We appreciate for your purchasing of Helio LP.
Please remind Caution before use the device for accurate use.
Quality Assurance is included in this user manual.
Manufactured by Good Doctors Co., Ltd.
VER.100
Good Doctors Co., Ltd. 2-706, IT Castle, 550-1, Gasan-dong, Geumcheon-gu, 153-768, Seoul, KOREA
Tel: +82-2-2082-6325 Fax: +82-2-2082-6326 Website: www.gooddrs.com E-mail: [email protected], [email protected]
TABLE OF CONTENTS
1. Table of contents
01
8. Maintenance
14
2. Instruction of Helio LP
02
9. Trouble shooting
15
10. Specification
16
11. Accessories
17
12. Quality Assurance
18
1
Feature
02
2
Intended use
02
3. Caution for safety
03
1
Safety information
03
2
Installation and recharging
04
3
For use
05
4. Contents
06
5. Part name and function
07
1
Name of part
07
2
Brightness control switch /
Indication of beep sound /
LCD screen
08
3
Line inlet / Power button / Belt clip
09
4
Buzzer / Charging indicator / DC inlet
10
6. Helio LP charging
1
Charging
7. Helio LP use
1
Direction for use
1. Table of contents
11
11
12
12
01
2. Instruction of Helio LP
Thank you for choosing our Helio LP.
This instruction provides all the necessary information that you might require for operation,
daily maintenance, and safety. In order to obtain full function and maintain best condition
of this device, please read this manual carefully prior to use. Always keep this manual
ready at hand for your reference.
Helio LP is medical device. Please do not use for the other purpose.
1 Feature
ⴘ Long time use (more than 20 hours at Max brightness)
ⴙ Pre-sterilized light probe
ⴚ Attachable with any medical instrument
ⴛ Lithium-polymer battery
ⴜ 5 level adjustable brightness
ⴝ Portable device
2 Intended use
Provide a direct light without shadow when operating and diagnosing.
02
2. Instruction of Helio LP
3. Safety information
1 Safety information
. Note the following items to use the product safely.
. This equipment has been tested and found to comply with the limits for
medical devices in IEC60601-1-2:1994. These limits are designed to provide
reasonable protection against harmful interference in a typical medical
installation.
5840
Type B applied part
On medical equipment.
To identify a type B applied part complying with
IEC60601-1
Note - B = Body
Description of warning and caution
Warning indicates a potential hazardous situation which, if not avoided,
Warning
could result in death or serious injury.
Caution indicates a potential hazardous situation which, if not avoided, may result
Caution
.
in minor or moderate injury or in property damage.
Warning
This equipment generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful interference
to other devices in the vicinity. However, there is no guarantee that interference will
not occur in a particular installation. If this equipment does cause harmful interference
to other devices, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one or more of the following
measures :
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment into an outlet on a circuit different from that to which
the other device(s) are connected.
- Consult the manufacturer or field service technician for help.
3. Caution for safety
03
2 Installation and recharging
Please read the following information for the safe and proper use of this unit.
Disconnect the power plug before connecting and disconnecting the
power cord to the Control unit.
- Failure to do so may result in an electric shock.
Keep away from sources of ignition.
- This may cause fire or explosion
Never expose the Control unit to any liquid.
- This may cause electrical leaks, electric shock or fire.
Warning
Keep the Control unit on a flat and stable table or inside of product box.
Do not drop the Control unit.
- Internal defect resulting from dropping may cause overheating or fire.
Never place the electrical cord close to the gas burner or sources of heat.
If the cord catches on fire, replace with the new one.
- Short-circuit may cause fire or electric shock.
Do not use or leave the unit in areas of high temperature with intense
direct sunlight, fire or a stove.
- Defect of internal circuit may cause overheat or fire.
Securely connect the dust-free power plug to the outlet. Use this device
in well-ventilated areas.
- Overheating may cause fire.
Caution
Do not touch the terminal areas of both control unit bottom inlet and AC
adapter with finger. Do not insert anything conductive, such as a metal
wire of safety-pin.
- It may cause electric shock or fire.
Do not use any adapter other than supplied.
- It may cause fire due to overheating.
04
3. Caution for safety
3 For use
Please read the following information for the safe and proper use of this unit.
Never handle the power plug with wet hands.
- This may result in an electric shock.
In case of fluid leak or deformation of the Control unit, immediately
discontinue use and have it repaired.
- This may cause fire.
Warning
Do not directly look at the light nor direct the light to the eyes of others.
- It is not familiar for eye.
Never use this product on a person having a pacemaker.
- This may cause malfunction on the pacemaker.
Disconnect the power cord by grasping the power plug.
- This may cause injuries or burns. Insulation degradation may cause
electric shock or fire.
If the battery fluid leaks stick to the skin or cloth, immediately wash it
off with generous amount of water.
- It may damage the skin.
Caution
If the Control unit is not used for an extended period of time,
recharge battery before using it.
- Insulation degradation may cause bad effect for battery life time.
Disconnect the power plug unless it charges.
Never use this product for any purposes other than the specifically
outlined this manual.
3. Caution for safety
05
4. Contents
Please check below contents without omission when you remove package of this
product.
Control unit
Connection cable
AC adapter
Manual
Option
Light probe
Light probe box
(Light probe 50 pcs)
06
4. Contents
5. Part name and function
1 Name of part
ྙ Brightness control switch
ྚ Indication of beep sound
ྛ LCD screen
ྜ Line inlet
ྜྷ Power button
ྞ Belt clip
ྟ Buzzer
ྠ Charging indicator
5. Part name and function
ྡ DC inlet
07
2 Brightness control switch / Indication of beep sound / LCD screen
Brightness control switch
User can adjust brightness by rotating
this switch.
Indication of beep sound
This shows beep sounds
:
:
:
:
:
beep
beep beep
beep beep beep
beep beep beep beep
bee~
LCD screen
It shows device’s condition for example
brightness, battery condition, charging,
and etc.
08
5. Part name and function
3 Line inlet / Power button / Belt clip
Line inlet
Insert the connection cable into the inlet.
Power button
Its function is turn on and off the device.
Belt clip
User can hang product on their belt or etc.
5. Part name and function
09
4 Buzzer / Charging indicator / DC inlet
Buzzer
Produces beeping sound.
Charging indicator
During charging, the red indicator is turned
on and turns green when it is finished.
DC inlet
It is used when charging.
10
5. Part name and function
6. Helio LP charging
1 Charging
Follow the warnings and cautions on charging.
(1) Securely insert the connector of the AC adapter
into the DC inlet on the bottom of the Control unit.
(2) Plug the AC adapter in the AC outlet.
(It can be used on AC 100V - 240V)
(3) Charging indicator is turned on, and started the
charging.
LCD screen on charging
When charging is done, the Charging indicator is turned to green.
If Power button is off, anything is not shown on LCD screen.
Use only AC adapter supplied.
Caution
Contents of LCD screen
Battery condition
Brightness
6. Helio LP charging
Charging condition
Charging
Battery
11
7. Helio LP use
1 Direction for use
1. Insert Connection cable
(1) Insert Connection cable to the line inlet on
Control unit.
2. Attach light probe on instrument
(2) Attach Light probe on instrument
by included band.
(1) Unseal Light probe package.
Example
Saliva ejector
Scaler
12
Mirror
7. Helio LP use
3. Connect Light probe
(1) Connect Light probe to Connection cable.
4. Activate Control unit
(1) Turn on Power button.
-“HELIO” is shown on the LCD screen for
a second.
(2) Select brightness level.
* When battery is almost discharged, device alarms for recharging. At this time battery
condition bar on LCD screen blinks.
7. Helio LP use
13
8. Maintenance
[Follow the warnings and cautions on maintenance and check.]
Cleaning
̻G
Clean the surface with a soft cloth moistened with mild detergent diluted with water.
Do not use solvent that contains oil or ethanol more than 30%.
Caution
Maintenance and replacement of battery
̻G
The power of battery in general is slowly consumed even when it is not being used.
Therefore, it is recommended to be recharged immediately before using.
If the running time extremely declines with the full-charged battery,
it is time to replace battery. Contact local dealer or Good Doctors Co., Ltd.
Use only the battery supplied.
- Any other battery may cause fluid leak or explosion.
Do not use Light probe twice. (It is single use only)
Warning
- Light probe is not autoclaveable.
- Re-using may cause cross infection.
- Using used Light probe may cause critical damage
to device.
Never disassemble or overhaul the product.
- Unusual performance may cause injuries, electric shock, or fire.
̻G
Caution
̻G
If the product malfunctions, reset plaese.
User must be in charge of operation, maintenance, and check.
* Reset: Turn off -> Turn on the Power button.
14
8. Maintenance
9. Trouble shooting
Trouble
Control unit is not
working
When brightness
is poor
Display does not
turn on
Cause and Solution
If Power button is on “OFF”, it is not working.
Move Power button to “ON”.
When brightness is poor, replace the Light probe with
a new one or recharge.
If battery is completely discharged, recharge the battery
or turn off and turn on the device.
If the power cord is not correctly connected to the control
unit, the battery can not be recharged.
Unrechargeable
Clean the DC inlet on control unit or the charging terminal,
if any dirty one is found.
If Helio LP does not operate normally after check up all above list, please contact to
local dealer or Good Doctors Co., Ltd.
9. Trouble shooting
15
10. Specification
Control unit
Product
Light source
Model
Helio LP
Measurement
Approx. H 100mm X L 60mm X T 33mm
Weight
Approx. 120g
Battery
Lithium-polymer battery
Connection cable
Length
Approx. 150cm
Weight
Approx. 15g
AC adapter
Input power
AC 100V ~ 240V
Output power
DC 5V
* Above figures are measured by Good Doctors Co., Ltd.
16
10. Specification
11. Accessories
AC adapter
Light probe
Connection cable
* If you want to buy additional accessories, contact local dealer or Good Doctors Co., Ltd.
Warranty
This product is shipped after strict inspection. If the product has any problem under
normal usage within one year from the date of purchase, repair will be free of charge.
However, the repair of consumables (Light probe, etc) that are not covered by the
warranty, will be charged. For more information, refer to the Quality Assurance
in this manual.
11. Accessories
17
18
- Address :
,
, 200
earthquake, flood, and etc.
4. The compensation method which compensates it based on consumer low.
and warranty claims cannot be accepted.
3. The manufacturer does not accept any liability resulting misuse
- If there is no assurance card.
- Malfunction and damage according to using incompatible product.
- Malfunction and damage according to user’s intention or accident.
Good Doctors Co.,Ltd.
Address : Good Doctors Co., Ltd. 2-706, IT Castle, 550-1, Gasan-dong, Geumcheon-gu, 153-768, Seoul, Korea
Tel: 82-2-2082-6325 Fax: 82-2-2082-6326 Website: www.gooddrs.com E-mail: [email protected], [email protected]
6. S/N :
Product is for 1 year except accessaries.
- Date of purchase : from
5. Assurance period
- Telephone :
- Malfunction and damage according to natural disaster such as fire,
- Malfunction and damage according to user’s carelessness.
4. Distributor
- Seller :
2. Below cases are not effective for the free remuneration.
with the same product.
fault itself with in the assurance period, it will be replaced
1. If there is any defect of manufacturer or product has any
<Assurance details>
We guarantee the quality of the product as follows
- Name :
- Name of dental clinic :
3. Customer information
2. Name od model :
1. Name of product :
Quality Assurance
12. Quality Assurance
12. Quality Assurance
Was this manual useful for you? yes no
Thank you for your participation!