U S E R M A N U A L
ASPIRATOR
Model No. PM63
(shown)
PM63E
CAUTION
SAVE THESE INSTRUCTIONS
Federal (USA) law restricts this device to sale by or on the order of a physician.
300 Held Drive
Northampton, PA 18067 USA
Tel: (+001) 610-262-6090
Fax: (+001) 610-262-6080 www.precisionmedical.com
CONTENTS
RECEIVING / INSPECTION .........................................................2
INTENDED USE ..........................................................................2
READ ALL INSTRUCTIONS BEFORE USING ............................2
ExPLANATION OF ABBREVIATION ...........................................2
SAFETY INFORmATION - WARNINGS AND CAUTIONS ...........2
SPECIFICATIONS ......................................................................4
COmPONENT DESCRIPTION ....................................................5
GROUNDING INSTRUCTIONS ...................................................6
OPERATING INSTRUCTIONS .....................................................6
CLEANING / mAINTENANCE .....................................................7
RETURNS ....................................................................................8
DISPOSAL INSTRUCTIONS .......................................................8
REPLACEmENT PARTS ..............................................................9
LImITED WARRANTY ................................................................10
ASPIRATOR
1
RECEIVING / INSPECTION
Remove the Precision medical, Inc. Power Vac + Aspirator from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider.
INTENdEd USE
To supply a continuous or intermittent vacuum source adequate within the stated operating vacuum range, to aspirate fluids in medical suctioning procedures by trained healthcare professionals.
REAd ALL INSTRUCTIONS BEFORE USING
This manual instructs a Professional to operate the Power Vac + Aspirator. This is provided for your safety and to prevent damage to the Aspirator. If you do not understand this manual, DO NOT USE the Aspirator and contact your Provider.
ExPLANATION OF ABBREVIATION
mmHg millimeters of mercury
SAFETY INFORMATION - WARNINGS ANd CAUTIONS
dANGER
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
CAUTION
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
Used without the safety alert symbol indicates a potentially hazardous situation which, if not avoided, may result in property damage.
CONSULT ACCOmPANYING DOCUmENTS
2/2/00
Type B Equipment
Protective Earth Ground
Fuse
Alternating Current
2
ASPIRATOR
dANGER
TO REdUCE THE RISk OF ELECTROCUTION:
• Always unplug Power Vac + Aspirator immediately after using.
• dO NOT use while bathing.
• dO NOT place or store Aspirator where it can fall or be pulled into a tub or sink.
• dO NOT place or drop into water or other liquid.
• dO NOT reach for the Aspirator that has fallen into water. Unplug immediately.
• dO NOT attempt to repair Aspirator. Any attempt could result in electrical shock.
• dO NOT allow the aspirator to exceed Operating or Storage
Temperature specifications.
• Never use near any type of flame or flammable/explosive substances.
• Caution must be exercised when using this Aspirator where oxygen is being administered.
WARNING
TO REdUCE RISk OF BURNS, ELECTROCUTION, FIRE, OR
INjURY TO PERSONS:
• The Power Vac + Aspirator should never be left unattended when running.
• Close supervision is necessary when the Aspirator is used by, on, or near children or invalids.
• Use the Aspirator only for its “Intended Use” as described in this manual.
• Adjust to desired setting prior to use.
• dO NOT use attachments not recommended by Precision medical, Inc.
• dO NOT use if Aspirator has a damaged cord or plug.
• dO NOT use if Aspirator has been dropped or damaged.
• dO NOT cover air vents. Restriction of air flow will cause unit to overheat, causing the thermal reset to shut down the Aspirator.
• dO NOT drop or insert any object into any opening or hole.
• Connect this Aspirator to a properly grounded outlet only. See
“GroUndInG InstrUctIons”.
• Keep the cord away from heated surfaces.
ASPIRATOR
3
SPECIFICATIONS
Operating Vacuum Range: 0 - 200 mmHg with “Full Vacuum”
Full Vacuum is 330.2 mmHg ± 51 mmHg
Vacuum Modes:
REG - Regulated (provides an adjustable, continuous vacuum level)
OFF - No Vacuum
INT - Intermittent (provides an adjustable, intermittent vacuum level)
Intermittent Cycle Time: Factory set at approximately sixteen (16) seconds “on”, eight (8) seconds “oFF”
Gauge Accuracy:
dimensions:
Length:
Width:
Height:
Weight:
Shipping Weight:
Electrical:
0 - 100 mmHg ±4 mmHg
101 - 200 mmHg ±6 mmHg
15.9 in (40.3 cm)
7.4 in (18.7 cm)
10.1 in (25.7 cm)
21.2 lbs
24.5 lbs
(9.6 kg)
(11.1 kg)
Pm63:
Pm63E:
108-132 VAC, 60 Hz, 5 Amps
207-253 VAC, 50 Hz, 2.2 Amps
Operating Temperature Range:
50°F to 104°F (10°c to 40°c)
Transport / Storage Requirements:
temperature range:
Humidity:
Filter:
-4°F to 104°F (-20°c to 40°c) max 95% Noncondensing
Inline, Hydrophobic
Safety Systems: High Temperature motor Shutdown, Class B,
Thermal Limit Automatic Reset
Suction Canister:
Equipment Classification:
Class I maximum Volume 2,000 cc
Do not exceed vacuum level of 25 inHg (640 mmHg)
Type B
IPx0
- with respect to protection from electrical shock.
- degree of protection against electrical shock.
- degree of protection against ingress of liquids.
mode of Operation - continuous
This Power Vac + Aspirator meets the EmC requirements of EN55011.
This Power Vac + Aspirator is not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen or nitrous oxide.
Specifications are subject to change without prior notice.
4
ASPIRATOR
COMPONENT dESCRIPTION
CAUTION
missing or illegible labels must be replaced, contact
Precision medical, Inc.
On/Off
Switch
Suction
Canister
Inline
Hydrophobic
Filter
Vacuum Control
Knob
Vacuum
Gauge
Selector
Knob
Air Inlet
Filters
AC Inlet /
Fuse Holder
ASPIRATOR
Power Cord
Retainer
5
Power
Cord
Hour meter
GROUNdING INSTRUCTIONS
The Power Vac + Aspirator must be plugged into an electrical outlet that is properly grounded. In the event of an electrical short circuit, grounding reduces the risk of electrical shock. This product is equipped with a grounding circuit.
dANGER
Improper use of the grounding plug can result in a risk of electrical shock.
WARNING
Ground reliability can only be achieved when the power cord is connected to an equivalent receptacle marked “Hospital Grade”.
OPERATING INSTRUCTIONS
WARNING
• Read this User manual before operating the Power Vac + Aspirator.
• Full Vacuum is present between settings.
CAUTION
Inspect the Power Vac + Aspirator for visual damage before use, DO
NOT USE if damaged.
1. Place Aspirator near a properly grounded electrical receptacle on a table, floor or other horizontal, smooth, hard surface.
Allow a minimum of 4 in (10 cm) clearance on all sides to ensure adequate air flow for cooling of Aspirator.
2. Ensure the on/off switch and selector Knob are in the “oFF” position.
3. Ensure the Inline Hydrophobic Filter is properly connected to the vacuum port on Suction Canister lid.
• the Inline Hydrophobic Filter prevents liquid from entering the vacuum pump
.
CAUTION
dO NOT run the Aspirator without Inline Hydrophobic Filter and canister. This will void the warranty.
4. Plug the Aspirator into a properly grounded receptacle.
5. Press the on/off switch to the “on” position.
6. turn the selector Knob to the “rEG” position.
6
ASPIRATOR
7. Block the patient connection on the Suction Canister and adjust the
Vacuum Control Knob to the desired vacuum.
To INCREASE vacuum - Turn knob CLOCKWISE
To DECREASE vacuum - Turn knob COUNTERCLOCKWISE
• Vacuum should be set as prescribed by your physician or healthcare professional.
NOTE: Intermittent cycle starts in the OFF phase, therefore a delay occurs before the ON (suction) cycle begins.
8. turn selector Knob to the “Int” position to obtain an intermittent, adjustable vacuum that cycles on and off.
9. Connect one end of Patient Suction Tubing to the port on the Canister marked Patient Tubing.
10. Connect other end of Patient Suction Tubing to suction catheter and begin procedure as instructed by healthcare provider.
11. When procedure is completed, press the on/off switch to the “oFF” position.
• DO NOT overfill the canister.
12. Empty the Suction Canister as follows:
• Disconnect all suctioning tubing.
• Remove the canister from its bracket and empty as instructed by your Physician or Healthcare Professional.
• Replace Suction Canister and reconnect all suctioning tubing.
CAUTION
Canister is for single patient use only.
CLEANING / MAINTENANCE
WARNING
• Turn the Power Vac + Aspirator off and disconnect from power source before cleaning.
• dO NOT attempt to repair. There are no serviceable parts inside.
The Inline Hydrophobic Filter cannot be cleaned. If it becomes contaminated or clogged, it must be replaced.
To replace Inline Hydrophobic Filter:
1. Disconnect filter assembly from Suction Canister and remove Suction
Canister from the Aspirator.
2. Remove filter assembly from the Aspirator.
3. Dispose of filter only.
4. Install new Inline Hydrophobic Filter (Part #1690) and reconnect tubing and elbow.
Tubing is single patient use only. DO NOT clean.
ASPIRATOR
7
To replace Air Filters:
CAUTION
• Excessive dust build-up on air filters will reduce performance of
Aspirator. If this occurs, filters should be cleaned or replaced.
• dO NOT use where saline mist could be drawn into the Aspirator.
The Air Filters do not filter out saline.
Saline causes rapid deterioration of the Aspirator and will void the warranty.
1. Check both Air Filters on top rear of cabinet for dust build-up.
2. If dust build-up occurs, unplug Aspirator, remove Air Filters from their recessed areas and wash in warm, soapy water and rinse.
3. Absorb excess water from filters with a dry towel.
4. Return filters to their original locations.
Cleaning for Exterior Components:
Use a cloth or sponge dampened with soap and water.
Wipe dry with a clean cloth or paper towel.
WARNING
• dO NOT immerse the Power Vac + Aspirator in liquid.
• If Aspirator is used for treatment of infectious disease, consult your
Physician or Healthcare Professional for recommended cleaning procedures.
RETURNS
Returned products require a Returned Goods Authorization (RGA) number, contact Precision medical, Inc. All returns must be packaged in sealed containers to prevent damage. Precision medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision medical, Inc. Return Policy available on the Internet, www.precisionmedical.com.
Manuals available on our Website; www.precisionmedical.com
NOTE:
When returning the Power Vac + Aspirator, disconnect the power
cord from the back of unit.
dISPOSAL INSTRUCTIONS
Dispose of the Power Vac + Aspirator in accordance with the local regulations.
Please Recycle
WARNING
Product should be cleaned before being disposed of. Potential for Biohazard.
8
ASPIRATOR
REPLACEMENT PARTS
Part # description
1690
1955
Inline Hydrophobic Filter
Suction Tubing, 1/4 ID x 6 ft, Clear Non-Conductive PVC
502391 Small Air Filter
502471 Large Air Filter
502687 Suction Canister - 2,000 cc (with lid)
502690 Inline Hydrophobic Filter & Tube Assy
503240 Fuse
TROUBLESHOOTING
If the Power Vac + Aspirator fails to function, consult the Troubleshooting
Guide below. If problem cannot be solved, consult your Provider.
Problem
Aspirator will not operate
Probable Cause
1. On/Off switch in
“oFF” position
2. Power Cord not connected properly
3. Thermal limit exceeded
Remedy
1. Push On/Off switch to
“on” position
2. Check plug at outlet
No vacuum on gauge / no vacuum at Patient
Connection
Tubing
4. Blown Fuse
1. Hydrophobic Filter assembly not connected properly
2. Suction Canister not connected or assembled properly
3. Defective Suction
Canister
4. Defective regulator
1. Defective vacuum gauge
No Vacuum on gauge/vacuum at Patient Tubing
Connection
Vacuum on gauge
/ No vacuum at
Patient Tubing
Unable to adjust vacuum
1. Clogged Hydrophobic
Filter
2. Canister overfilled
1. Defective regulator
3. shut Aspirator “oFF”, wait 15 minutes, be sure vents are clear
4. Replace Fuse
1. Confirm and tighten connections
2. Confirm and tighten connections/assembly
3. Replace Suction
Canister
4. Contact your Provider
1. Contact your Provider
1. Replace Hydrophobic
Filter
2. Empty canister
1. Contact your Provider
ASPIRATOR
9
LIMITEd WARRANTY
ANd
LIMITATION OF LIABILITY
Precision medical, Inc. warrants that Power Vac + Aspirator (the
Product) will be free of defects in workmanship and/or material for the following period:
Three (3) years from shipment
Should any failure to conform to this warranty appear within the applicable period, Precision medical, Inc. shall, upon written notification thereof and substantiation that the goods have been stored, installed, maintained and operated in accordance with Precision Medical, Inc.’s instructions and standard industry practice, and that no modifications, substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense.
ORAL STATEmENTS DO NOT CONSTITUTE WARRANTIES.
The representatives of Precision medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale. Thus, this writing is a final, complete and exclusive statement of the terms of that contract.
THIS WARRANTY IS ExCLUSIVE AND IS IN LIEU OF ANY WARRANTY
OF mERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR
OTHER WARRANTY OF QUALITY, WHETHER ExPRESS OR ImPLIED.
Precision medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property. Correction of non-conformities as provided above shall constitute fulfillment of all liabilities of Precision medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision medical, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or specifications without notice.
Precision medical, Inc. reserves the right to correct clerical or typographical errors without penalty.
Precision medical, Inc.
300 Held Drive
Northampton, PA 18067 USA
Toll Free Tel: 800-272-7285
Toll Free Fax: 800-352-1240
Tel: (+001) 610-262-6090
Fax: (+001) 610-262-6080
ISO 13485 Certified