Precision Medical PM65HG, PM65E Easy Go Vac Aspirator User manual
The Easy Go Vac Aspirator PM65HG, PM65E is a portable medical vacuum source that can be used in the homecare/healthcare environment. It is powered by a 12V DC battery, and it provides a vacuum of 2-21 inHg (51-533 mmHg) with a flow rate of 6.5 ± 0.4 lpm @ 11.8 inHg (300 mmHg). The device is equipped with a variety of features, including a built-in overfill protection device, an inline hydrophobic filter, and a vacuum control knob. The Easy Go Vac Aspirator is also equipped with a battery charger and a carry bag.
Advertisement
Advertisement
U S E R M A N U A L
Model No. PM65HG
PM65E
SAVE THESE INSTRUCTIONS
CAUTION
Federal (USA) law restricts this device to sale by or on the order of a physician.
300 Held Drive
Northampton, PA 18067
Tel: (+001) 610-262-6090
Fax: (+001) 610-262-6080 www.precisionmedical.com
CONTENTS
RECEIVING/INSPECTION ........................................................... 2
INTENDED USE .......................................................................... 2
READ ALL INSTRUCTIONS BEFORE USING ............................ 2
ExPLANATION OF ABBREVIATIONS ......................................... 2
SAFETY INFORmATION - WARNINGS AND CAUTIONS ........... 3
SPECIFICATIONS ....................................................................... 5
COmPONENT DESCRIPTION .................................................... 6
OPERATING INSTRUCTIONS ..................................................... 6
A. AC Electrical Recepticle Operation ............................. 7
B. Internal Battery Operation ............................................ 8
C. Auto DC Operation ..................................................... 9
CHARGING BATTERY ............................................................... 10
CLEANING/mAINTENANCE ..................................................... 10
RETURNS .................................................................................. 11
DISPOSAL INSTRUCTIONS ..................................................... 11
TROUBLESHOOTING GUIDE ................................................... 12
REPLACEmENT PARTS ............................................................ 13
LImITED WARRANTY ................................................................ 14
1
RECEIVING/INSPECTION
Remove the Precision medical Easy Go Vac Aspirator from packaging and inspect for damage. If there is any damage, DO
NOT USE and contact your Equipment Provider.
INTENdEd USE
The Easy Go Vac Aspirator provides a portable, AC/DC powered medical vacuum source. It is intended for use in the homecare/ healthcare environment.
REAd ALL INSTRUCTIONS BEFORE USING
This manual instructs a Professional to install and operate the Easy
Go Vac Aspirator. This is provided for your safety and to prevent damage to the Easy Go Vac Aspirator. If you do not understand this manual, DO NOT USE the Easy Go Vac Aspirator and contact your Equipment Provider.
ExPLANATION OF ABBREVIATIONS
mmHg inHg kPa lpm millimeters of mercury
Inches of mercury
Kilopascal
Liters per minute
2
SAFETY INFORMATION - WARNINGS ANd CAUTIONS
dANGER
WARNING
CAUTION
CAUTION
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
Used without the safety alert symbol indicates a potentially hazardous situation which, if not avoided, may result in property damage.
CONSULT ACCOmPANYING DOCUmENTS
.
Type BF equipment
.
Suction "OFF" (Battery Charging mode when plugged into charger.)
Suction "ON"
3
•
•
•
•
•
•
•
dANGER
Equipment not suitable for use in the presence of flammable anesthetic.
Caution must be exercised when operating the Easy Go
Vac Aspirator where oxygen is being administered.
dO NOT use near any type of flame or flammable/explosive
•
•
substances.
dO NOT allow Aspirator to exceed Operating or Storage
Temperature specifications.
TO REdUCE THE RISk OF ELECTROCUTION:
•
•
Always unplug the Easy Go Vac Aspirator after using, except when battery is being recharged.
dO NOT use while bathing.
dO NOT place or store the Easy Go Vac Aspirator where it can fall or be pulled into a tub or sink.
dO NOT immerse into water or other liquid.
dO NOT reach for the Easy Go Vac Aspirator that has fallen into water. Unplug immediately.
dO NOT attempt to repair the Easy Go Vac Aspirator. Any attempt could result in electrical shock.
WARNING
TO REdUCE RISk OF BURNS, ELECTROCUTION, FIRE, OR
•
•
•
INjURY TO PERSONS:
•
•
•
•
Close supervision is necessary when this Easy Go Vac
Aspirator is used by, on, or near children or invalids.
Use this Easy Go Vac Aspirator only for its "Intended Use" as described in this manual.
Keep the cord away from heated surfaces.
dO NOT use when sleepy or drowsy.
dO NOT use attachments not recommended by the
Precision medical, Inc.
dO NOT use if the Easy Go Vac Aspirator has a damaged cord or plug.
dO NOT use if the Easy Go Vac Aspirator has been dropped or damaged.
4
SPECIFICATIONS
dimensions (in Carry Bag): Length:
Width:
Height:
Weight (with Carry Bag and all Accessories): 9.6 lbs (4.4 kg)
Shipping Weight: 11 lbs (5 kg)
Suction Canister:
12” (30 cm)
9.5” (24 cm)
10” (25 cm)
800cc max Volume
Vacuum is not to exceed 25 inHg (84.7 kPa).
The Canister Lid has a float shut off valve, that protects the Aspirator from overfill.
Patient Tubing:
1/4" ID Non-Conductive PVC tubing, 6 ft. long, with 1/4" - 3/8" ID female connectors, or equivalent
Electrical:
Aspirator Unit:
Battery Charger/Power Supply:
12 Volts DC, 1.8 Amps
100 - 240 Volts AC, 50/60 Hz,
0.80-0.35 Amps
Internal Battery Type:
Vacuum:
12 Volt, Sealed Lead Acid
Flow:
2 - 21 inHg (51 - 533 mmHg)
6.5 ± 0.4 lpm @ 11.8 inHg
(300 mmHg)
Operating Temperature Range: 50°F to 104°F (10°C to 40°C)
Storage/ Transport Requirements
Temperature Range: maximum Humidity:
-4°F to 104°F (-20°C to 40°C)
95% Non-condensing
Equipment Classification:
Class I
Type BF
IPx0
- with respect to protection from electrical shock
- degree of protection against electrical shock
- degree of protection against ingress of liquids mode of Operation - continuous
The Easy Go Vac Aspirator meets the EmC requirements of EN 60601-1-2.
Specifications are subject to change without prior notice.
5
COMPONENT dESCRIPTION
CAUTION
If for any reason any label becomes illegible or lost, contact
Precision medical, Inc..
External
Power jack
Vacuum Control knob
Inline Hydrophobic Filter
On/Off Switch
Vacuum
Gauge
Yellow Low Battery
Light
Patient
Tubing
Connection
AC Power
Cord
Suction Canister
(Built in overfill protection device)
Battery Charger/
Power Supply
Patient
Suction
Tubing
Auto dC
Power Cord
(optional)
GROUNdING INSTRUCTIONS
The Battery Charger Power Supply should be grounded. In the event of an electrical short circuit, grounding reduces the risk of electrical shock by providing an escape wire for the electrical current. This Battery Charger Power Supply is equipped with a cord having a grounding wire with a grounding plug. The plug must be plugged into an outlet that is properly installed and grounded.
dANGER
Improper use of the grounding plug can result in a risk of electrical shock.
WARNING
Ground reliability can only be achieved when the power cord is connected to an equivalent receptacle marked
"Hospital Grade".
6
OPERATING INSTRUCTIONS
CAUTION
Inspect the Easy Go Vac Aspirator for visual damage, DO
NOT USE if damaged.
1. Place the Easy Go Vac Aspirator in the upright position on a level surface.
2. make sure the On/Off ( / ) Switch is in the “OFF” position.
3. Ensure the Inline Hydrophobic Filter and tubing is properly connected to the vacuum port on the Suction Canister lid.
CAUTION
dO NOT run the Easy GO Vac Aspirator without the Inline
Hydrophobic Filter. This will void the warranty.
4. Select appropriate mode of Operation.
A. AC ELECTRICAL RECEPTICLE OPERATION
(100-240 Volts AC) a. Insert the small plug on the cord from the Battery
Charger/Power Supply into External Power jack on the
Easy Go Vac Aspirator.
b. Connect the female end of the AC Power Cord into the
Battery Charger/Power Supply.
c. Plug the male end (3 prong plug) into a grounded receptacle.
5. Push the On/Off ( / ) Switch to the “ON” position.
6. Block the patient connection on the Suction Canister and adjust the Vacuum Control Knob to the desired vacuum.
To INCREASE vacuum - Turn knob CLOCKWISE
To DECREASE vacuum - Turn knob COUNTERCLOCKWISE
• Vacuum should be set as prescribed by your Physician or Healthcare Professional.
7. Connect one end of the Patient Suction Tubing to the port on the canister marked Patient Tubing.
8. Connect the other end of the Patient Suction Tubing to the suction catheter and begin procedure as instructed by
Healthcare Provider.
7
9. When procedure is completed, press the On/Off ( / )
Switch to the “OFF” position.
• DO NOT overfill the Canister.
10. Empty the Suction Canister as follows:
• Remove the canister from its bracket and empty as instructed by Physician or Healthcare Professional.
CAUTION
Canister is single patient use only.
11. Replace Suction Canister and reconnect all suction tubing.
B. INTERNAL BATTERY OPERATION
1. Steps 1 thru 4, see Operating Instructions on page 6.
CAUTION
•
•
•
Battery should be fully charged before using; to charge battery see Charging Battery.
When the yellow Low Battery Light is on continuously during normal use, the battery needs to be charged.
If automatic shut off occurs during battery operation, the
Easy Go Vac Aspirator cannot be restarted until the battery has been recharged, or until the Aspirator is connected to either of the external power source options.
5. Push the On/Off ( / ) Switch to the “ON” position. If yellow
Low Battery Light stays on continuously, battery needs charging. Charge battery or use another mode of Operation.
WARNING
Operating Easy Go Vac Aspirator after yellow Low Battery Light has been on continuously, will cause automatic shut down.
6. Block the patient connection on the Suction Canister and adjust the Vacuum Control Knob to the desired vacuum.
To INCREASE vacuum - Turn knob CLOCKWISE
To DECREASE vacuum - Turn knob COUNTERCLOCKWISE
• Vacuum should be set as prescribed by your Physician or Healthcare Professional.
7. Connect one end of the Patient Suction Tubing to the port on the canister marked Patient Tubing.
8
8. Connect the other end of the Patient Suction Tubing to the suction catheter and begin procedure as instructed by
Healthcare Provider.
9. When procedure is completed, press the On/Off ( / )
Switch to the “OFF” position.
• DO NOT overfill the Canister.
10. Empty the Suction Canister as follows:
• Remove the canister from its bracket and empty as instructed by Physician or Healthcare Professional.
CAUTION
Canister is single patient use only.
11. Replace Suction Canister and reconnect all suction tubing.
C. AUTO dC OPERATION
(Auto Power Port)
1. Steps 1 thru 4, see Operating Instructions on page 6.
a. Insert small plug on the Auto DC Power Cord into the
External Power jack on the Aspirator.
b. Plug the other end into the auto power port receptacle.
CAUTION
dO NOT leave the Aspirator connected to the auto power port when not in use. Failure to follow these instructions can cause severe battery failure and will VOID THE WARRANTY, see CHARGING BATTERY instructions.
5. Push the On/Off ( / ) Switch to the “ON” position. If yellow
Low Battery Light stays on continuously, battery needs charging. Charge battery or use another mode of operation.
WARNING
Operating Easy Go Vac Aspirator after yellow Low Battery Light has been on continuously, will cause automatic shut down.
6. Block the patient connection on the Suction Canister and adjust the Vacuum Control Knob to the desired vacuum.
To INCREASE vacuum - Turn knob CLOCKWISE
To DECREASE vacuum - Turn knob COUNTERCLOCKWISE
• Vacuum should be set as prescribed by your Physician or Healthcare Professional.
9
7. Connect one end of the Patient Suction Tubing to the port on the canister marked Patient Tubing.
8. Connect the other end of the Patient Suction Tubing to the suction catheter and begin procedure as instructed by
Healthcare Provider.
9. When procedure is completed, press the On/Off ( / )
Switch to the “OFF” position.
• DO NOT overfill the Canister.
10. Empty the Suction Canister as follows:
• Remove the canister from its bracket and empty as instructed by Physician or Healthcare Professional.
CAUTION
Canister is single patient use only.
11. Replace Suction Canister and reconnect all suction tubing.
CHARGING BATTERY
1. Insert small plug on cord from the Battery Charger/Power
Supply into External Power jack on the Easy Go Vac Aspirator.
2. Connect the female end of AC Power Cord into the Battery
Charger/Power Supply.
3. Plug male end (3 prong plug) into a grounded receptacle.
Amber light on Battery Charger/Power Supply will come on.
When amber light changes to green, battery is fully charged.
Average charging time is 2-4 hours.
(A new battery may take longer to fully charge the first time.)
Note: No damage will occur if the charger is left connected.
4. Disconnect AC Power Cord from AC receptacle.
5. Disconnect cord from External Power jack on Aspirator. When fully charged, battery will provide approximately 120 minutes of operation in 8, 15 minute intermittent duration periods.
CLEANING/MAINTENANCE
WARNING
•
•
Turn the Easy Go Vac Aspirator off and disconnect from any external power source before cleaning.
dO NOT clean the Easy Go Vac Aspirator with Alcohol,
Acetone or other solvents.
10
Inline Hydrophobic Filter:
This filter cannot be cleaned. If it becomes contaminated or clogged, it must be replaced.
To replace Inline Hydrophobic Filter:
1. Disconnect filter assembly from Suction Canister and remove Suction Canister from the Aspirator.
2. Unscrew filter assembly from the Aspirator.
3. Dispose of filter.
4. Install new filter (Part # 502438) by screwing filter into the
Aspirator, until hand tight. Then reconnect tubing and elbow.
CAUTION
dO NOT use sealing material on the filter threads.
Tubing is single patient use only. DO NOT clean.
Easy Go Vac Case:
Use a cloth or sponge dampened with soap and water.
Wipe dry with a clean cloth or paper towel.
RETURNS
Returned products require a Returned Goods Authorization
(RGA) number, contact Precision medical, Inc. All returns must be packaged in sealed containers to prevent damage. Precision medical, Inc. will not be responsible for goods damaged in transit.
Refer to Precision medical, Inc. Return Policy available on the
Internet, www.precisionmedical.com.
dISPOSAL INSTRUCTIONS
Dispose of Bio-waste products as described by your Healthcare
Professional or Healthcare Facility protocol.
•
•
WARNING
The Easy Go Vac Aspirator contains a sealed lead acid battery.
dO NOT dispose with normal trash. Consult your local waste management authority for proper disposal procedures.
If Aspirator is used for treatment of infectious disease, consult your Physician or Healthcare Professional for recommended procedures for proper disposal.
Please Recycle
11
TROUBLESHOOTING
If the Easy Go Vac Aspirator fails to function, consult the
Troubleshooting Guide.
There are no user serviceable parts inside. DO NOT open case.
If the problem cannot be corrected, consult your Equipment
Provider.
TROUBLESHOOTING GUIdE
Problem
Easy Go Vac
Aspirator will not operate on battery
Probable Cause Remedy
1.
On/Off ( / )
Switch in “OFF” position
2. Battery not charged
(Low battery light on)
1.
On/Off ( / )
Switch to “ON’’ position
2. Recharge battery
3. If battery will not charge, contact your
Equipment Provider
Easy Go Vac
Aspirator will not operate when connected to AC
Power Source
Easy Go Vac
Aspirator will not operate when connected to
Auto DC Power
Cord
1.
On/Off ( / )
Switch in “OFF” position
2. AC Power Cord not connected
1.
On/Off ( / ) in “OFF” position
2. Auto DC Power
Cord not connected properly
3. Defective auto power port receptacle
Switch
4. Defective Auto DC
Power Cord
1.
On/Off ( / ) to “ON” position
2. Check connection at both AC Power
Source and
Aspirator
Switch
1.
On/Off ( / )
Switch to “ON” position
2. Check connections at both ends of cord
3. Connect to a known working auto power port
4. Contact your
Equipment Provider
Continued on the next page.
12
Problem
Battery will not charge
Unable to adjust vacuum
Probable Cause
1. Not connected to
Battery Charger /
Power Supply
2. Defective Battery
Charger / Power
Supply
3. Defective battery
Remedy
1. Connect Battery
Charger/Power
Supply
2. Contact your
Equipment Provider
1. Filter assembly not connected properly
2. Suction Canister not connected properly
3. Filter clogged
3. Contact your
Equipment Provider
1. Confirm and tighten connections
2. Confirm and tighten connections
4. Defective Suction
Canister
5.
Defective Easy Go
Vac Aspirator
3. Replace filter assembly
4. Replace Suction
Canister
5. Contact your
Equipment Provider
REPLACEMENT PARTS description Part #
Filter
Filter Assembly
502438
503293
800cc Canister, (Case-10 ea.) 502519-10
Patient Suction Tubing 1955
Charger
Carry Bag
503974
502956
AC Power Cord (65)
Auto DC Power Cord
503106
502954
13
LIMITEd WARRANTY
ANd
LIMITATION OF LIABILITY
Precision medical, Inc. warrants that EASY GO VAC Aspirator
(the Product) and the following component parts thereof will be free of defects in workmanship and/or material for the following periods:
Easy Go Vac Aspirator Two (2) years from shipment
Battery Six (6) months from shipment
Should any failure to conform to this warranty appear within the applicable period, Precision medical, Inc. shall, upon written notification thereof and substantiation that the goods have been stored, installed, maintained and operated in accordance with Precision medical, Inc.’s instructions and standard industry practice, and that no modifications, substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense.
ORAL STATEmENTS DO NOT CONSTITUTE WARRANTIES.
The representatives of Precision medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale. Thus, this writing is a final, complete and exclusive statement of the terms of that contract.
THIS WARRANTY IS ExCLUSIVE AND IS IN LIEU OF ANY WARRANTY
OF mERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE
OR OTHER WARRANTY OF QUALITY, WHETHER ExPRESS OR
ImPLIED.
Precision medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property. Correction of non-conformities as provided above shall constitute fulfillment of all liabilities of Precision medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision medical, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or specifications without notice.
Precision medical, Inc. reserves the right to correct clerical or typographical errors without penalty.
14
manufacturer:
dECLARATION OF CONFORMITY
Precision medical, Inc.
300 Held Drive, Northampton, PA 18067, USA
CONTACT: Quality manager
Phone: 610-262-6090
Authorized European Representative: Emergo Europe molenstraat 15
2513 BH, The Hague
The Netherlands
Product: model(s): mDD Class:
Easy Go Aspirator
Pm65HG, Pm65E
IIa
Classification criteria: Clause 3.1 Rule 11 of Annex Ix of mDD
As delivered, the object of the declaration described above is in conformity with the requirements of mDD 93/42/EEC Annex II.3 and the following documents:
Document Title mDD-EEC/93/42 Council Directive Concerning medical Devices
Edition
1993
1990 BS EN 60601-1 medical Electrical Equipment-General requirements for safety
BS EN 60601-1-2 medical Electrical Equipment-Collateral Standard:
Electromagnetic compatibility
EN ISO 10079-1 medical Suction Equipment-Part 1: Electrically
Powered Suction Equipment-Safety Requirements
2002
2000
EN 1441 Risk Analysis 1998
Notified Body: TÜV Rheinland Products Safety GmbH CE 0197
EC Certificate No: HD60002285, SY60008447, 74_500_2415
There are no limitations on the validity of the declaration of conformity www.precisionmedical.com
ISO 13485 Certified
502598 Rev7(E) 01/05/07 (?m) Printed in USA