Clarus 500 GC Installation Guide

Add to my manuals
112 Pages

advertisement

Clarus 500 GC Installation Guide | Manualzz
Clarus 500 GC Installation Guide
Release History
Part Number
Release
09936590
B
Publication Date
August 2002
Any comments about the documentation for this product should be addressed to:
User Assistance
PerkinElmer, Inc.
710 Bridgeport Avenue
Shelton
Connecticut 06484-4794
U.S.A.
Or emailed to: [email protected]
Notices
The information contained in this document is subject to change without notice.
Except as specifically set forth in its terms and conditions of sale, PerkinElmer makes no
warranty of any kind with regard to this document, including, but not limited to, the implied
warranties of merchantability and fitness for a particular purpose.
PerkinElmer shall not be liable for errors contained herein for incidental consequential damages in
connection with furnishing, performance or use of this material.
Copyright Information
This document contains proprietary information that is protected by copyright.
All rights are reserved. No part of this publication may be reproduced in any form whatsoever or
translated into any language without the prior, written permission of PerkinElmer, Inc.
Copyright © 2002 PerkinElmer PerkinElmer, Inc.
Produced in the USA.
Trademarks
Registered names, trademarks, etc. used in this document, even when not specifically marked as such,
are protected by law.
PerkinElmer is a registered trademark of PerkinElmer, Inc.
Clarus 500 GC is a trademark of PerkinElmer, Inc.
e-ssentials is a trademark of PerkinElmer, Inc.
Table of Contents
Introduction...................................................................................................5
About This Manual .........................................................................................7
Conventions Used in this Manual ...................................................................8
Electromagnetic Compatibility (EMC)...........................................................9
Warning Signs on the Instrument..................................................................10
Safety Practices ...........................................................................................13
Chapter Overview .........................................................................................15
Precautions....................................................................................................17
Environmental Conditions ............................................................................18
Electrical Safety ............................................................................................20
Moving the Clarus 500 GC ...........................................................................23
ECD Radioactive Hazards ............................................................................24
Safe Handling of Gases.................................................................................27
Hazardous Chemicals....................................................................................30
Definitions in Warning for Hazardous Chemicals ........................................31
Preparing Your Laboratory.......................................................................33
Chapter Overview .........................................................................................35
Preparing the Laboratory ..............................................................................36
Clarus 500 GC Requirements .......................................................................39
Pre-Installation Checklist..............................................................................44
Installing the Clarus 500 GC .....................................................................45
Install the Autosampler .................................................................................47
Remove the Encoder Protector Block...........................................................48
Install the Autosampler Tower......................................................................49
Install the Autosampler Tower Cover...........................................................53
Install the Syringe Cover ..............................................................................55
Connecting the Gases and Electrical Supply ............................................59
Connect the Gases.........................................................................................61
Clarus 500 GC Gas Connections ..................................................................65
Common Injector/Detector Gas Connections ...............................................69
Connecting the Gas for Subambient Operation ............................................80
Connect the Electrical Supply.......................................................................83
Connecting the Accessories ........................................................................89
Connect the ElCD .........................................................................................91
Connect the ElCD Control Unit to the ElCD Components...........................93
Connect the ElCD Solvent System ...............................................................95
Make the ElCD/PID Series Connection........................................................97
Check the Calibration .................................................................................100
Install the NPD Bead Assembly..................................................................101
PPC Restrictor Information.....................................................................107
Index...........................................................................................................111
Introduction
1
Clarus 500 GC Installation Guide
About This Manual
This quick installation guide is divided into following chapters:
Chapter 1 Introduction
This chapter contains a brief introduction on the instrument, the conventions and warnings
used in the manual.
Chapter 2 Safety Practices
Important safety information for the Clarus 500 GC is provided in this chapter.
Chapter 3 Preparing Your Laboratory
The preparation of your laboratory needed for the Clarus 500 GC laboratory requirements for
the instrument are reviewed in this chapter.
Chapter 4 Installing the Clarus 500 GC
Information on connecting the Clarus 500 GC is provided.
Chapter 5 Connecting the Gases and Electrical Supply
Information on connecting the gases and electrical supply to the system is provided.
Chapter 6 Connecting the Accessories
Information on installing accessories such as the EICD, NPD Bead Assembly, recorder,
integrator and the printer is provided.
Chapter 7 PPC Restrictor Information
PPC restrictor information is provided in this chapter.
7
Introduction
Conventions Used in this Manual
Normal text is used to provide information and instructions.
Bold text refers to button or tab page that is displayed on the touch screen.
All eight digit numbers are PerkinElmer part numbers unless stated otherwise.
Notes, cautions and warnings
Three terms, in the following standard formats, are also used to highlight special
circumstances and warnings.
NOTE: A note indicates additional, significant information that is provided with some procedures.
We use the term CAUTION to inform you about situations that
could result in serious damage to the instrument or other
equipment. Details about these circumstances are in a box like
this one.
CAUTION
We use the term WARNING to inform you about situations that
could result in personal injury to yourself or other persons.
Details about these circumstances are in a box like this one.
WARNING
Customer Service
PerkinElmer Instruments
710 Bridgeport Avenue
Shelton, CT 06484-4794 U.S.A
Tel: 1 (800) 762-4000 or (203) 762-4000
Internet: http://www.perkinelmer.com
8
Clarus 500 GC Installation Guide
Electromagnetic Compatibility (EMC)
United States (FCC)
This product is classified as a digital device used exclusively as industrial, commercial, or
medical test equipment. It is exempt from the technical standards specified in Part 15 of the
FCC Rules and Regulations, based on Section 15.103 (c).
European Union
All information concerning EMC standards will be in the Declaration of Conformity and
these standards will change as the European Union adds new requirements.
European Union Industrial Environment
The 230 V/50 Hz. Clarus GC has been manufactured for use in the European Union and is
intended for the industrial environment. The instrument is to be connected to a mains power
network supplied from a high or medium-voltage transformer dedicated for the supply of an
installation feeding a manufacturing or similar plant.
Industrial environments are characterized by the existence of one or more of the following
conditions:
-industrial, scientific and medical (ISM) apparatus are present
-heavy inductive or capacitive loads are frequently switched
-currents and associated magnetic fields are high
These are the major contributors to the industrial electromagnetic environment and as such
distinguish the industrial from other environments. The instrument is not intended for
connection to a public mains network supplying residential, commercial and light-industrial
locations.
Susceptibility to RF Interference
With the exception of the Flame Ionization Detector (FID), a RF field strength of 10 V/m
between 80 MHz. and 1000 MHz. with 80% modulation at 1 kHz. may cause a deflection on
the chromatographic detector baseline that exceeds its normal pattern. This implies that if a
transmitting device, such as a walkie-talkie carried by a security guard, is use near the
detector, a spike or peak on the chromatographic baseline may occur. If you are concerned
that such an event may occur, PerkinElmer recommends that walkie-talkie restriction notices
be posted in the vicinity. Cell phones, beepers, and other similar devices operate in a much
higher frequency range and do not cause interference.
9
Introduction
Warning Signs on the Instrument
Alternating current.
Protective conductor terminal.
Off position of the main power source.
On position of the main power source.
Warning, hot surface
10
Clarus 500 GC Installation Guide
Warning, risk of electric shock.
Warning (refer to accompanying documents).
11
Introduction
Label locations on instrument
12
Safety Practices
2
Clarus 500 GC Installation Guide
Chapter Overview
This chapter describes the general safety practices and precautions that must be observed
when operating the Clarus 500 GC.
This advice is intended to supplement, not supersede, the normal safety codes in the user's
country. It is also a supplement to the PerkinElmer standard Safety and Health Policy. The
information provided does not cover every safety procedure that should be practiced.
Ultimately, maintenance of a safe laboratory environment is the responsibility of the analyst
and the analyst's organization.
Please consult all manuals supplied with the Clarus 500 GC and accessories before you start
working with the instrument. Carefully read the safety information in this chapter and in the
other manuals supplied. When setting up the instrument or performing analyses or
maintenance procedures, strictly follow the instructions provided. The Clarus 500 GC should
be used in accordance with the instructions provided in this manual. If used otherwise, the
protection provided by the instrument may be impaired.
Generic Warnings
Before installing or operating the Clarus 500 GC, read the following information concerning
hazards and potential hazards. You should ensure that anyone involved with installation
and/or operation of the Clarus 500 GC is knowledgeable in both general safety practices for
the laboratory and safety practices for the Clarus 500 GC. Get advice from your safety
engineer, industrial hygienist, environmental engineer, or safety manager before you install or
use this instrument.
Heated Zones
Heated zones should be treated with caution, for example, injector caps and detectors. Avoid
physical contact with the injector caps. The detector cover may get hot, especially if flame
ionization detectors are operated at high temperatures. As a general rule, allow heated zones
to cool before attempting to work in the oven, injector, or detector areas.
15
Safety Practices
CAUTION
THERMAL RUNAWAY PROTECTION: The Clarus 500 GC
software shuts down the instrument if any heated zone exceeds 470
°C. Should this occur, the following error message is displayed:
INSTRUMENT SHUTDOWN
xxx THERM RUNAWAY where xxx is the heated zone
Call your PerkinElmer Representative.
Instrument shutdown also occurs if there is a PRT (Platinum
Resistance Thermometer) or MPU (Micro Processor Unit) failure.
In these cases the following error message is displayed:
INSTRUMENT SHUTDOWN
xxx PRT ERROR where xxx is the failed zone.
Call your PerkinElmer Representative.
16
Clarus 500 GC Installation Guide
Precautions
Be sure that all instrument operators read and understand the
precautions listed below. It is advisable to post a copy of the
precautions near or on the instrument shelf.
WARNING
The following precautions must be observed when using the Clarus 500 GC:
•
Be sure that the power line voltage of the Clarus 500 GC corresponds to the voltage
used in your laboratory.
•
Never remove the side panels of the Clarus 500 GC without shutting down the
instrument and disconnecting the instrument power cord from line power.
•
Do not immerse the purge gas exit line in a liquid, as the liquid may be drawn back into
the sample holder.
•
Only high quality purge gases should be used with the Clarus 500 GC. Minimum purity
of 99.995% is recommended. A high quality filter-dryer accessory is recommended for
the removal of any moisture from the purge gases.
17
Safety Practices
Environmental Conditions
Operating Conditions
CAUTION
The Clarus 500 GC is designed for indoor use only.
Do not operate in a Cold Room or a refrigerated area. The Clarus 500
GC operates most efficiently under the following conditions:
CAUTION
•
Ambient temperature is 10 °C to 35 °C (50 °F to
95 °F). The GC will operate safely between 5 °C and 40 °C (41
°F and 104 °F).
•
Ambient relative humidity is 20% to 80% non-condensing.
•
Operating altitude is in the range of 0 to 2 000 m.
The Clarus 500 GC is not designed for operation in an explosive
environment.
WARNING
Installation Category
The Clarus 500 GC is able to withstand transient overvoltage according to Installation
Category II as defined in IEC 1010-1.
Pollution Degree
The Clarus 500 GC will operate safely in environments that contain nonconductive foreign
matter up to Pollution Degree 2 in IEC 1010-1.
18
Clarus 500 GC Installation Guide
Clarus 500 GC Touch Screen
For optimum performance, the Clarus 500 GC’s touch screen may require periodic recalibration. The interval between re-calibration may be affected by exposure to combined
heat and humidity conditions (ambient conditions between 30 °C / 50% RH and 35 °C / 80%
RH).
Storage Conditions
The Clarus 500 GC may be stored under the following conditions:
•
ambient temperature is -20 °C to +60 °C (-4 to 140 °F)
•
ambient relative humidity is 20 to 80%, non-condensing
•
altitude is in the range 0 to 12 000 m.
General Laboratory Safety
Your laboratory should have all equipment ordinarily required for the safety of individuals
working with chemicals (fire extinguishers, first-aid equipment, safety shower and eye-wash
fountain, spill cleanup equipment, etc.).
19
Safety Practices
Electrical Safety
The Clarus 500 GC contains high voltage. To prevent the risk of shock, unplug the line cord
from the AC outlet and wait at least one minute before opening or removing any instrument
panels.
The instrument has been designed to protect the operator from potential electrical hazards.
This section describes some recommended electrical safety practices.
CAUTION
This unit contains protective circuitry. Contact PerkinElmer Service
before performing any AC line tests.
WARNING
Connect the GC to an AC line power outlet that has a protective
ground connection. To ensure satisfactory and safe operation of the
GC, it is essential that the protective ground conductor (the
green/yellow lead) of the line power cord is connected to a true
electrical ground. Any interruption of the protective ground
conductor, inside or outside the GC, or disconnection of the
protective ground terminal may impair the protection provided by
the GC.
Do not operate the GC with any covers or parts removed.
WARNING
20
Clarus 500 GC Installation Guide
WARNING
To avoid electrical shock, disconnect the power cord from the AC
outlet before servicing. Servicing on the GC is to be performed
only by a PerkinElmer service representative or similarly trained
and authorized person.
Do not attempt to make adjustments, replacements or repairs to this
GC except as described in the user documentation.
WARNING
WARNING
CAUTION
For protection against fire hazard, only replace fuses with the same
type and rating. Servicing on the GC is to be performed only by
a PerkinElmer service representative or similarly trained and
authorized person.
To ensure adequate cooling of the instrument electronics, do not obstruct the
gap at the base of the GC, and leave at least a 6-inch clearance between
instruments.
Ensure that the power cord is correctly wired and that the ground leads of all electrical units
(for example, recorders, integrators) are connected together via the circuit ground to earth.
Use only three-prong outlets with common earth ground connections.
Servicing of incoming AC line components in your laboratory should be performed only by a
licensed electrician.
21
Safety Practices
WARNING
Lethal voltages are present at certain areas within the instrument.
Installation and internal maintenance of the instrument should
only be performed by a PerkinElmer service engineer or similarly
authorized and trained person. When the instrument is connected to
line power, opening the instrument covers is likely to expose live
parts. Even when the power switch is off, high voltages can still be
present. Capacitors inside the instrument may still be charged even if
the instrument has been disconnected from all voltage sources.
The instrument must be correctly connected to a suitable electrical supply. The supply must
have a correctly installed protective conductor (earth ground) and must be installed or
checked by a qualified electrician before connecting the instrument.
WARNING
Any interruption of the protective conductor (earth ground) inside or
outside the instrument or disconnection of the protective conductor terminal
is likely to make the instrument dangerous. Intentional interruption is
prohibited.
When working with the instrument:
•
Disconnect the instrument from all voltage sources before opening it for any adjustment,
replacement, maintenance, or repair. If afterwards, the opened instrument must be
operated for further adjustment, maintenance, or repair, this must only be done by a
PerkinElmer Service engineer.
•
Whenever it is possible that the instrument is no longer electrically safe for use, make
the instrument inoperative and secure it against any unauthorized or unintentional
operation. The electrical safety of the instrument is likely to be impaired if, for example,
the instrument shows visible damage, has been subjected to prolonged storage under
unfavorable conditions, or has been subjected to severe stress during transportation.
22
Clarus 500 GC Installation Guide
Moving the Clarus 500 GC
The Clarus 500 GC weighs 53.5 kg (118 lb). Improper lifting can cause injury to the back. If
the instrument must be moved, we recommend that at least two people carefully lift the
instrument in order to move it.
23
Safety Practices
ECD Radioactive Hazards
To assure that removable radioactive contamination on the external
parts of the ECD remains at a safe level, the United States Nuclear
Regulatory Commission requires that:
WARNING
•
The ECD be wipe tested at least once every six months.
•
A record of the results must be maintained for NRC inspection.
United States Government Regulations for ECDs
NOTE: To repair an Electron Capture Detector cell requires a specific license issued by the U.S.
Nuclear Regulatory Commission (NRC) and/or in some states by the equivalent state agency.
For further information on obtaining a license, contact the Customer Service Department at
PerkinElmer, Shelton, Connecticut, or the NRC Material Branch, Office of Nuclear
Materials, Safety and Safeguards, Washington, DC 20555.
All USNRC regulations can be obtained through the internet at www.nrc.gov/reading-rm/
NOTE: These instructions are for ECD cell purchasers who are not specifically licensed to handle
radioactive materials.
The Clarus 500 GC Electron Capture Detector model (P/N N610-0063) contains a maximum
of 15 mCi of Nickel 63 (Ni 63), a radioactive material. Your possession and use of this
detector is governed by 10 C.F.R. Section 31.5 which is reproduced in Appendix I. Under the
provisions of that regulation you are deemed a "General Licensee."
Your possession and use of the detector cell may also be regulated by the state where you are
located. The requirements of state regulatory agencies are substantially similar to those
contained in NRC regulation 10 C.F.R. Section 31.5, but they may differ in some respects. It
is suggested that you procure a copy of the regulations of your particular state. (Supplement 2
in Appendix I contains a list of the "Agreement States" which have been granted authority by
the U.S. Nuclear Regulatory Commission to regulate the possession and use of radioactive
material.)
It is required that you be familiar with regulation 10 C.F.R. Section 31.5 (Appendix 1 in the
Hardware Guide 0993-6590). Following are summaries of its requirements.
Labels
Do not remove any of the labels attached to the ECD cell or any of the labels attached to your
Clarus 500 Gas Chromatograph that refer to the ECD cell. Follow all instructions and abide
by all precautions provided by the labels and in user instruction manuals referred to by the
labels.
24
Clarus 500 GC Installation Guide
Leak Testing
You are obligated under U.S. federal and state regulations to make certain that the ECD cell
is wipe-tested for leakage of radioactive materials at intervals of no longer than six months,
and that the analysis of these wipe tests is conducted by a person specifically licensed to do
so, either by the U.S. Nuclear Regulatory Commission or by an Agreement State. The
analyses can be performed by the firm listed below:
National Leak Test Center
P.O. Box 486
North Tonawanda, New York 14120
ECD Cell Failure or Damage
If a leak test detects more than 0.005 µCi (microcurie) of removable radioactive material on
the surface of an ECD cell, or if the cell itself is damaged in such a way as to indicate that it
may no longer adequately shield the radioactive material inside, you must immediately
suspend operation of your chromatograph until the cell has been repaired or disposed of by a
person specifically licensed to do so. Any such incident must be reported by you to the
Regional Office, Inspection and Enforcement, U.S. Nuclear Regulatory Commission.
Reporting Radiation Incidents, Theft or Loss
Please read Regulation 10 C.F.R. Section 20.2201 and 20.2202. These describe your duties
should the radioactive material (Ni 63) in the ECD cell be lost, stolen, or released, or should
any person be exposed to radiation.
Other ECD Requirements
Regulation 10 C.F.R. Section 31.5 (see Appendix I) does not permit you to abandon the ECD
cell or export it. It may not be transferred except to a person specifically licensed to receive
it. Within thirty days of such a transfer, you must report to the Director of Nuclear Material
Safety and Safeguards, U.S. Nuclear Regulatory Commission, Washington, D.C. 20555, the
name and address of the transferee. However, no report is needed to transfer a defective ECD
cell to PerkinElmer in order to obtain a replacement.
You may transfer the ECD cell to another general licensee, like yourself, only when it
remains at the same location to which it was shipped by PerkinElmer. Give the transferee a
copy of these instructions and the regulations in Appendix I, and report to the commission as
required in Regulation C.F.R. Section 31.5.
25
Safety Practices
NEVER DISMANTLE THE ECD CELL!!
You can remove the ECD cell from the GC for repair.
WARNING
United Kingdom Regulations
In the U.K., registration is required under the Radioactive Substances Act of 1960, for anyone
keeping or using radioactive materials. Application should be made to any one of the
following governing bodies:
ENGLAND
Department of the Environment
Queen Anne's Chambers
Tothill Street
London, SW1H 9J4
SCOTLAND Scottish Development Department
21 Hill Street
Edinburgh, EH2 3J4
WALES
Welsh Office
Cathay's Park
Cardiff, CF1 3NG
NORTHERN Ministry of Development
IRELAND
Parliament Building
Storemont
Belfast, Northern Ireland
26
Clarus 500 GC Installation Guide
Safe Handling of Gases
When using hydrogen, either as the combustion gas for a flame ionization detector or as a
carrier gas, special care must be taken to avoid buildup of explosive hydrogen/air mixtures.
Ensure that all hydrogen line couplings are leak-free and do not allow hydrogen to vent
within the oven.
Ventilation
Adequate ventilation must be provided, particularly if a liquid nitrogen or carbon dioxide
subambient accessory is in constant use. When analyzing hazardous compounds, such as
pesticides, it may be necessary to arrange for venting of detector effluent into a fume hood.
Using Hydrogen
WARNING
Flame Ionization Detectors (FID) and Flame Photometric
Detectors (FPD) use hydrogen as fuel. If the hydrogen is turned
on without a column attached to the injector and detector fittings
inside the oven, hydrogen could diffuse into the oven creating the
possibility of an explosion. To avoid possible injury, DO NOT
TURN ON THE HYDROGEN UNLESS A COLUMN IS
ATTACHED AND ALL JOINTS HAVE BEEN LEAK TESTED.
Before disconnecting a column, make certain that the hydrogen has
been turned OFF.
If two FIDs or FPDs are installed and only one has a column
attached to it, make certain that you cap off the unused detector
inlet fitting with a 1/8-inch stainless steel plug (P/N N930-0061).
WARNING
Contact the gas supplier for a material safety data sheet (MSDS)
containing detailed information on the potential hazards associated
with the gas. Carefully use, store, and handle compressed gases in
cylinders. Gas cylinders can be hazardous if they are mishandled.
27
Safety Practices
NOTE: The permanent installation of gas supplies is the responsibility of the user and should
conform to local safety and building codes.
If liquid nitrogen is used, the gas cylinder must be fitted with an over-pressure regulator
which will vent the cylinder as necessary to prevent it from becoming a safety hazard.
Consult the following references for more detailed information and additional guidelines
about gas cyclinders.
•
Compressed Gas Association (USA), "Safe Handling of Compressed Gases in
Containers," pamphlet no. P-1, 1984.
•
Compressed Gas Association (USA), "The Inert Gases – Argon, Nitrogen and Helium,"
pamphlet no. P-9, 1992.
Identification of Gas Cylinders
•
Legibly mark cylinders to identify their contents. Use the chemical name or
commercially accepted name for the gas.
Storing Gas Cylinders
Review the following precautions with the customer to ensure the safe use and storage of gas
cylinders.
•
28
Cylinders should be stored in accordance with the regulations and standards applicable
to the customer’s locality, state, and country.
Clarus 500 GC Installation Guide
•
When cylinders are stored indoors in storage rooms, the storage room should be well
ventilated and dry. Ensure that the ventilation is adequate to prevent the formation of
dangerous accumulations of gas. This is particularly important in small or confined
areas.
•
Do not store cylinders near elevators, gangways, or in locations where heavy moving
objects may strike or fall against them.
•
Use and store cylinders away from exits and exit routes.
•
Locate cylinders away from heat sources, including heat lamps. Compressed gas
cylinders should not be subjected to temperatures above 52 °C (126 °F).
•
It is recommended that gas cylinders be stored and placed outside the laboratory and
connected to the instrument through copper lines.
Handling of Gas Cylinders
•
•
•
•
•
•
•
•
•
•
Do not allow ignition sources in the storage area and keep cylinders away from readily
ignitable substances such as gasoline or waste, or combustibles in bulk, including oil.
Store cylinders standing upright, fastened securely to an immovable bulkhead or
permanent wall.
When storing cylinders outdoors, they should be stored above ground on a suitable floor
and protected against temperature extremes (including the direct rays of the sun).
Arrange gas hoses where they will not be damaged or stepped on and where things will
not be dropped on them.
Take care not to kink or stress the gas lines. For safety, cylinders should be firmly
clamped in position.
If it becomes necessary to move the cylinders, do so with a suitable hand truck after
insuring that the container cap is secured and the cylinder is properly fastened to the
hand truck.
Use only regulators, tubing and hose connectors approved by an appropriate regulatory
agency.
Do not refill cylinders.
Check the condition of pipes, hoses and connectors regularly. Perform gas leak tests at
all joints and seals of the gas system regularly, using an approved gas leak detection
system.
When the equipment is turned off, close all gas cylinder valves tightly at the cylinder.
Bleed the remainder of the line before turning the exhaust vent off.
29
Safety Practices
Hazardous Chemicals
Before using samples, you should be thoroughly familiar with all hazards and safe handling
practices. Observe the manufacturer’s recommendations for use, storage and disposal. These
recommendations are normally provided in the Material Safety Data Sheets (MSDS) supplied
with the solvents.
Be aware that the chemicals that you use in conjunction with the GC may be hazardous. DO
NOT store, handle, or work with any chemicals or hazardous materials unless you have
received appropriate safety training and have read and understood all related Material Safety
Data Sheets (MSDS). MSDSs provide information on physical characteristics, precautions,
first aid, spill clean up and disposal procedures. Familiarize yourself with the information and
precautions contained in these documents before attempting to store, use or dispose of the
reagents. Comply with all federal, state, and local laws related to chemical storage, handling,
and disposal.
You must work under a suitable hood when handling and mixing certain chemicals. The room
in which you work must have proper ventilation and a waste collection system. Always wear
appropriate safety attire (full-length laboratory coat, protective glasses, gloves, etc.), as
indicated on Material Safety Data Sheets.
WARNING
30
Some chemicals used with this GC may be hazardous or may
become hazardous after completion of an analysis. The responsible
body (for example, the Lab Manager) must take the necessary
precautions to ensure that the GC operators and the surrounding
workplace are not exposed to hazardous levels of toxic substances
(chemical or biological) as defined in the applicable Material Safety
Data Sheets (MSDS) or OSHA, ACGIH, or COSHH documents.
Venting for fumes and disposal of waste must be in accordance with
all national, state and local health and safety regulations and laws.
Clarus 500 GC Installation Guide
Definitions in Warning for Hazardous Chemicals
Responsible body. “Individual or group responsible for the use and maintenance of
equipment, and for ensuring that operators are adequately trained.” [per IEC 1010-1].
Operator. “Person operating equipment for its intended purpose.”
[per IEC 1010-1].
OSHA:
Occupational Safety and Health Administration (United States)
ACGIH:
American Conference of Governmental Industrial Hygienists
COSHH:
Control of Substances Hazardous to Health (United Kingdom)
31
Safety Practices
32
Preparing Your
Laboratory
3
Clarus 500 GC Installation Guide
Chapter Overview
The items shown in the following checklist need to be considered when preparing the
laboratory for the instrument.
•
Environmental Conditions
•
Exhaust Vent Requirements (if required)
•
Laboratory Space Requirements
•
Cooling Water Requirements (if required)
•
Electrical Requirements
•
Pneumatic Requirements (if required)
35
Preparing Your Laboratory
Preparing the Laboratory
The following sections describe the laboratory requirements in detail for the Clarus 500 GC.
Environmental Conditions
The laboratory in which the Clarus 500 GC is located must meet the following conditions:
•
A corrosive-free environment.
•
The instrument will operate with a laboratory temperature between 10 and 35 °C (50- 95
°F). For optimum instrument preformance, the room temperature should be controlled at
20° ± 2 °C.
•
The environment should be relatively dust-free to avoid sample and instrument
contamination problems.
•
Free of excessive vibration.
The Clarus 500 GC has been designed for indoor use. Do not use the instrument in an area
where explosion hazards may exist.
Exhaust Vent Requirements
Exhaust venting is important for the following reasons.
•
It protects laboratory personnel from toxic vapors that may be produced by some
samples.
•
It helps to protect the instrument from corrosive vapors that may originate from the
sample(s).
•
It removes dissipated heat produced by the instrument and power supply.
36
Clarus 500 GC Installation Guide
The use of Clarus 500 GC without adequate ventilation to
outside air may constitute a health hazard.
WARNING
NOTE: Local electrical codes do not allow PerkinElmer Service Engineers to install the blower and
vent assembly.
The blower capacity depends on the duct length and number of elbows or bends used to
install the system. If an excessively long duct system or a system with many bends is used, a
stronger blower may be necessary to provide sufficient exhaust volume at the instrument.
Alternatively, smooth stainless-steel tubing should be used instead of flexible stainless steel
tubing where flexibility is not required to reduce system friction loss or "drag." If smooth
stainless steel is used, there must be a way to move the vent hood out of the way for
servicing. A length of smooth stainless steel ducting has 20-30% less friction loss than a
comparable length of flexible ducting. When smooth stainless steel tubing is used, elbows
must be used to turn corners. These elbows should turn at a centerline radius of 150mm with
a maximum bend angle of 45 degrees to reduce friction losses, and the number of elbows
should be minimized.
Additional recommendations on the venting system include the following items.
•
Make sure the duct casting is installed using fireproof construction. Route ducts away
from sprinkler heads.
•
The duct casing and venting system should be made of materials suitable for
temperatures greater than 70 °C (158 °F). It should be installed to meet local building
code requirements.
•
Locate the blower as close to the discharge outlet as possible. All joints on the discharge
side should be airtight, especially if toxic vapors are being carried.
•
Equip the outlet end of the system with a back draft damper and take the necessary
precautions to keep the exhaust outlet away from open windows or inlet vents. In
addition, extend it above the roof of the building for proper dispersal of the exhaust.
•
Equip the exhaust end of the system with an exhaust stack to improve the overall
efficiency of the system.
37
Preparing Your Laboratory
•
Make sure the length of the duct that enters into the blower is a straight length at least
ten times the duct diameter. An elbow entrance into the blower inlet causes a loss in
efficiency.
•
Provide make-up air in the same quantity as is exhausted by the system. An "airtight"
lab will cause an efficiency loss in the exhaust system.
•
Ensure that the system is drawing properly by using an air flow meter.
•
Equip the blower with a pilot light located near the instrument to indicate to the operator
when the blower is on.
38
Clarus 500 GC Installation Guide
Clarus 500 GC Requirements
Laboratory Space Requirements:
Size
GC:
66 cm (26 in.) wide x 40 cm (19 in.) high x
72 cm (28.5 in.) deep
Autosampler:
13 cm (5 in.) wide x 36 cm (14 in.) high x
24 cm (9.5 in.) deep
Weight
GC:
49 kg (108 lb)
Autosampler:
4.5 kg (10 lb)
Physical Configuration:
Single unit for use on standard laboratory bench
which can be interfaced to a computer.
Bench Space:
The laboratory bench should be sturdy enough to
support the full weight of the GC as well as
additional equipment (i.e., computer). Expect the
total weight of the GC and accessory equipment to
be at least 91 kg (200 lb).
Allow a minimum clearance of 10.2 cm (4 in.) on
each side, 15.2 cm (6 in.) at the rear of the GC, and
137.2 cm (54 in.) at the top of the GC. If this is not
possible, install the GC on a bench that has wheels.
Peripherals, Printers, etc.
Allow at least 61 cm (24 in.) on either side of the
GC to accommodate additional equipment.
39
Preparing Your Laboratory
Environmental Requirements:
Pollution Degree:
This instrument will operate safely in environments
that contain nonconductive foreign matter up to
Pollution Degree 2 as defined in IEC 1010-1.
Laboratory Environment:
Install the GC in an indoor laboratory environment
that is clean and is free of drafts, direct sunlight and
vibration.
The laboratory should be free of flammable,
explosive, toxic, caustic, or corrosive vapors or
gases and should be relatively free of dust.
The ambient laboratory temperature should be
between 10 °C and 35 °C (50 °F and 95 °F) with a
relative humidity between 20% and 80% with no
condensation. The GC will operate safely between 5
°C and 40 °C (41 °F and 104 °F).
Electrical Power Requirements:
40
Installation Category Statement:
This instrument is able to withstand transient
over-voltage according to Installation
Category II as defined in IEC 1010-1.
Power Consumption:
2 400 VA (volt-amps) for the GC. Add
100 VA for the computer and 108 VA for the
printer.
Clarus 500 GC Installation Guide
Power Specification:
All electrical supplies must be smooth, clean, and free of
line transients greater than 40 V peak to peak and must
meet and remain within the following tolerances:
120 VAC ±10% @ 50/60 Hz ±1%
230 VAC ±10% @ 50/60 Hz ±1%
Instruments and peripherals should not be connected to
circuits with large inductive or large and frequent loads
(i.e., large motors, discharge lamps, photocopy systems,
radio transmitters, etc.).
Power Outlets:
A minimum of one dedicated 120 VAC outlet at 20 A or
one 230 VAC outlet at 10 A (minimum) is required for
the GC. Additional equipment, such as computers and
printers, should be connected per their specifications.
Safety Requirements:
Gas Cylinders and
Gas Delivery Lines:
All gas cylinders should be firmly clamped to a
suitable surface. Care must be taken not to
kink or overstress the gas delivery lines.
Hydrogen:
Ensure that all hydrogen lines and connections
are leak-free. When using a hydrogen tank,
install an in-line hydrogen snubber (P/N 00090038) between the tank regulator and the
delivery tubing.
Ventilation:
Always provide adequate ventilation. When
analyzing hazardous compounds, such as
pesticides, it may be necessary to arrange for
venting the detector effluent into a fume hood.
41
Preparing Your Laboratory
Gas Requirements:
All gases must have a minimum purity of 99.995%. Gases used with the mass spectrometer
and ELCD detectors require a minimum purity of 99.999%. Gas cylinders should be located
outside of the laboratory whenever possible and should always be stored and operated in the
vertical position. Always use copper tubing that is free of grease, oil, and organic material
for all gases delivered to the Clarus 500 GC.
Helium, Nitrogen,
8.5% H2/91.5% Helium
95% Argon/5% Methane:
A number 1A (200 ft3) gas cylinder should be used for
all carrier gases. Filter all gases (except methane)
through a moisture filter and/or hydrocarbon trap and
de-oxo filter. Argon/methane should be filtered
through a moisture filter and a de-oxo filter. Gas
delivery pressure to the GC should be 60 – 90 psig
(414-621 kPa).
Air:
A number 1A (200 ft3) gas cylinder of compressed air
or an air compressor can be used. All air should be
filtered through a moisture filter. Do NOT use
“Breathing Air." When using manual pneumatics, gas
delivery pressure to the GC must not exceed 30 psi
(207 kPa). If this is not possible, secondary regulation
will be required. With PPC pneumatics, a delivery
pressure range of 70 – 90 psig (483 - 621 kPa) is
acceptable.
Hydrogen:
A number 2 (62 ft3) gas cylinder or a hydrogen
generator can be used. All hydrogen should be filtered
through a moisture filter. When using manual
pneumatics, gas delivery pressure to the GC must not
exceed 30 psi (207 kPa). If this is not possible,
secondary regulation will be required. With PPC
pneumatics, a delivery pressure range of 60 – 90 (414621 kPa) psig is acceptable.
42
Clarus 500 GC Installation Guide
Sample Preparation Requirements
Customer Responsibility
43
Preparing Your Laboratory
Pre-Installation Checklist
Model: ____________________________________________ Date:
________________
Customer:
SPO#
________________________________________
____________________________________________
Installation Requirements
Lab Space Requirements
Instrument
Lab Space Requirements
Peripherals
Environmental Requirements
Power Requirements
Safety Requirements
Gas Requirements
Sample Preparation
(Customer Responsibility)
Computer Configuration
Miscellaneous
44
OK
Needs Prior to Installation
Installing the
Clarus 500 GC
4
Clarus 500 GC Installation Guide
Install the Autosampler
This section describes how to install the autosampler tower, the autosampler tower cover, and
the autosampler syringe. The autosampler tower and the tower cover are shipped together in
one box. The autosampler tray is installed in the Clarus 500 GC at the factory.
47
Installing the
Clarus 500 GC
Remove the Encoder Protector Block
The Clarus 500 GC autosampler has been shipped with an encoder protector block. This is a
piece of antistatic polyethylene foam that protects the tower disk during shipment. Before you
install the autosampler tower onto your Clarus 500 GC, you must remove the encoder
protector block from the tower disk assembly.
To remove the encoder protector block:
1. Carefully remove the autosampler tower from the shipping box.
2. The piece of foam on top of the tower is the encoder protector block. Carefully slide
the block straight out from the tower disk.
CAUTION
48
Do not twist the block off. This will damage the tower disk.
Clarus 500 GC Installation Guide
Install the Autosampler Tower
Perform the following procedure to install the autosampler tower:
1. The autosampler tower mounting hardware is shipped already secured to the Clarus
500 GC top. Remove the three screws and flat washers securing the mounting
hardware and use them to secure the tower base to the Clarus 500 GC.
2. Position the tower interface harness through the autosampler stop, which is part of the
tower base. The harness should be fitted carefully into the hole in the autosampler
base assembly (see Note below).
NOTE: You may find it easier to feed the wires a few at a time. The two yellow tie wraps on this
harness indicate the portion of the harness that needs to be fed to the tower. Only one yellow
tie wrap should be visible at the autosampler stop.
3. Install the autosampler tower on the top of the Clarus 500 GC as shown in Figure 1.
The locating pin on the Clarus 500 GC fits into the slotted hole in the autosampler
tower base.
49
Installing the
Clarus 500 GC
Figure 1. Installing the autosampler tower.
4. Connect the two harness connectors to the tower interface P.C. board. Plug the large
connector into J2 and the small connector into J1 (see Figure 2).
These connectors are keyed and connect only one way.
50
Clarus 500 GC Installation Guide
Figure 2. Connecting the tower interface harness to the tower interface p.c. board.
5. Manually rotate the autosampler tower so that the front of the tower (this is where the
syringe will be installed) faces the rear of the Clarus 500 GC.
6. Secure the harness to the autosampler tower frame using the cable clamp attached to
the harness. Attach the cable clamp and the harness ground wire to the underside of
the casting as shown in Figure 3, using the screw provided. Make sure the harness is
routed correctly before you tighten the screw, or you may have difficulty installing the
autosampler tower cover.
NOTE: There are two holes in the autosampler tower frame casting in the area to which the cable
clamp should be secured. Secure the harness cable clamp and ground wire to the inner-most
hole.
51
Installing the
Clarus 500 GC
7. The screw is already attached to the casting. Use a long Phillips screwdriver to
remove the screw, then attach the cable clamp.
Figure 3. Securing the harness to the tower.
52
Clarus 500 GC Installation Guide
Install the Autosampler Tower Cover
Install the autosampler tower cover by referring to Figures 4 and 5 as you follow this
procedure:
1. The tower is shipped with two 3/8-in. long screws installed on the bottom of the
tower frame. Loosen but do not remove these screws.
2. Open the cover door and carefully lower the cover onto the tower, aligning the two
guides inside the cover with the sides of the tower (see Figure 5).
3. Pull the sides of the cover away from the tower frame just enough to slide the
cover tabs onto the two screws.
4. Tighten the two screws. Verify that the cover door opens and closes freely and that
it locks when closed. If not, realign the cover until the door locks when closed.
NOTE: If the door is hitting the door sensor, loosen the two screws that secure the door sensor
bracket to the tower, then adjust the sensor up or down until the door no longer hits the
sensor (see Figure 4).
Figure 4. Adjusting the position of the tower door sensor.
53
Installing the
Clarus 500 GC
Figure 5, on the following page, shows how to install the autosampler tower cover onto the
tower.
Figure 5. Installing the autosampler tower cover.
54
Clarus 500 GC Installation Guide
Install the Syringe Cover
To install the syringe, proceed as follows:
1. Before you install the syringe, move the autosampler tower so it faces the
front of the instrument; this is called the “Park" position (see Figure 6).
Figure 6. The autosampler tower in the “Park” position.
CAUTION
Always handle the syringe with extreme care. Be careful not to bend the
needle. Before installing the syringe, pull some solvent up into the barrel to
moisten the mechanism.
NOTE: Use only syringes sold by PerkinElmer for the Clarus 500 GC. Syringe plungers are not
interchangeable from one syringe to another.
55
Installing the
Clarus 500 GC
2. Lift and turn the plunger cap handle (located on top of the plunger assembly)
until the pin locks it in the up position as shown in Figure 7.
Figure 7. Lifting and turning the plunger cap handle.
3. Insert the syringe needle through the bottom of the carriage assembly and use your
finger to carefully guide the needle through the hole in the needle guide (see Figure 8).
CAUTION
56
Be careful not to bend the needle when inserting it into the hole in the needle
guide.
Clarus 500 GC Installation Guide
Figure 8. Installing the syringe.
4. Slide the top of the syringe plunger into the plunger cap slide, then lift and turn the
plunger cap handle so the pin drops into place. This will lock the syringe plunger into
place.
5. Hold the syringe nut with one thumb and, with your other thumb, turn the carriage
thumbscrew counterclockwise to tighten it (see Figure 8). DO NOT OVERTIGHTEN
THE THUMBSCREW.
6. Carefully guide the needle into the needle locator in the center of the vial locator as
you tighten the thumbscrew (see Figure 8).
Your autosampler is now installed.
57
Installing the
Clarus 500 GC
58
Connecting the
Gases and
Electrical Supply
5
Clarus 500 GC Installation Guide
Connect the Gases
The following sections describe how to connect the gas cylinders to the Clarus 500 GC and
how to check the connections for leaks. Use the fittings and copper tubing supplied in the
installation kit (P/N 0332-8000).
Gas Cylinder Connections
Use the following procedure to make the gas cylinder connections:
1. Using a large adjustable wrench (at least 1¼ in.), attach a dual-stage regulator, or two
single-stage regulators in series, to each gas cylinder. The regulator outlet fitting must
be a 1/8-in. Swagelok type.
The following applies to manual pneumatics configurations only:
CAUTION
1. Dual-stage regulation is required with the Clarus 500 GC. It permits
steady gas delivery at high or low flows and provides precise control
of gas pressure throughout the discharge of virtually the entire
contents of a compressed gas cylinder. It is not necessary for PPC
units.
2. If your Clarus 500 GC is equipped with a TCD and you are operating
more than one GC from the same carrier gas tank, and/or operating
the TCD at high sensitivity, install an additional single-stage
regulator (0 – 90 psig) before you install the gas to the TCD.
3. If your Clarus 500 GC is equipped with an ECD, you can control the
flow of make-up gas by installing an additional 30-psig single-stage
regulator before the ECD needle valve.
4. If your Clarus 500 GC is equipped with a NPD, we recommend
installing a toggle valve in the hydrogen line that enters the Clarus
500 GC. This allows you to turn off the hydrogen when the NPD is
not in use without changing the regulator setting on the Clarus 500
GC.
If your Clarus 500 GC is equipped with a FPD, we recommend installing
a toggle valve in the air line that enters the Clarus 500 GC. This allows
you to turn off the air when the FPD is not in use without changing the
regulator setting on the Clarus 500 GC.
61
Connecting the Gases and Electrical Supply
2. Connect a 1/4-in. to 1/8-in. adapter to the regulator outlet fitting on all your gas
cylinders except the hydrogen cylinder.
3. Insert a hydrogen snubber (P/N 0009-0038) in the hydrogen cylinder regulator outlet
fitting as shown in Figure 9. Use a 1/4-in. to 1/8-in. reducer (P/N 0990-3212) between
the snubber and the regulator.
NOTE: PerkinElmer offers a fixed 30-psig stainless steel diaphragm regulator (P/N N610-1473) for
use with manual pneumatics configurations. This provides a second stage of regulation
between the tank and the needle valves on the manual pneumatics version Clarus 500 GC.
Figure 9. Connecting a hydrogen snubber.
4. Cut a length of 1/8-in. copper tubing long enough to connect a line filter to each gas
supply line or gas cylinder.
62
Clarus 500 GC Installation Guide
5. Place a 1/8-in. nut, a 1/8-in. rear ferrule, and a 1/8-in. front ferrule over each end of
the copper tubing as shown in Figure 10.
Figure 10. Components of a 1/8-in. tubing connection.
6. Insert one end of the tubing with the nut and ferrules into the end fitting of a suitable
line filter (see Table 1), and insert the other end into the regulator outlet fitting or the
end fitting of the hydrogen snubber.
7. Insert the tubing into the fitting so that the tubing bottoms firmly, then tighten the nut
½-turn past finger-tight.
8. Cut a piece of 1/8-in. copper tubing long enough to connect the filter to the Clarus 500
GC. To allow for future instrument movement, cut the tubing 2 – 3 ft longer than
necessary.
63
Connecting the Gases and Electrical Supply
Table 1. Recommended Filters
Detectors
Carrier Gas
FID
NPD
FPD
Hydrocarbon Trap (N9301192) + Oxygen Filter (N9301179) if capillary column is
used
TCD**
Hydrocarbon Trap (N9301192) + Oxygen Filter (N9301179) if capillary column is
used
ECD**
Moisture Filter (i.e., Molecular
Sieve) (N930-1193) + Oxygen
Filter (N930-1179)
PID
Hydrocarbon Trap (N9301192) + Oxygen Filter (N9301179) if capillary column is
used
ElCD#
Hydrocarbon Trap (N9301192) + Oxygen Filter (N9301179) if capillary column is
used
Combustion Gas
Air
H2
Moisture Filter (i.e.,
Molecular Sieve)
(N930-1193)
Moisture Filter
(i.e., Molecular
Sieve) (N9301193)
Molecular Sieve
(N930-1193) +
Charcoal (N9301192)
*
There is a combination filter system that contains an Oxygen Filter, Oxygen Indicator, and a
Moisture Trap. Its Part No. is N930-6002.
**
Install a tee fitting on the carrier gas line for the TCD reference line or for the make-up line
for the ECD.
#
99.999%-purity helium and hydrogen must be used with the ElCD.
Stainless-steel diaphragm regulators must be used with the ElCD (hydrogen regulator P/N
0990-7128, helium regulator P/N 0990-7127).
64
Clarus 500 GC Installation Guide
Clarus 500 GC Gas Connections
Perform the following procedure to connect the gases to the Clarus 500 GC:
CAUTION
Do NOT connect any gases to the manual pneumatics bulkhead if there is a
PPC module installed for any channel. In this case, the manual pneumatics
bulkhead is NOT connected on the inside of the Clarus 500 GC. Therefore,
if a gas line is connected to the manual pneumatics bulkhead, the gas would
flow into the Clarus 500 GC itself. This must especially be avoided with the
hydrogen gas.
1. Make a tubing strain relief by forming an approximate 3-in. diameter loop in the
tubing so that it is approximately 3 in. from the end of the tubing closest to the Clarus
500 GC as shown in Figure 11. Make this strain relief on all gas lines.
Figure 11. Tubing strain-relief loop.
2. Insert a 1/8-in. nut, a 1/8-in. rear ferrule, and a 1/8-in. front ferrule over the end of
each piece of copper tubing as shown in Figure 10.
65
Connecting the Gases and Electrical Supply
3. Insert the tubing into the proper fitting on the Clarus 500 GC bulkhead for the gas that
you are connecting. Be sure the tubing bottoms out firmly in the fitting, then tighten
the nut ½-turn past finger-tight.
Figure 12 shows the manual pneumatics bulkhead and PPC bulkheads on the back of the
Clarus 500 GC. M/U = make-up gas, CARR. = carrier gas, and HYD and H2 = hydrogen
gas.
66
Clarus 500 GC Installation Guide
Figure 12. Location of the Clarus 500 GC bulkheads on the rear panel.
67
Connecting the Gases and Electrical Supply
Leak Test the Gas Connections
After connecting the gases to the Clarus 500 GC, check all connections for possible leaks by
following this procedure:
1. Set the regulator on the carrier gas cylinder between 60 and 90 psig.
2. Test the carrier gas connections by using either a gas leak-check solution of 50%
isopropyl alcohol in water or an electronic leak detector. Check for leaks (indicated
by bubbles) at each connection.
CAUTION
Do not use a soap solution to check for gas leaks. It could contaminate the
system.
3. To check a hydrogen or air cylinder for leaks, set the regulator to 30 psig for manual
pneumatics and for PPC systems between 60 and 90 psig for H2 and between 70 and
90 psig for air; test the connections in the same manner that you tested the carrier gas
connections.
68
Clarus 500 GC Installation Guide
Common Injector/Detector Gas Connections
Figures 13 through 21 shows how gases are connected to the Clarus 500 GC in several
common injector/detector configurations. Note that the strain relief, which is required on all
gas lines, is not shown in these figures.
Position incoming gas lines away from the oven exhaust duct so that the hot
exhaust will not blow directly on the gas lines.
WARNING
CAUTION
Thoroughly purge the incoming gas lines before measuring and adjusting the
gases. This is especially important when using hydrogen gas.
NOTE: For Clarus 500 GC’s that have a combination of both PPC and manual pneumatics, set the
detector gas delivery pressure between 60 and 90 psig and install the fixed 30-psig stainless
steel diaphragm regulator (P/N N610-1473), behind the Clarus 500 GC for the manual
pneumatics detector gases.
The manual pneumatics bulkhead always contains Swagelok nuts in all six locations whether
or not there is an injector or a detector connected to this bulkhead. The PPC modules will be
installed only if your Clarus 500 GC has been configured with PPC for an injector and/or
detector.
The Clarus 500 GC can be configured as any of the following:
•
All zones (injector/detector) as PPC.
•
All zones as manual pneumatics.
69
Connecting the Gases and Electrical Supply
•
A combination of PPC and manual pneumatics.
For example, the injectors could be controlled by PPC and the detectors could be
controlled by manual pneumatics.
The following nine figures show the connections for the injectors and detectors to either the
PPC or the manual pneumatics bulkheads. The connections for a PPC system are indicated by
solid lines and the connections for manual pneumatics are indicated by dashed lines.
70
Clarus 500 GC Installation Guide
Figure 13. TCD in Channel B (rear channel) with one packed injector.
71
Connecting the Gases and Electrical Supply
Figure 14. ECD in Channel A (front channel) using argon/methane or nitrogen as the
carrier gas, with one injector.
NOTE: If you are using the ECD with a capillary column at low flows (5 mL/min or less), you can
plumb the injector with helium and the make-up line with argon/methane or nitrogen. Helium
is the preferred carrier for capillary columns.
72
Clarus 500 GC Installation Guide
Figure 15. FID, NPD, or FPD in Channel A and ECD in Channel B using nitrogen as
the carrier gas, with two injectors.
73
Connecting the Gases and Electrical Supply
Figure 16. FID, NPD, or FPD in Channel A using helium or nitrogen as the carrier gas,
with one injector.
74
Clarus 500 GC Installation Guide
Figure 17. FID, NPD, or FPD in Channel A using helium carrier gas, ECD in Channel
B using argon/methane or nitrogen carrier gas, with two injectors.
NOTE: If you are using the ECD with a capillary column at low flows (5 mL/min or less), you can
plumb the injector with helium and the make-up line with argon/methane or nitrogen. Helium
is the preferred carrier gas for capillary columns. The FID, NPD, and FPD do not require
make-up gas.
75
Connecting the Gases and Electrical Supply
Figure 18. FID, NPD, or FPD in Channel A, TCD in Channel B, with two packed
injectors.
76
Clarus 500 GC Installation Guide
Figure 19. FID, NPD, or FPD in Channel A, PID in Channel B, with two packed
injectors.
77
Connecting the Gases and Electrical Supply
Figure 20. ECD in Channel A, ElCD in Channel B, with two packed injectors.
78
Clarus 500 GC Installation Guide
Figure 21. ECD in Channel A, PID/ElCD in Channel B, with two packed injectors.
79
Connecting the Gases and Electrical Supply
Connecting the Gas for Subambient Operation
This section describes how to connect a liquid nitrogen or liquid carbon dioxide supply to
your Clarus 500 GC for subambient operation.
For liquid nitrogen, use a supply with a liquid delivery pressure of 20 – 30 psig. For liquid
carbon dioxide, use a size 1A cylinder equipped with a fitting containing a full-length dip
(eductor) tube and approximately 27.3 kg (60 lb) of carbon dioxide.
The LN2 or CO2 supply tubing connects the supply tank to the 1/4-in. union on the rear of the
Clarus 500 GC. To install the supply tubing, select the procedure that corresponds to the
subambient kit you are installing and refer to Figure 22.
Connect the LN2 Supply Tubing
1. Locate the LN2 tank. Make sure the tank provides a liquid delivery pressure of 20 –
30 psig.
2. Connect the large nut on one end of the LN2 supply tubing (P/N N600-0403) to the
1/2-in. pipe to 1/4-in. Swagelok reducer labeled “LIQUID"
on the LN2 tank by turning the nut clockwise until it is finger-tight. To provide a leakfree connection, tighten the nut an additional half-turn with a large adjustable wrench.
3. Connect the 1/4-in. nut on the other end of the tubing to the 1/4-in. union (see Figure
22) on the rear of the Clarus 500 GC by turning the nut clockwise until it is fingertight. To provide a leak-free connection, place a 9/16-in. wrench on the nut and a 1/2in. wrench on the union, then tighten the nut an additional half-turn.
4. Turn on the valve and check for leaks.
Connect the CO2 Supply Tubing
WARNING
80
Protect the CO2 supply from undue heating. CO2 is a gas above 31°C (87.8°F) at a
pressure of 1069 psig. The pressure increases rapidly at temperatures above 31°C
(87.8°F).
Clarus 500 GC Installation Guide
1. Locate a size 1A CO2 cylinder equipped with a fitting containing a full-length dip
(eductor) tube and approximately 27.3 kg (60 lb) of carbon dioxide.
2. Connected to the CO2 supply tubing (P/N N600-0404) is an envelope containing a
Teflon gasket. This goes inside the nut on the tubing before attaching it to the tank.
Connect the large nut on one end of the CO2 supply tubing to the CO2 tank fitting by
turning the nut clockwise until it is finger-tight. To provide a leak-free connection,
tighten the nut an additional half-turn with a large adjustable wrench.
3. Connect the 1/4-in. nut on the other end of the tubing to the 1/4-in. union on the rear
of the Clarus 500 GC by turning the nut clockwise until it is finger-tight. To provide a
leak-free connection, place a 9/16-in. wrench on the nut and a 1/2-in. wrench on the
bulkhead fitting, then tighten the nut an addition half-turn.
WARNING
The CO2 supply system contains liquid CO2 pressurized at 6000 kPa (870 psig)
at 20°C (68°F). Before breaking connections to change cylinders, shut off the
liquid CO2 cylinder valve, then bleed the supply tubing by closing the oven door
and adjusting the oven temperature below 60°C (140°F). This opens the coolant
solenoid valve and allows the CO2 supply tubing to bleed.
4. Turn on the valve and check for leaks.
81
Connecting the Gases and Electrical Supply
Figure 22. Connecting the CO2 or LN2 supply.
82
Clarus 500 GC Installation Guide
Connect the Electrical Supply
This section describes how to connect the Clarus 500 GC to AC line power both in and
outside the U.S.
Only a PerkinElmer Service representative or similarly trained and
authorized personnel should ever change or replace the plug for the Clarus
500 GC.
WARNING
Connecting the Clarus 500 GC to Line Power in the U.S.
To connect the Clarus 500 GC to line power in the U.S., simply plug the power cord into the
proper AC outlet (see Table 2).
Connecting the Clarus 500 GC to Line Power Outside the U.S.
To connect the Clarus 500 GC to line power outside the U.S., connect a properly grounded
plug to the power cord and change the wiring on the AC Distribution P.C. board as described
in the following procedures.
Table 2 shows the available plugs, their electrical power, and the countries where they are
used.
83
Connecting the Gases and Electrical Supply
Table 2. Plugs Used in Different Countries
84
Clarus 500 GC Installation Guide
In Countries with 230V Single-Phase Power
Only a PerkinElmer Service representative or similarly trained and
authorized personnel should ever change or replace the plug for the
Clarus 500 GC.
WARNING
1. Select the appropriate plug and connect it to the three power cord wires as follows:
•
Connect the brown wire to line (AC).
•
Connect the blue wire to neutral (ACC).
•
Connect the green/yellow wire to chassis ground.
2. Remove the three screws that secure the right side panel to the Clarus 500 GC. Then
remove the panel.
3. Verify that the brown wire from the line filter is connected to E1, that the blue wire
from the line filter is connected to E2, and that the yellow/green wire is connected to
E3 on the AC Distribution P.C. board (see Figure 23).
4. Make sure that a 10-A Slow-Blow fuse is installed in the F3 fuse holder and a wire is
soldered across the contacts of the F3 fuse holder to short this connection.
5. Replace the right side panel and connect the power cord to a suitable electrical outlet.
In Countries with 230V Two-Phase Power
Only a PerkinElmer Service representative or similarly trained and
authorized personnel should ever change or replace the plug for the
Clarus 500 GC.
WARNING
1. Select the appropriate plug and connect it to the three power cord wires as follows:
85
Connecting the Gases and Electrical Supply
•
Connect the brown wire to line (AC Hot).
•
Connect the blue wire to line (AC Hot).
•
Connect the green/yellow wire to chassis ground.
2. Remove the three screws that secure the right side panel to the Clarus 500 GC and
remove the panel.
3. Verify that the brown wire from the line filter is connected to E1, that the blue wire
from the line filter is connected to E2, and that the green/yellow wire is connected to
E3 on the AC Distribution P.C. board (see Figure 23).
4. Cut and remove the wire that is used as a jumper across the F3 fuse holder. Make sure
that the 10-A Slow-Blow fuse remains installed in the F3 fuse holder.
5. Replace the right side panel and connect the power cord to a suitable electrical outlet.
86
Clarus 500 GC Installation Guide
Figure 23. AC Distribution P.C. board connections for countries that use single-phase
or two-phase power.
87
Connecting the Gases and Electrical Supply
88
Connecting the
Accessories
6
Clarus 500 GC Installation Guide
Connect the ElCD
If you ordered your Clarus 500 GC with either an EICD (also called a Hall detector) or an
EICD in series with a photoionization detector (PID), you must perform the following
procedures:
•
Connect the EICD Control Unit to the EICD Components
•
Connect the EICD solvent system
•
Make the EICD/PID series connection
•
Check the calibration
•
If necessary, convert the EICD Control Unit line voltage setting to 230-V operation.
Refer to the procedure described in the Tremetrics Instruction Manual that is shipped
with the Control Unit.
Never operate the Control Unit at an incorrect voltage setting.
WARNING
91
Connecting the Accessories
Figure 24. ElCD Control Unit front panel.
92
Clarus 500 GC Installation Guide
Connect the ElCD Control Unit to the ElCD
Components
To connect the ElCD Control Unit to the ElCD components, refer to Figure 25 and follow
this procedure:
Figure 25. Connection diagram of the ElCD Control Unit to ElCD components.
1. Plug the connector on the end of the Reactor Cable Assembly into the ElCD’s Reactor
Connector on the rear panel of the Control Unit.
2. Connect the Conductivity Cell Drive Cable and the Signal Cable to the following
connectors on the Control Unit rear panel:
•
Connect the Drive Cable red lead to CELL DRIVE and the black lead to SIG
COM.
93
Connecting the Accessories
•
Connect the Signal Cable red lead to CELL SIGNAL and the black lead to SIG
COM.
3. If you are installing an ElCD amplifier board (P/N N612-0060) in the Clarus 500 GC,
locate the External Amplifier Cable Assembly
(P/N N610-0187) supplied with the amplifier kit.
4. Connect one end of the External Amplifier Cable Assembly to connector J2 on the
ElCD amplifier board that is installed in the Clarus 500 GC, and at the other end of
the cable connect the red lead to 10V FS and the black lead to SIG COM on the
Control Unit rear panel (top two connectors).
5. Connect the solenoid vent valve cable to TE3 (J20) on the main board in the Clarus
500 GC electronics compartment behind the right side panel.
6. Secure all cables in the existing cable clamps.
Figure 26. ElCD solenoid valve connection to TE3 on the Main Board in the Clarus 500
GC.
94
Clarus 500 GC Installation Guide
Connect the ElCD Solvent System
Figure 27. Connecting the ElCD solvent system.
To connect the ElCD solvent system, refer to Figures 27 and 28 and follow the procedure
below. The lines have been connected to the detector assembly at the Clarus 500 in the
factory. These lines have been dressed to the back of the unit.
1. At the back of the ElCD Control Unit, connect the output of the pump (the elbow
fitting on the left side of the pump) to the top of the ion exchange tube. The ion
exchange tube is inside the solvent reservoir.
95
Connecting the Accessories
2. Connect the solvent reservoir to the fitting on the right side of the pump.
3. Connect the top of the ElCD reactor tube to the Conductivity Cell.
4. Place the drain tubing from the bottom of the Conductivity Cell into the solvent
reservoir.
Figure 28. Solvent Delivery System Connections
96
Clarus 500 GC Installation Guide
Make the ElCD/PID Series Connection
Use the following procedure to make the ElCD/PID series connection. After you make this
connection, you can operate either the ElCD or the PID without any loss of sensitivity.
NOTE: Always connect the analytical column to the PID, and the PID exit tube to the ElCD.
1. Cut a length of the supplied deactivated 0.53-mm. I.D. fused silica tubing (P/N N9301358) long enough to connect the PID exit line to the ElCD. Try to use the shortest
possible length (approximately 12 inches).
2. nsert a 1/16-in. column nut and a 1/16-in. graphite/Vespel ferrule
(P/N 0992-0301) over one end of the cut fused-silica tubing. Make sure the tapered
end of the ferrule faces the nut as shown below.
Figure 29. Placing a 1/16-in. nut and graphite ferrule over the fused silica line.
3. Cut about 1 cm (3/8 inch) from the end of the fused-silica tubing using a wafer scribe
(P/N N930-1386, pkg. of 10 scribes) or other column cutting tool. Break off the tubing
at the score mark and make sure that the break is clean and square. Examine the cut
with a magnifying glass and compare it to the examples below.
97
Connecting the Accessories
Figure 30. Examples of a good tubing cut and bad cuts.
4. Position the nut on the fused-silica tubing so that the back of the nut is
8.8 cm (3 1/2 inches) from the end of the fused-silica tubing. This positions the fusedsilica tubing inside the ElCD receiver.
5. Using typewriter "white-out" or a felt-tipped pen, make a mark on the fused-silica
tubing just beyond the back edge of the nut.
CAUTION
To avoid contaminating the system, make sure the nut does not come
into contact with the mark on the fused-silica tubing.
6. Connect the fused-silica tubing to the ElCD fitting inside the oven as follows:
a) Insert this end of the fused-silica tubing into the ElCD fitting. Slide the nut up to
the ElCD and turn it until it is fingertight.
b) Push the fused-silica tubing into the ElCD fitting until the mark on the tubing is
aligned with the back of the nut.
c) Using two 1/4-inch wrenches, tighten the nut only until the fused-silica tubing
cannot be pulled out of the ElCD fitting. Do not overtighten the nut!
98
Clarus 500 GC Installation Guide
7. Connect the other end of the fused-silica tubing to the PID exit tubing as follows:
a) Insert the nut on one end of the 1/16 to 1/16-in. union (P/N 0990-3075) over the
PID exit tube until it bottoms. Using two wrenches, tighten the nut 1/4 turn past
fingertight.
b) Remove the nut and metal ferrule from the other end of the 1/16 to 1/16-in.
union. Put the metal ferrule aside; you will not need it.
c) Insert the nut and a 1/16-inch graphite/Vespel ferrule (P/N 0992-0301) over the
free end of the fused-silica tubing.
d) Insert the end of the fused-silica tubing into the union until it bottoms. Slide the
nut up to the union and tighten it fingertight.
e) Using two 1/4-in. wrenches, tighten the nut only until the fused-silica tubing
cannot be pulled out of the ElCD fitting. Do not overtighten the nut!
You can operate the ElCD and the PID in series or operate either detector independently.
Always connect your analytical column to the PID.
Figure 31. Connecting the fused silica line to the ElCD and the PID.
99
Connecting the Accessories
Check the Calibration
After you have made all the proper connections, check the calibration by following this
procedure:
1. Connect the Control Unit to the AC line power.
2. Select SOLVENT FLOW. Then set the SOLVENT FLOW to 0.8 mL/min. You can
set the solvent flow to 0.8 mL/min by turning the knob to the right of SOLVENT
FLOW until 60 appears.
3. Start the flow by setting the switch to the right of SOLVENT FLOW to the On
position.
4. Measure the solvent flow at the exit of the Conductivity Cell and adjust the
SOLVENT FLOW until you get 0.8 mL/min.
5. Allow the ElCD to equilibrate at 0.8 mL/min for 4 to 5 hours.
6. Check the CALIBRATE display. It should indicate 2000 + 20.
7. If the CALIBRATE display does not indicate 2000 + 20, refer to Diagnostic Display
in the Operation Section of the ElCD Operator’s Manual that is supplied with the
ElCD.
100
Clarus 500 GC Installation Guide
Install the NPD Bead Assembly
If you purchased your Clarus 500 GC equipped with a Nitrogen Phosphorus Detector (NPD),
the NPD bead assembly is packaged separately and shipped inside the Clarus 500 GC oven.
To install the NPD bead assembly, use the following procedure and refer to Figure 32.
CAUTION
After installing a NPD bead, you must set up the NPD and condition a
new bead before you can use it. Refer to Setting Up a NPD in Chapter 8
of the Clarus 500 GC User’s Manual (P/N 0993-6073).
1. Open the detector cover.
2. Remove the bead assembly package from the Clarus 500 GC oven.
3. Loosen the knurled ring on the NPD collector assembly. Then remove the collector
assembly.
CAUTION
You may find it easier to first remove the coaxial cable from the
collector assembly before you remove the collector assembly.
4. Remove the screw that secures the bead transformer assembly to the top of the Clarus
500 GC oven.
5. Plug the bead assembly (P/N N612-0092) into the connector on the bead transformer.
The connector is keyed so that the bead assembly can only be inserted one way.
6. Carefully insert the bead portion of the bead assembly in the detector body as shown
in Figure 32.
7. Secure the bead transformer to the top of the oven with the screw removed in step 4.
8. Replace the collector assembly on the detector body and secure it by tightening the
knurled ring. If the coaxial cable was removed, connect it to the collector assembly.
101
Connecting the Accessories
CAUTION
Ensure that the polarizing wire is still connected to the polarizing pin on the
detector body before you set up and run the NPD.
Figure 32. Exploded view of the Nitrogen Phosphorus Detector.
Connect a Recorder or Integrator
This section describes how to connect an optional recorder or integrator to the Clarus 500
GC. To connect either one of these devices, use the following procedure.
102
Clarus 500 GC Installation Guide
1. To connect a device to amplifier Channel A or B, first configure output 1 or 2 for a
recorder or integrator as described in Chapter 3, Initial Setup Procedures, in the
Clarus 500 GC User's Manual.
2. Open the small access cover on top of the Clarus 500 GC. With a large flat bladed
screw driver loosen the large screw. On the main cover is a screw well. Loosen this
screw with the large flat bladed screwdriver. You will then be able to lift the entire
cover up and have access to the recorder or integrator.
3. To connect a recorder or integrator to Channel A, connect the ground lead of the
device to TB1-13 and the positive lead of the device to TB1-14. Make sure you route
the wires around the strain relief posts as shown in Figure 34.
4. To connect a recorder or integrator to Channel B, connect the ground lead to TB1-15
and the positive lead to TB1-16.
Figure 33. Recorder and integrator connections at TB1.
In addition to the amplifier Channel A and B contacts, TB1 has READY and START
contacts. Table 3 describes the functions of these contacts.
103
Connecting the Accessories
Table 3. READY and START Connections at TB1
Connection
Function
READY OUT
TB1-10 (C), 11 (NO), 12 (NC)
Instrument READY OUT Relay: These contacts are
used to tell an external device that the Clarus 500 GC
is ready. The normally open contact (NO) is closed
in the Ready state.
START OUT
TB1-7 (C), 8 (NO), 9 (NC)
Instrument START OUT Relay: These contacts are
used to start an external device, such as an integrator,
when the Clarus 500 GC starts a chromatographic
run. The normally open contact is closed for 1 sec
when a run is started.
EXT. READY
(TB1-5, 6)
TB1-6 is signal
External Ready In: The Clarus 500 GC requires that
these contacts be shorted together to become ready
and is shipped with a link across them. When using
an external device, such as an integrator, remove the
link and wire the device to provide a contact closure
indicating the Ready state. This will prevent the
instrument from becoming ready before the external
device is ready.
TB1-5 is ground
EXT. START
(TB1-3, 4)
TB1-4 is signal
TB1-3 is ground
104
External Start In: Shorting these contacts will cause
the Clarus 500 GC to start a chromatographic run. It
is equivalent to pressing the RUN key.
Clarus 500 GC Installation Guide
Figure 34. Routing recorder/integrator wires around the strain relief posts.
105
Connecting the Accessories
106
PPC Restrictor
Information
7
Clarus 500 GC Installation Guide
Available Restrictors
Restrictor
Number
Color
Helium Flow at 30 PSIG, vent to
atmosphere
1
Silver
1 mL/min
2
Red
3 mL/min
3
Yellow
10 mL/min
4
Black
30 mL/min
5
Green
100 mL/min
6
Blue
300 mL/min
Restrictors for Carrier Gas Control
Injector
Restrictor Number
Helium Flow (mL/min)
(with a 90 psig inlet pressure and a 10 psig
drop across the restrictor)
Packed
4 (std)
30
Split/Splitless or
PSS
6 (std)
300
Split/Splitless or
PSS
4 (acc)
30
POC
3 (std)
10
POC
2 (acc)
3
109
PPC Restrictor Information
Restrictors for Detector Gas Control
NOTE: Maintain an inlet pressure between 60 and 90 psig.
110
Detector
Gas
Restrictor
Number
Nominal Flow
Required (mL/min)
FID
Air
6
450
FID
Hydrogen
4
45
NPD
Air
5
100
NPD
Hydrogen
1
2
FPD
Air
5
90
FPD
Hydrogen
4
75
ECD
Argon/Methane
4
30
ECD
Nitrogen
4
30
TCD
Helium
4
30
TCD
Hydrogen
3
30
ElCD
Hydrogen
3
25
Index
A
About This Manual, 7
Autosampler
autosampler tower, 49
autosampler tower cover, 53
encoder protector block, 48
Autosampler Installation, 47
Autosampler Tower
installation, 49
syringe cover, 55
Autosampler Tower Cover
installation, 53
C
Calibration
check, 100
Chemicals
Definitions of Warnings, 32
Hazardous, 31
Compressed gases, safety practices, 27
Conventions
Notes, cautions and warnings, 8
text, 8
E
ECD
cell failure, 25
damage, 25
labels, 24
purchasers, 24
reporting radioactive exposure, 25
reporting theft or losses, 25
testing for radioactive leaks, 25
U.K. regulations, 26
EICD Connections, 91
EICD Solvent System Connections, 95
Electrical Power
connections, 83
Electrical Safety, 20
Electrical, safety practices, 21
Encoder Protector Block, 48
Environmental Conditions, 18, 37
Index
Exhaust Vent Requirements, 37
R
G
heated zones, safety practices, 16
Radioactive exposure, reporting, 25
Recorder
connection, 102
Requirements
electrical power, 41
environmental, 41
gas, 43
laboratory space, 40
safety, 42
Restrictors, 109
carrier gas control, 109
detector gas control, 110
I
S
Installation
NPD bead assembly, 101
Integrator
connection, 102
Safety
chemical use, 31
compressed gases, 27
ECD Radioactive Hazards, 24
electrical, 21
generic warnings, 15
heated zones, 16
high voltage, 20
moving the GC, 23
thermal runaway protection, 16
ventilation, 27
Safety Information, 17
electrical safety, 20
environmental conditions, 18
gas cylinders, 27
Gas Cylinders, 27
Gases
common injector/detector gas connections, 69
connections, 61
leak test, 68
subambient connections, 80
H
L
Labels
Warning Signs, 10
Laboratory Preparation, 37
vent system, 38
N
NPD Bead Assembly
installation, 101
P
Precautions, 17
Preparing Your Laboratory, 37
environmental conditions, 37
exhaust vent, 37
V
Vent
system, recommendations, 38
Ventilation, safety practices, 27
W
Warnings
Hazardous Chemical, 32
112

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement