User Manual

Add to my manuals
76 Pages

advertisement

User Manual | Manualzz
5000 Series Indicators
Instruction Manual
T51P Indicator
T51XW Indicator
Modification Sheet to this manual p/n 80251400
Section
2.2.1
3.4.1
3.4.4
3.4.5
3.4.9
3.4.10
3.5.2
3.9.1
3.10
5.2
Explanation
For connecting bases with a connector to a T51XW (which does not have the external connector), a Load Cell Cable Adapter
Kit p/n 80500736 is available as an accessory. This kit connects to the terminal block inside the T51XW and has an external
connector on the other end.
Reset the Setup menu to the factory defaults (except Range, Capacity and Graduation)
Note: Range 2 graduation is retained even under half capacity until the scale returns to zero
Note: Units oz, lb:oz and C (custom) will not be valid as Power On units when Range is set to Dual. The next available unit
will be displayed instead.
When Legal for Trade is on, the following Menu settings are effected:
Range, Graduation, Power On unit, Auto-Tare, Retain Zero, Gross Indication, and Print Output settings are locked; Zero
Range is locked at 2%; Stable Range is locked at 1d; AZT is set to 0.5d; Capacity is read-only; Continuous Print is disabled;
Lock Unit and Lock Mode are turned on; Calibration functions except for Cal Test are hidden; IP and CP RS232 commands
are disabled.
Beeper volume settings are Off, Low (default), High
Stable Range settings are: 0.5d, 1d, 2d, 3d, 5d
Note: If LFT is on, the following Print menu settings are not reset: Stable
Handshake default setting is “none”
Output Format:
Field:
Length:
Weight
9
Space*
1
Unit
5
Space*
1
Stability
1
Space*
1
G/N
N
Space*
1
Term. Char(s)
*
*Each field is followed by a single delimiting space (ASCII: 32)
Definitions:
Weight – up to 9 characters, right justified, “-“ at immediate left of most significant character (if negative).
Unit
– up to 5 characters
Stability – “?” character is printed if not stable. If weight is stable, neither “?” nor following space is printed.
G/N
– “NET” printed if weight is net weight, “G”, “B”, or nothing (depending on GROSS menu setting – Sec. 3.5.7) printed if weight is a gross weight.
*Terminating Character(s) – terminating character(s) printed depending on FEED menu setting (CR,LF / 4xCR,LF / ASCII:12, refer also to Sec. 3.9.5).
6.3
8.2
D51Pxxx scales utilize a load cell cable that is plugged onto the indicator. For EC and OIML type approved applications, the
plugged connection must be sealed using the Load Cell Sealing Collar p/n 80500737.
Table 8-3:
Interface Cable/PC 9-pin, T51P
80500525
Interface Cable/PC 25-pin, T51P
80500524
Interface Cable/Printer SF-42, T51P
80500571
Interface Cable/PC 9-pin, T51XW
80500552
Interface Cable/PC 25-pin, T51XW
80500553
Interface Cable/Printer SF-42, T51XW 80500574
Load Cell Cable Adapter Kit
80500736
Load Cell Cable Sealing Collar
80500737
Note: The next revision of this manual will be updated with these modifications.
5000 Series Indicators
EN-
TABLE OF CONTENTS
1.
1.1
1.2
1.3
INTRODUCTION...........................................................................................................................................EN-5
Safety Precautions......................................................................................................................................EN-5
1.1.1 Relay Option Safety Precautions.........................................................................................................EN-5
Overview of Parts and Controls . ..................................................................................................................EN-6
Control Functions......................................................................................................................................EN-10
2.
2.1
2.2
2.3
2.4
2.5
INSTALLATION..........................................................................................................................................EN-11
Unpacking...............................................................................................................................................EN-11
External Connections.................................................................................................................................EN-11
2.2.1 Scale Base with Connector to T51P .................................................................................................EN-11
2.2.2 RS232 Interface Cable to T51P........................................................................................................EN-11
2.2.3 AC Power to T51P .........................................................................................................................EN-11
2.2.4 AC Power to T51XW........................................................................................................................EN-11
2.2.5 Battery Power to T51P.....................................................................................................................EN-11
2.2.6 Mounting Bracket............................................................................................................................EN-12
Internal Connections..................................................................................................................................EN-12
2.3.1 Opening the Housing........................................................................................................................EN-12
2.3.2 Scale Base Without Connector to T51P or T51XW .............................................................................EN-12
2.3.3 RS232 Interface Cable to T51XW.......................................................................................................EN-13
2.3.4 Footswitch to T51P or T51XW...........................................................................................................EN-13
T51P Rear Housing Orientation..................................................................................................................EN-13
Mounting Bracket......................................................................................................................................EN-13
3. SETTINGS.................................................................................................................................................EN-14
3.1 Menu Structure.........................................................................................................................................EN-14
3.2 Menu Navigation......................................................................................................................................EN-16
3.3 Calibration Menu......................................................................................................................................EN-16
3.3.1 Zero Calibration.............................................................................................................................EN-17
3.3.2 Span Calibration............................................................................................................................EN-17
3.3.3 Linearity Calibration.......................................................................................................................EN-18
3.3.4 Calibration Test.............................................................................................................................EN-19
3.3.5 Geographical Adjustment Factor......................................................................................................EN-19
3.3.6 End Calibration..............................................................................................................................EN-19
3.4 Setup Menu..............................................................................................................................................EN-21
3.4.1 Reset............................................................................................................................................EN-21
3.4.2 Range..........................................................................................................................................EN-21
3.4.3 Capacity.......................................................................................................................................EN-21
3.4.4 Graduation....................................................................................................................................EN-22
3.4.5 Power On Unit...............................................................................................................................EN-22
3.4.6 Zero Range...................................................................................................................................EN-22
3.4.7 Auto-Tare......................................................................................................................................EN-23
3.4.8 Retain Weight Data........................................................................................................................EN-23
3.4.9 Legal for Trade..............................................................................................................................EN-23
3.4.10Beeper Volume..............................................................................................................................EN-24
EN-
5000 Series Indicators
TABLE OF CONTENTS (Cont.)
3.4.11 Beeper Signal................................................................................................................................EN-24
3.4.12Button Beeper................................................................................................................................EN-24
3.4.13 End Setup.....................................................................................................................................EN-24
3.5 Readout Menu..........................................................................................................................................EN-24
3.5.1 Reset............................................................................................................................................EN-25
3.5.2 Stabile Range................................................................................................................................EN-25
3.5.3 Filter.............................................................................................................................................EN-25
3.5.4 Auto-Zero Tracking.........................................................................................................................EN-25
3.5.5 Backlight......................................................................................................................................EN-26
3.5.6 Auto Off Timer...............................................................................................................................EN-26
3.5.7 Gross Indicator...............................................................................................................................EN-26
3.5.8 End Readout..................................................................................................................................EN-26
3.6 Mode Menu..............................................................................................................................................EN-26
3.6.1 Reset............................................................................................................................................EN-27
3.6.2 Weighing Mode.............................................................................................................................EN-27
3.6.3 Parts Counting Mode......................................................................................................................EN-27
3.6.4 Parts Counting Optimize.................................................................................................................EN-27
3.6.5 Percent Weighing Mode..................................................................................................................EN-27
3.6.6 Dynamic Weighing Mode...............................................................................................................EN-27
3.6.7 Check Weighing Mode...................................................................................................................EN-28
3.6.8 End Mode.....................................................................................................................................EN-28
3.7 Unit Menu................................................................................................................................................EN-28
3.7.1 Reset............................................................................................................................................EN-28
3.7.2 Kilogram Unit................................................................................................................................EN-28
3.7.3 Gram Unit.....................................................................................................................................EN-28
3.7.4 Pound Unit....................................................................................................................................EN-29
3.7.5 Ounce Unit....................................................................................................................................EN-29
3.7.6 Pound Ounce Unit..........................................................................................................................EN-29
3.7.7 Tonnes Unit...................................................................................................................................EN-29
3.7.8 Custom Unit..................................................................................................................................EN-29
3.7.9 End Unit.......................................................................................................................................EN-30
3.8 GMP Menu...............................................................................................................................................EN-30
3.8.1 Reset............................................................................................................................................EN-30
3.8.2 Date Type.....................................................................................................................................EN-30
3.8.3 Date Set........................................................................................................................................EN-31
3.8.4 Time Type.....................................................................................................................................EN-31
3.8.5 Time Set.......................................................................................................................................EN-31
3.8.6 User ID.........................................................................................................................................EN-32
3.8.7 Project ID......................................................................................................................................EN-32
3.8.8 Scale ID........................................................................................................................................EN-32
3.8.7 End GMP......................................................................................................................................EN-32
3.9 Print 1, Print 2 Menus...............................................................................................................................EN-33
3.9.1 Reset............................................................................................................................................EN-33
3.9.2 Print Stable data Only.....................................................................................................................EN-33
3.9.3 Auto Print......................................................................................................................................EN-33
3.9.4 Print Content Sub-menu..................................................................................................................EN-34
5000 Series Indicators
EN-
TABLE OF CONTENTS (Cont.)
3.9.5 Layout Sub-menu..........................................................................................................................EN-36
3.9.6 List Menu Settings..........................................................................................................................EN-36
3.9.7 End Print......................................................................................................................................EN-36
3.10 COM 1, COM 2 Menus..............................................................................................................................EN-36
3.10.1Reset............................................................................................................................................EN-36
3.10.2Baud............................................................................................................................................EN-37
3.10.3Parity...........................................................................................................................................EN-37
3.10.4 Stop Bit.........................................................................................................................................EN-37
3.10.5 Handshake...................................................................................................................................EN-37
3.10.6Address........................................................................................................................................EN-37
3.10.7Alternate Command Sub-menu........................................................................................................EN-38
3.10.8 End COM 1 or End COM 2..............................................................................................................EN-38
3.11 I/O Menu.................................................................................................................................................EN-38
3.11.1Reset............................................................................................................................................EN-38
3.11.2External Input................................................................................................................................EN-39
3.11.3Input Beep....................................................................................................................................EN-39
3.11.4Relay Output.................................................................................................................................EN-39
3.11.5End..............................................................................................................................................EN-40
3.12 Menu Lock Menu ....................................................................................................................................EN-40
3.12.1Reset............................................................................................................................................EN-40
3.12.2Lock Calibration............................................................................................................................EN-40
3.12.3 Lock Setup....................................................................................................................................EN-41
3.12.4Lock Readout................................................................................................................................EN-41
3.12.5Lock Mode....................................................................................................................................EN-41
3.12.6Lock Unit......................................................................................................................................EN-41
3.12.7Lock Print 1..................................................................................................................................EN-41
3.12.8Lock Print 2..................................................................................................................................EN-41
3.12.9Lock Com 1..................................................................................................................................EN-41
3.12.10 Lock Com 2................................................................................................................................EN-41
3.12.11 Lock GMP...................................................................................................................................EN-42
3.12.12 Lock I/O......................................................................................................................................EN-42
3.12.13 End Lock....................................................................................................................................EN-42
3.13 Key Lock Menu........................................................................................................................................EN-42
3.13.1Reset............................................................................................................................................EN-42
3.13.2Lock All Buttons.............................................................................................................................EN-42
3.13.3Lock Off Button..............................................................................................................................EN-42
3.13.4Lock Zero Button............................................................................................................................EN-42
3.13.5Lock Print Button...........................................................................................................................EN-43
3.13.6Lock Unit Button............................................................................................................................EN-43
3.13.7Lock Function Button......................................................................................................................EN-43
3.13.8Lock Mode Button..........................................................................................................................EN-43
3.13.9Lock Tare Button............................................................................................................................EN-43
3.13.10 Lock Menu Button........................................................................................................................EN-43
3.13.11 End Lock....................................................................................................................................EN-43
3.14 Security Switch.................................................................................................................................EN-43
EN-
5000 Series Indicators
TABLE OF CONTENTS (Cont.)
4.
4.1
4.2
4.3
4.4
4.5
4.6
4.7
4.8
OPERATION..............................................................................................................................................EN-44
Turning Indicator On/Off.............................................................................................................................EN-44
Zero Operation..........................................................................................................................................EN-44
Manual Tare.............................................................................................................................................EN-44
Pre-Set Tare.............................................................................................................................................EN-44
Auto-Tare.................................................................................................................................................EN-44
Changing Units of Measure........................................................................................................................EN-45
Printing Data............................................................................................................................................EN-45
Application Modes....................................................................................................................................EN-45
4.8.1 Weighing......................................................................................................................................EN-45
4.8.2 Parts Counting..............................................................................................................................EN-45
4.8.3 Percent Weighing..........................................................................................................................EN-46
4.8.4 Check Weighing............................................................................................................................EN-47
4.8.5 Dynamic Weighing........................................................................................................................EN-48
5.
5.1
5.2
5.3
SERIAL COMMUNICATION...........................................................................................................................EN-49
Interface Commands.................................................................................................................................EN-49
Output Format..........................................................................................................................................EN-50
Printouts..................................................................................................................................................EN-50
6.
6.1
6.2
6.3
LEGAL FOR TRADE....................................................................................................................................EN-52
Settings...................................................................................................................................................EN-52
Verification...............................................................................................................................................EN-52
Sealing....................................................................................................................................................EN-52
7.
7.1
7.2
7.3
7.4
MAINTENANCE..........................................................................................................................................EN-53
Model T51P Cleaning................................................................................................................................EN-53
Model T51XW Cleaning.............................................................................................................................EN-53
Troubleshootting.......................................................................................................................................EN-53
Service Information...................................................................................................................................EN-54
8.
8.1
8.2
8.3
8.4
TECHNICAL DATA......................................................................................................................................EN-55
Specifications...........................................................................................................................................EN-55
Accessories and Options............................................................................................................................EN-56
Drawings and Dimensions.........................................................................................................................EN-57
Compliance.............................................................................................................................................EN-58
5000 Series Indicators
1.
EN-
INTRODUCTION
This manual contains installation, operation and maintenance instructions for the T51P and T51XW Indicators. Please read this
manual completely before installation and operation.
1.1 Safety Precautions
For safe and dependable operation of this equipment, please comply with the following safety precautions:
•Verify that the input voltage range printed on the data label matches the local AC power to be used.
•Make sure that the power cord does not pose a potential obstacle or tripping hazard.
•Use only approved accessories and peripherals.
•Operate the equipment only under ambient conditions specified in these instructions.
•Disconnect the equipment from the power supply when cleaning.
•Do not operate the equipment in hazardous or unstable environments.
•Do not immerse the equipment in water or other liquids.
•Service should only be performed by authorized personnel.
•The T51XW is supplied with a grounded power cable. Use only with a compatible grounded power outlet.
1.1.1 Relay Option Safety Precautions This equipment may have an optional AC or DC Relay Option board installed. This option allows external devices to be
controlled by the Indicator.
CAUTION: ELECTRICAL SHOCK HAZARD. REMOVE ALL POWER CONNECTIONS TO THE INDICATOR
BEFORE SERVICING OR MAKING INTERNAL CONNECTIONS. THE HOUSING SHOULD ONLY BE OPENED BY
AUTHORIZED AND QUALIFIED PERSONNEL, SUCH AS AN ELECTRICAL TECHNICIAN.
Before making connections to the Relay terminals, remove power from the system. If the system contains an optional
rechargeable battery system, be sure that the ON/ZERO Off button is used to fully turn off the system after removing the AC power
plug.
More detailed installation instructions are included with the Relay Option Kit when purchased.
EN-
5000 Series Indicators
1.2 Overview of Parts and Controls
2
3
4
1
5
6
7
8
9
10
11
12 13 14
15
Figure 1-1. T51P Indicator.
16
TABLE 1-1. T51P PARTS.
Item
Description
1
Data Label
2
Front Housing
3
Control Panel
4
Adjusting Knob (2)
5
Mounting Bracket
6
Security Screw
7
Data Label 8
Rear Housing
9
Battery Cover
10
Screw (4)
11
Power Receptacle
12
Hole plug for option
13
Strain relief for alternate
load cell connection
14
Load Cell Connector
15
Hole plug for option
16
RS232 Connector
5000 Series Indicators
EN-
1.2 Overview of Parts and Controls (Cont.)
1
2
3
4
5
6
7
8
9
10
11
12 13
Figure 1-2. T51XW Indicator.
14
TABLE 1-2. T51XW PARTS.
Item
Description
1
Data Label
2
Front Housing
3
Control Panel
4
Adjusting Knob (2)
5
Mounting Bracket
6
Screw (4)
7
Rear housing
8
Data Label
9
Strain relief for option
10
Strain relief for RS232
11
Strain relief for option
12
Strain relief for option
13
Strain relief for Load Cell
Cable
14
Power cord
EN-
5000 Series Indicators
1.2 Overview of Parts and Controls (Cont.)
1
2
3
4
7
Figure 1-3. Main PC Board.
Item
TABLE 1-3. MAIN PC BOARD.
Description
1
Sense Jumper W1
2
Alternate Load Cell Terminal Block J4
3
Sense Jumper W2
4
Security Switch SW2
5
External input Terminal Block J9
6
RS232 Terminal Block J7 (T51XW only)
7
Load Cell Connector
5
6
5000 Series Indicators
EN-
1.2 Overview of Parts and Controls (Cont.)
1
27
26
25
24
23
22
21
20
4
3
2
56
78 9

19
18
17
16 15 14 13 12 11 10
Figure 1-4. Controls and Indicators.
TABLE 1-4. CONTROL PANEL.
No.
Designation
No.
Designation
1
UNDER LED
15
TARE Menu-Cal button
2
ACCEPT LED
16
Pieces symbol
3
OVER LED
17
FUNCTION Mode button
4
Capacity Label Window
18
PRINT Units button
5
Brackets (not used)
19
ON/ZERO Off button
6
Kilogram, gram symbols
20
7
Scale symbol (not used)
Pointer symbols (not
used)
8
Range symbol
21
Brutto, Gross symbols
9
Percent symbol
22
10
Pound, Ounce, Pound:
ounce symbols
Preset Tare, Tare
symbols
23
Stable weight Indicator
11
Tonne symbol
24
Negative symbol
12
Battery charge symbol
25
Center of Zero Indicator
13
Custom unit symbol
26
NET symbol
14
Dynamic symbol
27
7-segment Display
EN-10
5000 Series Indicators
1.3 Control Functions
TABLE 1-5. CONTROL FUNCTIONS.
Button
Primary Function
(Short Press)
Secondary Function
(Long Press)
PRINT
ON/ZERO
Turns the Indicator Sends the current value
on.
to the selected COM
ports if AUTOPRINT is
If Indicator is On, sets set to Off.
zero.
FUNCTION
Initiates an application
mode.
Off
Turns the Indicator off.
Mode
Allows changing the
application mode.
Units
Changes the weighing
Unit.
TARE
Performs a tare
operation.
Temporarily displays the
active mode’s reference
data.
Menu-Cal
Enter the User menu.
Press and hold allows
scrolling through modes.
Menu Function
(Short Press)
Yes
Accepts the current
setting on the display.
No
Advances to the next
menu or menu item.
Rejects the current
setting on the display
and advances to the
next available setting.
Increments the value.
Back
Moves Back to previous
menu item.
Decrements the value.
Exit
Exits the User menu.
Aborts the calibration in
progress.
5000 Series Indicators
EN-11
2. INSTALLATION
2.1 Unpacking
Unpack the following items:
•T51P or T51XW Indicator
•AC Power Cord (T51P only)
•Mounting Bracket
•Knobs (2)
•Capacity Label Sheet
•LFT Sealing kit
•Instruction Manual CD
•Warranty Card
2.2 External Connections
2.2.1 Scale Base with Connector to T51P
Ohaus bases with a connector can be attached to the external load cell connector (Figure 1-1, item 14). Refer to section 2.3.2
for bases without a connector.
2.2.2 RS232 interface Cable to T51P
Connect the optional RS232 cable to the RS232 connector (Figure 1-1, item 16).
Pin
Connection
1
N/C
2
TXD
3
RXD
4
N/C
5
GND
6
N/C
7
CTS
8
RTS
9
N/C
4
5
9
2
3
8
7
1
6
Figure 2-1. RS232 Pins.
2.2.3 AC Power to T51P
Connect the AC power cord (supplied) to the power receptacle (Figure 1-1, item 11), then connect the AC plug to an electrical
outlet.
2.2.4 AC Power to T51XW
Connect the AC plug to a properly grounded electrical outlet.
2.2.5 Battery Power to T51P
The indicator can be operated on alkaline batteries (not supplied) when AC power is not available. It will automatically switch
to battery operation if there is power failure or the power cord is removed. The indicator can operate for up to 80 hours on
battery power.
Remove the battery cover (Figure 1-1, item 9) and install 6 C-type (LR14)
alkaline batteries in the orientation specified. Re-install the battery cover.
During battery operation, the battery charge symbol indicates the battery status.
The indicator will automatically turn-off when the batteries are fully discharged.
EN-12
5000 Series Indicators
2.2.6 Mounting Bracket
Align the wall bracket over the threaded holes in the side of the indicator and install the knobs. Adjust the indicator to the
desired angle and tighten the knobs.
2.3 Internal Connections
Some connections require the housing to be opened.
2.3.1 Opening the Housing
CAUTION: ELECTRICAL SHOCK HAZARD. REMOVE ALL POWER CONNECTIONS TO THE INDICATOR
BEFORE SERVICING OR MAKING INTERNAL CONNECTIONS. The housing should only be opened
by authorized and qualified personnel, such as an Electrical Technician.
T51P
Remove the four Phillips head screws from the rear housing.
Remove the front housing being careful not to disturb the internal connections.
Once all connections are made, reattach the front housing.
T51XW
Remove the four hex head screws from the rear housing.
Open the housing by carefully pulling the front housing forward.
Once all connections are made, reattach the front housing.
The screws should be tightened fully to maintain a water tight seal.
2.3.2 Scale Base Without Connector to T51P or T51XW
Pin
Connection
Bases without a connector must be attached to the internal load cell connector on the main
PC board. Pass the load cell cable through the strain relief (Figure 1-1, item 13 or Figure 1-2,
item 13) and attach it to terminal block J4 (Figure 1-3, item 2). Tighten the strain relief to
maintain a watertight seal.
J4-1
+EXE
J4-2
+SEN
J4-3
+SIG
J4-4
GND
J4-5
-SIG
J4-6
-SEN
J4-7
-EXE
Jumper Connections
For a 4-wire load cell with no sense wires: Jumpers W1 and W2 must be left in place shorting
the two pins.
For a 6-wire load cell that includes sense wires, Jumpers W1 and W2 must be removed.
For load cells with an extra ground shield wire: Connect the shield to the center position (GND)
of J4. Figure 2-2. Jumper Connections.
After wiring is completed and jumpers are in place, replace the indicator housing screws. Make sure the liquid-tight connector is
properly tightened.
5000 Series Indicators
EN-13
2.3.3 RS232 Interface Cable to T51XW
Pass the optional RS232 cable through the strain relief (Figure 1-2, item 10) and attach it to
terminal block J7 (Figure 1-3, item 6). Tighten the strain relief to maintain a watertight seal.
2.3.4 Footswitch to T51P or T51XW
Pin
Connection
J7-1
RTS
J7-2
TXD
J7-3
RXD
J7-4
CTS
J7-5
GND
Pass the optional footswitch cable through the strain relief (Figure 1-1, item 15 or Figure 1-2, item 11) and attach it to terminal
block J9 (Figure 1-3, item 5).
2.4 T51P Rear Housing Orientation
The T51P is delivered in the wall mount orientation with the connections exiting below the display. The rear housing may be
reversed so the connections exit above the display when the T51P is placed horizontally on a bench. To reverse the rear housing,
remove the four Phillips head screws, carefully rotate the housing 180°, and reinstall the screws.
Figure 2-3. Wall Mount Configuration.
Figure 2-4. Bench Top Configuration.
2.5 Mounting Bracket
Attach the bracket to a wall or table using fasteners (not supplied) that are appropriate for the type of mounting surface. The
bracket will accommodate up to 6 mm (1/4”) diameter screws. Locate the mounting holes as shown in Figure 2-4.
Figure 2-5 Mounting Bracket Dimensions.
EN-14
5000 Series Indicators
3 SETTINGS
3.1 Menu Structure
TABLE 3-1. MENU STRUCTURE.
5000 Series Indicators
EN-15
3.1 Menu Structure (Cont.)
TABLE 3-1. MENU STRUCTURE (Cont.).
EN-16
5000 Series Indicators
3.2 Menu Navigation
Enter the menu by pressing the Menu-Cal button until MENU is displayed. The first menu is displayed. Press the No or Back
button to move to a different menu. Press the Yes button to enter the menu. Once in the menu, press the Yes button to view the
menu item setting or press the No or Back button to move to the next menu item. When viewing the setting, press the Yes button
to accept the setting, or press the No or Back button to change the setting. Once all settings have been made, press the Exit
button to return to the current application mode.
For menu items with alphanumeric settings such as Capacity, the current setting is displayed with all digits
flashing. Press the No button to begin editing.
The first digit is displayed flashing.
Press the No button to increment the digit or press the Yes button to accept the digit and move to the next
digit. Repeat this process for all digits.
Press the Yes button when the last digit has been set. The new setting is displayed with all digits flashing. Press the Yes button to accept the setting or press the No
button to resume editing.
This method also applies to setting Checkweigh under and over targets.
For End menu items, pressing the Yes button advances to the next menu, while pressing the No button returns
to the top of the current menu.
3.3 Calibration Menu
When CAL is displayed, press the Yes button to accept the Calibration menu
selection. Press the No button to advance to the desired calibration menu item. Zero Three calibration processes are available: Zero Calibration, Span Calibration and Span Linearity Calibration. Linearity NOTES:
Cal Test 1. Make sure that appropriate calibration masses are available before Geographic
beginning calibration. Adjustment 2. Make sure that the scale base is level and stable during the entire End Calibration
calibration process. 3. Calibration is unavailable with LFT set to On.
4. Allow the Indicator to warm up for approximately 5 minutes after stabilizing to room temperature.
5. To abort calibration, press the Exit button anytime during the calibration process.
6. When any selection within the GMP menu is enabled, calibration results are automatically printed.
Perform
Perform
Perform
Perform
Set 00…Set 12…Set 31
Exit CALIBRATE menu
5000 Series Indicators
EN-17
3.3.1 Zero Calibration
Zero calibration uses one calibration point. The zero calibration point is established with no weight on the
scale. Use this calibration method to adjust for a different pre-load without affecting the span or linearity
calibration. When ZErO is displayed, press the Yes button to initiate Zero Calibration.
The display flashes 0 and the calibration unit. Press the Yes button to establish the zero point.
The display shows --C-- while the zero point is established.
When zero calibration is completed, the display shows dONE. Then the scale exits to the active weighing mode and displays the actual weight value.

3.3.2 Span Calibration
Span Calibration uses two points to adjust the scale. The span calibration point is established with a
calibration mass placed on the scale. The zero calibration point is established with no weight on the scale.
When SPAN is displayed, press the Yes button to initiate Span Calibration.
The display flashes the span calibration point. Place the specified weight on the scale and press the Yes
button.
To choose a different span point or calibration unit, edit the setting as explained in Section 3.2 Menu
Navigation. When the desired setting is displayed, place the specified weight on the scale and press the Yes
button.
The display shows --C-- while the span point is established.
The display flashes 0.
With no weight on the scale, press the Yes button to establish the zero point.
The display shows --C-- while the zero point is established.
When span calibration is completed, the display shows dONE. Then the scale exits to the active weighing mode and displays the actual weight value.

EN-18
5000 Series Indicators
3.3.3 Linearity Calibration
Linearity calibration uses 3 calibration points. The full calibration point is established with a weight on the
scale. The mid calibration point is established with a weight equal to half of the full calibration weight on
the scale. The zero calibration point is established with no weight on the scale. The mid calibration points
cannot be altered by the user during the calibration procedure.
When LINEAr is displayed, press the Yes button to initiate Linearity Calibration.
The display flashes the full calibration point and calibration unit. Place the specified weight on the scale and
press the Yes button.
To choose a different full point or calibration unit (kg or lb), edit the setting as explained in Section 3.2 Menu
Navigation. When the desired setting is displayed, place the specified weight on the scale and press the Yes
button.
The display shows --C-- while the full point is established.
The display flashes the mid calibration point.
Place the specified weight on the scale and press the Yes button.
The display shows --C-- while the mid point is established.
The display flashes 0.
With no weight on the scale, press the Yes button to establish the zero point.
The display shows --C-- while the zero point is established.
When linearity calibration is completed, the display shows dONE. Then the scale exits to the active weighing mode and displays the actual weight value.

5000 Series Indicators
EN-19
3.3.4 Calibration Test
Calibration test is used to compare a known calibration weight against the stored span calibration data.
NOTE: Calibration Test is always available (even when LFT is set to ON).
When tESt is displayed, press the Yes button to initiate Calibration Test.
The display flashes 0. With no weight on the scale, press the Yes button to record the current zero point.
The display shows --t-- while the zero point is recorded.
The display flashes the span calibration weight using the value from the last calibration. The example shows
test weight of 30 kg.
Place the specified test weight on the scale and press the Yes button. The display shows --t-- while the data is processed.
The display flashes the actual difference between the calibration data and the test weight. The example shows a 0.010 kg difference. The result of the Calibration Test is printed.
After 5 seconds, Calibration Test ends, the scale returns to the active weighing mode and displays the current
weight.

3.3.5 Geographical Adjustment Factor
Refer to table 3-3 and set the GEO factor that corresponds to your location.
00 to 31
•
•
•
3.3.6 End Calibration
Advance to the next menu.
EN-20
5000 Series Indicators
TABLE 3-2. GEOGRAPHICAL ADJUSTMENT VALUES
latitude
Geographical
away from the equator,
(North
or South) in
and
minutes.
degrees
0°00’ - 5°46’ 5°46’ - 9°52’ 9°52’ - 12°44’ 12°44’ - 15°06’ 15°06’ - 17°10’ 17°10’ - 19°02’ 19°02’ - 20°45’ 20°45’ - 22°22’ 22°22’ - 23°54’ 23°54’ - 25°21’ 25°21’ - 26°45’ 26°45’ - 28°06’ 28°06’ - 29°25’ 29°25’ - 30°41’ 30°41’ - 31°56’ 31°56’ - 33°09’ 33°09’ - 34°21’ 34°21’ - 35°31’ 35°31’ - 36°41’ 36°41’ - 37°50’ 37°50’ - 38°58’ 38°58’ - 40°05’ 40°05’ - 41°12’ 41°12’ - 42°19’ 42°19’ - 43°26’ 43°26’ - 44°32’ 44°32’ - 45°38’ 45°38’ - 46°45’ 46°45’ - 47°51’ 47°51’ - 48°58’ 48°58’ - 50°06’ 50°06’ - 51°13’ 51°13’ - 52°22’ 52°22’ - 53°31’ 53°31’ - 54°41’ 54°41’ - 55°52’ 55°52’ - 57°04’ 57°04’ - 58°17’ 58°17’ - 59°32’ 59°32’ - 60°49’ 60°49’ - 62°09’ 62°90’ - 63°30’ 63°30’ - 64°55’ 64°55’ - 66°24’ 66°24’ - 67°57’ 67°57’ - 69°35’ 69°35’ - 71°21’ 71°21’ - 73°16’ 73°16’ - 75°24’ 75°24’ - 77°52’ 77°52’ - 80°56’ 80°56’ - 85°45’ 85°45’ - 90°00’ Elevation above sea level in meters
0 325 650 975 325 650 975 1300 Elevation above sea level in feet
0 1060 2130 3200 1060 2130 3200 4260 5 4 4 3 5 5 4 4 6 5 5 4 6 6 5 5 7 6 6 5 7 7 6 6 8 7 7 6 8 8 7 7 9 8 8 7 9 9 8 8 10 9 9 8 10 10 9 9 11 10 10 9 11 11 10 10 12 11 11 10 12 12 11 11 13 12 12 11 13 13 12 12 14 13 13 12 14 14 13 13 15 14 14 13 15 15 14 14 16 15 15 14 16 16 15 15 17 16 16 15 17 17 16 16 18 17 17 16 18 18 17 17 19 18 18 17 19 19 18 18 20 19 19 18 20 20 19 19 21 20 20 19 21 21 20 20 22 21 21 20 22 22 21 21 23 22 22 21 23 23 22 22 24 23 23 22 24 24 23 23 25 24 24 23 25 25 24 24 26 25 25 24 26 26 25 25 27 26 26 25 27 27 26 26 28 27 27 26 28 28 27 27 29 28 28 27 29 29 28 28 30 29 29 28 30 30 29 29 31 30 30 29 1300 1625 1950 2275 2600 2925 3250
1625 1950 2275 2600 2925 3250 3575
4260 5330 3 3 4 4 5 5 6 6 7 7 8 8 9 9 10 10 11 11 12 12 13 13 14 14 15 15 16 16 17 17 18 18 19 19 20 20 21 21 22 22 23 23 24 24 25 25 26 26 27 27 28 28 29 5330 6400 2 3 3 4 4 5 5 6 6 7 7 8 8 9 9 10 10 11 11 12 12 13 13 14 14 15 15 16 16 17 17 18 18 19 19 20 20 21 21 22 22 23 23 24 24 25 25 26 26 27 27 28 28 6400 7460 8530 9600 10660
7460 8530 9600 10660 11730
2 1 1 0 0
2 2 1 1 0
3 2 2 1 1
3 3 2 2 1
4 3 3 2 2
4 4 3 3 2
5 4 4 3 3
5 5 4 4 3
6 5 5 4 4
6 6 5 5 4
7 6 6 5 5
7 7 6 6 5
8 7 7 6 6
8 8 7 7 6
9 8 8 7 7
9 9 8 8 7
10 9 9 8 8
10 10 9 9 8
11 10 10 9 9
11 11 10 10 9
12 11 11 10 10
12 12 11 11 10
13 12 12 11 11
13 13 12 12 11
14 13 13 12 12
14 14 13 13 12
15 14 14 13 13
15 15 14 14 13
16 15 15 14 14
16 16 15 15 14
17 16 16 15 15
17 17 16 16 15
18 17 17 16 16
18 18 17 17 16
19 18 18 17 17
19 19 18 18 17
20 19 19 18 18
20 20 19 19 18
21 20 20 19 19
21 21 20 20 19
22 21 21 20 20
22 22 21 21 20
23 22 22 21 21
23 23 22 22 21
24 23 23 22 22
24 24 23 23 22
25 24 24 23 23
25 25 24 24 23
26 25 25 24 24
26 26 25 25 24
27 26 26 25 25
27 27 26 26 25
28 27 27 26 26
5000 Series Indicators
EN-21
3.4 Setup Menu
When the Indicator is used for the first time, enter this
menu to set the Range, Capacity and Graduation.
Reset Range Full Scale Capacity Graduation Power On unit Zero Range Auto-Tare Retain Weight Data Legal for Trade Beeper Volume Beeper Signal Button Beep End Setup No, Yes
Single, Dual
1…999950
0.00001…1000
Auto, kg, g, lb, oz, lb:oz
2%, 100%
Off, On, Accept
Off, On
Off, On
Off, Lo, Hi
Off, Accept, Under, Over, Under-Over
Off, On
Exit SETUP menu
3.4.1 Reset
Reset the Setup menu to the factory defaults.
No = not reset.
Yes = reset.
NOTE: If the Legal for Trade menu item is set to ON, the Range, Capacity, Graduation, Zero Range, Auto Tare,
Retain Weight Data and Legal For Trade settings are not reset. 3.4.2 Range
Set the number of weighing ranges.
SINGLE = one weighing range from zero to Capacity.
DUAL = two weighing ranges, Range 1 is from zero to half Capacity and Range 2 is from half
Capacity to Capacity.
3.4.3 Capacity
Set the scale capacity as explained in Section 3.2 Menu Navigation.
NOTE: If dUAL was selected in the rANGE menu item, the Capacity setting defines the Range 2 capacity. The
Range 1 capacity is automatically defined as half of the Capacity setting. For example, if Capacity is set to
15, the Range 1 capacity becomes 7.5.
After the capacity is set, select the Primary Unit.
kg = the primary unit is kilograms
lb. = the primary unit is pounds
EN-22
5000 Series Indicators
3.4.4 Graduation
Set the scale readability. 0.00001, 0.00002, 0.00005, 0.0001, 0.0002, 0.0005, 0.001, 0.002, 0.005, 0.01, 0.02,
0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50, 100, 200, 500, 1000.
NOTE: Graduation settings are limited to values from Capacity divided by 1000 to Capacity divided by
30000. Therefore, not all settings are available for each capacity. NOTE: If dUAL was selected in the rANGE menu item, the Graduation setting defines the Range 1 graduation. The Range 2 graduation is automatically defined as one step greater than the Graduation setting. For
example, if Graduation is set to 0.001, the Range 2 graduation becomes 0.002.
3.4.5 Power On Unit
Set the unit of measures displayed at startup
Auto = last unit in use when turned off
PWr.UN kg = kilograms
PWr.UN g = grams
PWr.UN lb = pounds
PWr.UN oz = ounces
PWr.UN lb:oz= pound ounces
PWr.UN t = tonnes
PWr.UN C = custom unit
3.4.6 Zero Range
Set the percentage of scale capacity that may be zeroed.
2% = zero up to 2 percent of capacity
100% = zero up to full capacity
•
•
•
5000 Series Indicators
3.4.7 Auto-Tare
Set the Automatic Tare functionality.
OFF = Automatic Tare is disabled.
ON = the first stable gross weight will be tared.
ACCEPt = when the application mode is CHECK, stable gross weight that is within the Checkweigh accept limits will be tared.
When Accept is selected, set the current delay time is displayed.
Settings:
OFF = automatic tare takes affect immediately
0.5, 2 or 5 = automatic tare takes affect after the selected delay period (in seconds).
3.4.8 Retain Weight Data
Set the Retain Weight Data functionality. OFF = Disabled.
ON = When power is turned on, the displayed weight is based on the last stored zero (Zero button or “Z” command). 3.4.9 Legal for Trade
Set the legal for trade status.
OFF = standard operation
ON = operation complies with weights and measures regulations
EN-23
EN-24
5000 Series Indicators
3.4.10 Beeper Volume
Set the beeper volume.
Low = soft.
Medium = medium.
High = loud.
3.4.11Beeper Signal
Set how the beeper responds in the Checkweigh mode.
Off = the beeper is disabled.
Accept = the beeper will sound when the weight is within the Accept range.
Under = the beeper will sound when the weight is below the Under setting.
Over = the beeper will sound when the weight is above the Over setting.
Under- Over = the beeper will sound when the weight is below the Under setting or above the Over setting.
3.4.12Button Beeper
Set how the beeper sounds when a button is pressed.
OFF = no sound
ON = sound
3.4.13End Setup
Advance to the next menu.
3.5 Readout Menu
Enter this menu to customize display functionality.
Reset: No, Yes
Stable Range 0.5d, 1d, 2d, 5d
Filter Level Lo, Med, Hi Auto Zero Tracking
Backlight Auto Shut Off Gross Indicator End Readout Off, 0.5d, 1d, 3d
Off, On, Auto (->Set 1, Set 2, Set 5)
Off, Set 1, Set 2, Set 5
Off, Gross, Brutto
Exit READOUT menu
5000 Series Indicators
3.5.1 Reset
Set the Readout menu to factory default settings.
No = not reset
Yes = reset
If the Legal for Trade menu item is set to ON, the Stable Range, Averaging Level, Auto Zero Tracking, Auto Off
and Gross settings are not reset.
3.5.2 Stable Range
Set the amount the reading can vary before the stability symbol turns off.
0.5d = 0.5 scale division
1d = 1 scale division
2d = 2 scale divisions
5d = 5 scale divisions
NOTE: When LFT is set to ON, the setting is forced to 1 d. The setting is locked when the hardware lock
switch is set to the ON position.
3.5.3 Filter
Set the amount of signal filtering.
LO = less stability, faster stabilization time (<1 sec.)
MEd = normal stability, stabilization time (<2 sec.)
HI = greater stability, slower stabilization time (<3 sec.)
3.5.4 Auto-Zero Tracking
Set the automatic zero tracking functionality.
OFF = disabled.
0.5 d = the display will maintain zero until a drift of 0.5 divisions per second has been exceeded.
1 d = the display will maintain zero until a drift of 1 division per second has been exceeded.
3 d = the display will maintain zero until a drift of 3 divisions per second has been exceeded.
NOTE: When the LFT menu item is set to ON, the selections are limited to 0.5d and 3d. The setting is
locked when the hardware lock switch is set to the ON position.
EN-25
EN-26
5000 Series Indicators
3.5.5 Backlight
Set the display backlight functionality.
OFF = always off.
ON = always on.
AUtO = turns on when a button is pressed or the displayed weight changes. When Auto is selected, set Backlight shut off time.
Settings:
SEt 1 = backlight turns off after 1 minute of no activity.
SEt 2 = backlight turns off after 2 minute of no activity.
SEt 5 = backlight turns off after 5 minute of no activity.
3.5.6 Auto Off Timer
Set the automatic shut off functionality.
OFF = disabled
SEt 1 = powers off after 1 minute of no activity.
SEt 2 = powers off after 2 minutes of no activity.
SEt 5 = powers off after 5 minutes of no activity.
3.5.7 Gross Indicator
Set the type of gross indicator.
OFF = disabled
G GrOSS = the G icon is lit when gross weights are displayed.
B brutto = the B icon is lit when gross weights are displayed.
3.5.8 End Readout
Advance to the next menu.
3.6 Mode Menu
Enter this menu to activate the desired application
modes.
Reset: No, Yes
Weigh: Count: Percent: Dynamic: Checkweigh:
End Mode Off, On
Off, On (-> Piece weight optimization (-> On, Off))
Off, On
Off, Manual (-> Set 0 … Set 60), Semi-automatic
(-> Set 0 … Set 60), Automatic (-> Set 0 … Set 60)
Off, On
Exit MODE menu
5000 Series Indicators
3.6.1 Reset
Set the Mode menu to the factory defaults.
No = not reset.
Yes = reset.
NOTE: If the Legal for trade menu item is set ON, the settings are not reset.
3.6.2 Weighing Mode
Set the status.
OFF = Disabled
ON = Enabled
3.6.3 Parts Counting Mode
Set the status.
OFF = Disabled
ON = Enabled
3.6.4 Parts Counting Optimize
Set the status.
OFF = Disabled
ON = Enabled
3.6.5 Percent Weighing Mode
Set the status.
OFF = Disabled
ON = Enabled
3.6.6 Dynamic Weighing Mode
Set the status.
OFF = Disabled
MAN = averaging and resetting are initiated manually by pressing the FUNCTION button.
SEMI = averaging is automatically initiated when the display is stable; resetting is initiated by pressing the FUNCTION button.
AUtO = averaging is automatically initiated when the display is stable; resetting is automatically initiated when the weight on the display is < 5 divisions.
If MAN, SEMI or AUtO is selected, the current level setting is displayed.
Set the averaging time.
SEt 0 = the first stable weight will be held on the display until it is reset (display hold).
SEt 1 = the weight readings will be averaged for 1 second. The average will be held on the display until it is reset.
SEt 60 = the weight readings will be averaged for 60 seconds. The average will be held on the display until it is reset.
EN-27
EN-28
5000 Series Indicators
3.6.7 Check Weighing Mode
Set the status.
OFF = Disabled
ON = Enabled
3.6.8 End Mode
Advance to the next menu.
3.7 Unit Menu
Enter this menu to activate the desired units. Default settings are
bold.
Note: Due to national laws, the indicator may not include some of
the units of measure listed.
3.7.1 Reset
Set the Unit menu to the factory defaults.
Settings:
NO = not reset.
YES =reset
If the Legal for Trade menu item is set ON, the settings are not reset.
3.7.2 Kilogram Unit
Set the status.
OFF = Disabled
ON = Enabled
3.7.3 Gram Unit
Set the status.
OFF = Disabled
ON = Enabled
Reset: Kilograms: Pounds: Grams: Ounces: Pounds:Ounces Tonnes: Custom: End Unit No, Yes
Off, On
Off, On
Off, On
Off, On
Off, On
Off, On
Off, On (-> Factor, Exponent, LSD)
Exit UNIT menu
5000 Series Indicators
EN-29
3.7.4 Pound Unit
Set the status.
OFF = Disabled
ON = Enabled
3.7.5 Ounce Unit
Set the status.
OFF = Disabled
ON = Enabled
NOTE: Ounce Unit is not available when Range is set to Dual.
3.7.6 Pound Ounce Unit
Set the status.
OFF = Disabled
ON = Enabled
NOTE: Pound Ounce Unit is not available when Range is set to Dual.
3.7.7 Tonnes Unit
Set the status.
OFF = Disabled
ON = Enabled
3.7.8 Custom Unit
Use Custom Unit to define an alternate unit of measure, where Custom unit = Factor x Exponent x grams. For example: 1 troy ounce = 373.2417216 grams, so Factor = 3.73242, Exponent = 2.
Set the status.
OFF = Disabled
ON = Enabled
NOTE: Custom Unit is not available when Range is set to Dual.
Factor
Set the conversion factor.
0.00001 to 9.99999
Refer to Section 3.2 Menu Navigation to enter settings. Exponent
Set the factor multiplier.
0 = Factor x 1 1 = Factor x 10
2 = Factor x 100
3 = Factor x 1000
-2 = Factor ÷ 100
-1 = Factor ÷ 10
•
•
•
EN-30
5000 Series Indicators
Least Significant Digit
Set the custom unit readability. 0.00001, 0.00002, 0.00005, 0.0001, 0.0002, 0.0005, 0.001, 0.002, 0.005, 0.01, 0.02,
0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 50, 100, 200, 500, 1000
NOTE: LSD settings are limited to values that result in a displayed resolution of 1000 to 30000 divisions.
•
•
•
3.7.9 End Unit
Advance to the next menu.
3.8 GMP Menu
Enter this menu to set the data for Good Manufacturing Practice.
Reset No, Yes
Date Time User ID Project ID
Scale ID End GMP
3.8.1 Reset
Set the GMP menu to factory defaults.
NO =not reset.
YES = reset.
3.8.2 Date Type
Set the date format.
MDY = Month.Day.Year
DMY = Day.Month.Year
YMD = Year.Month.Day
Type (->MDY, DMY, YMD)
Set 00.00.00 … 99.99.99
Type (-> 24 hr, 12 hr)
Set HH:MM or HH:MM A/P
000000 … 999999
000000 … 999999
000000 … 999999
Exit GMP menu
5000 Series Indicators
EN-31
3.8.3 Date Set
Set the date. 00 to 99 = year position
01 to 12 =month position
01 to 31 = day position
Refer to Section 3.2 Menu Navigation to enter settings.
3.8.4 Time Type
Set the time format.
24 hr = 24 hour format.
12 hr = 12 hour format.
3.8.5 Time Set
Set the time. 24 hour format
00 to 24 =hour position
00 to 59 = minute position
(current time blinking)
(Set hours 00 to 23)
(Set minutes 00 to 59)
12 hour format
12 hr = 12 AM to 12 PM hour position
= 00 to 59 minute position
Refer to Section 3.2 Menu Navigation to enter settings.
(current time blinking)
(Set hours 01 to 12 A or P)
(Set minutes 00 to 59)
EN-32
3.8.6 User ID
Set the user identification.
000000 to 999999
Refer to Section 3.2 Menu Navigation to enter settings.
3.8.7 Project ID
Set the Project identification.
000000 to 999999
Refer to Section 3.2 Menu Navigation to enter settings.
3.8.8 Scale ID
Set the Scale identification.
000000 to 999999
Refer to Section 3.2 Menu Navigation to enter settings.
3.8.9 End GMP
Advance to the next menu.
5000 Series Indicators
5000 Series Indicators
EN-33
3.9 Print 1 and Print 2 Menus
The table shows the items in the Print menu. Default settings are bold.
Enter this menu to define printing parameters.
NOTE: The Print 2 menu is only displayed if a second interface
(RS232 or RS422/RS485) is installed.
3.9.1 RESET
Set the Print menu to factory defaults. NO = not reset.
YES = reset.
NOTE: If the Legal for Trade menu item is set to ON, the following
settings are not reset: Stable, Auto Print
Reset Stable Only Auto Print Print Content
Layout feed)
List End Print1 (End Print2)
No, Yes
Off, On
Off,
On Stable (-> Load, Load and Zero),
Interval (-> 0…3600), Continuous,
On Accept
Result (-> Off, On, Numeric only),
Gross (-> Off, On),
Net (-> Off, On),
Tare (-> Off, On),
Header ( ->Off, On),
User ID (-> Off, On),
Project ID (-> Off, On),
Scale ID (-> Off, On),
Difference (-> Off, On),
Date and Time (-> Off, On),
Information (-> Off, On),
Application Mode ( Off, On),
Name (-> Off, On),
Format (-> Multiple, Single),
Feed (-> Line feed, 4 Line feed, Form
No, Yes
Exit PRINT1 menu
Exit PRINT2 menu
3.9.2 Print Stable Data Only
Set the print critera.
OFF = values are printed immediately.
ON = values are only printed when the stability criteria are met.
3.9.3 Auto Print
Set the automatic printing functionality.
OFF = disabled.
ON.StAb = printing occurs each time the stability criteria are met.
INtEr = printing occurs at the defined interval.
CONt = printing occurs continuosly.
ACCEPt = printing occurs each time the display is within the Checkweigh accept range and stability criteria are met.
EN-34
When INtEr is selected, set the Print Interval.
1 to 3600 (seconds)
3.9.4 Print Content Sub-menu
This sub-menu is used to define the content of the printed data.
Result
Set the status. OFF = Disabled
ON = the displayed reading is printed.
NUM = only the numeric portion of the displayed reading is printed.
Gross
Set the status. OFF = Disabled.
ON = the Gross weight is printed.
Net
Set the status.
OFF = Disabled.
ON = the Net weight is printed.
Tare
Set the status.
OFF = Disabled.
ON = the Tare weight is printed.
Header
Set the status. OFF = Disabled.
ON = the Header is printed.
User ID
Set the status. OFF = Disabled.
ON = the User ID is printed.
5000 Series Indicators
5000 Series Indicators
Project ID
Set the status.
OFF = Disabled.
ON = the Project ID is printed.
Scale ID
Set the status.
OFF = Disabled.
ON = the Scale ID is printed.
Time
Set the status.
OFF = Disabled.
ON = the Date and Time is printed.
Difference
Set the status.
OFF = Disabled.
ON = the Calibration Test difference is printed.
Reference Information
Set the status.
OFF = Disabled.
ON = the Reference Information is printed.
NOTE:
Mode
Set the status. OFF = Disabled.
ON = the Mode is printed.
Name
Set the status. OFF = Disabled.
ON = the Name line is printed.
EN-35
EN-36
5000 Series Indicators
3.9.5 Layout Sub-menu
This sub-menu is used to define format of data output to a printer or computer.
Format
Set the printing format.
MULtI = a multi-line (single column style) printout is generated. A CRLF is added after each item.
SINGLE = a single line printout is generated. (A TAB space is added between each item and a CLRF is used only after the very last item.)
Line Feed
Set the paper feed.
Settings:
LINE = move paper up one line after printing
4.LINE = move paper up four lines after printing
FOrM = a form feed is appended to the printout
3.9.6 List Menu Settings
Print the menu settings.
NO = do not print.
YES = print.
3.9.7 End Print
Advance to the next menu.
3.10 COM 1 and COM 2 Menus
The table shows the items in the communication menus. Default settings are bold.
Enter the menu to define communication parameters. NOTE: The COM2 menu is only displayed if a
second interface (RS232 or RS422/RS485) is
installed. 3.10.1Reset
Set the COM1 and COM2 menu to factory defaults.
NO = not reset.
YES = reset.
Reset: Baud Rate: Parity: Handshake: Address: Alt Command: Zero End Com1 (End Com2) No, Yes
300, 600, 1200, 2400, 4800, 9600, 19200
7 Even, 7 Odd, 7 None, 8 None
Off, XON/XOFF, Hardware
Off, 01,…, 99
Print (-> Off, A … P … Z), Tare (-> Off, A … T … Z),
(-> Off, A … Z)
Exit COM1 menu
Exit COM2 menu
5000 Series Indicators
EN-37
3.10.2Baud
Set the Baud rate.
300 600 1200 2400 4800 9600 19200 = 300 bps
= 600 bps
=1200 bps
= 2400 bps
= 4800 bps
= 9600 bps
= 19200 bps
3.10.3Parity
Set the data bits and parity. 7 EVEN = 7 data bits, even parity.
7 Odd = 7 data bits, odd parity.
7 NONE = 7 data bits, no parity.
8 NONE = 8 data bits, no parity.
3.10.4 Stop Bit
Set the number of stop bits.
1 = 1 stop bit.
2 = 2 stop bits.
3.10.5Handshake
Set the flow control method.
NONE = no handshaking.
ON-OFF = XON/XOFF software handshaking.
HArd = hardware handshaking.
3.10.6Address
Set the communication address.
NOTE: Address is only displayed in the COM2 menu if the RS422/RS485 option is installed.
OFF = no address.
01 to 99 = address 01 to 99
•
•
•
EN-38
5000 Series Indicators
3.10.7Alternate Command Sub-menu
Enter this sub-menu to set a different command character for the P (Print), T (Tare) and Z (Zero)
commands.
Alternate Print Command
set the alternate command character for Print.
A to Z. Alternate Tare
Set the alternate command character for Tare.
A to Z. Alternate Zero
Set the alternate command character for Zero.
A to Z. 3.10.8 End COM1 or End COM2
Advance to the next menu.
3.11 I/O Menu
Enter this menu to set the optional input and output device parameters.
Default settings are bold.
Reset No, Yes
External Input
Input Beep Relay Output
End.I-O 3.11.1 Reset
Set the I/O menu to factory defaults
NO = not reset.
YES = reset.
Off, Tare, Zero, Print, Function, Start-Stop, Tare-Start-Stop
Off, On
Type (-> Open, Closed),
Sequence (-> Normal, Hold),
Contact (-> Simultaneous, Break-
Before-Make, Make-Before-Break)
When Stable (-> Off, On)
Exit I-O menu
5000 Series Indicators
3.11.2 External Input
set the function to be controlled by an optional external input device such as a foot switch.
OFF = disabled.
tAre = Tare function.
ZErO = Zero function.
PrINt = Print function.
FUNCt = action specific to the current application mode.
Start-Stop = the first external input changes the state of the relay. The second external input (S-S) returns the relay to the original state. Tare-Start-Stop = the first external input initiates a Tare function, the second external input (t-S-S) changes the state of the relay. The third external input returns the relay to its original state. 3.11.4 Input Beep
Set the beeper response to an external input.
OFF = Disabled. ON = Enabled.
3.11.4 Relay Output
Set the relay output parameters.
NOTE: If the Relay option is not installed the OUTPUT menu and associated menu items are not available.
Type
Set the initial state of the relay.
OPEN = the relay output is normally open. CLOSEd = the relay output is normally closed.
CAUTION: The normally closed relay condition is only active while the Indicator is powered on. When powered off or when power is removed, the relay condition returns to a normally open condition. Restoring power to the Indicator will restore the closed condition of the relays.
Output Sequence
Set how the relay outputs react as the weight reading changes from under / accept / over.
NOrM = the previously enabled relay will be disabled as the next relay is enabled.
HOLd = the previously enabled relay will hold the same state as the next relay is enabled.
EN-39
EN-40
5000 Series Indicators
Contact
Set the timing of the relay contacts.
NOTE: A 100 ms delay or over-lap is used for the bbM and Mbb timing.
SIM = relays open or close at the same time. b-b-M = relay opens before the next relay closes.
M-b-b = relay closes before the next relay opens.
Stable
Set how the relay outputs react during instability. OFF = relay changes are immediate. ON = delays relay changes until weight reading is stable.
3.11.5 End
Advance to the next menu.
3.12 Menu Lock Menu
Enter this menu. Default settings are bold.
3.12.1Reset
Set the menu Lock menu to factory defaults.
NO = not reset.
YES = reset.
NOTE: Settings for LFT controlled menu items are not reset.
3.12.2Lock Calibration
Set the status.
OFF = Calibration menu is not locked.
ON = Calibration menu settings is locked.
Reset: Lock Calibration Menu Lock Setup Menu Lock Readout Menu Lock Mode Menu Lock Unit Menu Lock Print-1 Menu Lock Print-2 Menu Lock Com-1 Menu Lock Com-2 Menu Lock GMP Menu Lock I/O Menu End Lock Menu
No, Yes
Off, On
Off, On
Off, On
Off, On
Off, On
Off, On
Off, On
Off, On
Off, On
Off, On
Off, On
5000 Series Indicators
3.12.3Lock Setup
Set the status.
OFF = Setup menu is not locked.
ON = Setup menu is locked.
3.12.4Lock Readout
Set the status.
OFF = Readout menu is not locked.
ON = Readout menu is locked.
3.12.5 Lock Mode
Set the status.
OFF = Mode menu is not locked.
ON = Mode menu is locked..
3.12.6Lock Unit
Set the status.
OFF = Unit menu is not locked.
ON = Unit menu is locked.
3.12.7Lock Print 1
Set the status.
OFF = Print 1 menu is not locked.
ON = Print 1 menu is locked.
3.12.8Lock Print 2
Set the status.
OFF = Print 2 menu is not locked.
ON = Print 2 menu is locked.
3.12.9Lock Com 1
Set the status.
OFF = COM1 menu is not locked.
ON = COM1 menu is locked.
3.12.10 Lock Com 2
Set the status.
OFF = COM2 menu is not locked.
ON = COM2 menu is locked.
EN-41
EN-42
5000 Series Indicators
3.12.11 Lock GMP
Set the status.
OFF = GMP menu is not locked.
ON = GMP menu is locked.
3.12.12Lock I-O
Set the status.
OFF = I-O menu is not locked.
ON = I-O menu is locked.
3.12.13End Lock
Advance to the next menu.
3.13 Key Lock Menu
Enter this menu to lock buttons. Default settings are bold.
3.13.1Reset
Set the Key lock menu to factory defaults.
NO = not reset.
YES = reset.
3.13.2Lock all Buttons
Set the status.
OFF = all buttons unlocked.
ON = all buttons are locked.
3.13.3 Lock Off Button
Set the status.
OFF = Off button is unlocked.
ON = Off button is locked.
3.13.4Lock Zero Button
Set the status.
OFF = Zero button is unlocked.
ON = Zero button is locked.
Reset Lock All Buttons Lock Off Button Lock Zero Button Lock Print Button Lock Unit Button Lock Function Button
Lock Mode Button Lock Tare Button Lock Menu Button End Lock Button
No, Yes
Off, On
Off, On
Off, On
Off, On
Off, On
Off, On
Off, On
Off, On
Off, On
5000 Series Indicators
EN-43
3.13.5Lock Print Button
Set the status.
OFF = Print button is unlocked.
ON = Print button is locked.
3.13.6 Lock Unit Button
Set the status. OFF = Unit button is unlocked.
ON = Unit button is locked.
3.13.7Lock Function Button
Set the status.
OFF = Function button is unlocked.
ON = Function button is locked.
3.13.8Lock Mode Button
Set the status.
OFF = Mode button is unlocked.
ON = Mode button is locked.
3.13.9Lock Tare Button
Set the status.
OFF = Tare button is unlocked.
ON = Tare button is locked.
3.13.10 Lock Menu Button
Set the status.
OFF = Menu button is unlocked.
ON = Menu button is locked.
NOTE: When the Menu button is locked, the user may unlock this button by holding the Menu button for 10
seconds until UNLOCK is displayed. The hardware Lock Switch must be in the unlocked position.
3.13.11 End Lock
Advance to the next menu.
3.14 Security Switch
A slide switch is located on the Main PCB board. When the switch is set to the on position, user menu settings that were locked
in the Menu Lock and Key Lock menus can be viewed but not changed.
Open the housing as explained in Section 2.3.1. Set the position of security switch SW2 to ON as shown in Figure 1-3.
EN-44
5000 Series Indicators
4 OPERATION
4.1 Turning Indicator On/Off
To turn the Indicator on, press the ON/ZERO Off button. The Indicator performs a display test followed by a
series of informational displays, and then enters the active weighing mode.

To turn the Indicator off, press and hold the ON/ZERO Off button until OFF is displayed.
4.2 Zero Operation
Zero can be set under the following conditions:
•Automatically at Power On (initial zero).
•Semi-automatically (manually) by pressing the ON/ZERO Off button.
•Semi-automatically by sending the Zero command (Z or alternate zero command).

Press the ON/ZERO Off button to zero the weight display. The scale must be stable to accept zero operation.
4.3 Manual Tare
When weighing an item that must be held in a container, taring stores the container weight in memory. Place the empty container on the scale (example 0.5 kg) and press the TARE button. The display will show
the net weight.
To clear the Tare value, empty the scale and press the TARE button. The display will show the gross weight.



4.4 Pre-Set Tare
A Pre-set Tare (PT) is a known tare value entered using the xT command (example 1.234 kg).
The display will show the Pre-set Tare as a negative value, with the PT Indicator on.
NOTES: 1.The PT value will supersede any other Tare or PT value in memory. 2.When using Pre-Set Tare, make sure that Auto-Tare function is set off in the Setup menu.
3.If the Tare entry includes digits beyond the readability of the Indicator, the tare value is rounded off to the readability of the Indicator.
To clear a Pre-set Tare value, empty the scale then press the TARE button. The display will show the Gross
weight.
4.5 Auto-Tare
Auto-Tare automatically tares the initial weight (such as a container) placed on the empty scale, without
having to press the TARE button. The tare value is cleared automatically when the weight on the scale is fully
removed.
During Checkweighing operation, if the On Accept setting is selected in the Setup menu, weight values that are
within the accept range will be tared automatically.
NOTE: Auto-Tare supersedes any pre-set (PT) value in memory.

5000 Series Indicators
EN-45
4.6 Changing Units of Measure
Press and hold the PRINT Units button until the desired measuring unit appears. Only measuring units enabled in the Unit Menu
will be displayed (refer to Section 3.7).
4.7 Printing Data
Printing the displayed data to a printer or sending the data to a computer requires that the communication parameters in the Print
and Communication Menu are set (refer to Sections 3.9 and 3.10).
Press the PRINT Units button to send the displayed data to the communication port (the Auto-Print Mode in Section 3.9 function
must be Off). 4.8 Application Modes
Press and hold the FUNCTION button until the desired application mode appears. Only modes enabled in the
mode menu will be displayed (refer to Section 3-8).
4.8.1Weighing
Place the item to be weighed on the scale. The illustration indicates a sample of 1.5 kg, Gross weight.

NOTE: Press the FUNCTION button to temporarily display the weight in 10x expanded resolution.

4.8.2 Parts Counting
Use this mode to count parts of uniform weight. The Indicator determines the quantity based on the average
weight of a single part. All parts must be uniform in weight for accurate measurements.
Average Piece Weight (APW)
When the Mode button is released, CLr.PW Pcs is displayed.
Clearing a Stored APW
Press the Yes button to clear the stored APW.
Recalling a Stored APW
Press the No button to recall the existing APW.
NOTE: Press the FUNCTION button to temporarily display the APW value.
EN-46
5000 Series Indicators
Establishing the Average Piece Weight (APW) (Cont.)
The display shows the sample size Put10Pcs.
Establishing a New APW
Press the No button to increment the sample size. Choices are 5, 10, 20, 50 and 100. To establish the APW, place the specified quantity of samples on the scale and press the FUNCTION button to
capture the weight.
APW is displayed shortly followed by the APW value with the current unit of measure.
Begin Counting
Place the parts on the scale and read the count. If a container is used, be sure to tare the empty container
first.

4.8.3 Percent Weighing
Use this mode to measure the weight of a sample as a percentage of a reference weight. Reference Weight (Ref Wt)
When the Mode button is released, CLr.rEF% is displayed.
Clearing a Stored Reference Weight
Press the Yes button to clear the stored reference weight.
Recalling a Stored Reference Weight
Press the No button to recall the existing reference weight.
NOTE: Press the FUNCTION button to temporarily display the reference weight.
Establishing a New Reference Weight
The display shows Put.rEF %.
To establish the Ref Wt, place the sample on the scale and press the FUNCTION button to capture the weight. rEF.Wt is displayed shortly followed by the REF Wt value with the current unit of measure.
Begin Percent Weighing
Place the sample on the scale, and read the percent value. If a container is used, be sure to tare the empty
container first.

5000 Series Indicators
EN-47
4.8.4 Check Weighing
Use this mode to determine if the weight of a sample is within prescribed limits.
Checkweighing Limits
When the Mode button is released, CLr.rEF is displayed.
Clearing Stored Check Weighing Limits
Press the Yes button to clear the stored limits.
Recalling Stored Check Weighing Limits
Press the No button to recall the stored limits.
NOTE: Press the FUNCTION button to temporarily display the Under and Over Limit values. Editing the Under Setting
The display shows SEt LO. Press the Yes button to edit setting..
Settings:
-999950 to 999950
Refer to Menu Navigation Section 3.2 to enter settings.
to
NOTE: The minus sign is used together with the first digit to show a negative value. Editing the Over Setting
The display shows Set.HI.
Press the Yes button to edit the Over setting.
Settings:
-999950 to 999950
Refer to Menu Navigation Section 3.2 to enter settings.
Begin Check Weighing
The appropriate Under, Accept or Over LED lights to indicate Check Weigh status.

Place a sample on the scale and read the weight.
For loads less than the Under Limit, the yellow Under LED is lit.

For loads greater than the Under Limit and less than the Over limit, the green Accept LED is lit.

For loads greater than the Over Limit, the red Over LED is lit.

EN-48
5000 Series Indicators
4.8.5 Dynamic Weighing
Use this mode to weigh moving or oversized objects. The weight is held on the display until reset. Manual,
semi-automatic and automatic start/stop methods are available (refer to Section 3.6.6).
Begin Dynamic Weighing
The display alternates between the current weight and rEAdY.
Place the object on the scale. If the manual mode is in use, press the FUNCTION button to start
measurement. If the semi-automatic or automatic mode is in use, measurement is started automatically.
•
•
•
NOTE: When using manual mode, it is not necessary for the display to be at zero gross or net. When using
semi-automatic or automatic mode, the display must be at zero gross or net before placing the object on
the scale. The example is for a setting of 5 seconds. During the averaging period, the countdown timer
decreases in one second increments. NOTE: If SEt 0 was selected in the Dynamic menu item, the countdown timer is not displayed.
When the countdown has completed, the readings are averaged and held on the display. The averaged weight is displayed until reset.
If the manual or semi-automatic mode is in use, reset the countdown timer by pressing the FUNCTION button. The display alternates between the current weight and rEAdY.
If the automatic mode is in use, the held reading is shown on the display for 10 seconds after the object is
removed. The display alternates between zero and rEAdY.
The scale is now ready to accept a new object.


5000 Series Indicators
EN-49
5 SERIAL COMMUNICATION
The T51P and T51XW Indicators include an RS232 serial communication interface. The setup of RS232 operating parameters are more fully explained in Section 3.10. The physical hardware connection is
explained in in Section 2.6.
The interface enables display and GMP data to be sent to a computer or printer. A computer can be used to control some
functions of the indicator using the commands listed in Table 5-1.
5.1 Interface Commands
Communicate to the indicator using the command characters listed in Table 5-1.
TABLE 5-1. SERIAL INTERFACE COMMAND TABLE.
Command
Character 1)
Function
IP
Immediate Print of displayed weight (stable or unstable).
P 2)
Print displayed weight (stable or unstable).
CP
Continuous Print.
SP
Print on Stability.
xP
Interval Print x = Print Interval (1-3600 sec)
Z
2)
Same as pressing Zero button
T
2)
Same as pressing Tare button
xT
Download Tare value in grams (positive values only). Sending 0T clears tare (if allowed)
PU
Print current unit: g, kg, lb, oz, lb:oz, t, C (custom)
xU
Set scale to unit x: 1=g, 2=kg, 3=lb, 4=oz, 5=lb:oz, 6=t, 7=C
PV
Version: print name, software revision and LFT ON (if LFT is set ON).
H x “text”
Enter Header line , where x = line number 1 to 5, “text” = header text up to 24 alphanumeric
characters
Esc R
Global reset to reset all menu settings to the original factory defaults
NOTES:
1) Commands sent to the Indicator must be terminated with a carriage return (CR) or carriage return-line feed (CRLF).
2) Alternate command characters may be defined by the user (see Alternate Commands in Section 3.10). 3) Data output by the Indicator is always terminated with a carriage return-line feed (CRLF).
EN-50
5000 Series Indicators
5.2 Output Format
The default serial output format is shown below.
Field: Polarity Space Weight Space Units Stability CR LF
Length: 1 1
7 1 5 1
1 1
Definitions: Polarity, “-” sign if negative, blank if positive.
Weight, up to 6 numbers and 1 decimal, right justified, leading zero blanking.
Units, up to 5 characters.
Stability, “?” character is printed if not stable, blank if stable.
NOTE: If the Print Content-Numeric Only is set to On, the Units and Stability fields are omitted.
5.3 Printouts
The following sample print outs are generated by the Print button, “P” Command or alternate print command. The content of the
printout is defined in the Print Content menu item. A maximum of 24 characters can be printed on each line.
NOTE: Shaded areas = if set on in GMP menu.
Unshaded = typical
Weigh Mode Printout
OHAUS CORPORATION
MODEL T51P
Indicator
S/N 1234567890
-----------------------01/31/07 12:30 PM
SCALE ID 123456
USER ID 234567
PROJ ID 345678
NAME _____________
12.34 KG 12.34 KG G
11.11 KG NET
1.22 KG T
MODE: WEIGH
Count Mode Printout
OHAUS CORPORATION
MODEL T51P
Indicator
S/N 1234567890
-----------------------01/31/07 12:30 PM
SCALE ID 123456
USER ID 234567
PROJ ID 345678
NAME _____________
12.34 KG 12.34 KG G
11.11 KG NET
1.22 KG T
APW 0.012 kg
MODE: COUNT
Percent Mode Printout
OHAUS CORPORATION
MODEL T51P
Indicator
S/N 1234567890
-----------------------01/31/07 12:30 PM
SCALE ID 123456
USER ID 234567
PROJ ID 345678
NAME _____________
12.34 KG 12.34 KG G
11.11 KG NET
1.22 KG T
REF WT. 0.012 kg
MODE: PERCENT
5000 Series Indicators
Check Weighing Mode Printout
OHAUS CORPORATION
MODEL T51P
Indicator
S/N 1234567890
-----------------------01/31/07 12:30 PM
SCALE ID 123456
USER ID 234567
PROJ ID 345678
NAME _____________
12.34 kg 12.34 kg G
11.11 kg NET
1.22 KG T
UNDER TARGET 1.00 KG
OVER TARGET 2.00 KG
MODE: CHECKWEIGH
EN-51
Dynamic Mode Printout
OHAUS CORPORATION
MODEL T51P
Indicator
S/N 1234567890
01/31/07 12:30 PM
SCALE ID 123456
USER ID 234567
PROJ ID 345678
NAME _____________
12.34 kg 12.34 kg G
11.11 kg NET
1.22 kg T
LEVEL 10
MODE: DYNAMIC
Linearity Calibration Printout
OHAUS CORPORATION
MODEL T51P
Indicator
S/N 1234567890
--------LIN CAL-------01/31/07 12:30 PM
SCALE ID 123456
USER ID 234567
PROJ ID 345678
NAME ________________
_____
New Cal:50.00 kg
Old Cal: 49.99 kg
Diff: 0.01g
Wt. ID______________
___
----------END----------
Span Calibration Printout
--------SPAN CAL------New Cal:50.00 kg
Old Cal: 49.99 kg
Diff: 0.01g
Wt. Ref________________
01/31/07 12:30 PM
SCALE ID 123456
PROJ ID 345678
USER ID 234567
NAME ______________
____
----------END----------
EN-5
.
5000 Series Indicators
LEGAL FOR TRADE
.1 Settings
Enter the menu to verfy the settngs and perform a calbraton as explanedn Secton 3.
Set the LFT menu to ON.Ext the setup menu and power off thendcator.
Open the housng as explanedn Secton 2.3.1.
Set the poston of the securty swtch SW2 to LFT ON as shownn Fgure 1-3,tem 4.
Close the housng.
NOTE: When LFTs set to ON and the securty swtchs set to ON, the followng menu settngs cannot be changed:Zero
Calbraton, Span Calbraton, Lnearty Calbraton, GEO, LFT, Range, Capacty, Graduaton, Zero Range, Stable Range, AZT,
Modes, Unts.
. Verification
Before ths product can be usedn commercal trade,t must benspectedn accordance wth local weghts and measures
regulatons.It
It s the responsblty of the purchaser to ensure that all pertnent legal requrements are met.Please contact your
It
local weghts and measures offce for further detals.
. Sealing
The weghts and measures offcal can apply a wre or paper securty seal as shown below.
Fgure 6-1. T51P Wre SealFgure 6-2. T51P Paper Seal
SEAL
SEAL
Fgure 6-3. T51XW Wre SealFgure 6-4. T51XW Paper Seal
5000 Series Indicators
EN-53
7 MAINTENANCE
CAUTION: DISCONNECT THE UNIT FROM THE POWER SUPPLY BEFORE CLEANING.
7.1 Model T51P Cleaning •The housing may be cleaned with a cloth dampened with a mild detergent if necessary.
•Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean the housing or control panel.
7.2 Model T51XW Cleaning
• Use approved cleaning solutions for the stainless-steel Indicator housing and rinse with water. Dry thoroughly.
•Do not use solvents, chemicals, alcohol, ammonia or abrasives to clean the control panel.
7.3 Troubleshooting
TABLE 7-1. TROUBLESHOOTING.
SYMPTOM
PROBABLE CAUSE(s)
REMEDY
Power cord not plugged in or properly
connected.
Check power cord connections. Make sure
power cord is plugged in properly into the
power outlet.
Power outlet not supplying electricity.
Check power source.
Battery discharged (T51P).
Replace batteries (T51P).
Other failure.
Service required.
Load on Scale exceeds allowable limits.
Remove load on Scale.
Load on Scale is not stable.
Wait for load to become stable.
Load Cell damage.
Service required.
Lock Calibration Menu set to On.
Set Lock Calibration Menu to Off. Refer to Section 3.12 Menu Lock.
LFT menu set to On.
Set LFT menu to Off.
Incorrect value for calibration mass.
Use correct calibration mass.
Cannot display weight in desired weighing
unit.
Unit not set to On.
Enable unit in the Units Menu. Refer to Section 3.7 in the Unit Menu.
Cannot change menu settings.
Menu has been locked.
Set selected menu to Off in the Lock Menu. Lockout Switch on the circuit board may
need to be set to the Off position.
Error 8.1
Weight reading exceeds Power On Zero limit.
Remove load from scale. Recalibrate scale.
Error 8.2
Weight reading below Power On Zero limit.
Add load to scale. Recalibrate scale.
Error 8.3
Weight reading exceeds Overload limit.
Reduce load on scale.
Error 8.4
Weight reading below Underload limit.
Add load to scale. Recalibrate scale.
Error 8.6
Weight exceeds six digits. Display overflow.
Reduce load on scale.
Unit will not turn on.
Cannot zero the Scale, or will not zero when
turned on.
Unable to calibrate.
EN-54
5000 Series Indicators
TABLE 7-1. TROUBLESHOOTING (Cont.).
SYMPTOM
PROBABLE CAUSE(s)
REMEDY
Error 9.5
Calibration data not present.
Calibrate scale.
Battery symbol flashing
Batteries are discharged.
Replace batteries (T51P). CAL E
Calibration value outside allowable limits
Calibration Error
LOW REF WT
Average Piece Weight too small. (Warning)
Use correct calibration weight.
REF WT Err
Reference Weight too small. The weight on
the platform is too small to define a valid
reference weight.
Use a greater weight for sample.
7.4 Service Information
If the troubleshooting section does not resolve your problem, contact an authorized Ohaus Service Agent. For Service assistance
in the United States, call toll-free 1-800-526-0659 between 8:00 AM and 5:00 PM Eastern Standard Time. An Ohaus Product
Service Specialist will be available to assist you. Outside the USA, please visit our website www.ohaus.com to locate the Ohaus
office nearest you.
5000 Series Indicators
EN-55
8. TECHNICAL DATA
8.1 Specifications
Materials
T51XW Housing: stainless-steel
T51P Housing: ABS plastic
Display window: polycarbonate Keypad: polyester
Feet: Rubber
Ambient conditions
The technical data is valid under the following ambient conditions:
Ambient temperature: -10°C to 40°C / 14°F to104°F
Relative humidity: Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50%
relative humidity at 40°C.
Height above sea level: up to 2000m
Operability is assured at ambient temperatures between -10°C. and 40°C.
TABLE 8-1. SPECIFICATIONS
Indicator
T51P
T51XW
Maximum Displayed Resolution
1:30,000
Maximum Approved Resolution
1:10,000
Maximum Counting Resolution
1:300,000
Weighing Units
Functions
Display
kg, lb, g, oz, lb:oz, tonnes, custom
Static Weighing, Dynamic Weighing, Counting, Checkweighing, Percent Weighing
25 mm / 1 in High 6-digit, 7-segment LCD
Over/Accept/Under Indicators
Red, Green, Yellow LED
Backlight
White LED
Keypad
Ingress Protection
4-button membrane switch
---
IP66
Load Cell Excitation Voltage
5V DC
Load Cell Drive
up to 3 mV/V
Load Cell Input Sensitivity
Up to 8 x 350 ohm Load Cells
Stabilization Time
Within 2 Seconds
Auto-zero Tracking
Off, 0.5, 1 or 3 Divisions
Zeroing Range
2% or 100% of Capacity
Span Calibration
Overall Dimensions (W x D x H)
(mm/in)
Net Weight (kg/lb)
Shipping Weight (kg/lb)
Operating Temperature Range
Power
Interface
1 kg or 1 lb to 100% Capacity
260 x 71 X 166 / 10.2 x 2.7 x 6.5
262 x 76 x 149 / 10.3 x 3.0 x 5.8
1.5 / 3.3
3.5 / 7.7
2.3 / 5
4.3 / 9.5
-10°C to 40°C/14°F to 104°F
100-240 VAC / 50-60 Hz Internal Universal Power Supply,
6 C-type batteries (T51P)
Built-in RS232 and External Input
EN-56
5000 Series Indicators
8.2 Accessories and Options
TABLE 8-2. OPTIONS.
DESCRIPTION
PART NUMBER
AC Relay Kit
80500720
Base Mount Kit, T51P
80500722
Column Mount Kit, 35 cm painted steel
80500724
Column Mount Kit, 68 cm painted steel
80500723
Column Mount Kit, 35 cm stainless steel
80500725
Column Mount Kit, 68 cm stainless steel
80500726
DC Relay Kit
80500727
Rechargeable Battery Kit
80500729
RS422/485 Interface Kit
80500731
RS232 Interface kit
80500733
TABLE 8-3. ACCESSORIES.
DESCRIPTION
PART NUMBER
Foot Switch
71173378
Interface Cable/PC 25-pin
80500524
Interface Cable/PC 9-pin
80500525
Interface Cable/PC 9-pin
80500552
Interface Cable/PC 25-pin
80500553
Interface Cable/Printer SF42
80500571
Interface Cable/Printer SF42
80500574
SF42 Printer
SF42
The Rechargeable Battery Kit, RS232 Kit, RS422/485 Kit, AC Relay Kit, DC Relay kit and Foot switch must be
installed by a qualified technician. 5000 Series Indicators
EN-57
8.3 Drawings and Dimensions
Figure 8-1. T51P Indicator Overall Dimensions with Mounting Bracket.
Figure 8-2. T51XW Indicator Overall Dimensions with Mounting Bracket.
EN-58
8.4
5000 Series Indicators
Compliance
Compliance to the following standards is indicated by the corresponding marking on the product.
Marking
Standard
This product conforms to the EMC Directive 2004/108/EC, the Low Voltage Directive 2006/95/EC and the
Non-automatic Weighing Instruments Directive 90/384/EEC. The complete Declaration of Conformity is
available from Ohaus Corporation
UL60950-1 : 2003
AS/NZS4251.1, AS/NZS4252.1
FCC Note
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC
Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated
in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of
this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the
interference at his own expense.
Industry Canada Note
This Class A digital apparatus complies with Canadian ICES-003.
Cet appareil numérique de la classe A est conforme à la norme NMB-003 du Canada.
ISO 9001 Registration
In 1994, Ohaus Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International
(BVQI), confirming that the Ohaus quality management system is compliant with the ISO 9001 standard’s requirements. On May
15, 2003, Ohaus Corporation, USA, was re-registered to the ISO 9001:2000 standard.
5000 Series Indicators
EN-59
Important Notice for verified weighing instruments
Weighing Instruments verified at the place of manufacture bear one of the preceding marks on the packing label and the green ‘M’ (metrology) sticker on the descriptive plate. They may be put into service
immediately.
Weighing Instruments to be verified in two stages have no green ‘M’ (metrology) on the descriptive plate
and bear one of the preceding identification mark on the packing label. The second stage of the initial
verification must be carried out by the approved service organization of the authorized representative
within the EC or by the national weights & measures (W+M) authorities.
The first stage of the initial verification has been carried out at the manufacturer’s work. It comprises all tests according to the
adopted European standard EN 45501:1992, paragraph 8.2.2.
If national regulations limit the validity period of the verification, the user of the weighing instrument must strictly observe the
re-verification period and inform the respective W+M authorities.
Disposal
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic Equipment
(WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside
the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point specified for
electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the distributor from which you
purchased this device.
Should this device be passed on to other parties (for private or professional use), the content of this
regulation must also be related.
Thank you for your contribution to environmental protection.
LIMITED WARRANTY
Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of
the warranty period. During the warranty period Ohaus will repair, or, at its option, replace any component(s) that proves to be
defective at No charge, provided that the product is returned, freight prepaid, to Ohaus.
This warranty does Not apply if the product has been damaged by accident or misuse, exposed to radioactive or corrosive
materials, has foreign material penetrating to the inside of the product, or as a result of service or modification by other than
Ohaus. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the
authorized dealer. No other express or implied warranty is given by Ohaus Corporation. Ohaus Corporation shall Not be liable
for any consequential damages.
As warranty legislation differs from state to state and country to country, please contact Ohaus or your local Ohaus dealer for
further details.
Ohaus Corporation
19A Chapin Road
P.O. Box 2033 Pine Brook, NJ 07058, USA
Tel: (973) 377-9000
Fax: (973) 593-0359
www.ohaus.com
*80251400*
P/N 80251400 © 2007 Ohaus Corporation, all rights reserved.
Printed in China
Defender™ Series Base
Instruction Manual
Compliance
This product conforms to the EMC directive 2004/108/EC and the Low Voltage Directive 2006/95/EC.
The complete declaration of Conformity is available from Ohaus Corporation
Disposal
In
conformance with the European Directive 2002/96 EC on Waste Electrical and Electronic Equipment (WEEE)
this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their
specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point specified for electrical
and electronic equipment.
If you have any questions, please contact the responsible authority or the distributor from which you purchased
this device.
Should this device be passed on to other parties (for private or professional use), the content of this regulation
must also be related.
Thank you for your contribution to environmental protection.
ISO 9001 Registration
In 1994, Ohaus Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International
(BVQI), confirming that the Ohaus quality management system is compliant with the ISO 9001 standard’s requirements. On
May 15, 2003, Ohaus Corporation, USA, was re-registered to the ISO 9001:2000 standard.
EN-1
Defender Series Base
INTRODUCTION
This manual covers installation, and maintenance instructions for the Ohaus Defender TM Series Base. Please read this manual
completely before installation and operation.
SAFETY PRECAUTIONS
For safe and dependable operation of this product, please comply with the following precautions:
• Operate the base only under ambient conditions specified in these instructions
• Ensure that the load cell cable does not pose an obstruction or tripping hazard
• Do not operate the base in hazardous environments or unstable locations
• Do not drop loads on the base
• Do not lift the base by the top frame; always lift from the bottom frame when moving the base
• Service should only be performed by authorized personnel
INSTALLATION
Unpacking
Unpack and inspect the product to make sure that all components have been included. The package includes the following:
• Defender TM Series Base • Weighing Pan • Warranty Card • Instruction Manual
When purchased as a complete Defender Series Scale, the package will also include:
• Indicator • Column Assembly (optional)
Assembly
Weighing Pan
Place the weighing pan securely over the rubber load pads on the top frame of the base.
Wiring Connections
DxxxHx and DxxxQx Bases
When connecting the DxxxHx or DxxxQx base to an Ohaus indicator equipped with the matching circular connector, push the
base connector onto the indicator connector and turn the locking ring clockwise to lock it in place.
When connecting the DxxxHx or DxxxQx base to an indicator that does not have the matching connector, install the optional load
cell cable adapter (P/N 80500736 sold separately) to the indicator using the wiring codes in Table 1. Then attach the load cell
cable to the cable adapter.
Note: As an alternate wiring method, cut off the connector, strip the wires and attach them to the indicator using the wiring codes
in Table 1.
DxxxWx Bases
Install the load cell cable to an indicator using the wiring codes in Table 1.
Note: When purchased as a Defender Series Scale, the DxxxWx base is already pre-wired to the indicator.
TABLE 1. LOAD CELL CONNECTION
FUNCTION
WIRE COLOR
+ Excitation
Green
- Excitation
Black
+ Signal
Red
- Signal
White
+ Sense
Blue
- Sense
Brown
Shield
Yellow
EN-2
Defender Series Base
Selecting the Location
To ensure accuracy, proper performance and safety, locate and operate the base on a stable, level surface. Avoid locations with
rapid temperature changes or excessive dust, air currents, vibrations, electromagnetic fields or heat.
Level the base by adjusting the four leveling feet until the bubble in the level indicator (located at the rear of the base) is centered.
A wrench may be needed to loosen the locking nut above each leveling foot. When the base is level, retighten the locking nuts
up against the base to lock each foot into place.
Note: Ensure that the base is level each time its location is changed.
CORRECT
INCORRECT
MAINTENANCE
Cleaning
The base components should be kept clean and free of excessive material build up.
• Damp cloth with water and a mild detergent may be used to wipe clean the external surfaces—do not use acids,
alkalis, strong solvents or abrasive materials and chemicals
Troubleshooting
Aside from installing components and leveling adjustments, the Defender Series Base does not require any other adjustments as
shipped from the factory.
Operational difficulties that may be encountered can often be traced to simple causes such as:
• Loose or incorrect wiring connections
• Obstructions to the base frame
• Unstable environments
• Incorrect calibration or setup of the indicator
If the troubleshooting section does not resolve or describe your problem, contact your authorized Ohaus service agent. For
service assistance or technical support in the United States call toll-free 1-800-526-0659 between 8:00 AM and 5:00 PM EST.
An Ohaus product service specialist will be available to provide assistance. Outside the USA, please visit our web site,
www.ohaus.com to locate the Ohaus office nearest you.
EN-3
Defender Series Base
TECHNICAL DATA
Technical Specifications
The technical data is valid under the following ambient conditions:
Ambient temperature: -10ºC to +40 ºC
Relative humidity: 10% to 90% relative humidity, non-condensing
Height above sea level: Up to 4,000m
Operability is assured at ambient temperature -10ºC to +40 ºC
TABLE 2.1 SPECIFICATIONS
MODEL
D15HR
Capacity
15 kg / 30 lb
Approved Resolution
D60HR
30 kg / 60 lb
60 kg / 150 lb
OIML: 3000e, NTEP: 3000e
Safe Overload Capacity
150% of capacity
Pan Dimensions
305 x 355 mm / 12 x 14 in
Base Construction
Stainless Steel platform with painted steel frame and rubber leveling feet
Repeatability (std. deviation)
1d
Linearity
Load Cell Capacity
D30HR
±1d
30 kg
50 kg
100 kg
Load Cell Cable
2 m L x 6-wire
Load Cell Type
350 Ohm, aluminum, single point
Load Cell Excitation
5-15V DC/AC
Load Cell Rated Output
2mV/V
Load Cell Protection
IP67
Net Weight
9 kg / 19.8 lb
Shipping Weight
10.9 kg / 24 lb
TABLE 2.2. SPECIFICATIONS
MODEL
D60HL
D100HL
D150HX
Capacity
60 kg / 100 lb
100 kg / 250 lb
150 kg / 300 lb
Approved Resolution
OIML: 3000e,
NTEP: 3000e
OIML: 3000e,
NTEP: 2500e
Safe Overload Capacity
Pan Dimensions
400 x 500 mm / 15.7 x 19.7 in
420 x 550 mm / 16.5 x 21.6 in
Stainless Steel platform with painted steel frame and rubber leveling feet
Repeatability (std. deviation)
1d
Linearity
±1d
100 kg
150 kg
300 kg
Load Cell Cable
2 m L x 6-wire
Load Cell Type
350 Ohm, aluminum, single point
Load Cell Excitation
2mV/V
Load Cell Protection
Shipping Weight
500 kg
5-15V DC/AC
Load Cell Rated Output
Net Weight
300 kg / 600 lb
OIML: 3000e,
NTEP: 3000e
150% of capacity
Base Construction
Load Cell Capacity
D300HX
IP67
19.1 kg / 42.1 lb
24.5 kg
22 kg / 48.5 lb
27.1 kg / 59.7 lb
EN-4
Defender Series Base
TABLE 2.3 SPECIFICATIONS
MODEL
D10QR
D15QR
D25QR
D30QR
D50QL
Capacity
10 kg / 25 lb
15 kg / 30 lb
25 kg / 50 lb
30 kg / 60 lb
50 kg / 100 lb
Approved Resolution
NTEP: 5000e
OIML: 3000e
NTEP: 5000e
OIML: 3000e
NTEP: 5000e
Safe Overload Capacity
150% of capacity
Pan Dimensions
305 x 305 mm /
12 x 12 in
Base Construction
457 x 457 mm /
18 x 18 in
Stainless Steel platform with painted steel frame and rubber leveling feet
Repeatability (std. deviation)
1d
Linearity
Load Cell Capacity
±1d
30 kg
30 kg
50 kg
50 kg
Load Cell Cable
2.5 m L x 6-wire
Load Cell Type
350 Ohm, aluminum, single point
Load Cell Excitation
100 kg
5-15V DC/AC
Load Cell Rated Output
2mV/V
Load Cell Protection
IP67
Net Weight
12.2 kg / 26.9 lb
27.2 kg / 60 lb
Shipping Weight
14.2 kg / 31.3 lb
29 kg / 64 lb
TABLE 2.4 SPECIFICATIONS
MODEL
D60QL
D100QL
D150QL
D250QX
D300QX
Capacity
60 kg / 150 lb
100 kg / 250 lb
150 kg / 300 lb
250 kg / 500 lb
300 kg / 600 lb
OIML: 3000e
NTEP: 5000e
OIML: 3000e
NTEP: 5000e
OIML: 3000e
Approved Resolution
Safe Overload Capacity
150% of capacity
Pan Dimensions
457 x 457 mm /
18 x 18 in
Base Construction
Stainless Steel platform with painted steel frame and rubber leveling feet
Repeatability (std. deviation)
1d
Linearity
Load Cell Capacity
610 x 610 mm /
24 x 24 in
±1d
100 kg
150 kg
200 kg
Load Cell Cable
2.5 m L x 6-wire
Load Cell Type
350 Ohm, aluminum, single point
Load Cell Excitation
2mV/V
Load Cell Protection
Shipping Weight
500 kg
5-15V DC/AC
Load Cell Rated Output
Net Weight
500 kg
IP67
27.2 kg / 60 lb
35 kg / 77 lb
29 kg / 64 lb
41 kg / 90 lb
EN-5
Defender Series Base
TABLE 2.5 SPECIFICATIONS
MODEL
D10WR
D15WR
D25WR
D30WR
D50WL
Capacity
10 kg / 25 lb
15 kg / 30 lb
25 kg / 50 lb
30 kg / 60 lb
50 kg / 100 lb
Approved Resolution
NTEP: 5000e
OIML: 3000e
NTEP: 5000e
OIML: 3000e
NTEP: 5000e
Safe Overload Capacity
150% of capacity
Pan Dimensions
305 x 305 mm /
12 x 12 in
Base Construction
457 x 457 mm /
18 x 18 in
Stainless Steel platform with stainless steel frame and rubber leveling feet
Repeatability (std. deviation)
1d
Linearity
Load Cell Capacity
±1d
30 kg
30 kg
50 kg
50 kg
Load Cell Cable
2.5 m L x 6-wire
Load Cell Type
350 Ohm, stainless steel, single point
Load Cell Excitation
100 kg
5-15V DC/AC
Load Cell Rated Output
2mV/V
Load Cell Protection
IP67
Net Weight
12.2 kg / 26.9 lb
27.2 kg / 60 lb
Shipping Weight
14.2 kg / 31.3 lb
29 kg / 64 lb
TABLE 2.6 SPECIFICATIONS
MODEL
D60WL
D100WL
D150WL
D250WX
D300WX
Capacity
60 kg / 150 lb
100 kg / 250 lb
150 kg / 300 lb
250 kg / 500 lb
300 kg / 600 lb
OIML: 3000e
NTEP: 5000e
OIML: 3000e
NTEP: 5000e
OIML: 3000e
Approved Resolution
Safe Overload Capacity
150% of capacity
Pan Dimensions
457 x 457 mm /
18 x 18 in
Base Construction
Stainless Steel platform with stainless steel frame and rubber leveling feet
Repeatability (std. deviation)
1d
Linearity
Load Cell Capacity
610 x 610 mm /
24 x 24 in
±1d
100 kg
200 kg
200 kg
Load Cell Cable
2.5 m L x 6-wire
Load Cell Type
350 Ohm, stainless steel, single point
Load Cell Excitation
2mV/V
Load Cell Protection
Shipping Weight
500 kg
5-15V DC/AC
Load Cell Rated Output
Net Weight
500 kg
IP67
27.2 kg / 60 lb
35 kg / 77 lb
29 kg / 64 lb
41 kg / 90 lb
EN-6
Defender Series Base
Drawings
Figure 1. Defender Base Dimension Drawing.
A
TABLE 3. DIMENSIONS.
B
C
Height of Pan to
Pan Width
Surface of Table*
305 mm
95 mm
50 mm
E
Depth Including
Level Indicator
385 mm
105 mm
60 mm
530 mm
420 mm
125 mm
70 mm
580 mm
305 mm
305 mm
95 mm
52 mm
335 mm
457 mm
457 mm
115 mm
70 mm
487 mm
DxxxQX, DxxxWX
610 mm
610 mm
137 mm
*For leveling purposes, the feet may be extended up to an additional 15 mm.
90 mm
640 mm
Base
Pan Depth
DxxxHR
355 mm
DxxxHL
500 mm
400 mm
DxxxHX
550 mm
DxxQR, DxxxWR
DxxxQL, DxxxWL
Accessories
TABLE 4. ACCESSORIES.
Description
Part Number
Base Mount Kit, painted steel
80500722
Column Kit, 35 cm painted steel
80500723
Column Kit, 68 cm painted steel
80500724
Column Kit, 35 cm stainless steel
80500725
Column Kit, 68 cm stainless steel
80500726
Load Cell Cable Adapter Kit
80500736
D
Height of Pan
EN-7
Defender Series Base
LIMITED WARRANTY
Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the
warranty period. During the warranty period Ohaus will repair, or, at its option, replace any component(s) that proves to be
defective at no charge, provided that the product is returned, freight prepaid, to Ohaus. This warranty does not apply if the
product has been damaged by accident or misuse, exposed to radioactive or corrosive materials, has foreign material
penetrating to the inside of the product, or as a result of service or modification by other than Ohaus. In lieu of a properly returned
warranty registration card, the warranty period shall begin on the date of shipment to the authorized dealer. No other express or
implied warranty is given by Ohaus Corporation. Ohaus Corporation shall not be liable for any consequential damages.
As warranty legislation differs from state to state and country to country, please contact Ohaus or your local Ohaus dealer for
further details.
Ohaus Corporation
19A Chapin Road
P.O. Box 2033
Pine Brook, NJ 07058-2033, USA
Tel: (973) 377-9000
Fax: (973) 944-7177
With offices worldwide.
www.ohaus.com
*80251421*
P/N 80251421 © Ohaus Corporation 2007, all rights reserved.
Printed in China

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement