K-365 Dist Line
Operation Manual
Imprint
Product Identification:
Operation Manual (Original) K-365 Dist Line
11594193
Publication date: 11.2021
Version B
BÜCHI Labortechnik AG
Meierseggstrasse 40
Postfach
CH-9230 Flawil 1
E-Mail: [email protected]
BUCHI reserves the right to make changes to the manual as deemed necessary in the light of experience, especially with respect to structure, illustrations and technical details.
This manual is copyrighted. Information from it may neither be reproduced, distributed, or used for competitive purposes, nor made available to third parties. The manufacture of any component with the aid of this manual without prior written agreement is also prohibited.
Büchi Labortechnik AG Table of contents
Table of contents
Operation Manual K-365 Dist Line iii
Table of contents Büchi Labortechnik AG
Installing the recirculating chiller on the condenser (option)................................... 30
iv Operation Manual K-365 Dist Line
Büchi Labortechnik AG Table of contents
Selecting the determination type for a single determination................................... 53
Changing the sample quantity for single determination.......................................... 54
Selecting the method for a determination within a series ....................................... 58
Changing the sample quantity for determination within a series ............................ 58
Changing the amount unit for a determination within a series................................ 58
Changing the reaction detection for a method (option for MultiDist only)............... 61
Changing the H ₂ O volume for a method (MultiDist only) ........................................ 62
Changing the H ₂ SO ₄ volume for a method (MultiDist only) .................................... 62
Changing the NaOH volume for a method (MultiDist only)..................................... 63
Changing the acid volume for a method (BasicDist, MultiDist only) ....................... 63
Changing the reaction time for a method (BasicDist, MultiDist only)...................... 64
8.6.13 Changing the level detection sensor settings (option) for a method....................... 66
8.6.15 Changing the distillation stirring speed for a method (MultiDist T only).................. 67
8.6.18 Changing the dosing unit volume for a method (MultiDist T only) .......................... 69
8.6.21 Changing the aspiration time for the sample tube for a method (BasicDist, MultiDist,
8.6.22 Changing the aspiration time for the receiving vessel for a method (MultiDist T only)
Operation Manual K-365 Dist Line v
Table of contents Büchi Labortechnik AG
vi Operation Manual K-365 Dist Line
Büchi Labortechnik AG Table of contents
Operation Manual K-365 Dist Line vii
Table of contents Büchi Labortechnik AG
viii Operation Manual K-365 Dist Line
Büchi Labortechnik AG
1
1.1
1.2
1.3
About this document | 1
About this document
This operation manual is applicable for all variants of the instrument.
Read this operation manual before operating the instrument and follow the instructions to ensure safe and trouble-free operation.
Keep this operation manual for later use and pass it on to any subsequent user or owner.
BÜCHI Labortechnik AG accepts no liability for damage, faults and malfunctions resulting from not following this operation manual.
If you have any questions after reading this operation manual: u Contact BÜCHI Labortechnik AG Customer Service.
https://www.buchi.com/contact
Mark-ups and symbols
NOTE
This symbol draws attention to useful and important information.
R This character draws attention to a requirement that must be met before the instructions below are carried out.
u This character indicates an instruction that must be carried out by the user.
ð This character indicates the result of a correctly carried out instruction.
Mark-up
Window
Explanation
Software Windows are marked-up like this.
Tab
Tabs are marked-up like this.
Dialogs are marked-up like this.
Dialog
[Button]
[Field names]
[Menu / Menu item]
Status
Signal
Buttons are marked-up like this.
Field names are marked-up like this.
Menus or menu items are marked-up like this.
Status is marked-up like this.
Signals are marked-up like this.
Trademarks
Product names and registered or unregistered trademarks that are used in this document are used only for identification and remain the property of the owner in each case.
Connected devices
In addition to these operating instructions, follow the instructions and specifications in the documentation for the connected devices.
Operation Manual K-365 Dist Line 9/106
2 | Safety
2
2.1
2.2
2.3
Büchi Labortechnik AG
Safety
Proper use
The instrument is designed for steam distillation.
The instrument can be used in laboratories for the following tasks:
Distillation of steam-volatile substances.
Use other than that intended
The use of the instrument other than described in proper use and specified in technical data is use other than that intended.
The operator is responsible for damages or hazards that are caused by use other than that intended.
Specially the following uses are not permitted:
Use of the instrument in areas which require explosion-safe instruments.
Use of samples, which can explode or inflame (example: explosives, etc.) due to shock, friction, heat or spark formation.
Use of the instrument with other than original BUCHI glassware.
Staff qualification
Unqualified persons are unable to identify risks and are therefore exposed to greater dangers.
The device may only be operated by suitably qualified laboratory staff.
These operating instructions are aimed at the following target groups:
Users
Users are persons that meet the following criteria:
They have been instructed in the use of the device.
They are familiar with the contents of these operating instructions and the applicable safety regulations and apply them.
They are able on the basis of their training or professional experience to assess the risks associated with the use of the device.
Operator
The operator (generally the laboratory manager) is responsible for the following aspects:
The device must be correctly installed, commissioned, operated and serviced.
Only suitably qualified staff may be assigned the task of performing the operations described in these operating instructions.
The staff must comply with the local applicable requirements and regulations for safe and hazard-conscious working practices.
Safety-related incidents that occur while using the device should be reported to the manufacturer ([email protected]).
10/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG
2.4
2.5
2.6
Safety | 2
BUCHI service technicians
Service technicians authorized by BUCHI have attended special training courses and are authorized by BÜCHI Labortechnik AG to carry out special servicing and repair measures.
Personal protective equipment
Depending on the application, hazards due to heat and/or corrosive chemicals may arise.
u
Always wear appropriate personal protective equipment such as safety goggles, protective clothing and gloves.
u Make sure that the personal protective equipment meets the requirements of the safety data sheets for all chemicals used.
Warning notices in this document
Warning notices warn you of dangers that can occur when handling the device.
There are four danger levels, each identifiable by the signal word used.
Signal word
DANGER
WARNING
CAUTION
NOTICE
Meaning
Indicates a danger with a high level of risk which could result in death or serious injury if not prevented.
Indicates a danger with a medium level of risk which could result in death or serious injury if not prevented.
Indicates a danger with a low level of risk which could result in minor or medium-severity injury if not prevented.
Indicates a danger that could result in damage to property.
Warning and directive symbols
The following warning and directive symbols are displayed in this operation manual or on the instrument.
Symbol Meaning
Hot surface
Corrosive
General warning
Device damage
Dangerous electrical voltage
Operation Manual K-365 Dist Line 11/106
2 | Safety
Symbol Meaning
Wear protective gloves
Wear safety goggles
Büchi Labortechnik AG
2.7
Fig. 1: Location of the warning and directive symbols
Protection devices
1
3
12/106
2.8
2
Fig. 2: Protection devices
1 Protection shield
3 Stop button
2 Protection shield
Residual risks
The device has been developed and manufactured using the latest technological advances. Nevertheless, risks to persons, property or the environment can arise if the device is used incorrectly.
Appropriate warnings in this manual serve to alert the user to these residual dangers.
Operation Manual K-365 Dist Line
Büchi Labortechnik AG Safety | 2
2.8.1
Hot surfaces
The surfaces of the device can become very hot. If touched they can cause skin burns.
u Do not touch hot surfaces or else wear suitable protective gloves.
2.8.2
Faults during operation
If a device is damaged, sharp edges, moving parts or exposed electrical wires can cause injuries.
u
Regularly check device for visible damage.
u u
If faults occur, switch off the device immediately, unplug the power cord and inform the operator.
Do not continue to use devices that are damaged.
2.8.3
Glass breakage
Broken glass can cause severe cuts.
Minor damage to the ground joints impairs the sealing effect and may therefore diminish suction capacity.
Handle the glass components carefully and do not drop them.
Always place the glassware in a suitable holder when they are not in use.
Always visually inspect glass components for damage every time they are to be used.
Do not continue to use glass components that are damaged.
Always wear protective gloves when disposing of broken glass.
2.8.4
Not suitable titrators
Not suitable titrators can lead to malfunction and wrong results.
u Use recommended titrators only.
2.8.5
Dosing pump defect
A defect dosing pump can spill hazardous liquids on the installation site.
u
Set the On / Off master switch to Off.
u u u
Disconnect the power supply.
Use personal protective equipment to remove the liquid.
Dispose of the remains of the liquid according to the local legal regulations.
2.9
Modifications
Unauthorized modifications can effect safety and lead to accidents.
u u u
Use only genuine BUCHI accessories, spare parts and consumables.
Carry out technical changes only with prior written approval from BUCHI.
Only allow changes to be made by BUCHI service technicians.
BUCHI accepts no liability for damage, faults and malfunctions resulting from unauthorized modifications.
Operation Manual K-365 Dist Line 13/106
3 | Product description
3
3.1
Product description
Description of function
The instrument is suitable for determining steamvolatile substances (e.g. of alcohol, sulfur dioxide, volatile acids) and nitrogen by using Kjeldahl and
Devarda methods (MultiDist).
Steam is introduced into the sample solution to drive out volatile components.
After condensation in condenser the condensate is collected in a receiver solution.
Büchi Labortechnik AG
3.2
Configuration
3.2.1
Front view
NOTE
The interface differs according to the configuration: u
Interface see Chapter 6 "Description of the interface", page 39
ð EasyDist, BasicDist u
Interface Pro see Chapter 7 "Description of the interface pro", page 45
ð BasicDist (option), MultiDist, MultiDist T
14/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG Product description | 3
1 2
9
8
7
3
4
5
6
Fig. 3: Front view
1 Splash protector 2 Condenser
3
5
Ventilation slots
Ventilation slots
7 Sample tube
9 Interface
(according to the instrument configuration)
4 On/Off master switch
6 Receiving area
See Chapter 3.2.3 "Receiving area", page 16
8 Handle
Operation Manual K-365 Dist Line 15/106
3 | Product description
3.2.2
Rear view
1 2 3
Büchi Labortechnik AG
14
13
12
11
10
4
16/106
9 Cooling water in
11 Rear cable duct
13 Fuses
(resettable)
9 8 7 6
Fig. 4: Rear view
1 Communication connections
See Chapter 3.2.4 "Communication connections", page 18
3 Ventilation slots
5 Connections on the rear side
See Chapter 3.2.5 "Connections on the rear side", page 20
7 Cooling water out
5
2 Ventilation slots
4 Dosing pumps window
6 Steam valve
8 Drain connection
(according to the instrument configuration)
10 Ventilation slots
12 Power supply connection
14 Type plate
See Chapter 3.4 "Type plate", page 23
3.2.3
Receiving area
The receiving area is different for each instrument configuration.
Operation Manual K-365 Dist Line
Büchi Labortechnik AG
Receiving area
Product description | 3
1
2
3
4
Fig. 5: Receiving area
1 Rear cable duct
3 Condenser outlet
Receiving area (MultiDist T only)
1 2
2 Cable duct reaction detection sensor
4 Receiving vessel area
3
11
10
9
6
7
4
5
8
Fig. 6: Receiving area
1 Cable duct reaction detection sensor 2 Condenser outlet tubing
3 Rear cable duct 4 Tubing for boric acid aspiration
5 Boric acid dosing 6 Retainer stirrer
7
9
Retainer dosing tip titration
Retainer titration sensor
11 Retainer dosing tip back titration
8 Receiving vessel
10 Retainer temperature sensor
Operation Manual K-365 Dist Line 17/106
3 | Product description
3.2.4
Communication connections
Communication connection EasyDist
1
2
Büchi Labortechnik AG
7
3
4
6 5
Fig. 7: Communication connections
1 BUCHI standard communication port
(COM)
(marked COM)
2 BUCHI standard communication port
(COM)
(marked COM)
3 Level detection sensor port
(marked Level Detection)
5 Communication port
(marked RS-232)
7 NaOH tank level sensor port
(marked NaOH)
4
6
H₂O tank level sensor port
(marked H₂O)
Communication port
(marked RS-232)
18/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG
Communication connection BasicDist
1 2
Product description | 3
3
11
10
9
4
5
6
8 7
Fig. 8: Communication connection
1 LAN port
(marked LAN)
2 BUCHI standard communication port
(COM)
(marked COM)
3 BUCHI standard communication port
(COM)
(marked COM)
4 Level detection sensor port
(marked Level Detection)
5
7
9
H₂O tank level sensor port
(marked H₂O)
Communication port
(marked RS-232)
Acid tank level sensor port
(marked Acid)
11 USB port
(marked USB)
6 Waste tank level sensor port
(marked Sample Tube Waste)
8 Communication port
(marked RS-232)
10 NaOH tank level sensor port
(marked NaOH)
Operation Manual K-365 Dist Line 19/106
3 | Product description
Communication connection MultiDist
1 2
15
14
13
12
6
7
8
9
3
4
5
Büchi Labortechnik AG
20/106
11 10
Fig. 9: Communication connections
1 LAN port
(marked LAN)
3 BUCHI standard communication port
(COM)
(marked COM)
5 Stirrer port
(marked Stirrer)
2 BUCHI standard communication port
(COM)
(marked COM)
4 BUCHI standard communication port
(COM)
(marked COM)
6 Level detection sensor port
(marked Level Detection)
7
9
H₂O tank level sensor port
(marked H₂O)
Waste tank level sensor port
(marked Sample Tube Waste)
11 Titrator port
(marked Titrator)
13 Waste tank level sensor port
(marked Receiver Waste)
15 USB port
(marked USB)
8 H₃BO₃ tank level sensor port
(marked H₃BO₃)
10 Dispenser port
(marked Dosing Unit)
12 Acid tank level sensor port
(marked Acid)
14 NaOH tank level sensor port
(marked NaOH)
3.2.5
Connections on the rear side
The connections on the rear side are depending on the version of the instrument.
Operation Manual K-365 Dist Line
Büchi Labortechnik AG Product description | 3
Connections on the rear side EasyDist
The connections are located at the rear side of the instrument. See Chapter 3.2.2
1
Fig. 10: Connections on the rear side
1 H₂O supply for steam generation
Connections on the rear side BasicDist (option)
The connections are located at the rear side of the instrument. See Chapter 3.2.2
1
2
3
4
5
Fig. 11: Connections on the rear side
1 Reserve
3 Reserve
5 Acid supply for sample tube
2 H₂O supply for steam generation
4 Reserve
Connections on the rear side BasicDist (option)
The connections are located at the rear side of the instrument. See Chapter 3.2.2
Operation Manual K-365 Dist Line 21/106
3 | Product description Büchi Labortechnik AG
1
4
5
2
3
6
Fig. 12: Connections on the rear side
1 Waste pump for sample tube
3 H₂O supply for steam generation
5 Reserve
2 Reserve
4 Reserve
6 Acid supply for sample tube
Connections on the rear side MultiDist
The connections are located at the rear side of the instrument. See Chapter 3.2.2
1
2
3
4
5
6
Fig. 13: Connections on the rear side
1 Waste pump for sample tube
3 H₂O supply for steam generation
5 NaOH supply for sample tube
2 H₂O supply for sample tube
4 Boric acid supply
6 Acid supply for sample tube
22/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG Product description | 3
Connections on the rear side MultiDist T
The connections are located at the rear side of the instrument. See Chapter 3.2.2
1 2
3
4
5
6
3.3
3.4
7
Fig. 14: Connections on the rear side
1 Waste pump receiving vessel
3 H₂O supply for sample tube
5 Boric acid supply
7 Acid supply for sample tube
2 Waste pump for sample tube
4 H₂O supply for steam generation
6 NaOH supply for sample tube
Scope of delivery
NOTE
The scope of delivery depends of the configuration of the purchase order.
Accessories are delivered as per the purchase order, order confirmation, and delivery note.
Type plate
The type plate identifies the instrument. The type plate is located at the rear of the instrument.
Operation Manual K-365 Dist Line 23/106
3 | Product description Büchi Labortechnik AG
1 2 3
11
10
9
8
7
6
5 4
Fig. 15: Type plate
1 Symbol for "electronics recycling" 2 Initial product code
3 Approvals
5
7
9
Year of manufacture
Frequency
Serial number
11 Company name and address
4 Symbol for "Do not dispose of as household waste"
6 Power consumption maximum
8 Input voltage range
10 Instrument name
3.5
Technical data
3.5.1
K-365 Dist Line
EasyDist BasicDist MultiDist
Dimensions (W x D x H)
Weight
Connection voltage
Power consumption
Frequency
IP Code
Pollution degree
Overvoltage category
Tap water pressure
Tap water condition
(if the temperature is more than
25°C use a recirculating chiller)
Tap water consumption
Approvals
~ 1.2 L/min
CE, CSA
320 x 400 x 730 mm
23 kg
320 x 400 x 730 mm
23 kg
220 - 240 ± 10
% VAC
2100 W
50 / 60 Hz
IP20
2
II
1 - 10 bar
1- 25°C
220 - 240 ± 10
% VAC
2100 W
50 / 60 Hz
IP20
2
II
1 - 10 bar
1 - 25°C
320 x 400 x 730 mm
23 kg
220 - 240 ± 10
% VAC
2100 W
50 / 60 Hz
IP20
2
II
1 - 10 bar
1 - 25°C
~ 1.2 L/min
CE, CSA
~ 1.2 L/min
CE, CSA
24/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG Product description | 3
Analysis specifications
EasyDist BasicDist MultiDist
Measuring range
Analysis time for 30 mg N
(with Metrohm Eco Titrator without titrator preparation)
analyte specific analyte specific 0.02 – 220 mg N
3.5 min
Analysis time for 200 mg N
(with Metrohm Eco Titrator without titrator preparation)
-
Recovery
5 min analyte specific analyte specific N: >99.5% (direct distillation) analyte specific analyte specific N: <0.8% Measurement reproducibility
(Direct distillation of 1 mg N absolute and 300 sec. distillation )
Instrument specifications
EasyDist BasicDist MultiDist
Distillation capacity (100%)
Distillation capacity (10%)
Adjustable distillation power
~ 40 mL/min ~ 40 mL/min ~ 40 mL/min
~ 12.5 mL/min ~ 12.5 mL/min ~ 12.5 mL/min
10 - 100%
MaxAccuracy Mode
(Compensation of power grid fluctuations)
Yes
Distillate reproducibility (RSD)
(At 300 seconds distillation time)
<1%
AutoDist Mode
(Automatic detection of the condensation start)
No
10 - 100%
Yes
<1%
Yes
10 - 100%
Yes
<1%
Yes
NaOH pump
H
2
O pump
Acid pump
No
No
No
No
No
Yes
Yes
Yes
Yes
Boric acid pump
Sample disposal
Receiver disposal
Safety sensors
Titration vessel
Titrator connection
No
No
No
Yes
No
No
No
Optional
No
Yes
No
No
Yes
Yes
Yes
Yes
Optional
Optional
Operation Manual K-365 Dist Line 25/106
3 | Product description Büchi Labortechnik AG
Interface specifications
Display
Control elements
Max. number of methods
Max. number of determinations
Max. number of determinations per series
Interface
EasyDist, BasicDist
LCD color, 4.3 in
Buttons / control knob
8
36
16
Max. number of templates
Max. number of users
4
4
Max. number of determination results 40
Language en, de, fr, it, es, pt, ja, zh, ru, pl, ko, id
Interface pro
BasicDist (option),
MultiDist
LCD color, 7.0 in
Touch-screen / control knob
96
512
32
16
32
256 en, de, fr, it, es, pt, ja, zh, ru, pl, ko, id
3.5.2
Ambient conditions
For indoor use only.
Max. altitude above sea level
Ambient and storage temperature
Maximum relative humidity
2000 m
5‒40 °C
80% for temperatures up to 31 °C decreasing linearly to 50 % relative humidity at 40 °C
3.5.3
Materials
Component
Housing
Housing electric
Steam generator housing
Housing stirrer
Rear cover
Glass parts
Plastic splash protector
Protective cover
Sealing
NaOH hose
Three-way valve
Materials of construction
Polyurethane
Stainless steel
Stainless steel
Glass fiber reinforced polypropylene
Polyvinyl chloride
Borosilicate 3.3
Glass fiber reinforced polypropylene
PMMA
CSM
EPDM
PP / PE
26/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG Product description | 3
3.5.4
Installation site
The installation site meets the safety requirements. See Chapter 2 "Safety", page 10
The installation site has a firm, level and nonslip surface.
The installation site has no obstacles (e.g. water taps, drains, etc.).
The installation site has enough space that the canisters can be placed safely.
The installation site has an own mains outlet socket for the instrument.
The installation site allows that the power supply can be disconnected at any time in an emergency.
The installation site is not exposed to external thermal loads, such as direct solar radiation.
The installation site has enough space that cables / tubes can be routed safely.
The installation site meets the requirements for the connected devices. See related documentation
The installation site meets the specifications according to the technical data (e.g.
weight, dimension, etc.). See Chapter 3.5 "Technical data", page 24
Operation Manual K-365 Dist Line 27/106
4 | Transport and storage
4
4.1
Büchi Labortechnik AG
Transport and storage
Transport
NOTICE
Risk of breakage due to incorrect transportation
Make sure that the instrument is fully dismantled.
Pack every instrument components properly to prevent breakage. Use the original packaging whenever possible.
Avoid sharp movements during transit.
4.2
4.3
u u u
After transporting, check the instrument and all glass components for damage.
Damage that has occurred in transit should be reported to the carrier.
Keep packaging for future transportation.
Storage
u u u
Make sure that the ambient conditions are complied with (see Chapter 3.5
Wherever possible, store the device in its original packaging.
After storage, check the device, all glass components, seals and tubing for damage and replace if necessary.
Lifting the instrument
NOTICE
Dragging the instrument can damage the feet of the instrument.
u
Lift the instrument when positioning or re-locating.
u
Lift the instrument at the points indicated.
28/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG
5
5.1
Installation | 5
Installation
Before installation
NOTICE
Instrument damage due to switching it on too early.
Switching on the instrument too early after transportation can cause damage.
u
Climatize the instrument after transportation.
5.2
Establishing electrical connections
NOTICE
Risk of instrument damage because of not suitable power supply cables.
Not suitable power supply cables can cause bad performance or an instrument damage u
Use only BUCHI power supply cables.
5.3
Precondition:
R The electrical installation is as specified on the type plate.
R The electrical installation is equipped with a proper grounding system.
R The electrical installation is equipped with suitable fuses and electrical safety features.
R The installation site is as specified in the
technical date. See Chapter 3.5 "Technical data", page 24
u
Connect the power supply cable to the
connection on the instrument. See Chapter 3.2
u
Connect the mains plug to an own mains outlet socket.
Securing against earthquakes
The instrument has an earthquake fixing point to protect the device against falling.
u
Tie the lashing mount to a fixed point using strong cord or a wire.
Operation Manual K-365 Dist Line 29/106
5 | Installation
5.4
Installing the cooling water supply
5.4.1
Installing a tap water supply (option)
Precondition:
R The tap water supply complies with the
specified parameters. See Chapter 3.5
R Make sure that the instrument is not connected to the power supply.
u
Install the inlet hose to the connection marked
Cooling Water / Chiller IN.
u
Install the drain hose to the connection marked
OUT Cooling Water / Chiller.
Büchi Labortechnik AG u
Put the other end of the hose in a sink.
5.4.2
Installing the recirculating chiller on the condenser (option)
NOTE
Place the hose in the rear cable duct.
Precondition:
R The water supply complies with the tap water
specifications. See Chapter 3.5 "Technical data", page 24
u
Make sure that the instrument is not connected to the power supply.
u
Remove the connection at the condenser inlet.
u
Attach the cooling water hose from the chiller to the condenser.
30/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG u
Install the drain hose to the connection marked
OUT Cooling Water / Chiller on the rear of the instrument.
Installation | 5
5.5
u
Make sure that the hoses are not bend.
Installing the drain tubing (option)
u
Install the drain hose to the connection marked
Drain.
u
Place the other end of the hose in an aqueous waste container located lower than the instrument.
5.6
Installing the waste pump connection (BasicDist, MultiDist,
MultiDist T only)
u
Attach the waste hose onto the connection marked Pump Receiving Vessel OUT.
u
Secure the hose in place with a hose clip.
u
Put the other end of the hose in a suitable collection device.
Operation Manual K-365 Dist Line 31/106
5 | Installation
5.7
Büchi Labortechnik AG
Installing the sample waste pump connection (MultiDist T only)
u
Attach the waste hose to the connection marked Pump Sample Tube OUT.
u
Secure the hose in place with a hose clip.
5.8
u
Put the other end of the hose in a suitable collection device.
Installing the H₂O supply for steam generation
NOTE
MultiDist only
The water supply comes from one canister.
Precondition:
R The water complies with the specified
parameters. See Chapter 3.5 "Technical data", page 24
R Make sure that the instrument is not connected to the power supply.
u
Install the inlet hose to the connection marked
H₂O Steam Gen. IN.
u
Attach the inlet hose in place with a hose clip.
u
Moist the aspiration hose.
u
Push the hoses together.
20
m m u
Attach the hose grommet to the canister lid.
32/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG u
Moist the hose.
u
Insert the inlet hose through the opening of the cap nut and the screw cap of the water tank.
u
Submerse the inlet hose into the water.
5.9
Installing NaOH supply (MultiDist, MultiDist T only)
u
Install the inlet hose onto the connection marked NaOH Sample Tube IN.
u
Attach the inlet hose in place with a hose clip.
Installation | 5 u
Moist the aspiration hose.
u
Push the hoses together.
20
m m u
Attach the hose grommet to the canister lid.
u
Submerse the inlet hose into the NaOH.
5.10
Installing H₃BO₃ supply (MultiDist, MultiDist T only)
u
Install the inlet hose to the connection marked
H₃BO₃ Receiving Vessel IN.
u
Attach the inlet hose in place with a hose clip.
Operation Manual K-365 Dist Line 33/106
5 | Installation u
Moist the aspiration hose.
u
Push the hoses together.
u
Attach the hose grommet to the canister lid.
u
Submerse the inlet hose into the H₃BO₃.
Büchi Labortechnik AG
20
m m
5.11
Installing the H₂O supply for the sample tube (MultiDist,
MultiDist T only)
NOTE
MultiDist only
The water supply comes from one canister.
u
Install the inlet hose to the connection marked
H₂O Sample Tube IN.
u
Attach the inlet hose in place with a hose clip.
u
Moist the aspiration hose.
u
Push the hoses together.
20
m m u
Attach the hose grommet to the canister lid.
34/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG u
Submerse the inlet hose into the H₂O.
Installation | 5
5.12
Installing the level sensor on the canister (option)
NOTE
Do not expose the tank level sensor to an electromagnetic field in the frequency range of 2 to 10 MHz.
u
Attach the hook-and-loop fastener sticker to the sensor.
u
Attach the strap to the canister.
u
Fix the sensor on the canister.
u
Connect the sensor cable to the related
connection on the instrument. See Chapter 3.2
5.13
Installing a titrator with LAN (MultiDist T with Eco Titrator only)
Settings
NOTE
Make sure, that the IP on both instruments is 192.168.10.3
Operation Manual K-365 Dist Line 35/106
5 | Installation u
Connect the LAN cable to the related
connection on the instrument. See Chapter 3.2
Büchi Labortechnik AG
36/106
Specifications on the display:
Precondition:
R The titrator is prepared. See related manual u
Navigate to the specified submenu.
Peripherals
Titrator Model u
Select the specified action.
u
Tap the specified function on the function bar.
u
Select the titrator model you wish to use.
Titrator Model
EDIT
...
...
...
...
u
Tap the specified function on the function bar.
ð The titrator model is saved.
SAVE
5.14
Installing a titrator with RS232 (MultiDist T only)
Precondition:
R The dosing unit is prepared. See related documentation u
Attach the plug onto the connection marked
Titrator (RS-232).
u
Secure the plug in place.
...
Specifications on the display:
Precondition:
R The titrator is prepared. See related manual
R The instrument is connected to the LAN network.
u
Navigate to the specified submenu.
Peripherals
Titrator Model u
Select the specified action.
u
Tap the specified function on the function bar.
Titrator Model
EDIT
...
Operation Manual K-365 Dist Line
Büchi Labortechnik AG Installation | 5 u
Select the titrator model you wish to use.
u
Tap the specified function on the function bar.
ð The titrator model is saved.
Specifications on the display:
...
...
...
...
SAVE
5.15
Installing a dispenser (MultiDist T only)
Precondition:
R The dosing unit is prepared. See related
Documentation u
Attach the plug onto the connection marked
Dosing Unit.
u
Secure the plug in place.
5.16
Installing the reaction detection sensor (MultiDist T only)
u
Attach the reaction detection sensor to the instrument.
u
Connect the sensor cable to the related
connection on the instrument. See Chapter 3.2
5.17
Installing a SD card (Interface pro only)
u
Insert the SD card.
u
Restart the instrument.
ð The status bar shows the SD card symbol.
5.18
Installing the level detection sensor (option)
NOTE
Operation Manual K-365 Dist Line 37/106
5 | Installation u
Attach the metal plate to the magnetic area of the instrument.
u
Push the cable through the rear cable duct.
u
Connect the sensor cable to the related
connection on the instrument. See Chapter 3.2
5.19
Installing the stirrer (MultiDist only)
u
Attach the stirrer to the receiving vessel.
Büchi Labortechnik AG u
Push the cable through the rear cable duct.
u
Connect the sensor cable to the related
connection on the instrument. See Chapter 3.2
38/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG
6
6.1
Description of the interface
Layout of the interface
Description of the interface | 6
1
2
3
4
5
6
7
Fig. 16: Interface
No.
1
Description
Status bar
2
3
4
5
Menu bar
Content area
Function bar
Function buttons
Function
Shows the determination name and status of the instrument.
See Chapter 6.4 "Status bar interface", page 43
Shows symbols representing the menus.
See Chapter 6.3 "Menu bar interface", page 40
Shows current settings, submenus or actions depending on the current operation.
Shows functions that can be performed according to the current operation.
See Chapter 6.2 "Function bar interface", page 40
Pressing a function button performs the assigned function on the function bar.
6
7
Stop button
Navigation control
Using the navigation control
u
Select an entry.
Used for navigating the user interface.
u
Confirm the selected entry.
Operation Manual K-365 Dist Line 39/106
6 | Description of the interface
6.2
6.3
Function bar interface
Büchi Labortechnik AG
The function bar shows available functions according to the current operation.
The functions on the function bar are executed by tapping the relevant function buttons.
Symbol
OK
START
COPY
EDIT
READY
PREP
DELETE
DEL ALL
STANDBY
Description
[Back]
[Confirm]
[Start]
[Copy]
[Edit]
[Ready]
[Preparation]
[Template]
[Delete]
[Delete all]
[Standby]
Menu bar interface
Menu symbol Name
Home menu
Meaning
The display reverts to the previous view.
Confirms an entry.
Starts an extraction process.
Copies the selected method.
Allows the selected item to be edited.
Steam generator is ready to distill.
Carries out the method selected in the [Priming] submenu. See Chap-
ter 8.2 "Editing the priming function", page 52
Saves the selected series as a template.
Creates a series from the selected template.
Deletes the selected entry.
Deletes all entries.
The steam generator changes to standby mode.
Explanation
Process control parameters
Manual control menu
Process menu
Start various processes manually
Editing and saving:
◦ Methods
◦ Series
◦ Templates
◦ Determinations
40/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG Description of the interface | 6
Menu symbol Name
Configuration menu
Explanation
Change settings
Service menu
System information
View processed determinations
Determination data menu
6.3.1
Home menu
6.3.2
Manual control menu
The manual control menu contains the following submenus:
Submenu Explanation
Dose manually
See Chapter 8.13 "Dosing manually", page 74
Aspirate manually
See Chapter 8.14 "Aspirating manually", page 75
[Preheating]
See Chapter 8.1 "Editing the preheating function", page 52
[Priming]
[Cleaning]
See Chapter 8.2 "Editing the priming function", page 52
See Chapter 10.3 "Cleaning the glass components", page 83
[Stirring]
See Chapter 8.15 "Changing the stirring speed manually", page 76
[Pump Calibration]
See Chapter 10.14 "Calibrating the pumps", page 85
6.3.3
Method menu
The method menu contains the following submenus:
Submenu
[Single Determination]
[Series]
[Templates]
[Methods]
Explanation
See Chapter 8.3 "Editing the single determination", page 52
See Chapter 8.4 "Editing a series", page 55
See Chapter 8.5 "Editing a template", page 59
See Chapter 8.6 "Editing a method", page 60
6.3.4
Configuration menu
Settings submenu
Action
[Language]
Option
Choice of display language on the interface
[Current Date] Date input
Explanation
The following languages are available:
English/German/French/Italian/
Spanish/Japanese/Chinese/Russian/
Polish
Enter in sequence: Day, month, year.
Apply the settings by pressing [Save] .
Operation Manual K-365 Dist Line 41/106
6 | Description of the interface Büchi Labortechnik AG
Action
[Current Time]
Time zone Choice of time zone
[Keyboard Layout] Choice of display keyboard on the interface
[Key Tones] Off/On
[Beep on Finish]
[Beep on Error]
[Display Brightness]
Off/On
Off/On
Enter setting
[Zero Amount WarnOn/Off ing]
[Default Amount
Unit]
[MaxAccuracy
Mode]
[AutoDist Mode] On/Off
[Level Detect. Max.
Time]
[Pump Maintenance
Interval]
[H₂O Dosage per
Keypress]
[Acid Dosage per
Keypress]
[Steam Generator
Standby]
Option
Time input
Select value
On / Off
Enter value
Select value
Enter value
Enter value
[NaOH Dosage per
Keypress]
Enter value
[Aspiration Time Receiver]
Select value
Select value [Aspiration Time
Sample Tube]
[H₃BO₃ Dosage per
Keypress]
Enter value
Enter value
Explanation
Enter in sequence: Minutes, hours.
Apply the settings by pressing [Save] .
Select your time zone
Select the keyboard layout
Setting for audible signal in response to input controls.
Setting for audible signal by the end of a determination.
Setting for audible signal by the end of a determination.
Display illumination level in %: 0 -
100
The instrument gives a warning when the entry for the weight is zero.
The following values are selectable: g, mL
Compensation of the steam power caused by voltage fluctuations.
Conditioning and distillation are performed automatically.
The time after which the distillation stops without level detection sensor triggering.
Selecting a frequency to carry out a pump rinsing.
See Chapter 10.20 "Rinsing a pump", page 90
Dosage volume which is applied by pressing the related button.
Dosage volume which is applied by pressing the related button.
Dosage volume which is applied by pressing the related button.
Aspiration time for the related button.
Aspiration time for the related button.
Dosage volume which is applied by pressing the related button.
Enter the time until the instrument turns automatically into standby.
42/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG Description of the interface | 6
Action
[Demo Mode]
[Service Data Logger]
Option
On/Off
On/Off
Explanation
Simulate a distillation.
Saves all data on the instrument for
14 months.
Peripherals submenu
Shows information about the connected peripherals.
Users submenu
Creating and setting users. See Chapter 8.7 "Editing user setting", page 71
Network submenu
Action
[Network]
Option
Enter value
Explanation
The following parameters can be edited:
Device name/MAC address/DHCP/
System IP address/Subnet mask/
Gateway/DNS server/BUCHI Cloud/
Server IP address
Service submenu
NOTE
Only super user can carry out changes in the [Service] submenu.
System Information submenu
Technical information about the connected devices (e.g. serial number, firmware version).
6.3.5
Determination data menu
Shows the history of the determination data.
6.4
Status bar interface
The status bar shows actual information about the instrument.
View Status
The instrument is in standby mode.
Process is running.
The instrument is heating up.
The BLE dongle is connected to the instrument.
A level detection sensor is connected to the instrument.
Reaction detection sensor attached.
Operation Manual K-365 Dist Line 43/106
6 | Description of the interface
View
SD
USB
Büchi Labortechnik AG
Status
A tank level sensor is connected to the instrument.
Dosing unit attached.
Titration unit attached.
White: A SD card is in the instrument.
Red: The SD card in the instrument is in read only mode.
A USB input device is connected.
A USB memory stick is connected to the instrument.
A recirculating chiller is connected.
44/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG
7 Description of the interface pro
CAUTION
Risk of injury from glass splinters
Sharp objects can damage the display.
u
Keep sharp objects away from the display.
7.1
Layout of the interface pro
1
2
3
Description of the interface pro | 7
4
2
3
4
5
6
5
6
Fig. 17: Interface pro
No.
1
Description
Status bar
Menu bar
Content area
Function bar
Stop button
Navigation control
Function
Shows the determination name and status of the instrument.
See Chapter 7.4 "Status bar interface pro", page 50
Shows symbols representing the menus.
Shows current settings, submenus or actions depending on the current operation.
Shows functions that can be performed according to the current operation.
Used for navigating the user interface.
Operation Manual K-365 Dist Line 45/106
7 | Description of the interface pro
Using the navigation control
u
Select an entry.
u
Confirm the selected entry.
Büchi Labortechnik AG
7.2
Function bar interface pro
The function bar shows available functions according to the current operation.
The functions on the function bar are executed by tapping the relevant function buttons.
Symbol Description
[Back]
[Confirm]
Meaning
The display reverts to the previous view.
Confirms an entry.
OK
START
COPY
EDIT
READY
PREP
[Start]
[Copy]
[Edit]
[Ready]
[Preparation]
Starts an extraction process.
Copies the selected method.
Allows the selected item to be edited.
Steam generator is ready to distill.
[Template]
[Delete]
Carries out the method selected in the [Priming] submenu. See Chap-
ter 8.2 "Editing the priming function", page 52
Saves the selected series as a template.
Creates a series from the selected template.
Deletes the selected entry.
DELETE
DEL ALL
STANDBY
[Delete all]
[Standby]
Deletes all entries.
The steam generator changes to standby mode.
46/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG
7.3
Description of the interface pro | 7
Menu bar interface pro
Menu symbol
Name
Home menu
Explanation
Process control parameters
Manual control menu
Process menu
Configuration menu
Start various processes manually
Editing and saving:
◦ Methods
◦ Series
◦ Templates
◦ Determinations
Change settings
Service menu
System information
View processed determinations
Determination data menu
7.3.1
Home menu
1
2
3
4
5
Fig. 18: Home menu during a determination
1 Select and shows operation mode 2 Preview of next determination
3 Short-cut to determination data of the running determination
5 Remaining time
4 Status of the determination (incl. corresponding icons)
7.3.2
Manual control menu
The manual control menu contains the following submenus:
Submenu Explanation
Dose manually
See Chapter 8.13 "Dosing manually", page 74
Aspirate manually
See Chapter 8.14 "Aspirating manually", page 75
Operation Manual K-365 Dist Line 47/106
7 | Description of the interface pro Büchi Labortechnik AG
Submenu Explanation
[Preheating]
[Priming]
See Chapter 8.1 "Editing the preheating function", page 52
See Chapter 8.2 "Editing the priming function", page 52
[Cleaning]
[Stirring]
See Chapter 10.3 "Cleaning the glass components", page 83
See Chapter 8.15 "Changing the stirring speed manually", page 76
[Pump Calibration]
See Chapter 10.14 "Calibrating the pumps", page 85
7.3.3
Method menu
The method menu contains the following submenus:
Submenu
[Single Determination]
[Series]
[Templates]
[Methods]
Explanation
See Chapter 8.3 "Editing the single determination", page 52
See Chapter 8.4 "Editing a series", page 55
See Chapter 8.5 "Editing a template", page 59
See Chapter 8.6 "Editing a method", page 60
7.3.4
Configuration menu
Settings submenu
Action
[Language]
[Current Date]
[Current Time]
Option
Choice of display language on the interface
Date input
Time input
Explanation
The following languages are available:
English/German/French/Italian/
Spanish/Japanese/Chinese/Russian/
Polish
Enter in sequence: Day, month, year.
Apply the settings by pressing [Save] .
Enter in sequence: Minutes, hours.
Apply the settings by pressing [Save] .
Select your time zone
Select the keyboard layout
Time zone Choice of time zone
[Keyboard Layout] Choice of display keyboard on the interface
[Key Tones] Off/On
[Beep on Finish]
[Beep on Error]
[Display Brightness]
[Zero Amount Warning]
Off/On
Off/On
Enter setting
On/Off
Setting for audible signal in response to input controls.
Setting for audible signal by the end of a determination.
Setting for audible signal by the end of a determination.
Display illumination level in %: 0 -
100
The instrument gives a warning when the entry for the weight is zero.
48/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG Description of the interface pro | 7
Action
[Default Amount
Unit]
[MaxAccuracy
Mode]
[AutoDist Mode]
Option
Select value
On / Off
On/Off
[Level Detect. Max.
Time]
Enter value
[Pump Maintenance
Interval]
Select value
[H₂O Dosage per
Keypress]
[NaOH Dosage per
Keypress]
[Acid Dosage per
Keypress]
[H₃BO₃ Dosage per
Keypress]
Enter value
Enter value
Enter value
Enter value
[Aspiration Time Receiver]
Select value
Select value [Aspiration Time
Sample Tube]
[Steam Generator
Standby]
[Demo Mode]
Enter value
On/Off
On/Off [Service Data Logger]
[Export Device] Select value
Explanation
The following values are selectable: g, mL
Compensation of the steam power caused by voltage fluctuations.
Conditioning and distillation are performed automatically.
The time after which the distillation stops without level detection sensor triggering.
Selecting a frequency to carry out a pump rinsing.
See Chapter 10.20 "Rinsing a pump", page 90
Dosage volume which is applied by pressing the related button.
Dosage volume which is applied by pressing the related button.
Dosage volume which is applied by pressing the related button.
Dosage volume which is applied by pressing the related button.
Aspiration time for the related button.
Aspiration time for the related button.
Enter the time until the instrument turns automatically into standby.
Simulate a distillation.
Saves all data on the instrument for
14 months.
Select between the following data export devices:
USB stick/SD card
Peripherals submenu
Shows information about the connected peripherals.
User administration submenu
Creating and setting users. See Chapter 8.7 "Editing user setting", page 71
Operation Manual K-365 Dist Line 49/106
7 | Description of the interface pro Büchi Labortechnik AG
Network submenu
Action
[Network]
Option
Enter value
Explanation
The following parameters can be edited:
Device name/DHCP/MAC address/
System IP address/Subnet mask/
Gateway
Service submenu
NOTE
Only super user can carry out changes in the [Service] submenu.
System Information submenu
Technical information about the connected devices (e.g. serial number, firmware version).
7.3.5
Determination data menu
Shows the history of the determination data.
7.4
Status bar interface pro
The status bar shows actual information about the instrument.
View Status
The instrument is in standby mode.
SD
USB
Process is running.
The instrument is heating up.
The BLE dongle is connected to the instrument.
A level detection sensor is connected to the instrument.
Reaction detection sensor attached.
A tank level sensor is connected to the instrument.
Dosing unit attached.
Titration unit attached.
White: A SD card is in the instrument.
Red: The SD card in the instrument is in read only mode.
A USB input device is connected.
A USB memory stick is connected to the instrument.
50/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG
View
Description of the interface pro | 7
Status
A recirculating chiller is connected.
Operation Manual K-365 Dist Line 51/106
8 | Preparations for a determination
8
8.1
Preparations for a determination
Büchi Labortechnik AG
Editing the preheating function
The preheating function defines how long the instrument parts are conditioned with steam.
Specifications on the display: u
Navigate to the specified submenu.
u
Select the preheating function for which you want to change the preheating time.
u
Tap the specified function on the function bar.
u
Enter the preheating time in seconds.
Preheating
...
EDIT
...
...
52/106
8.2
u
Tap the specified function on the function bar.
ð The time is saved.
SAVE
Editing the priming function
NOTE
The method selected in this submenu is used for preparation.
Specifications on the display: u
Navigate to the specified submenu.
Priming u
Select the method you wish to use.
ð The priming method is selected.
...
...
...
...
8.3
Editing the single determination
8.3.1
Changing the name of the single determination
NOTE
The name of the single determination is displayed in the [Determination data] menu.
Specifications on the display: u
Navigate to the specified submenu.
Single Determination
Operation Manual K-365 Dist Line
Büchi Labortechnik AG Preparations for a determination | 8 u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter the name for the single determination.
Specifications on the display:
Determination Name ...
EDIT
...
u
Tap the specified function on the function bar.
ð The name for the determination is saved.
SAVE
8.3.2
Selecting the determination type for a single determination
The following determination types are selectable:
Determination type
[Blank]
[Reference Substance]
[Sample]
Explanation
For determinations without sample.
For determinations with a defined reference substance.
For determinations with unknown sample.
Specifications on the display: u
Navigate to the specified submenu.
Single Determination u
Select the specified action.
u
Tap the specified function on the function bar.
u
Select the determination type you wish to use.
ð The determination type is selected.
u
Tap the specified function on the function bar.
Type
SAVE
...
SAVE
...
...
...
...
8.3.3
Selecting the method for a single determination
Specifications on the display: u
Navigate to the specified submenu.
Single Determination u
Select the specified action.
u
Tap the specified function on the function bar.
u
Select the method you wish to use.
ð The method is selected.
Method Name
EDIT
...
...
...
...
...
Operation Manual K-365 Dist Line 53/106
8 | Preparations for a determination Büchi Labortechnik AG u
Tap the specified function on the function bar.
Specifications on the display:
SAVE
8.3.4
Changing the sample quantity for single determination
Specifications on the display:
Precondition:
R The determination type is set to [Sample] . See
Chapter 8.3.2 "Selecting the determination type for a single determination", page 53
u
Navigate to the specified submenu.
u
Select the specified action.
Single Determination
...
u
Tap the specified function on the function bar.
u
Enter a value for the sample weight.
EDIT
...
u
Tap the specified function on the function bar.
ð The sample weight is saved.
SAVE
8.3.5
Changing the unit for a single determination
The following units are available:
Unit
[g]
[ml]
Explanation
Enter the weight for the determination in gram.
Enter the weight for the determination in mL.
Specifications on the display:
Precondition:
R The determination type is set to [Sample] or
[Reference Substance]
"Selecting the determination type for a single determination", page 53
u
Navigate to the specified submenu.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Select the unit you wish to use.
ð The unit is saved.
u
Tap the specified function on the function bar.
Single Determination
Unit
EDIT
...
SAVE
...
...
...
...
54/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG Preparations for a determination | 8
8.4
Editing a series
8.4.1
Creating a new series
There are two possibilities to create new series:
Chapter "Creating a new series", page 55
Chapter "Creating a new series from a template", page 55
Creating a new series
Specifications on the display: u
Navigate to the specified submenu.
u
Tap the specified function on the function bar.
ð The new series is created.
Series
NEW
Creating a new series from a template
Specifications on the display: u
Navigate to the specified submenu.
u
Select the template from which you want to create a series.
u
Tap the specified function on the function bar.
ð The series is created.
8.4.2
Changing the name of a series
Templates
...
Specifications on the display: u
Navigate to the specified submenu.
Series u
Select the name of the series that you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter a name for the series.
...
Series Name
EDIT
...
...
u
Tap the specified function on the function bar.
ð The name is saved.
SAVE
Operation Manual K-365 Dist Line 55/106
8 | Preparations for a determination
8.4.3
Adding a determination to a series
Add the first determination to a series
u
Navigate to the specified submenu.
Büchi Labortechnik AG
Specifications on the display: u
Select the series you wish to edit.
u
Tap the specified function on the function bar.
ð The determination is created.
Series
...
NEW
Add more determinations to a series by copy existing
Specifications on the display: u
Navigate to the specified submenu.
Series u
Select the series you wish to edit.
u
Select the determination you wish to copy.
u
Tap the specified function on the function bar.
ð The determination is created.
8.4.4
Deleting a determination from a series
...
...
COPY
Specifications on the display: u
Navigate to the specified submenu.
Series u
Select the series you wish to edit.
u
Select the determination you wish to delete.
...
...
u
Tap the specified function on the function bar.
ð The determination is deleted.
DELETE
8.4.5
Changing the determination name for a series
Specifications on the display: u
Navigate to the specified submenu.
Series
56/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG Preparations for a determination | 8 u
Select the series you wish to edit.
u
Select the determination you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter the name for the single determination.
Specifications on the display:
...
...
Determination Name
EDIT
...
...
u
Tap the specified function on the function bar.
ð The name for the determination is saved.
SAVE
8.4.6
Selecting the determination type within a series
The following determination types are selectable:
Determination type
[Blank]
[Reference Substance]
[Sample]
Explanation
For determinations without sample.
For determinations with a defined reference substance.
For determinations with unknown sample.
Specifications on the display: u
Navigate to the specified submenu.
Series u
Select the series you wish to edit.
u
Select the determination you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Select the determination type you wish to use.
ð The determination type is selected.
u
Tap the specified function on the function bar.
ð The name for the determination is saved.
...
...
Type
EDIT
...
SAVE
...
...
...
...
Operation Manual K-365 Dist Line 57/106
8 | Preparations for a determination Büchi Labortechnik AG
8.4.7
Selecting the method for a determination within a series
Specifications on the display: u
Navigate to the specified submenu.
Series u
Select the series you wish to edit.
u
Select the determination you wish to edit.
...
...
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Select the determination type you wish to use.
ð The determination type is selected.
u
Tap the specified function on the function bar.
Method Name
EDIT
...
SAVE
...
...
...
...
8.4.8
Changing the sample quantity for determination within a series
Specifications on the display:
Precondition:
R The determination type is set to [Sample] or
[Reference Substance]
"Selecting the determination type within a series", page 57
u
Navigate to the specified submenu.
u
Select the series you wish to edit.
Series
...
u
Select the determination you wish to edit.
...
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter a value for the sample weight.
EDIT
...
...
u
Tap the specified function on the function bar.
ð The value for the sample weight is saved.
SAVE
8.4.9
Changing the amount unit for a determination within a series
The following units are available:
Unit
[g]
Explanation
58/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG
Unit
[ml]
Preparations for a determination | 8
Explanation
Enter the weight for the determination in mL. See Chapter 8.3.4
"Changing the sample quantity for single determination", page 54
Specifications on the display:
Precondition:
R The determination type is set to [Sample] or
[Reference Substance]
"Selecting the determination type within a series", page 57
u
Navigate to the specified submenu.
u
Select the series you wish to edit.
u
Select the determination you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Select the determination type you wish to use.
ð The determination type is selected.
u
Tap the specified function on the function bar.
Series
...
...
Unit
EDIT
...
SAVE
...
...
...
...
8.5
Editing a template
8.5.1
Creating a new template
There are two possibilities to create a template:
Chapter "Creating a new template", page 60
Chapter "Creating a new template from an existing series", page 59
Creating a new template from an existing series
Specifications on the display: u
Navigate to the specified submenu.
Series u
Select the series from which you want to create a template.
u
Tap the specified function on the function bar.
ð The template is created.
...
Operation Manual K-365 Dist Line 59/106
8 | Preparations for a determination
Creating a new template
u
Navigate to the specified submenu.
Büchi Labortechnik AG
Specifications on the display: u
Tap the specified function on the function bar.
ð The template is created.
8.5.2
Changing the name of a template
Templates
NEW
Specifications on the display: u
Navigate to the specified submenu.
Templates u
Select the name of the template that you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter a name for the template.
...
Template Name
EDIT
...
...
u
Tap the specified function on the function bar.
ð The template name is saved.
8.6
Editing a method
8.6.1
Creating a new method
SAVE
Specifications on the display: u
Navigate to the specified submenu.
Methods u
Select a method you wish to copy.
...
u
Tap the [Copy] function on the function bar.
ð A new method is created.
COPY
8.6.2
Deleting a method
NOTE
Default methods [Priming Method] and [Standard Method] cannot be deleted.
60/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG Preparations for a determination | 8
Specifications on the display: u
Navigate to the specified submenu.
Methods u
Select a method you wish to delete.
u
Tap the specified function on the function bar.
u
Confirm the secure question.
ð The method is deleted.
8.6.3
Changing the name of a method
...
DELETE
Specifications on the display: u
Navigate to the specified submenu.
Methods u
Select the method that you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter a name for the method.
...
Method Name
EDIT
...
...
u
Tap the specified function on the function bar.
ð The name is saved.
SAVE
8.6.4
Changing the reaction detection for a method (option for MultiDist only)
The reaction detection sensor monitors the alkalization and doses the NaOH accordingly. The instrument switches automatically off when the alkalization is completed.
Status
On
Off
Explanation
Automatic alkalization during the determination.
Enter the parameters for the alkalization manually.
Specifications on the display: u
Navigate to the specified submenu.
Methods
Operation Manual K-365 Dist Line 61/106
8 | Preparations for a determination Büchi Labortechnik AG u
Select the method that you wish to edit.
Specifications on the display:
...
u
Select the specified action.
Reaction Detection u
Tap the specified function on the function bar.
ð The status is changed.
EDIT
8.6.5
Changing the H₂O volume for a method (MultiDist only)
NOTE
The [Reaction Detection]
Manually entering the quantity of H₂O to dilute the sample.
The input values can be found in the corresponding "Application note" .
Specifications on the display: u
Navigate to the specified submenu.
Methods u
Select the method that you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter the volume quantity.
...
H₂O Volume
EDIT
...
...
62/106 u
Tap the specified function on the function bar.
ð The quantity for the dosing is saved.
SAVE
8.6.6
Changing the H₂SO₄ volume for a method (MultiDist only)
NOTE
The [Reaction Detection]
Enter manually the quantity of sulfuric acid.
The value is known from the previous digestion process.
Specifications on the display: u
Navigate to the specified submenu.
u
Select the method that you wish to edit.
Methods
...
Operation Manual K-365 Dist Line
Büchi Labortechnik AG Preparations for a determination | 8 u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter the quantity of the sulfuric acid.
Specifications on the display:
H₂SO₄ for Digest.
...
EDIT
...
u
Tap the specified function on the function bar.
ð The quantity for the sulfuric acid is saved.
SAVE
8.6.7
Changing the NaOH volume for a method (MultiDist only)
NOTE
The [Reaction Detection]
Enter manually the quantity of NaOH for alkalizing the sample.
For the correct quantity check related "Application note" or use the KjelOptimizer
App.
Specifications on the display: u
Navigate to the specified submenu.
Methods u
Select the method that you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter the quantity for the dosing.
...
NaOH Volume
EDIT
...
...
u
Tap the specified function on the function bar.
ð The quantity for the dosing is saved.
SAVE
8.6.8
Changing the acid volume for a method (BasicDist, MultiDist only)
Enter manually the quantity of reagent to be dosed.
Specifications on the display: u
Navigate to the specified submenu.
u
Select the method that you wish to edit.
Methods
...
Operation Manual K-365 Dist Line 63/106
8 | Preparations for a determination Büchi Labortechnik AG u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter the quantity for the dosing.
Specifications on the display:
Acid Volume ...
EDIT
...
u
Tap the specified function on the function bar.
ð The quantity for the dosing is saved.
SAVE
8.6.9
Changing the reaction time for a method (BasicDist, MultiDist only)
NOTE
The [Reaction Detection]
Specifications on the display: u
Navigate to the specified submenu.
Methods u
Select the method that you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter a value for the reaction time.
...
Reaction Time
EDIT
...
...
u
Tap the specified function on the function bar.
ð The reaction time is saved.
SAVE
8.6.10 Changing the number of steam steps for a method
Select steps to get the steam power gradually.
Selection Explanation
[None]
[2]
Steam power in % selectable. See Chapter 8.6.12 "Changing the steam power for a method", page 66
Steam power in % selectable. See Chapter 8.6.12 "Changing the steam power for a method", page 66
64/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG Preparations for a determination | 8
Selection Explanation
[3]
Steam power in % selectable. See Chapter 8.6.12 "Changing the steam power for a method", page 66 and
Specifications on the display: u
Navigate to the specified submenu.
Methods u
Select the method that you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Select the number of steps you wish to use.
...
Steam Steps
EDIT
...
...
...
...
...
u
Tap the specified function on the function bar.
SAVE
8.6.11 Changing the time a steam step is carried out for a method
NOTE
Calculating the starting time for a steam step
R Make sure that the distillation time is set longer than the calculated time for the
steam steps. See Chapter 8.6.14 "Changing the distillation time for a method", page 67
u
Starting time for steam step two:
ð The duration time of the first steam step.
u
Starting time for steam step three:
ð The duration time of the first steam step.
ð +
ð The duration time of the second steam step.
Specifications on the display:
Precondition:
R More than one steam step is selected. See
Chapter 8.6.10 "Changing the number of steam steps for a method", page 64
u
Navigate to the specified submenu.
u
Select the method that you wish to edit.
u
Select the specified action.
Methods
...
Calibration NaOH ...
Operation Manual K-365 Dist Line 65/106
8 | Preparations for a determination Büchi Labortechnik AG u
Tap the specified function on the function bar.
Specifications on the display:
EDIT u
Enter the time for the steam step.
...
u
Tap the specified function on the function bar.
ð The time for the steam step is saved.
8.6.12 Changing the steam power for a method
The steam power in % during a steam step.
SAVE
Specifications on the display: u
Navigate to the specified submenu.
Methods u
Select the method that you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter the level for the steam power.
...
Steam Power
EDIT
...
...
66/106 u
Tap the specified function on the function bar.
ð The level for the steam power is saved.
SAVE
8.6.13 Changing the level detection sensor settings (option) for a method
The level detection stops the distillation when a defined distillate volume is reached.
NOTE
The maximum level detection time can be changed in the Settings submenu.
Selection Explanation
On
Off
The level detection sensor is active.
The level detection sensor is off.
Specifications on the display: u
Navigate to the specified submenu.
u
Select the method that you wish to edit.
Methods
...
Operation Manual K-365 Dist Line
Büchi Labortechnik AG Preparations for a determination | 8 u
Select the specified action.
Specifications on the display:
Level Detection u
Tap the specified function on the function bar.
ð The status changed.
EDIT
8.6.14 Changing the distillation time for a method
Enter manually the time for the determination.
Specifications on the display:
Precondition:
R The [Level Detection] action is set to Off. See
Chapter 8.6.13 "Changing the level detection sensor settings (option) for a method", page 66
u
Navigate to the specified submenu.
u
Select the method that you wish to edit.
Methods
...
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter a value for the distillation time.
Distillation Time
EDIT
...
...
u
Tap the specified function on the function bar.
ð The distillation time is saved.
SAVE
8.6.15 Changing the distillation stirring speed for a method (MultiDist T only)
Specifications on the display: u
Navigate to the specified submenu.
Methods u
Select the method that you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter the stirring speed.
...
Stirrer Speed Dist.
EDIT
...
...
u
Tap the specified function on the function bar.
ð The stirring speed is saved.
SAVE
Operation Manual K-365 Dist Line 67/106
8 | Preparations for a determination Büchi Labortechnik AG
8.6.16 Changing the titration type for a method (MultiDist T only)
Titration type
[None]
[Boric Acid Titration]
[Back Titration]
Explanation
No titration
Titration
Back titration
Specifications on the display: u
Navigate to the specified submenu.
Methods u
Select the method that you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Select the titration type you wish to use.
...
Titration Type
EDIT
...
...
...
...
...
u
Tap the specified function on the function bar.
SAVE
8.6.17 Changing the H₃BO₃ volume for a method (MultiDist only)
NOTE
The availability depends on the selected titration type.
Specifications on the display: u
Navigate to the specified submenu.
Methods u
Select the method that you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter the quantity for the volume.
...
H₃BO₃ Volume
EDIT
...
...
u
Tap the specified function on the function bar.
ð The volume is saved.
SAVE
68/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG Preparations for a determination | 8
8.6.18 Changing the dosing unit volume for a method (MultiDist T only)
NOTE
The availability depends on the selected titration type.
Specifications on the display: u
Navigate to the specified submenu.
Methods u
Select the method that you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter the quantity for the volume.
...
Dosing Unit Vol.
EDIT
...
...
u
Tap the specified function on the function bar.
ð The volume is saved.
SAVE
8.6.19 Changing the titration stirring speed for a method (MultiDist T only)
NOTE
The availability depends on the selected titration type.
Specifications on the display:
Precondition:
R A titration type is activated. See u
Navigate to the specified submenu.
Methods u
Select the method that you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter the stirring speed.
...
Stirrer Speed Titr.
EDIT
...
...
u
Tap the specified function on the function bar.
ð The stirring speed is saved.
SAVE
8.6.20 Changing the titration start time for a method (MultiDist T only)
NOTE
The availability depends on the selected titration type.
Operation Manual K-365 Dist Line 69/106
8 | Preparations for a determination Büchi Labortechnik AG
Specifications on the display: u
Navigate to the specified submenu.
Methods u
Select the method that you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter a value for the titration start time.
...
Titration Start Time
EDIT
...
...
u
Tap the specified function on the function bar.
ð The start time is saved.
SAVE
8.6.21 Changing the aspiration time for the sample tube for a method
(BasicDist, MultiDist, MultiDist T only)
Specifications on the display: u
Navigate to the specified submenu.
Methods u
Select the method that you wish to edit.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter a value for the aspiration time.
...
Sample Tube Aspiration
EDIT
...
...
70/106 u
Tap the specified function on the function bar.
ð The aspiration time is saved.
SAVE
8.6.22 Changing the aspiration time for the receiving vessel for a method
(MultiDist T only)
Specifications on the display: u
Navigate to the specified submenu.
u
Select the method that you wish to edit.
u
Select the specified action.
Methods
...
Receiver Aspiration ...
Operation Manual K-365 Dist Line
Büchi Labortechnik AG Preparations for a determination | 8 u
Tap the specified function on the function bar.
Specifications on the display:
EDIT u
Enter a value for the aspiration time.
...
u
Tap the specified function on the function bar.
ð The aspiration time is saved.
SAVE
8.7
Editing user setting
NOTE
The user [Service User] is a default user and password protected.
8.7.1
Creating a new user
Specifications on the display: u
Navigate to the specified submenu.
u
Tap the specified function on the function bar.
u
Enter a name for the user.
User Administration
NEW
...
u
Tap the specified function on the function bar.
SAVE
8.7.2
Deleting a user
Specifications on the display: u
Navigate to the specified submenu.
User Administration u
Select the user you wish to delete.
u
Tap the specified function on the function bar.
u
Confirm the secure question.
ð The user is deleted.
...
DELETE
...
Operation Manual K-365 Dist Line 71/106
8 | Preparations for a determination
8.7.3
Changing the name of a user
u
Navigate to the specified submenu.
Büchi Labortechnik AG
Specifications on the display: u
Select the user that you wish to edit.
u
Tap the specified function on the function bar.
u
Enter a new name for the user.
User Administration
...
EDIT
...
...
u
Tap the specified function on the function bar.
ð The new name is saved.
8.7.4
Selecting a user
SAVE
Specifications on the display: u
Navigate to the specified submenu.
8.8
u
Select the user you wish to use.
u
Tap the specified function on the function bar.
ð The user is active.
ð The user is marked up green.
Installing the sample tube
NOTE
Removing is done in reverse sequence.
Precondition:
R Make sure, that the end stop is in the correct position for the sample tube in use. See
Chapter 8.10 "Preparing the end stop for the sample tube size", page 73
u
Hold the sample tube with the tongs.
User Administration
...
SET
...
72/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG u
Insert the sample tube.
Preparations for a determination | 8 u
Apply a light force to push the sample tube against the end stop and the sample tube sealing.
u
Secure the sample tube in place with the handle.
8.9
Installing a receiving vessel
u
Insert the sample receiving vessel to the instrument.
u
Place the sample receiving vessel on the receiving vessel area.
8.10
Preparing the end stop for the sample tube size
u
Pull the end stop.
u
Twist the end stop.
Operation Manual K-365 Dist Line 73/106
8 | Preparations for a determination
8.11
Preparing the level detection sensor (option)
Precondition:
R The installations for the level sensor are
completed. See Chapter 5.18 "Installing the level detection sensor (option)", page 37
u
Attach the magnetic level sensor to the metal plate.
u
Adjust the sensor near the sample receiving vessel.
Büchi Labortechnik AG
10 mm
74/106
8.12
Selecting the operation mode
Select one of the following operation modes:
Operation mode Explanation
[Single Determination] For carrying out one determination. Editing settings see
Chapter 8.3 "Editing the single determination", page 52
[Series] For planning and carrying out a series of determinations.
Editing settings see Chapter 8.4 "Editing a series", page 55
Specifications on the display: u
Navigate to the specified menu.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Select the operation mode according to your requirements.
u
Tap the specified function on the function bar.
Operation Mode
EDIT
...
...
...
...
SAVE
...
8.13
Dosing manually
8.13.1 Dosing H₂O manually
NOTE
The quantity of liquid per keystroke can be changed in the [Settings] submenu.
Operation Manual K-365 Dist Line
Büchi Labortechnik AG Preparations for a determination | 8
Specifications on the display:
Precondition:
R
A sample tube is installed. See Chapter 8.8
"Installing the sample tube", page 72
u
Navigate to the specified menu.
u
Tap the specified button.
ð The pump applies a defined quantity of H₂O.
H₂O
8.13.2 Dosing NaOH manually
NOTE
The quantity of liquid per keystroke can be changed in the [Settings] submenu.
Precondition:
R
A sample tube is installed. See Chapter 8.8
"Installing the sample tube", page 72
u
Navigate to the specified menu.
u
Tap the specified button.
ð The pump applies a defined quantity of NaOH.
Specifications on the display:
NaOH
8.13.3 Dosing H₃BO₃ manually
NOTE
The quantity of liquid per keystroke can be changed in the [Settings] submenu.
Specifications on the display:
Precondition:
R
A receiving vessel is installed. See Chapter 8.9
"Installing a receiving vessel", page 73
u
Navigate to the specified menu.
u
Tap the specified button.
ð The pump applies a defined quantity of
H₃BO₃.
H₃BO₃
8.14
Aspirating manually
8.14.1 Aspirating the sample tube manually
NOTE
The aspiration time per keystroke can be changed in the Settings submenu.
Specifications on the display: u
Navigate to the specified menu.
Operation Manual K-365 Dist Line 75/106
8 | Preparations for a determination Büchi Labortechnik AG u
Tap the specified button.
ð The pump aspirates liquid from the sample tube.
Specifications on the display:
[Sample Tube Aspiration]
8.14.2 Aspirating the receiving vessel manually
NOTE
The aspiration time per keystroke can be changed in the Settings submenu.
Specifications on the display: u
Navigate to the specified menu.
u
Tap the specified button.
ð The pump aspirates liquid from the receiving vessel.
[Receiver Aspiration]
8.15
Changing the stirring speed manually
Specifications on the display: u
Navigate to the specified submenu.
Stirring u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter the stirring speed.
Stirrer Speed
EDIT
...
...
u
Tap the specified function on the function bar.
ð The stirring speed is saved.
SAVE
8.16
Switching valve adjustment (MultiDist only)
NOTE
After switching the valve, the instrument automatically rinses the pumps.
The following valve positions are available:
Valve position
Acid
NaOH
Explanation
The system dose acid to the sample tube.
The system dose NaOH to the sample tube.
76/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG
Precondition:
R The On /Off master switch is set to on.
u
Rinse the pump you want to use. See
Chapter 10.20 "Rinsing a pump", page 90
u
Turn the valve in the correct position.
u
Follow the instructions on the interface.
Preparations for a determination | 8
Operation Manual K-365 Dist Line 77/106
9 | Carrying out a determination
9
9.1
9.2
Carrying out a determination
Büchi Labortechnik AG
Preparing the instrument
Navigation path
➔
Precondition:
R
All commissioning operations have been completed. See Chapter 5 "Installation", page 29
u
Set the On/Off master switch to On.
ð The instrument is starting up.
u u
Make sure that no defective sealings or glass parts are used.
Navigate to the Home menu according the navigation path.
u
Tap the [READY] button on the function bar.
ð The instrument is heating up.
u u u u u
Install the sample tube. See Chapter 8.8 "Installing the sample tube", page 72
Close the protection shield.
Flush the tubing. See Chapter 8.13 "Dosing manually", page 74
Aspirate the receiving and the sample vessels. See Chapter 8.14 "Aspirating manually", page 75
Starting a determination
Navigation path
➔
Precondition:
R
The instrument is prepared. See Chapter 9.1 "Preparing the instrument", page 78
R The sample is prepared.
R The connected instruments are prepared (MultiDist T only). See related documentation.
u u
Navigate to the Home menu according the navigation path.
Press the [PREP] button on the function bar.
ð Wait until the status bar shows the status READY.
u u
Select the operation mode. See Chapter 8.12 "Selecting the operation mode", page 74
Tap the [START] button on the function bar.
78/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG
9.3
9.4
9.5
9.6
Carrying out a determination | 9
Ending a determination
Navigation path
➔
Precondition:
R The display shows the Finished dialog.
u Confirm the message on the dialog.
Shutting down the instrument
Precondition:
R The extraction process has ended.
u u u u u
Shut down the connected instruments (MultiDist T only). See related documentation.
Clean the instrument. See Chapter 10.3 "Cleaning the glass components", page 83
Install an empty sample tube. See Chapter 8.8 "Installing the sample tube", page 72
Install an empty receiving vessel (EasyDist and BasicDist only). See Chapter 8.9
"Installing a receiving vessel", page 73
Set the On/Off master switch to off.
Filtering determination data
Specifications on the display: u
Navigate to the specified menu.
u
Select the specified action.
u
Select the filter you wish to use.
Filter by
...
...
...
...
...
Showing determination data
u
Navigate to the specified submenu.
u
Select the determination you wish to view.
ð The display shows the data of the determination.
9.7
Deleting determination data
9.7.1
Deleting one determination
...
u
Navigate to the specified menu.
u
Select the determination you wish to delete.
Specifications on the display:
...
Operation Manual K-365 Dist Line 79/106
9 | Carrying out a determination Büchi Labortechnik AG u
Tap the specified function on the function bar.
Specifications on the display:
DELETE u
Confirm the secure question.
ð The determination data is deleted.
9.7.2
Deleting all determination data
Specifications on the display: u
Navigate to the specified menu.
u
Tap the specified function on the function bar.
u
Confirm the secure question.
ð All determination data are deleted.
DEL ALL
9.8
Exporting determination data (Interface pro only)
9.8.1
Exporting one determination (Interface pro only)
Specifications on the display:
Precondition:
R A storage medium is available. See Chapter
u
Navigate to the specified menu.
u
Select the determination you wish to export.
...
u
Tap the specified function on the function bar.
EXPORT
9.8.2
Exporting all determination data (Interface pro only)
Specifications on the display: u
Navigate to the specified menu.
u
Tap the specified function on the function bar.
EXPORT
ALL
80/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG
10
Cleaning and servicing | 10
Cleaning and servicing
NOTE
Users may only carry out the servicing and cleaning operations described in this section.
Any servicing and repair work which involves opening up the casing may only be carried out by BUCHI service technicians.
u Use only genuine BUCHI consumables and spare parts in order to ensure correct operation of the device and preserve the warranty.
Operation Manual K-365 Dist Line 81/106
10 | Cleaning and servicing
10.1
Regular maintenance work
Büchi Labortechnik AG
82/106
Action
Cleaning and servicing the sample tube
Cleaning and servicing the hoses and hose connections
Cleaning and servicing the sample tube seal
Cleaning and servicing the splash protector
1
1
2
2
2
2
2
2
2
Yearly Additional information
Carry out this action before every instrument use.
Carry out this action after every instrument use.
Replace the sample tube seal twice per year
Replace the glass splash protector after ~ 3500 distillations (twice per year).
Replace the plastic splash protector once in 2 years.
Replace the tube once per year.
Cleaning and servicing the steam inlet and condenser outlet tube
Cleaning and servicing the warning and directive symbols
Cleaning and servicing the bridge splash protector to condenser
Cleaning and servicing the dosing pumps
Cleaning and servicing the condenser
Cleaning and servicing the steam generator
10.14.1 Calibrating the H₂O pump
10.14.2 Calibrating the acid pump
10.14.3 Calibrating the NaOH pump
1 - User; 2 - Operator
2
2
2
2
2
2
2
2
Replace the bridge once per year.
Replace the NaOH pump once per year
10.2
Cleaning and servicing the sample tube
u u
Before using, check the sample tube for defects (cracks/splintering).
Prevent temperature shocks on the sample tubes.
ð Temperature shocks may lead to breakage.
Operation Manual K-365 Dist Line
Büchi Labortechnik AG Cleaning and servicing | 10
Cleaning
If cleaning with a dishwashing machine: u Make sure, that there is no contact with other glassware.
Storage (e.g. drawer)
u
Prevent rolling and small hits.
ð This can cause hair cracks which may lead to breakages.
10.3
Cleaning the glass components
The following values for the cleaning can be adjusted (the list differs according to the instrument configuration):
[H₂O Volume]
[Steam Power]
[Distillation Time]
[Sample Tube Aspiration]
[Receiver Aspiration]
Specifications on the display:
Precondition:
R
A sample tube is installed. See Chapter 8.8
"Installing the sample tube", page 72
R A sample receiving vessel is installed. See
Chapter 8.9 "Installing a receiving vessel", page 73
u
Navigate to the specified submenu.
u
Adjust the values according to your needs.
Cleaning
...
u
Confirm the entries with the specified function in the function bar.
u
Tap the specified function on the function bar.
SAVE
START
10.4
Cleaning and servicing the hoses and hose connections
u u
Check the hoses and hose connections for defects (cracks, brittle areas).
If defective, replace the hoses.
10.5
Cleaning and servicing the sample tube seal
u Check the seal for the following damage that may cause leakage:
Deterioration (surface roughness)
Deformation (rubber shape)
Residues (dirt, white crystals) u
Rinse the seal with water.
Operation Manual K-365 Dist Line 83/106
10 | Cleaning and servicing Büchi Labortechnik AG u
If necessary, replace the seal. See Chapter 10.21 "Replacing the sample tube seal", page 90
10.6
Cleaning and servicing the splash protector
u u u
Check for leaks on the connectors.
Check for residues inside the splash protector.
Check the glass wall of the main body and the connectors.
ð If the part is dirty, clean it with detergents.
ð If the glass part is showing deterioration (transparency loss/reams in the glass wall) or leakages (white residues), replace it.
10.7
Cleaning and servicing the steam inlet and condenser outlet tube
u u
Check the tubes concerning dirt or residues.
Clean and replace periodically depending on the use.
10.8
Cleaning the housing
u u u
Wipe down the housing with a damp cloth.
If heavily soiled, use ethanol or a mild detergent.
Wipe down the display with a damp cloth.
10.9
Cleaning and servicing the warning and directive symbols
u u
Check that the warning symbols on the instrument are legible.
If they are dirty, clean them.
10.10 Cleaning and servicing the bridge splash protector to condenser
u u u
Check if it is tight.
Check for residues inside the tube.
If necessary, clean it with a brush and water.
10.11 Cleaning and servicing the dosing pumps
u Check the pump connectors in the dosing pump window for:
Leakages
Crystallization u If necessary, contact BUCHI customer service.
10.12 Cleaning and servicing the condenser
Decalcification solution:
160 g citric acid or 80 g amidosulfuric acid dissolved in 0.8 L water.
u
Check if the cooling media of the condenser is transparent.
84/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG Cleaning and servicing | 10 u
Check if the inside of the coil is showing condensation droplets.
ð If droplets appear:
Flush the condenser coil through the condensate expansion vessel with an 1% hydrochloric acid.
Flush the condenser coil through the condensate expansion vessel with water and water/ethanol mixture.
u Clean the condenser on the outside of the coil (water connectors) with decalcification solution to remove chalk, algae, and rust.
10.13 Cleaning and servicing the steam generator
u
Check the distillation volume.
ð If the condensate amount is less than 160 mL water/per 5 minutes distillation
10.14 Calibrating the pumps
10.14.
1
Calibrating the H₂O pump
Precondition:
R
A sample tube is installed. See Chapter 8.8
"Installing the sample tube", page 72
u
Navigate to the specified submenu.
Specifications on the display:
Pump Calibration
H₂O u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter the calibration volume you wish to use.
Nominal Volume
EDIT
...
...
u
Tap the specified function on the function bar.
u
Wait until the display shows a numeric dialog.
START
...
u
Remove the sample tube.
u
Measure the volume in the sample tube.
u
Enter the value in the numeric dialog.
u
Tap the specified function on the function bar.
SAVE
Operation Manual K-365 Dist Line 85/106
10 | Cleaning and servicing Büchi Labortechnik AG
10.14.
2
Calibrating the acid pump
Precondition:
R
A sample tube is installed. See Chapter 8.9
"Installing a receiving vessel", page 73
u
Navigate to the specified submenu.
u
Select the specified action.
u
Tap the specified function on the function bar.
u
Wait until the display shows a numeric dialog.
Specifications on the display:
Pump Calibration
Acid
START
...
u
Remove the sample tube.
u
Measure the volume in the sample tube.
u
Enter the value in the numeric dialog.
u
Tap the specified function on the function bar.
SAVE
10.14.
3
Calibrating the NaOH pump
Precondition:
R
A sample tube is installed. See Chapter 8.9
"Installing a receiving vessel", page 73
u
Navigate to the specified submenu.
Specifications on the display:
Pump Calibration
NaOH u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter the calibration volume you wish to use.
Nominal Volume
EDIT
...
...
u
Tap the specified function on the function bar.
u
Wait until the display shows a numeric dialog.
START
...
u
Remove the sample tube.
u
Measure the volume in the sample tube.
u
Enter the value in the numeric dialog.
u
Tap the specified function on the function bar.
SAVE
86/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG Cleaning and servicing | 10
10.14.
4
Calibrating the H₃BO₃ Pump
Precondition:
R
A receiving vessel is installed. See Chapter 8.9
"Installing a receiving vessel", page 73
u
Navigate to the specified submenu.
Specifications on the display:
Pump Calibration
H₃BO₃ u
Select the specified action.
u
Tap the specified function on the function bar.
u
Enter the calibration volume you wish to use.
Nominal Volume
EDIT
...
...
u
Tap the specified function on the function bar.
u
Wait until the display shows a numeric dialog.
START
...
u
Remove the sample receiving vessel.
u
Measure the volume in the sample receiving vessel.
u
Enter the measured volume.
u
Tap the specified function on the function bar.
SAVE
10.15 Decalcifying the instrument
Time required: approximately 2.5 hours
Decalcification solution:
160 g citric acid or 80 g amidosulfuric acid dissolved in 0.8 L water.
Precondition:
R The instrument temperature is the same as the ambient temperature.
u u u u u u u u u
Prepare the decalcification solution.
Install a suitable hose to the H₂O supply for steam generation.
Put the other end of the hose in the decalcification solution.
Set the On/Off master switch to On.
Press the [READY] button on the function bar.
Wait until the pumps stop working.
Set the On/Off master switch to Off.
Wait 30 minutes.
Operation Manual K-365 Dist Line 87/106
10 | Cleaning and servicing Büchi Labortechnik AG u u u u u u u u u
Remove the decalcification solution from the instrument. See Chapter 12.4
"Removing water from the steam generator", page 98
Do steps (5) thru (10) again.
Install the H₂O supply for steam generation.
Set the On/Off master switch to On.
Press the [READY] button on the function bar.
Wait until the pumps stop working.
Set the On/Off master switch to Off.
Do steps (11) thru (16) five times.
10.16 Replacing the splash protector
NOTE
Installing is done in reverse sequence.
u
Remove the protection shield. See
Chapter 10.18 "Attaching and removing the protection shield", page 89
u
Loosen the cap nut at the splash protector.
u
Open the splash protector nut.
u
Remove the splash protector.
u
Remove the steam connection.
u
Remove the NaOH/acid connection (according to the instrument configuration).
88/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG
10.17 Replacing the condenser
NOTE
Installing is done in reverse sequence.
u
Remove the protection shield. See
Chapter 10.18 "Attaching and removing the protection shield", page 89
u
Loosen the cap nut at the condenser.
Cleaning and servicing | 10 u
Remove the cooling water hoses.
u
Remove the rubber band.
10.18 Attaching and removing the protection shield
10.18.
1
Removing the protection shield
u
Pull the protection shield from the instrument.
Operation Manual K-365 Dist Line 89/106
10 | Cleaning and servicing
10.18.
2
Attaching the protection shield
u
Attach the protection shield onto the instrument.
Büchi Labortechnik AG
10.19 Cleaning the drip tray
u
Pull the drip tray.
u
Rinse the trip tray with water.
10.20 Rinsing a pump
Precondition:
R
A sample tube is installed. See Chapter 8.8 "Installing the sample tube", page 72
u u
Put the side of the hose in distilled water.
Apply 100 mL with the manual functions. See Chapter 8.13 "Dosing manually", page 74
10.21 Replacing the sample tube seal
NOTE
Installing is done in reverse sequence.
NOTE
Make the seal moist with water before installation.
u
Remove the protection shield. See
Chapter 10.18.1 "Removing the protection shield", page 89
u
Loosen the cap nut.
90/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG u
Open the splash protector nut.
u
Remove the splash protector.
u
Remove the seal.
Cleaning and servicing | 10
Operation Manual K-365 Dist Line 91/106
11 | Help with faults Büchi Labortechnik AG
11 Help with faults
11.1
Troubleshooting digestion
Problem
Crystallization after digestion
Samples do not get clear
Fume leakage
Boiling retardation/ bumping/foaming
Possible cause Action
False ratio of H
2
SO
4
to catalyst.
u
Use KjelOptimizer software to optimize the digestion
(download available on
BUCHI website).
Digestion time too long.
u
Decrease digestion time.
u
Compare to conditions to similar applications.
Suction capacity of scrubber too strong.
No or not enough catalyst used.
u
Lower the suction capacity on the scrubber. See
"Scrubber" operation manual u
Use KjelOptimizer software to optimize the digestion
(download available on
BUCHI website).
Digestion temperature is too low.
Temperature is too high.
u
Decrease digestion time.
u
Compare to conditions to similar applications.
u
Decrease digestion temperature.
u
Check the sealing.
Sealing material was flushed into the sample.
The seals are defective.
Suction capacity of scrubber is too weak.
u
Replace the seals.
u
Carry out the scrubber maintenance. See
"Scrubber" operation manual
Leakage in the system, e.g.
hose connector not tight.
Blocked hoses.
u u
Check connections between digester and scrubber.
Clean the hoses.
Reduced suction on the bypass valve.
Missing digestion rods or boiling stones.
u
Add boiling rods during digestion.
Missing antifoam tablet or other anti-foaming agent.
u
Add antifoam tablet.
92/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG Help with faults | 11
Problem Possible cause
Samples do not get dark blue/brown after addition of
NaOH
Air in NaOH hose.
No catalyst used for digestion
(only H
2
O
2
or Kjeldahl tablets
ECO).
Action u
Flush the system. See
Chapter 8.13 "Dosing manually", page 74
11.2
Troubleshooting distillation
Problem Possible cause Action
Splashing during distillation or addition
The sample tube is too small.
u
Select a bigger sample tube.
Sample volume too high.
u
Lower sample volume.
of chemicals
Amount of water used for dilution is too low.
u
Increase dilution volume.
Measured nitrogen content is too high
Air in titration system, burette or tubes.
u u
Tighten the connections.
Flush the tubing with titrant.
u
Refill burette.
Carry over during distillation.
u
Insert less sample.
u
Avoid big concentration differences in measurement series.
Titrant concentration too high.
u
Lower the titrant concentration.
Error in calculation.
u
Check calculation.
u
Check titration parameters.
u
Check the titer for the titrant in use.
pH electrode is defective.
The glassware is dirty.
u
Maintain electrode. See related documentation u
If necessary, replace it.
u
Clean the glassware. See
Chapter 10.2 "Cleaning and servicing the sample tube", page 82
Operation Manual K-365 Dist Line 93/106
11 | Help with faults Büchi Labortechnik AG
Problem
Measured nitrogen content is too low
Possible cause Action
Incomplete digestion.
u
Increase digestion time.
u
Use different Kjeldahl tablet.
The H₂SO₄ volume is too low.
u
Use KjelOptimizer software to optimize the digestion
(download available on
BUCHI website).
Kjeldahl Tablets and H₂SO₄ in wrong ratio.
u
Correct ratio of Kjeldahl
Tablets and H₂SO₄.
u
Use KjelOptimizer software to optimize the digestion
(download available on
BUCHI website).
Nitrogen content per sample tube is too high.
u
Not apply more than 200 mg
Nitrogen per sample tube.
u
Use KjelOptimizer software to optimize the digestion
(download available on
BUCHI website).
Not enough NaOH or incorrect concentration of NaOH used
(required is 32 %) u
Correct volume for complete alkalization of the digested sample.
Leakage during distillation.
u
Check connection between condenser and splash protector.
u
Tighten the connection.
u
If necessary, replace it.
Titrant solution pH electrode is defective.
The glassware is dirty.
Incorrect weighing.
u
Check titer of titrant.
u
Maintain electrode. See related documentation u
If necessary, replace it.
u
Clean the glassware. See
Chapter 10.2 "Cleaning and servicing the sample tube", page 82
u
Use weighing boats (easy sample transfer from balance to sample tube).
u
Use anti-static equipment.
u
Use larger sample sizes.
94/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG Help with faults | 11
Problem Possible cause Action
Poor repeatability Air bubbles in titration system, burette, tubes.
u
Tighten the connections.
u
Flush the tubing with titrant.
u
Refill burette.
Aspiration not working properly.
Sample is inhomogeneous.
u u u
Check for leaks.
Tighten the connections.
Homogenize the sample.
Sample weighing problems.
u
Use weighing boats (easy sample transfer from balance to sample tube).
u
Use anti-static equipment.
u
To decrease the degree of error, keep the weighed sample portion as high as possible.
Incomplete digestion, digestion time too short.
Stirrer is defective.
u
Choose digestion time accordingly.
u
Check color of samples during digestion.
ð Solution should be transparent by the end of the digestion.
u
Clean the stirrer.
u
If necessary, replace the stirrer.
Loose contact of the sensor cables.
u
Check the sensor cables.
11.3
Troubleshooting instrument
Problem
The instrument does not work
Possible cause
The installation site has no power supply.
The instrument is not connected to the power supply.
The power supply cable is defect.
The fuse was triggered.
The switch is defect.
Action u
Check the power supply of the installation site.
u
Connect the instrument to the power supply. See
Chapter 5.2 "Establishing electrical connections", page 29
u
Replace the power supply cable.
u
Reset the fuse u Contact BUCHI Customer
Service.
Operation Manual K-365 Dist Line 95/106
11 | Help with faults Büchi Labortechnik AG
Problem
Steam generator is not reaching the status Ready
Display is black
No cooling water flow
Dosing pump is not feeding
Aspiration is not working
Possible cause Action
Not enough water in the steam u
Make sure, that the generator.
installation H₂O supply for steam generation is correct.
u
Make sure, that there is enough water in the canister.
u
Make sure, that the hose is submerged in the water.
u
Check for leaks between instrument and canister.
u
Check water level status in the steam generator.
The connection between instrument and display is interrupted.
u
Check the connection cable from the instrument to the display.
u Contact BUCHI Customer
Service.
The cooling water supply is blocked.
u
Make sure, that the hoses are not bend.
u
Check for leaks between instrument and cooling water source.
Water flow sensor blocked.
u
Clean the hoses.
Not enough liquid for feeding.
u
Make sure, that the installation is correct.
u
Make sure, that there is enough liquid in the canister.
u
Make sure, that the hose is submerged in the liquid.
u
Check for leaks between instrument and canister.
u
Make sure that liquid is inside the pump.
Leaks u
Check the hoses connected with the pump for leaks and deterioration.
The aspiration pump is defect.
u
Check if the wheel inside the pump is turning.
u Contact BUCHI Customer
Service.
96/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG
11.4
Turning the splash protector nut
u
Turn the splash protector nut.
Help with faults | 11
Operation Manual K-365 Dist Line 97/106
12 | Taking out of service and disposal Büchi Labortechnik AG
12 Taking out of service and disposal
12.1
Taking out of service
u u u u u u u
Rinse all pumps. See Chapter 10.20 "Rinsing a pump", page 90
Remove all reagents and coolants.
Clean the instrument.
Set the On/Off master switch to Off.
Disconnect the power supply.
Remove all tubing and cables from the instrument.
12.2
Disposal
The operator is responsible for proper disposal of the instrument.
u
When disposing of equipment observe the local regulations and statutory u requirements regarding waste disposal.
When disposing, observe the disposal regulations of the materials used. Materials
used see Chapter 3.5 "Technical data", page 24
12.3
Returning the instrument
Before returning the instrument, contact the BÜCHI Labortechnik AG Service
Department.
https://www.buchi.com/contact
12.4
Removing water from the steam generator
Precondition:
R The On / Off master switch is set to Off.
u
Install a drain hose to the connection marked
Steam Generator OUT.
u
Put the other end of the hose in a sink.
u
Open the valve.
98/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG
13 Appendix
13.1
Spare parts and accessories
13.1.1 Accessories
Reaction detection sensor
Order no.
11072666
OnLevel sensor
Tank level sensor kit
11070270
11072294
Eco Titrator
StatusLight cpl., incl. communication cable
Indicates the status of the instrument (instrument is ready to use, has an error or is in operation).
11072748
11068959
BUCHI Bluetooth® Dongle, connects instrument to smartphone via Bluetooth®
11067770
Adapter for 3rd party sample tubes 11072398
Sample tube holder (4x 500 mL tubes)
Sample tube holder (6 x 300 mL tubes)
016951
043039
Sample tube holder (12 x 300 mL tubes) 043041
Cyanide caps
Caps for hermetical closing of all types of Kjeldahl sample tubes during sample preparation for cyanide in food
11067871
Appendix | 13
Image
Operation Manual K-365 Dist Line 99/106
13 | Appendix
Communication cableRJ45, 2 m
Connection between vacuum controller/interface and recirculating chiller, vacuum controller/interface and vacuum pump or steam distillation unit and Eco titrator.
Order no.
044989
Dispenser unit Titronic 300
13.1.2 Splash protector spare parts
11062956
Glass splash protector (EasyDist)
Order no.
11071015
Büchi Labortechnik AG
Image
Image
Plastic splash protector (EasyDist) 11072055
Glass splash protector
Plastic splash protector
Devarda splash protector
11071013
11070670
11071014
Sealing sample tube to splash protector 043068
Sealing sample tube to splash protector (acid resistant)
11063243
Steam inlet tubing to sample tube 043424
Steam inlet tubing to sample tube (750 mL) 043119
100/106 Operation Manual K-365 Dist Line
Büchi Labortechnik AG
Connecting piece
Connecting piece EPDM
Order no.
019002
11062478
Bridge splash protector to condenser
Cap nut GL14
11070620
033577
Hose barbs. set. 4 pcs, straight, GL14, FPM seal
Content: Hose barbs, cap nuts, seals.
040296
Seals, set. 10 pcs, for hose barbs GL14,
FPM, green
13.1.3 Tank spare parts
040040
Cap Set, 10 L tank
Canister 10L thin walled
Order no.
11072173
043410
Cap Set, 20 L tank
Canister 20L
11072174
043408
Tank labels 043434
Appendix | 13
Image
Image
13.1.4 Condenser spare parts
Condenser with check valve
Order no.
11072183
Image
Operation Manual K-365 Dist Line 101/106
13 | Appendix
Clamp for condenser
Rubber band
Check valve
Dist. Outlet PTFE L=300
13.1.5 Titration spare parts
Receiving vessel
Stirrer cpl.
pH electrode (refillable electrolyte) pH Electrode
Clamp cone pH-electrode
Hose Tygon Ø 8.0 x 4.8
Receiving support
Dist. Outlet PTFE L=212
102/106
Order no.
11066868
Büchi Labortechnik AG
Image
11070669
11071740
11071940
Order no.
043390
11070246
11065834
11056842
11069793
043364
11071003
11071941
Image
Operation Manual K-365 Dist Line
Büchi Labortechnik AG
Aspiration tubing receiving vessel
H₃BO₃ tubing receiving vessel
Order no.
11072589
11072637
Appendix | 13
Image
13.1.6 Sample tubes
Sample tubes micro (100 mL)
4 pcs.
Sample tubes 300 mL
20 pcs.
Sample tubes 300 mL
For sample volumes up to 200 mL or 5 g in weight
4 pcs.
Sample tubes graduated 300 mL
4 pcs.
Sample tube 500 mL
Order no.
11057442
11059690
037377
043049
026128
Sample tubes 500 mL
4 pcs.
Sample tube 750 mL including suction tube
043982
11058999
Image
13.1.7 Cable and tubing
Hose peristaltic pump out (2.5 m)
Hose clips peristaltic pump hose
Hose acid pump (2 m)
Hose clip D7.6
Operation Manual K-365 Dist Line
Order no.
11071630
043586
043588
043571
103/106
13 | Appendix
Hose to NaOH/ H₂O/H₃BO₃ pumps or drain (2 m)
Hose clip to NaOH/H₂O/ H₃BO₃ pumps or drain
Suction to tube tank (0.58 m)
Tap water hose cpl.
Cable to Metrohm Titrino Plus 877/848 titrator
Set of sealings tap water hose
Cable to SI-Analytics TitroLine Easy and TitroLine 5000 titrator
Cable to Metrohm (7-series)/Mettler (T-series) titrator
Silicone hose D6/9 L=3 m 048355
13.1.8 Other spare parts
Cap RJ-45 cap
Order no.
11055949
Büchi Labortechnik AG
Order no.
11072687
043841
043407
037780
11055333
040043
043618
043617
Image
USB cover
Level sensor capacitive
Tank level sensor (Velcro point)
Tank level sensor (Velcro strap)
Collecting pan
Pair of glass tongs
11069375
11065245
11070517
11070516
11066465
002004
104/106
13.1.9 Consumables
2% boric acid pH 4.65 +/- 0.15 with Sher indicator, 5 L
4% boric acid pH 4.65 +/- 0.15 with Sher indicator, 5 L
4% boric acid pH 4.65 +/- 0.15 with bromocresol green / methly red mixed indicator, 5 L
Sher indicator 100 mL
Order no.
11064972
11064973
11064976
003512
Operation Manual K-365 Dist Line
Büchi Labortechnik AG Appendix | 13
Ammonium phosphate monobasic, 25 g
13.1.1
0
Maintenance kit
Customer Kit BasicDist and MultiDist
Order no.
045527
Order no.
11073025
13.1.1
1
Upgrade kits
Upgrade BasicDist with aspiration pump
Upgrade Basic (base unit) with aspiration pump and I-300 Pro
Upgrade Multi (base unit) with titration vessel
Order no.
11072752
11072754
11072755
13.1.1
2
Documentation
Kjeldahl Knowledge Base (EN)
Comprehensive guide covering all aspects of the theoretical and practical knowhow.
Kjeldahl Practice Guide (EN)
Provides theoretical background information, useful hints and calculation tables for daily routine work
Kjeldahl Practice Guide (DE)
Kjeldahl Practice Guide (CN)
Order no.
11595478
11592548
11592547
11592549
Operation Manual K-365 Dist Line 105/106
Quality in your hands
BUCHI Affiliates:
Europe
Switzerland/Austria
BÜCHI Labortechnik AG
CH – 9230 Flawil
T +41 71 394 63 63
F +41 71 394 64 64 [email protected]
www.buchi.com
Italy
BUCHI Italia s.r.l.
IT – 20010 Cornaredo (MI)
T +39 02 824 50 11
F +39 02 575 12 855 [email protected]
www.buchi.com/it-it
Benelux
BÜCHI Labortechnik GmbH
Branch Office Benelux
NL – 3342 GT Hendrik-Ido-Ambacht
T +31 78 684 94 29
F +31 78 684 94 30 [email protected]
www.buchi.com/bx-en
Russia
BUCHI Russia/CIS
Russia 127287 Moscow
T +7 495 36 36 495 [email protected]
www.buchi.com/ru-ru
France
BUCHI Sarl
FR – 91140 Villebon-sur-Yvette
T +33 1 56 70 62 50
F +33 1 46 86 00 31 [email protected]
www.buchi.com/fr-fr
Spain
BUCHI Ibérica S.L.U.
ES – 08960 Barcelona
T +34 936 06 8010 [email protected]
www.buchi.com/es-es
Germany
BÜCHI Labortechnik GmbH
DE – 45127 Essen
T +800 414 0 414 0 (Toll Free)
T +49 201 747 49 0
F +49 201 747 49 20 [email protected]
www.buchi.com/de-de
United Kingdom
BUCHI UK Ltd.
GB – Suffolk CB8 7SQ
T +44 161 633 1000
F +44 161 633 1007 [email protected]
www.buchi.com/gb-en
NIR-Online
Germany
BÜCHI NIR-Online
DE – 69190 Walldorf
T +49 6227 73 26 60
F +49 6227 73 26 70 [email protected]
www.nir-online.de
America
Brazil
BUCHI Brasil Ltda.
BR – Valinhos SP 13271-200
T +55 19 3849 1201
F +55 19 3849 2907 [email protected]
www.buchi.com/br-pt
USA/Canada
BUCHI Corporation
US – New Castle, DE 19720
T +1 877 692 8244 (Toll Free)
T +1 302 652 3000
F +1 302 652 8777 [email protected]
www.buchi.com/us-en
Asia
China
BUCHI China
CN – 200233 Shanghai
T +86 21 6280 3366
F +86 21 5230 8821 [email protected]
www.buchi.com/cn-zh
India
BUCHI India Private Ltd.
IN – Mumbai 400 055
T +91 22 667 75400
F +91 22 667 18986 [email protected] www.buchi.com/in-en
Indonesia
PT. BUCHI Indonesia
ID – Tangerang 15321
T +62 21 537 62 16
F +62 21 537 62 17 [email protected]
www.buchi.com/id-in
Japan
Nihon BUCHI K.K.
JP – Tokyo 110-0008
T +81 3 3821 4777
F +81 3 3821 4555 [email protected]
www.buchi.com/jp-ja
Korea
BUCHI Korea Inc.
KR – Seoul 153-782
T +82 2 6718 7500
F +82 2 6718 7599 [email protected]
www.buchi.com/kr-ko
Malaysia
BUCHI Malaysia Sdn. Bhd.
MY – 47301 Petaling Jaya,
Selangor
T +60 3 7832 0310
F +60 3 7832 0309 [email protected]
www.buchi.com/my-en
BUCHI Support Centers:
South East Asia
BUCHI (Thailand) Ltd.
TH-Bangkok 10600
T +66 2 862 08 51
F +66 2 862 08 54 [email protected]
www.buchi.com/th-th
Middle East
BÜCHI Labortechnik AG
UAE – Dubai
T +971 4 313 2860
F +971 4 313 2861 [email protected]
www.buchi.com
Singapore
BUCHI Singapore Pte. Ltd.
SG – Singapore 609919
T +65 6565 1175
F +65 6566 7047 [email protected]
www.buchi.com/sg-en
Latin America
BUCHI Latinoamérica
S. de R.L. de C.V.
MX – Mexico City
T +52 55 9001 5386 [email protected]
www.buchi.com/es-es
Thailand
BUCHI (Thailand) Ltd.
TH – Bangkok 10600
T +66 2 862 08 51
F +66 2 862 08 54 [email protected]
www.buchi.com/th-th
We are represented by more than 100 distribution partners worldwide.
Find your local representative at: www.buchi.com