Varian 400-DS Automated Apparatus 7 dissolution apparatus Operator's Manual
Below you will find brief information for dissolution apparatus 400-DS Automated Apparatus 7. The Varian 400-DS Automated Apparatus 7 is ideal for extended-release products or any dosage form requiring release profiling at multiple pH levels. A report can be generated to provide hard-copy documentation of the testing progress and conditions. All parameters are controlled via the PC. Up to 4 units can be pre-programmed to withdraw samples at designated timepoints. Samples can be collected in pre-capped HPLC vials for direct transfer to an HPLC system.
Advertisement
Advertisement
Varian 400-DS Automated
Apparatus 7 Operator’s
Manual
P/N 70-9051
May 2007
Revision A
Limitation of Liability
The information in this document is subject to change without notice. Varian, Inc. makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties or merchantability and fitness for a particular purpose. Varian, Inc. shall not be liable for errors contained herein or for incidental consequential damages in connection with the furnishing, performance or use of this material.
All rights are reserved. Reproduction, adaptation or translation without prior written permission is prohibited, except as allowed under copyright laws.
First Edition (May 2007)
Varian 400-DS Automated Apparatus 7 Operator’s Manual
Part Number 70-9051 Revision A
Printed in the United States of America
The following terms are trademarks of Varian, Inc.:
• Benchsaver
TM
• BIO-DIS III
®
• Enhancer Cell
®
• Full Flow Filters
TM
• Peak Vessel
TM
• Practical Solutions
®
• QA and QAII
TM
• TruCenter
TM
• VanKel
®
• VK
®
All other trademarks are property of their respective owners.
Varian, Inc.
13000 Weston Parkway
Cary, North Carolina 27513-2250
800.229.1108
919.677.1108
Fax: 919.677.1138
www.varianinc.com
Copyright (c) 2007 by Varian, Inc.
Table of Contents
Chapter 1
Chapter 2
Safety Practices and Hazards 11
Electrical Hazards
12
Reciprocation Hazard
13
Other
13
Warning
13
General
15
WEEE Directive
17
Introduction 19
Conventions Used in this Manual
20
USP Physical Parameters
20
Specifications
21
Environmental Specifications
21
Engineering Specifications
21
Accessories
23
Varian, Inc.
Chapter 3
Chapter 4
Table of Contents
Installation and Setup 25
Unpacking Your Equipment
25
Leveling Your Apparatus
26
Electrical Connection
27
Setting up Your PC
28
Setting up Your PC
28
Cable Connections
29
Single System Setup
29
Multiple System Setup
29
Setting Up the Varian 400-DS
30
Initial Power Up
32
Software Installation
32
Attaching the Database
33
Local Security Policy
35
E-mail Notification Configuration (Optional)
37
Starting 400-DS Workstation
40
Dissolution Server Selection
41
Configuring the Application - Adding Users
43
Operation 47
Log on to the Varian 400-DS Workstation
47
Configuring Your System
50
Copying a System Configuration
54
Deleting a System Configuration
54
Serial Numbers
55
System Configuration Report
56
Editing an Existing System Configuration
56
Import/Export XML File
58
System Calibration
60
Cell Calibration Temperatures
62
Manual Control/Diagnostics
64
Varian, Inc.
Chapter 5
Chapter 6
Table of Contents
Verifying Sample Tray Control
65
Media Control
66
Verifying the Cell Temperature
67
Reciprocation
68
Controlling the Evaporation Cover
69
Method Editor
69
Creating a Method
70
Copying Methods
77
Deleting Methods
77
Editing an Existing Method
78
Method Report
78
Audit Trail
79
Import/Export XML File
80
Running the Method
82
Test Reports
87
Electronic Signatures
88
Clean System
88
Maintenance 89
Routine Maintenance (to be Performed Between Each Dissolution
Method
90
Sample Holders
90
Non-Routine Maintenance (to be Performed at Pre-Determined
Intervals)
91
Service and Warranty 93
Exclusions and Limitations
94
Obtaining Warranty Service
94
Warranty Limitations
94
Exclusive Remedies
95
Index 97
Tell Us How We Are Doing
101
Varian, Inc.
Table of Contents
This page was intentionally left blank, except for this message.
Varian, Inc.
List of Figures
FIGURE 1. Accessory Tray 1 24
FIGURE 2. Accessory Tray 2 24
FIGURE 3. Diagram of the electrical connection 27
FIGURE 4. Diagram of cable connections 29
FIGURE 5. License dongle 33
FIGURE 6. MSDE Manager screen 34
FIGURE 7. DissoService Manager screen 37
FIGURE 8. Tools menu 38
FIGURE 9. DissoService Options screen 39
FIGURE 10. Logon screen 41
Varian, Inc.
List of Figures
FIGURE 11. Add New Dissolution Server screen 42
FIGURE 12. Browse Dissolution Servers screen 42
FIGURE 13. Configuration Dialog screen 43
FIGURE 14. Local Users and Groups screen 44
FIGURE 15. Miscellaneous tab 45
FIGURE 16. Logon screen 47
FIGURE 17. 400-DS Workstation screen 48
FIGURE 18. System Configuration screen 50
FIGURE 19. System Editor screen 51
FIGURE 20. 400-DS Workstation screen 52
FIGURE 21. System Calibration screen 60
FIGURE 22. System Diagnostics screen 62
FIGURE 23. Sample Tray Control box 65
FIGURE 24. Reciprocation box 68
FIGURE 25. Methods screen 70
FIGURE 26. Method Editor screen 71
FIGURE 27. Automated App7 Sequence tab 73
FIGURE 28. Automated App7 Options tab 74
FIGURE 29. Notifications tab 76
Varian, Inc.
List of Figures
FIGURE 30. Select System Step 1 of 2 screen 82
FIGURE 31. Select Method Step 2 of 2 screen 83
FIGURE 32. System status screen 84
FIGURE 33. Method Start Options screen 85
FIGURE 34. Test Report Selection screen 87
Varian, Inc.
List of Figures
This page was intentionally left blank, except for this message.
Varian, Inc.
Chapter 1
Safety Practices and Hazards
The Varian 400-DS Automated Apparatus 7 has been carefully designed so that when used properly you have an accurate, fast, flexible, and safe instrument.
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
Operation of a Varian 400-DS Automated Apparatus 7 involves the use of solid dosage forms and aqueous liquids. Unskilled, improper, or careless use of this instrument can create shock hazards, fire hazards, or other hazards which can cause death, serious injury to personnel, or severe damage to equipment and property.
Information on safety practices is provided with your instrument and operation manuals.
Before using your instrument or accessories, you must thoroughly read these safety practices.
Observe all relevant safety practices at all times.
Varian, Inc.
Page 12
Safety Practices and Hazards
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Electrical Hazards
Warning
The tester contains electrical circuits, devices, and components operating at dangerous voltages. Contact with these circuits, devices, and components can cause death, serious injury, or painful electric shock.
Warning
The apparatus should be disconnected from AC power before conducting cleaning or maintenance.
Panels or covers that are retained by fasteners which require the use of a tool for removal may be opened only by Varian-trained, Varian-qualified, or Varian-authorized service engineers. Consult the manuals or product labels supplied with the instrument to determine which parts are operator accessible.
Application of the wrong supply voltage, connection of the instrument to an incorrectly wired supply outlet, or lack of proper electrical grounding can create a fire hazard or a potentially serious shock hazard and could seriously damage the instrument and any attached ancillary equipment.
Always use a three-wire outlet with ground connection which is adequately rated for the load. The installation must comply with local, state, and federal safety regulations.
Do not connect the instrument to the main power supply until you have made sure that the operating voltage is correctly set for the main power supply in the specific outlet in your laboratory to which the equipment will be connected.
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 13
Safety Practices and Hazards
Reciprocation Hazard
The Varian 400-DS Automated Apparatus 7 contains a magnetic plate that reciprocates up and down during operation. Trapping an appendage between the evaporation cover and this plate could cause injury.
Other
Other specific warnings and cautions appear in the manuals where appropriate. These notifications detail the specific hazard, describe how to avoid it, and specify the possible consequences of not heeding the warning or caution.
Warning
A ‘Warning’ message appears in the manual when failure to observe instructions or precautions could result in death or injury. Symbols depicting the nature of the specific hazard are also placed alongside warnings.
These symbols may be used on warning labels attached to the instrument. When you see one of these symbols you must refer to the relevant operation or service manual for the correct procedure referred to by that warning label.
Varian, Inc.
Page 14
Safety Practices and Hazards
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
The meaning of the symbols that appear alongside warnings in this manual are as follows:
Electrical shock
Pinch point
Caution
Refer to accompanying documents
Read all warnings and cautions carefully and observe them at all times.
Caution
A ‘Caution’ message appears in the manual when failure to observe instructions could result in damage to equipment (Varian supplied and/or other associated equipment).
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 15
Safety Practices and Hazards
A ‘Note’ appears in the manual to give advice or information.
Information Symbols
Switches main power on
I
0
Switches main power off
Indicates single-phase alternating current
Indicates the product complies with the requirements of one or more European Union (EU) directives
Indicates specific equipment meets consensus-based standards of safety to provide assurance, required by
OSHA, that these products are safe for use in the workplace for North America
Indicates that this product must not be disposed of as unsorted municipal waste (see “WEEE Directive” on page 17)
General
CE Compliant Products
The Varian 400-DS Automated Apparatus 7 has been designed to comply with the requirements of the Electro-magnetic Compatibility (EMC) Directive and the Low Voltage
Directive (LVD) of the EU.
Varian, Inc.
Page 16
Safety Practices and Hazards
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Varian, Inc. has confirmed that each product complies with the relevant directives by testing a prototype against the prescribed European Norm (EN) standards.
Proof that a product complies with the directives is indicated by:
•
the CE marking appearing on the rear of the product.
•
the documentation package that accompanies the product containing a copy of the declaration of conformity. This declaration is the legal declaration by Varian,
Inc. that the product complies with the directives, and also shows the EN standards to which the product was tested to demonstrate compliance. The declaration of conformity is signed by the representative of the manufacturing plant.
cTUVus - U.S. and Canadian Product Approvals
The Varian 400-DS Automated Apparatus 7 has been designed to comply with North
American safety requirements.
This product has been tested and certified for the North American market by TUV
Rheinland of North America, Inc.. The TUVus mark signifies that this product has been tested to U.S. standards and certified for the U.S. market. The cTUV mark signifies that this product has been tested to Canadian standards and certified for the Canadian market. When the two marks are coupled, the cTUVus mark signifies that this product has been tested to standards and certified for both markets.
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 17
Safety Practices and Hazards
WEEE Directive
All Varian products that are subject to the WEEE directive shipped after August 13, 2005 are compliant with the WEEE marking requirements. Such products are marked with the
“crossed out wheelie bin” WEEE symbol shown on page 15 in accordance with European
Standard EN 50419.
This symbol on the product or on its packaging indicates that this product must not be disposed of as unsorted municipal waste. The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protects human health and the environment.
For more information on collection, reuse, and recycling systems, please contact your local/regional waste administration, your local distributor, or Varian, Inc.
Varian, Inc.
Page 18
Safety Practices and Hazards
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
This page was intentionally left blank, except for this message.
Varian, Inc.
Chapter 2
Introduction
The patented Varian 400-DS Automated Apparatus 7 is ideal for extended-release products or any dosage form requiring release profiling at multiple pH levels. The 13 test samples traverse the two rows of corresponding cells filled with media. The Varian 400-
DS Automated Apparatus 7 fills and dispenses media automatically, without operator intervention. A report can be generated to provide hard-copy documentation of the testing progress and conditions.
All parameters are controlled via the PC. Up to 4 units can be pre-programmed to withdraw samples at designated timepoints. Samples can be collected in pre-capped
HPLC vials for direct transfer to an HPLC system. Septa are pierced using the exclusive needle manifold, which lowers and raises at each sample point. Sample trays are available in 1.5 mL and 4 mL sizes.
Caution
Panels or covers that are retained by fasteners which require the use of a tool for removal may be opened only by Varian-trained,
Varian-qualified, or Varian-authorized service engineers.
Varian, Inc.
Page 20
Introduction
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Conventions Used in this Manual
•
Items you are asked to press are in bold. For example, “press H on the keypad”.
•
Key sequences you are asked to press appear like this: MENU > 0 > 8.
Note
Remember to return the warranty card supplied with this manual.
Completing and returning the card ensures your right to protection under the terms and conditions of your warranty. It also enables us to better assist you in the event of any problems.
Additionally, it guarantees you will be informed of any issues that arise concerning your equipment, such as upgrades, retrofits, or regulatory changes.
USP Physical Parameters
In addition to the apparatus suitability test, you must monitor several physical parameters, such as stroke distance, dip speed, and temperature. A Certificate of Compliance is included with your Varian 400-DS. Contact the Laboratory Services Department for more information on USP physical parameters.
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 21
Introduction
Specifications
Your tester is designed for indoor use. It is suitable for the categories stated below:
Environmental Specifications
Temperature
Humidity (%RH non-condensing)
Physical site
Support weight
15 - 40
°
C not more than 80 clean dry level bench
150 lb. (68 kg) min.
Engineering Specifications
Variable
Voltage Requirements
Current Requirements
Circuit Breaker
Specification
115 VAC/60 Hz
230 VAC/50 Hz
115 VAC - 8.0 AMP
230 VAC - 8.0 AMP
115 VAC - 8.0 AMP, 250 VAC, 50-60 Hz
230 VAC - 8.0 AMP, 250 VAC, 50-60 Hz
600 W Power Consumption (Max
Loaded)
Reciprocation
Reciprocation Accuracy
Reciprocation Resolution
Speed Selection
Temperature Range
5 - 35 DPM
± 0.5%
± 0.1 DPM
Via PC software
Ambient to 50
°
C
Varian, Inc.
Page 22
Introduction
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Variable
Temperature Measurement
Accuracy
Communications
Sampling
Sample Holder Material
Status Indicator
Dimensions
Weight
± 0.2
°
C
Specification
RS232, USB, and Ethernet
Automated sampling driven through PC software
PEEK and titanium
LEDs
Height: 24.00 inches (60.9 cm)
Width: 20.00 inches (50.8 cm)
Depth: 22.00 inches (55.8 cm)
130.3 lbs (59.1 kg)
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 23
Introduction
Accessories
Following is a list of accessories that ship with the Varian 400-DS. All items are pictured in the protective foam packaging on the following page.
E
M
N
O
K
E
H
J
Label Item
Varian 400-DS
A
B
AC power cord
Sample cells
F
G
C
D
Cell heater jackets
Sample holders
25’ roll of tubing
Sample tray with 12 rows
P
Q
R
Additional sample tray row for calibration
Magnetic plate
Evaporation cover
1.5 mm Allen wrench
9/64” Allen wrench
Modified temperature probe
Sample vials, 12x32, Qty. 100
Sample vials, 15x35, Qty. 100
Level
Sample holder extraction tools
RS232 cable
User manual CD
2
1
1
1
1
1
1
1
1
1
1
1
1
1
13
13
Quantity
1
1
13
Varian, Inc.
Page 24
Introduction
Varian 400-DS Automated Apparatus 7
Operator’s Manual
FIGURE 1. Accessory Tray 1
Revision A, 05/07
P/N 70-9051
FIGURE 2. Accessory Tray 2
Varian, Inc.
Chapter 3
Installation and Setup
Unpacking Your Equipment
The Varian 400-DS is shipped in one crate. The crate contains the apparatus and accessories.
Step 1. Open the crate and check the contents for damage, which may have occurred during shipping. Shipping damage rarely occurs, but if it does contact both the carrier who delivered the instruments and the
Dissolution Systems Service Department. Though claims for damage should be filed with the carrier, we can help you file a claim.
Step 2. Remove the top of the crate.
Step 3. Remove one side of the crate and lift the accessories box out of the crate.
Step 4. Disassemble the other three sides of the crate.
Varian, Inc.
Page 26
Installation and Setup
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Step 5. Remove any cushioning material and tape. If any item is missing, contact the Dissolution Systems Service Department or your local representative for replacements.
Step 6. Carefully remove the apparatus from the base of the shipping crate.
Warning
Because of its heavy weight, two people should lift the apparatus.
Lift by holding underneath the base of the unit on each side.
Support feet are located at each corner of the unit. Hands should be positioned between the support feet to ensure they are not trapped when lowering.
Step 7. Place the apparatus on a clean, dry, and level section of the bench.
Warning
The electrical connection at the back of the apparatus is the primary disconnect for the instrument. The apparatus should be positioned to allow accessibility to the power cords and circuit breaker switch for easy disconnection.
Leveling Your Apparatus
Step 1. Place the level provided in the accessory kit on the vessel table.
Step 2. Check the level of the instrument in the front and rear center of the unit as well as the left and right sides.
Step 3. The apparatus has adjustable feet under each corner. These may be screwed in or out to raise or lower each leg. Adjust the feet so the bubble inside the level is within the circle at all four points.
Varian, Inc.
Revision A 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 27
Installation and Setup
Electrical Connection
Warning
Ensure the apparatus is configured for the voltage available in your laboratory. Check the serial number tag on the rear panel of the instrument to confirm the voltage requirement. The electrical connection at the back of the apparatus is the primary disconnect for the instrument.
Step 1. Verify the Varian 400-DS is off.
Step 2. Connect the power cord between the receptacle on the drive unit rear panel and an outlet of the appropriate voltage.
FIGURE 3. Diagram of the electrical connection
Varian, Inc.
Page 28
Installation and Setup
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Setting up Your PC
The 400-DS Workstation software requires Microsoft Windows® 2000 or Windows XP® to be installed on your computer. For instructions on installing the operating system, refer to the documentation supplied with Windows®.
Minimum PC Configuration
• Windows® 2000 SP6 or
Windows XP® SP1
• 10 GB free hard disk drive space
• 256 MB RAM
• 1024 x 768 pixel display resolution
• 2x USB port (optional)
If the PC is not equipped with a standard serial port, a USB-to-serial converter must be purchased.
If Varian is installing the 400-DS Workstation software for you, note that the installation of a Windows® operating system is not included as part of the standard instrument installation.
Setting up Your PC
The monitor, keyboard, mouse, and printer/plotter are connected to the PC via cables which plug into the back of the PC. Consult your monitor and printer manuals for details of their individual cabling requirements.
Varian, Inc.
Revision A 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Cable Connections
FIGURE 4. Diagram of cable connections
Page 29
Installation and Setup
Caution
The Varian 400-DS can be connected to the PC using an RS232,
USB, or Ethernet cable. Do not connect more than one type of cable. Communication errors may occur.
Single System Setup
Step 1. Connect a cable between COM1 on the PC and the left serial port (IN
COM A) on the rear of the Varian 400-DS Automated Apparatus 7.
Step 2. Ensure the loop back switch (located to the right of the right serial port) is switched to the left (ON).
Multiple System Setup
Step 1. Connect a cable between COM1 on the PC and the left serial port (IN
COM A) on the rear of the Varian 400-DS Automated Apparatus 7.
Step 2. Ensure the loop back switch (located to the right of the right serial port) is switched to the right (OFF).
Varian, Inc.
Page 30
Installation and Setup
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Step 3. Connect a 6-pin mini-DIN cable between the right serial port (OUT COM
B) of the first apparatus and the left serial port of the second apparatus.
Step 4. If this is the final Varian 400-DS Automated Apparatus 7 in the series, ensure the loop back switch (located to the right of the right serial port) is switched to the left (ON).
If this is not the final Varian 400-DS Automated Apparatus 7 in the series, ensure the loop back switch (located to the right of the right serial port) is switched to the right (OFF).
Step 5. Repeat steps 3 and 4 for each additional apparatus.
Setting Up the Varian 400-DS
Step 1. Locate the 1.5 mm Allen wrench to install the sample heater jackets.
Step 2. Install the first cell heater by placing it over the thermistor at the desired cell position of the Varian 400-DS.
Step 3. While holding the jacket in place, use the Allen wrench to tighten the screw.
Step 4. Insert the 2-pin plug for the heater in the appropriate location on the sample heater jacket.
Step 5. Repeat steps 2 - 4 to install the sample heater jackets for the remaining cell positions.
Step 6. Install the magnetic plate on the reciprocation guide rod, which is the shorter rod in the middle of the 3 rods on the Varian 400-DS.
Step 7. Tighten the bolt on each end of the magnetic plate using the 9/64" Allen wrench.
Varian, Inc.
Revision A 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 31
Installation and Setup
Step 8. Install the guide rod stabilizers on top of the magnetic plate using the two longer guide rods for alignment.
Step 9. Tighten the two bolts on each stabilizer plate using the 9/64" Allen wrench.
Step 10. Locate the sample cells and sample holders provided with your instrument.
Step 11. Remove the sample cells and holders from the packing material to ensure no damage occurred during shipment.
Step 12. To install a sample cell in the heater jacket, slide one end through the magnetic lift plate.
Step 13. Use the finger tabs on the heater jacket to spread it open.
Step 14. Continue to lower the sample cell until it is secure and completely covers the bottom cell seal.
Caution
Be careful when sliding the sample cells over the O-ring seals as damage can occur. Ensure the cell and seal are properly aligned before applying pressure.
Step 15. With the magnet at the top, insert the sample holder into the sample cell.
Step 16. Repeat steps 12 - 15 to install the remaining sample cells and holders.
Step 17. Replace the evaporation cover and tighten the two thumb screws.
Varian, Inc.
Page 32
Installation and Setup
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Initial Power Up
Step 1. Ensure the circuit breaker on the rear of the instrument is on.
Step 2. Turn on the power switch on the front panel and ensure the integrated red LED illuminates.
Step 3. Repeat steps 1 and 2 for each Varian 400-DS Automated Apparatus 7.
Step 4. Turn on the PC, monitor, and printer. Verify the PC power light illuminates and the Microsoft ® Windows®XP or Windows® 2000 initializes.
Software Installation
Note
To complete the installation of the 400-DS Workstation software, you must log on to the PC as an administrator. Attaching the
Database and Local Security Policy are included as part of the software installation procedures. The CD contains all of the requirements for an installation.
To install the software, complete the following steps:
Step 1. Install the 400-DS Workstation CD and access the files.
Step 2. Execute the file setup.exe and follow the on-screen prompts.
Step 3. When asked if you want to install the Microsoft .Net Framework V1.1, you must click Yes.
Step 4. After the installation has successfully completed, install the license dongle by plugging it into an open USB port.
Varian, Inc.
Revision A 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
FIGURE 5. License dongle
Page 33
Installation and Setup
Attaching the Database
Note
This section is only required when upgrading an existing installation.
Once the basic installation is complete, complete the following steps to attach the database files:
Step 1. Click Start > All Programs > Varian > Dissolution > MSDE Manager.
The MSDE Manager screen displays.
Varian, Inc.
Page 34
Installation and Setup
Varian 400-DS Automated Apparatus 7
Operator’s Manual
FIGURE 6. MSDE Manager screen
Revision A, 05/07
P/N 70-9051
Step 2. Select a machine name from the list of available servers. When connecting to a database installation on a local machine, the encrypted password is automatically entered in the password box. When connecting to a remote database, your system administrator may have to provide you with a password.
Step 3. Click Connect to connect to the server or local PC.
Varian, Inc.
Revision A 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 35
Installation and Setup
Step 4. Select the Attach tab on the MSDE Manager screen and attach the database by clicking Attach. Leave all the values at their default settings. After successful attachment, the Database Attach screen displays.
Step 5. Click OK.
Step 6. Close the MSDE Manager screen.
Local Security Policy
For 21 CFR Part 11 compliance purposes, you must ensure that the following minimum requirements are met by your system’s security policy.
Note
It may be necessary to coordinate the following configurations with the domain administrator.
Step 1. Click Start > Run to run the Local Security Settings Manager. Type
secpol.msc and press Enter. The Local Security Settings manager
displays.
Varian, Inc.
Page 36
Installation and Setup
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Step 2. Click Security Settings > Account Policies > Password Policy and set the following security policy configuration:
Policy
Enforce Password History
Maximum password age
Minimum password length
Password Must Meet
Complexity Requirements
Security Setting
3 passwords remembered
30 days
6 characters
Enabled
Step 3. Click Security Settings > Account Policies > Account Lockout
Policy and set the following configuration:
Policy
Account lockout duration
Account lockout threshold
Reset account lockout counter
Security Setting
0 minutes (infinite)
3 invalid login attempts
99999 minutes
Step 4. Click Security Settings > Local Policies > Audit Policy and set the following configuration:
Policy
Audit account logon events
Audit account management
Audit logon events
Audit policy change
Security Setting
Success, Failure
Success, Failure
Success, Failure
Success, Failure
Step 5. Close the Local Security Settings Management screen.
Varian, Inc.
Revision A 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 37
Installation and Setup
E-mail Notification Configuration (Optional)
Step 1. On the computer which is hosting the DissoService, click Start > All
Programs > Varian > Dissolution > Service Manager. The
DissoService Manager screen displays.
FIGURE 7. DissoService Manager screen
Step 2. On the toolbar, click Tools and select Stop Service. The Options item on the Tools menu enables after the service stops (see Figure 8, “Tools menu,” on page 38).
Varian, Inc.
Page 38
Installation and Setup
Varian 400-DS Automated Apparatus 7
Operator’s Manual
FIGURE 8. Tools menu
Revision A, 05/07
P/N 70-9051
Step 3. On the toolbar, click Tools > Options. The DissoService Options screen displays (see Figure 9, “DissoService Options screen,” on page 39).
Varian, Inc.
Revision A 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
FIGURE 9. DissoService Options screen
Page 39
Installation and Setup
Step 4. Configure the SMTP server address.
Note
The e-mail alert software assumes the use of an SMTP server which does not require authentication for sending. Most e-mail servers only require authentication for receiving (POP3), but allow sending from any system on the same domain without the use of a password.
If your corporate server does not require passwords, provide the address of the corporate SMTP server in order for Dissolution Server to send email. Enter the SMTP server address as either a name (for example, server.domain.com) or as an IP address (for example, 192.168.1.1).
If your corporate SMTP server requires passwords, consider installing
Microsoft Internet Information Server (IIS) to a server on your intranet through which other Dissolution Servers could send e-mails. IIS includes a SMTP server which allows unrestricted e-mail sending.
Varian, Inc.
Page 40
Installation and Setup
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Step 5. Once you have configured the SMTP server address, click OK to save the settings.
Step 6. On the toolbar, click Tools and select Start Service.
Starting 400-DS Workstation
Step 1. Double-click the 400-DS Workstation icon on the Windows desktop to start the software.
Step 2. If your system has Windows Firewall installed, the following displays:
Click Unblock to enable the program.
Varian, Inc.
Revision A 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 41
Installation and Setup
Dissolution Server Selection
Note
A dissolution server must be selected when logging into a server for the first time. The same user ID and password that was used to log on to Windows should be used to log on to the 400-DS
Workstation software.
Step 1. From the logon screen, enter a domain user identification and password.
Step 2. Click the drop-down arrow corresponding to Dissolution server and select Add New. The Add New Dissolution Server screen displays (see
Figure 11, “Add New Dissolution Server screen,” on page 42).
FIGURE 10. Logon screen
Varian, Inc.
Page 42
Installation and Setup
Varian 400-DS Automated Apparatus 7
Operator’s Manual
FIGURE 11. Add New Dissolution Server screen
Revision A, 05/07
P/N 70-9051
Step 3. If you know the domain name of the computer on which the Dissolution
Server is installed, enter the name in the Server Name box.
Alternately, click the network browse button (…) to select a server from a list of computers on your network (see Figure 12, “Browse Dissolution
Servers screen,” below).
Note
If your network has a large number of systems connected, browsing the network may take several minutes.
FIGURE 12. Browse Dissolution Servers screen
Step 4. The default server port is 8787, but may be changed to any other port.
Coordinate the selection of a port number with your I.T. department to satisfy any port restriction requirements they may have.
Varian, Inc.
Revision A 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 43
Installation and Setup
Step 5. Click OK.
Step 6. Click Logon on the Logon screen.
Step 7. After successfully logging onto the software, click Tools > Options.
The
Configuration Dialog screen displays (see Figure 13, “Configuration
Dialog screen,” on page 43).
Configuring the Application - Adding Users
Note
In order to configure the application, it is necessary to be logged on as an administrator.
FIGURE 13. Configuration Dialog screen
Varian, Inc.
Page 44
Installation and Setup
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Step 1. To add a user to a group, select the Security tab on the Configuration
Dialog screen.
Note
Only an administrator can add users.
Step 2. Click User Administration at the bottom of the screen. The Local Users and Groups screen displays.
FIGURE 14. Local Users and Groups screen
Step 3. Double-click the Groups folder to expand the list of groups.
Step 4. Click Add... to add users to the group. The Select Users, Computers, or
Groups screen displays.
Step 5. Enter the names of the users to add to the group and click OK.
Step 6. Repeat steps 3-5 as appropriate for each group that requires new users.
Varian, Inc.
Revision A 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 45
Installation and Setup
Step 7. Close the window.
Step 8. Select the Miscellaneous tab (see Figure 15, “Miscellaneous tab,” on page 45). To run the software without physically connecting instruments, check the Use Simulated Instruments box. The system operates in simulation mode.
Step 9. Enter an appropriate length of time in minutes the application can be idle before locking.
FIGURE 15. Miscellaneous tab
Step 10. Click OK. The application has been configured.
Varian, Inc.
Page 46
Installation and Setup
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
This page was intentionally left blank, except for this message.
Varian, Inc.
Chapter 4
Operation
Log on to the Varian 400-DS Workstation
Step 1. Double-click the Varian 400-DS Workstation icon on your desktop. The
Logon screen displays.
FIGURE 16. Logon screen
Varian, Inc.
Page 48
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Step 2. Enter your user identification and password. Verify the domain is correct and click Logon. The 400-DS Workstation screen displays.
FIGURE 17. 400-DS Workstation screen
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 49
Operation
Following is a description of the screen options listed on the navigation bar of the 400-DS
Workstation screen:
Option
System
Method
Status
Description
Configuration Use this option to configure the system. See “Configuring
Your System” on page 50.
Diagnostics Use this option to check the diagnostics of a dissolution apparatus or accessory within the system. See “Manual
Control/Diagnostics” on page 64.
Editor Use this option to create a method, change or delete method parameters, run a report of the method setup, view the method audit trail, and verify the integrity of the method.
See “Method Editor” on page 69.
Run Method Use this option to run a method. See “Running the Method” on page 82.
Test Reports Use this option to run a report of the completed method.
See “Test Reports” on page 87.
Security Change User To change the user, click Change User. The Logon screen displays. Enter the appropriate user identification and password and click Logon.
Lock
Application
To lock the 400-DS Workstation screen, click Lock
Application. The 400-DS Workstation Locked screen displays. Click the lock or the link to unlock the screen. The
Logon screen displays. Enter the appropriate user identification and password as applicable and click Logon.
Audit Trail Click Audit Trail. The Security Audit Trail screen displays.
Click Show Report to display the report. Change the start date, end date, and/or user identification as desired and click Retrieve Records to change the parameters for the information displayed.
Permissions Use this option to view permissions assigned to the current user and the descriptions of the corresponding privileges.
Click Permissions. The User Group Membership screen displays.
Click one of the pie-shaped graphics under status to view the status of the corresponding dissolution system.
Varian, Inc.
Page 50
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Configuring Your System
It is necessary to configure the components that will be used for automated sample collection. Systems can be added, modified, and removed from the database. All system configuration activity is recorded in the system audit log. The dissolution software allows the configuration of multiple systems. A maximum of four systems can be running methods at one time.
System configuration entails selecting the appropriate equipment and setting the communication and other physical properties of the system. Serial numbers are stored for each system to allow tracking of physical system changes.
Step 1. From the navigation bar, click Configuration. The System Configuration screen displays.
FIGURE 18. System Configuration screen
Step 2. Click Create. The System Editor screen displays (see Figure 19,
“System Editor screen,” on page 51).
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
FIGURE 19. System Editor screen
Page 51
Operation
Following is a description of the System Editor screen options:
Screen
Section
System
Parameter Description
System
Name
Enter a name for your system.
Laboratory Enter a laboratory name.
Varian, Inc.
Page 52
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Screen
Section
Restrict
Execution To
Specified
Server
Change
Management
Parameter Description
Click to display the name of the workstation or enter the name of the workstation connected to the system. Click
No Restriction to allow the system to be run from any workstation.
Note: The system must be physically connected.
If applicable, select the box under Change Management in order to restrict the system editing rights to the current user.
Note: This option prevents system configuration modification by anyone other than the system owner or a user with VkModifyOthersSystems privilege.
Step 3. Click Next. The 400-DS Workstation screen displays.
FIGURE 20. 400-DS Workstation screen
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 53
Operation
Following is a description of the 400-DS Workstation screen options:
Screen Section
Varian 400-DS
Automated
Apparatus 7
Cell Calibration
Temperatures
Moisture Sensor
Control
Parameter
Serial Number
Cell Size
Vial Size
Syringe Size
Manufacturer
Model Name
Serial Number
Set Point
Temperature (ºC)
Stop On Internal
Moisture Detection
Volume Calibration
Coefficient
Offset
Slope
Response
Click ... to display the serial numbers of all connected instruments. Highlight the appropriate serial number and click Select.
Click the drop-down arrow to indicate the appropriate cell size.
Click the drop-down arrow to indicate the appropriate vial size.
Click the drop-down arrow to indicate the appropriate syringe size.
Enter the manufacturer of the calibrated temperature probe.
Enter the model name of the calibrated temperature probe.
Enter the serial number of the calibrated temperature probe.
Enter the set point for the cell temperature.
Enter the temperature for each cell as determined from the diagnostic (see “Cell
Calibration Temperatures” on page 62).
Check this option to shut down the system when moisture is detected.
Note: Enabling this option is recommended.
The current offset displays in this area.
The current slope displays in this area.
Step 4. Click Next. The Serial Numbers screen displays.
Step 5. Enter the serial number for the sample cells, sample holders, etc. and click Add.
Step 6. Click Finish.
Varian, Inc.
Page 54
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Step 7. Repeat all the sections under “Configuring Your System” on page 50 for each additional system.
Step 8. Close the System Configuration screen.
Copying a System Configuration
To copy a system configuration, complete the following steps:
Step 1. From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 18, “System Configuration screen,” on page 50).
Step 2. Select the desired system configuration.
Step 3. Click Copy.
Step 4. Click Paste. A new system configuration displays. The description of the new system configuration is Copy of....
Step 5. Close the System Configuration screen.
Step 6. To edit the system configuration, see “Editing an Existing System
Configuration” on page 56.
Deleting a System Configuration
To delete a system configuration, complete the following steps:
Step 1. From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 18, “System Configuration screen,” on page 50).
Step 2. Select the desired system configuration.
Step 3. Click Delete.
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 55
Operation
Step 4. Click Yes.
Step 5. Close the System Configuration screen.
Note
The system configuration is never physically deleted. It is marked as deleted in the database.
Serial Numbers
To review or add serial numbers to the system configuration, complete the following steps:
Step 1. From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 18, “System Configuration screen,” on page 50).
Step 2. Select the desired system configuration.
Step 3. Click Serial Numbers. The Serial Number Editor screen displays.
Note
Alternately, you can double-click the desired system configuration and double-click Next until the Serial Numbers section of the System Editor displays or right-click the desired system configuration and select Serial Numbers. Serial numbers can be added or deleted as a result of any of these actions.
Step 4. Enter the serial number for the sample cells, sample holders, etc. and click Add.
Step 5. Click OK. The Serial Number Editor screen closes.
Varian, Inc.
Page 56
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
System Configuration Report
To display a report of the system configuration, complete the following steps:
Step 1. From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 18, “System Configuration screen,” on page 50).
Step 2. Select the desired system configuration.
Step 3. Click Report. By default, the most recent report version displays.
Note
Alternately, you can right-click the desired system configuration and select Report. The Version Selection screen displays as a result of either of these actions.
Step 4. Use the up and down arrows to indicate the desired report version and click OK.
Step 5. The system report displays. The report can be printed, exported, searched, or verified.
Editing an Existing System Configuration
To display and edit the properties of an existing system configuration, complete the following steps:
Step 1. From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 18, “System Configuration screen,” on page 50).
Step 2. Select the desired system configuration.
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 57
Operation
Step 3. Click Properties. The System Editor screen displays (see Figure 19,
“System Editor screen,” on page 51).
Note
Alternately, you can double-click the desired system configuration or right-click the desired system configuration and select Properties. The System Editor screen displays as a result of any of these actions.
Step 4. Select the appropriate tabs and change the relevant information in the same manner that the system was created.
Step 5. Close the System Configuration screen.
Show Audit Trail
To display the audit trail for a system configuration, complete the following steps:
Step 1. From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 18, “System Configuration screen,” on page 50).
Step 2. Select the system configuration and click Show Audit Trail. The System
Audit Trail screen displays.
Note
Alternately, you can right-click the desired system configuration and select Show Audit Trail. The System Audit Trail screen displays as a result of either of these actions.
Step 3. Select two or more versions and click Differences in the navigation bar.
A change report displays. The report can be printed or exported.
Varian, Inc.
Page 58
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Verify Integrity
To verify that the system configuration has not been changed outside of the 400-DS
Workstation program, complete the following steps:
Step 1. From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 18, “System Configuration screen,” on page 50).
Step 2. Select the system configuration and click Verify Integrity. Either the data is verified successfully or the user is directed to contact their system administrator.
Note
Alternately, you can right-click the desired system configuration and select Verify Integrity. The integrity of the system configuration is checked as a result of either of these actions.
Step 3. Click OK to close the Data Verification screen.
Step 4. Close the System Configuration screen.
Import/Export XML File
To use an existing system configuration from one 400-DS Workstation computer on a different 400-DS Workstation computer, you can export and import the system configuration as an XML file.
To export the system configuration, complete the following steps:
Step 1. From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 18, “System Configuration screen,” on page 50).
Step 2. Select the desired system configuration.
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 59
Operation
Step 3. Click Export. The Version Selection screen displays.
Note
Alternately, you can right-click the desired system configuration and select Export. The Version Selection screen displays as a result of either of these actions.
Step 4. If applicable, indicate which version to export and click OK. The Export
System to XML File screen displays.
Note
Each time a system configuration is saved, a new version is created. To export a version other than the most recently saved, indicate the appropriate version number on the Version Selection screen.
Step 5. Indicate the directory and file name and click Save. The code is saved as an XML file.
To import the XML file at another 400-DS Workstation, complete the following steps:
Step 1. From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 18, “System Configuration screen,” on page 50).
Step 2. Click Import. The Import System from XML File screen displays.
Note
Alternately, you can right-click the desired system configuration and select Import. The Import System from XML File screen displays as a result of either of these actions.
Step 3. Select the appropriate directory and file name and click Open. The system configuration displays on the System Configuration screen.
Step 4. Close the System Configuration screen.
Varian, Inc.
Page 60
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
System Calibration
Step 1. Place the Media A tube into a suitable container with deionized water.
Step 2. From the navigation bar, click Configuration.
Step 3. Select desired configuration.
Step 4. Click Calibration. The System Calibration screen displays.
FIGURE 21. System Calibration screen
Step 5. Label 13 empty 4 mL sample vials (Row 1 Vial 1 - Row 1 Vial 13).
Step 6. Label 13 empty 1.5 mL sample vials (Row 2 Vial 1 - Row 2 Vial 13).
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 61
Operation
Step 7. Weigh each of the vials and record the empty masses in the assigned cells in the table.
Step 8. Place the sample vials in row 1 (4 mL) and row 2 (1.5 mL) of the sample tray.
Step 9. Click Fill Calibration Volumes to fill the vials in the first row with 4 mL and the second row with 1 mL of deionized water.
Step 10. After the fill process is complete, weigh each of the vials and record their filled masses in the table. The vial volume calculates and displays in the table automatically.
Step 11. Click OK. The Reason for Change screen displays.
Note
The software performs a Least Squares Fit analysis. The volume deviation from the curve is 1.5% of the vial volume. An automated error message is generated if the calibration fails.
Step 12. Enter a reason code and change description and click OK.
Step 13. Highlight the appropriate system.
Step 14. Click Properties.
Step 15. Click Next and ensure the Volume Calibration Coefficients are visible.
Step 16. Click Finish.
Step 17. Close the System Configuration screen.
Varian, Inc.
Page 62
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Cell Calibration Temperatures
Step 1. Place the rinse media line into a suitable container holding deionized water.
Step 2. From the navigation bar, click Diagnostics. The Select Item screen displays.
Step 3. Select the desired system and click OK. The System Diagnostics screen displays.
FIGURE 22. System Diagnostics screen
Step 4. In the Cells box under Media Control, click Fill. Media moves into the cells.
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 63
Operation
Step 5. In the Cell Heat box, set the cell temperature to an appropriate temperature.
Step 6. Click . D
O THE SAMPLE CELLS CONTAIN LIQUID
? displays.
Step 7. Click Yes. The temperatures display in the Cell Temperature box.
Step 8. Set the reciprocation rate and click Run.
Step 9. Allow the media to equilibrate to the set temperature.
Step 10. In the Evaporation Cover box, click the up arrow to raise the evaporation cover.
Step 11. Remove the two thumb screws from the evaporation cover and remove the evaporation cover.
Step 12. Remove the sample holder for cell 1 and replace with the modified temperature probe.
Step 13. Allow 3 - 5 minutes for the temperature to stabilize.
Step 14. Note the temperatures for use at step 23.
Step 15. Repeat steps 12 - 14 for each additional cell.
Step 16. Replace the sample holders and the evaporation cover.
Step 17. Stop reciprocation and cell heating.
Step 18. Under the Media Control box, click Purge. The media in all cells purges.
Varian, Inc.
Page 64
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Step 19. Close the Diagnostics screen.
Step 20. From the navigation bar, click Configuration. The System Configuration screen displays (see Figure 18, “System Configuration screen,” on page 50).
Step 21. Select the appropriate system and click Next.
Step 22. Enter the set point temperature as appropriate based on the cell calibration temperatures.
Step 23. Enter the temperature values for each cell (as noted in step 14) into the cell calibration field.
Step 24. Click Finish. The Reason for Change screen displays.
Step 25. Enter a reason code and change description and click OK.
Step 26. Close the System Configuration screen.
Manual Control/Diagnostics
To check the diagnostics of a dissolution apparatus, complete the procedures on the following pages. These procedures are performed on one system at a time. Repeat the procedures as applicable for each additional system.
Step 1. From the navigation bar, click Diagnostics. The Select Item screen displays.
Step 2. Select the desired system and click OK. The System Diagnostics screen displays (see Figure 22, “System Diagnostics screen,” on page 62).
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 65
Operation
Verifying Sample Tray Control
Step 1. Click any vessel position corresponding to the desired row in the Sample
Tray Control box. The tray moves and the indicated row raises.
FIGURE 23. Sample Tray Control box
Step 2. Click under the Sample Tray Control box. The drive unit returns to the home position.
Step 3. Click to stop the movement.
Step 4. Click to eject or retract the sample tray.
Step 5. Click to raise the evaporation cover.
Varian, Inc.
Page 66
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Step 6. Click to lower the evaporation cover.
Step 7. Click to turn off any audible alarm.
Media Control
Step 1. Under the Media Control box, set the following parameters:
Option
Aspirate Flow Rate
(mL/s)
Dispense Flow Rate
(mL/s)
Aspirate Dwell Time
(s)
Description
Use the up and down arrows to indicate the rate at which the media is pulled into the syringe.
Use the up and down arrows to indicate the rate at which the media is dispensed from the syringe.
Use the up and down arrows to indicate the duration of time the plunger holds at the top of the stroke before dispensing.
Note: The time specified is for full syringe volume. For smaller syringe volumes, it is calculated proportionately.
Needle Clean
Cycles
Prime Volume (mL) Use the up and down arrows to indicate the volume to be pulled before taking measurement.
Media Source
Media Volume
Use the up and down arrows to indicate the number of times the unit performs the cleaning cycle.
Use the drop-down arrow to indicate the source of the media.
Use the up and down arrows to indicate the desired volume of the media.
Step 2. Under the Media Control box, click Prime.
Step 3. Under the Cells box, click Fill. The media moves into the cells.
Step 4. Under the Sample Vials box, use the up and down arrows to indicate the desired row number to sample into and the desired volume.
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 67
Operation
Step 5. Under the Sample Vials box, click Fill Row. The media moves from the cell into the sample vial.
Step 6. Under the Waste Pump box, click On to turn on the waste pump.
Step 7. Under the Waste Pump box, click Off to turn off the waste pump.
Verifying the Cell Temperature
Step 1. In the Cell Heat box, use the up and down arrows to set the media cell temperature to an appropriate temperature (see the sample screen below).
Step 2. Click . D
O THE SAMPLE CELLS CONTAIN LIQUID
? displays.
Step 3. Click Yes. The temperatures display in the Cell Temperature box.
Step 4. As applicable, click .
Step 5. Under the Cells box, click Purge to purge the remaining media from the tube.
Caution
Do not click Fill more than one time without purging in between.
The media will overflow the cells.
Varian, Inc.
Page 68
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Reciprocation
To set the dips per minute, complete the following steps:
Step 1. Click any vessel position corresponding to the desired row in the
Sample Tray Control box (see Figure 23, “Sample Tray Control box,” on page 65). The drive unit moves to the indicated row.
Step 2. In the Reciprocation box, use the up and down arrows in the box that corresponds to DPM, to set the dips per minute to 15 and click
Dipping begins.
FIGURE 24. Reciprocation box
.
Step 3. Click . The dipping stops.
Step 4. Click under the Sample Tray Control box. The drive unit returns to the home position.
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Controlling the Evaporation Cover
To raise the evaporation cover, click in the Evaporation Cover box.
To lower the evaporation cover, click in the Evaporation Cover box.
Page 69
Operation
Method Editor
All test parameters are entered via the Method Editor screen.
Function
Create a new method
Copy a method
Delete a method
Edit an existing method
Run a report of the method setup
View the method audit trail
Verify the integrity of a method
Import or export the method between two 400-DS Workstations
Procedure
See “Creating a Method” on page 70.
See “Copying Methods” on page 77.
See “Deleting Methods” on page 77.
See “Editing an Existing Method” on page 78.
See “Method Report” on page 78.
See “Show Audit Trail” on page 79.
See “Verify Integrity” on page 80.
See “Import/Export XML File” on page 80.
Varian, Inc.
Page 70
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Creating a Method
To create a new method, complete the following steps:
Step 1. From the navigation bar, click Editor. The Methods screen displays.
FIGURE 25. Methods screen
Step 2. Click New Method. The Method Editor screen displays.
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
FIGURE 26. Method Editor screen
Page 71
Operation
Varian, Inc.
Page 72
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Step 3. Select the Parameters tab.
Following is a description of the Parameters tab options:
Option
Sample
Information
Error Tolerance
(±)
Product Name
Notes
User Defined Labels
Label 1
Label 2
Label 3
Temperature
Profile Interval
Speed
Change Management
Description
Enter the name of the product.
Enter any relevant notes regarding the sample.
The fields in this area have default values of LOT,
BATCH, and GROUP. These fields are customizable; enter the information that best serves the needs of the method parameters.
Enter the desired temperature fluctuation limit (±).
If the tube temperature goes over or under the set temperature (see “Verifying the Cell Temperature” on page 67) by the amount of this tolerance, an error is recorded as part of the results.
Enter the desired speed fluctuation limit (±). If the
RPM goes over or under the set speed (see
“Reciprocation” on page 68) by the amount of this tolerance, an error is recorded as part of the results.
Enter the timepoint in hh:mm format at which the temperature and speed settings are recorded.
Note: Profile measurements are optional.
Values are always recorded at sample timepoints independent of this setting.
If applicable, select the box under Change
Management in order to restrict the method editing rights to the current user or any user with
VkModifyOthersMethod.
Step 4. Select the Automated App7 Sequence tab (see Figure 27, “Automated
App7 Sequence tab,” on page 73).
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
FIGURE 27. Automated App7 Sequence tab
Page 73
Operation
Following is a description of the Automated App7 Sequence tab options:
Option
Sample Row
Media
Sampled Volume (mL)
Dissolution Volume (mL)
Description
Use the up and down arrows to indicate the row number for which the parameters are being set. After clicking Add to List, use the up and down arrows to indicate the next row.
Use the drop-down arrows to indicate Media A, Media B,
Media C, Media D, or Media E.
Use the up and down arrows to indicate appropriate volume based on your vial size.
Use the up and down arrows to indicate appropriate volume based on your cell size.
Varian, Inc.
Page 74
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Option
Cell Temperature
Hold Time (mm:ss)
Dip Speed (DPM)
Dip Interval (hhh:mm:ss)
Description
Enter the desired tube temperature in degrees Celsius. This can be entered for the first timepoint.
Enter the desired duration for the dip to remain at the bottom of the stroke in mm:ss format.
Use the up and down arrows to indicate the desired dips per minute (DPM) for each applicable row of the 400-DS.
Enter the desired duration for dipping in hhh:mm:ss format.
Step 5. Select the Automated App7 Options tab.
FIGURE 28. Automated App7 Options tab
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 75
Operation
Following is a description of the Automated App7 Options tab options:
Screen
Section
Media
Description
Syringe
Properties
Parameter
Media A
Media B
Media C
Media D
Media E
Rinse Media
Aspirate Flow Rate
(mL/s)
Dispense Flow
Rate (mL/s)
Aspirate Dwell
Time (s)
Description
Enter descriptions as appropriate for the different medias.
Use the up and down arrows to indicate the rate at which the media is pulled into the syringe.
Use the up and down arrows to indicate the rate at which the media is dispensed from the syringe.
Use the up and down arrows to indicate the duration of time the plunger holds at the top of the stroke before dispensing.
Note: The time specified is for full syringe volume.
For smaller syringe volumes, it is calculated proportionately.
Prime Volume (mL) Use the up and down arrows to indicate the volume to be pulled before taking measurement.
Syringe Clean
Cycles
Use the up and down arrows to indicate the number of times to rinse the syringe between samples.
Sample Cell Clean
Cycles
Needles Clean
Cycles
Use the up and down arrows to indicate the number of times to rinse the sample cell between samples.
Use the up and down arrows to indicate the number of times to rinse the needles between samples
Step 6. Select the Notifications tab (see Figure 29, “Notifications tab,” on page 76).
Varian, Inc.
Page 76
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
FIGURE 29. Notifications tab
Revision A, 05/07
P/N 70-9051
Step 7. Enter appropriate e-mail address(es) to receive notification of system operation.
Step 8. Indicate which options would require notification.
Step 9. Click OK to close the Method Editor screen.
Step 10. Close the Methods screen.
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 77
Operation
Copying Methods
To copy a method, complete the following steps:
Step 1. From the navigation bar, click Editor. The Methods screen displays (see
Figure 25, “Methods screen,” on page 70).
Step 2. Select the desired method.
Step 3. Click Copy Method.
Step 4. Click Paste. A new method displays. The description of the new method is Copy of....
Step 5. To change any of the parameters of the method, see “Editing an Existing
Method” on page 78.
Deleting Methods
To delete a method, complete the following steps:
Step 1. From the navigation bar, click Editor. The Methods screen displays (see
Figure 25, “Methods screen,” on page 70).
Step 2. Select the desired method.
Step 3. Click Delete.
Step 4. Click Yes.
Varian, Inc.
Page 78
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Editing an Existing Method
To edit a method already entered on the 400-DS Workstation, complete the following steps:
Step 1. From the navigation bar, click Editor. The Methods screen displays (see
Figure 25, “Methods screen,” on page 70).
Step 2. Click the desired method.
Step 3. Click Properties on the navigation bar. The Method Editor screen displays (see Figure 26, “Method Editor screen,” on page 71).
Note
Alternately, you can double-click the desired method or right-click the desired method and select Properties. The Method Editor screen displays as a result of any of these actions.
Step 4. Select the appropriate tabs and change the relevant information in the same manner that the method was created.
Method Report
To display a report of the method parameters, complete the following steps:
Step 1. From the navigation bar, click Editor. The Methods screen displays (see
Figure 25, “Methods screen,” on page 70).
Step 2. Select the desired system configuration.
Step 3. Click Report. The Version Selection screen displays.
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 79
Operation
Step 4. If applicable, indicate which version and click OK.
Note
Each time a method is saved, a new version is created. To create a report of a version other than the most recently saved, indicate the appropriate version number on the Version Selection screen.
Step 5. The method report displays. The report can be printed, exported, searched, verified, and/or signed.
Audit Trail
Once a method has completed, the results are available for review, audited modification, and electronic signature. The software maintains complete history for all runs executed on the system. Results can be previewed and printed.
Show Audit Trail
To display the audit trail for a method, complete the following steps:
Step 1. From the navigation bar, click Editor. The Methods screen displays (see
Figure 25, “Methods screen,” on page 70).
Step 2. Select the method and click Show Audit Trail. The Method Audit Trail screen displays.
Note
Alternately, you can right-click the desired method and select
Show Audit Trail. The Method Audit Trail screen displays as a result of either of these actions.
Step 3. Select two or more versions and click Differences in the navigation bar.
A change report displays. The report can be printed or exported.
Varian, Inc.
Page 80
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Verify Integrity
To verify that the method has not been changed outside of the application, complete the following steps:
Step 1. From the navigation bar, click Editor. The Methods screen displays (see
Figure 25, “Methods screen,” on page 70).
Step 2. Select the method and click Verify Integrity. Either the data is verified successfully or the user is directed to contact their system administrator.
Note
Alternately, you can right-click the desired method and select
Verify Integrity. The integrity of the method is checked as a result of either of these actions.
Import/Export XML File
To use an existing method from one 400-DS Workstation on a different 400-DS
Workstation, you can export and import the method as an XML file.
To export the method, complete the following steps:
Step 1. From the navigation bar, click Editor. The Methods screen displays (see
Figure 25, “Methods screen,” on page 70).
Step 2. Select the desired method.
Step 3. Click Export. The Version Selection screen displays.
Note
Alternately, you can right-click the desired method and select
Export. The Version Selection screen displays as a result of either of these actions.
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 81
Operation
Step 4. If applicable, indicate which version to export and click OK. The Export
Method to XML File screen displays.
Note
Each time a method is saved, a new version is created. To export a version other than the most recently saved, indicate the appropriate version number on the Version Selection screen.
Step 5. Indicate the directory and file name and click Save. The code is saved as an XML file which displays in Notepad.
Step 6. Close the Notepad file.
To import the XML file at another 400-DS Workstation, complete the following steps:
Step 1. From the navigation bar, click Editor. The Methods screen displays (see
Figure 25, “Methods screen,” on page 70).
Step 2. Click Import. The Import Method from XML File screen displays.
Note
Alternately, you can right-click the desired method and select
Import. The Import Method from XML File screen displays as a result of either of these actions.
Step 3. Select the appropriate directory and file name and click Open. The method displays on the Method screen.
Varian, Inc.
Page 82
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Running the Method
Step 1. From the navigation bar, click Run Method. The Select System Step 1 of 2 screen displays (see Figure 30, “Select System Step 1 of 2 screen,” on page 82).
FIGURE 30. Select System Step 1 of 2 screen
Step 2. Select the desired system to run the method and click Next. The Select
Method Step 2 of 2 screen displays.
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
FIGURE 31. Select Method Step 2 of 2 screen
Page 83
Operation
Step 3. Select the desired method to run and click Finish. The system status screen displays.
Varian, Inc.
Page 84
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
FIGURE 32. System status screen
Revision A, 05/07
P/N 70-9051
Step 4. Click Start. The Method Start Options screen displays (see Figure 33,
“Method Start Options screen,” on page 85).
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
FIGURE 33. Method Start Options screen
Page 85
Operation
Following is a description of the Method Start Options screen options:
Screen Section Parameter
Sample
Information
Product Name
Notes
Label 1
Label 2
Label 3
Active Channels
Description
Enter the product name.
Enter any appropriate notation.
Enter the appropriate information based on the user-defined labels (see “User Defined Labels” on page 72).
Enter the number of active heating channels.
Note: Heating channels are counted from left to right.
Varian, Inc.
Page 86
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Screen Section Parameter
Injection Control Enable Injections
Temperature
Delayed Start
Inject Only (No
Sampling)
Vessel Temperature
Start
Bath Temperature
Start
Time Delayed
Start
Description
N/A
N/A
Use this option to program a delayed start.
Enter the desired date and time to start the method.
Step 5. Click OK. The Method Sample Information screen displays.
Step 6. Enter the sample information, if applicable.
Step 7. When the method is complete, click Test Report. The test report displays. The report can be printed, exported, searched, verified, and/or signed.
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 87
Operation
Test Reports
To display a report of the completed method, complete the following steps:
Step 1. From the navigation bar, click Test Reports. The Test Report Selection screen displays.
FIGURE 34. Test Report Selection screen
Note
By default, the test report dialog only shows tests done in the previous week. To view tests completed earlier than the previous week, click the
Start Date and/or End Date drop-down arrows to indicate date range for a test report selection and click Filter Date and System.
Additionally, you can select a system name from the System combo box to restrict the selection to a particular system.
To restrict the selection to a particular test identification, enter the value in the Test ID Filter box and click Filter Test ID.
Step 2. Click a test in the list to select it.
Step 3. From the navigation bar in the Test Report Selection dialog, click Show
Report. The report for the selected test displays.
Varian, Inc.
Page 88
Operation
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Step 4. Using the buttons on the report toolbar, you can do the following: first page previous page next page print the report skip to a specific page by number last page
Electronic Signatures
When the user is satisfied with the results, the results can be electronically signed. The software allows multiple electronic signings of a set of results. Each signing is accomplished using the signature dialog box shown. The user authenticity is determined by testing the user identification and password against the Windows security database.
Electronic signatures are permanently linked to the results. The software always requires the signature to be executed using all the signature components. Any attempts to sign a set of results using an invalid user identification, password, or any combination thereof that is incorrect is automatically recorded to the system audit trail.
Clean System
From the system status screen, click . All instruments initialize. Media is pulled from the Rinse Media Container and cells are filled and expelled through the valves. The cells are aspirated and purged.
Varian, Inc.
Chapter 5
Maintenance
Periodic Maintenance
Warning
The tester contains electrical circuits, devices, and components operating at dangerous voltages. Contact with these circuits, devices, and components can cause death, serious injury, or painful electric shock.
Warning
The apparatus should be disconnected from AC power before conducting cleaning or maintenance.
Periodic maintenance intervals may vary depending on frequency of instrument usage.
Varian, Inc.
Page 90
Maintenance
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Routine Maintenance (to be Performed
Between Each Dissolution Method
•
All parts of the 400-DS exposed to the dissolution medium should be cleaned after each use. Parts made from stainless steel are particularly susceptible to surface corrosion if not cleaned immediately after use. If any stainless steel parts show signs of surface discoloration, lightly wipe the surface with a soft cloth or nonabrasive pad to remove it.
•
Execute the Clean System cycle using the 400-DS Workstation software (“Clean
System” on page 88). Ensure an appropriate cleaning solution is in place to thoroughly rinse the media lines and sample cells.
•
If an alternate cleaning solution is used, it is recommended that all the media lines are rinsed with deionized water at the end of the cleaning procedure.
Sample Holders
•
Clean sample holders thoroughly after each use and place them in an ultrasonic bath for a few minutes, if necessary.
•
When using corrosive materials such as hydrochloric acid or medium containing salts, be sure to rinse the holders with deionized water immediately after use. Dry completely with a soft towel or cloth (preferably lint-free).
•
Do not clean the sample holders with abrasive cleansers or cloths. Use deionized water whenever possible. If you must use cleanser or solvent, be sure that it is as mild as possible, non-abrasive, and fully compatible with fluorocarbons and stainless steel before use.
•
It is suggested that the sample holders are cleaned by hand. Use of a laboratory dishwasher is NOT recommended.
•
Handle with care at all times.
•
Store the sample holders properly between each use.
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Page 91
Maintenance
Non-Routine Maintenance (to be Performed at Pre-Determined Intervals)
It is recommended that the following procedures be performed by a properly trained service technician.
•
Remove and inspect all sample cells for signs of damage. Replace as necessary.
Clean the sample cells thoroughly, especially the area that typically covers the seal.
•
Clean the entire external surface of the instrument including top cover and syringe pump window.
•
Inspect and clean the air filter(s). Replace if necessary.
•
Inspect all sample cell holder seals. Re-grease the o-rings, if necessary, using a pre-approved lubricant (e.g. Krytox®). Replace any o-rings that appear dried or cracked.
•
Inspect all external media lines. Replace if necessary.
•
Check the magnet plate for free play.
•
Check the evaporation mechanism for free play.
•
Inspect the evaporation cover seals.
•
Remove, inspect, and clean the needles. Place them in an ultrasonic bath for a few minutes, if necessary.
•
Verify the following limit switch/alarm operations: home position, top cover (open/ close), front door, and sample tray (lift empty sample row). See “Verifying Sample
Tray Control” on page 65.
•
Verify proper operation of all LEDs (4 total).
Open the front door (sample tray compartment) of the Varian 400-DS.
•
Lubricate reciprocation slide rail using a pre-approved lubricant. Using the
Diagnostics portion of the software (“Manual Control/Diagnostics” on page 64), move the sample tray from row to row to work in the lubricant.
•
Inspect the front door bumper.
•
Inspect waste line (front view).
Varian, Inc.
Page 92
Maintenance
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Power OFF the 400-DS and disconnect the power cord. Remove the back cover of the
400-DS to perform the following:
•
Lubricate the ball screw on syringe pump.
•
Lubricate all moving pieces of syringe pump.
•
Inspect the main drive belt.
•
Inspect/clean the waste row.
•
Inspect/clean the waste tubing.
•
Inspect all internal stainless steel components for signs of discoloration. Lightly wipe the surface with a soft cloth to remove, if necessary.
•
Lubricate the main rails.
•
Inspect the waste line (rear view).
•
Check lift blocks for free play.
Varian, Inc.
Chapter 6
Service and
Warranty
The warranty is provided by Varian, Inc. or one of its authorized representatives.
Service and Warranty Information
Varian dissolution products carry a one-year warranty on parts and labor. The Dissolution
Systems Service Department (or one of its representatives) will, at its option, either repair or replace any mechanical and electrical components in your instrument which prove to be defective. During the first year of warranty coverage, there is no charge for the labor to repair your unit. The Dissolution Systems Service Department (or one of its representatives) will determine the best site to repair the unit, either onsite or returned to
Varian, Inc. Any onsite warranty services are provided only at the initial installation point.
Installation and onsite warranty services are available only in Dissolution Systems service travel areas.
Varian, Inc.
Page 94
Service and Warranty
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
Exclusions and Limitations
Excluded from this warranty are expendable or consumable items such as, but not limited to, paddles, baskets, vessels, and acrylic water baths. Also excluded are defects from improper or inadequate maintenance by the customer, user-induced chemical action or contamination, unauthorized modification or misuse, and improper site preparation and maintenance.
Operation of software is not warranted to be uninterrupted or error-free.
Obtaining Warranty Service
To obtain warranty service in the United States, contact the Dissolution Systems Service
Department at 800.229.1108 to obtain authorization to return units for repair. At the option of the customer, onsite warranty service is available, but travel charges may be incurred.
The customer should prepay all shipping charges for products returned to the Dissolution
Systems Service Department (unless otherwise authorized), and Varian, Inc. will pay all charges for return to the customer.
Warranty Limitations
Varian, Inc. makes no other warranty, either express or implied, with respect to this product. Specifically disclaimed are any implied warranties of merchantability and fitness for a particular use. In no event will Varian, Inc. be liable for any indirect, incidental, or consequential damages arising from the use of this product. This warranty gives you specific legal rights which may vary from state to state or province to province, so you may have other rights and some of these exclusions may not apply to you.
Varian, Inc.
Revision A, 05/07
P/N 70-9051
Varian 400-DS Autoamted Apparatus 7
Operator’s Manual
Page 95
Service and Warranty
Exclusive Remedies
The remedies provided herein are the customer’s sole and exclusive remedies. In no event shall Varian, Inc. or its representatives be liable for any direct, indirect, special, incidental, or consequential damages, whether based on contract, tort, or any other legal theory. Some states or provinces do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you.
Varian, Inc.
Page 96
Service and Warranty
Varian 400-DS Automated Apparatus 7
Operator’s Manual
Revision A, 05/07
P/N 70-9051
This page was intentionally left blank, except for this message.
Varian, Inc.
Index
a
accessories
23 add new dissolution server screen
42 application, configuring
43 audit trail
49, 79 automated apparatus 7 workstation screen
48
b
browse dissolution servers screen
42
c
cable connections multiple system
29 single system
29 change management
72 change user
49 clean system
88 configuration dialog screen miscellaneous tab
45 security tab
43 configuring automated apparatus 7 workstation
50 configuring the application
43 conventions
20 copying a system configuration
54 copying methods
77 creating a method
70
d
database, attaching
33 deleting a system configuration
54 deleting methods
77 diagnostics
64 dissolution server selection
41 dissoservice manager screen
37 tools menu
38 dissoservice options screen
39
e
editing an existing method
78 editing an existing system configuration
56 electronic signatures
88
Varian, Inc.
Index
e-mail notification configuration
37 error tolerance
72 exclusions
94 exclusive remedies
95 export xml file
58, 80
h
hazards
11
i
import xml file
58, 80 initial power up
32 introduction
19
l
limitations
94 local security policy
35 local users and groups screen
44 lock application
49 logon
47 logon screen
41, 47
m
maintenance
89 manual control
64 method editor
69 method editor screen
71 method report
78 method start options screen
85 methods screen
70 minimum pc configuration
28 msde manager screen
34
o
obtaining warranty service
94
p
pc requirements, minimum
28 pc, setting up
28 periodic maintenance
89 permissions
49 physical parameters, usp
20 power up
32 profile interval
72
Varian, Inc.
r
reader comment form
101 report toolbar
88 run method
82
s
safety features
20 safety practices
11 sample information
72 screen options automated apparatus 7 workstation
49 method start options screen
85 system editor
51 security
49 select method step 2 of 2 screen
83 select system step 1 of 2 screen
82 selecting dissolution server
41 serial numbers
55 service
93 setting up your pc
28 show audit trail
57, 79 starting dissolution workstation
40 status
49 system configuration report
56 system configuration screen
50 system editing rights
52 system editor screen
51 system status screen
84
t
tab options method editor
Parameters tab 72 test report selection screen
87 test reports
87
u
user administration
47 usp physical parameters
20
v
verify integrity
58, 80
Index Index w
warranty
93 warranty limitations
94 windows security alert screen
40
Varian, Inc.
Index
This page was intentionally left blank, except for this message.
Varian, Inc.
Tell Us How We Are Doing
We listen to our customers. We work hard to make our technical documentation user friendly, and to make the information in our manuals easy to retrieve and use. We’d like you to tell us the kinds of additional information you’d find helpful in our documentation. Your feedback will be carefully considered when we prepare future editions of this manual.
This manual should contain the following additional information:
The most useful thing about this book is:
This manual would be more helpful if:
My general impressions of this book are:
May we contact you regarding your comments? ____ YES ____ NO (If yes, please write your name, address, and telephone number here.)
Please return this form via mail to: Technical Writing/Dissolution Systems, Varian, Inc.,
13000 Weston Parkway, Cary, North Carolina 27513-2250 USA. Optionally, you can return this form via fax at 1.919.677.1550. Always, feel free to telephone us to discuss your comments at 1.800.229.1108.
Varian 400-DS Automated Apparatus 7 Operator’s Manual P/N 70-9051 Revision A
Advertisement
Key features
- Automated sample collection
- Extended-release products
- Multiple pH levels
- Report generation
- PC control
- Up to 4 units
- HPLC vial collection