LumaSense IMPAC IPE 140/34 Manual
The LumaSense IMPAC IPE 140/34 is a digital pyrometer designed for non-contact temperature measurement of plastic films, particularly thin polyethylene or polypropylene films with a minimum thickness of 30 µm. It features a 3.43 µm spectral range, allowing for accurate measurements through plastic materials. The instrument offers adjustable settings like emissivity, exposure time, and analog output range. It is also equipped with a digital interface for data communication and remote parameter control.
PDF
Download
Document
Advertisement
Advertisement
MANUAL IMPAC Pyrometers IPE 140 • IPE 140/34 • IPE 140/39 • IPE 140/45 Confidential Information The material contained herein consists of information that is the property of LumaSense Technologies and intended solely for use by the purchaser of the equipment described in this manual. All specifications are subject to change without notice. Changes are made periodically to the information in this publication, and these changes will be incorporated in new editions. LumaSense Technologies prohibits the duplication of any portion of this manual or the use thereof for any purpose other than the operation or maintenance of the equipment described in this manual, without the express written permission of LumaSense Technologies. Copyright © LumaSense Technologies 2013. All rights reserved. Trademarks IMPAC is a trademark of LumaSense Technologies. All other trademarks are trademarks, registered trademarks, and/or service marks of their respective holders. Service Centers LumaSense Technologies GmbH Other Than North America Sales & Support Frankfurt, Germany Ph: +49 (0) 69 97373 0 Fax: +49 (0) 69 97373 167 LumaSense Technologies, Inc. North America Sales & Service Santa Clara, CA, USA Ph: +1 800 631 0176 Ph: +1 408 727 1600 Fax: +1 408 727 1677 Global and Regional Centers Our Headquarters LumaSense Technologies, Inc. Santa Clara, CA, USA Ph: +1 800 631 0176 Fax: +1 408 727 1677 Americas, Australia, & Other Asia LumaSense Technologies, Inc. Santa Clara, CA, USA Ph: +1 800 631 0176 Fax: +1 408 727 1677 Europe, Middle East, Africa LumaSense Technologies GmbH Frankfurt, Germany Ph: +49 (0) 69 97373 0 Fax: +49 (0) 69 97373 167 Brazil LumaSense, Vendas Brasil Campinas, Brasil Ph: +55 19 3367 6533 Fax: +55 19 3367 6533 India LumaSense Technologies, India Mumbai, India Ph: + 91 22 67419203 Fax: + 91 22 67419201 China LumaSense Technologies, China Shanghai, China Ph: +86 133 1182 7766 Fax: +86 21 5877 2383 E-mail [email protected] [email protected] [email protected] [email protected] Website http://www.lumasenseinc.com Part No 3 875 267- EN Revision B March 2013 Contents 1 General Information ......................................................................................................... 5 1.1 1.2 1.3 1.4 1.5 1.6 1.7 2 Introduction ....................................................................................................................... 9 2.1 2.2 2.3 2.4 2.5 2.6 2.7 3 Appropriate use ........................................................................................................ 9 Scope of delivery ...................................................................................................... 9 Technical Data ........................................................................................................... 9 Dimensions .............................................................................................................. 11 Physical User Interface ........................................................................................... 11 Accessories (Optional) ............................................................................................ 12 Transport, Packing, Storage .................................................................................. 13 Controls and Installation................................................................................................ 15 3.1 3.2 3.3 4 Information about the user manual ....................................................................... 5 1.1.1 Legend ......................................................................................................................5 1.1.2 Terminology .............................................................................................................5 Safety ......................................................................................................................... 5 1.2.1 Laser Targeting Light ............................................................................................... 5 1.2.2 Electrical Connection ...............................................................................................6 Limit of liability and warranty ................................................................................ 6 Unpacking the Instrument ....................................................................................... 6 Service Request, Repair, or Support ....................................................................... 7 Shipments to LumaSense for Repair ...................................................................... 7 Disposal / decommissioning .................................................................................... 7 Electrical Installation.............................................................................................. 15 3.1.1 Pin assignment of the connector ..........................................................................16 3.1.2 Connecting the pyrometer to a PC .......................................................................16 3.1.3 Connection to RS232 interface .............................................................................17 3.1.4 Connection to RS485 interface .............................................................................17 3.1.5 Connection of additional analyzing devices ........................................................17 Sighting ................................................................................................................... 18 3.2.1 Thru-lens view finder .............................................................................................18 3.2.2 Laser targeting light ..............................................................................................18 Optics ....................................................................................................................... 18 3.3.1 Spot Sizes ................................................................................................................18 3.3.2 Deviation from the focused measuring distance.................................................19 3.3.3 Adjusting the required measuring distance ........................................................19 Instrument settings ........................................................................................................ 21 4.1 4.2 4.3 Settings at the instrument .................................................................................... 21 Key panel operation ............................................................................................... 21 Selection of the serial interface ............................................................................ 22 Series IPE 140 Manual Contents iii 4.4 4.5 5 Settings / parameter descriptions ................................................................................. 23 5.1 5.2 5.3 5.4 5.5 5.6 5.7 5.8 5.9 5.10 5.11 5.12 6 Test function ........................................................................................................... 22 Factory settings ...................................................................................................... 22 Emissivity (Emi) ...................................................................................................... 23 Compensation of ambient temperature (tamb) .................................................... 24 Exposure time (t90) .................................................................................................. 24 Clear time of the maximum value storage (tClear) ............................................... 25 Analog output (mA) ............................................................................................... 26 Subrange (from / to) ............................................................................................... 26 Address (Adr) .......................................................................................................... 26 Baud rate (Baud) ..................................................................................................... 26 Temperature display (C / F) .................................................................................... 26 Wait time (tw) .......................................................................................................... 26 Maximum internal temperature (MaxIntTemp) .................................................. 26 Error status (Status) ............................................................................................... 26 Software InfraWin........................................................................................................... 27 6.1 6.2 6.3 6.4 Connecting the pyrometer to a PC ....................................................................... 27 Installation .............................................................................................................. 27 Program start .......................................................................................................... 27 Basic settings .......................................................................................................... 28 6.4.1 Open/Save...............................................................................................................28 6.4.2 1 measur….............................................................................................................. 28 6.4.3 Print ........................................................................................................................28 6.4.4 Close........................................................................................................................28 6.4.5 Test.......................................................................................................................... 28 6.5 Measurement online trend .................................................................................... 29 6.6 Listing (analyzing) .................................................................................................. 30 6.7 Output .TXT file (analyzing) ................................................................................... 30 6.8 Trend output (analyzing) ....................................................................................... 30 6.9 PC sampling rate (time interval between two measurements) ........................ 31 6.10 Spot size calculator ................................................................................................. 31 6.11 Search I/O Module ................................................................................................... 31 7 Maintenance .................................................................................................................... 33 7.1 7.2 Cleaning the window ............................................................................................. 33 Optics replacement ................................................................................................. 33 8 Data format UPP (Universal Pyrometer Protocol) ...................................................... 35 9 Reference Numbers ......................................................................................................... 39 9.1 9.2 Reference numbers instrument ............................................................................ 39 Reference numbers accessories ............................................................................ 39 10 Troubleshooting .............................................................................................................. 41 Index ....................................................................................................................................... 43 Series IPE 140 Manual Contents iv 1 General Information 1.1 Information about the user manual Congratulations on choosing this high quality and highly efficient IMPAC pyrometer. Please read this manual carefully, step by step, including all notes to security, operation, and maintenance before installing the pyrometer. For installation and operation of the instrument this manual is an important source of information and work of reference. To avoid handling errors, keep this manual in a location that you always have access to. When operating the instrument, it is necessary to follow general safety instructions (see Section 1.2 Safety). 1.1.1 Legend Note: The note symbol indicates tips and useful information in this manual. All notes should be read to effectively operate the instrument. Attention: This sign indicates special information which is necessary for a correct temperature measurement. Warnings and Cautions: The general warnings and cautions symbol signifies the potential for bodily harm or damage to equipment. MB 1.1.2 Shortcut for Temperature range (in German: Messbereich) Terminology The terminology used in this manual corresponds to the VDI- / VDE-directives 3511, Part 4. 1.2 Safety This manual provides important information on safely installing and operating an IMPAC Series 140 pyrometer. Several sections of this manual provide safety warnings to avert danger. These safety warnings are specified with a warning symbol. You must read and understand the contents of this manual before operating the instrument even if you have used similar instruments or have already been trained by the manufacturer. It is also important to always pay attention to all labels and markings on the instrument and to keep the labels and markings in a permanent readable condition. Warning: The pyrometer is only to be used as described in this manual. It is recommended that you only use accessories provided by the manufacturer. In addition, signs and markings on the device are to be observed and maintained in legible condition. 1.2.1 Laser Targeting Light For easy alignment to the measuring object, the pyrometer can be equipped with a laser targeting light. This is a visible red light has a wavelength between 630 and 680 nm and a maximum power of 1 mW. The laser is classified as product of laser class II. Series IPE 140 Manual General Information 5 Warning: To reduce the risk of injury to the eyes, do not look directly into the targeting laser and do not point the targeting laser into anyone's eyes. The instrument is equipped with a class II laser that emits radiation (IEC 60825-1-3-4). Safety regulations: Never look directly into the laser beam. The beam and spot can be watched safely from side. Make sure that the beam will not be reflected into eyes of people by mirrors or shiny surfaces. 1.2.2 Electrical Connection When connecting additional instruments to which the mains voltage is applied (for example transformers), general safety regulations for connecting to the 230 V power supply must be observed. Contact with mains voltage may have lethal consequences. Improper installation may cause extremely severe injuries, health problems, or material damage. Only qualified personnel are permitted to connect power supply units of this type. 1.3 Limit of liability and warranty All general information and notes for handling, maintenance and cleaning of this instrument are offered according to the best of our knowledge and experience. LumaSense Technologies is not liable for any damages that arise from the use of any examples or processes mentioned in this manual or in case the content of this document should be incomplete or incorrect. LumaSense Technologies reserves the right to revise this document and to make changes from time to time in the content hereof without obligation to notify any person or persons of such revisions or changes. All Series 140 instruments from LumaSense Technologies have a regionally effective warranty period. This warranty covers manufacturing defects and faults which arise during operation, only if they are the result of defects caused by LumaSense Technologies. The warranty is VOID if the instrument is disassembled, tampered with, altered, or otherwise damaged without prior written consent from LumaSense Technologies; or if considered by LumaSense Technologies to be abused or used in abnormal conditions. The Windows compatible software was thoroughly tested on a wide range of Windows operating systems and in several world languages. Nevertheless, there is always a possibility that a Windows or PC configuration or some other unforeseen condition exists that would cause the software not to run smoothly. The manufacturer assumes no responsibility or liability and will not guarantee the performance of the software. Liability regarding any direct or indirect damage caused by this software is excluded. 1.4 Unpacking the Instrument Thoroughly inspect the instrument upon receipt. Check all materials in the container against the enclosed packing list. LumaSense Technologies cannot be responsible for shortages against the packing list unless a claim is immediately filed with the carrier. Final claim and negotiations with the carrier must be completed by the customer. Save all packing materials, including the carrier’s identification codes, until you have inspected the pyrometer and find that there is no obvious or hidden damage. Before shipment, the pyrometer was examined and has been tested. If you note any damage or suspect damage, immediately contact the carrier and LumaSense Technologies, Inc. Series IPE 140 Manual General Information 6 1.5 Service Request, Repair, or Support Contact LumaSense Technologies Technical Support in case of a malfunction or service request. Provide clearly stated details of the problem as well as the instrument model number and serial number. Upon receipt of this information, Technical Support will attempt to locate the fault and, if possible, solve the problem over the telephone. If Technical Support concludes that the instrument must be returned to LumaSense Technologies for repair, you will receive a Return Material Authorization (RMA) number. Return the instrument upon receipt of the RMA number, transportation prepaid. Clearly indicate the assigned RMA number on the shipping package exterior. Refer to Section 1.6, Shipments to LumaSense for Repair, for shipping instructions. Technical Support can be contacted by telephone or email: Santa Clara, California Telephone: +1 408 727 1600 or +1 800 631 0176 Email: [email protected] Frankfurt, Germany Telephone: +49 (0) 69 97373 0 Email: [email protected] Erstein, France Telephone +33 (0)3 88 98 98 01 Email [email protected] 1.6 Shipments to LumaSense for Repair All RMA shipments of LumaSense Technologies instruments are to be prepaid and insured by way of United Parcel Service (UPS) or preferred choice. For overseas customers, ship units airfreight, priority one. The instrument must be shipped in the original packing container or its equivalent. LumaSense Technologies is not responsible for freight damage to instruments that are improperly packed. Contact us to obtain an RMA number (if one has not already been assigned by Technical Support). Clearly indicate the assigned RMA number on the shipping package exterior. Send RMA Shipments to your nearest technical service center: Santa Clara, California Frankfurt, Germany LumaSense Technologies, Inc. 3301 Leonard Court Santa Clara, CA 95054 USA Telephone: +1 408 727 1600 +1 800 631 0176 LumaSense Technologies GmbH Kleyerstr. 90 60326 Frankfurt Germany Telephone: +49 (0)69-97373 0 Email: [email protected] Email: [email protected] 1.7 Disposal / decommissioning Inoperable IMPAC pyrometers must be disposed of in compliance with local regulations for electro or electronic material. Series IPE 140 Manual General Information 7 To ensure consistent document formatting, this page was intentionally left blank. Series IPE 140 Manual General Information 8 2 Introduction 2.1 Appropriate use Series 140 pyrometers are digital pyrometers for non-contact temperature measurement: The IPE 140 is used for measurements on metals, graphite, or ceramics at low temperatures or if extreme long temperature ranges are required. The IPE 140/39 is used for non-contact temperature measurement of metal parts in flame heated furnaces. The narrow spectral range of 3.9 µm avoids the influence of humidity and CO2 and enables a correct measurement through flames and combustion gases. Also humidity and CO2 do not have any influence on measurements with long measuring distances. Another application is the measurement of glass if a small penetration into the glass is necessary (e.g. glass drop). Measurement errors caused by partially cooled down surfaces can be avoided. The IPE 140/34 is used for non-contact temperature measurement of plastic films; especially extremely thin Polyethylene or Polypropylene films, with a minimum material thickness of only 30 µm. The IPE 140/45 is used for the temperature measurement of CO2 containing gases above approx. 500 °C. As a gas is not a solid object, a certain penetration path into the gas is necessary for an exact temperature measurement. The path length is dependent on the temperature, the concentration of CO2 and the pressure. For example: on atmosphere condition, approximately 4% CO2, a penetration depth of 2-3 m is necessary, for concentrations of min 10% CO2, 0.5 m is sufficient. 2.2 Scope of delivery Device with thru-lens sighting or laser targeting light, selectable optics, PC software “InfraWin“, Hex key 3 mm, and operation manual. Note: A connection cable is not included with the instrument and has to be ordered separately (see Chapter 9, Reference numbers). 2.3 Technical Data Temperature ranges Sub Range IPE 140 5 to 350 °C (MB 3.5) 5 to 500 °C (MB 5) 30 to 1000 °C (MB 10) 50 to 1200 °C (MB 12) 50 to 400 °C (MB 4) IPE 140/34 75 to 500 °C (MB 5) IPE 140/39 20 ... 700 °C (MB 7) 75 ... 1200 °C (MB 12) 300 ... 1450 °C (MB 14.5S) 500 … 1800 °C (MB 18) 400 ... 2000 °C (MB 20) IPE 140/45 Any range adjustable within the temperature range, minimum span 51 °C Signal processing photoelectric current, digitized Spectral Range IPE 140 Series IPE 140 Manual 3 ... 5 µm IPE 140/39 3.9 µm Introduction 9 IPE 140/34 3.43 µm IPE 140/45 CO2 hot band PbSe 24 V AC or DC (14 to 30 V AC or DC) (AC: 48 to 62 Hz) Max. 6 W 0 to 20 mA or 4 to 20 mA (linear), switchable; Test current 10 mA or 12 mA by pressing test key Load 0 to 500 Digital Interface RS232 or RS485 addressable (half duplex), switchable; Baud rate 2400 up to 115200 Bd Resolution 0.1 °C on interface and display; < 0.03% of temperature range at the analog output Isolation Power supply, analog output, and digital interface are galvanically isolated from each other Operation Indication Green LED LC display Illuminated LC display for temperature indication or parameter settings Parameter Adjustable at the device or via interface: Emissivity , exposure time t90, 0 to 20 or 4 to 20 mA switch for analog output, sub range, clear times for maximum value storage, automatically or external deletion of maximum value storage, address, baud rate, wait time tW Readable at the device or via interface: Measuring temperature, internal instrument temperature. Emissivity 10 to 100% adjustable in the instrument or via interface in steps of 0.1% Exposure time t90 1.5 ms (with dynamical adaptation at low signal levels); adjustable at 0.01 s; 0.05 s; 0.25 s; 1 s; 3 s; 10 s Maximum Value Storage Built-in single or double storage. Clearing with adjusted time tclear (off; 10 ms; 50 ms; 250 ms; 1 s; 5 s; 25 s), via interface, automatically with the next measuring object, hold-function Switch contact max. 0.15 A (only active with automatic clear mode or tclear 0.25 s) Measurement Uncertainty: IP140, MB 3.5: up to 250 °C: 1 °C (with =1, t90=1 s, Tamb.=10...40 °C) above 250 °C: 0.4% of measured value in °C + 1 °C All other types: up to 400 °C: 2 °C above 400 °C: 0.4% of measured value in °C + 1 °C above 1200 °C: 0.6% of measured value in °C + 1 °C Repeatability IPE 140, MB 3.5: 0.1% of measured value in °C + 0.2 °C ( = 1, t90 = 1 s, TUmg. = 40 °C) All other types: 0.1% of measured value in °C + 1 °C Ambient Temperature 0 to 53 °C Storage Temperature -20 to 60 °C Protection Class IP65 (DIN 40050) Weight approximately 550 g CE-Label According to EU directives about electromagnetic immunity Laser targeting light (max. power level < 1 mW, Sighting = 630-680 nm, CDRH class II) or built-in optimized thru-lens view finder IR Detector Power Supply Power Consumption Analog Output Note: The calibration/adjustment of the instruments was carried out in accordance with VDI/VDE directive "Temperature measurement in industry, Radiation thermometry, Calibration of radiation thermometers", VDI/VDE 3511, Part 4.4. The directive can be ordered from: "Beuth Verlag GmbH" in D-10772 Berlin, Germany. Series IPE 140 Manual Introduction 10 2.4 Dimensions Pyrometer with thru-lens view finder Pyrometer with laser targeting light All dimensions in mm 2.5 Physical User Interface 7 8 1 9 2 10 3 4 1 2 3 4 5 6 7 5 Focusable optics (with label with optical data) 4 threads for fixing the pyrometer or accessory parts Mounting rail Type label LC display Extendable back cover Screws for rear cover (3 mm hex screws) 11 6 12 8 Laser targeting light on/off switch or parallax free view finder, dependent on the instrument’s type) 9 Operating status for targeting light / laser (only instruments with laser targeting light) 10 Male socket for electrical connections 11 LC display, extended 12 Setting keys Attention: The radiation temperature in the spot area of the pyrometer is not allowed to be higher than 150 °C above the end of the basic range because the temperature sensor can be damaged. Series IPE 140 Manual Introduction 11 2.6 Accessories (Optional) Numerous accessories guarantee easy installation of the pyrometer. The following overview shows a selection of suitable accessories. You can find the entire accessory program with all reference numbers in Chapter 9, Reference numbers. Mounting For mounting and aligning the pyrometer to the measured object a mounting angle or a ball and socket mounting is available. The ball and socket mounting is an easy way to align the pyrometer to the measured object. The quick-clamping-screws of the ball and socket mounting enable an easy and fast adjustment of the pyrometer in all directions. Ball and socket mounting Mounting angle Cooling The pyrometer can be used in ambient temperatures outside of the specifications if preventive maintenance is taken. The cooling plate is used to protect the pyrometer from heat coming from the front. The completely covered water cooling jacket made from stainless steel protects the pyrometer if exposed to a hot environment. It is designed for ambient temperatures up to 180 °C. Displays In addition to the built-in temperature indicator of the pyrometer, LumaSense offers several digital displays which can also be used for remote parametrizing of the pyrometer. Cooling plate Water cooling jacket Digital display DA 6000 LED large display Miscellaneous The air purge protects the lens from contamination with dust and moisture. It has to be supplied with dry and oil-free pressurized air and generates an air stream shaped like a cone. The scanning attachment SCA 140 moves the measuring beam of the pyrometer from 0 to 12°. This angle is adjustable to smaller values. The scanning frequency is as well adjustable from 1 to 5 Hz. In most cases the scanning attachment SCA 140 is used as a peak picker for measuring smaller objects like thin wires which may be moving. The 90°-mirror enables the capture of objects at an angle of 90° to the pyrometer axis. Air purge Scanning attachment SCA 140 90° mirror Example for a typical furnace mounting assembly: The flange system with air purge, window slide (easy and fast cleaning of sighting window) and cooling jacket is used for the mounting onto the wall of a combustion chamber. Flange system Series IPE 140 Manual Introduction 12 2.7 Transport, Packing, Storage With faulty shipping, the instrument can be damaged or destroyed. To transport or store the instrument, please use the original box or a box padded with sufficient shock-absorbing material. For storage in humid areas or shipment overseas, the device should be placed in welded foil (ideally along with silica gel) to protect it from humidity. The pyrometer is designed for a storage temperature of -20 to 60 °C with non-condensing conditions. Storing the instrument out of these conditions can cause damage or result in malfunction of the pyrometer. Series IPE 140 Manual Introduction 13 To ensure consistent document formatting, this page was intentionally left blank. Series IPE 140 Manual Introduction 14 3 Controls and Installation 3.1 Electrical Installation The series IPE 140 is powered by a voltage of 24 V (14 to 30 V) DC or AC (48 to 62 Hz). The instrument needs some time to warm up. This time depends on the ambient temperature and connected power supply. During this time, the display indicates “wait”. The analog output is 0 mA. After the warm up time, the display shows the measuring temperature. To switch off the instrument, interrupt the power supply or unplug the electrical connector. Green control light Internal Firmware version for approx. 1 s “wait“ indication during warm-up period After connecting the power supply, the display shows the internal firmware version for approximately 1 s. The green control light on the rear cover is swiched on. The control light blinks as long as the laser targeting light is switched on (only for instruments equipped with a targeting light). To meet electromagnetic requirements (EMV), a shielded connecting cable must be used. The shield of the connecting cable has to be connected only on the pyrometer's side. One side of the power supply (switch board) the shield must be open to avoid ground loops. LumaSense offers connecting cables, which are not part of standard scope of delivery.The connecting cable has wires for the power supply, interface, analog output, external laser switch, and external clear of maximum value storage via contact (see section 9, Reference numbers) and 12-pin connector. The cable includes a short RS232 adapter cable with a 9-pin SUB-D connector for direct PC communication. This adapter is not used in combination with the RS485 interface. Note: At object temperatures below 150 °C and emissivity smaller than 90%, the pyrometer needs another half an hour for warm-up until the final accuracy is reached. Otherwise a sufficient accuracy is reached after approximately 10 min. Series IPE 140 Manual Controls and Installation 15 3.1.1 Pin assignment of the connector Pin K A L B H Male socket E F G M D C L B H J A K Pin-assignment (side of male inserts) Color white brown green yellow gray J pink G F C D E red black violet gray/pink red/blue M orange Function + 24 V power supply (or 24 V AC) 0 V power supply + Ioutp. analog output – Ioutp. analog output external switch for targeting light (bridge to K) see below, operating mode contact or external clearing of maximum value storage or hold function DGND (Ground for interface) RxD (RS232) or B1 (RS485) TxD (RS232) or A1 (RS485) RxD (RS232) or B2 (RS485) (bridge to F) TxD (RS232) or A2 (RS485) (bridge to C) Screen only for cable extension, don’t connect at the switchboard Connector pin J The connector pin J can be used for 3 different functions: 1) Operating mode contact: During the warm-up of the pyrometer (after connection to the power supply, the LED display on the converter indicates “wait”), the pin J is connected to the power supply voltage. This voltage is connected by a relays switch (max. 0.2 A at 50 V) to pin K (power supply voltage). This relays switch opens when the warm-up is finished and the pyrometer is in operating mode. For that reason this pin can be used as operating mode contact. 2) External clearing of the maximum value storage: When the pyrometer is in operating mode, pin J can be used for external clearing of maximum value storage (see section 5.4). To clear the maximum value storage, connect pin J for a short time to pin K (power supply voltage). The function “external clearing” is activated with the following conditions: Warm-up is finished (no “wait“ indication on the display) The clear time is set to “extern“ (see section 5.4) 3) Hold function: when the hold function mode is activated the current temperature reading is frozen as long as J and pin K are connected (see section 5.4 clear time for the maximum value storage). 3.1.2 Connecting the pyrometer to a PC The pyrometers are equipped with a serial interface RS232 or RS485 (switchable at the pyrometer). Standard on a PC is the RS232 interface. Only one pyrometer can be connected if the interface is set to RS232. Only short distances can be transmitted with RS232 and electromagnetic interferences can affect the transmission. The RS485 transmission is relatively free of problems, allows long transmission distances, and enables you to connect several pyrometers to a bus system. If RS485 is not available at the PC, it can be realized with an external converter which converts the RS485 to RS232 for a standard connection to a PC. When using a converter RS485 RS232, ensure that the converter is fast enough to receive the pyrometer’s answer to an instruction of the master. Most of the commonly used converters are too slow for fast measuring equipment. It is recommended to use the LumaSense converter -7520 (order no. 3 852 430). With a slow Series IPE 140 Manual Controls and Installation 16 RS485 connection, it is also possible to set a wait time at the pyrometer to delay the response of a command to the pyrometer (see 5.10 Wait time tw). Connection to RS232 interface The transmission rate (in baud) of the serial interface is dependent on the length of the cable. Values between 2400 and 115200 Bd may be set. TxD (violet) RxD (black) DGND (red) E The baud rate has to be reduced by 50% when the transmission distance is doubled (see also 5.8 Baud rate). M H L B 1 J A 2 6 K 3 7 4 8 Connection to RS485 interface Terminator 120 Ohm Master A B C The transmission rate of the serial interface in Baud (Bd) is dependent on the length of the cable. Values between 2400 and 115200 Bd may be set. L B M C K L B Pyrometer 1 e.g. address 00 DGND F G M D H C K Pyrometer 2 e.g. address 01 B1 E J A B2 G A1 B1 F D H J A DGND E M D A2 G B2 F A1 B1 E DGND S B2 Half-duplex mode A1 and A2, as well as B1 and B2, are bridged in the 12-pin round connector of the connecting cable to prevent reflections due to long stubs. It also safeguards against the interruption of the RS485 bus system should a connecting plug be pulled out. The master labels mark the connections on the RS485 converter. 5 9 PC’s side (soldering side female cable connec Pyrometer’s side (soldering side female cable connector) A1 3.1.4 cable G A2 Typical cable length for RS232 at 19200 Bd is 7 m. F D C A2 3.1.3 L B J A Connection of additional analyzing devices Additional analyzing instruments, such as an LED digital display instrument, only need to be connected to a power supply and the analog outputs from the pyrometer. Instruments, like a controller or printer, can be connected to the display in series as shown above (total load of resistance max. 500 Ohm). white brown 230V ~ 24 V DC Power supply green °C LED digital display Controller Writer yellow Series IPE 140 Manual K Pyrometer 32 e.g. address 31 The baud rate is reduced by 50% when the transmission distance is doubled (see Section 5.8 Baud rate). Typical cable length for 19200 Bd is 2 km. 3.1.5 H Controls and Installation 17 3.2 3.2.1 Sighting Thru-lens view finder The view finder can be used to align the measured object through direct observation. The view finder is true-sided and parallax-free; a circle marks the position of the measuring area, but not the exact size. 3.2.2 Thru-lens view finder Laser targeting light The laser targeting light is a red laser beam used to align the pyrometer at a target. The laser marks the center of the measuring spot. The laser targeting light can be used during operation without affecting the measurement. When the laser targeting light is switched on, the green control light on the rear cover will blink and the display will show “PILT”. It can be switched off by pushing the targeting light button again. After 2 minutes, the targeting light switches off automatically. Laser targeting light push button (on/off) PILT display Blinking control light Warning: To reduce the risk of injury to the eyes, do not look directly into the targeting laser and do not point the targeting laser into anyone's eyes. The instrument is equipped with a class II laser that emits radiation (IEC 60825-1-3-4). 3.3 3.3.1 Optics Spot Sizes The instrument will be delivered with one of the following focusable optics. This allows the adjustment to the needed measuring distance between its smallest and largest value to offer the smallest possible spot sizes. Optics 3-PE Optics Optics Optics 2-PE 1-PE 0-PE Measuring distance a [mm] a = 71 mm a = 77 mm a = 90 mm a = 105 mm a = 120 mm a = 150 mm a = 200 mm a = 260 mm a = 440 mm a = 345 mm a = 580 mm a = 1000 mm a = 4300 mm Aperture D [mm] Series IPE 140 Manual IPE 140 Spot size M90 [mm] MB 3.5 2.4 3 3.6 3.6 4.4 6.2 6.3 8.6 17.1 10.2 34.5 48.5 158 MB 5 MB 10 1.6 2 2.4 2.4 2.9 4.1 4.2 5.7 11.4 6.8 12.6 23 105 14 … 17 1.1 1.3 1.7 1.8 2.4 4.6 2.9 5.1 9.2 42 *) MB 12 0.9 1.0 1.4 1.4 1.8 3.5 2.3 4 7.1 32 IPE 140/39 Spot size M90 [mm] Objective length MB 14.5S MB 7 MB 12 S [mm] MB 18 26 13 0 2.4 0.9 0.7 26 2.9 1.0 0.8 13 4.1 1.4 1.0 0 4.2 1.4 1.1 26 5.7 1.8 1.3 13 11.4 3.5 2.5 0 6.8 2.3 1.7 26 12.6 4 2.8 13 23 7.1 5 6.2 105 32 22 0 14 … 17 *) Controls and Installation 18 IPE 140/34 Spot size M90 [mm] Optics 3-PE Optics Optics 2-PE 1-PE Measuring distance a [mm] a = 100 mm a = 114 mm a = 142 mm a = 185 mm a = 235 mm a = 390 mm a = 305 mm a = 480 mm a = 700 mm a = 1900 mm Aperture D [mm] MB 4 MB 5 3.3 2.1 4.2 2.7 5.9 3.8 6 3.9 8 5.2 15.2 9.8 9.6 6.2 16.4 10.6 26.4 17 73 47 14 … 17 *) Objective length S [mm] 26 13 0 26 13 0 26 13 6.2 0 IPE 140/45 Spot size M90 [mm] MB 20 Measuring distance a [mm] a = 115 mm a = 135 mm a = 170 mm a = 210 mm a = 280 mm a = 500 mm a = 360 mm a = 625 mm a = 2000 mm a = 10000 mm 14 … 17 *) Objective length S [mm] 1.1 1.3 1.6 1.8 2.6 4.9 3 5,6 20 100 26 13 0 26 13 0 26 13 4 0 *) The aperture is the effective lens diameter. It is depending on the objective length. The biggest aperture value belongs to the fully extended objective (S = 26), the smallest aperture value for objective length S = 0. Intermediate values have to be interpolated. The spot sizes, mentioned in the table above, will be only achieved at the adjusted distance. Decreasing or increasing the measuring distance enlarges the spot size. Note: The InfraWin program includes a calculator (Spot Size calculator“) that roughly estimates the unknown values. Note: The measuring object has to be at least as big as the spot size of the pyrometer. 3.3.2 Deviation from the focused measuring distance Spot sizes for non-focused distances (shorter or longer than the focused distance) may be calculated by using the formula below. a2 a1 a Ø M2 Aperture D M2 Ø M1 ØM a2 (M D) D a M1 a1 (M D) D a Formula for Calculating Spot Sizes The InfraWin software also includes a Spot Size Calculator that calculates the data for the non-focused regions if you enter the values of aperture D, focused measurement distance a, and focused measuring field diameter M as found in the above table (see section 3.3.1). 3.3.3 Adjusting the required measuring distance The required measuring distance must be adjusted to achieve the spot sizes mentioned in the tables above. This can be done between the smallest and the biggest limit value. An exception is model IPE 140/45 as there is no solid object for alignment. In this case the optimal optical adjustment can only be determined by an experiment. If the pyrometer has to Series IPE 140 Manual Controls and Installation 19 measure through a sighting tube, the optics has to be adjusted that the field of view (FOV) does not touch the wall. For releasing the optics has to be turned anticlockwise. Then it can be pushed or pulled to find the correct measuring distance. For fixing the optics has to be turned clockwise. fix Adjusting the measuring distance with help of the table: The table mentions the minimum and maximum measuring distance for each optics. This corresponds to the longest or the shortest objective length. As an example a further value between max. and min. is shown. The objective length “S” can be measured with a caliper. push / pull release Objective length S The following methods can be used for other measuring distances: Adjusting the measuring distance with help of the thru-lens view finder The focusable optics is correctly adjusted to the required distance, if the measuring object is shown as a sharp image in the view finder. A circle marks the position of the measuring spot. Adjusting the measuring distance with help of the laser targeting light On the focused measuring distance the laser has its smallest spot size and is illustrated exactly. Series IPE 140 Manual Controls and Installation 20 4 Instrument settings The series IPE 140 pyrometers are equipped with a wide range of settings for optimal adaption to the required measuring condition and for getting the correct measuring temperature (description of all available parameters see section 5, Parameter descriptions / settings). The digital PC interface allows you to exchange data with a PC either by using the supplied InfraWin software or by using the Universal Pyrometer Protocol (UPP) commands with your own communication program (see Chapter 8 for the UPP Data Format commands). 4.1 Settings at the instrument Screw connection 3 mm hex screw The LC-display, as well as the push buttons for displaying and setting of the parameters, is found inside the unit. The pyrometer is opened by loosening four hex screws. If unscrewed, the rear cover can be pulled out along with the attached display and push buttons. The pullout is limited by the lengths of the screws. Illuminated Display Setting keys The backlight of the display is always powered in either status, opened or closed pyrometer. Max. pull-out Rear cover Note: Please make sure that the pyrometer is not contaminated while open. 4.2 1 2 3 Key panel operation PAR: With the PAR button, all available Temperature disparameters are displayed in the play or parameter following description (Chapter 5). short form Pushing the button again changes Push button the display to the next parameter short description and the display shows a corresponding short form (see Chapter 5, in brackets behind the 1 2 3 parameter names). : With the arrow keys and all parameter settings can be displayed. Pushing the button longer changes the settings in fast mode. ESC / ENT: Pushing the ESC button changes the pyrometer to measuring mode. If a parameter is changed with the arrow keys the indication of the ESC button changes to ENT. Pressing the button again confirms the value into the pyrometer. Changing the parameters again by pushing the PAR button doesn’t confirm this value in the pyrometer. If no button is pressed for 30 s, the pyrometer changes to the temperature indication without accepting the changed value. Series IPE 140 Manual Instrument settings 21 4.3 Selection of the serial interface Opposite to the display there is a switch to select a serial interface RS232 or RS485. The LC display shows as chosen either RS232 or RS485. Interface switch Display of the selected interface Test button 4.4 Test function 4.5 Factory settings Test function active The diagnostic push button test generates a current (here the analog output on the analog output which is used to check if a is set to 0 … 20 mA) connected external indicator shows the correct temperature value. The test current output is centered Display which also has to the chosen analog output span, consequently to be on an external indication instrument 10 mA is supplied if the analog output is adjusted to 0 (here temp. range to 20 mA and 12 mA is supplied if the analog output 75 ... 550°C) span is set from 4 to 20 mA. The LC display indicates the respective current along with the corresponding temperature. For example if a measuring range of 75 °C to 550 °C is selected the temperature shown in the display is 312.5 °C). This temperature must be reflected exactly by the indicator which is supplied by the respective current. If this is not the case, the selected analog input current span of the indicator is not equivalent to the chosen current output span of the pyrometer and one of the current spans or temperature range have to be modified. By pressing the test push button once again or by pressing any push button of the LC-display, the test current is switched off. Also after 30 seconds idle time, the test current is switched off. The unit will be in the measurement mode again. Emissivity (Emi) = 100% Exposure time (t90) = min Clear time (tClear) = off Analog output (mA) = 0 ... 20 mA Sub range (from / to) same as temperature range Series IPE 140 Manual Address (Adr) = 00 Baud rate (Baud) = 19200 Bd Temperature display (C / F) = °C Wait time (tw) (for RS485) = 10 Interface (RS485 / RS232) = RS232 Instrument settings 22 5 Settings / parameter descriptions 5.1 Emissivity (Emi) For a correct measurement, it is necessary to adjust the emissivity. Emissivity is the relationship between the emission of a real object and the emission of a black body radiation source (this is an object which absorbs all incoming rays and has an emissivity of 100%) at the same temperature. Different materials have different emissivities ranging between 0% and 100% (settings at the pyrometer between 10 and 100%, the set value is indicated on the display). Display of the adjusted emissivity Additionally, the emissivity depends on the surface condition of the material, the spectral range of the pyrometer, and the measuring temperature. The emissivity setting of the pyrometer has to be adjusted accordingly. Typical emissivity values of various common materials for the two spectral ranges of the instruments are listed below. The tolerance of the emissivity values for each material is mainly dependent on the surface conditions. Rough surfaces have higher emissivities. Emissivity for IPE 140 and IPE 140/39 Measuring object Emissivity [%] “Black body furnace“ 100 Extruded Aluminum 5 to 15 Brass 10 to 15 Brass, oxidized 70 (tarnished) Copper 3 Copper, oxidized 70 Inconel, oxidized 85 Emissivity for IPE 140/34 Film thickness PE PP 7 µm 76 15 µm 92 25 µm 95 96 75 µm 96 96 125 µm 96 96 250 µm 96 96 PET 90 96 PC 96 96 Measuring object Steel, oxidized Nickel Molybdenum Emissivity [%] 70 to 85 8 to 12 18 Black Carbon 95 Graphite Porcelain, glazed Iron, oxidized 80 to 90 60 80 to 90 Emissivity [%] PVC PA PS 96 96 96 96 96 96 96 Emissivity for IPE 140/45 Measuring object Hot CO2 containing gases Hot CO2 containing gases through sapphire window (1.6 mm thick) Series IPE 140 Manual 70 96 PUR PMMA Cell.Acetate 90 96 66 96 96 96 96 Emissivity [%] 100% 85% Settings / parameter descriptions 23 5.2 Compensation of ambient temperature (tamb) The compensation of the ambient temperature can be used at low measuring Settings: temperatures (below 100 °C). This compensation is used for a very few special auto applications only. The standard setting of this parameter is “auto”, because the 00 °C (32 °F) .. . temperature of the air around the pyrometer is normally the ambient temperature 53 °C (127 °F) of the measured object. Should the measured object be placed in an area with a higher wall temperature (e.g. inside a furnace), the measurement might be falsified (probably too high temperature indication). This influence can be compensated by pre-setting the ambient temperature of the object (pre-setting within the measuring range of the instrument). Attention: This method only improves the results if the ambient temperature at the place of the measured object is always constant and the emissivity is well known. 5.3 Exposure time (t90) The exposure time is the time interval when the measured temperature has to be present after an abrupt change so that the output value of the pyrometer reaches a given measurement value. This is the time it takes to reach 90% of the recorded temperature difference. In the “min” position, the device operates using its time constant. Settings: min 0.01 s 0.05 s . .. 10.00 s The dynamic exposure time adjustment prolongs the exposure time at the lower range limit, even if t90 is set to a lower value. The following table shows the dependence of the exposure time from the measured temperature. Switch points on increasing temperatures: Type, temperature range IPE 140, MB 3.5 IPE 140, MB 5 IPE 140, MB 10 IPE 140, MB 12 switch points for = 1 5 ms 5 ms 1.5 ms 63 °C 67 °C 111 °C 162 °C 124 °C 183 °C 35 ms switch points for = 0.5 35 ms 5 ms 5 ms 1.5 ms 67 °C 74 °C 133 °C 197 °C 150 °C 224 °C IPE 140/34, MB 5 147 °C 211 °C 176 °C 252°C IPE 140/39, IPE 140/39, IPE 140/39, MB 7 MB 12 MB 14.5S 113 °C 81 °C 168 °C 137 °C always 1.5 ms 95°C 205°C IPE 140/45, MB 20 always 1.5 ms If the maximum value storage is selected and the measuring object has a low emissivity and the measuring temperature is at the beginning of the range, a higher exposure time must be chosen to prevent measurement errors. Longer exposure times can be used for the measurement of objects which have rapidly fluctuating temperatures to achieve constant temperature reading. Series IPE 140 Manual Settings / parameter descriptions 24 5.4 Clear time of the maximum value storage (tClear) If the maximum value storage is always switched on, the highest last temperature value will be displayed and stored. As such, it may be beneficial to periodically clear and reset the stored maximum values in order to obtain new temperature readings. This feature is particularly useful when fluctuating object temperatures cause the display or the analog outputs to change too rapidly, or the pyrometer is not constantly viewing an object to be measured. In addition, it may also be beneficial to periodically delete and reset the stored maximum values. Settings: off 0.01 s . .. 25 s extern auto Hold The following settings are possible: off: At off, the maximum value storage is switched off and only momentary values are measured. 0.01...25 s: If any clear time between 0.01 s and 25 s is set, the maximum value is estimated and held in double storage mode. After the entered time, the storage will be deleted. extern: The external clearing can be activated and used within your own software (see section 8, Data format UPP®) or via an external contact (for connection see 3.1.1 Pin assignment of the male socket on the back of the pyrometer). In this case, the storage operates only in single storage, because only a single deletion mechanism is used. auto: The auto mode is used for discontinuous measuring tasks. For example, when objects are transported on a conveyer belt and pass the measuring beam of the pyrometer only for a few seconds. Here the maximum value for each object has to be indicated. In this mode, the maximum value is stored until a new hot object appears in the measuring beam. The temperature which has to be recognized as “hot“ is defined by the low limit of the adjusted sub range. The stored maximum value will be deleted when the temperature of the new hot object exceeds the low limit from of the sub range by 1% or at least 2 °C. If a lower limit is not entered, the maximum value storage will be deleted whenever the lower level of the full measuring range has been exceeded. Hold: The function hold enables you to freeze the current temperature reading at any moment. To use this function, anexternal push button or switch has to be connected (which holds the temperature reading as long as the contacts are closed). Operation note: depending on the settings, the maximum value storage either works in single storage mode or in double storage mode: Single storage: the single storage is used when you want to reset the stored value using an external impulse via one contact closure from an external relay (i.e. between two measured objects). The relay contact is connected directly to the pyrometer between pins J and K. This mode allows a new value to be established, after each impulse from the reset signal. Double storage: when entering the reset intervals via push buttons or PC interface, the double storage is automatically selected. This mode utilizes two memories in which the highest measured value is held and is deleted alternately in the time interval set (clear time). The other memory retains the maximum value throughout the next time interval. The disadvantages of fluctuations in the display with the clock frequency are thereby eliminated. Series IPE 140 Manual Settings / parameter descriptions 25 Note: 5.5 In the command structure, the maximum storage comes after the exposure time. This results in: clear time the adjusted response time is useless clear times must be at least 3 times longer than the response time only maxima with full maximum value can be recorded, which appear at least 3 times longer than the response time. Analog output (mA) The analog output has to be selected according to the signal input of the connected instrument (controller, PLC, etc.). 5.6 Subrange (from / to) Settings: 0 ... 20 mA 4 ... 20 mA You have the opportunity to choose a subrange (minimum 51 °C) within the basic measuring range of the pyrometer. This subrange corresponds to the analog output. From describes the beginning of this measuring range, to the end of the range. Additionally, with the setting of a subrange, it is possible to fulfill the requirements of the “auto” clear mode of the maximum value storage (see above). 5.7 Address (Adr) When connecting several pyrometers to one serial interface with RS485, it is Settings: necessary for each instrument to have its own device address for communication 00 .. . purposes. First, it is necessary to connect each instrument separately to give it an 97 address. After that, all instruments can be connected and addressed individually. If parameters may be changed simultaneously on all pyrometers, the global Address 98 can be used. This allows you to program all pyrometers at the same time, regardless of the addresses that have already been assigned. If the address of a pyrometer is unknown, it is possible to communicate with it using the global Address 99 (connect only one pyrometer). 5.8 Baud rate (Baud) The transmission rate of the serial interface in Baud (Bd) is dependent on the length of the cable. A standard cable length with RS232 for 19200 Bd is 7 m, and with RS485, it is 2 km. The baud rate is reduced by 50% if the transmission distance is doubled. 5.9 Settings: 2.4 .kBd .. 115.2 kBd Settings: °C °F Temperature display (C / F) The temperature can be displayed in °C or °F. 5.10 Wait time (tw) Using a pyrometer with RS485 it is possible that the connection is not fast enough Settings: to receive the pyrometer’s answer to a command of the master. In this case, a 00 .Bit .. minimum delay time (tw) can be set. The pyrometer waits this time until it answers 99 Bit a master inquiry (e.g.: tw = 02 at a baud rate 9600 means a wait time of 2/9600 sec). Note: the setting of a delay time (tw) does not guarantee an answer to some commands directly after this time. Certain commands require an internal operation time of max. 3 ms. 5.11 Maximum internal temperature (MaxIntTemp) Shows the maximum internal temperature the device ever reached. 5.12 Error status (Status) In case of a device error, the pyrometer displays a hex code which identifies this error to service. The standard display at this point is ok. Series IPE 140 Manual Settings / parameter descriptions 26 6 Software InfraWin The operating and analyzing InfraWin software is included with delivery of the pyrometer. In addition to allowing you to make parameter adjustments, the InfraWin software also provides temperature indication, data logging, and measurement analysis features. This section gives an overview about the functions of the software. It also provides a description of the individual icons found in the program's help menu. To access the help menu, press the F1 button on your keyboard after loading InfraWin or click on the ? in the menu bar. The following descriptions refer to software InfraWin version 5.0. The latest version is available for free as a download from the homepage www.lumasenseinc.com. 6.1 Connecting the pyrometer to a PC The program InfraWin can operate up to two devices. Two devices using RS485 may be operated simultaneously by the same interface, if two different addresses have been properly entered (see section 5.7 Device Address for more information). 6.2 Installation To install the InfraWin software, select setup.exe from the InfraWin-CD or from the downloaded and unpacked zip file from the internet and then follow the installation instructions. 6.3 Program start The first time you load InfraWin 5, you will be prompted to select a default language. The InfraWin software is available in German, English, Spanish, French, Portuguese, and Chinese. Once installed, click Language/Languages if you would like to select another language. On the start page, the screen shows the following start menu icons. Opens a saved file Storage of measured values for further processing Starts measurement for the selected device Setting of the parameters of the instruments Time interval between two measurements Listing of measured or stored values in tabular form Processing of measured (stored) readings in graph form Processing of measured (stored) readings in a text file Calculation of spot sizes in various measuring distances Only if available: controls the programmable controller PI 6000 Search I/O module Series IPE 140 Manual Software InfraWin 27 6.4 Basic settings All preset values for the device can be displayed and modified, if necessary under the Devices/Parameters window. Changing an existing pyrometer setting can be accomplished by typing a value in an input box or by selecting a preset value from the list field. Choose the correct settings for your application from the displayed options. This window contains the parameter settings described in Chapter 5, Parameters. 6.4.1 Open/Save 6.4.2 1 measur… The open and save buttons enable you to save customized configurations, which can then be opened and reused at any time. The 1 measur… function shows the current measuring temperature in the pyrometer parameters window for approximately 1 second. 6.4.3 Print The print icon will generate a screenshot of the parameter form and allow you to send it a printer. 6.4.4 Close 6.4.5 Test The close icon will close the pyrometer parameters window. Clicking on the Test icon will open a window that allows you to directly communicate with the pyrometer using the UPP (Universal Pyrometer Protocol) Data Format commands. After entering an interface command (00 is the adjusted address ex works, ms is the command reading temperature value) and clicking on Send, the window shown to the right will be opened. This window already shows the answer of the pyrometer in 1/10°. The actual temperature reading is 41.2 °C. Len indicates the length of the answered data string, incl. Carriage Return (Chr(13)). Series IPE 140 Manual Software InfraWin 28 In the lower part of the window, the connection with the preset baud rate can be checked. Here the command was sent 100 times with 19200 baud. It has taken 0.799 seconds without transmission errors. Emi: AutoFind: If the true temperature of the measured object is known, you can calculate the emissivity of the measured object using the Emi: AutoFind function: A measured temperature is displayed with the current set emissivity (in this example 100%) (here: 71.7 °C). If you press Emi: Autofind a window will open which allows you to enter the "true" temperature (here 100 °C). Once the temperature entry has been entered and confirmed with Calculate, InfraWin will then calculate the emissivity which occurs with the new temperature (here 33.0%). This is displayed immediately and can be used for further temperature measurement. 6.5 Measurement online trend The measurement function allows you to access a number of input tabs located on the left side of the screen. The main or home tab is the Output Screen. You can toggle the input tabs on and off by clicking them. This window displays: temperature as graphical diagram internal temperature of the instrument current temperature quantity of the measured values and file size of the current measurement The example shows a sample reading over the period of approximately 15 seconds with a temperature range between 40 and 105 °C. The final temperature (at the end of the reading) is 99.9 °C. The Mark Zone button allows you to color mark a temperature range for easier recognition. The Threshold button allows you to set a temperature value as a baseline to prevent recording values above or below the baseline temperature. This allows you to keep the output file size small. The Scaling Trend button allows you to scale temperature trend view. Series IPE 140 Manual Software InfraWin 29 Note: The measuring values of “measurement online trend” are automatically saved as "standard.i12". Should you need to edit the data later, you need to save the file as another .i12-file because old values are over-written when a new measurement is taken. Files from older program versions (.i10-files) can be opened and saved as .i12. 6.6 Listing (analyzing) For analyzing the measured values in this field, all measured data appears in a numeric list. The column between time and temperature provides a time resolution of milliseconds. The value specifies the time in seconds after midnight (0:00 h). The amount of data depends on the frequency that readings were taken (settings at 6.9 PC sampling rates). As the amount of data increases, so does the amount of storage space required to save it. In order to save room, all .i12 data files are stored by a binary code. 6.7 Output .TXT file (analyzing) The same file as under Output listing may be converted into a text file and can be easily opened, for example with Microsoft Excel. With the standard import settings, Excel automatically formats the columns accordingly (with tabs as separators). 6.8 Trend output (analyzing) The graph’s curve depicts the temperature change over time within the specified temperature range. Additionally, other information appears in this window; such as recorded time (x-axis) and temperature in degrees (y-axis) as well as the time and temperature at the vertical cursor line which can be dragged with the mouse. Selecting the Trend output initially causes all the saved data to be displayed. If the data exceeds an amount that can be reasonably represented, you may zoom in on a partial segment using the mouse (such as the segment represented in the example). Under Total you can return to the representation of the entire curve. Series IPE 140 Manual Software InfraWin 30 Note: The last reading is saved in the standard.i12 file and automatically appears in this form upon opening Listing or Trend output. Selecting file open with another file, the previous file will be overwritten and replaced by the standard.i12 file. 6.9 PC sampling rate (time interval between two measurements) This function sets a time interval. After each interval, one measured value is stored on the PC. Longer time intervals will result in creating smaller stored file sizes. This function is mainly used for long term measurements. 6.10 Spot size calculator The InfraWin Spot Size Calculator calculates the data for the nonfocused regions. To calculate data, enter the values of aperture D, focused measurement distance a, and focused measuring field diameter M as found in the documentation relating to the optics supplied with your specific pyrometer. After entering the aperture and the main spot size, the input of interim values calculates spot sizes in different measuring distances of the fixed optics. 6.11 Search I/O Module The I/O Module allows accessories to connect to the software and is used to trigger measurement externally or to send a signal (like a relay) under certain conditions. Series IPE 140 Manual Software InfraWin 31 To ensure consistent document formatting, this page was intentionally left blank. Series IPE 140 Manual Software InfraWin 32 7 Maintenance 7.1 Cleaning the window Because there are no moving parts in the Series 140 pyrometer, the only regular maintenance required is a periodic inspection of the front window for build-up of foreign particles which, in time, can influence the energy received by the instrument. The Series 140 pyrometer has a “Dirty Window” warning alarm feature that can measure the current window/optical path transmission and provide a contact closure alert when the window transmission falls below the user set point. The Series 140 pyrometer window is not water soluble and can be cleaned with standard lens tissue dampened with a camera-store lens-cleaning solution. A soft blower/brush (also at camera stores) should be used to remove any grit on the window before you rub the lens with lens tissue and solution. Attention: NEVER CLEAN A SERIES 140 PYROMETER WINDOW WITH A DRY TISSUE OF ANY KIND! The only time dry lens tissue may be used is to dry a window which has already been cleaned with wet lens tissue. 7.2 Optics replacement Series 140 pyrometers are equipped with a focusable optics. This optics can be changed against another. Replacement optics for different measuring distances can be used without recalibration of the instrument. However, replacement may be necessary if the lens is scratched or the pyrometer will be used for other measuring distances. Replacement: Only the lens will be replaced for changing the focusable optics. The fixing ring has to be removed with a suitable objective wrench. After removing the old lens, put in the new one with the convex side to the front. Fix the lens with a new fixing ring. On the inside of this ring is the sticker with the optics data. For differentiation they are marked with a color mark at the border of the lens: Focusable optics 0-PE Focusable optics 1-PE Focusable optics 2-PE Focusable optics 3-PE Series IPE 140 Manual Lens Fixing ring Sticker with optics data (without color mark) (red/green) (black/green) (yellow/red) Maintenance 33 To ensure consistent document formatting, this page was intentionally left blank. Series IPE 140 Manual Data format UPP (Universal Pyrometer Protocol) 34 8 Data format UPP (Universal Pyrometer Protocol) Software commands can be exchanged directly with the pyrometer through an interface using suitable communication software or by using the Test function located in the Pyrometer Parameters window of the InfraWin software package. The data exchange occurs in ASCII format with the following transmission parameters: • The data format is: 8 data bits, 1 stop bit, even parity (8,1,e) no handshake; • The device responds to the entry of a command with output (such as the measuring value) + CR (Carriage Return, ASCII 13), to pure entry commands with ok + CR, or no + CR. • Every command starts with the 2-digit device address AA followed by two lower case command letters and finished with CR. Example Read Command: Entry: “00em“ + CR The emissivity setting ( ) of the device with the address 00 is returned. Answer: “0970“ + CR means Emissivity = 0.97 or 97.0% • The ASCII parameter “X” indicates a change to be made in a parameter. When used, the command contains the new value. Example Write Command: Entry: “00emXXXX“ + CR The parameter used for the emissivity setting ( ) with the address 00 is changed. Answer: “00em0853“ + CR changes the Emissivity to 0.853 or 85.3% A “?” after the lower case command letters answers with the limits of the respective settings (only at setting commands, not at query commands). Example Read Limits Command: Entry: “00em?“ + ˜CR! Answer: Could be 00501000 + <CR>, which means that E can vary between 0.050 and 1.000 (or 5% and 100%) Description Reading temperature value: Command AAms Reading temperature value repeated: Emissivity: Ambient temperature compensation: AAmsXXX Series IPE 140 Manual AAemXXXX AAutXXXX Parameters Output: XXXXX (dec., in 1/10 °C or °F) (77770 = Instrument’s temperature too high or chopper wheel rotates with wrong speed 88880 = temperature overflow) XXX = 000...999 (XXX = number of measuring values) XXXX = (0010 ... 1000‰) (decimal) XXXX = value of ambient temperature, 4-digit, hex XXXX e.g. FFEC corresponds to -20 degrees - 99dez = FF9Dhex means: automatic, no manual compensation Data format UPP (Universal Pyrometer Protocol) 35 Exposure time t90: AAezX Clear times of the maximum value storage: AAlzX External clearing: AAlx Analog output: Reading basic temperature range: AAasX AAmb Reading temperature sub range: AAme Setting of temperature sub range: Address: AAm1XXXXYYYY Baud rate: AAbrX Temp. display °C / °F Wait time: AAfhX AAtwXX Internal temperature: Max. internal temperature: Error status: AAgt AAtm Laser targeting light: AAlaX Reading interface: Lock keyboard: AAin AAlkX Series IPE 140 Manual AAgaXX AAfs X = 0 ... 6 (decimal) 0 = intrinsic time constant of the device 1 = 0.01 s 3 = 0.25 s 5 = 3.00 s 2 = 0.05 s 4 = 1.00 s 6 = 10.00 s X = 0 ... 9 (dez.) 0 = Maximum value storage off 1 = 0.01 s 4 = 1.00 s 7 = external deletion 2 = 0.05 s 5 = 5.00 s 8 = automatically deletion 3 = 0.25 s 6 = 25.00 s 9 = Hold Clearing the max. value storage (only with lz = 7, external) X = 0...1 0 = 0...20 mA 1 = 4...20 mA Output: XXXXYYYY (hex 8-digit, °C or °F) XXXX = beginning of temp. range YYYY = end of temp. range Output: XXXXYYYY (hex 8-digit, °C or °F) XXXX = beginning of temp. range YYYY = end of temp. range XXXX (hex 4-digit) beginning of temp. range (°C or °F) YYYY (hex 4-digit) end of temp. range (°C or °F) XX = (00 ... 97) 00 ... 97 = regular device addresses 99 = Global address with response 98 = Global address without response (only setting commands!) X = 1...6 or 8 (dez.) 1 = 2400 baud 5 = 38400 baud 2 = 4800 baud 6 = 57600 baud 3 = 9600 baud (7 is not available) 4 = 19200 baud 8 = 115200 baud Output: X = 0: display in °C; X = 1: display in °F XX = 00 ... 99 (decimal, in bit time of the current baud rate) Output: XXX (dec. 000 ... 099°C or 032 ... 210 °F) Output: XXX (dec. 000 ... 099°C or 032 ... 210 °F) Output: XX; XX=00…FF (00 = no error) (01…FF: error code for LumaSense service) X = 0 switch off laser X = 1 switch on laser Output: 1 or 2 (1 = RS232, 2 = RS485) X = 0 ... 3 1 = lock lk1, removal with command lk0 or power offon 0 = removal of lock lk1 3 = continuous lock lk3, removal only with command lk2 2 = removal of lock lk3 Data format UPP (Universal Pyrometer Protocol) 36 Reading parameters: AApa Device type: AAna Serial number: Device type / software version: AAsn AAve Detailed Software version: AAvs Reference number: AAbn Output 11-digit, decimal: Digit 1 und 2 (10...99 or 00): Emissivity Digit 3 (0 ... 6): Exposure time Digit 4 (0 ... 8): Clear time max. storage Digit 5 (0 ... 1): Analog output Digit 6 und 7: (00 ... 98): Internal temperature Digit 8 und 9 (00 ... 97): Address Digit 10 (0 ... 6 or 8): Baud rate Digit 11 (0,1, 3): Keyboard lock Output: “IPE 140”, “IPE 140/34”, “IPE 140/39” or “IPE 140/45” (16 ASCII-characters) Output: XXXXX (dec. 5-digit) Output: XXYYZZ (6-digit decimal) XX = 63 (for all series IPE 140 pyrometers) YY = Month of software version ZZ = Year of software version tt.mm.yy XX.YY tt = day; mm = month; yy = year; XX.YY = software version Output: XXXXXX (hex 6-digit) Note: the letter “l” means the lower case letter of “L”. Additional instruction for the RS 485 interface: Requirements to the master system during half-duplex operation: 1. After an inquiry, the bus should be switched into a transmission time of 3 bits (some older interfaces are not fast enough for this). 2. The pyrometer's response will follow after 5 ms at latest. 3. If there is no response, there is a parity or syntax error and the inquiry has to be repeated. Series IPE 140 Manual Data format UPP (Universal Pyrometer Protocol) 37 To ensure consistent document formatting, this page was intentionally left blank. Series IPE 140 Manual Data format UPP (Universal Pyrometer Protocol) 38 9 Reference Numbers 9.1 Reference numbers instrument Type Temperature range IPE 140 IPE 140/34 IPE 140/39 IPE 140/45 MB 5: MB 10: MB 12: MB 5: MB 7: MB 12: MB 14.5S: MB 18: MB 20: 5 ... 500 °C 30 ... 1000 °C 50 ... 1200 °C 75 ... 500 °C 20 ... 500 °C 75 ... 1200 °C 300 ... 1450 °C 500 ... 1800 °C 400 ... 2000 °C With laser targeting light 3 875 740 3 875 720 3 875 700 3 875 800 3 875 840 3 875 760 3 875 860 3 875 960 3 875 880 With thru-lens view finder 3 875 750 3 875 730 3 875 710 3 875 810 3 875 850 3 875 770 3 875 870 – 3 875 890 Ordering note: When ordering, please select one focusable optics (0-PE, 1-PE, 2-PE or 3-PE) (see section 3.3 Optics). A connection cable is not included in scope of delivery and has to be ordered separately. 9.2 Reference numbers accessories 3 848 600 3 848 370 3 848 380 3 848 390 Replacement optics 0-PE Replacement optics 1-PE Replacement optics 2-PE Replacement optics 3-PE 3 820 340 3 820 530 3 820 540 3 820 830 3 820 840 3 820 550 3 820 330 3 820 500 3 820 510 3 820 810 3 820 820 3 820 520 3 820 740 3 820 750 Connection cable, length 5 m, 90° connector Connection cable, length 10 m, 90° connector Connection cable, length 15 m, 90° connector Connection cable, length 20 m, 90° connector Connection cable, length 25 m, 90° connector Connection cable, length 30 m, 90° connector Connection cable, length 5 m, straight connector Connection cable, length 10 m, straight connector Connection cable, length 15 m, straight connector Connection cable, length 20 m, straight connector Connection cable, length 25 m, straight connector Connection cable, length 30 m, straight connector Connection cable, length 5 m, straight connector, temperature resistant up to 200 °C Connection cable, length 5 m, 90° connector, temperature resistant up to 200 °C 3 834 280 3 834 270 3 835 230 3 837 290 3 835 060 3 834 140 3 837 240 3 835 460 3 843 530 Adjustable mounting angle Ball and socket mounting Air purge Cooling jacket, stainless steel Air purge for cooling jacket Heavy ball and socket mounting for cooling jacket Cooling plate 90° mirror with CaF2 glass window Rugged scanner SCA 140 with CaF2 glass window (scanning angle 0...12°, scanning frequency 1...5 Hz) Air purge for scanner SCA 140 3 835 290 Series IPE 140 Manual Reference Numbers 39 3 852 290 3 852 550 3 890 640 3 890 650 3 890 560 3 890 570 3 890 520 3 890 530 3 826 500 3 826 510 3 825 430 Power supply for DIN rail mounting NG DC (100 to 240 V AC 24 V DC, 1 A) Power supply NG 2D, as NG 0D: additionally with 2 limit switches (not for US sales) LED digital display DA 4000-N LED digital display DA 4000: with 2 limit switches LED digital display DA 6000-N: with possibility for pyrometer parameter settings for digital IMPAC pyrometers; RS232 interface LED digital display DA 6000-N: with RS485 interface LED digital display DA 6000; DA 6000-N additional with 2 limit switches and analog input and output, RS232 interface LED digital display DA 6000 with RS485 interface HT 6000, portable battery driven indicator and instrument for pyrometer parameter setting PI 6000: PID programmable controller, very fast, for digital IMPAC pyrometers -7520, RS232 RS485 converter Series IPE 140 Manual Reference Numbers 40 10 Troubleshooting Before sending the pyrometer for repair, try to find the error and to solve the problem with the help of the following list. Temperature indication too low Incorrect alignment of the pyrometer to the object New correct alignment to achieve the maximum temperature signal (see 3.2) Measuring object smaller than spot size check measuring distance, smallest spot size is at nominal measuring distance (see 3.3) Measuring object is not always in the measuring spot of the pyrometer Use max. value storage (see 5.4) Emissivity set too high Set lower correct emissivity corresponding to the material (see 5.1) Lens contaminated Clean lens carefully (see 7.1) Temperature indication too high Emissivity set too low Set lower correct emissivity corresponding to the material (see 5.1) The measurement is influenced by reflections of hot machine parts Use mechanical construction to avoid the influence of the interfering radiation Measuring errors Indicated temperature is decreasing during the use of the pyrometer, contamination of the lens Clean lens. Recommendation: use of air purge (see 7.1) Indicated temperature is decreasing during the use of the pyrometer, although the air purge unit is used. Probably compressed air is not clean or air failed Clean the lens and use clean, dry and oil free compressed air Air contamination in the sighting path between pyrometer and object Change position of the pyrometer with a clean sighting path (if necessary use a ratio pyrometer) HF-interferences Correct the connection of the cable shield (see 3.1) Instrument overheated Use cooling jacket with air or water cooling (see 3.1) Temperature Indication is fluctuating, probably caused by changing emissivity Wrong pyrometer type, use of ratio pyrometer recommended Laser targeting light Laser targeting light fails Instruments max. Temperature is exceeded. Use cooling jacket (see 2.6) Note: The wavelength band of the IPE 140 series reacts at low measuring temperatures to incandescent lamps or very bright daylight (not valid for fluorescent tube). For a correct measurement strong external light to the measured object should be avoided. Series IPE 140 Manual Troubleshooting 41 To ensure consistent document formatting, this page was intentionally left blank. Series IPE 140 Manual Troubleshooting 42 Index 1 1 measure… 28 F A Factory settings 22 Focusable optics 18 Accessories 12 Address 26 Ambient temperature compensation 24 Analog output 26 Appropriate use 9 G General Information 5 H B Hold function 16 Basic settings 28 Baud rate 26 I C Cleaning the front window 33 Clear time 25 Close 28 Color mark 33 Connecting the pyrometer to a PC 16, 27 Connection cable 9 Connection of additional analyzing devices 17 Connection to RS232 17 Connection to RS485 17 Converter 16 cooling plate 12 D I/O Module 31 Installation 27 Installation, Electrical 15 Instrument settings 21 Internal temperature 26 IPE 140 9, 23 IPE 140/34 9, 23 IPE 140/39 9, 23 IPE 140/45 9, 23 IR calculator 19 K Key panel operation 21 L Data format UPP 35 Dimensions 11 Disposal 7 double storage 25 Laser targeting light 18 Laser Targeting Light 5, 18 Legend 5 Liability 6 Listing (analyzing) 30 E M Electrical Connection 6 Electrical Installation 15 Emi: Autofind 29 Emissivity 23 Error Status 26 Exposure time 24 Maintenance 33 Mark Zone 29 Maximum value storage 16 Measurement 28 Measurement online trend 29 Measuring distance 19 Series IPE 140 Manual Index 43 O Open/Save 28 Operating mode contact 16 Optics 18 Optics replacement 33 P Packing 13 PC sampling rate 31 Physical User Interface 11 Pin assignment 16 Pin assignment of the male socket 25 Print 28 Program start 27 Trend output 30 Troubleshooting 41 TXT file 30 U Unpacking the Instrument 6 UPP data format 35 W Wait time 26 Warranty 6 R Reference numbers Accessories 39 Instrument 39 Reference Numbers 39 Repair 7 RS232 17 RS485 17 S Safety 5 Scaling Trend button 29 Scope of delivery 9 Serial interface 22 Service Request 7 Settings / parameter descriptions 23 Sighting 18 single storage 25 Software InfraWin 27 Spot size calculator 31 Spot sizes for non-focused distances 19 Storage 13 Subrange 26 Support 7 Switch points 24 T Technical Data 9 Temperature display 26 Test 28 Test function 22 Threshold button 29 Thru-lens view finder 18 Time interval 31 Transport 13 Series IPE 140 Manual Index 44 ">
Advertisement
Key features
- 3.43 µm spectral range
- Non-contact temperature measurement
- Adjustable emissivity
- Adjustable exposure time
- Digital interface (RS232 or RS485)
- Analog output (0 to 20 mA or 4 to 20 mA)
- Plastic film measurement
Frequently asked questions
The IPE 140/34 is designed to measure plastic films with a minimum thickness of 30 µm.
The IPE 140/34 has a spectral range of 3.43 µm.
Yes, the IPE 140/34 can be connected to a computer via a RS232 or RS485 interface.