Additel 914 Handheld Pneumatic Pressure Test Pump User Manual

Additel 914 Handheld Pneumatic Pressure Test Pump User Manual
Warnings and cautions
Low pressure gauges can very easily be over pressured if not careful.
Please take caution when applying pressures.
Do not exceed the safety pressure limit of 1000 psi (70bar).
This pump is not designed to be used in explosive or corrosive
environments.
Do not switch between pressure and vacuum mode under pressure.
Over tightening connectors may cause damage.
Store the pump in dry and non-corrosive environments.
Additel is not liable for any safety problems or damages caused by
misuse or incorrect operation.
All moisture and contaminants should be cleaned out of the liquid
isolator before creating a vacuum. This is accomplished by venting
the pump when at high pressure.
For the least risk in contamination of the reference gauge, the DUT
should be installed to the pressure port which is closest to the vent
valve.
Specification
ADT914
Handheld Pneumatic Pressure Test Pump
User's Manual
[Version number:1409V12]
Please download the latest version from www.additel.com
Pressure range: 95% vacuum to 375 psi ( 25 bar )
Temperature: 0~50 ℃
Humidity: < 85%RH
Adjusting fineness: 10 Pa ( 0.1 mbar )
Safety pressure: < 1000 psi ( 70 bar )
Pressure media: Air
Size: Height: 5.12'' ( 130 mm );
Base: 9.45'' ( 240 mm ) x 4.72'' ( 120 mm )
Weight: 3.3 lb ( 1.5 kg )
Configuration & Air routine
2
1
3
4
1 - Quick connector
Air routine
2 - Vent valve
3
(clockwise to close/ counterclockwise to open)
3 - Isolation valve
(isolates the calibration volumn from pressurizing system)
4 - Hand pump
7
2
5 - Fine adjust handle
(clockwise to increase pressure)
4
6 - Security bolt pin
(use it to ensure safety)
7 6
5
7 - Pressure/vacuum (P/V) selector
5
(Postive pressure: pull it out; Vacuum: push it in)
Troubleshooting
Problem
Solution
Cause
Hand pump will not generate pressure The isolation valve is not open.
Open the isolation valve.
A. Vent valve is not closed.
Close the vent valve.
It is difficult to increase pressure B. The O-ring seal is loose or damaged.
Hard to use the fine adjust
It is difficult to turn the quick connectors
Replace the O-ring seal.
C. P/V selector is in the wrong position.
Pressure: pull the P/V selector out; Vacuum: push the P/V selector in.
A. The isolation valve is not closed.
Close the isolation valve.
B. The gauges are not tightened.
Tighten the reference gauge or the gauge under test.
C. The O-ring seal is damaged.
Replace the O-ring seal.
D. The thread surface is not smooth.
Use Teflon tape on the thread and turn it in tight.
E. The connector is mismatched to the gauge pressure port.
Use the correct adapter.
F. Dust or contaminants are blocking the pressure flow
Repeat pressurizing the pump several times and then suddenly vent which will push any contaminants out.
A. Too much force was previously applied.
Secure the quick connection with less force.
B. The threads have no lubrication.
Apply lubrication to the threads.
O-Rings for pressure connector
P/N
Size
Connector
1611300004
4X1.5
M10X1, 1/8BSP, 1/8NPT
1611300220
6.5X3
M20X1.5, 1/2BSP, 1/2NPT
1611300024
6X2
M14X1.5, 1/4BSP, 1/4NPT, 3/8BSP
Basic Operation
Connection
A
DIGITAL PRESSURE CALIBRATOR
Units
A
0
25
50
75
100
SW
SW
V
mA
COM
V
mA
COM
24V
24V
Pressure Gauge
k
pressure switch
dial gauge
Vent
Zeroing should
be operated
under this
condition only.
B
pressure transmitter
Pre-pressure
Increasing pressure
and fine adjustment
C
Units
Units
A
A
A
0
25
50
75
D
DIGITAL PRESSURE CALIBRATOR
DIGITAL PRESSURE CALIBRATOR
DIGITAL PRESSURE CALIBRATOR
Units
flexible hose
75
50
25
0
Pressure Gauge
100
Pressure Gauge
100
75
50
25
0
100
⑤ push down to
Pressure Gauge
open
pre-pressurize to
desired value or until
pressurization stops
① close, do not use too much force
④ close, do not use too much force
③ open
it to the mid
② set
position of its travel
open
to higher
② Pressurize
pressure by turning
the handle and
make fine
adjustments
①
push the security bolt pin in
Decreasing pressure
E
Vacuum pressures
F
Repeat Vacuum pressure
DIGITAL PRESSURE CALIBRATOR
DIGITAL PRESSURE CALIBRATOR
DIGITAL PRESSURE CALIBRATOR
Units
Units
A
0.06
A
0
25
50
75
Pressure Gauge
100
0
25
50
75
② fine adjust
pull out
Units
0.04
A
Vacuum Gauge
100
(0~6)MPa
it slowly to get the desired
① open
value, and close!
G
②
08
0.02
- 0.1
0
0
pull up to achieve the
⑤ desired value
25
50
75
0.06
0.04
Vacuum Gauge
100
08
0.02
- 0.1
0
up to achieve
② pull
desired value
it must be open when
④ close(but
select the pressure/vacuum)
① open
① open
③
push in to select the vacuum position
Decreasing vacuum pressures
H
DIGITAL PRESSURE CALIBRATOR
Units
A
0.06
0.04
Vacuum Gauge
0
25
50
75
100
08
0.02
- 0.1
0
Remark:
it slowly to get the desired
② open
value, and close!
① close
③ fine adjust
A: Additel has made a concerted effort to provide complete and current
information for the proper use of the equipment. The product
specifications and other information contained this manual are subject
to change without notice.
B: Above pictures are just for reference.
Additel Corporation, USA
www.additel.com