Labconco Intelli-Sense MSB CS 10", CS 12", FRP 10", FRP 12", PVC 10", PVC 12" blower User's Manual
Intelli-Sense MSB CS 10", Intelli-Sense MSB CS 12", Intelli-Sense MSB FRP 10", Intelli-Sense MSB FRP 12", Intelli-Sense MSB PVC 10", Intelli-Sense MSB PVC 12" are designed to meet the demanding requirements of most laboratory ventilation situations associated with fume hood and enclosure exhaust. The outside steel housing of the blower encloses the UL listed ECM motor, shaft, junction box wiring, and bearings. The MSB also includes a fume hood blower control box to be located on top of the fume hood, inlet power control cable, three position switch replacement, and fume hood switch label replacement.
Advertisement
Advertisement
User’s Manual
Intelli-Sense
™
Multi-Speed Blowers
Coated Steel Models
Complete Intelli-Sense MSB
Blower with Controls
CS 10"
CS 12"
All 50/60 Hz
7061110-115v, -20-230v
7061112-115v, -22-230v
Fiberglass Models
Complete Intelli-Sense MSB
Blower with Controls
FRP 10"
FRP 12"
All 50/60 Hz
7181810-115v, -20-230v
7181812-115v, -22-230v
PVC Models
Complete Intelli-Sense MSB
Blower with Controls
All 50/60 Hz
PVC 10"
7183410-115v, -20-230v
PVC 12"
7183412-115v, -22-230v
Intelli-Sense MSB Blower – No Controls (used for replacement and configured fume hood specials)
All 50/60 Hz
7061111-115v, -21-230v
7061113-115v, -23-230v
Intelli-Sense MSB Blower – No Controls (used for replacement and configured fume hood specials)
All 50/60 Hz
7181811-115v, -21-230v
7181813-115v, -23-230v
Intelli-Sense MSB Blower – No Controls (used for replacement and configured fume hood specials)
All 50/60 Hz
7183411-115v, -21-230v
7183413-115v, -23-230v
To receive important product updates, complete your product registration card online at register.labconco.com
Labconco Corporation
8811 Prospect Avenue
Kansas City, MO 64132-2696
800-821-5525, 816-333-8811
FAX 816-363-0130
E-MAIL [email protected]
HOME PAGE www.labconco.com
Please read the User’s Manual before operating the equipment.
Copyright © 2013 Labconco Corporation. All rights reserved.
The information contained in this manual and the accompanying products are copyrighted and all rights reserved by Labconco Corporation. Labconco Corporation reserves the right to make periodic design changes without obligation to notify any person or entity of such change.
Warranty
Labconco provides a warranty on all parts and factory workmanship. The warranty includes areas of defective material and workmanship, provided such defect results from normal and proper use of the equipment.
The warranty for all Labconco products will expire one year from date of installation or two years from date of shipment from Labconco, whichever is sooner, except the following;
• Purifier
® Logic ® + Biological Safety Cabinets and PuriCare ® Procedure Stations carry a fiveyear warranty from date of installation or six years from date of shipment from Labconco, whichever is sooner.
• SteamScrubber
® & FlaskScrubber ® Glassware Washers carry a two-year warranty from date of installation or three years from date of shipment from Labconco, whichever is sooner.
• Blood Drawing Chairs carry a ten year warranty.
• Carts carry a lifetime warranty.
• Glassware is not warranted from breakage when dropped or mishandled.
This limited warranty covers parts and labor, but not transportation and insurance charges. In the event of a warranty claim, contact Labconco Corporation or the dealer who sold you the product. If the cause is determined to be a manufacturing fault, the dealer or Labconco Corporation will repair or replace all defective parts to restore the unit to operation. Under no circumstances shall Labconco
Corporation be liable for indirect, consequential, or special damages of any kind. This statement may be altered by a specific published amendment. No individual has authorization to alter the provisions of this warranty policy or its amendments. Lamps and filters are not covered by this warranty.
Damage due to corrosion or accidental breakage is not covered.
Returned or Damaged Goods
Do not return goods without the prior authorization from Labconco. Unauthorized returns will not be accepted. If your shipment was damaged in transit, you must file a claim directly with the freight carrier.
Labconco Corporation and its dealers are not responsible for shipping damages.
The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen (15) days of delivery.
Limitation of Liability
The disposal and/or emission of substances used in connection with this equipment may be governed by various federal, state, or local regulations. All users of this equipment are required to become familiar with any regulations that apply in the user’s area concerning the dumping of waste materials in or upon water, land, or air and to comply with such regulations. Labconco Corporation is held harmless with respect to user’s compliance with such regulations.
Contacting Labconco Corporation
If you have questions that are not addressed in this manual, or if you need technical assistance, contact
Labconco’s Customer Service Department or Labconco’s Product Service Department at 1-800-821-5525 or 1-816-333-8811, between the hours of 7:00 a.m. and 6:00 p.m., Central Standard Time.
Part #7115900, Rev. -
ECO H871
ORIGINAL INSTRUCTIONS
T
ABLE
O
F
C
ONTENTS
CHAPTER 1: INTRODUCTION
About This Manual
1
2
3
CHAPTER 2: PREREQUISITES 5
5
Support
Electrical
Controls Wiring
Connect Blower Inlet for Coated Steel Blowers
Connect Blower Inlet for Fiberglass Blowers
Connect Blower Inlet for PVC Blowers
6
6
6
CHAPTER
7
7
Install Blower on a Supporting Structure
Adjust Outlet
8
8
9
9
9
10
Connect Blower
Coated
Outlet
Steel
For Fiberglass Blowers
For PVC Blowers
Connect to the Electrical Supply Source
Modify the Fume Hood Controls and Install the
11
11
12
Control
Adjust
Box
Fan
CHAPTER 4: PERFORMANCE DATA & SAFETY PRECAUTIONS 21
Specifications
Blower Curves
Safety Precautions
CHAPTER 5: USING YOUR BLOWER
CHAPTER 6: MAINTAINING YOUR BLOWER
22
28
29
30
Motor
Bearings
30
31
Coupler 31
CHAPTER 7: MODIFYING YOUR BLOWER, CALCULATING
STATIC PRESSURE LOSS, AND BLOWER SIZING
Main
Additional Modifications by Adding Ductwork Accessories
32 for 32
33
Sizes and Pressure Losses in Thermoplastic Duct
Typical System
34
35
Thermoplastic Duct 36
Couplings,
Elbows
Zero Pressure Weathercaps
Spiral Tube
Manual Duct Dampers
Duct
Blower
36
37
37
37
T 38 for
Exhaust
47
Transition
Wash Rings 39
Accessories for Pathogens, Organic Vapors and Odor Control Applications 39
HEPA 39
Charcoal Packs
Backdraft Dampers
39
40
Bird Screens
CHAPTER 8: TROUBLESHOOTING
40
41
APPENDIX A: BLOWER REPLACEMENT PARTS
APPENDIX B: BLOWER DIMENSIONS
43
46
APPENDIX C: BLOWER ENVIRONMENTAL CONDITIONS,
WIRING DETAIL, AND WIRING DIAGRAMS
APPENDIX D: TYPICAL INTELLI-SENSE OPERATING CURVE
APPENDIX E: TYPICAL LABORATORY CONTROLS WITH
49
52
C
HAPTER
1
I
NTRODUCTION
Congratulations on your purchase of a Labconco Intelli-Sense Multi-Speed Blower
(MSB). Your MSB with electronically commutated motor (ECM) has been specifically engineered to meet the demanding requirements of most laboratory ventilation situations associated with fume hood and enclosure exhaust. The outside steel housing of the blower encloses the UL listed ECM motor, shaft, junction box wiring, and bearings. The MSB also includes a fume hood blower control box to be located on top of the fume hood, inlet power control cable, three position switch replacement, and fume hood switch label replacement.
The contaminated air housing of the Coated Steel Blower has a protective phenolic coating. The Fiberglass Blower features a fiberglass reinforced polyester housing and an injection molded polypropylene impeller. The PVC Blower housing is formed out of polyvinyl chloride (PVC) and the impeller wheel is injection molded polyvinylidene fluoride (PVDF). The MSB’s feature cooling vents and weather cover that allow for unrestricted exterior roof mounting. The blowers also feature forward curved impellers to ensure quiet operation and optimum air delivery.
All Multi-Speed Blowers utilize a direct drive ECM motor that uses 1/3 less energy than typical AC belt drive motors. ECM’s run approximately 95% efficient whereas AC motors run about 60% efficient. Multi-Speed Blowers make it easy to size fume hood blower systems as they cover a much broader speed range. For most applications, the Intelli-Sense Blower is typically used for only one fume hood. The Intelli-Sense Multi-Speed Blower with two or three position operation can be used in standby mode or night setback mode and set to exhaust at a minimum flow rate with the fume hood sash closed. Two or three speed blower systems should be considered when the total number of fume hoods does not financially substantiate the start-up costs of a variable air volume (VAV) system.
Two or three position blower systems are less expensive to install and still provide an air volume reduction. Please reference Appendix E for typical laboratory controls with individual Intelli-Sense Blowers. Studies have shown that fume hoods on two or three position systems can reduce air flow (CFM) by 50-70% per year. For example, a 6-ft fume hood operating 24 hours a day at 100fpm (735 CFM
Product Service 1-800-522-7658 1
2
Original instructions
Chapter 1: Introduction
at $7/CFM/Yr) at the 18-in sash height at constant volume can cost $5145 per year to operate, but with a MSB used 4 hours per day and the remaining 20 hours a day in standby mode would cost only $2578 per year to operate, saving $2567 per hood annually. For this example, the $2567 annual savings is 50%.
Your Multi-Speed Blower offers many unique features to enhance performance.
To take full advantage of them, acquaint yourself with this manual and keep it handy for future reference. If you are unfamiliar with how blowers operate, review
Chapter 4: Performance Data and Safety Precautions before you begin operating the blower. Even if you are experienced, review Chapter 5: Using your Blower.
!
If the unit is not operated as specified in this manual it may impair the protection provided by the unit.
Si l'appareil n'est pas utilisé comme spécifié dans ce manuel, il peut compromettre la protection offerte par l'appareil.
About This Manual
This manual will help you learn how to install, use, and maintain your blower.
Instructions for installing optional equipment on your blower are also included.
Chapter 1: Introduction provides a brief overview of the blower, explains the organization of the manual, and defines the typographical conventions used in the manual.
Chapter 2: Prerequisites explains what you need to do to prepare your site before you install your blower. Electrical and service requirements are discussed.
Chapter 3: Getting Started contains the information you need to properly unpack, inspect, install and certify your blower.
Chapter 4: Performance Data and Safety Precautions explains how the blower operates and the appropriate precautions you should take when using the blower.
Chapter 5: Using your Blower discusses basic operation.
Chapter 6: Maintaining Your Blower explains how to perform routine maintenance on your blower.
Chapter 7: Modifying Your Blower, Calculating Static Pressure Loss, and Blower
Sizing explains how to modify the blower or add ductwork accessories.
Chapter 8: Troubleshooting contains a table of situations you may encounter while using your blower including the probable causes of the problems and suggested corrective actions.
Product Service 1-800-522-7658
Original instructions
Chapter 1: Introduction
Appendix A: Blower Replacement Parts contains labeled diagrams of all of the components of the blowers.
Appendix B: Blower Dimensions contains comprehensive diagrams showing all of the dimensions for the various blowers.
Appendix C: Blower Environmental Conditions contains the environmental conditions to operate the blower.
!
)
CS
Typographical Conventions
Recognizing the following typographical conventions will help you understand and use this manual:
• Book, chapter, and section titles are shown in italic type (e.g., Chapter 3:
Getting Started).
• Steps required to perform a task are presented in a numbered format.
• Comments located in the margins provide suggestions, reminders, and references.
• Critical information is presented in boldface type in paragraphs that are preceded by the exclamation icon. Failure to comply with the information following an exclamation icon may result in injury to the user or permanent damage to the Blower.
•
L'information critique est présenté en caractères gras dans les paragraphes qui sont précédés de l'icône d'exclamation. Le non respect de l'information suite à un point d'exclamation peut entraîner des blessures à l'utilisateur ou des dommages permanents au ventilateur.
• Critical information is presented in boldface type in paragraphs that are preceded by the wrench icon. Only a trained certifier or contractor should perform these operations. Failure to comply with the information following a wrench icon may result in injury to the user or permanent damage to your
Blower.
•
L'information critique est présenté en caractères gras dans les paragraphes qui sont précédés de l'icône de la clé. Seul un certificateur ou un entrepreneur qualifié doit effectuer ces opérations. Le non respect de l'information suite à une icône de la clé peut entraîner des blessures à l'utilisateur ou des dommages permanents à votre ventilateur.
• Important information is presented in capitalized type in paragraphs that are preceded by the pointer icon. It is imperative that the information contained in these paragraphs be thoroughly read and understood by the user.
• A letter icon precedes information that is specific to a particular blower model.
The CS icon indicates the text is specific to the Coated Steel Blower.
Product Service 1-800-522-7658 3
FRP
PVC
!
Original instructions
Chapter 1: Introduction
• The FRP icon indicates the text is specific to the Fiberglass Blower.
• The PVC icon indicates the text is specific to the PVC Blower.
• CAUTION – See Manual. When this symbol is on the unit it indicates a caution that is detailed in this manual.
•
ATTENTION – Voir le manuel. Lorsque ce symbole est allumé l'unité, il indique une mise en garde qui est détaillée dans ce manuel.
• CAUTION – Hot Surface.
• AVERTIR – Surface Chaude
If you would like to review how blowers operate, or their differentiating features go to Chapter 4: Performance Data and Safety Precautions.
For information on the operational characteristics of the blower, go to Chapter 5:
Using your Blower.
If your blower is installed and you need to perform routine maintenance on the blower, proceed to Chapter 6: Maintaining Your Blower.
For information on making modifications to the configuration of your blower, go to Chapter 7: Modifying Your Blower, Calculating Static Pressure Loss, and
Blower Sizing.
Refer to Chapter 8: Troubleshooting if you are experiencing problems with your blower.
4 Product Service 1-800-522-7658
C
HAPTER
2
P
REREQUISITES
Before you install your blower, you need to prepare your site for installation. A dedicated source of electrical power must be located near the installation site.
Carefully read this chapter to learn:
• The location requirements for your installation site.
• The mounting support requirements for your installation site.
• The electrical power requirements for your installation site.
• The fume hood controls, controls wiring, and controls modifications for your installation site.
• The space requirements for your installation site.
Refer to Appendix B: Blower Dimensions for complete blower dimensions.
Refer to Appendix C: Blower Environmental Conditions for complete environmental conditions and wiring diagram.
Location Requirements
In positioning your blower, care should be taken to make sure that it is away from all other types of air handling equipment (intake fans, air conditioning units, etc.).
Your blower should also include a minimum of 7 ft. of ducting above the roofline to generate proper air dispersion of materials being exhausted through the blower.
Also consider the location of the blower inlet with respect to the hood you are exhausting. Proper planning and layout are essential in selecting a blower location.
Product Service 1-800-522-7658 5
Original instructions
Chapter 2: Prerequisites
Mounting Support Requirements
You must provide vibration isolators, vibration mounting pads, and/or a roof curb support for proper mounting of the blower. Vibration isolators or vibration mounting pads are available from many sources such as a local industrial supply company. Labconco recommends supporting the blower with 5/16" diameter mounting hardware.
Electrical Power Requirements
Please refer to the wiring diagrams in Appendix C. The exterior blower wiring should be terminated at the junction box located inside the blower housing. The outside blower weather cover is removed to access the junction box. The MSB blower is wired according to the specific voltage and terminations on the motor and blower model. Locate the specific wiring voltage for your blower motor on the instruction manual cover and in the charts in Chapter 4: Performance Data and
Safety Precautions. All MSB blowers operate at a maximum of 10 full load amps for 115V and 6 full load amps for 230V.
Controls Wiring Requirements
Please refer to the wiring diagrams in Appendix C. The MSB comes with the
Intelli-Sense MSB Control Box, controls cable, power inlet cable, 3-position switch, and 3-position switch label. The power for the MSB Control Box comes from the light circuit on the fume hood. The controls wiring for the MSB requires a fume hood modification to add the control cable from the 3-position switch/label.
A shielded cable from the roof blower junction box must be run to the MSB
Control Box per the wiring diagram and wiring detail in Appendix C. If needed, refer to the diagram in Appendix E that further depicts the controls needed for a typical laboratory with individual Intelli-Sense blowers.
Space Requirements
The dimensions for the different models are shown in Appendix B: Blower
Dimensions.
6 Product Service 1-800-522-7658
C
HAPTER
3
G
ETTING
S
TARTED
Now you are ready to unpack, inspect, and install the blower. Read this chapter to learn how to:
• Unpack and move your blower.
• Install the blower on a supporting structure.
• Adjust the blower outlet orientation.
• Install the exhaust run with vibration dampers.
• Connect to the blower inlet.
• Connect to the blower outlet.
• Connect the electrical supply source.
• Modify the fume hood controls and install the MSB Control Box.
• Adjust the fan speed and confirm blower performance.
• Connect the PVC Blower drain. (PVC Blowers only)
Unpacking Your Blower
Carefully remove the shipping carton from your blower. Inspect the blower for damage that may have occurred in transit. If the blower is damaged, notify the delivery carrier immediately and retain the entire shipment intact for inspection by the carrier.
)
DO NOT RETURN GOODS WITHOUT THE PRIOR
AUTHORIZATION OF LABCONCO. UNAUTHORIZED
RETURNS WILL NOT BE ACCEPTED.
The United
States Interstate
Commerce
Commission rules require that claims be filed with the delivery carrier within fifteen
(15) days of delivery.
Product Service 1-800-522-7658 7
8
Original instructions
Chapter 3: Getting Started
)
IF YOUR BLOWER WAS DAMAGED IN TRANSIT, YOU
MUST FILE A CLAIM DIRECTLY WITH THE FREIGHT
CARRIER. LABCONCO CORPORATION AND ITS
DEALERS ARE NOT RESPONSIBLE FOR SHIPPING
DAMAGE.
Do not discard the packing material for your blower until you have installed and tested the blower.
Install the Blower on a Supporting
Structure
Now that you have located your blower as instructed in Chapter 2: Prerequisites, you are ready to mount the blower on a roof curb support. Additionally, you may isolate the blower by mounting the blower on either vibration isolators or vibration mounting pads, which are available from a local industrial supply company. The supporting structure is custom for each installation. Labconco recommends supporting the blower with 5/16" diameter mounting hardware. See Appendix B:
Blower Dimensions for appropriate mounting hole locations for your particular blower.
Adjust the Blower Outlet Orientation
The housing on your blower can be rotated to facilitate your duct run configuration. By rotating the blower housing, you will be able to change the blower outlet location on your blower. The Coated Steel Blowers can be rotated to one of eight different positions and the Fiberglass or PVC Blowers can be rotated to one of three different positions. See the drawings in Appendix A to locate the parts to disassemble. To rotate the blower housing, proceed as follows:
1. Remove the upper weathercover panel, which has louvers.
2. Once removed, you have complete access to the 8 fasteners that hold your blower housing assembly in position. Removing these fasteners allows you to rotate the housing to one of the other desired outlet positions. Then insert the 8 fasteners and secure the blower housing in that specific position.
3. Reinstall the upper weathercover panel and you are ready for operation.
!
HINT: Most blowers are installed for upward discharge.
Product Service 1-800-522-7658
Original instructions
Chapter 3: Getting Started
Install the Exhaust Run with Vibration
Dampers
Rubber isolation sleeves may be used in your exhaust duct run prior to entering the blower inlet. These isolation sleeves dampen vibration that is being generated by the blower and decrease noise level at the fume hood. See the flexible duct connections listed in Chapter 7.
CS
Connect to the Blower Inlet for Coated
Steel Blowers
10" Coated Steel Blowers feature a 10-7/8" ID by 1-1/2" wide circular inlet ring.
This inlet ring is suitable for use with 10-inch diameter PVC ductwork. The PVC ductwork itself will fit inside the inlet ring and should be fastened by sheet metal screws through the inlet ring. A silicone sealant should also be used to seal between the ductwork and the blower inlet ring to prevent any air or moisture leakage.
12" Coated Steel Blowers feature a 12-1/4" OD by 1-1/2" wide inlet ring, which is suitable for use with 12-inch diameter PVC ductwork. The 12-inch diameter PVC ductwork will fit over the inlet ring on the blower and should be fastened in position by sheet metal screws into the metal ring.
FRP
Connect to the Blower Inlet for Fiberglass
Blowers
10" Fiberglass Blowers feature a 10-3/8" OD inlet ring. This inlet ring is suitable for use with 10-inch diameter PVC ductwork. The PVC ductwork will fit outside the inlet ring and should be fastened by sheet metal screws through the fiberglass inlet ring. A silicone sealant is to be used to seal between the ductwork and the blower inlet ring to prevent air or moisture leakage.
12" Fiberglass Blowers feature 12-3/8" OD inlet ring, which is suitable for use with 12-inch diameter PVC ductwork. The 12-inch diameter PVC ductwork will fit over the inlet ring on the blower and fastened by sheet metal screws through the fiberglass inlet ring.
Product Service 1-800-522-7658 9
PVC
PVC
Original instructions
Chapter 3: Getting Started
Connect to the Blower Inlet for PVC
Blowers
10" PVC Blowers feature a 10-3/8" OD inlet ring. This inlet ring is suitable for use with 10-inch diameter PVC ductwork. The PVC ductwork will fit outside the inlet ring through the fiberglass inlet ring. A silicone sealant should also be used to seal between the ductwork and the blower inlet ring to prevent air or moisture leakage.
12" PVC Blowers feature a 12-3/8" OD inlet ring, which is suitable for use with
12-inch diameter PVC ductwork. The 12-inch diameter PVC ductwork will fit over the inlet ring on the blower and should be fastened in position by sheet metal screws into the fiberglass ring.
Connect the PVC Blower Drain
The PVC Blower has a 1/2" NPT drain connection in the bottom of the housing.
When this connection is used, it should be directed into a proper drain or into the exhaust duct for proper disposal.
!
CAUTION: Draining the blower housing directly onto the roof may cause damage to your roof due to the corrosive chemicals exhausted.
ATTENTION: Vidange du carter de la soufflante directement sur le toit peut causer des dommages à votre toit en raison des produits chimiques corrosifs épuisé.
Connect to the Blower Outlet
Before proceeding with the blower outlet connection, read the two warnings listed below:
!
WARNING: Should your exhaust stack, on the outlet side of the exhaust blower, extend over 7 feet, both guy wires and additional structural supports are required to carry the weight of this ductwork. The guy wires must be substantial enough to support the exhaust stack against high wind velocities.
ATTENTION: Si votre cheminée d'échappement, sur le côté de sortie du ventilateur d'évacuation, s'étendent sur 7 pieds, les deux haubans et les supports structuraux supplémentaires sont nécessaires pour supporter le poids de cette canalisation. Les
10 Product Service 1-800-522-7658
Original instructions
CS
Chapter 3: Getting Started haubans doivent être suffisamment importants pour soutenir la cheminée d'échappement contre la vitesse des vents élevés.
HINT: An exhaust damper either before or after the MSB
Blower is very useful for fine tuning the airflow.
For Coated Steel Blowers
10" Coated Steel Blowers include a 10" by 5-1/2" rectangular outlet. A rectangular to round Transition Adaptor, Labconco part number 4722401, adapts the outlet connection to accept 10-inch diameter PVC ductwork Transition
Adaptor. Labconco part number 4722400 adapts the outlet on these blowers to accept 8-inch diameter PVC ductwork.
12" Coated Steel Blowers include a 13-1/2" by 7" rectangular outlet. Transition
Adaptor, Labconco part number 7003400, adapts the outlet of these blowers to accept 12-inch diameter PVC ductwork.
FRP
PVC
For Fiberglass Blowers
10" Fiberglass Blowers feature a 10-3/4" ID outlet connection. Ten-inch diameter ductwork will slip into this connection and should be held by sheet metal screws through the housing. Silicone sealant should be used to seal any air leaks between the duct and blower outlet connection.
12" Fiberglass Blowers feature a 12-3/4" ID outlet connection. Twelve-inch diameter ductwork will slip into this connection and should be held by sheet metal screws through the housing. Silicone sealant should be used to seal any air leaks between the duct and blower outlet connection.
For PVC Blowers
10" PVC Blowers feature a 10-3/4" ID outlet connection. Ten-inch diameter ductwork will slip into this connection and should be held by sheet metal screws through the housing. Silicone sealant should be used to seal any air leaks between the duct and blower outlet connection.
12" PVC Blowers feature a 12-3/4" ID outlet connection. Twelve-inch diameter ductwork will slip into this connection and should be held by sheet metal screws through the housing. Silicone sealant should be used to seal any air leaks between the duct and blower outlet connection.
Product Service 1-800-522-7658 11
Original instructions
Chapter 3: Getting Started
Connect to the Electrical Supply Source
The main electrical power supply connection for the MSB is made directly at the junction box underneath the outer weather cover and below the ECM motor. Refer to wiring detail and wiring diagrams in Appendix C. Remove the weather cover from the MSB and wire directly to the power wires inside the outside junction box located beneath the ECM motor. A knockout has also been provided on the side of the j-box for this purpose.
Access the j-box by removing the top weathercover of the blower base. This weathercover is held in position by machine screws, and once they have been removed, you will have access to the j-box of your blower.
CAUTION: Inspect the MSB blower nameplate voltage and model number carefully prior to connecting your electrical
! service. The wiring harness determines the voltage. The 115V harness is 7115300 and the 230V harness is 7115301 per wiring diagrams in Appendix C.
ATTENTION: Vérifiez la tension nominale ventilateur MSB et le numéro de modèle attentivement avant de brancher votre service électrique. Le câblage détermine la tension. Le harnais de 115V est 7115300 et le faisceau de 230V est 7115301 par les schémas de câblage à l'annexe C.
The blower base end panel is positioned above the blower base to allow for a continuous flow of cooling air through the motor cavity. Slots and louvers have also been placed on both the front and back of the blower base to provide airflow through this motor cavity. Because there is a gap between the end panel and the base, you will be able to run your electrical conduit directly underneath this panel.
No special cutouts or modifications are necessary.
)
NOTE: WHEN THE WIRING FOR YOUR BLOWER HAS
BEEN COMPLETED, CHECK FOR PROPER MOTOR
ROTATION. THE MOTOR SHOULD OPERATE IN A
CLOCKWISE ROTATION AS VIEWED FROM THE SHAFT
SIDE TO OBTAIN PROPER ROTATION OF THE IMPELLER
WHEEL. WHEN IN OPERATION, THE IMPELLER SHOULD
ROTATE COUNTERCLOCKWISE WHEN VIEWED FROM
THE BLOWER INLET.
12 Product Service 1-800-522-7658
Original instructions
Chapter 3: Getting Started
!
CAUTION: This blower contains an electrical motor, which requires proper electrical connection per the countries National
Electrical Code to prevent hazards. The National Electrical
Code and local codes may require that a circuit disconnect, overload protection, and short circuit protection be included in the installation. The unit should be connected by a licensed electrician.
ATTENTION: Ce ventilateur contient un moteur électrique, ce qui nécessite une bonne connexion électrique par les pays
National Electrical Code de prévention des risques. Le National
Electrical Code et les codes locaux peuvent exiger que la déconnexion du circuit, protection de surcharge, et la protection de court-circuit sont inclus dans l'installation. L'appareil doit
être raccordé par un électricien agréé.
Modify the Fume Hood Controls and
Install the MSB Blower Control Box
Labconco Intelli-Sense Blowers require control circuitry to vary the speed and set up the fume hood for two or three speed operation. If desired, the MSB can still be set up for simple ON/OFF operation. The three switch options are listed below and on the wiring diagrams in Appendix C.
Mode A
Switch Options
Mode B Mode C
MAX MAX MAX
MED MIN OFF
MIN OFF
Follow these steps to properly install the control circuitry per the wiring diagrams in Appendix C.
1. Ensure the power is turned off at the roof disconnect and at the fume hood lighting circuit.
2. With shielded wire, run the control wires from the roof mounted junction box beneath the ECM motor to the top of the fume hood.
3. With the two self-tapping screws provided, mount the MSB Control Box to the back of the light support per diagrams, Figure 3-1 through 3-8.
4. Crimp the shielded control wire from the roof mounted blower to the MSB
Control box per the wiring detail and wiring diagrams in Appendix C. As needed, reference Appendix E for installation.
Product Service 1-800-522-7658 13
Original instructions
Chapter 3: Getting Started
5. Run the MSB Control Box Power Cable with the IEC inlet from the control box to the junction box on the fume hood and connect to the fume hood light circuit per wiring detail and wirings diagram in Appendix C.
6. If two or three position blower setting is desired, remove the blower switch and blower switch label from the fume hood. For simple ON/OFF, skip this step.
7. Remove the wires to the blower switch. Important: Prevent the existing
wires to the blower switch from causing interference with the sash or
remove them.
8. Install the new blower switch label, blower switch, and blower switch retainer. If needed, blower switch cutout is 0.83" x 1.45" (2.1 cm x 3.7 cm).
9. Locate the MSB Control Cable and connect the wires from the blower switch to the MSB Control Box connector per wiring detail and wiring diagrams in Appendix C and reference Appendix E.
10. All electrical connections to the roof mounted blower, the MSB Control
Box, and switch wiring should be intact.
11. Restore power to the roof mounted blower and the fume hood lighting circuit.
Top
14
Front
Figure 3-1
MSB Control Box
Product Service 1-800-522-7658
Original instructions
Chapter 3: Getting Started
MSB
Control Box
Fume Hood
Junction Box
Figure 3-2
Top View of Typical Isometric View of MSB Control Box
Product Service 1-800-522-7658 15
Original instructions
Chapter 3: Getting Started
MSB
Control
Box
Figure 3-3
3' Installation Location
MSB
Control
Box
16
Figure 3-4
4' Installation Location
Product Service 1-800-522-7658
Original instructions
MSB
Control
Box
Chapter 3: Getting Started
Figure 3-5
5' Installation Location
MSB
Control
Box
Product Service 1-800-522-7658
Figure 3-6
6' Installation Location
17
Chapter 3: Getting Started
MSB
Control
Box
Original instructions
Figure 3-7
7' Installation Location
MSB
Control
Box
Figure 3-8
8' Installation Location
18 Product Service 1-800-522-7658
Original instructions
Chapter 3: Getting Started
Adjust the Fan Speeds and Confirm the
Performance of your Multi-Speed Blower/
Hood System
The MSB fan speeds need to be adjusted in the field to allow the hood to operate at the proper face velocity. In order to adjust the blower to the proper speeds, refer to the diagram in Figure 3-9 and follow the procedure below.
1. Ensure the blower is turned on at the roof and the control circuit is installed and powered up.
2. Refer to Figure 3-9 and place the blower switch on the hood to its maximum position.
3. Locate the Intelli-Sense MSB Control Box on top of the fume hood. With a small jewelers screw driver, adjust the MSB “MAX” setting and measure the fume hood airflow with a thermal anemometer at the proper sash height setting. Readjust the MSB “MAX” setting as required to set the face velocity. Counter-clockwise rotation increases the blower speed and corresponding fume hood airflow (CFM) increase.
4. It will be necessary to check the blower wheel shaft speed by removing the weathercover to confirm operation per the blower airflow charts in Chapter 4.
!
HINT: Many times a manual damper is used in the exhaust for fine tuning and operating in the proper portion of the blower airflow chart. For lower airflow volume (CFM) at any RPM, close the damper. Refer to Appendix E that shows a manual damper. Refer to Appendix D that shows a typical blower operating curve with the damper adjusted to different positions.
5. Repeat steps 3 and 4 for MSB “MED/MIN” speed. Decide if operation is for medium or minimum based on fume hood sash position. If “OFF” is desired, then turn “MED/MIN” to lowest setting.
6. Repeat steps 3 and 4 for MSB “MIN/OFF” speed. Decide if operation is for
“minimum” or “OFF” based on fume hood sash position and fume hood airflow requirements.
7. Cycle hood at each blower switch setting and confirm airflow requirements. If necessary, confirm blower shaft RPM on the roof and confirm with blower performance charts in Chapter 4.
8. With fume hood air monitor installed, calibrate airflow monitor and confirm airflow at each speed setting.
9. Replace weathercover over the ECM motor.
Product Service 1-800-522-7658 19
Original instructions
Chapter 3: Getting Started
Figure 3-9
Adjust the Fan Speeds
20 Product Service 1-800-522-7658
C
HAPTER
4
P
ERFORMANCE
D
ATA AND
S
AFETY
P
RECAUTIONS
Specifications and Performance Data
The specifications and performance data for your particular model are listed in detail in Chapter 4 and sub-grouped by Coated Steel, Fiberglass, and PVC
Blowers. Refer to Appendix D to view the effect damper position has on performance. All ECM blower motors feature thermal override protection and
RPM speeds are automatically reduced when motor is overloaded. Listed below is a quick selection guide.
Intelli-Sash MSB Blower Quick Selection Guide
Hood Size 3' 4' 5' 6' 7' 8' 10' 12' 16'
10" Blower
Airflow Range
(CFM)
175-495 175-735 175-955 175-1100 175-1100 175-1100 N/A N/A N/A
Note: 10" Blowers have a minimum achievable airflow of 175CFM and a maximum airflow of 1100CFM
12" Blower
Airflow Range
(CFM)
200-495 200-735 200-955 200-1180 200-1410 200-1410 200-1410 200-1410 200-1410
Note: 12" Blowers have a minimum achievable airflow of 200CFM and a maximum airflow of 1410CFM
18" Reduced
Sash Height
28" Full Open 10" or 12" 10" or 12" 10" or 12"
Sash Height
Sash Height and Airflow
Restrictions if any
10"
Blower
Blower
None
10"
Blower
Blower
None
10"
Blower
Blower
None
10" or 12" 10" or 12" 10" or 12"
Blower
12"
Blower
None
Blower
12"
Blower
None
Blower
12"
Blower
12"
Blower
12"
Blower
N/A
12"
Blower
N/A 12"
Blower
80fpm 60fpm 80fpm 60fpm max. at max. at max. at max. at full open full open 60% open 60% open
28" sash 28" sash 18" sash 18" sash
Note: For best results, always use a manual damper in the exhaust for fine tuning blower along with variable speed RPM motor.
Corrosion Resistance Guide
Coated Steel Blowers Mild to moderate corrosive chemicals
Moderate to highly corrosive chemicals
Highly corrosive acids
Fiberglass FRP Blowers
PVC Blowers
Product Service 1-800-522-7658
21
Original instructions
Chapter 4: Performance Data and Safety Precautions
2.5
2.25
Models 70611-10, -11, -20, -21
120- 240 Volts 10.0 Amps 1 HP
Blower Performance Curve with 10" Coated Steel inlet
2
1.75
1750
1.5
1.25
1500
Optimum
Blower
Performance
1
1250
SEE
NOTE
0.75
0.5
1000
0.25
750
Note: Ambient Temperatures
Inside Motor Housing Greater than 125 °F May Reduce
Maximum Blower Performance 500
0
150 250 350 450 550 650 750 850 950 1050 1150 1250 1350 1450
Airflow Rate (CFM)
22
Speed
(RPM)
1800 MAX
1500
1250
1000
750
500
300
Optimum Airflow (CFM) and Static Pressure (inches of water)
750 CFM @ 2.2" to 1050 CFM @ 1.9"
650 CFM @ 1.6" to 1250 CFM @ 0.7"
550 CFM @ 1.1" to 1050 CFM @ 0.5"
400 CFM @ 0.7" to 850 CFM @ 0.33"
275 CFM @ 0.37" to 650 CFM @ 0.20"
200 CFM @ 0.20" to 450 CFM @ 0.10"
175 CFM @ 0.06" to 350 CFM @ 0.08"
Practical range is 175 CFM, 0.06" to 1050 CFM, 1.9" or 1200 CFM, 1.0"
Product Service 1-800-522-7658
Original instructions
Chapter 4: Performance Data and Safety Precautions
2.25
2
Models 70611-12, -13, -22, -23
120- 240 Volts 10.0 Amps 1 HP
Blower Performance Curve with 12" Coated Steel inlet
1.75
1250 RPM
See
Note
1.5
1.25
1
1000
RPM
Optimum
Blower
Performance
0.75
750
0.5
0.25
500
Note: Ambient Temperatures
Inside Motor Housing Greater than 125 °F May Reduce
Maximum Blower Performance
0
150 250 350 450 550 650 750 850 950 1050 1150 1250 1350 1450 1550
Airflow Rate (CFM)
Speed
(RPM)
1125 MAX
1000
750
500
300
Optimum Airflow (CFM) and Static Pressure (inches of water)
1200 CFM @ 1.6" to 1350 CFM @ 1.5"
Practical range is 200 CFM, 0.08" to 1350 CFM, 1.5" or 1450 CFM @ 1.2"
Product Service 1-800-522-7658
1100 CFM @ 1.3" to 1450 CFM @ 1.2"
800 CFM @ 0.8" to 1150 CFM @ 0.7"
550 CFM @ 0.33" to 725 CFM @ 0.3"
200 CFM @ 0.08" to 450 CFM @ 0.1"
23
Chapter 4: Performance Data and Safety Precautions
2
Models 71818-10, -11, -20, -21
120- 240 Volts 10.0 Amps 1 HP
Blower Performance Curve with 10" FRP inlet
1.75
Original instructions
1.5
1.25
Optimum
Blower
Performance
1750
1800
1
0.75
1500
SEE
NOTE
1250
0.5
24
Speed
(RPM)
1800 MAX
1500
1250
1000
750
500
300
1000
0.25
750
Note: Ambient Temperatures
Inside Motor Housing Greater than 125 °F May Reduce
Maximum Blower Performance
500
0
150 250 350 450 550 650 750 850 950 1050 1150 1250 1350
Airflow Rate (CFM)
Optimum Airflow (CFM) and Static Pressure (inches of water)
750 CFM @ 1.7" to 1025 CFM @ 1.55"
600 CFM @ 1.2" to 1050 CFM @ 0.5"
500 CFM @ 0.8" to 850 CFM @ 0.37"
400 CFM @ 0.6" to 700 CFM @ 0.25"
300 CFM @ 0.3" to 475 CFM @ 0.12"
200 CFM @ 0.12" to 325 CFM @ 0.06"
175 CFM @ 0.06" to 275 CFM @ 0.04"
Practical range is 175 CFM, 0.06" to 1025 CFM, 1.5"
Product Service 1-800-522-7658
Original instructions
Chapter 4: Performance Data and Safety Precautions
2.25
Models 71818-12, -13, -22, -23
120- 240 Volts 10.0 Amps 1 HP
Blower Performance Curve with 12" FRP inlet
2
1.75
1.5
1.25
1
0.75
1300
RPM
1250 RPM
Optimum Blower
Performance
1000
1500
See Note
1400
0.5
750
0.25
500
Note: Ambient Temperatures
Inside Motor Housing Greater than 125 °F May Reduce
Maximum Blower Performance
0
150 250 350 450 550 650 750 850 950 1050 1150 1250 1350 1450 1550
Airflow Rate (CFM)
Speed
(RPM)
1400 MAX
1300
1250
1125
1000
750
500
300
Optimum Airflow (CFM) and Static Pressure (inches of water)
960 CFM @ 1.7" to 1100 CFM @ 1.7"
900 CFM @ 1.5" to 1150 CFM @ 1.4"
850 CFM @ 1.3" to 1400 CFM @ 0.87"
780 CFM @ 1.1" to 1310 CFM @ 0.75"
700 CFM @ 0.87" to 1180 CFM @ 0.62"
450 CFM @ 0.4" to 860 CFM @ 0.34"
300 CFM @ 0.20" to 575 CFM @ 0.12"
200 CFM @ 0.08" to 450 CFM @ 0.10"
Practical range is 200 CFM, 0.08" to 1400 CFM, 0.87" or 1150 CFM, 1.4"
Product Service 1-800-522-7658 25
Original instructions
Chapter 4: Performance Data and Safety Precautions
2
Models 71834-10, -11, -20, -21
120- 240 Volts 10.0 Amps 1 HP
Blower Performance Curve with 10" PVC inlet
1.75
1.5
1.25
Optimum
Blower
Performance
1750 RPM
1800 RPM
1
1500 RPM
0.75
1250 RPM
SEE
NOTE
0.5
1000 RPM
0.25
750 RPM
Note: Ambient Temperatures
Inside Motor Housing Greater than
125 °F May Reduce Maximum
Blower Performance
0
150
500 RPM
250 350 450 550 650 750 850 950 1050 1150 1250 1350
Airflow Rate (CFM)
Speed
(RPM)
1800 MAX
1500
1250
1000
750
500
300
26
Optimum Airflow (CFM) and Static Pressure (inches of water)
650 CFM @ 1.7" to 1100 CFM @ 1.3"
600 CFM @ 1.2" to 1050 CFM @ 0.5"
500 CFM @ 0.8" to 900 CFM @ 0.33"
375 CFM @ 0.6" to 700 CFM @ 0.2"
300 CFM @ 0.3" to 525 CFM @ 0.12"
200 CFM @ 0.12" to 350 CFM @ 0.06"
175 CFM @ 0.06" to 275 CFM @ 0.04"
Practical range is 175 CFM, 0.06" to 1100 CFM, 1.3"
Product Service 1-800-522-7658
Original instructions
Chapter 4: Performance Data and Safety Precautions
2.25
Models 71834-12, -13, -22, -23
120- 240 Volts 10.0 Amps 1 HP
Blower Performance Curve with 12" PVC inlet
2
1.75
1500
1.5
See Note
1.25
1
0.75
1250 RPM
Optimum Blower
Performance
1000
RPM
0.5
750
0.25
500
Note: Ambient Temperatures
Inside Motor Housing Greater than 125 °F May Reduce
Maximum Blower Performance
0
150 250 350 450 550 650 750 850 950 1050 1150 1250 1350 1450 1550
Airflow Rate (CFM)
Speed
(RPM)
1500 MAX
1400
1250
1000
750
500
300
Optimum Airflow (CFM) and Static Pressure (inches of water)
900 CFM @ 1.9" to 1025 CFM @ 1.8"
900 CFM @ 1.8" to 1180 CFM @ 1.7"
850 CFM @ 1.37" to 1325 CFM @ 0.8"
675 CFM @ 0.9" to 1100 CFM @ 0.6"
450 CFM @ 0.45" to 825 CFM @ 0.33"
300 CFM @ 0.20" to 550 CFM @ 0.12"
200 CFM @ 0.08" to 400 CFM @ 0.10"
Practical range is 200 CFM, 0.08" to 1180 CFM, 1.7"
Product Service 1-800-522-7658 27
Original instructions
Chapter 4: Performance Data and Safety Precautions
)
NOTE: FOR A COMPLETE PERSPECTIVE OF YOUR MSB,
REFER TO APPENDIX A REPLACEMENT PARTS.
Safety Precautions
!
!
Before attempting any service and/or maintenance on your blower, always disconnect the blower motor from its power source to prevent possible injury. Upon initial start-up, always wear protective eyewear. A qualified technician should certify the blower/hood system before it is initially used. The blower/hood system should be re-certified annually or whenever it is relocated.
Avant de procéder à un entretien et / ou de maintenance sur votre ventilateur, débranchez toujours le moteur du ventilateur de sa source d'alimentation pour éviter tout risque de blessure.
Après le démarrage initial, toujours porter des lunettes de protection. Un technicien qualifié doit certifier le système souffleur / capot avant est initialement utilisé. Le système ventilateur / hotte doit être re-certifié chaque année ou chaque fois qu'il est déplacé.
Do NOT contact blower wheel while still in motion.
NE PAS être en contact avec la roué du ventilateur tant qu’il est en marche.
28 Product Service 1-800-522-7658
C
HAPTER
5
U
SING
Y
OUR
B
LOWER
Normal Operation
Once your blower has been fully ducted and electrically wired, it is ready for operation. The blower is activated from a switch on the fume hood. For night set back to save energy, fully lower the fume hood sash and select the lowest switch setting. Always monitor your airflow constantly with a fume hood airflow monitor. For best results, use Labconco’s Guardian Digital Airflow Monitor and order from Chapter 7.
Laboratory work can resume when the blower is operational or placed in one of the two higher speeds, so that any prevailing fumes and/or odors can be exhausted from the fume hood effectively. Work must cease prior to turning the blower off or when placed at the minimum setting for night set back.
Product Service 1-800-522-7658
29
C
HAPTER
6
M
AINTAINING
Y
OUR
B
LOWER
Now we will review the suggested maintenance schedule and the common service operations necessary to maintain your blower for peak performance.
!
Only trained and experienced certification technicians should perform some of the service operations after the blower has been properly decontaminated. The wrench icon precedes the
service operations that require qualified technicians.
Routine Maintenance
!
CAUTION: Before attempting any service and/or maintenance on your blower, always disconnect the unit from its power supply source.
ATTENTION: Avant de procéder à un entretien et / ou de maintenance sur votre ventilateur, débranchez toujours l'appareil de sa source d'alimentation.
Motor
Under normal usage, the electronically commutated (ECM) motor will not require maintenance and is projected to last 50,000 hours at maximum speed or greater at lower speeds. Refer to Appendix A.
Product Service 1-800-522-7658
30
Original instructions
Chapter 6: Maintaining Your Blower
Bearings
The pillow block bearings on your blower are factory sealed and lubricated. Under normal operation, no further lubrication is required. Excessive lubricating may cause damage to the bearing seal and significantly shorten the life span of the bearing. Refer to Appendix A.
Coupler
The shaft coupler from the ECM motor to the blower wheel shaft can be inspected once a year as needed. Tightening of the coupler screws would indicate possible misalignment or an unbalanced blower wheel. Refer to Appendix A.
Common Service Operations (See drawings in Appendix A Replacement
Parts)
Pillow Block Bearing Replacement
If you need to replace a pillow block bearing on your blower, make sure that the locking collars on the bearings face one another. The setscrews used on the locking collars have been sealed with a removable thread sealant. Thread sealant should be applied when a pillow block bearing is replaced in the field. Refer to
Appendix A.
Motor Replacement
With the power disconnected, remove the upper weathercover, the mounting hardware that supports the ECM motor frame and remove the ECM motor.
Remove both of the wire lead connectors to the motor. Re-install the new motor in reverse order. Refer to Appendix A.
Intelli-Sense MSB Control Box Replacement
With the power disconnected, remove the existing Intelli-Sense MSB Control Box and replace it with the new one. Make all connections per electrical wiring diagram in Appendix C. Then follow the instructions in Chapter 3 to adjust the fan speeds and confirm the performance of your Intelli-Sense Multi-Speed Blower/
Hood System. Refer to Appendix A.
Product Service 1-800-522-7658 31
C
HAPTER
7
M
ODIFYING
Y
OUR
B
LOWER,
C
ALCULATING
S
TATIC
P
RESSURE
L
OSS, AND
B
LOWER
S
IZING
Airflow Monitors for Fume Hoods
Guardian™ Airflow Monitors continuously monitor face velocity through the fume hood opening and should always be used with Intelli-Sense Multi-Speed
Blowers.
Guardian™ Airflow Monitors
Sense and alert the operator to low airflow conditions. From the monitor’s face plate, the user can easily select and calibrate a set point between 30 and 250 fpm using a velocity meter and a screwdriver. Audible/visual alarm. Includes night setback, external alarm and alarm mute functions. Flush-mount design on Protector
Premier, XStream and XL Fume Hoods.
Catalog #
9413300
Ranges
100-115 volts, 50/60 Hz
9413301* 208-230 volts, 50/60 Hz
*International electrical configuration
For use with Hood
Premier, XStream, XL
Premier, XStream, XL
Shipping Wt. lbs. / kg.
6 / 2.7
6 / 2.7
Product Service 1-800-522-7658
32
Original instructions
Chapter 7: Modifying Your Blower
Guardian™ Digital Airflow Monitors
Guardian Digital Airflow Monitor senses and alerts the operator to low airflow conditions. LCD displays actual airflow in fpm or m/sed. Audible/visual alarm alerts the user to sustained low velocity condition. Calibration instructions displayed on LCD.
Each monitor also includes a temperature-compensated sensor, external alarm, night setback and alarm mute functions. Flushmount design on Protector Premier, XStream and XL Fume Hoods.
Contact Labconco for optional temperature sensor and optional
RS-485 port for Modbus** RTU communication.
Catalog #
9413400
Ranges
100-115 volts, 50/60 Hz
For use with Hood
Premier, XStream, XL
Shipping Wt. lbs. / kg.
6 / 2.7
9413401* 208-230 volts, 50/60 Hz
*International electrical configuration
Premier, XStream, XL
**Modbus is a registered trademark of Schneider Automation
6 / 2.7
7117700 Cable, Airflow Monitor – Night Setback
This cable is pre-configured to attach to the three position switch at minimum setting and send an input signal to the airflow monitor for night setback to disable the air monitor and prevent it from alarming.
Two Main Blower Modifications
There are two main ways to modify the performance of your blower as listed in
Chapter 3: Getting Started. One way is the adjustment of the blower outlet orientation. The other way is the adjustment of the fan speed to fine-tune the performance of your blower/hood system. Refer to Chapter 3 for these instructions. Additionally, all the performance data for your particular blower model number are listed in Chapter 4: Performance Data and Safety Precautions.
Additional Modifications by Adding Ductwork Accessories
There are additional ways to equip your blower by adding any of the accessories on the following pages. To ensure that your blower exhaust system will operate properly, the ductwork and accessories linking your hood and blower must be sized correctly. Along with the accessories listed next are the “equivalent resistance in feet of straight duct” for each accessory. It is necessary to compute the sum of the equivalent resistance factors for each accessory and ductwork length for your entire hood/blower system. Then the blower can be sized properly from the total equivalent resistance for your exhaust system. Blower Sizing
Example: You have selected a Labconco Protector Premier 48 Laboratory Hood at
100 fpm and 725 CFM. The static pressure of the Protector Premier 48 at 100 fpm is 0.22". The exhaust collar of this hood is sized to receive 1" diameter PVC duct directly. In this example, your fume removal system requires 30 feet of straight duct, two 90° elbows and one zero pressure weathercap to adequately exhaust the
Product Service 1-800-522-7658 33
Original instructions
Chapter 7: Modifying Your Blower
chemical fumes. You will be handling low to moderately corrosive materials, so you have selected a Coated Steel Blower.
The information following for the 90° elbow tells you that each 90°, 12" diameter elbow has the equivalent resistance of 25 feet of straight duct. The zero pressure weathercap has 5 feet of equivalent resistance. The total equivalent feet for the system is:
Feet of straight duct
2 elbows @ 25 feet each
30 Feet
50 Feet
1 zero pressure weathercap @ 5 feet 5 Feet
85 Feet
The chart following converts “equivalent resistance in feet of straight duct” to static pressure. So the static pressure of 85 equivalent feet of 12" nominal duct is equal to (85/10) x 0.011" = 0.094". Now this is added to the hood static pressure of 0.22" for a total system static pressure of 0.314". Look at the performance data in Chapter 4 for Coated Steel Blowers. Refer to the 0.25" to .50" range at 725
CFM since this is closest to 0.314". The chart will lead you to select a Coated
Steel Blower model that is 10" set at 900 RPM or a 12" set at 600 RPM. In this example, models 7061110, 7061112, 7061120 and 7061122 all would work. A manual damper would also provide additional fine tuning of the blower/hood exhaust system.
Sizes and Pressure Losses in Thermoplastic Duct
This chart provides static pressure losses for 10' long duct lengths of various diameters over a range of airflows in CFM for use in sizing hood/blower combinations at 100 feet per minute.
Nominal Diameter/Inches
Catalog Number
Shipping Wt./lbs.
Airflow/CFM
6 8 10 12 16
6.625 8.625 10.750 12.750 16.000
6.25 8.250 10.375 12.375 15.625
47086 47189 70272 56020 56050
20 35 50 65 80
Static Pressure Loss/Inches H
2
0
For Each 10 ft. of Duct Length
1000 .557 .005
1250 .855 .008
1500 -- .012
1750 -- .016
2000 -- .020
2500 -- .031
3000 -- .042
4000 -- .074
34 Product Service 1-800-522-7658
Original instructions
Chapter 7: Modifying Your Blower
Typical Fume Removal System
This diagram details the many components that are needed to complete a typical fume removal system. All of these components as well as others are available from Labconco.
Zero Pressure
Weathercap
Blower Transition Adapter
(for Coated Steel Blowers only)
3-5 duct diameter lengths
(approximately 3-5 feet) between elbow and blower is required
Thermoplastic Duct
Intelli-Sense Coated
Steel, Fiberglass or
PVC Blower
Speed
Control Box
Flexible Duct Connection
Thermoplastic Duct*
90° Elbow
Thermoplastic Duct*
Manual Duct Damper
(Highly recommended)
Protector
®
Laboratory Hood
SpillStopper™ Work Surface
Protector ® Standard,
Acid or Solvent
Storage Cabinets
*Minimum of 5 to 15 feet of duct in both locations is recommended for noise reduction.
Product Service 1-800-522-7658 35
Original instructions
Chapter 7: Modifying Your Blower
Thermoplastic Duct
PVC exhaust duct is Type 1, unplasticized, schedule 40, lightweight and corrosionresistant. A female duct coupling is required to join two sections. Connections are simple with solvent cement. This rigid duct may be cut without special tools.
Comes in 10' lengths.
Nominal Diameter/Inches 6 8 10 12 16
6.625 8.625 10.750 12.750 16.000
6.250 8.250 10.375 12.375 15.625
Duct Couplings, Female
PVC coupling makes connection between two sections of thermoplastic duct quick and easy.
Nominal Diameter/Inches 6 8 10 12 16
Equivalent Resistance in Feet of Straight Duct
Duct Couplings, Male
PVC duct in 6" length facilitates connections between Coated Steel Blowers and elbows, thermoplastic duct reducers and weather caps.
Nominal Diameter/Inches
Catalog Number
Actual OD/Inches
Actual ID/Inches
6 8
21447 47199
6.625 8.625
6.250 8.250
10
70278
12
70673
10.750 12.750
10.375 12.375
Shipping Wt./lbs.
Elbows
3 4 5 6
PVC elbows both 45° and 90°, are compatible with thermoplastic duct. Designed and engineered for quick installation and minimum pressure losses, they feature belled end connections to receive PVC duct directly.
Nominal
Diameter/Inches
6 8 10 12 16
47087 47190 70273 56021 56051
Shipping 8 10 12 14 17
Equivalent Resistance in feet of Straight Duct
12 15 20 25 36
Nominal
Diameter/Inches
6 8 10 12 16
47088 47191 70274 56022 56052
Approx. Height/Inches 8-3/4 10-3/4 12-1/2 15 17-1/2
Shipping 8 10 12 14 17
Equivalent Resistance in feet of Straight Duct
6 7.5 10 12.5 18
36 Product Service 1-800-522-7658
Original instructions
Chapter 7: Modifying Your Blower
Thermoplastic Duct Reducers
PVC coupling type reducers are designed for connecting thermoplastic duct of different diameters. Compare your blower inlet size with your duct size to see if one is necessary.
Nominal Size/Inches
Catalog Number
Shipping Wt./lbs.
Equivalent Resistance in feet of Straight Duct
6x8 8x10
56059 56060
2 5
10x12
56061
6
12x16
56307
8
0 0 0 0
Zero Pressure Weathercaps
The zero pressure weathercap is made of strong, corrosion-resistant PVC. The cap adds little static pressure to the exhaust system and allows for vertical discharge of the effluent air for dispersion away from the building.
Nominal
Diameter/Inches
Catalog Number
6 8 10 12 16
47222 47223 70951 56221 56222
Shipping Wt./lbs.
Equivalent Resistance in feet of Straight Duct
Spiral Tube
20 25 30 35 40
5 5 5 5 5
This spiral tube simplifies temporary installations. It is corrosion-resistant, neoprene-impregnated fiberglass reinforced with steel wire. Includes rigid duct connector and two clamps. Length is ten feet.
Nominal Diameter/Inches
Catalog Number
Shipping Wt./lbs.
Equivalent Resistance in feet of Straight Duct
7" for use with
6" fittings
19651
10
9" for use with
8" fittings
47194
15
11" for use with
10" fittings
70277
20
13" for use with
12" fittings
56223
25
Because this ductwork is flexible and may conform to various configurations, it is not possible to know the precise equivalent resistance.
Manual Duct Dampers
This damper fitting allows you to balance airflow. It may be used with exhaust and auxiliary air ducts, and is usually placed directly above the fume hood.
Nominal Diameter/Inches 6 8 10 12 16
Catalog Number 47242 47413 59834 59812 47264
Shipping Wt./lbs.
Approx. Height/Inches
10 12 15 20 25
19-1/2 24
Product Service 1-800-522-7658 37
Original instructions
Chapter 7: Modifying Your Blower
Flexible Duct Connections
This flexible connection reduces vibration between the blower and PVC ductwork.
It is supplied with two clamps for easy installation.
Nominal Diameter/Inches 9" for use with 8" fittings 11" for use with 10" fittings 13" for use with 12" fittings
Catalog Number 47265 70342 56214
Shipping Wt./lbs. 5 5 5
Blower Transition Adaptors
This epoxy-coated steel transition adaptor fits all Labconco Coated Steel Blowers.
This adaptor allows you to connect round thermoplastic duct to the exhaust side of the blower to create an exhaust stack. Nominal size PVC duct fits inside the adaptor opening.
Nominal Diameter/Inches
Catalog Number
8
47224
10
4722401
12
70034
Shipping Wt./lbs. 3 4 4
For use with Labconco Blowers 70680-
70687
70680-
70687
70688-
70697
Auxiliary-Air Transition Adaptor
The auxiliary-air transition adaptor is the same construction as the blower transition adaptor, but is designed to allow you to connect round thermoplastic duct to the regular auxiliary-air collar of Protector Fume Hoods.
Nominal Diameter/Inches 10
Catalog Number 48893
Shipping Wt./lbs. 4
T and Y Connections
PVC fittings shaped in T and Y configurations are compatible with thermoplastic duct. End connections receive PVC pipe directly. Contact Labconco for help in sizing blowers with these accessories.
Nominal Diameter/Inches
10x10x12
T’s
Catalog Number
Shipping Wt./lbs.
56304
20
Approx. Height/Inches 19
Y’s
Nominal Diameter/Inches
Catalog Number
Shipping Wt./lbs.
10x10x12
56301
19
12x12x16
56305
20
Approx. Height/Inches 12-3/4 23-1/4
38 Product Service 1-800-522-7658
Original instructions
Chapter 7: Modifying Your Blower
Accessory for Basic 47 Hoods
Exhaust Transition Adaptor
The exhaust transition adapts to 7" and 10" rectangular outlet on Basic 47 Hoods, model series 22473 and 22475, to receive 10" diameter PVC duct.
Nominal Diameter/Inches 10
Catalog Number 22648
Shipping Wt./lbs. 5
Accessory for Perchloric Acid Applications
Wash Rings
Wash rings are suited for use in Perchloric acid duct systems. Each features a wide-angle conical spray nozzle and wash water connector nipple fabricated into a
PVC coupling for use with 10" or 12" nominal duct.
Nominal Diameter/Inches 10 12
Catalog Number 47460 47461
Shipping Wt./lbs. 5 6
Accessories for Pathogens, Organic Vapors and Odor Control Applications
HEPA Filter Packs
High Efficiency Particulate Air Filter for non-radioactive particulate and pathogenic applications. Rated for 1000 CFM airflow with initial 1.0" static pressure drop. Replaceable HEPA filter media removes 99.97% of all particles 0.3 micron or greater. Furnished with clamping frame and duct connections. Unit measures 28" x 28" x 25" high.
Charcoal Filter Packs
Activated Charcoal Filter for non-radioactive organic vapors and odor control are rated for 1000 CFM airflow with 0.2" static pressure drop. Unit measures 28" x
28" x 25" high.
Nominal Diameter/Inches 8 10 12
HEPA Filter
Charcoal Filter
22400 22401 22442
100 100 100
22430 22431 22441
100 100 100
Product Service 1-800-522-7658 39
Original instructions
Chapter 7: Modifying Your Blower
Backdraft Dampers
Designed for use in buildings under negative pressure to keep outside air from entering the laboratory through the hood ventilation system. Damper is weighted to stay in down/resting position when the hood is not in use, and rises from the airflow exhausting when the blower is on. It mounts vertically on blower outlet.
The damper is made of PVC Type 1, unplasticized, schedule 40 duct.
Nominal Diameter/Inches 8 10 12
S304508 S304510 S304512
15 18 20
Bird Screens
Screen attaches easily with screws to auxiliary-air blower inlet to keep birds from nesting in blower.
Nominal Diameter/Inches
Catalog Number
Shipping Wt./lbs.
10 12
S122500 S122501
5 5
40 Product Service 1-800-522-7658
C
HAPTER
8
T
ROUBLESHOOTING
Refer to the following table if your blower fails to operate properly. If the suggested corrective actions do not solve your problem, contact Labconco for additional assistance.
PROBLEM CAUSE
Wires not connected
Remote blower won’t operate.
at junction boxes or switches.
CORRECTIVE ACTION
Check connection of switches.
Check connection to control box on top of fume hood, connection of switch control wires, and connection of shielded control wires from roof blower. If needed, replace control box on top of fume hood.
Reset circuit breakers. Circuit breakers tripped in building electrical supply.
Blower wiring is disconnected.
Blower motor is defective.
Shaft coupling is loose.
Inspect blower wiring and 2 or 3 position switch.
Check connection of ECM motor cables or replace ECM motor.
Tighten shaft coupling with lockwashers between ECM motor and blower wheel.
Product Service 1-800-522-7658 41
Original instructions
Chapter 8: Troubleshooting
PROBLEM CAUSE
Contamination outside of fume hood.
Remote blower has excessive vibration.
Fume hood has
improper face velocity.
Fume hood has improper face velocity
Improper motor mount.
Inspect wheel for damage and appropriate clearance.
Check for objects in blower wheel.
Improper inlet connection.
Blower not sized properly.
RPM adjustment or a manual damper is needed in the exhaust ductwork.
CORRECTIVE ACTION
Have fume hood re-certified and check remote blower exhaust system. Hood should have average face velocity of 60-100 fpm depending on application.
Review Chapter 2: Prerequisites and Chapter 3: Getting Started.
The blower should be mounted on vibration isolators or vibration mounting pads to isolate vibration.
Replace damaged wheel or adjust clearance.
Remove objects in blower wheel.
Review Chapter 3: Getting
Started. The blower inlet should be installed with a vibration damper or flexible duct connection.
Review Chapter 7: Modifying
Your Blower. Blower Sizing Size the blower properly with equivalent resistance method.
Refer to Chapters 4 and 7.
Review Chapter 3: Getting Started
Adjust the fan speed and confirm blower performance. See Chapters
3, 4, and 7.
42 Product Service 1-800-522-7658
1B
1C
2A
2B
2C
2D
2E
2F
2G
2H
2I
2J
2K
2M
2N
3A
3B
3C
3D
3E
3F
4
4
1
1
1
1
1
1
1
1
1
3
1
3
3
1
1
1
1
1
1
A
PPENDIX
A
B
LOWER
R
EPLACEMENT
P
ARTS
W
ITH
W
IRING
S
HOWN
The following replacement parts are organized for Intelli-Sense MSB’s for low pressure coated steel, low pressure fiberglass, and low pressure PVC. See next page for illustrations.
Item Quantity Part No. Description – Roof Blower Components
1889012 Coupler Socket Head Screw 10-32 x .75"
1910110 Coupler, Lock Washer #10 stainless
Fan, Cooling
7116201 ECM Motor, 1 HP - 10" Coated Steel, 1900 RPM
7116202 ECM Motor, 1 HP - 12" Coated Steel, 1250 RPM
7116203 ECM Motor, 1 HP - 10" Fiberglass FRP, 1900 RPM
7116204 ECM Motor, 1 HP - 12" Fiberglass FRP, 1500 RPM
7116205 ECM Motor, 1 HP - 10" PVC, 1900 RPM
7116206 ECM Motor, 1 HP - 12" PVC, 1500 RPM
7116500 Motor Mounting Bracket
7115300, 01 Motor Harness to Blower J-Box (115V-00, 230V-01)
3843500 Motor Clamp Bracket
7117000 Vibration Foot Support, Motor
1880924 Screw, Motor Clamp
Nut, Clamp
1881020 Screw, Vibration Mount
1906921 Nut, Vibration Mount
7117800 Impeller Wheel Assembly, 10" Coated Steel
7117801 Impeller Wheel Assembly, 12" Coated Steel
7117200 Impeller Wheel Assembly, 10" Fiberglass FRP
7117201 Impeller Wheel Assembly, 12" Fiberglass FRP
7117202 Impeller Wheel Assembly, 10" PVC
7117203 Impeller Wheel Assembly, 12" PVC
5B
5C
1
6
7116800
1885806
Product Service 1-800-522-7658
Internal Rain Guard, Motor Cover
Weather Cover, Screw
43
7B
7C
8
9A
9B
10
11A
11B
12
13
14A
14B
15
16
Original instructions
Appendix A: Blower Replacement Parts
6A
6B
1
2
1860500
1881712
Flanged Pillow Block Bearing 5/8" Bore
Screw – Cap Hex Head 3/8 – 16 x 0.75" SS
1
1
1
1
1
1
2
1
1
1
1
1
1
1
9816800
4450100
7115100
7115000
7115001
7116000
1302401 Blower Switch, 3 Position
1305200 Retainer, Blower Switch
9486200
3207100
7115200
7115201
7117500
7117700
Cover, J-Box Blower
Screw, Cover J-Box
Cable, Blower Switch to MSB Control Box
MSB Control Box, 115V
MSB Control Box, 230V
Cable, Control Box Power Inlet from hood lighting
Label, 3 Position Blower Switch for Hood Corner Post
ECM Blower Motor control Board
Internal Harness Power, Control Box 115V
Internal Harness Power, Control Box 230V
Power Supply, 24V – Control Box
Accessory Cable, Airflow Monitor Night Set Back
(not included)
44 Product Service 1-800-522-7658
Original instructions
6
5
Not Shown
9
13, 14, 15
Not Shown
4
1
1D
2
10
Appendix A: Blower Replacement Parts
3
8
12
Not Shown
Figure A-1
11
16
Not Shown
2H
7
Product Service 1-800-522-7658 45
A
PPENDIX
B
B
LOWER
D
IMENSIONS
46 Product Service 1-800-522-7658
Original instructions
Appendix B: Blower Dimensions
Product Service 1-800-522-7658 47
Appendix B: Blower Dimensions
Original instructions
48 Product Service 1-800-522-7658
A
PPENDIX
C
B
LOWER
E
NVIRONMENTAL
C
ONDITIONS,
W
IRING
D
ETAIL &
W
IRING
D
IAGRAMS
Environmental Conditions
• Maximum altitude: 9843 feet (3000 meters).
• Ambient temperature range: -30° to 130°F (-34° to 54°C).
• Main supply voltage fluctuations not to exceed ±10% of the nominal voltage.
• Transient over-voltages according to Installation Categories II (Overvoltage Categories per IEC 1010). Temporary voltage spikes on the AC input line that may be as high as 1500V for 115V models and 2500V for
230V models are allowed.
• Used in an environment of Pollution degrees 2 (i.e., where normally only non-conductive atmospheres are present). Occasionally, however, a temporary conductivity caused by condensation must be expected, in accordance with IEC 664.
Product Service 1-800-522-7658 49
Original instructions
Appendix C: Blower Environmental Conditions & Wiring Diagrams
Wiring Detail
50 Product Service 1-800-522-7658
Original instructions
Appendix C: Blower Environmental Conditions & Wiring Diagrams
115V Wiring Diagram
230V Wiring Diagram
Product Service 1-800-522-7658 51
A
PPENDIX
D
T
YPICAL
I
NTELLI
-S
ENSE
B
LOWER
O
PERATING
C
URVE
Damper 50% Open (Red)
• High Data Point, H1 895 CFM, 1.8" S.P., 1250 RPM
• Medium Data Point, M1 675 CFM, 1.0" S.P., 895 RPM
• Low Data Point, L1 260 CFM, 0.12" S.P., 330 RPM
Damper 75% Open (Green)
• High Data Point, H2 955 CFM, 1.0" S.P., 875 RPM
• Medium Data Point, M2 565 CFM, 0.37" S.P., 515 RPM
• Low Data Point, L2 330 CFM, 0.12" S.P., 315 RPM
Product Service 1-800-522-7658
52
Original instructions
Appendix D: Typical Intelli-Sense Blower Operating Curve
Product Service 1-800-522-7658 53
A
PPENDIX
E
T
YPICAL
L
ABORATORY
C
ONTROLS
W
ITH
I
NDIVIDUAL
I
NTELLI-
S
ENSE
B
LOWERS
Product Service 1-800-522-7658 54
Advertisement
Key features
- UL listed ECM motor
- Three position switch replacement
- Fume hood blower control box
- Quiet operation with forward curved impeller
- Cooling vents and weather cover for exterior roof mounting
- Direct drive ECM motor
- Corrosion resistant housing