Agilent 2200 TapeStation DNA Analysis Technical Overview
Below you will find brief information for DNA Analysis 2200 TapeStation. The Agilent 2200 TapeStation system offers scalable throughput and automation, making it an ideal solution for quality control of biological samples in next-generation sequencing (NGS), microarray, and quantitative PCR workflows. The 2200 TapeStation system is straightforward to use; simply place the sample tubes and ScreenTape consumable in the instrument and let it load, separate, image, analyze, and present the results in approximately 1-2 minutes per sample.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
Good Measurement Practices for
DNA Analysis with the Agilent
2200 TapeStation System
Technical Overview
Introduction
The Agilent 2200 TapeStation system is a complete solution for automated, fast, and reliable electrophoresis. It is made up of three elements: the ScreenTape consumable, the 2200 TapeStation instrument, and the analysis software. The 2200 TapeStation system is straightforward to use; simply place the sample tubes and ScreenTape consumable in the instrument and let it load, separate, image, analyze, and present the results in approximately 1–2 minutes per sample.
The 2200 TapeStation system offers scalable throughput and rapid results, making it an ideal solution for quality control of biological samples in next-generation sequencing
(NGS), microarray, and quantitative PCR workfl ows. During assay preparation, the instructions regarding reagent preparation and instrument maintenance must be strictly followed. Important technical points are described in the TapeStation User Manual 1 and other supporting documentation. This Technical Overview describes techniques for ensuring reliable quantifi cation and sizing results using DNA assays on the
2200 TapeStation system.
Experimental
Materials
The 2200 TapeStation system
(G2964AA/G2965AA), D1000 ScreenTape and Reagents (5067-5582 and 5067-5583), and Genomic DNA ScreenTape and reagents (5067-5365 and 5067-5366), were obtained from Agilent Technologies
(Waldbronn, Germany). The IKA MS 3 basic vortexer with PCR plate adapter
(4674100) was purchased from
IKA GmbH & Co. KG (Staufen, Germany www.ika.com).
Quantifi cation
Acquiring DNA concentration data is essential for sample QC. To ensure accurate quantifi cation from the
2200 TapeStation system and its DNA
Assays (D1000, and High Sensitivity
D1000), it is important to note the following:
Sample mixing
In D1000 and High Sensitivity D1000 assays, the protocol states that sample and sample buffer are vortex-mixed using the IKA vortexer and adaptor at 2,000 rpm for 1 minute to ensure proper mixing.
After vortex mixing, it is recommended that the samples be collected at the bottom of the tube by brief centrifugation.
The effects of insuffi cient mixing were investigated using D1000 ScreenTape and reagents, and results are presented in
Figures 1 and 2, as well as Table 1.
It is clearly shown that poor mixing dramatically affects the reported sample concentration, and that following the recommended procedure is the best way to attain accurate quantifi cation results.
A
1,000
800
Vortex mixing
Pipette
Mixing
No mixing
600
400
200
0
120
100
80
60
40
20
Size (bp)
B
[bp]
1,500
1,000
700
500
300
100
25
Figure 1. A) shows the electropherogram and B) the gel image of the ScreenTape mixing tests. In both panels the green trace shows the recommended protocol of vortex mixing using the IKA vortexer and adaptor at 2,000 rpm for 1 minute followed by brief centrifugation. Blue shows results for pipette mixing only. Red shows the effect of no mixing. Images were taken from the Agilent TapeStation Analysis
Software.
0
Vortex mixing then spin down
Pipette mixing only No mixing
Figure 2. Chart of reported concentrations for the ScreenTape mixing tests, using the D1000 Assay.
Concentrations are expressed as a percentage of the theoretical for the three mixing methods. As above, the green bar represents the recommended protocol of vortex mixing using the IKA vortexer and adaptor at 2,000 rpm for 1 minute followed by brief centrifugation; blue - pipette mixing only; red - no mixing.
Table 1. ScreenTape Reagent mixing tests. Quantifi cation values obtained from the TapeStation Analysis
Software and the Agilent D1000 ScreenTape Assay when using the correct protocol (vortex mix followed by brief centrifugation) and two incorrect mixing protocols (pipette mixing only and no mixing) as illustrated in Figure 2.
Measured concentration
(ng/µL)
Vortex mixing then centrifugation 68.5
Pipette mixing only 21.2
No mixing 6.5
Theoretical concentration
(ng/µL)
70
70
70
2
Peak integration
For D1000 and High Sensitivity D1000
ScreenTape, the concentration values are calculated using the area of the sample peak compared to the known concentration of the top marker. It is therefore essential that all peaks are integrated correctly in the TapeStation
Analysis Software. Ensure that both the upper marker and sample peaks are properly integrated by manually adjusting the peak when necessary. Figure 3 shows examples of correct marker peak integration.
Figure 4 demonstrates the effect of peak integration on the quantitative results. Incorrect peak integration can signifi cantly bias the determined DNA sample concentration.
Magnification of the chart on the left
1
0
3
2
×10 3
6
5
4
Lower 258 Upper
Size (bp)
Figure 3. Correct upper marker integration for the 2200 TapeStation DNA Assays. Figure shows D1000
Assay.
2
1
0
4
3
A
×10 3
6
5
Lower
Correct peak integration concentration: 41.2 ng/µL
258 Upper
Size (bp)
2
1
4
3
0
B
×10
6
3
5
Lower
Incorrect peak integration concentration: 14.9 ng/µL
258 Upper
Figure 4. Example of correct (A) and incorrect (B) sample peak integration, and their effect on reported sample concentration.
Size (bp)
NOTE: NGS libraries should be quantifi ed using the region mode of the TapeStation Analysis software.
Please see the section on Peak maxima versus average molecular weight sizing for further details.
3
Use the correct protocol
Each ScreenTape type is designed for use with its corresponding Reagent kit. It is important that the correct reagents are used with the selected ScreenTape. It is also important that the correct sample preparation protocol 2,3 is followed exactly, using the correct volumes of sample and sample buffer as seen in the Figure 5.
Any variations in the volumes shown can adversely affect the quantifi cation results generated from the system.
In addition, it is important to choose the correct assay based on the concentration of the sample. Using sample concentrations outside the specifi ed quantitative ranges (as detailed in Table 2) will lead to inaccurate quantifi cation.
2200 TapeStation foil cover
(p/n 5067-5154) to prevent the sample from leaving the plate during vortexing.
After vortexing, use an appropriate centrifuge for either 96-well plates, or
8-way strips to ensure that all of the samples are at the bottom of the tube before placing in the 2200 TapeStation.
A Agilent D1000 ScreenTape assay
1 µL DNA or
Ladder
3 µL
Sample Buffer
CAUTION: To avoid damage to the
2200 TapeStation instrument and to ensure correct results, use only the recommended consumables and reagents with the 2200 TapeStation system as listed in the TapeStation
User Manual 1 . The use of 96-well plate covers from other suppliers could result in a failure to pierce the fi lm potentially leading to an instrument crash.
Use the correct tools for the job
Use pipettes that are calibrated and suffi cient for the volume to pipette. A
20-µL pipette will not be as accurate at pipetting 1–2 µL as a 2-µL or 10-µL pipette. Ensuring correct pipetting technique assists in ensuring that the volumes used in each assay are precise, and that the concentrations can be calculated correctly.
Use a vortexer designed for mixing 8-way tube strips, or 96-well plates. TapeStation systems are supplied with an optional IKA
MS3 vortexer, which includes a 96-well plate adaptor suitable for both 96-well plates and 8-way strips. This vortexer is recommended for use with the D1000 and High Sensitivity D1000 ScreenTape assays. For all other assays, please consult the assay Quick Guide for mixing recommendations.
It is recommended that TapeStation users obtain the IKA MS3 vortexer for best results. This vortexer can only be obtained directly from IKA (www.ika.com) by quoting the part number 4674100.
Agilent Technologies, Inc. will not sell these parts separately. If an IKA MS3 vortexer is not available, ensure thorough mixing by vortex for 10 seconds on maximum speed.
With a 96-well plate, use the
Vortex 1 minute
Spin down
B Agilent High Sensitivity D1000 and High Sensitivity D5000 ScreenTape assays
2 µL DNA or
Ladder
2 µL
Sample Buffer
Vortex 1 minute
Spin down
Figure 5. Agilent D1000 ScreenTape Assay (A) and High Sensitivity D1000 ScreenTape Assay (B) sample preparation protocols.
Table 2. Specifi ed quantitative ranges for the Agilent D1000 and High Sensitivity D1000 ScreenTape
Assays.
Quantitative range
Agilent D1000 ScreenTape 0.1 ng/µL–50 ng/µL
Agilent High Sensitivity D1000 ScreenTape 10 pg/µL–1,000 pg/µL
4
TapeStation analysis in the Agilent
SureSelect workfl ow
The 2200 TapeStation system has been verifi ed for use within the Agilent
SureSelect protocol. However, tests have shown that the purifi cation step, where DNA libraries are purifi ed using
AMPure XP beads, can have an effect on quantifi cation. It is postulated that carryover of the beads can cause issues, as the beads could be retained at the top of the separation matrix and contribute to the area of the upper marker (Figure 6).
This causes a lower relative reported value for sample concentration. The phenomenon can be identifi ed when there is a visible signal above the upper marker, as seen in Figure 6. A similar profi le can also occur when overamplifi cation of samples causes the signal to run concurrently with the upper marker.
This artifact can be avoided by increasing the time for which the samples are incubated on the magnetic plate to
10 minutes, thereby removing a higher percentage of the beads.
Sizing
Peak maxima versus average molecular weight sizing
The TapeStation analysis software contains options to display both an
Electropherogram view and a Region view. The Region view is designed for analyzing samples that appear as a smear (for example, NGS libraries) and gives slightly different information to that displayed in the electropherogram view (Figure 7). With this in mind, it is important to use the correct function for your samples.
200
150
100
50
Upper
designed for use with discrete peaks, and the default size reported is that of the highest point of the peak. whole smear or region, and reports size as that of the center of the regions’ mass. This gives the user an idea of the distribution of sizes within that sample.
0
Figure 6. Enlarged image of the upper marker showing additional signal from AMPure beads.
Lower
Electropherogram view
179 Upper Lower 70
Region view
700 Upper
Size (bp)
Size: 179 bp
Size (bp)
Average size: 232 bp
Figure 7. The sizing data obtained in Electropherogram and Region views of the Agilent TapeStation
Analysis Software.
5
Identifying the correct markers
It is important to ensure that the correct upper and lower markers are assigned in the TapeStation analysis software. The markers are used within the software as internal references in order to determine the molecular weight of each sample peak. Incorrect identifi cation can lead to miscalculations in the reported sizing values.
A
Figure 8 shows an example of incorrect lower marker identifi cation next to the corrected fi le. In each of the images, lanes 2 and 3 are the same sample at different concentrations. As seen in
Figure 8A, incorrect identifi cation of the lower marker has caused misalignment of the sample; therefore all the fragments are reported with incorrect sizes.
Manually assigning the correct lower marker (Figure 8B) provides accurate sizing information.
B
Flicking the ScreenTape
Due to the nature of the ScreenTape consumables, bubbles can form in the buffer chamber. If bubbles form at the gel/buffer interface, a loss of performance can be observed in that lane.
It is important therefore always to ‘fl ick’ the ScreenTape before placing it into the
2200 TapeStation instrument to move the bubble to the top of the chamber where it will no longer affect sample loading
(Figure 9).
X Incorrectly assigned as the lower marker Correctly assigned as the lower marker
Figure 8. Screenshots of the Agilent TapeStation Analysis Software, showing A) incorrect lower marker identifi cation in the middle lane of the gel image and
B) corrected lower marker in the middle lane. The lower marker is always highlighted with green in the gel image.
X
Figure 9. Flicking the ScreenTape consumable removes bubbles from the gel interface.
6
Bubbles at the gel interface can affect the performance of the run since the bands appear smudged or diagonal (Figure 10).
In turn, this can affect how the bands are identifi ed in the TapeStation Analysis
Software.
[bp]
A1 (L) A2
X
1,500
1,000
700
500
300
B2
X
100
25
Figure 10. The effect of bubbles at the gel interface on the DNA separation profi le
(lanes A2 and B2).
Molarity
In the TapeStation Analysis Software, molarity is determined from both size and quantity. Ensure that good measurement practices for sizing and quantifi cation have been followed to ensure accurate molarity values.
Genomic DNA Assay
When using the Genomic DNA Assay, it is important to note the following recommendations to ensure correct sizing.
Equilibrate reagents to room temperature
Genomic DNA Reagents must be equilibrated to room temperature for
30 minutes before use. Failure to do so can affect sizing results as cold reagents will overestimate the size of genomic
DNA samples (Table 3).
Table 3. The effect of room temperature (RT) equilibrated Genomic DNA reagents as well as cold reagents on the sizing accuracy of the Genomic DNA ScreenTape assay.
MW
Accuracy
Expected
17,000 bp
–
Reagents at RT
18,867 bp
+ 11 %
Reagents at 4 °C
24,369 bp
+ 43 %
Always use fresh genomic
DNA ladder
When using the Genomic DNA
ScreenTape Assay, it is important that the
Genomic DNA Ladder is freshly prepared for each run. For tube strips or 96-well plate, the ladder must always be the fi rst well selected to run. Failure to do so will affect the sizing results obtained from the assay.
It is also important to note that there is no Genomic DNA software saved ladder in the TapeStation Analysis Software.
The effect of shaking the genomic
DNA ladder vial
The Genomic DNA Ladder must only be vortex mixed as shaking the ladder vial can degrade the top fragment (Figure 11).
Therefore, before pipetting the ladder into the TapeStation approved tube strips or
96-well plate, the ladder vial should be handled carefully, then gently vortexed for 5 seconds to maintain accurate performance.
To minimize shaking during transit, the
Genomic DNA reagents are shipped frozen, on dry ice. Once received, these should be kept at 2–8 °C in the refrigerator.
700
600
500
400
300
200
100
0
A
Lo w er
250 40
0
A1:Ladder
600 900 1,
20
0
1,
50
0
2,
00
0 0
2,50 3,
00
0
4,
00
0
7,
00
0
15
,0
00
48
,5
00
Size (bp)
48,500
15,000
7,000
4,000
3,000
2,500
2,000
1,500
1,200
900
600
400
250
100 bp A1 (L)
600
500
400
300
200
100
0
B
Lo w er
250 40
0
A1:Ladder
600 900 1,
20
0
1,
50
0
2,
00
0 0
2,50 3,
00
0
4,
00
0
7,
00
0
15
,0
00 bp A1 (L)
!
Size (bp)
15,000
7,000
4,000
3,000
2,500
2,000
1,500
1,200
900
600
400
250
100
Figure 11. The effect of shaking the Genomic DNA Ladder. A) Genomic DNA Ladder has been vortex mixed for 5 seconds prior to analysis on the Genomic DNA
ScreenTape assay. B) the Genomic DNA Ladder vial was shaken by manually inverting the tube 30 times. Degradation of the top fragment (48,500 bp) is clearly shown.
7
Degradation of the Genomic DNA Ladder as shown in Figure 11, can result in inaccurate sizing results. In instances where the degradation has resulted in failure of the software to assign the top ladder fragment, no DIN score will be presented for the lanes in this fi le. Where the ladder failure has resulted in a loss of data, a software alert will be presented in the sample table and above the gel lane.
Conclusion
The Agilent 2200 TapeStation system provides scalable throughput and automation, making it an ideal solution for quality control of biological samples in next-generation sequencing (NGS), microarray, and quantitative PCR workfl ows.
Following the good measurement practices described in this Technical
Overview ensures reliable and accurate
DNA quantifi cation and sizing.
References
1. Agilent 2200 TapeStation User Manual,
Agilent Technologies, publication number G2964-90003, 2013.
2. Agilent D1000 ScreenTape System
Quick Guide, Agilent Technologies, publication number G2964-90032,
2013.
3. Agilent High Sensitivity D1000
ScreenTape System Quick Guide,
Agilent Technologies, publication number G2964-90131, 2013.
4. Agilent Genomic DNA ScreenTape
System Quick Guide, Agilent
Technologies, publication number
G2964-90040, 2012.
5. Agilent D5000 ScreenTape System
Quick Guide, Agilent Technologies, publication number G2964-90050,
2015.
6. Agilent High Sensitivity D5000
ScreenTape System Quick Guide,
Agilent Technologies, publication number G2964-90150, 2015.
www.agilent.com/genomics/ tapestation
For Research Use Only.
Not for use in diagnostic procedures.
This information is subject to change without notice.
© Agilent Technologies, Inc., 2015
Published in the USA, September 1, 2015
5991-5153EN
advertisement
Key Features
- Scalable throughput
- Automated DNA analysis
- Fast and reliable electrophoresis
- Quality control for NGS, microarray, and qPCR samples
- Easy to use
- Rapid results
- Detailed technical guide