GE Series 35 IS Single-Channel Hygrometer Moisture Monitor User’s Manual
Below you will find brief information for Moisture Monitor Series 35 IS Single-Channel Hygrometer. The Series 35 IS, a microprocessor-based, single-channel hygrometer, measures moisture content in gases. It is suitable for use in a wide range of process conditions requiring real-time moisture measurement. The unit measures dew/frost points over a temperature range of –110 to +60°C (–166 to +140°F).
Advertisement
Advertisement
GE
Sensing & Inspection Technologies
Moisture Monitor Series 35 IS
Single-Channel Hygrometer
User’s Manual
GE
Sensing & Inspection Technologies
Moisture Monitor Series 35 IS
Single-Channel Hygrometer
User’s Manual
910-182D
March 2008
March 2008
Warranty
Each instrument manufactured by GE Sensing, Inc. is warranted to be free from defects in material and workmanship. Liability under this warranty is limited to restoring the instrument to normal operation or replacing the instrument, at the sole discretion of GE. Fuses and batteries are specifically excluded from any liability. This warranty is effective from the date of delivery to the original purchaser. If
GE determines that the equipment was defective, the warranty period is:
•
one year from delivery for electronic or mechanical failures
•
one year from delivery for sensor shelf life
If GE determines that the equipment was damaged by misuse, improper installation, the use of unauthorized replacement parts, or operating conditions outside the guidelines specified by GE, the repairs are not covered under this warranty.
The warranties set forth herein are exclusive and are in lieu of all other warranties whether statutory, express or implied (including warranties or merchantability and fitness for a particular purpose, and warranties arising from course of dealing or usage or trade).
iii
March 2008
Return Policy
If a GE Sensing, Inc. instrument malfunctions within the warranty period, the following procedure must be completed:
1. Notify GE, giving full details of the problem, and provide the model number and serial number of the instrument. If the nature of the problem indicates the need for factory service, GE will issue a RETURN AUTHORIZATION NUMBER
(RAN), and shipping instructions for the return of the instrument to a service center will be provided.
2. If GE instructs you to send your instrument to a service center, it must be shipped prepaid to the authorized repair station indicated in the shipping instructions.
3. Upon receipt, GE will evaluate the instrument to determine the cause of the malfunction.
Then, one of the following courses of action will then be taken:
•
If the damage is covered under the terms of the warranty, the instrument will be repaired at no cost to the owner and returned.
•
If GE determines that the damage is not covered under the terms of the warranty, or if the warranty has expired, an estimate for the cost of the repairs at standard rates will be provided. Upon receipt of the owner’s approval to proceed, the instrument will be repaired and returned.
iv
March 2008
Table of Contents
Chapter 1: Features & Capabilities
The Electronics Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Moisture Probes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
The Sample System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
The User Program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Chapter 2: Installing the Series 35 IS
Choosing a Measurement Site . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Moisture Probe Considerations . . . . . . . . . . . . . . . . . . . . . . . 2-3
Sample System Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Mounting an External Sample System. . . . . . . . . . . . . . . . . 2-6
The Self-Contained Sample System . . . . . . . . . . . . . . . . . . . 2-7
Installing a Probe into a Sample System. . . . . . . . . . . . . . . 2-8
External Sample System . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Self-Contained Sample System . . . . . . . . . . . . . . . . . . 2-10
Making External Probe Connections. . . . . . . . . . . . . . . . . . 2-11
Using Modified or Non-GE Sensing Cables . . . . . . . . 2-14
Performing a Calibration Adjustment. . . . . . . . . . . . . 2-15
v
March 2008
Table of Contents (cont.)
Chapter 3: Operating the Series 35 IS
Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Starting Up and Shutting Down the Sample System 3-3
Powering Up the Series 35 IS. . . . . . . . . . . . . . . . . . . . . . 3-4
Using the Keypad to Enter and Change Data . . . . . . . . . . 3-5
How to Use the Programming Keys . . . . . . . . . . . . . . . 3-5
How to Enter and Exit the User Program. . . . . . . . . . . 3-6
How to Move Through the User Program . . . . . . . . . . 3-7
How to Enter Numeric Data. . . . . . . . . . . . . . . . . . . . . . . 3-8
Verifying and Changing Factory Setup Data . . . . . . . . . . . 3-9
Entering the Dew Point Range . . . . . . . . . . . . . . . . . . . 3-10
Entering Calibration Data. . . . . . . . . . . . . . . . . . . . . . . . 3-11
Entering the Probe Serial Number . . . . . . . . . . . . . . . . 3-12
Entering High and Low Reference Values. . . . . . . . . 3-13
Changing the Measurement Display . . . . . . . . . . . . . . . . . 3-14
Entering an Offset Value for Dew/Frost Point . . . . . . . . . 3-15
Entering a Constant Pressure Value . . . . . . . . . . . . . . . . . . 3-16
Entering a PPMv Constant Multiplier . . . . . . . . . . . . . . . . . 3-18
Entering an Interval for Automatic Calibration. . . . . . . . 3-19
Viewing the Battery Status . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
Entering a Backlight-On Time Interval. . . . . . . . . . . . . . . . 3-22
Setting Up Computer-Enhanced Response . . . . . . . . . . . 3-23
Setting Up Range Error Processing . . . . . . . . . . . . . . . . . . . 3-25
Selecting AC Line Frequency (60/50 Hz) . . . . . . . . . . . . . . 3-26
Adjusting the Contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-27
Setting the Clock Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-28
Setting Up the Data Logger. . . . . . . . . . . . . . . . . . . . . . . . . . 3-30
Setting the Data Logger Time Interval . . . . . . . . . . . . 3-31
Entering the Parameter(s) to Log . . . . . . . . . . . . . . . . . 3-32
Viewing Log Status, Summary and Data. . . . . . . . . . 3-34
Using the Built-In RS232 Interface. . . . . . . . . . . . . . . . 3-39
Dumping Logged Data Via the RS232 Interface . . . 3-40
vi
March 2008
Table of Contents (cont.)
Chapter 4: Troubleshooting and Maintenance
Common Problems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Screen Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Calibration Error Processing . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Recharging and Replacing the Battery . . . . . . . . . . . . . . . . 4-8
Recharging the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Replacing the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Replacing the User Program . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Removing the Circuit Board. . . . . . . . . . . . . . . . . . . . . . 4-14
Replacing the EPROM. . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
Replacing the Circuit Board . . . . . . . . . . . . . . . . . . . . . . 4-19
Replacing and Recalibrating the Moisture Probe . . . . . . 4-20
Chapter 5: Specifications
Electrical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Appendix A: Outline and Installation Drawings
Appendix B: Menu Map
Appendix C: Data Information Sheet
Series 35 IS Data Information Sheet. . . . . . . . . . . . . . . . . . . . -2
Appendix D: Series 35 IS Hygrometer Spare Parts
vii
March 2008
Chapter 1
Features & Capabilities
The Series 35 IS, a microprocessor-based, single-channel hygrometer, measures moisture content in gases. The Series 35 IS is suitable for use in a wide range of process conditions requiring real-time moisture measurement. It measures dew/frost points over a temperature range of –110 to +60°C (–166 to +140°F).
The following topics are discussed in this chapter:
•
The Electronics Unit [page 1-2]
•
•
•
Features & Capabilities 1-1
March 2008
The Electronics Unit
The Series 35 IS portable moisture monitor is available in two configurations:
•
with a self-contained sample system
•
without a self-contained sample system.
Both Series 35 IS configurations display measurement data on a one-line, 16-character alpha-numeric LCD display screen. All probe information is entered into the unit using the programming
keys on the front panel keypad (see Figure 1-1 below). The
Series 35 IS operates on an internal 6 VDC battery, and the unit can operate while being charged with a 12 VDC charger.
Caution!
Do not use AC power in hazardous areas. Charge the battery only in a non-hazardous area.
POWER
ESC
P
1-2
Figure 1-1: Series 35 IS Front Panel
Features & Capabilities
March 2008
Moisture Probes
The moisture probe is the part of the system that comes in contact with the process, and is usually installed in a sample system. The
Series 35 IS uses any M Series probe to measure dew point temperature in °C or °F. A sensor assembly is secured to the probe mount and protected with a sintered stainless steel shield
(see Figure 1-2 below). Other types of shields are available.
Figure 1-2: The M Series Probe
The Sample System
The sample system conditions or controls a sample stream to the specifications of the measurement probe. Typically, the sample system is kept very simple, with as few components as possible located upstream of the measurement probe. The sample system may include a filter to remove particulates from the sample stream and/or a pressure regulator to reduce or control the pressure of the sample stream. In general, stainless steel is the preferred material for all the sample system’s wetted parts.
If it was specified at the time of purchase, your Series 35 IS will include a self-contained sample system.
Features & Capabilities 1-3
March 2008
The User Program
The Series 35 IS contains a user program that enables you to change moisture calibration data and select a number of userdefined program functions.
All functions and features of the Series 35 IS user program are discussed in Chapter 3, Operating the Series 35 IS.
1-4 Features & Capabilities
March 2008
Chapter 2
Installing the Series 35 IS
This chapter discusses installing the Series 35 IS in all its configurations. Use the following list of procedures to help you install your unit.
Depending on the type of unit you have, refer to the appropriate section(s) that follow to install your Series 35 IS correctly.
Installing your Series 35 IS consists of the following procedures:
•
Choosing a Measurement Site [page 2-2]
•
Moisture Probe Considerations [page 2-3]
•
Sample System Guidelines [page 2-5]
•
Mounting an External Sample System [page 2-6]
•
The Self-Contained Sample System [page 2-7]
•
Installing a Probe into a Sample System [page 2-8]
•
Making External Probe Connections [page 2-11]
Proceed to the appropriate section to install your Series 35 IS.
Installing the Series 35 IS 2-1
March 2008
Choosing a Measurement Site
You should have discussed environmental factors with a GE
Sensing applications engineer or field sales person before you receive your Series 35 IS. The equipment should be suited to both the application and the measurement site.
Read the following guidelines to verify that you have selected the best measurement site.
•
Choose a measurement site for the probe and sample system as close to the process line as possible. Avoid long lengths of connecting tubing. If long distances are unavoidable, a fastsampling bypass loop is recommended.
•
Do not install any other components, such as filters, upstream of the probe or sample system unless instructed to do so by GE
Sensing. Many common components, such as filters and pressure regulators, are not suitable for sample systems. These components contain wetted parts that may absorb or release materials (e.g., moisture) into the sample stream or allow ambient contamination to enter the sample stream. In general, use stainless-steel for all parts in contact with the sample.
•
Observe all normal safety precautions. Use the probes within their maximum pressure and temperature ratings.
•
Do not expose the electronics unit to high temperatures, strong electrical transients, mechanical vibrations, corrosive atmospheres, or any other conditions that could damage or interfere with the Series 35 IS operation. See Chapter 5,
Specifications, for limits.
•
If you are using an external sample system, you should observe the proper cable restrictions for the probes. The M
Series probes require specially shielded cable. For units with separate sample systems, you can locate the M Series probes up to 100 meters (330 feet) from the Series 35 IS.
•
Protect the probe cables from excessive strain (bending,
2-2
pulling, twisting, etc.). In addition, do not subject the cables to temperatures above 65°C (149°F) or below –50°C (–58°F).
Installing the Series 35 IS
March 2008
Moisture Probe Considerations
The M Series probes consist of an aluminum oxide sensor located on a connector and covered by a protective stainless-steel shield.
The probe sensor materials and housing maximize durability and minimize water absorbing surfaces in the vicinity of the aluminum oxide surface. A sintered stainless-steel shield protects the sensor from high flow rates and particulate matter (other shields are available). Do not remove the shield except upon advice from GE Sensing.
The sensor design permits the device to withstand normal shock and vibration. Be sure to avoid touching the active sensor surface or allowing it to come into direct contact with foreign objects.
Such contact may adversely affect performance.
Observing these precautions should ensure a long and useful probe life. GE Sensing recommends that you check probe calibration routinely, at one-year intervals, or as recommended by the applications engineers for your particular application.
Because the probe measures water vapor pressure in its immediate vicinity, proximity to the system walls, materials of construction, and other environmental factors can influence readings. You can operate the sensor under vacuum or pressure and under dynamic or static conditions. Observe the following environmental precautions.
1. Temperature Range: the standard probe is operable from
–110°C to +70°C (–166°F to +158°F).
2. Moisture Condensation: be sure the temperature is at least
10°C higher than the dew/frost point temperature. If this condition is not maintained, moisture condensation could occur on the sensor or in the sample system and cause reading errors. If reading errors occur, refer to “Aluminum Oxide
Probe Maintenance” in Appendix A.
Installing the Series 35 IS 2-3
March 2008
Moisture Probe Considerations (cont.)
3. Static or Dynamic Use: the sensor performs equally well in still air or where considerable flow occurs. Its small size makes it ideal for measuring moisture conditions within completely sealed containers or dry boxes. It also performs well under gas flow linear velocities as high as 10,000 cm/sec, and liquid flow linear velocities to 10 cm/sec. Refer to Tables
A-2 and A-3 in Appendix A for maximum gas and liquid flow rates.
4. Pressure: the moisture probe always senses the existing water vapor pressure regardless of the total ambient pressure. The moisture sensor measures water vapor under vacuum or high pressure conditions from as little as a few microns Hg to as high as 5000 psi total pressure.
5. Long-Term Storage & Operational Stability: continuous abrupt humidity changes do not affect the sensor, and exposure to saturation conditions does not damage the sensor, even when stored.
6. Freedom from Interference: the presence of a wide variety of gases or organic liquids does not affect the sensor. Also, large concentrations of hydrocarbon gases, Freon®, carbon dioxide, carbon monoxide, and hydrogen do not affect sensor water vapor indications. The sensor functions properly in a multitude of gaseous or non-conductive liquid environments.
7. Corrosive Materials: avoid all materials that are corrosive or otherwise damaging to aluminum or aluminum oxide. These include strongly acidic or basic materials and primary amines.
Freon is a registered trademark of E.I. du Pont de Nemours and Company.
2-4 Installing the Series 35 IS
March 2008
Sample System Guidelines
A sample system, although not mandatory, is highly recommended for moisture measurement. The purpose of a sample system is to condition or control a sample stream to within the specifications of the probe. The application requirements determine the design of the sample system. GE
Sensing applications engineers can make recommendations based on the following general guidelines:
•
Typically, keep the sample system simple. Include as few components as possible and locate all or most of the components downstream of the measurement probe. A simple sample system consists of an explosion-proof housing with a sample cell, a filter, and inlet and outlet needle valves.
•
Do not use any material for sample system components that can affect measurements. A sample system may include a filter to remove particulates from the sample stream and/or a pressure regulator to reduce or control the pressure of the stream. However, most common filters and pressure regulators are not suitable for sample systems because their wetted parts may absorb or release components (such as moisture) into the sample stream. They may also allow ambient contamination to enter the sample stream. In general, use stainless-steel for all wetted parts.
Note: The actual sample system design depends on the
application requirements.
Installing the Series 35 IS 2-5
March 2008
Mounting an External Sample System
Note: If you have a Series 35 IS with a self-contained sample
system, disregard this section and proceed to The Self-
Contained Sample System on page 2-7.
The sample system is usually fastened to a metal plate that has four mounting holes. GE Sensing can provide the sample system in an enclosure, if requested. Sample system outline and dimension drawings are included in your shipment, if you ordered them.
Use the following steps to mount the external sample system and connect it to the process:
1. Fasten the sample system plate or enclosure to a solid stable surface with bolts and washers in each of its four corners.
2. Connect the process supply and return lines to the sample system inlet and outlet using the required stainless-steel fittings and tubing.
Caution!
Do not start a flow through the system until the probe has been properly installed.
Note: For common applications, GE Sensing can provide a
standard sample system. If applicable, additional instructions on how to start up and shut down the sample system may be included.
2-6 Installing the Series 35 IS
March 2008
The Self-Contained Sample System
The Portable Series 35 IS can include a self-contained sample system as shown in Appendix B. This sample system is constructed with 1/4" Swagelok® inlet and outlet compression fittings and 1/4" stainless-steel tubing.
The sample system also includes inlet and outlet needle valves and a 7-micron sintered stainless-steel filter with a replaceable element. All wetted parts are Series 300 stainless steel, Viton®, or Teflon®. The maximum allowable pressure for the sample
system is 4500 psig. See Figure 2-1 below.
Probe Outlet
Sample Cell
Inlet
Figure 2-1: Self-Contained Sample System with Probe
Swagelok is a registered trademark of Crawford Fitting Company.
Teflon is a registered trademark of E.I. du Pont de Nemours and Company, Inc.
Viton is a registered trademark of DuPont Dow Elastomers, L.L.C.
Installing the Series 35 IS 2-7
March 2008
Installing a Probe into a Sample System
The sample system design protects the aluminum oxide moisture probe from particulates and is recommended for process gases such as dry instrument/plant air or other similar process gas.
Other sample handling equipment (pressure regulators, cooling coils, additional filters, pressure gauges, rota-meters, etc.) may be required for some applications. Contact GE Sensing or see
Appendix A for general information about using the GE Sensing
Aluminum Oxide Moisture Probe. Proceed to the appropriate probe installation section for your system.
Caution!
If you mount the probe directly into the process line, consult GE Sensing for proper installation instructions and precautions.
External Sample System
Use the following steps to install the probe into the external
sample cell. Refer to Figure 2-2 below.
Probe
Inlet
Sample Cell
Outlet
2-8
Figure 2-2: A Typical Probe Installation
Installing the Series 35 IS
March 2008
External Sample System (cont.)
1. Insert the probe into the sample cell so that it is perpendicular to the sample inlet.
IMPORTANT:
For maximum protection of the aluminum oxide sensor, always leave the shield in place.
2. Screw the probe into the receptacle fitting, making sure not to cross the threads.
Note: M2 probes have 3/4-16 straight threads with an o-ring
seal for installation either into the sample system or directly into the process line. Other mounts are available for special applications.
3. Hand-tighten the probe in a clockwise direction, then tighten it an additional 1/8 turn with a wrench.
Figure 2-2 on page 2-8 shows a typical probe installation with the
probe mounted into a sample cell.
Installing the Series 35 IS 2-9
March 2008
Self-Contained Sample System
Use the following steps to install the probe into the sample cell of a self-contained sample system:
1. Loosen the two quarter-turn screws on the back of the
Series 35 IS cover.
2. Remove the cover by gently lifting it off the unit from back to front. Be careful not to pull on the cables that connect to the cover.
3. Insert the probe into the sample cell.
4. Screw the probe into the receptacle fitting, making sure not to cross the threads.
Note: Insert the probe into the cylindrical shaped container
called the sample cell, which is included as part of your
sample system. M2 probes have 3/4-16 straight threads and an o-ring seal for installation of the probe into the
standard sample cell (see Figure 2-1 on page 2-7).
5. Hand-tighten the probe in a clockwise direction, then tighten it an additional 1/8 turn with a wrench.
6. Connect the blue probe cable to the probe by inserting the bayonet-type connector onto the probe and twisting the shell clockwise until it snaps into a locked position.
Note: The connector on the probe cable must be rotated until it
aligns with the pins in the probe connector, before the cable may be properly inserted into the probe and secured.
7. Close the Series 35 IS cover and tighten the two quarter-turn screws on the back of the cover.
2-10 Installing the Series 35 IS
March 2008
Making External Probe Connections
Probe connections may differ, depending on the type of sample system you are using. Connect the moisture probe using a continuous run of GE Sensing two-wire shielded cable (see
Figure 2-3 below).
Red
Shield
Green
Figure 2-3: Two-Wire Shielded Cable
Protect all cables from excessive strain (bending, pulling, etc.).
Do not subject cables to temperatures above +65°C (149°F) or below –50°C (–58°F). You can order standard cable assemblies
(including connectors) from GE Sensing in any length up to 100 meters (330 feet).
Use the following steps to make probe connections:
Note: Probe connections for an internal sample system have
been described in the previous section. The following instructions are for an external sample system only.
1. Turn off the Series 35 IS.
2. Connect the cable to the probe by inserting the bayonet-type connector onto the probe and twisting the shell clockwise until it snaps into a locked position.
Installing the Series 35 IS 2-11
March 2008
Making External Probe Connections (cont.)
Note: The connector on the probe cable must be rotated until it
aligns with the pins in the probe connector, before the cable may be properly inserted into the probe and secured.
3. Connect the other end of the probe cable to the terminal block
on the Series 35 IS side panel (see Table 2-1 below and
IMPORTANT:
To maintain good contact at each terminal block and to avoid damaging the pins on the connector, pull the connector straight off (not at an angle), make cable connections while the connector is away from the unit, and push the connector straight on (not at an angle) when the wiring is complete.
Caution!
The M Series probe may not be capable of withstanding the 500 V insulation test required by clause 5.7 of EN50 020 when installed in the process media. This must be taken into account in any installation in which it is used.
(See Cert. #Ex95C2002X in its entirety.)
Table 2-1: Probe Connections
Connect:
To PROBE
Terminal Block
Red (H2) wire
Shield
Green (H1) wire pin 1 pin 2 pin 3
2-12 Installing the Series 35 IS
March 2008
Figure 2-4: Probe Terminal Block Location
Installing the Series 35 IS 2-13
March 2008
Using Modified or Non-GE Sensing Cables
In some installations, it may be desirable to use a pre-existing moisture cable or to modify the standard GE Sensing moisture cable. If this approach is taken, it is essential that the following precautions be observed. Furthermore, after connecting the moisture probe, a calibration adjustment must be performed to compensate for any electrical offsets introduced by the nonstandard cable.
Caution!
GE Sensing cannot guarantee operation of the Series 35 IS to the specified accuracy unless a standard GE Sensing hygrometer cable is used.
1. Use cable that matches the electrical characteristics of GE
Sensing cable (contact the factory for specific information on cable characteristics).
2. If possible, avoid all splices. Splices will impair the performance of the instrument. When possible, instead of splicing, coil the excess cable.
3. If you must splice cables, be sure the splice introduces minimum resistive leakage or capacitive coupling between conductors.
4. Carry the shield through any splice. A common mistake is to not connect the shields over the splice. If you are modifying a
GE Sensing cable, the shield will not be accessible without cutting back the cable insulation. Also, do not ground the shield at both ends. The shield should be grounded only at the
hygrometer end, as described in Table 2-1 on page 2-12.
2-14 Installing the Series 35 IS
March 2008
Performing a Calibration Adjustment
After completing the installation of a modified or non-standard moisture cable, it is necessary to perform a calibration adjustment on the Series 35 IS. This procedure will adjust the electronics to compensate for any error or offset introduced by splices or long cable lengths. It is also recommended that a calibration be performed after the installation of a new GE Sensing cable, to verify the accuracy of the unit.
Use the following steps to perform a calibration adjustment:
1. Power up the Series 35 IS.
2. To check the new cable, set up the screen to display “MH”.
See Chapter 3, Operating the Series 35 IS, for instructions on how to navigate through the user program.
3. Note the high, low and zero reference values that are recorded on the sticker located on the inside of the Series 35 IS chassis.
4. Disconnect the cable from the probe only (leave the cable connected to the Series 35 IS), and verify that the displayed
MH value equals the zero reference value within ±0.0003 MH.
If this reading is within specifications, no further testing is necessary.
5. If the zero reference value reading above was out of range, the low reference value must be adjusted. For readings more than
0.0003 below the previously recorded zero reference value, add the difference to the low reference value. For readings more than 0.0003 above the previously recorded zero reference value, subtract the difference from the low reference value. Record the final corrected low reference value.
6. Reprogram the Series 35 Is with the new, corrected low reference value. See Entering High and Low Reference Values in Chapter 3.
7. Make sure the cable is still disconnected from the probe but is connected to the Series 35 IS.
Installing the Series 35 IS 2-15
March 2008
Performing a Calibration Adjustment (cont.)
8. Recheck the zero reference reading and verify that it is now within ±0.0003 MH of the previously recorded value.
9. Fill out a new reference sticker with the final low reference value and/or record the information on the Data Information
Sheet in Appendix D. Make sure the following information is included:
•
HIGH REF = Original Value
•
LOW REF = New Corrected Value
•
ZERO REF = Original Recorded Value
10.Reconnect the cable to the probe.
Note: If the cable is changed in any way, repeat the above
procedure for maximum accuracy of the unit.
The Series 35 IS is now ready for operation. Proceed to
Chapter 3, Operating the Series 35 IS.
2-16 Installing the Series 35 IS
Operating the Series 35 IS
March 2008
Chapter 3
Your factory-programmed Series 35 IS can begin taking measurements as soon as you turn the power on. The user program enables you to enter and change setup information.
Note: The Series 35 IS suspends taking measurements when you
enter the user program listed below. Refer to the menu map in Appendix C to navigate the user program.
This section explains the following procedures:
•
•
Using the Keypad to Enter and Change Data [page 3-5]
•
Verifying and Changing Factory Setup Data [page 3-9]
•
Changing the Measurement Display [page 3-14]
•
Entering an Offset Value for Dew/Frost Point [page 3-15]
•
Entering a Constant Pressure Value [page 3-16]
•
Entering a PPMv Constant Multiplier [page 3-18]
•
Entering an Interval for Automatic Calibration [page 3-19]
•
Viewing the Battery Status [page 3-21]
•
Entering a Backlight-On Time Interval [page 3-22]
•
Setting Up Computer-Enhanced Response [page 3-23]
•
Setting Up Range Error Processing [page 3-25]
•
Selecting AC Line Frequency (60/50 Hz) [page 3-26]
•
Adjusting the Contrast [page 3-27]
•
Setting the Clock Values [page 3-28]
•
Setting Up the Data Logger [page 3-30]
Operating the Series 35 IS 3-1
March 2008
Getting Started
Your unit is completely set up at the factory. To begin taking measurements, you only need to apply power to the unit and start up the sample system. Complete the instructions in the following sections to get your unit up and running:
•
Starting Up and Shutting Down the Sample System
•
Powering up the Series 35 IS
3-2 Operating the Series 35 IS
March 2008
Starting Up and Shutting Down the Sample System
If your Series 35 IS has a self-contained sample system, use this section to properly start up the sample system before you begin taking measurements, and to shut down the sample system when you have finished taking measurements.
Note: If you are using an external sample system, refer to the
separate manual supplied with your order.
Starting Up the Sample System
In order to operate the self-contained sample system correctly, first determine if you need to make measurements at atmospheric pressure or at process pressure. After determining the pressure, use the appropriate section that follows to start up and shut down the sample system.
Note: The maximum allowable pressure for the self-contained
sample system for the Series 35 IS is 4500 psig.
At Atmospheric Pressure:
1. Begin with the Series 35 IS inlet and outlet valves in the closed position, then open the outlet valve completely.
2. Open the inlet valve slightly to allow only a small amount of flow from the process.
At Process Pressure:
1. Begin with the Series 35 IS inlet and outlet valves in the closed position, then slowly open the inlet valve fully.
2. Open the outlet valve slightly to allow only a small amount of flow.
Operating the Series 35 IS 3-3
March 2008
Shutting Down the Sample System
At Atmospheric or Process Pressure:
1. Close the inlet valve completely.
2. Wait for the flow from the outlet valve to stop.
3. Close the outlet valve to keep the probe sensor from being affected by ambient moisture.
Powering Up the Series 35 IS
To turn on the Series 35 IS, press the POWER switch (located on the left side of the front panel) to the ON position. (If the battery is depleted, see Chapter 4, Troubleshooting, to recharge the battery.)
The Series 35 IS displays “Loading. . . .”, followed by a series of additional screens. This indicates that the system is loading calibration and reference data. After this data is loaded, the Series
35 IS automatically calibrates (Autocal) the moisture circuitry, and then begins taking measurements.
3-4 Operating the Series 35 IS
March 2008
Using the Keypad to Enter and Change Data
The Series 35 IS user program enables you to change factory setup data and user-defined program functions.
Use the appropriate sections that follow to learn how to:
•
Use the programming keys
•
Enter and exit the user program
•
Move through the user program
•
Enter numeric data
How to Use the Programming Keys
The membrane keypad to the right of the display window on the front panel of the Series 35 IS contains six (6) keys:
•
ENTER - Use this key to: enter a menu option; switch from viewing to editing data; move to the next digit position during numeric entry; confirm an entry.
•
ESC - Use this key to: cancel an entry; move back up one level in the menu sequence.
•
Δ - This key scrolls upward through the menu options and increases the value during numeric entry.
•
∇
- This key scrolls downward through the menu options and decreases the value during numeric entry.
•
HYGRO - Press this key in measurement mode to display the dew/frost point temperature in the selected units.
•
P const
- Press this key in measurement mode to display the pressure in the selected units.
Operating the Series 35 IS 3-5
March 2008
How to Enter and Exit the User Program
To enter programming mode, perform the following sequence within five (5) seconds, or the Series 35 IS will time out and return to measurement mode.
P: +0.00 Kcmg
While in measurement mode, press the [ESC] key.
ESC
PROGRAM MENU
Within 5 seconds of the appearance of this display, press the [ENTER] key followed by the [ESC] key.
This will display for 1 second.
DP RANGE You are now in the user program menu.
From the main menu, use the arrow keys to scroll through the options to your selection. Use the [ENTER] key to confirm your selection or the [ESC] key to cancel your selection.
To exit the user program, press the
[ESC]
key until
RUN ?
displays, then press the [ENTER] key to return to displaying measurements.
DP RANGE
RUN?
Exiting......
P: +0.00 Kcmg
Press the [ESC] key.
Press the [ENTER] key.
This will display for 1 second.
You have now returned to measurement mode.
3-6 Operating the Series 35 IS
March 2008
How to Move Through the User Program
Use the arrow keys to scroll through the following menu options.
Refer to the Menu Map (Figure C-1) in Appendix C as a guide, while moving through the user program.
Note: While in the programming mode, the Series 35 IS
suspends taking measurements.
•
DP RANGE - Use to enter high and low dew points for the calibration curve. This input determines the number of points in the Moisture Probe Calibration Data entry, and shows the maximum/minimum dew point values in degrees celsius (°C), as shown on the Moisture Probe Calibration Data Sheet.
•
MH CURVE - Use to enter a value for each point in the calibration curve. This Moisture Probe Calibration Data Entry accepts only MH values. The dew point automatically increases at 10°C intervals.
•
SERIAL # - Input the moisture probe serial number.
•
HYGRO REFERENCE - Use to enter high and low reference values for moisture measurement.
•
USER - Use to set up user-defined program functions, such as offset value, constant pressure, PPMv multiplier, Autocal interval, battery, backlight interval, computer enhanced response (optional), range error handling, selecting 60/50 Hz, contrast, and clock.
•
FACTORY SETUP - Used for GE Sensing field service and factory use only. A passcode is required to access this function.
•
LOGGER - This optional section lets you program the unit to log data at pre-selected time intervals and access the logged data for viewing or for sending to a dump terminal. It includes log interval, log setup, log display, and log dump settings.
Operating the Series 35 IS 3-7
March 2008
How to Enter Numeric Data
To enter numbers one digit at a time, use the arrow keys to scroll to the desired number, then press the [ENTER] key to move to the next digit position. Repeat this procedure until all numbers have been entered.
Note: In the programming mode, pressing an arrow key lets you
enter the edit mode. Pressing
[ESC]
terminates the edit mode without changing any data. In edit mode, check all characters before pressing the
[ENTER]
key and moving to the next digit position.
3-8 Operating the Series 35 IS
March 2008
Verifying and Changing Factory Setup Data
Use this section to confirm, or make any necessary changes to, the factory setup data. The setup data includes information regarding the moisture probe as listed below:
•
Dew point range
•
Calibration data
•
Probe serial number
•
High and low reference values
Note: Remember to record all setup data in Appendix D, Data
Information Sheet, in the back of this manual.
The factory enters all the necessary data into your Series 35 IS.
Therefore, you should not need to re-enter this data unless you
•
Purchase a new probe
•
Recalibrate an existing probe
•
Have reason to believe the data is corrupted.
Use the appropriate sections that follow to verify and/or change the setup data.
Operating the Series 35 IS 3-9
March 2008
Entering the Dew Point Range
Use the
DP RANGE
to enter high and low dew point values. The
Series 35 IS uses this range to determine the number of points in the calibration curve.
Note: Enter the user program as described on page 3-6.
DP RANGE
Hi DP
Use the arrow keys to scroll to
DP RANGE
and press the
[ENTER]
key.
Scroll to this screen with the arrow keys and press the
[ENTER]
key.
Hi DP +20°C Use the arrow keys to scroll to the desired dew point value, then press the [ENTER] key.
Note: The high and low dew points are listed on the Moisture
Probe Calibration Data Sheet located in the probe box.
The default values are: Low = -110°C, High = +20°C.
Hi DP Press the down arrow key.
Low DP
Press the [ENTER] key.
Low DP -110°C
Use the arrow keys to scroll to the desired value. Then, press the [ENTER] key.
Low DP
Press [ESC] to return to the main menu, and proceed to the next section.
IMPORTANT:
If you change the dew point settings, you must change the calibration curve also.
3-10 Operating the Series 35 IS
March 2008
Entering Calibration Data
After entering the high and low dew point values, use the
MH
CURVE option to enter calibration data for the moisture probe in accordance with the following procedure.
Note: Remember to record all set-up data in Appendix D, Data
Information Sheet, in the back of this manual.
Moisture probe calibration data is taken at fixed dew point values in 10°C intervals. After the high and low dew point values have been entered, the appropriate number of data points for the moisture probe is automatically determined. Starting with the minimum dew point value, the MH values must be entered in
10°C increments until the maximum dew point is reached.
Note: Changing the dew point setting affects the MH curve.
MH CURVE Use the arrow keys to scroll to
MH CURVE
and press the
[ENTER]
key.
-110°C MH 0.1890
Press the arrow keys to scroll through the
MH
values. Press
[ENTER] then the arrow keys to edit a value. ( ED displays in edit mode and MH displays in view mode.) Using the arrow keys and the
[ENTER]
key, edit the value one digit at a time, and then press [ENTER] .
Note: To abort the editing function at any time without changing
the value, press the
[ESC]
key.
Repeat the above procedure for each point in the MH Curve.
Then, press [ESC] until you return to main menu and proceed to the next section.
Operating the Series 35 IS 3-11
March 2008
Entering the Probe Serial Number
The probe serial number is listed on the Calibration Data Sheet and is inscribed on the hex nut of the moisture probe.
Note: Remember to record all set up data in Appendix D, Data
Information Sheet, in the back of this manual.
USER
SERIAL # xxxxxx S/N
From the main menu, use the arrow keys to scroll to the
USER prompt, then press the
[ENTER] key to view the options.
Use the arrow keys to scroll to
SERIAL # and press the
[ENTER] key.
The unit displays the current serial number.
123456 S/N To edit the serial number, press an arrow key and a blinking cursor appears at the left-most digit. Use the arrow keys to scroll to the desired value and press
[ENTER]
to move the cursor to the next digit. Repeat until you have entered all six serial number digits. Press the
[ENTER] key to confirm the change or the
[ESC]
key to cancel the change.
SERIAL # Press the
[ESC]
key to return to the main menu.
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press
[ESC]
until the RUN ?
prompt appears and then press the [ENTER] key.
3-12 Operating the Series 35 IS
March 2008
Entering High and Low Reference Values
The Series 35 IS requires high and low reference values for its moisture measurement circuitry. These factory calibration values, specific to each unit, are listed on a label inside the unit.
IMPORTANT:
Do not change the high and/or low reference values unless instructed to do so by GE Sensing.
Note: Remember to record all setup data in Appendix D, Data
Information Sheet, in the back of this manual.
HYGRO REFERENCE
High REF
Use the arrow keys to scroll to
HYGRO REFERENCE and press the [ENTER] key.
Press either arrow key to scroll to the High REF option, then press the
[ENTER]
key.
HX.XXXX
Use the arrow and [ENTER] keys to change the value one digit at a time, then press
[ENTER]
.
Note: To abort the editing function at any time without changing
the value, press the
[ESC]
key.
Low REF
Press either arrow key to scroll to the Low REF option, then press the [ENTER] key.
LX.XXXX
Use the arrow and [ENTER] keys to change the value one digit at a time, then press
[ENTER]
.
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press [ESC] until the RUN ?
prompt appears and then press the [ENTER] key.
Operating the Series 35 IS 3-13
March 2008
Changing the Measurement Display
The front panel of the Series 35 IS contains two keys that let you quickly change the L.C.D. display mode during operation of the unit. These display keys are labeled HYGRO and P const
.
•
HYGRO - Press this key to display and scroll through all the available moisture measurement units (DP/°C, DP/°F, PMv,
MH).
•
P const
- Press this key to display and scroll through all available constant pressure units (PSG, Bar, KPAg, Kcmg).
After you select a specific display mode, that display remains on the screen until you choose another display mode or until you enter the programming mode. Upon exiting the programming mode and returning to measurement mode, the screen automatically returns to the previously selected display mode.
3-14 Operating the Series 35 IS
March 2008
Entering an Offset Value for Dew/Frost Point
Use this option to adjust the displayed dew/frost point reading. A positive number increases the reading and a negative number decreases the reading. The offset value always displays in dew/ frost point degrees C.
Note: Enter the user program as described on page 3-6.
USER
OFFSET
OFFSET +5.0 °C
From the main menu, use the arrow keys to scroll to the
USER prompt, then press the [ENTER] key to view the options.
Use the arrow keys to scroll to
OFFSET and press the [ENTER] key.
The current offset value is displayed.
OFFSET +10.0 °C Press either arrow key to delete the current value and enter the edit mode. Use the arrow and
[ENTER] keys to change the value one digit at a time, then press the
[ENTER]
key.
Note: To abort the editing function at any time without changing
the value, press the
[ESC]
key.
OFFSET
Press the [ESC] the main menu.
key to return to
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press
[ESC]
until the
RUN ?
prompt appears and then press the
[ENTER]
key.
Operating the Series 35 IS 3-15
March 2008
Entering a Constant Pressure Value
This option lets you enter a fixed value for the pressure of the sample gas at the moisture probe location. The value is used to calculate the moisture content in PPM v
.
Note: Enter the user program as described on page 3-6.
USER
From the main menu, use the arrow keys to scroll to the USER prompt, then press the [ENTER] key to view the options.
CONSTANT PRESSUR
Use the arrow keys to scroll to
CONSTANT PRESSUR
and press the
[ENTER]
key.
KP PSIg
The current units choice is displayed. Use the arrow keys to scroll to the desired pressure units, then press the [ENTER] key. See Table 3-1 below for the available choices.
Table 3-1: Constant Pressure Units
Available Units Description of Units
KP psig
KP bar
KP KPag
KP Kcmg pounds per square inch gauge bars absolute kilopascals gauge kilograms per square centimeter gauge
KP: +XXX.XX PSG The current constant pressure value is displayed.
Note: To abort the editing function at any time without changing
the value, press the
[ESC]
key.
3-16 Operating the Series 35 IS
Entering a Constant Pressure Value (cont.)
March 2008
KP: +500.00
KP:
PSG
+XXX.XX PSG
Press either arrow key to delete the current value and enter edit mode. Use the arrow and
[ENTER]
keys to change the value one digit at a time, then press [ENTER] .
The new constant pressure value is displayed. Press [ESC] .
KP PSIg
CONSTANT PRESSUR
The current units choice is displayed. Press
[ESC]
to return to the options menu.
Press the
[ESC]
key to return to the main menu.
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press [ESC] until the RUN ?
prompt appears and then press the [ENTER] key.
Operating the Series 35 IS 3-17
March 2008
Entering a PPMv Constant Multiplier
Use this option to apply a user-defined constant multiplier to the the PPMv value.
Note: Enter the user program as described on page 3-6.
USER
From the main menu, use the arrow keys to scroll to the USER prompt, then press the
[ENTER] key to view the options.
PPMv MULTIPLIER
Use the arrow keys to scroll to
PPMv MULTIPLIER and press the
[ENTER]
key.
KPPMv 1.0000
The current value is displayed.
Press either arrow key to delete the current value and enter the edit mode.
KPPMv X.XXXX
Use the arrow and [ENTER] keys to enter the new multiplier, one digit at a time. Then, press
[ENTER]
twice to return to the options menu.
PPMv MULTIPLIER
Press the [ESC] key to return to the main menu.
Note: PPMv constant multiplier values up to 999.9999 may be
entered, but the value should be set at 1.0000 unless otherwise advised by GE Sensing. Be sure to record all
entered data in Appendix D, Data Information Sheet, in
the back of this manual.
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press
[ESC]
until the RUN ?
prompt appears and then press the [ENTER] key.
3-18 Operating the Series 35 IS
March 2008
Entering an Interval for Automatic Calibration
The Series 35 IS automatically calibrates itself (Autocal) at user- programmable time intervals, in order to compensate for any drift in the electronics. An Autocal interval value between 0 and 1440 minutes (24 hours) may be entered. GE Sensing recommends setting the Autocal interval to eight hours (480 minutes). If the
Series 35 IS is exposed to extreme temperatures or weather conditions, a shorter Autocal interval may be more appropriate.
Note: Enter the user program as described on page 3-6.
USER
From the main menu, use the arrow keys to scroll to the USER prompt, then press the [ENTER] key to view the options.
AUTOCAL INTERVAL
ACAL (Mins) 30
Use the arrow keys to scroll to
AUTOCAL INTERVAL
and press the
[ENTER]
key.
The current Autocal interval value is displayed.
ACAL (Mins)
AUTOCAL INTERVAL
XX
Press either arrow key to delete the current value and enter the edit mode. Use the arrow and
[ENTER]
keys to change the value one digit at a time, then press [ENTER] to return to the options menu.
Press the [ESC] key to return to the main menu.
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press
[ESC]
until the
RUN ?
prompt appears and then press the
[ENTER]
key.
Operating the Series 35 IS 3-19
March 2008
Entering an Interval for Automatic Calibration
(cont.)
After you confirm the Autocal interval and return to the main menu, the Series 35 IS immediately performs an Autocal. The next time Autocal occurs depends on the length of the time interval setting.
The Series 35 IS determines the times of subsequent Autocals by establishing a fixed schedule, beginning at midnight, using the specified interval. For example, if you enter a 90-minute time interval, Autocal occurs 16 times per day
(1 day = 1440 minutes ÷ 90 minutes = 16).
The following fixed schedule applies:
1. 1:30 a.m.
2. 3:00 a.m.
3. 4:30 a.m.
4. 6:00 a.m.
5. 7:30 a.m.
9. 1:30 p.m.
10. 3:00 p.m.
11. 4:30 p.m.
12. 6:00 p.m.
13. 7:30 p.m.
6. 9:00 a.m.
7. 10:30 a.m.
14. 9:00 p.m.
15. 10:30 p.m.
8. 12:00 p.m.(noon) 16. 12:00 a.m.(midnight)
If you enter a time interval not evenly divisible into 1440 minutes, the Series 35 IS rounds up to the next acceptable interval. For example, if you set the 90-minute Autocal interval at
6:10 p.m., the next Autocal occurs at 7:30 p.m. (excluding the
Autocal performed when you exit the Autocal menu).
3-20 Operating the Series 35 IS
March 2008
Viewing the Battery Status
Make sure that the battery charger is disconnected before viewing the battery status.
Note: Enter the user program as described on page 3-6.
USER
BATTERY
DISCONNECT A.C.
Press ENTER
BAT: +6.89 V
BAT: 100%
BATTERY
From the main menu, use the arrow keys to scroll to the USER prompt, then press the
[ENTER] key to view the options.
Use the arrow keys to scroll to the
BATTERY
option and press the
[ENTER]
key.
The display alternates between the two messages shown. Press the
[ENTER]
key when the AC power has been disconnected.
The display alternates between the two messages shown. Press the [ESC] key to return to the options menu.
Press the [ESC] key to return to the main menu.
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press
[ESC]
until the
RUN ?
prompt appears and then press the
[ENTER]
key.
Operating the Series 35 IS 3-21
March 2008
Entering a Backlight-On Time Interval
Your Series 35 IS is equipped with a backlight, which can be programmed to turn off automatically after a pre-determined time. Values between 0 and 1440 minutes (24 hours) may be entered.
Note: Enter the user program as described on page 3-6.
USER
BACKLIGHT
BLITE (Mins) 30
From the main menu, use the arrow keys to scroll to the
USER prompt, then press the
[ENTER] key to view the options.
Use the arrow keys to scroll to
BACKLIGHT
and press the
[ENTER]
key.
The current backlight interval value is displayed.
BLITE (Mins) XX
Press either arrow key to delete the current value and enter the edit mode. Use the arrow and
[ENTER]
keys to change the value one digit at a time, then press
[ENTER]
to return to the options menu.
BACKLIGHT Press the
[ESC]
key to return to the main menu.
Note: Constant use of the backlight shortens the battery life. All
units are shipped with a default backlight-on time interval of 5 minutes.
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press
[ESC]
until the
RUN ?
prompt appears and then press the
[ENTER]
key.
3-22 Operating the Series 35 IS
March 2008
Setting Up Computer-Enhanced Response
Note: This option may not be installed on your Series 35 IS.
Using a dynamic moisture calibration technique, Computer-
Enhanced Response extrapolates the moisture level to the end point, when making measurements in abrupt “dry down” conditions. Response time depends on the relative change in dew point. For example, the Series 35 IS can respond in three to five minutes to reflect a change from ambient moisture to trace levels.
The accuracy of Computer-Enhanced Response equals ±2°C of the sensor reading, assuming an equilibrium with the gas. To ensure accuracy, your application should:
•
Have a reasonably constant flow rate and final dew point.
•
Maintain a minimum flow rate of one standard cubic foot per hour (SCFH).
•
Operate at atmospheric pressure (0 psig)
•
The actual moisture content must be above –85°C frost point and must be stable.
You can enable or disable the Computer-Enhanced Response feature. If the feature is not available, the display reads Option
Not Avail . Otherwise, the Enhance Response display appears.
Note: Enter the user program as described on page 3-6.
USER
ENHANCE RESPONSE
Operating the Series 35 IS
From the main menu, use the arrow keys to scroll to the
USER prompt, then press the
[ENTER] key to view the options.
Use the arrow keys to scroll to the ENHANCE RESPONSE option and press the [ENTER] key.
3-23
March 2008
Setting Up Computer-Enhanced Response (cont.)
ENHANCE OFF
ENHANCE ON
Use the arrow keys to scroll to one of the option screens shown and press the
[ENTER]
key.
ENHANCE RESPONSE Press the
[ESC]
key to return to the main menu.
Note: An activated Computer Enhanced Response displays a
reverse video “E” on the left side of the display. After determining the final value, the reverse video “E” changes to a regular “E”.
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press [ESC] until the RUN ?
prompt appears and then press the [ENTER] key.
3-24 Operating the Series 35 IS
March 2008
Setting Up Range Error Processing
Range errors occur when an input signal within the capacity of the analyzer is outside the range of the probe calibration data. The
Series 35 IS displays Range Errors with an OVER RANGE or
UNDER RANGE message. The error condition extends to all displayed measurements of that mode. For example, if dew point displays
OVER RANGE
, then moisture in PPMv also displays
OVER RANGE
.
There are two available Range Error processing options:
•
No Action - the range error is not displayed.
•
Display - the range error is displayed.
Note: Enter the user program as described on page 3-6.
USER
RANGE ERROR
R_ERR = Display
R_ERR = No Action
From the main menu, use the arrow keys to scroll to the USER prompt, then press the
[ENTER] key to view the options.
Use the arrow keys to scroll to the
RANGE ERROR
option and press the
[ENTER]
key.
Use the arrow keys to scroll to one of the options shown and press the
[ENTER]
key.
RANGE ERROR Press the
[ESC]
key to return to the main menu.
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press
[ESC]
until the
RUN ?
prompt appears and then press the
[ENTER]
key.
Operating the Series 35 IS 3-25
March 2008
Selecting AC Line Frequency (60/50 Hz)
Select this setting according to the AC line frequency in the country where the Series 35 IS is installed.
Note: Enter the user program as described on page 3-6.
USER
SELECT 60/50 HZ
60HZ
50HZ
From the main menu, use the arrow keys to scroll to the USER prompt, then press the
[ENTER] key to view the options.
Use the arrow keys to scroll to the
SELECT 60/50 HZ
option and press the
[ENTER]
key.
Use the arrow keys to scroll to one of the options shown and press the
[ENTER]
key.
SELECT 60/50 HZ Press the
[ESC]
key to return to the main menu.
Note: For reference, record all entered data in Appendix D,
Data Information Sheet, in the back of this manual.
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press [ESC] until the RUN ?
prompt appears and then press the [ENTER] key.
3-26 Operating the Series 35 IS
March 2008
Adjusting the Contrast
Use this option to adapt to the lighting conditions of the Series 35
IS installation.
Note: Enter the user program as described on page 3-6.
USER
CONTRAST
Adj. Contrast
CONTRAST
From the main menu, use the arrow keys to scroll to the USER prompt, then press the
[ENTER] key to view the options.
Use the arrow keys to scroll to the
CONTRAST
option and press the
[ENTER]
key.
Press the up-arrow key to increase contrast or the downarrow key to decrease, contrast.
Hold the key down for large changes or press repeatedly for small changes. Press the
[ENTER]
key to confirm the new setting.
Press the
[ESC]
key to return to the main menu.
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press [ESC] until the
RUN ?
prompt appears and then press the
[ENTER]
key.
Operating the Series 35 IS 3-27
March 2008
Setting the Clock Values
The Series 35 IS has no backup power source, and the time/date must be reset whenever the battery has been totally depleted.
Note: Enter the user program as described on page 3-6.
USER
From the main menu, use the arrow keys to scroll to the USER prompt, then press the
[ENTER] key to view the options.
CLOCK
Use the arrow keys to scroll to the CLOCK option and press the
[ENTER]
key.
Read Time
Use the arrow keys to scroll to
READ TIME then press the
[ENTER] key.
hh:mm mm/dd/yy
The current time and date are displayed for 2 seconds, before the CLOCK prompt returns.
To edit the time, use the following steps:
CLOCK
Set Time hh:mm
Set Time
3-28
Press the [ENTER] key.
Use the arrow keys to scroll to
SET TIME
. Then, press the
[ENTER] key.
Use the arrow and
[ENTER]
keys to scroll to the correct hour and minutes values. Then, press the
[ENTER] key.
Press
[ESC]
to return to the
CLOCK
prompt.
Operating the Series 35 IS
March 2008
Setting the Clock Values (cont.)
To edit the date, use the following steps:
CLOCK Press the
[ENTER]
key.
Set Date
Use the arrow keys to scroll to
SET DATE . Then, press the
[ENTER]
key.
MM/DD/YY
Use the arrow and [ENTER] keys to scroll to the correct month, day and year values. Then, press the
[ENTER]
key.
Note: The Series 35 IS clock is year 2000 compliant.
Set Date
CLOCK
Press [ESC] to return to the
CLOCK prompt.
Press the
[ESC]
key to return to the main menu.
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press
[ESC]
until the
RUN ?
prompt appears and then press the
[ENTER]
key.
Operating the Series 35 IS 3-29
March 2008
Setting Up the Data Logger
The Series 35 IS optional data logging feature enables you to internally view and store measurement data. The main menu displays the LOGGER prompt only if this option is available on your unit.
Data is logged at user-selected time intervals and stored in the unit’s FRAM. Up to three different parameters may be logged at any one time.
Note: Because the Series 35 IS contains a fixed amount of
memory, the more parameters you select and the more frequently you log data, the shorter the total available logging time.
The Series 35 IS assigns a record and a header to all logged data.
The record consists of logged measurement values. The header includes the time the log started, the constant pressure value, the log time intervals, and the selected measurement modes. By assigning each log a header, the Series 35 IS distinguishes one log from another.
The Series 35 IS can store more than one log. Each time the log setting is changed, the Series 35 IS queries if you want to overwrite or append to the existing log(s). If overwrite mode is chosen, the Series 35 IS writes over the logs currently stored in memory. If append mode is chosen, the Series 35 IS adds new log data to the end of the previously stored log data. In either case, the Series 35 IS continues logging measurement data until it runs out of memory.
Note: Before beginning setup of the data logger, check to make
sure that the clock is set to the correct time and date. See
“Setting the Clock Values” on page 3-28 for the
procedure to follow.
3-30 Operating the Series 35 IS
March 2008
Setting the Data Logger Time Interval
The Series 35 IS logs measurement data at a time interval that is specified by the user.
Note: Enter the user program as described on page 3-6.
LOGGER
LOG INTERVAL
LINTV (Mins)
LINTV (Mins)
LINTV (Mins)
LOG INTERVAL
15
XX
30
From the main menu, use the arrow keys to scroll to the
LOGGER
prompt, then press the
[ENTER]
key.
Use the arrow keys to scroll to the LOG INTERVAL option and press the [ENTER] key.
The current log interval is displayed. Press either arrow key to delete the current value and enter the edit mode.
Use the arrow and
[ENTER]
keys to change the value one digit at a time, then press [ENTER] to confirm the new value.
The new log interval is displayed. Press
[ENTER]
to return to the options menu.
Press the
[ESC]
key to return to the main menu.
Note: Data logging is suspended when you enter the
programming mode. If you remain in the programming mode longer than the log intervals, the System 35 IS appends a new log when you exit in order to track the log time properly. Also, a new log is appended to the current log on power up.
Operating the Series 35 IS 3-31
March 2008
Entering the Parameter(s) to Log
Use this option to enter the parameters to be logged. You may log up to three different parameters at the same time.
Note: Enter the user program as described on page 3-6.
LOGGER
LOG SETUP
From the main menu, use the arrow keys to scroll to the
LOGGER
prompt, then press the
[ENTER]
key.
Use the arrow keys to scroll to the
LOG SETUP
option and press the
[ENTER]
key.
Note: Before beginning setup of the data logger, check to make
sure that the clock is set to the correct time and date. See
“Setting the Clock Values” on page 3-28 for the
procedure to follow.
L/MODE# X
LM=HYG
LM=HYG DP/°C
Select the desired parameter number (1, 2, or 3) with the arrow keys and then press the
[ENTER] key.
Use the arrow keys to select the desired mode ( HYG , NONE , or
BAT ). Press the [ENTER] key.
Use the arrow keys to select the desired units setting ( MH , PMv ,
DP/°F
,
DP/°C
, or
DVM
). Then, press the
[ENTER]
key.
Note: If the NONE or BAT mode is chosen, the units setting
screen will not appear. Simply press
[ENTER]
to return to the parameter number screen.
3-32 Operating the Series 35 IS
March 2008
Entering the Parameter(s) to Log (cont.)
L/MODE# X
If desired, repeat the above procedure to set up the other two parameters. When finished, press the
[ESC]
key.
Note: Because the Series 35 IS contains a fixed amount of
memory, the more parameters you select and the more frequently you log data, the shorter the total available logging time.
LOG SETUP Press the
[ESC]
key to return to the main menu.
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press [ESC] until the
RUN ?
prompt appears and then press the
[ENTER]
key.
Operating the Series 35 IS 3-33
March 2008
Viewing Log Status, Summary and Data
Use this option to view the log status, log summary and log data.
Be aware that the nature of the 16-character L.C.D. display will limit the ability to view the log data.
Note: Enter the user program as described on page 3-6.
LOGGER From the main menu, use the arrow keys to scroll to the
LOGGER
prompt, then press the
[ENTER] key.
LOG DISPLAY Use the arrow keys to scroll to the LOG DISPLAY option and press the [ENTER] key.
Viewing Log Status
Note: The log status applies only to the current log.
VIEW LOG STATUS
Records: 109
Rem: 10h 10m
VIEW LOG STATUS
Use the arrow keys to scroll to
VIEW LOG STATUS and press the [ENTER] key.
The number of data points logged and the remaining logging time screens will be displayed for about 2 seconds each.
Press the
[ESC]
key to return to the options menu.
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press [ESC] until the RUN ?
prompt appears and then press the [ENTER] key.
3-34 Operating the Series 35 IS
Viewing the Log Summary
Follow these instructions to view the log summary.
March 2008
VIEW LOG SUMMARY
Total #Log: xx
LOG# x
Log Started at: hh:mm mm/dd/yy
P:
LINTV (Mins)
+x.xx PSG x
Use the arrow keys to scroll to
VIEW LOG SUMMARY
and press the
[ENTER]
key.
The total number of logs in memory will be displayed for about 2 seconds.
For each log, the five screens shown will be displayed for about 2 seconds each.
Note: The Series 35 IS does not log time against each data
point, but logs the log time interval at the start of a log.
Use the log start time and log interval to determine the data log time.
The above sequence of five screens will repeat automatically until all of the log summaries have been displayed.
Note: The summary display may be terminated at any time by
pressing the
[ESC]
key between any pair of the five-screen sequences. This will immediately return you to the options menu.
Operating the Series 35 IS 3-35
March 2008
Viewing the Log Summary (cont.)
VIEW LOG #:
VIEW LOG SUMMARY xx
After the log summary has been completed, the display shown will appear. Press the
[ESC]
key or proceed to the next section for instructions.
Press
[ESC]
to return to the options menu.
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press [ESC] until the RUN ?
prompt appears and then press the [ENTER] key.
3-36 Operating the Series 35 IS
March 2008
Viewing Log Data
Follow these instructions to view the log data.
VIEW LOG DATA
VIEW LOG #: xx
Use the arrow keys to scroll to
VIEW LOG DATA and press the
[ENTER] key.
After the log summary has been completed, the display shown will appear. Press the [ESC] key or proceed to the next section for instructions
Log Started at: hh:mm mm/dd/yy
P:
LINTV (Mins)
+x.xx PSG x
For the chosen log, the four screens shown will be displayed for about 2 seconds each.
#x +x.xx PMV The list of items in the log may now be viewed. Use the arrow keys to scroll through the list.
Note: If more than one parameter is being logged, use the
[HYGRO]
and
[P const
]
keys to scroll through the parameters for each item number.
Table 3-2 on page 3-38 shows an example of a log containing
three items and three parameters per item. Scroll through the table as follows:
•
[
↑]: scroll up a column of items
•
[
↓]: scroll down a column of items
•
[HYGRO] : scroll right across a row of parameters
Operating the Series 35 IS 3-37
March 2008
•
[P const
] : scroll left across a row of parameters
Viewing Log Data (cont.)
Item #
1
2
3
Table 3-2: Log Layout Example
Parameter
PMV DP/°F
#1 0.00PMV
#2 0.00PMV
#3 0.00PMV
MH
#1 -166.0DP/°F #1 -0.0019MH
#2 -166.0DP/°F #2 -0.0019MH
#3 -166.0DP/°F #3 -0.0019MH
Note: You can press
[ESC]
to stop viewing the log and return to the
VIEW LOG DATA
menu at any time.
If you scroll all the way to either side of the table, you will not be able to scroll any further in that direction. However, you may scroll back in the opposite direction. Vertical scrolling to either the top or bottom edge of the table, respectively, will lead to one of the following screens:
Skip to next ?
Skip to prev ?
Press
[ENTER]
to view the next or previous log. Press [ESC] to stop viewing the log data.
#
VIEW LOG DATA
<END> Press
[ESC]
to exit.
Press [ESC] to return to the options menu.
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press
[ESC]
until the
RUN ?
prompt appears and then press the
[ENTER]
key.
3-38 Operating the Series 35 IS
March 2008
Using the Built-In RS232 Interface
The Series 35 IS data log can be transferred to a dump terminal via the built-in RS232 interface. Use the following interface settings for the transfer:
•
8 data bits
•
1 stop bit
•
no parity
•
9600 baud
IMPORTANT:
The RS232 interface setup and data transfer must be done in a non-hazardous area.
One of the GE Sensing RS232 cables listed in Table 3-3 below
must be used to connect the RS232 connector on the
Series 35IS to the serial port of the computer.
Table 3-3: RS232 Cable Assemblies
Part Number # Pins on PC End Length (ft)
704-693-02
704-693-03
704-693-04
704-693-05
704-693-12
704-693-13
704-693-14
704-693-15
25
25
25
25
9
9
9
9
5
10
15
20
5
10
15
20
IMPORTANT:
The end of the above cables with the “MMS 35
IS” label must be connected to the hygrometer and not to the PC. Also, standard RS232 cables will not work with the Series 35 IS. Refer to
Figure B-4 in Appendix B for details of the
RS232 cable construction.
Operating the Series 35 IS 3-39
March 2008
Dumping Logged Data Via the RS232 Interface
Use the following procedure to transfer logged data to a dump terminal via the built-in RS232 interface.
IMPORTANT:
Refer to “Using the Built-In RS232 Interface” on
page 3-39 to make sure the Series 35 IS is set up
properly for the transfer.
LOGGER
LOG DUMP
In the main menu, use the arrow keys to scroll to
LOGGER
and press the
[ENTER]
key.
Use the arrow keys to scroll to
LOG DUMP
and press the
[ENTER]
key.
LOG DUMP ON Use the arrow keys to scroll to
LOG DUMP ON
and press the
[ENTER] key.
Dumping Log
Dump Completed
The entire log memory is transferred to the PC. The screens shown will display the status of the data dump.
Note: The log dump may be terminated at any time by pressing
the
[ESC]
key.
LOG DUMP Press the
[ESC]
key to return to the main menu.
3-40 Operating the Series 35 IS
March 2008
Dumping Logged Data Via the RS232 Interface (cont.)
The format of the dump consists of a header, modes with the
TAB delineator between titles, and data with the TAB delimiter between values. The information contained in a typical dump
screen is shown in Figure 3-1 on page 3-42.
To enter more setup data, proceed to the appropriate section of this manual. To return to measurement mode, press [ESC] until the RUN ?
prompt appears and then press the [ENTER] key.
Operating the Series 35 IS 3-41
March 2008
3-42
Log Started at: 01/09/1999 10:13
Constant Pressure = +0.00 PSG
Logging Interval = 1 minutes
LOG#=1
Mode being Logged
Time H:DP/C
10:13
10:14
10:15
-110
-110
-110
10:16
10:17
10:18
10:19
10:20
10:21
-110
-110
-110
-110
-110
-110
H:DP/F
-166
-166
-166
-166
-166
-166
-166
-166
-166
Log Started at: 01/09/1999 10:26
Constant Pressure = +0.00 PSG
Logging Interval = 1 minutes
LOG#=2
Mode being Logged
Time H:DP/C
10:26 -110
H:DP/F
-166
Log Started at: 01/09/1999 10:29
Constant Pressure = +0.00 PSG
Logging Interval = 1 minutes
LOG#=3
Mode being Logged
Time H:DP/C
10:29
10:30
-110
-110
H:DP/F
-166
-166
Log Started at: 01/09/1999 10:31
Constant Pressure = +0.00 PSG
Logging Interval = 1 minutes
LOG#=4
Mode being Logged
Time H:PMv
10:31
10:32
10:33
10:34
10:35
0
0
0
0
0
Dump Completed
H:DP/F
-166
-166
-166
-166
-166
0
0
0
0
0
0
0
0
H:PMv
0
H:PMv
0
H:PMv
0
0
H:MH
-0.0019
-0.0019
-0.0019
-0.0019
-0.0019
Figure 3-1: Example of a Dump Screen
Operating the Series 35 IS
March 2008
Chapter 4
Troubleshooting and Maintenance
The Moisture Monitor Series 35 IS is designed to be maintenance and trouble free. Due to process conditions and other factors, however, minor problems may occur. This chapter discusses some of the most commonly encountered problems and the procedures for correcting them. If you cannot find the information you need in this chapter, contact GE Sensing.
Caution!
Do not attempt to troubleshoot the Series 35 IS beyond the instructions in this chapter. If you do, you may damage the unit and void the warranty.
This section includes the following information:
•
•
•
Calibration Error Processing [page 4-7]
•
Recharging and Replacing the Battery [page 4-8]
•
Replacing the User Program [page 4-13]
•
Replacing and Recalibrating the Moisture Probe [page 4-20]
Common Problems
Incorrect Series 35 IS measurement readings may indicate a
measurement and ways that you can rectify the problems.
Troubleshooting and Maintenance 4-1
Symptom
Table 4-1: Troubleshooting Guide
Possible Cause System Response Action
Accuracy of moisture sensor is questioned
Insufficient time for system to equilibrate
Dew point at sampling point is different than the dew point of the main stream
Probe reads too wet during dry down conditions, or too dry in wet up conditions
Change the flow rate. A change in dew point indicates the sample system is not at equilibrium, or there is a leak. Allow sufficient time for sample system to equilibrate and moisture reading to become steady. Check for leaks.
Probe reads too wet or too dry
Probe reads too wet or too dry
Readings may be correct if the sampling point and main stream do not run under the same process conditions. The different process conditions cause the readings to vary. Refer to
Appendix A for more information. If sampling point and main stream conditions are the same, check sample system pipes and any pipe between the sample system and main stream for leaks.
Also, check sample system for adsorbing water surfaces, such as rubber or plastic tubing, paper-type filters, or condensed water traps. Remove or replace contaminating parts with stainless steel parts.
Clean the sensor and the sensor shield as described in Appendix
A. Then reinstall the sensor.
Sensor or sensor shield affected by process contaminant (refer to
Appendix A)
Sensor is contaminated with conductive particles (refer to
Appendix A)
Probe reads high dew point
Sensor is corroded (refer to
Appendix A)
Probe reads too wet or too dry
Clean the sensor and the sensor shield as described in Appendix
A. Then reinstall the sensor. Also, install a proper filter (i.e. sintered or coalescing element).
Return the probe to factory for evaluation.
Symptom
Table 4-1: Troubleshooting Guide (cont.)
Possible Cause System Response Action
Accuracy of moisture sensor is questioned
Sensor temperature is greater than 70°C (158°F)
Stream particles causing abrasion
Probe reads too dry
Probe reads too wet or too dry
N.A.
Return the probe to factory for evaluation.
Return the probe to factory for evaluation.
Slow response
Accuracy of moisture sensor is questioned
Slow outgassing of system
Sensor is contaminated with non-conductive particles
(refer to Appx. A)
Insufficient time for system to equilibrate
Shorted circuit on sensor
N.A.
Replace the system components with stainless steel or electropolished stainless steel
Clean the sensor and the sensor shield as described in Appendix
A. Then, reinstall the sensor.
Probe reads too wet during dry down conditions, or too dry in wet up conditions
N.A.
Change the flow rate. A change in dew point indicates the sample system is not at equilibrium, or there is a leak. Allow sufficient time for sample system to equilibrate and moisture reading to become steady. Check for leaks.
Run “dry gas” over sensor surface. If high reading persists, then probe is probably shorted and should be returned to the factory for evaluation.
Sensor is contaminated with conductive particles (refer to
Appendix A)
Improper cable connection
N.A.
N.A.
Clean the sensor and the sensor shield as described in Appendix
A. Then, reinstall the sensor.
Check the cable connections to both the probe and the Series 35
IS.
Symptom
Table 4-1: Troubleshooting Guide (cont.)
Possible Cause System Response
Screen always reads the driest (lowest) programmed moisture calibration value while displaying dew/frost point
Open circuit on sensor
Non-conductive material is trapped under contact arm of sensor
Improper cable connection
N.A.
N.A.
N.A.
Action
Return the probe to the factory for evaluation.
Clean the sensor and the sensor shield as described in Appendix
A. Then reinstall the sensor. If the low reading persists, return the probe to the factory for evaluation.
Check the cable connections to both the probe and the Series 35
IS.
March 2008
Screen Messages
Several Series 35 IS screen messages may display during operation of the unit. Table 4-2 lists these messages, possible causes of them and recommended solutions to them.
Troubleshooting and Maintenance 4-5
Display
Blank
Possible Cause
Loss of Power
Table 4-2: Screen Messages
System Response
System shutdown
Action
Turn off unit and recharge or replace the battery. (See Recharging and
Replacing the Battery, page 4-8.)
Call GE Sensing if this happens more than 5 times within ten minutes.
“Loading...” (reinitializes)
“Autocal...” (displays measurement)
“CAL ERROR”
Watchdog reset
Internal reference components may be out of specifications.
(Only occurs when unit is set to measure DP/°C, DP/°F, or PPM v
.)
Reads over or under range Signal received is lower or higher than calibration data supplied with probe
System resets because watchdog signal is not generated within 1.6 seconds. Fault alarm is on for approximately 20 seconds.
Measurement stops for affected modes.
System defaults to lowest or highest dew point found in calibration data
“Supply Low” displays for 1 hour, then “Shutdown
Please”
Battery power is low Loss of power
Check wiring. Call GE Sensing.
Check probe for open circuit or shorts if probe is not subjected to extreme dry or wet conditions. Contact GE Sensing regarding a higher calibrated probe.
Turn unit off and recharge or replace battery (See Recharging & Replacing
March 2008
Calibration Error Processing
A Calibration Error indicates a failure during the measurement of the internal moisture references. During Autocal, internal references are read repeatedly and the Series 35 IS compares measured values to a table of acceptable factory calibration values. The system calculates and corrects any deviation from the factory values. If a reference falls outside the acceptable range, a
CAL ERROR
message appears.
If you attempt to display data after the system experiences a calibration error, the display reads “ CAL ERROR DP/°C .”
Note: The units displayed, (e.g., °C) depends on the previously
selected measurement units.
Troubleshooting and Maintenance 4-7
March 2008
Recharging and Replacing the Battery
Portable battery-powered units, with and without internal sample systems, require a battery charger that accepts either 110 VAC or
220 VAC input, and outputs 12 volts DC at a maximum current of
500 mA.
A fully charged Series 35 IS battery pack provides continuous operation for between 12 hours (with the backlight on) and 120 hours (with the backlight off). When the battery charge is low
(usually below 5.85 volts), the screen blinks and displays the
“Supply Low” prompt. This display remains on the screen for approximately one (1) hour before the Series 35 IS shuts down the electronics.
4-8 Troubleshooting and Maintenance
March 2008
Recharging the Battery
GE Sensing supplies a suitable charger with each Series 35 IS battery-powered unit.
Caution!
Make all AC connections to charge the battery in a non-hazardous, general-purpose area.
If the battery is low, move the Series 35 IS to a non-hazardous area and connect the charger to the unit. While the battery is being charged, the unit may be turned back on. The Series 35 IS displays “Loading....” while it initializes, then returns to whatever it displayed when it was turned off.
The Series 35 IS begins charging when you plug it into AC power. A full charge takes 16 hours. Use the following steps to recharge the battery.
1. Turn the unit off.
2. Plug the charger into the line power.
3. Plug the other end of the charger into the connector on the left side of the electronics unit.
4. Charge the unit for 16 hours.
Note: The unit can continue to operate while its batteries are
being charged.
!WARNING!
Do not connect the Series 35 IS to a charger in a hazardous environment.
Troubleshooting and Maintenance 4-9
March 2008
Replacing the Battery
If the battery of your Series 35 IS does not hold a charge for a full
12 hours after 16 hours of recharging, the battery needs replacement. Contact the factory for a new one.
Always store the Series 35 IS lead acid battery at room temperature (
≈25°C). Table 4-3 below illustrates how periods of
storage deplete battery capacity.
Table 4-3: Battery Capacity in Storage
Storage Time at
Room Temperature Capacity
3 months
6 months
12 months
91%
82%
64%
Refer to Figure 4-1 on page 4-11 and complete the following
steps to remove the old battery:
Note: The only tools needed to change the Series 35 IS battery
are a standard screwdriver to open the case and a 1/8"
Allen wrench to remove the battery cover.
1. Loosen the two quarter-turn screws on the back of the Series
35 IS cover. Remove the cover by lifting it off the unit from back to front.
2. Disconnect the ribbon cable that connects the encapsulated charger board to connector J3 on the back of the main printed circuit board, which is mounted in the cover of the Series 35
IS. See Figure 4-1 on page 4-11 and Figure 4-2 on page 4-15.
Note: Pin 1 of connector J3 is clearly labeled on the circuit
board, and the side of the ribbon cable that must align with pin 1 has been marked as such at the factory.
4-10 Troubleshooting and Maintenance
Replacing the Battery (cont.)
Ribbon Cable
March 2008
Cover
Foam
Battery
Encapsulated
Charger Board
Black Lead
Red Lead
Figure 4-1: Battery Pack with Charger Board
3. Use the Allen wrench to remove the four socket head cap screws that secure the battery cover to the base plate.
4. Lift the battery cover out of the electronics unit and slide the battery and the encapsulated charger board out of the battery cover.
5. Disconnect the leads from the encapsulated charger board by removing the two crimp pins from the battery terminals (take care not to damage the wires).
Troubleshooting and Maintenance 4-11
March 2008
Replacing the Battery (cont.)
Refer to Figure 4-1 on page 4-11 and Figure 4-2 on page 4-15 and
use the following steps to install the new battery:
1. Connect the two (2) leads from the encapsulated charger board to the new battery by pressing the crimp pins on the ends of the leads onto the terminals of the new battery.
IMPORTANT:
Make sure the red lead is connected to the “+” battery terminal and the black lead is connected to the “–” battery terminal.
2. Insert the new battery and the encapsulated charger board into the battery cover. Make sure the ribbon cable is threaded through the slot in the battery cover.
3. Place the battery pack assembly into position in the electronics unit, and secure the battery cover to the base plate with the four (4) screws previously removed.
4. Reconnect the ribbon cable from the encapsulated charger board to connector J3 on the back of the main printed circuit board. Make sure that the side of the ribbon cable with the factory marking is aligned with the end of connector J3 that is labeled as pin 1.
5. Replace the Series 35 IS cover by attaching it to the unit from front to back. Tighten the two quarter-turn screws on the back of the cover.
4-12 Troubleshooting and Maintenance
March 2008
Replacing the User Program
The user program is stored on an EPROM (Erasable
Programmable Read Only Memory) chip. The EPROM is located on the main printed circuit board (No. 703-1262), which is mounted in the cover of the Series 35 IS electronics unit.
You must do the following to replace the user program:
•
Remove the main printed circuit board.
•
Replace the EPROM.
•
Reinstall the main printed circuit board.
Refer to the following sections to replace the user program.
Contact GE Sensing if your unit contains a main circuit board with a registration number other than 703-1262.
Troubleshooting and Maintenance 4-13
March 2008
Removing the Circuit Board
See Figure 4-2 on page 4-15 to locate the cable connections on
the back (non-component side) of the main printed circuit board.
Note: The main printed circuit board is mounted on the inside of
the electronics cover.
Caution!
Remove the circuit board to a non-hazardous area.
1. Turn the power off and, if applicable, unplug the unit.
!WARNING!
You must turn off and unplug the Series 35 IS before continuing with the following steps.
2. Discharge static electricity from your body before touching the Series 35 IS enclosure.
3. Open the Series 35 IS enclosure by loosening the two quarterturn screws on the back of the unit and gently lifting the cover off from back to front.
There are five (5) detachable cables plugged into the back (non-
component side) of the printed circuit board. Refer to Table 4-4
below for a description of these cables.
Table 4-4: Printed Circuit Board Cables
Cable Type Connector Source
2-Wire
16-conductor Ribbon
14-conductor Ribbon
2-Wire
9-conductor Ribbon
J2
J3
J4
J5
J6
ON/OFF Switch
Battery Pack
Display Board
Display Backlight
Keypad
4-14 Troubleshooting and Maintenance
March 2008
Removing the Circuit Board (cont.)
Display Cable
Keypad Cable
S1
3
1
GRN
RED
J2
J5
1 J4 1 J6
Backlight Cable
Power Cable
Probe Cable
(Internal Probe only)
Charger Cable
J3
1
Mounting Screw
4 places
Figure 4-2: Rear of PC Board - Cable Connections
Note: Be careful not to pull on the cables while removing the
cover.
IMPORTANT:
Before disconnecting the ribbon cables, mark the side of each cable that is near pin 1 of its circuit board connector. Pin 1 is clearly labeled on the
circuit board for J3, J4 and J6. See Figure 4-2
above.
Troubleshooting and Maintenance 4-15
March 2008
Removing the Circuit Board (cont.)
4. Disconnect the five (5) cables from the J2, J3, J4, J5, and J6 connectors on the back (non-component side) of the main printed circuit board.
IMPORTANT:
To maintain good contact at each terminal block and to avoid damaging the pins on a connector, pull the connector straight off (not at an angle).
5. If your Series 35 IS includes a self-contained sample system, you must disconnect the moisture probe cable from the probe at this time (refer to Installing a Probe into a Sample System in Chapter 2).
6. Remove any external connections to the Series 35 IS. These may include: the battery charger, an RS232 interface cable and/or a moisture probe cable from an external sample system.
7. Remove the main printed circuit board from inside the cover by unscrewing the four screws that fasten the circuit board to its mounting bracket.
8. Turn the circuit board over to access the EPROM on the component side of the circuit board.
Note: To replace the circuit board, refer to Replacing the Circuit
4-16 Troubleshooting and Maintenance
March 2008
Replacing the EPROM
Caution!
EPROMs can be damaged by static electricity. Be sure to discharge any static electricity from your body before handling the EPROM.
Replace the EPROM by carefully completing the following steps in the order presented.
1. Refer to Figure 4-3 on page 4-18 to locate the EPROM
(labeled U18) on the component side of the main printed circuit board.
2. Use a chip puller to remove the EPROM. If you do not have a chip puller, use a small screwdriver to carefully wedge the chip out of its mounting. Be sure none of the EPROM legs breaks off in the socket.
3. Insert the new EPROM into the socket labeled U18, making sure the notch on the EPROM aligns with the notch on the
socket. See Figure 4-3 on page 4-18.
4. If all the EPROM legs do not align with the corresponding holes in the socket, gently remove the EPROM and place it on its side (where the legs are located) on a flat surface. Then, gently roll the EPROM toward the legs to bend the row of legs slightly inward.
Caution!
The EPROM legs are very delicate and may snap off if they are bent too far or too many times.
5. Repeat step 4 for the legs on the opposite side of the EPROM.
Then, insert the EPROM back into its socket, making sure that the notch on the EPROM aligns with the notch on the socket.
Troubleshooting and Maintenance 4-17
March 2008
Replacing the EPROM (cont.)
TB1
1
R6
R3 R4
C1
C9
U1
3
R1 R2
U6
C3
C2
U31 U23
C5
U3
C4
U7
C6
C7
C15
C13
U8
C8
C14
U2
C21
+
U9
U5
R39
C20
C18
C19
C16
+
C10
C12
Y1
C11
R15
R16
U28
C17
U4
R38
J1
C24
U24 U25 U26 U27
C28
U29 U30
R37
U19
U10
R35
R36
U22
6
1 U13
R31
9
J7
5
R32
C25
+
703-1262
C26
U14
R33
R34
R30
R29
Y2
C27
C22
C23
Y3
U11
U15 U21
U17
U16 U12
U20
1
EPROM
RA1
U18
Notch
Figure 4-3: EPROM (U18) Notch Location
6. Gently press the EPROM into place until it is fully seated in the socket, making sure not to bend or break any of the legs.
Repeat steps 4 and 5 if necessary.
Caution!
Do not force the EPROM into the socket. If the EPROM does not seat easily, check for bent legs or legs that do not properly align with the corresponding holes in the socket.
4-18 Troubleshooting and Maintenance
March 2008
Replacing the Circuit Board
1. Turn the circuit board over so that the component side (the side with the EPROM) of the board faces the inside of the
Series 35 IS cover, and position the circuit board on the mounting bracket in the cover.
Note: When orienting the circuit board within the electronics
unit cover, make certain that the RS232 connector, battery charger jack and external probe cable connector are properly inserted into the cutouts in the side panel of the cover.
2. Install and tighten the four (4) previously removed bracket screws and washers in the corners of the circuit board.
3. If your unit includes a self-contained sample system, reconnect the moisture probe cable to the probe at this time
(see Chapter 2, Installing a Probe into a Sample System).
4. Reconnect the five (5) cables to the J2, J3, J4, J5, and J6 connectors on the back (non-component side) of the circuit board. The cables that connect to J2 and J5 are polarized and can only be inserted in the correct orientation. However, the three (3) ribbon cables could be rotated 180° from the proper alignment. Therefore, make sure that the previously applied pin 1 marking on each ribbon cable aligns with the pin 1 label on the circuit board at the J3, J4 and J6 connectors.
IMPORTANT:
To maintain good contact at each terminal block and to avoid damaging the pins on a connector, push the connector straight on (not at an angle).
5. Replace the Series 35 IS cover by attaching it to the unit from front to back. Tighten the two quarter-turn screws on the back of the unit.
6. Reconnect any external cables such as the battery charger, an
RS232 interface cable and/or the moisture probe cable from an external sample system.
Troubleshooting and Maintenance 4-19
March 2008
Replacing and Recalibrating the Moisture
Probe
For maximum accuracy you should send probes back to the factory for recalibration every six months to one year, depending on the application. Under severe conditions you should send the probes back for recalibration more frequently. Contact a GE
Sensing applications engineer for the recommended calibration frequency for your application.
When you receive a new or recalibrated probe, make sure to install and connect it as described in Chapter 2, Installing a Probe
into a Sample System. Once you have installed and connected the probes, enter the calibration data as described in Chapter 3,
Entering Calibration Data.
Note: An individualized Calibration Data Sheet, containing the
corresponding probe serial number, accompanies each probe.
4-20 Troubleshooting and Maintenance
March 2008
Chapter 5
Specifications
The Moisture Monitor Series 35 IS has been designed to meet or exceed the specifications listed in this chapter. The information has been divided into the following two categories:
•
Electrical Specifications [page 5-2]
•
General Specifications [page 5-3]
Specifications 5-1
March 2008
Electrical Specifications
Functions:
Dew Point:
PPMv in gases at constant pressure
(pressure by programmable constant).
Inputs:
Moisture:
Single input via M-Series probe connected to main board.
Probe may be remotely located up to 100 m (330 ft) from electronic console, except for units with self-contained internal sample systems.
Intrinsic Safety:
BASEEFA-certified intrinsically safe probe and cable.
ISSEP-certified intrinsically safe hygrometer (Code: EEx ia
IICT4)
Computer-Enhanced Response:
Optional
Display Units:
Dew/frost point temperature: ºC, ºF, PPMv, MH
Pressure: psi(g), bar, kPa(g), kg/cm
2
(g)
Power Requirements:
AC power supply:
Voltage: Specify as 100, 120, 230, or 240 VAC
Frequency: Specify as 50/60 Hz
12V Charger Input Power: 5 watts, maximum
Rechargeable Battery Pack: Battery power for 12-hour operation (with backlight) or 120-hour operation (without backlight)
5-2 Specifications
March 2008
General Specifications
Temperature:
Operating: 0° to 60°C (32º to 140ºF)
Storage: -20° to 70°C (-22º to 158ºF).
Configurations:
Battery Powered
Battery Powered with self-contained Sample System.
Dew/Frost Point Temperature:
Overall calibration range capability:
+60° to -110°C (+140º to -166ºF).
Available Calibration Range Options:
Standard Calibration Range:
20 o
to -80 o
C with data to -110 o
C
(68º to -112ºF with data to -166ºF)
Extended High Calibration Range:
60 o
to -80 o
C with data to -110 o
C
(140º to -112ºF with data to -166ºF)
Accuracy:
±2°C from 60° to -65°C (140º to -85ºF)
±3°C from -66° to -110°C (-86º to -166ºF)
Repeatability:
±0.5°C from 60° to -65°C (140º to -85ºF)
±1.0°C from -66° to -110°C (-86º to -166ºF)
Specifications 5-3
March 2008
General Specifications (cont.)
Response Time:
5 sec for 63% of a step change in moisture content in either wet up or dry down cycle
Gas Flow Range:
From static to 10,000-cm/s linear velocity at 1 atm
Sensor Type:
GE Sensing M-Series thin-film aluminum oxide
Traceability:
Every moisture probe is individually computer-calibrated against known moisture concentration standards, traceable to
National Institute of Standards and Technology (NIST) or
National Physical Lab, U.K. (NPL) as accredited by Irish
National Accreditation Board (INAB).
Temperature:
Operating and Storage:
-110º to +70ºC (-166º to 158ºF)
@LEVEL4A = Pressure:
Operating:
5 microns Hg to 5000 psig
5-4 Specifications
March 2008
Appendix A
Outline and Installation Drawings
This appendix contains the following outline and installation drawings:
•
MMS35 IS with Sample System (ref. dwg #712-940)
•
MMS35 IS without Sample System (ref. dwg #712-940)
•
Internal Sample System Assembly (ref. dwg #750-297)
•
RS232 Cable Assembly (ref. dwg #704-693)
Outline and Installation Drawings A-1
10.78
(273.8)
March 2008
Outline and Installation Drawings
5.40
(137.0)
10.98 (278.8)
A-2
10.78
(273.8)
March 2008
Outline and Installation Drawings
5.40
(137.0)
10.98 (278.8)
A-3
2
4.88
MIN
3
1
4
7
6
9
Outline and Installation Drawings
2.67
MAX
March 2008
8
6
7
4
5
1
2
3
8
9
10
PART NO.
412-722-03
421-882
421-638
2530
304SS
SS-400-2-2
SS-400-1-2
SS-400-61
SS-4TF-7
SS-1RS4-A
PARTS LIST
DESCRIPTION
Screw, PH, FH 6-32 x3/16"
Base Plate
Bracket, Sample Cell
Sample Cell
Tubing 1/4" x .035
Elbow Fitting
Fitting
Fittings
Filter
Valves
MANUFACTURER QTY
GE Panametrics
2
GE Panametrics
1
GE Panametrics
1
GE Panametrics
1
Seamless Tubing
A/R
Swagelok
Swagelok
Swagelok
1
1
Nupro
Whitey
2
1
2
10
5
8
A-4
March 2008
1
-02
-03
-04
-05
-12
-13
-14
-15
DASH
NO.
TABLE 1
CONFIGURATION
PC END
CONN.
PINS
9
9
9
9
25
25
25
25
LENGTH, L
FT.
5
10
15
20
5
10
15
20
1/2"TYP
REMOVE 1/2" OF
CABLE SHEATH & SHIELD
FOIL, BOTH ENDS
19
6
9
1
5
2--RED
3--GRN
5--BLK & DRAIN
9 CONTACT
CONNECTIONS
MMS35IS END
WIRING END
PART NO.
704-556-02
THRU
704-556-05
704-556-12
THRU
704-556-15
TABLE 2
WIRE RUN LIST
PIN NO.
2
3
5
2
3
5
WIRE
COLOR
RED
GRN
BLK/DRAIN
RED
GRN
BLK/DRAIN
CONNS (PC END)
PIN NO.
WIRE
COLOR
RED
3
2
5
GRN
BLK
2
3
7
RED
GRN
BLK
MMS35IS END
SEE TABLE 1 FOR LENGTH
8 9 10 11
SEE TABLE 1
PC END
9 CONTACT
1
25 CONTACT
5
15
SEE NOTE 4
13
BOTH ENDS
PUSH OVER WIRES
AFTER MAKING CONNECTIONS
AND SHRINK IN PLACE
2" 2"
TRIM FOIL & WHITE WIRE
FLUSH WITH CABLE JACKET
STRIP & TIN 1/4"
ON 3 WIRES
EACH END
TRIM FOIL,WHITE WIRE &
DRAIN FLUSH WITH
CABLE JACKET
14
1
2--RED
3--GRN
7--BLK
6
1
25 13 9
5
25 CONTACT 9 CONTACT
CONNECTIONS
PC END
WIRING END
2--GRN
3--RED
5--BLK
2
16
MMS35IS END ONLY
9 CONTACT 2
25 CONTACT
AS SHOWN
6
14
1/4"±1/16"
(3 PLCS EACH END)
TYP
SEE NOTE 2
NOTES:
1. NO BROKEN OR NICKED STRANDS
PERMITTED (TIN ALL LEADS).
2. SHRINK ITEM 14 ON EACH CONTACT
AFTER SOLDERING WIRE TO CONTACT.
3. TOLERANCE: SLEEVING ± 1/4"
LENGTH ± 6"
4. AFTER SHRINKING SLEEVING (ITEM 13)
IN PLACE, POSITION AND SHRINK THE
OTHER SLEEVE (ITEM 15) OVER THE
FIRST SLEEVE IN ORDER TO PROVIDE
ADEQUATE STRAIN RELIEF WHEN
CONNECTOR IS CLOSED UP.
Outline and Installation Drawings A-5
March 2008
Appendix B
Menu Map
Figure C-1 is a top-level diagram of the Series 35 IS menu. Once you are familiar with how the Series 35 IS operates, use this diagram as a reference for moving through the user program.
To enter the programming mode, perform the following sequence within five (5) seconds, or the Series 35 IS will time out and return to the measurement mode.
P:
ESC
+0.00 Kcmg While in run mode, press the
[ESC]
key.
PROGRAM MENU
DP RANGE
Within 5 seconds of the appearance of this display, press the
[ENTER]
key followed by the
[ESC]
key.
This will display for 1 second.
You are now in the user program menu.
From the main menu, use the arrow keys to scroll through the options to your selection. Use the
[ENTER]
key to confirm your selection or the [ESC] key to cancel your selection.
Menu Map B-1
March 2008
Menu Map
DP RANGE
MH CURVE
SERIAL #
HYGRO REFERENCE
USER
FACTORY SETUP
LOGGER
Hi DP
Low DP
XX° MH X.XXXX
XX° ED X.XXXX
_ _ _ _ _ 0 S/N
Low REF
High REF
_ _ 0 _ _ Passcode
LOG INTERVAL
LOG SETUP
LOG DISPLAY
LOG DUMP
LX.XXXX
HX.XXXX
LINTV (Mins) _ _ _ _ X
L/MODE # X
LOG DUMP OFF/ON
OFFSET
CONSTANT PRESSUR
PPMv MULTIPLIER
AUTOCAL INTERVAL
BATTERY
BACKLIGHT
ENHANCE RESPONSE
RANGE ERROR
SELECT 60/50 HZ
CONTRAST
CLOCK
LM = HYG
LM = NONE
LM = BAT
VIEW LOG STATUS
VIEW LOG SUMMARY
VIEW LOG DATA
ENHANCE OFF
ENHANCE ON
R_ERR=No Action
R_ERR=Display
60HZ
50HZ
Set Date
Read Time
Set Time
LM=HYG _ _ _ (units)
B-2
March 2008
Appendix C
Data Information Sheet
Use this sheet to record all of the data that you enter into the user program. If data is lost for any reason, use this sheet to reprogram your unit. Store this sheet and any other related documents in a safe place for future reference.
Tables are provided for entry of the following information:
•
Dew Point Measurement Range [page -2]
•
Calibration Data for Moisture [page -2]
•
•
MH Reference Values for Moisture [page -3]
•
User Defined Functions [page -3]
•
Data Information Sheet C-1
March 2008
Series 35 IS Data Information Sheet
Notes:
Date: ___________________________________________
Unit Serial Number: _______________________________
Application Description: ____________________________
________________________________________________
High
Table C-1: Dew Point Measurement Range
Low
Point #
1
2
5
6
3
4
11
12
13
14
7
8
9
10
Table C-2: Calibration Data for Moisture
MH Value
C-2 Data Information Sheet
March 2008
Series 35 IS Data Information Sheet (cont.)
Table C-3: Serial Number
___ ___ ___ ___ ___ ___
High
Table C-4: MH Reference Values for Moisture
Low
Table C-5: User-Defined Functions
Function
Offset Value
Setting
Constant Pressure
PPMv Constant Multiplier
Autocal Interval
Battery
Backlight-On Time Interval
Computer-Enhanced Response
Range Error Handling
Selecting 60/50 Hz
Data Information Sheet C-3
March 2008
Series 35 IS Data Information Sheet (cont.)
Function
Log Interval
Log Setup
Log Display
Log Dump
Table C-6: Logger
Setting
LINTV (Mins):
________
L/Mode #1:
+
HYG NONE BAT
(units = ______ )
L/Mode #2:
L/Mode #3:
HYG NONE BAT
(units = ______ )
HYG NONE BAT
(units = ______ )
View Log Status
View Log Summary
View Log Data
OFF ______
____________
____________
____________
ON ______
C-4 Data Information Sheet
March 2008
Appendix D
Series 35 IS Hygrometer Spare Parts
Spare parts for the portable Moisture Monitor Series 35 IS are
available directly from GE Sensing. Table D-1 below lists the
ordering information for the most commonly needed spare parts.
If you require parts that are not listed in Table D-1, contact GE
Sensing for assistance.
P/N
Table D-1: Spare Parts List
Qty* Description
705-799
200-050
M2L
201-055
201-066
1
1
1
1
1
Display Assembly
Battery Pack
Moisture Probe
120V Battery Charger
076 1**
240V Battery Charger
7-micron Sintered SS Filter Elements
(pkg. of 4)
* recommended for 0-2 years operation
**required only for unit with internal sample system
Series 35 IS Hygrometer Spare Parts D-1
March 2008
Index
A
AC Line Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-26
Automatic Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
B
Backlight-On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22
Battery Charger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Battery Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
Battery, Recharging and Replacing . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
C
Cables, Modified . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Calibration
Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Reference Sticker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Calibration Data
Data Sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Entering. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Probe Serial Number. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Circuit Board
Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
Removing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
Replacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Clock Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-28
Computer-Enhanced Response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23
Constant Multiplier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Constant Pressure Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Contrast Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-27
D
Data Information Sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -1
Data Logger. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-30
Parameter(s) to Log. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-32
Viewing Log Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-32
Viewing Log Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-34
Viewing Log Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-35
Dew Point Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
1
2
March 2008
Index (cont.)
E
Electronics Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
EPROM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Replacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
Error Processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
F
Factory Setup Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
G
Getting Started. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
I
Installation
Mounting the Sample System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Probe Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Wiring Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
K
Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Displaying Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Entering Numeric Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Using . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
L
Logged Data, Dumping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-40
M
M Series Probes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Measurement Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Menu Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Moisture Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3, 2-3
Fluid Velocities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
M Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Moisture Condensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Replacing and Recalibrating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Temperature Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
March 2008
Index (cont.)
O
Offset Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
P
Power Switch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Powering Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Pressure
Constant Value, Entering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Installing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Self-Contained . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Probe Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Programming Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
R
Range Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-25
Reference Values, Entering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
RS232 Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-39
Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-39
Dumping Logged Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-40
Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-39
3
4
March 2008
Index (cont.)
S
Sample System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3, 2-10
Cable length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
External . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Mounting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Pressure Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Probe Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Self-Contained . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Valves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Screen Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Serial Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Shield. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Shutting Down. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Site Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Spare Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
Specifications
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Starting Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
T
Terminal Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Troubleshooting and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
U
User Program
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Entering and Exiting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
EPROM. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Navigating. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Replacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
W
Wiring Connections, Description . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Wiring, Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
GE Infrastructure
Sensing
DECLARATION
OF
CONFORMITY
We,
Panametrics Limited
Shannon Industrial Estate
Shannon, County Clare
Ireland
declare under our sole responsibility that the
Moisture Monitor Series 35IS Portable Analyzer
PM880 Portable Moisture Monitor
to which this declaration relates, are in conformity with the following standards:
• EN 50014:1997+A1+A2:1999
• EN 50020:1994
• EN50284:1999
• II 1 G EEx ia IIC T4
PM880: BASEEFA02ATEX0191
Baseefa (2001) Ltd/EECS, Buxton SK17 9JN, UK
MMS 35IS: ISSeP01ATEX035 X; ISSeP, B7340 Colfontaine, Belgium
• EN 61326:1998, Class A, Annex C, Continuous Unmonitored Operation
(for EN 61000-4-3, the MMS 35IS meets performance Criteria A and, in a limited number of frequencies, performance Criteria B per EN 61326.) following the provisions of the 89/336/EEC EMC Directive and the 94/9/EC ATEX Directive.
The units listed above and any sensors and ancillary sample handling systems supplied with
them do not bear CE marking for the Pressure Equipment Directive, as they are supplied in accordance with Article 3, Section 3 (sound engineering practices and codes of good workmanship) of the Pressure Equipment Directive 97/23/EC for DN<25.
Shannon - July 1, 2003
Mr. James Gibson
GENERAL MANAGER
CERT-DOC-H3
TÜV
TÜV ESSEN
ISO 9001
U.S.
August 2004)
GE Infrastructure
Sensing
DECLARATION
DE
CONFORMITE
Nous,
Panametrics Limited
Shannon Industrial Estate
Shannon, County Clare
Ireland
déclarons sous notre propre responsabilité que les
Moisture Monitor Series 35IS Portable Analyzer
PM880 Portable Moisture Monitor
rélatif á cette déclaration, sont en conformité avec les documents suivants:
• EN 50014:1997+A1+A2:1999
• EN 50020:1994
• EN50284:1999
• II 1 G EEx ia IIC T4
PM880: BASEEFA02ATEX0191
Baseefa (2001) Ltd/EECS, Buxton SK17 9JN, UK
MMS 35IS: ISSeP01ATEX035 X; ISSeP, B7340 Colfontaine, Belgium
• EN 61326:1998, Class A, Annex C, Continuous Unmonitored Operation
(for EN 61000-4-3, the MMS 35IS meets performance Criteria A and, in a limited number of frequencies, performance Criteria B per EN 61326.) suivant les régles de la Directive de Compatibilité Electromagnétique 89/336/EEC et d’ATEX
94/9/EC.
Les matériels listés ci-dessus, ainsi que les capteurs et les systèmes d'échantillonnages
pouvant être livrés avec ne portent pas le marquage CE de la directive des équipements sous pression, car ils sont fournis en accord avec la directive 97/23/EC des équipements sous pression pour les DN<25, Article 3, section 3 qui concerne les pratiques et les codes de bonne fabrication pour l'ingénierie du son.
Shannon - July 1, 2003
Mr. James Gibson
DIRECTEUR GÉNÉRAL
CERT-DOC-H3
TÜV
TÜV ESSEN
ISO 9001
U.S.
August 2004)
GE Infrastructure
Sensing KONFORMITÄTS-
ERKLÄRUNG
Wir,
Panametrics Limited
Shannon Industrial Estate
Shannon, County Clare
Ireland
erklären, in alleiniger Verantwortung, daß die Produkte
Moisture Monitor Series 35IS Portable Analyzer
PM880 Portable Moisture Monitor
folgende Normen erfüllen:
• EN 50014:1997+A1+A2:1999
• EN 50020:1994
• EN50284:1999
• II 1 G EEx ia IIC T4
PM880: BASEEFA02ATEX0191
Baseefa (2001) Ltd/EECS, Buxton SK17 9JN, UK
MMS 35IS: ISSeP01ATEX035 X; ISSeP, B7340 Colfontaine, Belgium
• EN 61326:1998, Class A, Annex C, Continuous Unmonitored Operation
(for EN 61000-4-3, the MMS 35IS meets performance Criteria A and, in a limited number of frequencies, performance Criteria B per EN 61326.) gemäß den Europäischen Richtlinien, Niederspannungsrichtlinie EMV-Richtlinie
Nr.: 89/336/EG und ATEX Richtlinie Nr. 94/9/EG.
Die oben aufgeführten Geräte und zugehörige, mitgelieferte Sensoren und
Handhabungssysteme tragen keine CE-Kennzeichnung gemäß der Druckgeräte-Richtlinie, da sie in Übereinstimmung mit Artikel 3, Absatz 3 (gute Ingenieurpraxis) der Druckgeräte-
Richtlinie 97/23/EG für DN<25 geliefert werden.
Shannon - July 1, 2003
Mr. James Gibson
GENERALDIREKTOR
CERT-DOC-H3
TÜV
TÜV ESSEN
ISO 9001
U.S.
August 2004)
USA
1100 Technology Park Drive
Billerica, MA 01821-4111
Web: www.gesensing.com
Ireland
Sensing House
Shannon Free Zone East
Shannon, County Clare
Ireland

Download
Advertisement
Key features
- Measures moisture content in gases
- Real-time moisture measurement
- Dew/frost point measurement
- Temperature range of –110 to +60°C (–166 to +140°F)
- Microprocessor-based
- Single-channel
- User-programmable
- Data logging capabilities