POL-EKO SMART ZLN-UT 500 Instruction Manual

Add to My manuals
92 Pages

advertisement

POL-EKO SMART ZLN-UT 500 Instruction Manual | Manualzz
Instruction manual SMART PRO
Laboratory freezers
models:
ZLN 85
ZLN-T 125, ZLN-T 200, ZLN-T 300
ZLW-T 200, ZLW-T 300
Ultra-low freezers
models:
ZLN-UT 130 VIP, ZLN-UT 200 VIP, ZLN-UT 300 VIP, ZLN-UT 500 VIP
Before using the equipment, please read carefully this instruction manual!
Version 1.28
Issued 4.01.2022
1
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Manufacturer's address:
POL-EKO-APARATURA sp. j.
ul. Kokoszycka 172 C
44-300 Wodzisław Śląski
Country of origin: Polska
As a manufacturer, we inform you that we took the necessary measures to ensure that this device fully meets your
expectations and is reliable for a long period of use. Due to the continuous improvement of our products, as well as the
expansion of our offer, any suggestions regarding additional functions and equipment functioning are welcome. Visit
our homepage www.pol-eko.com.pl/home-en/
Equipment disposal
This equipment is marked with the crossed out wheeled bin symbol to indicate that this equipment must not be disposed of with unsorted waste. Instead it's your responsibility to correctly dispose of your equipment at lifecycle -end by
handling it over to an authorized facility for separate collection and recycling. It's also your responsibility to
decontaminate the equipment in case of biological, chemical and/or radiological contamination, so as to protect from
health hazards the persons involved in the disposal and recycling of the equipment. For more information about where
you can drop off your waste of equipment, please contact your local dealer from whom you originally purchased this
equipment.
By doing so, you will help to conserve natural and environmental resources and you will ensure that your equipment is
recycled in a manner that protects human health.
Thank you!
2
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Contents:
1.
INTENDED USE AND IMPORTANT INFORMATION FOR THE USER ......................................................... 6
2.
PACKAGE CONTENTS .................................................................................................................................. 7
3.
BEFORE THE FIRST USE .............................................................................................................................. 7
4.
5.
6.
3.1.
Installation of shelves ............................................................................................................................. 9
3.2.
Remarks on the placement of samples................................................................................................. 12
3.3.
Closing chamber door........................................................................................................................... 12
3.4.
Anchoring two-chamber equipment ...................................................................................................... 13
DESCRIPTION OF THE DEVICE.................................................................................................................. 14
4.1.
Design of ZLN 85 .................................................................................................................................. 14
4.2.
Design of ZLN-T 125, ZLN-T 200, ZLN-T 300, ZLW 200, ZLW 300 devices ........................................ 16
4.3.
Design of ZLN-UT 130 VIP, ZLN-UT 200 VIP, ZLN-UT 300 VIP, ZLN-UT 500 VIP devices ................. 18
DEVICE EQUIPMENT (standard and optional) .......................................................................................... 20
5.1.
Internal door (standard for ZLN-UT VIP)............................................................................................... 20
5.2.
Door lock (standard for all units) ........................................................................................................... 20
5.3.
Access port for external sensor (standard for all units) ......................................................................... 20
5.4.
Open door alarm (standard for all units) ............................................................................................... 20
5.5.
USB port (standard for all units)............................................................................................................ 21
5.6.
Display battery backup (optionally for ZLN, ZLN-T, ZLW-T, standard for ZLN-UT VIP)........................ 22
DEVICE OPERATION ................................................................................................................................... 22
6.1.
External memory (USB flash drive)....................................................................................................... 22
6.2.
First boot ............................................................................................................................................... 22
6.3.
Using the keypad .................................................................................................................................. 23
6.4.
User logging in ...................................................................................................................................... 23
6.5.
6.5.1.
6.5.2.
6.5.3.
Main screen ................................................................................................................................. 24
Information panel.............................................................................................................................. 25
The meaning of icons and symbols .................................................................................................. 27
Upper expandable and configurable menu ....................................................................................... 29
6.5.4.
6.5.5.
Quick Note – user’s message ................................................................................................. 30
Alarm bar .......................................................................................................................................... 31
6.6.
Quick Program ...................................................................................................................................... 31
6.7.
6.7.1.
6.7.2.
6.7.3.
6.7.4.
6.7.5.
6.7.6.
Programs..................................................................................................................................... 33
Creating / editing a program ............................................................................................................. 33
Segments edition.............................................................................................................................. 34
Summary of segments...................................................................................................................... 36
Priority .............................................................................................................................................. 37
Loop ................................................................................................................................................. 37
Defrosting of ZLN, ZLN-T, ZLN-UT VIP devices............................................................................... 37
6.8.
Starting the program ............................................................................................................................. 38
6.8.1. The first way ..................................................................................................................................... 38
6.8.2. The second way ............................................................................................................................... 39
3
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
6.9.
Quick Change of parameters ................................................................................................................ 40
6.9.1. Quick change of set temperature ..................................................................................................... 40
6.9.2. Quick change of set time .................................................................................................................. 41
6.10.
Schedules ................................................................................................................................... 42
6.10.1. Creating / editing a schedule ............................................................................................................ 43
6.10.2. Starting a schedule ........................................................................................................................... 46
6.11.
Statistics ...................................................................................................................................... 48
6.12.
Data record.................................................................................................................................. 49
6.12.1. Graph ............................................................................................................................................... 50
6.13.
Event log ..................................................................................................................................... 51
6.14.
Info .............................................................................................................................................. 54
6.15.
Users ........................................................................................................................................... 55
6.15.1. Creating / editing a user ................................................................................................................... 56
6.15.2. Account types and their limits ........................................................................................................... 57
6.16.
User settings panel ...................................................................................................................... 59
6.16.1. Unlocking the touch screen .............................................................................................................. 60
6.17.
Time ............................................................................................................................................ 61
6.18.
Alarms ......................................................................................................................................... 62
6.18.1.
7.
6.19.
Network ....................................................................................................................................... 64
6.20.
E-mail reports .............................................................................................................................. 65
6.21.
Automatic defrosting function (standard for ZLW-T) .................................................................... 67
6.22.
Temperature – additional temperature sensor Pt 100 (option).............................................................. 68
6.23.
Corrections................................................................................................................................... 69
6.24.
Archives ..................................................................................................................................... 70
INTERFACE .................................................................................................................................................. 71
7.1.
8.
Mute option ............................................................................................................................. 63
MODBUS TCP ...................................................................................................................................... 71
co2 back-up SYSTEM (OPTIONALLY FOR ZLN-UT VIP) ........................................................................... 71
8.1.
Indications for using pressure cylinders ................................................................................................ 72
8.1.1. Characteristics of carbon dioxide ..................................................................................................... 72
8.1.2.
8.1.3.
8.1.4.
8.1.5.
Hazards when working with CO2 ...................................................................................... 73
Personal protection .......................................................................................................................... 73
Control devices ................................................................................................................................. 73
First aid............................................................................................................................................. 73
4
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
8.2.
User's responsibility .............................................................................................................................. 74
8.3.
Emergency shutdown of the CO2 back-up system ............................................................................... 74
8.4.
Connecting the freezer to a CO2 cylinder ............................................................................................. 75
8.5.
Disconnecting the freezer from the CO2 cylinder ................................................................................. 78
8.6.
Starting the device ................................................................................................................................ 78
9.
CONNECTING THE DEVICE TO A COMPUTER ......................................................................................... 79
10.
CLEANING AND MAINTENANCE OF THE DEVICE ................................................................................... 79
10.1.
Exterior cleaning ................................................................................................................................... 79
10.2.
Interior cleaning .................................................................................................................................... 79
10.3.
Cleaning the touch screen .................................................................................................................... 80
10.4.
Consumables ........................................................................................................................................ 81
11.
ADVICE ON HOW TO SAFELY STORE THE DEVICE ................................................................................ 81
12.
TROUBLESHOOTING .................................................................................................................................. 81
12.1.
Possible defects.................................................................................................................................... 82
13.
WARRANTY CONDITIONS .......................................................................................................................... 83
14.
RATING PLATE ............................................................................................................................................ 84
15.
TECHNICAL DATA ....................................................................................................................................... 85
15.1.
ZLN, ZLN-T, ZLW-T devices ................................................................................................................. 85
15.2.
ZLN, ZLN-UT VIP devices .................................................................................................................... 87
16.
DECLARATIONS OF CONFORMITY ........................................................................................................... 90
5
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
1. INTENDED USE AND IMPORTANT INFORMATION FOR THE USER
Laboratory freezers are laboratory equipment intended for freezing and storage of frozen samples below 0°C:
•
laboratory freezer ZLN: -25°C … 0°C
•
laboratory freezer ZLN-T: -40°C … 0°C
•
laboratory freezer ZLW-T: -40°C … 0°C
•
ultra-low freezer ZLN-UT VIP: -86°C...-50°C
All devices are controlled by a precise SMART PRO controller, thanks to which the set temperature is maintained with
good fluctuation and variation. The ZLN 85 freezer is also available in a 2-chamber version with ST thermostatic
cabinet or CHL laboratory refrigerator, where each chamber is independently controlled – for service and operation of
ST and CHL devices, see the relevant instruction manual. In 2-chamber devices, the freezer is always at the bottom.
For devices with forced air convection, the symbol
appears on the screen.
The meaning of information symbols
This symbol means that failure to follow the instructions could endanger people's health or life, or
damage the device. The manufacturer is not liable for damages resulting from non-compliance with
the instructions contained in the manual.
A flammable coolant is used in the cooling system. If the cooling system is damaged, ventilate the
room carefully and remove all open flames near the unit.
There are extremely low temperatures inside the chamber so do not touch the samples and the
interior of the chamber without suitable protective gloves.
This symbol indicates helpful tips.
To guarantee your security and the longevity of the unit, please comply with the following rules:
1.
2.
3.
The unit cannot be installed:
• outside,
• in damp places or places which can be easily flooded,
• near flammable or volatile substances,
• near acids or in corrosive environments.
It is forbidden to:
• store inflammable or volatile substances inside the unit,
• touch live parts of the unit,
• operate the unit with wet hands,
• put water vessels on the unit,
• climb on the unit,
• overload the shelves (the maximum load is described in technical data).
You should:
•
place samples in such a way to provide proper air circulation in the chamber,
•
open the door for the shortest period of time to reduce temperature fluctuations,
•
secure samples from being blown out by the chamber fan e.g powdery samples,
•
always check that the doors are closed correctly,
•
use only mains with earth to avoid electric shocks,
•
unplug the power cable holding the protective cover and not the cable itself,
•
disconnect the unit from the mains before undertaking any repairs or maintenance work (in order to not
lose the warranty during its duration, all repairs should be carried out by an authorized service),
6
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
•
•
•
•
protect the power cable and the plug from any damage,
disconnect the power plug before moving the unit,
disconnect the power plug if the device will not be used for a long period of time,
disconnect the unit and protect it from reconnecting if it has any visual fault.
Failure to comply with the above recommendations may result in damage to the device or deterioration of
technical parameters, as well as loss of warranty.
2. PACKAGE CONTENTS
Laboratory freezers in SMART PRO version are delivered with:
Device
ZLN
ZLN-T
ZLW -T
ZLN-UT VIP
Capacity
85
125
200
300
200
300
130
200
300
500
Shelves [pcs.]
2
2
2
3
2
3
1
2
2
4
Slides [pcs.]
4
4
4
6
4
6
x
x
x
x
Power cord [pcs.]
1
1
1
1
1
1
1
1
1
1
2
2
2
2
Silicone cap [pcs.]
2
2
2
2
2
2
Key for door lock [pcs.]
2
2
2
2
2
2
2
2
2
2
Ethernet cable [pcs.]
1
1
1
1
1
1
1
1
1
1
Lab Desk program (in the internal memory of the
equipment)
1
1
1
1
1
1
1
1
1
1
Wrench (13mm) for wheels adjustment [pcs.]
x
x
x
1
x
1
1
1
1
1
Quality Control Certificate [pcs.]
1
1
1
1
1
1
1
1
1
1
3. BEFORE THE FIRST USE
The manufacturer sends the device protected by cardboard profiles and foil. The device should be transported in an
upright position and the package should be secured against sliding during transport.
After receiving the device, visually assess its condition and equipment in the presence of the
person delivering the goods. A courier company is responsible for any damage caused during
transport.
While carrying the unit, do not tilt it to one side more than 45° from the upright position, as there is
a high probability of damaging the compressor. If it is necessary to tilt it to one side more than 45°,
then after placing it, please wait about 3 hours before connecting the unit to the mains.
After transporting the device at a temperature below 10°С, wait at least 2 hours before connecting it
to the mains.
On the surface of unit’s components made of stainless steel, slight discoloration may occur. It is a result of the
technologies used in the production of metal sheet in accordance with the requirements of PN-EN 10088-2 standard
and it is not a defect of the unit.
7
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
The place of installation of the unit should meet the following conditions:
•
ambient temperature +10°C...+28°C,
•
recommended relative humidity of the ambient air up to 60%,
•
the unit has not been designed to work in highly dusty environments,
•
ensure proper ventilation in the room,
•
the device should be placed on a hard and stable surface,
•
the unit should be placed at least 100mm away from the walls,
•
the height of the room must be at least 300mm greater than the height of the unit,
•
the unit is not designed to be built-in,
•
the place of installation of the device should be equipped with a socket with parameters suitable for the
device.
If you don’t comply with the above recommendations, the unit may get broken or it may worsen the technical
parameters such as:
•
temperature fluctuation,
•
temperature variation,
•
power consumption,
•
icing of evaporator
and may result in loss of warranty.
Wheels / leveling feet
If the device has been equipped with wheels or leveling feet:
•
in the case of wheels after placing the unit at its destination, secure the device against
movement by locking the wheels
•
in the case of feet, after placing the device in the destination, they should be leveled.
If the device is equipped with wheels, they must be locked. For this purpose use the red
knob mounted in the wheel housing. At the beginning, the knob can be turned by hand,
if it encounters resistance, use a 13 size wrench.
Leveling wheels are ONLY for positioning the device at its destination. They can not be used to
transport the device!
Electric installation
Power supply of the equipment is indicated on a rating plate on the unit. Connect the device to a
socket with ground in order to avoid electric shocks in case of the unit’s failure.
The installation should be protected by a 16A slow-blow fuse and equipped with a residual current device.
8
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
3.1. Installation of shelves
In the ZLN 85 freezer
To install the shelf or to change its position, follow these steps:
Install the shelf slide at the selected height by inserting it into proper slots on the wall of the device. Do the same with
the slide on the opposite wall.
Slide the shelf into the installed shelf slides. Now, the shelf is correctly installed!
To remove a shelf, perform the above steps in reverse order. To remove the shelf slide from the slots, lift it up and slide
it towards the rear of the chamber.
9
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
In the ZL-T 125, ZL-T 200, ZL-T 300 models
To install the shelf or to change its position, follow these steps:
Install the shelf slide at the selected height by inserting it into proper slots on the wall of the device. Do the same with
the slide on the opposite wall.
Slide the shelf into the installed shelf slides. Now, the shelf is correctly installed!
In the freezers: ZLN-T 125, ZLN-T 200, ZLN-T 300 there are full shelves with a hole in the middle in standard
equipment.
In the freezers: ZLW-T 200 and ZLW-T 300 there are perforated shelves (as shown above) in standard equipment.
To remove a shelf, perform the above steps in reverse order.
10
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
In the ZLN-UT 130 VIP, ZLN-UT 200 VIP, ZLN-UT 300 VIP and ZLN-UT 500 VIP models
In the ultra-low freezers ZLN-UT 200 VIP and ZLN-UT 300 VIP each separately closed compartment is equipped with a
shelf. In the ZLN-UT 500 VIP freezer, in each of the two separately closed compartments there are two shelves. In the
model ZLN-UT 130 VIP there is one compartment with one shelf.
ZLN-UT 500 VIP
Compartments in ultra-low freezers ZLN-UT 130 VIP, ZLN-UT 200 VIP, ZLN-UT 300 VIP, ZLN-UT 500 VIP
11
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
3.2. Remarks on the placement of samples
To provide proper air circulation and stable conditions in which the samples are stored in the chamber, it is necessary
to keep the following rules:
•
the max height of the samples should not exceed 1/3 of the space between the shelves,
•
approx. 1/3 of the width and depth of the shelf should remain empty, while the distances between the
samples, as well as between the samples and the wall should be approximately equal.
The picture below is an example of the placement of samples in the chamber.
H
Hmax
Front view
Shelf
Hmax
H
Hmax = 2/3H
w
Top view
y
u+v+w > 1/3 a
u
a
z
v
x
b
x+y+z > 1/3 b
u ≅v ≅w
x≅y≅z
Following the above rules will provide best optimal parameters of temperature fluctuation and variation.
3.3. Closing chamber door
Freezers door have been equipped with a gasket and open door sensor. The external door of the ZL-T and ZL-UT
freezers are equipped with a lock. In ultra-low freezers each compartment is closed with a separate door.
If the door has not been closed properly, an audible and visual alarm will start. You can set delay door alarm by: 30s, 1
min, 2 min, 5 min or 10 min (see Section 6.18).
12
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
3.4. Anchoring two-chamber equipment
In the case of high multi-chamber units (ST/ZLN 85 or CHL/ZLN 85) they must be anchored to the wall
with suitable fixings (the anchoring kit is supplied with the device). This prevents the unit from falling over.
Lack of anchoring may result in damage to the device and in extreme cases may endanger the
health or life of the person who using the device.
Rear view
1
2
1.
2.
3.
Install the mounting brackets (2) to the holes marked (1) on the back of the device using the supplied screws and nuts. Handles can be set up or
sideways.
Place the equipment at the destination.
Use the  6 mm wall plugs adapted to the
anchoring point to attach the holders. Elongated
mounting hole in the bracket 10x10.
13
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
4. DESCRIPTION OF THE DEVICE
SMART PRO models are equipped with a PID microprocessor temperature controller and a 7 inch colour touch screen
with a resolution of 800x480.
4.1. Design of ZLN 85
Below there’s a picture of ZLN 85 model with a description of the important components of the device.
Front view
6
5
4
1
3
2
1)
2)
7
3)
4)
5)
6)
7)
rating plate
access port ø20mm for external
sensor
external solid door
key lock
touch control panel
USB port
adjustable feet
14
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Rear view
2
1
3
4
5
1)
2)
3)
4)
5)
fuse
main power socket C14
main switch
condenser
cooling system
15
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
4.2. Design of ZLN-T 125, ZLN-T 200, ZLN-T 300, ZLW 200, ZLW 300 devices
Below there’s a picture of ZLN-T 200 model with a description of the important components of the device.
Front view
7
1
2
6
3
5
4
1)
2)
3)
4)
5)
6)
7)
rating plate
handle with bolt lock with a key
main switch
wheels
condenser housing
external solid door
touch control panel
16
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Rear view
3
2
1
1)
main power socket C20
2)
3)
fuse
balancing valve
17
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
4.3. Design of ZLN-UT 130 VIP, ZLN-UT 200 VIP, ZLN-UT 300 VIP, ZLN-UT
500 VIP devices
Below there’s a picture of ZLN-UT 500 VIP model with a description of the important components of the device.
Front view
1
9
2
8
7
6
3
5
4
1)
2)
3)
4)
5)
6)
7)
8)
9)
rating plate
handle with bolt lock with a key
main switch
wheels
condenser housing
emergency back-up system switch
external solid door
internal door
touch control panel
18
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Rear view
5
4
3
2
1
1)
2)
3)
4)
5)
main power socket C20
fuse
potential-free
alarm
contact
(option)
CO2 back-up system (option)
balancing valve
19
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
5. DEVICE EQUIPMENT (STANDARD AND OPTIONAL)
5.1. Internal door (standard for ZLN-UT VIP)
The interior of the ultra-low freezers ZLN-UT 130 VIP, ZLN-UT 200 VIP, ZLN-UT 300 VIP and ZLN-UT 500 VIP has
been divided into compartments, which are additionally isolated by means of the internal doors. This allows samples to
be removed without the risk of temperature rise in other compartments.
5.2. Door lock (standard for all units)
All devices have a key lock. In the model ZLN 85 the key lock is situated above the door. In the freezers ZLN-T, ZLW-T
and ZLN-UT VIP the bolt lock is located in the door handle.Two keys are supplied with the device.
5.3. Access port for external sensor (standard for all units)
A Ø20 mm access port can be used to insert an external temperature sensor for independent temperature control
inside the device. The access port has been secured with a silicone plug. The plug should cover the access port while
the unit is operating. If multiple cables have been inserted through the access port and it is not possible to use the plug,
secure the access port with adhesive tape. If you leave the access port open, it may affect temperature fluctuation and
variation inside the chamber.
5.4. Open door alarm (standard for all units)
All units have been equipped with an open door sensor. If you open the door, the icon:
will appear (the number
above the icon presents open door counter. Press the icon to cancel the counter. The counter is also cancelled by
turning of the device). If the door remains open longer than the time set by the user (30s, 1 min, 2 min, 5 min, 10 min)
an acoustic signal, red pulsating alarm bar and „open door” alarm with active status will appear.
20
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Open door sensor
in ZLN 85
Open door sensor in
ZLN-T and ZLN-UT
5.5. USB port (standard for all units)
The USB port on the front panel is used only to transfer data from the device's internal memory to the flash drive. To do
this insert the flash drive into the USB port on the front panel and then:
•
go to the main menu
•
go to the data record
•
press
•
press
,
,
and choose type of the file: *.csv, or *.plkx.
. Data has been copied.
After copying the data to the USB flash drive, before removing it from the USB socket, it should be unmounted by pressing the icon
in the top drop-down list (Figure 1). If the pendrive is not unmounted after
connecting to the computer, a message about pendrive damage may be displayed with a repair proposal,
when actually the pendrive is not damaged
Figure 1 Unmounting USB flash drive
Data saved in the * .csv file can be opened in a spreadsheet. Data saved as * .plkx can be opened in the Lab Desk
program (standard equipment). This program allows, among others, for data preview in the form of a table or a graph. It
also allows you to prepare a statistic report for a selected range of data. More information Section 6.1.
21
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
5.6. Display battery backup (optionally for ZLN, ZLN-T, ZLW-T, standard for
ZLN-UT VIP)
Units in the SMART PRO version can be optionally equipped with a battery backup of the display. This is a standard
equipment for the ZLN-UT VIP ultra-low freezers.
The power loss and switching to the battery backup display mode is signaled by a pulsating red frame around the display and a sound signal (if it is turned on). In the battery backup display mode, all parameters are displayed, ie temperature. Other alarms, e.g. exceeding the temperature range, are also signaled.
In order to extend the battery life, the display is dimmed all the time. Batteries are automatically charged in AC mode.
Batteries should be replaced every 12 months. During the warranty period, the replacement should be
performed by an authorized service. Otherwise, you will lose your warranty.
6. DEVICE OPERATION
This symbol means that a given window can be moved in the direction shown in the picture.
6.1. External memory (USB flash drive)
The external memory (USB flash drive) allows to copy: instruction manual, data record, event log and service log from
the device memory. Before first use the USB flash drive should be formatted in the FAT 32 file system. Insert the
device in the USB slot on the front of the device next to the display. Wait a few seconds, the correct reading is
indicated by the message „USB flashdrive connected” at the bottom of the screen.
USB slot is used to connect only a flash memory – a flash drive or a card reader with a memory card.
Connecting any other device (e.g. external hard drive) without consultation is not authorized by the
manufacturer and may damage the USB slot.
After copying the data to the USB flash drive, before removing it from the USB socket, it should be unmounted (see Section 5.5.).
6.2. First boot
During the first boot, the screen (Figure 2) will display information about saving the "Download" folder (with instruction
manual in pdf format) on the USB flash drive. In order to do it, insert the USB flash drive and wait a second to detect
the hardware, then press
.
If you press
you quit downloading the folder. The window will appear again during the next boot. You can tick
„Don’t show again” so that the window will not be displayed after switching on the device. You can always download
the „Download” folder in the Info submenu. More information Section 6.14.
Figure 2 Downloading files
22
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
6.3. Using the keypad
When operating the device, sometimes it is necessary to enter alphanumeric characters (e.g. when logging into the
system, entering the user name, etc.). In such cases, a keyboard appears on the display screen (Figure 3). In addition
to the standard letters, it contains symbols that correspond to the computer keyboard.
Figure 3 Keypad
Deleting the entire text
Changing to capital letters (it´s matters when entering login and password).
Changing to numbers and special characters.
Changing to letters.
Deleting the entered character.
Confirming the entered text / closing the keypad.
Sometimes entered characters can be hidden (then they are replaced with „* ”). This happens when
classified information is entered (e.g. when a user enters a password).
6.4. User logging in
Setting device parameters is only possible by the logged in user. To log in, press
window will appear (Figure.4):
login:
from 1 to 10 characters
password:
from 1 to 10 characters
Factory default login parameters:
login: admin
password: leave the password field blank
in the main screen. The login
Figure 4 Login panel
It is recommended during the first boot to set the password of the Admin account and write it down in a safe place to
avoid tampering with the device settings by unauthorized persons.
23
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
The password should be remembered or noted down because it is not possible to delete the password of
the admin account. If you lose your password, please contact the manufacturer's service. Deleting the
password is not covered by the warranty.
For information on user account types, see Section 6.15.2.
Logging out: press
6.5.
in the main menu. For automatic logout, go to the Section 6.16.
Main screen
After switching on the device, the main screen (Figure 5) appears. It contains the information about the device status.
After starting the program, additional information appears on the screen (Figure 6). For devices with forced air
convection (ZLW-T 200, ZLW-T 300) symbol
appears on the screen.
Figure 5 Main screen
From this point, any action requires logging in.
24
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 6 Main screen – running program in ZL
main
menu
set
temperature
segment
duration time
turning off alarm
sound
start/stop
program
current
temperature
in the
chamber
alarm bar
side-scrolling
information
panel
logged in user
active
panel
current date/hour
6.5.1. Information panel
There are four different windows in the information panel. Switching between them is done by swiping the finger left or
right.
Figure 7 Information panel
The icon
indicates information about which window is active.
25
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
6.5.1.1.
Alarms panel
On the first page of the information panel there’s alarms panel (Figure 8).
Figure 8 Icon: Alarms panel
alarms panel
In the alarms panel there’s a list with active alarms or the alarms that have occurred but have not been confirmed. When
the alarm is active, the alarm bar is red and the alarm event is displayed in the list with the status "active". When the alarm
event stops, the state changes to "inactive".
•
“delete” button – confirms and removes the alarm from the list (only inactive alarms can be deleted),
•
“confirm” button – confirms an alarm,
•
“details” button – displays a preview of all instances of selected alarm (Figure 9).
Figure 9 Alarm details
With more alarms, a button
appears on the right side of the list and allows you to enlarge the view to full screen.
26
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
6.5.1.2.
Status panel
The status of the device is displayed descriptively on the third page of the information panel (Figure 10).
Figure 10 Icon: Status panel
program name
user
priority
current segment
current loop
status
time set
time elapsed
time remaining
the name of running program
name of the user to whom the program is assigned
of time or parameters
currently running segment / total number of segments in the program
currently performing cycle/ total number of cycles to perform
stage of device operation, e.g. reaching or maintaining of set temp.
set time of running segment
elapsed time since the segment has started
remaining time until the end of the segment
6.5.2. The meaning of icons and symbols
The icon allows you to go to the main screen.
Automatic return to the home screen. Factory setting: disabled.
The icon allows you to go to the main menu.
Automatic logout. Factory setting: disabled.
Automatic screen lock. Factory setting: disabled.
Unmounting the USB flash drive before removing it from the USB socket.
Closed door, open door. The number above the icon presents open door counter.
Press the icon to cancel the counter. The counter is also cancelled by turning of
the device.
Fan icon. Rotating icon shows that the fan is running (for ZLW-T).
Ramp status: Chamber is currently cooling down.
Set temperature is reached.
27
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
The program will start on the given date / time.
Start delay activated.
Icon is visible only when the chamber is cooling down.
Available when the program is running Clicking the icon allows you to quickly
change the set temperature (Quick Change function).
When the program is running, click the icon to quickly change the time of
program duration (Quick Change function).
Indicates the time that has elapsed from the program start.
Countdown of the time remaining to the end of the program.
.
The arrow icon allows navigation between: segments, program parameters and
summary.
Starting the selected program. In the list of programs - the program is running.
Stopping the program.
Adding a new program to the program list. The user can create up to 40
programs.
Editing the selected program from the list. In the program list, a new program has
been created but not approved yet.
Removing selected program from the list.
Going to alarms panel.
Going to the status panel with information about the program parameters.
Going to the menu to create, edit, delete and start programs.
Canceling adding or editing of the program. Canceling changes.
Editing individual program segments (the program can have max. 100
segments).
Immediate start of the program selected from the program list.
Delayed start of the program from the list of programs. The program starts according to the set date and time.
Going to the SMART program (Quick Program function).
28
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Turning off of the alarm sound (open door alarm, exceeding temperature range).
Critical alarms (i.e. damage to the temperature sensor, temperature protection,
etc.) continue emitting a sound.
User message. Clicking on the icon allows you to enter a message.
The icon appears in the event log and symbolizes entered user message.
6.5.3. Upper expandable and configurable menu
When the program is running, in the upper part of the main screen there’s a bar menu with parameter icons
(temperature, time, mute function). These parameters can be quickly changed (Quick Change).
Depending on the model, after swiping your finger down (Figure 11) you will see icons for all parameters which can be
quickly changed (see Section 6.9). Among the options available in the bar you will find the following icons:
•
USB flash drive unmounting – more information Section 5.5.
•
Quick Note – more information Section 6.5.4.
•
mute option. Critical alarms e.g. damage to the temperature sensor, temperature protection (if available)
will be still emitted.
•
Quick Change (more information Section 6.9.) of:
o
program duration time
o
set temperature
Figure 11 Upper expandable menu when the program is running
29
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 12 Upper expandable menu when the program is stopped
Positions available on the upper bar can be personalized. Just drag the selected icon to a new location (Figure 13).
Figure 13 Changing icon’s position
6.5.4.
Quick Note – user’s message
During equipment’s operation, the user can save messages in unit’s memory with information about: the date of
inserting a new sample, observed changes in the samples, the place of sampling, etc. To enter a message you must
first log in and then press the icon
in the main screen in upper menu. Click on „Write a message” (Figure 14).
Using the keypad, enter the message and confirm it with the button
. Once entered, a message cannot be
changed. Entered notes can be seen in the event log, they are symbolized by the icon
6.13.
. More information Section
Figure 14 User’s message
30
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
6.5.5. Alarm bar
The alarm bar is a quick visual information about the device status. The colour of the bar indicates the status of the
device:
– blue - the device is working properly
– red bar and pulsating frame– active alarm
6.6. Quick Program
Quick program allows you to quickly start the program from the main screen position without having to enter to the
menu
.
The Quick program has several features that guarantee its uninterrupted operation:
•
•
•
•
you can not set the duration of the program - time is always set to infinity,
if the display fails, the program continues,
after the power supply is resumed (after its failure), the program continues,
to prevent the program from stopping accidentally, the STOP button was removed from the main window
(Figure 15).
Figure 15 SMART program
In order to go to Quick program click the icon
−
in the main screen. By clicking the appropriate icon you can set:
temperature (Figure 16),
Clicking the icon
starts the program in continuous mode (time set to infinity).
31
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 16 Starting the Quick Program
Press
Set
Start
Stopping a Quick program has been made difficult on purpose (this prevents the program from being stopped accidentally) – to stop a program, you have to go to the menu
button
, click the program window
and keep pressing STOP
for 5 seconds.
After configuriation of the Quick Program, it appears in the programs list (Figure 17). Quick Program is displayed at the
top of the list by default. Moreover, it cannot be deleted and cannot be assigned to a user of the User type.
Figure 17 Quick program on programs list
In Quick Program editing mode, you can change:
•
settings of the data recording interval,
•
settings of the protection class.
When the program is running you can change the temperature by pressing the icon
Quick program, your previous settings will be remembered.
. During next launch of the
32
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
6.7.
Programs
Press the icon of main menu
, and then press
. In this panel (Figure 18) you can run the selected program,
add a new one, edit the program, delete it or share it with another user and download the program from a USB flash
drive. The number of programs that can be created depends on the limit assigned by the Super Admin user. More
information on the rights and configuration of account types (Super Admin, Admin, User) see Section 6.15.2.
Start the selected program.
Stop the program.
Add a new program.
Edit the selected program.
Delete the selected program.
Download program from USB flash drive.
Share the selected program with the user type account.
Figure 18 List of programs
Information on the number of created programs / the maximum number of programs that the user can create is at the
top of the screen (programs: 1/10).
6.7.1. Creating / editing a program
Press the button
or
and a panel with program parameters will appear (Figure 19).
In this panel you can set:
•
Program name – after clicking, the keypad will appear and you will be able to enter the program name,
•
Segments number – max. 100 segments
•
Interval – frequency of saving the data in the data record (1 min, 2 min, 5 min, 10 min, 15 min, 30 min, 1 h),
•
Priority – the priority of time or parameters, more information Section 6.7.4.
•
Loop – the number of program repetitions, more information Section 6.7.5.
33
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 19 Program parameters
Cancels adding or editing of the program.
Going to the edition of program segments.
With more parameters, the window can be scrolled up and down.
6.7.2. Segments edition
For each program, you can set maximum 100-segment time-temperature profiles that allow you to gradually increase
or decrease the incubation temperature of the samples. This can e.g. protect the sample from so-called thermal shock.
Example of program operation with programmed segments (parameters priority):
Program 1 (for ZLN-T/ZLW-T)
segment1: temp. -5C, time 2 hours (after reaching the temperature -5C, it is maintained for 2 hours)
segment2: temp. -10C, time 3 hours (after reaching the temperature -10C, it is maintained for 3 hours)
segment3: temp. -15C, time 3 hours (after reaching the temperature -15C, it is maintained for 3 hours)
segment4: temp. -20C, time 2 hours (after reaching the temperature -20C, it is maintained for 2 hours)
segment5: temp. -30C, time 2 hours (after reaching the temperature -30C, it is maintained for 2 hours)
segment6: temp. -40C, time  (after reaching the temperature -40C, it is maintained in continuous way)
Press the buton
and the first program segment will appear (Figure 20).
In this window you can set:
•
temperature – target temperature which the device is to achieve in this segment
•
time – the time of maintaining the set temperature ([d hh:mm]) in days, hours and minutes. It is possible to
select continuous work ∞ in the last segment,
•
ramp time – the time of reaching the set temperature ([d hh:mm]) in days, hours, minutes.
The active value is highlighted in blue. The item highlighted in red means that the value is out of range and you should
enter another one, e.g. the temperature is above / below the operating range of the device or the protection
temperature.
In ZLW-T 200 and ZLW-T 300 freezers the fan is set to 100% by default and cannot be changed.
34
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Ramp time - setting a short time will not accelerate reaching the ramp, but the ramp will be reached in the
shortest possible time depending on the set temperature, ambient conditions and the possibilities of the
cooling system in the device.
The ramp parameters are factory set in accordance with the manufacturer's instructions. If it is necessary to set
individual parameters when reaching the segment temperature, activate the ramp edition field
your own values.
and set
With more parameters, the window can be scrolled up and down.
Figure 20 Program segment edition
The navigation between: segments, program parameters and summary is done by swiping your finger to the side or by
touching the icon
.
If, when editing a program, you automatically return to the home screen or you are automatically logged
out, the edited program will not be lost, but saved as a draft (see below).
After switching to the program edition, the information about the possibility of continuing changes in the program
settings appears (Figure 21).
35
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 21
6.7.3. Summary of segments
In the segments summary (Figure 22) all segments can be seen along with introduced parameters:
•
number of segments,
•
temperature,
•
duration time of the segment,
•
target time of reaching temperature of a given segment (ramp).
Figure 22 The summary of the segment
Confirms and saves the changes.
Cancels the entered changes in the segments and goes to program parameters.
36
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
6.7.4. Priority
Can be set in terms of:
Parameters:
In the program without ramp – the device starts the countdown of the segment time when the set temperature is
reached.
In the program with ramp – first, the device counts down the time of the ramp and then proceeds to the segment
countdown when the set temperature is reached. Regardless of whether the time of ramp elapsed.
It may happen that the device failed to reach the set temperature within the set time because the reaching time was too short. In such situation the reaching time will be prolonged and the segment’s time
countdown will start when the set temperature will be reached.
Time:
In the program without ramp – the device starts counting down the segment time when the program is started. Regardless of whether the temperature has been reached.
In the program with ramp – first, the device counts down the ramp time and after its expiry it proceeds to the countdown
of the segment time. Regardless of whether the temperature has been reached.
It may happen that the time of reaching was too short and the device failed to reach the set temperature
within the set time. Then the countdown of the segment time will start before reaching the set temperature. Thus, the actual time of device operating in the set temperature will be shortened.
6.7.5. Loop
The option is available if the number of segments is equal to 2 or more (maximum 100). When the program finishes the
last segment, the device starts the program again from the first segment. You can define if the program should be
carried out once (loop: 1) or multiple times (loop: 2 to 255). In order to set the program to be carried out continuously,
tick the „∞” option. If the time of the last segment is set to infinity, it will be treated as infinite only in the last cycle. In
other cycles it will be treated as 0.
Example:
Loop:3
segment1: temp. -10C, time 2 h
segment2: temp. -30C, time 2 h,
segment3: temp. -40C, time „∞”
The device will run segment1 and segment2 three times and then will go to segment3 which will last indefinitely.
Due to the fact that laboratory freezers do not have a heating system, when switching from a lower to a higher
temperature, the compressor turns off, and the temperature rises naturally, which greatly extends the cycle. The cycle
duration can be determined empirically.
6.7.6. Defrosting of ZLN, ZLN-T, ZLN-UT VIP devices
During operation, a layer of ice may form inside the freezer’s compartment. The speed of ice or frost formation depends
on several factors: ambient conditions (temperature, humidity), how often the door is opened, and the type of samples.
When an ice or frost layer covers the entire chamber, defrosting is necessary. Defrosting is performed manually in the
following order:
37
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
1.
turn off the device (unplug the power cable from the socket),
2.
open the door and allow the device to defrost (do not accelerate the defrosting process),
3.
wipe dry the water accumulating on the bottom and the walls of the chamber,
4.
turn on the device (insert the power plug into the socket).
6.8. Starting the program
The created program can be started in two ways:
6.8.1. The first way
•
Go to the main menu
•
Then select the program you want to activate and press „start” button
and press the icon „programs”
(Figure 23).
(Figure 24).
Figure 23 Main menu
Figure 24 Program management menu
If the program is running, the symbol
appears next to the program name on the list. The symbol
the program has been edited, but the changes have not been confirmed (Figure 25).
means that
38
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 25 List of programs with the selected status
6.8.2. The second way
•
In the main screen (Figure 26) press the icon in the upper right corner
•
Select the program you want to start (Figure 27). You have two additional options to start the program:
.
Immediate start of the program.
Scheduled program start according to the set date and time.
Figure 26 Main screen
39
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 27 Selection and launch of the program
A delayed start of backdated program is possible (up to 7 days back). This is possible for the programs
with time priority. Program segments that would last from the back date to the current date will be skipped.
If the program is running, the symbol
appears next to the program name on the list.
6.9. Quick Change of parameters
Although the ramp time has been included in the program, the Quick Change of parameters will take place
immediately while the temperature is being reached.
6.9.1. Quick change of set temperature
In order to quickly change the value of set temperature of a running program, press the icon
in the main screen
(Figure 28). The value of the temperature should be selected by scrolling the list up or down (Figure 29). Click
confirm the change.
to
Figure 28 Quick change of set temperature
40
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 29 Quick change of set temperature
If the set temperature is higher than the current temperature inside the chamber, the compressor turns off
and the device reaches the set temperature naturally (freezers do not have a heating system).
6.9.2. Quick change of set time
In order to quickly change the duration time of a running program, press the icon
in the main screen (Figure 30).
Select the number of days, hours and minutes by scrolling the list up or down (Figure 31). Click
change. To set the continuous work press
.
To change the way of displaying the time, press::
to confirm the
– to display the elapsed time
– to display the remaining time
To change only the way of displaying, you do not have to confirm it by
.
Figure 24 Quick change of set time
41
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 31 Quick change of set time
6.10.
Schedules
The option allows creating a list of programs to be implemented in a given time. You can create several independent
schedules. The schedules window contains a list of all created schedules of the logged-in user (Figure 32).
Before you start creating a schedule, you must create programs that you want to include in it.
On the upper part of the screen there is information about the number of created schedules / the maximum number of
schedules to be created (1/10).
Figure 32 List of schedules
Start the schedule
Stop the schedule
Add the schedule
42
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Edit the schedule
Delete the schedule
6.10.1.
Creating / editing a schedule
To create / edit a schedule, press the button
or
. The panel with schedule parameters will appear on the
screen (Figure 33). Press "Schedule" and use the keypad to enter the schedule name. The schedule may consist of up
to 10 programs.
Figure 33 Creating / editing a schedule
Add a new program to the queue (program must be previously created)
Edit and make changes to the selected position
Delete the selected position from the list
Save the schedule
Cancel instroduced changes
After pressing
(Figure 33) or
its start and end (Figure 34).
a window appears allowing you to select the program and the date and time of
43
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 33 Adding a program to the schedule
Select a program from the drop-down list and press on the field next to the inscription "program" (Figure 35). Information about the selected program will be displayed (Figure 36): number of segments, number of cycles, priority, temperature protection, upper protection, lower protection. This is only a preview of the parameters - it is not possible to change
them in this window.
Figure 35 Selection of the program
44
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 36
Figure 37 Information about the program
Press the 'start date' field and then set the date and time of program start.
Press the 'stop date' field and then set the date and time of the program end.
45
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
You can assign up to 10 programs to one schedule. In total you can create ten schedules.
When creating a schedule, consider the following restrictions:
• the start time of the first program on the list cannot be earlier than the current date and time,
• the start time of the next program on the list cannot be earlier than the end time of the previous program,
• the program end time cannot be later than the start time of the next program,
• the end time of the next program does not have to coincide with the start time of the next one, there may be a break
between them,
• if the program is not fully completed (due to setting a too short time of a schedule), it will be interrupted.
When choosing time intervals, consider whether they are long enough for the selected program to be
implemented. The duration of the program can be affected by: ambient conditions, samples and the
program carried out immediately before it.
6.10.2.
Starting a schedule
The schedule can be started in two ways:
6.10.2.1.
The first way
•
Press the icon of the main menu
•
Then select the schedule you want to activate and press the start button.
and then press the icon „schedule”
(Figure 38).
(Figure 39).
Figure 38 Main menu
46
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 39 List of schedules
Start the schedule
6.10.2.2.
The second way
•
In the main screen press the icon
(Figure 40), then press the SCHEDULE inscription. The schedule selection window will be displayed (Figure 41).
•
Then select the schedule you want to activate and press the button
.
Please note that it is not possible to run a schedule in which all parameters refer to the past time.
Figure 40 Main screen
47
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 41 Selection of the schedule
6.11.
Statistics
Press the icon of the main menu
and then press
. In this panel (Figure 42) statistics of the currently running
program or program that has ended are displayed. Statistics are calculated separately for each segment. Data logging
for calculation starts after 30 seconds from reaching the set temperature in the segment. Further data is registered
every 1 minute.
The following information is available:
•
set temperature [°C] – set temperature in the segment,
•
minimum temperature [°C] – the lowest recorded temperature,
•
maximum temperature [°C] – the highest recorded temperature,
•
average temperature [°C] – average temperature,
•
segment – status of the segment:
in progress – currently running segment (data is being constantly updated),
finished – the segment has been completed,
interrupted – the segment was interrupted by the user before the set time has elapsed,
•
segment 1/2 – the number of the currently overviewing segment / number of the currently performed or
completed segment. Navigating between the segments is done by swiping your finger up or down.
•
loop 1/1 – the number of the currently overviewing cycle / number of the currently performed or completed
cycle. Navigating between the cycles is done by swiping your finger left or right.
You cannot overview the segment / cycle data that has not started yet.
48
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 42 Statistics
6.12.
Data record
Press the icon of the main menu
and then press
. Data record window (Figure 43) contains the following
information:
•
time and date of sample registration [date],
•
temperature value measured with the main sensor in the chamber [temp.].
Each user can register 10 000 data records for the max period of 12 months. If all the memory cells are full, the oldest
ones are overwritten. Data is displayed in the chronological order from the oldest to the latest record. It is possible to
segregate data according to the selected column. To do so, click the column heading. The samples are only registered
when the program is running. The frequency of registration depends on the program parameters settings.
When opening the data record, data from the last 3 days is downloaded. Press
the data.
to display the rest of
Figure 43 Data record
49
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Press to display the rest of the data.
Recording data onto the USB flash drive. .csv files are available - separated by semicolon when opening
e.g. with a spreadsheet, .plkx - opening with the Lab Desk application
Before removing the USB flash drive from the USB port, it must be unmounted, see Section 5.5.
Displaying data as a graph, see Section 6.12.1.
Deleting data. Users with Super Admin privileges can delete all data, including those registered by other
users (Figure 44).
If there is a lot of data, a progress bar appears on the display:
Figure 44 Deleting data
6.12.1.
Graph
A graph can be generated from the data stored in the data register (Figure 45). The time during which the graph opens
depends on the number of saved samples data. If the unit is equipped with additional sensors, press the selected
graph twice.
Figure 45 Temperature graph
50
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Returns to displaying the entire chart (undo all magnifications) / returns to the list of charts.
Returns to data register
The opening time of the chart depends on the number of saved data samples. The greater the number of
saved samples, the longer this window will open.
You can enlarge a fragment of the chart. Press the graph anywhere and drag to the right and down simultaneously
(Figure 46) - enlarge in the same way as it is done on a smartphone. Swipe left to return the chart to normal size.
Figure 46 Enlarging a part of the chart
6.13.
Event log
Data available for the following users (see Section 6.15 Users):
•
Super Admin who can overview, download to USB flash drive and delete all data,
•
Admin who can overview, download to USB flash drive all data,
•
User who can only overview all data.
51
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Press the icon of the main menu
alarms and errors.
and then press
. This window displays information about registered events,
Figure 47 Event log
Recording data onto the USB flash drive. .csv files are available - separated by semicolon when
opening e.g. with a spreadsheet, .plkx - opening with the LabDesk application
Before removing the USB flash drive from the USB port, it must be unmounted, see Section 5.5.
Deleting data
QR code – opens smart4lab.eu (in the "Support" tab there are explanations of some of the
information appearing in the event log). Press the symbol
then scan it with your smartphone
and enlarge the code QR, and
Figure 48 QR code
Before removing the USB flash drive from the USB port, it must be unmounted, see Section 5.5.
52
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Information signs in the event log:
Information event
Message entered by the user
Alarm event
Error
Warning
Possible events:
Program Start
starting the program
Program Stop
stopping the program
Program Edit
changing the program parameters
Program End
program is completed
DeviceOn
the device is switched on (on the main switch)
DeviceOff
the device is switched off (on the main switch)
Door opened
the door is opened
Open door alarm start
open door alarm has been activated
Door closed
the door is closed
Open door alarm stop
open door alarm has been deactivated
Program Restarted
program has been reasumed after power failure
Date/time change
date/time has been changed
Lower temp. alarm Start
activation of the alarm of exceeding the temperature below the set
temperature
Lower temp. alarm End
deactivation of the alarm of exceeding the temperature below the
set temperature
Upper temp. alarm Start
activation of the alarm of exceeding the temperature above the set
temperature
Upper temp. alarm End
deactivation of the alarm of exceeding the temperature above the
set temperature
Deleted Measurement
user measurements have been deleted
Deleted All Mesurement
all measurements have been deleted
User added
new user has been added
User updated
user has been changed
User deleted
user has been deleted
Program deleted
program has been deleted
Program updated
program has been updated
Time Zone Changed
in the time settings the time zone has been changed
Temperature Correction Changed
main sensor temperature correction has been changed
Emergency stop of the program
the program has been automatically stopped – there was a
situation that didn’t allow the program to be continued. PLEASE
CONTACT THE SERVICE
Defrosting Start
starting the defrosting process
Defrosting Stop
Power Fail Start
Power Fail Stop
User login
User logout
stopping the defrosting process
power failure / device fuse blown out.
power reasumed, returned to maintain program parameters
date and time of login
date and time of logout
53
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Quick note
To view message details, click
. In this window (Figure 49) you can see the content of the message, the
name of the user who entered it and the name of the program during which the message was written.
Figure 49 Details of user’s message
6.14.
Info
Press the icon of the main menu
•
name of device,
•
serial number,
•
temperature range,
•
Software version,
•
manufacturer’s address,
•
manufacturer’s website,
•
QR code.
and then press
.The panel contains the following information:
Figure 50 Info window (example)
54
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Press icon
to save the “Download” folder (with instruction manual) on the USB flash drive. After inserting the flash
drive into USB port wait few seconds until the information “Flashdrive connected” will appear on the display - for more
information go to the Section 6.1. Press the icon
service department for more information.
to write the service data on the USB flash drive – contact the
Press
to go to the main screen. If a USB flsh drive is connected to the device, when entering the "Info" panel, a
proposal to save configuration file will appear (Figure 51). This file is used to create an offline program in the Lab Desk
application.
Figure 51 Saving configuration file
Before removing the USB flash drive from the USB port, it must be unmounted, see Section 5.5.
6.15.
Users
Go to the main menu
one or delete it.
and press the icon
. In this panel (Figure 52) you can add a new user, edit an existing
55
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 52 Users list
Adding a new user
Editing selected user
Deleting selected user - his programs and data register will be deleted with the user.
At the top of the screen (Figure 52) you can see information about:
•
users: number of created users / total number of users to create (users 2/5),
•
available programs: the number of free programs to be assigned to users.
6.15.1.
Creating / editing a user
To add or change user settings, press the button
or
, a panel with user data will be displayed (Figure 53).
You have to enter:
•
login – user name,
•
password – account password,
•
password confirmation – you must enter the password again to confirm it,
•
type – account type (Super Admin , Admin, User), for more information see Section 6.15.2.,
•
programs limit – number of programs that can be created by the user / number of available programs (it’s not
possible to set a limit to the User).
56
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 53 Editing a user
Confirms and saves the user
Cancels introduced changes and returns to the users list
The device can have maximum 5 users. There are 40 programs available which can be freely distributed
among users.
6.15.2.
Account types and their limits
Three different types of users (accounts) are available: Super Admin, Admin, User. Each user has their rights and
limitations described below in the table 1.
Table 1. Right and limitations of the users.
Super Admin
Admin
User
Creating programs
✔
✔
✘
Editing programs
✔
✔
✘
Stopping your own program
✔
✔
✔
Stopping a program of another user
✔
✘
✘
Quick program
✔
✔
✘
Quick change of set time
✔
✔
✘
Quick change of set temperature
✔
✔
✘
Assigning a program to a user of the user type
✔
✔
✘
Creating a schedule
✔
✔
✘
Editing a schedule
✔
✔
✘
Defrost control
✔
✘
✘
Management of the illumination shelves in the unit (FIT version)
✔
✘
✘
Setting temperature measurement parameters
✔
✘
✘
Temperature value correction
✔
✘
✘
Setting the alarms
✔
✘
✘
57
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Temporarily silencing the alarms
✔
Enabling / disabling the sound
✔
✘
✘
Saving a Quick Note
✔
✔
✔
Creating users accounts
✔
✘
✘
Changing user's settings
✔
✘
✘
Changing equipment's name
✔
✘
✘
Setting a time zone
✔
✘
✘
Changing the language
✔
✘
✘
Setting the automatic logout time
✔
✘
✘
System information preview
✔
✔
✔
Statistics preview
✔
✔
✔
WiFi settings
✔
✘
✘
LAN settings
✔
✘
✘
Setting e-mail reports
✔
✘
✘
Access to the archive
✔
✘
✘
Events preview
✔
✔
✔
Deleting events
✔
✘
✘
Copying data to a pendrive
✔
✔
✘
Data preview
✔
✔
✔
Copying data to a pendrive
✔
✔
✔
Displaying data as a graph
✔
✔
✔
Deleting your own data
✔
✔
✔
Deleting all data
✔
✘
✘
Reseting the open door counter
✔
✘
✘
✔
✔
Super Admin account
The Super Admin account has no limits. Has access to the program management menu and to the settings menu, go
to the table 1.
Figure 54 Menu available for Super Admin
Information about the currently displayed window indicates
.
58
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Admin account
Has access to programs menu
and has rights and limitations in accordance with table 1.
Figure 55 Menu available for Admin
User account:
•
has access to programs menu
, where User can start programs previously assigned to him, check their
statistics (statistics, data register), check events history of the equipment (event log) and the information
about the system (info),
•
can´t create his own programs and schedules but start those which has been assigned to him by Super
Admin,
•
can´t stop or edit a program or schedule which wasn´t started by him,
•
doesn´t have access to create or edit schedules,
•
the program started by the User can be stopped by a user with Super Admin privileges.
Other rights and limitations of the User type account are shown in table 1.
Figure 56 Menu available for User
6.16.
User settings panel
Press the icon of the main menu
and then press
. In this panel (Figure 57) you can:
59
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Change the name of the equipment – by default, the device serial number is entered.
Change the language in the equipment´s menu.
Set the time after which the screen will be dimmed.
Turn on/off the sound . Critical alarms will continue emitting a sound.
Set the time after which the user will be automatically logged out. Available settings: off, 1 min, 3
min, 5 min, 10 min. Factory setting: enabled.
Set the time after which the user will return to the home screen. Available settings: off, 1 min, 3 min,
5 min, 10 min.
Set the automatic screen lock. Available settings: off, 5 min, 15 min, 30 min, 60 min. Factory setting:
disabled.
Only one feature can be enabled at the same time: automatic logout or automatic screen lock.
Figure 57 User settings panel
Confirms changes
Cancels the entered changes
6.16.1.
Unlocking the touch screen
When the automatic touch screen lock is enabled (Section 6.16), slide the blue circle into the white circle to unlock the
screen.
60
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 58 Unlocking the touch screen
6.17.
Time
Go to the main menu
zone.
and press the icon
. In this panel you can change the date and system time and time
Change of the date / system time
If the date / system time is changed to the later date / time comparing with the data and events which are
stored in the memory, they will remain in the register. If the date / system time is changed to the earlier
date than the date / time which is stored in the memory, they will be transferred to the archive
After changing the date/system time the device will be restarted.
To change the date / system time it is necessary to press
you will be able to make changes (Figure 60).
in the window (Figure 59). The window will appear and
Figure 59 Time change
61
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 60 Date / time change
Confirms changes and restarts the device
Cancels the entered changes
Change of time zone
The change of time zone will not affect the date / time in data and events previously saved.
To change time zone, you have to press the buton
in the window (Figure 59). Select the time zone from the dropdown list. After changing only the time zone, the device is not restarted.
Confirms changes
Cancels the entered changes
The same time zones on the device and computer are required for correct operation of the programs.
6.18.
Alarms
Go to the main menu
and press the icon
. You can set parameters related to alarms (Figure 61).
•
lower alarm – an alarm will be activated if the temperature drops below the value given in this field,
•
upper alarm – an alarm will be activated if the temperature rises above the value given in this field.
The lower and upper alarm can only be activated after reaching the set temperature.
•
delay temp alarm:
•
the alarm will be activated with a delay (1 min, 2 min, 5 min, 10 min, 15 min) after exceeding the permitted
temperature.
delay door alarm:
the door alarm will be activated when the door is opened for the time selected by the user (30 s, 1 min, 2 min,
5 min, 10 min).
62
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 61 Alarms
Confirms changes
Cancels the entered changes
In the field "lower alarm temp" you can set a value of -0,5°C to - 5°C and in the field "upper alarm temp"
you can set a value of +0,5°C to +5°C.
6.18.1.
Mute option
The icon
in the main screen in the upper menu allows temporary switching off of the alarms sound (open door
alarm, exceeding temperature range), e.g. to avoid door alarm during planned loading of the samples into the chamber.
To set the mute time, press the icon
and choose: 5, 10 or 15 minutes (Figure 62), however, the sounds of critical
alarms (e.g. damage to the temperature sensor) will be still emitted.
Figure 62 Mute function
63
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
6.19.
Network
Go to the main menu
and press the icon
between LAN / WiFi network by pressing
or
. In this panel you can change the settings for LAN or WiFi. Switch
.
LAN settings:
•
IP – the device’s IP address
•
Mask – an Ethernet network mask to which the device is connected
•
Gate – Server’s IP address or router’s that manages the Ethernet network
•
DNS – IP address of the domain name system
•
MAC – the address of the network card, read-only
•
DHCP – you can select if the server that allocates IP addresses is running on the local network. You can then
skip setting IP, Masks, Gates
Icon
A
indicates the connection status:
Device connected to the network
Device disconnected from the network
Figure 63 LAN settings
A
Confirms changes
Cancels the entered changes
64
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
WiFi settings:
•
•
•
•
•
– press to refresh network list,
SSID – press to select network from the drop-down list,
PSK – network password,
IP, Mask, Gate, DNS – after a successful connection to the network these fields are automatically completed,
MAC – physical address of the network card, read-only.
Figure 64 WiFi settings
Confirms changes
Cancels the entered changes
6.20.
E-mail reports
Go to the main menu
mail notifications.
and press the icon
. In this window you can set the parameters needed to activate e-
In the panel there are three windows:
•
•
•
•
•
•
Selection of event types for which notifications should be sent (Figure 65).
Activate – turning on/off e-mail notifications,
Doors alerts – alarms triggered by an open door,
Critical alerts – critical alarms (e.g. sensor damage),
Temperature alerts – alarms caused by too high or too low temperature,
Programs events – program-related events (e.g. adding, editing, deleting a program),
Users events – events related to editing user settings (e.g. adding, editing, deleting users).
65
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 65 E-mail: events
Confirms changes
Cancels the entered changes
If the "activate" option at the top of the panel is not enabled, emails will not be sent!
Sender and recipients (Figure 66)
•
•
Sender – sender’s e-mail address
Recipients – recipients e-mail addresses, maximum 3
Figure 66 E-mail: Sender – Recipients
Confirms changes
Cancels the entered changes
66
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Configuration of the sender's e-mail account (Figure 67)
In this window, enter your e-mail account details:
•
SMTP server user ID
•
SMTP server password
•
SMTP server host
•
SMTP server port
You can also choose the TLS or SSL encryption method (get more information from your email account provider).
Figure 67 E-mail: email account configuration
Connection test
Confirms changes
Cancels introduced changes
Before testing the connection, make sure that the device is connected to the network and has a properly
configured network connection, see Section 6.19.
6.21.
Automatic defrosting function (standard for ZLW-T)
Press the icon of the main menu
and then press
. In this panel (Figure 68) you can control the defrosting of
the interior of the device. This is a standard equipment for the freezers with forced air convection: ZLW-T 200 and
ZLW-T 300. ZLW units are additionally equipped with evaporator temperature measurement and heating of the
condensate drain system. Defrosting has additional parameters to set:
•
preheat time of the tray [s] - time before defrosting during which the condensate drain system should be
preheated,
•
evaporator temperature [°C] – temperature (on the evaporator) for the end of defrosting. Defrosting ends
when the evaporator reaches the set temperature or after the time set in the time parameter has elapsed,
depending on what occurs first.
67
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
User can set parameters such as: time, preheat time of the tray, evaporator temperature.
CAUTION: too low values may cause that the accumulated ice will not melt during the defrost cycle, which
may result in increased icing. However, higher values can cause an unnecessary increase of the
temperature in the chamber.
Pressing the button
– restores the default defrost settings (period [h]: 2, time [s]: 120, delay [m]: 0,
preheat time [s]: 90, evaporator temperature [C]: 0.
NOTE: the option
must be selected.
Figure 68 Defrosting program
Confirms changes
Cancels the entered changes
6.22. Temperature – additional temperature sensor Pt 100 (option)
Go to the main menu
and press the icon
. In this panel (Figure 69). You can set parameters related to the
temperature measurement in the equipment using an additional temperature sensor.
Additional sensor
The following parameters can be set relative to the additional temperature sensor:
•
mode – sensor operation mode:
disable – the sensor is switched on,
measure – displaying in the main window and recording temperature from the additional temperature
sensor,
compensation - displaying in the main window and recording temperature from the additional
temperature sensor + compensation,
•
rate [C°/min] – determines how fast the equipment is to respond to differences in temperature indications
between the additional temperature sensor and the main sensor (higher value = faster response of the
equipment, can affect the stability of temperature maintenance),
•
limit – parameter which defines the acceptable difference in temperature compensation during operation; the
limit range is from 2°C to 8°C.
68
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Compensation
The compensation value is a correction for the regulation of the main sensor in order to obtain the set temperature at
the place of measurement of the additional sensor. To control the equipment according to the additional temperature
sensor’s indications, select the "compensation" operating mode. An additional temperature sensor can be used to
control the temperature in the chamber according to e.g. sample temperature.
Figure 69 Temperature measurement settings with an additional sensor
Confirms changes
Cancels the entered changes
6.23.
Corrections
Press the icon of the main menu
and then press
. In this window (Figure 70) you can correct temperature
value indicated on the display by adding the correction value. The set correction value applies to the whole temperature
range of the device. For example, if the average temperature displayed by the device indicates -20,0°C and the
average temperature measured by independent, external sensor indicates -20,5°C, the correction should be set on 0,5°C. The average temperature should be calculated from chosen period of time e.g. 30 min. The correction available
range is between -5°C to +5°C.
The device has been calibrated by the manufacturer in accordance with applicable norms. The temperature
shown on the display corresponds with a great accuracy to the temperature near chamber’s sensor. For the
correct operation of the device it is not necessary to use user’s calibration. The user is performing
temperature correction on his own responsibility and must be aware of consequences of changing
manufacturer’s settings. If the equipment was calibrated, calibration certificate loses its validity.
69
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 70 User’s correction
Confirms changes
Cancels the entered changes
6.24.
Archives
Press the icon of the main menu
and then press
. The archive folder contains data and events transferred
there while changing the date / time to an earlier one than the date / time with which they have been stored in the
device memory.
To download the data, insert the pendrive into the USB port on the front panel, select "archive" and press the icon
.
At the same time, data and event files are downloaded. All files are saved to the folder named by the downloaded
archive.
Figure 71 Archive
Save the selected archive onto USB flash drive
70
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Delete the selected archive
Before removing the USB flash drive from the USB port, it must be unmounted, see Section 5.5.
7. INTERFACE
7.1. MODBUS TCP
The device allows status monitoring using the MODBUS TCP communication interface.
Connection parameters:
•
IP address: same as device’s (set in the panel Section 6.18.)
•
port: 502
register INPUT REGISTERS
function READ_INPUT_REGISTERS (0x30001)
Address
Type
Multiplier
Description
0
int
10
temperature from the main sensor
3
bool
-
Open door
4
bit
-
b0 – door alarm
b1 – upper temperature alarm
b2 – lower temperature alarm
b5 – main sensor error
b6 – additional sensor error
b8 – temperature sensors error
b10 – hardware error
b11 – MRW error
8. CO2 BACK-UP SYSTEM (OPTIONALLY FOR ZLN-UT VIP)
The CO2 back-up emergency power supply system is used in ZLN-UT ultra-low freezers and provides protection against temperature rise in the chamber and maintains a safe temperature of the samples in the event of a power outage.
An increase of temperature above 10°C from the set temperature and loss of power supply causes the dosing of liquid
CO2 to the chamber. The amount of dispensed liquid CO2 is regulated by a microprocessor controller. The minimum
operation temperature for the CO2 back-up system is -70°C and results from the physical properties of carbon dioxide.
If the chamber door is open during CO2 dosing, the gas will be automatically cut off.
Time of temperature maintenance by CO2 back-up system depends on:
•
capacity of CO2 cylinder,
•
ambient temperature,
•
type and temperature of samples at the moment of power failure,
•
degree of freezer filling.
Theoretically, a full 50l high-pressure cylinder is enough for 8 hours of CO2 back-up operation in the ZLN-UT 300 VIP
freezer (set temp. -80°C).
71
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
• During the operation of the CO2 back-up emergency power supply system, a significant amount
of CO2 enters the room, which in high concentration can be life-threatening. The device does not
emit CO2 when working with the power supply switched on.
• The CO2 back-up system can’t be operated near concentrated acids or corrosive vapors.
• The key to turn off the CO2 back-up emergency power supply system must always be in an
easily accessible place.
• The workplace must be properly ventilated!
8.1. Indications for using pressure cylinders
The liquid CO2 pressure cylinder is NOT supplied with the freezer with CO2 backup system. The purchase,
transport and connection is the responsibility of the user.
•
•
•
•
Together with a pressure cylinder with compressed or liquefied gas, the user should be given a safety data
sheet for the gas and instructions for its use.
The ventilation performance in the room must ensure levels of gas concentrations in the rooms below the limit
values.
Opening the pressure cylinder valves is only permitted if the cylinders have been connected to the receiving
installation first.
During gas intake from the gas cylinder, check the pressure level regularly.
8.1.1. Characteristics of carbon dioxide
Carbon dioxide (CO2) is a colorless, odorless, non-flammable and slightly acidic liquid gas. CO2 is heavier than air and
soluble in water.
Warning signs on the bottle containing CO2
Non-flammable and non-toxic
gases
Pressurised gas
Product ID: Carbon dioxide (100%)
Chemical formula: CO2
Application of the substance: General industrial applications
Supplying gases other than CO2 to the freezer is UNACCEPTABLE.
72
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
8.1.2.
1.
2.
Hazards when working with CO2
The bottle contains compressed gas that may cause cryogenic burns or injury (extremely cold liquid and high
pressure gas).
Carbon dioxide in elevated concentration causes shortness of breath, circulation problems and ultimately death.
Avoid breathing gas.
8.1.3. Personal protection
1)
2)
Hand protection
Wear loose fitting thermal-insulated gloves or gloves for handling cryogenic liquids. Norm EN 511 - protective
gloves against the cold.
Eye or face protection
Wearing safety glasses is recommended when working with the bottle. Protect eyes, face and skin from splashes
of liquid. Use safety goggles and face shield when reloading the product or disconnecting the transmission
connections. Norm EN 166 – Personal eye protection.
3)
Skin and body protection
Never touch with exposed body parts uninsulated elements of the CO2 backup system. Extremely cold metal can
cause the body to stick quickly and to damage it (detach) when attempting to retract.
4)
Respiratory protection
In an atmosphere where there is a lack of oxygen (excess of carbon dioxide), use a self-contained breathing
apparatus or a face mask with a positive pressure air supply. Respirators with air filters will not provide protection.
8.1.4. Control devices
To prevent the accumulation of gas above the permissible concentrations in the room, natural or mechanical ventilation
should be provided. In addition, a natural or mechanical installation must be provided to prevent the oxygen
concentration from falling below 19.5%.
According to the recommendations, the highest acceptable concentration of CO2, that affects an employee during the
8-hour daily and average weekly working time specified in the Labor Code is: 9000mg / m3. Maximum permissible
instantaneous concentration: 27,000mg / m3
It is recommended to use control and measurement equipment for measuring CO2 concentration in rooms
with a CO2 backup system.
8.1.5. First aid
Contact with eyes: In case of contact with eyes, rinse them immediately with plenty of water and seek medical advice.
Keep eye wide open while rinsing.
Skin contact: In the event of frostbite, seek medical assistance immediately. Whenever possible, immerse the affected
area in a warm bath not exceeding 40°C (105°F). Do not rub frostbitten parts of the body as it may cause tissue
damage. Apply a sterile dressing to the wound.
Inhalation: Move the victim to fresh air. CO2 gas is heavier than air and accumulates e.g. at the floor. If breathing has
stopped or is difficult, give aided respiration. Oxygen administration may be indicated. In the event of cardiac arrest, a
trained person should immediately begin cardiopulmonary resuscitation.
73
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
in life-threatening emergencies:
1) protect yourself with a breathing apparatus,
2) remove the victim from the contaminated area,
3) call a doctor,
4) start CPR.
In the event of a gas leak:
Leave the room immediately, warn other people and ventilate the room if possible.
Emergency shutdown of the CO2 back-up system:
1)
2)
close the gas cylinder valve,
turn off the freezer by turning the rotary switch on the front panel of the device to the OFF position (Figure 37).
Figure 72 Rotary switch
8.2. User's responsibility
The user is obliged to:
1) be familiar and comply with applicable health and safety rules, regulations and to train operators properly,
2) secure the device against access by unauthorized persons,
3) keep the device in perfect condition,
4) follow the maintenance schedule,
5) ensure that operators use appropriate personal protective equipment,
6) share the user manual and CO2 safety data sheet with operators working with the freezer.
8.3. Emergency shutdown of the CO2 back-up system
In each ZLN-UT VIP ultra-low freezer at the front in the bottom there is an emergency switch for emergency temperature maintenance system. To turn off the device, turn the rotary switch key (Figure 72) of the CO2 Backup system to the
"OFF" position.
74
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
8.4. Connecting the freezer to a CO2 cylinder
The CO2 back-up system is located in the rear of the equipment (on the back).
To CO2
cylinder
1
1)
2)
2
Steel hose for connecting CO2 cylinder with 1/4'' SAE internal thread
CO2 back-up system
The manufacturer supplies the device with a specialist steel hose with a length of 1500 mm. This hose is used to connect CO2 cylinder. On one side the hose is screwed to the CO2 back-up system - DO NOT unscrew it. On the other
side, it should be screwed to a CO2 cylinder (see below "Connecting the equipment to a CO2 cylinder"). The kit comes
with a reduction on cylinder W21,8 x 1/4 '' SAE. The minimum bend radius of the hose is 110 [mm].
75
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
The bending radius of the steel hose should not be smaller than specified in the documentation, otherwise
the hose or its teflon inner coating may be damaged. Prevent:
•
sharp hose bends near the ends,
•
twisting, cutting, rubbing, stretching and squeezing the hose,
•
exceeding the maximum solenoid working pressure of 70 bar,
•
if you notice any signs of damage of the components, replace it.
To connect the freezer to a CO2 cylinder you have to:
1) place the CO2 cylinder vertically and attach it to the wall with a special clamp, rope or chain,
2) screw the supplied reduction W21,8 x 1/4 '' SAE on the cylinder connector,
3) screw the hose to the reduction,
4) slowly unscrew the valve while checking low leakage connection,
5)
after connecting the cylinder correctly, press the main menu icon
the button
and then press
(Figure 73); press
(Figure 74) and carry out the solenoid valve opening test.
When working on CO2 back-up system is recommended to lock the freezer’s door with a key.
Figure 73 CO2 Test
76
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Figure 74
When the button is pressed, the CO2 valve opens, accompanied by a loud noise and a drop in temperature.
Connection diagram of a CO2 cylinder with a freezer:
3
4
2
5
CO2
back-up
system
1
6
7
CO2
1)
2)
3)
4)
5)
6)
7)
ZLN-UT
cylinder containing CO2
CO2 cylinder valve
reducer with pressure gauge (option)
CO2 back-up system
pressure reducer of the freezer’s chamber
frezer
steel hose for connecting a CO2 cylinder ended with an internal thread W21,8 according to DIN477
77
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
CO2 back-up system should be connected to a siphon cylinder.
The manufacturer does not provide the equipment with a CO2 cylinder.
Water condensation on the solenoid valve and steel hose is a natural phenomenon (only during operation
of the CO2 back-up system).
Connection to a CO2 cylinder should be made by a person who has undergone appropriate training.
8.5. Disconnecting the freezer from the CO2 cylinder
To disconnect the freezer from the CO2 cylinder:
1) close the cylinder’s valve,
2) press the main menu icon
3)
, and then press
(Figure 73); using the button
perform the solenoid valve opening test three times to empty the gas in the connection,
unscrew the connection and reduction of W21,8 x 1/4 '' SAE from the CO2 cylinder.
(Figure 73)
8.6. Starting the device
To start the device, turn the key of the rotary switch (Figure 72) of the CO2 back-up system to the "ON" position, and
then turn the main switch to the "I" position.
It is not recommended to open the freezer door during operation of the CO2 back-up system because it
causes a sudden increase of the temperature in the chamber of the equipment, and thus an increase in
the amount of CO2 dosed.
During the system operation, frost will accumulate in the device - this is normal. CO2 dosing is done through the
nozzles inside the freezer.
The manufacturer does not recommend placing the samples directly under the CO2 nozzle. Covering the
perforations on the device shelves reduces the efficiency of the system.
78
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
9. CONNECTING THE DEVICE TO A COMPUTER
Each device in the SMART PRO version can be connected to an Ethernet network or directly to a computer with a LAN
cable (standard equipment). Using the Lab Desk program (standard equipment), you can program and monitor the
operation of multiple devices with the SMART PRO controller. The features of the software have been described in a
separate instruction manual.
10. CLEANING AND MAINTENANCE OF THE DEVICE
Disconnect the device from the power supply before carrying out any activities related to the
cleaning! In the case of the battery back-up of the controller, also turn it off.
On the internal walls of the device (in particular the new one) made of stainless steel, discoloration (spots) may appear
- which are not caused by factory defects, but only by the steel production process. They can be cleaned using
extraction gasoline.
INOX products are manufactured with stainless steel. When used in standard laboratory conditions they do not rust.
However it is possible that stains (which may look like rust) form on the steel surface (e.g. due to the kind of samples
that are incubated in the chamber). In such case we recommend using cleaning solution (to clean the stains) which is
dedicated to this particular application, e.g. Pelox.
When cleaning stainless steel product with dedicated cleaning solution, one should pay attention
to the suggestions and recommendations given in the instruction manual or in the safety data sheet of the cleaning solution.
10.1. Exterior cleaning
1.
The housing of the device should be cleaned at least once a week, depending on the working conditions.
2.
The housing and door should be cleaned with caution using a soft cloth dampened with water.
3.
Only mild cleaning products should be used to clean the device.
4.
Electrical parts should not get in contact with water or detergent.
5.
Clean the touch screen using a soft cloth or a foam for cleaning touch screens.
6.
USB port can be cleaned with a vacuum cleaner to prevent accumulation of dirt inside the port.
10.2. Interior cleaning
1.
Before cleaning the interior of the device, empty the chamber.
2.
Open the door of the device, if necessary wait till the chamber has cooled down, take out the shelves and start
cleaning of the device
3.
Only water or water with mild detergent should be used.
4.
Having finished cleaning, you should allow the device to dry fully and instal all parts removed before cleaning.
5.
During cleaning you should make sure not to damage the temperature sensors which are located inside the
79
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
chamber.
At least once a month clean the condenser with a vacuum cleaner, dry cloth or a soft brush. Placement of the
condenser in the freezers it’s at the front in the bottom of the device. To access it, pull the ventilation cover (a)
towards you and then pull it up (b). After cleaning the condenser (1), install the cover.
6.
Failure to clean regularly may result in damage to the compressor and loss of the rights for repair
under warranty.
10.3. Cleaning the touch screen
The touch screen is exposed to dirt, so it must be cleaned regularly. To clean the touch screen, use a clean and dry
microfiber cloth. It is a very delicate material and collects dirt well.
Before using the cloth, make sure that on the surface there are no crumbs or particles. During
cleaning, they can act like sandpaper and scratch the surface of the screen.
If the stains cannot be removed by dry cleaning, the cloth can be lightly dampened with water.
Do not use paper towels to clean the screen as it may cause microdamages.
Before cleaning, lock the screen by pressing
on the top drop-down list (Figure 75).
Figure 75 Locking the screen
80
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
The screen is ready to be cleaned.
To unlock the touch screen, slide the blue circle into the white circle
Figure 76 Unlocking the screen
10.4. Consumables
Consumables during normal operation are:
•
silicone door seal - in all units,
•
chamber fan - in equipment with forced air convection,
•
interior lighting bulb - in units with the option of interior lighting.
11. ADVICE ON HOW TO SAFELY STORE THE DEVICE
1.
Remove all objects from the chamber.
2.
Disconnect the device from the mains. If the unit is equipped with battery back-up of the controller (optional), also
turn it off.
3.
Clean and dry the chamber.
4.
Leave the door open to avoid unpleasant odors.
5.
Store in temperatures between 0°C and 50°C and relative humidity maximum 70%.
12. TROUBLESHOOTING
Before you contact Service Department:
1.
Make sure that the operation complies with the instruction manual of the device.
2.
Restart the device to make sure that the unit is not functioning properly. If it still does not work, disconnect the unit
again from the mains and repeat the operation after one hour. Do the same with optional battery back-up of the
controller.
Service
Visit the POL-EKO-APARATURA website at: www.pol-eko.com.pl in order to:
• get full contact details of technical service
• access to POL-EKO-APARATURA online catalogue, and information about accessories and related products
• receive additional product information and special offers
81
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
To receive information or technical assistance, contact the Service Department or visit the website: www.poleko.com.pl
12.1. Possible defects
Malfunction
The unit is not working
What to check?
Check if the unit is plugged in correctly
What to do?
Plug in the unit correctly
Check if the circuit-breaker has tripped
Press the circuit breaker on the
back of the device
Connect the device to a different
socket, preferably from a different
electrical circuit. Call a licensed
electrician to check the electrical
installation.
Change the cable
Clean the condenser
Change the location of the unit
Change the location of the unit
Clean the gasket
Remove other objects
Check the voltage in the socket
The unit is not cooling
down
The unit is working too
loud
The door has dropped or
is skewed
No battery backup of the
display
Check if the power cable is broken
Check if the condenser is dirty
Check if the unit is exposed to direct sunlight
Check if there is a heat emitter near the device
Check if the door is closed properly
Check if the unit is not touching other objects or
furniture etc.
Check if the door is properly leveled
Check if the door is properly leveled
Is the battery exhausted?
Level the device
Level the device. If this does not
help, contact the service.
Replace the battery (replace the
battery every 12 months)
Gurgling sound of the refrigerant fluid flowing in the refrigerant circuit is normal.
82
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
13. WARRANTY CONDITIONS
POL-EKO-APARATURA warrants that this product will be free from defects in material and workmanship for a period of
two (2) years from date of the invoice. If a defect is present, POL-EKO-APARATURA will, at its option and cost, repair,
replace, or refund the purchase price of this product to the customer, provided it is returned during the warranty period.
This warranty does not apply if the product has been damaged by accident, abuse, misuse, or misapplication, or from
ordinary wear and tear. If the required maintenance and inspection services are not performed according to the manuals and any local regulations, such warranty turns invalid.
The device that is being returned must be secured by the customer in the event of any damage or loss. The warranty
will be only limited to the situations listed above. IT IS EXPRESSLY AGREED THAT THIS WARRANTY WILL BE IN
LIEU OF ALL WARRANTIES OF FITNESS AND IN LIEU OF THE WARRANTY OF MERCHANTABILITY.
All complaints should be reported to the following address:
POL-EKO-APARATURA Sp.j.
ul. Kokoszycka 172 C, 44-300 Wodzisław Śl.
Tel: +48 / 32 453 91 96, 32 453 91 70, 32 453 90 30
E-mail: [email protected]
Support form and warranty conditions are specified on the manufacturer's website:
http://www.pol-eko.com.pl/en/service
Compliance with local laws and regulations
The user is responsible for obtaining any approvals or authorizations required to launch and use the product. POLEKO-APARATURA shall not be liable for any negligence in the above matter except when the refusal to obtain
authorization is caused by a product defect.
83
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
14. RATING PLATE
1
11
2
10
3
9
4
8
5
6
7
Manufacturer’s data
Type of device
Serial number (the two marked digits indicate the year of manufacture of the device)
Temperature protection class according to DIN 12880
Degree of protection against electric shock (class I: protection against indirect contact) and IP enclosure protection
rating
6. Disposal of used device according to WEEE2
7. CE marking as confirmation of compliance with the directives
8. Information about cooling system (gas type and quantity)
9. Temperature range of the device
10. Maximum power consumption, and capacity of device
11. Acceptable range of voltage and frequency of mains supply
1.
2.
3.
4.
5.
84
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
15. TECHNICAL DATA
Technical data are given with a tolerance of ± 5%, the working capacity of the chamber is always smaller. All the below
technical data refers to standard units (without optional accessories).
15.1. ZLN, ZLN-T, ZLW-T devices
Paremeter
Air Convection
Chamber capacity [I]
Working capacity [I]
Door
Temperature range [oC]
Temperature range pracy [oF]
Temperature resolution [oC]
Controller
P Smart
Interior
PS Smart
P Smart
Housing
PS Smart
Overall dims1
A widith
[mm]
B height
C depth
Internal dims
D widith
[mm]
D΄ widith
E height
F depth
F΄ depth
G depth
H height
Max shelf
workload2 [kg]
version
PW3
Max unit workload [kg]
version W4
Nominal power [W]
Weight [kg]
Refrigerant gas
Shelves fitted/max
Warranty
Manufacturer
1.
2.
3.
4.
ZLN 85
ZLN-T 125
85
73
130
109
ZLN –T 200
ZLN- T 300
ZLW- T 200
310
262
210
140
310
213
-40…0
-25,6…32
every 0,1
microprocessor with external touch screen 7’
acid-proof stainless steel according to DIN 1.4301
acid-proof stainless steel according to DIN 1.4301
powder coated sheet
polished stainless steel
720
820
820
820
1190
1380
1730
1380
810
810
810
810
370
450
450
450
420
520
520
520
600
770
1120
770
520
520
520
520
530
530
530
530
550
10
10
10
10
50
50
50
50
820
1730
810
450
520
1120
520
530
900
10
50
natural
ZLW-T 300
forced
210
180
solid
- 25…0
-13…32
610
930
650
380
420
590
400
440
230
380
10
30
200
62
R290 /
GWP=36
2/4
50
100
450
105
R290 /
GWP=3
2/3
65
80
130
160
450
450
120
185
R290 /
R290 /
GWP=3
GWP=3
2/4
3/6
24 months
POL - EKO - APARTURA
65
160
450
120
R290 /
GWP=3
2/4
80
160
450
185
R290 /
GWP=3
3/6
depth does not include 50mm of power cable
on uniformly loaded surface
reinforced shelf
reinforced version
85
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
ZLN 85
ZLW-T 200, ZLW-T 300
ZLN-T 125, ZLN-T 200, ZLN-T 300
86
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
15.2. ZLN, ZLN-UT VIP devices
Parametr
ZLN-UT 130 VIP
ZLN-UT 200 VIP
130
96
Air convection
Chamber capacity [I]
Number of boxes
133x133x50mm[pcs]
Door
Temperature range [0C]
Temperature range [0F]
Temperature resolution [o C]
Controller
P SMART PRO
Interior
Housing
ZLN- UT 500 VIP
259
345
482
192
256
352
natural
solid
-86 …-50
-122,8 …-58
every 0,1
microprocessor with external touch screen 7’
stainless steel according to DIN 1.4301
P SMART PRO
Overall
dims1[mm]
A widith
B height
C depth
Internal dims
D widith
[mm]
E height
F depth
G height
Max unit workload [kg] 2
Max shelf workload [kg]
Nominal power [W]
Weight [kg]
Refrigerant gas
powder coated sheet
850
940
960
620
360
580
45
10
2100
147
R290 / GWP=3 |
R170 / GWP=6
880
1390
960
620
770
580
360
65
10
2100
180
R290 / GWP=3 |
R170 / GWP=6
1|1
2|2
Number of compartments
Number of shelves fitted/max
Warranty
Manufacturer
1.
2.
ZLN- UT 300 VIP
880
1620
960
620
1000
580
480
65
10
2100
220
R290 / GWP=3 |
R170 / GWP=6
880
2000
960
620
1380
580
670
65
10
2100
243
R290 / GWP=3 |
R170 / GWP=6
1
2|2
24 months
POL – EKO – APARATURA
4|4
depth does not include 50mm of power cable
on uniformly loaded surface
ZLN-UT 130 VIP
87
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
ZLN-UT 200 VIP
ZLN-UT 300 VIP
88
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
ZLN-UT 500 VIP
89
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
16. DECLARATIONS OF CONFORMITY
90
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
91
Instruction manual ZLN, ZLN-T, ZLW-T, ZL-UT VIP SMART PRO
Manufacturer of laboratory equipment
and authorized distributor of:
WTW, Thermo Scientific, Knick.
POL-EKO-APARTATURA sp.j.
A. Polok-Kowalska, S. Kowalski
ul. Kokoszycka 172 c, 44-300 Wodzisław Śląski
tel. 32 453 91 70, fax 32 453 91 85
e-mail: [email protected]
web: https://www.pol-eko.com.pl/home-en/ * https://smart4lab.eu/smart-en/
We produce:
❑ thermostatic cabinets
❑ laboratory refrigerators
❑ laboratory incubators
❑ devices with photoperiod and phytotron system
❑ drying ovens and sterilizers
❑ drying ovens with nitrogen blow
❑ laboratory freezers
❑ ultra-low freezers
❑ climatic chambers
❑ Caldera fluid and blanket warmers
❑ colony counters
❑ laboratory shakers
❑ stationary samplers
❑ Hydromat water dispensers
❑ Eurodrop stations
❑ FEKO+ waste water receipt station
❑ heating ovens
❑ cooled incubators
❑ Compact lab designing
❑ fume hoods
We offer portable, laboratory and on-line equipment:
❑ pH-meters
❑ ionmeters
❑ dissolved oxygen meters
❑ conductivity meters
❑ photometers and spectrophotometers
❑ thermo reactors
❑ turbidity metres
❑ pH electrodes
❑ conductivity sensors
❑ oxygen probes
❑ heavy metals trace analyzers
❑ water baths
❑ autoclaves
❑
pH buffer solutions
❑ conductivity standards
❑ photometric tests
❑ laboratory accessories
❑ consumables
We organize:
❑ regional trainings
❑ individual trainings
❑ seminars
We provide:
❑ warranty and post-warranty service
❑ consultancy in the selection, maintenance and
operation of laboratory equipment
POL-EKO LAB is Accredited by the Polish Centre for
Accreditation (a member of ILAC) and provides accredited calibration of:
❑
❑
❑
❑
❑
❑
❑
❑
thermostatic and climatic chambers (incubators, drying ovens,
thermostatic cabinets, climatic chambers, freezers)
water baths and thermo reactors
autoclaves
electric and electronic thermometers
data loggers
high temperature laboratory furnaces
thermohygrometers
laboratory sieves
Calibration is confirmed with the issue of 'Calibration Certificate'.
Services outside the scope of accreditation:
❑
❑
❑
checking equipment for physicochemical measurements (meters and probes),
carrying out IQ, OQ, PQ qualification procedures,
mapping of temperature and humidity in the rooms
92