Portable Multi-logger

ZR-RX40 / ZR-RX20

O

P

E

N

Fund

Portable Multi-logger

ZR-RX

40

A-E

Standard

20 Channels

Insulated

Up to 200 isolated channels

Multifunction input (thermocouples, resistance thermometer, voltage etc.)

LAN connection Compatible with USB flash drive Internal 12 MB flash-memory

5.7 inch TFT color LCD Easy-to-navigate menus 9 hour battery (option)

M3 screw terminal blocks Standard PC software

amentals, Versatility and Usability!

Secure

Isolated Multi-Input

Multi-input is possible for temperature, humidity, voltage, logic and pulse. Each channel is supplied with its own isolated circuit giving zero inter-channel noise interference.

Maximum 200 Channel

Expandability

The terminal block, just connect and go, comes with 20 channels in standard configuration. A maximum of 10 units can be connected to each logger, easily creating a multi-channeled measurement system to be used according to the number of channels needed.

Simple

LAN Connection

The embeded Web server function enables remote control from a web browser. A USB connection means a greater array of measurements is possible. Standard PC software is also included .

Supper Light Mobile Model Also Available!

Portable Multi-Logger

ZR-RX

20

A-E

10 isolated channels

Multi-channel input (thermal resistance, voltage etc.)

Compatible with USB flash drive Internal 3.5MB flash memory

3.5 inch TFT color LCD Easy-to-navigate menus

6 hour battery (option) M3 screw terminal blocks

Standard PC software

10 insulated channels

Anyone- Anytime- Anyplace

Portable Multi-logger

ZR-RX

40

Input signal

Temperature Voltage

Logic Pulse

Humidity

List of features

Maximum

200 channels

All channels isolated

12MB internal memory

5.7 inch

TFT LCD

Thermocouples

K,J,E,T,R,S,B,N,W

Easy-tonavigate menus

Resistance thermometer

Pt100,JPt100,Pt1000

9 hour battery

LAN function

Display switch

USB flash drive

16bit resolution

Detachable terminal blocks

Expandable to 10 units (option)

* Optional parts needed when detaching or expanding.

2

Advanced Fundamentals

All channels isolated

Each circuit is electrically isolated so that the different channels do not affect each other. Consequently, there is no noise interference between the channels giving highly accurate and stable measurements.

Multi-function input

The temperature input supports thermocouples and resistance thermometer. Furthermore, in addition to temperature, an array of measurements including humidity and voltage can be taken at once. It also supports logic or pulse input, input from external triggers, as well as alarm signals output.

Expandable up to 200 channels

Maximum in its class

Each detachable terminal block has a standard 20 channels per unit. Up to 10 units can be connected making it possible to extend up to as much as 200 channels. It can also be used in car data logging where multi-measurements of temperature, voltage etc. on many channels is needed. Its seismic resistance corresponds to Class 1-A car parts(JIS standards)which means that it can be used stably even when there are vibrations and jolts during data-gathering.

Base unit cable

(Length 60 cm)

Maximum 10 unit connection

MAX

200CH

External input terminal

Logic alarm cable

Analog signal input terminal

*Switches between logic and pulse

Logic/ pulse input

Temperature Humidity Voltage

Alarm output

Trigger input

GND

Palm-top

Miniature, light-weight

480

g

990 g

*

Outstanding Versatility

Portable size - only 990 grams

Lightest in its class

Its light weight of 990 gram* makes handheld measurement possible. It is quite portable and can easily be slipped into your brief-case when setting out on a business trip.

*Including 1 terminal unit block. Not including battery or AC adapter.

Easy-to-see 5.7 inch color TFT LCD.

Comes with bright, easy-to-see, high-intensity

5.7 inch TFT large-scale color LCD panel. Its wide field of vision means it can even be seen at an angle.

USB flash drive and internal

12MB flash memory

Inside has a 12MB flash memory. It is also possible to connect it to commercially sold USB flash drive which means that the recording capacity can be selected as required according to application.

*Do not use USB flash drive with security functions.

Potential data recording time

(when 10 channels are used)

Recording interval (sampling speed) 100ms 200ms 500ms 1s 10s

12MB embedded flash memory 13hours 1day 3hours 2days 21hours

256MB USB flash drive 12days 25days 62days

5days 58days

125days 1256days

LAN port enabling remote control from web browser.

Equipped with LAN port for Ethernet connection. By connecting the ZR-RX 40 to a PC with a LAN cable, it is possible to set the logger up with your web browser. You can take measurements from your own desktop.

Remote control

LAN cable

(10BASE-T/100BASE-TX)

Portable Multi-logger

ZR-RX

20

*Not including battery,

AC adapter

Input signal

Temperature Voltage Humidity

Logic Pulse

List of features

Maximum

10 channels

All channels isolated

Thermocouples

K,J,E,T,R,S,B,N,W

USB flash drive

3.5MB

internal memory

3.5 inch

TFT LCD

Easy-tonavigate menus

6 hour battery

Display switch

16bit resolution

'Easy to operate' and 'easy to handle' means that data can be gathered at ease even from equipment jammed into narrow spaces!

Small but fully-fledged! multifunction input of 10 insulated channels

First in its class

Even with its small size, the 10 isolated channel input can provide accurate and stable measurements. The ability to input voltage, humidity, pulse and logic, as well as temperature, makes it practical in every field.

Pulse signal input terminal

Pulse

Logic signal input terminal

Logic

Analog signal input terminal

Temperature Humidity Voltage

3

Comfortable Usability

4

Step1

Connection

No need For Wiring

Detachable terminal blocks - no need for wiring to the main body

ZR-RX40 only

The ZR-RX40 has removable terminal blocks which means that it can be wired up securely in narrow places just by detaching the blocks. Also, if the terminal blocks are left in the location they have been wired into, there is no need to rewire them back to the main body when you want take repeat measurements.

* Optional parts needed

Connection with just one screwdriver

It uses M3 terminal blocks for analog signal input. The ring terminals and tongue terminals can be easily connected with just one screw driver.

There is no need for special tools so it can be used immediately.

Step3

Configuration- Measurement

No need for operation manuals

Operates just like a mobile phone so you use it just by following your instincts rather than by following the manual. Touching the NAVI button just once brings up a help screen so that you can operate it without the ponderous burden of browsing a manual.

ring terminal tongue terminal

Step4

Operation- Maintenance

No need for time-consuming work

Easy display and easy editing with PC software

Using the standard PC software provided, graphs can be displayed on a PC, or saved as Excel data.

Furthermore, using the optional special PC software, in addition to displaying graphs or saving Excel data, it is possible to monitor multiple measurement data divided into groups, so preparing data for reports is simple.

Step2

Installation

No need for AC Adapter

With optional batteries, there is no need for the AC adapter. When fully charged, the ZR-RX20 and ZR-

RX40 can run for 6 hours and 9 hours respectively. In the unlikely even that the batteries should cut-out, there is no need to worry as the preceding data is saved with the data back-up function.

ZR-RX20 ZR-RX40

No need for initial set-up with 3 way PC link

ZR-RX40 only

3 ways to link to a PC are available through a USB flash drive, a USB connection or a LAN connection. It is also possible to operate with wireless LAN with a commercially available wire LAN to wireless LAN adapter to support your application needs.

Runs for

6 hours

Runs for

9 hours

USB cable

LAN cable

USB flash drive

ZR-XRB 1 Battery pack (one) ZR-XRB 1 Battery pack (two)

Calibration service (charged option)

Calibrations and certification on request.

Freely compare and quickly decide!

Smart grouping & multiple display

3 step simple waveform observation

Step1

Smart search

Search selected channels in one operation!

After the search the channels have selected status.

Step2 Grouping

Grouping is simple by drag and drop into the created groups.

Step3

Multiple display

For each group a waveform screen can be displayed, so observation of comparisons between groups is possible.

Search all the

Voltage channels

Temperature

Display for each input signal

Compare multiple boards (target)

Voltage

Drag and drop into the

"IC Chip Voltage" group

Free movement, and extremely intuitive operation!

Free adjustment function

new concept

Click on wave form, scale, or cursor, and move

Easy to view recording setting screen patent pending

Direct input

Pull-down selection

Setting and editing can all be completed from this screen!

Move scale left or right

Move waveform up or down

Insert comment at an arbitrary position on the waveform

Change the position of the wave form with a single click of an icon

Display side by side Display superimposed

Extremely easy to use compared with conventional logger software

Wave Inspire RX

Common Options

ZR-SX10

* 30 days´ trial software is provided on the

utility disk in a standard set.

Using the new special PC software "Wave Inspire RX", in a simple 3 step operation it is possible to compare each measurement, or observe the waveforms for each input signal.

Observation of wave forms is very convenient when there is much measurement data, such as multi-channel measurements, multi-measurements, etc. Also, functions not available in standard software are provided, such as directly dragging waveforms, scales, and cursors with a mouse, so that extremely intuitive operability that far exceeds that of conventional logger software is achieved.

Major functions

Quick call up of the cursor

Special search function

Screen save function/CSV save function

Report printing function

5

System Configuration

Standard set

ZR-RX20A-E

Options

Common options

Special PC Software

ZR-SX10

DC cable(2m)

ZR-XRD1

Battery Pack

ZR-XRB1

Humidity Sensor(3m)

ZR-XRH1

AC adaptor/AC cable

Start-up guide

Utility disk*

*Content of utility disk

– Special PC software(Trial version)

– Standard PC software

– manual PDF file

ZR-RX40A-E

*USB flash drive can

be used usually.

ZR-RX40 Special options

Base Set

ZR-XRE1

Extension Terminal Set

ZR-XRT1

Utility disk*

*Content of utility disk

Special PC software(Trial version)

Standard PC software

– manual PDF file

Logic Alarm Cable

ZR-XRL1

AC adaptor/AC cable

Start-up guide

ZR-RX40 Necessary option when channel is enhanced

Item

ZR-RX40 Main unit

Base Set(ZR-XRE1)

Extension Terminal Set(ZR-XRT1)

20CH

1

0

0

40CH

1

1

1

100CH

1

1

4

200CH

1

1

9

Standard Specifications

Main unit

Model

Input method

Input terminal shape

Number of input channels

Sampling interval

A/D resolution

Measurement ranges

Voltage

Thermocouples

Resistance temperature detector

Humidity

Logic

Maximum input voltage

Pulse

Reference contact compensation

Reference contact compensation accuracy

Channel Terminals

Logic input/Pulse input/Trigger input

Input impedance

Allowable signal source resistance

Alarm output Number of channels

Output format

Output conditions

Trigger input

PC I/F Type

PC I/F Functions

Data save functions

Measured data

Other

Display

Display Language

Operating environment

Withstand voltage

Power supply/Voltage

Power consumption

Weight

ZR-RX20A

all channels isolated

10CH

Max 100ms

Uncorrespondence

0 to100%(Voltage 0V to 1V Scaling conversion)*ZR-XRH1(Refer to the option.)

1CH

1CH(Revolutions mode, Counts mode, Inst. Mode switch method)

Switch between logic4CH and pulse4CH pulse type(Revolutions mode, Counts mode, Inst. Mode switch method)

Internal / External switching

±0.5 ˚C(23˚C±2 ˚C, when the input terminal temperature is balanced)

60Vp-p

1CH

ZR-RX40A

all channels isolated

Resistance temperature detector is non-isolated

M3 screw type terminal

Standard: 20ch Max: 200ch(When the terminal unit is connected)

100ms/10CH max

16bit

20/50/100/200/500mV,1/2/5/10/20/50V,1-5V F.S.

K/J/E/T/R/S/B/N/W(WRe5-26)

24V

Pt100/JPt100/Pt1000(IEC751)

Open collector output

Level judgment, window judgment, logic pattern judgment, pulse judgment

USB(1.1)

Data transfer to the PC(realtime memory)

1M

Ω±5%

Less than 300

1CH

4CH

USB(2.0),LAN(10BASE-T/100BASE-TX)

Data transfer to the PC(realtime memory)

Web server function(Operation of ZR-RX40)

FTP function(Transfers and deletes measured data from internal memory via network)

Internal memory(3.5MB Flash memory) or direct capture in the USB memory device

Internal memory(12MB Flash memry) or direct capture in the USB memory device

Setup conditions, copy of data screen can be saved in the internal memory or USB memory device

3.5-inch TFT color LCD(QVGA: 320 x 240 dots) 5.7-inch TFT color LCD(QVGA: 320 x 240 dots)

Japanese/English(US)/English(UK)/French/German/Chinese/Korean

0 to 40˚C, 30 to 80 % RH(15 to 40˚C when the battery) 0 to 45˚C, 5 to 85 % RH(15 to 40˚C when the battery)

350 Vp-p(between input channel/GND; between each chs)1 minute

AC adapter: 100 to 240 VAC/50 to 60 Hz(*9)DC input: 8.5 to 24VDC Battery pack(opitin)

28 VA or less(when the AC adapter is used)

480g(Excluding the AC adapter and battery) 990g(Including 1 terminal units,Excluding the AC adapter and battery)

Measurement accuracy

(ZR-RX20,ZR-RX40 Commonness)

Voltage

±0.1% of F.S.

Thermocouple Measurement Temperature Range (˚C)

R/S

0

≤TS ≤100

B

100<TS

≤300

R : 300<TS

≤1600˚C

S : 300<TS

≤1760˚C

400

≤TS ≤600

K

E

600<TS

≤1820˚C

–200

≤TS ≤–100

–100<TS

≤1370˚C

–200

≤TS ≤–100

T

J

–100<TS

≤800˚C

–200

≤TS ≤–100

–100<TS

≤400˚C

–200

≤TS ≤–100

–100<TS

≤100

N

W

100<TS

≤1100˚C

0

≤TS ≤1300˚C

0

≤TS≤2315˚C

Measureme

±5.2˚C

±3.0˚C

±(0.05%of rdg+2.0˚C)

±(0.05%of rdg+2.0˚C)

±3.5˚C

±(0.05%of rdg+2.0˚C)

±(0.05%of rdg+2.0˚C)

±(0.05%of rdg+1.0˚C)

±(0.05%of rdg+2.0˚C)

±(0.05%of rdg+1.0˚C)

±(0.1%of rdg+1.5˚C)

±(0.1%of rdg+0.5˚C)

±2.7˚C

±1.7˚C

±(0.05%of rdg+1.0˚C)

±(0.1%of rdg+1.0˚C)

±(0.1%of rdg+1.5˚C)

Resistance

Temperature

Detector

Pt100

JPt100

Pt1000

Measurement Temperature

Range(˚C)

–200 to 850˚C(FS=1050˚C)

–200 to 500˚C(FS=700˚C)

–200 to 500˚C(FS=700˚C)

Applied

Current

1mA

1mA

0.2mA

Features under the following measurement parameters

Operating environment 23 ˚C

±3 ˚C

Left for at least 30 minutes after the power supply is turned on

Sampling interval 1 s(20 ch)

Filter ON(Average: 10 times)

GND connection

Thermocouple used is T: 0.32f, other: 0.65f

Only ZR-RX40 corresponds to the resistance temperature detector.

Measurement

Accuracy

±1.0˚C

±0.8˚C

±0.8˚C

Humidity Sensor ZR-XRH1

(Option)

Item

Ambient temperature

Ambient humidity

Relative humidity measurement accuracy

Response time

Sensor output

External dimensions

Cable length

Description

–25 to +80˚C

0 to 100%RH

±3%RH(5 to 98%RH at 25˚C)

15 sec(90 % response when membrane filter is installed)

DCO to 1V

14 x 80 mm(excluding cable)

3m

PC Software

Special PC Software ZR-SX10

Wave Inspire RX

(OPTION; Sold separately,Trial version enclosed in the ZR-RX standard set.)

Standard PC Software

Smart Viewer RX20/RX40

(Enclosed in the ZR-RX standard set.)

Standard

Wave

Operation

Logging-data display,Real-time PC data logging,

ZR-RX setting,CSV file conversion

Direct drug & move

Scale batch change of all channels

Other Intuitive functions including mouse wheel operation

Scale change by numeric input.

(RX20)

Scale change of each channel by icon.

(RX40)

Wave

Display

Setting

Display

CSV

Conversion

Multiwindow display

All channel multi-scale display

Scroll(UP/Down/left/right)

"Smart List view"

"Smart grouping"

Single window division display.

Y-T, X-Y, and meter display selectable.

Tab form

Between cursors,All data,File batch conversions

Other functions

"Smart Search"

Cursor View,

Comment input,

I/F for Real-time logging

USB

OS

Windows Vista / XP / 2000

Cursor View,

Comment input,

Excel transfer,Alarm mail.

USB,LAN

Windows XP/2000

Model

Standard set

Item

Portable Multi Logger

Portable Multi Logger

Model

ZR-RX20A-E

ZR-RX40A-E

Option

Item

Battery Pack

Humidity Sensor

DC Cable

Base Set(ZR-RX40 Special option goods)

Extension Terminal Set(ZR-RX40 Special option goods)

Logic Alarm Cable(ZR-RX40 Special option goods)

Special PC Software Wave Inspire RX(ZR-RX20/40 commonness)

The temperature detector also abundantly does the lineup.

Model

ZR-XRB1

ZR-XRH1

ZR-XRD1

ZR-XRE1

ZR-XRT1

ZR-XRL1

ZR-SX10

External Dimensions

ZR-RX20 ZR-RX40

(Unit: mm)

121.5

1.5

38

1.5

194

32

120

48

2

232

185

ZR-RX40 Main body installation screw hole

2-M4 Depth 6

23.5

60

Portable Multi-logger

ZR-RX20/RX40

This document provides information mainly for selecting suitable models. Please read the document User's Manual (Z257/Z260) carefully for information that the user must understand and accept before purchase, including information on warranty, limitations of liability, and precautions.

OMRON Corporation

I ndustrial Automation Company

Sensing Devices Division H.Q.

Application Sensors Division

Shiokoji Horikawa, Shimogyo-ku,

Kyoto, 600-8530 Japan

Tel: (81) 75-344-7068/Fax: (81) 75-344-7107

Regional Headquarters

OMRON EUROPE B.V.

Sensor Business Unit

Carl-Benz-Str. 4, D-71154 Nufringen,

Germany

Tel: (49) 7032-811-0/Fax: (49) 7032-811-199

OMRON ELECTRONICS LLC

One Commerce Drive Schaumburg,

IL 60173-5302 U.S.A.

Tel: (1) 847-843-7900/Fax: (1) 847-843-7787

OMRON ASIA PACIFIC PTE. LTD.

No. 438A Alexandra Road # 05-05/08 (Lobby 2),

Alexandra Technopark, Singapore 119967

Tel: (65) 6835-3011/Fax: (65) 6835-2711

OMRON (CHINA) CO., LTD.

Room 2211, Bank of China Tower,

200 Yin Cheng Zhong Road,

Pu Dong New Area, Shanghai, 200120, China

Tel: (86) 21-5037-2222/Fax: (86) 21-5037-2200

Authorized Distributor:

Note: specifications subject to change without notice.

Cat. No E387-E1-01A

Printed in Japan

1107(0807)(KW)

Read and Understand This Catalog

Please read and understand this catalog before purchasing the products. Please consult your OMRON representative if you have any questions or comments.

Warranty and Limitations of Liability

WARRANTY

OMRON's exclusive warranty is that the products are free from defects in materials and workmanship for a period of one year (or other period if specifi ed) from date of sale by OMRON.

OMRON MAKES NO WARRANTY OR REPRESENTATION, EXPRESS OR IMPLIED, REGARDING NON-INFRINGEMENT, MERCHANTABILITY, OR

FITNESS FOR PARTICULAR PURPOSE OF THE PRODUCTS. ANY BUYER OR USER ACKNOWLEDGES THAT THE BUYER OR USER ALONE

HAS DETERMINED THAT THE PRODUCTS WILL SUITABLY MEET THE REQUIREMENTS OF THEIR INTENDED USE. OMRON DISCLAIMS ALL

OTHER WARRANTIES, EXPRESS OR IMPLIED.

LIMITATIONS OF LIABILITY

OMRON SHALL NOT BE RESPONSIBLE FOR SPECIAL, INDIRECT, OR CONSEQUENTIAL DAMAGES, LOSS OF PROFITS, OR COMMERCIAL

LOSS IN ANY WAY CONNECTED WITH THE PRODUCTS, WHETHER SUCH CLAIM IS BASED ON CONTRACT, WARRANTY, NEGLIGENCE, OR

STRICT LIABILITY.

In no event shall responsibility of OMRON for any act exceed the individual price of the product on which liability is asserted.

IN NO EVENT SHALL OMRON BE RESPONSIBLE FOR WARRANTY, REPAIR, OR OTHER CLAIMS REGARDING THE PRODUCTS UNLESS

OMRON'S ANALYSIS CONFIRMS THAT THE PRODUCTS WERE PROPERLY HANDLED, STORED, INSTALLED, AND MAINTAINED AND NOT

SUBJECT TO CONTAMINATION, ABUSE, MISUSE, OR INAPPROPRIATE MODIFICATION OR REPAIR.

Application Considerations

SUITABILITY FOR USE

OMRON shall not be responsible for conformity with any standards, codes, or regulations that apply to the combination of products in the customer's application or use of the product.

At the customer's request, OMRON will provide applicable third party certifi cation documents identifying ratings and limitations of use that apply to the products. This information by itself is not suffi cient for a complete determination of the suitability of the products in combination with the end product, machine, system, or other application or use.

The following are some examples of applications for which particular attention must be given. This is not intended to be an exhaustive list of all possible uses of the products, nor is it intended to imply that the uses listed may be suitable for the products:

• Outdoor use, uses involving potential chemical contamination or electrical interference, or conditions or uses not described in this catalog.

• Nuclear energy control systems, combustion systems, railroad systems, aviation systems, medical equipment, amusement machines, vehicles, safety equipment, and installations subject to separate industry or government regulations.

• Systems, machines, and equipment that could present a risk to life or property.

Please know and observe all prohibitions of use applicable to the products.

NEVER USE THE PRODUCTS FOR AN APPLICATION INVOLVING SERIOUS RISK TO LIFE OR PROPERTY WITHOUT ENSURING THAT THE

SYSTEM AS A WHOLE HAS BEEN DESIGNED TO ADDRESS THE RISKS, AND THAT THE OMRON PRODUCT IS PROPERLY RATED AND

INSTALLED FOR THE INTENDED USE WITHIN THE OVERALL EQUIPMENT OR SYSTEM.

Disclaimers

CHANGE IN SPECIFICATIONS

Product specifi cations and accessories may be changed at any time based on improvements and other reasons.

It is our practice to change model numbers when published ratings or features are changed, or when signifi cant construction changes are made.

However, some specifi cations of the product may be changed without any notice. When in doubt, special model numbers may be assigned to fi x or establish key specifi cations for your application on your request. Please consult with your OMRON representative at any time to confi rm actual specifi cations of purchased product.

DIMENSIONS AND WEIGHTS

Dimensions and weights are nominal and are not to be used for manufacturing purposes, even when tolerances are shown.

ERRORS AND OMISSIONS

The information in this catalog has been carefully checked and is believed to be accurate; however, no responsibility is assumed for clerical, typographical, or proofreading errors, or omissions.

PERFORMANCE DATA

Performance data given in this catalog is provided as a guide for the user in determining suitability and does not constitute a warranty. It may represent the result of OMRON’s test conditions, and the users must correlate it to actual application requirements. Actual performance is subject to the OMRON

Warranty and Limitations of Liability.

PROGRAMMABLE PRODUCTS

OMRON shall not be responsible for the user's programming of a programmable product, or any consequence thereof.

COPYRIGHT AND COPY PERMISSION

This catalog shall not be copied for sales or promotions without permission.

This catalog is protected by copyright and is intended solely for use in conjunction with the product. Please notify us before copying or reproducing this catalog in any manner, for any other purpose. If copying or transmitting this catalog to another, please copy or transmit it in its entirety.

Cat. No. E387-E1-01A

OMRON Corporation

Industrial Automation Company http://www.ia.omron.com/

2007.11

In the interest of product improvement, specifications are subject to change without notice.

(c)Copyright OMRON Corporation 2007 All Rights Reserved.

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement