Physio Range User Manual

PULSEPRESS
DYNAMIC COMPRESSION THERAPY
OPERATING INSTRUCTIONS
0120
ENGLISH
CONTENTS:ENGLISH
SAFETY INSTRUCTIONS
ELECTROMAGNETIC INTERFERENCE
DESCRIPTION OF CONTROLS
CARE AND MAINTENANCE
INDICATIONS AND CONTRA-INDICATIONS
TECHNICAL SPECIFICATIONS
FRANÇAIS
INSTRUCTIONS DE SECURITE
INTERFERENCE ELECTROMAGNETIQUE
DESCRIPTION DES COMMANDES
ENTRETIEN ET MAINTENANCE
INDICATIONS ET CONTRE-INDICATIONS
SPECIFICATIONS TECHNIQUES
DEUTSCH
SICHERHEITSHINWEISE
ELEKTROMAGNETISCHE STÖRUNG
BESCHREIBUNG DER STEUERELEMENTE
PFLEGE UND WARTUNG
ANZEIGEN UND GEGENANZEIGEN
TECHNISCHE DATEN
ITALIANO
ISTRUZIONI DI SICUREZZA
INTERFERENZA ELETTROMAGNETICA
DESCRIZIONE DEI COMANDI
CURA E MANUTENZIONE
INDICAZIONI E CONTROINDICAZIONI
SPECIFICHE TECNICHE
PORTUGUESE
INSTRUÇÕES DE SEGURANÇA
INTERFERÊNCIA ELECTROMAGNÉTICA
DESCRIÇÃO DOS CONTROLOS
CUIDADOS E MANUTENÇÃO
INDICAÇÕES E CONTRA-INDICAÇÕES
ESPECIFICAÇÕES TÉCNICAS
ESPAÑOL
INSTRUCCIONES DE SEGURIDAD
INTERFERENCIA ELECTROMAGNÉTICA
DESCRIPCIÓN DE LOS MANDOS
CUIDADO Y MANTENIMIENTO
INDICACIONES Y CONTRAINDICACIONES
ESPECIFICACIONES TÉCNICAS
PULSEPRESS
DYNAMIC COMPRESSION THERAPY
ENGLISH
SAFETY INSTRUCTIONS
Read all of these instructions and save them for later reference. Follow all
warnings and instructions marked on the devices.
•
•
•
•
•
•
•
•
•
•
These units must only be operated by qualified medical staff.
Operation of these units by any other person is entirely the
responsibility of that person.
This manual must be read and fully understood before these units are
applied to patients.
Equipment not suitable for use in the presence of Flammable Anesthic
mixture with air or with oxygen or nitrous oxide.
Unplug the pumps before cleaning. Only ever use a damp cloth. Do
not spill liquid onto the pump enclosure.
Always place pumps on a flat, stable surface before operation. The
pumps will not operate properly if tilted or at an angle.
Do not cover the pumps whilst in operation.
Use only the type of power source indicated on the pump’s rear label.
Do not use a damaged or frayed power cord.
Unplug the pumps and refer servicing to qualified personnel under the
following conditions:o If the power cord or plug is damaged.
o If liquid has entered the pumps.
o If the plastic outer enclosure is damaged.
o If the pump is not operating normally, continuous running,
noisy running or low performance.
Always switch off the pump before unzipping the garments.
ELECTROMAGNETIC
INTERFERENCE
This equipment has been constructed and tested to comply with EN60601-12 (EMC regulations for Medical devices). This test was designed to provide
reasonable protection against harmful interference. This equipment may
generate and radiate radio frequency energy and, if not installed correctly may
cause interference to radio communications. If this equipment does cause
interference then the user is encouraged to try to correct the interference by
one or more of the following measures:•
•
•
•
Re-orientate or relocate the receiving antenna.
Increase the separation between the equipment and the receiver.
Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
Consult the manufacturer of the device.
PULSEPRESS
DYNAMIC COMPRESSION THERAPY
PHYSIO 3 CONTROLS
1
2
3
1. Pressure Control: This sets the air pressure in the garments to
that stated on the front panel in mmHg. 20 - 60 mmHg
2. Air Outlet: This is where the garment air tube is attached.
3. Power Switch: This turns the unit on and off. The unit starts
cycling as soon as power in turned on.
The system automatically sets the inflation to that needed by the
garment being used. Once the pressure in the garnent hits the set
pressure the system deflates for a fixed time period.
4. Power Inlet: This is where the power cable is plugged in.
4
PHYSIO 3 PRO CONTROLS
1
2
3
1. Pressure Control: This sets the air pressure in the garments. To
set the pressure attach a garment and set the inflation to 120 secs.
Switch the pump on and wait for the pressure meter reading to settle
on a number. Now adjust the pressure, clockwise to increase, to the
desired level, giving the system time to settle. 2. Pressure Meter: This displays the pressure in the garment at all
times in the cycle.
3. Deflate Timer Control: This sets the deflation time cycle, in
seconds.
4. Air Outlet: This is where the air tube is attached. The unused air
outlet is plugged. Unplug for second garment.
5. Power Inlet: This is where the power cable is plugged
6. Power Switch: This turns the unit on and off. The unit starts
cycling as soon as power in turned on.
4 56
7
7. Inflate Timer Control: This sets the inflation time cycle, in seconds.
NB if the inflate time is set too low then the system will not have time
to reach the desired pressure.
PHYSIO 6 CONTROLS
1
1. Power Switch: This turns the unit on and off. The unit starts
cycling as soon as power in turned on.
2
2. Pressure Display: This displays the pressure set by the user.
When the set pressure is reached in the garment the compressor switches off and waits for the cycle time to complete prior to
deflating the garment. When this happens 0mmHg is displayed.
3. Air Outlet: This is where the air tube is attached. For use with
a single garment the top bung must be in place. Remove the
bung for use with two garments.
4. Power Inlet: This is where the power cable is plugged
5. Pressure Control: This sets the air pressure in the garments
to that stated on the pressure display from 30-100mmHg
34
5
6
6. Deflate Timer Control: This sets the deflation time cycle, in
seconds.
7
7. Inflate Timer Control: This sets the inflation time cycle, in
seconds.
PHYSIO 12 PRO CONTROLS
1
2
98
1. Power Switch: This turns the unit on and off. The unit starts
cycling as soon as power in turned on.
2. Pressure Display: This displays the pressure set by the user.
When the set pressure is reached in the garment the compressor switches off and waits for the cycle time to complete prior to
deflating the garment. When this happens 0mmHg is displayed.
3. Air Outlet: This is where the air tube is attached. For use with
a single garment the top bung must be in place. Remove the
bung for use with two garments.
4. Power Inlet: This is where the power cable is plugged
5. Pressure Control: This sets the air pressure in the garments
to that stated on the pressure display from 30-100mmHg
6. Deflate Timer Control: This sets the deflation time cycle, in
seconds.
34
5
6
7
7. Inflate Timer Control: This sets the inflation time cycle, in
seconds. Auto mode sets the inflation and deflation setting to
the optimal for the garment. Just select a pressure and the rest
is done by the internal microprocessor
8. Therapy Control. This sets the therapy time, in minutes. The
black start button needs to be pushed to start the unit. Once
complete the unit bleeps and vents.
9. Start button. Press this to start the unit running. Once selected therapy time is completed press this to restart the system
ENGLISH
opERAting instRuctions
WARNING: These units must only be operated by qualified medical staff.
Operation of these units by any other person is entirely the responsibility of that
person.
WARNING: This manual must be read and fully understood before these units are
applied to patients.
WARNING: Refer to section on indications and contra-indications prior to the
unit being applied to patients.
Application to Patient
These units are designed to provide compression to the Pulse Press range of
Solo and Multi pressure garments. Each unit can be connected to one or two
garments simultaneously. If only one garment is being used then the other
outlet is sealed.
For the Multi pumps the blanking plug must be removed to use a second
garment. The hoses connection is unable to be the wrong way round. The
other end of the air feed hose simply pushes onto the garment.
For the Solo pumps the hose latches into the outlet with a push and twist
action. The other end of the air feed hose pushes into the garment inlet elbow.
Operational Use
1. Connect the mains lead to the IEC power inlet socket. Insert the
mains plug of the lead into a standard 220-240 V mains socket.
2. Fit the air feed hose(s) to the air outlets, as described above. Ensure
that all connections to the pump and garment make a good seal.
3. Place / fit the garment(s) onto the limb(s) requiring treatment.
4. Turn the unit on. A visual mains indicator will illuminate.
5. Select the required pressure using the pressure control.
6. Set up the inflate and deflate times as required.
7. Note the time the treatment starts so that the unit can be switched off
when the desired time has elapsed.
8. DO NOT unzip the garment when the unit is switched on. This may
cause damage to the garment. Always turn off the power before
unzipping the garment.
9. Extra comfort can be obtained by first placing a Pulse Press Garment
Liner on the limb under the pressure garment.
PULSEPRESS
DYNAMIC COMPRESSION THERAPY
ENGLISH
CARE AND MAINTENANCE
PUMP
Before cleaning the pump unit, always ensure it has been disconnected from
the mains power supply. Clean the unit with a light detergent, ensuring that
only a damp, and not wet, cloth is used. If in doubt as to the action of the
detergent, try cleaning the underside of the unit first, where damage will be
unseen.
There are no service items which need replacing or cleaning out on a regular
basis. All technical information can be found at ‘www.mjsgrp.com’ .
GARMENT
The garments can be sponged with a detergent. Allow to dry fully before use.
DO NOT FORCE DRY.
In the event of heavy soiling the garments can be machine washed. See wash
label. Always insert the air entry blanking plug, to prevent water ingress, and
turn the garment inside out prior to machine washing.
inDICATIONS AND CONTRA-INDICATIONS
INDICATIONS
VENOUS
Post phlebetic syndrome, varicose veins,
varicosities of pregnancy, stasis dermatitis,
leg ulcers.
WOUND RECOVERY
Venous ligation or stripping, hand surgery,
total knee replacement.
D.V.T.
For prevention of deep vein thrombosis.
ARTERIAL
Raynaud’s disease, arteriosclerosis, acro
cyanosis, blue leg.
EDEMA
Gravitational, lymphatic, post mastectomy,
inguinal lymphadenectomy, after cast
removal, rheumatoid, multiple sclerosis.
FLEXION CONTRACTURES
Chronic, injury, burns and skin graft.
CONTRA-INDICATIONS
Acute thrombophlebitis, acute pulmonary edema, congestive heart failure,
known or suspected deep vein thrombosis.
PULSEPRESS
DYNAMIC COMPRESSION THERAPY
ENGLISH
technical specifications
Power Source 220-250V AC
Frequency
50Hz
Internal Fuse
T500mA
Type BF
Caution
!
Equipment not suitable for use in the presence of
Flammable Anesthic mixture with air or oxygen
or nitrous oxide.
Read instructions prior to using equipment.
Water Resistance
Ordinary Equipment
Mode Intermittent Operation
Storage / Operation
Equipment to be stored / used between 20 and 30 degrees C
Equipment suitable for use 0-50% humidity.
Pressure Regulation
+_
Range 30 - 60 / 100mmHg 5mmHg
Pressure Meter
+_
LCD Display. Range 0-100mmHg 2mmHg
The following data applies to Physio 3, Physio 3 Pro, Physio 6
Power Consumption
20W
Class II (Double Insulated Device)
The following data applies to the Physio 12 Pro
Power Comsumption
40W
Class II ( Double Insulated Device)
PULSEPRESS
DYNAMIC COMPRESSION THERAPY