Data sheet
DIGITAL LINK MEASURING INSTRUMENTS
PORTABLE 2.5G/10G ANALYZER
MP1580A
For 2.5G/10G Jitter/Wander Measurements
The MP1580A is a unique and powerful solution for analyzing jitter
at the standard OC-48/192 or STM-16/64 bit rates. It can measure
jitter of 2.5G/10G electrical interfaces (clock signal) with a simple
operation. In addition, when used in combination with the MP1570A
SONET/SDH/PDH/ATM Analyzer, evaluation of jitter characteristics
in digital transmission lines, systems and devices, such as — jitter
tolerance, jitter transfer, jitter generation, etc., can be performed
easily.
Application
• Output jitter measurement
The MP1580A can easily measure the jitter clock signal (electrical interface only) by just inputting the output clock of DUT directly.
Functions
• Complies with
the latest ITU-T O.172 and Bellcore GR-1377
standards
The MP1580A conforms to both the OC-192/STM-64 jitter measurement standards and supports required jitter modulation amplitude of
4000 UIp-p and 80 MHz jitter bandwidth.
• Supports 10 GHz wander measurement according to the latest
ITU-T G.813 standard (option)
The MP1580A can generate and measure various types of wander.
It can generate wander in the frequency range of 10 µHz to 10 Hz at
400,000 UIp-p max. In addition, MTIE/TDEV can be measured in
real-time using an external PC and optional application software
(MX150002A).
• Single cabinet support for both 2.5G and 10G jitter/wander
measurements
Just one MP1580A is required for 2.5G and 10G jitter generation and
analysis. When combined with the MP1570A and MU150000A, jitter
can be added to SONET/SDH signals and measured.
• Differences from existing instrument (MP1777A)
Anritsu launched the MP1777A 10 GHz Jitter Analyzer in February
of 1998, as a jitter measurement solution for OC-192/STM-64
(9953M). The new MP1580A Portable 2.5G/10G Jitter Analyzer is
providing more convenience in measurement without the need for
ancillary equipment (network analyzer, external E/O-O/E converter).
Anritsu has also developed a Wide Band O/E Converter
(MU150017A/B) for the MP1570A to support jitter measurement of
80 MHz at 9953.28 Mbit/s as required by ITU-T standard in conjunction with the MP1580A. Although it uses two cabinets, the compact
size makes the system ideal for R&D, manufacturing, installation and
maintenance. In addition, the MP1570A can be controlled from the
MP1580A for performing automatic measurements, such as Jitter
Tolerance and Jitter Transfer.
176
For product ordering information, see pages 5 – 12
Optical signals can be measured easily by combining the MP1580A
with the MP1570A, MU150000A, MU150001A and MU150017A/B.
• Jitter tolerance measurement
When the MP1580A is used with the MP1570A (send/receive jittered
clock), jitter tolerance tests can be performed on OC-192/STM-64
and OC-48/STM-16 signals of electrical and optical interfaces.
DIGITAL LINK MEASURING INSTRUMENTS
• Jitter transfer measurement
• Wander generation and measurement
When the MP1580A is used with the MP1570A (send/receive jittered
clock), jitter transfer tests can be performed on OC-192/STM-64 and
OC-48/STM-16 signals of electrical and optical interfaces.
The MP1580A can generate and measure of wander conforming to
ITU-T O.172 and also generation of TDEV conforming to ITU-T
G.813. It also can measure TIE (Time Interval Error) by itself and
measure MTIE and TDEV by connection of an external PC in which
MX150002A is installed.
Specifications
Frequency
Range: 9953.28, 2488.32 MHz
Offset range: ±100 ppm
Resolution: 0.1 ppm
Accuracy: ±0.1 ppm (calibrate after 60 min warm-up, 23 ±5˚C)
Generation function: Clock signal output, data signal output (with MP1570A), jitter on, wander on/off
Modulation source: Internal (sine wave), external (for jitter generation function only)
Modulation frequency accuracy: fm ±100 ppm (0.1 Hz to 80 MHz)
Jitter generation: Conform to ITU-T O.172
Jitter amplitude [UIp-p]
4000 Ul range
Slope : –20 dB/dec
80 Ul range
Slope : –20 dB/dec
A2'
0.5 Ul range
A1
f0
f1 f2
Bit rate
(bit/s)
f0
(Hz)
f1
(Hz)
9953.28M
0.1
15
f3
Modulation frequency (Hz)
f4
f5
f2
(Hz)
f3
(kHz)
f4
(MHz)
f5
(MHz)
A1
(UIp-p)
A2'
(UIp-p)
A2
(UIp-p)
A3'
(UIp-p)
A3
(UIp-p)
600
100
2
80
0.5
4
80
100
4000
0.5 UI range: 0.000 to 0.505 UIp-p (0.001 UIp-p steps)
80 UI range: 0.00 to 80.80 UIp-p (0.05 UIp-p steps)
4000 UI range: 0 to 4040 UIp-p (2 UIp-p steps)
1000 Ul range
Jitter amplitude [UIp-p]
Jitter generation
A3
A3'
A2
A3
A3'
A2
Slope : –20 dB/dec
20 Ul range
Slope : –20 dB/dec
A2'
0.5 Ul range
A1
f0
f1 f2
f3
Modulation frequency (Hz)
Bit rate
(bit/s)
f0
(Hz)
f1
(Hz)
2488.32M
0.1
15
f4
f5
f2
(Hz)
f3
(kHz)
f4
(MHz)
f5
(MHz)
A1
(UIp-p)
A2'
(UIp-p)
A2
(UIp-p)
A3'
(UIp-p)
A3
(UIp-p)
600
100
2
20
0.5
1
20
25
1000
0.5 UI range: 0.000 to 0.505 UIp-p (0.001 UIp-p steps)
20 UI range: 0.00 to 20.20 UIp-p (0.01 UIp-p steps)
1000 UI range: 0 to 1010 UIp-p (1 UIp-p steps)
Continued on next page
http://www.anritsu.com
177
2
DIGITAL LINK MEASURING INSTRUMENTS
Wander generation: 10 µHz to 10 Hz, 0 to 400,000 UIp-p (1 UIp-p steps), conform to ITU-T O.172
Wander generation
Wander amplitude [UIp-p]
A0
Slope : –20 dB/dec
A1
F0
F1
Modulation frequency (Hz)
F2
Bit rate
(bit/s)
F0
(µHz)
F1
(mHz)
F2
(Hz)
A0
(UIp-p)
A1
(UIp-p)
Steps
(UIp-p)
2488.32M
10
400
10
400,000
16,000
1
9953.28M
10
400
10
400,000
16,000
1
Jitter amplitude [UIp-p]
Measurement functions: UIp-p, UI + peak, UI – peak, UIrms, hit count, hit second, %F second, peak jitter
Measurement mode: Repeat, single, manual
Display: Current, last
Measurement interval: 1 to 99 s, 1 to 99 min, 1 to 99 h, 1 to 99 day
Jitter measurement: Conform to ITU-T O.172
20 Ul range
20
2 Ul range
2
0.2
F1
Bit rate
(bit/s)
9953.28M
Jitter amplitude [UIp-p]
F2
F2'
Modulation frequency (Hz)
Range
(UI)
2488.32M
Jitter measurement
F1'
F1
(Hz)
F1'
(Hz)
F3
F4
F2
(kHz)
F2'
(kHz)
F3
(MHz)
F4
(MHz)
20
2
—
100
—
100
1
20
10
—
10
—
1
20
2
—
100
—
400
4
80
20
10
—
40
—
4
80
F2
(Hz)
F2'
(Hz)
F3
(kHz)
4000 Ul range
4000
1000 Ul range
1000
2
F1
F1'
F2 F2'
Modulation frequency (Hz)
F1'
(Hz)
F3
Bit rate
(bit/s)
Range
(UI)
F1
(Hz)
2488.32M
1000
—
1
—
12.1
5
9953.28M
4000
1
—
12.1
—
20
UIp-p measurement
2 UI range: 0.000 to 2.020 UIp-p (0.001 UIp-p steps)
20 UI range: 0.00 to 20.20 UIp-p (0.01 UIp-p steps)
1000 UI range: 0 to 1010 UIp-p (1 UIp-p steps, 2488.32 Mbit/s only)
4000 UI range: 0 to 4040 UIp-p (2 UIp-p steps, 9953.28 Mbit/s only)
UI rms measurement
2 UI range: 0.000 to 0.714 UIrms (0.001 UIrms steps)
20 UI range: 0.00 to 7.17 UIrms (0.01 UIrms steps)
Filters:
Confirming to ITU-T O.172 and Bellcore GR1377
LP, HP0 + LP, HP1 + LP, HP1' + LP, HP2 + LP, HP + LP, HP' + LP, LP' (1000/4000 UI range only),
HP0 + LP' (1000/4000 UI range only)
178
Bit rate
(bit/s)
HP0
(Hz)
2488.32M
10
9953.28M
10
HP1
(kHz)
HP1'
(kHz)
HP2
(MHz)
5
—
10
20
For product ordering information, see pages 5 – 12
HP'
(kHz)
HP
(kHz)
LP
(MHz)
LP'
(kHz)
1
—
12
20
5
4
50
12
80
20
DIGITAL LINK MEASURING INSTRUMENTS
Off: Able to set non-modulated status∗
TDEV mask:
The 37 types of TDEV masks that are regulated by ITU-T, ETSI, ANSI, and Bellcore standards are available as default.
It is possible to add the wander modulation on the user specified TDEV mask.
Transient: It is possible to change the A (1 – e –63.7t ) phase by the timing of the start.
Signal off: It is possible to disconnect the standard signal.
Wander tolerance (TDEV) measurement: Evaluation by the various TDEV mask generations
Reference wander
generation
(Option 03)
2
Conform to ITU-T O.172
Reference input: 2.048M (HDB3, clock), 1.544M (AMI/B8ZS, clock), 64k + 8 kHz, 5 MHz, 10 MHz
Sampling frequency: 40 Hz, 1 Hz, 0.1 Hz (select by MX150002A)
Measurement range
P-P: 0.0 to 2E10 ns, +P/–P: 0.0 to 1E10 ns, TIE: 0.0 to ±1E10 ns
Measurement time: 10 to 1 x 108 s (max. 120,000 s; MP1570A only)
Wander application (requires MX150002A Wander Application Software)
TIE: Max. 1 x 108 s, MTIE: Max. 1 x 108 s, TDEV: Max. 1 x 106 s
Frequency offset: Measurement conforms to ANSI TI.105.09
Frequency drift rate: Measurement conforms to ANSI TI.105.09
Wander tolerance (TDEV) measurement: Evaluation by the various TDEV mask generations
Wander
measurement
(Option 02)
Other measurement
Jitter transfer, frequency measurement, jitter tolerance, jitter sweep, frequency sweep, wander sweep (with MP1570A)
Dimensions and mass
320 (W) x 100 (H) x 350 (D) mm, ≤10 kg (with MU150018A)
Power
≤250 VA
Temperature range
0˚ to +40˚C (operating), –20˚ to +60˚C (storage)
EMC
EN61326: 1997/A1, 1998 (Class A)
EN61000-3-2: 1995/A2, 1998 (Class A)
EN61326: 1997/A1, 1998 (Annex A)
LVD
EN610101-1: 1993/A2, 1995 (Installation Category ΙΙ, Pollution degree 2)
∗: Only non-modulated status can be set without this option.
Ordering information
Please specify model/order number, name and quantity when ordering.
Model/Order No.
MP1580A
Name
Main frame
Portable 2.5G/10G Analyzer
J1074
J1075
Standard accessories
AC power cord:
1 pc
Fuse, 6.3 A:
1 pc
Front cover:
1 pc
MP1580A operation manual
(Vol 1 Jitter/wander):
1 copy
MP1580A operation manual
(Vol 2 Remote control):
1 copy
Wander Measurement Application Software
(MTIE/TDEV) ∗Supplied with MU150018A-02: 1 pc
MX150002A operation manual (wander application)
∗Supplied with MX150002A:
1 copy
Semirigid cable Tx (for connection to MP1570A): 1 pc
Semirigid cable Rx (for connection to MP1570A): 1 pc
MU150018A
Plug-in unit
2.5G/10G Jitter Unit
MP1580A-01
MP1580A-02
MP1580A-04
MU150018A-02
MU150018A-03
Options
RS-232C
GPIB
VGA
Wander measurement
Wander reference output phase modulation
MP1580A-90
MU150018A-90
Maintenance service
Extended three year warranty service
Extended three year warranty service
F0093A
B0489
W1889AE
W1890AE
MX150002A
W1892AE
MP1570A
MP1570A-02
MP1570A-10∗
MP1570A-11∗
MU150000A
MU150001A
MU150001B
MU150001A/B-01
MU150001A/B-02
MU150001A/B-03
MU150017A
MU150017B
MP9677B
MU967701A
MP35A
Model/Order No.
J0661A
J0006
J0007
J0008
J0696A
J0696C
J0900E
J0162A
J0162C
J0845A
J0775D
J0776D
B0490
B0491
B0492
Name
Optical accessories
RS232C cable (cross cable with D-sub 9 pin connector
at both ends), 2 m
GPIB cable, 0.5 m
GPIB cable, 1 m
GPIB cable, 2 m
Coaxial cord (AA-165-500), 0.5 m
Coaxial cord (AA-165-1000), 1 m
Coaxial cord (AA-165-1500), 1.5 m
Balanced cord (Siemence 3P · Siemence 3P) 1 m
Balanced cord (Siemence 3P · Siemence 3P), 2 m
Balanced cord, (Bantam 3P · Bantam 3P), 6 ft
Coaxial cord (BNC-P620 · 3C-2WS · BNC-P620, 75 Ω),
2m
Coaxial cord (BNC-P-3W · 3D-2W · BNC-P-3W, 50 Ω),
2m
Joint plate (to mount MP1580A and MP1570A in a stack)
Soft case
Hard carrying case
∗: Must specify SDH (Option 10) or SONET (Option 11) when ordering depends on your systems. The option price is included in the MP1570A.
These two options can be installed simultaneously. But in this case, one
option is charged.
Peripherals
SONET/SDH/PDH/ATM Analyzer
GPIB (requires to combine with MP1580A)
SDH
SONET
2.5G/10G Unit (electrical for MP1570A)
Optical 10G Tx (1.55) Unit ∗2 km, for MP1570A
Optical 10G Tx (1.55) Unit ∗40 km, for MP1570A
2.5G (1.31, option for MP1570A)
2.5G (1.55, option for MP1570A)
2.5G (1.31/1.55, option for MP1570A)
Optical 10G Rx (Wide) Unit ∗For MP1570A
Optical 2.5G/10G Rx (Wide) Unit ∗For MP1570A
E/O, O/E Converter
Clock Recovery Unit (9953.28 MHz) ∗For MP9677B
Matching Transformer
(BNC-J/Siemence, C42334-A282, 75/120 Ω)
http://www.anritsu.com
179
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF