GE Aurora H2O Moisture Analyzer User’s Manual
Below you will find brief information for Moisture Analyzer Aurora H2O. The GE Aurora H2O is designed to continuously monitor the moisture content in natural gas. It utilizes a tunable diode laser absorption spectroscopy (TDLAS) method, which allows for a fast response time of less than 2 seconds after purging the sample cell. The device also includes an integrated sample system with components like a filter, pressure regulator, and optional heater. The system can directly measure moisture parameters including dew point temperature, volume ratio, and absolute humidity, all while offering both analog and digital signal transmission. The Aurora H2O also features an auxiliary analog input for connecting to a process pressure transmitter, enabling the display and transmission of the process dew point.
Advertisement
Advertisement
GE Measurement & Control Moisture Aurora H2O User’s Manual 910-284 Rev. E April 2012 GE Measurement & Control Aurora H2O Moisture Analyzer for Natural Gas User’s Manual 910-284 Rev. E April 2012 www.ge-mcs.com ©2012 General Electric Company. All rights reserved. Technical content subject to change without notice. [no content intended for this page] ii Preface Information Paragraphs • Note paragraphs provide information that provides a deeper understanding of the situation, but is not essential to the proper completion of the instructions. • Important paragraphs provide information that emphasizes instructions that are essential to proper setup of the equipment. Failure to follow these instructions carefully may cause unreliable performance. • Caution! paragraphs provide information that alerts the operator to a hazardous situation that can cause damage to property or equipment. • Warning! paragraphs provide information that alerts the operator to a hazardous situation that can cause injury to personnel. Cautionary information is also included, when applicable. Safety Issues WARNING! It is the responsibility of the user to make sure all local, county, state and national codes, regulations, rules and laws related to safety and safe operating conditions are met for each installation. Auxiliary Equipment Local Safety Standards The user must make sure that he operates all auxiliary equipment in accordance with local codes, standards, regulations, or laws applicable to safety. Working Area WARNING! Auxiliary equipment may have both manual and automatic modes of operation. As equipment can move suddenly and without warning, do not enter the work cell of this equipment during automatic operation, and do not enter the work envelope of this equipment during manual operation. If you do, serious injury can result. WARNING! Make sure that power to the auxiliary equipment is turned OFF and locked out before you perform maintenance procedures on the equipment. Qualification of Personnel Make sure that all personnel have manufacturer-approved training applicable to the auxiliary equipment. Personal Safety Equipment Make sure that operators and maintenance personnel have all safety equipment applicable to the auxiliary equipment. Examples include safety glasses, protective headgear, safety shoes, etc. Unauthorized Operation Make sure that unauthorized personnel cannot gain access to the operation of the equipment. Aurora H2O User’s Manual iii Preface Environmental Compliance Waste Electrical and Electronic Equipment (WEEE) Directive GE Measurement & Control is an active participant in Europe’s Waste Electrical and Electronic Equipment (WEEE) take-back initiative, directive 2002/96/EC. The equipment that you bought has required the extraction and use of natural resources for its production. It may contain hazardous substances that could impact health and the environment. In order to avoid the dissemination of those substances in our environment and to diminish the pressure on the natural resources, we encourage you to use the appropriate take-back systems. Those systems will reuse or recycle most of the materials of your end life equipment in a sound way. The crossed-out wheeled bin symbol invites you to use those systems. If you need more information on the collection, reuse and recycling systems, please contact your local or regional waste administration. Visit http://www.ge-mcs.com/en/about-us/environmental-health-and-safety/1741-weee-req.html for take-back instructions and more information about this initiative. iv Aurora H2O User’s Manual Contents Chapter 1. Features and Capabilities 1.1 Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.3 Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.4 System Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.5 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 1.5.1 1.5.2 1.5.3 1.5.4 1.5.5 1.5.6 1.5.7 1.5.8 1.5.9 1.5.10 1.5.11 1.5.12 1.5.13 1.5.14 1.5.15 1.5.16 1.5.17 1.5.18 1.5.19 Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Response Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Operating Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 Flowrate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 Inputs/Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 Enclosure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 Laser Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11 Weight and Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11 Hazardous Area Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11 European Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11 Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11 Sampling System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11 Wetted Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11 Chapter 2. Installation 2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13 2.2 Bill of Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13 2.3 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13 2.4 Choosing A Site for Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14 2.5 Low Voltage Directive. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16 2.6 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16 2.7 Optional Insertion Probe/Regulator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17 2.8 2.7.1 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17 2.7.2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17 2.7.3 Setting the Pressure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25 Making Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26 Chapter 3. Operation and General Programming 3.1 Using the Aurora H2O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33 Aurora H2O User’s Manual v Contents 3.2 Sample System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33 3.3 3.2.1 Startup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33 3.2.2 Shut Down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33 3.2.3 Purge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33 Keypad Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35 3.4 3.3.1 Indicator Lights. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36 3.3.2 The Magnetic Stylus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36 3.3.3 The Default Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36 3.3.4 Unlocking the Keypad. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37 3.3.5 Keypad Lock-Out Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37 3.3.6 Accessing the Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38 3.3.7 Entering Numeric Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38 3.3.8 Starting Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39 Setting Up the Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39 3.5 3.4.1 Selecting Primary Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39 3.4.2 Selecting Alt 1 and Alt 2 Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40 3.4.3 Setting Decimal Places. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40 3.4.4 Data/Scan. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40 3.4.5 Adjust . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41 3.4.6 Reverse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41 Setting Up Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41 3.6 3.5.1 Selecting an Output for Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41 3.5.2 Selecting Output Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42 3.5.3 Selecting an Output Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42 3.5.4 Changing the Upper Output Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42 3.5.5 Changing the Lower Output Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43 3.5.6 Testing the Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43 3.5.7 Trimming the Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44 Setting Up Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45 3.6.1 3.6.2 3.6.3 3.6.4 3.6.5 3.6.6 3.6.7 vi Selecting an Alarm Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45 Selecting Alarm Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46 Selecting Alarm Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46 Selecting an Alarm Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47 How the Alarm Types Work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47 Changing the Upper Alarm Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48 Changing the Lower Alarm Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48 Aurora H2O User’s Manual Contents Chapter 4. Programming Advanced Features 4.1 Comm Port Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49 4.2 4.1.1 Selecting a Comm Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49 4.1.2 Setting the Baud Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49 4.1.3 Setting Parity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50 4.1.4 Selecting Protocol. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50 4.1.5 Setting the Network ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50 Adjust Offset Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51 4.3 4.2.1 Adjusting the PPMv Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51 4.2.2 Adjusting the Smoothing Filter Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51 4.2.3 Setting the Dew Point Calculation Method. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52 4.2.4 Adjusting the Enh Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52 Set Up the Background Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53 4.4 4.3.1 Selecting the Type of Gas. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53 4.3.2 Setting the Gas Composition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54 4.3.3 Setting the Z Factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54 4.3.4 Adjusting the Gas Offset. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55 Clock Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55 4.5 4.4.1 Resetting the Hour . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55 4.4.2 Resetting the Minutes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56 4.4.3 Resetting the Month . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56 4.4.4 Resetting the Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56 4.4.5 Resetting the Year . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57 Pressure Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57 4.6 4.5.1 Setting the Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57 4.5.2 Changing the Constant. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57 4.5.3 Editing Pressure Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58 Regional Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59 4.7 4.6.1 Setting the Country Code. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59 4.6.2 Setting the Decimal Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60 4.6.3 Setting the Date Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60 4.6.4 Setting the Unit System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60 User Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61 4.8 Service Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63 4.9 Aurora H2O Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64 4.9.1 4.9.2 4.9.3 4.9.4 4.9.5 Checking the ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64 Checking the System Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64 Checking the Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65 Checking the Gas Composition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65 Checking the Alternate Gas Composition. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66 Aurora H2O User’s Manual vii Contents 4.10 Custom Gas Composition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67 4.11 Locking/Unlocking the Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68 Chapter 5. AuroraView Interface Software 5.1 Capabilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69 5.2 Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69 5.3 Installing AuroraView. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70 5.4 Starting AuroraView . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77 5.5 Using the Main Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .79 5.6 Datalogging with AuroraView . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85 5.7 Working with Trend Plots, Trend Tabular Data & Scan Plots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86 Chapter 6. Maintenance 6.1 Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .95 6.2 Recommended Factory Verification Period . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .95 6.3 Cleaning the Mirror . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .95 6.4 Replacing the Filter Element . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101 Chapter 7. Troubleshooting 7.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105 7.2 Blank Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105 7.3 Display Dim or Hard to Read . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105 7.4 Status Messages and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105 7.5 No Flow Measurement Indicated on Aurora H2O Measurement Cell Outlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 7.6 Verifying Aurora H2O Performance in the Field. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 7.7 7.6.1 Using A Portable Hygrometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 7.6.2 Using a Moisture Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108 Background Selection Lockout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109 Appendix A. MODBUS RTU/RS485 Communications A.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111 A.2 Modbus Functions and Data Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111 A.3 Modbus Registers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112 A.4 Modbus System Status Codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116 viii Aurora H2O User’s Manual Chapter 1. Features and Capabilities Chapter 1. Features and Capabilities 1.1 Overview GE’s Aurora H2O Moisture Analyzer for Natural Gas makes it possible for natural gas processing and transportation facilities to monitor moisture content in real-time with high precision and reliability. The Aurora H2O analyzer uses tunable diode laser absorption spectroscopy (TDLAS) to measure moisture in natural gas at the speed of light. The analyzer is suitable for installation in hazardous areas and operates over a wide range of environmental conditions. Aurora’s fast response quickly alerts and documents when moisture concentrations are out of compliance. Once process upsets are corrected and the gas dries out, the fast response quickly enables natural gas to be cleared for entry into the “energy grid”. CLASS 1 LASER PRODUCT WARNING! Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous laser exposure. 1.2 Features • Optical response: <2 seconds once flow cell is purged. • No cross sensitivity to glycols or amines. • Direct readout in lbs/mmscf, mg/m3 or ppm. • Reads process pressure dew point (with user programmable constant or live auxiliary input for process pressure). • Turnkey sampling system specifically designed for natural gas applications ensures measurement integrity. • Magnetic Stylus enables through-glass programming – hot permit not required to field program. • Explosionproof/Flameproof design • 4-20 mA signals and RS-232/485 MODBUS RTU for connection to SCADA or plant monitoring system. • Supplied with AuroraView software for remote configuration, data logging and data retrieval capacity. • NIST traceable calibration. • In conformance with IEC 60825-1 Edition 2.0, Safety of Laser Products. Aurora H2O User’s Manual 1 Chapter 1. Features and Capabilities 1.3 Theory of Operation The GE Aurora H2O (Tunable Diode Laser Absorption Spectroscopy) hygrometer for natural gas is a system designed to continuously monitor the moisture content in natural gas. It fundamentally measures the partial pressure of water vapor (water in the gas state), and with the simultaneous measurement of pressure and temperature, provides readings and both analog and digital signal transmission of user-selectable moisture parameters including dew point temperature, volume ratio and absolute humidity (lbs/MMSCF or mg/m3) in addition to temperature and pressure. The Aurora H2O is supplied with an integrated sampling system which includes an optional pipe-mounted liquid separator and pressure reducing valve/regulator, and the following components mounted within a stainless steel enclosure: isolation valve, coalescing filter, coalescer flow control valve, sample cell control valve, second pressure reducing valve/regulator, and flow indicator (rotameter), as well as an optional electrical heater and thermostat. The fundamental water vapor pressure measurement is based on the Beer-Lambert Law: Io A = In --- = SLN I where: A = Absorbance I = Light intensity transmitted through a sample gas Io = Incident light intensity S = Absorption coefficient* L = Absorption path length (a constant) N = Concentration of the water vapor in the absorption cell * The absorption coefficient is a constant for a specific gas composition at a given pressure and temperature. The concentration of the water is directly related to the partial pressure. At certain specific frequencies, light energy will be absorbed by water molecules. As the concentration of water increases, the absorption also increases. The Aurora H2O sweeps the diode laser output across a narrow band in the near infrared spectrum and, by measuring the light intensity with a photo detector, is able to provide a direct indication of the partial pressure of water. The partial pressure, multiplied by 106 and divided by the total pressure, yields the volume ratio in ppmv (parts per million by volume). The laser diode is housed in a hermetically sealed and dry housing. The light is transmitted through a window made of proprietary transparent material. The light travels through a stainless steel cell and is reflected off a gold-plated mirror and returned to a photo detector, where the light intensity is measured. 2 Aurora H2O User’s Manuall Chapter 1. Features and Capabilities 1.3 Theory of Operation (cont.) Since only light comes in contact with the sample of natural gas, and all of the wetted materials are made of non-corrosive and inert materials, this technology does not exhibit the drift associated with gas contacting sensor-based hygrometers. The diode laser emits low energy light, and therefore the system will not ignite the gas. The complete system is rated explosion-proof, and the Aurora H2O provides very fast response time. Once the absorption cell (see Figure 1 below) is purged, the response time is a matter of a few seconds. Photodetector Optical Window Mirror Gas Inlet Gas Outlet Stainless Steel Absorption Cell Tunable Diode Laser Hermetically Sealed Housing Figure 1: Laser Absorption Cell, Basic Elements The control of the laser, power supply and signal conditioning circuitry are housed in an explosion-proof transmitter enclosure. A backlit, three-parameter, LCD display provides digital indication of user-programmable parameters. The Aurora H2O has three user-programmable (4-20 mA) analog outputs and two programmable digital ports that may be configured as either RS-485 or RS-232 with Modbus protocol. The analyzer has an auxiliary analog input (4-20 mA) input that is used for connection to an optional process pressure transmitter. Measurement of the process pressure enables the Aurora H2O to display and transmit the process dew point. AuroraView software is provided and enables remote readings, programming data logging and data logging with a personal computer. The Aurora H2O hygrometer is calibrated against an NIST (or other national metrological institute) traceable reference dew/point generator and hygrometer. Each system is supplied with a certificate of calibration with functional test data. Aurora H2O User’s Manual 3 Chapter 1. Features and Capabilities 1.4 System Components Note: Refer to Table 1 on page 6 for part descriptions. 1 3 2 12 13 4 5 14 7 6 11 8 16 9 10 15 Figure 2: Aurora H2O System Assembly with Optional USA/CAN Heater 4 Aurora H2O User’s Manuall Chapter 1. Features and Capabilities 1.4 System Components (cont.) Note: Refer to Table 1 on page 6 for part descriptions. 1 2 3 12 13 4 5 14 7 6 11 8 16 9 10 15 Figure 3: Aurora H2O System Assembly with Optional EU/ATEX Heater Aurora H2O User’s Manual 5 Chapter 1. Features and Capabilities 1.4 System Components (cont.) Table 1: Aurora H2O System Parts List Description No. No. 6 Description 1 Conduit I/O 9 Process or Purge Gas Selector 2 Display and Magnetic Stylus Keypad 10 Purge Gas Inlet Needle Valve 3 Wiring Terminals 11 Absorption Cell 4 Isolation Valve (ball valve) 12 Temperature Sensor 5 Coalescing Filter 13 Pressure Sensor 6 Coalescing Filter Drain & Fast Loop Vent 14 Rotameter 7 Pressure Regulator w/ 0-10 psig outlet pressure gauge 15 Optional Heater, thermostat, junction box 8 Flow Control Needle Valve 16 Magnetic Stylus Aurora H2O User’s Manuall Chapter 1. Features and Capabilities 1.4 System Components (cont.) IMPORTANT: The filter installed within the Aurora is designed to function as a secondary filter. A primary filter or filter train should be used upstream of the analyzer. Do not sample directly from the process gas to the Aurora without the use of a primary filter and pressure reduction. The pressure inlet to the Aurora sample inlet should not exceed 500 psig. 2.25 (57.1) 2 1.20 - 1.75 (30.48 - 44.45) 3 3.42 (86.8) 1 5.60 (142.2) 4 Overall Length (See Detail A) 1 2 3 4 Probe/filter/regulator assembly Pressure Gauge, 0 to 600 psig 2-way ball valve Tube fitting NOTES: 1. Sample Process Connection: 3/4" NPTM 2. Sample Outlet Connection: 1/4" Compression Fittings 3. Dimensions are in inches (millimeters). 3/4" NPTM Detail A Probe Length (See Detail A) 0.92 01 02 03 04 05 06 Part Number 750-633 750-634 750-635 750-633 750-634 750-635 Probe Length 4.00" 7.00" 9.00" 4.00" 7.00" 9.00" Overall Length 11.80" 14.80" 16.80" 11.80" 14.80" 16.80" Figure 4: Sample System (ref. dwg #733-737) For natural gas, a pipeline insertion membrane filter is recommended as shown in Figure 4 above. The installation of the insertion filter is described in this manual. The insertion filter also is equipped with a pressure regulator which functions to reduce the pressure. If it is not possible to install the insertion filter, consult GE application engineers for information about a sample conditioning system. Aurora H2O User’s Manual 7 Chapter 1. Features and Capabilities 1.4 System Components (cont.) THERMOSTAT BROWN J1 16 AWG BROWN WIRE (2") STRIP 1/4" LINE NEUTRAL 230 VAC 120 VAC L1 L L2 N G G J1 GROUND BLUE ENCLOSURE WALL GEMT JUNCTION BOX (SHOWN FOR REFERENCE) GREEN/YELLOW NOTES: 1. HEATER - USE GE PART NUMBER FROM 227-102 705-1238-01 AND -06. 2. DO NOT "DAISY-CHAIN" HEATER WIRING TO OR FROM AURORA ELECTRONICS Figure 5: USA/CAN Heater Wiring Diagram ATEX BARTEC JUNCTION BOX GREEN/YELLOW CUSTOMER GROUND CABLE FROM ATEX HEATER CUSTOMER CABLE BROWN LINE 1 RED PLUG BLUE NEUTRAL CUSTOMER NEUTRAL CUSTOMER LINE NOTE: 705-1238-02 ATEX HEATER. Figure 6: EU/ATEX Heater Wiring Diagram 8 Aurora H2O User’s Manuall Chapter 1. Features and Capabilities 1.5 Specifications 1.5.1 Power Analyzer: 100-240 VAC, 50-60Hz, 10W 18-32 VDC (24 VDC nominal), 10W Optional Electrical Heater: 120VAC, 120W 230VAC, 75W 1.5.2 Range Part Per Million by Volume 5 to 5000 ppmv Dew/Frost Point* –65.5°C to –2.6°C (–85.9°F to 27.3°F) Process Dew/Frost Point* Process or equivalent dew point/frost point by calculation with process pressure signal (4-20 mA) or constant Absolute Humidity 3.8 to 3803 mg/m3 (0.24 to 237 lbs/MMSCF) *Readings below 0°C (32°F) are in “frost point” temperature, and above 0°C (32°F) are in “dew point” temperature. 1.5.3 Accuracy Parts per Million by Volume ±2% of reading in ppmv or 4 ppmv (Accuracy of other parameters derived from ppmv) 1.5.4 Response Time Optical Response <2 seconds System Response The system response is dependent upon the length of sample tubing, sample system components, flowrate and pressure, as well as the change in moisture concentration. 1.5.5 Operating Pressure Operating Sample Cell Pressure: 69 to 172 kPa (10 to 25 psia) Maximum Pressure 1380 kPa (200 psi) Process Pressure 10,342 kPa (1500 psig) maximum [Higher pressure available using additional sampling system components] Aurora H2O User’s Manual 9 Chapter 1. Features and Capabilities 1.5.6 Temperature Operating –20° to +65°C (–4° to 149°F) Storage –20° to +70°C (–4° to +158°C) Optional Heater/Thermostat Setpoint 20°±5°C (68°±9°F) for U.S. and Canada 10°±5°C (50°±9°F) for EU and elsewhere 1.5.7 Flowrate Sample Cell Flowrate 10 to 60 SLH (0.4 to 2 SCFH); 30 SLH (1 SCFH) nominal Coalescer By-pass Fast Loop 5 to 10X of flowrate through sample cell 1.5.8 Display Backlit LCD, three programmable simultaneous parameters, alphanumeric status and diagnostic display, LEDs for power, laser temperature stability, keypad lockout 1.5.9 Inputs/Outputs Analog Outputs Three programmable 0/4-20 mA, 500 Ω maximum load Analog Input Loop-powered 4-20 mA input for remote pressure transmitter (Aurora H2O supplies 24 VDC) Digital Interface Two programmable digital communications ports RS-232 and RS-485 with multidrop capability and assignable address, Modbus RTU protocol 1.5.10 Enclosure Ingress Protection IP-66 1.5.11 Laser Certification Class 1 product, conforms to IEC 60825-1 Edition 2.0, Safety of Laser Products WARNING! Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous laser exposure. 10 Aurora H2O User’s Manuall Chapter 1. Features and Capabilities 1.5.12 User Interface Programmable “through-the-glass” via magnetic stylus 1.5.13 Weight and Dimensions Net Weight 45 kg (100 lb) Overall Dimensions 34”H x 18”W x 14”D (87 cm x 46 cm x 36 cm) [see drawings] 1.5.14 Hazardous Area Certification USA/Canada Explosion-Proof for Class I, Division 1, Groups C&D EU and Elsewhere ATEX and IECEx: Ex de IIB T6; Tamb = –20°C to +65°C. Flameproof with increased safety compartment 1.5.15 European Certification CE Complies with EMC Directive 2004/108/EC, Low Voltage Directive 2006/95/EC and Pressure Directive 97/23/EC for DN<25 1.5.16 Calibration Recommended factory Verification/Service Five years 1.5.17 Software Displays all key parameters. Provides time-base graphing, data logging ability, export data as ASCII text. Software has lockout/passcode. 1.5.18 Sampling System Integrated sampling system to filter out physical contaminants, remove glycol carry over, control temperature condition, regulate pressure and flow rate. Optional thermostatically controlled heater. 1.5.19 Wetted Components 316/316L stainless steel for tubing and fittings. Other wetted components such as the optical window and mirror are proprietary materials and are compatible with natural gas and typical contaminants found in natural gas. Other sample system components wetted parts include PTFE, PFA, Inconel, Hastelloy, PVDF, glass, Viton. Aurora H2O User’s Manual 11 Chapter 1. Features and Capabilities [no content intended for this page] 12 Aurora H2O User’s Manual Chapter 2. Installation Chapter 2. Installation 2.1 Introduction The Aurora H2O analyzer provides direct indication of moisture concentration in natural gas. Temperature and pressure sensors are used to provide high precision enhancement. It may be installed in a wide variety of environmental conditions, and meets the requirements for operation in hazardous areas. Be sure that the ambient temperature is at least 10°C higher than the maximum dew/frost point temperature you expect to measure. This will ensure that you will not have liquid condensation in the sample transport line or the Aurora H2O sampling system components. Heat tracing the sample line will aid in elevating the sample temperature above the dew point. An optional heater installed within the enclosure is also available. 2.2 Bill of Materials The following items should have been received with your shipment: • Aurora H2O Unit • Aurora H2O Calibration Data Sheet • Aurora H2O User’s Manual on CD ROM • Maintenance/Accessories Kit • AuroraView Software on CD ROM • Optional: Pipeline Insertion Filter/Regulator Assembly 2.3 Unpacking The Aurora H2O is shipped in a packing box with a plywood base (see Figure 7 to the right), and the Aurora H2O analyzer is secured to the plywood base with mounting bolts. Transport the shipping package with the plywood base on the bottom, and according to the warning labels on the exterior packaging. Open the packing box from the top, and remove the foam packing material. Collect any separate components such as the AuroraView CD, the User Manual CD, and other items contained in the shipment. Remove the mounting studs at the bottom of the enclosure. Using two people, lift the Aurora H2O from the bottom of the enclosure and from the top of the enclosure where the analyzer electronics is located. A typical Aurora H2O weighs approximately 100 pounds (45 kg). Use a proper lifting technique to avoid injury. Check all the received components and record the model numbers and serial numbers for your records. If anything is missing, contact GE immediately. Figure 7: Unpacking the Aurora H2O Aurora H2O User’s Manual 13 Chapter 2. Installation 2.4 Choosing A Site for Installation You should have discussed environmental and installation factors with a GE Sales, Application or Service Engineer by the time you receive the analyzer. Before installing the analyzer, read the guidelines below on installation recommendations: 1. Choose an installation site for the Aurora H2O analyzer as close to the actual sample point (sample take-off point) as possible, to minimize transport time to the analyzer. 2. Avoid unnecessarily long lengths of sample transport tubing to minimize transport time to the analyzer. 3. Avoid dead-legs in the sample transport tubing to minimize the possibility of liquid build-up. 4. Use stainless steel tubing. Avoid using copper tubing, as the water molecule has greater absorption capabilities for copper compared to stainless steel. Avoid rubber tubing at all costs, as water molecules interact with the rubber, and ambient moisture can permeate through the tube wall into the sample gas. 5. Mount the Aurora H2O analyzer at grade, or at a location that is easily accessible for maintenance (on a platform or other structure). 6. Be sure that the ambient temperature is at least 10°C higher than the maximum dew/frost point temperature you expect to measure. This will ensure that you will not have liquid condensation in the sample transport line nor in the Aurora H2O. Heat tracing the sample line will aid in elevating the sample temperature above the dew point. An optional heater installed within the enclosure is also available. A typical Aurora H2O system for monitoring moisture off a natural gas pipeline is shown in Figure 8 on page 15. 14 Aurora H2O User’s Manuall Aurora H2O User’s Manual Input/Output Conduit entering 2nd of 3 inlets. Note: There is no orientation preference on vertical pipe runs. Power Conduit entering 1 of 3 inlets. Sample Transport Line Minimize distance. Install at a 12 o'clock, or optionally 3 o'clock or 9 o'clock position, on a horizontal pipe run. Optional Insertion Regulate Sample Tap Tie to Vent/Flare (Do not restrict.) Conduit or Armored Cable suitable for area rating Power Input/Output Chapter 2. Installation 2.4 Choosing A Site for Installation (cont.) Figure 8: Aurora H2O Analyzer Monitoring Moisture Off a Natural Gas Pipeline 15 Chapter 2. Installation 2.5 Low Voltage Directive To comply with the Low Voltage Directive, you must install a switch or circuit breaker on the input power line. For greatest safety, locate the circuit breaker or power switch near the electronics console. IMPORTANT: Installation must be done in accordance with the National Electrical Code, the Canadian Electric Code, and/or any other applicable local codes. 2.6 Mounting Use the four mounting tabs to mount the Aurora H2O System Assembly in the desired location (see Figure 31 on page 31). IMPORTANT: The Aurora H2O should only be mounted vertically. 16 Aurora H2O User’s Manuall Chapter 2. Installation 2.7 Optional Insertion Probe/Regulator 2.7.1 Description For natural gas applications that may have entrained particulate and liquid contaminants (especially TEG carry-over from TEG dryers), GE recommends the use of a Insertion Probe/Regulator to serve as the first part of particulate and condensate filtration where the gas is sampled off the pipe. The device combines the features of a sample tap, a membrane filter and an integral pressure regulator (0-500psig outlet), adjustable at the sample take-off point. The sample tap housing includes a foot-valve so that, once installed, the assembly can be removed from a line under pressure for membrane filter replacement as needed. Figure 9: Insertion Probe/Regulator Note: A 29/32” (0.907 inch, 23.1 mm) minimum bore-through clearance is needed on the customer nozzle for insertion of a 3/4” NPT insertion membrane liquid separator/pressure regulator. 2.7.2 Installation Note that the Optional Insertion Probe/Regulator can be installed only on depressurized lines. There is no hot-tap installation for this device. This procedure is broken down into two steps: • Installing the Housing • Installing the Insertion Probe/Regulator Assembly Aurora H2O User’s Manual 17 Chapter 2. Installation 2.7.2a Installing the Housing The housing, as shown in Figure 10 below, is supplied with a Locking Mechanism to ensure that housing can only be removed intentionally. Vertical Set Screws Slots Locking Mechanism Horizontal Set Screw Horizontal Set Screw Foot Valve Locking Mechanism Figure 10: Installing the Housing and Locking Mechanism 1. Turn the locking mechanism counter-clockwise until it is at its extreme upper position. Apply thread sealant to the threaded area below vertical slots in the threads. Do not allow the thread sealant to enter the slot, as it may interfere with the mechanism. Thread Sealant Figure 11: Applying Thread Sealant 18 Aurora H2O User’s Manuall Chapter 2. Installation 2.7.2a Installing the Housing (cont.) 2. Confirm that the pipeline has been depressurized. Insert the housing into the pipeline through a ¾” NPTF thread-o-let (The minimum ID of the thread-o-let is 0.91”). Housing ¾” NPTF Thread-O-Let Figure 12: Inserting Housing into Pipeline 3. Using a wrench on the wrench flats, turn the housing until it is secure and sealed. This will require between three and five turns. DO NOT OVERTIGHTEN! The housing may be damaged if it is over-tightened, causing the housing to deform. Wrench Flats 180° apart Figure 13: Tightening the Housing Aurora H2O User’s Manual 19 Chapter 2. Installation 2.7.2a Installing the Housing (cont.) 4. Turn the locking mechanism clockwise until it first touches the top of the thread-o-let. 5. Turn the locking mechanism counterclockwise until the Allen screw is aligned with the thread slot. Using a 1/8” Allen wrench, tighten the Allen screw until its tip is tight against the slot. DO NOT OVERTIGHTEN THE ALLEN SCREW as this will cause the housing wall to be indented. Figure 14: Aligning and Tightening the Allen Screw 6. Using a 3/32” Allen wrench, tighten the Allen screws on the locking mechanism's surface until their tips are firmly set into the thread-o-let’s upper surface. Figure 15: Tightening the Allen Screws The housing is now installed. The locking mechanism should prevent the housing from becoming unintentionally unscrewed from the thread-o-let. The pipeline my now be pressurized. 20 Aurora H2O User’s Manuall Chapter 2. Installation 2.7.2b Installing the Insertion Membrane Liquid Separator/Pressure Regulator Assembly 1. Confirm that the allen cap screw head’s hex cavity is clean and free from foreign material. Confirm that the allen cap screw that holds the membrane ferrule is torque-wrench tight. The torque value should be 10 inch-lbs. If the allen cap screw is only hand-tight, the screw may protrude excessively, which could result in foot valve actuation when the probe is installed into the housing. Allen Cap Screw Figure 16: Checking/Tightening the Allen Cap Screw 2. Rotate the pressure adjustment screw on the pressure regulator fully counterclockwise until it rotates freely. Close the isolation ball valve. Pressure Adjustment Screw Pressure Gauge Isolation Ball Valve Figure 17: Rotating the Pressure Adjustment Screw Aurora H2O User’s Manual 21 Chapter 2. Installation 2.7.2b Installing the Insertion Membrane Liquid Separator/Pressure Regulator Assembly (cont.) 3. Position the membrane end of the probe above the installed housing. Slowly lower the probe into the housing. Avoid membrane contact with the upper section of the housing. DO NOT APPLY DOWNWARD FORCE. The probe should slide easily into the housing. Lower the probe only enough to thread the insertion nut one complete turn of thread engagement. Insertion Nut Membrane Housing First Vertical Slot Membrane Spring Guard Insertion Washer Second Vertical Slot Housing Figure 18: Installing the Probe 4. Thread the insertion nut down by hand, lowering the probe until the insertion washer pins slide to the bottom of the first vertical slot. Note: The threaded nut on the housing ensures that if all other safety procedures are disregarded, it is mechanically impossible to remove the probe. Insertion Nut Pin Housing Insertion Washer Figure 19: Threading the Insertion Nut 22 Aurora H2O User’s Manuall Chapter 2. Installation 2.7.2b Installing the Insertion Membrane Liquid Separator/Pressure Regulator Assembly (cont.) 5. Rotate the probe counterclockwise until the pins are to the far right in the horizontal slot. At this point, the probe is sealed against the housing interior wall. The pins will be in the middle of the second vertical slot. Figure 20: Rotating the Probe 6. Loosen the insertion nut until it is above the top of the second vertical slot. The probe should not rise to the top of the second vertical slot. If the probe rises in the slot, the foot valve o-ring may have been damaged or attacked by the process. Note: The threaded nut on the housing ensures that if all other safety procedures are disregarded, it is mechanically impossible to remove the probe. Perform the next step regardless of the status of the foot valve o-ring. Figure 21: Loosening the Insertion Nut Aurora H2O User’s Manual 23 Chapter 2. Installation 2.7.2b Installing the Insertion Membrane Liquid Separator/Pressure Regulator Assembly (cont.) 7. Tighten the insertion nut by hand until it is against the insertion washer again. Using a wrench, tighten the insertion nut against the insertion washer so that the pins are at the bottom the second vertical slot. At this point the foot valve opens and the insertion process is complete. Figure 22: Tightening the Insertion Nut 24 Aurora H2O User’s Manuall Chapter 2. Installation 2.7.3 Setting the Pressure Note: Perform these steps only after the entire system has been plumbed, including the Aurora H2O. 1. Ensure the inlet sample gas isolation valve on the Aurora H2O sample system is closed. Open the isolation ball valve at the Optional Pipeline Insertion Membrane Liquid Separator/Pressure Regulator. 2. Turn the pressure adjustment screw clockwise to increase the pressure. The optional pipeline insertion membrane liquid separator/pressure regulator is the FIRST STAGE pressure reduction of the system when used. Depending upon your source pressure, you should step down the pressure to a value in the range given in Table 2 below. Table 2: Pressure Settings Source Pressure Outlet Pressure Setting 750 psig < source < 1500 psig 400-500 psig 500 psig < source < 750 psig 300 psig < 500 psig 50% of the average source pressure Pressure Adjustment Screw Lock Nut Pressure Gauge Figure 23: Turning the Pressure Adjustment Screw 3. Tighten the lock nut down to the top of the pressure regulator to avoid future possible changes in pressure regulator setting, once the pressure is set. Aurora H2O User’s Manual 25 Chapter 2. Installation 2.8 Making Electrical Connections Refer to Figure 32 on page 32 for wiring connections. 1. Aurora H2O has three ¾” NPT conduit inlet ports for power and I/O. These will normally be shipped plugged from the factory. Follow the applicable wiring code and requirements for wiring the unit. Figure 24: Conduit Inlet Ports Note: Use one conduit inlet for power. Use the two other conduit inlets for input/output as needed. All unused conduit inlet ports should be sealed with suitable blanking elements. 2. Select one conduit for inlet power to the Aurora H2O based on your configuration. The Aurora H2O comes with a universal power supply, or optionally, as a 24VDC powered unit. Remove the wiring cover to view the wiring terminal block. Figure 25: Wiring Terminal Blocks 26 Aurora H2O User’s Manuall Chapter 2. Installation 2.8 Making Electrical Connections (cont.) Note: Supply connection wiring shall be rated at least 10°C above the rate maximum service temperature of 85°C, be stripped back 5/16 in. (8 mm) and torqued to a minimum of 4.4 in. lb. (0.5 Nm). 3. Run the AC power connections to the Power Terminal Block shown in Figure 26 below. It is recommended to use 12-18 AWG (3.3 - 0.82 mm2) power wiring. Figure 26: Power Terminal Block 4. Use wiring conduit runs, separate from the Aurora H2O main power, for all I/O (Input/Output) leads. Wire up to three 4-20mA outputs to the terminals labeled A, B, and C. The three analog outputs A, B and C (0-20mA or 4-20mA) are internally powered by the Aurora H2O. Use shielded 18-22 AWG (0.82–0.33 mm2) twisted pair wire, and ground the shield at one end only. Wire up digital communications to Port 1 and/or Port 2 as labeled. 5. Either digital port may be configured for RS-232 or RS-485. Port 1 is designated as “SCADA.” Port 2 is designated as “SERVICE.” For operation on RS-485, 2-wire, half-duplex bus, attach the RS-485(+) to (+), and the RS-485(–) to (–). Make no connection to RTN. Note: For an RS-485 Multidrop Network, a terminating resistor must be installed across the Aurora H2O RS-485 terminals, or an internal terminating resistor can be applied. See below. • When using the Aurora H2O in RS-485 mode, and to prevent signal reflections on the high-speed RS-485 connections, it is recommended that the far end of the RS-485 lines be terminated properly. The termination can be accomplished in one of two ways: — Connect 120Ω ¼W leaded resistors across the + and – terminals of ports 1 and 2 (both ports or whichever one will be in use), or — Using long-nose pliers, move jumpers J15 and J16 from pins 2 and 3 (default setting from factory) to pins 1 and 2 (see Figure 27 on page 28). J16 is the termination for port 1 and J15 is the termination for port 2. It is also recommended that basic ESD precautions such as grounded wrist straps be used for this procedure. Aurora H2O User’s Manual 27 Chapter 2. Installation 2.8 Making Electrical Connections (cont.) Multi-drop RS-485: For multiple Aurora H2O units connected in daisy-chain fashion to the RS485 interface, it is important that the farthest unit away from the transmitting device be the only unit incorporating any termination (see Figure 27 below). All other units must have jumpers J15 and J16 in positions 2 & 3 (default setting from the factory). For more details on RS-485 wiring or operation, refer to TIA/EIA-485-A Specification. 13 J16 2 13 J15 2 Figure 27: RS-485 Mode Termination - Install Jumpers at J15 and J16, Pins 1 and 2 Note: Terminations are NOT required when using ports in RS-232 mode. • For operation on RS-232, connect RS-232(TXD) to (+), RS-232(RXD) to (–), and RS-232(GND) to RTN (see Figure 28 below). Figure 28: Input/Output Connections 28 Aurora H2O User’s Manuall Chapter 2. Installation 2.8 Making Electrical Connections (cont.) 6. For connection to a PC to interface with AuroraView software, you may use the supplied 704-688 cable (RS-232 w/ SUB-D-9 connector to tinned leads). Wire the cable as follows: Color Code Aurora H2O Terminal White Tx + Red Rx – Green Ground RTN Note: The default configuration is as shipped: BAUD Rate: Parity: ID Note: 115,200 Even 1 for Port 1, 2 for Port 2 7. Use a separate wiring conduit run for any 4-20mA pressure transmitter input. This input is used when a live input pressure reading for the main process pressure is desired, to determine an equivalent dew point by the Aurora H2O analyzer. Wire the 4-20mA pressure transmitter to the Pressure Transmitter terminal block (see Figure 29 below). The Aurora H2O supplies 24VDC for use with a loop-powered, 2-wire pressure transmitter. Figure 29: Pressure Transmitter Connections IMPORTANT: Use of an external pressure transmitter is not covered by the Aurora H2O hazardous area certifications. The external pressure transmitter should be suitably rated for the area classification. Its associated wiring should be done in accordance with local codes and regulations, and suitably rated for the area classification. 8. If the Aurora H2O has been supplied with an optional electrical heater, there are two possible configurations: USA/CAN or EU. Connect the AC power using a separate conduit from the power for the Aurora H2O analyzer. The heater is equipped with a thermostat preset to 25°C (77°F) nominal. Use 12-18 AWG (3.3 - 0.82 mm2) wires. The heater terminals are located within a junction box (item 15 in Figure 2 on page 4 or Figure 3 on page 5). Aurora H2O User’s Manual 29 Chapter 2. Installation 2.8 Making Electrical Connections (cont.) 9. Lastly, the Aurora H2O analyzer requires a connection to ground from the electronics explosion-proof/flame-proof enclosure. There are two external ground connections available for the user (on the left and right hand sides of the enclosure). Wire this connection to earth ground, local to the Aurora H2O analyzer installation site. Figure 30: Earth Ground Connection 30 Aurora H2O User’s Manuall Chapter 2. Installation 15.66 1.622 13.08 .125 9.29 TYP 3.18 TYP 33.12 24.26 24.91 26.16 .95 18.31 Aurora H2O User’s Manual Figure 31: Aurora H2O Outline and Mounting (ref. dwg #712-1456) 10.92 10.24 31 Chapter 2. Installation L1 L2/N G 50-60Hz 0.5A 100-240 VAC + 2 4 V [OUTPUT] 4-20mA RS-485 A B C PORT 1 PORT 2 + – + – + – + – + – AC OPTION SHOWN DC WIRING + – SEE NOTE 2 SEE NOTE 1 CUSTOMER GROUND I/O CONNECTIONS SEE NOTE 1 Figure 32: Aurora H2O Electronic Assembly Wiring Diagram (ref. dwg #702-8976) GROUND BAR NOTES: 1. I/O CONNECTIONS AND PRESSURE TRANSMITTER WIRE GAUGE RANGE 12-24 AWG. 2. Aurora H2O User’s Manual AC AND DC CONNECTION WIRE GAUGE RANGE 12-18 AWG. 32 Chapter 3. Operation and General Programming Chapter 3. Operation and General Programming 3.1 Using the Aurora H2O Follow the information in this chapter to operate the Aurora H2O system. CLASS 1 LASER PRODUCT WARNING! Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous laser exposure. 3.2 Sample System See the instructions below and Figure 33 on page 34 to operate the Aurora H2O sample system. 3.2.1 Startup 1. Start with all valves in closed position and the pressure regulator turned fully counter-clockwise. 2. Turn the Sample/Purge 3-way ball valve to point towards the sample needle valve. 3. Sample Inlet gas pressure should be <400 psig (2760 kPa). 4. Confirm that there are no pressure restrictions downstream of the sample outlet flowmeter. 5. Open the inlet isolation ball valve. 6. Crack the bypass valve ¼ turn to establish coalescer bypass flow (fast loop). Establish a flow rate of approximately 10x the sample flow through the bypass (10 SCFH/5 LPM nominal). The gas from this vent can be piped to a process or flare. 7. Turn the pressure regulator clockwise until the pressure gauge reads approximately 3-5 psig. 8. Open the sample needle valve until the flowmeter reads 30 SLPH (1 SCFH). 3.2.2 Shut Down 1. Close the inlet isolation ball valve. 2. Turn the pressure regulator fully counter-clockwise. 3. Close the sample needle valve. 4. Close the bypass needle valve. 3.2.3 Purge 1. Hook up the purge gas. 2. Regulate the pressure externally to 3-5 psig. 3. Turn the sample/purge 3-way ball valve towards the purge inlet. 4. Crack the purge inlet needle valve to set the to 30 SLPH (1 SCFH). Aurora H2O User’s Manual 33 Chapter 3. Operation and General Programming 3.2 Sample System (cont.) INLET ISOLATION BALL VALVE PRESSURE REGULATOR SAMPLE NEEDLE VALVE BYPASS NEEDLE VALVE SAMPLE/PURGE 3-WAY BALL VALVE Figure 33: Aurora H2O Sample System Quick Startup Guide (ref. dwg #902-004) 34 Aurora H2O User’s Manual Chapter 3. Operation and General Programming 3.3 Keypad Features Menu Power Indicator Main Display Fault Indicator Laser Indicator Information Indicator Keypad Lock Indicator Cancel Enter Up, Left, Right, Down Arrows Figure 34: Aurora H2O Keypad The Aurora H2O has seven keys: a Menu key, four Arrow keys, a Cancel key, and an Enter key. • Use the Menu key to open the main menu on the display. • Use the Arrow keys to navigate among menu choices and to increment/decrement numeric entries. • Use the Cancel • Use the Enter key to cancel a numeric entry change, or exit a menu. key to accept a numeric entry or select a menu option. Aurora H2O User’s Manual 35 Chapter 3. Operation and General Programming 3.3.1 Indicator Lights If the Fault Indicator is lit, an instrument fault is detected. A message will be displayed in the Main Display, top/right. If the Information Indicator is lit, the instrument is still operating, but a message will appear in the Main Display top/right, with information about the instrument. The Keypad Lock Indicator will be lit if either: A) the Keypad Lock-Out Switch, internal to the instrument, has been engaged, or B) the instrument keypad has not been used for a period of several minutes, engaging a software feature to lock-out inadvertent key usage. Type (B) keypad lock-out is overcome by pressing Cancel, Enter, Cancel in sequence. If the Laser Indicator is lit, the laser is powered and operating normally. This indicator will be off if there is a laser-specific fault. This indicator will also be off for a brief period when the instrument is first powered. After initial power-up, this indicator may blink several times as the laser temperature is stabilized. The laser indicator will be lit constantly in normal operation. The Power Indicator is normally lit when the instrument is powered. 3.3.2 The Magnetic Stylus Each of the keys can be selected using a hand-held magnet called a Magnetic Stylus (see Figure 35 below), which is included with the meter. By touching the clear window at a key location, that key will be selected and will flash a red light to verify the contact. Figure 35: Magnetic Stylus 3.3.3 The Default Display Figure 36 below shows the default display of the Aurora H2O window. PPMv H2O Primary Alt 2 Alt 1 Figure 36: Default Display 36 Aurora H2O User’s Manual Chapter 3. Operation and General Programming 3.3.4 Unlocking the Keypad After power-on, the Aurora H2O keypad is locked as indicated by the symbol , lit up with a red backlight. It is necessary to enter the keypad unlock sequence to make any changes to the Aurora H2O. Similar to a mobile phone, the Aurora H2O will prompt the operator to unlock if any key is pressed. A passcode is required to use certain factory service features only. To unlock the keypad, press Cancel , Enter , Cancel in sequence. 3.3.5 Keypad Lock-Out Switch Keypad Lock-Out Switch Keypad Lock Indicator Figure 37: Keypad Lock-Out Switch Location Note: If the Keypad Lock-Out Switch is in the “down” position (towards the Aurora H2O sample system), the keypad is locked out and the RED LED on the Keypad Lock Indicator is on all the time. WARNING! Do not open or remove the cover with the power on, unless the area is non-hazardous. Aurora H2O User’s Manual 37 Chapter 3. Operation and General Programming 3.3.6 Accessing the Menus After successfully unlocking the keypad, press the Menu key. The Aurora H2O will display the Main Menu (see Figure 38 below). Use the arrow keys to highlight the menu item desired. Refer to Menu Map, Figure 73 on page 93. Press Enter to select the highlighted item. Many menu items will display another menu. Use Cancel to return to the previous menu page. Pressing Cancel from the Main Menu will return the screen to the Measurement Display. Note: Menu items displayed with an ellipsis (shown as a series of three dots after the menu item) will bring up more choices, while those without take immediate action. Figure 38: Main Menu 3.3.7 Entering Numeric Values Since the Aurora H2O has no numeric keypad, numeric values are entered using a “combination lock” style of entry: Use the left and right Use the up and down arrow keys to select the digit to change. The digit selected will be indicated with a . arrow keys to increment or decrement the digit. Note: If incrementing or decrementing a digit would cause the numeric value to exceed its allowable range (maximum/minimum value), the digit will not change. Press Enter to save the new value and return, or Cancel to return, leaving the original value intact. Figure 39: Numeric Entry 38 Aurora H2O User’s Manual Chapter 3. Operation and General Programming 3.3.8 Starting Up After proper installation, the Aurora H2O Transmitter can be set up to accommodate the user’s requirements. Typically, the user may need to configure the analog outputs, trim the analog outputs, and program the digital outputs. Refer to the Menu Map, Figure 73 on page 93, and complete the following steps. Upon startup, the Aurora H2O proceeds through several displays until a screen similar to the following appears: PPMv H2O After startup, the screen will need to be unlocked. To unlock the screen, select Cancel, Enter, Cancel Note: In most instances; use the Enter key to save an entry and/or move ahead to the following screen; use the Cancel key to reject an entry and/or return to the previous screen. 3.4 Setting Up the Display When the screen is unlocked, touch the Menu key and the Main Menu appears with several options. To set up the display, select Display... and press Enter. The following screen appears: 3.4.1 Selecting Primary Units To select units for the primary display, select Primary and press Enter. The following screen appears: Use the arrow keys to highlight the desired units and press Enter. The screen returns to the Display Menu. Aurora H2O User’s Manual 39 Chapter 3. Operation and General Programming 3.4.2 Selecting Alt 1 and Alt 2 Units To set the units for Alt 1 and/or Alt 2, use the arrow keys to highlight the one to be set, and press Enter. The following screen appears: Use the arrow keys to highlight the desired unit type (Hygro, Temperature or Pressure) and press Enter. If Temp is selected, the following screen appears. t 1 Units: Use the arrow keys to highlight the desired unit and press Enter. The screen returns to the Display Menu. Use the same procedure to change other units. 3.4.3 Setting Decimal Places To set the decimal places for unit values, from the Display Menu use the arrow keys to highlight Decimal and press Enter. Then select the type of display and press Enter. The decimal places setting determines the number of digits displayed for the value to the right of the decimal symbol (“.”), if possible. Use the arrow keys to change the number of decimal places and press Enter, or press Cancel if no changes are necessary. The screen returns to the Display Menu. 3.4.4 Data/Scan To toggle the display between showing the numeric values (data), and a graphic plot of the 2f waveform (scan), from the Display Menu use the arrow keys to highlight Data/Scan and press Enter. A screen similar to the following appears. Note:The scan can be used for diagnostic purposes when a PC with AuroraView is not readily available. 40 Aurora H2O User’s Manual Chapter 3. Operation and General Programming 3.4.5 Adjust To modify the display contrast and brightness, from the Display Menu use the arrow keys to highlight Adjust and press Enter. The following screen appears. Use the Up/Down arrow keys to increase/decrease display brightness. Use the Right/Left arrow keys to increase/decrease display contrast. Press Enter to save the changes, or press Cancel to return to the previous setup. The screen returns to the Display Menu. 3.4.6 Reverse To reverse the text and background shades, from the Display Menu use the arrow keys to highlight Reverse and press Enter. The following screen appears. To return to the previous shade setup, select Reverse and press Enter. The previous Data/Scan screen appears. Reverse 3.5 Setting Up Outputs 3.5.1 Selecting an Output for Setup To set up outputs, from the Main Menu choose Outputs... and press Enter. The following screen appears. From the Output Menu choose Select and press Enter. The following screen appears. Use the arrow keys to select the output (A, B or C) to be set up, and press Enter. Aurora H2O User’s Manual 41 Chapter 3. Operation and General Programming 3.5.2 Selecting Output Units From the Output Menu, select Units and press Enter. Use the arrow keys to select the unit type and press Enter. A screen similar to the following appears: Use the arrow keys to select a new unit. Press Enter to save (or Cancel to keep the previous value), and return to the Output Menu. 3.5.3 Selecting an Output Type To change the output type, from the Output Menu select Type and press Enter. A screen similar to the following appears: Use the arrow keys to select a new output type. Press Enter to save (or Cancel to keep the previous value), and return to the Output Menu. Note: If the ALM option is selected in the above menu, the analog output will function as an alarm indicator. In this mode, the analog output will remain at 0 mA during normal operating conditions, but the analog output will be driven to 24 mA whenever there is an alarm condition. 3.5.4 Changing the Upper Output Span To adjust the upper output span, from the Output Menu select Upper and press Enter. A screen similar to the following appears. Use the left and right arrow keys to select each digit to be changed and the up and down arrow keys to increase or decrease its value. Press Enter to save (or Cancel to keep the previous value), and return to Output Menu. 42 Aurora H2O User’s Manual Chapter 3. Operation and General Programming 3.5.5 Changing the Lower Output Span To adjust the lower output span, from the Output Menu select Lower and press Enter. A screen similar to the following appears. Use the left and right arrow keys to select each digit to be changed and the up and down arrow keys to increase or decrease its value. Press Enter to save (or Cancel to keep the previous value), and return to Output Menu. 3.5.6 Testing the Output The Test Menu causes the Aurora H2O to generate a 0- or 4-20mA output at the percent of scale selected. For example, in 4-20 operation, 0% = 4mA, 50% = 12mA, 100% = 20mA. This allows the proper function of recording or SCADA equipment to be verified. In 0-20 operation, 0% = 0mA, 50% = 10mA, 100% = 20mA. To test system output, from the Output Menu select Test and press Enter. The Aurora H2O will proceed to check the settings, and a screen similar to this display will appear. Use the left and right arrow keys to select each digit to be changed, and the up and down arrow keys to increase or decrease its value. Press Enter to save (or Cancel to keep) the previous value, and return to the Output Menu. Check your output wiring. If the reading on your SCADA or DCS is off slightly, then you may use the Trim feature to trim the output zero or span. Aurora H2O User’s Manual 43 Chapter 3. Operation and General Programming 3.5.7 Trimming the Outputs The Trim Menu enables the operator to compensate for differences in measurement of the 0/4-20 mA outputs by connected recorders or SCADA equipment. To trim the output: Select Trim from the Output Menu and press Enter. The following screen appears. To select an output to be trimmed, highlight Select Trim Output and press Enter. The following screen appears Use the left and right arrow keys to select an output (A, B or C) and press Enter. The screen returns to the previous display. When performing a Trim operation, the Aurora H2O unit requires you to first reset the trim. To reset the trim output, highlight Reset Trim and press Enter. The following screen appears. Use the left or right arrow keys to highlight YES and press Enter. This cancels any previous trim values, and returns the Aurora H2O to its factory adjustment. The display returns to the previous screen with Trim Zero highlighted. To trim the zero value, press Enter. A screen similar to the following appears. This will cause the Aurora H2O to output 4.000 mA on the output being trimmed. The output value should then be read using the connected recorder, SCADA equipment, or DVM. Enter the value read from the connected equipment as the Zero Trim value, as follows: 44 Aurora H2O User’s Manual Chapter 3. Operation and General Programming 3.5.7 Trimming the Outputs (cont.) Note: Since you cannot trim 0 mA for negative offsets, trim for the lower end of the scale is at the 4 mA output level. Use the left and right arrow keys to select each digit to be changed, and the up and down arrow keys to increase or decrease its value. Press Enter to save (or Cancel to keep the previous value). The Trim Menu returns with Trim Span highlighted. To change the span value, press Enter. A screen similar to the following appears. This will cause the Aurora H2O to output 20.000 mA on the output being trimmed. The output value should then be read using the connected recorder, SCADA equipment, or DVM. Enter the value read from the connected equipment as the Span Trim value. Use the left and right arrow keys to select each digit to be changed, and the up and down arrow keys to increase or decrease its value. Press Enter to save (or Cancel to keep the previous value). Trimming is complete. Accuracy can be verified using the Test Menu, above. Example: Trim is reset, then Trim Zero is selected. The SCADA input reports 3.977 mA. The operator enters “3.977” as the Zero Trim value. Trim Span is selected. The SCADA input reports 19.985 mA. The operator enters “19.985” as the Span Trim value. Aurora H2O will adjust the output accordingly to true the output as read by the customer recorder, SCADA or DVM. Using the Test Menu, the operator verifies that a test value of 0% now reads 4.000 mA at the SCADA equipment, and a test value of 100% now reads 20.000 mA. 3.6 Setting Up Alarms Note: The Aurora H2O is not equipped with alarm relays. The Alarm function is useful only when reading the alarm status via Modbus. 3.6.1 Selecting an Alarm Output To set up alarm outputs, on the Main Menu choose Alarm and press Enter. From the Alarm Menu choose Select and press Enter. A screen similar to the following appears. Use the arrow keys to select the output (A, B or C) to be set up and press Enter. The display returns to the Alarm Menu. Aurora H2O User’s Manual 45 Chapter 3. Operation and General Programming 3.6.2 Selecting Alarm Status To select the alarm status, from the Alarm Menu select Status and press Enter. The following screen appears: Use the arrow keys to select OFF or ON and press Enter. The display returns to the Alarm Menu. 3.6.3 Selecting Alarm Units To select alarm units, from the Alarm Menu select Units and press Enter. Use the arrow keys to select the unit type and press Enter. If Hygro was selected, this display appears. Use the arrow keys to select a unit. Press Enter to save (or Cancel to keep the previous value), and return to the Alarm Menu. If Temperature was selected, this display appears. Use the arrow keys to select a unit. Press Enter to save (or Cancel to keep the previous value), and return to the Alarm Menu. If Pressure was selected, this display appears. Use the arrow keys to select a unit. Press Enter to save (or Cancel to keep the previous value), and return to the Alarm Menu. 46 Aurora H2O User’s Manual Chapter 3. Operation and General Programming 3.6.4 Selecting an Alarm Type To change the alarm type, from the Alarm Menu select Type and press Enter. A screen similar to the following appears: Use the arrow keys to select an alarm type. Press Enter to save (or Cancel to keep the previous value), and return to the Alarm Menu. • SetPoint: Alarm activates when parameter exceeds upper limit, and deactivates when parameter is less than lower limit. • Inner Band: Alarm activates when parameter is between upper and lower limits. • Outer Band: Alarm activates when parameter is outside upper and lower limits. 3.6.5 How the Alarm Types Work Trip Upper Hysteresis Setpoint Lower Reset Reset In Band Trip Trip Reset Trip Out Band Reset Reset Trip Figure 40: Example of Alarm Types Aurora H2O User’s Manual 47 Chapter 3. Operation and General Programming 3.6.6 Changing the Upper Alarm Span To adjust the upper alarm span, from the Alarm Menu select Upper and press Enter. A screen similar to the following appears. Use the left and right arrow keys to select each digit to be changed and the up and down arrow keys to increase or decrease its value. Press Enter to save (or Cancel to keep the previous value), and return to Output Menu. 3.6.7 Changing the Lower Alarm Span To adjust the lower alarm span, from the Alarm Menu select Lower and press Enter. A screen similar to the following appears. Use the left and right arrow keys to select each digit to be changed and the up and down arrow keys to increase or decrease its value. Press Enter to save (or Cancel to keep the previous value), and return to Output Menu. 48 Aurora H2O User’s Manual Chapter 4. Programming Advanced Features Chapter 4. 4.1 Programming Advanced Features Comm Port Settings To access the communication port settings, from the Main Menu select Settings and press Enter. The following screen appears: To access the communications port settings, select Comms... and press Enter. The following screen appears: 4.1.1 Selecting a Comm Port There are two physical comm ports in the Aurora H2O. Comm Port 1 is aligned to SCADA in the instrument program and Comm Port 2 is aligned to SERVICE. This setup enables the user to have Comm Port 1 set up for the primary digital output (for example, RS-485 to the customer SCADA system), and Comm Port 2 to be used for service (for example, to enable a service engineer to interface with the Aurora H2O using an RS-232 cable connected to a laptop PC in the field, running AuroraView software). To select a communication port, use the arrow keys to highlight Select and press Enter. The following screen appears. Select SCADA or SERVICE and press Enter. The screen returns to the Comm Port Menu. 4.1.2 Setting the Baud Rate To set the baud rate, from the Comm Port Menu select Baud Rate and press Enter. The following screen appears. Use the arrow keys to highlight the desired baud rate and press Enter. The screen returns to the Comm Port Menu. IMPORTANT: If you are using version 1A of the Aurora H2O firmware, do not select the 1200 or 2400 baud rate. Aurora H2O User’s Manual 49 Chapter 4. Programming Advanced Features 4.1.3 Setting Parity To set parity, from the Comm Port Menu select Parity and press Enter. The following screen appears. Use the arrow keys to highlight the desired parity and press Enter. The screen returns to the Comm Port Menu. 4.1.4 Selecting Protocol To choose the protocol, from the Comm Port Menu select Protocol and press Enter. The following screen appears. Use the arrow keys to highlight the desired protocol and press Enter. The screen returns to the Comm Port Menu. 4.1.5 Setting the Network ID To set the network ID, from the Comm Port Menu select Network ID and press Enter. The following screen appears. Use the left and right arrow keys to select each digit to be changed. Use the up and down arrow keys to change the value. When finished, press Enter. The screen returns to the Comm Port Menu. 50 Aurora H2O User’s Manual Chapter 4. Programming Advanced Features 4.2 Adjust Offset Values To adjust offset values, from the Settings Menu select Adjust... and press Enter. The following screen appears. 4.2.1 Adjusting the PPMv Offset User Adjustments PPM Level Filter Dew Pt Calc Enh Factor To adjust the PPMv offset, select PPM Level and press Enter. The following screen appears. Use the left and right arrow keys to select each digit to be changed. Use the up and down arrow keys to change the value. When finished, press Enter. The screen returns to the User Adjustments Menu. 4.2.2 Adjusting the Smoothing Filter Offset User Adjustments PPM Level Filter Dew Pt Calc Enh Factor To adjust the smoothing filter offset, from the User Adjustments Menu select Filter and press Enter. The following screen appears. The smoothing filter setting is used to change the system responsiveness. It is a moving average filter to smooth the moisture readings. 1 sample = 1 scan. Generally, Aurora H2O can perform up to 12 samples per second. The minimum setting is 10 samples, and the maximum setting is 200 samples. The default, as set at the factory, is typically 40. Use the left and right arrow keys to select each digit to be changed. Use the up and down arrow keys to change the value. When finished, press Enter. The screen returns to the User Adjustments Menu. Aurora H2O User’s Manual 51 Chapter 4. Programming Advanced Features 4.2.3 Setting the Dew Point Calculation Method • The dew point is the temperature at which the air is saturated with respect to water vapor over a liquid surface. • The frost point is the temperature at which the air is saturated with respect to water vapor over an ice surface. There can be a difference of several degrees C between the dew point and the frost point. • When set for Dew/Frost, the Aurora H2O will report the Dew Point if the reading is above freezing, and will report the Frost Point if the reading is below freezing. • When set for Dew Point, the Aurora H2O will calculate the Dew Point temperature, even if that temperature is below freezing. User Adjustments PPM Level Filter Dew Pt Calc Enh Factor To set the dew point calculation method, from the User Adjustments Menu select Dew Pt Calc and press Enter. The following screen appears. Use the arrow keys to highlight the desired dew point setting and press Enter. The screen returns to the User Adjustments Menu. • The Dew calculation should be used for compatibility with ASTM-1142/IGT-8. The tables and calculations in those reports require measurements and provide results in dew point, regardless of the actual phase (dew or frost). • The Dew/Frost calculation should be used for compatibility with ISO-18453:2004, or when using a chilled mirror apparatus as a check standard. 4.2.4 Adjusting the Enh Factor User Adjustments PPM Level Filter Dew Pt Calc Enh Factor User Adjustments Enhancement Factor: OFF ON The Enhancement (Enh) Factor is used to adjust the calculated Pressure Dew Point (Equivalent Dew Point) in nitrogen for the effects of pressure.To select whether an enhancement factor is applied to the dew point calculation, from the User Adjustments Menu select Enh Factor and press Enter. The following screen appears. The Enhancement (Enh) Factor can be turned OFF or ON as required. Use the left or right arrow key to select your choice and press Enter. The screen returns to the User Adjustments Menu. X = Cancel Note: The Enhancement Factor does not apply when operating in methane/natural gas, as those calculations are designed to be consistent with the methods described in ASTM D-1142. 52 Aurora H2O User’s Manual Chapter 4. Programming Advanced Features 4.3 Set Up the Background Gas 4.3.1 Selecting the Type of Gas Beginning with firmware version H2O.001.C,.the background gas is selectable from the Settings Menu. For normal operation in natural gas service, Methane should be selected as the background gas. For verification testing, it may be desirable to use Nitrogen with a known moisture concentration. In this application, Nitrogen should be selected as the background gas. The Aurora H2O provides a positive indication if it is operating in Nitrogen mode. An N2 indicator will appear in the upper right corner of the LCD, adjacent to the status message. In normal Methane operation, only the status message is displayed. Unless otherwise requested, the Aurora H2O is shipped from the factory configured for Methane operation. To change the type of background gas, from the Settings Menu select Gas and press Enter. The following screen appears. From the Gas Data menu, select Background and press Enter. The following screen appears. Z Factor Use the arrow keys to select the desired background gas, and press Enter. The background gas selection is now complete. Press Cancel to return to the display page. Aurora H2O User’s Manual 53 Chapter 4. Programming Advanced Features 4.3.2 Setting the Gas Composition Note: The Gas Composition option is available only if Methane is selected as the Background Gas. To set the gas composition, from the Gas Data Menu select Composition and press Enter. The following screen appears. Z Factor The first choice is STD, the composition of our standard calibration mixture (90.0% CH4, 6.0% N2 and 4.0% CO2. To set the standard mixture, select STD and press Enter. The screen returns to the previous menu. The second choice is CH4, a composition of 100.0% Methane (CH4), for use when using bottled gas for verification. To set the Methane composition, select CH4 and press Enter. The screen returns to the previous menu. Note: An additional choice or two will be available above only if the customer has requested optional custom gas compositions. 4.3.3 Setting the Z Factor The Z factor is a number that accounts for the non-ideal compressibility of natural gas, and is vital for accurate calculation of mass/volume (lbs/MMSCF, mg/m3). To set the Z factor, from the Gas Data Menu, select Z Factor and press enter. The following screen appears. Use the left and right arrow keys to select each digit to be changed. Use the up and down arrow keys to change the value. When finished, press Enter. The screen returns to the Settings Menu. 54 Aurora H2O User’s Manual Chapter 4. Programming Advanced Features 4.3.4 Adjusting the Gas Offset The input information for gas molecular weight is not currently used for any moisture calculations and is reserved for future use. • lbs/MMSCF is calculated using IGT Research Bulletin #8 and ASTM D-1142-95 referenced at 60°F, 1 ATM. • mg/cm3 is based on ideal gas law derivation referenced at 15°C, 1.01325 kPa. To adjust the gas molecular weight offset, from the Settings Menu select Gas and press Enter. From the Gas Data menu select Mol. Weight and press Enter. The following screen appears. Use the left and right arrow keys to select each digit to be changed. Use the up and down arrow keys to change the value. When finished, press Enter. The screen returns to the Settings Menu. 4.4 Clock Settings The clock settings are for informational purposes. They are used to keep track of the test analyzer start time and the laser operational time. To reset the clock, from the Settings Menu select Clock and press Enter. The following screen appears. 4.4.1 Resetting the Hour To reset the hour, from the Clock Menu select Hour and press Enter. The following screen appears. Use the left and right arrow keys to select each digit to be changed. Use the up and down arrow keys to change the value. When finished, press Enter. The screen returns to the Clock Menu. Aurora H2O User’s Manual 55 Chapter 4. Programming Advanced Features 4.4.2 Resetting the Minutes To reset the minutes, from the Clock Menu select Minutes and press Enter. The following screen appears. Use the left and right arrow keys to select each digit to be changed. Use the up and down arrow keys to change the value. When finished, press Enter. The screen returns to the Clock Menu. 4.4.3 Resetting the Month To reset the month, from the Clock Menu select Month and press Enter. The following screen appears. Use the left and right arrow keys to select each digit to be changed. Use the up and down arrow keys to change the value. When finished, press Enter. The screen returns to the Clock Menu. 4.4.4 Resetting the Date To reset the date, from the Clock Menu select Date and press Enter. The following screen appears. Use the left and right arrow keys to select each digit to be changed. Use the up and down arrow keys to change the value. When finished, press Enter. The screen returns to the Clock Menu. 56 Aurora H2O User’s Manual Chapter 4. Programming Advanced Features 4.4.5 Resetting the Year To reset the year, from the Clock Menu select Year and press Enter. The following screen appears. Use the left and right arrow keys to select each digit to be changed. Use the up and down arrow keys to change the value. When finished, press Enter. The screen returns to the Clock Menu. 4.5 Pressure Settings To reset the pressure settings, from the Settings Menu, select Pressure... and press Enter. The following screen appears. 4.5.1 Setting the Source To reset the source, from the Pressure Menu, select Source and press Enter. The following screen appears. Use the left and right arrow keys to select the line pressure source. Live accepts a pressure value from a pressure transmitter and Remote accepts a value via a Modbus connection. To set a constant value, select Constant and press Enter. The screen returns to the Pressure Menu. 4.5.2 Changing the Constant If the pressure source selected is Constant, to reset its value, select Constant from the Pressure Menu and press Enter. The following screen appears. Aurora H2O User’s Manual 57 Chapter 4. Programming Advanced Features Use the left and right arrow keys to select each digit to be changed. Use the up and down arrow keys to change the value. When finished, press Enter. The screen returns to the Pressure Menu. Note: Data entry for this setting is only in kPa. 4.5.3 Editing Pressure Calibration Pressure input in this section is used only for equivalent dew point calculations. Equivalent dew point is the dew point of the process gas at the process pressure. Input a “constant” value if the line pressure is at a normal pressure, or use an external pressure transmitter to input a “live” pressure input into the Aurora H2O analyzer. To edit the pressure calibration, from the Line Pressure Source Menu, select Live and press Enter. The following screen appears. To edit the Calibration Data, use the arrow keys to select Cal Data and press Enter. The following screen appears. To select the Calibration Point, use the up and down arrow keys to highlight Select Cal Point and press Enter. The following screen appears. Use the left and right arrow keys to select Zero or Span and press Enter. The screen returns to the previous menu. To edit the Pressure Value, use the up and down arrow keys to select Edit Pressure Value and press Enter. The following screen appears. Use the left and right arrow keys to select each digit to be changed. Use the up and down arrow keys to change the value. When finished, press Enter. The screen returns to the previous menu. 58 Aurora H2O User’s Manual Chapter 4. Programming Advanced Features 4.5.3 Editing Pressure Calibration (cont.) To edit the Input Value, use the up and down arrow keys to select Edit Input Value and press Enter. The following screen appears. Use the left and right arrow keys to select each digit to be changed. Use the up and down arrow keys to change the value. When finished, press Enter. The screen returns to the previous menu. 4.6 Regional Settings This section enables the setting of regional information, depending on the location of the Aurora H2O. To reset the regional settings, from the Settings Menu, select Locale... and press Enter. The following screen appears. Note: Locale settings for your order have been set at the factory and are access code protected. If you determine a need to access Regional Settings, contact the factory for assistance. 4.6.1 Setting the Country Code To edit the country code, from the Regional Settings Menu select Country Code and press Enter. The following screen appears. A passcode is required to make changes. Use the left and right arrow keys to select each digit to be changed. Use the up and down arrow keys to change the value. When finished, press Enter. The screen returns to the Regional Settings Menu. • Country Codes = international telephone country codes. • Default = 1 for U.S. • Option = 81 for Japan is available to conform to METI requirements. Aurora H2O User’s Manual 59 Chapter 4. Programming Advanced Features 4.6.2 Setting the Decimal Format The Decimal Format option determines whether a decimal [.] or a comma [,] is used as the decimal separator. To edit the decimal format, from the Regional Settings Menu select Decimal Format and press Enter. The following screen appears. Use the left and right arrow keys to select a decimal [.] or a comma [,] as the decimal separator and press Enter. The screen returns to the Display Menu. 4.6.3 Setting the Date Format To edit the date format, from the Regional Settings Menu select Date Format and press Enter. The following screen appears. Use the left and right arrow keys to select the desired date format and press Enter. The screen returns to the previous display. 4.6.4 Setting the Unit System IMPORTANT: If the Country Code (see “Setting the Country Code” on page 59) has been set to 81 (Japan) the Unit System menu is not available. To select the unit system to be used for measurements, select Unit System and press Enter. The following screen appears. Use the left and right arrow keys to select the Unit System desired [SI = only metric units are available in the selection menus; SI + US = both metric and English units are available in the selection menus] and press Enter. The screen returns to the Regional Settings Menu. Note: Changing the Unit System setting from SI+US to SI in the above menu will NOT immediately change the current Display or Outputs units. This is because it is not possible for the Aurora H2O to know the SI measurement units desired by the operator to replace any currently displayed or output English units. However, the next time the operator selects different Display (see “Selecting Primary Units” on page 39) or Outputs (see “Selecting Output Units” on page 42) units, only SI units will be available in the menu. 60 Aurora H2O User’s Manual Chapter 4. Programming Advanced Features 4.7 User Calibration Beginning with software revision H2O.001.F, the Aurora H2O supports a two-point User Calibration for use in methane/natural gas. This is provided so that a field calibration can be easily performed to cause the Aurora H2O measurement to correspond to a calibration gas or other moisture device. The User Calibration is independent from, and does not affect, the Factory Calibration. The Aurora H2O can be quickly switched between User and Factory calibration at any time. Before the calibration can be performed, the Aurora H2O must be operating correctly in methane or natural gas, with methane selected as the Background Gas. Choose Settings... from the Main Menu. A User Cal... selection is now provided. Note: If the Aurora H2O is set for Nitrogen as the background gas, the User Cal selection will be disabled/grayed out. The Aurora H2O will prompt for the User Passcode [2719] By default, the Aurora H2O is operating with the Factory calibration, so the User Calibration Menu is disabled with the exception of the Status selection. Select Status, and choose User as the Active Calibration. The User Calibration Menu is now enabled. Aurora H2O User’s Manual 61 Chapter 4. Programming Advanced Features 4.7 User Calibration (cont.) The selections on the User Calibration Menu include these functions: Item Purpose Zero Gas User specifies the water concentration, in PPMv, of the “Zero”, or dry gas. Span Gas User specifies the water concentration, in PPMv, of the “Span”, or Cal/Wet gas. Read Zero The Aurora H2O will read the signal while exposed to the Zero gas. Read Span The Aurora H2O will read the signal while exposed to the Span gas, then determine the CH4 Offset and Scalar calibration values. CH4 Offset Permits the user to view/edit the calculated offset term. CH4 Scalar Permits the user to view/edit the calculated scalar (gain) term. First select the Zero Gas and Span Gas items, to specify the cal gases to be used. The zero gas must be in the range of 0.0 to 50.0 PPMv water. It is recommended that the zero gas be 10.0 PPMv or higher for best results. The Span gas must be in the range of 0.0 to 5000.0 PPMv water. It is recommended that the Span gas be 500.0 PPMv or lower for best performance. Apply the Zero gas, and wait for the Aurora H2O reading to stabilize. Note:The user can exit to the display and return to the User Cal Menu without affecting the Calibration procedure. When the reading is stable, select Read Zero. The Aurora H2O performs multiple measurements for approximately 15 seconds, indicating its progress. When the readings are complete, you will be prompted to apply the span gas, and continue. If there is doubt about the stability of the zero gas, the reading can be repeated at this point. 62 Aurora H2O User’s Manual Chapter 4. Programming Advanced Features 4.7 User Calibration (cont.) Apply the Span gas, and wait for the Aurora H2O reading to stabilize. When the reading is stable, select Read Span. The Aurora H2O uses the current measurement, and immediately calculates the new calibration. If the calculated values are determined to be valid, the Aurora H2O will display “Calibration OK” and prompt you to Save the results, or Cancel. If the values are not valid, the Aurora will display “Calibration *FAILED*”, and prompt you to continue. On a cal failure or Cancel, no change is made to the User Calibration values. The CH4 Offset and CH4 Scalar items can be used to view or modify the calculated calibration points, if desired. Note: The default values for the CH4 Offset is 0.0 and the CH4 Scalar is 1.0. These defaults are NOT the same as the Factory Calibration! Making the User Cal active without performing the actual calibration will adversely affect the accuracy of the Aurora H2O. If there is any doubt about the quality/accuracy of the gases used, the Aurora H2O should be returned to the Factory calibration using the Status item. 4.8 Service Settings CAUTION! The Service Settings Menu should be used by factory-trained personnel only! Aurora H2O User’s Manual 63 Chapter 4. Programming Advanced Features 4.9 Aurora H2O Information To check the Aurora H2O information, from the Main Menu select About and press Enter. The following screen appears. 4.9.1 Checking the ID To check identification information, select ID and press Enter. A screen similar to the following appears. Gas Composition To return to the About Menu, press Enter. 4.9.2 Checking the System Status To view the status of the Aurora H2O system, from the About Menu select System Status and press Enter. A screen similar to the following appears. Gas Composition Uptime: is the elapsed time since the Aurora H2O was powered on or reset. Started: is the date and time that the Aurora H2O was last powered on/reset. Start Temp: is the laser housing temperature as measured at the last startup/reset. Laser Hours: indicates the total lifetime that the laser has been energized. To return to the About Menu, press Enter. 64 Aurora H2O User’s Manual Chapter 4. Programming Advanced Features 4.9.3 Checking the Software To view the software versions being used, from the About Menu select Software Versions and press Enter. A screen similar to the following appears. Gas Composition To return to the About Menu, press Enter. 4.9.4 Checking the Gas Composition To view the gas content, from the About Menu select Gas Composition and press Enter. A screen similar to the following appears. To return to the About Menu, press Enter. Aurora H2O User’s Manual 65 Chapter 4. Programming Advanced Features 4.9.5 Checking the Alternate Gas Composition The Aurora H2O TDLAS is normally calibrated to a standard gas mixture that is representative of “typical” natural gas. The primary components and concentrations of this gas mixture are: Component Concentration Methane (CH4) 90.0% Nitrogen (N2) 6.0% Carbon Dioxide (CO2) 4.0% For special applications, where the composition of the gas to be measured differs significantly from the standard, GE can provide an alternate calibration. If this service has been ordered, the Aurora H2O will be shipped from the factory with both the standard and a custom calibration installed. The calibration in use can be verified at any time using the Aurora H2O About… menu. From the Main Menu, select About and press Enter. The following screen appears. From the About Aurora menu, select Gas Composition and press Enter. The following screen appears. An identifier label for the gas composition will be displayed above the components: Gas Composition Identifier 66 Aurora H2O User’s Manual Chapter 4. Programming Advanced Features 4.10 Custom Gas Composition If a custom Gas Composition has been installed, an additional Composition menu selection will appear in the Settings/Gas Data menu. Note: If no alternate composition is available, the menu selection will be disabled and “grayed out”. Gas Composition Identifier The new composition can be verified by selecting About… Gas Composition: Note: Unless otherwise requested, the Aurora H2O is shipped from the factory configured to use the custom gas composition. Aurora H2O User’s Manual 67 Chapter 4. Programming Advanced Features 4.11 Locking/Unlocking the Display To lock the Aurora H2O against any future changes, from the Main Menu select Lock and press Enter. The screen returns to the standard display. Note: This menu option is the same as exiting the programming menu and waiting for a keypad time-out to lock the keypad. To unlock the Aurora H2O for changes, press Cancel, Enter, Cancel as instructed in Unlocking the Keypad on page 37. 68 Aurora H2O User’s Manual Chapter 5. AuroraView Interface Software Chapter 5. AuroraView Interface Software 5.1 Capabilities Your Aurora H2O Analyzer is shipped with a CD which includes a PC-Software Application called AuroraView. With AuroraView, you can: • View Aurora H2O Configuration Items such as Alarms & Outputs. • Use the DataLog function to copy data to a comma delimited .txt file which can be opened by spreadsheet applications such as Microsoft Excel. • Plot real-time data for one or more Aurora H2O parameters • Manipulate plotted data in a variety of ways such as: color, line type, zoom in/out, etc. • Trend tabular data in real-time. • Show Scan Plots of the moisture absorption spectra. • Copy plots from AuroraView to other Window applications like Microsoft Powerpoint or Word. AuroraView does not provide functionality for the following: • Aurora H2O software updates. • Save the Aurora H2O Configuration. The Aurora H2O has a robust design which allows the meter to recover from fault conditions without the need to upload the configuration of the meter using external software. 5.2 Requirements AuroraView leverages a National Instruments Run-Time environment. This environment is supported on the following operating systems with the minimum installation requirements indicated: • 260 MB of available hard disk space • 64 MB of RAM or more • 300 MHz Pentium CPU • Windows NT 4.0 SP6 or higher, Windows ME, Windows 2000, Windows XP • Internet Explorer v5.0 or higher AuroraView supports the following serial interfaces: • RS232 • RS485 Modbus Aurora H2O User’s Manual 69 Chapter 5. AuroraView Interface Software 5.3 Installing AuroraView 1. Insert the Installation CD into the optical drive on your PC. 2. The installation program should launch automatically. If it does not, select Start → Run → Browse. Figure 41: Initial Screen 70 Aurora H2O User’s Manual Chapter 5. AuroraView Interface Software 5.3 Installing AuroraView (cont.) 3. Browse to the file named “setup.exe” in the root directory of the CD. Double-click the file to start the installation. Figure 42: AuroraView Installer Aurora H2O User’s Manual 71 Chapter 5. AuroraView Interface Software 5.3 Installing AuroraView (cont.) 4. Exit all other programs before running the installer. Figure 43: Installation Recommendation 72 Aurora H2O User’s Manual Chapter 5. AuroraView Interface Software 5.3 Installing AuroraView (cont.) 5. The next screen provides the opportunity to change installation locations if desired. When ready, click Next. Figure 44: Destination Directory 6. The next screen shows the Software License Agreement. Select “I accept the License Agreement” and click Next. Figure 45: Software License Agreement Aurora H2O User’s Manual 73 Chapter 5. AuroraView Interface Software 5.3 Installing AuroraView (cont.) 7. The next screen provides instructions for initiating the installation. When ready, click Next to begin. Figure 46: Starting the Installation 74 Aurora H2O User’s Manual Chapter 5. AuroraView Interface Software 5.3 Installing AuroraView (cont.) Figure 47: Overall Progress Aurora H2O User’s Manual 75 Chapter 5. AuroraView Interface Software 5.3 Installing AuroraView (cont.) 8. The following screen appears when the installation is complete. Figure 48: Installation Is Complete 76 Aurora H2O User’s Manual Chapter 5. AuroraView Interface Software 5.4 Starting AuroraView 1. From the Start menu, click Programs → AuroraView → AuroraView. Figure 49: AuroraView in Programs Menu Aurora H2O User’s Manual 77 Chapter 5. AuroraView Interface Software 5.4 Starting AuroraView (cont.) 2. AuroraView boots up and displays a screen similar to Figure 50 below. Command Line Alarms Status Trend Plots Tab Scan Plots Tab Trend Tabular Data Tab Reading Interval Datalog Interval Current Readings Pane Figure 50: AuroraView Main Screen 78 Aurora H2O User’s Manual Chapter 5. AuroraView Interface Software 5.5 Using the Main Menus 1. Click Measurements → Config Figure 51: Configuration Measurements • Unit String: Set this value to the value you want to read, plot or datalog. • Digits of Precision: Set a numerical value (typically 0, 1, 2). This sets the resolution of the displayed measurement units to the right of the decimal place (i.e., “20.78” would result from a setting of “2”). • Read: Check this box if you want to show the value in the current Readings pane. • Plot: Check this box if you want to show the value in the Trend Plots graph AND the Trend Tabular Data tab. Aurora H2O User’s Manual 79 Chapter 5. AuroraView Interface Software 5.5 Using the Main Menus (cont.) Note: The other options under MEASUREMENTS are for individual unit types and perform the same function as checking a box under the CONFIG pop-up window. Figure 52: Other Measurement Options 2. Click Alarms → Config This window enables the user to configure the alarm status within the AuroraView application. This feature allows you to remotely configure Aurora H2O alarms, which are used only with Modbus RTU digital output. The AuroraView Alarms are shown below. Figure 53: Alarms Configuration Set Alarm Trigger drop-down Toggle Alarms ON/OFF Alarm Type: Set Point, Out of Band, In Band Limits Alarm “Read” Indication (see below) Datalog Alarm Condition On/Off Figure 54: Other Measurements Options Figure 55: Alarm Status Indicators 80 Aurora H2O User’s Manual Chapter 5. AuroraView Interface Software 5.5 Using the Main Menus (cont.) 3. Click Outputs → Config Figure 56: Outputs Configuration Set Zero or Lower Range; Set Span or Upper Range Set Output Parameter Set Output Type 0-20mA or 4-20mA Figure 57: Other Output Options Aurora H2O User’s Manual 81 Chapter 5. AuroraView Interface Software 5.5 Using the Main Menus (cont.) 4. Click Scan This section will enable you to pick the type of scan you want to see. The default scan is the SPECTRA scan, which shows the 2f spectral scan. This is the processed signal waveform that the Aurora H2O uses to determine the moisture concentration. Viewing this scan may be helpful in certain troubleshooting situations. A typical 2f spectra scan is shown in Figure 59 on page 83. You may select the scan interval in minutes. This will be the refresh rate at which AuroraView updates the scan plot. To enter a scan interval, click on the Click to Save Scans Periodically button, and the following screen appears. Enter the interval and click on Continue to save or Cancel to reject the change. Figure 58: Save Scans Periodically 82 Aurora H2O User’s Manual Chapter 5. AuroraView Interface Software 5.5 Using the Main Menus (cont.) Figure 59: Scan Plot Tab Aurora H2O User’s Manual 83 Chapter 5. AuroraView Interface Software 5.5 Using the Main Menus (cont.) 5. Click Comms This window enables you to configure communication options. If you have more than one Aurora H2O on your network, you will have to establish different NETWORK ID's for each analyzer using the main keypad on the Aurora H2O. For your PC system, you will have to select which comm port to use. This is typically COM1. The default baud rate is 115200 baud. Figure 60: Configure Communication Options 6. Click Help This screen indicates the revision level of AuroraView. Figure 61: Software Information 84 Aurora H2O User’s Manual Chapter 5. AuroraView Interface Software 5.6 Datalogging with AuroraView 1. In the main view, click on the button Click to Datalog. Figure 62: Dataloging with AuroraView 2. AuroraView will request a file location. Pick a file location and a file name to save your data log file. All data log files are comma delimited .txt files by default. 3. Once you pick a file location, AuroraView will write any parameter that has a check box with Datalog checked in the main Config window at the time interval set in the Datalog Interval box, and the button in the main window will change to Datalogging…Click to Stop. 4. When you are done datalogging, click the button to stop logging. You may now open your .txt datalog file in any application, such as Microsoft Excel, so that you can analyze that data. Note: When datalogging multiple parameters at intervals of five seconds or less, it is recommended to use baud rates of 57.6K or 115.2K. Aurora H2O User’s Manual 85 Chapter 5. AuroraView Interface Software 5.7 Working with Trend Plots, Trend Tabular Data & Scan Plots 1. Trend Plots is a powerful graphing feature in AuroraView. You can graph many parameters at the same time. Clear Trend Data Button Graph Window Width Graph Tools Secondary Y-axis Primary Y-axis Current parameters being trended legend Secondary Y-axis Example Figure 63: Using Trend Plots Note: If you use the secondary y-axis, you may see “–” tick marks preceding the value. These are tick marks from the graphical applet and not an indication of negative values. 86 Aurora H2O User’s Manual Chapter 5. AuroraView Interface Software 5.7 Working with Trend Plots, Trend Tabular Data & Scan Plots (cont.) 2. If you right-click on any series of data within the graph, or you click on the current parameter being-trended item in the legend, you will see a variety of options for graphing data. You can change to a variety of common plots and adjust color, line style, and line width. For some data sets with lots of finite points, you may want to click Anti-Aliased which will smooth the plot line. You can also change bar plots, fill base line, interpolation and point style. X-Scale adjusts the x-scale. Y-Scale adjusts the y-scale and enables the secondary y-axis. Figure 64: Options for Graphing Data 3. There is a series of Graph Tools available at the top left of the trend plot area. Figure 65: Graph Tools • Pointer • Zoom Tool - gives you six options as shown in Figure 66. Zoom X-axis Zoom Y-axis Zoom window Zoom in Zoom all Zoom out Figure 66: Zoom Tool • Hand Tool - Enables you to graph the trend plot area and move it around without rescaling. Aurora H2O User’s Manual 87 Chapter 5. AuroraView Interface Software 5.7 Working with Trend Plots, Trend Tabular Data & Scan Plots (cont.) 4. Copying and Pasting a Trend Plot can be done from AuroraView. One way to do this quickly is to simply right click over the data area and choose Copy. In another application, like Microsoft Word, simply paste. [no content intended for this page] Figure 67: Copying a Trend Plot Figure 68: Pasting a Trend Plot 88 Aurora H2O User’s Manual Chapter 5. AuroraView Interface Software 5.7 Working with Trend Plots, Trend Tabular Data & Scan Plots (cont.) Another option is to right-click and chose the option Export Simplified Image. When you do this, a variety of image file formats will appear. A good universal option is Enhanced Metafile. Pasting an enhanced metafile will give you the ability to paste an image with an inverted color scheme as shown in the second example posted into Word (see Figure 71 on page 90). Figure 69: Exporting a Simplified Image Aurora H2O User’s Manual 89 Chapter 5. AuroraView Interface Software 5.7 Working with Trend Plots, Trend Tabular Data & Scan Plots (cont.) Figure 70: Selecting Enhanced Metafile Figure 71: Pasting an Enhanced Metafile 90 Aurora H2O User’s Manual Chapter 5. AuroraView Interface Software 5.7 Working with Trend Plots, Trend Tabular Data & Scan Plots (cont.) Working with Trend Tabular Data, you will be able to see data in tabular format as shown in Figure 72 below. You can adjust column widths to see data more easily with full titles in the header row. Figure 72: Trend Data in Tabular Format Aurora H2O User’s Manual 91 Chapter 5. AuroraView Interface Software [no content intended for this page] 92 Aurora H2O User’s Manual Chapter 4. Programming Advanced Features Main Menu Primary Alt 1 Alt 2 Outputs Decimal Adjust Reverse Set Decimal for: Primary ppmv Alarm Select Select Output: Pw, kPa Dew Pt °F Eq DP °C B C ESC = Cancel 1 decimals ENT = Save ESC = Cancel ESC = Cancel C ESC = Cancel Select Unit Type: Temp ppmv Eq DP °F Press Sample Line Set Alarm Status OFF Press ON 115.2 k 9600 57.6 k 4800 38.4 k 2400 19.2 k 1200 Temp °C kPa Select Output Type: Select Unit Type: Pw, kPa Temp °F mPa 4-20 mA Hygro Dew Pt °C Temp K bci 0-20 mA Temp Dew Pt °F Dew Pt °C kg/cm2 Dew Pt °F Note: The selection of "English" or non-metric measurements is dependent on the Regional Settings from the LOCALE menu. The unit system choices are SI (from the French Le Systeme International d'Unites), commonly referred to as the "metric" system, or SI+US. Measurement Units Hygro Temp DP °C, Eq. DP °C, T °C, DP °C mg/m , PPMv, Pw kPa Eq. DP °F, Lbs/MMSF T °F, DP °F Set Output A, B or C: Upper Span Lower Zero mmHg PSIa PSIg Press Type Select Alarm Type: i.e. 00250.0 PPMv Setpoint Set Output A, B or C Test Value Test In Band i.e. ±050.00 % Out Band Set Output A, B or C Pressure Trim... Reset Trim Trim Zero Bar, kg/cm All SI units plus: PSIa, PSIg Set Parity EVEN Units mg/m3 Eq DP °F Aurora H2O User’s Manual Baud Rate Parity Eq DP °C SI+US Status Temp Type SI Comm Port: Select Select Comm Port SCADA / Service ESC = Cancel Units Hygro Hygro Comms... A B Alt 2 Dew Pt °C Select Select Output: A Alt 1 mg/m 3 Settings Trim Span Enter A, B or C Min/Max Upper Alarm Value Lower i.e. ± 000.0 Td °C ODD NONE Protocol Service Factory-trained personnel only Adjust Adjust Offset Values: ppm Level Set Hour [0-23]: Set Minutes [0-59]: Set Month [1-12]: Set Date: Set Year: Dew Frost RS-485 Network ID Set Node ID: i.e. 001 Lock Lock Display About Aurora: ID GE Copyright, etc. Unit SN: Sensor SN: System Status Uptime: i.e. 0 d 00 h Laser Hours: i.e. 85 Year Month BOOT: PROG: Standard Methane Gas Set Molecular Weight Pressure Pressure Settings User Calibration Status Source Factory Select Pressure Source: Constant User Background Select Background Gas Nitrogen Select Protocol RS-232 Set Clock: Hour Date Minutes Filter Smoothing Filter: i.e. 200 Samples Dew Clock About Methane Composition Select Composition Identifier STD Live Constant Enter Line Pressure: i.e. ± 0101.325 kPa Figure 73: Programming Menu Map Display Cal Data Edit Pressure Cal: Select Cal Point Select Calibration Point: Zero Passcode Protected Locale Regional Settings Span CH4 Z Factor Edit Pressure Value Enter Line Pressure: i.e. 0000.0000 kPa Edit Input Value Enter Line Signal: i.e. 0001.000 mA Date Format Unit System 93 Chapter 4. Programming Advanced Features [no content intended for this page] Aurora H2O User’s Manual 94 Chapter 6. Maintenance Chapter 6. Maintenance CAUTION! CLASS 1M INVISIBLE LASER RADIATION WHEN OPEN. DO NOT VIEW DIRECTLY WITH OPTICAL INSTRUMENTS. WARNING! Use of controls or adjustments or performance of procedures other than those specified herein may result in radiation exposure that is more hazardous than specified. 6.1 Spare Parts Part No. Table 3: Aurora H2O Spare Parts List Description Qty. 704-688-12 RS-232 Cable; SUB-9-F to Tinned Leads; 12 ft. 1 Aurora H2O Maintenance Kit N/A Plastic Case with Foam Inserts 1 421-3230 Magnetic Stylus 1 240-199 Air Blower 1 403-161 Lens Tissue Package 1 463-030 Replacement Membrane Filter Elements 5 240-201 Allen Wrench, 5/32” 1 240-200 Allen Wrench, 3/32” 1 S40046393 Small Screw Driver 1 Gloves 4 403-163 6.2 Recommended Factory Verification Period Aurora H2O technology is designed for long life without calibration. There are no wetted sensing surfaces, which might degrade over time due to direct sample gas contamination. The optical components are designed to be stable over the span of many years. GE recommends that Aurora H2O analyzers may be returned to the factory for verification on a five (5) year periodic basis. GE will inspect, clean, replace filter elements and calibrate the unit to traceable standards as part of the factory service for Aurora H2O analyzers. 6.3 Cleaning the Mirror The Aurora H2O analyzer may provide the message Weak Signal Return - Check Mirror on the top line of the display, along with the red light “!” to the left of the main display. If this occurs, the mirror and/or the optical window of the measuring cell may be contaminated due to liquid or particulate coating/deposition. Note: Reagent grade Acetone (Cas no. 67-64-1) may be needed for the cleaning process. The reagent is not supplied as part of the Aurora maintenance kit, and you must obtain the reagent from a local chemical supplier. Aurora H2O User’s Manual 95 Chapter 6. Maintenance 6.3 Cleaning the Mirror (cont.) If the error condition Check Mirror is indicated, the first step is to clean the mirror. The process is as follows: 1. Turn the power to the Aurora H2O OFF. CAUTION! CLASS 1M INVISIBLE LASER RATIATION WHEN OPEN. DO NOT VIEW DIRECTLY WITH OPTICAL INSTRUMENTS. WARNING! Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous laser exposure. 2. Shut down the flow through the measuring cell. a. Turn the inlet isolation ball valve (item 4 in Figure 2 on page 4 or Figure 3 on page 5) to the closed position. b. Ensure that the sample flow rotatmeter indicates zero flow. 3. Wear a pair of disposable latex gloves. Note: A 5/32 inch allen wrench is required to accomplish this task. 4. Remove the 6 hex bolts securing the mirror (see Figure 74 below). A 5/32 inch allen wrench is required to accomplish this task. Hold the stainless steel base of the mirror with one hand while removing the last retaining bolts. The mirror is removed by pulling it straight down. There is one alignment pin to facilitate proper removal. Alignment Pin Slot Figure 74: Removing the Aurora H2O Mirror 96 Aurora H2O User’s Manual Chapter 6. Maintenance 6.3 Cleaning the Mirror (cont.) CAUTION! Handle the mirror with extreme care. Performance of the analyzer is dependent upon the mirror integrity. Do not touch the mirror surface with any tools, objects, hands or fingers. Alignment Pin Mirror Figure 75: Handling the Mirror Assembly 5. Make a visual inspection of the mirror surface. Record any observation of gross contamination. If the mirror appears to be clean, do not clean it. Instead, reinstall it into the system. Figure 76: Examining the Mirror Aurora H2O User’s Manual 97 Chapter 6. Maintenance 6.3 Cleaning the Mirror (cont.) 6. If the mirror is contaminated, the first step in cleaning it is to use the air blower to blow away any particulate matter on the mirror surface. Place the mirror assembly on a flat surface and repeatedly puff the air blower to force dry clean air onto the mirror surface (see Figure 77 below). Figure 77: Blowing Air onto the Mirror 7. Using a piece of lens tissue, wet a small area with a very small amount of analytical grade acetone (barely wet the lens tissue). Typically, one drop of acetone will suffice. Tilt the lens tissue so that the acetone drop is absorbed by the lens paper along the length of the piece (see Figure 78 below). Figure 78: Using Analytical Grade Acetone 98 Aurora H2O User’s Manual Chapter 6. Maintenance 6.3 Cleaning the Mirror (cont.) 8. Place the wet lens tissue on top of the mirror and drag it horizontally across the surface of the mirror (see Figure 79 below). Figure 79: Dragging the Wet Lens Tissue Across the Mirror 9. Use the air blower to blow dry air onto the surface of the mirror. Puff repeatedly until the mirror surface appears dry (see Figure 80 below). Figure 80: Drying the Mirror Aurora H2O User’s Manual 99 Chapter 6. Maintenance 6.3 Cleaning the Mirror (cont.) 10. Repeat steps 7-9 at least three times, using a new lens tissue each time. 11. Visually inspect the mirror, and record any observation of gross contamination. 12. If the mirror does not appear to be clean, contact GE for further assistance. 13. If the mirror does appear to be clean, re-install it onto the measurement cell. 14. As you line up the mirror to the measurement cell, note the “key” for properly connecting the two pieces. The mirror will align for mounting in only one position. Rotate the mirror assembly to line up the alignment pin with the alignment pin slot (see Figure 81 below). Alignment Pin Slot Figure 81: Reinstalling the Mirror Assembly 15. Loosely reinstall the six retaining bolts by hand. 16. Use an Allen wrench to tighten the six retaining bolts in an alternating or star pattern to achieve a uniform metal-to-metal contact. Tighten the bolts just until they are snug - DO NOT overtighten these bolts. 17. Re-establish power to the Aurora H2O analyzer. 18. Re-establish the flow through the sample system. 19. If the Aurora H2O still indicates a Check Mirror, contact GE for further assistance. 100 Aurora H2O User’s Manual Chapter 6. Maintenance 6.4 Replacing the Filter Element The Aurora H2O uses a membrane filter (see photograph to the left) as the secondary filter. This filter is intended to prevent liquid or particulate contamination from entering the absorption cell. The Aurora should not be operated without a filter train upstream of the unit. The membrane filter is equipped with “Flow Block” which shuts off the flow if the filter element is heavily loaded with contamination. A spring loaded check valve closes the outlet flow of the sample if the differential pressure across the filter element exceeds a threshold limit. At any given time the flow through the Aurora H2O can be checked by observing the rotameter. If the “flow block” feature shuts off the flow, do not increase the pressure. Replace the filter element and clean the filter. If the flow is being shut off too frequently, additional sample condition schemes, or the combination of the following, will have to be employed: • Bypass flow - needed to sweep liquids or contamination off the filter. A 10:1 bypass flow rate should be maintained. • Additional upstream filtration • Heat – Heat tracing of the sample line and sample system components, sufficiently above both the water and hydrocarbon dew point, will keep the sample in the gas phase. To replace the filter element, complete the following steps: 1. Close the inlet isolation ball valve and allow the system to fully depressurize. 2. Turn the filter cap counter-clockwise (see Figure 82 below). You might need to use a channel-lock wrench to assist in loosening the cap. Figure 82: Removing the Filter Cap Aurora H2O User’s Manual 101 Chapter 6. Maintenance 6.4 Replacing the Filter Element (cont.) 3. Place the filter cap on a horizontal flat surface, filter side up. 4. Carefully remove the large O-ring. Note: Because the O-rings are reused, replacement O-rings are not included in the maintenance kit. Figure 83: Orient the Filter Cap and Remove the Large O-Ring 5. Remove the white membrane filter element and the membrane backing plate. Figure 84: Remove the White Filter Element and the Backing Plate 102 Aurora H2O User’s Manual Chapter 6. Maintenance 6.4 Replacing the Filter Element (cont.) 6. Remove the small O-ring. Figure 85: Remove the Small O-ring 7. Using a tissue, clean the filter components. Large O-Ring Membrane Filter Backing Plate Small O-Ring Figure 86: O-Rings, Membrane Filter and Backing Plate Removed 8. Re-assemble the filter, and reinstall the cap hand–tight. Aurora H2O User’s Manual 103 Chapter 6. Maintenance [no content intended for this page] 104 Aurora H2O User’s Manual Chapter 7. Troubleshooting Chapter 7. Troubleshooting 7.1 Introduction The following are possible Aurora H2O analyzer conditions with details on how to deal with them. 7.2 Blank Display 1. Is the green POWER LED lit? a. Yes - proceed to Step 2. b. No - Check the wiring and the fuse 2. Are the four arrow keys illuminated? a. Yes - If the keys remain illuminated for more than 12 seconds, the Boot Loader cannot find a valid Instrument Program to run. b. No - contact the factory for service. 7.3 Display Dim or Hard to Read 1. Adjust the LCD brightness and contrast using the Display/Adjust menu. 7.4 Status Messages and Indicators 1. The Aurora H2O categorizes status messages as either Faults, Warnings, or Information. Status messages are displayed in the upper right corner of the display. Messages that are longer than the message area continuously scroll from right to left. 2. A fault is a non-recoverable condition that can affect the quality of measurement by the Aurora H2O. Fault messages are accompanied by a slow flashing (!) indicator. 3. A warning is a recoverable condition that can affect the quality of measurement by the Aurora H2O. Warning messages are accompanied by a rapid flashing (!) indicator. 4. Information messages alert the operator to a condition that is abnormal, but does not affect the quality of measurements. Information messages are accompanied by a slow flashing (i) indicator. 5. Aurora H2O fault and status messages are prioritized; in case of more than one fault/status condition, the condition with the highest priority is displayed. When that condition is resolved, the next highest priority condition is displayed. Aurora H2O User’s Manual 105 Chapter 7. Troubleshooting 7.4 Status Messages and Indicators (cont.) Message Table 4: Status Messages and Indicators Condition Description Status OK Info Aurora H2O is operating normally, no faults or other indications. No CH4 detected Info Aurora H2O is reading moisture, but cannot detect the presence of methane. H2O Under Range Info The moisture level is below the system detection limits. Warning - System Overheating Fault The temperature inside the electronics module exceeds 85°C or the air temperature inside the sample system enclosure exceeds 68°C. The laser is powered off until the electronics module temperature is below 80°C, and the sample system enclosure temperature is below 65°C. FAULT: Temperature Fault The temperature transducer is operating out of limits, is disconnected, or has failed. FAULT: Sample Pressure Fault The internal (sample) pressure transducer is operating out of limits, is disconnected, or has failed. FAULT: Line Pressure Fault The external (line) pressure transmitter is operating out of limits, is disconnected, or has failed. Occurs if source of the line pressure measurement is set to "Live", and no pressure transmitter is attached. Laser Temp Unstable... Warning Laser Adjust at Limits Info Aurora H2O has reached the limit for adjusting the laser power. Contact the factory for assistance. Laser Reference Fail Fault Aurora H2O could not detect any signal from the laser. Contact the factory for assistance. Weak Signal Return Check Mirror Info Aurora H2O could not detect a signal returned from the sample cell, or the signal is below allowed limits. Check mirror for contamination. FAULT: TEC FAIL Fault Aurora H2O has detected a failure in the laser temperature control. Contact the factory for assistance. WARNING Sample Pressure TOO HIGH Warning ERROR: TEC Setpoint out of Range Fault The Aurora H2O temperature controller is was set beyond its operating limits. Contact the factory for assistance. Service Req: ### Fault Aurora H2O has detected a fault condition that has no associated status message. Contact the factory for assistance. 106 The temperature of the laser is not stable. This warning occurs briefly at power on, as the Aurora H2O sets the correct operating temperature. The laser is powered off until the temperature has stabilized. The pressure in the Aurora H2O sample cell is greater than 212 kPa (30.75 psia). Verify regulator and flow settings; check for blocked vent line or excessive back pressure. Aurora H2O User’s Manual Chapter 7. Troubleshooting 7.5 No Flow Measurement Indicated on Aurora H2O Measurement Cell Outlet Check to make sure that the outlet of the Aurora H2O is venting to atmospheric pressure. Ensure that the sample system valves are configured correctly and that the Aurora H2O internal pressure regulator is capable of a barely positive pressure setting. Check/Replace the filter element in the coalescer/filter as detailed in Chapter 5, Maintenance. 7.6 Verifying Aurora H2O Performance in the Field There are two methods for verifying the performance of the Aurora H2O in the field. The first method is to use a portable hygrometer such as GE's PM880 portable hygrometer. The second method is to use a moisture generator or standard gas cylinder. 7.6.1 Using A Portable Hygrometer One quick spot check that can be done, that is relatively easy to do and requires minimal set-up time and no consumables, is to use a second hygrometer. GE recommends using a PM880 Portable Hygrometer for this requirement with a recently calibrated Aluminum Oxide Moisture Probe. Figure 87: PM880 Portable Hygrometer The PM880 can be connected to the outlet of the Aurora H2O analyzer, using the portable sample system with the aluminum oxide moisture probe, to verify the system. The response time of the portable hygrometer is limited, as the sensor is typically exposed to air during movement to the sample point. So, it is recommended that the sample gas remains flowing through the portable sample system on the sample outlet of the Aurora H2O until equilibrium with the moisture concentration of the sample gas is reached. You can use the data-logging capability of the PM880 to determine when steady-state conditions are reached. Consult GE if you have application questions regarding this process. Aurora H2O User’s Manual 107 Chapter 7. Troubleshooting 7.6.2 Using a Moisture Standard The best moisture standard to use is one generated by a moisture generator as a flowing reference, due to the fact that static moisture standards in cylinders have limited capability and reliability. A moisture generator such as GE's MG101 moisture generator can be used for this purpose. Figure 88: MG101 Moisture Generator However, a moisture generator is typically used only in an indoor location with reasonably good temperature control. In the field this is not always practical, so a more convenient option would be to use a static moisture standard in a gas cylinder. Consult with your local specialty gas supplier about standards for moisture. Based on application experience, GE recommends the following guidelines regarding moisture standards in gas cylinders: • Use passivated, aluminum gas cylinders only. • Do not use the cylinder when the pressure falls below 50% of the original pressure supplied by the vendor (typically 1500-1800psig). • Use the gas cylinder for moisture values of 50–100 ppm. • Use a moisture standard having a background gas of nitrogen (N2). • Mix the moisture standard for 10 minutes prior to use, following the manufacturer's guidelines (rolling the cylinder is typical). • Use the gas cylinder at the nominal temperature at which the cylinder was tested by the manufacturer. Whether using a moisture generator or a standard moisture gas cylinder, the source gas should be connected to the Aurora H2O using the PURGE INLET on the analyzer. Ensure that the sample gas pressure is regulated to a just barely positive pressure and establish gas flow from the PURGE INLET to the Aurora H2O measurement cell. 108 Aurora H2O User’s Manual Chapter 7. Troubleshooting 7.7 Background Selection Lockout To prevent accidental selection or tampering, the Background menu option can be disabled using a mechanical switch located behind the Aurora H2O display. Access to the switch requires removal of the cover, and should be performed only in non-hazardous conditions. Note: Unless otherwise requested, the Aurora/H2O is shipped from the factory with the Background selection unlocked. The Lockout switch is positioned to the right of the Laser Indicator (see Figure 89 below). When the switch is in the UP position, the Background Selection menu is unlocked. When the switch is in the DOWN position, the Background Selection menu is locked. Background Selection Lockout Switch (shown in unlocked position) Figure 89: Background Selection Lockout Switch Location Attempting to access the Background Gas Selection menu with the switch in the Locked Out (down) position will result in the following message being displayed: Aurora H2O User’s Manual 109 Chapter 7. Troubleshooting [no content intended for this page] 110 Aurora H2O User’s Manual Appendix A. MODBUS RTU/RS485 Communications Appendix A. MODBUS RTU/RS485 Communications A.1 Introduction The Aurora H2O supports digital communications using the Modbus/RTU protocol, with 2-wire RS-485 or 3-wire RS-232C as the physical layer. Data rate can be specified ar rates from 1200 to 115200 bits per second (bps), with selectable parity. Aurora H2O has two physically separate communications ports. Both ports can be selected for either RS-232 or RS-485 operation, and Aurora H2O can communicate with both ports simultaneously. Note: Aurora H2O supports the Modbus/RTU protocol as defined in: MODBUS Application Protocol Specification, V1.1b & MODBUS over Serial Line Specification and Implementation Guide V1.02. Note: The above referenced specifications are available from the Modbus Organization at http://modbus-ida.org/ A.2 Modbus Functions and Data Types The Modbus functions supported by Aurora H2O include: • • • • • • (0x03) Read Holding Registers (0x04) Read Input Registers (0x08) Diagnostics (Serial Line only) - only supports Echo subcommand (0x10) Write Multiple registers (0x11) Report Slave ID (Serial Line only) (0x2B/0x0E) Read Device Identification - only supports Basic Device Identification tags, which are: • • • VendorName Product code Revision number Aurora H2O supports data types of Integer and Double/Float. Integer data types are always four (4) bytes long and should be read with a request for two registers (two bytes per each register, two registers total) at the address. Double/Float data types provide eight (8) byte double precision data or four (4) byte single precision data, depending on how many registers are requested. Four registers are used for double data types, and two registers are used for single precision data types. Aurora H2O User’s Manual 111 Appendix A. MODBUS RTU/RS485 Communications A.3 Modbus Registers All registers denoted with a bullet (•) in the Read-Only column are read-only registers and should be read with the function “Read Input Registers.” All other registers can be read and written with “Read Holding Registers” or “Write Multiple Registers.” Note: The Modbus Register Address map supported by Aurora H2O is shown in Table 5 below. Table 5: Modbus Register Map Function Parameter Description Range/State Addr 0 Data Type System Status1 Status Register Status Register, Latched Write 0 to clear 1000 Integer Analog Output Output 1 Trim Sequence Start/Resume Live Output2 2100 Integer Units Reg. address of Meas. 2110 Integer Type 0 = 4-20mA, 1 = 0-20mA 2120 Integer Trim Reading Zero 3.0 ~ 5.2 2140 Double/Float Trim Reading Span 10.0 ~ 22.2 2150 Double/Float Upper of Value –10000 ~ 10000 2160 Double/Float Lower of Value –10000 ~ 10000 2170 Double/Float % value of output, 0~100 2180 Double/Float Test Output 2 Output 3 112 Integer Read Only • 2200 Integer 2 Trim Sequence Start/Resume Live Output Units Reg. address of Meas. 2210 Integer Type 0 = 4-20mA, 1 = 0-20mA 2220 Integer Trim Reading Zero 3.0 ~ 5.2 2240 Double/Float Trim Reading Span 10.0 ~ 22.2 2250 Double/Float Upper of Value –10000 ~ 10000 2260 Double/Float Lower of Value –10000 ~ 10000 2270 Double/Float Test % value of output, 0~100 2280 Double/Float 2300 Integer Trim Sequence Start/Resume Live Output2 Units Reg. address of Meas. 2310 Integer Type 0 = 4-20mA, 1 = 0-20mA 2320 Integer Trim Reading Zero 3.0 ~ 5.2 2340 Double/Float Trim Reading Span 10.0 ~ 22.2 2350 Double/Float Upper of Value –10000 ~ 10000 2360 Double/Float Lower of Value –10000 ~ 10000 2370 Double/Float Test % value of output, 0~100 2380 Double/Float Aurora H2O User’s Manual Appendix A. MODBUS RTU/RS485 Communications Table 5: Modbus Register Map (cont.) Function Alarm Parameter All Alarm Status Alarm 1 Alarm 2 Alarm 3 Settings Description Adjust Clock 0 ~ 7 (Bitfield) Addr Data Type Read Only 3000 Integer • Status 0 = Not tripped, 1 = Tripped 3100 Integer • Switch 0 = OFF, 1 = ON 3110 Integer Units Reg. address of Meas. 3120 Integer Type Set Point = 0, In Band = 1, Out Band = 2 3130 Integer Upper Depends on unit type 3140 Double/Float Lower Depends on unit type 3150 Double/Float Status 0 = Not tripped, 1 = Tripped 3200 Integer Switch 0 = OFF, 1 = ON 3210 Integer Units Reg. address of Meas. 3220 Integer Type Set Point = 0, In Band = 1, Out Band = 2 3230 Integer Upper Depends on unit type 3240 Double/Float Lower Depends on unit type 3250 Double/Float Status 0 = Not tripped, 1 = Tripped 3300 Integer Switch 0 = OFF, 1 = ON 3310 Integer Units Reg. address of Meas. 3320 Integer Type Set Point = 0, In Band = 1, Out Band = 2 3330 Integer Upper Depends on unit type 3340 Double/Float Lower Depends on unit type 3350 Double/Float PPM Level offset adjust –25.00 ~ +25.00 5210 Double/Float Moisture reading average filter size 10 ~ 200 samples 5230 Integer Hour 0~23 5410 Integer Minutes 0~59 5420 Integer Month 1~12 5430 Integer Date 1~28/29/30/31 5440 Integer Year 2000~2099 5450 Integer 0 ~ 3500.00 kPa 5510 Double/Float Pressure Zero Calibration, mA 0~22 mA 5520 Double/Float Pressure Span Calibration, mA 0~22 mA 5525 Double/Float Pressure Zero Calibration, kPa 0~3500 kPa 5530 Double/Float Pressure Span Calibration, kPa 0~3500 kPa 5535 Double/Float Pressure Source Constant Value = 0, Live Sensor = 1 5540 Integer External Pressure Constant Aurora H2O User’s Manual Range/State • • 113 Appendix A. MODBUS RTU/RS485 Communications Table 5: Modbus Register Map (cont.) Function Device ID Measurements Parameter Description Range/State Addr Data Type Read Only Aurora H2O Serial Number ------ ------ 8100 8 Character Bytes • Laser Serial Number ------ ------ 8200 8 Character Bytes • Calibration Date Month 1~12 8310 Integer • Date Depends on month 8320 Integer • Year 2000~2100 8330 Integer • System Up Time MSDate Uptime, in days 8400 Double/Float • Dew Point Dew Point °C ------ 9110 Double/Float • Dew Point °F ------ 9120 Double/Float • Equivalent Dew Point °C ------ 9130 Double/Float • Equivalent Dew Point °F ------ 9140 Double/Float • Sample Temperature °C ------ 9210 Double/Float • Sample Temperature °F ------ 9220 Double/Float • External Pressure kPa ------ 9510 Double/Float • MPa ------ 9512 Double/Float • Temperature PSIa ------ 9520 Double/Float • PSIg ------ 9530 Double/Float • kg/cm2 ------ 9540 Double/Float • Bars ------ 9550 Double/Float • mmHg ------ 9560 Double/Float • Internal Pressure kPa ------ 9610 Double/Float • H2O Concentration Vapor Pressure MPa ------ 9612 Double/Float • PSIa ------ 9620 Double/Float • PSIg ------ 9630 Double/Float • kg/cm2 ------ 9640 Double/Float • Bars ------ 9650 Double/Float • mmHg ------ 9660 Double/Float • PPM ------ 9710 Double/Float • Lbs MMSCF ------ 9720 Double/Float • mg/sm3 ------ 9730 Double/Float • kPa ------ 9800 Double/Float • 1 Address 0 is the System Status register, and address 1000 is the latching version of the System Status register. That is, both registers will show the error bit if the error is currently present, but only the latching register will show it if the condition is no longer present. Writing 0 to the latching register will clear the error code it contains. 114 Aurora H2O User’s Manual Appendix A. MODBUS RTU/RS485 Communications A.3 Modbus Registers (cont.) 2The “Trim Sequence Start/Resume Live Output” registers for the three outputs, (address 2100, 2200, 2300) accept certain values through the Write Multiple Register to trim the output current: 1. Write 0 to 2x00 to select the normal mA output (proportional to measurement). 2. Write 1 to 2x00 to reset the mA output trim to factory defaults. 3. Write 2 to 2x00 to output the “zero” current (~4.000 mA) and accept a calibration value written to 2x40. 4. Write 3 to 2x00 to output the “span” current (~20.000 mA) and accept a calibration value written to 2x50. Note: Attempting to write to Trim registers 2x40/2x50 without first writing to Trim State register 2x00 will fail with Modbus exception 4. At the end of calibration, write 0 to 2x00 to make Aurora H2O exit trim mode. Aurora H2O User’s Manual 115 Appendix A. MODBUS RTU/RS485 Communications A.4 Modbus System Status Codes Table 6 below lists the System Status codes and descriptions. It is possible for multiple status codes to be present; the hexadecimal values represent the bit set for a given condition. Table 6: System Status Codes Description Status 0x00000000 Aurora H2O is operating normally, no faults or other indications. 0x00000008 Aurora H2O is reading moisture, but cannot detect the presence of methane. 0x00000010 The moisture level is below the system detection limits. 0x00000020 The temperature inside the electronics module exceeds 85°C. The laser is powered off until the temperature drops below 80°C. 0x00000040 The temperature transducer is operating out of limits, is disconnected, or has failed. 0x00000080 The internal (sample) pressure transducer is operating out of limits, is disconnected, or has failed. 0x00000100 The external (line) pressure transmitter is operating out of limits, is disconnected, or has failed. Occurs if source of the line pressure measurement is set to “Live”, and no pressure transmitter is attached. 0x00000200 Power supply under voltage 0x00000400 System ground fault 0x00000800 The temperature of the laser is not stable. This warning occurs briefly at power on, as the Aurora H2O sets the correct operating temperature. The laser is powered off until the temperature has stabilized. 0x00001000 Aurora H2O has reached the limit for adjusting the signal gain. Contact the factory for assistance. 0x00002000 Aurora H2O has reached the limit for adjusting the laser power. Contact the factory for assistance. 0x00004000 Aurora H2O could not detect any signal from the laser. Contact the factory for assistance. 0x00010000 Aurora H2O could not detect a signal returned from the sample cell, or the signal is below allowed limits. Check the mirror for contamination. 0x00020000 Aurora H2O has detected a failure in the laser temperature control. Contact the factory for assistance. 116 0x00040000 The pressure in the Aurora H2O sample cell is greater than 212 kPa (30.75 psia). Verify regulator and flow settings; check for blocked vent line or excessive back pressure. 0x00000000 The Aurora H2O temperature controller is was set beyond its operating limits. Contact the factory for assistance. 0x1yyyzzzz Extended Error Code Aurora H2O User’s Manual Index A AuroraView Capabilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69 Datalogging with . . . . . . . . . . . . . . . . . . . . . . . . . 85 Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70 Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . 69 Scan Plots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86 Starting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 Trend Plots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86 Trend Tabular Data . . . . . . . . . . . . . . . . . . . . . . . 86 Using Main Menus . . . . . . . . . . . . . . . . . . . . . . . . 79 B Bill of Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 C Clock, Resetting . . . . . . . . . . . . . . . . . . . . . . . . . . . 55 Comm Port Settings . . . . . . . . . . . . . . . . . . . . . . . . . 49 Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 6 Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 D Default Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Display Blank . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105 Dim or Hard to Read . . . . . . . . . . . . . . . . . . . . . 105 Locking/Unlocking . . . . . . . . . . . . . . . . . . . . . . . 68 Setting Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Display, Default. . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Insertion Probe/Regulator Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Setting the Pressure . . . . . . . . . . . . . . . . . . . . . . . 25 Installation Choosing a Site . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Installation, System . . . . . . . . . . . . . . . . . . . . . . . . . 13 K Keypad Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 Lock-Out Switch . . . . . . . . . . . . . . . . . . . . . . . . . 37 Unlocking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 L Lights, Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Low Voltage Directive . . . . . . . . . . . . . . . . . . . . . . . 16 M Magnetic Stylus. . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95 Menu Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93 Menus, Accessing . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Mirror, Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . 95 Modbus Register Map . . . . . . . . . . . . . . . . . . . . . . . . . . . 112 RTU/RS485 Communications . . . . . . . . . . . . . . . 111 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16, 31 E Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . 26 N Numeric Values, Entering . . . . . . . . . . . . . . . . . . . . 38 F Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Filter Element, Replacing . . . . . . . . . . . . . . . . . . 7, 101 O Offset Values, Adjusting . . . . . . . . . . . . . . . . . . . . . 51 G Gas Molecular Weight, Adjusting . . . . . . . . . . . . . . . 53 H Heater EU/ATEX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5, 8 USA/CAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 8 Operation Sample System . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Theory of . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Outline . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 P Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95 Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Pressure Settings, Resetting . . . . . . . . . . . . . . . . . . . 57 I Indicator Lights . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Aurora H2O User’s Manual 117 Index Programming Advanced Features . . . . . . . . . . . . . . . . . . . . . . . . 49 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Menu Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93 Wiring Diagram EU/ATEX Heater . . . . . . . . . . . . . . . . . . . . . . . . . . 8 USA/CAN Heater . . . . . . . . . . . . . . . . . . . . . . . . . 8 Q Quick Startup Guide . . . . . . . . . . . . . . . . . . . . . . . . 34 R Regional Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . 59 Replacing the Filter Element . . . . . . . . . . . . . . . 7, 101 S Sample System Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Quick Startup Guide . . . . . . . . . . . . . . . . . . . . . . . 34 Settings, Regional . . . . . . . . . . . . . . . . . . . . . . . . . . 59 Site Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Starting Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Stylus, Magnetic . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 System Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 T Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . 105 Blank Display . . . . . . . . . . . . . . . . . . . . . . . . . . 105 Display Dim or Hard to Read . . . . . . . . . . . . . . . 105 No Flow Measurement . . . . . . . . . . . . . . . . . . . . 107 Status Messages and Indicators . . . . . . . . . . . . . . 105 Using a Moisture Standard . . . . . . . . . . . . . . . . . 108 Using a Portable Hygrometer . . . . . . . . . . . . . . . 107 Verifying Performance . . . . . . . . . . . . . . . . . . . . 107 U Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 V Verification Period . . . . . . . . . . . . . . . . . . . . . . . . . . 95 W Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26, 32 118 Aurora H2O User’s Manual Warranty Warranty Each instrument manufactured by GE Sensing is warranted to be free from defects in material and workmanship. Liability under this warranty is limited to restoring the instrument to normal operation or replacing the instrument, at the sole discretion of GE Sensing. Fuses and batteries are specifically excluded from any liability. This warranty is effective from the date of delivery to the original purchaser. If GE Sensing determines that the equipment was defective, the warranty period is: • one year from delivery for electronic or mechanical failures • one year from delivery for sensor shelf life If GE Sensing determines that the equipment was damaged by misuse, improper installation, the use of unauthorized replacement parts, or operating conditions outside the guidelines specified by GE Sensing, the repairs are not covered under this warranty. The warranties set forth herein are exclusive and are in lieu of all other warranties whether statutory, express or implied (including warranties or merchantability and fitness for a particular purpose, and warranties arising from course of dealing or usage or trade). Return Policy If a GE Sensing instrument malfunctions within the warranty period, the following procedure must be completed: 1. Notify GE Sensing, giving full details of the problem, and provide the model number and serial number of the instrument. If the nature of the problem indicates the need for factory service, GE Sensing will issue a RETURN AUTHORIZATION NUMBER (RAN), and shipping instructions for the return of the instrument to a service center will be provided. 2. If GE Sensing instructs you to send your instrument to a service center, it must be shipped prepaid to the authorized repair station indicated in the shipping instructions. 3. Upon receipt, GE Sensing will evaluate the instrument to determine the cause of the malfunction. Then, one of the following courses of action will then be taken: • If the damage is covered under the terms of the warranty, the instrument will be repaired at no cost to the owner and returned. • If GE Sensing determines that the damage is not covered under the terms of the warranty, or if the warranty has expired, an estimate for the cost of the repairs at standard rates will be provided. Upon receipt of the owner’s approval to proceed, the instrument will be repaired and returned. Aurora H2O User’s Manual 119 Warranty [no content intended for this page] 120 Aurora H2O User’s Manual GE Sensing DECLARATION OF CONFORMITY DOC-0007, Rev. A GE Sensing 1100 Technology Park Drive Billerica, MA 01821 USA We, declare under our sole responsibility that the Aurora H2O Moisture Analyzer to which this declaration relates, is in conformity with the following standards: • EN 60079-0: 2006 • EN 60079-1: 2007 • EN 60079-7: 2007 • EN 60529: 1991 +A1: 2000 • II 2 G Ex de IIB T6, Ta = -20°C to +65°C, IP66; FM09ATEX0065X (FM Global, UK) • EN 61326-1: 2006, Class A, Table 2, Industrial Locations • EN 61326-2-3: 2006 • EN 61010-1: 2001, Overvoltage Category II, Pollution Degree 2 • IEC 60825-1 following the provisions of the 2004/108/EC EMC, 2006/95/EC Low Voltage and 94/9/EC ATEX Directives. The unit listed above and any ancillary equipment supplied with it do not bear CE marking for the Pressure Equipment Directive, as they are supplied in accordance with Article 3, Section 3 (sound engineering practices and codes of good workmanship) of the Pressure Equipment Directive 97/23/EC for DN<25. Billerica - August 2010 Issued Mr. Gary Kozinski Certification & Standards, Lead Engineer [no content intended for this page] Customer Support Centers U.S.A. The Boston Center 1100 Technology Park Drive Billerica, MA 01821 U.S.A. Tel: 800 833 9438 (toll-free) 978 437 1000 E-mail: [email protected] Ireland Sensing House Shannon Free Zone East Shannon, County Clare Ireland Tel: +353 (0)61 470291 E-mail: [email protected] An ISO 9001:2008 Certified Company www.ge-mcs.com/en/about_us/quality.html www.ge-mcs.com ©2012 General Electric Company. All rights reserved. Technical content subject to change without notice. 910-284 Rev. E ">

Public link updated
The public link to your chat has been updated.
Advertisement
Key features
- Fast Response Time: <2 seconds
- Direct Readout in lbs/mmscf, mg/m3 or ppm
- Process Dew/Frost Point Measurement
- Integrated Sample System for Natural Gas
- Magnetic Stylus for Through-Glass Programming
- Explosionproof/Flameproof Design
- 4-20 mA Signals and RS-232/485 MODBUS RTU
- AuroraView Software for Remote Configuration and Data Logging
- NIST Traceable Calibration
- Conformance with IEC 60825-1 Edition 2.0