Golden Engineering XRS-3 is a portable x-ray generator that produces high-quality images with minimal radiation exposure. With its compact size and lightweight design, it's easy to transport and operate, making it ideal for on-site inspections and field applications. The XRS-3 can penetrate up to an inch of steel, making it suitable for a wide range of industrial and security tasks, including weld inspection, corrosion detection, and foreign object detection.
Advertisement
Advertisement
XRS-3
X-RAY SOURCE
OPERATOR’S MANUAL
CONTENTS
ITEM PAGE
1.0 INTRODUCTION................................................................................. 2
2.0 WARNINGS......................................................................................... 2
DUTY CYCLE……………………………………………………………. 2
3.0 PHYSICAL DESCRIPTION................................................................. 3
HIGH VOLTAGE PULSER/TUBEHEAD.................................
...…….. 3
BASE.................................................................................................. 3
BATTERY PACK……........................................................................ 3
BATTERY CHARGER........................................................................ 3
CONTROL MODULE…………………………..................................... 4
CABLE CONNE CTOR DIAGRAM………………………………………. 5
4.0 DESCRIPTION OF OPERATION........................................................ 6
BLOCK DIAGRAM………………………………………………………… 6
5.0 OPERATING INSTRUCTIONS............................................................ 7
OPERATING PRECAUTIONS.................................................
........…. 7
EXCLUSION ZONE………………………………………………………. 7
PULSE SELE
CTION …………………………………………………….. 7
REMOTE CABLE OPTION……………………………………………… 8
DELAY MODE OPTION…………………………………………............ 8
REAL TIME
IMAGING OPTION………………………………………… 8
PULSE SETTINGS............................................................................. 8
6.0 SOFTWARE........................................................................................ 9
7.0 MAINTENANCE.................................................................................. 9
DOSE MEASUREMENT…………………….…………………………. 9
TUBE RE PLACEMENT………………………………………………... 9
8.0 TROUBLE-SHOOTING....................................................................... 10
9.0 INSTRUCTIONS FOR REPAIR......................................................... 10
HANDL E REPLACEMENT…………………………………………… 10
FUSE REPLACEMENT………………………………………………. 11
REMOVING BOARDS………………………………………………… 11
BOARD INSTALL ATION……………………………………………… 11
HEAD REPLACEMENT………………………………………………. 12
BATTE
RY DISPOSAL………………………………………………… 12
10.0 WARRANTY...................................................................................... 13
RETURNING REPAIR INSTRUCTION S ……………………………. 13
11.0 SPECIFICATIONS............................................................................. 14
PHYSICAL DIMENSIONS................................................................ 14
X-RAY OUTPUT............................................................................... 14
ELECTRICAL & THERMAL CHARACTERISTICS …...................…. 14
12.0 SPAR
E PARTS………………………………………………………… 14
- 1 -
1.0 INTRODUCTION
The XRS-3 produces high levels of radiation and must be operated by qualified personnel who have read the Warning and Operations section of the manual before operating the device.
The XRS-3 is a small, lightweight x-ray generator that operates on its own removable battery pack. The
XRS-3 is a pulsed x-ray device that produces x-ray pulses of very short duration (50 nanoseconds). It produces a relatively low dose rate comparable to a 0.25 ma constant potential machine. The energy produced by the XRS-3 is up to 270KVP, which makes it possible to radiograph up to one (1) inch (2.54 cm) of steel.
XRS-3 standard accessories are two keys, two battery packs, and one battery charger. Remote cable, carrying case, and film developing equipment are also common accessories.
2.0 WARNINGS
The XRS-3 is a pulsed X-ray generator that emits hazardous ionizing radiation when pulsing. The XRS-3 should only be operated by authorized personnel who are properly trained to safely operate the generator. The XRS-3 must be registered with proper authorities prior to use and should not be used to intentionally expose humans.
Develop and closely follow a safe operating system for using the XRS-3. The safe operating system must ensure that no one is exposed to radiation above the permissible limits which are 2 mR (0.02 mSv) per hour for a member of the public. The safe operating system must ensure the XRS-3 is used within federal and state guidelines.
All operators and users of the XRS-3 x-ray machine must wear a personal radiation monitoring device, such as a TLD (thermoluminescent dosimeter), film badge, and/or a pocket dosimeter consistent with the appropriate federal, territorial or provincial standards (note: an electronic dosimeter will not detect the XRS-3 radiation pulses).
Due to the short pulse width of the XRS-3, survey meters of the Geiger-Mueller and scintillator type do not accurately detect the radiation emitted from the x-ray source.
Survey meters should be of the ionization type and should be used in the integration mode.
Survey meters must not be used in the rate mode because the XRS-3 does not produce constant radiation. The XRS-3 produces very high rates of radiation for very short periods of time resulting in either unrealistically high readings or no readings for a survey meter in rate mode.
The XRS-3 has no explosion proof rating and should not be used in an explosive atmosphere.
The Spark Gap is vented to the air and could be a source of ignition.
DUTY CYCLE WARNING.
The XRS-3 is a light duty machine that is not made to pulse continuously. The maximum duty cycle for the XRS-3 is 200 pulses every four minutes (3000 pulses per hour). Two consecutive pulse trains of 99 pulses can be fired then the unit should rest at least four minutes. Exceeding the duty cycle will shorten the life of the tube and head.
3.0 PHYSICAL DESCRIPTION
HEAD
COLLIMATOR
HANDLE
KEY
CONTROL MODULE
BEAM
ANGLE
LABEL
TRIPOD MOUNT
RELEASE BUTTON
RADIATION WARNING LABEL
BATTERY
BASE
BACK PLATE
Figure 1: XRS-3 X-ray Unit
HIGH VOLTAGE PULSER/TUBEHEAD.
The main body of the XRS-3 is the tube head which contains the tube cavity, cold cathode type X-ray tube, spark gap, high voltage capacitor, and transformer. The standard collimator located on the front of the head limits the X-ray beam to 40 degrees. Special order collimators up to 85 degrees are available.
BASE . The base of the XRS-3 contains the base plate, tripod mount, tripod mount release button, and identification label. The tripod mount contains threaded ¼-20 insert that can be attached to any standard camera tripod. Press the release button to remove the tripod mount from the base. A label identifying the model, manufacturer, and serial number is located on the bottom of the XRS-3 base.
TRIPOD MOUNT
Figure 2: Base
BATTERY PACK . The standard battery pack is a DeWalt 14.4V nickel-cadmium battery. Optional battery chemistry or voltage may be available.
BATTERY CHARGER : The standard battery charger is the DeWalt DW9116 110V charger or DE9108
220V charger. Battery charge time is one hour. See battery charger manual for additional instructions and warnings.
3
3.5 CONTROL MODULE
X-RAY PULSING LIGHT
LCD
DELAY SWITCH
RANGE SWITCH
BACK PLATE
KEY
EMERGENCY STOP \
PULSE DEFAULT
UNITS SWITCH
POWER ON LIGHT
CABLE CONNECTOR
Figure 3: Control Module
POWER ON LIGHT: Illuminates when battery voltage is applied to control module.
RED X-RAY PULSING LIGHT: Blinks after time delay button or remote cable button is pressed to warn that the XRS-3 is going to pulse. The light stays on continuously while the XRS-3 is pulsing. This is a failsafe warning light. If the light does not work the X-ray unit will not pulse .
LIQUID CRYSTAL DISPLAY (LCD): Two digit LCD displays selected pulse, time before unit pulses, software version, and total number of pulses on the unit.
RANGE SWITCH: Used to alternate LCD between tens digit and ones digit.
UNITS SWITCH: Used to change the value of the tens digit or units digit from 0 to 9. The UNITS
SWITCH is also used with the EMERGENCY STOP SWITCH to alter the default pulse setting.
DELAY SWITCH: Initiates delay mode.
EMERGENCY STOP SWITCH: Stops the unit before it begins pulsing or stops the unit in the middle of a pulse train. This switch can also be pressed with the UNITS SWITCH to alter the default pulse setting when the XRS-3 is first powered up.
CABLE CONNECTOR: Lemo connector located on the back of the control module beneath the battery receives the remote cable or imaging system cable. See Rear View Diagram on page 5 for details.
BACK PLATE: Covers the Oscillator board and contains battery terminal connectors.
4
XRS-3 REAR VIEW/CABLE CONNECTOR
PIN #
1
2
3
4
5
DESCRIPTION
+5 VOLTS 100 ma MAXIMUM
REMOTE SWITCH
REMOTE SWITCH – NO DELAY
X-RAY ON SIGNAL
COMMON 0 VOLTS
1
2 5
3 4
REMOTE CONNECTOR: LEMO EPG.0B.305.HLN
MATING CABLE PLUG: LEMO FGG.0B.305.CLAD 56Z
PIN 2
PIN 5
Remote switch inputs are activated when grounded.
5
4.0 DESCRIPTION OF OPERATION
The block diagram below illustrates how the XRS-3 functions. The following sequence of events takes place each time the XRS-3 is fired:
1. User initiates operation of the machine.
2. The control section sends a signal to the converter section to begin oscillating.
3. Once oscillating, the converter section changes the 14.4 volts DC to 22Khz AC.
4. The transformer charges the High Voltage Capacitor to about 9000 volts.
5. The spark gap arcs after the High Voltage Capacitor reaches proper voltage.
6. The pulse detector signals the control block that the unit has pulsed.
7. As the High Voltage Switch is closed, a high voltage transient of approximately 270,000 volts and
50 nanoseconds in duration is applied across the x-ray tube generating x-rays.
The closing of the High Voltage Switch produces an audible pulsing sound. The XRS-3 cannot produce x-rays without the pulsing sound so it serves as an additional warning the XRS-3 is functioning.
This unit generates x-rays through high voltage bombardment of a tungsten target. The XRS-3 does
not contain radioactive materials.
All the high voltage is contained within the aluminum canister and as long as the canister is not punctured the operator is not exposed to dangerous voltages.
BLOCK DIAGRAM
X-RAY
TUBE
SPARK
GAP
HIGH VOLTAGE
CAPACITOR
USER
INTERFACE
PULSE
DETECTOR
CONTROLLER
CONVERTER
BATTERY
6
5.0 OPERATING INSTRUCTIONS
OPERATING PRECAUTIONS: The operator should always stand at least 10 feet behind the X-ray unit and clear all personnel at least 10 feet behind the unit or at least 100 ft. from the front of the unit before pulsing. The exclusion zone (below) should be a controlled area free of all personnel while X-ray pulses.
20’
10’
X-ray Beam
100’
20’
Figure 4: Exclusion Zone
PULSE SELECTION
1. Attach a charged battery and turn on the X-ray generator.
2. Press the RANGE BUTTON. The one’s digit of the LCD will blink twice and then go blank.
3. Press the UNITS BUTTON to adjust the ones digit from 0 to 9.
4. Press the RANGE BUTTON again and the tens digit of the LCD will blink twice and go blank.
5. Press the UNITS BUTTON to change the tens digit from 0 to 9.
6. Press the RANGE or UNITS BUTTON again to accept the new pulse setting.
7. Both tens digit and units digit will blink to indicate acceptance of the new pulse setting. (If
RANGE or UNITS BUTTON is not selected after entering the pulse count the generator will automatically accept the entered pulse setting after six seconds).
8. Press the UNITS BUTTON and EMERGENCY STOP BUTTON simultaneously to lock in pulse setting. XRS-3 will retain locked in pulse setting until it is changed.
Select ones digit Enter ones value (0-9) Select tens digit
Enter tens value (0-9) Change pulse default (optional)
7
DELAY MODE OPTION
1. Press Green DELAY BUTTON on top of the control module.
2. LCD displays 60 and X-ray begins 60 second time delay.
3. Hold DELAY BUTTON down for 1.5 seconds and time delay goes from 60 to 15 seconds.
4. X-ray makes audible beep and red pulsing light blinks as generator counts down to pulse.
5. Audible warning and red warning light stay on while X-ray generator pulses.
REMOTE CABLE OPTION
1. Attach remote cable to back of control module.
2. Retreat distance of the cable behind the unit.
3. Press and hold down button on the end of the remote cable until generator completes pulse train.
4. There is 5 second safety delay before generator begins pulsing. Audible beep and red pulsing light blink.
5. Audible warning and red pulsing light stay on while X-ray generator pulses.
REAL TIME IMAGING OPTION
1. Change pulse setting default to 99 pulses if it is not already set to 99 pulses.
2. Attach imager cable to back of the control module unless using wireless option.
3. Refer to imaging system operating instructions for details on pulse setting and pulsing the generator.
4. There is no delay in this mode so the X-ray generator pulses immediately unless there is a time delay on the imaging system .
SUGGESTED PULSE SETTINGS
The chart below lists approximate pulses necessary to penetrate various materials. Settings vary greatly depending on imaging system used. Refer to imaging system instructions for more information.
MATERIAL PULSE SETTING
24 inches between x-ray & imager
CARDBOARD / LIGHT WOOD / PLASTIC
LIGHT METAL
STEEL 1/4”
STEEL ½”
STEEL 1”
BRASS 1/8” to ¼”
2-5
10
25
50
99
50- 99
The following is true when using film or digital systems that generate a positive image. If the radiograph is too dark, the film is underexposed. If the radiograph is too light the film is overexposed.
Underexposure can be corrected by increasing the number of pulses and/or decreasing the distance between the imaging medium and the XRS-3. Overexposure can be corrected by reducing the number of pulses and/or increasing the distance between the imaging medium and XRS-3.
Underexposed Overexposed Correct exposure (pulse setting)
8
6.0 SOFTWARE
The software program that controls the microcontroller can be identified by turning the key switch on while both push button switches (RANGE & UNITS) below the LCD are depressed. The LCD displays the software version “65”. After “65” is displayed the total number of pulses on the XRS-3 will be displayed in the LCD. Each digit represents 10,000 pulses. Example: If the LCD reads “04” the total number of pulses on the XRS-3 is between 40,000 and 50,000 pulses. After the total number of pulses is displayed the LCD will read “00” or the default pulse setting that was last stored on the unit.
The software program is capable of determining the state of battery charge based on the time between each pulse. As the battery loses charge the XRS-3 pulse rate slows, with more time between each pulse. If there is more than 0.33 seconds between two consecutive pulses the following will occur:
➢ The XRS-3 continues the current pulse train to “00”.
➢ After the XRS-3 stops pulsing. The LCD goes back to the original pulse setting. The left and right digits blink alternately.
➢ The condition indicates a low battery.
➢ The XRS-3 will be inoperable until the key switch is turned off and on, or the battery is replaced.
If there is more than one second between two consecutive pulses.
➢ The XRS-3 stops pulsing immediately and the LCD displays 00.
➢ This function prevents XRS-3 from pulsing continuously if there is a failure in detecting circuitry.
➢ This condition may indicate a low battery, electrical noise, or failure in detecting circuitry.
➢ The operator may need to replace the battery pack, turn key switch off and on, or send the XRS-3 back for repair.
7.0 MAINTENANCE
X-RAY DOSE MEASUREMENT Using a dosimeter, the average X-ray dose for new tube can be established.
➢ With the dosimeter located 1 foot from the front of the case and in line with the center of the beam angle label, the reading for 10 pulses should be 26 mR to 36 mR.
➢ The leakage sheet illustrates the X-ray dose and maximum allowable radiation leakage levels for each X-ray unit. A completed copy of this form accompanies each X-ray.
TUBE REPLACEMENT If you have a tube replacement kit refer to instructional disk included with the kit. If you do not have a kit the unit must be sent back to Golden Engineering or an Authorized
Distributor for tube replacement.
Tube life is approximately 100,000 pulses. Under normal conditions the tube’s output will decrease slowly with use. If the tube is broken or the glass cracks the tube output will cease immediately.
9
8.0 TROUBLESHOOTING
SYMPTOM
No “power on” light
Power on lights, but X-ray does not pulse.
Power on lights, X-ray pulsing light does not illuminate, X-ray dose not pulse
X-ray pulses, but no image or black image.
Unit stops pulsing in the middle of a pulse train and
LCD displays 00.
TEST
-Check battery voltage
-Check battery connection
-Check the battery voltage.
-Check the 2amp fuse.
-Test X-ray output.
ACTION
- Replace or charge battery
- Make sure battery is securely
attached and battery clips are
not bent or broken.
-Charge or replace the battery.
-Replace the fuse if necessary.
- Replace the X-ray pulsing light
-Replace the tube.
Unit makes loud popping noise while pulsing.
Oil leaking from unit.
-Check the battery voltage.
-Check 15 amp fuse.
-Check feedback line
connection.
- Charge battery if necessary.
- Replace the fuse if blown.
- Make sure the screw holding
wire to the oscillator board is tight.
- Stop immediately and return
for repair .
Return for repair.
9.0 INSTRUCTIONS FOR REPAIR
HANDLE REPLACEMENT Requires T-10 torx driver.
The Handle has two sides that are screwed together. Three screws in the handle hold it together and one screw in the base holds the handle strap together under the head.
1. Remove the 3 screws on the side of the handle.
2. Remove the base plate on the bottom of the unit.
3. Remove the screw on the bottom of the unit that holds the handle strap together.
SCREWS
Figure 5: Handle Figure 6: Handle strap base
4. Remove the half of the handle that is broken.
5. Place a small amount of glue on the protrusion from the control module.
6. Install new half of the handle.
7. Insert and tighten 3 screws in the handle and one screw in the handle strap. Do not over tighten the screw in the handle strap (bottom) as this will crack the strap.
8. Attach the base plate and insert all the screws that hold it in place.
10
FUSE REPLACEMENT Requires T-10 Torx driver & needle nose pliers.
1. Remove the back plate. Remove the 5 screws in the back plate then pull the back plate off slowly maneuvering the battery terminal connecting wires through the opening in the oscillator board.
2. The 15 amp fuse is the white one inch long fuse on the left side of the oscillator board. The 2 amp fuse is a small green fuse just to the right of the 15 amp fuse. See diagram on page 12.
The 15 amp fuse can be removed with fingers. The 2 amp fuse may require needle nose pliers to pull it out of the board. It should be pulled in a downward direction to remove from the board.
REMOVING THE BOARDS Refer to the diagram on page 12 for steps 1-4.
1. Remove the back plate (see FUSE REPLACEMENT step 1). After the terminals are through the oscillator board, disconnect the red and black battery wires.
2. There are three terminals on the lower left side of the oscillator board and one on the lower right.
Disconnect the two blue wires, one red wire, and one green signal wire using Philips head screw driver.
3. There are three socket head cap screws holding the oscillator board in place. Two are in the middle of the board and one is at the bottom. Remove these three screws. Be aware of the ½ inch stand offs behind the oscillator board.
4. Remove the flat head screw in the upper left side of the oscillator board between the two fuses.
5. Pull the board out.
6. Disconnect the oscillator board from the counter board by pulling the bottom of the oscillator board up and away from the counter board. Disconnect the two pin white key switch connector, white three pin remote connector and black touch pad connector.
7. Unscrew the LED bulb cover on the X-ray on LED.
8. Tilt the back of the counter board down until the LED clears the housing then pull the board out of the housing.
BOARD INSTALLATION
1. Slide counter board back into the top of the Control Module housing just below the screw receptacles.
2. Push the counter board up so LED goes through the appropriate hole in the control module housing.
3. Connect the three white and one black connector.
4. Screw the LED cover back on.
5. Put the t hree cap screws through the oscillator board and then put the ½” offsets on the back of the screws.
6. Plug the oscillator board back into the counter board and push it in position.
7. Tighten the three screws holding the board in place.
8. Insert the flat head screw in upper left corner of the oscillator board.
9. Attach the two blue wires, one red wire, and green signal line.
10. Reinstall the back plate.
11
STEP 4
2 AMP FUSE
STEP 1
15 AMP
FUSE POWER
TRANSISTORS
STEP 2
STEP 2
SIGNAL WIRE
X-RAY ON /
FEEDBACK
SWITCH
Figure 7:Oscillator Board
STEP 3
HEAD REPLACEMENT Requires a small flat head screw driver.
1. Remove the boards as instructed above.
2. Remove the two screws on the bottom that attach the head to the control module.
3. Remove the screw on the upper left side of the head. Must rotate head slightly and put the screw driver through the left LED hole to remove the screw.
4. Pull the head straight out of the control module. You might need to loosen the screw in the handle to get the head out.
12
INSTRUCTIONS FOR BATTERY DISPOSAL Follow all federal, state, and local laws for disposal of nickel-cadmium batteries. Batteries may be returned to Golden Engineering for disposal.
10.0 WARRANTY
Golden Engineering, Inc. warrants XRS-3 X-ray unit made and sold by it or its authorized representatives to be free of defects in materials and workmanship for a period of twelve (12) months from the date of shipment to the end user. Warranty does not cover maintenance required due to life . To make a claim under this limited warranty, customer must ship the entire unit (or the component believed to be defective) to Golden Engineering, post-paid. Golden Engineering, Inc. assumes no liability for units or components shipped until they are actually in the custody of Golden Engineering, Inc. Provided Golden
Engineering, Inc. in its sole discretion, is satisfied that the failure is not the result of excessive use, abuse, misuse, accident, modification or improper disassembly or repair, Golden Engineering will provide parts and labor required to repair the unit. Golden Engineering reserves the right to use reconditioned and remanufactured components that meet original specifications. The unit or component will be return shipped to customer at customer's expense. THIS EXPRESS LIMITED WARRANTY IS IN LIEU OF ALL
OTHER WARRANTIES AND GUARANTEES, EITHER EXPRESS OR IMPLIED OR CREATED BY
OPERATION OF LAW.
THE XRS-3 X-Ray Source is manufactured by:
GOLDEN ENGINEERING, INC.
PO BOX 185
CENTERVILLE, IN 47330 USA
Phone: 1-765/855-3493
Fax: 1-765/855-3492
WEB: www.goldenengineering.com
RETURNING UNIT FOR REPAIR
➢ Complete the repair form at www.goldenengineering.com/technical.html
and include a copy of the printed form with the repair. If you do not have internet access prior to sending repair then include a letter containing a brief description of the problem, contact name, phone number, and return address.
➢ Remove battery before shipping the unit.
➢ Accessories are not necessary with units shipped back for repair.
➢ Be sure the unit is securely packaged for shipment and wrap in plastic bag if there is an oil leak.
➢ Ship to address:
Golden Engineering
6364 Means Road,
Centerville, IN 47330 USA
Phone: 1-765-855-3493
EMAIL: [email protected]
Serial Number:______________
Delivery Date:_______________
13
11.0 SPECIFICATIONS
LENGTH
WIDTH
HEIGHT
WEIGHT
PHYSICAL DIMENSIONS INCLUDING BATTERY PACK
16 inches (40.6 cm)
4.5 inches (11.5 cm)
7.5 inches (10 cm)
12.5 pounds (5.7 Kg) with battery
X-RAY OUTPUT
2.6 mR to 3.6 mR (12 inches in front of unit) X-ray dose per pulse
Pulses per battery charge
Pulses per second
Expected tube life (glass tube)
X-ray source size
Maximum Photon Energy
X-ray pulse width
4000
15 (Nominal)
100,000 pulses
1/8 in. (3mm)
270 KVP
20 nanoseconds
Battery voltage
Battery type
Battery recharge time
Current draw
Current flow
Temperature range
Maximum duty cycle
Warm-up
ELECTRICAL AND THERMAL CHARACTERISTICS
14.4 volts
Nickel Cadmium sub C cells
One Hour
17 amps @ 14.4 volt battery
0.25 mA
-10 to 120 degrees F (-23 to 50 degrees C)
200 pulses every 4 minutes (3000 pulses per hour)
None required
12.0 SPARE PARTS AND ACCESSORIES FOR THE XRS-3
Thumbwheel Key
Flat key
DeWalt Battery 14.4v DC 9091
ITEM
DeWalt Battery 18v PLUS 9096
DeWalt Battery Charger (110V) DW9116
DeWalt Battery Charger (220V) DE9108
Remote Cable
Tripod Mount
Handle – Left Side
Handle – Right Side
Carrying case (holds X-ray, 2 batteries, charger, cable)
PART NUMBER
2002000
5951020
1800102
1800103
1800156
1800158
1809013
4000350
4000210
4000200
1701655
14
15