Below you will find brief information for Shielding Effectiveness Measurement EMC32 K48. This document explains how to use the EMC32-K48 option of the R&S EMC32 software to measure shielding effectiveness. It covers setting up instrumentation, hardware, and test templates for measuring shielding effectiveness, as well as running tests and printing reports.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
Shielding Effectiveness
Measurement
Application Note
Products:
ı
R&S
®
EMC32-K48
This application note shows all necessary settings for measuring shielding effectiveness using the R&S
®
EMC32-K48 option via R&S
®
EMC32
Measurement Software.
Note:
Please find the most up-to-date document on our homepage http://www.rohde-schwarz.com/appnote/
Table of Contents
Table of Contents
5 Test Template Configurations ......................................................... 10
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
2
Overview
1 Overview
This document describes the functionalities for R&S
®
EMC32-K48 option in
R&S
®
EMC32 platform which have to be done to support the shielding effectiveness test method.
The R&S
®
EMC32 software offers the following applications:
▪
▪
▪
▪
▪
Provide control for instruments (RF generator, amplifier, switch units, spectrum analyzer, network analyzer)
Perform reference level testing of system and measurement protocol as recommended by test standard
Perform EUT Test and Measurement automatically
Evaluate and display real-time value of the measurement
Generate report
The R&S
®
EMC32-K48 option requires R&S
®
EMC32-S Main Option (EMS Scan
Template) and R&S
®
EMC32-K4 option (EMS Auto Test).
Multi-user licensee should purchase R&S
®
EMC32MK48 option.
▪
▪
▪
▪
▪
▪
▪
▪
▪
▪
▪
The following abbreviation are used in the following text:
R&S
®
EMC32 software is referred to as EMC32
R&S
®
EMC32-S software option is referred to as EMC32-S
R&S
®
EMC32-K4 software option is referred to as EMC32-K4
R&S
®
EMC32-K48 software option is referred to as EMC32-K48
R&S
®
EMC32MK48 software option is referred to as EMC32MK48
Shielding Effectiveness is referred to as SE
Equipment under test is referred to as EUT
Radio frequency is referred to as RF
Electromagnetic interference is referred to as EMI
Electromagnetic susceptibility is referred to as EMS
R&S
®
refers to Rohde & Schwarz GmbH & Co. KG
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
3
Introduction
2 Introduction
Refer to the general block diagram below on the setup for SE system according to
IEEE STD 299, EN 50147-1 and MIL-STD-188-125-1. EMC32 software is used.
1.00
Fig. 2-1: Typical SE system
The system above consists of the following:
ı
ı
ı
RF generator as RF signal source generation
Amplifiers to magnify signal to increase system dynamic range
Antenna sets for transmitting and receiving
ı
Spectrum analyzer measuring the received level at a given level of signal source generated
ı
Switching unit which can be used to switch to different amplifiers of different frequency range capabilities
ı
Network analyzer for SE measurement of an enclosure
Rohde & Schwarz Shielding Effectiveness Measurement
4
EMC32 Setup
3 EMC32 Setup
Follow the instructions below to setup EMC32. The steps are:
ı
ı
ı
Installation of software and drivers iKey requirements
Online help
3.1 Installation of Software and Drivers
This test is programmed to work with EMC32 version 9.20 and above. Follow the installation procedures below:
1. Install National Instruments GPIB driver with NI-VISA.
2. Install EMC32 version 9.20 or higher. It is important to check all options for EMI and EMS.
3. Install iKey application
4. Install VISA and drivers for relevant R&S device (e.g. SMC100A, SMF100A, NRP-
USB, and FSP)
The software can now be launched.
3.1.1 EMC32 on the Web
Do check for the latest version of EMC32 via the help menu. In the main toolbar, select
“?” and click on “EMC32 on the Web”. Alternatively, you may also find the latest update info at www.emc32.rohde-schwarz.com
.
1.00
Fig. 3-1: EMC32 on the web
Rohde & Schwarz Shielding Effectiveness Measurement
5
EMC32 Setup
3.1.2 Update Manager
The EMC32 integrated update manager will automatically prompt whenever there is a new service patch or version update. You can either enable or disable this update manager via the help menu by selecting “?” and clicking on “Update Manager” for its settings.
Fig. 3-2: Update manager
3.2 iKey Requirements
EMC32 uses a physical USB dongle referred to as iKey to run test simulations and control real equipment. Without the iKey, the EMC32 software can only run test simulations.
The required iKeys options for SE are EMC32-S (EMS Scan template), EMC32-K4
(EMS Auto-Test) and EMC32-K48 (Shielding Effectiveness Test). EMC32 allows merging of several options onto one iKey using EMC32 iKey Merge Tool from the EMC program group. You may refer to the EMC32 installation manual chapter 9 for more information on using the iKey Merge Tool.
3.3 Online Help
Online help is available on the CD and on the software after installation. Help can be accessed at any time via the main toolbar by selecting "?" or by pressing the F1 key.
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
6
Test Configuration
4 Test Configuration
ı
ı
Before performing SE measurements in EMC32, setup the test configuration as described in the following sections:
Instrumentation
Hardware Setup
4.1 Instrumentation
Refer to chapter 3.3 "Online Help" on page 6. In HTML Help, select the Index tab, search for "Device list" to show a detailed description on setting up the instruments.
Fig. 4-1: Online help for device list
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
7
Test Configuration
The EMC32 software supports a wide range of spectrum analyzer models, antenna mast controller, amplifiers and their interlock, and OSP switch units.
Fig. 4-2: Device list dialog box
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
8
Test Configuration
4.2 Hardware Setup
Refer to chapter 3.3 "Online Help" on page 6. In HTML Help, select the Index tab, search for "Hardware Setup" to show a detailed description on setting up the hardware.
Fig. 4-3: Online help for hardware setup
Hardware setup can be configured for splitting into different frequency subranges to suit different antennas, amplifiers and generator models. It is recommended to conduct splitting according to the antenna subranges.
A typical SE system setup consist of a generator, power amplifier, power sensors, transmit antenna, receive antenna, switching unit and receiver (see chapter 2.
"Introductions" on page 4). This is the same setup used in EMC32-S (Susceptibility) hardware setup for SE test.
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
9
Test Template Configurations
5 Test Template Configurations
ı
ı
ı
ı
Before performing SE measurements in EMC32, setup the test template configuration as described in the following sections:
Reference Level Test Template
EUT Test Template
EUT Monitoring Test Template
EUT Auto Test Template
5.1 Reference Level Test Templates
Refer to chapter 3.3 "Online Help" on page 6. In HTML Help, select the Index tab, search for "EMS test new" to show a detailed description on setting up the reference level test template.
1.00
Fig. 5-1: Online help for Reference level test template
Reference level test template, known as reference calibration test template in EMC32, is required to calibrate a known level at the output of the transmit cable, and to save the result to a reference calibration table.
The purpose of reference level test for SE is to set a reference for generator level to achieve SE maximum RF level which is known as reference calibration in EMC32.
Rohde & Schwarz Shielding Effectiveness Measurement
10
Test Template Configurations
This template is used for the first calibration without EUT. The test template is configured with several sub-ranges according to the different antennas and also the antenna sub-ranges.
In EMC32, reference level test is created in EMS scan test template in order to select the correct hardware setup and run as reference calibration test method to perform reference level testing.
5.1.1 Reference Level Test Template for SE
This chapter includes some of the parameters that is needed for the reference level test templates (known as reference calibration in EMC32) to perform the SE test.
For SE reference level measurement, SE test standard will be created in the EMC Test
Standard dropdown box. EMC32-K48 option will be needed to activate the SE test standard
In Fig. 5-2, select General Settings tab. Under EMC Test Standard, select "SE REF
CAL" from the dropdown list.
1.00
Fig. 5-2: Test standard selection for SE
Rohde & Schwarz Shielding Effectiveness Measurement
11
Test Template Configurations
In Fig. 5-3, select Frequency tab. In the area Start Frequency and Stop Frequency, enter the appropriate antenna frequency subranges.
Fig. 5-3: Antenna frequency subranges setting
In Fig. 5-4, select Leveling Mode Tab. Under the section Common Ref. Cal. File
Name, enter desired filename in the field below.
1.00
Fig. 5-4: Filename entry
Rohde & Schwarz Shielding Effectiveness Measurement
12
Test Template Configurations
From Fig. 5-5, select Device Setups tab. Click "Span" to bring up the dialog box as shown in Fig. 4-4. Set Device Mode to "Receiver" to activate zero span measurement and Detector to "Average".
Fig. 5-5: Span dialog box
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
13
Test Template Configurations
5.2 EUT Test Template
Refer to chapter 3.3 "Online Help" on page 6. In HTML Help, select the Index tab, search for "EMS test new" to show a detailed description on setting up the EUT test template.
Fig. 5-6: Online help for creating a new EMS test
This template is used for measurement on the actual EUT. It sets the output transmit level according to previous reference level results and measure the difference in the output level from within the EUT. This difference is the shielding effectiveness.
The hardware setup should be preset accordingly (see chapter 4.2 "Hardware Setup on page 9) before the EUT test sequence can be created in the EMS scan test template. The EUT test carries out the antenna coupling test using the substitution method, based on the saved reference calibration table.
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
14
Test Template Configurations
5.2.1 EUT Test Template for SE
For SE test configuration, SE test standard will be created in the EMC Test Standard dropdown box. EMC32-K48 option will be needed to activate the SE test standard
In Fig. 5-7, select General Settings tab. Under EMC Test Standard, select "SE
TEST" from the dropdown list.
1.00
Fig. 5-7: SE test standard selection
For SE requirement, (nominal) immunity level in test measurement should follow its reference level table to maintain a nearly fixed generator level output for both horizontal and vertical polarization.
Rohde & Schwarz Shielding Effectiveness Measurement
15
Test Template Configurations
In Fig. 5-8, select Leveling Mode tab, input any reference calibration table created in
SE REF CAL test template. Target level generator output will follow its reference calibration generator output hence the Power Control dropdown box will be greyed out.
Take note that SE tests are not run via a normal EMS Scan Test but in EMS Auto Test.
Fig. 5-8: Reference calibration table selection
5.3 EUT Monitoring Template
Refer to chapter 3.3 "Online Help" on page 6. In HTML Help, select the Index tab, search for "EUT Monitoring" to show a detailed description on setting up the EUT
Monitoring template.
1.00
Fig. 5-9: Online help for EUT monitoring
Rohde & Schwarz Shielding Effectiveness Measurement
16
Test Template Configurations
The main purpose of a EUT measurement is to arrest unwanted failures during operation by stressing the EUT with a signal. For this purpose, the EUT monitoring template is necessary to ensure that certain parameters of the EUT are under stress still behaving as usual. It can also provide the EUT’s worst-case results.
5.3.1 EUT Monitoring Template for SE
The settings of each EUT monitoring channel that use the spectrum analyzer are configured as follow:
In Fig. 5-10, select Display tab and set the Units to be displayed as "dB". Under the section Value Conversion, enter the evaluation formula as:
IMMLVL{Imm Lvl dB}-107-MEASVAL{Meas Value}
This is to convert the measured value in dBm and tabulate the effective results in dB.
Fig. 5-10: Evaluation formula for value conversion
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
17
Test Template Configurations
In Fig. 5-11, select Options tab. Under EMS Auto Test Evaluation, select checkbox option "Worst Case Analysis. Under Evaluation Mode, select "Max. Peak".
Fig. 5-11: EMS auto test evaluation options
5.3.2 Average Detector for EUT Monitoring
For SE testing, select "Average Detector" as shown in Fig. 5-12, instead of "Max Peak
Detection". This option should be selected regardless of whether spectrum analyzer, or test receiver is used. This option is only available with EMC32-K48.
In Fig. 5-12, select Hardware tab. Under Detector, select from dropdown list
"Average
”.
1.00
Fig. 5-12: Hardware tab in EUT monitoring window
Rohde & Schwarz Shielding Effectiveness Measurement
18
Test Template Configurations
5.3.3 Limit Line Input for EUT Monitoring
In EMC32, the NoGo in EUT monitoring defines the limit line for SE. The value of the limit line must be input in the NoGo tab as shown in Fig. 5-13. The value can either be a constant value (e.g. 120 dB), or a shape table which consists of different values at different frequencies.
1.00
Fig. 5-13: NoGo tab in EUT Monitoring window
In the NoGo tab, the criteria for pass or fail are defined, with four NoGo types to choose from:
ı
Above Limit: The EUT has failed if the measured (and converted) value is bigger than a limit value or the value from a limit shape to be defined by the user. This type will be the preferred setting for SE measurement
ı
Below Limit: The EUT has failed if the measured (and converted) value is smaller than a limit value or the value from a limit shape to be defined by the user.
ı
Outside Value Range: The EUT has failed if the measured (and converted) value is outside of a window of values to be defined by the user.
Rohde & Schwarz Shielding Effectiveness Measurement
19
Test Template Configurations
ı
None: The EUT will never be considered to have failed, the channel is only used for recording the EUT's parameter.
The limits defined here will be displayed in the graphics window associated to the channel.
5.4 EUT Auto Test Template
Refer to chapter 3.3 "Online Help" on page 6. In HTML Help, select the Index tab, search for "EMS Auto Test Template Editor" section to show a detailed description on setting up the EUT Auto Test template.
Fig. 5-14: Online help for EMS auto test template editor
EUT auto test (known as EMS auto test in EMC32), further enhances the automation capability of SE test. It allow users to repeat frequency sweeping of EUT measurement for multiple location, multiple subranges and different polarization when EMC32-K48 option is used. In addition, EUT monitoring template can be used together with EUT auto test to calculate SE for each location. Worst case analysis feature is also available to obtain the worst case result over all locations for every frequency point.
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
20
Test Template Configurations
5.4.1 EUT Auto Test for SE
This section shows the configuration for EUT Auto Test for SE Test.
From Fig. 5-15, left-click on Measurement Settings. Enter the same setup as the
EMS Scan template for EUT test.
Fig. 5-15: Measurement settings for EUT auto test
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
21
Test Template Configurations
From Fig. 5-16 left-click on Loop Settings, add polarization and auto test subranges.
Under User Definition Loop Settings, enter the number of antenna positions. The step number corresponds to the antenna position number. For example, step 1 refers to antenna position 1.
Select checkbox Visible Column in the Report to display each loop column in the test report. Select checkbox Show Trace for each Loop Result to show loop result graphics in the test report.
Fig. 5-16: Loop settings for EUT auto test
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
22
Test Template Configurations
From Fig. 5-17, left-click on Evaluation Settings to show its dialog box. Under EUT
Worst Case Analysis, select checkbox for "Do Worst Case Analysis for EUT
Monitoring Channels".
Fig. 5-17: Evaluation settings for EUT auto test
From Fig. 5-18, left-click on General Settings to open its dialog box. Set the "EUT
Monitoring" file to be used.
1.00
Fig. 5-18: General settings for EUT auto test
Rohde & Schwarz Shielding Effectiveness Measurement
23
Running of Test
6 Running of Test
6.1 Reference Level Test for SE
The objective of a reference calibration is to set a known calibrated level at the
connection point to the transmitting antenna. This calibrated level will then be used again with the EUT to get the SE of the shielded enclosure.
As shown in Fig. 6-1 below, the output cable is directly connected to the transmitting antenna; and the receiving antenna output cable is directly connected to the spectrum analyzer. The minimum respective distances between the transmitting antenna and receiving antenna are 0.6 m at 9 kHz to 20 MHz, 2 m at 20 MHz to 1 GHz and 1 m above 1 GHz.
1.00
Fig. 6-1: Reference level test for SE
Rohde & Schwarz Shielding Effectiveness Measurement
24
Running of Test
Under Test Template > EMS Scan > SE - Reference calibration, right-click on the appropriate reference calibration template and select "New Test".
Fig. 6-2: New test selection
In Fig. 6-3, select Test Definition tab. Under Test method, select "Reference
Calibration" from the dropdown list and click OK.
1.00
Fig. 6-3: Test method selection
Rohde & Schwarz Shielding Effectiveness Measurement
25
In Fig. 6-4, click play/start button to begin the reference calibration.
Running of Test
Fig. 6-4: Reference level test
A prompt window message will appear to announce completion of the reference calibration process. If multiple frequency subranges were inputted in the calibration, a prompt window as shown in Fig. 6-5 will appear. Choose the corresponding antenna position for the subrange that will be measured.
1.00
Fig. 6-5: Reference calibration data
Rohde & Schwarz Shielding Effectiveness Measurement
26
Running of Test
The naming convention for saving the reference calibration table will be
SE_RC_Name_SR0x_POS0y; where SE_RC_Name is the reference calibration name, x is the subrange number and y is the antenna position.
Fig. 6-6: Reference calibration filename convention
Save desired reference calibration for all subranges and antenna positions that are to be tested. Commence with the EUT testing. With the calibration results, the EUT test can now begin.
1.00
Fig. 6-7: Reference calibration result parameters
Rohde & Schwarz Shielding Effectiveness Measurement
27
Running of Test
6.2 EUT Auto Test for SE
The receiving antenna is placed inside the EUT to be measured. Using EMC32, run the Auto Test template as described in the actions that follows.
Fig. 6-8: EUT test for SE
Under Test Template > EMS Auto Test, right-click on the appropriate test template and select "New Test".
1.00
Fig. 6-9: New test selection
Rohde & Schwarz Shielding Effectiveness Measurement
28
Running of Test
In Fig. 6-10, select Test Definition tab. Under Test method, select "EMS Auto Test" from the dropdown list. Under EUT Monitoring Parameters, select the appropriate
EUT monitoring file and click OK.
Fig. 6-10: Test method selection
In Fig. 6-11, click play/start button to begin the EMS Auto Test for SE.
1.00
Fig. 6-11: EUT auto test
Rohde & Schwarz Shielding Effectiveness Measurement
29
Running of Test
In the left window toolbox under User Definition, right-click the corresponding set position. Select Set as next Loop Position to move to the next frequency range or polarization.
Fig. 6-12: Next loop position
The worst case analysis will be reflected when EUT Monitoring Template is used with
EMS Auto Test.
1.00
Fig. 6-13: Worst case analysis
Rohde & Schwarz Shielding Effectiveness Measurement
30
The results are also made available in table format.
Running of Test
Fig. 6-14: Worst case result in table format
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
31
Printing Report
7 Printing Report
Refer to chapter 3.3 "Online Help" on page 6. In HTML Help, select the Index tab, search for "Report" to show a detailed description on setting up the Report template and print it out.
Fig. 7-1: Online help for report overview
7.1 Printing Report for SE
This section provides a guide to setting up and saving a report for SE.
When the test is complete, click on "Finish the Measurement" measurement mode.
to exit
1.00
Fig. 7-2: Measurement mode exit
Rohde & Schwarz Shielding Effectiveness Measurement
32
Printing Report
From the “Test Components” tab, right-click on the folder “Report Setups” and select
“Add Report Setup”.
Fig. 7-3: Report setup
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
33
Select the appropriate report template and click OK
Printing Report
Fig. 7-4: Report template selection
The new setup will be shown in the folder "Report Setups" (see Fig. 7-5).
Fig. 7-5: Report setup creation
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
34
Printing Report
From Fig. 7-6, double-click the designated report filename for more details.
Fig. 7-6: Report details
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
35
Printing Report
In the left window toolbar under Selected Components (Fig. 7-6), double-click on
"Information" to enter the test report information description.
Fig. 7-7: Information details
In the left window toolbar under Selected Components (Fig. 7-6), double-click on
"Hardware Setup" report.
to select the hardware setup required in the
1.00
Fig. 7-8: Hardware setup option
Rohde & Schwarz Shielding Effectiveness Measurement
36
Printing Report
In the left window toolbar under Selected Components (Fig. 7-6), double-click on
"Test Template" to select the test template required in the report.
1.00
Fig. 7-9: Test template details
Rohde & Schwarz Shielding Effectiveness Measurement
37
Printing Report
In the left window toolbar under Selected Components (Fig. 7-6), double-click on
"Table" to select the type of tables required in the report.
Under Table Name, click on to select table types.
Fig. 7-10: Table selection
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
38
Printing Report
In the left window toolbar under Selected Components (Fig. 7-6), double-click on
"Graphics" to select the type of graphs required in the report.
More than one graph can be added to the report if the option is available.
From Graphics Display, under the field Graphics Arrangement, we recommend selecting "2 rows x 1 column" for optimum display.
Fig. 7-11: Graphics details
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
39
Printing Report
In the left window toolbar under Selected Components (Fig. 7-6), double-click on
"Image/Photo" report.
to select any required image or photo for SE into the
Fig. 7-12: Image/photo selection
1.00 Rohde & Schwarz Shielding Effectiveness Measurement
40
Printing Report
In the left window toolbar under Available Components (Fig. 7-6), click on "Export the report" to export and save the final test report. Three types of file formats are available: PDF, RTF and HTML.
Select ‘Save to the selected directory” and save the report to your desired file location.
Click OK to save the report.
1.00
Fig. 7-13: Saving of report
Rohde & Schwarz Shielding Effectiveness Measurement
41
Printing Report
Rohde & Schwarz
The Rohde & Schwarz electronics group offers innovative solutions in the following business fields: test and measurement, broadcast and media, secure communications, cybersecurity, radiomonitoring and radiolocation. Founded more than 80 years ago, this independent company has an extensive sales and service network and is present in more than 70 countries.
The electronics group is among the world market leaders in its established business fields. The company is headquartered in Munich, Germany. It also has regional headquarters in Singapore and
Columbia, Maryland, USA, to manage its operations in these regions.
Regional contact
Europe, Africa, Middle East
+49 89 4129 12345 [email protected]
North America
1 888 TEST RSA (1 888 837 87 72) [email protected]
Latin America
+1 410 910 79 88 [email protected]
Asia Pacific
+65 65 13 04 88 [email protected]
China
+86 800 810 82 28 |+86 400 650 58 96 [email protected]
Sustainable product design
ı
ı
ı
Environmental compatibility and eco-footprint
Energy efficiency and low emissions
Longevity and optimized total cost of ownership
This application note and the supplied programs may only be used subject to the conditions of use set forth in the download area of the Rohde &
Schwarz website.
R&S
®
is a registered trademark of Rohde & Schwarz GmbH & Co.
KG; Trade names are trademarks of the owners.
1.00
Rohde & Schwarz GmbH & Co. KG
Mühldorfstraße 15 | 81671 Munich, Germany
Phone + 49 89 4129 - 0 | Fax + 49 89 4129
– 13777
Rohde & Schwarz Shielding Effectiveness Measurement www.rohde-schwarz.com
42
advertisement
Key Features
- Shielding effectiveness measurement
- R&S EMC32-K48 option
- Reference level testing
- EUT test and measurement
- Report generation
- Test template configuration
- Hardware setup
- Instrumentation setup
- Multiple frequency subranges
- Worst case analysis
Frequently Answers and Questions
What is the purpose of the R&S EMC32-K48 option?
What are the main steps involved in setting up the test configuration?
How is the reference level test performed?
What is the purpose of the EUT test?
Related manuals
advertisement
Table of contents
- 3 Overview
- 4 Introduction
- 5 EMC32 Setup
- 5 Installation of Software and Drivers
- 5 EMC32 on the Web
- 6 Update Manager
- 6 iKey Requirements
- 6 Online Help
- 7 Test Configuration
- 7 Instrumentation
- 9 Hardware Setup
- 10 Test Template Configurations
- 10 Reference Level Test Templates
- 11 Reference Level Test Template for SE
- 14 EUT Test Template
- 15 EUT Test Template for SE
- 16 EUT Monitoring Template
- 17 EUT Monitoring Template for SE
- 18 Average Detector for EUT Monitoring
- 19 Limit Line Input for EUT Monitoring
- 20 EUT Auto Test Template
- 21 EUT Auto Test for SE
- 24 Running of test
- 24 Reference Level Test for SE
- 28 EUT Auto Test for SE
- 32 Printing Report
- 32 Printing Report for SE