MEGADYNE™ Smoke Evacuator
Optimized Device Performance Guide
Before you begin
1.
Plug in power cord: Insert power
plug into the receptacle on the
MEGADYNE™ Smoke Evacuator.
2. Recommended: Plug in connect
cable. Insert the male connector of
the connect cable into the
receptacle on the MEGADYNE™
Smoke Evacuator and the female
connector into the receptacle on the
electrosurgical generator.
Generator
Smoke
evacuator
Connect
cable
3. A
lternate: Connect RF sensor to monopolar cord. When connecting smoke
evacuator and electrosurgical generator via connect cable is not possible:
a. Insert RF sensor (sold separately)
into the RF sensor port on
MEGADYNE™ Smoke Evacuator.
b. Insert
monopolar pencil cord or
return electrode cord into the
receptacle on RF sensor.
Before you begin (cont.)
4. Insert the filter securely into the
filter port located on the front of the
MEGADYNE™ Smoke Evacuator.
5. Turn
on the machine: Press the
power switch button to turn on the
smoke evacuator.
6. Check the FILTER life indicator:
a. G
reen
Ready for use:
100% - 25% filter life.
b. Y
ellow
Ready for use:
25% - 0% filter life.
If the filter is not securely seated,
all filter lights will blink until proper
installation is completed.
c. R
ed
Once the red light is illuminated,
the smoke evacuator can only be
activated within a 3 hour window
to complete the case.
One indicator light will be
illuminated when the filter is secure
and indicate remaining filter life.
Using in open procedures
1.Press the OPEN button to switch to
OPEN mode.
2.Adjust FLOW settings to userpreferred flow-rate levels.
118LPM
3. Adjust RUN TIME settings to
user-preferred time-delay intervals.
30 sec.
106LPM
15 sec.
94LPM
10 sec.
82LPM
6 sec.
71LPM
3 sec.
RUN TIME options control how
long flow continues after the
energy device is deactivated when
utilizing one of the automatic
activation methods. This feature
enables evacuation of residual
smoke after the surgeon discontinues activating the energy device.
4. Ensure smoke evacuation tube
is connected to MEGADYNE™
Smoke Evacuator.
Using in LAP procedures
1.Press the LAP button to switch to
LAP mode.
2.Adjust FLOW settings to userpreferred flow-rate levels.
41LPM
3. Adjust RUN TIME settings to
user-preferred time-delay intervals.
4. Ensure smoke evacuation tube
is connected to MEGADYNE™
Smoke Evacuator.
10 sec.
29LPM
8 sec.
18LPM
6 sec.
11LPM
4 sec.
5LPM
2 sec.
RUN TIME options control how
long flow continues after the
energy device is deactivated when
utilizing one of the automatic
activation methods. This feature
enables evacuation of residual
smoke after the surgeon discontinues activating the energy device.
Manual activation of smoke evacuation
If preferred, press the MANUAL
START/STOP button to activate
smoke evacuation in OPEN or LAP
mode. Smoke evacuation remains
activated until the MANUAL
START/STOP button is pressed again.
Please always refer to the Instructions for Use / Package Insert that come
with the device for the most current and complete instructions
Johnson & Johnson Medical Limited
Baird House, 4 Lower Gilmore Bank,
Edinburgh, EH3 9QP, UK
www.jnjmedicaldevices.com
© Johnson & Johnson Medical Limited 2020, 154866-201002 UK
">