R&S®FSW-K6 Pulse Measurement Application

R&S®FSW-K6 Pulse Measurement Application
FSW-K6_dat-sw_en_5214-6097-22_cover.indd 1
Data Sheet | 03.00
Test & Measurement
R&S®FSW-K6
Pulse Measurement
Application
Specifications
26.02.2015 13:32:09
Version 03.00, February 2015
CONTENTS
Definitions ....................................................................................................................................................................... 3
Specifications .................................................................................................................................................................. 4
Frequency ......................................................................................................................................................................................... 4
Level.................................................................................................................................................................................................. 4
Signal acquisition (standard) .............................................................................................................................................................. 4
Signal acquisition (segmented) .......................................................................................................................................................... 5
Measurement capability (nom.) .......................................................................................................................................................... 6
Measurement accuracy (nom.) .......................................................................................................................................................... 7
Ordering information ...................................................................................................................................................... 9
Upgrades ......................................................................................................................................................................................... 10
2
Rohde & Schwarz R&S®FSW-K6 Pulse Measurement Application
Version 03.00, February 2015
Definitions
General
Product data applies under the following conditions:




Three hours storage at ambient temperature followed by 30 minutes warm-up operation
Specified environmental conditions met
Recommended calibration interval adhered to
All internal automatic adjustments performed, if applicable
Specifications with limits
Represent warranted product performance by means of a range of values for the specified parameter. These specifications are
marked with limiting symbols such as <, ≤, >, ≥, ±, or descriptions such as maximum, limit of, minimum. Compliance is ensured by
testing or is derived from the design. Test limits are narrowed by guard bands to take into account measurement uncertainties, drift
and aging, if applicable.
Specifications without limits
Represent warranted product performance for the specified parameter. These specifications are not specially marked and represent
values with no or negligible deviations from the given value (e.g. dimensions or resolution of a setting parameter). Compliance is
ensured by design.
Typical data (typ.)
Characterizes product performance by means of representative information for the given parameter. When marked with <, > or as a
range, it represents the performance met by approximately 80 % of the instruments at production time. Otherwise, it represents the
mean value.
Nominal values (nom.)
Characterize product performance by means of a representative value for the given parameter (e.g. nominal impedance). In contrast to
typical data, a statistical evaluation does not take place and the parameter is not tested during production.
Measured values (meas.)
Characterize expected product performance by means of measurement results gained from individual samples.
Uncertainties
Represent limits of measurement uncertainty for a given measurand. Uncertainty is defined with a coverage factor of 2 and has been
calculated in line with the rules of the Guide to the Expression of Uncertainty in Measurement (GUM), taking into account
environmental conditions, aging, wear and tear.
Device settings and GUI parameters are indicated as follows: “parameter: value”.
Typical data as well as nominal and measured values are not warranted by Rohde & Schwarz.
Rohde & Schwarz R&S®FSW-K6 Pulse Measurement Application
3
Version 03.00, February 2015
Specifications
The specifications of the R&S®FSW-K6 pulse measurement application are based on the specifications in the data sheet for the
R&S®FSW signal and spectrum analyzer. They have not been checked separately and are not verified during instrument calibration.
Measurement uncertainties are given as 95 % confidence intervals. They apply to the specified center frequencies and measurement
bandwidths. The specified measurement uncertainties do not take into account systematic errors due to reduced signal-to-noise ratio
(S/N).
Frequency
Frequency range
frequency range same as R&S®FSW
Level
Level range
Level setting
RF input
70 dBm 1 to +30 dBm
manual
Signal acquisition (standard)
Input
standard
Output
Measurement bandwidth
1
2
4
RF
iq-tar file
MSRA I/Q data capture
MSRT I/Q data capture
with R&S®FSW-B160 or B320 option
and R&S®FSW-K160R option
with R&S®FSW-B17 option
digital baseband
with R&S®FSW-B71 option
analog baseband
with R&S®FSW-B21 option
external mixer
standard
iq-tar file
with R&S®FSW-B17 option
digital baseband
flat acquisition filter
standard
100 Hz to 10 MHz
with R&S®FSW-B28 option
100 Hz to 28 MHz (nom.) 2
with R&S®FSW-B40 option
100 Hz to 40 MHz (nom.) 2
with R&S®FSW-B80 option
100 Hz to 80 MHz (nom.) 2
with R&S®FSW-B160 option
100 Hz to 160 MHz (nom.) 2
with R&S®FSW-B320 option
100 Hz to 320 MHz (nom.) 2
®
with R&S FSW-B500 option
100 Hz to 500 MHz (nom.) 2
with R&S®FSW-B2000 option
100 Hz to 2000 MHz (nom.) 2
Gaussian acquisition filter (3 dB bandwidth)
Note that all Gaussian filters above 10 MHz are limited in total I/Q bandwidth according
to the installed bandwidth option. For certain bandwidths this can result in a “Gaussian
top” filter shape as described in the user manual.
standard
100/200/400/800 Hz,
1.6/3.2/6.4/12.5/25/50/100/200/400/800 kHz
1.6/3/5/8/10 MHz
with R&S®FSW-B40 option
as above and 18 MHz 2
with R&S®FSW-B80 option
as above and 28 MHz, 40 MHz 2
with R&S®FSW-B160 option
as above and 80 MHz, 160 MHz 2
with R&S®FSW-B320 option
as above and 250 MHz 2
®
with R&S FSW-B500 option
as above and 320 MHz 2
with R&S®FSW-B2000 option
as above and 500 MHz, 1000 MHz 2
Requires R&S®FSW-B24 RF preamplifier option.
YIG preselector off for f ≥ 8 GHz.
Rohde & Schwarz R&S®FSW-K6 Pulse Measurement Application
Version 03.00, February 2015
Measurement time
Triggering
maximum for measurement bandwidth (RF input, flat filter)
10 MHz
36.9 s
28 MHz (R&S®FSW-B28 option)
13.1 s
40 MHz (R&S®FSW-B40 option)
8.92 s
80 MHz (R&S®FSW-B80 option)
4.6 s
160 MHz (R&S®FSW-B160 option)
2.3 s
320 MHz (R&S®FSW-B320 option)
492.8 ms
320 MHz (R&S®FSW-B500 option)
1.15 s
500 MHz (R&S®FSW-B500 option)
786.9 ms
2000 MHz (R&S®FSW-B2000 option)
up to 80 ms 3
RF input
free run, external, I/Q power, IF power,
RF power, frequency mask trigger 4
Signal acquisition (segmented)
Input
Measurement bandwidth 5
Measurement time (compressed)
Triggering
3
4
5
standard
RF
with R&S®FSW-B21 option
external mixer
flat acquisition filter
standard
100 Hz to 10 MHz
with R&S®FSW-B28 option
100 Hz to 28 MHz (nom.) 2
with R&S®FSW-B40 option
100 Hz to 40 MHz (nom.) 2
®
with R&S FSW-B80 option
100 Hz to 80 MHz (nom.) 2
with R&S®FSW-B160/-B320 option 5
100 Hz to 160 MHz (nom.) 2
®
with R&S FSW-B500 option
100 Hz to 500 MHz (nom.) 2
Gaussian acquisition filter (3 dB bandwidth)
Note that all Gaussian filters above 10 MHz are limited in total I/Q bandwidth according
to the installed bandwidth option. For certain bandwidths this can result in a “Gaussian
top” filter shape as described in the user manual.
standard
100/200/400/800 Hz,
1.6/3.2/6.4/12.5/25/50/100/200/400/800 kHz
1.6/3/5/8/10 MHz
with R&S®FSW-B40 option
as above and 18 MHz 2
®
with R&S FSW-B80 option
as above and 28 MHZ, 40 MHz 2
with R&S®FSW-B1600/-B320 options
as above and 80 MHz and 160 MHz 2
®
with R&S FSW-B500 option
as above and 320 MHz 2
maximum for measurement bandwidth (RF input, flat filter)
10 MHz
30.8 s
28 MHz (R&S®FSW-B28 option)
11.0 s
40 MHz (R&S®FSW-B40 option)
7.7 s
80 MHz (R&S®FSW-B80 option)
3.85 s
160 MHz (R&S®FSW-B160/-B320)
1.92 s
500 MHz (R&S®FSW-B500 option)
641.7 ms
RF input
external, IF power, RF power
no. of trigger events
2 to 100 000
The maximal value depends on the R&S®RTO memory configuration. 200 Msample can be reached with a R&S®RTO-B104 memory extension and a
sampling rate of 2.5 GHz.
Using MSRT I/Q data capture; R&S®FSW-K160R option and one of the options R&S®FSW-B160/-B320 is required, not available for R&S®FSW-B500.
The R&S®FSW-B320 option is limited to 160 MHz of I/Q bandwidth when using segmented capture.
Rohde & Schwarz R&S®FSW-K6 Pulse Measurement Application
5
Version 03.00, February 2015
Measurement capability (nom.)
Pulse detection
System rise time
Pulse parameters
Result displays
6
6
measured pulses
1 to 100 000
min. pulse width for measurement bandwidth (flat)
10 MHz
400 ns
28 MHz (R&S®FSW-B28 option)
150 ns
40 MHz (R&S®FSW-B40 option)
100 ns
80 MHz (R&S®FSW-B80 option)
50 ns
160 MHz (R&S®FSW-B160 option)
25 ns
320 MHz (R&S®FSW-B320 option)
12.5 ns
500 MHz (R&S®FSW-B500 option)
8 ns
measurement bandwidth (flat acquisition filter)
10 MHz
< 110 ns
28 MHz (R&S®FSW-B28 option)
< 40 ns
40 MHz (R&S®FSW-B40 option)
< 28 ns
80 MHz (R&S®FSW-B80 option)
< 14 ns
160 MHz (R&S®FSW-B160 option)
< 7 ns
320 MHz (R&S®FSW-B320 option)
< 3.5 ns
500 MHz (R&S®FSW-B500 option)
< 2.2 ns
measurement bandwidth (Gauss acquisition filter)
10 MHz
< 73 ns
28 MHz (R&S®FSW-B80 option)
< 26 ns
40 MHz (R&S®FSW-B80 option)
< 23 ns
80 MHz (R&S®FSW-B160 option)
< 12 ns
160 MHz (R&S®FSW-B320 option)
< 6 ns
250 MHz (R&S®FSW-B320 option)
< 4 ns
320 MHz (R&S®FSW-B500 option)
< 3 ns
timing
timestamp, settling time, rise time, fall time, pulse width, off time, duty ratio, duty cycle,
pulse repetition interval, pulse repetition frequency
amplitude
top power, base power, average on power,
average transmitted power, minimum power, peak power, peak-to-average on power
ratio, peak-to-average transmitted power ratio, peak-to-min power ratio, droop, ripple,
overshoot, power (at point), pulse-to-pulse power ratio (at point)
frequency
frequency (at point), pulse-to-pulse frequency difference (at point), frequency deviation,
frequency error (peak), frequency error (RMS), chirp rate
phase
phase (at point), pulse-to-pulse phase difference (at point), phase deviation, phase
error (peak), phase error (RMS)
time sidelobe 6
peak-to-sidelobe level, integrated sidelobe level, mainlobe 3 dB width, sidelobe delay,
compression ratio, mainlobe power (integrated) mainlobe power (average), peak
correlation, mainlobe phase, mainlobe frequency
inter-pulse analysis
table with numeric values per pulse, table with statistics (average, standard deviation,
max., min.), trend plot of parameter versus time, scatter plot of parameter versus
parameter, spectrum of parameter versus time, histogram of parameter distribution
intra-pulse analysis
traces aligned to pulse for magnitude versus time, frequency versus time, phase
(wrapped or unwrapped) versus time, power spectrum, correlated magnitude 6,
frequency error 6, phase error 6
Requires R&S®FSW-K6S time sidelobe measurement upgrade option.
Rohde & Schwarz R&S®FSW-K6 Pulse Measurement Application
Version 03.00, February 2015
Measurement accuracy (nom.)
Amplitude and timing parameters
The statistical measurement uncertainty is given below as a 95 % confidence interval (Gauss acquisition filter) 7.
Pulse width
±2 %
Duty cycle
±2 %
Average on power
±0.2 dB + absolute amplitude accuracy
Average transmitted power
±0.2 dB + absolute amplitude accuracy
Peak power
±0.3 dB + absolute amplitude accuracy
Frequency and phase parameters (CW pulse modulation)
The total frequency accuracy is comprised of absolute frequency accuracy and a statistical uncertainty due to measurement noise.
The absolute frequency accuracy is given in the R&S®FSW data sheet.
The statistical measurement uncertainty is given below as a 95 % confidence interval at stated center frequencies and measurement
bandwidths (flat acquisition filter) for a pulse modulated CW carrier 8.
2 GHz center frequency
Measurement bandwidth (flat)
10 MHz
28 MHz (R&S®FSW-B28)
40 MHz (R&S®FSW-B40)
80 MHz (R&S®FSW-B80)
160 MHz (R&S®FSW-B160)
320 MHz (R&S®FSW-B320)
500 MHz (R&S®FSW-B500)
Measured parameter
Frequency error (RMS) 9
± 1.5 kHz
± 5.5 kHz
± 9.5 kHz
± 43 kHz
± 85 kHz
± 260 kHz
± 430 kHz
Pulse-to-pulse frequency 10
± 2.5 kHz
± 12 kHz
± 22 kHz
± 90 kHz
± 200 kHz
± 650 kHz
± 1100 kHz
Pulse-to-pulse phase 10
±0.11°
±0.13°
±0.15°
±0.20°
±0.29°
±0.39°
±0.45°
Measured parameter
Frequency error (RMS) 9
± 2 kHz
± 5 kHz
± 8.5 kHz
± 40 kHz
± 80 kHz
± 230 kHz
± 370 kHz
Pulse-to-pulse frequency 10
± 3 kHz
± 11.5 kHz
± 21 kHz
± 70 kHz
± 190 kHz
± 625 kHz
± 900 kHz
Pulse-to-pulse phase 10
± 0.15°
± 0.18°
± 0.18°
± 0.20°
± 0.30°
± 0.36°
± 0.43°
Measured parameter
Frequency error (RMS) 9
± 3 kHz
± 8.5 kHz
± 14.5 kHz
± 60 kHz
± 100 kHz
± 300 kHz
± 500 kHz
Pulse-to-pulse frequency 10
± 6 kHz
± 20 kHz
± 35 kHz
± 130 kHz
± 230 kHz
± 750 kHz
± 1325 kHz
Pulse-to-pulse phase 10
± 0.35°
± 0.40°
± 0.40°
± 0.45°
± 0.55°
± 0.70°
± 0.90°
8 GHz center frequency
Measurement bandwidth (flat)
10 MHz
28 MHz (R&S®FSW-B28)
40 MHz (R&S®FSW-B40)
80 MHz (R&S®FSW-B80)
160 MHz (R&S®FSW-B160)
320 MHz (R&S®FSW-B320)
500 MHz (R&S®FSW-B500)
20 GHz center frequency
Measurement bandwidth (flat)
10 MHz
28 MHz (R&S®FSW-B28)
40 MHz (R&S®FSW-B40)
80 MHz (R&S®FSW-B80)
160 MHz (R&S®FSW-B160)
320 MHz (R&S®FSW-B320)
500 MHz (R&S®FSW-B500)
7
8
9
10
Pulse width > 10/measurement bandwidth, pulse rise/fall time > 3 × system rise time.
Signal level ≥ –10 dBm, RF level and attenuator: auto, 100 MHz external reference with 300 Hz bandwidth locked to sender, PRI ≤ 10 ms.
Measurement range: 50 % of pulse top, pulse width ≥ 100/measurement bandwidth.
Pulse-to-pulse measurement point occurs at least 10/measurement bandwidth after the rising edge (i.e. 50 % level crossing) and 10/measurement
bandwidth before the falling edge (i.e. 50 % level crossing), pulse-to-pulse measurement point time difference ≤ 10 ms.
Rohde & Schwarz R&S®FSW-K6 Pulse Measurement Application
7
Version 03.00, February 2015
Frequency parameters (linear FM pulse modulation)
The total frequency accuracy is comprised of absolute frequency accuracy and a statistical uncertainty due to measurement noise.
The absolute frequency accuracy is given in the R&S®FSW data sheet
The statistical measurement uncertainty is given below as a 95 % confidence interval at stated center frequencies and measurement
bandwidths (flat acquisition filter) for a pulsed and linearly frequency modulated carrier 11.
2 GHz center frequency
Measurement bandwidth (flat)
10 MHz
28 MHz (R&S®FSW-B28)
40 MHz (R&S®FSW-B40)
80 MHz (R&S®FSW-B80)
160 MHz (R&S®FSW-B160)
320 MHz (R&S®FSW-B320)
500 MHz (R&S®FSW-B500)
Measured parameter
Frequency error (RMS) 12
± 2 kHz
± 6 kHz
± 8 kHz
± 29 kHz
± 75 kHz
± 230 kHz
± 390 kHz
Pulse-to-pulse frequency 13
± 5 kHz
± 21 kHz
± 28 kHz
± 110 kHz
± 190 kHz
± 625 kHz
± 1100 kHz
Measured parameter
Frequency error (RMS) 12
± 2.5 kHz
± 6 kHz
± 9 kHz
± 36 kHz
± 85 kHz
± 250 kHz
± 410 kHz
Pulse-to-pulse frequency 13
± 6.5 kHz
± 28 kHz
± 31 kHz
± 110 kHz
± 230 kHz
± 725 kHz
± 1075 kHz
measured parameter
Frequency error (RMS) 12
± 3 kHz
± 10 kHz
± 16 kHz
± 50 kHz
± 120 kHz
± 370 kHz
± 675 kHz
Pulse-to-pulse frequency 13
± 8.5 kHz
± 31 kHz
± 49 kHz
± 160 kHz
± 1175 kHz
± 1100 kHz
± 1975 kHz
8 GHz center frequency
Measurement bandwidth (flat)
10 MHz
28 MHz (R&S®FSW-B28)
40 MHz (R&S®FSW-B40)
80 MHz (R&S®FSW-B80)
160 MHz (R&S®FSW-B160)
320 MHz (R&S®FSW-B320)
500 MHz (R&S®FSW-B500)
20 GHz center frequency
Measurement bandwidth (flat)
10 MHz
28 MHz (R&S®FSW-B28)
40 MHz (R&S®FSW-B40)
80 MHz (R&S®FSW-B80)
160 MHz (R&S®FSW-B160)
320 MHz (R&S®FSW-B320)
500 MHz (R&S®FSW-B500)
11
12
13
8
Signal level ≥ –10 dBm, RF level and attenuator: auto, 100 MHz external reference with 300 Hz bandwidth locked to sender, PRI ≤ 10 ms.
Measurement range: 50 % of pulse top, pulse width ≥ 1000/measurement bandwidth.
Pulse-to-pulse measurement point occurs at least 10/measurement bandwidth after the rising edge (i.e. 50 % level crossing) and 10/measurement
bandwidth before the falling edge (i.e. 50 % level crossing), pulse-to-pulse measurement point time difference ≤ 10 ms.
Rohde & Schwarz R&S®FSW-K6 Pulse Measurement Application
Version 03.00, February 2015
Ordering information
Designation
Pulse Measurement Application
Time Sidelobe Measurements
Type
R&S®FSW-K6
R&S®FSW-K6S
Order No.
1313.1322.02
1325.3783.02
Retrofittable
yes
yes
Remarks
user-retrofittable
user-retrofittable,
R&S®FSW-K6 required
Recommended options and extras
Real-Time Spectrum Analyzer, 160 MHz
R&S®FSW-K160R
1313.5340.02
yes
OCXO Precision Frequency Reference
RF Preamplifier, 100 kHz to 13.6 GHz
R&S®FSW-B4
R&S®FSW-B24
1313.0703.02
1313.0832.13
yes
yes
RF Preamplifier, 100 kHz to 26.5 GHz
R&S®FSW-B24
1313.0832.26
yes
RF Preamplifier, 100 kHz to 43.5 GHz
R&S®FSW-B24
1313.0832.43
yes
RF Preamplifier, 100 kHz to 50 GHz
R&S®FSW-B24
1313.0832.50
yes
RF Preamplifier, 100 kHz to 67 GHz
R&S®FSW-B24
1313.0832.67
yes
28 MHz Analysis Bandwidth
40 MHz Analysis Bandwidth
80 MHz Analysis Bandwidth
160 MHz Analysis Bandwidth
320 MHz Analysis Bandwidth
500 MHz Analysis Bandwidth
2000 MHz Analysis Bandwidth
Digital Baseband Interface
Analog Baseband Inputs,
40 MHz Analysis Bandwidth
Analog Baseband Inputs,
40 MHz Analysis Bandwidth
Analog Baseband Inputs,
40 MHz Analysis Bandwidth
Analog Baseband Inputs,
80 MHz Analysis Bandwidth
Highpass Filter for Harmonic
Measurements
LO/IF Connections for external mixers
R&S®FSW-B28
R&S®FSW-B40
R&S®FSW-B80
R&S®FSW-B160
R&S®FSW-B320
R&S®FSW-B500
R&S®FSW-B2000
R&S®FSW-B17
R&S®FSW-B71
1313.1645.02
1313.0861.02
1313.0878.02
1313.1668.02
1313.7172.02
1313.4296.02
3591.7436.02
1313.0784.02
1313.1651.13
yes
yes
yes
yes
yes
yes
yes
yes
yes
R&S®FSW-B71
1313.1651.26
yes
R&S®FSW-B71
1313.1651.67
yes
R&S®FSW-B71E
1313.6547.02
yes
R&S®FSW-B13
1313.0761.02
yes
one of the options
R&S®FSW-B160/-B320 is
required, not available for
R&S®FSW-B500
user-retrofittable
for R&S®FSW8/13;
contact service center
for R&S®FSW26;
contact service center
for R&S®FSW43;
contact service center
for R&S®FSW50;
contact service center
for R&S®FSW67;
contact service center
user-retrofittable
user-retrofittable
user-retrofittable
contact service center
contact service center
contact service center
contact service center
user-retrofittable
for R&S®FSW8/13;
contact service center
for R&S®FSW26/43/50;
contact service center
for R&S®FSW67;
contact service center
R&S®FSW71 required;
user-retrofittable
user-retrofittable
R&S®FSW-B21
1313.1100.26
yes
LO/IF Connections for external mixers
R&S®FSW-B21
1313.1100.43
yes
Harmonic Mixer, 40 GHz to 60 GHz
R&S®FS-Z60
1089.0799.02
yes
Harmonic Mixer, 50 GHz to 75 GHz
R&S®FS-Z75
1048.0271.02
yes
Harmonic Mixer, 60 GHz to 90 GHz
R&S®FS-Z90
1048.0371.02
yes
Harmonic Mixer, 75 GHz to 110 GHz
R&S®FS-Z110
1048.0471.02
yes
for R&S®FSW26;
contact service center
for R&S®FSW43/50/67;
contact service center
user-retrofittable,
R&S®FSW-B21 required
user-retrofittable,
R&S®FSW-B21 required
user-retrofittable,
R&S®FSW-B21 required
user-retrofittable,
R&S®FSW-B21 required
Oscilloscope and options needed for bandwidth extension R&S®FSW-B2000
Digital Oscilloscope, 4 GHz, 20 Gsample/s, R&S®RTO1044
1316.1000.44
20/80 Msample, 4 channels
OCXO 10 MHz
R&S®RTO-B4
1304.8305.02
Oscilloscope extras
Memory Upgrade, 50 Msample per channel R&S®RTO-B101
1304.8428.02
Memory Upgrade, 100 Msample per
R&S®RTO-B102
1304.8434.02
channel
Memory Upgrade, 200 Msample per
R&S®RTO-B103
1304.8440.02
channel (instruments with Windows 7
Embedded operating system)
Memory Upgrade, 400 Msample per
R&S®RTO-B104
1304.8457.02
channel (instruments with Windows 7
Embedded operating system)
Rohde & Schwarz R&S®FSW-K6 Pulse Measurement Application
9
Version 03.00, February 2015
Upgrades
Designation
Analysis Bandwidth Upgrade
from 28 MHz to 40 MHz
Analysis Bandwidth Upgrade
from 40 MHz to 80 MHz
Type
R&S®FSW-U40
Order No.
1313.5205.02
Retrofittable
yes
R&S®FSW-U80
1313.5211.02
yes
Analysis Bandwidth Upgrade
from 80 MHz to 160 MHz
R&S®FSW-U160
1313.5754.02
yes
Analysis Bandwidth Upgrade
from 160 MHz to 320 MHz
R&S®FSW-U320
1313.7189.02
yes
Analysis Bandwidth Upgrade
from 80 MHz to 500 MHz
R&S®FSW-U500
1321.6320.02
yes
For R&S®FSW signal and spectrum analyzer data sheet, see PD 5214.5984.22 and
for R&S®FSW product brochure, see PD 5214.5984.12
10
Rohde & Schwarz R&S®FSW-K6 Pulse Measurement Application
Remarks
user-retrofittable,
R&S®FSW-B28 required
user-retrofittable,
R&S®FSW-B40 or
R&S®FSW-U40 required
contact service center,
R&S®FSW-B80 or
R&S®FSW-U80 required
contact service center,
R&S®FSW-B160 or
R&S®FSW-U160 required
contact service center,
R&S®FSW-B80 or
R&S®FSW-U80 required
Version 03.00, February 2015
Rohde & Schwarz R&S®FSW-K6 Pulse Measurement Application
11
Service that adds value
Worldwide
Local and personalized
❙ Customized and flexible
❙ Uncompromising quality
❙ Long-term dependability
❙
❙
About Rohde & Schwarz
The Rohde & Schwarz electronics group is a leading
­supplier of solutions in the fields of test and ­measurement,
broadcast and media, secure communications, cyber­
security, and radiomonitor­ing and radiolocation. Founded
more than 80 years ago, this independent global ­company
has an extensive sales network and is present in more
than 70 countries. The company is headquartered in
­Munich, Germany.
Sustainable product design
❙❙ Environmental compatibility and eco-footprint
❙❙ Energy efficiency and low emissions
❙❙ Longevity and optimized total cost of ownership
Certified Quality Management
ISO 9001
Certified Environmental Management
ISO 14001
Rohde & Schwarz GmbH & Co. KG
www.rohde-schwarz.com
R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG
Trade names are trademarks of the owners
PD 5214.6097.22 | Version 03.00 | February 2015 (as)
R&S®FSW-K6 Pulse Measurement Application
Data without tolerance limits is not binding | Subject to change
© 2011 - 2015 Rohde & Schwarz GmbH & Co. KG | 81671 Munich, Germany
5214.6097.22 03.00 PDP 1 en
Regional contact
❙❙ Europe, Africa, Middle East | +49 89 4129 12345
[email protected]
❙❙ North America | 1 888 TEST RSA (1 888 837 87 72)
[email protected]
❙❙ Latin America | +1 410 910 79 88
[email protected]
❙❙ Asia Pacific | +65 65 13 04 88
[email protected]
❙❙ China | +86 800 810 82 28 | +86 400 650 58 96
[email protected]
5214609722
FSW-K6_dat-sw_en_5214-6097-22_cover.indd 2
26.02.2015 13:32:09
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertisement