BC Biomedical SA-2005, -INTL, -AUS Electrical Safety Analyzer User Manual
Below you will find brief product information for Electrical Safety Analyzer SA-2005, Electrical Safety Analyzer SA-2005-INTL, Electrical Safety Analyzer SA-2005-AUS. The SA-2005 Series is a Microprocessor based Electrical Safety Analyzer. It allows for a multitude of tests to be performed on a device using the same unit and lead connections.
Advertisement
Advertisement
ELECTRICAL
SAFETY ANALYZER
SA-2005
SA-2005-INTL SA-2005-AUS
USER MANUAL
BC BIOMEDICAL
SA-2005 SERIES
TABLE OF CONTENTS
WARNINGS, CAUTIONS, NOTICES ............................................................................ ii
DESCRIPTION ............................................................................................................. 1
LAYOUT ....................................................................................................................... 4
TESTING ........................................................................................................................ 9
COMMUNICATION PROTOCOL ................................................................................ 14
MANUAL REVISIONS .................................................................................................. 16
LIMITED WARRANTY .................................................................................................. 16
SPECIFICATIONS ....................................................................................................... 17
NOTES ......................................................................................................................... 20
This User Manual covers the following units:
SA-2005 & SA-2005-R
SA-2005-INTL & SA-2005-INTL-R
SA-2005-AUS & SA-2005-AUS-R i
WARNING - USE
The SA-2005 is intended for testing only and should never be used in diagnostics, treatment or any other capacity where they would come in contact with a patient.
WARNING - CONNECTIONS
All connections to patients must be removed before connecting the DUT to the SA-2005. A serious hazard may occur if the patient is connected when testing with the SA-2005. Do not connect any leads from the patient directly to the SA-2005 or DUT while it is powered by the
SA-2005.
WARNING - MODIFICATIONS
The SA-2005 is intended for use within the published specifications. Any application beyond these specifications or any unauthorized user modifications may result in hazards or improper operation.
WARNING - CLEANING
Disconnect Line Power to the SA-2005 before attempting to clean it. Do not immerse. The SA-
2005 should be cleaned by wiping gently with a damp, lint-free cloth. A mild detergent can be used if desired.
WARNING - LIQUIDS
Do not submerge or spill liquids on the SA-2005.
In the event of a spill onto the SA-2005, Do not operate the SA-2005 regardless of fluid type.
ii
WARNING - VOLTAGE
When the SA-2005 is in Lead Isolation mode and the ISO key is depressed, 110% of line voltage is applied to the Patient lead connectors and/or
External test leads. Although this voltage is applied through an internal current limiting resistance of 121 kΩ (per standard test specifications), Do not touch the test leads, connections or DUT while the ISO key is depressed.
CAUTION - USAGE
The SA-2005 is not a continuous duty device, it is intended for short duration testing within the current limits and duty periods specified. Do not leave the DUT connected to the SA-2005 for extended time periods.
Do not to drop the SA-2005.
CAUTION - ENVIRONMENT
Exposure to environmental conditions outside the specifications can adversely affect the performance of the SA-2005. Allow SA-2005 to acclimate to specified conditions for at least 30 minutes before attempting to operate it.
CAUTION - INSPECTION
The SA-2005 should be inspected before each use for obvious signs of abuse or wear. The SA-
2005 should not be used and should be serviced if any parts are in question.
CAUTION - SERVICE
The SA-2005 is intended to be serviced only by authorized service personnel. Troubleshooting and service procedures should only be performed by qualified technical personnel.
iii
CAUTION - FUSE
Only replace the SA-2005 fuse with the specified type and rating.
iv
NOTICE
– SYMBOLS
Symbol Description
Caution
(Consult Manual for Further Information)
Electrical Caution
(Consult Manual for Further Information)
Per European Council Directive 2002/95/EC, do not dispose of this product as unsorted municipal waste.
Ω
PC
Lbs
RH
RMS
USA
V
VA
VAC
W
C cm
°
DUT
Euro ft
FS
Hz
IEC
ISO kg kHz kΩ
LED
MAP
MHz
µA mA mm
NOTICE
– ABBREVIATIONS
Amp
AAMI
NEMA
Ampere(s)
Association for the Advancement of
Medical Instrumentation
Celsius centimeter(s) degree(s)
Device Under Test
European feet
Full Scale hertz
International Electrotechnical Commission
Isolation kilogram(s) kilohertz kilohm(s)
Light Emitting Diode
Mains on Applied Parts
Megahertz microampere(s) milliampere(s) millimeter(s)
National Electrical Manufacturers
Association
Ohm(s)
Personal Computer pounds
Relative Humidity
Root Mean Square
United States of America
Volt(s)
Volt-Ampere(s)
Volt(s) Alternating Current
Watt(s)
v
NOTICE
– PERFORMING TESTS
REFER TO DUT MANUFACTURER’S SERVICE MANUAL FOR
TEST PROCEDURES AND MEASUREMENT LIMITS
NOTICE
– DISCLAIMER
BC GROUP INTERNATIONAL, INC. WILL NOT BE
RESPONSIBLE FOR ANY INJURIES SUSTAINED DUE TO
UNAUTHORIZED EQUIPMENT MODIFICATIONS OR
APPLICATION OF EQUIPMENT OUTSIDE OF THE
PUBLISHED INTENDED USE AND SPECIFICATIONS.
NOTICE
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT
TO MAKE CHANGES TO ITS PRODUCTS OR
SPECIFICATIONS AT ANY TIME, WITHOUT NOTICE, IN
ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND
TO SUPPLY THE BEST POSSIBLE PRODUCT. THE
INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY
CHECKED AND IS BELIEVED TO BE ACCURATE. HOWEVER,
NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES.
NOTICE
– DISCLAIMER
.
– CONTACT INFORMATION
BC BIOMEDICAL
BC GROUP INTERNATIONAL, INC.
3081 ELM POINT INDUSTRIAL DRIVE
SA-2005 Series User Manual www.bcgroupintl.com
08/12
ST. CHARLES, MO 63301
USA
1-800-242-8428
1-314-638-3800 www.bcgroupintl.com
Copyright © 2012
Made in the USA
Rev 01
vi
BC BIOMEDICAL
SA-2005 SERIES
ELECTRICAL SAFETY ANALYZER
The Model SA-2005 Series is a Microprocessor based Electrical Safety Analyzer. It allows for a multitude of tests to be performed on a device using the same unit and lead connections. The following are highlights of some of the main features:
SA-2005:
LED STATUS INDICATORS
AUDIO FEEDBACK
TOUCH CONTROL KEYS
– NO KNOBS
5 UNIVERSAL PATIENT LEAD INPUTS
HIGH IMPACT PLASTIC CASE
LINE VOLTAGE MEASUREMENT
DEVICE UNDER TEST CURRENT MEASUREMENT
EARTH / GROUND LEAD RESISTANCE
EARTH / GROUND LEAKAGE CURRENT
ENCLOSURE / CHASSIS LEAKAGE CURRENT
EXTERNAL RESISTANCE
EXTERNAL LEAKAGE CURRENT
SOURCE WIRING INTEGRITY MONITOR
TRUE RMS MEASUREMENTS
AAMI ES1-1993 or IEC 601 SELECTABLE TEST LOADS
90 TO 264 VAC OPERATION
20 AMP RATING (SA-2005 and SA-2005-INTL Models)
10 AMP RATING (SA-2005-AUS Models)
SELF TEST POINTS
EXTERNALLY REPLACEABLE GROUND FUSE
AUTOMATIC LOAD REVERSAL DELAY
PATIENT LEAD TO LEAD LEAKAGE CURRENT
PATIENT LEAD TO EARTH / GROUND LEAKAGE CURRENT
PATIENT ISOLATION LEAKAGE CURRENT
EXTERNAL ISOLATION LEAKAGE CURRENT
1
AVAILABLE MODELS:
SA-2005
– STANDARD MODEL WITH HOSPITAL-GRADE NEMA 5-15P LINE
PLUG AND HOSPITAL-GRADE NEMA 5-20R DUT TEST RECEPTACLE FOR
USE IN THE US AND OTHER COMPATIBLE COUNTRIES
SA-2005-INTL
– INTERNATIONAL MODEL, IEC C20 RECEPTACLE PIGTAIL
(MUST USE COUNTRY-SPECIFIC LINE CORD
– SEE ACCESSORIES
SECTION) AND UNIVERSAL DUT TEST RECEPTACLE THAT WORKS WITH
THE FOLLOWING COUNTRY-SPECIFIC PLUGS: o NEMA 5-15P, NEMA 5-20P, NEMA 6-15P and NEMA 6-20P (US/NORTH
AMERICA) o UK1-13P and UK3-5P (UK) o SW1-10P (SWITZERLAND) o IT1-10P (ITALY) o IS1-16P (ISRAEL) o JA1-15P (JAPAN) o EU1-16P (EURO) CEE
7/7 “SCHUKO” (NOTE: MUST USE SCHUKO
GROUNDING ADAPTER TO PERFORM LEAKAGE MEASUREMENTS,
SEE OPTIONAL ACCESSORIES SECTION) o DE1-13P (DENMARK) (NOTE: NO EARTH/GROUND PIN, THEREFORE
LEAKAGE MEASUREMENTS NOT APPLICABLE) o EUROPLUG CEE 7/16 (NOTE: NO EARTH/GROUND PIN, THEREFORE
LEAKAGE MEASUREMENTS NOT APPLICABLE)
SA-2005-AUS
– AUSTRALIAN MODEL, AU1-10P LINE PLUG AND AU1-10R
DUT TEST RECEPTACLE FOR USE IN AUSTRALIA/NEW ZEALAND AND
OTHER COMPATIBLE COUNTRIES
OPTIONS
-R ADDS RS-232 SERIAL COMMUNICATIONS
AVAILABLE MODELS: o SA-2005-R o SA-2005-INTL-R o SA-2005-AUS-R
STANDARD ACCESSORIES
BC20-20110
BC20-30107
UF-0250-01
BC20-204XX
OPTIONAL ACCESSORIES
8 FT CHASSIS TEST LEAD (BLACK)
SOFT CARRYING CASE
REPLACEMENT GROUND LEG FUSE
(SA-2005-INTL ONLY)
BC20-20111
BC20-20112
BC20-20113
8 FT EXTERNAL TEST LEAD (RED)
16 FT CHASSIS TEST LEAD (BLACK)
16 FT EXTERNAL TEST LEAD (RED)
2
BC20-20200
BC20-20221
BC20-41337
BC20-41339
CS-2000-U
USB TO RS-232 ADAPTER
1 AMP CURRENT SOURCE - USA/NORTH AMERICA
CS-2000-E 1 AMP CURRENT SOURCE - EURO (SCHUKO)
SA-2005-INTL LINE CORDS: (REFER TO PAGE 8 FOR DETAILS)
BC20-20400
BC20-20401
BC20-20402
BC20-20403
BC20-20409
BC20-20410
BC20-20412
BC20-20416
REPLACEMENT SA-2005-INTL DUT INTERNATIONAL
RECEPTACLE ADAPTER
SCHUKO GROUNDING ADAPTER
RS-232 COMMUNICATIONS CABLE (DB-9M to DB-9F)
NEMA 5-20P PLUG LINE CORD (USA/NORTH AMERICA)
JA1-15P PLUG LINE CORD (JAPAN)
UK1-13P PLUG LINE CORD (UK)
CEE 7/7 “SCHUKO” PLUG LINE CORD (EURO)
NOTE: INCLUDES BC20-20221 GROUNDING ADAPTER
BSS546A PLUG LINE CORD (INDIA/SOUTH AFRICA)
SW1-10P PLUG LINE CORD (SWITZERLAND)
IT1-10P PLUG LINE CORD (ITALY)
AUSTRALIA PLUG LINE CORD
3
LAYOUT
This section looks at the layouts of the SA-2005 Models and gives descriptions of the elements that are present.
For all models, the user controls all Analyzer functions via light touch keys. There is an audio click when any key is depressed
, while a “Razz” or error tone consisting of a rapid succession of beeps is sounded if an invalid key is depressed. Analyzer modes and status are indicated by 29 LEDs as well as 3 Neon lamps. A 3-1/2 digit LCD display conveys DUT test results to the user.
3 Neon indicators for
Incoming Power status
External Lead
Connector
5 Universal Patient Lead Inputs
Test Snaps:
100 uA
1.0 Ω
Earth/Ground
Fuse
Chassis Connector
3-1/2 digit LCD
Display
LCD Unit Indicators:
Volts
Amps
µAmps
Ohms
Load Selection switch:
AAMI, IEC
Analyzer Mode keys & indicators:
Mains Voltage
Device (DUT) Current
Patient Lead
Selection keys &
Earth Resistance
Earth/Ground Leakage indicators:
Enclosure Leakage
Lead to Earth Leakage
Lead to Lead Leakage
ISO Test key
& indicator
Lead Isolation Leakage
External Leakage
NEMA 5-20R
Receptacle for
DUT
RS-232 Communication Port
(Note: Only with Option
–R)
Analyzer Receptacle Control keys & indicators:
Hot
Neutral
Ground
Polarity
SA-2005 Layout (SA-2005-R shown)
4
Universal International
Receptacle for DUT
Australian Receptacle for DUT
SA-2005-INTL Layout SA-2005-AUS Layout
(SA-2005-INTL-R shown)
The SA-2005-INTL and SA-2005-AUS Model Layouts are shown above, note that the only differences from the standard model are the Line cords and the DUT test
Receptacles.
Note: The RS-
232 connector isn’t shown for the SA-2005-AUS above, but is an available option (see previous section).
5
Display
The main information in the system is presented in the 3-1/2 digit LCD display. This data is provided as simple meter readings with the measurement units indicated to the right by one of four LED indicators.
Function Selection
Nine LEDs and two keys make up the Function Selection Section. The keys are up and down arrows. When depressed, they step the Analyzer through the available options.
The LED next to the currently selected option is illuminated.
Load Selection
The unit may either use the AAMI ES1-1993 or IEC 601 Test load for measurements.
This is selected by the Load Selection switch.
Analyzer Test Receptacle Control
There are four keys and 8 LEDs in the Analyzer Test Receptacle Control Section. They allow the manual control of the power connections that are made to the DUT. Internally, a series of relays are switched by the microprocessor based on the keys that are depressed. The LEDs indicate the current state of the power connections to the
Receptacle.
Note: The Forward/Reverse Polarity key has a safety delay feature, preventing damage to the internal relays and the DUT. When the key is depressed, the DUT power is disabled and the safety delay is activated. When this delay is complete, the internal relays switch the polarity to the DUT and apply power. This delay allows any reactive power stored in the DUT to self-discharge before the polarity is reversed.
Note: The unit will power up with the Neutral and Ground Closed, in Forward Polarity and with the Hot Open. It is recommended that the unit be returned to this condition when plugging and unplugging the DUT.
Patient Lead Control
There are five patient leads. During testing, it is necessary to select between these leads, select all of them and apply High Voltage to them. This section provides the control keys to do these test configurations and the LEDs to indicate the current state.
There is one LED and an internal relay for each lead. The LEDs and the markings below each LED indicate when that relay is on, thus selecting the indicated lead. The
Up and Down arrow keys sequentially select each lead in order and scroll through from
None, to 1-5 to All and around again.
To apply High Voltage to the leads, the
“ISO” Test key is depressed. It is only active in the Lead Isolation mode. It is a momentary key so that High Voltage is only applied while the key is held down.
6
Power Outlet Indicators
Three Neon indicators help verify the polarity and wiring of the wall receptacle that the
Safety Analyzer is plugged into.
Self-Test Snaps
There are two snaps on the side of the unit that allow for a quick self-test of the
Analyzer. They provide a fixed 1.0 Ohm resistance to Earth/Ground and a 100 µAmp leakage current source to Earth/Ground when the Chassis lead is applied and the
Analyzer set to the proper mode.
Connectors
There are two connectors for test cables on the unit. One is for the Chassis lead and the other is for one of two different leads used for external testing. The test cables simply plug into the sockets. There is a release pin on the cable plug that must be depressed to remove the cable.
Fuse
There is a fuse in the ground leg of the Analyzer Test Receptacle. This is to help prevent damage from excess ground current. It is located on the face for ease of replacement.
Test Receptacle
This receptacle is for the connection of the DUT. The Receptacle Rating depends on the specific Analyzer model. SA-2005 models use a Hospital Grade North
American/USA standard NEMA 5-20R receptacle rated 20 Amps @ 125 VAC. SA-
2005-INTL models use a universal international receptacle rated 20 Amps @ 250 VAC.
SA-2005-AUS models use an Australian-specific AU1-10R receptacle rated 10 Amps @
240 VAC. An external patch cord may be necessary to connect devices utilizing different types of plugs to the Analyzer receptacle.
Power Cord
The Power Cord, which is connected internally, provides power to both the Safety
Analyzer and the DUT through the Test Receptacle. The Power Cord varies depending on the Analyzer model. SA-2005 models use a NEMA 5-15P plug designed to plug into a NEMA 5-15R or 5-20R Receptacle. SA-2005-INTL models have a short IEC C20 plug for which a country-specific adapter cable must be connected
– See the following section for this information. SA-2005-AUS models use an Australian AU1-10P plug intended for operation with Australian and New Zealand AU1-10R Receptacles.
7
Universal Power Cord which
Connects to Standard Product
Plug
Adapter Cable with Standard
Product Plug and Required
International Connector
(See Options Below)
Communications (Optional)
On Analyzer models with the
–R option, a DB-9 RS-232 Communications port is provided to interface the analyzer with a PC. Data from the Analyzer display may be transferred over this communications link as well as full control over all analyzer functions and settings. For further details, see the Communication Protocol section.
8
TESTING
The SA-2005 Models allow the user a great deal of flexibility in testing. Any of the basic tests can be run and in almost any sequence. The information in this section presents a systematic approach that is just one way to proceed. It is only presented as a guide and it is the responsibility of the user to establish which tests are required.
NOTICE
– PERFORMING TESTS
REFER TO DUT MANUFACTURER’S SERVICE MANUAL FOR
TEST PROCEDURES AND MEASUREMENT LIMITS
.
The Analyzer requires a good Earth/Ground connection for operation. It should be plugged into a “Hospital Grade” receptacle where available. This is necessary for both valid test results and personal safety.
Power Receptacle Confirmation
Once plugged in, the first step is to ensure that the wall receptacle the Safety Analyzer is plugged into is wired properly (120 VAC Non-Isolated Power Systems Only). There are three neon indicators in the unit that provide this confirmation. The REV indicator is red and the other two are green. If the two green indicators are on, the receptacle is wired correctly. If not, utilize the following patterns to help determine the problem. Do not proceed with any testing until you get only the two green lights.
NOTE: Neutral/Ground Reversal is not checked.
NOTE: These Indicators are valid only for 120 VAC Non-Isolated Power Systems.
Mains Voltage
With the Mains Voltage function selected, the display will show the Voltage that is present on the incoming power lines. This is measured from Line to Neutral. Note that the voltage may drop when the DUT is turned on. Ensure that this value is within the
DUT specifications.
9
Device Current
With the Device Current function selected, the display will show the current draw of the
DUT. The Receptacle should be configured with HOT-CLOSED, NEUTRAL-CLOSED,
GROUND-CLOSED and POLARITY-FWD. Refer to the Specifications section for current capacity and permitted duty cycle for this test mode.
Earth Resistance
With the Earth Resistance function selected, the display will show the resistance between the Chassis Test lead and Receptacle Earth/Ground. This resistance is a combination of the resistance within the DUT enclosure and the resistance of the
Earth/Ground Lead in the DUT power cord.
NOTE: This test has no meaning for equipment that does not use a grounded cord.
The test requires that the Chassis Test lead be plugged into the Chassis Connector.
The other end should be connected to a solid ground point on the DUT.
The display is in hundredths of
Ω and will read to 19.99 Ω. Over-range shows as
“1_ _ _”.
Earth/Ground Leakage Current
With the Earth/Ground Leakage function selected and the Ground-Open, the display will show the leakage current in the ground wire of the DUT.
NOTE: This test has no meaning for equipment that does not use a grounded cord.
Selecting this function automatically opens the connection to Earth/Ground and passes any leakage current through a 1000
Ω load with either AAMI ES1-1993 or IEC 601 frequency compensation as selected by the Load Selection switch.
Enclosure Leakage
With the Enclosure function selected, the display will show the leakage current between the Enclosure (Chassis) and Earth/Ground.
The test requires that the Chassis Test lead be plugged into the Chassis Connector.
The other end should be connected to a solid ground point on the DUT.
NOTE: If a non-conductive enclosure is used, a 200 cm
2
conductive foil pad should be used. This foil is to be placed in close contact with the enclosure and connected to the
Chassis Test lead.
Any leakage current will flow through the Chassis Test lead and then through a 1000
Ω load with either AAMI ES1-1993 or IEC 601 frequency compensation as selected by the
Load Selection switch.
10
Lead to Earth/Ground Leakage
With the Lead to Earth/Ground function selected, the display will show the leakage current between the selected lead and Earth/Ground.
Attach the patient leads to the connectors on the top of the Safety Analyzer. The Up and Down arrow keys may then be used to select any individual lead or all of the leads.
This test should be done for each lead individually and all leads together.
This test measures the leakage current that would flow through the leads if the patient were to come into contact with Earth/Ground.
Lead to Lead Leakage
With the Lead to Lead function selected, the display will show the leakage current between the selected Patient Lead and all other patient leads.
Attach the patient leads to the connectors on the top of the Safety Analyzer. The Up and Down arrow keys may then be used to select any individual lead. Internally, relays connect the leads as necessary. The LEDs indicate the selected lead.
This test should be done for each lead individually.
This test measures the current that would flow from a lead to other leads. Normally these are Auxiliary currents from bias, measurement and sensing circuits.
Lead Isolation
With the Lead Isolation function selected and the
“ISO” Test key depressed, the display will show the leakage current between the selected Patient Lead(s) and Earth/Ground.
WARNING
The SA-2005 applies 110% of line voltage to the
Patient leads or External test leads during the
Isolation test. Although this is current limited by a 121 kΩ internal resistor, per standard test specifications, care should be taken to prevent contact with this voltage. Do not touch the test leads, connections or DUT while depressing the
Isolation Test key.
Attach the patient leads to the connectors on the top of the Safety Analyzer. The Up and Down arrow keys may then be used to select any individual lead and all of the leads.
11
As each lead and then All leads are selected, depress and hold the
“ISO” Test key.
This will apply 110% of the line voltage through a 121 kΩ resistor to the selected lead(s) and measure the current that flows to Earth/Ground through a 1000
Ω load with either
AAMI ES1-1993 or IEC 601 frequency compensation as selected by the Load Selection switch.
This test is to be done for each lead individually and All leads together.
This test measures the leakage current that would flow through the lead(s) if the patient were to come into contact with Line voltage. This is referred to as MAP (MAINS on
Applied Parts).
Point to Point Measurements
The unit has the ability to measure Leakage Current, Isolation Leakage Current and
Resistance between two points, utilizing two test leads. These tests are separated because they use a slightly different setup than the previous tests.
Point to Point Leakage Current
With the External function selected, the display will show the leakage current between the test leads.
The test requires that the Chassis Test lead be plugged into the Chassis Connector and the External Test lead be plugged into the External Connector. The other ends of the leads are then attached to the points of interest.
Any current flowing between the test points is passed through a 1000
Ω load with either
AAMI ES1-1993 or IEC 601 frequency compensation as selected by the Load Selection switch.
Point to Point Isolation Leakage Current
WARNING
The SA-2005 applies 110% of line voltage to the
Patient leads or External test leads during the
Isolation test. Although this is current limited by a 121 kΩ internal resistor, per standard test specifications, care should be taken to prevent contact with this voltage. Do not touch the test leads, connections or DUT while depressing the
Isolation Test key.
With the External function selected and the
“ISO” Test key depressed, the display will show the isolation leakage current between the test leads.
12
The test requires that the Chassis Test lead be plugged into the Chassis Connector and the External Test lead be plugged into the External Connector. The other ends of the leads are then attached to the points of interest.
Depress and hold the Isolation Voltage key. This will apply 110% of the line voltage through a 121 kΩ resistor to the test leads and measure the current that flows through a
1000
Ω load with either AAMI ES1-1993 or IEC 601 frequency compensation as selected by the Load Selection switch.
Point to Point Resistance
With the Earth Resistance function selected, the display will show the resistance between the two test leads.
The test requires two Chassis Test leads, one in the Chassis Connector and the second in the External Connector. The opposite cable ends should be connected to the points of interest.
NOTE: Remove any device plugged into the Analyzer Test Receptacle. If there is a DC voltage present between the two test points, the reading may contain an error. This can be checked by reversing the connections. If the readings differ, average the two to get the actual resistance value.
13
COMMUNICATION PROTOCOL
The communication protocol provides a means to completely configure and use the
Analyzer from a PC or other device with RS-232 communications interface. This provides for hands free or automated operation of the equipment.
Communication Port
The Serial port is configured as 115,200 Baud Rate, 8 Data Bits, 1 Stop Bit, and No
Parity.
Command Syntax
The command description is broken into columns; the KEYWORD, the NODE and the
VALUE.
The KEYWORD provides the name of the command. The actual name of the command consists of one or more keywords since SCPI commands are based on a hierarchical structure, also known as a tree system.
In such a system, associated commands are grouped together under a common node in the hierarchy, analogous to the way leaves at a same level are connected at a common branch. This and similar branches are connected to fewer and thicker branches, until they meet at the root of the tree. The closer to the root, the higher a node is considered in the hierarchy. To activate a particular command, the full path to it must be specified.
This path is represented in the following tables by placing the highest node in the leftmost position. Further nodes are indented one position to the right, below the parent node.
The highest level node of a command is called the Keyword, followed by the Node, and then the value.
Some commands allow for reading and writing data and some commands are Read
Only. To indicate a read function, a question mark (?) is placed at the end of the command path. For example, to change the mode to Earth Resistance measurement,
“CONFigure:MODE ERESistance<cr>”, where <cr> indicates a carriage-return. To read the current operating mode of the device, a mode read command would be
“CONFigure:MODE?<cr>”, which would return the current operating mode of the device.
Lowercase letters indicate the long-form of the command (for example,
CONFigure:MODE MVOLtage) and can be omitted for simplification. Uppercase letters indicate the abbreviated, or short-form, of the commands and must be included
(for example, CONF:MODE MVOL).
All commands sent to the unit are terminated with a Carriage Return.
NOTE: Commands can be entered in either upper or lowercase or a mixture of the two,
14
uppercase and lowercase. Commands sent to the device are not case sensitive. Upper case characters in the command summary table below are used to show the shortened versions of the commands if available.
SA-2005-R Communication Command Summary
Keyword Node
Value
Description
CONFigure MODE MVOLtage
DCURrent
ERESistance
EGROund
ENCLosure
LEARth
LLEAd
LISOlation
EXTernal
Mains Voltage
Device Current
Earth Resistance
Earth / Ground Leakage
Enclosure Leakage
Lead to Earth Leakage
Lead to Lead Leakage
Lead Isolation
External Measurement
SYSTem
LEADSelect
HOT
NEUtral
GROUnd
POLarity
ISO
FILter
UNITs?
NONe n (1 to 5)
ALL
No Leads Selected n Lead only
All Leads Active
OPEN
CLOSed
OPEN
CLOSed
OPEN
CLOSed
FWD
REV
Receptacle Hot Lead Open
Receptacle Hot Lead Closed
Receptacle Neutral Lead Open
Receptacle Neutral Lead Closed
Receptacle Ground Lead Open
Receptacle Ground Lead Closed
Receptacle Polarity Forward
Receptacle Polarity Reversed (reverse
HOT/NEUTRAL)
ON
OFF
Isolation Test ON
Isolation Test OFF
AAMI
IEC
Set AAMI test load
Set IEC test load
Returns the units of Measure (Volts, Amps, uA, Ohms)
[Read Only]
MEASurement? Returns the latest measurement [Read Only]
KEY
MODel?
VERsion?
MUP
MDN
LECW
LECCW
HOT
NEUtral
GROUnd
POLarity
Changes Mode Up one position
Changes Mode Down one position
Changes Output / Lead Select Clockwise one position
Changes Output / Lead Select Counter-
Clockwise one position
Duplicates the HOT key press
Duplicates the Neutral Key press
Duplicates the Ground key press
Duplicates the Polarity key press
ISO Duplicates the ISO Test key press
Returns Safety Analyzer Model [Read Only]
Returns Firmware Version [Read Only]
15
MANUAL REVISIONS
Revision #
Rev 01
Revisions Made
Origination
LIMITED WARRANTY
WARRANTY
: BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE
FREE FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR
WHICH THEY ARE INTENDED. THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM
THE DATE OF SHIPMENT.
EXCLUSIONS: THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR
IMPLIED, INCLUDING, BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY
OR FITNESS FOR A PARTICULAR PURPOSE.
BC GROUP INTERNATIONAL, INC. IS NOT LIABLE FOR ANY INCIDENTAL OR
CONSEQUENTIAL DAMAGES.
NO PERSON OTHER THAN AN OFFICER IS AUTHORIZED TO GIVE ANY OTHER WARRANTY
OR ASSUME ANY LIABILITY.
REMEDIES: THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR
OR REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE
OPTION OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.
P:\MANUALS\BCGroup\
…\SA2005_Series_UM_Rev01.docx
16
SPECIFICATIONS
RANGE
ACCURACY
RANGE
ACCURACY
RANGE
ACCURACY
MAINS VOLTAGE
90 to 264 VAC, 50/60 Hz
± 3% of Reading, ± 1 digit
DEVICE CURRENT
0 to 19.99 Amps, RMS
± 5% of Reading, ± 1 digit
EARTH RESISTANCE
0 to 19.99 Ω
0 to 1.99 Ω
2.00 to 19.99 Ω
± 1% Range
± 1% Range
CURRENT SOURCE
RANGE
ACCURACY
LEAKAGE LOAD
LEAD ISOLATION TEST
(MAP)
FUSE
10 mA
(Note: 1 Amp Current Source Available as an optional accessory, see description section for ordering information on the CS-2000 Series)
LEAKAGE CURRENT
DC
0 to 1999 µA, RMS
± 1% Reading, ±1 digit
25 Hz up to 1 kHz
1 kHz up to 100 kHz
± 1% Reading, ±1 digit
± 2.5% Reading, ±1 digit
100 kHz to 1 MHz ± 5% Reading, ± 1 digit
1000 Ω
AAMI ES1-1993 or IEC 601
User Selectable
110% Line Voltage, ± 5%
Internally Current-Limited by 121 k
Ω resistor
250 mA, 250 V, 5x20 mm, Fast Acting
(Receptacle Ground Leg)
BC Part No. UF-0250-01
17
OPERATING LINE
VOLTAGE
LINE PLUG
DUT CURRENT CAPACITY
DUT RECEPTACLE
ELECTRICAL
90 to 264 VAC, 50/60 Hz
SA-2005 MODELS
SA-2005-INTL MODELS
SA-2005-AUS MODELS
SA-2005 & SA-2005-INTL
MODELS
SA-2005-AUS MODELS
SA-2005 MODELS
SA-2005-INTL MODELS
NEMA 5-15P
Universal IEC C20
Receptacle Pigtail Cord
Must use Country-specific
Line Cord (see
Accessories section)
AU1-10P (AS/NZS 3112)
15 Amps, 30 minutes
20 Amps, 5 minutes
10 Amps, 30 minutes
20 Amps @ 125 VAC
Capacity
Hospital Grade
NEMA 5-20R
Compatible with:
NEMA 5-15P and 5-20P
Plugs
20 Amps @ 250 VAC
Capacity
International Receptacle
Compatible with: o NEMA 5-15P, NEMA
5-20P, NEMA 6-15P and NEMA 6-20P
(US/NORTH
AMERICA) o UK1-13P and UK3-5P
(UK) o SW1-10P
(SWITZERLAND) o IT1-10P (ITALY) o IS1-16P (ISRAEL) o JA1-15P (JAPAN)
18
DUT RECEPTACLE
(continued)
ELECTRICAL (continued)
SA-2005-INTL MODELS o EU1-16P
(EURO)
CEE 7/7 “SCHUKO”
(NOTE: MUST USE
BC20-20221
GROUNDING
ADAPTER) o DE1-13P
(DENMARK) o EUROPLUG
CEE 7/16
SA-2005-AUS MODELS
10 Amps @ 240 VAC
Capacity
AU1-10R (AS/NZS 3112)
Compatible with:
AU1-10P Plugs
POWER CONSUMPTION 5 VA (5 W)
PHYSICAL & ENVIRONMENTAL
DISPLAY
Non-Backlit 3½ Digit LCD
0.5 inches (12.7 mm) Digit Height
ENCLOSURE ABS Plastic
CONSTRUCTION
OVERLAY Back-printed Lexan
SIZE
WEIGHT
OPERATING RANGE
STORAGE RANGE
8.65 x 5.73 x 2.40 Inches (219.7 x 145.5 x 61.0 mm)
≤ 3.0 Lbs (1.36 kg)
15 to 40 °C (59 to 104 °F)
10 to 90% RH, Non-Condensing
-20 to 65 °C (-4 to 149 °F)
19
NOTES
20
NOTES
21
22
BC GROUP INTERNATIONAL, INC.
3081 ELM POINT INDUSTRIAL DRIVE
ST. CHARLES, MO 63301
USA
1-800-242-8428
1-314-638-3800 www.bcgroupintl.com
SA-2005 Series User Manual
08/12
– Rev 01
Made in the USA
Copyright © 2012
Download
Advertisement
Key features
Microprocessor based
Multitude of tests
Universal Patient Lead Inputs
High Impact Plastic Case
Line Voltage Measurement
Device Under Test Current Measurement
Earth / Ground Lead Resistance
Earth / Ground Leakage Current
Enclosure / Chassis Leakage Current
External Resistance
Frequently asked questions
The SA-2005 Series can perform a variety of tests, including: Line Voltage Measurement, Device Under Test Current Measurement, Earth / Ground Lead Resistance, Earth / Ground Leakage Current, Enclosure / Chassis Leakage Current, External Resistance, External Leakage Current, Source Wiring Integrity Monitor, Patient Lead to Lead Leakage Current, Patient Lead to Earth / Ground Leakage Current, Patient Isolation Leakage Current, External Isolation Leakage Current.
The SA-2005 and SA-2005-INTL Models have a 20 AMP rating. The SA-2005-AUS Models have a 10 AMP rating.
The SA-2005 Series operates from 90 to 264 VAC, 50/60 Hz.
The SA-2005 Series has an operating temperature range of 15 to 40 °C (59 to 104 °F).