Extech Instruments EX210 8 Function Mini Digital MultiMeter User guide


Add to my manuals
15 Pages

advertisement

Extech Instruments EX210 8 Function Mini Digital MultiMeter User guide | Manualzz

User Guide

Multimeter with IR Thermometer

Extech EX210 and EX210T (True RMS)

Introduction

Thank you for selecting the Extech EX210/EX210T True RMS (EX210T only) Autoranging Multimeter.

This meter measures AC/DC Voltage, AC/DC Current, Resistance, Diode Test, and Continuity plus non-contact InfraRed Temperature. Proper use and care of this meter will provide many years of reliable service. Please visit the Extech Instruments website ( www.extech.com

) to check for the latest version of this User Guide. Extech Instruments is an ISO-9001 certified company.

Safety

This symbol adjacent to another symbol, terminal or operating device indicates that the operator must refer to an explanation in the Operating Instructions to avoid personal injury or damage to the meter.

WARNING

This WARNING symbol indicates a potentially hazardous situation, which if not avoided, could result in death or serious injury.

CAUTION

MAX

600V

This CAUTION symbol indicates a potentially hazardous situation, which if not avoided, may result damage to the product.

This symbol advises the user that the terminal(s) so marked must not be connected to a circuit point at which the voltage with respect to earth ground exceeds (in this case) 600 VAC or VDC.

This symbol adjacent to one or more terminals identifies them as being associated with ranges that may, in normal use, be subjected to particularly hazardous voltages. For maximum safety, the meter and its test leads should not be handled when these terminals are energized.

This symbol indicates that a device is protected throughout by double insulation or reinforced insulation.

PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY

OVERVOLTAGE CATEGORY I

Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measures are taken to limit the transient overvoltages to an appropriate low level.

Note – Examples include protected electronic circuits.

OVERVOLTAGE CATEGORY II

Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation.

Note – Examples include household, office, and laboratory appliances.

OVERVOLTAGE CATEGORY III

Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations.

Note – Examples include switches in the fixed installation and some equipment for industrial use with permanent connection to the fixed installation.

OVERVOLTAGE CATEGORY IV

Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation.

Note – Examples include electricity meters and primary over-current protection equipment

2 EX210_EX210T-en-US_v1.5 9/15

CAUTIONS

 Improper use of this meter can cause damage, shock, injury or death. Read and understand this user manual before operating the meter.

 Always remove the test leads before replacing the battery or fuses.

 Inspect the condition of the test leads and the meter itself for any damage before operating the meter.

 Use great care when making measurements if the voltages are greater than 25 VAC rms or 35

VDC. These voltages are considered a shock hazard.

 Warning! This is a Class A device. This device can cause interference in residential areas.

 Always discharge capacitors and remove power from the device under test before performing

Diode, Resistance or Continuity tests.

 Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty of connection to the recessed electrical contacts. Other means should be used to ensure that the terminals are not "live".

 If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.

 This device must not be used by children. It contains hazardous objects as well as small parts that the children could swallow.

 Do not leave batteries and packing material lying around unattended; they can be dangerous to children.

 In the event that this device will be stored for an extended period of time, remove the batteries.

 Expired or damaged batteries can leak and can be hazardous to skin. Use suitable hand gloves in such cases.

 Do not short-circuit the battery. Do not place the battery near a fire.

 Never mix battery types. Always install new batteries of the same type.

3 EX210_EX210T-en-US_v1.5 9/15

SAFETY INSTRUCTIONS

This meter has been designed for safe use, but must be operated with caution. The rules listed below must be carefully followed for safe operation.

1. NEVER apply voltage or current to the meter that exceeds the specified maximum:

Input Protection Limits

V AC/DC, Resistance, Diode Test, Continuity

μA or mA AC/DC

A AC/DC

2. USE EXTREME CAUTION when working with high voltages.

600 VDC/AC rms

200mA fused

10A fused

3. DO NOT measure voltage if the voltage on the "COM" input jack exceeds 600V above earth ground.

4. NEVER connect the meter leads across a voltage source while the function switch is in the current, resistance, or diode mode. Doing so can damage the meter.

5. ALWAYS discharge filter capacitors in power supplies and disconnect the power when making resistance or diode tests.

6. ALWAYS turn off the power and disconnect the test leads before opening the covers to replace the fuse or batteries.

7. NEVER operate the meter unless the back cover and the battery and fuse covers are in place and fastened securely.

8. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.

4 EX210_EX210T-en-US_v1.5 9/15

Controls and Jacks

3. 2000 count LCD

7. mA, µA and 10A input jacks

8. IR Thermometer button

10. RANGE button

11. Positive input jack

12. COM input jack

Note: Tilt stand, test lead holders, and battery compartment are on rear of unit.

Symbols and Enunciators

Continuity test

-6 ) (amps)

A Amps k kilo

) (ohms)

 Ohms

V Volts current

Fahrenheit

MAX Maximum

AUTO Autoranging hold

SCAN IR temperature in progress

5 EX210_EX210T-en-US_v1.5 9/15

Operating Instructions

WARNING: Risk of electrocution. High-voltage circuits, both AC and DC, are extremely dangerous and should be measured with great care.

1. ALWAYS turn the function switch to the OFF position when the meter is not in use.

2. If “OL” appears in the display during a measurement, the value exceeds the range you have selected. Change to a higher range.

AC/DC VOLTAGE MEASUREMENTS

CAUTION: Do not measure DC voltages if a motor on the circuit is being switched ON or OFF.

Large voltage surges may occur that can damage the meter.

1. Rotate the function switch to the V AC or V DC position.

2. Insert the black test lead banana plug into the negative COM jack. Insert the red test lead banana plug into the positive V jack.

3. Touch the black test probe tip to the negative side of the circuit.

Touch the red test probe tip to the positive side of the circuit.

4. Read the voltage in the display.

AC/DC CURRENT MEASUREMENTS

1. Insert the black test lead banana plug into the negative COM jack.

2. Press the MODE button to indicate “DC” or “AC” on the display.

3. For current measurements up to 2000µA, set the function switch to the µA position and insert the red test lead banana plug into the µA/mA jack.

4. For current measurements up to 200 mA DC, set the function switch to the mA position and insert the red test lead banana plug into the µA/mA jack.

5. For current measurements up to 10A DC, set the function switch to the 10A position and insert the red test lead banana plug into the 10A jack.

6. Connect the test leads in series with the circuit under test.

7. Apply power to the circuit.

8. Read the current in the display.

6 EX210_EX210T-en-US_v1.5 9/15

RESISTANCE MEASUREMENTS

WARNING: To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any resistance measurements. Remove the batteries and unplug the line cords.

1. Rotate the function switch to the position.

2. Insert the black test lead banana plug into the negative COM jack.

Insert the red test lead banana plug into the positive  jack.

3. Touch the test probe tips across the device under test.

4. Read the resistance in the display.

CONTINUITY CHECK

WARNING: To avoid electric shock, never measure continuity on circuits or wires that have a voltage potential.

1. Rotate the function switch to the position.

2. Insert the black lead banana plug into the negative COM jack.

Insert the red test lead banana plug into the positive jack.

3. Press the MODE button to indicate“ " on the display

4. Touch the test probe tips to the circuit or wire you wish to check.

5. If the resistance is less than the continuity threshold, the audible signal will sound.

DIODE TEST

1. Rotate the function switch to the green position.

2. Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the positive jack.

3. Press the MODE button to indicate “ “and “V” on the display.

4. Touch the test probes to the diode under test. Forward voltage will typically indicate

0.400 to 0.700V. Reverse voltage will indicate “OL”. Shorted devices will indicate near 0V and an open device will indicate “OL” in both polarities.

7 EX210_EX210T-en-US_v1.5 9/15

NON-CONTACT INFRARED TEMPERATURE MEASUREMENTS

1. Rotate the function switch to the IR position. The upper right four digit temperature display will switch on.

3. Aim the infrared sensor (top of meter) at the surface to be measured.

4. Press and Hold the IRT button to turn on the IR thermometer and laser pointer. The laser pointer identifies the surface spot to be measured and “SCAN” will flash in the display.

5. Read the temperature in the display.

6. The area of the surface to be measured must be larger than the spot size as determined by the distance to spot size values listed in the specification table. the button to turn the IR thermometer and laser pointer off. “HOLD” and the final measured value will remain in the display.

WARNING: Do not directly view or direct the laser pointer at an eye. Low power visible lasers do not normally present a hazard, but may present some potential for hazard if viewed directly for extended periods of time.

IR Spot to Distance Diagram

The 4:1 spot to distance ratio determines the size of the measured surface area with respect to the distance the meter is held away from the surface.

1”

@ 4”

2”

@ 8”

4”

@ 16”

20mm

@ 160mm 80mm

IR Measurement Notes

1. The object under test should be larger than the spot (target) size calculated by the field of view diagram.

2. If the surface of the object under test is covered with frost, oil, grime, etc., clean before taking measurements.

3. If an object's surface is highly reflective, apply masking tape or flat black paint to the surface before measuring.

4. The meter will not make measurements through glass steam, dust, and smoke.

5. To find a hot spot, aim the meter outside the area of interest then scan across (in an up and down motion) until the hot spot is located.

8 EX210_EX210T-en-US_v1.5 9/15

AUTORANGING/MANUAL RANGE SELECTION

When the meter is first switched on, it automatically enters the AutoRanging mode. This automatically selects the best range for the measurements being made and is generally the best mode for most measurements. For measurement situations requiring that a range be manually selected, perform the following:

1. Press the RANGE button. The “AUTO” display indicator will turn off.

2. Press the RANGE key to step through the available ranges until the range desired is selected.

3. Press and hold the RANGE button for 2 seconds to exit manual ranging.

Note: Manual ranging does not apply to the Temperature, Diode and Continuity functions.

MAX (MAXIMUM READING) MODE

1. Press the MAX button to activate the MAX mode. The display icon "MAX" will appear. The meter will display and hold the maximum reading and will update only when a new maximum value is detected.

2. Press the MAX button again to exit the mode.

Note: Max does not apply to the Resistance, Diode and Continuity functions.

DISPLAY BACKLIGHT

Press and hold the IRT button for 2 second to turn the backlight on. The backlight will automatically turn off after approximately 10 seconds or press and hold the button again for 2 seconds to manually turn it off.

HOLD

The HOLD function freezes the reading in the display. Press the HOLD button momentarily to activate or to exit the HOLD function.

Note: The HOLD button does not function in the IR measurement mode.

AUTO SLEEP

If no button is pressed the auto sleep feature will place the meter in sleep mode after approximately

15 minutes of operation. If this happens, press any button to wake the meter or switch the meter to

OFF if it no longer in use.

9 EX210_EX210T-en-US_v1.5 9/15

Maintenance

WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the back cover or the battery or fuse covers.

WARNING: To avoid electric shock, do not operate the meter until the battery and fuse covers are in place and fastened securely.

This MultiMeter is designed to provide years of dependable service, if the following care instructions are followed:

1. KEEP THE METER DRY. If it gets wet, wipe it off.

2. USE AND STORE THE METER IN NORMAL TEMPERATURES. Temperature extremes can shorten the life of the electronic parts and distort or melt plastic parts.

3. HANDLE THE METER GENTLY AND CAREFULLY. Dropping it can damage the electronic parts or the case.

4. KEEP THE METER CLEAN. Wipe the case occasionally with a damp cloth. DO NOT use chemicals, cleaning solvents, or detergents.

5. USE ONLY FRESH BATTERIES OF THE RECOMMENDED SIZE AND TYPE. Remove old or weak batteries so they do not leak and damage the unit.

6. IF THE METER IS TO BE STORED FOR A LONG PERIOD OF TIME, the batteries should be removed to prevent damage to the unit.

BATTERY and FUSE Replacement

WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the battery cover.

1. Turn power off and disconnect the test leads from the meter.

2. Lift the meter stand to expose the battery cover.

3. Open the rear battery/fuse compartment by removing the Phillips head screw on the lower rear of the meter.

4. Carefully pull down on the cover just enough to release the latch and then lift it off. The cover will not pull down all the way.

5. Remove the old battery or fuse and install a new one of the correct rating.

6. Place the battery/fuse cover back in place. Secure with the screw.

WARNING: To avoid electric shock, do not operate the meter until the battery cover is in place and fastened securely.

NOTE: If the meter does not work properly, check the fuses and replace the battery to make sure that they are still good and that they are properly inserted.

Battery Safety Reminders o Please dispose of batteries responsibly; observe local, state, and national regulations. o Never dispose of batteries in a fire; batteries may explode or leak. o Never mix battery types; install new batteries of the same type.

10 EX210_EX210T-en-US_v1.5 9/15

Specifications

Function Range

DC Voltage 200 mV

Resolution

0.1 mV

Accuracy

(0.8% reading + 6 digits)

2V 0.001V

20V 0.01V

(0.5% reading + 2 digits)

200V 0.1V

600V 1V

(0.8% reading + 2 digits)

AC Voltage

(50/60Hz)

200 mV 0.1 mV

2V 0.001V

20V 0.01V

True RMS

(EX210T only)

200V 0.1V

600V 1V

(1.5% reading + 6 digits)

DC Current

All AC voltage ranges are specified from 5% of range to 100% of range

200 A 0.1

2000 A 1

20 mA

200 mA

0.01 mA

0.1 mA

(1.5% reading + 5 digits)

(2.5% reading + 5 digits)

10 A 0.01 A

Note:10A for 30 sec max

AC Current

(50/60Hz)

True RMS

(EX210T only)

200 A 0.1

2000 A 1

20 mA

200 mA

0.01 mA

0.1 mA

(1.5% reading + 8 digits)

(3.0% reading + 5 digits)

10 A 0.01 A

Note: 10A for 30 sec max

NOTE: Accuracy is stated at 18 o C to 28 o C (65 o F to 83 o F) and less than 75% RH.

11 EX210_EX210T-en-US_v1.5 9/15

Function Range Resolution

Resistance 200  0.1

2 k  0.001

Accuracy

(0.8% reading + 5 digits)

20 k  0.01 (0.8% reading + 2 digits)

200 k  0.1

2 M  0.001

(2.5% reading + 8 digits)

Temperature IR

20 M  0.01

-20 to 0

0 to 93

C

C

93 to 230 C

-5 to 32 F

0.1

C/F

4C

(2.0%rdg or 2C)

(3.0%rdg or 3C)

 8F

32 to 200°F (2.0%rdg or 4F)

200 to 446°F (3.0%rdg or 6F)

Enclosure

Diode Test

Double molded

Test current of 0.9mA maximum, open circuit voltage 2.8V DC typical

Continuity Threshold 20 to 50 , test current <1.5mA

Input Impedance 10MΩ VDC/VAC

AC Response

ACV Bandwidth

True rms (EX210T only)

40Hz to 1000Hz

IR Spectral response 6 to 16µm

IR Emissivity 0.95 fixed

IR distance ratio

Laser pointer

4:1

Class 2 laser < 1mW power; Wavelength is 630 to 670nm

Display 2,000 count backlit liquid crystal display

Overrange indicationOL” is displayed

Auto Power Off 15 minutes (approximately)

Polarity

Measurement Rate

Battery

Fuses

Automatic (no indication for positive); Minus (-) sign for negative

2 times per second, nominal

One 9 volt (NEDA 1604) battery mA, µA ranges; 200mA 250V ceramic fast blow

‘A’ range; 10A 600V ceramic fast blow

Operating Temperature 14ºF to 122ºF (-10ºC to 40ºC)

Storage Temperature 14 o F to 140 o F (-10 o C to 60 o C)

Operating Humidity

Storage Humidity

Max 80% up to 31ºC (87ºF) decreasing linearly to 50% at 40ºC (104ºF)

<80%

Operating Altitude

Weight

Size

Safety

7000 ft (2000m) maximum

9.17 oz. (260 g)

5.8x2.9x1.6” (147x76x42 mm)

Approvals

This meter is intended for origin of installation use and protected, against the users, by double insulation per IEC/EN 61010-1:2001 and IEC/EN

61010-031:2002 to Category III 600V; Pollution Degree 2.

CE

12 EX210_EX210T-en-US_v1.5 9/15

Warranty 

FLIR Systems, Inc. warrants this Extech Instruments brand device to be free of defects in parts and workmanship for one year  from date of shipment (a six month limited warranty applies to sensors and cables). If it should become necessary to return the  instrument for service during or beyond the warranty period, contact the Customer Service Department for authorization. Visit  the website www.extech.com for contact information. A Return Authorization (RA) number must be issued before any product is  returned. The sender is responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transit. 

This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation outside of  specification, improper maintenance or repair, or unauthorized modification. FLIR Systems, Inc. specifically disclaims any implied  warranties  or  merchantability  or  fitness  for  a  specific  purpose  and  will  not  be  liable  for  any  direct,  indirect,  incidental  or  consequential damages. FLIR’s total liability is limited to repair or replacement of the product. The warranty set forth above is  inclusive and no other warranty, whether written or oral, is expressed or implied. 

Calibration, Repair, and Customer Care Services 

FLIR Systems, Inc. offers repair and calibration services for the Extech Instruments products we sell. 

NIST  certification  for  most  products  is  also  provided.  Call  the  Customer  Service  Department  for  information  on  calibration  services  available  for  this  product.  Annual  calibrations  should  be  performed  to  verify  meter  performance  and  accuracy.  Technical  support  and  general  customer  service is also provided, refer to the contact information provided below. 

Support Lines: U.S. (877) 439‐8324; International: +1 (603) 324‐7800 

Technical Support: Option 3; E‐mail: [email protected] 

Repair & Returns: Option 4; E‐mail: [email protected] 

Product specifications are subject to change without notice 

Please visit our website for the most up‐to‐date information  www.extech.com 

FLIR Commercial Systems, Inc., 9 Townsend West, Nashua, NH 03063 USA  

ISO 9001 Certified 

Copyright © 2013‐2015 FLIR Systems, Inc. 

All rights reserved including the right of reproduction in whole or in part in any form 

www.extech.com 

13 EX210_EX210T-en-US_v1.5 9/15

Garantie 

FLIR Systems, Inc. garantit que cet appareil Extech Instruments est exempt de défauts matériaux et de fabrication pendant un  an à partir de la date d’envoi (une garantie limitée de six mois s’applique aux capteurs et aux câbles). Si le renvoi de l’appareil  pour  réparation  devient  nécessaire  durant  ou  après  la  période  de  garantie,  contactez  le  service  client  pour  autorisation.  Pour  obtenir les coordonnées, visitez le site Web suivant : www.extech.com. Un numéro d’autorisation de retour (AR) doit être délivré  avant tout retour de produit. L’expéditeur prend à sa charge les frais d’expédition, le fret, l’assurance et l’emballage correct de  l’appareil afin de prévenir toute détérioration durant le transport. Cette garantie ne s’applique pas aux dommages imputables à  l’utilisateur,  tels  que  l’usage  impropre  ou  abusif,  un  mauvais  câblage,  une  utilisation  non  conforme  aux  spécifications,  un  entretien ou une réparation incorrecte, ou toute modification non autorisée. FLIR Systems, Inc. déclinera spécifiquement toute  garantie  ou  qualité  marchande  ou  aptitude  à  l’emploi  prévu,  et  ne  sera  en  aucun  cas  tenu  responsable  pour  tout  dommage  conséquent,  direct,  indirect  ou  accidentel.  La  responsabilité  totale  de  FLIR  est  limitée  à  la  réparation ou  au  remplacement du  produit. La garantie définie ci‐dessus est inclusive et aucune autre garantie, écrite ou orale, n’est exprimée ou implicite.

 

Calibrage, réparation et services après‐vente 

FLIR  Systems,  Inc.  offre  des  services  de  calibrage  et  de  réparation  pour  les  produits  Extech 

Instruments que nous commercialisons. Nous fournissons également une certification NIST pour la  plupart  des  produits.  Contactez  notre  service  client  pour  toute  information  sur  les  services  de  calibrage disponibles pour ce produit. Un calibrage doit être effectué chaque année pour vérifier les  performances  et  la  précision  du  mètre.  Nous  offrons  également  une  assistance  technique  et  un  service à la clientèle. Veuillez vous reporter aux coordonnées fournies ci‐dessous. 

Lignes d’assistance: États‐Unis (877) 439‐8324; international: +1 (603) 324‐7800 

Service d’assistance technique : Option 3 ; E‐mail : [email protected] 

Réparations et retours : Option 4 ; E‐mail : [email protected] 

Les spécifications produit sont sujettes à modifications sans préavis. 

Pour les toutes dernières informations, veuillez visiter notre site Web.  www.extech.com 

FLIR Commercial Systems, Inc., 9 Townsend West, Nashua, NH 03063 USA 

Certifié ISO 9001 

Copyright © 2013‐2015 FLIR Systems, Inc. 

Tous droits réservés, y compris la reproduction partielle ou totale sous quelque forme que ce soit. 

www.extech.com 

14 EX210_EX210T-en-US_v1.5 9/15

Garantía 

FLIR Systems, Inc., garantiza este dispositivo marca Extech Instruments para estar libre de defectos en partes o mano de obra  durante un año a partir de la fecha de embarque (se aplica una garantía limitada de seis meses para cables y sensores). Si fuera  necesario regresar el instrumento para servicio durante o después del periodo de garantía, llame al Departamento de Servicio a 

Clientes  para  obtener  autorización.    Visite  www.extech.com  para  Información  de  contacto.  Se  debe  expedir  un  número  de 

Autorización de Devolución (AD) antes de regresar cualquier producto. El remitente es responsable de los gastos de embarque,  flete,  seguro  y  empaque  apropiado  para  prevenir  daños  en  tránsito.  Esta  garantía  no  se  aplica  a  defectos  resultantes  de  las  acciones  del  usuario  como  el  mal  uso,  alambrado  equivocado,  operación  fuera  de  las  especificaciones,  mantenimiento  o  reparación  inadecuada  o  modificación  no  autorizada.  FLIR  Systems,  Inc.,  rechaza  específicamente  cualesquier  garantías  implícitas  o  factibilidad  de  comercialización  o  idoneidad  para  cualquier  propósito  determinado  y  no  será  responsable  por  cualesquier daños directos, indirectos, incidentales o consecuentes. La responsabilidad total de FLIR está limitada a la reparación  o reemplazo del producto.  La garantía precedente es inclusiva y no hay otra garantía ya sea escrita u oral, expresa o implícita. 

Servicios de calibración, reparación y atención a clientes 

 

 

 

 

 

 

 

 

FLIR Systems, Inc., ofrece servicios de reparación y calibración para los productos que vendemos  de  Extech  Instruments.  Además  ofrecemos  certificación  NIST  para  la  mayoría  de  los  productos. 

Llame  al  Departamento  de  Servicio  al  Cliente  para  solicitar  información  de  calibración  para  este  producto. Para verificar el funcionamiento y precisión se debe realizar la calibración anual. Además 

 

 

 

 

 

  se provee Soporte Técnico y servicios generales al cliente, consulte la información de contacto en  seguida. 

Líneas de soporte: EE.UU. (877) 439‐8324; Internacional: +1 (603) 324‐7800 

Soporte Técnico Opción 3; correo electrónico: [email protected] 

Reparación / Devoluciones: Opción 4; correo electrónico: [email protected] 

Las especificaciones del producto están sujetas a cambios sin aviso 

Por favor visite nuestra página en Internet para la información más actualizada  www.extech.com 

FLIR Commercial Systems, Inc., 9 Townsend West, Nashua, NH 03063 USA 

Certificado ISO 9001 

 

 

Copyright © 2013‐2015 FLIR Systems, Inc. 

Reservados todos los derechos, incluyendo el derecho de reproducción total o parcial en cualquier medio

 

www.extech.com 

15 EX210_EX210T-en-US_v1.5 9/15