O2XPRESS ™
PNEUMATIC DEMAND OXYGEN CONSERVER
Model 8511/8512
Operating Instruction Manual
TABLE OF CONTENTS
Introduction .................................................................................. ......................................1
Important Safeguards ................................................................... ......................................1
System Components .................................................................... ......................................3
Setup and Operation .................................................................... ......................................4
Connecting to the Cylinder ..................................................... ......................................4
Starting Operation .................................................................. ......................................5
Ending Operation ................................................................... ......................................6
Disconnecting from the Cylinder ............................................ ......................................6
Care and Maintenance ................................................................. ......................................6
Cleaning ................................................................................. ......................................6
Storage .................................................................................. ......................................7
Maintenance ........................................................................... ......................................7
Specifications ............................................................................... ......................................7
Accessories .................................................................................. ......................................7
Warranty ....................................................................................... .........................................8
DEVICE INFORMATION
Record the following information for easy reference regarding your O2XPRESS™
Conserver.
Physician Name
Physician Phone
Prescribed Flow Rates
Rest
Exercise
Dealer Name
Dealer Phone
Purchase Date
i
INTRODUCTION
Salter Labs O2XPRESS™ Pneumatic Demand Oxygen Conserver is lightweight, easy to use
demand oxygen conserver designed for the ambulatory patient. When used with a portable
oxygen cylinder, the O2XPRESS™ Conserver delivers a discrete burst of oxygen during inhalation and shuts off oxygen delivery during exhalation when you do not need it.
The O2XPRESS™ Conserver incorporates an integral high pressure regulator and connects
directly to a medical oxygen cylinder. The conserver is pneumatically operated, so it does not
require batteries or any other external power source.
The O2XPRESS™ Conserver will typically provide up to 3:1 oxygen saving over a continuous
flow device at equivalent flow rates. This estimate is based on an inspiration: expiration (I:E)
ratio 1:2 at 20 breaths per minute.
To maximize your benefit from the O2XPRESS™ Conserver, it is important for you to read and
understand the information in this instruction manual. By following your doctor’s advice and
these simple instructions, your O2XPRESS™ Conserver will become a valuable part of your
daily routine.
CAUTION
Use only as prescribed by your physician. Read all warnings and instructions
before use. Improper use can result in serious or fatal illness or injury.
CAUTION
Federal law (USA) restricts this device to sale by or on the order of a physician.
IMPORTANT SAFEGUARDS
READ ALL INSTRUCTIONS BEFORE USING
DANGER - To reduce the risk of fire, burns, or injury:
´
DO NOT smoke when oxygen is being used. DO NOT allow flames or sparks anywhere
near oxygen equipment or the cannula during use or storage
´
DO NOT cover or carry the device under clothing. The conserver vents a small amount of
oxygen during operation.
´
DO NOT use oil, grease, or other petroleum-based materials on the device. The conserver
does not require lubrication.
´
Rapid pressurization can increase the risk of fire. SLOWLY open the cylinder valve to start
the flow of oxygen with the flow rate selector on the conserver set to the OFF position
´
CLOSE the cylinder valve and turn the conserver off when not in use. ALWAYS close the
cylinder valve first before disconnecting the conserver from the cylinder.
´
If a leak is detected, CLOSE the cylinder valve and reposition the conserver on the cylinder
post.
´
Replace a dirty or worn inlet seal washer with a new Viton rubber/bass washer supplied by
Salter Labs. Do not use any other washer as it could increase the risk of fire. Ensure that
the inlet and surrounding area are clean and free of debris.
1
WARNING - To reduce the risk of injury:
´
Use this product only for its intended use as described in this manual.
´
Use only oxygen from an approved oxygen source. This equipment delivers supplemental oxygen and is not to be considered life-supporting.
•
Use only cannulas and accessories specified by Salter Labs. Do not use a mask with the
conserver
´
Momentarily open the cylinder valve to clear the cylinder outlet of debris before connecting the conserver.
´
Inspect the inlet connection and washer on the conserver to ensure it is clean. Replace
the washer if necessary. If the oxygen inlet is clogged or contaminated, DO NOT use
the conserver. Contact your Salter Labs dealer or authorized service center
´
DO NOT connect the conserver to an oxygen source that has a pressure greater than
3000 psig.
´
Allow the conserver to return to room temperature before operation if it has been stored
in extremely hot or cold temperatures.
´
DO NOT use the conserver with a humidifier.
´
The conserver should not be used while sleeping unless approved by your physician
´
DO NOT allow any liquid to enter the device. Clean the conserver only as instructed.
´
DO NOT operate the conserver if it is not working properly, if it has been dropped or
damaged, or if it has been immersed in any liquid. Contact your Salter Labs dealer
or authorized service center if your device needs repair.
´
DO NOT turn or attempt to remove the pressure gauge. The gauge is locked into place
and not threaded.
´
DO NOT tamper with or attempt to repair the device. There are no user serviceable
parts inside the unit. Except for instructed user maintenance, all device servicing must
be performed by a Salter Labs authorized service center.
´
Use care in handling and storing oxygen cylinders. Do not allow cylinders to fall. Do
not store cylinders near any source of heat or flame.
Professional Instructions for Use
The O2XPRESS pneumatic demand oxygen conserver has been tested and demonstrated to
maintain oxygen saturations equivalent to steady flow at all settings. However, the actual
clinical results obtained in individual patients may vary with the settings utilized and their
individual breathing patterns. Fro this reason, it is highly recommended that, as with oxygen
delivery device, patients be individually tested to titrate dosing and to determine the suitability of this device to meet their unique and varying individual needs and dosage requirements.
2
SYSTEM COMPONENTS
Flow rate setting
Pressure gauge
T-handle
Oxygen inlet with seal washer
Figure 1
Flow rate selector
Barb connector
Figure 2
3
SETUP AND OPERATION
CONNECTING TO THE CYLINDER
1. Remove the seal from the cylinder if present.
2. Point the outlet of the oxygen cylinder away from you.
Momentarily open and quickly close the cylinder valve to clear the outlet of any debris with
a short blast of oxygen.
3. Check the conserver to ensure the inlet seal washer at the oxygen inlet is in place and in
good condition. Replace a dirty or worn washer with a new Viton rubber/brass washer.
Check that the inlet is clean and free of debris (Figure 1).
4. Loosen the T-handle counterclockwise and slide the conserver over the cylinder post. Align
the two pins on the conserver with the matching holes in the cylinder post.
5. Secure the conserver to the cylinder post by turning the T-handle clockwise until hand tight
(Figure 3).
6. Place the cylinder and conserver in its carrying case or cart.
T-handle
Cylinder post
Figure 3
4
STARTING OPERATION
WARNING - Use only cannulas specified by Salter Labs. Do not use a mask with the
conserver.
1. Firmly press the two ends of the cannula onto the two barb connectors on the bottom of the
conserver. The orientation of the cannula makes no difference. The cannula lines may be
attached to either barb connector (Figure 2 and Figure 4).
2. Turn the flow rate selector to 0 (OFF).
3. Open the oxygen supply by SLOWLY turning the cylinder valve until you hear oxygen flow
into the conserver. The pressure gauge on the conserver now reads the oxygen pressure
remaining in the cylinder.
4. Listen for leaks around the cylinder connection. If a leak is detected, close the cylinder
valve and reposition the conserver on the cylinder post. Then slowly reopen the cylinder
valve and recheck for leaks.
5. Turn the flow rate selector to the prescribed flow rate for your desired level of activity.
Selector settings represent oxygen delivered in liters per minute (lpm).
6. Unwind the cannula and ensure there are no kinks or twists in the tubing. Place the cannula over your ears and position the prongs in your nose.
Flow rate setting
Flow rate selector
Figure 4
5
Figure 5
7. Breathe in normally through your nose. Notice that the conserver provides a pulse of
oxygen during each inhalation cycle and stops oxygen flow during exhalation. The oxygen
pulse is delivered in two phases: a short pulse is first given at the start of inhalation, followed by a continuous flow for the remainder of inhalation. You may also notice that oxygen is delivered through only one cannula line. To change which side oxygen is delivered,
switch the cannula tubing connections on the conserver.
8. Periodically check the pressure gauge on the conserver. Install a new cylinder when the
gauge reading drops into the red “Refill” zone to continue operation.
ENDING OPERATION
1. Turn the cylinder valve clockwise until it is completely closed to shut off the oxygen supply.
2. Purge the oxygen in the conserver by inhaling from the cannula several times until oxygen
is no longer delivered in response to inhalation.
3. Turn the flow rate selector to 0 (OFF).
4. Remove the cannula from your head. Coil the cannula tubing and store with equipment.
DISCONNECTING FROM THE CYLINDER
1. Ensure that the cylinder valve is closed and the conserver is turned off as instructed in
ENDING OPERATION.
2. Disconnect the cannula from the conserver.
3. Turn the T-handle counterclockwise to disconnect the conserver from the cylinder.
4. Store the equipment as instructed in CARE AND MAINTENANCE.
5. Replace the cylinder if the pressure gauge was reading in the red “Refill” zone. Refill the
empty cylinder.
CARE AND MAINTENANCE
CLEANING
If necessary, wipe the outside of the conserver with a clean cloth lightly dampened with
warm water. Do not let water enter any holes or joints on the conserver. If extensive cleaning is required, contact your Salter Labs dealer or authorized service center.
6
WARNING - DO NOT immerse the conserver in any liquid. DO NOT use liquid cleaners
or spray fluid directly on the conserver. Fluids can cause a fire hazard or may damage
internal components that could lead to a malfunction.
STORAGE
The conserver may remain connected to the oxygen cylinder for short term storage. Ensure
that the cylinder valve is closed and the conserver is turned off as instructed in ENDING
OPERATION. Store the equipment securely in a clean, cool enviroment.
For long term storage, disconnect the conserver from the cylinder as instructed in DISCONNECTING FROM THE CYLINDER. Place the conserver in its original package and store in
a clean, dry environment away from oils, grease, dirt, or other contaminants. Store oxygen
cylinders securely in a clean, cool environment
MAINTENANCE
Check to ensure the inlet seal washer at the oxygen inlet is in place and in good condition.
Replace a dirty or worn washer with a new Viton rubber/brass washer supplied by Salter Labs.
Unless instructed otherwise, all other maintenance must be performed by a Salter Labs authorized service center. There are no user serviceable parts inside the unit. Contact your Salter
Labs dealer or authorized service center if your unit needs repair.
WARNING - DO NOT tamper with or attempt to repair the device.
SPECIFICATIONS
Specifications and appearance are subject to change without notice
Operating pressure
Oxygen cylinder connection
Flow rate selections
Operating temperature
Storage temperature
Size
Weight
200-3000 psig (1,400-20,700 kPa)
CGA 870
0, .5, .75, 1, 1.5, 2, 2.5, 3, 4, 5, 6 lpm
0 - 45°C (32 - 113°F)
-40 - 70°C (-40 - 158°F)
3.5 in L x 2.1 in W x 5.0 in H
(8.9 cm L x 5.3 cm W x 12.7 cm)
1.0 lb (0.45 kg)
ACCESSORIES
Model
4804
4807
1248
1250
1251
1252
Description
Adult Nasal Divided Cannula, 4 ft. Tubing
Adult Nasal Divided Cannula, 7ft. Tubing
Inlet Seal Washers (Viton rubber/brass) (5 per pkg)
E Oxygen Cylinder with CGA 870 Connector
M6 Oxygen Cylinder with CGA 870 Connector
M9 Oxygen Cylinder with CGA 870 Connector
7
WARRANTY
Salter Labs warrants the O2XPRESS ™ Conserver to be free from defects in workmanship and materials for a period of one year from the date of purchase. Check with your
dealer for more details. This warranty is limited to the original purchaser of the device.
Any defective part or assembly will be repaired or replaced, at the sole discretion and
determination of Salter Labs, if the unit has not been misused or tampered with during
the period. Normal maintenance items, as outlined in this manual, and disposable
components are not covered by this warranty. Shipping charges, if any, shall be paid
by the purchaser.
THERE IS NO OTHER EXPRESS WARRANTY, IMPLIED WARRANTIES, INCLUDING THOSE OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE LIMITED TO THE DURATION OF THE EXPRESS LIMITED WARRANTY
AND TO THE EXTENT PERMITTED BY LAW AND ANY OTHER IMPLIED WARRANTIES ARE EXCLUDED. THIS WARRANTY DOES NOT COVER PROVIDING A
LOANER, COMPENSATING FOR COST INCURRED FOR RENTAL, OR FOR LABOR COST INCURRED IN REPAIRING OR REPLACING A DEFECTIVE PART(S).
Thank you for selecting the O2XPRESS ™ Conserver. Salter Labs is an innovative,
industry-leading manufacturer of respiratory care devices. Contact your local Salter
Labs dealer for information about additional products.
®
100 W. Sycamore Road
Arvin, California 93203 USA
Tel:
661-854-3166
800-421-0024
800-235-4203
Fax: 661-854-3850
800-628-4690
Internet: www.salterlabs.com
Salter Labs, O2XPRESS™, and the Salter Labs logo are trademarks of Salter Labs.
801521 Rev C (Printed in the USA)