The a.max AM-I Capsule Mixer is a small, compact device that provides a quick and efficient way to mix dental capsules. The mixer has two different modes and a timer function that can be set from 1-30 seconds. It also features a safety switch that will automatically stop the operation if the protective shield is opened.
Advertisement
Advertisement
User Manual
CAPSULE MIXER
Disclaimer:
1. First of all, we would like to thank you for purchasing Monitex
AM-I Optimum Capsule Mixer. Please carefully read
the user manual and operate the apparatus accordingly. Please
keep this user manual well preserved for future reference.
2. Monitex shall not assume any responsibility if any malfunction
or damage was caused by improper operation, removal,
modification or maintenance.
3. We shall not assume any responsibility if any malfunction or
damage was caused by operating under unsuited environment.
4. We shall not assume any responsibility if any malfunction or
damage was caused by forceful majeure.
5. We reserve the right to explain for the given cause of the
Defective products which might have been caused by natural
disasters or abnormal temperature, humidity that is inconcurrent
to the guidance instructed in this manual.
Content
Product parts & accessories.......................................2
Important tips before operation...................................3
Operating steps.........................................................4
Important tips on product maintenance........................5
Important tips on product cleaning..............................5
Trouble Shooting.......................................................6
Product Warranty & Repair.........................................6
Product Specification................................................7
1
Disclaimer:
1. First of all, we would like to thank you for purchasing Monitex
AM-I Optimum Capsule Mixer. Please carefully read
the user manual and operate the apparatus accordingly. Please
keep this user manual well preserved for future reference.
2. Monitex shall not assume any responsibility if any malfunction
or damage was caused by improper operation, removal,
modification or maintenance.
3. We shall not assume any responsibility if any malfunction or
damage was caused by operating under unsuited environment.
4. We shall not assume any responsibility if any malfunction or
damage was caused by forceful majeure.
5. We reserve the right to explain for the given cause of the
Defective products which might have been caused by natural
disasters or abnormal temperature, humidity that is inconcurrent
to the guidance instructed in this manual.
Content
Product parts & accessories.......................................2
Important tips before operation...................................3
Operating steps.........................................................4
Important tips on product maintenance........................5
Important tips on product cleaning..............................5
Trouble Shooting.......................................................6
Product Warranty & Repair.........................................6
Product Specification................................................7
1
Product Parts & Accessories
1
7
5
4
2
6
8
3
9
1 Protective Shield:
The safety switch will be activated when the Protective Shield is not closed.
The machine will automatically stop its operation in case the protective
shield is opened.
2 Holding Arms:
Where to place the capsule.
3 Power Button:
The Mode indicator & Time Display indicator will start to blink (showing
display of “10” ) when the power button is pressed ON.
4 Mode Set-up Buttons:
Press Mode button to set up required Mode setting.
5 Mode Indicator:
When the required mode is selected, the corresponding indicator will be
turned on.
To set the mixing time, adjustable up to 1~30 seconds.
7 Time Display indicator:
Display the selected mixing time. The timer will blink during operation as
it counts down.
8 STORT/STOP Button:
Press the button to start operation. Press once again to stop the operation
at anytime.
9 Adaptor: 100V ~ 240V 50/60Hz
2
Important tips before Operation
1. Due to transportation requirement, all the motors inside each unit is
secured with 2 fixing screw bolts. Please observe below diagram for
removing these screws before place the capsule mixer on the cabinet.
1) Please unscrew the two screw
bolts, which are pre-installed at
the bottom of each unit, like shown
in the Diagram 1.
Please maintain the balance
of the apparatus while removing
the screw bolts.
Please also remove the two screw
bolts simultaneously as unscrewing
only 1 bolts first could result in
difficult position for the other
screw bolt to be loosen.
2) Please safe-keep these two bolts
1
to the screw holes pre-designated
in the Diagram 2 for future needs.
When the apparatus requires to
be transported from one place to
another, please screw back these
two screw bolts before removal
or delivery to prevent possible
damage.
3) Insert the “L” shape plug into the
amalgamator power socket, then
connect the power plug with the
adaptor. Please make sure the
power plug is connected securely.
(Like shown in Diagram 3)
2
3
3
Product Parts & Accessories
1
7
5
4
2
6
8
3
9
1 Protective Shield:
The safety switch will be activated when the Protective Shield is not closed.
The machine will automatically stop its operation in case the protective
shield is opened.
2 Holding Arms:
Where to place the capsule.
3 Power Button:
The Mode indicator & Time Display indicator will start to blink (showing
display of “10” ) when the power button is pressed ON.
4 Mode Set-up Buttons:
Press Mode button to set up required Mode setting.
5 Mode Indicator:
When the required mode is selected, the corresponding indicator will be
turned on.
To set the mixing time, adjustable up to 1~30 seconds.
7 Time Display indicator:
Display the selected mixing time. The timer will blink during operation as
it counts down.
8 STORT/STOP Button:
Press the button to start operation. Press once again to stop the operation
at anytime.
9 Adaptor: 100V ~ 240V 50/60Hz
2
Important tips before Operation
1. Due to transportation requirement, all the motors inside each unit is
secured with 2 fixing screw bolts. Please observe below diagram for
removing these screws before place the capsule mixer on the cabinet.
1) Please unscrew the two screw
bolts, which are pre-installed at
the bottom of each unit, like shown
in the Diagram 1.
Please maintain the balance
of the apparatus while removing
the screw bolts.
Please also remove the two screw
bolts simultaneously as unscrewing
only 1 bolts first could result in
difficult position for the other
screw bolt to be loosen.
2) Please safe-keep these two bolts
1
to the screw holes pre-designated
in the Diagram 2 for future needs.
When the apparatus requires to
be transported from one place to
another, please screw back these
two screw bolts before removal
or delivery to prevent possible
damage.
3) Insert the “L” shape plug into the
amalgamator power socket, then
connect the power plug with the
adaptor. Please make sure the
power plug is connected securely.
(Like shown in Diagram 3)
2
3
3
4
Operating STEPS
1. After pressing the Power button on,
the mode indicator as well as the time
indicator will start blinking. There
should be a sign of “10” seconds
blinking in the display.
2. Depending on each characteristic of different capsules, the time
equired for mixing might vary from others. Please refer to the
specified capsule instruction for further details.
※ The Amalgam capsule & Filling material in general require an
average time of 10 seconds to gain a consistent mixture. Material
with special specification may vary from different capsules.
※ Mixing time can be reset at any given time setting between 1 to
30 seconds. There are 2 separate time setting memories that can
allow user to set up two different time setting accordingly.
Regarding individual capsule time setting, please review the
instruction manual accompanied with the capsule products for
further details.
1) Please make sure the Protective glass is securely closed.
2) Press Set button to select the designated mode.
to add or reduce the mixing time by one second.
4) Press “SET” button to confirm the setting. The time indicator will
change from blinking display to fix display.
3. Open the Protective shield, and place
the capsule into the holding arms as
shown in diagram 4.
4
4. Securely close the protective glass and press “START/STOP”
button to begin mixing operation. The timer display on the screen
will start to countdown as time proceeds.
※ This apparatus will not operate if
the protective glass is not securely
shut. The time indicator will display
“OF” as a result.
5. After Mixing complete, please open
the protective glass and take the
capsule from the holding arms.
Please reivew diagram 5 for removing
the capsule.
Important tips on Product maintenance
1. Do not operate when the capsule is not in place. Operation without
capsule might result malfunction under human error. The possible
damage is not covered with warranty.
2. The location of this product should be avoided under direct sun light.
Please DO NOT place this equipment under dusty environment.
3. Perchance when liquid inside the capsule escaped inside the
mixing chamber, please immediately wipe clean to prevent possible
damage of color-change or metallic distortion caused by Dental-use
liquid or solvent.
Important tips on Product cleaning
Please apply with soft clean clothe soaked with warm water or disconcentrated neutral cleaning agent to wipe clean the surface of this equipment. One should avoid using Volatile solvent when cleaning.
5
4
Operating STEPS
1. After pressing the Power button on,
the mode indicator as well as the time
indicator will start blinking. There
should be a sign of “10” seconds
blinking in the display.
2. Depending on each characteristic of different capsules, the time
equired for mixing might vary from others. Please refer to the
specified capsule instruction for further details.
※ The Amalgam capsule & Filling material in general require an
average time of 10 seconds to gain a consistent mixture. Material
with special specification may vary from different capsules.
※ Mixing time can be reset at any given time setting between 1 to
30 seconds. There are 2 separate time setting memories that can
allow user to set up two different time setting accordingly.
Regarding individual capsule time setting, please review the
instruction manual accompanied with the capsule products for
further details.
1) Please make sure the Protective glass is securely closed.
2) Press Set button to select the designated mode.
to add or reduce the mixing time by one second.
4) Press “SET” button to confirm the setting. The time indicator will
change from blinking display to fix display.
3. Open the Protective shield, and place
the capsule into the holding arms as
shown in diagram 4.
4
4. Securely close the protective glass and press “START/STOP”
button to begin mixing operation. The timer display on the screen
will start to countdown as time proceeds.
※ This apparatus will not operate if
the protective glass is not securely
shut. The time indicator will display
“OF” as a result.
5. After Mixing complete, please open
the protective glass and take the
capsule from the holding arms.
Please reivew diagram 5 for removing
the capsule.
Important tips on Product maintenance
1. Do not operate when the capsule is not in place. Operation without
capsule might result malfunction under human error. The possible
damage is not covered with warranty.
2. The location of this product should be avoided under direct sun light.
Please DO NOT place this equipment under dusty environment.
3. Perchance when liquid inside the capsule escaped inside the
mixing chamber, please immediately wipe clean to prevent possible
damage of color-change or metallic distortion caused by Dental-use
liquid or solvent.
Important tips on Product cleaning
Please apply with soft clean clothe soaked with warm water or disconcentrated neutral cleaning agent to wipe clean the surface of this equipment. One should avoid using Volatile solvent when cleaning.
5
6
Trouble Shooting
When there are some problems or troubles, occuring on the capsule mixer, please check against the below listed SOP. If none of the provided possible cause matches your situation, please contact your local dealer or Monitex for solution or repair. Thank you.
Scenario Possible Cause
The plug of the power adaptor might be loosen.
Unable to operate when pressing
START / STOP button
The L shaped power plug might be loosen.
The protective shield is open.
The apparatus moves during operation.
Please check if the apparatus is placed on uneven surface.
Abnormal Noise
2 Screw bolts have not been removed.
Possible Solution
Please make sure the power adaptor plug-in to the unit.
Please make sure the L-shape plug is plug-in properly.
Make sure the protective shield is firmly shut.
Please place a securely stable plastic pad under the equipment.
Unscrew the two screw bolts at the bottom of the units.
Product Warranty & Repair
All Monitex Products were manufactured under thorough quality control. Perchance the apparatus were to be broken under normal operation in accordance to the Instruction Manual within warranty period . The broken unit is subject to repair and maintenance free of charge. Please contact your nearest local dealer or Monitex
Industrial Co., Ltd. for your inquiry & best solution.
Our warranty period extends 1 year after the date of purchasing.
Specification
Product Classification:
Protection against power strike
Voltage
Input Power
Specification on the power adaptor
CLASS I equipment
DC24V
50VA
(Input) AC110V~240V 50/60Hz
(Output) DC24V
Specification:
Timer Set up
Modes
Operating temperature
Size
Weight
1~30 Seconds Time adjustable
2 modes
10º C~40º C
Machine-L 210 x W 200 x H 150 mm
Adaptor-L 60 x W 120 x H 40 mm
3.1kgs (Power adaptor included)
** Product specification is subject to implementation of change
or further upgrades without prior notice.
7
6
Trouble Shooting
When there are some problems or troubles, occuring on the capsule mixer, please check against the below listed SOP. If none of the provided possible cause matches your situation, please contact your local dealer or Monitex for solution or repair. Thank you.
Scenario Possible Cause
The plug of the power adaptor might be loosen.
Unable to operate when pressing
START / STOP button
The L shaped power plug might be loosen.
The protective shield is open.
The apparatus moves during operation.
Please check if the apparatus is placed on uneven surface.
Abnormal Noise
2 Screw bolts have not been removed.
Possible Solution
Please make sure the power adaptor plug-in to the unit.
Please make sure the L-shape plug is plug-in properly.
Make sure the protective shield is firmly shut.
Please place a securely stable plastic pad under the equipment.
Unscrew the two screw bolts at the bottom of the units.
Product Warranty & Repair
All Monitex Products were manufactured under thorough quality control. Perchance the apparatus were to be broken under normal operation in accordance to the Instruction Manual within warranty period . The broken unit is subject to repair and maintenance free of charge. Please contact your nearest local dealer or Monitex
Industrial Co., Ltd. for your inquiry & best solution.
Our warranty period extends 1 year after the date of purchasing.
Specification
Product Classification:
Protection against power strike
Voltage
Input Power
Specification on the power adaptor
CLASS I equipment
DC24V
50VA
(Input) AC110V~240V 50/60Hz
(Output) DC24V
Specification:
Timer Set up
Modes
Operating temperature
Size
Weight
1~30 Seconds Time adjustable
2 modes
10º C~40º C
Machine-L 210 x W 200 x H 150 mm
Adaptor-L 60 x W 120 x H 40 mm
3.1kgs (Power adaptor included)
** Product specification is subject to implementation of change
or further upgrades without prior notice.
7
INDUSTRIAL CO., LTD
6F, 70 GUANG-FU ROAD SEC. 1,
SAN-CHONG CITY, TAIPEI, TAIWAN
TEL : 886-2-29993869 FAX: 886-2-29993647
E-mail : [email protected]
http://www.monitex.com.tw
Copyright© 2010 MONITEX Industrial CO., LTD. All Rights Reserved. 2010.12 (Ver. 1)