Zoll Powerheart Software User's guide
ZOLL Powerheart Software helps you manage your Powerheart AEDs, review rescues, and customize settings to match your specific needs. Use it to select the RescueCoach voice and visual prompts, customize CPR protocols, analyze patient ECGs, add patient and rescue information, export rescue data, and print reports.
Advertisement
Advertisement
Using AED Manager
70-00974-02 E
Copyright © 2014 Cardiac Science Corporation. All rights reserved.
Contents u
u u
Setting AED and rescue parameters
u
1
Using AED Manager
About AED Manager
AED Manager assists Medical Directors, Automated External Defibrillator (AED) administrators, and emergency medical technicians with configuring Powerheart® G5
AEDs and reviewing rescue incidents of victims of sudden cardiac arrest.
With AED Manager, you can…
Select the RescueCoach™ audio and visual rescue prompts to match the coaching needs and language of a typical user*
See page…
Customise many aspects of a rescue protocol—for example, lengths and types of CPR, and shockable conditions*
Review rescue incidents, including the ECG of the patient
Customise and add patient and rescue-specific information to the data recorded by an AED
View self-test and other event messages for each AED
Export saved demographic and rescue data for use in other applications
Print and save PDF Reports
Configure remote AEDs*
Note: *Any rescue setting should be designated only by authorised, trained medical personnel.
2 70-00974-02 E
Overview of the AED Manager window
3
Overview of the AED Manager window
6
AED list
1
2
4 5
70-00974-02 E
AED list
1
2
3
•
Click the arrow to hide or show the AED list.
Search box. Type the name or serial number of the AED to show in the list.
List of AEDs that have been or are currently connected to the computer. Select an AED to view information for it.
Click + to show the list of connected or unconnected AEDs. When the list is expanded, click to hide the list of AEDs.
5
6
AED configuration and information tabs
4 General information about the AED, including name, type, and battery status.
Configure settings of the AED here.
View messages saved by the AED.
View recordings of rescue sessions.
3
Using AED Manager
Running AED Manager
AED Manager can be used on Windows XP®, Windows Vista®, and Windows 7®. You need to be logged into Windows as either an Administrator or a normal user. You cannot run AED Manager when logged in as a guest.
For the best performance, close all programs, especially those that use audio and video files, before using AED Manager.
To open AED Manager:
1.
Double-click the AED Manager icon on the desktop.
2.
When the splash screen appears, choose either Administrator or Normal User .
You need to enter the password for Administrator .
• Administrator allows you not only to view AED information, but change and update AED settings.
Important: Administrator level access should be limited to those who have medical training and authority to make changes to rescue protocols or other therapy settings of an AED.
3.
• Normal User allows you only to view, export, and print AED information.
Click Continue .
Setting AED Manager application preferences
Select Settings | Preferences .
On the General panel, set these preferences: u Language : The language in which to use AED Manager.
Note: Restart AED Manager to have the language selection take effect.
u Password : If you started AED Manager as Administrator, you can click
Change Password to update the Administrator password.
On the Export panel, select a folder to save exports of rescue data. To select a folder, click Browse , then navigate to the folder you want to use.
4 70-00974-02 E
Transferring AED data to AED Manager
Transferring AED data to AED Manager
You can transfer rescue and event data with a USB flash drive from one or more AEDs or directly with a USB cable from an AED to AED Manager . You can also use flash drives to re-configure AEDs where they are kept instead of transporting the units to the computer.
!
WARNING! Electric Shock and Fire Hazard
Do not connect any telephones or unauthorised connectors to the socket on this equipment.
!
CAUTION! Possible loss of data.
Do not remove the flash drive or disconnect the AED while data is being transferred.
Data may be lost or corrupted.
USB flash drive—
Note: For better reliability of the data transfer, use one of these brands of flash drive:
Lexar JumpDrive, SanDisk Cruzer or HP.
1.
Disconnect the defibrillation pads.
2.
3.
•
•
Attach the flash drive to the USB port.
The AED copies information to the flash drive.
When finished, the AED prompts to remove the flash drive.
Remove the flash drive.
4.
5.
Connect the pads to the AED and close the lid.
Copy the AED’s rescue data from the flash drive to AED Manager .
70-00974-02 E 5
Using AED Manager
To view the data: u Plug the flash drive into a USB port on the computer. If a Windows Explorer window opens, close it.
Each AED that has data on the drive appears in the AED list under Not
Connected. You can view, print and export the data as you would with a connected AED.
USB cable—
1.
Disconnect the defibrillation pads.
2.
Attach the USB cable to the USB port.
3.
4.
5.
Connect the other end to the computer’s USB port. If a Windows
Explorer window opens, close it.
•
Start AED Manager .
The AED prompts, “Communications Mode”.
• The AED appears in the AED list and the data is copied into AED
Manager’s database.
Disconnect the USB cable
After you disconnect the AED, its icon moves from Connected to Not
Connected in the AED list.
6.
Connect the pads.
6 70-00974-02 E
Setting AED and rescue parameters
Setting AED and rescue parameters
Important: You must have Administrator access to change rescue parameters. When starting AED Manager, log in as Administrator. If you are currently running as User, exit and restart.
To display the current configuration of an AED, click its icon or name in the AED list.
Under the Summary tab, you can customise many attributes of each AED in your inventory. The panels on the Summary window set different features of the AED:
General settings
Time settings
Language and rescue prompting
Date and time of the AED’s clock
Shock settings
CPR settings
Thresholds for delivering therapy
CPR protocols for Adult and Paediatric therapy
CPR Device settings Thresholds for CPR prompting when adult pads with CPR
Device assistance are attached to the AED.
After you finish setting the parameters, click Save Configuration or the Save to AED toolbar icon. AED Manager sends the configuration to the connected AED.
To revert all configuration settings on all configuration tabs to their original values since the last save, click the Undo All Changes toolbar icon.
To revert the AED’s settings to the values that it was shipped with, click Restore
Factory Defaults and save.
70-00974-02 E 7
Using AED Manager
General Settings: Set language and prompting
Use the GENERAL SETTINGS panel to set these preferences.
Preference
AED Identification
Description
The name of the AED as it appears in the AED list and on reports. A name can be up to 16 characters long. Use any of these characters: A-Z, a-z, 0-9, dash, space.
If you manage more than one AED, choose a naming convention, such as one based on the location of the units.
Choose the language the AED uses when its lid opens. The user can change to the other listed language during a rescue.
Primary Language
(For dual-language models only)
AED Overall Prompting Level Basic , Standard or Advanced
See the Powerheart G5 User’s Guide for descriptions and the prompts used by each level.
Volume Level Low —Use for an office or other quiet areas.
Maximum —use for out of doors or noisy areas.
CPR First
Startup Prompt
The AED can begin a rescue either by analysing the patient’s ECG or by instructing the rescuer to begin CPR.
Enabled —The AED provides prompts for performing CPR first.
Disabled —The AED analyses the patient’s ECG first.
•
•
The AED provides an initial prompt when the lid is opened.
No Prompt : The AED does not play a startup prompt.
Otherwise choose one of these prompts:
Call 911 (Call Triple Zero for Australia)
Call emergency services
8 70-00974-02 E
Setting AED and rescue parameters
Time Settings: Set the clock
To set the AED's clock:
1.
2.
Select your time zone from the Time Zone/UTC Offset menu.
Click Synchronise Date/Time .
The AED’s time will be updated to the computer’s time when you click Save
Configuration .
Note: Times shown in AED Manager are not adjusted for Daylight Saving Time.
Shock Settings: Set rhythm detection thresholds
To set rhythm detection parameters, use the preferences in the SHOCK SETTINGS panel.
The AED can hold different protocols for adult and paediatric pads.
Preference
Same Energy After
Conversion
Description
When enabled, the AED delivers the previous defibrillation energy the next time a shock is required if the patient’s rhythm becomes non-shockable rhythm.
The number of shocks that can be delivered before the AED enters CPR mode.
Maximum Shocks Per
Sequence
Shock Energy Protocol Select the protocol for the energy delivered by a shock. If the AED has paediatric defibrillation pads attached, it uses the number in the Paediatric field.
VF/VT Rate Minimum heart rate at which the AED detects ventricular fibrillation and determines the rhythm is shockable. If the AED has paediatric defibrillation pads attached, it uses the number in the Paediatric field.
70-00974-02 E 9
Using AED Manager
Preference
SVT rate
Description
Adds discrimination for SVT (supraventricular tachycardia) to the shockable/ non-shockable decision. If the AED has paediatric defibrillation pads attached, it uses the number in the Paediatric field.
OFF (no monitoring for SVT):
•
•
•
•
•
Heart rates below the VF/VT rate are never shockable
Heart rates above the VF/VT rate are always shockable
No Therapy (monitoring for SVT, but therapy delivery depends on the following):
•
• Heart rates below the VF/VT rate are never shockable
Heart rates above the VF/VT rate are always classified as shockable or nonshockable based on their waveform morphologies
160-300 :
Heart rates below the VF/VT rate are never shockable
Heart rates above the set SVT rate are always shockable
Heart rates between the VF/VT rate and the set SVT rate are classified as shockable or non-shockable based on their waveform morphologies
10 70-00974-02 E
Setting AED and rescue parameters
CPR Settings: Set rescue parameters
The AED can provide different rescue protocols for adults and children (paediatric).
Preference
Session Type
CPR Timeout
CPR Metronome Prompt
CPR Metronome Rate
End session when
Number of Sets
Breaths per set
Compressions per set
Description
Traditional CPR (compressions and breaths) and Compressions only use the same settings. Compressions only uses CPR Timeout and the metronome settings only.
Length of time, in seconds, that each CPR session lasts. The countdown timer on the AED tracks the remaining time.
Marks the pace for the rescuer to match for chest compressions.
Press —Spoken prompt of “press” given at the selected CPR
Metronome rate.
Ping —AED provides a mechanical metronome sound at the CPR
Metronome rate.
No Metronome —No sound
The pace of the metronome prompt in beats per minute.
This preference determines the total time of a session.
Sets complete —The CPR session ends when the number of CPR sets expires. Each set includes the number of compressions per set and number of breaths per set.
Time expires —The session ends when the CPR timer reaches zero.
The final set of compressions and breaths might be only partially complete.
If using traditional CPR, the number of times the rescuer is prompted to perform the combination of compressions and breaths.
Number of breaths to give before beginning another round of compressions.
Number of chest compressions to give before providing breaths.
70-00974-02 E 11
Using AED Manager
CPR Device Settings: Set CPR Device thresholds
To set the thresholds for CPR prompting for rescues that use the CPR Device, modify the preferences in the CAD SETTINGS panel.
Preference
Compression rate
(compressions/minute)
Compression Depth (cm)
Description
Minimum : If the rescuer performs compressions under this rate, the
AED prompts, “Press faster”.
Maximum : If the rescuer performs compressions above this rate, the
AED prompts, “Press slower”.
Minimum : If the rescuer does not reach this compression depth, the
AED prompts, “Press harder and fully release”.
Maximum : If the rescuer presses to a depth greater than this, the AED prompts, “Press softer”.
12 70-00974-02 E
Reviewing a rescue
Reviewing a rescue
The Summary tab under the Rescues tab shows the time, number of defibrillation shocks given and the length of the rescue.
1
View the AED Snapshot tab to see all the settings with which the AED was programmed at the time of the rescue. See
Setting AED and rescue parameters on page 7
for descriptions of the settings.
70-00974-02 E 13
Using AED Manager
14
Viewing the ECG waveform and rescue events
Click the Waveform & Events tab under the Rescue Centre tab.
The event list shows all AED operations the AED performs during the rescue.
The graph below the event list shows a recording of the patient’s ECG and CPR
Device-assisted CPR with rescue events.
u Select an item in the event list.
The graph moves to the place in the rescue when the event occurred.
u u u
Drag and release the mouse over the waveform to scroll backward and forward through time.
Add notes to the waveform. Right-click on the chart where you want a note to appear. From the contextual menu, select Add User Annotation . In the Create/
Edit Event dialogue box, type the name and comment then click Save .
Use the player controls to replay the rescue in real time (one second of the rescue takes one second to play) and move forward and backward through the waveform and rescue.
Move back one page Select which items to view: show CPR graph, show actual time in the event list, show user added events, show event markers
Move back one second
Set the vertical scale of the
ECG graph
(5 - 25 mm/mV)
Play the recording
Stop the play back
Set the horizontal scale of the
ECG and CPR graphs
(10 - 40 mm/second)
Set the vertical scale of the
CPR graph
(1- 3 mm/g; g = m/sec
2
)
70-00974-02 E
70-00974-02 E
Move forward one second
Move forward one page
Print the rescue waveform
Reviewing a rescue
Click to open the waveform tab in a separate window.
You can maximise this window to view the recording full screen.
Drag the pointer up to zoom in or down to zoom out. Click a magnifying glass to zoom one step at a time.
Drag the position pointer left and right to view different parts of the rescue.
15
Using AED Manager
Viewing and updating incident information
Click the Rescues tab. Select the rescue to view. Click the Incident tab.
Use the Incident form to add demographic information about the patient and general information about the rescue.
The AED does not have to be connected to modify or save this information.
To complete the form:
1.
Type or select the data from the drop-down menus to update or add.
2.
When finished, click Save to Database .
Printing incident data and rescue waveforms
3.
4.
1.
2.
You can print the information displayed on the Incident and the Waveform tabs.
To print only the waveform: u Click the Print icon under the Rescues | Waveform & Events tab on the icon bar.
To print the waveform, rescue events, and incident information:
Right-click on a rescue option in the rescue list.
Select Print Rescue Report .
Select the information to print.
Click Print .
16 70-00974-02 E
Reviewing event messages
Reviewing event messages
You can view details about AED events and errors in the Event Log tab.
The AED records the times different actions happen, such as system self-tests and its lid opening or closing.
1) Event Log :
Lists all self-tests recorded by the
AED.
2) Current
Errors : Lists errors that are generated by the
AED.
1
2
70-00974-02 E 17
Using AED Manager
Administering your AEDs
With a connected AED, you can review self-tests, print its configuration, import and export AED data, and update its firmware.
Printing AED information
1.
2.
Select an AED from the AED list. It does not have to be connected.
Select File | Generate Report | Print AED Report .
Importing and exporting AED data
If you have a previously saved AED data file, you can load this file for viewing in AED
Manager.
To import an AED data file:
1.
2.
Select File | Import .
Choose a file to import.
The imported AED appears in the list of not-connected AEDs.
You can export the entire set of AED data to a compressed file format. This file can be used with other applications for further analysis.
1.
2.
To export AED data:
Select the AED to export from the connected or not connected list.
From the File menu, select Export | Selected AED Data to a Single File .
All the rescues, including the ECG waveform and recorded rescue events, are saved to a single compressed file.
18 70-00974-02 E
70-00974-02 E
Administering your AEDs
Configuring remote AEDs (Administrator level only)
You can update the configuration of AEDs at their locations instead of carrying them to the AED Manager computer. Using a single USB flash drive, you can configure your entire inventory of AEDs. Update each AED in-place by plugging the flash drive into an AED’s USB port. When you plug in the flash drive, the AED copies the configuration parameters that you selected.
To update remote AEDs with a new configuration:
1.
Select an AED from either the connected or the not-connected list. You can select any AED.
Note: You must select an AED because its configuration is needed to prepare the flash drive.
5.
6.
2.
3.
4.
7.
8.
Connect a USB flash drive to the computer.
Select File | Prepare USB Flash Drive .
In the USB Flash Drive preparation window that appears, select the flash drive.
CAUTION! Formatting erases all data from the flash drive.
Click Next to format (prepare) the flash drive.
In the configuration window, choose the parameters you want.
The panels in the configuration dialogue box are similar to those under the
Summary tab.
•
Review the configuration.
Click Save if your changes are complete.
After you click Save , AED Manager writes configuration files to the USB flash drive.
Remove the flash drive from the USB port.
19
!
Updating AED software (Administrator level only)
Periodically, there may be new software released for the AED.
To update the software:
5.
6.
3.
4.
1.
2.
Connect the AED to the computer. (Software cannot be updated with a flash drive).
Log in to AED Manager as Administrator.
The AED appears under Connected in the AED list.
Select the AED.
Insert the update CD in the computer.
Select File | Update AED Software .
Click Update .
Caution: Device made inoperable.
Keep the AED connected while the firmware is loading. The AED can become inoperative if it is disconnected before the load completes.
When the update finishes, you see the message Update complete . It is now safe to disconnect the AED.
Cardiac Science Corporation • N7 W22025 Johnson Drive, Waukesha, WI 53186 USA • 262.953.3500 • US toll-free 800.426.0337
• Fax: 262.953.3499 • [email protected]
Technical Support • (US) Fax: 262.798.5236 • [email protected]• (International) [email protected]