advertisement
Intended Use
Product Overview
Background
Product Components
Names of Parts
Caution
Contents
Power Supply
Main Menu
Code Registration
Patient Test (Basic Measurement)
Quick Test (Quick Measurement)
Memory
QC Test (Quality Control)
Device Settings
10
11
7
9
3
3
4
5
12
14
19
25
31
42
SelexOn .1
ID Input
Printer
Barcode Scanner
Battery
PC Sync Software
Troubleshooting
Appendix
Contents
58
60
63
49
53
55
57
Intended Use
SelexOn is an in-vitro diagnostic device that measures the concentration of biomarkers within blood. It is used for diagnosis.
This device uses an anticoagulant-treated (EDTA-treated) venous whole blood sample and should be operated by experts in the laboratory or hospital.
<IMPORTANT>
Please read this manual carefully prior to the initial use of the product.
Product Overview
SelexOn meter is easy to use. After injecting the collected venous blood sample into the sensor on the test strip, the strip changes its color depending on the concentration of the analyte. This reaction is then quantitatively measured to calculate the results.
The meter should be used only with the provided adapter. If a different adapter is used, incorrect results may be obtained.
Product Principle
The test contains two monoclonal antibodies specific to analyte. The antibodies form a sandwich complex with the analyte in the blood.
Following removal of erythrocytes from the sample, plasma passes through the detection zone in which the gold –labelled analyte sandwich complexes and the positive signal is displayed as a reddish line (the test line). Excess gold-labelled antibodies accumulate along the control line, signaling that the test was valid. The intensity of the test line increases in proportion to the analyte concentration. The optical system of the
SelexOn instrument detects the two lines and measures the intensity of the test line. The integrated software converts the signal intensity to a quantitative result and showed it in the display.
SelexOn .3
SelexOn .2
Background
Lateral Flow Immunoassay
The lateral flow immunoassay is also known as a sandwich assay.
In this quantitative measurement method, by using an antigen, a gold nanoparticle-labelled detection antibody is combined with an antibody specific to disease markers.
The antigen concentration within blood appears as the thickness of the red band on the test line. Through this correlation, the antigen concentration can be measured by using the standard curve.
Product Components
Components
Meter Components
SelexOn Meter 1 Meter
1 User's Manual
1 External power adapter
1 Fan filter
1 Standard strip
SelexOn .4
SelexOn .5
Product Components
Optional
Optional Components
SelexOn test strip
SelexOn printer
Barcode scanner
SelexOn battery
Data cable
Quantity
1 Box
1 ea.
1 ea.
1 ea.
1 ea.
You can purchase the optional components from your nearest dealer.
For details, please contact the store or visit our web site
(www.infopia21.com).
Names of Parts
SelexOn Meter
SelexOn Test Strip
Strip Handle
Measuring Window
Blood Inlet area
Strip Code
SelexOn .6
SelexOn .7
Names of Parts
Port Panel
Name
1. Printer port
2. Barcode scanner port
3. USB port
4. Power adapter port
5. Power switch
<CAUTION!>
The USB port is exclusively used by the manufacturer. If this port is used for other than the intended use, the product may have a problem.
SelexOn .8
Caution
Precautions
1. Please read this manual carefully prior to use.
2. Prior to the initial use, please set the current year, month and day in
'SetUp' mode.
If the date is incorrect, an error related to exceeding the expiration date of a strip may occur.
3. In order to prevent any infection or contamination, please wear gloves.
4. The meter must be used on a flat, stable surface and should not be moved during its operation.
5. If the RFID information on the sensor container and the code information of the sensor do not match, an error will occur. Please replace the strip with the same code information on both sensor container and sensor.
6. Please routinely run QC (Quality Control) mode.
7. Wipe off any contaminants such as blood spots or dust on the blood inlet prior to use.
8. Since a strong impact may cause the device to fail, avoid external shock.
9. Use the meter at a stable temperature range (18 - 32
℃).
Storage After Use and Maintenance
1. Do not arbitrarily disassemble or alter the device.
2. Store the device in a dry place.
3. Avoid dust and high humidity when storing the device.
4. Store the device in a stable, level location without vibration or shock.
5. Do not store near chemical storage or a place that generates gas.
6. Avoid direct sunlight.
7. Clean the meter with a soft cloth or tissue only if there are any contaminants such as blood spots or grease on the device after use.
SelexOn .9
Power Supply
Meter Power Supply
1. Check that the "Power Switch" is Off.
2. Plug the power plug into the wall outlet.
3. Connect the "External
Power Adapter" to the
"Power Adapter Port" on the back of the
SelexOn meter.
4. Flip the
"Power Switch" 'ON'.
5. Briefly press the
"Power Button" on top of the meter.
6. After the system initialization, the Main
Menu appears.
SelexOn .10
Main Menu
After the system initialization, the Main Menu appears on the meter LCD.
Icons for each mode appear on the LCD.
Patient Test
Runs the basic measurement mode.
Quick Test
Runs the quick measurement mode.
Memory
Runs the memory mode.
QC Test
Runs the QC Test mode.
SetUp
Runs the SetUp mode.
SelexOn .11
Code Registration
RFID Code Registration
The RFID must be registered prior to use when performing a test with a new strip.
The RFID code needs to be registered once prior to use of a new strip box.
Step1
In 'SetUp' mode, press the 'RFID' icon in order to run the RFID communications.
Code Registration
Step 3
Hold the label until the RFID information appears.
When the RFID code registration succeeds, the screen below appears.
Confirm if the "Marker", "Code" and "Exp. Date" information on the RFID label of the strip box match with the RFID information displayed on the
LCD.
If they do not match, rerun the code registration. If the problem persists, contact the nearest dealer.
Step2
Place the RFID label of the strip box onto the "RFID Reader" on top of the meter.
SelexOn .12
<NOTE>
Please refer to 'Troubleshooting' if an error message appears.
<CAUTION!>
Ensure that the RFID code is registered when using a new strip box.
SelexOn .13
Patient Test (Basic Measurement)
Step 1
Press 'Patient Test' on the Main Menu.
Step 2
A screen to insert a strip appears.
SelexOn .14
Patient Test (Basic Measurement)
Step 3
Take one strip out of the strip storage box.
The test strip must be used immediately after opening the pouch.
Strip Handle
Measuring Window
Blood Inlet
Strip Code
<CAUTION!>
Avoid any foreign substances contaminating the strip code label, blood inlet and measuring window.
Contaminants may cause an error in the results.
Step 4
Hold the strip handle and insert the strip into the strip insertion port on the left side of the meter.
SelexOn .15
Patient Test (Basic Measurement)
Step 5
The strip marker and code are displayed on the LCD.
Confirm if the marker and code information on the RFID label of the strip box match with those on the screen.
When the strip is recognized, the meter automatically runs 'System check'.
Step 6
When the 'System check' is complete, the meter automatically runs
'Heating'.
Patient Test (Basic Measurement)
Step 7
When the LCD instructs to inject blood, inject 100 ul of whole blood into the blood inlet by using a pipette. Press 'OK'. (
①-
>
②->③)
1
3
2
Step 8
The measurement runs automatically and the measurement time is displayed on the LCD.
SelexOn .16
SelexOn .17
Patient Test (Basic Measurement)
Step 9
When the measurement is complete, the results appear on the LCD.
Remove the strip.
The results are automatically saved. (Please refer to 'Device Settings' for changing the settings.)
Press 'Delete' to delete the results.
Press 'Print' to print the results on the 'SelexOn Printer'.
Press or to return to the Main Menu.
Single Strip
Triple Strip
<NOTE>
If the results for CK-MB are displayed with an inequality sign such as '>
40 ng/mL' or '< 3 ng/mL', the results are out of the measurement range.
Please refer to the strip instruction for the measurement range and
Cutoff concentration.
SelexOn .18
Quick Test (Quick Measurement)
Step 1
Take one test strip out of the strip storage box.
The test strip must be used immediately after opening the pouch.
Strip Handle
Measuring Window
Blood Inlet
Strip Code
<CAUTION!>
Avoid any foreign substances contaminating the strip code label, blood inlet and measuring window.
Contaminants may cause an error in the results.
Step 2
Inject 100 ul of whole blood into the blood inlet on the strip by using a pipette.
SelexOn .19
Quick Test (Quick Measurement)
Step 3
Press'Quick Test' to display the timer selection screen.
SelexOn .20
Quick Test (Quick Measurement)
Icon
10 Min
5 Min
SKIP
Description
Select if the strip reaction time is 10 minutes
Select if the strip reaction time is 5 minutes
Select to proceed to the next step without setting the timer
<NOTE>
Select an appropriate timer after confirming the reaction time on the strip manual.
If the timer is selected, the remaining time is displayed in the 'Remaining
Time' field. When the time is up, the next step is automatically executed.
If 'SKIP' is pressed, the timer stops and the next step is executed.
<CAUTION!>
You must proceed to the next step after the reaction time specified in the strip manual is passed by using the timer.
SelexOn .21
Quick Test (Quick Measurement)
Step 4
A screen to insert a strip appears.
Quick Test (Quick Measurement)
Step 6
The strip marker and code are displayed on the LCD.
Confirm if the marker and code information on the RFID label of the strip box match with those on the screen.
The measurement will automatically start.
Step 5
Hold the strip handle and insert the strip into the strip insertion port on the left side of the meter.
Push the strip to the end.
SelexOn .22
Step 7
When the measurement is complete, the results appear on the LCD.
Remove the strip.
The results are automatically saved. (Please refer to 'Device Settings' for changing the settings.)
Press 'Delete' to delete the results.
Press 'Print' to print the results on the 'SelexOn Printer'.
Press 'Continue' to return to Step 4 and to measure another strip.
Press or to return to the Main Menu.
SelexOn .23
Quick Test (Quick Measurement)
Single Strip
Triple Strip
<NOTE>
If the results for CK-MB are displayed with an inequality sign such as '>
40 ng/mL' or '< 3 ng/mL‘, the results are out of the measurement range.
Please refer to the strip instruction for the measurement range and
Cutoff concentration.
SelexOn .24
Memory
Press
' ' Memory’ on the Main Menu and press ' ' Patient’ in order to view the result list.
The user can view and manage the results.
Up to 2000 measurements are saved.
The latest results appear at the top of the list.
By using the ▲/▼ icons, the previous or next result can be viewed.
SelexOn .25
Memory
Press each icon in order to sort the data accordingly.
Displays the current page / all pages.
Displays the test type.
[P]: Patient Test / [Q]: Quick Test
Memory
Press 'Memory' on the Main Menu and press 'QC' in order to view the
QC result list.
Press 'Control Solution' in order to view the Control Solution Test results.
Selects the results.
Moves the page.
Icon Function
All Select
Del
Selects all results on the current page.
Prints the selected results on the SelexOn Printer.
Deletes the selected results.
SelexOn .26
Up to 1000 results can be saved.
The latest results appear at the top of the list.
By using the
▲/▼ icons, the previous or next result can be viewed.
SelexOn .27
Memory
Press each icon in order to sort the data accordingly.
Displays the current page / all pages.
Memory
Press 'Memory' on the Main Menu and press 'QC' in order to view the
QC result list.
Press 'Standard Strip' in order to view the Standard Strip Test results.
Selects the results.
Moves the page.
Icon Function
All Select
Selects all results on the current page.
Del
Prints the selected results on the SelexOn Printer.
Deletes the selected results.
SelexOn .28
Up to 1000 results can be saved.
The latest results appear at the top of the list.
By using the ▲/▼ icons, the previous or next result can be viewed.
SelexOn .29
Memory
Press each icon in order to sort the data accordingly.
Displays the current page / all pages.
1
[ 1]
Selects the results.
Moves the page.
Icon Function
All Select
Selects all results on the current page.
Del
Prints the selected results on the SelexOn Printer.
Deletes the selected results.
SelexOn .30
QC Test
Meter Quality Control Management
Please inspect if the SelexOn meter operates properly by using the standard strip.
Standard Strip
-SelexOn Standard Strip Level 1: Standard1 (LOW)
-SelexOn Standard Strip Level 2: Standard2 (HIGH)
Storage
-Store the standard strip in its container.
-Avoid direct sunlight.
-Store the standard strip between 10 - 35
℃ (50 - 95℉).
-Store the standard strip between 10% - 85% humidity.
-Since a standard strip damaged from external shock may affect the results, safely store the standard strip.
<CAUTION!>
Do not use an expired standard strip.
Do not use a standard strip damaged due to external shock, scratch, or contamination.
SelexOn .31
QC Test
When to test the standard strip
-Prior to the sample measurement
-When the results are deemed to be inaccurate
-When the QC alarm appears (Weekly, Monthly)
<NOTE>
Store the standard strip in its container when it is not in use.
(It may be deformed due to direct sunlight or other sources of light.)
QC Test
How to use the standard strip
Step 1
Press 'QC Test' on the Main Menu.
Step 2
Press 'Standard Strip Test'.
SelexOn .32
SelexOn .33
QC Test
Step 3
Select the level of the standard strip.
- LEVEL 1: Standard1 (LOW)
- LEVEL 2: Standard2 (HIGH)
QC Test
Step 5
The results are displayed.
- PASS: indicates that the meter operates properly.
- FAIL: indicates that the meter may be used but the results may be inaccurate.
Please rerun the QC test if FAIL appears. If FAIL persists, stop using the meter and contact your local representative for technical support.
Step 4
Insert the standard strip of the relevant level.
Push the strip to the end.
If an RFID is required, register the RFID tag on the standard strip storage container.
Step 6
After completing the test, remove the standard strip from the meter.
Store the standard strip in its storage container while avoiding direct sunlight.
SelexOn .34
SelexOn .35
QC Test
Strip Quality Control Management
SelexOn Control Solution verifies the accuracy of the SelexOn test strip.
Control Solution
The control solution is a solution that verifies the accuracy of the test strip.
For details regarding the control solution, please refer to the manual provided with the control solution.
When to run the control solution test
- Prior to the sample measurement
- When the results are deemed to be inaccurate
- When the QC alarm appears (Weekly, Monthly)
<CAUTION!>
Do not reuse the strip that was used for the control solution test.
Do not use an expired control solution.
<NOTE>
For details regarding the control solution, please refer to the control solution manual.
QC Test
How to use the control solution
Step 1
Press 'QC Test' on the Main Menu.
Step 2
Press 'Control Solution Test'
SelexOn .36
SelexOn .37
QC Test
Step 3
Select a level for the control solution.
- LEVEL 1: Low concentration (LOW)
- LEVEL 2: High concentration (HIGH)
QC Test
Step 5
The strip marker and code are displayed on the LCD.
Confirm if the marker and code information on the RFID label of the strip box match with those on the screen.
When the strip is recognized, the meter automatically runs 'System
check'.
Step 4
Insert an unused strip that has the same marker as the control solution to be tested. Push the strip to the end. If an RFID is required, register the RFID tag on the strip box.
Step 6
When the 'System check' is complete, the meter automatically runs
'Heating
' ' .
SelexOn .38
SelexOn .39
QC Test
Step 7
When a screen for injecting the control solution appears on the LCD, inject 100 ul of the solution into the blood inlet. Press 'OK'
QC Test
Step 9
When the test is complete, the results are displayed on the LCD.
Remove the used strip from the meter.
- PASS: indicates that the strip is usable.
- FAIL: indicates that the strip is usable but may affect the results.
Please rerun the QC test if FAIL appears. If FAIL persists, stop using the strip and contact the nearest dealer for service.
Step 8
The test runs automatically and the measurement time is displayed on the LCD.
SelexOn .40
SelexOn .41
Device Settings
Entering SetUp Mode
Press 'SetUp' in Main Menu mode.
Selecting SetUp Menu
BASIC
ID Setup
BEEP
DEVICE
RFID
QC Period
Date, Time, LCD Light Period, LCD Brightness
Operator ID or Patient ID Input options
Beep on/off, Touch sound effect on/off
Printer on/off, Barcode reader on/off
Runs RFID communication mode
QC test alarm settings
SelexOn .42
Device Settings
Date & Time
Select year, month, day, hour, and minute and press the ▲/▼ icons in order to set the values.
Press '24 H' to change the time display.
Press or in order to not save the changes.
Press 'OK' to save the changes.
<CAUTION!>
Please operate the meter after accurately setting the current year, month and day.
If the date is incorrect, an error related to exceeding the expiration date of a strip may occur.
LCD
Power Saving
LCD Contrast
Sets the screen saver time (15 minutes, 30 minutes, 1 hour). Touch the LCD in order to turn the screen back on when the screen is off.
Sets the LCD brightness (1 - 5).
Press or in order to not save the changes.
Press 'OK' to save the changes.
SelexOn .43
Device Settings
ID Setup
Device Settings
BEEP
Press the ▲/▼ icons in order to set the ID input option.
NONE: no ID input
Optional: selects the ID input option
Required: ID input required
Press or in order to not save the changes.
Press 'OK' to save the changes.
<NOTE>
The measurement order varies depending on either the Operator ID or
Patient ID on the ID Setup.
Please refer to "ID Input" for details.
BEEP Available
BEEP on Touch
BEEP LEVEL
Set to OFF in order to disable the beep.
Both BEEP on Touch and BEEP LEVEL are deactivated.
Set to OFF in order to disable the touch sound.
Press the
▲/▼ icons in order to set the volume.
Press or in order to not save the changes.
Press ‘OK’ to save the changes.
SelexOn .44
SelexOn .45
Device Settings
DEVICE
Device Settings
RFID
Thermal Printer
Auto Print
Barcode reader
Auto Save
Sets the printer.
Set to OFF in order to disable the Auto Print settings.
Sets the Auto Print feature.
If the SelexOn Printer is connected, the results are automatically printed after the test.
Sets the barcode reader.
Sets the Auto Save feature.
The results are automatically saved after the test.
Press or in order to not save the changes.
Press 'OK' to save the changes.
SelexOn .46
Runs the RFID communication mode.
Ensure that the RFID code is registered when using a new strip box.
The RFID code needs to be registered only once.
<CAUTION!>
If the RFID is not recognized within 30 seconds, an error message appears. Please refer to 'Troubleshooting' for details.
SelexOn .47
Device Settings
QC Period
Sets the QC Test alarm.
The period is set either weekly or monthly.
Weekly: runs the test alarm 7 days after the setup date.
Monthly: runs the test alarm 30 days after the setup date.
Press or in order to not save the changes.
Press 'OK’ to save the changes.
SelexOn .48
ID Input
ID Input Settings
The SelexOn meter can manage the results by ID through entering either the Operator or Patient ID.
Press 'ID Setup' in 'SetUp' mode in order to allow either the Operator
ID or Patient ID to be entered. The measurement order changes depending on this setup.
3
4
1
2
Operator ID
NONE
Optional
Required
NONE
Optional
Required
Patient ID
NONE
NONE
Optional
Required
Optional
Required
SelexOn .49
ID Input
The figure below shows the measurement order by setup.
SelexOn .50
ID Input
Operator ID: Optional or Required setup
Optional: performs a test by pressing Enter without entering an ID.
Required: must enter the ID.
An Operator ID input message appears after booting.
Press ‘ENTER’ after entering the ID in order to display the Main Menu.
The ‘OPID’ icon to change the Operator ID appears on the right corner of the Main Menu.
If the Operator ID is set to 'NONE' on the ‘ID Setup’ menu, the ‘OPID’ icon disappears.
SelexOn .51
ID Input
Patient ID: Optional or Required setup
Optional: performs a test by pressing Enter without entering an ID.
Required: must enter the ID.
Press 'Patient Test' and the Patient ID input message appears.
Press 'ENTER' after entering the ID in order to go to the next step.
Printer (Optional)
SelexOn printer
Directions for Use
1. While both the meter and the printer are off, connect the printer to the printer port (1) on the rear panel of the meter.
2. Turn on the printer and then turn on the SelexOn meter.
Press 'Device' in 'Setup' mode to turn ON the Thermal Printer.
3. After completing the measurement, press 'Print' in order to print the results.
<NOTE>
The Patient ID may be entered by using the barcode scanner.
Please refer to 'Barcode Scanner' for details.
SelexOn .52
SelexOn .53
Printer (Optional)
4. Select the results to print out in Memory mode and press 'Print' in order to print the selected results.
Barcode Scanner (Optional)
Barcode Scanner
The barcode scanner is used to enter
The operator or patient ID.
Directions for Use
1. While both the meter and the barcode scanner are off, connect the barcode scanner to the "Barcode Scanner Port (2)" on the rear panel of the meter.
2. Turn on the meter.
Press 'Device' in 'Setup' mode to turn ON the Barcode Reader.
3. Press 'Barcode Reader' on the 'ID Input' screen.
When the barcode reader is ready to use, the message "Please use the barcode reader" appears.
SelexOn .54
SelexOn .55
Barcode Scanner (Optional)
4. Scan a barcode with the scanner in Barcode Input Standby mode in order to display the barcode information in the ID input field.
<NOTE>
Up to 16 characters can be entered in the barcode information field.
If the barcode input standby time is exceeded, press 'Barcode Reader' again.
It is important that the labels be printed to the required specifications.
Reading errors may occur if any of the following conditions exist :
- The narrow bar width is too small.
- The barcode length is too small.
- The height is too small.
- The reader is held too far from the label.
- The background reflection is too high or low
If the reader is not able to consistently read your labels, apply a test label of the format being used to a new specimen tube and perform the barcode test.
If the reader is able to read the test label, the quality of your labels may be inadequate.
If the test label cannot be read, the reader itself could be the problem.
Contact with our local representatives for technical assistance.
SelexOn .56
Battery (Optional)
SelexOn Battery
The SelexOn meter can be operated by using the SelexOn Battery without the power adapter.
How to use the battery
1. Open the battery cover under the meter.
2. Insert the battery to fit the charging terminal of the battery with the other terminal of the meter.
3.Align the battery cover with the battery cover groove. Press and click the cover.
How to charge the battery
1. Plug the power cable of the meter into the wall outlet and turn the power switch 'ON'.
2. Briefly press the "Power Button" on top of the meter.
3. The "Charging LED" on the top of the meter is red while charging the battery.
4. When the battery is charged, the "Charging LED" turns off.
<CAUTION!>
The battery must be removed after the meter is turned off.
If the battery is removed while the power is on, the meter may be damaged.
SelexOn .57
PC Sync Software
PC Sync Software
This program transfers the results from the meter to a PC.
You can download this program from our web site (www.infopia21.com).
PC Sync Software
4. When the PC Sync program and the meter are connected, the screen below appears on the LCD.
Directions for Use
1. While both the meter and the PC are off, connect the PC to the
"Printer Port (1)" on the rear panel of the meter by using the data cable that is sold separately.
5. When executing the data transfer command by using the PC Sync program, the results stored on the meter are saved on the PC.
6. When you exit the PC Sync program after transferring the data, the meter returns to 'Main Menu' mode.
<CAUTION!>
Do not remove the data cable while running the PC Sync program.
2. After turning on the PC, turn on the meter.
3. When the meter is in "Main Menu" mode, run the PC Sync program from the PC.
SelexOn .58
SelexOn .59
Troubleshooting
Error Cause Solution
Er1 Memory test failed
Reboot the meter.
If the problem persists, contact the nearest dealer for the service.
Er2 Test strip expired
Check if the date on the meter is properly set to the current date. If the date is incorrect, set the current date from the setup menu.
Check for the expiration date of the strip.
Er3 Unmatched code
Er4 RFID recognition time has passed
Hold the RFID tag on the strip box as close as possible to the RFID reader on the meter.
If the problem persists, contact the nearest dealer.
Er5 Barcode cannot be read
Check if the RFID label code and the strip code match.
The same RFID code must be recognized.
Check if the strip code label is contaminated. If the label has no problem, slowly insert the strip again.
If the problem persists, contact the nearest dealer.
SelexOn .60
Troubleshooting
Error Cause Solution
Er6 Test strip was moved
Do not move or remove the strip during the measurement.
If the problem persists, contact the nearest dealer.
Er7 Temperature too high or too low
Er10 Test strip may be contaminated
Leave the meter at a temperature between 18 - 32
℃ for 10 minutes prior to use.
Er8 Blood inlet cover is open
The "Blood Inlet Cover" must be kept closed during the measurement. Close the cover and rerun the measurement.
Er9 Strip insertion timeout
The strip must be inserted within
3 minutes of the strip insertion screen appearing on the LCD.
Rerun the measurement.
The strip is contaminated or used.
Use a new strip and rerun the measurement.
SelexOn .61
Troubleshooting
Error Cause Solution
Er11 Test strip can not be used
A test strip has used for the standard strip test.
Rerun the test by using the standard strip.
Er12 Standard strip can not be used
Do not use the standard strip for tests other than the standard strip test.
Rerun the test by using the test strip.
Er13 Level discordance The level selected from the standard strip test and the standard strip level do not match.
When the Level 1 is selected,
Standard1 should be inserted.
When the Level 2 is selected,
Standard2 should be inserted.
Appendix
1. Maintenance
2. Check upon Failure
3. Specifications and Characteristics
4. Symbol & Description
SelexOn .62
SelexOn .63
Maintenance
Never disassemble the meter.
If the meter does not operate properly, unplug the power cable and contact the authorized service center.
Cleaning
Turn OFF the meter prior to cleaning and separate the external power adapter and the meter.
Wipe off dust on the meter with a soft, clean cloth.
Do not use any liquid or gas cleaning agents.
When cleaning the blood inlet, remove the "Blood Inlet Cover" and clean only contaminated areas by using a cotton swap.
If needed, use an alcohol wipe to clean the contaminated area. Do not use other chemicals. Other chemicals may damage the meter surface.
Replacing Fan Filter
Routinely inspect the fan filter on the right side of the meter. If the fan filter is dirty due to dust, replace with a new filter that is provided with
SelexOn.
Check upon Failure
When the meter does not boot
- Check for the "Power Switch" on the rear panel.
When the screen displays nothing
- When the meter is in Power Saving mode, a screen saver runs. Touch the LCD in order to exit Power Saving mode.
When the strip is expired
- Check if the date and time on the meter is set to the current date and time.
1. Open the fan filter cover on the right side of the meter.
2. Carefully remove the used fan filter.
3. Replace it with a new fan filter.
4. Close the fan filter cover.
SelexOn .64
SelexOn .65
Specifications and Characteristics
Sample Type Anticoagulant-treated (EDTA-treated) venous whole blood
Sample Volume 100ul
Measurement Range Refer to the strip manual.
Number of Stored Results 2000 results
Power Source DC 9V / 3A
Dimensions / Weight 168 x 190 x 95(mm) / 0.8kg
Storage Temperature 10 ~ 35
℃ (50 ~ 95℉)
Operational Temperature 18 ~ 32
℃ (64 ~ 90℉)
Humidity 10% ~ 85%
Peripheral Devices Thermal Printer
Barcode Scanner
Battery (7.4V / 4400mAh)
Data Cable
SelexOn .66
Symbol & Description
Symbol Description
This symbol indicates that the product complies with the applicable directives of the
European Union.
Consult instruction for use.
Used by
Caution,consult accompanying documents
Authorised representative in the European
Community.
in vitro diagnostic medical device.
Batch code
Catalog number
Serial number
Manufacturer
SelexOn .67
Symbol & Description
Symbol Description
Temperature Limitation
Do not reuse
Date of Manufacturer
Keep away from sunlight
Direct Current
Waste Electrical and Electronic Equipment
This symbol indicates that the system is type
A equipment, which provides a particular degree of protection against electric shock.
SelexOn .68
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project