Data Sheet | Wayne Kerr 6430A, 6440A Precision Component Analyzer Datasheet 5 Pages
Data Sheet | Wayne Kerr 6430A, 6440A Precision Component Analyzer Datasheet
Add to My manuals5 Pages
The Wayne Kerr 6430A & 6440A Precision Component Analyzers are designed to allow complete and accurate characterization of any component. They offer a wide range of measurement functions, industry-leading accuracy and a wide frequency range, providing the user with a complete solution for component testing. The 6430A & 6440A are suitable for use in production environments as well as design laboratories.
advertisement
Precision Component Analyzers
6430A & 6440A
The 6430A and 6440A
Precision Component
Analyzers have been designed to allow complete and accurate characterization of any component.
A full complement of measurement functions, industry-leading accuracy and a wide frequency range provides the user with a complete solution for component testing.
The 6430A upper frequency limit is 500kHz while the 6440A extends to 3MHz and adds the ability to graphically display measurements in real time versus frequency.
If your business is design or manufacture of components, or you are a user of components as part of your equipment design, these products have been specifically developed with you in mind and at an affordable price.
●
●
●
Key Features include
●
●
●
●
●
0.02% Basic measurement accuracy
Real time graphical sweep on all measurement functions
Characterize components to 3MHz
Fast measurement speed
More measurement functions, including -
C, L, Z, Y, X, B, Rac, Rdc, G, Q, D, o & Resonant Frequency
High measurement accuracy on Dissipation Factor
Large LCD display and intuitive user interface
Unbeatable price
Industry leading basic accuracy
Wayne Kerr Electronics has over 50 years experience in developing accurate measurement instruments. The 6430A and 6440A continue this tradition by providing an industry leading accuracy of
0.02% combined with the measurement integrity and functionality demanded by modern industry.
Capacitor testing requires accurate, repeatable and stable Dissipation Factor measurements. The 6430A and 6440A provide the performance required while displaying measurements at high resolution.
All the measurement functions you need!
Whatever measurement you need the
6430A and 6440A will more than likely have it, and more besides! Providing all the well-known measurement parameters of Impedance (Z), Phase Angle (o),
Inductance (L), Capacitance (C), DC
Resistance (Rdc), AC Resistance (Rac),
Quality Factor (Q) and Dissipation Factor
(D). In addition Admittance (Y),
Conductance (G), Reactance (X) &
Susceptance (B) are available.
The Resonant Frequency can be automatically calculated for any component together with its equivalent series or parallel circuit at that frequency.
If you want to look at component performance over a fixed number of frequencies, use Multi Frequency mode.
In this mode the operator decides which parameter is to be measured and at what frequency. The 6430A and 6440A does the rest, creating an easy to read table on the large LCD display. It will even tell you if each test has passed or failed.
Real time graphical sweep on all measurement functions up to
3MHz
Characterize a component graphically and in real time over the full frequency band of the 6440A.
The user can select any of the available measurement functions and graph the parameter in real time against frequency. The user is also able to toggle between the major and minor term, for example
Impedance and Phase Angle. Both axes can be selected as either linear or logarithmic.
Following the first sweep the user can hit the FIT key, which automatically scales the vertical axis to provide the optimum display for the component under test. A marker is also displayed which can be positioned at any point on the graph using the navigation keys. Use the MAX and MIN keys to position the marker at the peak or trough of the graph.
The marker position is displayed at the bottom of the screen showing the vertical axis value, its minor term and frequency.
Characterize components to
3MHz
The 6440A can test any component up to 3MHz while the 6430A has a maximum frequency of
500KHz. Both instruments start at 20Hz.
Frequencies can be set with a resolution of < 1% on the 6440A.
Precision
Component Analyzers
6430A & 6440A
Accurate measurements guaranteed, and at high speed!
In some applications absolute measurement accuracy is a must while for others high speed is critical. The 6430A and 6440A provide both industry leading accuracy and extremely fast measurement speed. The 20 measurements per second performance makes the instruments suitable for the production environment as well as the design laboratory.
Printing results and external control
Need to produce a hard copy of your test results? Whether it’s a single parameter, table of results in Multi Frequency or a logarithmic graph of component performance, the 6430A and 6440A will output the results directly to a printer. If you need to control the instrument automatically, use the GPIB interface and you can achieve complete external control.
Output lines are also available to control a bin handler interface. Component tolerance levels may be pre-set and stored. A 25-way D connector mounted on the rear panel gives easy access for connection to external component handling equipment.
Function
Frequency Range
Basic Accuracy
Measurement Function
‘Multi Frequency’ mode
‘Graph’ mode
Printer Output
GPIB Interface
Measurement Speed
6430A
20Hz - 500kHz
0.02%
Z, o, L, C, R, Q, D, Y, G, B, X &
Resonant Frequency
Yes
-
Yes
Yes up to 20 measurements/sec
6440A
20Hz - 3MHz
0.02%
Z, o, L, C, R, Q, D, Y, G, B, X &
Resonant Frequency
Yes
Yes
Yes
Yes up to 20 measurements/sec
Excellent performance and unbeatable price
As the leader in our field and with over 50 years experience of developing component analyzers you would expect
Wayne Kerr to provide the best performance in the industry, and you would be right. But what you might not be aware of, is that Wayne Kerr is able to offer all this performance, whilst delivering it at the most competitive selling price in the market.
Large LCD display and intuitive user interface
Whatever environment you work in production test, design, quality, service or goods inward inspection - presentation of the measured data and instrument control are a basic requirement of all users. Wayne Kerr have responded to this need.
Using a large bright LCD display, all the measurement and set up information is made available to the user, with measured values displayed in large characters. If the user needs to change the set up this is straightforward too. Using a soft key menu system you are led through intuitive commands which prompt at each point for the relevant information. No matter what your knowledge or skill level these products make set up and testing easy.
MEASUREMENT PARAMETERS
Any of the following parameters can be measured and displayed:
Inductance(L), Impedance(Z),
Rdc and Capacitance(C)
Series or Parallel Equivalent Circuit
C+R, C+D, C+Q, L+R, L+Q
Series Equivalent Circuit Only
X+R
Parallel Equivalent Circuit Only
C+G, B+G
Polar Form
Z + Phase Angle, Y + Phase Angle
TEST CONDITIONS
Frequency Range 6430A
20Hz to 500kHz > 1000 steps
Accuracy of set frequency ±0.005%
Frequency Range 6440A
20Hz to 3MHz > 1800 steps
Accuracy of set frequency ±0.005%
Pre-set frequencies
20, 25, 30, 40, 50, 60, 80, 100, 120,
150 repeats for each decade
Drive Level (Rdc)
100mV or 1V with 100
Ω source resistance
Drive Level (AC Measurements)
Open circuit voltage 1mV to 10V rms
Signal source impedance 50
Ω
Automatic Level Control (ALC) maintains constant voltage or current
DC Bias Voltage (Internal)
2V with rapid charge capacitor bias with settling time < 150mSec (to 99%)
DC Bias Voltage (External)
External supply of up to ±60V may be connected via rear panel
MEASUREMENT SPEEDS
Four selectable speeds for all measurement functions Up to 20 measurements per second for test frequencies
≥
100Hz
MEASUREMENT RANGES
R, Z 0.01m
Ω to > 2G
Ω
G, Y 1nS to > 2kS
L 0.1nH to > 2kH
C
D
1fF to > 1F
0.00001 to >1000
Q 0.00001 to >1000
Rdc 0.1m
Ω to > 10M
Ω
BASIC ACCURACY
L/C
Z
Q
±0.05%
±0.02%
± 0.05%(Q + 1/Q)
D ± 0.0002(1+ D
Rdc ± 0.1%
2
)
Accuracy varies with component range measurement speed and frequency
PRECISION COMPONENT ANALYZER 6430A/6440A SPECIFICATION
MODES OF OPERATION
Measurement
Selection of any measurement parameter and test condition
Single level function menu controlled by keyboard and soft keys
Single and repetitive measurements displaying major and minor terms
Deviation
As measurement mode but relative or percentage deviation from nominal value displayed for major or minor term
Multi-Frequency
Measurement parameters and test conditions set using measurement mode
Up to 8 frequencies with absolute or percentage limits on major term PASS/FAIL indication
Frequency Sweep (6440A only)
Measurement parameters and test conditions set using measurement mode
Graphical sweep vs frequency with selection of start, stop, step size, units and linear/log
Binning (Optional)
Measurement parameters and test conditions set using measurement mode
9 bins with absolute or percentage limits
Up to 99 sets of limits may be saved
Bin outputs 0-5V (nominal) with > 10mA current sink capability
25 way D-type interface connector
MEASUREMENT CONNECTIONS
4 front panel BNC sockets
4-wire (Kelvin) measurements with screen at ground potential
GENERAL
Power Requirements
115V or 230V AC ±10% (selectable)
150VA Max
Mains Frequency
50/60Hz
Mains Fuse Rating
230V operation - 1A ’T’ type
115V operation – 2A ’T’ type
Display
High contrast black and white LCD module
320 x 240 with CPL back lighting visible area 115 x 86mm
Printer Output
Centronics/parallel printer port
Remote Control
Designed to GPIB IEEE-488.2 and
SCPI 1992.0
Mechanical
Height 150mm (5.9”)
Width 440mm (17.37”)
Depth 525mm (20.5”)
Weight 11kg (24.25lbs)
ENVIRONMENTAL CONDITIONS
Temperature Range
Storage -40˚C to +70˚C
Operating 0˚C to 40˚C
Full Accuracy 15˚C to 35˚C
Relative Humidity
Up to 80% non-condensing
Altitude
Up to 2000m
Installation Category
II in accordance with IEC664
Pollution Degree
2 - mainly non-conductive
This equipment is intended for indoor use only in a non-explosive and non-corrosive atmosphere
SAFETY
Complies with the requirements of
EN61010-1
EMC
Complies with EN61326 for emissions and immunity
ORDER CODES
1J6440A
Precision Component Analyzer 6440A (20Hz
- 3MHz) with User Manual, 2m AC power cable and Kelvin clips 1EVA40100
1J6430A
Precision Component Analyzer 6430A
(20Hz - 500kHz) with User Manual, 2m AC power cable and Kelvin clips 1EVA40100
OPTIONS
/B Binning
ACCESSORIES
Kelvin Clips
1EVA40100 (fine jaw)
1EVA40180 (large jaw)
Component Fixture
1EV1006
Surface Mount Tweezers
A40120
Rack Mounting Kit
1EXA20230 3U (full width)
Wayne Kerr Electronics reserves the right to change specification without notice
Wayne Kerr
Electronics Ltd
USA
Wayne Kerr Inc
11 Sixth Road
Woburn MA 01801-1744
Tel: 781 938 8390
Fax: 781 933 9523
Sales: (800) 933 9319 email: [email protected]
An Advance International Group Company
UK
Durban Road Bognor Regis
West Sussex PO22 9RL
Tel: +44 1243 825811
Fax: +44 1243 824698 email: [email protected]
www.wayne-kerr.co.uk
GERMANY
Advance Power Electronic GmbH
Gutenbergring 5
63110 Rodgau-Nieder-Roden
Tel: +49 06106 7080 30
Fax: +49 06106 7080 40 email: [email protected]
Distributors worldwide contact UK office
HUNGARY
Advance Power Elektronikai KFT
Czobor u. 84/b
Budapest
H-1147
Tel/Fax: +36 1 220 9401
Fax: +36 1 251 5520 email: [email protected]
advertisement
Key Features
- 0.02% Basic measurement accuracy
- Real time graphical sweep on all measurement functions
- Characterize components to 3MHz
- Fast measurement speed
- High measurement accuracy on Dissipation Factor