DL429
True RMS Digital Clamp-On
Wireless with Dual Temperature
INSTRUCTION MANUAL
ENGLISH
DL429
μA
Temp
/
°F°C
NCV
LINK
MAX/
MIN
LRA Inrush
V
Hz/D uty
MF
D
True RMS
RANGE
BT RF
INRUSH
T
1
-T
2
T
1
T
2
CATIII
600V
MAX
30V
TEMP
1-800-547-5740 • Fax: (503) 643-6322 www.ueitest.com • email: [email protected]
WARRANTY
The DL429 is warranted to be free from defects in materials and workmanship for a period of two years from the date of purchase.
For more information on warranty and service: www.ueitest.com • Email: [email protected]
1-800-547-5740 • FAX: (503) 643-6322
This warranty gives you specific legal rights. You may also have other rights which vary from state to state.
FCC INFORMATION
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
CAUTION : Changes or modifications not expressly approved by the manufacturer responsible for compliance could void the user’s authority to operate the equipment.
BATTERY REPLACEMENT
• When indicator is displayed on the LCD, batteries must be replaced.
• Remove the back screw and replace 6 x AAA batteries.
CLEANING
Turn instrument off and disconnect test leads. Clean the instrument by using a damp cloth. Do not use abrasive cleaners or solvents.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period of time. Do not expose to high temperatures or humidity. After a period of storage in extreme conditions exceeding the limits mentioned in the
Specifications section, allow the instrument to return to normal operating conditions before using it.
DISPOSAL / RECYCLE
Caution : This symbol indicates that equipment and its accessories shall be subject to a separate collection and correct disposal.
Copyright © 2015 UEi. All Rights Reserved.
iPhone ® and iPad ® are trademarks of Apple, Inc., registered in the U.S. and other countries. Android™ is a trademark of Google Inc. Other third party trademarks referenced are the property of their respective owners.
FEATURES
• True RMS
• Direct function access with push-button operation
• Dual Display
• Dual Temperature with Differential
• Locked Rotor Inrush Measurement
• Non-contact voltage
• Min / Max / Hold
• Backlit display with Clamp Work-light
• Magnetic mount for hands free use
• RF or Bluetooth ® Wireless : BT app available for Android and iOS
• Accepts CH3 : The Hook, or Fieldpiece accessory heads
• Input Jack Safety Lockout
• Auto-ranging measurement with manual ranging capability
• Auto Power Off : After 30 minutes of non-use.
• Low battery lockout : Displays “BATT” in display blocking potentially
inaccurate readings.
• Visible High-Voltage alert
MEASURES
• Amps AC to 600A
• µA DC 0.1µA~ 2000µA
• Volts DC 0.1mV ~ 1,000V
• Volts AC 0.1mV ~ 750V
• Resistance 0.1Ω ~ 60MΩ
• Capacitance 0.01nF ~ 2,000µF
• Temperature -22° ~ 752°F (-30° ~ 400°C) with Differential
• Frequency 0.01Hz ~ 99.99kHz
• Duty Cycle
GENERAL SPECIFICATIONS
• Altitude: Operating - up to 2000m (6,561 ft.)
Storage - 10,000m (32,808 ft.)
• Humidity: 80% max
• Operating Temperature: 32°F to 122°F (0°C to 50°C) at < 75% R.H
• Storage Temperature: -4°F to 140°F (-20°C to 60°C) at < 80% R.H
• Relative humidity: 0% to 80% at 32°F to 95°F (0°C to 35°C),
0% to 70% at 32°F to 131°F (0°C to 55°C)
• Temperature Coefficient: Nominal 0.1 x (Specified accuracy) / °C
(<18°C or >28°C ; <64°F or >82°F)
• Pollution degree: 2
• Display: 3-5/6 digits 6000 counts dual LCD display(s)
• Refresh Rate: 3 times/sec
• Overrange: "OL" is displayed
• Polarity: Automatic(no indication for positive polarity) ; Minus(-) sign for negative polarity
• Dimensions: 10.2" x 2.5" x 1.5"
• Weight: 15.2oz.
• Calibration: Accurate for one year
• CAT Rating: CAT III 600V
• Certifications: ETL & cETL IEC61010-2-032
• Battery type: 6 x 1.5V AAA or LR03
• Silicon Test Lead: IEC61010-2-031
• Accuracy: ± (% of reading + # of least significant digits)
WARNINGS
To ensure safe operation and service of the tester, follow these instructions.
Failure to observe these warnings can result in severe injury or death.
• Before each use, verify meter operation by measuring a known voltage or current.
• Never use the meter on a circuit with voltages that exceed the category based rating of this meter.
• Do not use the meter during electrical storms, or in wet weather.
• Do not use the meter or test leads if they appear to be damaged.
• Ensure meter leads are fully seated, and keep fingers away from the metal probe contacts when making measurements.
• Do not open the meter to replace batteries while the probes are connected.
• Use caution when working with voltages above 60V DC, or 25V AC RMS.
Such voltages pose a shock hazard.
• To avoid false readings that can lead to electrical shock, replace batteries if a low battery indicator appears.
• Unless measuring voltage or current, shut off and lock out power before measuring resistance or capacitance.
• Always adhere to local and national safety codes. Use Personal
Protective Equipment (PPE) to prevent shock and arc blast injury.
SYMBOLS USED ON LCD
~
AC Measurement
Negative DC
O.L.
Overload: Range Exceeded
Low Battery
DC Measurement
AT
Auto Ranging
Apo
Auto Power-Off Active
HOLD
Display not updating
MIN
Minimum measured value displayed
%
Units for duty cycle
V
Voltage measurement mode
MAX
A
Maximum measured value displayed
Units for frequency
Amperage measurement mode
Ω
Resistance measurement mode
Diode test mode
MFD
Capacitance measurement mode in nano farads or microfarads
M
mega (x 10
m
milli (x 10 -3
VALUE MULTIPLIER
6 or 1,000,000)
or 0.001)
µ
micro (x 10 -6 or 0.000001)
k
Kilo prefix (x 10 3 or 1,000)
n
nano (x 10 -9 or 0.000000001)
INTERNATIONAL SYMBOLS
~
AC Alternating Current Warning or Caution
DC Direct Current Dangerous levels
DC/AC Voltage or Current
Ground
Double Insulated Class II
Safe for disconnect from live conductors
NAVIGATION
DL429
μA
Temp
/
°F°C
NCV
LINK
MAX/
MIN
LRA Inrush
V
Hz/D uty
MF
D
RANGE
True RMS
• Press briefly to turn the meter on
to disable auto power off.
INRUSH
• Press and hold to turn the meter off
• The new user interface allows direct access from any mode.
• Wireless app and meter display update simultaneously.
• From wireless app press '
T
1
T
2
CATIII
600V
MAX
30V
HOLD
DL429
μA
Temp
/ °F°C
NCV
LINK
MAX/
MIN
LRA Inrush
V
Hz/D uty
MF
D
True RMS
RANGE
BT RF
INRUSH
T
1
-T
2
T
1
T
2
CATIII
600V
MAX
30V
TEMP
• Press to select DC µA measurement mode.
• Press to select T1.
• Press a second time for T2.
• Press a third time for T1 - T2.
• Press and hold to change from °F to °C.
MF
D
• Press to select continuity.
• Press a second time for resistance.
• Press a third time for capacitance and a fourth time for diode.
• Press to select VAC or VDC
In VAC mode
• Press and hold for Hz
• Press and hold again for Duty Cycle.
NCV
• Non-Contact Voltage Detection key is used to detect power with a sensor located at the tip of the clamp head and indicates positive response with an audible alarm and visual LED indicator light just above the “NCV” button.
• Do not use non-contact voltage detector to determine if there is current in the wire. Detection operation could be affected by socket design, insulation thickness, type and other factors.
• Voltage indicator light may also light when voltage is present on the meter’s input jack or from external interference sources such as motors, flashlights etc.
• From wireless app press ‘ ’ first
MAX/
MIN
LRA Inrush
RANGE
• Press to enter Max / Min mode; the largest and smallest values will be saved while in this mode.
• Press repeatedly to alternate between the maximum and minimum readings.
• Press for 2 seconds to return to live reading and clear the stored maximum and minimum values.
Note: Select range prior to selecting Min/Max to capture large values
The UEi LRA Inrush is programmed to properly capture the starting current for compressor motors.
• Select AC Amps
• Select the range capable of capturing the maximum value
• Press and hold LRA Inrush for two seconds – INRUSH will now be shown on the screen
• Activate compressor and read value on the display
• Press and hold LRA Inrush for two seconds to exit
• Press repeatedly to cycle through manual ranges.
• Press for 2 seconds to return to auto ranging mode.
• AT is displayed on LCD only during auto ranging mode.
Note: Select range prior to Min/Max for best results.
LINK
HOLD
x1 x1
• Press to activate Bluetooth®
• Press again to activate RF wireless
◊ Press and hold to turn radio off
• The RF will work with specific transmitter, while the BT will work with a tablet or smartphone with app installed
• Press to hold the reading on the display. Press again to return to live reading.
• Press and hold "work light" key for 2 seconds to turn on. Press and hold again to turn off.
10097 02/15
AC Amps: < 600A
MAX/
MIN
V V
Hz/D uty
Temp
/ °F°C / °F°C
MF
D D
Single
Conductor
Only
T
2
CATIII
600V
When the meter is powered on, the upper display will always show amperage, or adapter (ADP) output
• Center wire in guides for best accuracy.
• Opposing currents cancel (use line-splitter when necessary).
Keep hands below line when measuring high current levels.
Features:
NCV
LINK
MAX/
MIN
LRA Inrush
RANGE HOLD
AC Amps Measurement - Jaw input
Range
60A
600A
Resolution
0.01A
0.1A
Accuracy
± (2.9% + 15 digits)
± (1.9% + 8 digits)
True RMS: 45Hz to 400Hz
* Minimum Current for Clamp Measurement: 0.2A
Overload Protection
600V RMS
DC Low Amps: < 2000uA
DL429
μA
Temp
/ °F°C
NCV
LINK
MAX/
MIN
LRA Inrush
V
Hz/D uty
MF
D
True RMS
RANGE
BT RF
INRUSH
T
2
T
1
T
2
CATIII
600V
MAX
30V
TEMP
High
Voltage
Alert
DL429
μA
Temp
/
°F°C
NCV
LINK
MAX/
MIN
LRA Inrush
V
Hz/D uty
MF
D
True RMS
RANGE
BT RF
INRUSH
T
1
-T
2
T
1
T
2
CATIII
600V
MAX
30V
TEMP x1 =
• Press for DC micro amps
Features:
NCV
LINK
MAX/
MIN
LRA Inrush
RANGE
DC Low Amps Measurement -Test lead input
HOLD
Range
600µA
2000µA
Resolution
0.1µA
1µA
Accuracy
± (1.2% + 3 digits)
Overload Protection
2000µA / 600V RMS
Hz/Duty
V
Hz/D uty
x1 =
Press and hold =
Hz/Duty Cycle x2 =
Features:
NCV
LINK
MAX/
MIN
LRA Inrush
DC Voltage Measurement
Range
600mV
6V
60V
600V
1000V
Resolution
0.1mV
1mV
10mV
100mV
1V
AC Voltage Measurement
Range
600mV
6V
60V
600V
750V
Resolution
0.1mV
1mV
10mV
100mV
1V
True RMS: 45Hz to 400Hz
Accuracy
± (0.5% + 4 digits)
± (0.8% + 10 digits)
Accuracy
RANGE
± (2.0% + 5 digits)
HOLD
Overload Protection
1000V
Overload Protection
750V RMS
Frequency Measurement
Range Resolution
99.99Hz
999.9Hz
0.01Hz
0.1Hz
9.999kHz
1Hz
99.99kHz
10Hz
Sensitivity: 1.8V RMS
Duty Cycle Measurement
Range
1.0 ˜ 99.0%
Accuracy
± (0.1% + 4 digits)
Accuracy
±(0.2% per kHz + 0.1% + 5 digits)
Overload Protection
600V RMS
Overload Protection
600V RMS
/ / MFD /
MF
D
Audible Continuity
• Press for continuity
• Display shows resistance.
• Buzzer sounds if less than 40Ω.
Features:
NCV
LINK
MAX/
MIN
LINK
HOLD
DL429
μA
Temp
°F°C
NCV
LINK
MAX/
MIN
V
Hz/D uty
True RMS
RANGE
BT RF
INRUSH
T -T
2
T
1
T
2
CATIII
600V
MAX
30V
TEMP
Overload Protection
600V RMS
Open Circuit Voltage
< 0.44V
Threshold Approx : < 40Ω
Resistance
• Press again for resistance
• Do not measure resistance on a live circuit.
Features:
NCV
LINK
MAX/
MIN
LRA Inrush
RANGE HOLD
Range
600 Ω
6k Ω
60k Ω
600k Ω
6M Ω
60M Ω
Resolution
0.1
Ω
1 Ω
10 Ω
100 Ω
1k Ω
10k Ω
Accuracy
± (1.0% + 4 digits)
± (2.0% + 4 digits)
Capacitance
• Press again for capacitance
Features:
NCV
LINK
HOLD x1 = x2 = x3 =
MFD
x4 =
Overload Protection
600V RMS
Range
60.00
nF
600.0
nF
6.000
μF
60.00
μF
600.0
μF
2000 μF
Resolution
0.01
nF
0.1
nF
0.001
μF
0.01
μF
0.1
μF
1 μF
Diode
• Press again for diode test
GOOD DIODE
Accuracy
± (3.5% + 6 digits)
Overload Protection
600V RMS
BAD DIODE
Open Diode
Displays "OL"
Both directions
• Forward voltage drop if forward biased.
• “O.L.” if reverse biased.
Features:
NCV
LINK
MAX/
MIN HOLD or
'0' Both directions
(Shorted)
Diode Test
Range Open Circuit Voltage Test Current (Typical) Overload Protection
2.0V
< 1.6V DC 0.25mA
600V RMS
TEMP/ °F °C°
• Press for T1
• Press again for T2
• Press again for T1-T2
• Press and hold to change scale
Features:
NCV
LINK
MAX/
MIN HOLD
DL429
μA
Temp
°F°C
NCV
LINK
MAX/
MIN
LRA Inrush
V
Hz/D uty
True RMS
RANGE
BT RF
INRUSH
T
1
-T
2
T
1
T
2
CATIII
600V
MAX
30V
TEMP
Temp
/ °F°C
Range
-22°F ~ 14°F
(-30°C ~ -10°C)
15°F ~ 752°F
(-9°C ~ 400°C)
Resolution
0.1°F
(0.1°C)
0.1°F
(0.1°C)
Accuracy
± (1.0% + 5.4°F)
± (1.0% + 3.0°C)
± (1.0% + 3.6°F)
± (1.0% + 2.0°C)
Overload Protection
30V RMS
Sensor: K-Type Thermocouple, sensor accuracy not included
Connecting and Using the App
• Search for App as, “UEi Wireless”
• Compatible with iPhone ® 4X and up running iOS ® 7 or higher, Galaxy S4 ® ,
Nexus 5™, HTC One ® running Android™ 4.4 or higher.
• To install or search on iPad ® use "iPhone ® only" to find App.
• Press "Link" button on meter to activate wireless "BT"
• Open app. Meter will connect automatically.
Menu
• Press “ ” to connect, disconnect, and
access settings.
Settings
• General settings adjust button sound, vibrate
and refresh rate.
Recording settings
• Continuous reading
• Number of samples
• Sampling interval
Record
• Press to start, stop.
• The number of samples will show in real time
Logs
• Press to view
recorded data.
• Press the entry you wish
to view (yyyy-mm-dd
hh:mm:ss)
• Functions are noted
underneath respectively
AMP-AMP (TOP-
BOTTOM) Display
• Press “ “ button for summary
• Press “ “ button for sample data
• Press “
formats
“ button to export data via email in (.CSV .PNG or .JPG)
Use CAT III rated leads or higher.
Do not attempt to measure more than 1000V DC or 750V AC.
• Press for AC Volts
Note: This is the default setting when first powered on.
◊ Press and hold in AC volts for Hz/Duty Cycle
• Press again for DC volts
• High Volt Visible Alert - Any input exceeding 30V (AC or DC) will light
the NVC LED to alert users to potentially dangerous voltage levels
Reverse Bias
Displays "OL"
Forward Bias
Displays approx. voltage drop
DL429
μA
Temp
/ °F°C
NCV
LINK
MAX/
MIN
LRA Inrush
V
Hz/D uty
MF
D
True RMS
RANGE
BT RF
INRUSH
T
1
-T
2
T
1
T
2
CATIII
600V
MAX
30V
TEMP
MF
D x1 = x2 = x3 =
MFD
x4 =
Graph
• Press “ Graph ” to view trending data
in real time during measurement.
Graph