Air Techniques ProVecta HD Intraoral X-Ray Owner's Manual
Air Techniques ProVecta HD Intraoral X-Ray is a versatile and reliable dental imaging solution designed to meet the diverse needs of modern dental practices. With its advanced features and exceptional image quality, the ProVecta HD empowers clinicians to make confident diagnostic decisions and enhance patient care.
Key capabilities of the ProVecta HD include:
-
16" C to C Wall Mount Plate for secure and stable installation on wooden wall studs.
-
Precise and accurate positioning with the 16-inch on-center mount installation method.
-
Robust construction capable of withstanding significant shear and withdrawal forces, ensuring reliable operation even in demanding environments.
Advertisement
Advertisement
Intraoral Dental X-ray System
Wall Mounting Plate, 16” C to C,
Installation Kit
Part Number: A6330
INSTRUCTION SHEET
General
The ProVecta HD 16 Center to Center Wall Mount Plate is used to secure the ProVecta HD X-ray unit onto a facility wall via two wooden 2” X 4” (1 1/2” x 3 1/2”) wall studs located 16 inches on center apart. This kit is intended for use with wood wall studs and provides appropriate hardware. Use is prohibited on installations with metal wall studs. Make sure to perform the installation using and in accordance with local code.
16 Inch on Center Mount Installation
Install the Provecta HD using the 16 inch on center mounting method by performing the following procedures.
1. Locate the wall studs to mount 16 inch on center wall plate (P/N A6341). Install power and control wiring electrical boxes using template P/N A6343 for proper positioning.
2. Mark template holes at screw holes (A, B, C, D) making sure each is centered on the wooden wall studs.
3. Drill 1/4-inch pilot holes of the wooden wall studs.
4. Mount the 16 inch on center wall plate to wall studs using four 3/8 inch x 3 1/2 inches lag bolts with associated spring washers and flat washers into screw holes (A, B, C, D).
Make sure plate is level before tightening bolts.
5. Cover each screw hole by installing four screw hole plugs.
6. Mount the X-Ray Control Unit onto 16 inch on center wall plate by aligning screw holes and installing four 1/4 inch x 20 inches hex head bolts with associated spring washers and flat washers into screw holes (1, 2, 3, 4) .
7. Level plate and tighten bolts making sure plate stays level.
8. Continue with Provecta HD installation.
A
D
Mounting
Screw Hole
1
Cable/Wiring
Feed Through
3
Mounting
Screw Hole
Mounting
Screw Hole
2
Cable/Wiring
Feed Through
Mounting
Screw Hole
4
C
B
Caution: Load Capacity
The wall support and mounting hardware for installing the Provecta HD using the
16 center to center wall mount plate must withstand a shear load 240 pounds (109 kg) and a withdrawal force of 1560 lbf (708 kgf).
The wall fabrication and attachments to the building structure must be capable of withstanding a moment load of 1474 lbf ft (204 kgf m)
A
D
C
B
Caution
Kit is designed for use with wood wall studs and not intended to be used on any other installations.
Not for use with metal wall studs.
Never reuse old screw holes.
Supplied Contents of Kit P/N A6330
Part No.
Description
A6341
A6343
31547
Wall Mounting Plate, 16” C To C and fastening hardware
Installation Template,16” C To C Wall Mount
Screw hole plugs
Qty
1
1
4
1295 Walt Whitman Road | Melville, New York 11747- 3062
Phone: 800-247-8324 | Fax: 888-247-8481
Website: www.airtechniques.com
3
1
2
4
A6427
Included for Customer Convenience
5MM Power Ball-Hex T-Key Wrench
For instructions for use, refer to the ProVecta HD
Installation Guide (A6385) section titled
“Scissor Arm Tension Alignment”
© Air Techniques, Inc. Copyright 2012
•
P/N A6344, Rev. D - July 2021