advertisement
N E W
Portable Multi-logger
ZR-RX40 / ZR-RX20
O
P
E
N
Fund
Portable Multi-logger
ZR-RX
40
A-E
Standard
20 Channels
Insulated
Up to 200 isolated channels
Multifunction input (thermocouples, resistance thermometer, voltage etc.)
LAN connection Compatible with USB flash drive Internal 12 MB flash-memory
5.7 inch TFT color LCD Easy-to-navigate menus 9 hour battery (option)
M3 screw terminal blocks Standard PC software
amentals, Versatility and Usability!
Secure
Isolated Multi-Input
Multi-input is possible for temperature, humidity, voltage, logic and pulse. Each channel is supplied with its own isolated circuit giving zero inter-channel noise interference.
Maximum 200 Channel
Expandability
The terminal block, just connect and go, comes with 20 channels in standard configuration. A maximum of 10 units can be connected to each logger, easily creating a multi-channeled measurement system to be used according to the number of channels needed.
Simple
LAN Connection
The embeded Web server function enables remote control from a web browser. A USB connection means a greater array of measurements is possible. Standard PC software is also included .
Supper Light Mobile Model Also Available!
Portable Multi-Logger
ZR-RX
20
A-E
10 isolated channels
Multi-channel input (thermal resistance, voltage etc.)
Compatible with USB flash drive Internal 3.5MB flash memory
3.5 inch TFT color LCD Easy-to-navigate menus
6 hour battery (option) M3 screw terminal blocks
Standard PC software
10 insulated channels
Anyone- Anytime- Anyplace
Portable Multi-logger
ZR-RX
40
Input signal
Temperature Voltage
Logic Pulse
Humidity
List of features
Maximum
200 channels
All channels isolated
12MB internal memory
5.7 inch
TFT LCD
Thermocouples
K,J,E,T,R,S,B,N,W
Easy-tonavigate menus
Resistance thermometer
Pt100,JPt100,Pt1000
9 hour battery
LAN function
Display switch
USB flash drive
16bit resolution
Detachable terminal blocks
Expandable to 10 units (option)
* Optional parts needed when detaching or expanding.
2
Advanced Fundamentals
All channels isolated
Each circuit is electrically isolated so that the different channels do not affect each other. Consequently, there is no noise interference between the channels giving highly accurate and stable measurements.
Multi-function input
The temperature input supports thermocouples and resistance thermometer. Furthermore, in addition to temperature, an array of measurements including humidity and voltage can be taken at once. It also supports logic or pulse input, input from external triggers, as well as alarm signals output.
Expandable up to 200 channels
Maximum in its class
Each detachable terminal block has a standard 20 channels per unit. Up to 10 units can be connected making it possible to extend up to as much as 200 channels. It can also be used in car data logging where multi-measurements of temperature, voltage etc. on many channels is needed. Its seismic resistance corresponds to Class 1-A car parts(JIS standards)which means that it can be used stably even when there are vibrations and jolts during data-gathering.
Base unit cable
(Length 60 cm)
Maximum 10 unit connection
MAX
200CH
External input terminal
Logic alarm cable
Analog signal input terminal
*Switches between logic and pulse
Logic/ pulse input
Temperature Humidity Voltage
Alarm output
Trigger input
GND
Palm-top
Miniature, light-weight
480
g
990 g
*
Outstanding Versatility
Portable size - only 990 grams
Lightest in its class
Its light weight of 990 gram* makes handheld measurement possible. It is quite portable and can easily be slipped into your brief-case when setting out on a business trip.
*Including 1 terminal unit block. Not including battery or AC adapter.
Easy-to-see 5.7 inch color TFT LCD.
Comes with bright, easy-to-see, high-intensity
5.7 inch TFT large-scale color LCD panel. Its wide field of vision means it can even be seen at an angle.
USB flash drive and internal
12MB flash memory
Inside has a 12MB flash memory. It is also possible to connect it to commercially sold USB flash drive which means that the recording capacity can be selected as required according to application.
*Do not use USB flash drive with security functions.
Potential data recording time
(when 10 channels are used)
Recording interval (sampling speed) 100ms 200ms 500ms 1s 10s
12MB embedded flash memory 13hours 1day 3hours 2days 21hours
256MB USB flash drive 12days 25days 62days
5days 58days
125days 1256days
LAN port enabling remote control from web browser.
Equipped with LAN port for Ethernet connection. By connecting the ZR-RX 40 to a PC with a LAN cable, it is possible to set the logger up with your web browser. You can take measurements from your own desktop.
Remote control
LAN cable
(10BASE-T/100BASE-TX)
Portable Multi-logger
ZR-RX
20
*Not including battery,
AC adapter
Input signal
Temperature Voltage Humidity
Logic Pulse
List of features
Maximum
10 channels
All channels isolated
Thermocouples
K,J,E,T,R,S,B,N,W
USB flash drive
3.5MB
internal memory
3.5 inch
TFT LCD
Easy-tonavigate menus
6 hour battery
Display switch
16bit resolution
'Easy to operate' and 'easy to handle' means that data can be gathered at ease even from equipment jammed into narrow spaces!
Small but fully-fledged! multifunction input of 10 insulated channels
First in its class
Even with its small size, the 10 isolated channel input can provide accurate and stable measurements. The ability to input voltage, humidity, pulse and logic, as well as temperature, makes it practical in every field.
Pulse signal input terminal
Pulse
Logic signal input terminal
Logic
Analog signal input terminal
Temperature Humidity Voltage
3
Comfortable Usability
4
Step1
Connection
No need For Wiring
Detachable terminal blocks - no need for wiring to the main body
ZR-RX40 only
The ZR-RX40 has removable terminal blocks which means that it can be wired up securely in narrow places just by detaching the blocks. Also, if the terminal blocks are left in the location they have been wired into, there is no need to rewire them back to the main body when you want take repeat measurements.
* Optional parts needed
Connection with just one screwdriver
It uses M3 terminal blocks for analog signal input. The ring terminals and tongue terminals can be easily connected with just one screw driver.
There is no need for special tools so it can be used immediately.
Step3
Configuration- Measurement
No need for operation manuals
Operates just like a mobile phone so you use it just by following your instincts rather than by following the manual. Touching the NAVI button just once brings up a help screen so that you can operate it without the ponderous burden of browsing a manual.
ring terminal tongue terminal
Step4
Operation- Maintenance
No need for time-consuming work
Easy display and easy editing with PC software
Using the standard PC software provided, graphs can be displayed on a PC, or saved as Excel data.
Furthermore, using the optional special PC software, in addition to displaying graphs or saving Excel data, it is possible to monitor multiple measurement data divided into groups, so preparing data for reports is simple.
Step2
Installation
No need for AC Adapter
With optional batteries, there is no need for the AC adapter. When fully charged, the ZR-RX20 and ZR-
RX40 can run for 6 hours and 9 hours respectively. In the unlikely even that the batteries should cut-out, there is no need to worry as the preceding data is saved with the data back-up function.
ZR-RX20 ZR-RX40
No need for initial set-up with 3 way PC link
ZR-RX40 only
3 ways to link to a PC are available through a USB flash drive, a USB connection or a LAN connection. It is also possible to operate with wireless LAN with a commercially available wire LAN to wireless LAN adapter to support your application needs.
Runs for
6 hours
Runs for
9 hours
USB cable
LAN cable
USB flash drive
ZR-XRB 1 Battery pack (one) ZR-XRB 1 Battery pack (two)
Calibration service (charged option)
Calibrations and certification on request.
Freely compare and quickly decide!
Smart grouping & multiple display
3 step simple waveform observation
Step1
Smart search
Search selected channels in one operation!
After the search the channels have selected status.
Step2 Grouping
Grouping is simple by drag and drop into the created groups.
Step3
Multiple display
For each group a waveform screen can be displayed, so observation of comparisons between groups is possible.
Search all the
Voltage channels
Temperature
Display for each input signal
Compare multiple boards (target)
Voltage
Drag and drop into the
"IC Chip Voltage" group
Free movement, and extremely intuitive operation!
Free adjustment function
new concept
Click on wave form, scale, or cursor, and move
Easy to view recording setting screen patent pending
Direct input
Pull-down selection
Setting and editing can all be completed from this screen!
Move scale left or right
Move waveform up or down
Insert comment at an arbitrary position on the waveform
Change the position of the wave form with a single click of an icon
Display side by side Display superimposed
Extremely easy to use compared with conventional logger software
Wave Inspire RX
Common Options
ZR-SX10
* 30 days´ trial software is provided on the
utility disk in a standard set.
Using the new special PC software "Wave Inspire RX", in a simple 3 step operation it is possible to compare each measurement, or observe the waveforms for each input signal.
Observation of wave forms is very convenient when there is much measurement data, such as multi-channel measurements, multi-measurements, etc. Also, functions not available in standard software are provided, such as directly dragging waveforms, scales, and cursors with a mouse, so that extremely intuitive operability that far exceeds that of conventional logger software is achieved.
Major functions
Quick call up of the cursor
Special search function
Screen save function/CSV save function
Report printing function
5
System Configuration
Standard set
ZR-RX20A-E
Options
Common options
Special PC Software
ZR-SX10
DC cable(2m)
ZR-XRD1
Battery Pack
ZR-XRB1
Humidity Sensor(3m)
ZR-XRH1
AC adaptor/AC cable
Start-up guide
Utility disk*
*Content of utility disk
– Special PC software(Trial version)
– Standard PC software
– manual PDF file
ZR-RX40A-E
*USB flash drive can
be used usually.
ZR-RX40 Special options
Base Set
ZR-XRE1
Extension Terminal Set
ZR-XRT1
Utility disk*
*Content of utility disk
–
Special PC software(Trial version)
–
Standard PC software
– manual PDF file
Logic Alarm Cable
ZR-XRL1
AC adaptor/AC cable
Start-up guide
ZR-RX40 Necessary option when channel is enhanced
Item
ZR-RX40 Main unit
Base Set(ZR-XRE1)
Extension Terminal Set(ZR-XRT1)
20CH
1
0
0
40CH
1
1
1
100CH
1
1
4
200CH
1
1
9
Standard Specifications
Main unit
Model
Input method
Input terminal shape
Number of input channels
Sampling interval
A/D resolution
Measurement ranges
Voltage
Thermocouples
Resistance temperature detector
Humidity
Logic
Maximum input voltage
Pulse
Reference contact compensation
Reference contact compensation accuracy
Channel Terminals
Logic input/Pulse input/Trigger input
Input impedance
Allowable signal source resistance
Alarm output Number of channels
Output format
Output conditions
Trigger input
PC I/F Type
PC I/F Functions
Data save functions
Measured data
Other
Display
Display Language
Operating environment
Withstand voltage
Power supply/Voltage
Power consumption
Weight
ZR-RX20A
all channels isolated
10CH
Max 100ms
Uncorrespondence
0 to100%(Voltage 0V to 1V Scaling conversion)*ZR-XRH1(Refer to the option.)
1CH
1CH(Revolutions mode, Counts mode, Inst. Mode switch method)
Switch between logic4CH and pulse4CH pulse type(Revolutions mode, Counts mode, Inst. Mode switch method)
Internal / External switching
±0.5 ˚C(23˚C±2 ˚C, when the input terminal temperature is balanced)
60Vp-p
1CH
ZR-RX40A
all channels isolated
Resistance temperature detector is non-isolated
M3 screw type terminal
Standard: 20ch Max: 200ch(When the terminal unit is connected)
100ms/10CH max
16bit
20/50/100/200/500mV,1/2/5/10/20/50V,1-5V F.S.
K/J/E/T/R/S/B/N/W(WRe5-26)
24V
Pt100/JPt100/Pt1000(IEC751)
Open collector output
Level judgment, window judgment, logic pattern judgment, pulse judgment
USB(1.1)
Data transfer to the PC(realtime memory)
1M
Ω±5%
Less than 300
Ω
1CH
4CH
USB(2.0),LAN(10BASE-T/100BASE-TX)
Data transfer to the PC(realtime memory)
Web server function(Operation of ZR-RX40)
FTP function(Transfers and deletes measured data from internal memory via network)
Internal memory(3.5MB Flash memory) or direct capture in the USB memory device
Internal memory(12MB Flash memry) or direct capture in the USB memory device
Setup conditions, copy of data screen can be saved in the internal memory or USB memory device
3.5-inch TFT color LCD(QVGA: 320 x 240 dots) 5.7-inch TFT color LCD(QVGA: 320 x 240 dots)
Japanese/English(US)/English(UK)/French/German/Chinese/Korean
0 to 40˚C, 30 to 80 % RH(15 to 40˚C when the battery) 0 to 45˚C, 5 to 85 % RH(15 to 40˚C when the battery)
350 Vp-p(between input channel/GND; between each chs)1 minute
AC adapter: 100 to 240 VAC/50 to 60 Hz(*9)DC input: 8.5 to 24VDC Battery pack(opitin)
28 VA or less(when the AC adapter is used)
480g(Excluding the AC adapter and battery) 990g(Including 1 terminal units,Excluding the AC adapter and battery)
Measurement accuracy
(ZR-RX20,ZR-RX40 Commonness)
Voltage
±0.1% of F.S.
Thermocouple Measurement Temperature Range (˚C)
R/S
0
≤TS ≤100
B
100<TS
≤300
R : 300<TS
≤1600˚C
S : 300<TS
≤1760˚C
400
≤TS ≤600
K
E
600<TS
≤1820˚C
–200
≤TS ≤–100
–100<TS
≤1370˚C
–200
≤TS ≤–100
T
J
–100<TS
≤800˚C
–200
≤TS ≤–100
–100<TS
≤400˚C
–200
≤TS ≤–100
–100<TS
≤100
N
W
100<TS
≤1100˚C
0
≤TS ≤1300˚C
0
≤TS≤2315˚C
Measureme
±5.2˚C
±3.0˚C
±(0.05%of rdg+2.0˚C)
±(0.05%of rdg+2.0˚C)
±3.5˚C
±(0.05%of rdg+2.0˚C)
±(0.05%of rdg+2.0˚C)
±(0.05%of rdg+1.0˚C)
±(0.05%of rdg+2.0˚C)
±(0.05%of rdg+1.0˚C)
±(0.1%of rdg+1.5˚C)
±(0.1%of rdg+0.5˚C)
±2.7˚C
±1.7˚C
±(0.05%of rdg+1.0˚C)
±(0.1%of rdg+1.0˚C)
±(0.1%of rdg+1.5˚C)
Resistance
Temperature
Detector
Pt100
JPt100
Pt1000
Measurement Temperature
Range(˚C)
–200 to 850˚C(FS=1050˚C)
–200 to 500˚C(FS=700˚C)
–200 to 500˚C(FS=700˚C)
Applied
Current
1mA
1mA
0.2mA
Features under the following measurement parameters
Operating environment 23 ˚C
±3 ˚C
Left for at least 30 minutes after the power supply is turned on
Sampling interval 1 s(20 ch)
Filter ON(Average: 10 times)
GND connection
Thermocouple used is T: 0.32f, other: 0.65f
Only ZR-RX40 corresponds to the resistance temperature detector.
Measurement
Accuracy
±1.0˚C
±0.8˚C
±0.8˚C
Humidity Sensor ZR-XRH1
(Option)
Item
Ambient temperature
Ambient humidity
Relative humidity measurement accuracy
Response time
Sensor output
External dimensions
Cable length
Description
–25 to +80˚C
0 to 100%RH
±3%RH(5 to 98%RH at 25˚C)
15 sec(90 % response when membrane filter is installed)
DCO to 1V
14 x 80 mm(excluding cable)
3m
PC Software
Special PC Software ZR-SX10
Wave Inspire RX
(OPTION; Sold separately,Trial version enclosed in the ZR-RX standard set.)
Standard PC Software
Smart Viewer RX20/RX40
(Enclosed in the ZR-RX standard set.)
Standard
Wave
Operation
Logging-data display,Real-time PC data logging,
ZR-RX setting,CSV file conversion
Direct drug & move
Scale batch change of all channels
Other Intuitive functions including mouse wheel operation
Scale change by numeric input.
(RX20)
Scale change of each channel by icon.
(RX40)
Wave
Display
Setting
Display
CSV
Conversion
Multiwindow display
All channel multi-scale display
Scroll(UP/Down/left/right)
"Smart List view"
"Smart grouping"
Single window division display.
Y-T, X-Y, and meter display selectable.
Tab form
Between cursors,All data,File batch conversions
Other functions
"Smart Search"
Cursor View,
Comment input,
I/F for Real-time logging
USB
OS
Windows Vista / XP / 2000
Cursor View,
Comment input,
Excel transfer,Alarm mail.
USB,LAN
Windows XP/2000
Model
Standard set
Item
Portable Multi Logger
Portable Multi Logger
Model
ZR-RX20A-E
ZR-RX40A-E
Option
Item
Battery Pack
Humidity Sensor
DC Cable
Base Set(ZR-RX40 Special option goods)
Extension Terminal Set(ZR-RX40 Special option goods)
Logic Alarm Cable(ZR-RX40 Special option goods)
Special PC Software Wave Inspire RX(ZR-RX20/40 commonness)
The temperature detector also abundantly does the lineup.
Model
ZR-XRB1
ZR-XRH1
ZR-XRD1
ZR-XRE1
ZR-XRT1
ZR-XRL1
ZR-SX10
External Dimensions
ZR-RX20 ZR-RX40
(Unit: mm)
121.5
1.5
38
1.5
194
32
120
48
2
232
185
ZR-RX40 Main body installation screw hole
2-M4 Depth 6
23.5
60
Portable Multi-logger
ZR-RX20/RX40
This document provides information mainly for selecting suitable models. Please read the document User's Manual (Z257/Z260) carefully for information that the user must understand and accept before purchase, including information on warranty, limitations of liability, and precautions.
OMRON Corporation
I ndustrial Automation Company
Sensing Devices Division H.Q.
Application Sensors Division
Shiokoji Horikawa, Shimogyo-ku,
Kyoto, 600-8530 Japan
Tel: (81) 75-344-7068/Fax: (81) 75-344-7107
Regional Headquarters
OMRON EUROPE B.V.
Sensor Business Unit
Carl-Benz-Str. 4, D-71154 Nufringen,
Germany
Tel: (49) 7032-811-0/Fax: (49) 7032-811-199
OMRON ELECTRONICS LLC
One Commerce Drive Schaumburg,
IL 60173-5302 U.S.A.
Tel: (1) 847-843-7900/Fax: (1) 847-843-7787
OMRON ASIA PACIFIC PTE. LTD.
No. 438A Alexandra Road # 05-05/08 (Lobby 2),
Alexandra Technopark, Singapore 119967
Tel: (65) 6835-3011/Fax: (65) 6835-2711
OMRON (CHINA) CO., LTD.
Room 2211, Bank of China Tower,
200 Yin Cheng Zhong Road,
Pu Dong New Area, Shanghai, 200120, China
Tel: (86) 21-5037-2222/Fax: (86) 21-5037-2200
Authorized Distributor:
Note: specifications subject to change without notice.
Cat. No E387-E1-01A
Printed in Japan
1107(0807)(KW)
Read and Understand This Catalog
Please read and understand this catalog before purchasing the products. Please consult your OMRON representative if you have any questions or comments.
Warranty and Limitations of Liability
WARRANTY
OMRON's exclusive warranty is that the products are free from defects in materials and workmanship for a period of one year (or other period if specifi ed) from date of sale by OMRON.
OMRON MAKES NO WARRANTY OR REPRESENTATION, EXPRESS OR IMPLIED, REGARDING NON-INFRINGEMENT, MERCHANTABILITY, OR
FITNESS FOR PARTICULAR PURPOSE OF THE PRODUCTS. ANY BUYER OR USER ACKNOWLEDGES THAT THE BUYER OR USER ALONE
HAS DETERMINED THAT THE PRODUCTS WILL SUITABLY MEET THE REQUIREMENTS OF THEIR INTENDED USE. OMRON DISCLAIMS ALL
OTHER WARRANTIES, EXPRESS OR IMPLIED.
LIMITATIONS OF LIABILITY
OMRON SHALL NOT BE RESPONSIBLE FOR SPECIAL, INDIRECT, OR CONSEQUENTIAL DAMAGES, LOSS OF PROFITS, OR COMMERCIAL
LOSS IN ANY WAY CONNECTED WITH THE PRODUCTS, WHETHER SUCH CLAIM IS BASED ON CONTRACT, WARRANTY, NEGLIGENCE, OR
STRICT LIABILITY.
In no event shall responsibility of OMRON for any act exceed the individual price of the product on which liability is asserted.
IN NO EVENT SHALL OMRON BE RESPONSIBLE FOR WARRANTY, REPAIR, OR OTHER CLAIMS REGARDING THE PRODUCTS UNLESS
OMRON'S ANALYSIS CONFIRMS THAT THE PRODUCTS WERE PROPERLY HANDLED, STORED, INSTALLED, AND MAINTAINED AND NOT
SUBJECT TO CONTAMINATION, ABUSE, MISUSE, OR INAPPROPRIATE MODIFICATION OR REPAIR.
Application Considerations
SUITABILITY FOR USE
OMRON shall not be responsible for conformity with any standards, codes, or regulations that apply to the combination of products in the customer's application or use of the product.
At the customer's request, OMRON will provide applicable third party certifi cation documents identifying ratings and limitations of use that apply to the products. This information by itself is not suffi cient for a complete determination of the suitability of the products in combination with the end product, machine, system, or other application or use.
The following are some examples of applications for which particular attention must be given. This is not intended to be an exhaustive list of all possible uses of the products, nor is it intended to imply that the uses listed may be suitable for the products:
• Outdoor use, uses involving potential chemical contamination or electrical interference, or conditions or uses not described in this catalog.
• Nuclear energy control systems, combustion systems, railroad systems, aviation systems, medical equipment, amusement machines, vehicles, safety equipment, and installations subject to separate industry or government regulations.
• Systems, machines, and equipment that could present a risk to life or property.
Please know and observe all prohibitions of use applicable to the products.
NEVER USE THE PRODUCTS FOR AN APPLICATION INVOLVING SERIOUS RISK TO LIFE OR PROPERTY WITHOUT ENSURING THAT THE
SYSTEM AS A WHOLE HAS BEEN DESIGNED TO ADDRESS THE RISKS, AND THAT THE OMRON PRODUCT IS PROPERLY RATED AND
INSTALLED FOR THE INTENDED USE WITHIN THE OVERALL EQUIPMENT OR SYSTEM.
Disclaimers
CHANGE IN SPECIFICATIONS
Product specifi cations and accessories may be changed at any time based on improvements and other reasons.
It is our practice to change model numbers when published ratings or features are changed, or when signifi cant construction changes are made.
However, some specifi cations of the product may be changed without any notice. When in doubt, special model numbers may be assigned to fi x or establish key specifi cations for your application on your request. Please consult with your OMRON representative at any time to confi rm actual specifi cations of purchased product.
DIMENSIONS AND WEIGHTS
Dimensions and weights are nominal and are not to be used for manufacturing purposes, even when tolerances are shown.
ERRORS AND OMISSIONS
The information in this catalog has been carefully checked and is believed to be accurate; however, no responsibility is assumed for clerical, typographical, or proofreading errors, or omissions.
PERFORMANCE DATA
Performance data given in this catalog is provided as a guide for the user in determining suitability and does not constitute a warranty. It may represent the result of OMRON’s test conditions, and the users must correlate it to actual application requirements. Actual performance is subject to the OMRON
Warranty and Limitations of Liability.
PROGRAMMABLE PRODUCTS
OMRON shall not be responsible for the user's programming of a programmable product, or any consequence thereof.
COPYRIGHT AND COPY PERMISSION
This catalog shall not be copied for sales or promotions without permission.
This catalog is protected by copyright and is intended solely for use in conjunction with the product. Please notify us before copying or reproducing this catalog in any manner, for any other purpose. If copying or transmitting this catalog to another, please copy or transmit it in its entirety.
Cat. No. E387-E1-01A
OMRON Corporation
Industrial Automation Company http://www.ia.omron.com/
2007.11
In the interest of product improvement, specifications are subject to change without notice.
(c)Copyright OMRON Corporation 2007 All Rights Reserved.
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project