The result of dividing the delivered tidal volume by the peak airway pressure.
The change in pressure per unit change in flow.
The rate of expiratory flow occur‐ ring at the end of exhalation.
The pressure at the end of the expir‐ atory phase of the previous breath
(also applies in BiLevel).
The pressure at the end of the inspir‐ atory phase of the current breath
(also applies in BiLevel).
Range: 0 mL to 6000 mL
Resolution:
0.1 mL for 0 mL to 9.9 mL;1 mL for values 10 mL to 6000 mL
Range: 0 cmH
2
O to 90 cmH
Resolution: 0.1 cmH
2
2
O
O for 0 cmH to 9.9 cmH
2
1.0 cmH
2
90 cmH
2
O
2
O
O;
O for values 10 cmH
2
O to
Range: 0 mL to 6000 mL
Resolution:
1 mL for values
<10 mL;
1 mL for values 10 mL to 6000 mL
Range:
0 mL/cmH
2
O to 200 mL/cmH
Resolution:0.1 mL/cmH ues
<10 mL/cmH
2
2
O
O for val‐
1 mL/cmH
2
≥10 mL/cmH
2
O;
O for values
2
O
Range: 0.0 cmH
2
O/L/s to
100 cmH
2
O/L/s
Resolution: 0.1 cmH
<10 cmH
1 cmH
2
≥10 cmH
2
2
O/L/s;
O/L/s
2
O/L/ for values
O/L/s for values
Range: 0 L/min to 150 L/min
Resolution:
0.1 L/min for values <20 L/min
1 L/min for values ≥20 L/min
Range:
–20.0 cmH
2
O to 130 cmH
2
O
Resolution:
0.5 cmH
2
O between
–10.0 cmH
2
O and +10.0 cmH
1 cmH
2
2
O;
O for values ≤ –10 cmH and ≥10 cmH
2
O
2
O
Range:
–20.0 cmH
2
O to 130 cmH
2
O
Resolution:
0.1 cmH
2
O for –20.0 cmH
2
O to
9.9 cmH
1.0 cmH
10 cmH
2
2
2
O;
O for values
O to 130 cmH
2
O
254
11
Table 65.
Patient Data Range and Resolution (continued)
Data value
Exhaled mandatory tidal volume
(V
TE MAND
)
Exhaled minute volume (
Exhaled spontaneous minute vol‐ ume ( ⩒
E SPONT
)
⩒
E TOT
)
Description
The exhaled volume of the last man‐ datory breath. When the mode is
SPONT, and no mandatory breaths have occurred for a time period ≥2 minutes, the V
TE MAND
indicator is hidden. Mandatory breaths can occurs during SPONT mode via manual inspiration.
A calculated sum of the volumes exhaled by the patient for manda‐ tory and spontaneous breaths for the previous 1-minute interval (also applies in BiLevel).
The sum of exhaled spontaneous volumes per minute (also applies in
BiLevel).
Range and resolution
Range: 0 mL to 6000 mL
Resolution:
0.1 mL for 0 mL to 9.9 mL;
1 mL for 10 mL to 6000 mL
Range: 0.00 L/min to 99.9 L/min
Resolution: 0.01 L/min for 0.00 to
9.99 L/min; 0.1 L/min for 10.0 L/min to 99.9 L/min
Exhaled spontaneous tidal volume
(V
TE SPONT
)
The exhaled volume of the last spontaneous breath.
Exhaled tidal volume (V
TE
)
Leak Sync exhaled tidal volume (V
TE
) The volume exhaled by the patient for the previous mandatory or spon‐ taneous breath during Leak Sync
(also applies in BiLevel).
I:E ratio
The volume exhaled by the patient for the previous mandatory or spon‐ taneous breath (also applies in BiLe‐ vel).
Inspiratory compliance (C
20
/C)
The ratio of the inspiratory time to expiratory time for the previous breath.
the ratio of compliance of the last
20% of inspiration to the compli‐ ance of the entire inspiration.
Intrinsic PEEP (PEEP
I
) A calculated estimate of the pres‐ sure above PEEP at the end of exha‐ lation.
Mean circuit pressure (P
MEAN
)
Range: 0 L/min to 99.9 L/min
Resolution:
0.01 L/min for 0.00 to 9.99 L/min;
0.1 L/min for 10.0 to 99.9 L/min
Range: 0 mL to 6000 mL
Resolution: 0.1 mL for 0 mL to
9.9 mL; 1 mL for 10 mL to 6000 mL
Range: 0 mL to 6000 mL
Resolution:
0.1 mL for 0 mL to 9.9 mL;
1 mL for 10 mL to 6000 mL
Range: 0 mL to 6000 mL
Resolution:
0.1 mL for 0 mL to 9.9 mL;
1 mL for 10 mL to 6000 mL
Range: 1:599 to 149:1
Resolution: 0.1 for 9.9:1 to 1:9.9; 1 for
149:1 to 10:1 and 1:10 to 1:599
Range: 0 to 1.00
Resolution: 0.01
The calculated average circuit pres‐ sure for an entire breath cycle including both inspiratory and expiratory phases (whether the breath is mandatory or spontane‐ ous).
Range:
–20.0 cmH
2
O to +130 cmH
2
O
Resolution:
0.1 cmH
2
O between -9.9
and +9.9 cmH
2
1 cmH
2
≥10 cmH
2
O
O;
O ≤–10 cmH
2
O and
Range:
–20.0 cmH
2
O to 100 cmH
2
O
Resolution:
0.1 cmH
1 cmH
2
2
O for –20.0 to 9.9 cmH
O for 10 to 100 cmH
2
O
2
O;
11
255
Table 65.
Patient Data Range and Resolution (continued)
Data value
Negative inspiratory force (NIF)
O
P
2
% (monitored)
0.1
Description
The negative pressure generated during a maximally forced inspira‐ tory effort against an obstruction to flow.
The monitored percentage of oxy‐ gen in the gas delivered to the patient, measured at the ventilator outlet upstream of the inspiratory filter.
The inspiratory depression of airway pressure after 100 ms of occlusion.
P
0.1
measures respiratory drive.
Range and resolution
Range: ≤0 cmH
2
O to ≥–50 cmH
Resolution: 1 cmH
≤–10 cmH
2
0.1 cmH
2
2
O for values
O;
O for values >–10 cmH
2
O
2
O
Range: 0% to 103%
Resolution: 1%
PAV based intrinsic PEEP (PEEP
I PAV
) The estimated intrinsic PEEP during a PAV+™ breath. Intrinsic PEEP is an estimate of the pressure above PEEP at the end of every pause exhala‐ tion.
PAV-based lung compliance (C
PAV
The calculated change in pulmo‐ nary volume for an applied change in patient airway pressure when measured under conditions of zero flow during a PAV+™ plateau maneuver. When PAV+™ is selected, the ventilator displays the current filtered value for patient compli‐ ance, and updates the display at the successful completion of each esti‐ mation. C
PAV
can be displayed in the vital patient data banner. See
Sec‐ tion 3.8.2.7, Vital Patient Data, page 79 .
PAV-based lung elastance (E
PAV
)
For a PAV+™ breath, E ted as the inverse of C above). E
PAV
PAV
PAV
is calcula‐
(see
can be displayed in the vital patient data banner. See
Sec‐ tion 3.8.2.7, Vital Patient Data, page 79 .
PAV-based patient resistance (R
PAV
The difference between estimated total resistance R
TOT
and the simul‐ taneously estimated resistance of the artificial airway. When PAV+™ is selected, the ventilator displays the current filtered value for patient resistance, and updates the display at the successful completion of each estimation. R
PAV can be dis‐ played in the vital patient data ban‐
Range: ≥–20 cmH
2
O to 0 cmH
2
O
Resolution:
1 cmH
2
O when <–10 cmH
2
O;
0.1 cmH
2
O when ≥–10 cmH
2
O
Range: 0 cmH
2
O to 130 cmH
2
Resolution: 0.1 cmH
<10 cmH
2
≥10 cmH
2
O; 1 cmH
2
O
2
O
O for values
O for values
Range: 2.5 mL/cmH
2
O to 200 mL/cmH
2
O
Resolution: 0.1 mL/cmH for values <10 mL/cmH
2 for values ≥10 mL/cmH
2
2
O
O; 1 cmH
2
O
O
Range: 5.0 cmH
2
O/L to
400 cmH
2
O/L
Resolution: 0.1 cmH
<10 cmH
2
1 cmH
2
O/L;
2
O/L ≥10 cmH
O/L for values
2
O/L
Range:
0.0 cmH
2
O/L/s to 60 cmH
2
O/L/s
Resolution:
0.1 cmH
2
O/L/s for values
<10 cmH
2
1 cmH
2
O/L/s;
O/L/s for val‐ ues≥10 cmH
2
O/L/s
256
11
Table 65.
Patient Data Range and Resolution (continued)
Data value
PAV-based total airway resistance
(R
TOT
PAV-based work of breathing
(WOB
TOT
)
Description ner. See
Section 3.8.2.7, Vital Patient
.
R
TOT
is an estimated value captured just past peak expiratory flow and is equal to the pressure loss across the patient plus respiratory system
(patient + ET tube + expiratory limb of the VBS)/expiratory flow. This pressure loss is divided by the expir‐ atory flow estimated at the same moment, yielding the estimate for
R
TOT
.The complete operation is orchestrated and monitored by a software algorithm. When PAV+™ is selected, the ventilator displays the current filtered value for total resist‐ ance, and updates the display at the successful completion of each cal‐ culation. R
TOT
can be displayed in the vital patient data banner. See
Section 3.8.2.7, Vital Patient Data, page 79 .
The estimated effort needed for patient inspiration including both patient and ventilator.
Range and resolution
Range: 1.0 cmH
2
O/L/s to
80 cmH
2
O/L/s
Resolution: 0.1 cmH ues < 10 cmH
2
O/L/s; 1 cmH for values ≥10 cmH
Resolution: 0.1 J/L
2
2
O/L/s for val‐
O/L/s
Range: 1.0 J/L to 10.0 J/L
2
O/L/s
Peak expiratory flow (PEF)
Peak circuit pressure (P
Plateau pressure (P
PL
)
PEAK
)
Peak spontaneous flow (PSF)
Proximal exhaled tidal volume (V
TEY
The maximum speed of exhalation. Range: 0 L/min to 150 L/min
Resolution:
0.1 L/min for PEF <20 L/min;
1 L/min for PEF ≥20 L/min
The maximum pressure during the previous breath, relative to the patient wye, including inspiratory and expiratory phases.
The maximum flow rate sampled during a spontaneous inspiration.
Range:
–20.0 cmH
2
Resolution: 0.1 cmH ues –20.0 cmH
2
1.0 cmH
130 cmH
2
2
O to 130 cmH
2
2
O
O for val‐
O to 9.9 cmH
2
O;
O for values 10 cmH
2
O
Range: 0 L/min to 200 L/min
Resolution:
0.1 L/min for values <20 L/min;
1 L/min for values ≥20 L/min
O to
The pressure measured during an inspiratory pause maneuver.
) For neonatal patients, the exhaled volume of the previous breath
Range:
–20.0 cmH
2
O to 130 cmH
2
O
Resolution:
0.1 cmH
2
O for values –20.0 to
9.9 cmH
2
1.0 cmH
2
O;
O for values ≥10 cmH
2
O
Range: 0 mL to 500 mL
11
257
Table 65.
Patient Data Range and Resolution (continued) ume ( ⩒
E TOTY
Data value
Proximal exhaled total minute vol‐
)
Description measured by the proximal flow sen‐ sor) (if installed).
For neonatal patients, the exhaled minute volume measured by the proximal flow sensor (if installed).
Proximal inspired tidal volume (V
TIY
) For neonatal patients, the exhaled volume of the previous breath measured by the proximal flow sen‐ sor) (if installed).
Spontaneous inspiratory time (T
I
SPONT
)
Spontaneous inspiratory time ratio
(T
I
/T
TOT
)
Spontaneous rapid shallow breath‐ ing index (f/V
T
)
The duration of the inspiratory phase of a spontaneous breath.
The fraction of the total spontane‐ ous breath time used by inspiration.
Static compliance (C
STAT
)
A calculated value using exhaled spontaneous tidal volume. High val‐ ues indicate the patient is breathing rapidly, but with little vol‐ ume/breath. Low values indicate the inverse scenario.
An estimate of the patient’s lungthorax static compliance or elastic‐ ity.
Resistance (R
STAT
Total PEEP (PEEP
)
TOT
Vital capacity (VC)
)
Total respiratory rate (f
TOT
)
Range and resolution
Resolution: 0.1 mL for values 0 mL to
9.9 mL; 1 mL for values 10 mL to
500 mL
Range: 0.00 L/min to 99.9 L/min
Resolution: 0.01 L/min for
0.00 L/min to 9.99 L/min;
0.1 L/min for 10.0 L/min to
99.9 L/min
Range: 0 mL to 500 mL
Resolution: 1 mL
Range: 0 s to 10 s
Resolution: 0.01 s
Range: 0 to 1
Resolution: 0.01
Range: 0.1 1/min-L to 600 1/min-L
Resolution: 0.1 1/min-L for values
<10 1/min-L; 1 1/min-L; for values
≥10 1/min-L
An estimate of the restrictiveness of the patient’s lungs and the artificial airway.
The estimated pressure at the circuit wye during an expiratory pause maneuver.
The number of mandatory or spon‐ taneous breaths/min delivered to the patient.
The maximum amount of air that can be exhaled after a maximum inhalation.
Range: 0 mL/cmH mL/cmH mL/cmH
2
2
2
O to 500
O Resolution: 0.1
mL/cmH
2 mL/cmH
2
O for values <10 mL/; 1
O for values ≥10
O
Range:
0 cmH
2
O/L/s to 500 cmH
2
O /L/s
Resolution:
0.1 cmH
2
O/L/s for values
<10 cmH
2
O/L/s; 1 cmH values ≥10 cmH
2
O/L/s
2
O/L/s for
Range:
–20.0 cmH
2
O to +130 cmH
2
O
Resolution:
0.1 cmH
2
O for values <10 cmH
1 cmH
2
O for values ≤–10 cmH and ≥10 cmH
2
O
2
2
O;
O
Range: 1 to 200 1/min Resolution:
0.1 1/min for values <10 1/min; 1
1/min for 10 1/min to 200 1/min
Range: 0 mL to 6000 mL
Resolution: 0.1 mL for values
<10 mL; 1 mL for values ≥10 mL
258
11
Table 65.
Patient Data Range and Resolution (continued)
Data value
V
LEAK
%LEAK
LEAK
LEAK
Y
Description
Inspiratory leak volume. The total volume delivered during inspiration to compensate for the leak.
Percent leak.The percentage of total delivered volume during inspiration attributed to the leak calculated as
(leak volume during inspiration / total delivered inspiratory volume)
×100.
Exhalation leak. The leak rate at PEEP during exhalation.
Exhalation Leak at PEEP during Leak
Sync measured by the proximal flow sensor.
Range and resolution
Range: 0 mL to 9000 mL
Resolution: 1 mL
Range: 0% to 100%
Resolution: 1%
Range: 0 L/min to 200 L/min
Resolution: 0.1 L/min
Range: 0 L/min to 200 L/min
Resolution: 0.1 L/min a If the estimated value of C
PAV
, E
PAV
, R
PAV
, or R
TOT
violates expected (PBW-based) limits, parentheses around the value indicate the value is questionable. If the estimated value exceeds its absolute limit, the limit value flashes in parentheses.
Table 66.
Delivery Accuracy
Parameter
Inspiratory pressure (P
I
)
End expiratory pressure (PEEP)
Pressure support (P
SUPP
)
Tidal volume (V
T
)
O
P
P
2
H
L
% (delivered)
Accuracy
±(3.0+2.5% of setting) cmH
2
O
±(2.0+4% of setting) cmH
2
O
±(3.0+2.5% of setting) cmH
2
O
For adult and pediatric circuit type settings:
For T
I
<600 ms: ±(10 + 10% of setting
×600 ms/T
I
ms) mL For T
I
≥600 ms
±(10 +10% of setting) mL
For neonatal circuit type settings:
For setting of 2 mL (VC+ only):
±(1 +10% of setting) mL
For setting of 3 mL to 4 mL: ±(2 +
10% of setting) mL (delivered vol‐ ume shall be ≥1 mL
For setting of 5 mL to 20 mL ±(3 +
15% of setting)
For setting of ≥20 mL: ±(4+10% of setting) mL
±3%
±(2.0 +4% of setting) cmH
2
O
±(2.0 +4% of setting) cmH
2
O
Range
5 cmH
2
O to 90 cmH
2
O
0 cmH
2
O to 45 cmH
2
O
0 cmH
2
O to 70 cmH
2
O
For adult and pediatric circuit type settings:
25 mL to 2500 mL
For neonatal circuit type settings:
2 mL to 310 mL
21% to 100%
5 cmH
2
O to 90 cmH
2
O
0 cmH
2
O to 45 cmH
2
O
Table 67.
Monitoring (Patient Data) Accuracy
Parameter
Peak circuit pressure (P
PEAK
)
Mean circuit pressure (P
MEAN
)
End expiratory pressure (PEEP)
Accuracy
±(2 +4% of reading) cmH
2
O
±(2 +4% of reading) cmH
2
O
±(2 +4% of reading) cmH
2
O
Range
5 cmH
2
O to 90 cmH
2
O
3 cmH
2
O to 70 cmH
2
O
0 cmH
2
O to 45 cmH
2
O
11
259
Table 67.
Monitoring (Patient Data) Accuracy (continued)
Parameter
End inspiratory pressure (P
I END
)
Inspired tidal volume (V
TI
)
Exhaled tidal volume (V
TE
)
Inspired tidal volume during Leak
Sync
Accuracy
±(2 +4% of reading) cmH
2
O
±(4 mL +15% of actual) mL
±(4 mL +10% of actual) mL
For adult and pediatric circuit type settings:
For T
I
≤600 ms: ±(10 +20%×600
ms of reading) mL ms/T
I
For T
I
>600 ms: (10 +20% of reading)
mL
For neonatal circuit type setting:
±(10 +20% of reading) mL
For readings <100 mL, the accuracy shall apply when the percentage of inspiratory leak volume is less than
80%
Exhaled tidal volume (V
Leak Sync
TE
) during For adult and pediatric circuit type settings:
For T
E ms/T
E
≤600 ms: ±(10+20% ×600 ms of reading) mL
For T
E
>600 ms: ±(10 + 20% of read‐ ing) mL
For neonatal circuit type settings:
±(10+20% of reading) mL
For readings <100 mL, the accuracy shall apply when the percentage of inspiratory leak volume is less than
80%
Proximal exhaled tidal volume (V
TEY
) ±(1 +10% of reading) mL
Proximal inspired tidal volume (V
TIY
) ±(1 +10% of reading) mL
O
2
% (monitored) ±3%
Respiratory Rate (f ) ±0.8 1/min
Range
5 cmH
2
O to 90 cmH
2
O
2 mL to 2500 mL
2 mL to 2500 mL
For adult and pediatric circuit type settings:
25 mL to 2500 mL
For neonatal circuit type settings:
2 mL to 310 mL
For adult and pediatric circuit type settings:
25 mL to 2500 mL
For neonatal circuit type settings:
2 mL to 310 mL
2 mL to 310 mL
2 mL to 310 mL
15% to 100%
1 1/min to 150 1/min
Table 68.
Computed Value Accuracy
Parameter Accuracy Range
PAV-based lung compliance (C
PAV
) ±(1+20% of measured value) mL/cmH
2
O
PAV based total airway resistance
(R
TOT
)
±(3 +20% of measured) cmH
2
10 to 100 mL/cmH
O/L/s 5.0 to 50 cmH
2
2
O/L/s
O
PAV based work of breathing
(WOB
TOT
)
±(0.5 +10% of measured work) J/L with a percent support setting of
75%
0.7 J/L to 4 J/L
Warning: The ventilator accuracies listed in this chapter are applicable under the operating conditions identified
Operation outside specified ranges cannot guarantee the accuracies listed in the tables above, and may supply incorrect information.
260
11
11.8 Manufacturer’s Declaration
The following tables contain the manufacturer’s declarations for the ventilator system electromagnetic emissions, electromagnetic immunity, separation distances between ventilator and portable and mobile RF communications equipment and a list of compliant cables.
Warning: Portable and mobile RF communications equipment can affect the performance of the ventilator system. Install and use this device according to the information contained in this manual.
Warning: The ventilator system should not be used adjacent to or stacked with other equipment, except as may be specified elsewhere in this manual. If adjacent or stacked used is necessary, the ventilator system should be observed to verify normal operation in the configurations in which it will be used.
Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas should be used no closer than 30 cm (12 in) to any part of the ventilator, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Caution: This equipment is not intended for use in residential environments and may not provide adequate protection to radio communication services in such environments.
Note: The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals
(CISPR 11 class A).If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the equipment.
Table 69.
Electromagnetic Emissions
The ventilator is intended for use in the electromagnetic environment specified below. The customer of the operator of the ventilator should assure that it is used in such an environment.
Emissions Test Compliance
Electromagnetic environment – guidance
Radiated RF emissions
CISPR 11
Group 1
Class A
The ventilator uses RF energy only for its internal functions.
The ventilator is intended to be used only in hospitals and not be connected to the public mains net‐ work.
Conducted emissions
CISPR 11
The ventilator is intended to be used only in hospitals and not be connected to the public mains net‐ work.
Harmonic emissions IEC 61000-3-2 Class A
Voltage fluctuations/flicker IEC
61000-3-3
Complies
The ventilator is intended to be used only in hospitals and not be connected to the public mains net‐ work.
Table 70.
Electromagnetic Immunity
The ventilator is intended for use in the electromagnetic environment specified below. The customer or the operator of the ventilator should assure that it is used in such an environment.
EMC test Test standard Test levels Remarks Electromagnetic envi‐ ronment—guidance
ESD
IEC 60601-1-2, Edi‐ tion 3.0:2007
IEC 60601-1-2, Edi‐ tion 4.0:2014
±2,4,6,8 kV contact discharge
±2,4,8,15 kV air dis‐ charge
N/A
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the rel‐
261
11
Table 70.
Electromagnetic Immunity (continued)
The ventilator is intended for use in the electromagnetic environment specified below. The customer or the operator of the ventilator should assure that it is used in such an environment.
EMC test Test standard Test levels Remarks Electromagnetic envi‐ ronment—guidance
IEC 61000-4-2 ative humidity should be at least 30%.
Radiated immunity
IEC 60601-1-2, Edi‐ tion 3.0:2007
IEC 61000-4-3
IEC 60601-1-2, Edi‐ tion 4.0:2014
IEC 61000-4-3
10 V/m
3 V/m
Modulation:
80% AM, 2 Hz
Modulation:
80% AM, 1 kHz
N/A
EFT burst
Surge
IEC 60601-1-2, Edi‐ tion 3.0:2007
IEC 61000-4-4
IEC 60601-1-2, Edi‐ tion 4.0:2014
IEC 61000-4-4
IEC 60601-1-2, Edi‐ tion 3.0:2007
IEC 60601-1-2, Edi‐ tion 4.0:2014
IEC 61000-4-5
±1 kV (I/O)
±2 kV (AC Mains)
±05 kV,1 kV line to line
±05 kV,1 kV & 2 kV line to earth
5 kHz pulse repetition rate
100 kHz pulse repetition rate
N/A
Mains power quality should be that of a typical hospital environment.
IEC 60601-1-2, Edi‐ tion 3.0:2007
IEC 61000-4-6
3 V RMS
10 V RMS in the fol‐ lowing frequency
);
• 6.765–6.795 MH z
• 13.553–13.567
MHz
Modulation:
80% AM, 2 Hz
Portable and mobile RF communications equip‐ ment should be used no closer to any part of the ventilator system, includ‐ ing cables, than the sepa‐ ration distance calculated from the equation appli‐ cable to the frequency of the transmitter. See
.
Conducted immunity
IEC 60601-1-2, Edi‐ tion 4.0:2014
IEC 61000-4-6
• 26.957–27.283
MHz
• 40.66–40.70 MH z
3 V RMS
6 V RMS in the fol‐ lowing frequency
);
• 6.765–6.795 MH z
• 13.553–13.567
MHz
• 26.957–27.283
MHz
• 40.66–
40.70 MHz
Modulation:
80% AM, 1 kHz
262
11
Table 70.
Electromagnetic Immunity (continued)
The ventilator is intended for use in the electromagnetic environment specified below. The customer or the operator of the ventilator should assure that it is used in such an environment.
EMC test Test standard Test levels Remarks Electromagnetic envi‐ ronment—guidance
Magnetic immunity
IEC 60601-1-2, Edi‐ tion 3.0:2007
IEC 60601-1-2, Edi‐ tion 4.0:2014
IEC 61000-4-8
30 A/m N/A
Power frequency mag‐ netic fields should be at levels characteristic of a typical hospital environ‐ ment.
NOTE: U
T
is the AC mains voltage prior to application of the test level.
• 95% minimum voltage reduc‐ tion for 0.5 peri‐ ods (10 ms)
Voltage dips
IEC 60601-1-2, Edi‐ tion 3.0:2007
IEC 61000-4-11
IEC 60601-1-2, Edi‐ tion 4.0:2014
IEC 61000-4-11
• 60% minimum voltage reduc‐ tion for 5 peri‐ ods (100 ms)
• 30% minimum voltage reduc‐ tion for 25 peri‐ ods (500 ms)
• U
T
=0%, 0.5
cycle (0, 45, 90,
135, 180, 225,
270, and 350°)
• U
T
=0%; 1 cycle
• U
T
=70%; 25/30 cycles (@0°)
N/A
Interrupts
Proximity field from
RF wireless commu‐ nication equipment
IEC 60601-1-2, Edi‐ tion 3.0:2007
IEC 60601-1-2, Edi‐ tion 4.0:2014
IEC 61000-4-11
IEC 60601-1-2, Edi‐ tion 4.0:2014
IEC 61000-4-3
• U
See
T
=0%; 250/300 cycles
.
Modulation:
See
RFID immunity
AIM Standard
7351731 Rev. 2.00
2017
IEC 61000-4-3
See
.
See section 7 in
AIM Standard
7351731 for more details on execution of the different
RFID specifica‐ tions.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Mains power should be that of a typical hospital environment. If the opera‐ tor of the ventilator requires continuous oper‐ ation during power mains interruptions, it is recom‐ mended that the ventila‐ tor be powered from an uninterruptible power supply or a battery.
N/A
N/A
11
263
Table 70.
Electromagnetic Immunity (continued)
The ventilator is intended for use in the electromagnetic environment specified below. The customer or the operator of the ventilator should assure that it is used in such an environment.
EMC test Test standard Test levels Remarks Electromagnetic envi‐ ronment—guidance
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 to 6,795 MHz; 13.553 MHz to
13.567 MHz; 26.957 MHz; and 40.66 MHz to 40.70 MHz. The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz are intended to decrease the likelihood mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 is used in calculating the separation distance for transmitters in these frequency ranges.
Table 71.
Immunity to Proximity Fields RF Wireless Communications Equipment
Test fre‐ quency
(MHz)
Band (MHz) Service Modulation Maximum power (W)
Distance (m) Immunity test level
(V/m)
385 380–390
TETRA 400 Pulse modula‐ tion
18 Hz
1,8 0,3 27
450 430–470
• GMRS
460
• FRS 460
FM
±5 kHz deviation
1 kHz sine
2 0,3 28
710
745
780
810
870
930
704–787
800–960
LTE Band 13,
17
• GSM
800/900
• TETRA
800
• iDEN 820
• CDMA
850
• LTE Band
5
Pulse modula‐ tion
217 Hz
Pulse modula‐ tion
18 Hz
0,2
2
0,3
0,3
9
28
264
11
Table 71.
Immunity to Proximity Fields RF Wireless Communications Equipment (continued)
Test fre‐ quency
(MHz)
1720
1845
1970
2450
Band (MHz)
1700–1990
2400–2570
Service
• GSM
1800
• CDMA
1900
• GSM
1900
• GSM
1900
DECT
• LTE Band
1, 3, 4, 25
• UMTS
• Bluetoot h
• WLAN,
802.11
b/g/n
• RFID
2450
• LTE Band
7
Modulation
Pulse modula‐ tion
217 Hz
Pulse modula‐ tion
217 Hz
Maximum power (W)
2
2
Distance (m)
0,3
0,3
Immunity test level
(V/m)
28
28
5240
5500
5785
5100–5800
WLAN
802.11a/n
Pulse modula‐ tion
217 Hz
0,2 0,3 9
Table 72.
AIM Standard Test Levels
RFID specification
ISO 14223
ISO/IEC 14443-3 (Type A)
ISO/IEC 14443-4 (Type B)
ISO/IEC 15693 (ISO 18000-3 Mode 1)
ISO 18999-3 Mode 3
ISO/IEC 18000-7
ISO/IEC 18000-63 Type C
ISO/IEC 18000-4 Mode 1
Frequency
134.2 kHz
13.56 MHz
13.56 MHz
13.56 MHz
13.56 MHz
433 MHz
860–960 MHz
2.45 GHz
Test level (RMS)
65 A/m
7.5 A/m
7.5 A/m
5 A/m
12 A/m
3 V/m
54 V/m
54 V/m
11
265
Table 73.
Recommended Separation Distances for RF
The ventilator is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the operator of the ventilator can help prevent electromagnetic inter‐ ference by maintaining a minimum distance between portable and mobile RF communications equip‐ ment (transmitters) and the ventilator as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter (W)
150 kHz to 80 MHz outside of ISM bands
150 kHz to 80 MHz inside of ISM bands
80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01
0.1
1
10
100
0.117
0.37
1.17
3.7
11.7
0.12
0.38
1.2
3.8
12
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, as determined by an electromagnetic site survey the compliance level in each frequency range the following symbol:
, should be less than
. Interference may occur in the vicinity of equipment marked with
a The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to
2.5 GHz are intended to decrease the likelihood mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 is used in calculating the separation distance for transmitters in these frequency ranges.
b Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the 980 Series Ventilator is used exceeds the applicable
RF compliance level above, the 980 Series Ventilator should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the ventilator.
c Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Warning: The use of accessories and cables other than those specified with the exception of parts sold by
Covidien as replacements for internal components, may result in increased emissions or decreased immunity of the ventilator system.
Table 74.
Recommended Cables
Part number and description
10087151, Power cord, 10A, RA, ANZ
10087159, Power cord, 10A, RA, UK
10087155, Power cord, 10A, RA, EU
10087157, Power cord, 10A, RA, Japan
10 ft (3 m)
10 ft (3 m)
10 ft (3 m)
10 ft (3 m)
Cable length
266
11
Table 74.
Recommended Cables (continued)
Part number and description
10087152, Power cord, 10A, RA, British
10087154, Power cord, 10A, RA, Swtzrlnd
10081056, Power cord, 10A, RA, USA
10087156, Power cord, 10A, RA, Israel
10087160, Power cord, 10A, RA, Brazil
10087153, Power cord, 10A, RA, China
10 ft (3 m)
10 ft (3 m)
10 ft (3 m)
10 ft (3 m)
10 ft (3 m)
10 ft (3 m)
Cable length
11.9 Safety Tests
All safety tests should be performed by qualified service personnel at the interval specified. See
11.10 Essential Performance Requirements
Per ISO/EN 80601-2-12: 2011, Medical electrical equipment Part 2-12: Particular requirements for basic safety and essential performance of critical care ventilators, the ventilator’s essential performance requirements are given in
Ventilator Settings, Alarm Settings, and Patient Data tables earlier in this chapter. Alarms, including Oxygen level alarms and gas failure alarms, are identified in Chapter
. AC and battery backup power information is
given in Chapter Chapter 3 , and gas failure cross flow information is given in Chapter Chapter 3
.
If essential performance is lost or degraded due to exposure of electromagnetic disturbance levels higher than
, the following may occur:
• Component failures
• Changes in programmable parameters or settings
• Reset to default settings
• Changes to operating mode
• Initiation of an unintended operation
• Error in delivered volume of individual breaths greater than 35%
• Error in delivered minute volume greater than 25%
• False positive alarm condition
• Failure to alarm
11
267
268
12 Appendix BiLevel 2.0
12.1 Overview
This appendix describes the operation of the BiLevel 2.0 ventilation mode on the Puritan Bennett™ 980 Series
Ventilator.
BiLevel is a mixed mode of ventilation that combines attributes of mandatory and spontaneous breathing, with the breath timing settings determining which breath type is favored. In BiLevel Mode, mandatory breaths are always pressure-controlled, and spontaneous breaths can be pressure-supported (PS) or tube compensated (TC).
Figure 95.
Spontaneous Breathing at P
L
1 P
CIRC
(cmH
2
O)
2 T
H
3 T
L
4 P
H
5 P
L
6 Spontaneous breaths
BiLevel resembles SIMV mode, except that BiLevel establishes two levels of positive airway pressure. Cycling between the two levels can be triggered by BiLevel timing settings or by patient effort.
Figure 96.
BiLevel Mode
12
1 Pressure (y-axis)
2 P
L
3 P
H
4 Spontaneous breath
5 Synchronized transitions
6 Pressure support
7 Time-based transitions
The two pressure levels are called Low Pressure (P
L
) and High Pressure (P
BiLevel monitors mandatory and spontaneous tidal volumes separately.
H
). At either pressure level, patients can breathe spontaneously, and spontaneous breaths can be assisted with tube compensation or pressure support.
Inspiratory time and expiratory time in BiLevel become Time high (T
H these inspiratory and expiratory times, P
H
is maintained during T
H
) and Time low (T
and P
L
L
), respectively. During
is maintained during T
L
.
12
269
12.2 Intended Use
BiLevel is intended for adult, pediatric, and neonatal patients.
12.3 Safety Term Definitions
This section contains safety information for users, who should always exercise appropriate caution while using the ventilator.
Table 75.
Safety Term Definitions
Symbol
WARNING
CAUTION
NOTE
Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse events) to the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the prod‐ uct.
Note
Notes provide additional guidelines or information.
Warning: The ventilator offers a variety of breath delivery options. Throughout the patient’s treatment, the clinician should carefully select the ventilation mode and settings to use for that patient based on clinical judgment, the condition and needs of the patient, and the benefits, limitations and characteristics of the breath delivery options. As the patient’s condition changes over time, periodically assess the chosen modes and settings to determine whether or not those are best for the patient’s current needs.
12.4 Setting Up BiLevel
BiLevel is a ventilatory mode (along with A/C, SIMV, and SPONT).
To set up BiLevel
1. At the ventilator setup screen, enter PBW or gender and height.
2. Touch BiLevel . After selecting BiLevel mode, the ventilator uses the PC mandatory breath type, which cannot be changed.
3. Choose the spontaneous type (PS or TC).
4. Choose trigger type (P
TRIG
or ⩒
TRIG
).
5. Select desired ventilator settings. The default settings for BiLevel mode appear. To change a setting, touch its button and turn the knob to set its value. P
H
must always be at least 5 cmH
2
O greater than P
L
.
6. Set T
L
, T
H
, or the ratio of T
H
to T
L
. To select settings that would result in a T
H
:T
L
ratio greater than 1:1 or 4:1, you must touch Continue to confirm after reaching the 1:1 and 4:1 limits.
270
Figure 97.
BiLevel Setup Screen
12
7. Touch Start.
8. Set apnea and alarm settings by touching their respective tabs at the side of the ventilator settings screen and changing settings appropriately.
Note: The rise time% setting determines the rise time to reach target pressure for transitions from P
L spontaneous breaths, even when pressure support (P spontaneous breaths.
SUPP
) =0. Expiratory sensitivity (E
SENS
to P
) applies to all
H
and for
12.5 Using Pressure Support with BiLevel
Spontaneous breaths in BiLevel mode can be assisted with pressure support according to these rules (see
):
• Pressure support (P to P
L
SUPP
) can be used to assist spontaneous breaths at P
L
. Target pressure = P
L
+ P
SUPP
.
and P
H
. P
SUPP
is always set relative
• Spontaneous patient efforts at P
H
are not pressure supported unless P
SUPP
> (P
H
– P
L
). All spontaneous breaths
(whether or not they are pressure supported) are assisted by a pressure of 1.5 cmH
2
O.
• If P
SUPP
+ P
L
is greater than P
H
+1.5 cmH
2 all spontaneous breaths at P
H
O, all spontaneous breaths at P
L
are assisted by P
SUPP
–(P
H
–P
L
).
are assisted by the P
SUPP
setting, and
• All spontaneous breaths not supported by PS or TC (for example, a classic CPAP breath) are assisted with an inspiratory pressure of 1.5 cmH
2
O.
For example, if P
L
=5 cmH
2
O, P
H
=15 cmH
2
O, and P
SUPP
=20 cmH
2
O:
• All spontaneous breaths at P
L
25 cmH
2
O, and
are assisted by 20 cmH
2
O of pressure support (P
L
+ P
SUPP
) for a total pressure of
• All spontaneous breaths in P
H pressure of 25 cmH
2
O.
are assisted by 10 cmH
2
O of pressure support (P
SUPP
–(P
H
– P
L
)) for the same total
12
271
Figure 98.
BiLevel with Pressure Support
1 Pressure (y-axis)
2 P
H
Pressure support = 10 cmH
2
O
3 P
L
Pressure support = 20 cmH
2
O
4 P
H
5 P
L
During spontaneous breaths, the pressure target is calculated with respect to P
L
.
12.6 Manual Inspirations in BiLevel Mode
Pressing the manual inspiration key during BiLevel mode causes the ventilator to:
• Cycle to P
H
, if the current pressure level is P
L
.
• Cycle to P
L
, If the current pressure level is P
H
.
To avoid breath stacking, the ventilator does not cycle from one pressure level to another during the earliest stage of exhalation.
12.7 Respiratory Mechanics Maneuvers in BiLevel
In BiLevel, respiratory mechanics maneuvers are limited to inspiratory pause and expiratory pause maneuvers.
12.8 Specifications
See
Table 63, page 246 for the following specifications:
1. Low pressure (P
L
)
2. High pressure (P
H
)
3. Low pressure time (T
L
)
4. High pressure time (T
H
)
5. T
H
:T
L
ratio
6. Respiratory rate (f )
7. Rise time%
12.9 Technical Description
BiLevel is a mode of ventilation that alternately cycles between two operator-set pressure levels, P
L
and P
H pressure durations are defined by operator-set timing variables T
L levels, P
L
and P
H
, are analogous to breath phase transitions in PC.
and T
H
. The
. Transitions between the two pressure
At the extreme ranges of T
L
and T
H
, BiLevel can resemble the single breath type mode A/C —PC, or the more complex breath type mode, an “inverted-like” IMV. If T
H
and T
L
assume “normal” values with respect to PBW (for
272
example T
H
:T
L
>1:2 or 1:3), then BiLevel assumes a breathing pattern similar to, if not qualitatively identical to A/C
—PC. However, as T
L
begins to shorten with the T
H
and abrupt release to P
L
would match the pattern patented by John Downs
H
:T
L
ratio extending beyond 4:1, the breathing pattern assumes a distinctly different shape. In the extreme, the exaggerated time at P
In between the A/C—PC-like pattern and the APRV-like pattern, there would be patterns with moderately long T
H and T
L
intervals, allowing the patient sufficient time to breathe spontaneously at both P
H
and P
L spontaneous breath types. In this sense, BiLevel and SIMV are classified as mixed modes.
In these types of breathing patterns, (but less so with APRV) BiLevel, like SIMV, can be thought of as providing both mandatory and
Direct access to any of the three breath timing parameters in BiLevel is accomplished by touching the Padlock icon associated with the T
H
period, T
L
period or the T
H
:T
L
ratio displayed on the breath timing bar in the setup screen.
While in BiLevel mode, spontaneously triggered breaths at either pressure level can be augmented with higher inspiratory pressures using Pressure Support (PS) or tube compensation (TC) breath types.
12.9.1 Synchrony in BiLevel
Just as BiLevel attempts to synchronize spontaneous breath delivery with the patient’s inspiratory and expiratory efforts, it also attempts to synchronize the transitions between pressure levels with the patient’s breathing efforts.
This allows T
H the T
L
to be extended to prevent transitions to P
L
during the patient’s spontaneous inspiration. Likewise,
interval may be extended to prevent a transition to P
H
during the patient’s spontaneous exhalation.
The trigger sensitivity setting (P
SENS
P
H
T
L
down to P
and T pattern.
H
L
or ⩒
SENS
) is used to synchronize the transition from P
intervals as necessary to synchronize the transitions between P
L
and P
H
L
to P
H
. The transition from
is synchronized with the patient’s spontaneous expiratory effort. The BiLevel algorithm will vary the
to match the patient’s breathing
The actual durations of T
H
and T
L
vary according to whether or not the patient makes any spontaneous inspiratory efforts during those periods.
To manage synchrony with the patient’s breathing pattern, the BiLevel algorithm partitions the T
H
into spontaneous and synchronous intervals as shown in Figure 99
.
and T
L
periods
Figure 99.
Spontaneous and Synchronous Intervals
12
1 Pressure (y-axis)
2 T
H
3 P
H
4 P
L
5 T
L
6 Synchronous interval
7 Spontaneous interval
By partitioning T
H
and T
L
into spontaneous and synchronous phases, BiLevel responds to patient efforts (or lack of them) in a predictable pattern:
12
4 Downs, JB, Stock MC. Airway pressure release ventilation: A new concept in ventilatory support. Crit Care Med
1987;15:459–461
273
• During the spontaneous interval of each pressure level, successful inspiratory efforts cause the ventilator to deliver spontaneous breaths.
• During T
L synchronous intervals, successful inspiratory efforts cause the ventilator to cycle from P
L there is no spontaneous (patient) effort, this transition takes place at the end of the T
L
period.
to P
H
. If
• During T
H synchronous intervals, successful expiratory efforts cause the ventilator to cycle from P
H
to P
L
. If there is no spontaneous exhalation, the transition to the P
L
level takes place at the end of the T
H
period.
12.9.2 Patient Monitoring in BiLevel
If the patient breathes spontaneously at either pressure level, BiLevel monitors and displays the total respiratory rate, including mandatory and spontaneous breaths. BiLevel also displays the exhaled tidal volume and total exhaled minute volume for both mandatory and spontaneous breaths.
12.9.3 APRV Strategy in BiLevel
Lengthening the T
H
period and shortening the T breath volume, results in an inverse T
H
:T
L
L
period to only allow incomplete exhalation of the mandatory
ratio. In this breath timing configuration with T
4:1, BiLevel becomes A irway P ressure R elease V entilation (APRV).
H
:T
L
ratios of greater than
APRV is characterized as longer T
H
periods, short T
L
periods (usually less than 1 second), and inverse T
H
:T
Since, at these breath timing settings, all of the patient-triggered spontaneous breaths occur during the T
H
L
ratios.
period,
APRV resembles CPAP ventilation with occasional, short periods of incomplete exhalation referred to as “releases,
“ which are controlled by the f setting.
Figure 100.
APRV With Spontaneous Breathing at PH
1 P
CIRC
(cmH
2
O)
2 Lengthened inspiratory time (T
H
)
3 Shortened release time (T
L
)
In APRV, the P
H
level is set to optimize pulmonary compliance for spontaneous breathing while maintaining an elevated mean airway pressure to promote oxygenation, the P
L
level is set, along with the T
L
, to control the expiratory release volume of mandatory breaths to help manage CO controls the number of releases per minute which are used to help manage the patient’s CO also impacts the mean airway pressure.
2
and alveolar ventilation, and the f setting
2
levels. The f setting
In APRV the operator can configure the BiLevel settings to allow direct control of T f setting will not inadvertently lengthen the T
L volume. With the T
L
period locked, changes in set f will change the T
H while maintaining the set T
L
period.
L
to assure that changes in the
period resulting in destabilization of end-expiratory alveolar
period to accommodate the new f setting
12.9.4 Technical Structure of BiLevel
In BiLevel, the ventilator establishes two levels of baseline pressure. One level is essentially the same as the standard PEEP level set for all common modes of ventilation. The second pressure level is the level established at
274
T
H
. Both pressure levels permit CPAP, TC and PS breaths. The breath timing settings determine whether the patient can initiate any of these breath types.
12.10 Mode Changes
Changing to BiLevel mode from other modes follows the general guidelines for mode changes:
• The change is made as soon as possible without compromising inspiration or exhalation.
• Breaths are not stacked during inspiration.
12
275
12
276
13
13 Appendix Leak Sync
13.1 Overview
This appendix describes the operation of the Puritan Bennett™ 980 Series Ventilator Leak Sync function. Leak Sync enables the ventilator to compensate for leaks in the breathing circuit while accurately detecting the patient’s effort to trigger and cycle a breath. Because Leak Sync allows the ventilator to differentiate between flow due to leaks and flow due to patient respiratory effort, it provides dynamic compensation and enhances patient-ventilator synchrony. See
for general parameter and operational information.
13.2 Intended Use
Leak Sync is designed to compensate for leaks in the breathing circuit during Non-Invasive or Invasive Ventilation.
Leak Sync accurately quantifies instantaneous leak rates, therefore detecting patient respiratory phase transitions correctly and may affect work of breathing. Leak Sync is intended for neonatal, pediatric, and adult patients.
13.3 Safety Term Definitions
This section contains safety information for users, who should always exercise appropriate caution while using the ventilator.
Table 76.
Safety Term Definitions
Term
WARNING
Caution
Note
Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse events) to the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the product.
Note
Notes provide additional guidelines or information.
Warning: The ventilator offers a variety of breath delivery options. Throughout the patient’s treatment, the clinician should carefully select the ventilation mode and settings to use for that patient based on clinical judgment, the condition and needs of the patient, and the benefits, limitations and characteristics of the breath delivery options. As the patient’s condition changes over time, periodically assess the chosen modes and settings to determine whether or not those are best for the patient’s current needs.
13.4 Leak Sync
Breathing circuit leaks can cause the ventilator to erroneously detect patient inspiratory efforts (called autotriggering) or delay exhalation in pressure support. Patient interfaces such as masks are particularly prone to significant leaks. Inaccurately declaring inspiration or exhalation can result in patient-ventilator dysynchrony and increased work of breathing.
Changing inspiratory or expiratory sensitivity settings can temporarily correct the problem, but requires continued frequent clinical intervention to ensure that sensitivity is adjusted appropriately as conditions change
(for example, if the patient moves or the circuit leak changes).
Leak Sync adds flow to the breathing circuit to compensate for leaks. The maximum Leak Sync flow applies to the maximum base flow compensation during exhalation. During pressure-based inspirations, the total delivered flow (leak flow plus inspiratory flow) is limited by the maximum total flow.
The following table shows the maximum leak rates at set PEEP pressure that Leak Sync will compensate based on patient type.
277
13
Table 77.
Maximum Leak Compensation Flow Based on Patient Type
Patient type
Neonatal
Pediatric
Adult
Maximum Leak compensation flow at PEEP
Invasive: 15 L/min
NIV: 30 L/min (25 L/min if compres‐ sor is the air source)
40 L/min (25 L/min if the compres‐ sor is the air source)
65 L/min (25 L/min if the compres‐ sor is the air source)
Maximum total flow
50 L/min
120 L/min
200 L/min
Warning: With significant leaks, pressure targets may not be reached due to flow limitations.
13.5 Setting Up Leak Sync
For more information on setting up the ventilator, see Chapter 4 .
To enable Leak Sync
1. At the ventilator setup screen, touch the More Settings tab.
2. Touch Enabled in the Leak Sync area.
3. Touch Accept ALL to enable Leak Sync.
Figure 101.
Enabling Leak Sync
Note: The default value for Leak Sync is Disabled when the circuit type is pediatric or adult and the ventilation type is Invasive. Otherwise the default value for Leak Sync is Enabled .
Note: Leak Sync is not allowed for tube compensated (TC) and proportional assist ventilation (PAV+™) breath types.
278
13
13.6 When Leak Sync is Enabled
See
for an example showing the GUI screen when Leak Sync is enabled.
• The vent setup button on the GUI screen indicates Leak Sync is active.
• D
SENS
is displayed in units of L/min, rather than %.
• If the ventilator detects a leak during a respiratory mechanics maneuver, the message Leak Detected is displayed.
• A new leak or change in leak rate is typically quantified and compensated within three breaths. Monitored patient data stabilizes within a few breaths.
• Select inspiratory sensitivity settings as usual. if the ventilator auto-triggers, try increasing flow sensitivity
( ⩒
SENS
).
Note: The absence of the Leak Detected message does not mean there is no leak.
Note: Leak Sync is automatically enabled when ventilation type is NIV or if New patient is selected and circuit type is neonatal, regardless of the ventilation type. If Leak Sync is disabled while the ventilation type is Invasive but the ventilation type is changed to NIV, it remains disabled. Leak Sync becomes disabled when ventilation type is set to Invasive and circuit type is adult or pediatric.
Figure 102.
GUI Screen when Leak Sync is Enabled
13
1 LS appears on vent setup button notifying the operator that Leak Sync is enabled
13.6.1 Adjusting Disconnect Sensitivity (D
SENS
)
When Leak Sync is enabled, the CIRCUIT DISCONNECT alarm becomes active based on the D
SENS is the maximum allowable leak rate at set PEEP.
setting, which
When Leak Sync is disabled, D
SENS
is automatically set to 75%.
Warning: When ventilation type = NIV and Leak Sync is disabled, D
SENS
is automatically set to OFF.
See
SENS maximum Leak Sync flow.
settings when Leak Sync is enabled. Note that it is possible to set D
SENS
below
279
Table 78.
D
SENS
settings
Breathing circuit type
Neonatal
Pediatric
Adult
D
SENS
setting
Range:
Invasive: 1 L/min to 15 L/min
NIV: 1 L/min to 30 L/min
Default: 2 L/min (Invasive Ventila‐ tion) 5 L/min (NIV)
Range: 1 L/min to 40 L/min
Default: 20 L/min
Range: 1 L/min to 65 L/min
Default: 40 L/min
Maximum total flow
50 L/min
120 L/min
200 L/min
Warning: Setting D
SENS
higher than necessary may prevent timely detection of inadvertent extubation.
13.6.2 Monitored Patient Data
When Leak Sync is enabled, three additional parameters are displayed on the More Patient Data screen and updated for each breath. Display the More Patient Data screen by swiping the tab on the patient data banner.
These leak parameters may also be configured on the patient data banner and the large font patient data panel.
Figure 103.
Leak Sync Monitored Patient Data
1 Leak Sync Parameters
See
Table 65, page 253 for information regarding the following monitored patient data parameters:
• V
LEAK
• % LEAK
• LEAK
Displayed values for exhaled tidal volume (V
TE
) and inspired tidal volume (V the estimated inspired or exhaled lung volume. The accuracies for V
TL
) are leak-compensated, and indicate
TE
and V
TL
also change when Leak Sync is
280
13
enabled (see Section 13.7, Technical Discussion, page 281
for more information). Graphic displays of flow during
Leak Sync indicate estimated lung flows.
13.7 Technical Discussion
Managing breathing circuit leaks is important to ensure appropriate breath triggering and cycling, ventilation adequacy, and valid patient data. Detecting and monitoring leaks can improve treatment, reduce patient work of breathing, and provide more accurate information for clinical assessments.
Leak Sync recognizes that changing pressures lead to varying deflection of interface materials and leak sizes. The
Leak Sync leak model includes a rigid leak orifice whose size remains constant under changing pressures, combined with an elastic leak source whose size varies as a function of applied pressure. This algorithm provides a more accurate estimate of instantaneous leak to improve patient-ventilator synchrony under varying airway pressures.
Leak Sync allows the ventilator to determine the leak level and allows the operator to set the flow trigger and peak flow sensitivities to a selected threshold. The base flow during exhalation is set to:
• Flow triggering: 1.5 L/min + estimated leak flow at PEEP + flow sensitivity.
• Pressure triggering: 1.0 L/min + estimated leak flow at PEEP.
13.7.1 Inspired Tidal Volume (V
TL
) Accuracy During Leak Sync
See
TL
parameter, for V
TL
accuracy.
For readings <100 mL, accuracy ranges apply when the percentage of inspiratory leak volume is <80%, where the percentage of leak volume is:
(Leak volume during inspiration / total delivered inspiratory volume)×100
Note: Inspired tidal volume is labeled as V
TL
when Leak Sync is enabled, and as V
TI
when Leak Sync is disabled.
13.7.2 Exhaled Tidal Volume (V
TE
) Accuracy During Leak Sync
See
TE
parameter, for accuracy when Leak Sync is enabled.
where T
E
= time to exhale 90% of volume actually exhaled by the patient.
For readings <100 mL, accuracy ranges apply when the percentage of inspiratory leak volume is <80%, where the percentage of leak volume is:
(Leak volume during inspiration/total delivered inspiratory volume) ×100
13.7.3 %LEAK Calculation
See
Table 65, page 253 , % LEAK parameter, for specifications.
13.7.4 Circuit Disconnect Alarm During Leak Sync
The CIRCUIT DISCONNECT alarm is activated if the overall leak volume during the whole breath exceeds the maximum leak volume derived from the D if the end-inspiratory pressure falls below (set PEEP +1 cmH this alarm message:
SENS
setting. During VC, the CIRCUIT DISCONNECT alarm is also activated
2
O) for three consecutive breaths. The screen shows
13
281
Figure 104.
Circuit Disconnect During VC
If the compressor is in use and the D
SENS
setting >25 L/min, a D
SENS
of 25 L/min is used to determine Circuit
Disconnect. If LEAK >25 L/min, the alarm banner shows the following message:
Check patient. Reconnect circuit. Leak may exceed maximum compensation value for compressor.
Normal operation resumes if the ventilator detects a patient connection.
282
14
14 Appendix PAV™+
14.1 Overview
This appendix describes the operation of PAV+™ software for the Puritan Bennett™ 980 Ventilator.
Proportional Assist™ Ventilation (PAV+™) is designed to improve the work of breathing of a spontaneously breathing patient by reducing the patient’s increased work of breathing when pulmonary mechanics are compromised.
The PAV+™ breath type differs from the pressure support (PS) breath type in the following way:
PAV+™ acts as an inspiratory amplifier; the degree of amplification is set by the % Support setting (% Supp). PAV+™ software continuously monitors the patient’s instantaneous inspiratory flow and instantaneous lung volume, which are indicators of the patient’s inspiratory effort. These signals, together with ongoing estimates of the patient’s resistance and compliance, allow the software to instantaneously compute the necessary pressure at the patient wye to assist the patient’s inspiratory muscles to the degree selected by the % Supp setting. Higher inspiratory demand yields greater support from the ventilator.
PAV+™ software reduces the risk of inadvertent entry of incompatible settings, such as small predicted body weight (PBW) paired with a large airway.
14.2 Intended Use
PAV+™ is intended for use in spontaneously breathing adult patients whose ventilator predicted body weight
(PBW) setting is at least 25.0 kg (55 lb). Patients must be intubated with either endotracheal (ET) or tracheostomy
(Trach) tubes of internal diameter (ID) 6.0 mm to 10.0 mm. Patients must have satisfactory neural-ventilatory coupling, and stable, sustainable inspiratory drive.
14.3 Safety Term Definitions
This section contains safety information for users, who should always exercise appropriate caution while using the ventilator.
Table 79.
Safety Term Definitions
Term
WARNING
Caution
Note
Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse events) to the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the product.
Note
Notes provide additional guidelines or information.
Warning: The ventilator offers a variety of breath delivery options. Throughout the patient’s treatment, the clinician should carefully select the ventilation mode and settings to use for that patient based on clinical judgment, the condition and needs of the patient, and the benefits, limitations and characteristics of the breath delivery options. As the patient’s condition changes over time, periodically assess the chosen modes and settings to determine whether or not those are best for the patient’s current needs.
Warning: PAV+™ is not an available breath type in Non-Invasive Ventilation (NIV). Do not use non-invasive patient interfaces such as masks, nasal prongs, uncuffed ET tubes, etc. as leaks associated with these interfaces may result in over-assist and patient discomfort.
14
283
Warning: Breathing circuit and artificial airway must be free from leaks. Leaks may result in ventilator over-assist and patient discomfort.
Warning: Ensure high and low tidal volume alarm thresholds are set appropriately because an overestimation of lung compliance could result in an under-support condition resulting in the delivery of smaller than optimal tidal volumes.
14.4 PAV+™
Warning: Ensure that there are no significant leaks in the breathing circuit or around the artificial airway cuff.
Significant leaks can affect the performance of PAV+™ and the accuracy of resistance (R) and elastance (E) estimates.
Warning: Do not use silicone breathing circuits with PAV+™: the elastic behavior of a silicone circuit at the beginning of exhalation can cause pressure-flow oscillations that result in underestimates of patient resistance.
The act of inspiration requires the patient’s inspiratory muscles to develop a pressure gradient between the mouth and the alveoli sufficient to draw in breathing gas and inflate the lungs. Some of this pressure gradient is dissipated as gas travels through the artificial airway and the patient’s conducting airways, and some of the pressure gradient is dissipated in the inflation of the lungs and thorax. Each element of pressure dissipation is characterized by a measurable property: the resistance of the artificial and patient airways, and the compliance (or elastance) of the lung and thorax.
PAV+™ software uses specific information, including resistance of the artificial airway, resistance of the patient’s airways, lung-thorax compliance, instantaneous inspiratory flow and lung volume, and the % Supp setting to compute the instantaneous pressure to be applied at the patient connection port (patient wye). PAV+™ software randomly estimates patient resistance and compliance approximately every four to ten breaths. Every 5 ms, the software estimates lung flow, based on an estimate of flow at the patient wye, and lung volume, based on the integral of the value of estimated lung flow.
PAV+™ begins to assist an inspiration when flow (generated by the patient’s inspiratory muscles) appears at the patient wye. If the patient ceases inspiration, the assist also ceases. Once inspiratory flow begins, PAV+™ software monitors instantaneous flow and volume every 5 ms and applies the pressure calculated to overcome a proportion (determined by the % Supp setting) of the pressure losses dissipated across the resistances of the artificial and patient airways and lung/thorax compliance.
Because the PAV+™ algorithm does not know the patient’s mechanics when the PAV+™ breath type is selected, the software performs a startup routine to obtain initial data. At startup, PAV+™ software delivers four consecutive
PAV+™ breaths, each of which includes an end-inspiratory pause maneuver that yields estimates of the patient’s resistance and compliance. The first breath, however, is delivered using the predicted resistance for the artificial airway and conservative estimates for patient resistance and compliance, based on the patient’s PBW.
Each of the next three PAV+™ breaths averages stepwise decreased physiologic values with the estimated resistance and compliance values from the previous breath, weighting earlier estimates less with each successive breath, and yielding more reliable estimates for resistance and compliance. The fifth PAV+™ breath (the first non-startup breath) is delivered using the final estimates with the clinician-set % Supp setting. Once startup is complete, the PAV+™ software randomly applies a maneuver breath every four to ten breaths after the last maneuver breath to reestimate patient resistance and compliance. New values are always averaged with former values.
PAV+™ graphically displays estimates of patient lung pressure (intrinsic PEEP), patient compliance, patient resistance, total resistance, total work of inspiration, patient work of inspiration, inspiratory elastic work (an indicator of lung-thorax work), and inspiratory resistive work.
The % Supp setting ranges from a minimum of 5% (the ventilator performs 5% of the work of inspiration and the patient performs 95%) to a maximum of 95% (the ventilator performs 95% of the work and the patient performs
5%), adjustable in 5% increments.
284
14
PAV+™ also includes alarm limits, safety checks, and logic checks that reject non-physiologic values for patient resistance and compliance as well as inappropriate data.
Humidification type and volume can be adjusted after running SST, however the ventilator makes assumptions when calculating resistance and compliance if these changes are made without rerunning SST. For optimal breath delivery, run SST after changing humidification type and humidifier volume.
14.4.1 Setting Up PAV+™
To set up PAV+™
1. At the ventilator setup screen, enter the patient’s gender and height or the patient’s PBW.
2. Touch Invasive ventilation type.
3. Touch SPONT mode.
4. Touch PAV+™ to select Spontaneous type.
5. Touch the desired trigger type (P
TRIG
or V
TRIG
).
6. Select tube type.
7. Select the tube ID. Initially, a default value is shown based on the PBW entered at ventilator startup. If this
ID is not correct for the airway in use, turn the knob to adjust the ID setting.
8. Continue setting up the ventilator as described in Chapter 4
.
Figure 105.
Ventilator Setup Screen
14
Note: If the operator selects an internal diameter that does not correspond to the PBW/tube ID range pairs listed in the following table, touch Continue to override the tube ID setting. If attempts are made to choose a tube ID less than 6.0 mm or greater than 10 mm, a hard bound limit is reached, as PAV+™ is not intended for use with tubes smaller than 6.0 mm or larger than 10.0 mm. When touching Dismiss, the setting remains at the last tube ID selected. Touch Accept or Accept ALL to accept changes, or touch Cancel to cancel changes.
Note: If Leak Sync is currently enabled, it becomes disabled when PAV+™ is selected.
285
Note: When the ventilator is used on the same patient previously ventilated using PAV+™, the GUI displays an attention icon and the tube type and tube ID previously used, as a reminder to the clinician to review those settings during ventilator setup.
14.4.2 PBW and Tube ID
The ventilator uses “soft bound” and “hard bound” values for estimated tube inside diameters based upon PBW.
Soft bounds are ventilator settings that have reached their recommended high or low limits. When adjusting the tube size, if the inside diameter does not align with a valid predicted body weight, a Continue button appears.
Setting the ventilator beyond these soft bounds requires the operator to acknowledge the prompt by touching
Continue before continuing to adjust the tube size. The limit beyond which the tube ID cannot be adjusted is called a hard bound, and the ventilator emits an invalid entry tone when a hard bound is reached.
Warning: Ensure that the correct artificial airway ID size is entered. Because PAV+™ amplifies flow, entering a smaller-than-actual airway ID causes the flow-based pressure assistance to over-support the patient and could lead to transient over-assist at high values of % Supp. Conversely, entering a larger-than-actual ID results in under-support. PAV+™ software monitors the settings for the PBW and artificial airway. If the PBW and tube ID settings do not correspond to allowable values, confirm or correct the settings. Confirming or correcting the actual ID size minimizes the likelihood that PAV+™ will over-support or under-support.
To apply new settings for the artificial airway follow these steps
1. Touch the vent setup button at the lower left of the GUI screen.
2. Touch Tube Type and turn the knob to select Trach or ET to set the tube type.
3. Touch Tube ID and turn the knob to set the tube ID.
4. Touch Accept or Accept ALL to apply the new settings, or Cancel to cancel.
To apply new humidifier settings
1. Touch the More Settings tab.
2. Touch the appropriate button for Humidification Type.
3. For non-HME humidification types, touch Humidifier Volume, then turn the knob to adjust the (empty) humidifier volume.
4. Touch Accept ALL to apply the changes.
Warning: To ensure the accuracy of PAV+™ breaths and spirometry measurements, run SST following any change to the humidification type or humidification volume settings. Ensure that the intended circuit is used with the SST.
14.4.3 Apnea Parameters Adjustment
After accepting the PAV+™ settings, touch the apnea setup screen. Adjust the apnea parameters as required.
14.4.4 Alarm Settings Adjustment
PAV+™ includes the high inspired tidal volume ⤒ V
TI
) and low exhaled spontaneous tidal volume alarm ( ⤓ V
TE
SPONT
) alarm limit settings. See
Section 14.4.8, PAV+™ Alarms, page 288 .
Note: Because of the breathing variability that PAV+™ allows, the ⤓ V
TE SPONT
alarm, by default, is turned OFF to minimize nuisance alarms. To monitor adequate ventilation, use the ↓ ⩒
E TOT
alarm condition instead.
To adjust alarm settings
1. Touch the Alarm tab to view the current alarm settings.
2. Touch the button for each alarm limit requiring a change.
286
14
3. Turn the knob to adjust the value of the alarm limit. Proposed values are highlighted. You can change more than one alarm limit before applying the changes.
4. Touch Accept or Accept All to apply the changes, or Cancel to cancel.
14.4.5 PAV+™ Ventilator Settings
See
for a summary of PAV+™ ventilator settings for the following parameters:
• %
Supp
• Expiratory sensitivity (E
SENS
)
• Tube type
• Tube ID
• Trigger type
14.4.6 PAV+™ Alarm Settings
See
Table 64, page 252 for a summary of the following alarm settings available when PAV+™ is active:
• High inspired tidal volume limit ( ⤒ V
TI
)
• Low exhaled spontaneous tidal volume ( ⤓ V
TE SPONT
)
14.4.7 Monitored Data
For the following monitored data associated with PAV+™, see Table 65, page 253 :
• PAV-based lung compliance (C
PAV
)
• PAV-based lung elastance (E
PAV
)
• PAV-based lung resistance (R
PAV
)
• PAV-based total airway resistance (R
TOT
)
• Inspired tidal volume (V
TI
)
• Driving Pressure (P
DRIVE
)
See
Table 80 for monitored data absolute limits.
Table 80.
Absolute limits for PAV+™ Monitored Data
PBW (kg)
25
35
45
55
65
75
85
95
105
115
125
135
145
150
R
PAV
(cmH
2
O/L/s)
0 to 50
0 to 44
0 to 31
0 to24
0 to 20
0 to 18
0 to 17
0 to 16
0 to 15
0 to 15
0 to 14
0 to 14
0 to 14
0 to 14
C
PAV
(mL/cmH
2
O)
2.5 to 29
3.5 to 41
4.5 to 52
5.5 to 64
6.4 to 75
7.4 to 87
8.4 to 98
9.4 to 110
10 to 121
11 to 133
12 to 144
13to 156
14 to 167
15 to 173
E
PAV
(cmH
2
O/L)
34 to 400
24 to 286
19 to 222
16 to 182
13 to 156
11 to 135
10 to 119
9.1 to 106
8.3 to 100
7.5 to 91
6.9 to 83
6.4 to 77
6.0 to 71
5.8 to 67
14
287
14.4.8 PAV+™ Alarms
See
Table 36, page 144 for a summary of the following alarms associated with PAV+™:
• HIGH CIRCUIT PRESSURE ( ↑ P
PEAK
)
• HIGH VENTILATOR PRESSURE ( ↑ P
VENT
)
• PAV STARTUP TOO LONG
• PAV R&C NOT ASSESSED
• ↑ V
TI
14.5 Ventilator Settings and Guidance
Warning: For optimal performance of PAV+™, it is important to select the humidification type, tube type, and tube size that match those in use on the patient.
The instantaneous pressure generated at the patient wye during inspiration is a function of the patient effort, %
Supp setting, tube type and size, patient resistance and elastance, and the instantaneously measured gas flow and lung volume. Set ⤒ P
PEAK
to a safe circuit pressure, above which truncation and alarm annunciation are appropriate.
Note:
↑ P
PAV+™ has a built-in high pressure compensation ( minus 5 cmH
COMP
2
O or 35 cmH
2
↑ P
COMP
) limit that is determined by the ⤒ P
O, whichever is less. If the inspiratory pressure at the patient wye (P
Iwye
PEAK
for more details regarding ↑ P
COMP
and ↑ P
PEAK
.
setting
) reaches the
14.5.1 Specified Performance
Performance using PAV+™ is ±0.5 Joules/liter (J/L), compared to measured, work during inspiration at the 75% support (% Supp) level. Work is computed over the entire inspiratory interval. In ventilation terms, work (W) is expressed as: i
W
P ith sample interval (5 ms)
Work [J/L]
Synchronous and combined pressures developed by the ventilator and by the patient (P
MUS
), [cmH
2
O]
V k
Flow [L/s] conversion constant (0.098) [J/cmH
2
O × L)
14.5.2 Graphics Displays in PAV+™
When PAV+™ is active (the mode is SPONT and the spontaneous breath type is PAV+™), a work of breathing (WOB) graphic is automatically displayed. See
, which shows:
• an indicator showing the proportion of patient inspiratory work to overcome the elastance (E) of the lung-thorax and the combined resistance (R) of the artificial airway and the patient.
• estimates of work of breathing relative to normal, subnormal, and above-normal values, including:
– the estimated work of breathing (in Joules/L) during inspiration (WOB
PT
) and
– the estimated total work of breathing (in Joules/L) of the patient and ventilator during inspiration
(WOB
TOT
)
Additional information in the graphics screen includes:
288
14
• a “shadow” trace of the estimated lung pressure, shown as a solid area superimposed on the circuit pressure waveform, and
• PAV-based patient data estimates, including patient resistance (R
PAV
PEEP (PEEP
I PAV
).
), lung compliance (C
PAV
), and intrinsic
Note: Graphic displays of lung pressure and patient work of breathing are not actual measurements, and are derived from equations using filtered estimates of pressure and flow.
The WOB graphic is only available when SPONT mode and the PAV+™ breath type are selected. The shadow trace can be enabled or disabled when selecting the graphic display, or after a display is paused.
The act of pausing does not affect the WOB graphic, but does store the shadow trace. Once paused, the operator can enable or disable the shadow trace, then view the paused waveform again with or without the shadow trace.
14.5.3 WOB Terms and Definitions
The following table provides a definition and description of each of the Work of Breathing (WOB) terms.
Table 81.
PAV+™ Work of Breathing terms
WOB term
WOB
TOT
Definition
Total Work of Inspiration
WOB
PT
WOB
PT ELASTIC
ELASTIC
WOB
PT RESISTIVE
Patient Work of Breathing
Inspiratory Elastic Work
Inspiratory Resistive Work
Description
With the PAV+™ breath type active, the patient and the ven‐ tilator always share the in the work of breathing. The percent
WOB
TOT
performed by the ventilator always equals the %
Supp setting and the percent WOB
TOT
performed by the patient always equals (100 minus the % Supp setting).
WOB
TOT
is the sum of the work to move the breathing gas through the artificial airway and the patient’s own airways plus the work to inflate the patient’s elastic lung-thorax.
That part of WOB
TOT
performed by the patient.
That part of WOB
PT lung-thorax.
attributed to inflating the patient’s elastic
That part of the WOB
PT
attributed to moving breathing gas through resistive elements in the gas path.
14
289
Figure 106.
Graphics displays in PAV+™
1 Total work of breathing (WOB
TOT
)
2 Patient’s work of breathing (WOB
PT
)
3 Shadow trace
14.5.4 Technical Description
When PAV+™ is selected, the ventilator acts as an inspiratory amplifier, proportionally assisting the pressure generating capability of the inspiratory muscles (P
MUS
).
14.5.4.1 Pressure Gradient Equation of Motion
During spontaneous breathing, P
MUS
generates a pressure gradient that drives breathing gas through the artificial airway and the patient’s airways and into the elastic lung-thorax, and is described by the equation of motion:
Figure 107.
EQUATION 1
P
⩒
MUS
L
Pressure generating capability of patient’s inspiratory muscles
Flow through the resistance elements and into the lungs
Insufflation volume of the lung
R
E
LUNG-
THORAX
Resistance elements (artificial plus patient airways)
Elastance of the lung and thorax (1/C
THORAX
)
LUNG-
V
L
14.5.4.2 Estimates of Patient Resistance and Elastance
If the PAV+™ software estimates of patient resistance and elastance (R
PAV could be rewritten as:
and (E
PAV
) remain stable, this equation
Figure 108.
EQUATION 2
290
14 i Instantaneous value of pressure, flow, or air‐ way resistance, R i flow airway
being a function of
E
PAV
K
1
R
PAV
K
2
P i
MUS
could then be estimated at every control period if V i
L
, R i airway, and V i
L
were also known.
14.5.4.3 Valid Individual Pressure Measurements
Throughout any inspiration, the individual pressure elements that make up P
MUS
can be expressed as:
Figure 109.
EQUATION 3
P p p p
MUS
FLOW
ARTIFICIAL AIRWAY
FLOW
PATIENT
VOLUME
PATIENT
Pressure generating capability of patient’s inspiratory muscles
Flow based pressure drop across the artificial airway
Flow based pressure drop across the patient
Volume based pressure to overcome the lung-thorax elastance
Equations 2 and 3 provide the structure to explain how PAV+™ operates. The clinician enters the type and size of artificial airway in use, and the software uses this information to estimate the resistance of the artificial airway at any lung flow.
Applying a special pause maneuver at the end of selected inspirations provides the information the software needs to estimate patient resistance (R
PAV
) and compliance (C
PAV
, which is converted to elastance, E
PAV
).
Immediately following the end of the pause event, software captures simultaneous values for P which yield an estimate for R
TOT
at the estimated flow.
LUNG
, P wye
, and ⩒
E
All raw data are subjected to logic checks, and the estimates of R
PAV
and C
PAV checks. The estimates of R
PAV
R
PAV
is also rejected.
and C
PAV
are further subjected to physiologic
are discarded if any of the logic or physiologic checks fail. If C
PAV
is rejected,
Valid estimates of R
PAV
If new values for R
PAV
and C
PAV
are required for breath delivery, and are constantly updated by averaging new values with previous values. This averaging process smooths data and avoids abrupt changes to breath delivery.
refresh.
and C
PAV
are rejected, the previous values remain active until valid new values are obtained.
PAV+™ software monitors the update process and generates an escalating alarm condition if the old values do not
14.5.4.4 Maneuver Breaths and % Supp
During PAV+™, maneuver breaths are randomly performed every four to ten breaths after the last maneuver breath. A maneuver breath is a normal PAV+™ inspiration with a pause at end inspiration. Because muscle activity is delayed for at least 300 ms following the end of neural inspiration, the patient’s respiratory control center does not detect the pause. With this approach, maneuver breaths are delivered randomly so that their occurrence is neither consciously recognized nor predictable.
A PAV+™ breath begins, after the recognition of a trigger signal, with flow detection at the patient wye. The sample and control cycle of the ventilator (the value of i in Equation 2) is frequent enough to yield essentially constant tracking of patient inspiration. At every ith interval, the software identifies instantaneous lung flow (V of instantaneous lung volume, (V i
L
, which is impeded by the elastic recoil of the lung and thorax).
i
L
, which is impeded by the resistances of the artificial airway and patient airways) and integrates this flow to yield an estimate
Using the values for instantaneous lung flow and lung volume, PAV+™ software calculates each of the pressure elements in Equation 2, which gives the value of P
MUS
at each ith interval.
14
291
At this point, Equation 2 and the subsequent analysis identifies that an appropriate patient, supported by PAV+™ and with an active P
MUS
(an absolute requirement) will, within a few breaths, enable the algorithm to obtain reasonable estimates of R
PAV
and E
PAV
. Once these physiologic data are captured (and over a relatively brief time they are improved and stabilized), the PAV+™ algorithm mirrors the patient’s respiratory mechanics, which then allows the ventilator to harmoniously amplify P
MUS
. The key point to recognize is that patient’s continuous breathing effort “drives“ the PAV+™ support — no effort, no support.
The % Supp setting specifies the amount of resistance- and elastic-based pressure to be applied at each ith interval at the patient wye.
By taking all of the above information into consideration, EQUATION 2 can be rewritten to include the % Supp setting recognizing that V i
L
and V i
L
are driven by the patient, not by the ventilator. (It is important to note that the ventilator is not amplifying its own flow — only the flow generated by P
MUS
).
Figure 110.
EQUATION 4
P i wye
Pressure generated by the ventilator in response to the instantaneous values of lung flow and lung volume at the wye. This value is the sum of the three individual pres‐ sure elements (in parentheses) in Equation
4
S % Supp setting/100 (ranges from 0.05 to
0.95)
14.5.4.5 Resulting Pressure Gradient
The pressure gradient driving breathing gas into the patient’s lungs is given by the sum of P inspiratory effort, therefore: i wye
and the patient’s
Figure 111.
EQUATION 5
DP i
GRADI‐
ENT p i wye
Instantaneous pressure gradient
Pressure generated by the ventilator in response to the instantaneous values of lung flow and lung volume at the wye
P i
MUS
Instantaneous pressure generating capa‐ bility of patient’s inspiratory muscles
14.5.5 Protection Against Hazard
PAV+™ software is designed to reduce the risk that hyperinflation may occur. The potential for hyperinflation could arise if the software were to overestimate actual patient resistance or underestimate actual patient lung-thorax compliance (that is, to overestimate actual elastance). If the software cannot generate valid estimates of R
PAV
C
PAV
, PAV+™ cannot start. If, after startup, the values of R
PAV
and C
PAV previous values become less reliable.
and
cannot be updated with valid new values, the
The stability of PAV+™ is primarily determined by the relationship between the true lung elastance [E the true lung volume [V
L the elastic component.
(true)]. Although P i
L
(true)] and
wye (resistive) also plays a part, the following discussion focuses on
At all lung volumes, the true state of the lung and thorax is expressed by:
P i
L recoil
= V i
L
(true) * E
L
(true)
292
14
P
V i
L recoil i
L (true)
True lung recoil pressure
True instantaneous volume of the lung
E
L (true)
True lung elastance
Over-inflation will not occur as long as P i wye (elastic) < P i
L recoil
, which is equivalent to the inequality:
S[V i
L
(estimated) × K
2
] < V i
L
(true) × E
L
(true) where:
K
2
= E
PAV
As long as E
PAV
(estimated) = E
PAV
(true) and V
% Supp (that is, between 85% and 95%).
i
L
(estimated) = V i
L
(true) then P i recoil
> P i wye even at high values of
This means that if the pressure applied to the lung-thorax is never greater than E collapse if wye flow vanishes. As long as E
PAV
≤ R
L
(true), P
MUS
is the modulator of P i wye.
(estimated) ≤ E
L
(true), V i
L
L
(true) × V
(estimated) ≤ V i
L
L
, lung volume will
(true), and R
PAV
(estimated)
Hyperinflation could occur if the estimated E
PAV wye (elastic) could exceed P i
L recoil
were greater than the true value of E
L
. At a high % Supp setting, P i
, causing a self-generating flow at the patient wye, which in turn would cause a self-generating inflation of the lungs. This is part of the reason that the % Supp setting is limited to 95%.
Likewise, if the estimated R
PAV
were to exceed the true value of R
L
at a high % Supp setting, P
Iwye
(resistive) could exceed the value necessary to compensate for pressure dissipation across the artificial and patient airways, resulting in early hyperinflation of the lungs. As flow declines after the first third of inspiration, however, the hyperinflating effect would most likely disappear.
PAV+™ software includes these strategies to minimize the possibility of hyperinflation of the lungs:
1. The maximum % Supp setting is limited to 95%.
2. The raw data for R
PAV
and C
PAV
are checked for graph/math logic, and estimated mechanics values are checked against PBW-based physiologic boundaries. These checks reduce the possibility of overestimating patient resistance or underestimating patient compliance, which could lead to potential over-inflation.
3. The high inspiratory tidal volume limit ( ⤒ V
TI leak flow), which equals lung volume. If the value of V and immediately transitions to exhalation.
) places an absolute limit on the integral of lung flow (including
TI
reaches this limit, the ventilator truncates inspiration
4. The ⤒ V
TI
3 and 4). At the beginning of each new inspiration, PAV+™ software calculates a value for P follows:
setting places an upper limit on the value of the P VOLUME
PATIENT
component of Piwye (see Equations
VOLUME
PATIENT
as
P*wye (elastic threshold limit) = 0.75 × (V
TI
× E
PAV
) where P*wye is the unique value for the elastic threshold limit of Piwye that will cause the lung volume to expand to 75% of ⤒ V
TI
limit.
. When P i wye
(elastic) = P*wye (elastic threshold limit), the software stops increasing
Piwye (elastic). This means that any further increase in lung volume must be accomplished by the patient, which tends to hasten the conclusion of inspiratory effort and avoid truncation due to lung volume reaching the ⤒ V
TI
5. The high inspiratory pressure limit ( ⤒ P
PEAK
) applies to all breaths, and is used by PAV.+ software to detect the high compensation pressure condition ( ↑ P
COMP
):
↑ P
COMP
= ⤒ P
PEAK
–5 cmH
2
O or 35 cmH
2
O, whichever is less
If the user-adjustable ⤒ P
PEAK transitions to exhalation. If P i
limit is reached, the ventilator truncates inspiration and immediately wye
(the targeted wye pressure calculated in Equation 4) equals the ↑ P
COMP
for
14
5 see equations 2 and 4
293
500 ms, the inspiration is truncated and exhalation begins. Further, when P
↑ P
COMP
. Although this freezes the value of P
⤒ P
PEAK
, causing inspiration to end.
i wye i wye
= ↑ P
COMP
, P i wye
is limited to
, patient activity such as coughing could drive Piwye to
The rapid rise of P i wye
to the ↑ P
COMP were overestimated and % Supp were set above 85%. The ↑ P due to overestimation of R
PAV
.
limit would likely occur in the first third of inspiration, and only if R
COMP
PAV
condition guards against over-inflation
6. The% Supp setting ranges from 5 to 95% in 5% increments. Reducing the level of support decreases the possibility of over-inflation. A significant decrease could produce a sensation of inadequate support, and the patient would absorb the additional work of inspiration or require an increase in the level of support.
A significant increase could cause a surge in the ventilator generated value for P wye cause P i wye
to reach ⤒ P
COMP
, which in turn could
and lead to temporary patient-ventilator disharmony. To minimize this possibility, PAV+™ software limits the actual increase in support to increments of 10% every other breath until the new setting is reached.
7. Spirometry remains active during PAV+™ operation. ⤒ V
TI breaths, while ⤓⩒
E TOT
and ⤒⩒
E TOT
can be set high enough to allow spontaneous sigh
remain active to reveal changes in minute ventilation.
Because PAV+™ cannot operate without valid estimates of R
PAV unknown when PAV+™ starts, a startup routine obtains these values during four maneuver breaths that include an end inspiratory pause that provides raw data for R
PAV
and C
and C
PAV
PAV
maneuver breath at the next breath. See Section 14.4, PAV+™, page 284 .
, and because those values are
, and both estimated values must be valid. If either value is invalid during any of the four startup breaths, the software schedules a substitute
A low-priority alarm becomes active if a 45-second interval elapses without valid estimates for R
C
PAV
. If the condition persists for 90 seconds, the alarm escalates to medium-priority. If the condition persists for 120 seconds, the alarm escalates to high priority. The ↓ ⩒
E TOT this condition.
and ↑ f
TOT
PAV
and
alarms are also associated with
Similarly, if R
PAV and C
PAV
cannot be updated with valid values after a successful PAV+™ startup, a low-priority alarm is activated if the condition persists for 15 minutes. If the values still cannot be updated with valid values after 30 minutes, the alarm escalates to medium priority.
8. If PAV+™ estimates a high lung resistance following a sharp spike in the expiratory flow waveform, then a
PBW-based resistance value is used. See
Figure 112.
Use of Default Lung Resistance
1 Flow ( ⩒ )
2 Expiration
3 Inspiration
4 Exhalation with slow, restricted return to zero flow
5 High peak expiratory flow
6 Exhalation with normal return to zero flow
7 Normal peak expiratory flow
294
14
37
38
39
40
33
34
35
36
41
42
29
30
31
32
25
26
27
28
Table 82.
Default PBW-based Resistance Values
(cmH
2
O/L/s) PBW (kg)
18.1
43
17.7
17.4
44
45
17.1
16.8
16.5
16.2
15.9
15.7
15.4
15.2
14.9
14.7
14.5
14.3
14.1
13.9
13.7
46
47
48
49
54
55
56
57
50
51
52
53
58
59
60
Resistance
(cmH
2
O/L/s) PBW (kg)
13.5
61
13.3
13.2
62
63
13.0
12.9
12.7
12.6
64
65
66
67
12.4
12.3
12.2
12.1
12.0
11.8
11.7
11.6
11.5
11.4
11.3
76
77
78
72
73
74
75
68
69
70
71
Resistance
(cmH
2
O/L/s) PBW (kg)
11.3
79
11.2
11.1
80
81 to 150
11.0
10.9
10.9
10.8
10.7
10.7
10.6
10.5
10.5
10.4
10.4
10.3
10.3
10.2
10.2
Resistance
(cmH
2
O/L/s)
10.1
10.1
10
14
295
296
15
15 Appendix NeoMode 2.0
15.1 Overview
This appendix describes how to use NeoMode 2.0 software on the Puritan Bennett™ 980 Series Neonatal
Ventilator. NeoMode 2.0 enables the use of the ventilator with neonatal patients and is included with all
Puritan Bennett™ 980 Neonatal Ventilators and Puritan Bennett™ 980 Universal Ventilators. For a Puritan Bennett™
980 Pediatric–Adult ventilator to be used with neonatal patients, the NeoMode 2.0 software option must be installed.
15.2 Intended Use
NeoMode 2.0 is intended to provide respiratory support to neonatal patients with predicted body weights as low as 0.3 kg (0.66 lb). It is intended to cover a wide variety of clinical patient conditions in hospitals and hospital-type facilities, and may be used during intra-hospital and intra-hospital-type facility transport. It supports delivered tidal volumes as low as 2 mL.
15.3 Description
The ventilator determines values for operational variables and allowable settings based on breathing circuit type and predicted body weight (PBW). The PBW range for neonates is 0.3 kg to 7.0 kg (0.66 lb to 15 lb). Software controls prevent inadvertent mismatching of patient size and breathing circuit type. A neonatal breathing circuit connects to a neonatal exhalation filter that must be used with the neonatal adapter door assembly.
Note: To enable NeoMode 2.0, select the neonatal breathing circuit type in SST. Breathing circuit type can only be changed during SST.
15.4 Safety Information
Warning: The Puritan Bennett™ 980 Series Ventilator contains phthalates. When used as indicated, very limited exposure to trace amounts of phthalates may occur. There is no clear clinical evidence that this degree of exposure increases clinical risk. However, to minimize risk of phthalate exposure in children and nursing or pregnant women, this product should only be used as directed.
Warning: The ventilator offers a variety of breath delivery options. Throughout the patient’s treatment, the clinician should carefully select the ventilation mode and settings to use for that patient based on clinical judgment, the condition and needs of the patient, and the benefits, limitations, and characteristics of the breath delivery options. As the patient’s condition changes over time, periodically assess the chosen modes and settings to determine whether or not those are best for the patient’s current needs.
Warning: Neonatal patients are at risk for hypercarbia or hypoxemia during ↓ ⩒
E TOT
alarm conditions.
Warning: Disabling the low exhaled minute volume ( ↓ ⩒
E TOT hypoxemia.
) alarm increases the patient’s risk of hypercarbia or
Warning: When using NIV, the patient’s exhaled tidal volume (V patient data value for V
TE
) could differ from the ventilator’s monitored
TE
due to leaks around the interface. To avoid this, ensure Leak Sync is enabled.
15.5 Neonatal Door and Filter Installation
Warning: Removing the exhalation filter while the patient is connected to the ventilator can cause a loss of circuit pressure, ventilator autotriggering, or direct contact with liquid.
Caution: Do not pull on door while exhalation filter latch is closed, as damage to the ventilator can result.
Note: See the inspiratory filter and exhalation filter IFU for information on filtration efficiency and filter resistance.
To install the neonatal adapter door
15
297
1. Remove the expiratory limb of the patient circuit from the exhalation filter.
2. Lift the exhalation filter latch. See
3. Remove the existing exhalation filter door by lifting it off of the pivot pins.
4. Fit the neonatal adapter door onto the pivot pins.
To install neonatal exhalation filter assembly
1. With the door still open, push the neonatal filter straight up into the adapter.
2. Close the door.
3. Lower the exhalation filter latch.
4. Reattach the expiratory limb of the patient circuit to the filter.
Figure 113.
Installing the Neonatal Filter and Door
1 Neonatal exhalation filter
2 Neonatal adapter door
3 Exhalation filter latch
4 Filter door pivot pin
Warning: To ensure all breathing circuit connections are leak-tight, perform a circuit-leak test by running SST every time the filter is installed. The circuit-leak test can be performed as an individual test from the SST startup screen.
Warning: Empty the condensate vial before the liquid level reaches the maximum fill line. Condensate vial overflow can enter the filter or the breathing circuit, and can cause increased expiratory flow resistance. Change the filter if it appears to be saturated.
Warning: The neonatal exhalation filter and condensate vial is a single unit and is for single-patient use, only. Do not attempt to sterilize the filter assembly.
Warning: Adding accessories to the ventilator can change the pressure gradient across the ventilator breathing system (VBS) and affect ventilator performance. Ensure any changes to the recommended ventilator circuit
. If adding accessories to the patient circuit, always run SST to establish circuit compliance and resistance prior to ventilating the patient.
Note: If the ventilator has not reached operating temperature from recent usage, allow it to warm up for at least
15 minutes before running SST to ensure accurate testing.
Note: Check the inspiratory and expiratory limbs of the breathing circuit and in-line water traps regularly for water buildup. Under certain conditions, they can fill quickly. Empty and clean the in-line water traps as necessary. See the manufacturer’s IFU for additional information.
15.6 How to Empty the Condensate Vial
The condensate vial may accumulate liquid, especially if a non-heated wire patient circuit is in use.
298
15
Warning: To avoid liquid entering the ventilator, empty the condensate vial before the liquid level reaches the maximum fill line.
The condensate vial assembly is integrated with the neonatal exhalation filter and does not contain a drain port.
Empty the condensate vial when liquid reaches the maximum fill line.
To empty the condensate vial
1. While holding the exhalation filter, twist the condensate vial clockwise approximately one quarter turn to remove it.
2. Remove the condensate vial by carefully lowering the vial all the way down to the base of the exhalation compartment and then sliding it out.
3. Quickly empty the condensate vial.
4. Replace the condensate vial by carefully sliding the vial into position, lifting it upward to the filter assembly, and turning counterclockwise until it reaches the stop.
Note: Condensate vial removal may cause the loss of system pressure and a disconnect alarm to sound.
15.7 How to Connect the Breathing Circuit
Warning:
Use one of the ventilator breathing circuits listed. Table 83, page 300
or their equivalent. This ensures that maximum pressure and flow values specified by EN794-1 are not exceeded. Using a circuit with a higher resistance does not prevent ventilation, but can cause an SST fault or compromise the patient’s ability to breathe through the circuit.
See
to connect the breathing circuit.
Figure 114.
How to Connect the Breathing Circuit
15
299
1 Humidifier
2 Patient circuit inspiratory limb
3 Patient circuit wye
4 Patient circuit expiratory limb
5 Condensate vial
6 From patient port
7 Neonatal exhalation filter (installed in adapter door)
8 To patient port
9 Inspiratory filter
Table 83.
Recommended Breathing Circuits
Patient circuit
Assy, patient circuit, neonatal, single heated wire, dis‐ posable, incubator use, for F&P MR850–(Medtronic /
DAR) use with adapter 111/1146
Assy, patient circuit, neonatal, single heated wire, dis‐ posable, not for incubator use, for F&P
MR850–(Medtronic / DAR) use with adapter 111/1146
Ventilator breathing circuit, neonatal, dual heated sys‐
tem, disposable, Fisher & Paykel–(Fisher & Paykel a
)
307S9910
307/8682
RT265
Part number a The part numbers listed reflect the breathing circuit manufacturer part numbers and are subject to change. Refer to the breathing circuit manufacturer for exact circuit details regarding ordering information.
Table 84.
Allowable NeoMode Ventilator Settings and Ventilation Type
Ventilation type
Mode
Mandatory type
Spontaneous type
Trigger type
A/C, SIMV, SPONT, BiLevel
PC, VC, VC+
PS, VS
⩒
TRIG
Invasive
A/C, SIMV, SPONT, CPAP
PC, VC
PS
⩒
TRIG
NIV
15.8 Ventilation Features
using a Puritan Bennett™ 980 Series Neonatal Ventilator, NeoMode 2.0 is already in use and a neonatal patient circuit is the only choice available when performing SST. If using a Puritan Bennett™ 980 Series Universal Ventilator, the NeoMode 2.0 software option is already installed and SST must be run using a neonatal circuit to use the option. If using a Puritan Bennett™ 980 Series Pediatric–Adult Ventilator, the NeoMode 2.0 software option must be installed and SST must be run using a neonatal patient circuit.
15.8.1 SST (Short Self Test)
Warning: Always run SST prior to patient ventilation, ensuring that all accessories used during ventilation are in the ventilator breathing system when SST is run. This ensures correct calculation of compliance and resistance. See
Section 15.8.1, SST (Short Self Test), page 300
for more information.
.
Note: The manufacturer’s published chamber or column compressible volumes typically represent the container volume when empty and may not result in correct system compliance compensation for breath delivery and spirometry.
300
15
Table 85.
Humidifier Volumes—Neonatal Patients
Manufacturer
Fisher & Paykel
Model
MR290
Description
Neo/adult, disposable, autofill humidification chamber
SST humidifier volume setting (mL)
550
a If the following neonatal patient circuits are used with Fisher & Paykel MR290 humidification chamber, enter 500mL as the humidifier volume:• “DAR neonatal patient circuit with single heated wire (DAR 307S9910)–for incubator use • “DAR neonatal patient circuit with single heated wire (DAR 307/8682)–not for incubator use
15.8.2 Elevate O
2
In NeoMode 2.0, the elevate O
2
control works as described in Chapter 3 .
15.8.3 CPAP Mode
When using NeoMode 2.0 and ventilating with Non-Invasive Ventilation (NIV), a separate CPAP mode allows spontaneous breathing with a desired PEEP level. To limit inadvertent alarms associated with the absence of returned volumes in CPAP breathing, CPAP does not make volume alarms available. As some neonates don’t trigger breaths, the default apnea interval, T
A
, is set to OFF. Also, some settings changes will initiate a PEEP restoration breath before phasing in those changes.
Note: In CPAP, apnea interval,T icons are also displayed.
A
, can be adjusted, if desired. It merely defaults to OFF to avoid inadvertent alarms. The message “APNEA DETECTION DISABLED” is displayed at the bottom of the GUI screen. The attention
To set the ventilator for CPAP
1. Select New Patient from the ventilator’s startup screen or touch the Vent Setup button.
15
2. Touch PBW, and turn the knob to set the PBW.
3. Select NIV as the ventilation type.
4. Touch CPAP.
5. Touch each ventilator setting and turn the knob to select the appropriate ventilator settings. When finished, touch START or Accept ALL .
6. Complete the setup by setting the apnea parameters and alarm limits from their respective tabs.
301
Figure 115.
CPAP Setup Screen
15.8.4 Entering CPAP From Other Ventilation Modes
Entering CPAP mode from other ventilation modes or ventilation types requires using the NIV ventilation type.
Section 4.7, Non-Invasive Ventilation (NIV), page 105
explains how the ventilator transitions from Invasive to NIV ventilation type.
To enter CPAP mode for an existing patient
1. Touch the Vent Setup button at the lower left of the GUI screen.
2. Touch NIV ventilation type. CPAP is only allowed during NIV.
3. Touch CPAP. The ventilator enters CPAP mode.
4. Set an apnea interval, T
A
, if appropriate, as it defaults to OFF in CPAP.
Note: Exhaled minute volume ( ⩒
E TOT disabled upon entry into CPAP.
), exhaled tidal volume (V
TE SPONT
), and inspired tidal volume (V
TI
) alarms are
15.8.5 Exiting CPAP Mode
When changing the mode from CPAP to any other mode, several transition rules take effect:
• The ⩒
E TOT
, V
TE MAND
, V
TE SPONT
, and V
TI
alarms are set to their respective new patient defaults.
• The apnea interval T
A immediately.
changes from OFF to an apnea interval of 10 s and the new setting is phased in
• The ⩒
E TOT
, V
TE MAND
, V
TE SPONT
, and V applicability to the selected mode.
TI
alarm sliders appear in the alarm settings screen according to their
302
15
15.8.6 Compliance Compensation
See
Section 10.6, Compliance and BTPS Compensation, page 199
for a complete discussion of compliance compensation. Compliance compensation in NeoMode 2.0 is implemented as described in the aforementioned reference.
Note: If the patient’s compliance decreases beyond the limits of compliance compensation, the ventilator relies on the ⤒ P
PEAK
alarm setting to truncate the breath and switch to exhalation.
15.8.7 Settings, Alarms, and Monitored Patient Data
Warning: Monitor the patient closely if alarms are disabled. There are no audible or visual annunciations for out-of-range conditions when volume, pressure, or apnea alarms are disabled (turned OFF).
See
,and
Table 65 for the minimum and maximum ranges for each ventilator setting or data value.
Most settings, however, are also limited by other settings or conditions (for example, a low alarm limit is always limited by the corresponding high alarm limit). Review the prompt area when making settings changes.
Volume accuracy testing in VC+ was conducted to demonstrate performance of delivered and monitored parameters.
2 mL to 25 mL collected during test execution.
The first column (set tidal volume) represents the desired volume setting in milliliters (mL). The second column represents the total number of test points for test cases executed at that specific setting. The third column is the mean (mean value) of the ventilators and test cases executed for the setting listed. The fourth column represents the standard deviation (SD) of measurements taken for the ventilators and test cases executed at the setting listed.
The measurements were taken using instrumentation located at the patient-connection port. Accessories such as filters and humidifiers were in the circuit during the test. Values were BPTS and compliance compensated. A sample size of five ventilators was used to conduct the testing. Testing was conducted at ambient temperatures of 22°C ±5°C.
Table 86.
Delivered Volume Accuracy
Set tidal volume (mL) Number of test points
2 150
5
15
25
270
240
270
Mean value (mL)
2.061
4.853
15.108
24.608
SD
0.198
0.324
0.383
0.607
Table 87.
Monitored Inspired Volume (V
TI
) Accuracy
Set tidal volume (mL) Number of test points
2
5
150
270
15
25
240
270
Mean value (mL)
2.055
4.872
15.235
24.633
Table 88.
Monitored Exhaled Tidal Volume (V
TE
) Accuracy
Set tidal volume (mL) Number of test points
2 150
5
15
25
270
240
270
Mean value (mL)
2.212
5.892
16.145
25.492
SD
0.192
0.346
0.379
0.566
SD
0.274
0.607
0.851
0.819
15
303
304
16
16 Appendix Proximal Flow
16.1 Overview
This appendix describes the operation of the Proximal Flow Option for the Puritan Bennett™ 980 Series Ventilator.
The Proximal Flow Option is solely used for monitoring flows, pressures, and tidal volumes and does not control these parameters in any way.
The proximal flow sensor is designed to measure the lower flows, pressures and tidal volumes at the patient wye typically associated with invasively ventilated neonatal patients.
For general parameter and general ventilator setup information, see
.
16.2 Intended Use
The Proximal Flow option is used for measuring flows, pressures, and tidal volumes of invasively ventilated neonatal patients with predicted body weights (PBW) of 0.3 kg (0.66 lb) to 7.0 kg (15.4 lb) using ET tube sizes from
2.5 mm to 4.0 mm. The NeoMode 2.0 software option must also be installed on the ventilator.
16.3 Proximal Flow Option Description
The Proximal Flow option measures pressure, flow, and volume at the patient wye. A printed circuit board assembly (PCBA) containing the electronics and pneumatics for the Proximal Flow option is installed in the ventilator on the option host PCBA. Data measured by the Proximal Flow Sensor are displayed on the GUI for monitoring purposes, not for ventilator control. When the ventilator has a Proximal Flow Sensor installed, both proximal flow and proximal pressure measurements are obtained and displayed on the GUI.
A manual purge control is also provided to clear pneumatic lines for accurate pressure measurements. When a manual purge is requested, the ventilator will not allow another purge for at least 30 seconds. See
Sensor Calibration and Sensor Line Purging, page 309
for more information on the purge function.
16.3.1 Proximal Flow Option components
The Proximal Flow option consists of the following components:
Proximal flow option PCBA — Installed on the option host PCBA in the BDU, this printed circuit board assembly contains a pressure sensor to measure the pressure difference between the flow sensor lines and the interfaces required to convert analog measurements from the proximal flow sensor into digital data displayed by the ventilator. The PCBA also contains valves and an accumulator for purging the sensor lines from blockages.
Proximal flow sensor — The proximal flow sensor is required for use with the Proximal Flow option. The sensor is installed near the patient circuit wye. The other end of the sensor connects to the ventilator’s front panel behind a clear door designed to protect the connection point from exposure to spills or from sprayed liquids during cleaning and disinfection.
16
305
Figure 116.
Proximal Flow Sensor
VEN_10297_A
16.4 Safety Term Definitions
This section contains safety information for users who should always exercise appropriate caution while using the ventilator.
Table 89.
Safety Term Definitions
Term
WARNING
Caution
Note
Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse events) to the patient, user, or environment.
Caution
Cautions alert users to exercise appropriate care for safe and effective use of the product.
Note
Notes provide additional guidelines or information.
16.5 Software and Hardware Requirements
The Proximal Flow option requires installation of the NeoMode 2.0 software option or a Puritan Bennett™ 980
Neonatal Ventilator must be used. Details regarding NeoMode 2.0 can be found in the
16.6 Safety Information
Warning: The Puritan Bennett™ 980 Series Ventilator contains phthalates. When used as indicated, very limited exposure to trace amounts of phthalates may occur. There is no clear clinical evidence that this degree of exposure increases clinical risk. However, in order to minimize risk of phthalate exposure in children and nursing or pregnant women, this product should only be used as directed.
Warning: The ventilator offers a variety of breath delivery options. Throughout the patient’s treatment, the clinician should carefully select the ventilation mode and settings to use for that patient based on clinical
306
16 judgment, the condition and needs of the patient, and the benefits, limitations and characteristics of the breath delivery options. As the patient’s condition changes over time, periodically assess the chosen modes and settings to determine whether or not those are best for the patient’s current needs.
Warning: Inspect the proximal flow sensor prior to use, and do not use it if the sensor body, tubing, or connector are damaged, occluded, or broken.
Warning: Do not use the proximal flow sensor if there are kinks in the tubing.
Warning: Prior to patient ventilation with the Proximal Flow option, run SST with the exact configuration that will be used on the patient. This includes a neonatal patient circuit, proximal flow sensor, and all accessories used with the patient circuit. If SST fails any proximal flow sensor test, check the patient circuit and the proximal flow sensor for leaks or occlusions and replace the flow sensor, if necessary. If SST continues to fail, it may indicate a malfunction or a leak within the proximal flow hardware which could compromise accuracy or increase the likelihood of cross-contamination; thus, replace the Proximal Flow hardware.
Warning: Changing ventilator accessories can change the system resistance and compliance. Do not add or remove accessories after running SST.
Warning: If the Proximal Flow option fails to respond as described in this appendix, discontinue use until correct operation is verified by qualified personnel.
Warning: The proximal flow sensor measures gas flow at the patient wye. The actual volume of gas delivered to the patient may be affected by system leaks between the patient and the proximal flow sensor, such as a leak that could occur from the use of an uncuffed endotracheal tube.
Warning: Position the proximal flow sensor exactly as described in this appendix or the IFU provided with the sensor.
Warning: Do not position the proximal flow sensor cables or tubing in any manner that may cause entanglement, strangulation or extubation which could lead to hypercarbia or hypoxemia. Use the cable management clips supplied to mitigate this risk.
Warning: To reduce the risk of extubation or disconnection, do not apply tension to or rotate the proximal flow sensor by pulling on the proximal flow sensor’s tubing.
Warning: Do not install the proximal flow sensor in the patient circuit if the sensor is not also connected to the
BDU.
Warning: Excessive moisture in the proximal flow sensor tubing may affect the accuracy of the measurements.
Periodically check the sensor and tubing for excessive moisture or secretion build-up.
Warning: The proximal flow sensor is intended for single use only. Do not reuse the sensor. Attempts to clean or sterilize the sensor may result in bioincompatibility, infection, or product failure risks to the patient.
Warning: Install the proximal flow sensor as shown. See
. Improper orientation of the flow sensor could lead to misinterpretation of data or incorrect ventilator settings.
Caution: Do not use aerosolized medications with the proximal flow sensor. Such medications may damage the sensor.
Caution: To prevent damage to pneumatic lines, use supplied cable management clips.
Caution: Use only Covidien-branded proximal flow sensors with the Proximal Flow option.
16.7 On-screen symbols
When using the Proximal Flow option, flow, pressure, and volume waveform data, along with delivered and exhaled volumes are derived from proximal flow sensor measurements at the patient circuit wye. Proximal flow data are displayed on the waveform plot with a Y appearing in inverse video next to the measurement symbol.
16
307
Figure 117.
Sample GUI screen Showing Proximal Flow Data
1 Data measured using proximal flow sensor PY– Monitored circuit pressure throughout the breath cycle measured at the proximal airway ⩒
Y
–
Inspiratory and expiratory patient flow
Inspired and exhaled flows and volumes at the patient wye are measured and identified by the symbols shown in
configured. See
Section 3.8.2.7, Vital Patient Data, page 79
and
.
Table 90.
Proximal Flow Option Data Symbols
Data Symbol
V
TIY
V
V
V
TEY
TE SPONTY
TE MANDY
⩒
E TOTY
V
⩒
Y
TLY
Description
Inspired tidal volume (mandatory or spontaneous at patient circuit wye)
Exhaled tidal volume (at patient circuit wye)
Exhaled spontaneous tidal volume (at patient circuit wye)
Exhaled mandatory tidal volume (at patient circuit wye)
Exhaled total minute volume (at patient circuit wye)
Flow throughout the breath cycle (at patient circuit wye)
Inspired tidal volume (at patient circuit wye with Leak
Sync enabled)
Note:
In the patient data symbols shown above, the “Y” appears in inverse video, as shown. See Figure 117
.
Note: When the Proximal Flow and Leak Sync functions are enabled, the following parameters are available for display:
• V
TLY
and V
TL
• LEAK and LEAK
Y
308
When only the Proximal Flow option is enabled, VTI
Y
and V
TI
are available for display.
When a “Y” appears in the symbol, the data are measured with the proximal flow sensor. When a “Y” is absent from the symbol, the data are measured by the ventilator’s internal flow sensors.
16.8 Sensor Calibration and Sensor Line Purging
To ensure accurate pressure and flow measurements, the ventilator performs an autozero function to calibrate the proximal flow sensor. It does this by periodically opening the pressure sensor on the Proximal Flow option PCBA to atmosphere during exhalation, and uses the resulting measurements as offset corrections.
The purge function is designed to clear the pneumatic lines of fluids that may collect, and is performed periodically by sending a brief flow of air through the sensor lines. Autozero and purge functions are only active during exhalation which limits the effect of the purge gas on delivered oxygen concentration.
During the autozero or automatic purge processes, the measurement and display of proximal flow data is not shown in real time and a brief message appears on the GUI indicating the purge process is occurring.
During autozero or automatic purge processes, the pressure waveforms, when shown display the current PEEP value and the flow waveform, when shown, displays a value of 0.
Figure 118.
Message During Autozero and Purge Processes
16
16.9 SST Requirements
SST must be run prior to ventilation and all circuit components and accessories must be installed in the configuration to be used on the patient in order for the ventilator to calculate the correct compliance and resistance. This includes a neonatal patient circuit, proximal flow sensor, and other accessories used during ventilation. See
located in that section.
provides a listing of the test sequence when running SST with the Proximal Flow option.
309
16
Table 91.
Proximal Flow Option SST Tests
Test Step
SST Flow Sensor Cross Check
SST EV Performance
SST Circuit Pressure
SST Leak
SST Exhalation Filter
Tests O
Calibrates the exhalation valve and creates a table for use during calcu‐ lations.
2
Function
and Air Flow Sensors
Exercises delivery PSOL. Checks inspiratory and expiratory autozero solenoids. Cross-checks inspiratory and expiratory pressure transducers at various pressures.
Tests ventilator breathing system for leaks.
Checks for exhalation filter occlu‐ sion and exhalation compartment occlusion.
SST Circuit Resistance
SST Circuit Compliance
SST Prox
N/A
N/A
N/A
N/A
Comments
Checks for inspiratory and expira‐ tory limb occlusions, and calculates and stores the inspiratory and expir‐ atory limb resistance parameters.
Calculates the attached patient cir‐ cuit compliance.
Verifies functionality of Proximal
Flow System.
Ventilator prompts the user to block the proximal flow sensor outlet dur‐ ing Leak test. When prompted to reconnect the patient to the expir‐ atory filter during the Exhalation Fil‐ ter test, resume blocking the proxi‐ mal flow sensor outlet.
N/A
N/A
Includes tests of barometric pres‐ sure, autozero, purge, and pressure cross check functions.
Note: Failure of the Proximal Flow option to pass SST does not prevent ventilation, but will prevent measurement with the Proximal Flow option. The ventilator will use its internal flow sensors for measurement instead of the
Proximal Flow option.
16.9.1 Attaching the Proximal Flow Sensor for SST
During SST the ventilator prompts to attach the proximal flow sensor.
To attach the proximal flow sensor to the patient circuit
1. Verify the proximal flow sensor, pneumatic lines, and connector are not damaged.
2. Open the connector panel door and firmly attach the sensor connector to the receptacle in the BDU’s front connector port labeled Prox Flow.
310
Figure 119.
Attaching the Proximal Flow Sensor to the Ventilator
16
1 Proximal flow sensor connector insertion port
2 Proximal flow sensor connector
3. When prompted, block the breathing circuit wye.
4. When prompted to attach the proximal flow sensor, unblock the circuit wye and insert the smaller end of the sensor into the wye.
5. When prompted, cap or seal the larger end of the sensor (marked with “UP” and an arrow).
6. Follow the prompts to complete SST.
If SST fails, check the patient circuit and flow sensor connections for leaks or occlusions and replace the proximal flow sensor, if necessary. Replace the Proximal Flow option hardware if SST continues to fail, then repeat SST to determine circuit compliance and resistance. See the Puritan Bennett™ 980 Series Ventilator Hardware
Options Installation Instructions , p/n 10084704 for instructions on replacing the Proximal Flow option hardware.
16.10 Disabling and Enabling the Proximal Flow Option
The proximal flow sensor can function in the Enabled state only if the circuit type is neonatal. Assuming the
Proximal Flow option is available and the ventilation type is Invasive, the new patient default value is Enabled.
After SST has been performed, the clinician may disable the Proximal Flow option, if desired.
To disable or enable the Proximal Flow option
1. In the constant access icons area, touch the configure icon.
16
A menu containing tabs appears.
2. Touch the Options tab. A screen appears containing the Installed Options and Prox tabs.
3. Touch Enabled or Disabled to enable or disable the Proximal Flow option.
311
Figure 120.
Enabling and Disabling the Proximal Flow Sensor
Note: If the Proximal Flow option has been disabled or enabled, SST does not have to be rerun unless the breathing circuit or other breathing system accessories have been changed, removed, or added.
16.11 Using the Proximal Flow Sensor
Review and follow all warnings prior to patient ventilation with the proximal flow sensor. See
.
To connect the proximal flow sensor to the ventilator:
1. Verify the proximal flow sensor, pneumatic lines, and connector are not damaged in any way.
2. Open the connector panel door and firmly attach the sensor connector to the right-most receptacle in the
BDU’s front connector port labeled Prox Flow. See Figure 119, page 311 .
To attach the proximal flow sensor between the endotracheal tube and patient circuit, and ensure the
Proximal Flow option is enabled.
1. Connect the larger end of the sensor (marked with “UP” and an arrow) to the endotracheal tube. See
little effort.
312
Figure 121.
Attaching the Proximal Flow Sensor
16
1 Endotracheal tube
2 Breathing circuit wye
2.
Note: If using a heat-moisture exchanger (HME) on the endotracheal tube, place the proximal flow sensor between the HME and the breathing circuit wye.
Connect the smaller end of the sensor to the breathing circuit wye.
3. Ensure the sensor tubing is positioned in an upward direction, as shown in Figure 121
. If the sensor needs repositioning, DO NOT rotate it by pulling on the tubing. Reposition as follows: a. Grasp the sensor’s plastic body with one hand and the breathing circuit wye with the other hand.
b. Rotate the sensor body and wye towards each other until the sensor tubing is upright.
c. Confirm a tight connection between the sensor and breathing circuit wye.
4. Use the three cable management clips provided with the sensor to attach the sensor tubing to the breathing circuit tubing. Space the clips evenly along the length of the sensor tubing. Twist the ends of each clip to close.
Note: When the ventilator is set up for Proximal Flow option operation, the proximal flow sensor can be switched as necessary. There is no need to run SST after switching sensors unless the breathing circuit or other ventilator accessories have been changed.
16.11.1 How to Perform a Manual Purge
A manual purge may be performed any time the sensor lines contain excessive condensation, moisture, or secretions.
To perform a manual purge:
1. Touch the Configure icon on the in the constant access icons area of the GUI.
16
2. Touch the Options tab. A screen appears containing the Installed Options and Prox tabs.
3. Touch the Prox tab. The Prox Setup screen appears.
313
4. Touch START that appears next to the text “Prox Manual Purge: To begin touch the Start button”. During the purge, a message appears in the GUI prompt area stating the purge process is being performed. See
.
Figure 122.
Manual Purge
16.12 Alarms
If the Proximal Flow option becomes inoperable during ventilation, the ventilator annunciates an alarm and flow sensing reverts to the ventilator’s internal delivery and exhalation flow sensors. This switch-over may be triggered by any of the following events:
• The proximal flow sensor is not detected
• Pressure and flow readings are out of range
• Hardware problems are reported by the proximal flow option PCBA
• There is a communication failure between the ventilator and the Proximal Flow option
the information contained in the remedy message to troubleshoot the alarm.
314
Figure 123.
Alarm Message—Prox Inoperative
16
16.13 Ranges, Resolutions, and Accuracies
for proximal exhaled tidal volume, proximal inspired tidal volume, proximal exhaled minute volume, and proximal flow patient data parameters.
16.13.1 Proximal Flow Sensor Specifications
Table 92.
Proximal Flow Sensor Volume Accuracy
Measurement
Exhaled tidal volume
Inspired tidal volume
±(1.0 mL +10% of reading)
±(1.0 mL +10% of reading) a The conditions under which the accuracy values apply are as follows: Sensor is used as described in this appendix or the instructions for use provided with the sensor
Table 93.
Proximal Flow Sensor Specifications
Parameter
Weight
Dead space
Pressure drop
Specification
6.6 g
<1 mL
1.5 cmH
2
O at 10 L/min
16.14 Part Numbers
The following table lists the part numbers for the Proximal Flow option kit and individual components.
Table 94.
Proximal Flow Option and Component Part Numbers
Proximal Flow option kit
Includes:
Item Part Number
10084331
16
315
Table 94.
Proximal Flow Option and Component Part Numbers (continued)
Item
Installation hardware and accessories
Proximal flow sensor, neonatal (package of 10)
NOTE : Includes 3 cable management clips
Proximal flow sensor module
Interconnect PCBA
Purge control cable
Purge supply line
PCBA mounting screws
Proximal Flow option label
Part Number
10047078
10087622
10083941
10083940
10083966
10083963
10005748
316
17
17 Appendix Trending
17.1 Overview
This appendix describes the operation of the Puritan Bennett™ 980 Series Ventilator Trending function.
Trending is a graphically-based ventilator function allowing a combination of a total of six patient data parameters or ventilator settings to be plotted vs. time. Viewing these data allows the clinician to determine the effectiveness of the patient’s therapy.
17.2 Intended Use
The Trending feature is intended to trend a patient’s respiratory parameters and ventilator settings over time, to aid physicians in assessing the effectiveness of current therapy. It is not intended to determine the course of treatment.
17.3 Safety Reminder
Warning: The ventilator offers a variety of breath delivery options. Throughout the patient’s treatment, the clinician should carefully select the ventilation mode and settings to use for that patient based on clinical judgment, the condition and needs of the patient, and the benefits, limitations, and characteristics of the breath delivery options. As the patient’s condition changes over time, periodically assess the chosen modes and settings to determine whether or not those are best for the patient’s current needs.
17.4 Trending Description
Trending enables a combination of up to six patient data parameters and ventilator settings to be plotted at one time. The user can choose from one of eight time scales.
17.5 Setting Up Trending
To set up trending
.
2. Swipe the Menu tab on the left side of the GUI and touch Trending.
The default layout appears with two trended parameters displayed on a 2-hour time scale. Alternatively, touch the layout icon and touch Trending on the Graphs tab. The default layout appears with two trended parameters displayed on a 2-hour time scale.
17
317
Figure 124.
Accessing Trending via the Menu Tab
1 Trending button appears after swiping the menu tab
3. Select the parameters to be trended by double-tapping the trended parameter name at the top of the graph or choose Presets, which automatically populates trended items with preset trended parameters. See
Section 17.8, Trending Presets, page 324
. If you desire a different layout than the two-parameter default, touch
Custom 2 and a drop-down list appears with layouts for one, two, three, four, or six parameters. Choose the trended parameters for each graph by double-tapping the parameter. A list of buttons appears with arrows to let you know more parameters may be selected. Touch the desired button for the parameter to be trended.
To exit Trending
Touch the layout icon, then touch any of the waveform layout buttons (1 through 5).
The screen will exit the Trending layout and return to displaying the selected waveforms.
17.6 Trend Parameters
Both ventilator settings and patient data parameters can be trended. Trended ventilator settings are identified with brackets around the setting. For example, if respiratory rate is chosen as a trended ventilator setting, it would appear on the GUI as [f ]. Trended patient data parameters are not bracketed. If circuit pressure were trended, its value would appear as P subject to change.
PEAK
.
and
Table 96 include trended ventilator settings and parameters and are
Table 95.
Trended Ventilator Settings
Ventilator setting
Expiratory Sensitivity
Respiratory Rate
Peak Inspiratory Flow
I:E Ratio
Oxygen Percentage
PEEP
High Pressure (in BiLevel)
Symbol
[E
SENS
[f ]
]
[ ⩒
MAX
]
[I:E]
[O
2
%]
[PEEP]
[P
H
]
318
17
Table 95.
Trended Ventilator Settings (continued)
Ventilator setting
Low Pressure (in BiLevel)
Inspiratory Pressure
Pressure Support Level
Rise Time%
Flow Sensitivity
Pressure Sensitivity
High Time (in BiLevel)
Low Time (in BiLevel)
T
H
:T
L
Ratio
Inspiratory Time
Expiratory Time
Tidal Volume (in VC, VC+)
Volume Support
Percent Support - PAV
Percent Support - TC
High Spontaneous Inspiratory Time Limit
Apnea Interval
Predicted Body Weight
Alarm Volume
Expiratory Sensitivity - PAV
Tube Size
Plateau Time
Increase O
2
%
IE Sync Trigger Threshold
Tidal Volume/PBW Ratio
Support Volume per kg
Humidifier Volume
Table 96.
Trended Patient Data Parameters
Patient data parameter
Dynamic Compliance
PAV-based Lung Compliance
PAV-based Lung Elastance
End Expiratory Flow
Peak Expiratory Flow Rate
Peak Spontaneous Flow Rate
Total Respiratory Rate
I:E Ratio
Negative Inspiratory Force
Oxygen Percentage (monitored)
Occlusion Pressure
319
Symbol
[P
L
]
[P
I
]
[P
SUPP
]
[ ]
[ ⩒
SENS
]
[P
SENS
]
[T
H
]
[T
L
]
[T
H
:T
L
]
[T
I
]
[T
E
]
[V
T
]
[V
T SUPP
]
[% Supp]
[% Supp]
[ ⤒ T
I SPONT
]
[T
A
]
[PBW]
[Alarm Volume]
[E
SENS PAV
]
[Tube Size]
[T
PL
]
[O
2
%]
[I
SYNC
]
[V
T
/PBW]
[V
T SUPP
/PBW]
[Humid Vol]
Symbol
C
DYN
C
PAV
C
STAT
E
PAV
EEF
PSF f
TOT
I:E
NIF
O
2
%
P
0.1
17
Table 96.
Trended Patient Data Parameters (continued)
Patient data parameter
End Expiratory Pressure
Intrinsic PEEP
PAV-based Intrinsic PEEP
Total PEEP
Mean Circuit Pressure
Peak Circuit Pressure
Plateau Pressure
Spontaneous Rapid Shallow Breathing Index
Dynamic Resistance
PAV-based Patient Resistance
PAV-based Total Airway Resistance
Driving Pressure
Spontaneous Inspiratory Time
Spontaneous Inspiratory Time Ratio
Vital Capacity
Exhaled Total Minute Volume
Exhaled Spontaneous Minute Volume
Exhaled Tidal Volume
Exhaled Spontaneous Tidal Volume
Exhaled Mandatory Tidal Volume
Inspired Tidal Volume
Work of Breathing
Percent Leak
Inspiratory Leak
Exhaled Tidal Volume per kg PBW
End Tidal CO
2
Compliance Ratio
Inspiratory time constant
Estimated Inspiratory volume during Leak Sync
Leak Rate at PEEP
Exhalation leak at PEEP during Leak Sync measured by the proximal flow sensor
Inspired tidal volume measured by the proximal flow sensor
Inspired tidal volume measured by the proximal flow sensor per kg PBW
Estimated inspiratory volume during Leak Sync per kg
PBW
Exhaled tidal volume measured by the proximal flow sensor
Exhaled tidal volume measured by the proximal flow sensor per kg PBW
320
Symbol
PEEP
PEEP
I
PEEP
I PAV
PEEP
L
P
MEAN
P
PEAK
P
PL f/V
T
R
DYN
R
STAT
R
TOT
P
DRIVE
T
I SPONT
T
I
/T
TOT
VC
⩒
E TOT
⩒
E SPONT
V
TE
V
TE SPONT
V
TE MAND
V
TI
WOB
TOT
%LEAK
V
LEAK
V
TE
/PBW
ETCO
2
C
20
/C
3Tau
I
V
TL
LEAK
LEAK
Y
V
TIY
V
TIY
/PBW
V
TL
/PBW
V
TEY
V
TEY
/PBW
17
Table 96.
Trended Patient Data Parameters (continued)
Patient data parameter
Exhaled spontaneous tidal volume measured by the proximal flow sensor
Exhaled spontaneous minute volume measured by the proximal flow sensor
Exhaled mandatory tidal volume when the proximal flow sensor is enabled
Inspiratory tidal volume during Leak Sync measured by the proximal flow sensor
Estimated inspiratory volume per kg PBW during Leak
Sync measured by the proximal flow sensor
Exhaled minute volume when the proximal flow sensor is enabled
Symbol
V
TE SPONTY
⩒
E SPONTY
V
TEY MAND
V
TLY
V
TLY
/PBW
⩒
E TOTY
17.7 Viewing Trended Parameters
The cursor plays an important role when using Trending. Use the cursor to determine the parameter value vs. time and details regarding events. Touch the cursor button and turn the knob to move the cursor. The cursor moves along the waveform with the y-axis displaying the parameter’s value and the x-axis showing the time. As the waveform changes, each value is shown surrounded by a highlighted box as the cursor hovers over the waveform.
If the cursor is at its left-most position on the graph, it remains there and the graph displays the earliest time stamp.
If the cursor is in between left- and right-most positions on the graph, the cursor tracks its time stamp as the graph moves and new data arrive.
If the cursor is at its right-most position on the graph, it remains there and the graph displays the latest time stamp.
Note: If a trend parameter is not selected for a particular graph, the minimum and maximum values appear as dashes (- -).
17.7.1 Time Scales
Eight time scales are available. Time scales of 1, 2, 4, 8, 12, 24, 48, and 72 hours can be viewed. The time scale is indicated by the slider track at the bottom of the Trending screen. The complete slider track represents a 72-hour time interval, and the blue shuttle that slides along the track represents the selected time scale. The selected time scale applies to all displayed trend graphs.
Although data are sampled at periodic intervals, as shown in Table 97
, the GUI screen is refreshed every minute for any time scale selected.
Table 97.
Sampling Periods for Selected Time Scales
Time scale
1 hour
2 hours
4 hours
8 hours
12 hours
24 hours
48 hours
72 hours
Sampling period
10 seconds
20 seconds
40 seconds
80 seconds
2 minutes
4 minutes
8 minutes
12 minutes
17
321
To select a time scale
1. Touch the x-axis of the graph. The time values are surrounded by a highlighted box, indicating the time scale is ready to be changed.
2. Turn the knob to select a time scale. Turning the knob clockwise reduces the time scale, and turning it counter-clockwise increases the time scale. The relative size of the shuttle indicates the time interval selected, and the time interval is displayed along the x-axis.
3. When finished, touch the x-axis again to dismiss the box.
After each time scale change, the graphs refresh with updated parameter values for that time scale.
17.7.2 Events
Events are either automatic or manual and appear as vertical tick marks on the trend graph according to their time of occurrence. When the cursor hovers over a tick mark, Event Details changes from unselectable (gray) to a button containing the event ID numbers associated with the tick mark. Touching this button causes a dialog to appear with the event ID and its description. If many IDs are present for a selected time stamp, the user is notified by an ellipsis (…) indicating more IDs are present.
When the operator modifies the real time clock setting, the system places an event marker in the trend log to denote a time or date change. The time stamp of this automatic event will be the new time setting.
Note:
are subject to change.
Table 98.
Events
Event ID Description
Manual events
13
14
15
16
9
10
11
12
7
8
5
6
3
4
1
2
17
18
19
20
21
22
Suction
Rx: Bronchodilator
Rx: Antihistamine
Rx: Steroid
Rx: Antibiotic
Rx: Muco/Proteolytic
Blood Gas
Circuit Change
Start Weaning
Stop Weaning
Bronchoscopy
X-ray
Other 1
Other 2
Other 3
Surfactant administration
Prone position
Right side position
322
17
Table 98.
Events (continued)
Event ID
23
24
25
26
71
72
73
74
67
68
69
70
75
76
77
78
63
64
65
66
58
59
60
61
55
56
57
51
52
53
54
83
84
85
86
87
79
80
81
82
Description
Start transport
Stop transport
Start N.O Rx
Stop N.O Rx
Automatic events
Changed ventilation type to Invasive
Changed ventilation type to NIV
Changed mode to A/C
Changed mode to SIMV
Changed mode to SPONT
Changed mode to BiLevel
Changed mandatory type to VC
Changed mandatory type to VC+
Changed mandatory type to PC
Changed spontaneous type to PS
Changed spontaneous type to VS
Changed spontaneous type to PAV+™
Changed spontaneous type to TC a
Time (real-time clock) changed by user
Same Patient selected
Occlusion
Circuit Disconnect
Apnea Ventilation
NIF Accepted
P
0.1
Accepted
VC Accepted
Inspiratory Pause Maneuver Completed
Expiratory Pause Maneuver Completed
Elevate O
2
Alarm volume change
Proximal Flow Sensor state Enabled b
Proximal Flow Sensor state Disabled b
Manual Inspiration
Leak Sync Enabled
Leak Sync Disabled
Changed trigger type to Pressure
Changed trigger type to Flow
Changed trigger type to IE Sync a
Enter Stand-By Ventilation
Exit Stand-By Ventilation
323
17
Table 98.
Events (continued)
Event ID
88
89
90
91 a c
Not available for neonatal circuit types.
b Only for neonatal circuit types.
Adult only.
To record a manual event
1. Touch the Manual Event text below the home icon on the GUI screen.
Description
Backup Ventilation
Changed humidifier type to Heated
Changed humidifier type to Non-heated
Changed humidifier type to HME
The manual event screen appears with arrows allowing scrolling through the available manual events. See
2. Touch Accept to confirm the event or Cancel to cancel the action.
3. View the event by hovering the cursor over the vertical tick mark and touching Event Details which now appears as a button containing event IDs. After touching the button, a dialog appears showing the event
ID and its description.
17.8 Trending Presets
The trending function enables the clinician to view a combination of up to six patient data parameters and ventilation settings that are plotted over time on a clinician-selected time scale. There are two options for how trended values are selected for display on the trending screen:
• The clinician may select each individual trended value that is displayed on the trending screen.
• The clinician may select a trending preset (a preselected group of trended values) for display.
The ventilator offers presets for adult and pediatric patients and a different set of presets for neonatal patients. The trending presets are intended to aid clinicians in assessing the effectiveness of the current therapy but are not intended to determine the course of treatment.
17.8.1 Adult and Pediatric Trending Presets
Adult and pediatric trending presets include but are not limited to:
Weaning — f/V
T
, P
0.1
, NIF, [V
T SUPP
], [V
T SUPP
], C
STAT
ARDS — P
PL
, [PEEP], V
TE
, R
STAT
, C
STAT
, [V
T
]
COPD — R
DYN
, EEF, f
TOT
, V
TE SPONT
, ⩒
E TOT
, C
STAT
VC+ — P
PEAK
, V
TE
, C
STAT
, [V
T
], [T
I
], R
STAT
PAV+™ — [% Supp], P
PEAK
, WOB
TOT
, f
TOT
, V
TE
, R
PAV
BiLevel — C
STAT
, [P
H
], [P
L
], P
MEAN
, PEEP
I
, f
TOT
LRM — P
PEAK
, PEEP, C
DYN
, V
TE
, C
STAT
, [T
I
]
17.8.2 Neonatal Trending Presets
Neonatal trending presets include, but are not limited to:
324
17
VCV — P
PEAK
, PEEP, [V
T
], P
MEAN
, V
TE
, O
2
%
PCV — P
PEAK
, V
TE
, P
MEAN
, [T
I
], PEEP, O
2
%
BPD — P
PEAK
, C
DYN
, P
MEAN
, PEEP, R
DYN
, O
2
%
SURF — V
TE
, C
DYN
, R
DYN
, P
MEAN
, [T
I
], O
2
%
Weaning — f
TOT
, V
TE
, C
DYN
, R
DYN
, [f ], O
2
%
N SIMV — f
TOT
, [f ], [P
I
], PEEP, P
MEAN
, O
2
%
N SPONT — f
TOT
, PEEP, O
2
%, I:E ratio, PSF,%LEAK
Leak Sync — %LEAK, LEAK, V
LEAK
, V
TL
, V
TE
, ⩒
E TOT
To select a trending preset
1. Touch Presets. A dialog appears with available choices. Arrows on the dialog indicate more available choices.
2. Touch the desired trending preset. As all trending presets include six parameters, the trend graph changes to a six-graph layout, populated with the six trended preset parameters described above. The chosen trended preset appears on the Presets button.
Note: If a trended preset is selected, it is not possible to change the trended parameters. You must first select a custom layout, by touching Custom, then double-tapping the parameters you desire to change.
17.9 Data Gaps
Data gaps are shown during Apnea ventilation, occlusion, circuit disconnect, Stand-By ventilation, and backup ventilation. Gaps also appear for trended parameters that are not applicable, such as parameters associated with a non-installed option or those that are not active in the current ventilator settings.
17
325
326
18
18 Appendix Glossary
18.1 Glossary
Table 99.
Glossary of Ventilation Terms analysis message A message displayed on the GUI screen during an alarm condition, identifying the root cause of the alarm.
assist breath assist-control A/C mode audio paused
A mandatory breath triggered by patient inspiratory effort in A/C and SIMV modes.
A ventilation mode where only mandatory VC, PC, or VC+ breaths are delivered to the patient.
The 2-minute period that begins after the audio paused key is pressed, where the audible portion of an alarm is muted.
augmented alarm The initial cause of an alarm has precipitated one or more related alarms. When an alarm occurs, any subsequent alarm related to the cause of this initial alarm “augments” the initial alarm.
autotriggering background checks Continuously running tests during ventilation that assess the ventilator’s electronics and pneumatics hardware.
backup ventilation
(BUV)
A safety net feature that is invoked if a system fault in the mix subsystem, inspiratory subsystem, or expiratory subsystem occurs compromising the ventilator’s ability to ven‐ tilate the patient as set.
base flow
The ventilator delivers repeated, unintended breaths triggered by fluctuating flows or pressures as opposed to patient demand. Patient circuit leaks and low flow or pressure sensitivity settings are common causes of autotriggering.
A constant flow of gas through the patient circuit during the latter part of exhalation during flow triggering ( ⩒ -Trig). The value of this base flow is 1.5 L/min greater than the operator selected value for flow sensitivity.
base message batch changes battery back-up sys‐ tem
A message given by the ventilator during an alarm condition, identifying the alarm.
Changes to multiple settings that go into effect at the same time.
The system for supplying battery back-up power to a device. The ventilator’s battery back-up system consists of a single primary battery to provide up to 1 hour of battery power to the ventilator. An optional extended battery with the same characteristics as the primary battery is available.
BD, BDU
BiLevel mode
Breath delivery or breath delivery unit. The ventilator component that includes inspira‐ tory and expiratory pneumatics and electronics.
A mixed ventilation mode combining mandatory and spontaneous breaths, where two levels of pressure are delivered (P
and T
H
.
L
and P
H
) corresponding to expiratory and inspiratory times T
L
British Oxygen Company. A standard for high pressure gas inlet fittings.
BOC breath stacking
BTPS
The delivery of a second inspiration before the previous exhalation is complete.
Body temperature and pressure, saturated, 37°C, at ambient barometric pressure, at 100% relative humidity.
compliance volume The volume of gas that remains in the patient circuit and does not enter the patient’s respiratory system.
compressor The compressor provides compressed air, which can be used in place of wall or bottled air.
constant during rate change
One of three breath timing variables (inspiratory time, I:E ratio, or expiratory time) the operator can hold constant when the respiratory rate setting changes. Applies only to the pressure control (PC) mandatory breath type (including VC+ and BiLevel).
327
18
Table 99.
Glossary of Ventilation Terms (continued) control breath
CPU
A ventilator-initiated mandatory breath delivered in A/C mode
Central processing unit. The electronic components of the ventilator (BD and GUI) responsible for interpreting and executing instructions entered by the operator.
dependent alarm An alarm that arises as a result of another primary alarm (also referred to as an augmen‐ tation).
D
SENS
Disconnect sensitivity. A setting that specifies the allowable loss (percentage) of delivered tidal volume, which if equaled or exceeded, causes the ventilator to declare a DISCON‐
NECT alarm. The greater the setting, the more returned volume must be lost before
DISCONNECT is detected. If the Leak Sync function is in use, D allowable leak rate and is expressed in terms of L/min.
SENS
is the maximum
DISS
E
SENS
EST
Diameter index safety standard. A standard for high pressure gas inlet fittings.
Expiratory sensitivity. A setting that determines the percent of peak inspiratory flow (or flow rate expressed in L/min in a PAV breath) at which the ventilator cycles from inspi‐ ration to exhalation for spontaneous breaths. Low settings will result in longer sponta‐ neous inspirations.
Extended self test. A comprehensive test of ventilator function, intended to be run by qualified service personnel.
The exhalation flow sensor assembly.
EVQ expiratory pause an operator-initiated maneuver that closes the inspiration (proportional solenoid) and exhalation valves during the expiratory phase of a mandatory breath. The maneuver can be used to determine intrinsic (auto) PEEP (PEEP
I
).
exhalation valve (EV) The valve in the expiratory limb of the ventilator breathing system that controls PEEP.
f, f
TOT failure
Respiratory rate, as a setting (f ) in A/C, SIMV, and BiLevel the minimum number of man‐ datory breaths the patient receives per minute. As a monitored value (f total number of breaths delivered to the patient.
TOT
), the average
A category of condition detected during SST or EST that causes the ventilator to enter the safety valve open state. A ventilator experiencing a failure requires removal from clinical use and immediate service.
A setting that determines the gas flow pattern of mandatory volume-controlled breaths.
Test circuit designed for use with EST.
flow pattern gold standard test circuit
GUI hard bound
Graphical user interface. The ventilator’s touch screen used to enter patient settings. and alarm settings, including off-screen keys, soft keys, and knobs.
A ventilator setting that has reached its minimum or maximum limit.
high-priority alarm As defined by international standards organizations, an alarm that requires immediate attention to ensure patient safety. When a high-priority alarm is active, the red highpriority LED indicator flashes and the high-priority audible alarm sounds (a repeating sequence of five tones that repeats twice, pauses, then repeats again), and the alarm banner on the GUI screen shows an alarm message with the ( !!! ) symbol.
HME
I:E ratio
Heat-moisture exchanger. A humidification device, also called an artificial nose.
humidification type A setting for the type of humidification system (HME, non-heated expiratory tube, or heated expiratory tubing) in use on the ventilator.
The ratio of inspiratory time to expiratory time. Also, the operator- set timing variable that applies to PC and VC+ mandatory breaths.
inspiratory pause An operator-initiated maneuver that closes the inspiration (proportional solenoid) and exhalation valves at the end of the inspiratory phase of a mandatory breath. The maneu‐ ver can be used to determine static compliance (C
STAT
) and static resistance (R
STAT
).
328
18
Table 99.
Glossary of Ventilation Terms (continued) invasive ventilation Patient ventilation while intubated with an endotracheal (or tracheostomy) tube.
latched alarm An alarm whose visual alarm indicator remains illuminated after the alarm has autoreset.
low-priority alarm As defined by international standards organizations, an alarm that indicates a change in the patient-ventilator system. During a low-priority alarm, the yellow low-priority LED indicator lights, the low-priority audible alarm (one tone) sounds, and the GUI screen shows an alarm banner with the ( ! ) symbol.
lockable alarm maintenance
An alarm that does not terminate an active audio paused function.
All actions necessary to keep equipment in, or restore it to, serviceable condition. Includes cleaning, servicing, repair, modification, overhaul, inspection, and performance verifica‐ tion.
mandatory breath A breath whose settings and timing are preset; can be triggered by the ventilator, patient, or operator.
mandatory type The type of mandatory breath: volume control (VC), VC+, or pressure control (PC).
manual inspiration An operator-initiated mandatory (OIM) breath.
medium-priority alarm
As defined by international standards organizations, an abnormal condition that requires prompt attention to ensure the safety of the patient. When a medium-priority alarm is active, the yellow medium-priority LED indicator flashes, the medium- priority audible alarm (a repeating sequence of three tones) sounds, and the GUI screen shows an alarm banner with the ( !! ) symbol.
Ventilatory mode. The algorithm that determines type and sequence of breath delivery.
Non-interchangeable screw thread. A standard for high pressure gas inlet fittings.
mode
NIST non-invasive ventila‐ tion (NIV)
Patient ventilation without the use of an endotracheal tube; instead using interfaces such as masks, nasal prongs, or uncuffed endotracheal tubes.
non-technical alarm An alarm caused due to a fault in the patient-ventilator interaction or a fault in the elec‐ trical or gas supplies that the practitioner may be able to alleviate.
normal ventilation The state of the ventilator when breathing is in progress and no alarms are active.
O
2
%
OIM ongoing back‐ ground checks
OSC
Both a ventilator setting and a monitored variable. The O centage of oxygen in the delivered gas. The O
2
setting determines the per‐
2
% monitored data is the percentage of oxygen in the gas delivered to the patient, measured at the ventilator outlet upstream of the inspiratory filter.
Operator-initiated mandatory breath. A breath delivered when the operator presses the manual inspiration key.
Continuously running tests during ventilation that assess the ventilator’s electronics and pneumatics hardware.
Occlusion status cycling. A ventilation mode in effect during a severe occlusion. In this mode, the ventilator periodically attempts to deliver a pressure-based breath while monitoring the inspiratory and expiratory phases for the continuing existence of the occlusion.
OVERRIDDEN patient circuit
The final status of an SST or EST run in which the operator used the override feature. (The ventilator must have ended the test with an ALERT condition.) Failures cannot be over‐ ridden.
The entire inspiratory-expiratory conduit, including tubing, humidifier, and water traps.
patient data alarm An alarm condition associated with an abnormal condition of the patient’s respiratory status.
patient problems A definition used by the ventilator’s safety net. Patient problems are declared when patient data are measured equal to or outside of alarm thresholds and are usually selfcorrecting or can be corrected by a practitioner. The alarm monitoring system detects and
329
18
Table 99.
Glossary of Ventilation Terms (continued)
PBW
PC
P
DRIVE
PE
PEEP
PEEP
P
PI
P
P
P
P
I
I END
PIM
MEAN
PEAK primary alarm
PS
SENS
PSOL
P
SUPP
P-Trig
I announces patient problems. Patient problems do not compromise the ventilator’s per‐ formance.
Predicted body weight, a ventilator setting that specifies the patient’s body weight assuming normal fat and fluid levels. Determines absolute limits on tidal volume and peak flow, and allows appropriate matching of ventilator settings to the patient.
Pressure control. A mandatory breath type in which the ventilator delivers an operator-set inspiratory pressure for an operator- set inspiratory time. Available in A/C and SIMV modes, and for operator-initiated mandatory (OIM) breaths in SPONT mode.
While ventilating with PAV+™, Driving Pressure (P
Pressure and estimated PEEP
(V
T
/C
PAV
).
TOT
DRIVE
) is the difference between Plateau
and represents the amount of pressure required to over‐ come the elastic recoil of the lungs. P
DRIVE
can also be calculated by dividing V
T
by C
PAV
PAV+™ software automatically performs a 300 ms end-inspiratory hold in a random pat‐ tern every four to ten PAV+™ breaths. Each time the end-inspiratory hold is conducted this pressure change is measured from the end expiratory lung pressure at the beginning of the breath to the pressure at the end of the plateau maneuver.
Expiratory pressure transducer.
Positive end expiratory pressure. The measured circuit pressure (referenced to the patient wye) at the end of the expiratory phase of a breath. If expiratory pause is active, the displayed value reflects the level of any active lung PEEP.
Intrinsic PEEP. Indicates a calculated estimate of the pressure above the PEEP level at the end of exhalation. Determined during an expiratory pause maneuver.
Inspiratory pressure. The operator-set inspiratory pressure at the patient wye (above
PEEP) during a pressure control (PC) mandatory breath.
Inspiratory pressure transducer.
End inspiratory pressure. The pressure at the end of the inspiratory phase of the current breath. If plateau is active, the displayed value reflects the level of end-plateau pressure.
Patient-initiated mandatory breath. A mandatory breath triggered by patient inspiratory effort.
Mean circuit pressure, a calculation of the measured average patient circuit pressure over an entire respiratory cycle.
Maximum circuit pressure, the maximum pressure during the inspiratory and expiratory phases of a breath.
An initial alarm.
Pressure support, a spontaneous breath type in which the ventilator delivers an opera‐ tor-set pressure (in addition to PEEP) during the inspiratory phase. Available in SPONT,
SIMV, and BiLevel modes.
Pressure sensitivity. The operator-set pressure drop below PEEP (derived from the patient’s inspiratory flow) required to begin a patient-initiated breath when pressure triggering is selected.
Proportional solenoid valve.
Pressure support. A setting of the level of inspiratory assist pressure (above PEEP) at the patient wye during a spontaneous breath (when spontaneous breath type is PS).
Pressure triggering, a method of recognizing patient inspiratory effort in which the ven‐ tilator monitors pressure in the patient circuit. The ventilator triggers a breath when the airway pressure drops by at least the value selected for pressure sensitivity (P
SENS
).
330
18
Table 99.
Glossary of Ventilation Terms (continued) remedy message A message displayed on the GUI during an alarm condition suggesting ways to resolve the alarm.
resistance restricted phase of exhalation
The flow-dependent pressure drop across a conduit. Measured in cmH
2
O/L/s or hPa/L/s.
The time period during the expiratory phase where an inspiration trigger is not allowed.
The restricted phase of exhalation is defined as the first 200 ms of exhalation, or the time it takes for expiratory flow to drop to ≤50% of the peak expiratory flow, or the time it takes for the expiratory flow to drop to ≤0.5 L/min (whichever is longest). The restricted phase of exhalation will end after 5 seconds of exhalation have elapsed regardless of the meas‐ ured expiratory flow rate.
rise time% A setting that determines the rise time to achieve the set inspiratory pressure in pressurecontrolled (PC), VC+, BiLevel, volume supported (VS) or pressure supported (PS) breaths.
The larger the value, the more rapid the rise of pressure.
safety net safety valve (SV) safety ventilation service mode
SIMV
The ventilator’s strategy for responding to patient problems and system faults.
A valve residing in the ventilator’s inspiratory module designed to limit pressure in the patient circuit. When open, it allows the patient to breathe room air if able to do so.
A mode of ventilation active if the patient circuit is connected before ventilator startup is complete, or when power is restored after a loss of 5 minutes or more.
A ventilator mode providing a set of services tailored to the needs of testing and main‐ tenance personnel. When in the service mode, the ventilator does not provide ventilation.
Synchronized intermittent mandatory ventilation. A ventilatory mode in which the ven‐ tilator delivers one mandatory breath per breath cycle and as many spontaneous breaths as the patient can trigger during the remainder of the breath cycle.
SIS soft bound
SPONT Spontaneous. A ventilatory mode in which the ventilator delivers only spontaneous breaths. In SPONT mode, the patient triggers all breaths delivered by the ventilator with no set mandatory respiratory rate. The patient controls the breath variables, potentially augmented by support pressure.
spontaneous type A setting that determines whether spontaneous breaths are pressure-supported (PS), tube-compensated (TC), volume-supported (VS), or proportionally assisted (PAV).
SST Short self test. A test that checks circuit integrity, calculates circuit compliance and filter resistance, and checks ventilator function. Operator should run SST at specified intervals and with any replacement or alteration of the patient circuit.
STPD
Sleeved index system. A standard for high pressure gas inlet fittings
A ventilator setting that has reached its recommended high or low limit, accompanied by an audible tone. Setting the ventilator beyond this limit requires the operator to acknowl‐ edge a visual prompt to continue.
SVO system fault
Standard temperature and pressure, dry. Defined as dry gas at a standard atmosphere
(760 mmHg, 101.333 kPa, approximately 1.0 bar) and 0°C.
Safety valve open. An emergency state in which the ventilator opens the safety valve so the patient can breathe room air unassisted by the ventilator (if able to do so). An SVO state does not necessarily indicate a ventilator inoperative condition. The ventilator enters an SVO state if a hardware or software failure occurs that could compromise safe ventilation, with the loss of the air and oxygen supplies, or if the system detects an occlusion.
A definition used by the ventilator’s safety net. System faults include hardware faults
(those that originate inside the ventilator and affect its performance), soft faults (faults momentarily introduced into the ventilator that interfere with normal operation), inad‐ equate supply (AC power or external gas pressure), and patient circuit integrity (blocked or disconnected circuit).
331
18
Table 99.
Glossary of Ventilation Terms (continued)
T
A
Tb
T
T
Ts
⩒
V
I
E technical alarm
Tm
T
PL
T
E TOT
VBS
VC
Ventilation Assur‐ ance
Ventilator Inopera‐ tive (vent inop)
VIM
⩒
⩒
MAX
SENS
Apnea interval, the operator-set variable that defines the breath-to-breath interval which, if exceeded, causes the ventilator to declare apnea and enter apnea ventilation.
Breath time cycle during mechanical ventilation.
Expiratory time. The expiratory interval of a breath. Also the operator-set timing variable that determines the expiratory period for pressure-controlled (PC) or VC+ mandatory breaths.
An alarm occurring due to a violation of any of the ventilator’s self monitoring conditions, or detected by background checks.
Inspiratory time, the inspiratory interval of a breath. Also, the operator-set timing variable that determines the inspiratory interval for pressure-controlled (PC) or VC+ mandatory breaths.
Mandatory interval portion of SIMV breath cycle; it is reserved for a PIM.
Plateau time. The amount of time the inspiratory phase of a mandatory breath is extended after inspiratory flow has ceased and exhalation is blocked. Increases the residence time of gas in the patient’s lungs.
Spontaneous interval portion of SIMV breath cycle; it is reserved for spontaneous breath‐ ing throughout the remainder of the breath cycle.
Minute volume, the expiratory tidal volume normalized to unit time (L/min). The dis‐ played value is compliance- and BTPS-compensated.
Ventilator breathing system. Includes the gas delivery components of the ventilator the patient circuit with tubing, filters, humidifier, and other accessories; and the ventilator’s expiratory metering and measurement components.
Volume control, a mandatory breath type in which the ventilator delivers an operator-set tidal volume, peak flow, and flow pattern. Available in A/C and SIMV modes, and for operator-initiated mandatory (OIM) breaths in SPONT mode.
A safety net feature which is invoked if a system fault in the mix subsystem, inspiratory subsystem, or expiratory subsystem occurs compromising the ventilator’s ability to ven‐ tilate the patient as set.
An emergency state the ventilator enters if it detects a hardware failure or a critical soft‐ ware error that could compromise safe ventilation. During a ventilator inoperative con‐ dition, the safety valve opens to allow the patient to breathe room air (if able to do so) unassisted by the ventilator. Qualified service personnel must power up the ventilator and run EST before normal ventilation can resume.
Ventilator-initiated mandatory breath. A breath that is delivered at a time determined by the ventilator.
Peak flow. A setting of the peak (maximum) flow of gas delivered during a VC mandatory breath. (Combined with tidal volume, flow pattern, and plateau, constant peak flow defines the inspiratory time.) To correct for compliance volume, the ventilator automat‐ ically increases the peak flow.
Flow sensitivity. A setting that determines the rate of flow inspired by the patient that triggers the ventilator to deliver a mandatory or spontaneous breath (when flow trig‐ gering is selected).
Tidal volume. A setting that determines the volume inspired and expired with each breath. The V
T
delivered by some Puritan Bennett ventilators is an operator-set variable that determines the volume delivered to the patient during a mandatory, volume-based
332
18
Table 99.
Glossary of Ventilation Terms (continued)
⩒ -TRIG breath. V
T
is compliance-compensated and corrected to body temperature and pressure, saturated (BTPS).
Flow triggering. A method of recognizing patient inspiratory effort in which the ventilator monitors the difference between inspiratory and expiratory flow measurements. The ventilator triggers a breath when the difference between inspiratory and expiratory flows increases to a value that is at least the value selected for flow sensitivity ( ⩒
SENS
).
s
V
mL
ms
VA
cm
cmH
2
O ft
hPa
Hz
kg
kPa
L lb
L/min
m
Table 100.
Units of Measure
Centimeter. A unit of length.
Centimeters of water. A unit of pressure approximately equal to 1 hPa.
Feet. A unit of length.
Hectopascal. A unit of pressure, approximately equal to 1 cmH
2
O.
Hertz. A unit of frequency, indicating cycles per second.
Kilogram. A unit of weight.
Kilopascal. A unit of pressure approximately equal to 10 cmH
2
O.
Liter. A unit of volume.
Pound. A unit of weight.
Liters per minute. A unit of flow.
Meter. A unit of length.
Milliliter. A unit of volume.
Millisecond. A unit of time.
Second. A unit of time.
Volts. A unit of voltage.
Volt-amperes. A unit of power.
Table 101.
Technical Abbreviations
AC, also ac
ASCII
CE
CSA
CRC
DC, also dc
EMC
EN
ETO
IEC
ISO
LCD
LED
MRI
Alternating current. The movement of electrical charge that periodically reverses direc‐ tion.
American Standard Code for Information Interchange. A standard character encoding scheme.
A certification mark issued under the authority of the European Common Market that indicates compliance with the Medical Device Directive, 93/42/EEC.
Canadian Standards Association.
Cyclic Redundancy Check or Code. An algorithm or a computational result based on the remainder of a division defined over the ring of polynomials in the Galois field GF(2). CRC algorithms are the basis for data integrity checks.
Direct current. The movement of electrical charge flowing in a single direction.
Electromagnetic compatibility.
European norm (referring to the European Common Market).
Ethylene oxide.
International Electrotechnical Commission. A standards organization.
International Standards Organization. A standards organization.
Liquid crystal display. A type of visual equipment-operator Interface.
Light-emitting diode. A means of providing visual indications.
Magnetic resonance imaging.
18
333
Table 101.
Technical Abbreviations (continued)
NVRAM, also Nov‐
Ram
POST
RAM
Non-volatile random access memory. Memory that is kept active across resets and power cycles and is not normally initialized at startup.
Power-on self test. Software algorithms to verify the integrity of application software and the hardware environment. Power-on self test generally occurs at power on, after power loss, or when the device detects an internal fault.
Random access memory.
334
18
Index
A accessory
alarm
DEVICE ALERT 157 high circuit pressure 157
% 158 high exhaled minute volume 158 high exhaled tidal volume 158 high inspired tidal volume 158
low circuit pressure 159 low delivered O
alarm settings range and resolution 252
335
B
background diagnostic system 237
battery
BDU indicators
breath triggers
C
compliance compensation 199 compliance compensation in volume-based breaths 199
component cleaning and disinfection 170
configurable features
mL/kg ratio 78 new patient setup defaults 78
pressure units 77 screen brightness and keyboard backlight 77
connectivity to external patient monitoring systems 130
constant timing variable for rate changes 104
18
D
detecting occlusion and disconnect 220
display
brightness adjustement 94 lock 94
E
EMC
recommended separation distances 266
exhalation
time cycling method 199 time limit (backup method) 199
exhalation — detection and initiation 198
exhalation flow sensor assembly (EVQ)
expiratory module
exhalation flow sensor assembly (EVQ) removal, disinfection, reassembly 172
expiratory pause maneuvers 111
extended battery installation 69
F
flow pattern 231 flow sensitivity ( ⩒
G
336
double-tap 95 drag 95 drag and drop 95
graphical user interface (GUI) 32
GUI indicators
H
high spontaneous inspiratory time limit setting 108
high spontaneous inspiratory time limit ( ⤒
how to connect the gas supplies 57
how to connect the patient circuit 64
how to connect the ventilator to AC power 55
how to store the ventilator for an extended time period 182
how to use the ventilator system 95
how to use ventilator’s user interface 92
how to view ventilator logs 186
I icons
low priority alarm 36 maximize waveform 36 medium priority alarm 36
18
pause 36 restore waveform 36 screen capture 36 unread items 36
inspiration — detection and initiation 195
installation testing (testing prior to ventilating a patient) 82
L
M
manufacturer’s declaration 261
Medtronic Technical Services
Solv-IT Center knowledge base 20
N
NeoMode
NIV
ventilator settings and ventilation type 300
alarm settings 108 apnea settings 108 high spontaneous inspiratory time limit setting 108
337
Non-Invasive Ventilation (NIV) 105
O
on-screen symbols and abbreviations 37
oxygen sensor
P
patient data range and resolution 253
percent support in PAV+ TM 234
preparing the ventilator for use 76
primary battery installation 67
Q
R respiratory maneuvers
inspiratory pause maneuver 223
18
respiratory mechanics maneuvers
negative inspiratory force maneuver (NIF) 225
maneuver 226 vital capacity maneuver (VC) 226
RS-232 commands
S
serial number interpretation 20
settings
specifications
environmental 245 performance 245
spontaneous breath delivery 205
SST
surface cleaning of exterior surfaces 169
338 symbols
BDU rear panel label symbols and descriptions 30
shipping label symbols and descriptions 13
T
TC
alarms 209 monitored patient data 209
PBW and tube ID 209 technical description 209
tube type, tube ID, humidification 100
U
V
VC+
pressure adjustment limits according to patient body weight (PBW) 205
ventilating a new patient 96 ventilating the same patient 96
ventilator
BDU controls and indicators 40
BDU rear label symbols and descriptions 30
18
patient data log 186 service log 186
ventilator protection strategies 114
ventilator settings
217, 227 circuit type and PBW 227
disconnect sensitivity (D expiratory sensitivity (E
humidification type 235 humidifier volume 235
inspiratory time (T low pressure (P
PEEP 233 percent support in TC 233
VS
pressure adjustment limits according to patient body weight (PBW) 208
W
18
339
Covidien llc
15 Hampshire Street, Mansfield, MA 02048 USA
Covidien Ireland Limited, IDA Business and Technology Park, Tullamore, Ireland.
www.covidien.com
© 2021 Covidien
PT00147890 A
2021-11-22
*PT00147890*
Advertisement