Jerome - 431X Manual

Add to my manuals
69 Pages

advertisement

Jerome - 431X Manual | Manualzz

JEROME

®

431-X

MERCURY VAPOR ANALYZER

OPERATION MANUAL

March 2005

Arizona Instrument LLC

1912 W 4 th

Street

Tempe, AZ 85281

(800) 528-7411

(602) 470-1414

Fax (480) 804-0656 http://www.azic.com email: [email protected] - General [email protected] - International [email protected] - Customer Support

700-0046-B1

JEROME 431-X

Mercury Vapor Analyzer

Operation Manual

PROPRIETARY RIGHTS NOTICE for use with the Jerome 431-X Mercury Vapor Analyzer. No part of this manual can be re produced or transmitted in any form or by any means, electronic, mechanical or otherwise .

This includes photocopying and recording or in connection with any information storage or retrieval system without the express written permission of Arizona Instrument LLC.

ALL RIGHTS RESERVED

©

Copyright 2005, Arizona Instrument LLC

Resisorb™ is a registered trademark of J.T. B aker Chemical Company.

Tygon™ is a registered trademark of Norton.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 2 of 69

Table of Contents

1.

FOR THOSE WHO CAN'T READ THE WHOLE MANUAL NOW................................................5

2.

INTRODUCTION ...................................................................................................................................7

3.

PRINCIPLE OF OPERATION..............................................................................................................8

Zero Air Filter ............................................................................................................................................9

4.

INSTRUMENT OPERATION ...............................................................................................................10

LCD Codes .................................................................................................................................................10

Daily Operations.........................................................................................................................................11

Sensor Regeneration...................................................................................................................................12

Zero Adjust.................................................................................................................................................13

Sample Mode..............................................................................................................................................14

Survey Mode ..............................................................................................................................................15

Operating on AC Power or Generator ........................................................................................................16

Operating on Internal Battery Power..........................................................................................................16

External battery power ...............................................................................................................................16

Charging Batteries ......................................................................................................................................17

Obtaining Maximum Battery Life ..............................................................................................................17

5.

MAINTENANCE.....................................................................................................................................18

Preventive Maintenance Calendar ..............................................................................................................18

Flow System ...............................................................................................................................................19

.25 inch Fritware Filter........................................................................................................................19

Internal Filters .....................................................................................................................................20

Replacing the Battery Pack ........................................................................................................................21

Setting the Input Voltage and Frequency ...................................................................................................21

Display Nanograms or Milligrams Per Cubic Meter ..................................................................................22

Changing the Fuse ......................................................................................................................................22

6.

CALIBRATION ......................................................................................................................................23

Verification of Calibration and Quality Control.........................................................................................23

7.

431-X TROUBLESHOOTING...............................................................................................................24

8.

JEROME 431-X TECHNICAL SPECIFICATIONS ...........................................................................27

Optional Communications Capability ........................................................................................................27

Instrument I/O Interface .............................................................................................................................28

Potential Interferences ................................................................................................................................29

9.

ACCESSORIES & MAINTENANCE PARTS .....................................................................................30

Spare Parts..................................................................................................................................................32

10.

FACTORY CALIBRATION SERVICE ...............................................................................................36

11.

APPENDIX A - 431-X FUNCTIONAL TEST KIT ..............................................................................37

Preparation..................................................................................................................................................37

Mercury Transfer........................................................................................................................................38

Vessel Disassembly ....................................................................................................................................38

Replacing Mercury .....................................................................................................................................39

Syringe Technique......................................................................................................................................39

Functional Test Procedure ..........................................................................................................................41

431-X Temperature Conversion Chart .......................................................................................................42

Functional Test Troubleshooting................................................................................................................43

AZI Customer Service 800-528-7411 or 602-470-1414 Page 3 of 69

12.

APPENDIX B - PERSONAL MERCURY DOSIMETER ...................................................................44

Dosimeter Technical Specifications ...........................................................................................................44

Before Sampling with the Dosimeter ........................................................................................................45

Dosimeter Analysis ....................................................................................................................................46

Non-Standard Flow Rates and Dilution Modules ......................................................................................48

Dilution Module Ratio Check.....................................................................................................................49

Loading the Dosimeter ...............................................................................................................................50

Dilution Module Ratio Calculations ..........................................................................................................51

Analysis with a Dilution Module................................................................................................................52

13.

APPENDIX C - INTERNAL DIP SWITCH SETTINGS ....................................................................55

14.

APPENDIX D - JEROME COMMUNICATIONS SOFTWARE.......................................................56

JCS Kit Contents ........................................................................................................................................57

System Requirements .................................................................................................................................57

Data Logger Option....................................................................................................................................57

15.

APPENDIX E - JEROME 431-X OPTION BOARD ...........................................................................58

Auto-Zero ...................................................................................................................................................58

Instrument Zeroing .....................................................................................................................................58

Timed Regeneration ...................................................................................................................................59

Auto-Sample...............................................................................................................................................60

4-20 MA Analog Output ............................................................................................................................60

SW101 Functions: ...............................................................................................................................60

Connection and Setup:.........................................................................................................................61

Fresh Air Solenoid......................................................................................................................................63

DC Power Operation ..................................................................................................................................64

DC Power Adaptor Kit, AZI P/N Y031 0902 .....................................................................................64

16.

APPENDIX F: CALCULATION OF DYNACALIBRATOR MERCURY

CONCENTRATION ...............................................................................................................................66

17.

WARRANTY ...........................................................................................................................................68

AZI Customer Service 800-528-7411 or 602-470-1414 Page 4 of 69

1. FOR THOSE WHO CAN'T READ THE WHOLE MANUAL NOW

This manual contains details that will optimize the results and the life of your instrument. Read and refer to the manual for complete details on operation, maintenance and troubleshooting, special voltage inputs and data output.

The Jerome 431-X is easy to operate and ready for use upon receipt from the factory.

• Remove the instrument from the packing material.

Retain all packaging materials for any future shipment of the instrument.

If the instrument is returned to AZI for any reason, it must be placed in the original packaging materials that have been tested and proven to be effective protection during shipment.

• Call AZI Customer Service at 800-528-7411 or 602-470-1414 for Return Material

Authorization (RMA) information prior to returning a unit.

• For all shipments, boxes and packing materials are available from AZI.

• Pack the Jerome instrument only in a Jerome shipping container.

AZI WILL NOT BE RESPONSIBLE FOR SHIPPING DAMAGE.

IF YOU RETURN THE INSTRUMENT IMPROPERLY PACKAGED OR

SHIPPED,

YOU SHOULD INSURE IT FOR FULL VALUE.

• Check for any damage and confirm receipt of all parts on your packing list. Contact Arizona Instrument Customer

Service at (800) 528-7411 or (602) 470-1414 if you have any questions.

• Press the ON button. The display should read 000 in less than one second.

A LO BAT message appears briefly in the upper left corner.

If the LO BAT message persists, recharge the battery.

See page 17.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 5 of 69

• Check the voltage setting (110 or 220 VAC) on the back of the instrument. Ensure that it is set to the correct voltage. If the pointer is not aligned to the local voltage, turn the selector to point to the correct voltage.

• Perform a sensor regeneration by following these steps:

Connect the line cord between the connector on the back of the 431-X and an AC power outlet.

Press the ON switch and then press the REGEN button.

The instrument will begin a 10 minute regeneration cycle, indicated by .H.H.H flashing on the display. Do not interrupt this cycle. For a complete description of

this process, see page 12.

If any error message, such as .P.P.P, appears on the display, see the

“Troubleshooting” section beginning on page 24.

• When regeneration is complete, zero the sensor by pressing the ZERO button and turning the zero adjust screw, located under the handle, until the display reads 0.

• The instrument is now ready to sample.

• To ensure the input to the instrument contains no Mercury Vapor or mercaptans, use a

Zero Air Filter, AZI P/N Z2600 3905. The Zero Air Filter cleans the air sample and should produce sample readings of less than 0.003 mg/m

3

. Therefore, use the filter to:

Equilibrate the instrument to temperatures that are higher or lower than the instrument.

Sample with filter installed until the reading is below 0.003 mg/m

3

.

Identify contamination within the unit.

Confirm the presence of Mercury Vapor when readings are elevated. Install filter and verify that the readings go down with filter installed.

• When the instrument measures Mercury Vapor, the zero display will be replaced with a value.

CAUTION

Do not adjust the ZERO after the instrument has measured Mercury Vapor or before the next regeneration. (Occasionally the display may drop to

.L.L.L (indicating low) between the initial zeroing and the first sample. It is

OK to readjust the ZERO if the instrument has not measured Mercury

Vapor.)

• The instrument is designed for ambient air monitoring. DO NOT allow the probe or the

instrument’s intake to be exposed to any liquid.

• The instrument is not explosion proof.

• Press the SAMPLE button to start a 10 second sampling cycle.

• Perform sensor regeneration after each day’s testing.

• Perform another sensor regeneration and re-zero the instrument before each day’s use.

• Perform sensor regeneration after 30 days of storage or inactivity.

Call AZI Customer Service, at (800) 528-7411 from the United States and Canada or (602)

470-1414 if you have any questions. If you prefer, you may send e-mail to [email protected]

AZI Customer Service 800-528-7411 or 602-470-1414 Page 6 of 69

2. INTRODUCTION

The Jerome 431-X Mercury Vapor Analyzer is an ambient air analyzer with a range of 0.001 to

.999 milligrams of mercury vapor per cubic meter (mg/m

3

Hg).

CAUTION:

The Jerome 431-X is for vapor use only.

DO NOT allow the probe or the instrument's intake to be exposed to any liquid, dust or other foreign material.

The 431-X is designed to be easy to operate for quick and accurate analysis of mercury vapor levels. It has few maintenance requirements. However, please take a moment to read this manual before attempting operation. If you have any questions about your application or operation, please call AZI Customer Service at (800) 528-7411 or (602) 470-1414 or e-mail [email protected]

for assistance.

431-X Features

• Automatic sensor regeneration when equipped with the communications option and used with the Jerome Communication Software (JCS) program and the Jerome data logger.

• Regulated film heat voltage during sensor regeneration. This allows the sensor to clean properly with voltages from 100-130 VAC (or 200-260 VAC).

• Survey mode can be locked in.

• DIP switch setting can change the digital meter readings from mg/m

3

Hg to nanograms

(ng) of Hg (see page 22).

• The Jerome 431-X can be operated from 100-130 or 200-260 VAC. To change the default

voltage range, refer to Setting the Input Voltage, page 21.

Accessories and Maintenance Parts

The Accessories and optional items available to support the 431-X are listed and pictured

beginning on page 30.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 7 of 69

Applications

• Ambient air analysis

• Odor nuisance monitoring

• Regulatory compliance

• Control room corrosion monitoring

• Quality control

• Scrubber efficiency testing

• Accuracy check for other Mercury Vapor monitors and control systems

• Mercury Vapor source detection

• Leak detection

• The Jerome 431-X can be operated from 100-120 or 200-240 VAC. To change the default

voltage range, See “Setting the Input Voltage” on page 21.

3. PRINCIPLE OF OPERATION

A thin gold film, in the presence of Mercury Vapor, undergoes an increase in electrical resistance proportional to the mass of Mercury Vapor in the sample.

When the SAMPLE button is pressed, an internal pump pulls ambient air through a scrubber filter and into the flow system.

After 2 seconds, the sample solenoid bypass opens, closing off the scrubber filter from the flow system.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 8 of 69

T he sample air passes through a filter (removing any acidic gases which interfere with the sensor's response to mercury) and is drawn over the gold film sensor. The sensor absorbs t he

Mercury Vapor. Nine seconds after starting, the sample solenoid bypass closes and the remainder of the sample is drawn through the scrubber filter and the flow system. The in strument determines the amount absorbed and displays the measured concentration on the digital meter in milligrams per cubic meter (mg/m

3

) of mercury. An internal DIP switch can be used to change the digital meter display from mg/m

3

to nanograms of mercury (see page 22).

T he instrument’s microprocessor automatically re-zeroes the digital meter at the start of each sample cycle and freezes the meter reading until the next sample cycle is activated, thus eliminating drift between samples.

D uring the sample mode cycle, bars on the LCD represent the percentage of sensor saturation.

Depending on the concentrations, approximately sixty-five samples containing 0.1 mg/m

3

Hg may be taken before the sensor reaches saturation. After absorbing approximately 500 nanograms of mercury, the sensor becomes saturated and needs to be cleaned. This is accomplished by a manually activated 10-minute heat cycle, or sensor regeneration tha t burns the mercury from the sensor. This mercury is absorbed on internal filters to prevent any external contamination. The solenoid bypass closes during the sensor regeneration cycle, causing the a ir to pass through the scrubber filter, providing clean air for the regeneration process. The flow system's final scrubber prevents contamination to the atmosphere from the desorbed mercury.

T he heat generated during the regeneration may cause some low level thermal drift. To ensure maximum sample accuracy, wait 30 minutes after regeneration before zeroing and using the instrument.

Zero Air Filter

The Zero Air Filter removes mercury vapor, mercaptans, and mercury vapor from the air sample.

Readings with the filter installed should be near zero.

B ecause air that is cooler than the instrument will cause low readings and warmer air will cause higher readings, the Zero Air Filter should be used to equilibrate the unit to ambient air.

Continuous sampling with clean air will not cause saturation of the gold film sensor but w ill equalize temperatures faster to allow accurate analysis to begin sooner.

T he Zero Air Filter can also be used to identify contamination within the instrument. If the readings do not reduce to near zero with the filter installed, contamination should be suspecte d.

If the readings do drop to near zero with the filter installed but elevate with the filter removed, the presence of Mercury Vapor at the sampled location is confirmed.

F or more information on the use of the Zero Air Filter, contact customer service at 1-800-528-

7411, 1-602-470-1414, or visit our web site at http://www.azic.com.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 9 of 69

LCD Codes

LCD CODE

000

.000

00.0

EXPLANATION

Ready to sample

No Mercury Vapor reading

No Mercury Vapor reading, display in nanograms

.H.H.H

.L.L.L

.P.P.P

.H.L.P

Sensor regeneration in progress (.H.H.H flashes)

Re-zero needed (refer to page 13)

Power cord required or low line power, <100 VAC (or 200 VAC)

(see pages 16 and 17, Changing the Fuse, if .P.P.P remains on after

the cord is connected.)

High line power, greater then 130 AVC in 110 operation or 260

VAC in 220 operation

.-

.--

.---

-.---

.LO BAT

.E.E.E

Recharge batteries (refer to page 17)

Same as LO BAT, automatically shuts off

.HL

DURING SAMPLING

Very high concentration has been detected. Refer to your safety policy for additional direction to confirm the concentrations."

0-25% sensor saturation

25-50% sensor saturation

50-75% sensor saturation

75-100% sensor saturation

0

H

L

DURING SAMPLING, USING SURVEY MODE

- Survey sampling (minus sign flashes continuously)

WHEN ZERO IS

DEPRESSED

Adjust to 0 only after sensor regeneration. It is normal for the display to read H after sampling has started.

Zero, ready to sample

High, turn Zero potentiometer counterclockwise

Low, turn Zero potentiometer clockwise

AZI Customer Service 800-528-7411 or 602-470-1414 Page 10 of 69

Daily Operations

Before each day's use of the Jerome 431-X, perform the following steps to verify proper instrument operation:

• Press the power ON button.

The digital meter displays 000.

(Disregard the digital meter's initial momentary reading.)

Recharge or replace the battery pack if the LO BAT indicator REMAINS ON.

Refer to “Charging Batteries” on page

17 and/or “Replacing the Battery

Pack” on page 21.

To ensure the instrument's electronics have stabilized, allow a 1-minute warm up before beginning the next step.

• Use the Zero Air Filter to equilibrate the instrument to ambient air temperature.

Install the Zero Air Filter in the instrument’s intake.

Sample continuously until the readings stabilize.

• Perform sensor regeneration. Refer to page 12 for the procedure.

• Thirty minutes after sensor regeneration is complete, zero the instrument.

NOTE: For maximum accuracy, such as when testing with the Functional Test

Kit, wait 30 minutes after the sensor regeneration cycle to re-zero the unit. For immediate use, the unit can be re-zeroed immediately after sensor regeneration.

• Press the SAMPLE button.

During the sample cycle, the digital meter displays bars (-, --, or ---) to indicate the amount of sensor saturation.

• At the end of the sampling cycle, read the digital meter.

The number shown on the digital meter is the Mercury Vapor concentration in mg/m3.

This value remains on the display until the next sample is taken.

The digital meter automatically zeroes at the start of each sample.

• At the end of each day's use, perform sensor regeneration as described in the next section.

DO NOT ALLOW MERCURY VAPOR TO STAY ON

THE GOLD FILM SENSOR OVERNIGHT.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 11 of 69

Sensor Regeneration

Sensor regeneration is needed to clear the 431-X sensor of any accumulated Mercury Vapor and to prolong the life of the sensor. This simple procedure should be done:

• At the beginning of the day on which the instrument is to be used.

• During the day when the sensor becomes saturated.

• At the end of the day before storing the instrument.

• At a minimum of 30 day intervals while the instrument is in storage.

CAUTION:

Ensure the voltage selector on the back of the instrument, near the power

cord inlet connector, points to the local AC power value. See “Setting the

Input Voltage” on page 21.

To clean and protect the sensor, the supplied AC power must be 100 to 120

VAC or 220 to 240 VAC, depending on the available power source.

Once sensor regeneration is initiated, DO NOT interrupt the cycle.

• Attach the power cord to the 431-X and plug it into AC power. AC power is required to thermally regenerate the sensor.

• Press the power ON button.

• Press the REGEN button.

The digital meter flashes .H.H.H for the duration of the 10-minute cycle and displays

.0.0.0 when the cycle is completed.

DO NOT INTERRUPT THIS CYCLE.

Wait until the cycle is completed before continuing with the next step.

A minimum 30-minute wait after the sensor regeneration cycle is complete ensures

maximum sample accuracy. However, the unit can be used immediately following the sensor regeneration if necessary. When the sensor regeneration is complete, press

ZERO and adjust the ZERO ADJUST pot until 0 appears on the display. Install the zero air filter in the intake and take several samples or lock the instrument into survey

mode (see page 15). After approximately one minute, stop sampling and check the

ZERO. Adjust to 0. Repeat sampling through the zero air filter until reading remains on 0.

NOTE: The digital meter will read .P.P.P after REGEN is activated if the power cord is not plugged in or if the instrument's fuse needs to be replaced. Connect the power cord, or if

necessary, replace the fuse. See “Changing the Fuse” on page 22.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 12 of 69

Zero Adjust

• To ensure air entering the instrument is clean, install the zero air filter in the instrument’s intake and sample until the readings stabilize.

• While pressing the ZERO button, turn the

ZERO ADJUST potentiometer (shown at right) using the trimmer tool until the digital meter reads 0.

If the LCD reads H, turn the ZERO

ADJUST counter-clockwise;

If the LCD reads L, turn the ZERO

ADJUST clockwise.

NOTE: A minimum 30-minute wait after the sensor regeneration cycle is complete ensures maximum sample accuracy. The unit can be used immediately following the sensor regeneration if necessary. When the sensor regeneration is complete, press ZERO and adjust the ZERO ADJUST pot until

0 appears on the display. Install the zero air filter in the intake and take several

samples or lock the instrument into survey mode (see page 15). After

approximately one minute, stop sampling and check the ZERO. Adjust to 0 if necessary. Repeat sampling through the zero air filter until sensor remains on

0.

NOTE: When ZERO is pressed, and depending upon internal configuration, a number between 00 and 100 may appear on the display instead of H, L, or O. If

the instrument is configured with an Option Bard, see APPENDIX E -

JEROME 431-X OPTION BOARD beginning on page 58.

CAUTION:

Do not turn the ZERO ADJUST potentiometer between samples.

Turn the ZERO ADJUST only after a sensor regeneration cycle otherwise invalid readings will result.

• Press the power OFF button and disconnect the power cord.

• The Jerome 431-X is ready for sampling.

CAUTION:

The Jerome 431-X is intended for vapor use only. DO NOT allow the probe or the instrument's intake to be exposed to liquids, dust or other foreign material. Moisture or liquids drawn into the instrument can damage the sensor and flow system.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 13 of 69

Sample Mode

This mode, used for standard operation, produces optimum accuracy (+/-5% at 0.100 mg/m

3

Hg) with the Jerome 431-X.

• Press the power ON button.

The digital meter displays 000. If the unit is set to display in ng, the digital meter displays 00.0. (Disregard the digital meter's initial momentary readings.) Recharge or

replace the battery pack if the LO BAT indicator REMAINS ON. Refer to pages 17

and/or 21 for the procedure.

• To ensure the instrument's electronics have stabilized, allow a 1 minute warm up before beginning the next step.

• Press the SAMPLE button.

During the sampling cycle, the bar (or bars) shown on the digital display indicate the

current percentage of sensor saturation. (Refer to Meter Display Codes, page 10, for

code descriptions.)

The bar (or bars) flash after 2 seconds and again after an additional 7 seconds. This flashing signals the opening and closing of the solenoid sample bypass. (See the

Principles of Operation on page 8 for details.)

• At the end of the 12 second cycle, read the digital meter.

The number shown on the digital meter is the mercury concentration in mg/m

3

(or ng).

This value remains displayed until the next sample is taken. The digital meter automatically zeroes at the start of each sample.

• When the sensor is completely saturated, the digital meter displays .8.8.8 instead of a value. No further operation is possible until a sensor regeneration is performed. (Refer to

page 12 for the Sensor Regeneration procedure.)

• Press the power OFF button when not in use. Install the zero air filter in the instrument intake during storage.

Sampling Notes

The Jerome 431-X is intended for vapor use only. DO NOT allow the probe or the instrument's intake to come in contact with liquids, dust or other foreign material. Moisture or liquids drawn into the instrument can damage the sensor and flow system.

The Jerome 431-X operates a minimum of 6 hours on a fully charged battery.

Use the probe (AZI P/N1400-2002) to locate mercury vapor in hard to reach places. Plug the probe directly into the instrument's intake.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 14 of 69

Survey Mode

The survey mode takes samples every 3 seconds automatically. Use this mode to locate mercury spills or to assess areas of potentially high mercury concentrations. Sampling in the survey mode is not as accurate. Due to the decreased sample volume, the accuracy of the instrument is reduced to +/- 20% @ .100 mg/m

3

.

• Press the power ON button.

The digital meter displays 000. If the unit is set to display in ng, the digital meter displays 00.0. (Disregard the digital meter's initial momentary readings.) Recharge or

replace the battery pack if the LO BAT indicator REMAINS ON. Refer to pages 17

and/or 21 for the procedure.

To ensure the instrument's electronics have stabilized, allow a 1 minute warm up before beginning the next step.

• Press and hold the SAMPLE button.

The instrument takes a normal 12 second sample, displays the concentration at the end of the cycle and then goes into the survey mode sampling every 3 seconds. The display flashes the measured concentrations at the end of each 3 second sample cycle.

• When you are finished surveying, release the SAMPLE button.

The final survey value remains displayed until the next sample is taken.

NOTE: Approximately 65 samples at .1 mg/m

3

may be taken before a sensor regeneration is required.

• To lock the instrument in a survey mode:

Hold the SAMPLE button down until the sensor status indicator bar(s) "_" begins flashing on the display.

Press the ZERO button, then release the SAMPLE button.

The pump should continue to run and the display should update every 3 seconds.

The instrument remains in the survey mode until one of the following occurs:

The sensor is saturated

A LO BAT (low battery) signal is encountered

An HL (high mercury level) is encountered

The instrument is turned OFF.

Press the power OFF button when not in use.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 15 of 69

Operating on AC Power or Generator

• For stationary use, the 431-X may be operated on AC power.

Operating the instrument on AC power at all times eliminates the need for the battery pack and its necessary maintenance.

The battery may be unplugged or removed completely whenever the instrument is operating on AC power.

• When a generator is used to power the Jerome 431-X, a high quality line conditioner or voltage regulator is required to ensure a pure sine wave and regulated voltage is applied to the instrument. The gold film sensor may be damaged by voltage that varies in amplitude or by surges, spikes, and/or noise on the power line. This is especially true during the

sensor regeneration.

Operating on Internal Battery Power

For portable use, the 431-X may be operated on Battery power.

• When you operate the instrument on battery power, please be aware of the following:

A fully charged battery pack, AZI P/N Z4000-0907, provides power for a minimum of six (6) hours of operation.

For operating more than six (6) hours, an extra fully charged battery pack is needed.

Complete battery recharging takes 14 hours. Refer to Charging Batteries on page 17.

The 431-X uses a rechargeable Nickel Cadmium (NiCad ) battery. Dispose of worn-out batteries properly when you are replacing the battery pack.

External battery power

A special version of the Jerome 431-X and a DC Power Kit are available to operate the instrument from a secondary DC source. The source may be a car/truck battery or a storage cell used in conjunction with solar panels.

Call AZI Customer Service at 800-528-7411, 602-470-1414, or e-mail [email protected] for additional information.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 16 of 69

Charging Batteries

• Press the power OFF button.

• Connect the AC power cord between the 431-X power receptacle and an AC power source.

Complete battery recharging takes 14 hours.

The 431-X contains a trickle charger so it may be continually plugged into an AC power source without damaging the battery pack.

• The battery pack may be charged outside the instrument with an optional AZI IDC Battery

Charger. (AZI P/N 4000-1011, for 115 VAC, P/N 4000-1012, for 230 VAC)

Obtaining Maximum Battery Life

There are certain inherent limitations to NiCad batteries. The primary limitation is a memory effect that occurs when the batteries are partially discharged and then recharged, repeatedly.

This memory leads to a drastic reduction in the usable battery life. To prevent this memory effect, periodically allow the battery pack to discharge completely, and then recharge the battery pack.

• To obtain maximum battery life, follow these three (3) steps:

At least once a month wait until LO BAT appears on the digital meter before recharging the battery pack.

Charge the battery pack when the LO BAT indicator comes on. Excessive discharge can damage the battery pack.

Before storing the instrument verify the power is OFF.

• When the batteries fail to hold a charge, the battery pack should be replaced.

Battery life under normal usage is approximately 1 year, depending on the number of charge and discharge cycles.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 17 of 69

5. MAINTENANCE

Preventive Maintenance Calendar

To keep the Jerome 431-X operating at peak performance, follow the maintenance schedule below as a guide. Since maintenance is more a function of application and amount of use rather than time, your requirements may be different from the listed schedule. Call AZI Customer

Service at 800-528-7411, 602-470-1414, or e-mail [email protected] for additional guidance for your environment and operation.

PART/COMPONENT MAINTENANCE CYCLE

REFER TO

PAGE

Charge batteries

At least once per month, after 1 month's storage, or when LO BAT appears

Weekly or as needed

Page 17

Change .25 inch fritware

Change internal filters and tubing

1

Replace zero air filter

2

Factory calibration

Calibration check

After 6 months of use or as needed

Page 19

Page 20

Annually

Annually

Page 23

Monthly or as needed

Appendix

A, Page 37

Page 21

Replace batteries

Annually or as needed.

The battery pack contains NiCad batteries.

Dispose of properly.

NOTE: Install the zero air filter into the instrument's intake during storage.

1

C/M filters contain Mallcosorb ™. For safety information see the supplier’s Material Safety Data Sheet (MSDS) or call AZI Customer Service at 800-528-7411, 601-470-1414, or e-mail [email protected] for a copy. Dispose of all filters properly.

2

Zero air filters and scrubber filters contain Resisorb™. For safety information see the supplier’s Material Safety

Data Sheet (MSDS) or call AZI Customer Service at 800-528-7411, 601-470-1414, or e-mail [email protected] for a copy. Dispose of all filters properly.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 18 of 69

Flow System

The Jerome 431-X's flow system is the crucial link between the sensor and the sample. For the instrument to perform correctly, the flow system must be properly maintained. The user maintainable components of this system are the intake filter (.25 inch fritware), a C/M Filter, two scrubber filters and connecting tubing.

Check the Preventive Maintenance Calendar on page 18, for a suggested schedule for changing

fritware and scrubber filters. The Tygon™ tubing in the system must be free of crimps for proper flow.

Part

Scrubber Filter

C/M Filter

Part Number

Z2600 3930

Z2600 3928

.25 inch Fritware Filter

Tygon™ Tubing - 1/8" I.D. (1')

2600 3039

2500 3001

.25 inch Fritware Filter

Replace the .25 inch fritware filter once each week or as needed. In dusty environments, the fritware filter may need to be replaced as often as once a day. Replacement .25-inch fritware filters are available from AZI, Consumable Sales at 800-528-7411 or 602-470-1414.

• Unscrew and remove the intake.

• Push the old fritware filter disc out of the intake with your trimmer tool.

• Avoid touching the new fritware disc with fingers. Use tweezers to insert the new fritware.

• Use the blunt end of the trimmer tool to seat the fritware disc firmly against the inner ledge of the intake.

• Screw the intake back on the Jerome 431-X.

CAUTION:

The stem coming from the instrument onto which the outer intake housing is attached must be securely held in place. If loose, the tubing inside the instrument can become twisted when the intake housing is replaced. It may be necessary to open the instrument and tighten the hold-down nuts inside the instrument. Call AZI Customer Service at 800-528-7411, 601-470-1414, or e-mail [email protected] if you have any questions

AZI Customer Service 800-528-7411 or 602-470-1414 Page 19 of 69

Internal Filters

• Replace the internal filters after six (6) months of use, or as needed.

• Press the power OFF button and unplug the power cord.

• Remove the two (2) side screws from the intake end of the instrument and open the case.

• Carefully disconnect the Tygon tubing from both ends of the filters and discard the old filters.

CAUTION:

Filters contain either Mallcosorb or Resisorb. Used filters will contain trace amounts of Mercury also. Use proper methods when disposing of used filters. Call AZI Customer Service at 800-528-7411, 601-470-1414, or e-mail [email protected] for a copy of the Resisorb MSDS or for other questions.

• Connect the new filters to the Tygon tubing, ensuring all straight hose barbs point toward the intake/pump corner of the case and elbow hose barbs point toward the sensor housing as shown in the illustration.

Push the Tygon as far as it will go onto the filter fittings.

• Push the filters into the mounting clips.

• Remove any crimps or twists in the tubing and ensure that tubing connections are secure.

If the tubing is loose, readings may not be accurate. Replace any tubing that has deteriorated due to heat and/or age.

• Close the case and replace the screws.

• Dispose of all filters in accordance with state and federal environmental regulations.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 20 of 69

Replacing the Battery Pack

• Press the power OFF button.

• Unplug the power cord.

• Remove the two (2) side screws from the intake end of the instrument and open the case lid.

• Disconnect the battery connector from the board.

• Loosen the two (2) captive screws holding the battery bracket and remove the bracket.

• Remove the old battery pack and replace with a new battery pack.

• Replace the battery bracket and tighten the captive screws.

• Connect the new battery connector to the board.

• Close the case and replace the two (2) side screws.

• Dispose of the old NiCad battery in accordance with state and federal regulations.

Setting the Input Voltage and Frequency

Instruments are factory set and calibrated to use the power setting requested on the order.

However, the voltage setting is easily changed to use either 110 or 220 VAC.

• Ensure the instrument is turned OFF and unplugged.

• Locate the voltage selector on the rear of the instrument.

• Insert a small screwdriver in the voltage selector slot and turn the selector until the arrow points toward your setting choice and a click is heard.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 21 of 69

• Remove the two screws toward the front of the instrument and open the lid.

• Locate the DIP Switch SW2 at the top of the main circuit board. This is the red on illustrated at right

• Set SW2 position #1 and #6 as follows.

Position #1 OFF OFF

Position #6 OFF ON

Display Nanograms or Milligrams Per Cubic Meter

The instrument is factory set to display mg/m

3

(milligrams per cubic meter) of mercury (.XXX).

For some applications, including dosimeter analysis, the instrument’s display can be converted to nanograms.

• Ensure the instrument is turned OFF.

• Remove the two screws near the front of the instrument and open the lid.

• Locate DIP Switch SW2 at the top of the main circuit board.

• Place Position #2 to OFF for nanograms display.

• Return Position #2 to ON for the normal milligram display.

Changing the Fuse

If the instrument display reads .P.P.P when the instrument is connected to AC power or when

REGEN is pressed, or if the battery will not charge, the fuse may need to be replaced. The AC line power could also be less than 100 VAC (220 VAC). Check the fuse with an ohmmeter and the AC line power with a voltage meter.

• Locate the power receptacle on the rear of the instrument.

• Insert a small screwdriver in the slot, located in the power receptacle, and gently slide the fuse compartment out.

• If the fuse in the open-sided clip is open, remove and discard it.

• Replace the discarded fuse with the spare fuse located in the slide-out spare fuse compartment.

• Replace the fuse compartment in the power receptacle.

As soon as possible, replace the spare fuse with another 1A, 250V, time delay fuse, AZI

P/N 5100 1012).

AZI Customer Service 800-528-7411 or 602-470-1414 Page 22 of 69

6. CALIBRATION

The Jerome 431-X's gold film sensor is inherently stable and does not require frequent calibration. The interval between calibrations depends upon the application and frequency of use; however, the recommended interval is every 12 months.

The Jerome 431-X has been factory calibrated using laboratory equipment containing NIST traceable permeation tubes. These permeation tubes have a rated accuracy of +/- 2%. In order to calibrate the Jerome 431-X, a sophisticated calibration system is required that ensures stability of the calibration gas source, eliminates any pressure in the calibration gas stream, and controls the temperature of the calibration environment.

We strongly recommend you take advantage of our calibration and maintenance service at

Arizona Instrument. Call Customer Service at (800) 528-7411 or (602) 470-1414 to arrange recalibration. A certificate of calibration is issued from AZI when your instrument is factory calibrated.

Verification of Calibration and Quality Control

The Functional Test Kit (FTK), AZI P/N Y431 0902, is used to determine if your instrument is within calibration tolerances between recommended annual factory calibrations. It allows you to have complete confidence in the sample results. This test verifies proper instrument operation through the introduction of a known concentration of Mercury Vapor into the Jerome analyzer.

THIS TEST DOES NOT CALIBRATE THE INSTRUMENT.

If your application requires frequent verification of instrument function, this test demonstrates the unit’s operation, calibration, and function. Recording FTK results in an instrument log provides a quality control/quality assurance record of instrument function between regular calibrations. If test results fall within the expected range, you may assume the instrument is functioning correctly.

See APPENDIX A - 431-X FUNCTIONAL TEST KIT on page 37 for more information about

the FTK procedures. Complete instructions for use are supplied with the test kit, AZI P/N Y431

0902.

To order the FTK, contact your AZI Sales Representative at (800) 528-7411 or (602) 470-1414.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 23 of 69

Symptom

Power Problems

Possible Cause

Unit does not turn ON. Unit turns on when power cord is plugged in. LCD displays 000 when instrument is operating on

AC power.

Discharged battery or

Dead battery.

Unit does not turn on when connected to AC power cord.

Open fuse.

Insufficient power.

Internal component failure.

Solution

Recharge battery for a minimum of

14 hours. Refer to page 17.

Replace battery. Refer to page 21.

Replace fuse. Refer to page 22.

Use a voltmeter to verify there is power to the AC outlet.

Call AZI Customer Service for information at 800-528-7411 or

602-470-1414.

Regeneration & Zero Problems

LCD displays .8.8.8. Sensor saturated. Do not attempt to adjust zero pot.

The sensor must be regenerated.

Refer to page 12 for information.

LCD displays .L.L.L when taking first sample.

Changes in temperature.

Readjust zero pot. See page 13 for

information.

LCD displays H at finish of sensor regeneration when zero is pressed.

Internal contamination may redeposit Mercury Vapor from flow system onto gold film sensor.

Zero adjust pot cannot be adjusted to 0.

Pot not turned sufficiently.

Sensor may be ruptured or pot may be broken.

Remove and replace fritware filter, intake filter disk, scrubber filters

and Tygon tubing. Refer to “Flow

System” on Page 19.

Check tubing for kinks or crimps.

Repeat regeneration cycle. Refer to

page 12.

1. Turn zero adjust up to 20 times to reach the end. Pot will

“click” softly.

2. If no “0”, turn pot slowly in opposite direction until display reads “0”.

3. If still unchanged, call AZI

Customer Service at (800) 528-

7411 or 602-470-1414.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 24 of 69

Sampling Problems

Airflow is restricted during the sensor regeneration cycle, causing possible permanent damage.

High erratic results.

High/erratic results

High/erratic results

Readings vary more than 0.05 mg/m3 when in survey mode.

Low response or erratic readings after a long period of non-use.

Kinks and crimps in the

Tygon tubing.

Periodically check the Tygon tubing inside the instrument. Refer to page

20.

Internal Mercury Vapor contamination.

1. Install zero air filter in intake and tighten intake nut. Press

SAMPLE button. After three samples, if readings are over

0.003 mg/m

3

, replace fritware filter and Tygon tubing. Refer

to page 19.

2. Perform sensor regeneration with the zero air filter in intake.

Refer to page 12. Retest if

necessary. Replace scrubber filters and Tygon™ tubing.

Refer to page 20.

Intake and internal filters may become clogged and need replacement when sampling in a dusty or humid area.

1. Open instrument and check for pinched, crimped or disconnected internal tubing.

2. In extreme conditions, an additional particle filter may be installed on the intake.

Loose connections to gold film sensor.

Place a zero air filter into the intake.

Place the instrument in survey mode. Move the unit as samples are being taken. Call AZI

Customer Service at 800-528-7411 or 602-470-1414 for assistance.

May need a second regeneration cycle.

1. Wait 30 minutes and perform another sensor regeneration.

2. Test with FTK. Refer to page

37.

3. If still unresponsive, call AZI

Customer Service at 800-528-

7411 or 602-470-1414 for assistance.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 25 of 69

False readings, may go to .8.8.8 or

.L.L.L.

Extremely cold or extremely warm air sampled into unit.

If sampling under these conditions, install zero air filter in intake.

Sample until display reads 0.003 mg/m3 or less. This equilibrates sensor temperature with the temperature of the sample air stream. Remove filter and take samples.

Miscellaneous Problems

Display reads .P.P.P when regeneration is attempted.

Display reads .E.E.E

Power cord not attached.

Blown fuse.

Line voltage less than 100

VAC (or less than 200 VAC for 220 unit).

Cycles dipswitch set incorrectly.

Check power cord for connection

Replace fuse. Refer to page 22.

Check line voltage settings. Refer

to page 21.

Check input cycle settings. Refer to

55.

If fuse and line voltage are OK, it may be circuit board adjustment or component failure. Call AZI

Customer Service at 800-528-7411 or 602 470-1414.

Very low battery.

Recharge battery. Refer to page 17.

Replace battery. Refer to page 21.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 26 of 69

8. JEROME 431-X TECHNICAL SPECIFICATIONS

Range 0.001 to 0.999 mg/m

3 standard deviation at 0.100 mg/m

3

Hg

Response time-sample mode

Response time-survey mode

Flow rate

Power requirements

Fuse

Internal battery pack

Operating environment

Case construction

Dimensions

Weight

Digital meter display

Certification

12 seconds

3 seconds

750cc/min (0.75 liters/min)

100-120 VAC, 50/60 Hz, 1 A, or

220-240 VAC, 50/60 Hz, 1 A

F1A 250V, 5mm X 20mm

Rechargeable Nickel Cadmium

0

E to 40EC, non-condensing, non-explosive

Aluminum alloy

15 cm x 33 cm x 10 cm (6" w x 13" l x 4" h)

3.18 kilos (7 pounds)

Liquid crystal display (LCD)

CE mark on 220-240 V~, 431-XE model only.

Optional Communications Capability

Alarm output

Dosimeter power output

Data output

30V DC, 100mA

For dosimeter analysis and regeneration

1. RS-232 Serial, Baud Rate 1200 for use with data logger, and/or Jerome communication program.

2. RS-232 Serial data format with 0 & 20mA current logic levels; Baud Rate 1200 (special industrial applications) and Analog 20 mA output.

"With Option Board" - See APPENDIX E - JEROME 431-X OPTION BOARD on page 58.

Analog output 0 to 2V or 4 to 20 mA

Auto sample interval 5, 15, 30, or 60 minutes

Auto regeneration interval 6, 24 or 36 hours

AZI Customer Service 800-528-7411 or 602-470-1414 Page 27 of 69

Instrument I/O Interface

The 431-X I/O port (25 pin D-sub) provides the following functions:

• Serial data communication

Interface type: RS-232C full duplex, DCE

Parameters: 1200 Baud, 1 start bit, 8 data bits, 2 stop bits, no parity

Pin assignments:

Pin 1 Protective ground

Pin 2 Data in

Pin 3 Data out

Pin 7 Data ground

• Serial current loop

Interface type: 20mA current loop, full duplex

Parameters - 1200 Baud, 1 start bit, 8 data bits, 2 stop bits, no parity

Pin assignments:

Pin 1 Protective ground

Pin 4 Data out (+)

Pin 5 Data in (+)

Pin 14 Data out (-)

Pin 16 Data in (-)

• Alarm output

Maximum voltage 30 VDC

Maximum current 100mAmp

Pin assignments:

Pin 9 Switched battery (+)

Pin 10 Alarm output (open collector, active low)

Pin 7 Battery ground (-)

Pin 23 Battery ground (-)

• Dosimeter Power

Pin Assignments:

Pin 22 Dosimeter Enable – 24 to 28 Volts AC

Pin 23 Battery Ground

Pins 12 $ 24 Tied – Dosimeter Power

Pins 13 & 25 Tied – Dosimeter Power

• Switched battery connection for data logger

Pin assignments:

Pin 9 Battery (+)

Pin 7 Battery ground (-)

Pin 23 Battery ground (-)

• Unswitched battery connection for external battery pack pin assignments

Pin assignments:

Pin 15 Battery (+)

Pin 19 Battery (+)

Pin 7 Battery ground (-)

Pin 23 Battery ground (-)

NOTE: Pins 6, 8, 11, 17, 18, 20 and 21 are non-standard and should not be connected.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 28 of 69

Potential Interferences

Potential interferences to the Jerome mercury vapor analyzers are rare and most of these can be eliminated with proper maintenance procedures. However, erroneously high readings can sometimes occur. Here are a few things to be aware of when using the instruments.

The gold film sensors used in the Jerome mercury vapor analyzers do not respond to the following compounds:

• Hydrocarbons

• CO, CO

2

, and SO

2

• Water vapor (Note that water vapor condensation on the gold film can cause irreparable harm to the sensor and must be avoided.)

The acidic gas filter, contained in the internal filter system, removes the following compounds that cause the gold film sensor to respond:

• Chlorine

• NO

2

• Hydrogen Sulfide (H

2

S)

• Most mercaptans (organic sulfur compounds or “thiols”)

In areas containing these highly volatile compounds, the filter can become quickly saturated. In such situations, it is recommended that these gases be allowed to dissipate before sampling for the less volatile, more persistent mercury vapor. A special filter designed to remove chlorine gas is available from Arizona Instrument and may be ordered as Chlorine Filter, AZI P/N Z2600-

3940. Collection of air samples with Jerome gold coil dosimeters for analysis by the Jerome mercury vapor analyzers will also eliminate interferences.

Ammonia in very high concentrations can cause an offgassing of accumulated acidic fumes from the internal acidic gas filter, resulting in positive readings on the instrument. In these cases, the ammonia odors are very strong. Again, either allow the vapors to dissipate or use the dosimeters. Filter replacement at regular intervals, or when unexpectedly high readings are encountered in areas of these potential interferents, may resolve these problems.

Volatile mercury compounds in general will cause the gold film to respond. Alkyl organic mercuries such as methyl mercury (and other “straight chained” compounds) are typically extremely volatile and change the electrical resistance of the gold film sensor. Any such responses should be considered “qualitative,” not quantitative. The instruments are designed and calibrated to elemental mercury vapor only.

Inorganic mercury salts such as mercuric chloride are not very volatile. They may, however, generate some minute level of elemental mercury vapor to which the instruments will respond.

This response, again, should be considered a qualitative response only.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 29 of 69

9. ACCESSORIES & MAINTENANCE PARTS

PART #

Y431 0901

ITEM DESCRIPTION

431 Accessory Kit (See pictures beginning on page 32)

Y431 0902

1400 3010 Tubing Adapter, 1/4” to 1/8”

2300 0001 Trimmer Tool

2600 3039 .25 Fritware

6000 4003 Line Cord, 115 VAC - USA and Canada

Alt. 200-0003 Line Cord, 220-240 VAC - England

Alt. 200-0008 Line Cord, 220-240 VAC - Europe

Z2600 3905 Zero Air Filter

431 Functional Test Kit (See pictures beginning on page 32)

A2600 0902 Stopper Assembly

A2600 0903 Syringe Assembly

A2600 0904 Mercury Vial

2600 0022 Syringe Needles, 22 Ga. Reusable

2600 0030 Calibration Vessel, Thermos

3200 0011 Septa (20)

Y431 0903 or

Y431 0904

Y990-0175 or

Y990-0176

431 Maintenance Kit (110 VAC) (See pictures beginning on page 32)

431 Maintenance Kit (220 VAC) (See pictures beginning on page 32)

2500 3001 1' of 1/8" Tygon tubing

2600 3039 .25 inch fritware

Z2600 3905 Zero Air Filter

Z2600 3928 C/M Filter

Z2600 3930 Scrubber Filter

Z4000 0907 Battery Pack Assembly

Dosimeter Analysis Kit (110 VAC applications)

Dosimeter Analysis Kit (220 VAC applications)

990-0177 Pocket Pump, calibrated to 5 ml/min

Converter, 220VAC to 110VAC, 200 WATT (supplied with

990-0159

Y990-0176 for locations using 220 VAC power)

X412 0901 Personal Mercury Dosimeter (2)

2500 3001 2' of 1/8" Tygon tubing

2500 3010 1' of 3/16" Tygon tubing

2100 6017 Dosimeter Lead Set

1300 0031 3/16” to 1/8” Reducer

Z2600 3905 Zero Air Filter

AZI Customer Service 800-528-7411 or 602-470-1414 Page 30 of 69

Y6100 0057

Jerome Data Logger

Includes the Jerome Data Logger and JCS Software Kit.

Y990-0175

Personal Dosimeter Kit

Includes the Pocket Pump, two mercury dosimeters, zero air filter, regeneration cable, connecting tubing and adaptor.

Y6100 0054

Y990-0170

Jerome Communication

Software Kit

[TO BE REPLACED BY THE

JEROME COMMUNICATION

SOFTWARE (JCS) ON CD

ROM]

Y411 0904

Hard Side Carry Case

Includes a molded case with die cut foam rubber inserts to hold the

Jerome 431-X and accessories.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 31 of 69

1400 0052

Soft Field Carrying Case

Hand/shoulder case with pockets for accessories.

Spare Parts

X421 0901 Personal Mercury Dosimeter

A2600 0902 Stopper Assembly

Z2600 3914 Septum Holder

3200 0011 Septa

A2600 0903 Syringe Assembly

2600 0022 Syringe Needles, 22 Ga.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 32 of 69

1400 2002 Probe

2300 0001 Trimmer

1300 0031 1/8 x 3/16 reducer

Z4000 0907 Battery Pack Assembly

Z2600 3905 Zero air filter

Z2600 3928 C/M filter

Z2600 3930 Scrubber filter

Z2600 3940 Chlorine Filter

1400 3010 Tubing adapter

PS 151 Tube Nut

2600 3039 .25 inch fritware

AZI Customer Service 800-528-7411 or 602-470-1414

Includes mounting hardware

Page 33 of 69

2500 3001 Tygon tubing 1/8" I.D. (1 foot)

2500 3002

Tygon tubing 1/16" I.D. (1 foot)

4000 1011

115 VAC IDC battery charger

Used to charge an uninstalled battery

4000 1012

230 VAC IDC battery charger

Used to charge an uninstalled battery

6000 4003 100-120 VAC Line Cord

AZI Customer Service 800-528-7411 or 602-470-1414 Page 34 of 69

Alternate –

200-0003

220-240 VAC Line Cord for

England

Alternate –

200-0008

220-240 VAC Line Cord for

Europe

Z2600 3911 10 to 1 Dilution Module

AZI Customer Service 800-528-7411 or 602-470-1414 Page 35 of 69

5100 1012 Spare Fuse

6000 1055 Jerome Communication Cable

6000 1045 Adaptor, RS-232, 9M/25F

For current prices and delivery information, call AZI Customer Service at

(800) 528-7411 or (602) 470-1414.

10. Factory Calibration Service

Service includes filter replacement, component testing, and instrument calibration to NIST traceable standards.

For scheduling and shipping authorization, call AZI Customer Service at

(800) 528-7411 or (602) 470-1414.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 36 of 69

11. APPENDIX A - 431-X FUNCTIONAL TEST KIT

If your application requires frequent verification of instrument functionality, this test will benefit you. If the test results fall within the expected range, you may assume the instrument is functioning properly. This test does not calibrate the instrument.

NOTE: Perform the functional test ONLY after a sensor regeneration.

The 431-X Functional Test Kit contains all accessories necessary to perform the functional test.

See the complete list on page 30 and verify that all the parts to the kit are present.

CAUTION

:

The vial and thermometer contain liquid mercury and are possible sources of mercury contamination. Follow the instructions for handling or transferring the mercury into the

Functional Test Kit Vessel carefully.

For safety information, see the supplier’s Material Safety Data Sheets (MSDS) or call AZI

Customer Service at 1-800-528-7411 or 1-602-470-1414 for assistance in obtaining the

MSDS.

Preparation

• Carefully unpack and inspect the parts of the kit.

• ENSURE that the mercury shipping container and mercury filled thermometer are not broken.

• In a ventilated area, preferably under a fume hood, remove the mercury vial from its shipping container.

• Place the functional test kit vessel and the mercury vial close to each other and open the mercury vial.

CAUTION:

The edge between the plastic case and the glass inner vessel of the functional test kit vessel are not sealed tight enough to prevent mercury from entering the area between the inner and outer vessels. ENSURE the mercury, transferred in the next step, does not contact the

seal where the glass and plastic portions join.

NOTE:

The vessel may be disassembled to transfer the mercury and better prevent contamination of the outer portion of the vessel. Instructions to disassemble the vessel

can be found on page 38.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 37 of 69

Mercury Transfer

• CAREFULLY pour the mercury into the center of the functional test kit vessel’s opening.

ENSURE that no mercury residue is on the outside of the vessel. See the supplier’s

Material Safety Data Sheets (MSDS) or call AZI Customer Service at 1-800-528-7411 or 1-602-470-1414 for clean-up instructions.

• INSTALL the stopper assembly into the functional test kit vessel carefully, to prevent breakage of the thermometer.

PRESS the stopper assembly into the vessel to achieve a good seal.

• USE the 431-X instrument to verify that the outside of the vessel is not contaminated and the mercury vapor emission level, if any, is below the OSHA TLV for mercury.

• ALLOW the kit to adjust to room temperature for at least two (2) hours before using.

The temperature range for the test is 18-22

EC. Avoid temperature fluctuations.

CAUTION:

Do not use the calibration vessel as a portable container. If the calibration vessel is upset or greatly agitated, mercury droplets will cling to the thermometer stem, the rubber stopper, the mouth of the calibration vessel and the needle guide.

Vessel Disassembly

CAUTION:

The inner portion of the vessel is made of glass. Handle the vessel carefully to prevent breakage.

• LOOSEN, BUT DO NOT REMOVE the base of the vessel. The base unscrews from the body.

• SET the vessel on a firm surface.

• HOLD the base stationary and unscrew the body from the base.

• HOLD the base and the inner glass vessel with one hand while removing the body and gasket with the other hand.

• After the mercury is transferred into the glass inner vessel, reassemble in the reverse order.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 38 of 69

Replacing Mercury

An oxide coating will form on the drop of mercury and will cause lower readings in your testing.

Gently swirl the vessel to disturb the outer oxidized surface of the droplet. If this does not restore higher readings, it may be necessary to replace the mercury.

• Carefully remove the stopper assembly from the calibration vessel.

CAUTION:

BE SURE NEEDLE GUIDE IS FREE OF LIQUID MERCURY.

• Carefully pour the mercury into a disposal vessel. Refer to Vessel Disassembly

Instructions on page 38.

Mercury can become trapped between the plastic calibration vessel and the glass innerliner.

• Replace the oxidized mercury with approximately ½ cc fresh mercury. (AZI P/N A2600-

0904)

Do NOT use the syringe for measuring liquid mercury. Dispose of oxidized mercury properly.

• Reassemble the calibration vessel.

• Reinstall the stopper assembly.

Syringe Technique

• Pull and hold the syringe plunger against the bar-stop.

• Verify that the black mark on the syringe plunger aligns with the 1cc mark on the syringe barrel.

If it does not, the holder assembly must be adjusted. Call AZI customer service at 602-281-

1745 or 800-528-7411 for assistance.

• Insert the needle into the needle-guide of the bottle stopper.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 39 of 69

• Operate the plunger two or three times to pump mercury vapor into the syringe. On the final stroke, pull and hold the plunger against the bar-stop.

• Holding the plunger against the bar-stop, remove the syringe from the bottle and move it to the septum attached to the instrument intake.

• Continue to hold the plunger against the bar-stop and insert the syringe needle into the septum.

• Press “SAMPLE” on the instrument.

• When the display flashes, release the plunger and allow gravity to feed the mercury vapor into the airstream. If the plunger stops, gently press it completely closed.

• Remove the syringe needle from the septum.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 40 of 69

Functional Test Procedure

NOTE: Perform the functional test ONLY after sensor regeneration.

• Allow the calibration vessel to remain stable at room temperature for at least 2 hours.

The temperature range for the test is 18

E - 22EC.

Temperature fluctuations during the test procedure will produce erratic results.

• Replace the .25mm fritware.

Refer to page 19 for instructions.

• Replace the septum on the septum holder assembly.

• Plug the tubing adapter end of the septum assembly into the instrument's intake and tighten the intake tube nut.

NOTE: To check for a tight seal, gently pull on the septum holder assembly. If it comes out of the intake, it may be necessary to remove the intake tube from the instrument and firmly press the tubing adapter through the intake. Tighten the intake tube firmly to the intake stem.

• Attach a zero air filter to the septum assembly.

• Press power ON.

• Take 3 samples.

If the average meter reading is greater than .005, stop here. The instrument may be

contaminated. See 431-X TROUBLESHOOTING on page 24.

If the average meter reading is less than .005, continue to the next step.

• Note the temperature of the calibration vessel.

• Press the SAMPLE button, wait 2 seconds and when the display flashes, inject 1 cc of

mercury vapor according to the syringe technique described on page 39. Be sure all

mercury vapor has been injected before the solenoid closes (second click and display flash).

CAUTION:

Carefully follow these instructions to minimize error.

• Record the meter reading.

• Repeat the instructions for mercury injection three more times.

The readings obtained for the last three 1cc injections should be within +/- 5% of each other.

• Refer to the Temperature Conversion Chart, page 42, for the acceptable range.

The average of the last three readings should fall within the range shown on the chart.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 41 of 69

If the average is within range, the JEROME 431-X is functioning properly.

• If the last three readings are not within +/-5% of each other,

Perform sensor regeneration. Press ZERO and turn the ZERO ADJUST (refer to page

12 and 13 for the complete sensor regeneration procedure).

Wait 1 hour before proceeding to the next step.

Repeat the mercury injection test procedure.

If the average of the last three readings is still not within range, refer to the section on

Functional Test Troubleshooting below.

431-X Temperature Conversion Chart

Temperature in

EC

Digital Meter Response

19

20

21

22

16

17

18

23

24

.091 to .123

.100 to .135

.108 to .146

.118 to .159

.129 to .174

.138 to .187

.151 to .204

.164 to .222

.177 to .240

AZI Customer Service 800-528-7411 or 602-470-1414 Page 42 of 69

Functional Test Troubleshooting

If you don’t achieve good results with the functional test procedure, check the following:

Results Solution

Typically too high Ensure the calibration vessel temperature is stable.

Be sure to inject the Hg vapor ONLY after the display flashes

(approximately 2 seconds after SAMPLE is pressed).

Too Low

Ensure there is no oxidation on the mercury drop in the calibration vessel. Gently swirl the mercury drop in the calibration vessel.

Replace if necessary.

Ensure the instrument’s intake is not blocked with foreign matter.

Check flow with a flow meter.

Ensure syringe is calibrated to 1cc. Use a new syringe needle.

Straighten or replace crimped or blocked internal tubing.

If you find the above does not solve your problem, please call AZI Customer Service at 800-528-

7411 or 602-470-1414.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 43 of 69

12. APPENDIX B - PERSONAL MERCURY DOSIMETER

The gold coil personal mercury dosimeter is a unique collection device for mercury vapor. The

Jerome 431-X Gold Film Mercury Vapor Analyzer and the Personal Mercury Dosimeter determine personal exposure levels and ambient air concentrations, as well as low levels of mercury in natural and stack gases

For personal sample collection, the dosimeter is worn as close to the wearer's breathing zone as possible and is connected by tubing to a pump usually worn on a belt. The dosimeter can also be used for multiple point area monitoring by placing a dosimeter, with pump attached, in various strategic locations.

We recommend a pump flow rate of 5 cc/minute for the most accurate results when sampling in an atmosphere that for eight hours may contain an average of .025 mg/m

3

Hg. If you are

considering using any other flow rate, see page 48, Non-Standard Flow Rates and Dilution

Modules .

After sample collection is completed, the dosimeter is inserted in the Jerome 431-X's intake. A dosimeter lead set is connected between the dosimeter and the DB-25 connector on the back of specially equipped instruments. The instrument supplies power to volatilize the accumulated mercury from the dosimeter to the gold film sensor. The Jerome 431-X determines the mass of mercury collected by the dosimeter in a 17 second analysis. The dosimeter is ready for immediate re-use after a mercury measurement has been performed.

Dosimeter Technical Specifications

Sensitivity

Precision

Less than 0.5 nanograms of mercury

15% RSD @ 0.100 mg/m

3

Hg

Accuracy 15% @ 0.100 mg/m

3

Hg

Recommended flow rate 5 cc/min (0.005 liters/min) for atmospheres up to 0.025 mg/m

3

Construction Nylon and glass housing a gold film coil

Dimensions

Capacity

Analysis Time

0.5" dia. x 4.5"

1000 nanograms of mercury

Less than two 2 minutes

AZI Customer Service 800-528-7411 or 602-470-1414 Page 44 of 69

Before Sampling with the Dosimeter

The personal mercury dosimeter adsorbs mercury vapor over a set period of time. Therefore, before each day's use it is necessary to ensure the dosimeter is mercury free. Perform the following steps to remove any accumulated mercury.

• Connect the system as shown.

Insert the dosimeter's large end in the 431-X's intake and gently tighten the intake tube nut to ensure an airtight seal.

Connect the Dosimeter Lead Set clips as shown, Short red lead to the rear and long black lead to the far end.

• Attach the power cord to the 431-X and plug it into AC power. AC power is required to heat the dosimeter.

• Attach the Dosimeter Lead Set 25 pin connector to the respective 25 pin communications port.

• Press the instrument’s power ON button.

• Press the instrument’s SAMPLE button.

The digital meter reading will appear in 15 seconds.

• Wait 60 seconds and press the SAMPLE button again.

The digital meter should display less than 0.005, verifying all mercury has been removed from the dosimeter coil.

• Wait 2 minutes to cool sensor to prevent false positive response.

• The dosimeter is ready for sample collection.

NOTE: For best results, dosimeter analysis should be performed immediately after collection. If analysis cannot take place immediately, place the red end caps on the dosimeter. For accurate results, perform dosimeter analysis no later than five days after sampling.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 45 of 69

Dosimeter Analysis

NOTE: Wait a minimum of 30 minutes after a sensor regeneration before continuing.

• Connect the system as shown.

Insert the dosimeter's large end in the 431-X's intake and gently tighten the intake tube nut to ensure an airtight seal.

Connect the Dosimeter Lead Set clips as shown, Short red lead to the rear and long black lead to the far end.

• Attach the power cord to the 431-X and plug it into AC power. AC power is required to heat the dosimeter.

• Attach the Dosimeter Lead Set 25 pin connector to the respective 25 pin communications port.

• Press the instrument’s power ON button

• Reading the dosimeter (dosimeter desorption)

Press the SAMPLE button.

The digital meter reading appears in 15 seconds.

Record the digital meter reading (include the decimal point).

Wait 30 seconds.

Press SAMPLE again and record this digital meter reading. Repeating the heating/reading process ensures complete release of mercury from the dosimeter coil.

• Add the two digital meter readings together. The sum of the two digital meter readings is the figure you will use in your calculations and is referred to as the meter response (MR).

NOTE: A third dosimeter desorption (pressing the SAMPLE button) should give a reading of

.005 mg/m

3

or less.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 46 of 69

Perform the following calculation to obtain the mercury concentration in mg/m

3

based on a time weighted average; or alternately, DIP switch #2 can be set to OFF and the digital meter will

display nanograms Hg directly (refer to diagram, page 55 ).

Working Formula and Units of Measure

Meter Re sponse X ConversionFactor

=

Pump FlowRate X Sampling Time

SampleConcentration

Where:

Meter Response = Total of the two digital meter readings in mg/m

3

Conversion Factor =

87.5 ng/mg/m

3

(a constant which changes the meter response to nanograms of Hg)

Pump flow rate = 5.0cc per minute (calibrated value of the supplied SKC Pocket Pump)

Sampling time = Duration of the sample in minutes

Sample concentration = In ng/cc mg/m

3

EXAMPLE: To calculate a time weighted average during an 8 hour period using the following values:

Meter response = 0.600 mg/m

3

(sum of the two meter response readings)

Conversion factor = 87.5 ng/mg/m

3

(constant)

Pump flow rate

Sampling time

= 5 cc/min

= 8 hours (480 min)

• Convert the meter response (the total of the two digital meter readings) to nanograms of mercury.

0.600 x 87.5 = 52.5 nanograms of Hg

• Determine the total volume of air sampled.

5 cc/min x 60 min/hr x 8 hr = 2400 cc

• Determine the Hg concentration (time weighted average) of the dosimeter.

52.5ng

2400cc

=

0.022

ng / cc of Hg

=

0.022

mg/m

3 of Hg

AZI Customer Service 800-528-7411 or 602-470-1414 Page 47 of 69

Check the sensor status after each dosimeter analysis.

IMPORTANT:

To prevent the loss of a sample, perform sensor regeneration as soon as the display shows “- - - -” (four bars).

This indicates that the sensor is 75-100 percent saturated.

• Seal the dosimeter with caps after analysis to prevent mercury contamination during storage.

• If your readings exceed 75 nanograms or more, try the recommendations described next or call AZI Customer Service at 800-528-7411 or 602-470-1414 for assistance.

Non-Standard Flow Rates and Dilution Modules

You may use a pump with a flow rate up to 50 cc/min, but be aware that there are certain limitations. If your pump flow rate exceeds 5 cc/min and your average dosimeter analysis produces nanogram levels of 75 or more, it may be easy to collect enough mercury to saturate the

431-X sensor. You thus risk over ranging your instrument and losing your collection data.

Higher flow rates may also impair the capture efficiency of the dosimeter.

We recommend that you drop your flow rate or use a dilution module* (AZI P/N Z2600-3911).

Lowering the flow rate to decrease the sample volume provides the greatest accuracy. Using a dilution module introduces an additional 15% inaccuracy to your analysis. As an alternative to the dilution module, sample for shorter time periods.

Dilution Module Specifications

Accuracy +/- 15% of 10:1 ratio

Input concentration range

Housing Nylon

Dimensions 1" w x 2.7" l x 3" h

The dilution module is factory set to a 10:1 ratio. The mass of mercury entering the dilution module is reduced by 90%, leaving a 10% (X10 dilution) concentration to be introduced into the

Jerome 431-X. Since this ratio can change slightly with use, it is important to occasionally determine the current dilution module ratio to ensure accurate results. For normal applications a

X8 to X12 ratio is recommended. The 431-X Functional Test Kit contains all accessories necessary to determine the current dilution module ratio.

Call Customer Service at 800-528-7411 or 602-470-1414 if you have questions about flow rates or applications.

*The dilution module contains Resisorb™, mercury vapor adsorbent. For safety information, see the supplier’s Material Safety Data Sheets (MSDS) or call AZI Customer Service at 1-800-528-

7411 or 1-602-470-1414 for assistance in obtaining the MSDS.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 48 of 69

Dilution Module Ratio Check

The 10:1 dilution module was manufactured and calibrated to produce a 10:1 ratio in the sample delivered to the instrument. Over time the ratio may change slightly. To accurately determine the exact dilution offered by the 10:1 dilution module, perform the following tests and calculate the exact dilution ratio. The calculated ratio can then be used in all final calculations where the dilution module is used.

NOTE: Wait a minimum of 30 minutes after sensor regeneration before starting this procedure.

Direct 431-X Readings:

• Connect the instrument, septum holder assembly and zero air filter, with arrow pointing to the instrument, as shown.

• Press the Jerome 431-X power ON button.

• Inject 1 cc of mercury saturated vapor into the septum, according to the S yringe

Technique described on page 37,

Appendix A.

• Make 3 additional 1 cc injections sample readings and record the displayed values

(include the decimal points).

• Average the results of the last 3 injections.

• Remove the septum assembly and zero air filter from the instrument.

• Connect the 10:1 dilution module to the instrument, connect the septum assembly and zero air filter to the dilution module.

• Inject 1 cc of mercury vapor as above into the septum.

• Make three additional injections and average the three readings.

Divide this result into the average from the three direct injections.

• Use the result as the dilution module ratio in your dosimeter analysis.

Most Accurate Method

Perform the above test, however attach the dosimeter, septum holder assembly and zero air filter to the sampling pump that will be used. The technique is described in the next section.

Collection efficiencies should be approximately 100%.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 49 of 69

Loading the Dosimeter

• Connect your pump, dosimeter, septum holder assembly and zero air filter together with 1/8" an d 3/16" Tygon™ tubing as shown to the right.

• Turn on the pump.

• Inject 10 cc’s of Hg into the septum, 1 cc at a time.

A total of ten, 1cc injections, one after another.

• Wait 30 seconds after the last injection then turn off the pump.

• Remove the dosimeter, septum assembly and zero air filter from the pump.

• Connect the instrument, dilution module, dosimeter, zero air filter and dosimeter lead set as shown.

• Attach the power cord to the 431-X and plug it into AC power.

AC power is required to heat the dosimeter.

• Press the Jerome 431-X Power ON button and then press the SAMPLE button.

The digital meter reading appears in approximately 15 seconds.

• Record the digital meter reading (include decimal point). Wait 60 seconds, then press

SAMPLE again and record this reading.

• Repeating the heating process ensures complete release of mercury from the dosimeter coil.

• Add the two digital meter readings together.

The sum of the two digital meter readings is the figure you will use in your calculations and is referred to as the meter response (MR).

• Repeat this procedure two more times.

• Average the three meter responses you obtained in this section.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 50 of 69

Dilution Module Ratio Calculations

• Multiply the average obtained for Direct 431-X Readings by 10 (the number of 1 cc injections).

Divide this result by the average obtained in the section “Loading the Dosimeter” on

page 50.

• Use the result as the dilution module ratio in your dosimeter analysis.

EXAMPLE:

Direct 431-X readings

0.102 mg/m

3

0.103 mg/m

3

0.104 mg/m

3

0.103 mg/m

3

average

Loading the dosimeter

0.100 mg/m

3

Step 1 (above)

0.103 x 10 = 1.030 mg/m

3

Step 2 (above)

(1.030 ) / (0.111 mg/m

3

= 9.4

Dilution module ratio 9.4:1

NOTE:

For normal applications a X8 to X12 ratio is recommended. If your ratio is not within this range, call Customer Service at 800-528-7411 for assistance.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 51 of 69

Analysis with a Dilution Module

NOTE:

Wait a minimum of 30 minutes after sensor regeneration before starting this procedure.

• Connect the system as shown in the figure below.

• Attach the power cord to the 431-X and plug it into AC power.

AC power is required to heat the dosimeter.

• Press the Jerome 431-X power ON button and then press SAMPLE button. The digital meter reading appears in 12 seconds.

• Record the digital meter reading (include the decimal point). Wait 30 seconds, then press

SAMPLE button again and record this reading.

Repeating the heating process ensures complete release of mercury from the dosimeter coil.

• Add the two digital meter readings together.

The sum of the two digital meter readings is the figure you will use in your calculations and is referred to as the meter response.

• The following calculations will provide the mercury concentration in mg/m

3

based on a time weighted average.

(Meter Response converted to nanograms (Ng) of mercury) X

Dilution Module Ratio

= Sample Concentration

Sample Volume

• Alternately, DIP switch #2 can be set to OFF and the digital meter will display nanograms

Hg directly.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 52 of 69

MR

87.5

(meter response)

ng/mg/m

3 total of the two digital meter readings in mg/m

3 conversion factor, a constant which changes the meter response to nanograms of Hg

DM

dilution module ratio

the ratio determined on page 51

SV

(sample volume) pump flow rate (in cc/min) multiplied by sample time (in minutes) ng/cc = mg/m

3

Sample concentration

EXAMPLE:

Assume the following values.

Meter Response (MR) 0.600 mg/m

3

(sum of the two meter response readings)

Conversion Factor 87.5 ng/mg/m

3

(constant)

Dilution Module Rate 9.4

Pump Flow Rate 5 cc/min

Sampling time 8 hours = 480 min

Sample volume 5 cc/min X 480 min = 2400cc

A time weighted average during an 8 hour period is calculated by:

(

0.600

mg/m

3

) (

87.5

ng/mg/m

3

)

=

0.021875

ng/cc

2400 cc

• Convert the meter response (the total of the two digital meter

readings) to nanograms of mercury.

The MR (0.600) multiplied by the conversion factor (87.5) equals nanograms of mercur y

0.600 x 87.5 = 52.5 nanograms of Hg

Determine the actual mass of Hg collect ed by the dosimeter.

Nanograms of mercury multiplied by the dilution module ra tio.

52.5 nanograms x 9.4 = 493.5 nanograms

Determine the total volume of air sampled.

The pump flow rate multiplied by 60 min/ h r multiplied by 8 hours.

5 cc/min x 60 min/hr x 8 hr = 2400 cc

Determine the Hg concentration (time w eighted average) of the dosim eter.

The mass of Hg collected by the dosimeter divided by the total volume of air sampled.

Check the sensor status after each dosimeter analysis.

• Seal the dosimeter with tubing after analysis to prevent

excessive mercury contam ination during storage.

IMPORTANT:

Perform a sensor regeneration as s oon as the meter display shows “- - - -” (four bars) to prevent the loss of sample.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 53 of 69

Dosimeter Reference Chart 431-X

= Indicates the optimum meter response for that concentration with the corresponding volume

431-X reading in mg/m

3

0.5 0.429 0.857

Estimated concentration in mg/m

3

0.1 0.086 0.171 0.343 HL HL HL HL HL HL

0.05 0.043 0.086 0.171

0.025 0.021 0.043 0.086

0.343

0.171

HL HL HL HL HL

0.343 HL HL HL HL

0.005 0.004 0.009 0.017 0.034 0.069 0.137 0.274 0.549 HL HL

0.001 LOW LOW 0.003 0.007 0.014 0.027 0.055 0.110 0.219 0.439

Volume at 5ml/min 75 150 300 600 1200 2400 4800 9600 19200 38400

Use the following formula for calculating the concentration of mercury in air:

(Meter Response) X 87.5

(Flow Rate of Sampling Pump) X Time

=

Concentrat ion (mg/m

3

)

The following table shows the relation of flow rate and time to total volume collected. cc/min Total volume collected in cc

100 6000 12000 24000 48000 72000 144000 288000

7200 28800 43200 86400 259200

9600 14400 28800 57600

4800

2400

7200 14400 28800

3600 7200 14400

1 2 4 8

HOURS

12 24

AZI Customer Service 800-528-7411 or 602-470-1414 Page 54 of 69

13. APPENDIX C - IN TERNAL DIP SWITCH SET TINGS

Main c ircu it board RED DIP switches (SW2) top, center of the instrument’s main circuit board.

The 431-X both 50 Hz also pro pr an vides d e s t a unregulated f ilm hea t (100-1 2 0/200-2 4 0 volt and 110-130/22026 0 vol t ra n ges). T h e two ranges are av ailab le to redu ce the e f fec ts of chr onic l ow l lta ge.

Note:

The ranges are doubled when the AC line selector switch is set to the 220V position.

DIP switch positions 1 and 6 must be properly set.

The

DIP 1 DIP 6 Function

OFF

OFF

OFF

ON

60 Hz regulated film heat (100-130/200-260VAC)

50 Hz regulated film heat (102-130/205-260 VAC)

ON OFF 50/60 Hz pres et film heat (110-130/220-260 VAC)

O N ON 50/60 Hz preset film heat (100-120/200-240 VAC)

Regulated film heat should normally be used (D IP 1 O FF) excep t in the fe w c ases wh ere extremely irty l ine v oltage cond itio ns ma y exi st. Th ese c ondit ions might be found w he re large motors are being controlled or other situations may exist where the voltage may vary outside the

10013 0 VAC range with regularity. In those ca ses the tw o preset heat ranges will allow some d egree of satisfactory operation.

Switch

Number

Normal

Position

Action

3

4

5

ON Displays relative (not true)voltage during regen (0-255)

OFF Display L-O-H when “zero” button pressed

ON

ON

Display 00-99 when “zero” button pressed

Locks into 0-10mg/m

3

range (survey mode)

AZI Customer Service 800-528-7411 or 602-470-1414 Page 55 of 69

14. APPENDIX D - JEROME COMMUNICATIONS SOFTWARE

The Jerome Communications Software (JCS) is used with 431-X Mercury Vapor Analyzers that fe ature the communications configuration option.

• The JCS allows the user to program t he instrument for unattended monitoring and to d ownload recorded data stored in the Jerome data logger.

• Automatic sampling can be initiated every one (1) to sixty (60) minutes with programmable audible alarm levels.

The Jerome Communications Software (JCS) operates with the Jerome 431-X Mercury Vapor and Jerome 631-X Hydrogen Sulfide A nalyzers that have the “Communications Configuration” option installed. The software can control ins trument sampling for unattended continuous operation, collect data, graph this data in real tim e and perform statistical analysis.

T he software can also program the Jerome Data Logger, AZI P/N 6100-0010. This optional accessory enables data storage during manual sampling or portable automatic sampling withou t bein g attached to a computer. The data logger initiates automatic sampling, triggers alarms and st ores data. The logged data may then be downloaded to the computer when it is convenient.

T he JCS is m enu-dr iven and easy t o use. Each display screen is design ed for clarity with selfe xplana tory menu optio ns, such as “Operate

Instrument” or “D isplay Stored Data.” Select menu o ption s usin g eithe r a m ouse or a track ball pointing device or a s tandard k ey board. The user creates records, or f iles, f or com puter storage of collected d ata. Data is easily retrieved for later viewing, g raphing, printing or editing with spreadsheet or word processing software, (not provided). Data can be used for ongoing record keeping or for fulfilling local regulatory requirements.

Before using this software, famili arization with the o peration of the Jerome Hydrogen Sulfide Analyzer or Mercury Vapor Analyzer is important.

Al so, prior to installation of this software you should be fam iliar with the personal computer and operating syst you have any questions about how to proceed, call AZI

Custom er Service at (800) 528-7411 or (602) 470-1414 or send an e-m ail to [email protected]

for assistance.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 56 of 69

JCS Kit Contents

• One disk containing the Jerome Communication Software

• Jerome Communication Cable, AZI P/N 6000 1055

• Cable Adaptor, SR-232, 9M/25F, AZI P/N 6

• User’s manual

000 1045

System Requirements

Jerome 431-X with the “Communications Option.”

Windows 98 Second Edition, ME, NT, 2000 or XP

At least one free serial port

One free USB port

O ptional equipment:

Jerome Data Logger, AZI P/N 6100-0010 (to capture data without a computer nearby )

Data Logger Option

The software can also program the Jerome Data Logger (AZI P/N 6100 0010) used with the

Jerome analyzer. The computer programs the data logger that then attaches to the DB-25 connector on the rear of the instrument. The data log ger initiates automatic sampling, triggers alarms and stores data. This optional access ory enables portable automatic sampling without a dedicated computer.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 57 of 69

15.

APPENDIX E - JEROME 431-X OPTION BOA RD

Pro per use of this board requires that the base instrument be f ully functional and set correctly for the intended operation.

Auto-Zero

With the option board installed, the 431-X has a limited auto-zero function. This function cann ot be disabled and is transparent to the user. The instrument can be manually zeroed as described i n

“Zero Adjust” on pag e 13. However, if the instrument is to be operated by personnel not

familiar with the procedure or if it is operated unattended, the auto-zero function should satisfactorily zero the unit after each sensor regeneration.

Instrument Zeroing

The Jerome 431-X has essentially three ways to zero the sensor reading before samples are taken if the option board is installed.

• The instrument automatically re-zeroes between samples so that each sample is a unique reading. To take a sample, simply press the SAMPLE button.

• The manually adjusted zero, using the switch on the top of the 431-X is used to reestablish a baseline between the reference and sensor gold films only after a sensor

regeneration

. This zero is manually adjusted by pressing the ZERO button and turning the potentiometer on the top of the instrument until the display reads 0. Adjust only after

sensor regeneration;

it is normal for H to be displayed after sampling.

• The 431-X option board provides an auto-zero feature following regeneration that is invisible to the user.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 58 of 69

• In some cases, the instrument cannot resume sampling after regeneration. .L.L.L appears on the display when the ZERO button is pressed and the error message “manual bridge a djust needed” is added to the notes column of the JCS text file wh en the JCS is used. If t his problem persists, it may be necessary to re-set the auto-zero.

• When necessary to re-adjust the auto-zero point:

Turn the instrument off.

Make a note of the original DIP switch settings of SW100 on the option board.

On red DIP switch on the control board, SW2, turn DIP switch 4 to ON.

Set the switches on the option board’s blue DIP box, SW100, to 1,2,6 OFF; 3,4,5 ON.

Turn the instrument ON.

Press the Zero button and adjust the potentiometer on top of the instrument until the numbers read between 5 and 7.

Switch option board DIP #1 OFF and ON three times, leaving it ON.

While pressing the ZERO button, turn the potentiometer on the option board until the numbers read between 5 and 7. Note the display will flicker one digit.

Return all switches to their original position.

NOTE:

The higher the auto-zero number, the lower the sensor capacity and the more sensor regenerations are needed.

Timed Regeneration

If the unit is to be operated unattended for extended periods, AZI recommends that the sensor be regenerated regularly. Operation under JCS or data logger control automatically regenerates saturated sensors. The option board can control regeneration on a regular basis, every 6, 12 or 24 hours.

T he regeneration intervals are set through a combination of sw itch settings as shown in the following table:

------- SW100---------

Switch #1 Switch #2

REGENERATION

Interval (Hrs.)

OFF OFF

ON OFF

OFF ON

ON ON

OFF

6

12

24

AZI Customer Service 800-528-7411 or 602-470-1414 Page 59 of 69

Auto-Sample

If a data logger is either not connected or is operating in the manual sampling mode, the follow ing automatic sampling rate s may be selected with option board’s SW100 dip switch settin gs:

Dip switch settings

Sampling frequency

3 4 5

The switches have no effect if a data logger is connected and operating in automatic sampler mode as programmed throu gh the JCS.

4-20 MA Analog Output

The analog output signal at pin 18 of the 25 pin connector can be configured to provide the instrument's native mode 0-2 Volt output or the optional 4-20 mA output by setting the option b oard jumper (SW101) to the "V" position for voltage, or the "I" position for current. (Pin 23 is the ground pin for the analog output function. Pin 18 is positive with respect to the ground p in).

• The 0-2 Volt output circuit can drive loads of 10 k ohms or higher.

• The 4-20 mA output is a passive transm itter and requires the connected receiver to supply be tween 10 and 28 volts o f excitation potential.

The analog output signal is based on the entire range (0-.999 mg/m3).

Note that ne ither analog output circuit is flo ati ng. The negative terminals of both circuits are connected to the instrum ent's common ground buss.

SW101 Functions:

V = 0-2V analog output

I = 4-20 ma analog output

AZI Customer Service 800-528-7411 or 602-470-1414 Page 60 of 69

Jerome 431-X ins truments shipped after early 1995, are capable of providing 0-2 volts analog output. Instruments shipped before that tim e can be upgraded by a firmware update and a djustment.

Instruments that are capable of 0-2 volt output can be upgraded to the 4-20 mA output with the addition of an option board upgrade. T his must be installed at the factory.

Connection and Setup:

• 0

-2 vo lt dev ices c onnec t as s hown in Figure 1. If the instrume nt inc ludes an o ption board, be sure its analog jump er (S W101 ) is s et to t he “V” posi tion.

• The 4-20 mA active receivers connect as shown in

Figure 2. The active receiver contains a voltage source to power the loop current. The receiver must have an isolated input circuit. That is, it must not be connected to ground or to a voltage source referenced to ground. Be sure that jumper SW101 is set to the “I” position before power is applied.

• 431X reading

Current

0.000mg/m3

1.000mg/m3

4mA

20mA

• The 431X current formula is:

16mA

∗ Display Reading + 4mA = (Current)

Exam ple: ou tput urrent of,

16mA

∗ 0.1 +4mA = 5.6mA

Page 61 of 69

• The 4-20 mA passive receivers do not contain a voltage source to power the loop current. They requir e the addition of a separate isolated power supply. Typically a supply that delivers 15 to 20 volts DC at 50 mA is sufficient. Wire these as i n

Figure 3. Note that some 12-volt DC wall transformers (as used on portable equipment) may de liver 15 to 20 volts when they are lightly loaded.

The Digi-Key T5.9-PIP-ND is a commonly available example of a 12 volt 200mA supply that will deliver around 18 volts nominal when loaded below 20 mA.

Be sure that both the power supply used and the passive receiver are floating (not connected to earth ground). If either is not floating, the circuit will not work and damage may occur.

• Ensure that jumper SW101 is set to the “I” position before power-up.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 62 of 69

Fresh Air Solenoid

An ex ternal three-way solenoid can be used to provide fr esh air or conditioned air during sensor regene ration. This may be necessary if the sample stream lacks molecular oxygen. A low curren t six volt DC solenoid, connected between p ins 19 and 11 of the 25 pin rear panel conne ctor, will be energized during the regeneration cycle if the option board SW100 switch 6 is placed in the OFF position.

If need ed, the circuit may be built from the following com ponents and configured as shown in the fol lowing diagram. It will only function if the option board is installed in the 431-X instrum ent.

Requi red Parts:

1 sole noid, 6volt 3way

1/8" to 1/16" tubing adaptor

1/2 " clamp, adhesive mount

1/8" tu be to instrument adaptor

3" 1/8" tubing

A/R 1/16" tubing

1 25 pin male DB-25 connector

Solder-cup style

Suggested Part

Angar P/N 407569

Any

Any

Any

Tygon Formula R3603

Tygon Formula R3603

Similar AZI P/N

1300 1004

1300 0025

6000 0013

1400 3010

2500 3001

2500 3002

AMP 747912-2 None *

1 connector hood AMP 749626-2 None *

* These are types not stocked by AZI, but should be available overnight from many AMP stocking distributors such as Digi-Key Corporation. There are multiple suitable alternatives such as Radio Shack’s 276-1547 and 276-1549.

1/8" TUBE

1/16" TUBE

1/16" TUBE

TO INSTRUMENT

TO SAMPLE

SOURCE

PIN 1

REAR OF CONNECTOR,

CONNECT TO PINS 11 AND 19

MOST INSTRUMENTS

1/16" TUBE

TO FRESH AIR

SOURCE

PIN 1

REAR OF CONNECTOR,

CONNECT TO PINS 19 AND 21

631-X SHIPPED IN 1994 ONLY

AZI Customer Service 800-528-7411 or 602-470-1414 Page 63 of 69

DC Power Operation

Instruments with the 431-X option board modification can be used with any +12 VDC source for continuous operation, if the AZI Power Inverter Kit, P/N Y031 0902 is installed along with the option board. To preserve the life of the DC power source, usually a car or truck battery, th e power inverter will switch on automatically to supply the AC necessary for regeneration only.

The external switch on the inverter should always be OFF to preserve battery life during normal sampling.

To work with the power inverter kit, place option board SW100 DIP Switch #6 to the ON position.

W hen the instrument starts a regeneration with option board SW100 DIP Switch #6 ON, the instrument sends a signal to close the relay on the DC Power Adaptor, AZ I P/N 1000 0089, mounted between the da ta logger and the instrument. Th is switches the power inverter ON using the inverter’s internal switch.

NOTE:

When this mode is enab led, the in strument does NOT check for 115 VAC for the regeneration. I f there is no AC powe r to the instrument, and a regeneratio n is initiated , the instrument will fla sh .H.H.H (rather than .P.P.P), however the se nsor will not heat, no r will the sensor be cleaned.

DC Power Ad aptor Kit, AZI P/ N Y031 0902

• The DC power adaptor kit consists of:

DC Power Adaptor, P/N 1000 0089

DC Power Inverter, P/N 4000 1021

DC Power Cable Assembly, P/N 6000 1093

Installation

• Ensure that the instrument’s option board switches are set correctly for the intended operation with the option board’s SW100 DIP Switch #6 set to “ON” for DC operation.

• Mount the interface board to the rear of the instrument. Tighten the mounting screws.

• Place or mount the DC/AC power inverter in a secure position near the instrument.

• Connect the cable from the DC/AC power inverter to the matching connecter on the interface board. Note that the connectors are keyed to prevent improper connection.

• Plug the instrument’s AC power cord into the power inverter and connect it to the instrument.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 64 of 69

E nsure that the inverter’s power switch is in the “OFF” position. LEAVE the power switch in the “OFF” position at all tim es. The interface board will activate the inverter when necessary. If t he inverter power switch is placed in the “ON” position, it will cause a continuous drain on the external 12-volt power system.

• Remove the screws from the rear cover of the inverter and remove the cover.

• Place the wires from the external DC source (battery) and the wires from the DC pow er cable through the holes in the end plate.

• Connect cables from

the external 12-volt power source and the DC power cable assembly

• to the appropriate positive (+) and negative (-) terminals on the back of the inverter and tighten the hold down screws.

• Connect the yellow wire from the DC power cable to the “REMOTE” terminal on the power inverter and tighten the hold dow n screw.

Reinstall the cover.

If the external 12volt lines are not pow ered, power them now. (Connect them to the battery)

• Connect the instrument’s AC power cord between the instrument and the front o f the power inverter.

• Turn the instrument “ON.”

Press the “REGEN” switch on the instrument. Inverter operation can be verified in either of two ways:

Immediately after pressing “REGEN” the inverter will intermittently

“sing.” This tone slowly becomes nearly continuous and then ends after 64 seconds.

If the area is noisy, use a voltmeter or test lamp to verify that approxima tely 115 volts is present for about 64 seconds, starting when the “REGEN” switch is pressed.

• Allow the instrument to complete its regeneration before turning it off.

With the instrument turned off, complete the installation (i.e. connect data logger, communications cables, or other devices and ensure that the DIP switches for the instrument and option board are set correctly.

Page 65 of 69

MERCURY CONCENTRATION

UNIT DEFINITIONS

R (L*atm/mol/°C)

MV

NTP

(L/mol)

MW

HG

(g/mol)

K

HG

(L/g)

P

SITE

(mm Hg)

T

SITE

(°K)

FCarrierTop

NTP

(SLM)

FCarrierBot

NTP

(SLM)

PermRate (ng/min)

0.082057 Universal gas constant

24.453 Molar volume at NTP

200.59

0.122

730

298

Molar weight

Molar constant at site

Atmospheric pressure

Air temperature in rotometer

0.173

0.173

745

Carrier flow at NTP, top float

Carrier flow at NTP, bottom float

Permeation rate at 100°C (+/- 2%)

FDilutionTop

CAL

(SLM)

FDilutionBot

CAL

(SLM)

FDilutionTop

NTP

(SLM)

FDilutionBot

NTP

(SLM)

FTotalTop

NTP

(SLM)

Calibrated dilution flow at NTP, top float

Calibrated dilution flow at NTP, bottom float

Corrected dilution flow at NTP, top float

Corrected dilution flow at NTP, bottom float

Corrected total flow at NTP, top float

FTotalBot

NTP

(SLM)

ConcTop

NTP

(ng/m

3

)

Corrected total flow at NTP, bottom float

Concentration at NTP, top float

ConcBot

NTP

(ng/m

3

) Concentration at NTP, bottom float

Normal Temperature and Pressure (NTP) is at 25°C and 760 mm Hg

CALCULATIONS

Equations

Perfect gas law

Correct dilution flow rotometer to NTP

P * V = n * R * T

V / n = (R * T) / P

MV = (R * T) / P

FDilution

NTP

= FDilution * sqrt (P

NTP

/ 760) * sqrt (298 / T

NTP

)

Correction for carrier flow is not required because flow is not set by rotometer

Total mass flow at NTP

FTotal

NTP

= FCarrier

NTP

+ FDilution

NTP

Concentration at NTP

Conc

NTP

= PermRate / FTotal

NTP

AZI Customer Service 800-528-7411 or 602-470-1414 Page 66 of 69

Flowmeter Calibration Data At Site Total Mass Flo w At NTP Concentration At NTP

FD ilutionTop

NTP

(SLM)

FDilution Bot

NTP

(SLM )

FTotalTop

NTP

(SLM)

FTotalBot

NTP

(SLM)

ConcTop

(mg/m

3

NTP

)

ConcBot

(mg/m

3

NTP

)

0.000 0.000 0.173 0.173 4.306 4.306

0.308 0.247 0.481 0.420 1.550 1.772

0.615 0.495 0.788 0.668 0.945 1.115

1.004 1.098 1.177 1.271 0.633 0.586

1.392 1.701 1.565 1.874 0.476 0.397

1.777 2.348 1.950 2.521 0.382 0.295

2.162 2.995 2.335 3.168 0.319 0.235

2.555 3.610 2.728 3.783 0.273 0.197

2.947 4.225 3.120 4.398 0.239 0.169

3.315 4.869 3.488 5.042 0.214 0.148

3.683 5.513 3.856 5.686 0.193 0.131

4.054 6.156 4.227 6.329 0.176 0.118

4.425 6.799 4.598 6.972 0.162 0.107

4.781 7.461 4.954 7.634 0.150 0.098

5.137 8.123 5.310 8.296 0.140 0.090

5.501 8.799 5.674 8.972 0.131 0.083

5.865 9.474 6.038 9.647 0.123 0.077

6.219 10.172 6.392 10.345 0.117 0.072

6.572 10.869 6.745 11.042 0.110 0.067

6.931 11.566 7.104 11.739 0.105 0.063

7.289 12.264 7.462 12.437 0.100 0.060

7.642 13.005 7.815 13.178 0.095 0.057

7.994 13.746 8.167 13.919 0.091 0.054

8.351 14.527 8.524 14.700 0.087 0.051

8.707 15.308 8.880 15.481 0.084 0.048

9.037

16 .069 9.210 16.242 0.081 0.046

9.367 16.830 9.540 17.003 0.078 0.044

9.707 17.635 9.880 17.808 0.075 0.042

10.047 18.441 10.220 18.614 0.073 0.040

10.405 19.300 10.578 19.473 0.070 0.038

10.763 20.159 10.936 20.332 0.068 0.037

AZI Customer Service 800-528-7411 or 602-470-1414 Page 67 of 69

17. WARRANTY

Arizona Instrument LLC (seller) warrants to buyer that Jerome products delivered pursuant to this agreemen t shall, at the tim e o f delivery, an d fo r a period of o ne ( 1) year. The reafte r (the

In ternal B attery Pa ck, whe re applicable, is warranted for a period of ninety [90] days only), to be free from defects in materi al or work manshi p and shal l confo rm to selle r's spec ifications or such other specifications as sell er has agre ed to in writing. Seller's obligations with respect to claims under this warranty shall be limited, at seller's option, either to the repla cement of defecti ve or non-conforming product or to an appropriate credit fo r the pu rchase pric e there of subject to the p rovision s of seller 's Warr anty Polic y as am ended from time to time, sa id Polic y being incorporated herein by reference.

Returned products under warranty cl aims w ill be shipped to seller’s pla nt by bu yer at buy er's expense and shall b e accom panied by a statement of th e reaso n for the r eturn a nd an appr oved

Return Material Authorization Number issued by seller. Buyer remains respon sible for p ayment fo r produ cts not ac cepted f or warran ty adjus tment, han dling c osts, and f reight c osts assoc iated therewith.

Notwithstanding the foregoing, no warranty shall be e nforcea ble in the event th at produc t has been subjected to environmental or stress testing by buyer or any third party without writ ten approval of seller prior to such testin g. Furt her, no wa rranty shall be enforceable if the al leged defect is found to have occurred because of misuse, ne glect, i mproper in stallati on, repair, alteration, accident, or improper return handling procedure by buyer.

Discontinued product is warranted only for a credit or replacement at seller's option.

THE EXPRESS WARRANTIES GRANTED ABOV E SHAL L EXTEN D DIR ECTLY T O

B UYER AND NO T TO BU YER'S C USTO MERS, A GENTS , OR REP RESEN TATIVES AND,

EXCEPT FOR WARRANTY OF TITLE, IS IN LIEU OF ALL OTHER WAR RANTIES ,

W HETH ER EXPR ESSED OR IMPL IED, I NCLUDI NG ANY IMPLIE D WAR RANTIE S OF

FITNESS FOR A PARTIC ULAR PU RPOS E AND M ERCHA NTABILITY, SUCH OTH ER

WARRANTIES BEING SPECIFICA LLY D ISCLAIM ED BY SELLER . IN N O EVENT

SHALL EITHER PARTY'S LIABILITY FOR ANY BREAC H OR AL LEGED BREACH OF

THIS AGREEMENT EXCEED THE TOTAL EXTENDED PRICE OR PRICES SHOW N ON

UNFILLED ORDE RS, NO R SHAL L EITHER PART Y BE L IABLE F OR AN Y SPECIA L,

INCIDENTAL OR CONSEQUENTIAL DAMAGES RESUL TING FR OM BR EACH O R

A LLEGE D BREA CH.

Notwithstanding the foregoing, if any produc t covered by order(s) placed hereunder is designated as “developm ental,” “prototype” or “experimental,” no warranty whatsoever except a warranty of title to component materials, will be applicable thereto and buyer shall indemnify seller for any claims for liability asserted seller in connection therewith.

The foregoing state the entire liability of seller in connection with products supplied hereunder.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 68 of 69

TRADEMARK AND COPYRIGHT PROTECTION

Jerome Instrument, Arizona Instrument, AZI and the stylized AZI are all registered trademarks o f

Arizona Instrument LLC.

Copyright 1990-2005, Arizona Instrument LLC. All Rights Reserved.

Resisorb™ is a registered trademark of J.T. Baker Chemical Company

Tygon™ is a registered trademark of Norton.

Arizona I nstrument LLC

Jerome 431-X Mercury Vapor Analyzer Operation Manual

Part Number 700-0037

March 2005

If you have any questions regarding the operation of this instrument, p lease call our toll free n umber (800) 528-7411. Internationally, call (602) 470-1414 or fax (602) 470-1888.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 69 of 69

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement

Table of contents