SureSelect Target Enrichment System for Illumina Paired

Add to my manuals
90 Pages

advertisement

SureSelect Target Enrichment System for Illumina Paired | Manualzz
SureSelectXT Target
Enrichment System for
Illumina Paired-End
Sequencing Library
Illumina HiSeq and MiSeq
Multiplexed Sequencing
Platforms
Protocol
Version 1.8, October 2014
Before you begin, view hands-on
videos of SureSelect procedures at
http://www.agilent.com/genomics/protocolvideos.
SureSelect platform manufactured with Agilent
SurePrint Technology
For Research Use Only. Not for Use in Diagnostic
Procedures.
Notices
© Agilent Technologies, Inc. 2014
Warranty
No part of this manual may be reproduced
in any form or by any means (including
electronic storage and retrieval or translation into a foreign language) without prior
agreement and written consent from Agilent Technologies, Inc. as governed by
United States and international copyright
laws.
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection
with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Manual Part Number
G7530-90000
Edition
Version 1.8, October 2014
Agilent Technologies, Inc.
5301 Stevens Creek Rd 
Santa Clara, CA 95051 USA
Acknowledgment
Oligonucleotide sequences © 2006, 2008,
and 2011 Illumina, Inc. All rights reserved.
Only for use with the Illumina sequencer
systems and associated assays.
Technology Licenses
Technical Support
For technical product support, contact your
local Agilent Support Services representative.
For US and Canada, call (800) 227-9770
(option 3,4,4). For other countries, find your
support center telephone numbers at
www.agilent.com/chem/contactus.
Or send an e-mail to: 
[email protected]
Notice to Purchaser
SureSelect capture libraries and reagents
must be used within one year of receipt.
The hardware and/or software described in
this document are furnished under a
license and may be used or copied only in
accordance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Software and technical data rights granted to
the federal government include only those
rights customarily provided to end user customers. Agilent provides this customary
commercial license in Software and technical data pursuant to FAR 12.211 (Technical
Data) and 12.212 (Computer Software) and,
for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial
Items) and DFARS 227.7202-3 (Rights in
Commercial Computer Software or Computer Software Documentation).
Safety Notices
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated conditions are fully understood and met.
WA R N I N G
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and
met.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
In this Guide...
This guide describes the recommended operational procedures
to capture the genomic regions of interest using the Agilent
SureSelectXT Target Enrichment Kit for Illumina Multiplex
Sequencing. This protocol is specifically developed and
optimized to use biotinylated RNA oligomer libraries to enrich
targeted regions of the genome from repetitive sequences and
sequences unrelated to the research focus, specifically adjusted
to provide high performance with SureSelect.
This guide uses an optimized protocol for Illumina paired-end
multiplexed library preparation.
1
Before You Begin
This chapter contains information (such as procedural notes,
safety information, required reagents and equipment) that you
should read and understand before you start an experiment.
2
Sample Preparation (3 µg DNA Samples)
3
Sample Preparation (200 ng DNA Samples)
These chapters describe the steps to prepare the DNA samples
for target enrichment.
4
Hybridization
This chapter describes the steps to prepare and hybridize
samples.
5
Addition of Index Tags by Post-Hybridization Amplification
This chapter describes the steps to amplify, purify, and assess
quality of the sample libraries. Samples are pooled by mass
prior to sequencing.
6
Reference
This chapter contains information on reagent kit content,
SureSelect index sequences, and alternative equipment that can
be used with this protocol.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
3
What’s new in 1.8
• Support for SureSelectXT Focused Exome and Exome Plus 1
capture libraries.
What’s new in 1.7
• Support for SureSelectXT Clinical Research Exome and
SureSelectXT Inherited Disease capture libraries.
• D1000 ScreenTape and Reagents replace D1K ScreenTape
and Reagents. High Sensitivity D1000 ScreenTape and
Reagents replace High Sensitivity D1K ScreenTape and
Reagents.
What’s new in 1.6
• Support for V5+lncRNA SureSelect capture libraries.
• Support for 200-ng DNA Samples input option.
• Details added for storage of on-bead captured DNA samples.
What’s new in 1.5
• Support for SureSelect Human All Exon V5, V5 Plus, and
V5+UTRs capture libraries.
• More details provided for the Agilent 2200 TapeStation for
DNA quantitation and qualification.
• Option to hybridize for 16 hours.
What’s new in 1.4
• On-bead PCR procedure replaces in-solution PCR procedure
for post-hybridization.
• Support for Covaris E-series Sample Preparation system and
SonoLab 7 software.
4
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
Content
1
Before You Begin
7
Procedural Notes 8
Safety Notes
8
Required Reagents 9
Optional Reagents
11
Required Equipment
12
Optional Equipment
13
2
Sample Preparation (3 µg DNA Samples)
15
Step 1. Shear DNA 19
Step 2. Purify the sample using Agencourt AMPure XP beads
Step 3. Assess quality (optional) 22
Step 4. Repair the ends 25
Step 5. Purify the sample using Agencourt AMPure XP beads
Step 6. Add ‘A’ Bases to the 3' end of the DNA fragments 27
Step 7. Purify the sample using Agencourt AMPure XP beads
Step 8. Ligate the paired-end adaptor
29
Step 9. Purify the sample using Agencourt AMPure XP beads
Step 10. Amplify adaptor-ligated library
31
Step 11. Purify the sample with Agencourt AMPure XP beads
Step 12. Assess quality and quantity 35
3
Sample Preparation (200 ng DNA Samples)
21
26
28
30
34
39
Step 1. Shear DNA 43
Step 2. Assess quality (optional) 45
Step 3. Repair the ends 46
Step 4. Purify the sample using Agencourt AMPure XP beads 47
Step 5. Add ‘A’ Bases to the 3' end of the DNA fragments 48
Step 6. Purify the sample using Agencourt AMPure XP beads 49
Step 7. Ligate the paired-end adaptor
50
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing5
Contents
Step 8. Purify the sample using Agencourt AMPure XP beads
Step 9. Amplify adaptor-ligated library
52
Step 10. Purify the sample with Agencourt AMPure XP beads
Step 11. Assess quality and quantity 55
4
Hybridization
54
59
Step 1. Hybridize the library
62
Step 2. Prepare magnetic beads 68
Step 3. Select hybrid capture with SureSelect
5
51
69
Addition of Index Tags by Post-Hybridization Amplification
71
Step 1. Amplify the captured library to add index tags
72
Step 2. Purify the sample using Agencourt AMPure XP beads 76
Step 3. Assess quality
77
Step 4. Assess the quantity of each index-tagged library by QPCR
(optional)
80
Step 5. Pool samples for Multiplexed Sequencing 81
Step 6. Prepare sample for cluster amplification 82
6
Reference
83
SureSelect Reagent Kit Content
84
Other Reagent Kits Content 86
SureSelectXT Indexes for Illumina
87
Alternative Capture Equipment Combinations
6
88
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
SureSelectXT Target Enrichment System for Illumina Paired-End
Sequencing Protocol
1
Before You Begin
Procedural Notes 8
Safety Notes 8
Required Reagents 9
Optional Reagents 11
Required Equipment 12
Optional Equipment 13
Make sure you have the most current protocol. Go to genomics.agilent.com and
search for G7530-90000.
Make sure you read and understand the information in this chapter and have
the necessary equipment and reagents listed before you start an experiment.
NOTE
This protocol differs from the Illumina Multiplexed Paired-End sequencing manual and
other SureSelect protocols at several steps. Pay close attention to the primers used for
each amplification step and the blocking agents used during hybridization.
NOTE
Agilent cannot guarantee the use of the SureSelect Target Enrichment kits with
non-Agilent protocols nor provide technical support for non-Agilent protocols to process
samples for enrichment.
7
1

Before You Begin
Procedural Notes
Procedural Notes
• To prevent contamination of reagents by nucleases, always wear
powder-free laboratory gloves and use dedicated solutions and pipettors
with nuclease-free aerosol-resistant tips.
• Maintain a clean work area.
• Do not mix stock solutions and reactions containing gDNA on a vortex
mixer. Instead, gently tap the tube with your finger to mix the sample.
• When preparing frozen reagent stock solutions for use:
1 Thaw the aliquot as rapidly as possible without heating above room
temperature.
2 Mix briefly on a vortex mixer, then spin in a centrifuge for 5 to
10 seconds to drive the contents off of walls and lid.
3 Store on ice or in a cold block until use.
• In general, follow Biosafety Level 1 (BL1) safety rules.
Safety Notes
CAUTION
8
Wear appropriate personal protective equipment (PPE) when working in the
laboratory.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
Before You Begin
Required Reagents
1
Required Reagents
Table 1
Required Reagents for Library Prep and Post-Hybridization Amplification
Description
Vendor and part number
For use with 2100 Bioanalyzer:
DNA 1000 Kit
Agilent p/n 5067-1504
High Sensitivity DNA Kit
Agilent p/n 5067-4626
QPCR NGS Library Quantification Kit (Illumina GA)
Agilent p/n G4880A or equivalent
For use with 2200 TapeStation System:
D1000 ScreenTape
Agilent p/n 5067-5582
D1000 Reagents
Agilent p/n 5067-5583
High Sensitivity D1000 ScreenTape
Agilent p/n 5067-5584
High Sensitivity D1000 Reagents
Agilent p/n 5067-5585
Herculase II Fusion DNA Polymerase 
(includes dNTP mix and 5x Buffer)
200 reactions (processes 100 XT libraries)
400 reactions
Agilent 

p/n 600677
p/n 600679
Nuclease-free Water (not DEPC-treated)
Ambion Cat #AM9930
Agencourt AMPure XP Kit
5 mL
60 mL
450 mL
Beckman Coulter Genomics
p/n A63880
p/n A63881
p/n A63882
1X Low TE Buffer (10 mM Tris-HCl (pH 8.0), 0.1 mM EDTA)
Life Technologies p/n 12090-015 or
equivalent
Qubit dsDNA HS Assay Kit or
Life Technologies p/n Q32851
Qubit dsDNA BR Assay Kit
100 assays, 2-1000 ng
500 assays, 2-1000 ng
1000 assays, 2-1000 ng
Life Technologies p/n Q32850
Life Technologies p/n Q32853
Life Technologies p/n Q33130
Qubit assay tubes
Life Technologies p/n Q32856
Buffer EB (10mM Tris-Cl, ph 8.5)
Qiagen p/n 19086
Ethanol, 100% for molecular biology
Sigma-Aldrich p/n E7023
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
9
1

Before You Begin
Required Reagents
SureSelect Reagent Kit*
Table 2
Reagent Kits
16 Reactions
96 Reactions
480 Reactions
Reagent Kit, HSQ
G9611A
G9611B
G9611C
SureSelectXT Reagent Kit, MSQ
G9612A
G9612B
G9612C
SureSelect
XT
* SureSelect reagents must be used within one year of receipt.
SureSelect Capture Library (select one)*
Table 3
10
Capture Libraries
16 Reactions
96 Reactions
480 Reactions
SureSelectXT Focused Exome
5190-7787
5190-7788
SureSelectXT Focused Exome Plus 1
5190-7790
5190-7791
SureSelectXT Clinical Research Exome
5190-7338
5190-7339
SureSelectXT Human All Exon V5
5190-6208
5190-6209
SureSelectXT Human All Exon V5+UTRs
5190-6213
5190-6214
SureSelectXT Human All Exon V5+lncRNA
5190-6446
5190-6447
SureSelectXT Human All Exon V5 Plus
5190-6211
5190-6212
SureSelectXT Human All Exon V4
5190-4631
5190-4632
5190-4634
SureSelectXT Human All Exon V4+UTRs
5190-4636
5190-4637
5190-4639
SureSelectXT Mouse All Exon
5190-4641
5190-4642
5190-4644
SureSelectXT Inherited Disease
5190-7518
5190-7519
SureSelectXT Inherited Disease Plus
5190-7521
5190-7522
SureSelectXT DNA Kinome
5190-4646
5190-4647
5190-4649
SureSelectXT X-chromosome
5190-4651
5190-4652
5190-4653
SureSelectXT Custom 1 kb up to 499 Kb
5190-4806
5190-4807
5190-4809
(reorder)
5190-4811
5190-4812
5190-4814
SureSelectXT Custom 0.5 Mb up to 2.9 Mb
5190-4816
5190-4817
5190-4819
(reorder)
5190-4821
5190-4822
5190-4824
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
Before You Begin
Optional Reagents
Table 3
1
SureSelect Capture Library (select one)* (Continued)
Capture Libraries
16 Reactions
96 Reactions
480 Reactions
SureSelectXT Custom 3 Mb up to 5.9 Mb
5190-4826
5190-4827
5190-4829
(reorder)
5190-4831
5190-4832
5190-4834
SureSelectXT Custom 6 Mb up to 11.9 Mb
5190-4836
5190-4837
5190-4839
(reorder)
5190-4841
5190-4842
5190-4844
SureSelectXT Custom 12 Mb up to 24 Mb
5190-4896
5190-4897
5190-4899
(reorder)
5190-4901
5190-4902
5190-4904
* SureSelect capture libraries must be used within one year of receipt.
Table 4
Required Reagents for Hybridization
Description
Vendor and part number
Dynabeads MyOne Streptavidin T1
Life Technologies
2 mL
10 mL
100 mL
Cat #65601
Cat #65602
Cat #65603
Nuclease-free Water (not DEPC-treated)
Ambion Cat #AM9930
Optional Reagents
Table 5
Optional Reagents
Description
Vendor and part number
Ethylene glycol
American Bioanalytical p/n AB00455
Tween 20
Sigma-Aldrich p/n P9416-50ML
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
11
1

Before You Begin
Required Equipment
Required Equipment
Table 6
Required Equipment for Library Prep and Post-Hybridization Amplification
Description
Vendor and part number
2100 Bioanalyzer
Agilent p/n G2938C
or
2200 TapeStation System
Agilent p/n G2964AA or G2965AA
Mx3005P Real-Time PCR System
Agilent p/n 401449 or equivalent
Thermal cycler
Agilent SureCycler, Life Technologies Veriti
Thermal Cycler, BioRad (MJ Research) DNA
Engine PTC-200, or equivalent
Covaris Sample Preparation System, S-series or
E-series model
Covaris
Covaris microTUBE with AFA fiber and snap cap
Covaris p/n 520045
Microcentrifuge
Eppendorf Microcentrifuge Model 5417C or
equivalent
1.5-mL LoBind Tube
Eppendorf p/n 022431021 or equivalent
Qubit Fluorometer
Life Technologies p/n Q32857
Dynal DynaMag-2 magnetic stand
Life Technologies p/n 123-21D or equivalent
P10, P20, P200 and P1000 pipettes
Pipetman P10, P20, P200, P1000 or
equivalent
Vacuum concentrator
Savant SpeedVac or equivalent
Ice bucket
Powder-free gloves
PCR tubes, strips, or plates
Sterile, nuclease-free aerosol barrier pipette tips
Timer
Vortex mixer
Heat block at 37°C
12
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
Before You Begin
Optional Equipment
Table 7
1
Required Equipment for Hybridization
Description
Vendor and part number
Mx3000P/Mx3005P 96-well tube plates
Agilent p/n 410088 or equivalent
Mx3000P/Mx3005P optical strip caps
Agilent p/n 401425 or equivalent
MicroAmp Clear Adhesive Film
Life Technologies p/n 4306311
BD Clay Adams Nutator Mixer
BD Diagnostics p/n 421105 or equivalent
Dynal DynaMag-2 magnetic stand
Life Technologies p/n 123-21D or equivalent
P10, P20, P200 and P1000 pipettes
Pipetman P10, P20, P200, P1000 or
equivalent
Pipet-Light Multichannel Pipette, 12 channels
Rainin p/n L12-20 or equivalent
PCR tubes, strips, or plates
Sterile, nuclease-free aerosol barrier pipette tips
Thermal cycler
Agilent SureCycler, Life Technologies Veriti
Thermal Cycler, BioRad (MJ Research) DNA
Engine PTC-200, or equivalent
Timer
Vortex mixer
Water bath or heat block set to 65°C
Optional Equipment
Table 8
Optional Equipment for Hybridization
Description
Vendor and part number
Tube-strip capping tool
Agilent p/n 410099
PlateLoc Thermal Microplate Sealer and
Peelable Aluminum Seal
Agilent p/n G5402A and
Agilent p/n 24210-001
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
13
This page intentionally left blank.
14
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
SureSelectXT Target Enrichment System for Illumina Paired-End
Sequencing Protocol
2
Sample Preparation
(3 µg DNA Samples)
Step 1. Shear DNA 19
Step 2. Purify the sample using Agencourt AMPure XP beads
Step 3. Assess quality (optional) 22
Step 4. Repair the ends 25
Step 5. Purify the sample using Agencourt AMPure XP beads
Step 6. Add ‘A’ Bases to the 3' end of the DNA fragments 27
Step 7. Purify the sample using Agencourt AMPure XP beads
Step 8. Ligate the paired-end adaptor 29
Step 9. Purify the sample using Agencourt AMPure XP beads
Step 10. Amplify adaptor-ligated library 31
Step 11. Purify the sample with Agencourt AMPure XP beads
Step 12. Assess quality and quantity 35
21
26
28
30
34
This chapter contains instructions for prepped library production specific to
the Illumina paired-read sequencing platform. For each sample to be
sequenced, individual library preparations, hybridizations, and captures are
performed. The samples are then tagged by PCR with an index (barcode)
sequence. Depending on the target size of the SureSelect capture, up to 16
samples can be pooled and sequenced in a single lane using Illumina multiplex
index tags.
CAUTION
This protocol contains an option for 3-µg DNA samples (standard input) and an option
for 200-ng DNA samples (low input). Make sure you follow the steps in the
appropriate chapter.
15
The steps in this chapter differ from the Illumina protocol in the use of the
Covaris system for gDNA shearing, smaller target shear size, elimination of
size selection by gel purification, and implementation of AMPure XP beads for
all purification steps, and primers used for PCR.
Refer to the Illumina protocol Preparing Samples for Multiplexed Paired-End
Sequencing (p/n 1005361 Rev. C) for more information.
NOTE
Make sure genomic DNA samples are of high quality with an OD 260/280 ratio ranging
from 1.8 to 2.0. Use the Qubit system to quantify genomic DNA before library preparation.
Sample Preparation (3 µg DNA Samples)
2
Genomic DNA Sample 1, 2…n
Shear
DNA fragments with a base pair peak of 150 to 200 bp
Repair ends
Blunt-ended fragments with 5'-phosphorylated ends
Add Klenow and dATP
3'-dA overhang
Ligate indexing-specific
adaptors
Genomic Locations
Adaptor-modified ends
Bait Design in eArray
SureSelect Capture
Library
AMPure XP bead purification
Removal of unligated adaptors
PCR with SureSelect Primer
and SureSelect Pre-capture
Reverse PCR primers
Prepped Library
Library Hybridization (1 per sample)
16 or 24 hours at 65°C
Hybrid Capture Selection
Magnetic bead selection
Amplification and Index Tagging
PCR and purification
Quality Assessment of each indexed sample
Bioanalyzer and Quantification by QPCR
Pool samples 1, 2...n
Sequencing
Figure 1
Overall sequencing sample preparation workflow.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
17
Table 10
Overview and time requirements
Step
Time
Illumina Prepped library Production
1 day
Library Hybridization
16 or 24 hours
Bead preparation
30 minutes
Capture Selection and Washing
2 hours
DNA purification
30 minutes
Post-Hybridization Amplification
1 hour
PCR purification
30 minutes
Bioanalyzer QC and QPCR
2 to 3 hours
Pool indexed samples by mass
< 1 hour
Sample Preparation (3 µg DNA Samples)
Step 1. Shear DNA
2
Step 1. Shear DNA
For each DNA sample to be sequenced, prepare 1 library.
1 Use the Qubit dsDNA Assay to determine the concentration of your gDNA
sample. Make sure the gDNA is of high quality (non-degraded, A260/A280 is
1.8 to 2.0).
Follow the instructions for the instrument.
2 Set up the Covaris instrument.
a Check that the water in the Covaris tank is filled with fresh deionized
water to fill line level 12 on the graduated fill line label.
b Check that the water covers the visible glass part of the tube.
c Set the chiller temperature to between 2°C to 5°C to ensure that the
temperature reading in the water bath displays 5°C.
d Optional. Supplement the circulated water chiller with ethylene glycol to
20% volume to prevent freezing.
e On the instrument control panel, push the Degas button. Degas the
instrument for least 30 minutes before use.
Refer to the Covaris instrument user guide.
3 Dilute 3 μg of high-quality gDNA with 1X Low TE Buffer in a 1.5-mL LoBind
Tube to a total volume of 130 μL.
4 Put a Covaris microTube into the loading and unloading station.
Keep the cap on the tube.
5 Use a tapered pipette tip to slowly transfer the 130 μL of DNA sample
through the pre-split septa.
Be careful not to introduce a bubble into the bottom of the tube.
6 Secure the microTube in the tube holder and shear the DNA with the
settings in Table 11 or Table 12. The target peak for base pair size is 150 to
200 bp.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
19
Table 11
Shear settings for Covaris instruments that use SonoLab 7 or newer
Setting
Value
Duty Factor
10%
Peak Incident Power (PIP)
175
Cycles per Burst
200
Treatment Time
360 seconds
Bath Temperature
4° to 8° C
Table 12
Shear settings for Covaris instruments that use SonoLab software previous to
SonoLab 7
Setting
Value
Duty Cycle
10%
Intensity
5
Cycles per Burst
200
Time
6 cycles of 60 seconds each
Set Mode
Frequency sweeping
Temperature
4° to 7° C
7 Put the Covaris microTube back into the loading and unloading station.
8 While keeping the snap-cap on, insert a pipette tip through the pre-split
septa, then slowly remove the sheared DNA.
9 Transfer the sheared DNA into a new 1.5-mL LoBind Tube.
Sample Preparation (3 µg DNA Samples)
Step 2. Purify the sample using Agencourt AMPure XP beads
2
Step 2. Purify the sample using Agencourt AMPure XP beads
1 Let the AMPure XP beads come to room temperature for at least 30
minutes.
2 Mix the reagent well so that the reagent appears homogeneous and
consistent in color. Do not freeze.
3 Add 180 μL of homogeneous AMPure XP beads to a 1.5-mL LoBind Tube,
and add the sheared DNA library (~130 μL). Mix well on a vortex mixer and
incubate for 5 minutes.
4 Put the tube in the magnetic stand. Wait for the solution to clear
(approximately 3 to 5 minutes).
5 Keep the tube in the magnetic stand. Do not touch the beads while you
carefully discard the cleared solution from the tubes.
6 Continue to keep the tube in the magnetic stand while you dispense 500 μL
of 70% Ethanol in each tube.
Use fresh 70% Ethanol for optimal result.
7 Let the tube sit for 1 minute to allow any disturbed beads to settle, and
remove the Ethanol.
8 Repeat step 6 and step 7 step once.
9 Spin briefly in a centrifuge. Return the tube to the magnetic stand for 30
seconds. Remove residual Ethanol with a P20 pipette.
10 Dry the samples on the 37°C heat block for 5 minutes or until the residual
Ethanol completely evaporates.
Do not dry the bead pellet to the point that the bead pellet appears cracked.
Elution efficiency is significantly decreased when the bead pellet is
excessively dried.
11 Add 50 μL of Nuclease-free Water, mix well on a vortex mixer, and incubate
for 2 minutes at room temperature.
12 Put the tube in the magnetic stand and leave for 2 to 3 minutes, until the
solution is clear.
13 Remove approximately 50 μL of the supernatant to a fresh 1.5-mL LoBind
Tube. You can discard the beads at this time.
Stopping Point
If you do not continue to the next step, store the samples at -20°C.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
21
Step 3. Assess quality (optional)
Quality assessment can be done with either the 2100 Bioanalyzer instrument
or the 2200 TapeStation.
2100 Bioanalyzer System with DNA 1000 Assay
For the Standard Input option, use the DNA 1000 Assay.
1 Check that the 2100 Bioanalyzer electrodes have been cleaned as instructed
in the reagent kit guide.
2 Open the Agilent 2100 Expert Software (version B.02.02 or higher), turn on
the 2100 Bioanalyzer and check communication.
3 Prepare the chip, samples and ladder as instructed in the reagent kit guide.
4 Load the prepared chip into the 2100 Bioanalyzer and start the run within
five minutes after preparation.
5 Within the instrument context, choose the appropriate assay from the drop
down list.
6 Start the run. Enter sample names and comments in the Data and Assay
context.
7 Verify the results.
Check that the electropherogram shows a distribution with a peak height
between 150 to 200 bp.
Sample Preparation (3 µg DNA Samples)
Step 3. Assess quality (optional)
Figure 2
2
Analysis of sheared DNA using a DNA 1000 Bioanalyzer assay. The electropherogram shows a distribution with a peak size between 150 to 200 bp.
2200 TapeStation and D1000 ScreenTape
You can use the 2200 TapeStation for rapid analysis of multiple samples. Use
the D1000 ScreenTape and D1000 Reagents. For more information to do this
step, see the Agilent 2200 TapeStation User Manual.
1 Prepare the TapeStation samples as instructed in the Agilent 2200
TapeStation User Manual. Use 1 μL of each sheared DNA sample diluted
with 3 μL of D1000 sample buffer for the analysis.
CAUTION
Make sure that you thoroughly mix the combined DNA and D1000 sample buffer on a
vortex mixer for 5 seconds or more for accurate quantitation.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
23
2 Load the sample plate or tube strips from step 1, the D1000 ScreenTape,
and loading tips into the 2200 TapeStation as instructed in the Agilent 2200
TapeStation User Manual. Start the run.
3 Verify that the electropherogram shows an average DNA fragment size of
about 185 bp. A sample electropherogram is shown in Figure 3.
Stopping Point
If you do not continue to the next step, seal the sheared DNA sample plate and
store at 4°C overnight or at –20°C for prolonged storage.
Figure 3
Analysis of sheared DNA using the 2200 TapeStation with a D1000 ScreenTape. The electropherogram shows an average DNA fragment size between
150 to 200 bp.
Sample Preparation (3 µg DNA Samples)
Step 4. Repair the ends
2
Step 4. Repair the ends
To process multiple samples, prepare master mixes with overage at each step,
without the DNA sample. Master mixes for preparation of 12 samples
(including excess) are shown in each table as an example.
Use the SureSelect Library Prep Kit, ILM.
Prepare the master mix on ice.
1 Prepare reaction mix and DNA sample (prepare on ice):
a Prepare the reaction mix in Table 13. Mix well on a vortex mixer.
b Add 52 μL of the reaction mix to each well or tube.
c Add 48 μL of each DNA sample to each well or tube. Mix by pipetting.
Change pipette tips between samples.
2 Incubate in a thermal cycler for 30 minutes at 20°C. Do not use a heated lid.
If the heated lid of your thermal cycler cannot be turned off, incubate the
sample with lid open.
Table 13
End Repair Mix
Reagent
Volume for 1 Library
Volume for 12 Libraries
(includes excess)
Nuclease-free Water
35.2 µL
440 µL
10× End Repair Buffer (clear cap)
10 µL
125 µL
dNTP Mix (green cap)
1.6 µL
20 µL
T4 DNA Polymerase (purple cap)
1 µL
12.5 µL
Klenow DNA Polymerase (yellow cap)
2 µL
25 µL
T4 Polynucleotide Kinase (orange cap)
2.2 µL
27.5 µL
Total Volume
52 µL
650 µL (52 µL/sample)
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
25
Step 5. Purify the sample using Agencourt AMPure XP beads
1 Let the AMPure XP beads come to room temperature for at least 30
minutes.
2 Mix the reagent well so that the reagent appears homogeneous and
consistent in color. Do not freeze.
3 Add 180 μL of homogeneous AMPure XP beads to a 1.5-mL LoBind Tube,
and add the end-repaired DNA library (~100 μL). Mix well on a vortex
mixer and incubate for 5 minutes.
4 Put the tube in the magnetic stand. Wait for the solution to clear
(approximately 3 to 5 minutes).
5 Keep the tube in the magnetic stand. Do not touch the beads while you
carefully discard the cleared solution from the tubes.
6 Continue to keep the tube in the magnetic stand while you dispense 500 μL
of 70% Ethanol in each tube.
Use fresh 70% Ethanol for optimal result.
7 Let the tube sit for 1 minute to allow any disturbed beads to settle, and
remove the Ethanol.
8 Repeat step 6 and step 7 step once.
9 Spin briefly in a centrifuge. Return the tube to the magnetic stand for 30
seconds. Remove residual Ethanol with a P20 pipette.
10 Dry the samples on the 37°C heat block for 5 minutes or until the residual
Ethanol completely evaporates.
Do not dry the bead pellet to the point that the bead pellet appears cracked.
Elution efficiency is significantly decreased when the bead pellet is
excessively dried.
11 Add 32 μL of Nuclease-free Water, mix well on a vortex mixer, and incubate
for 2 minutes at room temperature.
12 Put the tube in the magnetic stand and leave for 2 to 3 minutes, until the
solution is clear.
Remove the supernatant (~32 μL) to a fresh 1.5-mL LoBind Tube. You can
discard the beads at this time.
Stopping Point
If you do not continue to the next step, store the samples at -20°C.
Sample Preparation (3 µg DNA Samples)
Step 6. Add ‘A’ Bases to the 3' end of the DNA fragments
2
Step 6. Add ‘A’ Bases to the 3' end of the DNA fragments
Use the SureSelect Library Prep Kit, ILM.
1 Prepare reaction mix and DNA sample (prepare on ice):
a Prepare the reaction mix in Table 14. Mix well on a vortex mixer.
b Add 20 μL of the reaction mix to each well or tube.
c Add 30 μL of each DNA sample to each well or tube. Mix by pipetting.
Change pipette tips between samples.
Table 14
Adding “A” Bases
Reagent
Volume for 1
Library
Volume for 12 Libraries
(includes excess)
Nuclease-free Water
11 µL
137.5 µL
10× Klenow Polymerase Buffer (blue cap)
5 µL
62.5 µL
dATP (green cap)
1 µL
12.5 µL
Exo(-) Klenow (red cap)
3 µL
37.5 µL
Total Volume
20 µL
250 µL (20 µL/sample)
2 Incubate in a thermal cycler for 30 minutes at 37°C.
If you use a heated lid, make sure that the lid temperature does not exceed
50°C.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
27
Step 7. Purify the sample using Agencourt AMPure XP beads
1 Let the AMPure XP beads come to room temperature for at least 30
minutes.
2 Mix the reagent well so that the reagent appears homogeneous and
consistent in color. Do not freeze.
3 Add 90 μL of homogeneous AMPure XP beads to a 1.5-mL LoBind Tube, and
add the A-tailed DNA library (~50 μL). Mix well on a vortex mixer and
incubate for 5 minutes.
4 Put the tube in the magnetic stand. Wait for the solution to clear
(approximately 3 to 5 minutes).
5 Keep the tube in the magnetic stand. Do not touch the beads while you
carefully discard the cleared solution from the tubes.
6 Continue to keep the tube in the magnetic stand while you dispense 500 μL
of 70% Ethanol in each tube.
Use fresh 70% Ethanol for optimal result.
7 Let the tube sit for 1 minute to allow any disturbed beads to settle, and
remove the Ethanol.
8 Repeat step 6 and step 7 step once.
9 Spin briefly in a centrifuge. Return the tube to the magnetic stand for 30
seconds. Remove residual Ethanol with a P20 pipette.
10 Dry the samples on the 37°C heat block for 5 minutes or until the residual
Ethanol completely evaporates.
Do not dry the bead pellet to the point that the bead pellet appears cracked.
Elution efficiency is significantly decreased when the bead pellet is
excessively dried.
11 Add 15 μL of Nuclease-free Water, mix well on a vortex mixer, and incubate
for 2 minutes at room temperature.
12 Put the tube in the magnetic stand and leave for 2 to 3 minutes, until the
solution is clear.
13 Remove the supernatant (~15 μL) to a fresh 1.5-mL LoBind Tube. You can
discard the beads at this time.
14 Proceed immediately to the next step, “Step 8. Ligate the paired-end
adaptor”.
Sample Preparation (3 µg DNA Samples)
Step 8. Ligate the paired-end adaptor
2
Step 8. Ligate the paired-end adaptor
This step uses a 10:1 molar ratio of adaptor to genomic DNA insert, based on a
starting quantity of 3 μg of DNA before shearing.
Use the SureSelect Library Prep Kit, ILM.
1 Prepare reaction mix and DNA sample (prepare on ice):
a Prepare the reaction mix in Table 15.
b Add 37 μL of the reaction mix to each well or tube.
c Add 13 μL of each DNA sample to each well or tube. Mix by pipetting.
Change pipette tips between samples.
Table 15
Ligation master mix
Reagent
Volume for 1 Library
Volume for 12 Libraries
(includes excess)
Nuclease-free Water
15.5 µL
193.75 µL
5× T4 DNA Ligase Buffer
10 µL
125 µL
SureSelect Adaptor Oligo Mix (brown cap) 10 µL
125 µL
T4 DNA Ligase (red cap)
1.5 µL
18.75 µL
Total Volume
37 µL
462.5 µL (37 µL/sample)
2 Incubate for 15 minutes at 20°C on a thermal cycler. Do not use a heated
lid.
Stopping Point
If you do not continue to the next step, store the samples at -20°C.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
29
Step 9. Purify the sample using Agencourt AMPure XP beads
1 Let the AMPure XP beads come to room temperature for at least 30
minutes.
2 Mix the reagent well so that the reagent appears homogeneous and
consistent in color. Do not freeze.
3 Add 90 μL of homogeneous AMPure XP beads to a 1.5-mL LoBind Tube, and
add the ligated library (~50 μL). Mix well on a vortex mixer and incubate
for 5 minutes.
4 Put the tube in the magnetic stand. Wait for the solution to clear
(approximately 3 to 5 minutes).
5 Keep the tube in the magnetic stand. Do not touch the beads while you
carefully discard the cleared solution from the tubes.
6 Continue to keep the tube in the magnetic stand while you dispense 500 μL
of 70% Ethanol in each tube.
Use fresh 70% Ethanol for optimal result.
7 Let the tube sit for 1 minute to allow any disturbed beads to settle, and
remove the Ethanol.
8 Repeat step 6 and step 7 step once.
9 Spin briefly in a centrifuge. Return the tube to the magnetic stand for 30
seconds. Remove residual Ethanol with a P20 pipette.
10 Dry the samples on the 37°C heat block for 5 minutes until the residual
Ethanol is completely evaporated.
Do not dry the bead pellet to the point that the bead pellet appears cracked.
Elution efficiency is significantly decreased when the bead pellet is
excessively dried.
11 Add 32 μL of Nuclease-free Water, mix well on a vortex mixer, and incubate
for 2 minutes at room temperature.
12 Put the tube in the magnetic stand and leave for 2 to 3 minutes, until the
solution is clear.
13 Remove the supernatant (~32 μL) to a fresh 1.5-mL LoBind Tube. You can
discard the beads at this time.
Stopping Point
If you do not continue to the next step, store the samples at -20°C.
Sample Preparation (3 µg DNA Samples)
Step 10. Amplify adaptor-ligated library
2
Step 10. Amplify adaptor-ligated library
Use reagents from these kits:
• SureSelect Library Prep Kit, ILM
• SureSelect Target Enrichment Kit ILM Indexing Hyb Module Box #2
• Herculase II Fusion DNA Polymerase (Agilent)
This protocol uses half of the adaptor-ligated fragments for amplification. The
remainder can be saved at -20°C for future use, if needed.
CAUTION
To avoid cross-contaminating libraries, set up PCR reactions (all components except
the library DNA) in a dedicated clean area or PCR hood with UV sterilization and
positive air flow.
1 Prepare reaction mix and DNA sample:
a Prepare the reaction mix in Table 16, on ice. Mix well on a vortex mixer.
b Add 35 μL of the reaction mix to each well or tube.
c Add 15 μL of each DNA sample to each well or tube. Mix by pipetting.
Change pipette tips between samples.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
31
Table 16
Components for PCR mix
Reagent
Volume for 1
reaction
Volume for 12 reactions
(includes excess)
Nuclease-free Water
21 μL
262.5 μL
SureSelect Primer (brown cap)*
1.25 µL
15.625 µL
SureSelect ILM Indexing Pre Capture PCR
Reverse Primer (clear cap)†
1.25 µL
15.625 µL
5× Herculase II Rxn Buffer (clear cap)‡
10 μL
125 μL
0.5 μL
6.25 μL
1 μL
12.5 μL
35 μL
437.5 µL 
(35 µL/reaction)
100 mM dNTP Mix (green cap)‡
Herculase II Fusion DNA Polymerase (red
cap)‡
Total
* Included in the SureSelect Library Prep Kit, ILM.
† Included in the SureSelect Target Enrichment Kit ILM Indexing Hyb Module Box #2.
‡ Included in the Herculase II Fusion DNA Polymerase (Agilent) kit. Do not use the buffer or dNTP
mix from any other kit.
2 Run the program in Table 17 in a thermal cycler.
Table 17
PCR program
Step
Temperature
Time
Step 1
98°C
2 minutes
Step 2
98°C
30 seconds
Step 3
65°C
30 seconds
Step 4
72°C
1 minute
Step 5
Repeat Step 2 through Step 4 for a total of 4 to 6 times.
Step 6
72°C
10 minutes
Step 7
4°C
Hold
Sample Preparation (3 µg DNA Samples)
Step 10. Amplify adaptor-ligated library
NOTE
2
Different library preparations can produce slightly different results, based on varying DNA
quality. In most cases, 5 cycles will produce an adequate yield for subsequent capture
without introducing bias or non-specific products. If yield is too low or non-specific high
molecular weight products are observed, adjust the number of cycles accordingly with the
remaining extra library template.
As an alternative, you can prepare one PCR master mix as outlined in Table 16. Split the
master mix into three small-scale 10-µL PCR reactions and run for 4, 5 or 6 cycles. Clean
these PCR reactions using the AMPure XP protocol outlined in “Step 2. Purify the sample
using Agencourt AMPure XP beads” with these modifications: Use 30 µL of AMPure XP
beads and elute with 20 µL of Nuclease-free Water. Run these cleaned samples on a
DNA1000 chip on the Bioanalyzer, as described in “Step 3. Assess quality (optional)”.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
33
Step 11. Purify the sample with Agencourt AMPure XP beads
1 Let the AMPure XP beads come to room temperature for at least 30
minutes.
2 Mix the reagent well so that the reagent appears homogeneous and
consistent in color. Do not freeze.
3 Add 90 μL of homogeneous AMPure XP beads to a 1.5-mL LoBind Tube, and
add the amplified library (~50 μL). Mix well on a vortex mixer and incubate
for 5 minutes.
4 Put the tube in the magnetic stand. Wait for the solution to clear
(approximately 3 to 5 minutes).
5 Keep the tube in the magnetic stand. Do not touch the beads while you
carefully discard the cleared solution from the tubes.
6 Continue to keep the tube in the magnetic stand while you dispense 500 μL
of 70% Ethanol in each tube.
Use fresh 70% Ethanol for optimal result.
7 Let the tube sit for 1 minute to allow any disturbed beads to settle, and
remove the Ethanol.
8 Repeat step 6 and step 7 step once.
9 Spin briefly in a centrifuge. Return the tube to the magnetic stand for 30
seconds. Remove residual Ethanol with a P20 pipette.
10 Dry the samples on the 37°C heat block for 5 minutes or until the residual
Ethanol completely evaporates.
Do not dry the bead pellet to the point that the bead pellet appears cracked.
Elution efficiency is significantly decreased when the bead pellet is
excessively dried.
11 Add 30 μL of Nuclease-free Water, mix well on a vortex mixer, and incubate
for 2 minutes at room temperature.
12 Put the tube in the magnetic stand and leave for 2 to 3 minutes, until the
solution is clear.
13 Remove the supernatant (~30 μL) to a fresh 1.5-mL LoBind Tube. You can
discard the beads at this time.
Stopping Point
If you do not continue to the next step, store the samples at -20°C.
Sample Preparation (3 µg DNA Samples)
Step 12. Assess quality and quantity
2
Step 12. Assess quality and quantity
Quality assessment can be done with either the 2100 Bioanalyzer instrument
or the 2200 TapeStation.
2100 Bioanalyzer System and DNA 1000 Assay
1 Check that the 2100 Bioanalyzer electrodes have been cleaned as instructed
in the reagent kit guide.
2 Open the Agilent 2100 Expert Software (version B.02.02 or higher), turn on
the 2100 Bioanalyzer and check communication.
3 Prepare the chip, samples and ladder as instructed in the reagent kit guide.
4 Load the prepared chip into the 2100 Bioanalyzer and start the run within
five minutes after preparation.
5 Within the instrument context, choose the appropriate assay from the drop
down list.
6 Start the run. Enter sample names and comments in the Data and Assay
context.
7 Verify the results. Check that the electropherogram shows a distribution
with a peak size approximately 225 to 275 bp. Measure the concentration of
the library by integrating under the peak.
Figure 4
Analysis of amplified library DNA using a DNA 1000 assay. The electropherogram shows a single peak in the size range of 225 to 275 bp.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
35
2200 TapeStation and D1000 ScreenTape
You can use the 2200 TapeStation to analyze the amplified libraries. Use the
D1000 ScreenTape and D1000 Reagents. For more information to do this step,
see the Agilent 2200 TapeStation User Manual.
1 Prepare the TapeStation samples as instructed in the Agilent 2200
TapeStation User Manual. Use 1 μL of each amplified library DNA sample
diluted with 3 μL of D1000 sample buffer for the analysis.
CAUTION
Make sure that you thoroughly mix the combined DNA and D1000 sample buffer on a
vortex mixer for 5 seconds or more for accurate quantitation.
2 Load the sample plate or tube strips from step 1, the D1000 ScreenTape,
and loading tips into the 2200 TapeStation as instructed in the Agilent 2200
TapeStation User Manual. Start the run.
3 Verify that the electropherogram shows a distribution with a peak size
approximately 225 to 275 bp. A sample electropherogram is shown in
Figure 5.
Stopping Point
If you do not continue to the next step, seal the sheared DNA sample plate and
store at 4°C overnight or at –20°C for prolonged storage.
Sample Preparation (3 µg DNA Samples)
Step 12. Assess quality and quantity
Figure 5
2
Analysis of amplified library DNA using the 2200 TapeStation with a D1000
ScreenTape. The electropherogram shows a single peak in the size range of
225 to 275 bp.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
37
This page intentionally left blank.
38
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
SureSelectXT Target Enrichment System for Illumina Paired-End
Sequencing Protocol
3
Sample Preparation
(200 ng DNA Samples)
Step 1. Shear DNA 43
Step 2. Assess quality (optional) 45
Step 3. Repair the ends 46
Step 4. Purify the sample using Agencourt AMPure XP beads
Step 5. Add ‘A’ Bases to the 3' end of the DNA fragments 48
Step 6. Purify the sample using Agencourt AMPure XP beads
Step 7. Ligate the paired-end adaptor 50
Step 8. Purify the sample using Agencourt AMPure XP beads
Step 9. Amplify adaptor-ligated library 52
Step 10. Purify the sample with Agencourt AMPure XP beads
Step 11. Assess quality and quantity 55
47
49
51
54
This chapter contains instructions for prepped library production specific to
the Illumina paired-read sequencing platform. For each sample to be
sequenced, individual library preparations, hybridizations, and captures are
performed. The samples are then tagged by PCR with an index (barcode)
sequence. Depending on the target size of the SureSelect capture, up to 16
samples can be pooled and sequenced in a single lane using Illumina multiplex
index tags.
CAUTION
This protocol contains an option for 3-µg DNA samples (standard input) and an option
for 200-ng DNA samples (low input). Make sure you follow the steps in the
appropriate chapter.
39
The steps in this chapter differ from the Illumina protocol in the use of the
Covaris system for gDNA shearing, smaller target shear size, elimination of
size selection by gel purification, and implementation of AMPure XP beads for
all purification steps, and primers used for PCR.
Refer to the Illumina protocol Preparing Samples for Multiplexed Paired-End
Sequencing (p/n 1005361 Rev. C) for more information.
NOTE
Make sure genomic DNA samples are of high quality with an OD 260/280 ratio ranging
from 1.8 to 2.0. Use the Qubit system to quantify genomic DNA before library preparation.
Sample Preparation (200 ng DNA Samples)
3
Genomic DNA Sample 1, 2…n
Shear
DNA fragments with a base pair peak of 150 to 200 bp
Repair ends
Blunt-ended fragments with 5'-phosphorylated ends
Add Klenow and dATP
3'-dA overhang
Ligate diluted indexingspecific adaptors
Genomic Locations
Adaptor-modified ends
Bait Design in eArray
SureSelect Capture
Library
AMPure XP bead purification
Removal of unligated adaptors
PCR with SureSelect Primer
and SureSelect Pre-capture
Reverse PCR primers
Prepped Library
Library Hybridization (1 per sample)
16 or 24 hours at 65°C
Hybrid Capture Selection
Magnetic bead selection
Amplification and Index Tagging
PCR and purification
Quality Assessment of each indexed sample
Bioanalyzer and Quantification by QPCR
Pool samples 1, 2...n
Sequencing
Figure 6
Overall sequencing sample preparation workflow.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
41
Table 18
Overview and time requirements
Step
Time
Illumina Prepped library Production
1 day
Library Hybridization
16 or 24 hours
Bead preparation
30 minutes
Capture Selection and Washing
2 hours
DNA purification
30 minutes
Post-Hybridization Amplification
1 hour
PCR purification
30 minutes
Bioanalyzer QC and QPCR
2 to 3 hours
Pool indexed samples by mass
< 1 hour
Sample Preparation (200 ng DNA Samples)
Step 1. Shear DNA
3
Step 1. Shear DNA
For each DNA sample to be sequenced, prepare 1 library.
1 Use the Qubit dsDNA Assay to determine the concentration of your gDNA
sample. Make sure the gDNA is of high quality (non-degraded, A260/A280 is
1.8 to 2.0).
Follow the instructions for the instrument.
2 Set up the Covaris instrument.
a Check that the water in the Covaris tank is filled with fresh deionized
water to fill line level 12 on the graduated fill line label.
b Check that the water covers the visible glass part of the tube.
c Set the chiller temperature to between 2°C to 5°C to ensure that the
temperature reading in the water bath displays 5°C.
d Optional. Supplement the circulated water chiller with ethylene glycol to
20% volume to prevent freezing.
e On the instrument control panel, push the Degas button. Degas the
instrument for least 30 minutes before use.
Refer to the Covaris instrument user guide.
3 Dilute 200 ng of high-quality gDNA with 1X Low TE Buffer in a 1.5-mL
LoBind Tube for a total of 50 μL.
4 Put a Covaris microTube into the loading and unloading station.
Keep the cap on the tube.
5 Use a tapered pipette tip to slowly transfer the 50 μL of DNA sample
through the pre-split septa.
Be careful not to introduce a bubble into the bottom of the tube.
6 Secure the microTube in the tube holder and shear the DNA with the
settings in Table 19 or Table 20. The target peak for base pair size is 150 to
200 bp.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
43
Table 19
Shear settings for Covaris instruments that use SonoLab 7 or newer
Setting
Value
Duty Factor
10%
Peak Incident Power (PIP)
175
Cycles per Burst
200
Treatment Time
360 seconds
Bath Temperature
4° to 8° C
Table 20
Shear settings for Covaris instruments that use SonoLab software previous to
SonoLab 7
Setting
Value
Duty Cycle
10%
Intensity
5
Cycles per Burst
200
Time
6 cycles of 60 seconds each
Set Mode
Frequency sweeping
Temperature
4° to 7° C
7 Put the Covaris microTube back into the loading and unloading station.
8 While keeping the snap-cap on, insert a pipette tip through the pre-split
septa, then slowly remove the sheared DNA.
9 Transfer the sheared DNA into a new 1.5-mL LoBind Tube.
Sample Preparation (200 ng DNA Samples)
Step 2. Assess quality (optional)
3
Step 2. Assess quality (optional)
Quality assessment can be done with the 2100 Bioanalyzer instrument.
For the 200-ng option, use the High Sensitivity DNA assay.
1 Check that the 2100 Bioanalyzer electrodes have been cleaned as instructed
in the reagent kit guide.
2 Open the Agilent 2100 Expert Software (version B.02.02 or higher), turn on
the 2100 Bioanalyzer and check communication.
3 Prepare the chip, samples and ladder as instructed in the reagent kit guide.
4 Load the prepared chip into the 2100 Bioanalyzer and start the run within
five minutes after preparation.
5 Within the instrument context, choose the appropriate assay from the drop
down list.
6 Start the run. Enter sample names and comments in the Data and Assay
context.
7 Verify the results.
Check that the electropherogram shows a distribution with a peak height
between 120 to 150 bp.
Figure 7
Analysis of sheared DNA using a High Sensitivity DNA assay. The electropherogram shows a distribution with a peak size between 120 to 150 bp.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
45
Step 3. Repair the ends
To process multiple samples, prepare master mixes with overage at each step,
without the DNA sample. Master mixes for preparation of 12 samples
(including excess) are shown in each table as an example.
Use the SureSelect Library Prep Kit, ILM.
Prepare the master mix on ice.
1 Prepare reaction mix and DNA sample (prepare on ice):
a Prepare the reaction mix in Table 21. Mix well on a vortex mixer.
b Add 52 μL of the reaction mix to each well or tube.
c Add 48 μL of each DNA sample to each well or tube. Mix by pipetting.
Change pipette tips between samples.
2 Incubate in a thermal cycler for 30 minutes at 20°C. Do not use a heated lid.
If the heated lid of your thermal cycler cannot be turned off, incubate the
sample with lid open.
Table 21
End Repair Mix
Reagent
Volume for 1 Library
Volume for 12 Libraries
(includes excess)
Nuclease-free Water
35.2 µL
440 µL
10× End Repair Buffer (clear cap)
10 µL
125 µL
dNTP Mix (green cap)
1.6 µL
20 µL
T4 DNA Polymerase (purple cap)
1 µL
12.5 µL
Klenow DNA Polymerase (yellow cap)
2 µL
25 µL
T4 Polynucleotide Kinase (orange cap)
2.2 µL
27.5 µL
Total Volume
52 µL
650 µL (52 µL/sample)
Sample Preparation (200 ng DNA Samples)
Step 4. Purify the sample using Agencourt AMPure XP beads
3
Step 4. Purify the sample using Agencourt AMPure XP beads
1 Let the AMPure XP beads come to room temperature for at least 30
minutes.
2 Mix the reagent well so that the reagent appears homogeneous and
consistent in color. Do not freeze.
3 Add 180 μL of homogeneous AMPure XP beads to a 1.5-mL LoBind Tube,
and add the end-repaired DNA library (~100 μL). Mix well on a vortex
mixer and incubate for 5 minutes.
4 Put the tube in the magnetic stand. Wait for the solution to clear
(approximately 3 to 5 minutes).
5 Keep the tube in the magnetic stand. Do not touch the beads while you
carefully discard the cleared solution from the tubes.
6 Continue to keep the tube in the magnetic stand while you dispense 500 μL
of 70% Ethanol in each tube.
Use fresh 70% Ethanol for optimal result.
7 Let the tube sit for 1 minute to allow any disturbed beads to settle, and
remove the Ethanol.
8 Repeat step 6 and step 7 step once.
9 Spin briefly in a centrifuge. Return the tube to the magnetic stand for 30
seconds. Remove residual Ethanol with a P20 pipette.
10 Dry the samples on the 37°C heat block for 5 minutes or until the residual
Ethanol completely evaporates.
Do not dry the bead pellet to the point that the bead pellet appears cracked.
Elution efficiency is significantly decreased when the bead pellet is
excessively dried.
11 Add 32 μL Nuclease-free Water, mix well on a vortex mixer, and incubate for
2 minutes at room temperature.
12 Put the tube in the magnetic stand and leave for 2 to 3 minutes, until the
solution is clear.
Remove the supernatant (~32 μL) to a fresh 1.5-mL LoBind Tube. You can
discard the beads at this time.
Stopping Point
If you do not continue to the next step, store the samples at -20°C.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
47
Step 5. Add ‘A’ Bases to the 3' end of the DNA fragments
Use the SureSelect Library Prep Kit, ILM.
1 Prepare reaction mix and DNA sample (prepare on ice):
a Prepare the reaction mix in Table 22. Mix well on a vortex mixer.
b Add 20 μL of the reaction mix to each well or tube.
c Add 30 μL of each DNA sample to each well or tube. Mix by pipetting.
Change pipette tips between samples.
Table 22
Adding “A” Bases
Reagent
Volume for 1
Library
Volume for 12 Libraries
(includes excess)
Nuclease-free Water
11 µL
137.5 µL
10× Klenow Polymerase Buffer (blue cap)
5 µL
62.5 µL
dATP (green cap)
1 µL
12.5 µL
Exo(-) Klenow (red cap)
3 µL
37.5 µL
Total Volume
20 µL
250 µL (20 µL/sample)
2 Incubate in a thermal cycler for 30 minutes at 37°C.
If you use a heated lid, make sure that the lid temperature does not exceed
50°C.
Sample Preparation (200 ng DNA Samples)
Step 6. Purify the sample using Agencourt AMPure XP beads
3
Step 6. Purify the sample using Agencourt AMPure XP beads
1 Let the AMPure XP beads come to room temperature for at least 30
minutes.
2 Mix the reagent well so that the reagent appears homogeneous and
consistent in color. Do not freeze.
3 Add 90 μL of homogeneous AMPure XP beads to a 1.5-mL LoBind Tube, and
add the A-tailed DNA library (~50 μL). Mix well on a vortex mixer and
incubate for 5 minutes.
4 Put the tube in the magnetic stand. Wait for the solution to clear
(approximately 3 to 5 minutes).
5 Keep the tube in the magnetic stand. Do not touch the beads while you
carefully discard the cleared solution from the tubes.
6 Continue to keep the tube in the magnetic stand while you dispense 500 μL
of 70% Ethanol in each tube.
Use fresh 70% Ethanol for optimal result.
7 Let the tube sit for 1 minute to allow any disturbed beads to settle, and
remove the Ethanol.
8 Repeat step 6 and step 7 step once.
9 Spin briefly in a centrifuge. Return the tube to the magnetic stand for 30
seconds. Remove residual Ethanol with a P20 pipette.
10 Dry the samples on the 37°C heat block for 5 minutes or until the residual
Ethanol completely evaporates.
Do not dry the bead pellet to the point that the bead pellet appears cracked.
Elution efficiency is significantly decreased when the bead pellet is
excessively dried.
11 Add 15 μL of Nuclease-free Water, mix well on a vortex mixer, and incubate
for 2 minutes at room temperature.
12 Put the tube in the magnetic stand and leave for 2 to 3 minutes, until the
solution is clear.
13 Remove the supernatant (~15 μL) to a fresh 1.5-mL LoBind Tube. You can
discard the beads at this time.
14 Proceed immediately to the next step, “Step 7. Ligate the paired-end
adaptor”.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
49
Step 7. Ligate the paired-end adaptor
Use the SureSelect Library Prep Kit, ILM.
1 Dilute SureSelect Adaptor Oligo Mix (brown cap) 1:10 in Nuclease-free
Water immediately before use. Use the diluted Oligo mix in the next step.
2 Prepare reaction mix and DNA sample (prepare on ice):
a Prepare the reaction mix in Table 23.
b Add 37 μL of the reaction mix to each well or tube.
c Add 13 μL of each DNA sample to each well or tube. Mix by pipetting.
Change pipette tips between samples.
Table 23
Ligation master mix
Reagent
Volume for 1 Library
Volume for 12 Libraries
(includes excess)
Nuclease-free Water
15.5 µL
193.75 µL
5× T4 DNA Ligase Buffer
10 µL
125 µL
Diluted SureSelect Adaptor Oligo Mix
10 µL
125 µL
T4 DNA Ligase (red cap)
1.5 µL
18.75 µL
Total Volume
37 µL
462.5 µL (37 µL/sample)
3 Incubate for 15 minutes at 20°C on a thermal cycler. Do not use a heated
lid.
Stopping Point
If you do not continue to the next step, store the samples at -20°C.
Sample Preparation (200 ng DNA Samples)
Step 8. Purify the sample using Agencourt AMPure XP beads
3
Step 8. Purify the sample using Agencourt AMPure XP beads
1 Let the AMPure XP beads come to room temperature for at least 30
minutes.
2 Mix the reagent well so that the reagent appears homogeneous and
consistent in color. Do not freeze.
3 Add 90 μL of homogeneous AMPure XP beads to a 1.5-mL LoBind Tube, and
add the ligated library (~50 μL). Mix well on a vortex mixer and incubate
for 5 minutes.
4 Put the tube in the magnetic stand. Wait for the solution to clear
(approximately 3 to 5 minutes).
5 Keep the tube in the magnetic stand. Do not touch the beads while you
carefully discard the cleared solution from the tubes.
6 Continue to keep the tube in the magnetic stand while you dispense 500 μL
of 70% Ethanol in each tube.
Use fresh 70% Ethanol for optimal result.
7 Let the tube sit for 1 minute to allow any disturbed beads to settle, and
remove the Ethanol.
8 Repeat step 6 and step 7 step once.
9 Spin briefly in a centrifuge. Return the tube to the magnetic stand for 30
seconds. Remove residual Ethanol with a P20 pipette.
10 Dry the samples on the 37°C heat block for 5 minutes until the residual
Ethanol is completely evaporated.
Do not dry the bead pellet to the point that the bead pellet appears cracked.
Elution efficiency is significantly decreased when the bead pellet is
excessively dried.
11 Add 32 μL of Nuclease-free Water, mix well on a vortex mixer, and incubate
for 2 minutes at room temperature.
12 Put the tube in the magnetic stand and leave for 2 to 3 minutes, until the
solution is clear.
13 Remove the supernatant (~32 μL) to a fresh 1.5-mL LoBind Tube. You can
discard the beads at this time.
Stopping Point
If you do not continue to the next step, store the samples at -20°C.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
51
Step 9. Amplify adaptor-ligated library
Use reagents from these kits:
• SureSelect Library Prep Kit, ILM
• SureSelect Target Enrichment Kit ILM Indexing Hyb Module Box #2
• Herculase II Fusion DNA Polymerase (Agilent)
CAUTION
To avoid cross-contaminating libraries, set up PCR reactions (all components except
the library DNA) in a dedicated clean area or PCR hood with UV sterilization and
positive air flow.
1 Prepare reaction mix and DNA sample:
a Prepare the reaction mix in Table 24, on ice. Mix well on a vortex mixer.
b Add 20 μL of the reaction mix to each well or tube.
c Add 30 μL of each DNA sample to each well or tube. Mix by pipetting.
Change pipette tips between samples.
Sample Preparation (200 ng DNA Samples)
Step 9. Amplify adaptor-ligated library
Table 24
3
Components for PCR mix
Reagent
Volume for 1
reaction
Volume for 12 reactions
(includes excess)
Nuclease-free Water
6 μL
75 μL
SureSelect Primer (brown cap)*
1.25 µL
15.625 µL
SureSelect ILM Indexing Pre Capture PCR
Reverse Primer (clear cap)†
1.25 µL
15.625 µL
5× Herculase II Rxn Buffer (clear cap)‡
10 μL
125 μL
0.5 μL
6.25 μL
1 μL
12.5 μL
20 μL
250 µL 
(20 µL/reaction)
100 mM dNTP Mix (green cap)‡
Herculase II Fusion DNA Polymerase (red
cap)‡
Total
* Included in the SureSelect Library Prep Kit, ILM.
† Included in the SureSelect Target Enrichment Kit ILM Indexing Hyb Module Box #2.
‡ Included in the Herculase II Fusion DNA Polymerase (Agilent) kit. Do not use the buffer or dNTP
mix from any other kit.
2 Run the program in Table 25 in a thermal cycler.
Table 25
PCR program
Step
Temperature
Time
Step 1
98°C
2 minutes
Step 2
98°C
30 seconds
Step 3
65°C
30 seconds
Step 4
72°C
1 minute
Step 5
Repeat Step 2 through Step 4 for a total of 10 times.
Step 6
72°C
10 minutes
Step 7
4°C
Hold
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
53
Step 10. Purify the sample with Agencourt AMPure XP beads
1 Let the AMPure XP beads come to room temperature for at least 30
minutes.
2 Mix the reagent well so that the reagent appears homogeneous and
consistent in color. Do not freeze.
3 Add 90 μL of homogeneous AMPure XP beads to a 1.5-mL LoBind Tube, and
add the amplified library (~50 μL). Mix well on a vortex mixer and incubate
for 5 minutes.
4 Put the tube in the magnetic stand. Wait for the solution to clear
(approximately 3 to 5 minutes).
5 Keep the tube in the magnetic stand. Do not touch the beads while you
carefully discard the cleared solution from the tubes.
6 Continue to keep the tube in the magnetic stand while you dispense 500 μL
of 70% Ethanol in each tube.
Use fresh 70% Ethanol for optimal result.
7 Let the tube sit for 1 minute to allow any disturbed beads to settle, and
remove the Ethanol.
8 Repeat step 6 and step 7 step once.
9 Spin briefly in a centrifuge. Return the tube to the magnetic stand for 30
seconds. Remove residual Ethanol with a P20 pipette.
10 Dry the samples on the 37°C heat block for 5 minutes or until the residual
Ethanol completely evaporates.
Do not dry the bead pellet to the point that the bead pellet appears cracked.
Elution efficiency is significantly decreased when the bead pellet is
excessively dried.
11 Add 30 μL of Nuclease-free Water, mix well on a vortex mixer, and incubate
for 2 minutes at room temperature.
12 Put the tube in the magnetic stand and leave for 2 to 3 minutes, until the
solution is clear.
13 Remove the supernatant (~30 μL) to a fresh 1.5-mL LoBind Tube. You can
discard the beads at this time.
Stopping Point
If you do not continue to the next step, store the samples at -20°C.
Sample Preparation (200 ng DNA Samples)
Step 11. Assess quality and quantity
3
Step 11. Assess quality and quantity
Quality assessment can be done with either the 2100 Bioanalyzer instrument
or the 2200 TapeStation.
2100 Bioanalyzer System and DNA 1000 Assay
1 Check that the 2100 Bioanalyzer electrodes have been cleaned as instructed
in the reagent kit guide.
2 Open the Agilent 2100 Expert Software (version B.02.02 or higher), turn on
the 2100 Bioanalyzer and check communication.
3 Prepare the chip, samples and ladder as instructed in the reagent kit guide.
4 Load the prepared chip into the 2100 Bioanalyzer and start the run within
five minutes after preparation.
5 Within the instrument context, choose the appropriate assay from the drop
down list.
6 Start the run. Enter sample names and comments in the Data and Assay
context.
7 Verify the results. Check that the electropherogram shows a distribution
with a peak size approximately 225 to 275 bp. Measure the concentration of
the library by integrating under the peak.
Figure 8
Analysis of amplified prepped library DNA using a DNA 1000 assay. The electropherogram shows a single peak in the size range of 225 to 275 bp.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
55
2200 TapeStation and D1000 ScreenTape
You can use the 2200 TapeStation to analyze the amplified libraries. Use the
D1000 ScreenTape and D1000 Reagents. For more information to do this step,
see the Agilent 2200 TapeStation User Manual.
1 Prepare the TapeStation samples as instructed in the Agilent 2200
TapeStation User Manual. Use 1 μL of each amplified library DNA sample
diluted with 3 μL of D1000 sample buffer for the analysis.
CAUTION
Make sure that you thoroughly mix the combined DNA and D1000 sample buffer on a
vortex mixer for 5 seconds or more for accurate quantitation.
2 Load the sample plate or tube strips from step 1, the D1000 ScreenTape,
and loading tips into the 2200 TapeStation as instructed in the Agilent 2200
TapeStation User Manual. Start the run.
3 Verify that the electropherogram shows a distribution with a peak size
approximately 225 to 275 bp. A sample electropherogram is shown in
Figure 9.
Stopping Point
If you do not continue to the next step, seal the sheared DNA sample plate and
store at 4°C overnight or at –20°C for prolonged storage.
Sample Preparation (200 ng DNA Samples)
Step 11. Assess quality and quantity
Figure 9
3
Analysis of amplified library DNA using the 2200 TapeStation with a D1000
ScreenTape. The electropherogram shows a single peak in the size range of
225 to 275 bp.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
57
This page intentionally left blank.
58
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
SureSelectXT Target Enrichment System for Illumina Paired-End
Sequencing Protocol
4
Hybridization
Step 1. Hybridize the library 62
Step 2. Prepare magnetic beads 68
Step 3. Select hybrid capture with SureSelect
69
This chapter describes the steps to combine the prepped library with the
hybridization reagents, blocking agents and the SureSelect capture library.
Each DNA library sample must be hybridized and captured individually prior
to addition of the indexing tag by PCR.
CAUTION
The ratio of SureSelect capture library to prepped library is critical for successful
capture.
59
4

Hybridization
GENOMIC SAMPLE
(Set of chromosomes)
SureSelect™
Target Enrichment System
Capture Process
S¢ŽEEcE3™ Kit
+
GENOMIC SAMPLE (PREPPED)
+
SureSelect HYB BUFFER
SureSelect
BIOTINYLATED RNA LIBRARY
“BAITS”
Hybridization
STREPTAVIDIN COATED MAGNETIC BEADS
+
Bead capture
UNBOUND FRACTION
DISCARDED
Wash Beads
and
Digest RNA
gyc`N§–
<<—¢c™`ycE¦—!U
Sequencing
ƒ¢`goc!Ž—yooc`lU—
oN—g¢c™`ycE¦E<—‘!gycE‘
Figure 10
60
SureSelect Target Enrichment System Capture Process
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
Hybridization
4
Refer to “SureSelect Reagent Kit Content” on page 84 for a complete content
listing of each SureSelect Target Enrichment kit.
CAUTION
You must avoid evaporation from the small volumes of the capture during the 16 or 24
hour incubation.
If you want to use a different combination of thermal cycler, lid temperature, plates or
strips, and sealing method (strip caps or sealing tape), first test the conditions.
Incubate 27 µL of SureSelect Hybridization Buffer (without DNA) at 65°C for 24 hours
as a test. Include buffer in each well that you might use, including those in the center
and those on the edges. Check that you do not get extensive evaporation. Evaporation
should not exceed 3 to 4 µL.
For a partial list of tested options showing minimal evaporation, refer to “Alternative
Capture Equipment Combinations” on page 88.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
61
4
Hybridization
Step 1. Hybridize the library

Step 1. Hybridize the library
For each sample library prepared, do one hybridization and capture. Do not
pool samples at this stage.
The hybridization reaction requires 750 ng of DNA with a maximum volume of
3.4 μL.
1 If the prepped library concentration is below 221 ng/μL, use a vacuum
concentrator to concentrate the sample at  45°C.
a Add the entire 30 μL of prepped library to an Eppendorf tube. Poke one
or more holes in the lid with a narrow gauge needle.
You can also break off the cap, cover with parafilm, and poke a hole in
the parafilm.
b Completely lyophilize. Use a vacuum concentrator on low heat (less than
45°C) to dehydrate.
c Reconstitute with nuclease-free water to bring the final concentration to
221 ng/μL (or greater if sample recovery is of concern). Pipette up and
down along the sides of the tube for optimal recovery.
d Mix well on a vortex mixer and spin in a microfuge for 1 minute.
2 Optional. To test recovery after lyophilization, reconstitute the sample to
greater than 221 ng/μL and check the concentration on a Bioanalyzer DNA
1000 chip. See “Step 12. Assess quality and quantity” on page 35. After
quantitation, adjust the sample to 221 ng/μL.
Alternatively, concentrate a 750 ng aliquot at  45°C down to 3.4 μL. If the
sample dries up completely, resuspend in 3.4 μL of water and mix on a
vortex mixer. If processing multiple samples, adjust to equivalent volumes
before concentrating.
3 Mix the components in Table 26 at room temperature to prepare the
hybridization buffer.
62
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
Hybridization
Step 1. Hybridize the library
Table 26
4
Hybridization Buffer
Reagent
Volume for 1
Volume for 6
capture (µL),
captures (µL),
includes excess includes excess
Volume for 12
captures (µL),
includes excess
SureSelect Hyb #1 (orange cap, or
bottle)
25
125
250
SureSelect Hyb #2 (red cap)
1
5
10
SureSelect Hyb #3 (yellow cap)
10
50
100
SureSelect Hyb #4 (black cap, or
bottle)
13
65
130
Total
49 (40 µL
needed)
245 (40
µL/sample)
490 (40
µL/sample)
4 If precipitate forms, warm the hybridization buffer at 65°C for 5 minutes.
5 In a PCR plate, prepare the SureSelect capture library mix for target
enrichment:
a Keep capture library mix tubes on ice during preparation and until they
are used in step 10.
b For each sample, add the amount of SureSelect capture library as listed
in Table 27, based on the Mb target size of your design.
c Use nuclease-free water to prepare a dilution of the SureSelect RNase
Block (purple cap) as listed in Table 27.
Prepare enough RNase Block dilution for all samples, plus excess.
d Add the amount of diluted SureSelect RNase Block (purple cap) listed in
Table 27 to each capture library, and mix by pipetting.
Table 27
SureSelect Capture Library.
Capture Size
Volume of SureSelect
Library
RNase Block Dilution
(Parts RNase block:
Parts water)
Volume of RNase
Block Dilution to Add
 3.0 Mb
2 µL
1:9 (10%)
5 µL
 3.0 Mb
5 µL
1:3 (25%)
2 µL
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
63
4

Hybridization
Step 1. Hybridize the library
6 Mix the contents in Table 28 to make the correct amount of SureSelect
Block mix for the number of samples used.
Table 28
SureSelect Block Mix
Reagent
Volume for 1 reaction
Volume for 12
reactions (includes
excess)
SureSelect Indexing Block #1 (green cap)
2.5 µL
31.25 µL
SureSelect Block #2 (blue cap)
2.5 µL
31.25 µL
SureSelect Indexing Block #3 (brown cap)
0.6 µL
7.5 µL
Total
5.6 µL
70 µL
7 In a separate PCR plate, prepare the prepped library for target enrichment.
a Add 3.4 μL of 221 ng/μL prepped library to the “B” row in the PCR plate.
Put each sample into a separate well.
b Add 5.6 μL of the SureSelect Block Mix to each well in row B.
c Mix by pipetting up and down.
d Seal the wells of row “B” with caps and put the PCR plate in the thermal
cycler. Do not heat the Hybridization Buffer or capture library yet, only
the prepped library with blockers.
e Start the thermal cycler program in Table 29.
Prepped Library
Figure 11
64
Prepped library shown in red
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
Hybridization
Step 1. Hybridize the library
Table 29
4
PCR program
Step
Temperature
Time
Step 1
95°C
5 minutes
Step 2
65°C
Hold
8 Use a heated lid on the thermal cycler at 105°C to hold the temperature of
the plate on the thermal cycler at 65°C.
CAUTION
The lid of the thermal cycler is hot and can cause burns. Use caution when working
near the lid.
9 Add the hybridization buffer to the PCR plate:
a Maintain the plate at 65°C while you load 40 μL of hybridization buffer
per well into the “A” row of the PCR plate. The number of wells filled in
Row A is the number of libraries prepared.
The example in Figure 12 is for 12 captures.
Hybridization Buffer
Prepped Library
Figure 12
Hybridization buffer shown in blue
b Seal the wells with strip caps. Use a capping tool to make sure the fit is
tight.
c Keep the plate at 65°C for a minimum of 5 minutes before you go to
step 10.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
65
4

Hybridization
Step 1. Hybridize the library
10 Add the capture library mix from step 5 to the PCR plate:
a Add the capture library mix (7 μL) to the “C” row in the PCR plate.
For multiple samples, use a multi-channel pipette to load the capture
library mix into the “C” row in the PCR plate.
Keep the plate at 65°C during this time.
b Seal the wells with strip caps. Use a capping tool to make sure the fit is
tight.
c Incubate the samples at 65°C for 2 minutes.
11 Maintain the plate at 65°C while you use a multi-channel pipette to take 13
μL of Hybridization Buffer from the “A” row and add it to the SureSelect
capture library mix contained in row “C” of the PCR plate for each sample.
(See Figure 13.)
step 11
step 12
Hybridization Buffer
Prepped Library
SureSelect
Capture Library
Figure 13
SureSelect Capture Library, or “Baits”, shown in Green
12 Maintain the plate at 65°C while you use a multi-channel pipette to transfer
the entire contents of each prepped library mix in row “B” to the
hybridization solution in row “C”. (See Figure 13.) Mix well by slowly
pipetting up and down 8 to 10 times.
The hybridization mixture is now 27 to 29 μL, depending on degree of
evaporation during the preincubations.
13 Seal the wells with strip caps or double adhesive film. Make sure all wells
are completely sealed.
Use new adhesive seals or strip caps. The structural integrity of the seals
and caps can be compromised during the previous incubation steps.
66
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
Hybridization
Step 1. Hybridize the library
CAUTION
4
Wells must be adequately sealed to minimize evaporation, or your results can be
negatively impacted. You can use the PlateLoc Thermal Microplate Sealer to minimize
evaporation.
If you use strip tubes, test for evaporation before you do the first
experiment to make sure the tube/capping method is appropriate for the
thermal cycler. Check that no more than 3 to 4 μL is lost to evaporation.
14 Incubate the hybridization mixture for 16 or 24 hours at 65°C with a heated
lid at 105°C.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
67
4
Hybridization
Step 2. Prepare magnetic beads

Step 2. Prepare magnetic beads
Use these reagents from the SureSelect Target Enrichment Kit Box #1:
• SureSelect Binding Buffer
• SureSelect Wash 2
1 Prewarm SureSelect Wash 2 at 65°C in a circulating water bath or heat
block for use in “Step 3. Select hybrid capture with SureSelect”.
2 Vigorously resuspend the Dynabeads MyOne Streptavidin T1 on a vortex
mixer. Magnetic beads settle during storage.
3 For each hybridization, add 50 μL of Dynabeads MyOne Streptavidin T1 to a
1.5-mL LoBind Tube.
4 Wash the beads:
a Add 200 μL of SureSelect Binding Buffer.
b Mix the beads on a vortex mixer for 5 seconds.
c Put the tubes into a magnetic device, such as the Dynal magnetic
separator (Life Technologies).
d Remove and discard the supernatant.
e Repeat step a through step d for a total of 3 washes.
5 Resuspend the beads in 200 μL of SureSelect Binding Buffer.
68
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
Hybridization
Step 3. Select hybrid capture with SureSelect
4
Step 3. Select hybrid capture with SureSelect
Use these reagents from the SureSelect Target Enrichment Kit Box #1:
• SureSelect Wash 1
• SureSelect Wash 2
1 Estimate and record the volume of hybridization that remained after 16 or
24 hour incubation.
2 Keep the PCR plate or tubes at 65°C in the PCR machine while you add the
hybridization mixture directly from the thermal cycler to the bead solution.
Invert the tube to mix 3 to 5 times.
Excessive evaporation, such as when less than 20 μL remains after
hybridization, can indicate suboptimal capture performance. See Table 43
on page 88 for tips to minimize evaporation.
3 Incubate the hybrid-capture/bead solution on a Nutator or equivalent for 30
minutes at room temperature.
Make sure the sample is properly mixing in the tube.
4 Briefly spin in a centrifuge.
5 Separate the beads and buffer on a magnetic separator and remove the
supernatant.
6 Resuspend the beads in 500 μL of SureSelect Wash 1 by mixing on a vortex
mixer for 5 seconds.
7 Incubate the samples for 15 minutes at room temperature. Occasionally mix
on a vortex mixer.
8 Briefly spin in a centrifuge.
9 Separate the beads and buffer on a magnetic separator and remove the
supernatant.
10 Wash the beads:
a Resuspend the beads in 500 μL of 65°C prewarmed SureSelect Wash 2
and mix on a vortex mixer for 5 seconds to resuspend the beads.
b Incubate the samples for 10 minutes at 65°C in a recirculating water
bath, heat block or equivalent. Occasionally mix on a vortex mixer.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
69
4
Hybridization
Step 3. Select hybrid capture with SureSelect

Do not use a tissue incubator. It cannot properly maintain temperature.
c Briefly spin in a centrifuge.
d Separate the beads and buffer on a magnetic separator and remove the
supernatant.
e Repeat step a through step d for a total of 3 washes.
Make sure all of the wash buffer has been removed.
11 Mix the beads in 30 μL of nuclease-free water on a vortex mixer for
5 seconds to resuspend the beads.
x
NOTE
70
Captured DNA is retained on the streptavidin beads during the post-capture amplification
step.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
SureSelectXT Target Enrichment System for Illumina Paired-End
Sequencing Protocol
5
Addition of Index Tags by
Post-Hybridization Amplification
Step 1. Amplify the captured library to add index tags 72
Step 2. Purify the sample using Agencourt AMPure XP beads 76
Step 3. Assess quality 77
Step 4. Assess the quantity of each index-tagged library by QPCR
(optional) 80
Step 5. Pool samples for Multiplexed Sequencing 81
Step 6. Prepare sample for cluster amplification 82
This chapter describes the steps to add index tags by amplification, purify,
assess quality and quantity of the libraries, dilute the sample appropriately for
cluster amplification, and pool indexed samples for multiplexed sequencing.
71
5
Addition of Index Tags by Post-Hybridization Amplification
Step 1. Amplify the captured library to add index tags

Step 1. Amplify the captured library to add index tags
Use reagents from:
• Herculase II Fusion DNA Polymerase (Agilent)
• SureSelect Target Enrichment Kit ILM Indexing Hyb Module Box #2
• SureSelect Library Prep Kit, ILM
CAUTION
Do not use amplification enzymes other than Herculase II Fusion DNA Polymerase.
Other enzymes have not been validated.
CAUTION
To avoid cross-contaminating libraries, set up PCR reactions (all components except
the library DNA) in a dedicated clean area or PCR hood with UV sterilization and
positive air flow.
Prepare 1 amplification reaction for each hybrid capture. Include a negative
no-template control.
To see the nucleotide sequence in each of the index included in SureSelect
reagent kits, see “SureSelectXT Indexes for Illumina” on page 87.
1 Prepare reaction mix and DNA samples:
a Prepare the reaction mix in Table 30, on ice. Mix well on a vortex mixer.
b Add 35 μL of the reaction mix to each well or tube.
c Add 1 μL of the appropriate index PCR Primer Index 1 through Index 16
(clear caps) from the SureSelect Library Prep Kit, ILM to each well and
mix by pipetting.
72
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
Addition of Index Tags by Post-Hybridization Amplification
Step 1. Amplify the captured library to add index tags
5
Use a different index primer for each sample to be sequenced in the same
lane. Use Table 31 as a guide to determine the number of indexes to pool
per sequencing lane.
d Pipette each DNA sample up and down to make sure that the bead
solution is homogeneous.
e Use a pipette to add 14 μL of each DNA sample to each well or tube. Mix
by pipetting. Change pipette tips between samples to avoid
cross-contamination.
If needed, store the remaining bead at -20°C for future use.
Table 30
Herculase II Master Mix
Reagent
Volume for 1
reaction
Volume for 12 reactions
(includes excess)
22.5 µL
281.25 µL
10 µL
125 µL
1 µL
12.5 µL
0.5 µL
6.25 µL
SureSelect ILM Indexing Post Capture Forward
PCR Primer (orange cap)†
1 µL
12.5 µL
Total
35 µL
437.5 µL
(35 µL/reaction)
Nuclease-free water
5× Herculase II Rxn Buffer (clear cap)
*
Herculase II Fusion DNA Polymerase (red cap)*
100 mM dNTP Mix (green
cap)*
* Included in the Herculase II Fusion DNA Polymerase (Agilent). Do not use the buffer or dNTP mix
from any other kit.
† Included in the SureSelect Target Enrichment Kit ILM Indexing Hyb Module Box #2.
Table 31
Sequencing data requirement guidelines
Capture size
Optimal sequencing output per index
1 kb up to 0.5 Mb
0.1 to 50 Mb*
0.5 Mb up to 2.9 Mb
50 to 290 Mb*
3 Mb up to 5.9 Mb
300 to 590 Mb*
6 Mb up to 11.9 Mb
600 to 1190 Mb*
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
73
5

Addition of Index Tags by Post-Hybridization Amplification
Step 1. Amplify the captured library to add index tags
Table 31
Sequencing data requirement guidelines
Capture size
Optimal sequencing output per index
12 Mb up to 24 Mb
1.2 to 2.4 Gb*
Human All Exon v5
4 Gb
Human All Exon v5 + UTRs
6 Gb
Human All Exon 50 Mb
5 Gb
Human DNA Kinome
320 Mb
Mouse All Exon
5 Gb
* For custom libraries, Agilent recommends analyzing 100× amount of sequencing data compared to
the Capture Library size for each sample. Pool samples according to your expected sequencing output.
2 Put the tubes in a thermal cycler and run the program in Table 32.
Store the unused bead-bound captured DNA samples at 4°C overnight, or at
-20°C long-term.
Table 32
PCR program
Step
Temperature
Time
Step 1
98°C
2 minutes
Step 2
98°C
30 seconds
Step 3
57°C
30 seconds
Step 4
72°C
1 minute
Step 5
74
• Repeat Step 2 through Step 4 for a total of 12
to 16 times.
Step 6
72°C
10 minutes
Step 7
4°C
Hold
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
Addition of Index Tags by Post-Hybridization Amplification
Step 1. Amplify the captured library to add index tags
5
As with the pre-capture PCR amplification, minimize the number of PCR
cycles used to enrich the captured DNA. The use of only half of the captured
DNA for amplification lets you adjust the number of cycles by repeating the
PCR if needed.
Table 33
Number of cycles
Capture Size
Cycles
1 kb up to 0.5 Mb
16 cycles
0.5 Mb up to 1.49 Mb
14 cycles
> 1.5 Mb
12 cycles
All Exon
10 to 12 cycles
As an alternative, you can prepare one PCR master mix as outlined in
Table 30. Split this master mix into three small-scale 10-μL PCR reactions and
cycle for 10, 12, 14 or 16 cycles. Clean these PCR reactions using the AMPure
XP protocol outlined in “Step 2. Purify the sample using Agencourt AMPure
XP beads” on page 21 with these modifications: use 30 μL of AMPure XP beads
and elute with 20 μL of nuclease-free water. Run these cleaned samples on a
DNA 1000 chip on the Bioanalyzer, as described in “Step 3. Assess quality
(optional)” on page 22.
Use the optimal cycle number to repeat PCR at the 50-μL reaction scale. See
Table 33 for approximate number of cycles for a given library size. Results may
vary based on library content.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
75
5
Addition of Index Tags by Post-Hybridization Amplification
Step 2. Purify the sample using Agencourt AMPure XP beads

Step 2. Purify the sample using Agencourt AMPure XP beads
1 Let the AMPure XP beads come to room temperature for at least 30
minutes.
2 Mix the reagent well so that the reagent appears homogeneous and
consistent in color. Do not freeze.
3 Add 90 μL of homogeneous AMPure beads to a 1.5-mL LoBind Tube, and
add amplified library (~50 μL). Mix well on a vortex mixer and incubate for
5 minutes.
4 Put the tube in the magnetic stand. Wait for the solution to clear
(approximately 3 to 5 minutes).
5 Keep the tube in the magnetic stand. Do not touch the beads while you
carefully discard the cleared solution from the tubes.
6 Continue to keep the tube in the magnetic stand while you dispense 500 μL
of 70% Ethanol in each tube.
Use fresh 70% Ethanol for optimal results.
7 Let the tube sit for 1 minute to allow any disturbed beads to settle, and
remove the Ethanol.
8 Repeat step 6 and step 7 once.
9 Spin briefly in a centrifuge. Return the tube to the magnetic stand for 30
seconds. Remove residual Ethanol with a P20 pipette.
10 Dry the samples on the 37°C heat block for 5 minutes or until the residual
Ethanol is completely evaporated.
Do not dry the bead pellet to the point that the bead pellet appears cracked.
Elution efficiency is significantly decreased when the bead pellet is
excessively dried.
11 Add 30 μL of Nuclease-free Water or 1X Low TE Buffer, mix well on a vortex
mixer, and incubate for 2 minutes at room temperature.
12 Put the tube in the magnetic stand and leave for 2 to 3 minutes, until the
solution is clear.
13 Remove the supernatant (~30 μL) to a fresh 1.5-mL LoBind Tube. You can
discard the beads at this time.
Stopping Point
76
If you do not continue to the next step, store the samples at 4°C for up to a
week, or at -20°C for longer periods.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
Addition of Index Tags by Post-Hybridization Amplification
Step 3. Assess quality
5
Step 3. Assess quality
Use a Bioanalyzer High Sensitivity DNA Assay or the 2200 TapeStation to
assess the quality and size range.
2100 Bioanalyzer High Sensitivity DNA Assay
You may need to dilute your sample accordingly. Refer to the Agilent High
Sensitivity DNA Kit Guide at http://www.chem.agilent.com/en-US
/Search/Library/_layouts/Agilent/PublicationSummary.aspx?whid=59504.
1 Check that the 2100 Bioanalyzer electrodes have been cleaned as instructed
in the reagent kit guide.
2 Open the Agilent 2100 Expert Software (version B.02.07 or higher required
to run the High Sensitivity Kit), turn on the 2100 Bioanalyzer and check
communication.
3 Prepare the chip, samples and ladder as instructed in the reagent kit guide.
4 Load the prepared chip into the 2100 Bioanalyzer and start the run within
five minutes after preparation.
5 Within the instrument context, choose the appropriate assay from the drop
down list.
6 Start the run. Enter sample names and comments in the Data and Assay
context.
7 Verify the results.
Determine the concentration of the sample by integration under the peak.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
77
5
Addition of Index Tags by Post-Hybridization Amplification
Step 3. Assess quality
Figure 14

Analysis of Amplified Capture DNA using the High Sensitivity DNA Kit. The
electropherogram shows a peak in the size range of approximately 250 to
350 bp.
2200 TapeStation and High Sensitivity D1000 ScreenTape
Use the 2200 TapeStation to analyze the barcoded DNA. Use the High
Sensitivity D1000 ScreenTape and High Sensitivity D1000 Reagents. For more
information to do this step, see the Agilent 2200 TapeStation User Manual.
1 Prepare the TapeStation samples as instructed in the Agilent 2200
TapeStation User Manual. Use 2 μL of each amplified library DNA sample
diluted with 2 μL of High-Sensitivity D1000 sample buffer for the analysis.
CAUTION
Make sure that you thoroughly mix the combined DNA and High-Sensitivity D1000
sample buffer on a vortex mixer for 5 seconds or more for accurate quantitation.
2 Load the sample plate or tube strips from step 1, the High Sensitivity D1000
ScreenTape, and loading tips into the 2200 TapeStation as instructed in the
Agilent 2200 TapeStation User Manual. Start the run.
3 Verify that the electropherogram shows an average DNA fragment size of
about 300 bp. A sample electropherogram is shown in Figure 15.
Stopping Point
78
If you do not continue to the next step, seal the sample plate and store at 4°C
overnight or at –20°C for prolonged storage.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
Addition of Index Tags by Post-Hybridization Amplification
Step 3. Assess quality
Figure 15
5
Analysis of amplified library DNA using the 2200 TapeStation with a High
Sensitivity D1000 ScreenTape. The electropherogram shows a peak size range
of 250 to 350 bp.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
79
5
Addition of Index Tags by Post-Hybridization Amplification
Step 4. Assess the quantity of each index-tagged library by QPCR (optional)

Step 4. Assess the quantity of each index-tagged library by QPCR
(optional)
Refer to the protocol that is included with the QPCR NGS Library
Quantification Kit (Illumina GA) for more details to do this step.
1 Use the QPCR NGS Library Quantification Kit (Illumina GA) to determine
the concentration of each index-tagged captured library.
2 Prepare a standard curve using the quantification standard included in the
kit, according to the instructions provided in the user guide.
3 Dilute each index-tagged captured library such that it falls within the range
of the standard curve.
Typically this corresponds to approximately a 1:1000 to 1:10,000 dilution of
the captured DNA.
4 Prepare the QPCR master mix with Illumina adaptor-specific PCR primers
according to instructions provided in the kit.
5 Add an aliquot of the master mix to PCR tubes and add template.
6 On a QPCR system, such as the MX3005P, run the thermal profile outlined
in the QPCR NGS Library Quantification kit user guide. Use the SYBR
Green instrument setting.
7 Use the standard curve to determine the concentration of each unknown
index-tagged library, in nM.
The concentration will be used to accurately pool samples for multiplexed
sequencing.
80
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
Addition of Index Tags by Post-Hybridization Amplification
Step 5. Pool samples for Multiplexed Sequencing
5
Step 5. Pool samples for Multiplexed Sequencing
1 Combine the libraries such that each index-tagged sample is present in
equimolar amounts in the pool. For each library, use the formula below to
determine the amount of index sample to use.
Vf  Cf
Volume of Index = ------------------------------ where
#  Ci
V(f) is the final desired volume of the pool,
C(f) is the desired final concentration of all the DNA in the pool, for
example, 10 nM for the standard Illumina protocol.
# is the number of index, and
C(i) is the initial concentration of each index sample.
Table 34 shows an example of the amount of 4 index-tagged (of different
concentrations) and 1X Low TE Buffer needed for a final volume of 20 μL at
10 nM.
Table 34
Example of index volume calculation for a total volume of 20 µL
Component
V(f)
C(i)
C(f)
#
Volume to use (µL)
Sample 1
20 µL
20 nM
10 nM
4
2.5
Sample 2
20 µL
10 nM
10 nM
4
5
Sample 3
20 µL
17 nM
10 nM
4
2.9
Sample 4
20 µL
25 nM
10 nM
4
2
1X Low TE Buffer
7.6
2 Adjust the final volume of the pooled library to the desired final
concentration.
• If the final volume of the combined index-tagged samples is less than the
desired final volume, V(f), add 1X Low TE Buffer to bring the volume to the
desired level.
• If the final volume of the combined index-tagged samples is greater than
the final desired volume, V(f), lyophilize and reconstitute to the desired
volume.
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
81
5
Addition of Index Tags by Post-Hybridization Amplification
Step 6. Prepare sample for cluster amplification

3 If you store the library before sequencing, add Tween 20 to 0.1% v/v and
store at -20°C short term.
4 Proceed to template denaturation and flow cell preparation. Refer to the
appropriate Illumina protocol.
Exact library pool dilution and processing can vary based on the flow cell
capacity and analysis pipeline versions being used. Refer to the appropriate
Illumina user guide for instructions. This protocol has been validated with
100-base paired-end reads. However, read length can be adjusted to achieve
the desired research goals.
Step 6. Prepare sample for cluster amplification
• Use the appropriate Illumina Paired-End Cluster Generation Kit to do
cluster amplification.
Refer to the instructions that are included with the Illumina Paired-End
Cluster Generation Kit. The optimal seeding concentration for SureSelectXT
libraries is 6 to pM, depending on the desired output and data quality.
82
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
SureSelectXT Target Enrichment System for Illumina Paired-End
Sequencing Protocol
6
Reference
SureSelect Reagent Kit Content 84
Other Reagent Kits Content 86
SureSelectXT Indexes for Illumina 87
Alternative Capture Equipment Combinations
88
This chapter contains reference information.
83
6

Reference
SureSelect Reagent Kit Content
SureSelect Reagent Kit Content
NOTE
SureSelect capture libraries and reagents must be used within one year of receipt.
Each SureSelect Reagent Kit contains one or more of each of these kits:
Table 35
SureSelect Reagent Kit Contents
Product
Storage
Condition
16
96
Reactions Reactions
480
Reactions
SureSelect Target Enrichment Kit
Box #1
Room
Temperature
5190-4393 5190-4394
5190-4395
SureSelect Target Enrichment Kit ILM
Indexing Hyb Module Box #2
-20°C
5190-4455 5190-4456
5190-4457
SureSelect Library Prep Kit, ILM
-20°C
5500-0105 5500-0075
The content of each of these kits are described in the next tables.
Table 36
SureSelect Target Enrichment Kit Box #1
Kit Component
SureSelect Hyb #1 (orange cap, or bottle)
SureSelect Hyb #2 (red cap)
SureSelect Hyb #4 (black cap, or bottle)
SureSelect Binding Buffer
SureSelect Wash 1
SureSelect Wash 2
SureSelect Elution Buffer*
SureSelect Neutralization Buffer*
* These reagents are not used in this protocol.
84
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
Reference
SureSelect Reagent Kit Content
Table 37
6
SureSelect Target Enrichment Kit ILM Indexing Hyb Module Box #2
Kit Component
SureSelect Hyb #3 (yellow cap)
SureSelect Indexing Block #1 (green cap)
SureSelect Block #2 (blue cap)
SureSelect Indexing Block #3 (brown cap)
SureSelect RNase Block (purple cap)
SureSelect ILM Indexing Pre Capture PCR Reverse Primer (clear cap)
SureSelect ILM Indexing Post Capture Forward PCR Primer (orange cap)
Table 38
SureSelect Library Prep Kit, ILM
Kit Component
10× End Repair Buffer (clear cap)
T4 Polynucleotide Kinase (orange cap)
10× Klenow Polymerase Buffer (blue cap)
T4 DNA Ligase (red cap)
Exo(-) Klenow (red cap)
T4 DNA Polymerase (purple cap)
Klenow DNA Polymerase (yellow cap)
dATP (green cap)
dNTP Mix (green cap)
SureSelect Adaptor Oligo Mix (brown cap)
SureSelect Primer (brown cap)
PCR Primer Index 1 through Index 16 (clear caps)
5× T4 DNA Ligase Buffer
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
85
6
Reference
Other Reagent Kits Content

Other Reagent Kits Content
These reagents are from kits other than the SureSelect Reagent kit. Make sure
you use only the reagents listed here.
Table 39
Herculase II Fusion DNA Polymerase (Agilent)
Component
DMSO (green cap)
5× Herculase II Rxn Buffer (clear cap)
100 mM dNTP Mix (green cap)
Herculase II Fusion DNA Polymerase (red cap)
Table 40
D1000 Reagents (Agilent p/n 5067-5583)
Components
D1000 ladder
D1000 sample buffer
Table 41
High Sensitivity D1000 Reagents (Agilent p/n 5067-5585)
Components
High-Sensitivity D1000 ladder
High-Sensitivity D1000 sample buffer
86
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
Reference
SureSelectXT Indexes for Illumina
6
SureSelectXT Indexes for Illumina
The nucleotide sequence of each of the SureSelectXT index is listed in
Table 42.
Table 42
SureSelectXT Indexes 1-16
Index Number
Sequence
1
ATCACG
2
CGATGT
3
TTAGGC
4
TGACCA
5
ACAGTG
6
GCCAAT
7
CAGATC
8
ACTTGA
9
GATCAG
10
TAGCTT
11
GGCTAC
12
CTTGTA
13
AAACAT
14
CAAAAG
15
GAAACC
16
AAAGCA
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
87
6

Reference
Alternative Capture Equipment Combinations
Alternative Capture Equipment Combinations
Table 43 lists combinations of thermal cycler, lid temperature, plates or strips,
and sealing method (strip caps or sealing tape) other than those used in this
protocol that have shown minimal evaporation.
Refer to this list for additional of equipment combination options for
hybridization. Note that minimal evaporation is needed to ensure good
capture results.
Table 43
Tested options that show minimal evaporation
PCR Machine
Plate/Strips
Cover
Comments
Agilent Mx3005P
QPCR
Mx3000P Strip Tubes
(401428)
MX3000P Optical Strip
Caps (401425)
Heated Lid
Agilent Mx3005P
QPCR
MicroAmp Optical
96-well reaction plate
(N801-0560)
MicroAmp Clear
Adhesive Film
(4306311)
Heated Lid;
ABI compression pad
(4312639)
Use two layers of film.
ABI GeneAmp 9700 MicroAmp Optical
96-well Reaction Plate
(N801-0560)
MicroAmp Caps
(8caps/strip)
(N801-0535)
Heated Lid
ABI Veriti
(4375786)
MicroAmp Clear
Adhesive Film
(4306311)
Heated Lid;
ABI compression pad
(4312639)
MicroAmp Optical
96-well Reaction Plate
(N801-0560)
Use two layers of film.
88
Eppendorf
Mastercycler
Eppendorf 8-Tube PCR
Tubes
Attached lids
Lid heating set to
75°C
BioRad (MJ
Research) PTC-200
Agilent strip tubes
410022 (Mx4000)
Agilent Optical cap
410024 (Mx4000)
Heated Lid
BioRad (MJ
Research) PTC-200
Agilent strip tubes
410022 (Mx4000)
Agilent Optical cap
401425 (Mx3000/3005)
Heated Lid
BioRad (MJ
Research) PTC-200
Agilent 96-well Plate
410088 (Mx3000/3005)
Agilent Optical cap
401425 (Mx3000/3005)
Heated Lid
SureSelectXT Target Enrichment System Kit for Illumina Multiplexed Sequencing
www.agilent.com
In This Book
This guide contains
information to run the
SureSelectXT Target
Enrichment System for
Illumina Paired-End
Sequencing protocol with
the Illumina HiSeq and
MiSeq Multiplexed
Sequencing Platforms.
Agilent Technologies, Inc. 2014
Version 1.7, July 2014
*G7530-90000*
G7530-90000
Revision A9
Agilent Technologies

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement