Xcalibur 4.0 XReport User Guide Version A

Xcalibur 4.0 XReport User Guide Version A
Thermo
XReport
Reporting Tool for the Xcalibur Data System
and the LCquan application
User Guide
XCALI-97783 Revision A
August 2015
© 2015 Thermo Fisher Scientific Inc. All rights reserved.
Xcalibur is a registered trademark and LCquan is a trademark of Thermo Fisher Scientific Inc. in the United
States.
Microsoft, Windows, and Excel are registered trademarks of Microsoft Corporation in the United States and
other countries. Adobe, Acrobat, and Reader are registered trademarks of Adobe Systems Incorporated in the
United States and other countries.
The following are registered trademarks in the United States and possibly other countries: Oracle is a registered
trademark of Oracle Corporation and/or its affiliates.
All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
Release history: Revision A, August 2015
Software version: Xcalibur 4.0 and later and LCquan 2.9 and possibly later.
For Research Use Only. Not for use in diagnostic procedures.
C
Contents
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Related Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Safety and Special Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .viii
Contacting Us . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ix
Thermo Scientific
Chapter 1
Introduction to XReport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
The XReport Window. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Report Templates Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Report Template . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Report Sections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Report Objects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Data Source Files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Configuring XReport Objects Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Configuring a Repeating Section Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Configuring a Table Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Using the Chromatogram Wizard Overview . . . . . . . . . . . . . . . . . . . . . . . . . 10
Configuring a Component Calibration Curve Overview . . . . . . . . . . . . . . . . 10
Using the Data Source Browser Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Sample XReport Templates Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Using XReport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Starting XReport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Selecting a Report Template . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Creating and Generating Reports with XReport . . . . . . . . . . . . . . . . . . . . . . 13
Exiting XReport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Chapter 2
Creating a Template . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Opening a New or an Existing XReport Template . . . . . . . . . . . . . . . . . . . . . . 16
Adding, Ordering, and Configuring Sections . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Adding, Deleting, or Reordering Sections . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Resizing Sections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Configuring a Qual or Quan Repeating Section . . . . . . . . . . . . . . . . . . . . . . 18
Adding and Configuring Report Objects. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Adding, Deleting, Repositioning, and Resizing Objects. . . . . . . . . . . . . . . . . 19
Aligning Objects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
XReport User Guide
iii
Contents
Centering Objects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Using the Report Template Grid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Using Columns . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Using Page Breaks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Configuring Objects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Adding User Text to a Text Object . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Adding a Formula to a Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Changing Font Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Changing the Page Layout. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Previewing and Saving the Template . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
iv
XReport User Guide
Chapter 3
Generating a Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Creating a Report Within XReport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Choosing Data Sources for a Template . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Generating, Saving, and Printing a Report . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Creating a Report from the Xcalibur Data System. . . . . . . . . . . . . . . . . . . . . . . 35
Sample Reports and Summary Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Generating a Report From Sequence Setup . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Generating a Report From Quan Browser . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Creating a Signed Report. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Chapter 4
Configuring Objects and Sections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
Configuring a Repeating Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Configuring a Text Object . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Configuring Text Fields. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Adding Custom Text Fields. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Configuring Text Attributes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Setting Text Fonts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Configuring an Annotation Object . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Configuring a Table Object. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Choosing Table Columns . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Configuring Custom (User) Columns . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Specifying Table Sort Order . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Setting Table Fonts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Setting Additional Table Properties . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Configuring a Bitmap Object . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Configuring a Chromatogram Object . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Use Container Supplied Quan Peak Page . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Use Container Supplied Qual Peak Page . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Select Chromatogram Type Page. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Select Demo File Page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Select the Quan Peak Page. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Select the Qual Peak Page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Select Mass Spectrum Quan Traces to Display Page . . . . . . . . . . . . . . . . . . . 88
Configure How the Peak Window Will Look Page . . . . . . . . . . . . . . . . . . . . 90
Thermo Scientific
Contents
Select Plot Details Page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Select Enhance Details Page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Select Chromatogram Style Page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Select the Peak Labeling Page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Select How the Chromatogram Axes Will Be Configured Page . . . . . . . . . . 101
Configure Normalization for the Chromatogram Page . . . . . . . . . . . . . . . . 102
Finishing the Chromatogram Wizard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Configuring a Component Calibration Curve . . . . . . . . . . . . . . . . . . . . . . . . . 103
Configuring a Spectrum Object. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Configure Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Configure Plot Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Enhance Spectra . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Configure Spectrum Style . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Configure Spectrum Peak Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Configure Axes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Select Normalization Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Specify a Peak . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Configuring Columns . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Set Sample Repeat Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Set Summary Repeat Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Appendix A Sample XReport Template Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .117
Appendix B Types of Report Objects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .119
Text Object Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Table Object Descriptions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Graphic Object Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Formatting Object Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Appendix C Sample Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .125
Peak Integration Report, Page 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
Peak Integration Report, Page 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Peak Integration Report, Page 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Peak Integration Report, Page 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Appendix D XReport Application Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .131
Reports View. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
XReport Toolbar Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132
XReport Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
XReport Dialog Boxes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138
Repeating Section Properties Dialog Box. . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Chromatogram Properties Dialog Box and Wizard . . . . . . . . . . . . . . . . . . . 141
Spectrum Properties Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Table Properties Wizard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 162
Edit Formula For Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166
Thermo Scientific
XReport User Guide
v
Contents
Configure Function Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
Spectrum List Table Properties Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . 168
Average Scan Filter Selection Dialog Box. . . . . . . . . . . . . . . . . . . . . . . . . . . 172
Library Search Graphics Table Properties Dialog Box . . . . . . . . . . . . . . . . . 172
Library Search Results Table Properties Dialog Box . . . . . . . . . . . . . . . . . . 172
Sample Table Properties Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173
Text Object Properties Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173
Bitmap Properties Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
Column Properties Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
Component Calibration Curve Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . 180
Data Sources Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180
Electronic Signature Properties Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . 181
Font Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181
Preferences Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181
Preview Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182
Save As Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183
Template Configuration Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184
Page Setup Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184
Report Sections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186
Nonrepeating Section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186
Report Header and Footer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187
Repeating Sections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
Report Objects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
Graphic and Formatting Objects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
Table Objects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192
Text Objects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 209
Report Templates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 221
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .223
vi
XReport User Guide
Thermo Scientific
P
Preface
Thermo™ XReport is the custom reports application for the Thermo Xcalibur™ mass
spectrometry data system and the LCquan application. Use XReport to create custom reports.
Contents
• Related Documentation
• Safety and Special Notices
• Contacting Us
To provide us with comments about this document, please click the link below. Thank you in
advance for your help.
Related Documentation
Thermo Fisher Scientific provides the following documentation for the Xcalibur data system:
• Xcalibur Quick Start
• Xcalibur Quantitative Analysis Getting Started Guide
• Xcalibur Data Acquisition and Processing User Guide
• Xcalibur Quan Browser User Guide
• Xcalibur Qual Browser User Guide
• Xcalibur Library Browser User Guide
• XReport User Guide
• Help from within the data system and the XReport application
Thermo Scientific
XReport User Guide
vii
Preface
Safety and Special Notices
Make sure you follow the precautionary statements presented in this guide. The safety and
other special notices appear in boxes.
Safety and special notices include the following:
IMPORTANT Highlights information necessary to prevent damage to software, loss of
data, or invalid test results; or might contain information that is critical for optimal
performance of the system.
Note Highlights information of general interest.
Tip Helpful information that can make a task easier.
viii
XReport User Guide
Thermo Scientific
Preface
Contacting Us
There are several ways to contact Thermo Fisher Scientific for the information you need. You
can use your smartphone to scan a QR code, which opens your email application or browser.
Contact us
Customer Service and Sales
Technical Support
(U.S.) 1 (800) 532-4752
(U.S.) 1 (800) 532-4752
(U.S.) 1 (561) 688-8731
(U.S.) 1 (561) 688-8736
us.customer-support.analyze
@thermofisher.com
us.techsupport.analyze
@thermofisher.com
 To find global contact information or customize your request
1. Go to www.thermoscientific.com.
2. Click Contact Us, select the Using/Servicing a Product option, and then
type the product name.
3. Use the phone number, email address, or online form.
 To find product support, knowledge bases, and resources
Go to www.thermoscientific.com/support.
 To find product information
Go to www.thermoscientific.com/lc-ms.
Note To provide feedback for this document:
• Send an email message to Technical Publications ([email protected]).
• Complete a survey at www.surveymonkey.com/s/PQM6P62.
Thermo Scientific
XReport User Guide
ix
1
Introduction to XReport
The XReport application is the custom reports feature of the Xcalibur data system. Use
XReport to create customized reports containing all of the results and important information
produced with the application. In addition, use XReport to format a report to a style, print
the report, and save the report in text (TXT), Microsoft™ Word document (DOC), Microsoft
Excel™ workbook (XLS), rich text (RTF), portable display (PDF), or web page (HTM or
HTML) formats. Use these options to email reports or display them on the Internet.
Create Xcalibur reports containing the following:
• Textual information (for example, sample information)
• Tabulated information (for example, a calibration file table)
• Graphical information (for example, a spectrum or a chromatogram)
Contents
• The XReport Window
• Report Templates Overview
• Configuring XReport Objects Overview
• Using XReport
Thermo Scientific
XReport User Guide
1
1
Introduction to XReport
The XReport Window
The XReport Window
The XReport window is split in four different panes:
• Sections pane
• Objects pane
• Template pane
• Report Template Outline pane
See Figure 1 for the location of the panes.
Figure 1.
XReport window
Sections
pane
Objects
pane
Template pane
Template outline pane
The Sections pane lists the available sections for the selected application (for example, the
Xcalibur data system). To add a section to a report, drag it from the Sections pane to the
Template pane.
2
XReport User Guide
Thermo Scientific
1
Introduction to XReport
The XReport Window
The Objects pane lists the objects that are available for the currently selected section. To add
an object to a report, drag it from the Objects pane to the currently selected section in the
Template pane.
Use the Template pane to configure a report layout by dragging and arranging objects.
• To configure a report object, right-click it and choose Properties from the shortcut
menu.
• To configure a repeating section, right-click the QL or QN icon in the left margin of the
page at the top of the sections pane and choose Properties from the shortcut menu. The
size of the page in this pane corresponds to the size of a standard 8.5  11 in. piece of
paper in either portrait or landscape mode.
• Adjust the height of sections by dragging the bottom border of the section downward. A
selected section has a bright blue border around it.
The Report Template Outline pane lists the sections that are in the current report. You can
move or delete sections from this pane, as well as undock, reposition, or close it.
• To undock the pane, drag the top gray portion away from the pane.
• To redisplay the pane, choose View > Report Template Outline pane.
• To delete a section from a template, click it in the Report Template Outline pane and
press DELETE.
• To move a section, drag it up or down in the Report Template Outline pane list. You
cannot move the header and footer sections.
• To select and scroll to a section, double-click it.
Thermo Scientific
XReport User Guide
3
1
Introduction to XReport
Report Templates Overview
Report Templates Overview
This section defines the basic terminology used throughout the software and this manual for
new users of the XReport application and the Xcalibur data system. This section defines the
following key terms:
• Report Template
• Report Sections
• Report Objects
• Data Source Files
• Report
Report Template
An XReport template is an XReport document (XRT file) that defines the content, layout,
and style of a report. Use the template to include sections and report objects (for example,
graphs, tables, annotation text, or graphics) in a report template. The software generates an
Xcalibur report after combining a report template with Xcalibur data source files.
XReport provides several example report templates for you to use. These are located in the
Xcalibur\Templates folder. For details about the preconfigured report templates, see
Appendix A, “Sample XReport Template Files.”
Report Sections
Divide the report template in sections to hold report objects. Add sections to a report
template using the Sections pane in the XReport window.
The basic types of section used in XReport are as follows:
• Header or Footer Sections
• Repeating Sections
• Nonrepeating Sections
A report object that is placed in a repeating section can behave differently from the same
report object placed in a nonrepeating section when the software combines a report template
with data. For information on how a report object functions when it is placed inside or
outside a repeating section, see Appendix B, “Types of Report Objects.”
Header or Footer Sections
A header or footer section is a section of the report that is automatically repeated at the top
and bottom of each page in the resolved report. Header and footer sections can contain
objects such as a company logo, report information, page numbers, and date stamps.
4
XReport User Guide
Thermo Scientific
1
Introduction to XReport
Report Templates Overview
Repeating Sections
A repeating section is a section of a report template that is automatically repeated for every
component or peak in a resolved document. Change the properties of a repeating section to
customize its behavior.
Two types of repeating section can be used in an XReport template:
Qual repeating section
Use a Qual repeating section to display details relating to a
single Qual peak within some data. The section is repeated for
all Qual peaks found.
Quan repeating section
Use a Quan, or component, repeating section to display
details relating to a single Quan peak in some data or a
component configured in a processing method. The section is
repeated for all Quan peaks or components.
Nonrepeating Sections
A nonrepeating section usually displays static details associated with data or methods that
apply once and globally within the system (for example, processing method settings,
instrument method settings, filename, confirmation ion information, and so on).
Report Objects
Report objects are customized tables, graphics, or text that indicate what kind of data appears
in the report. When the Xcalibur data system combines a report template with data, it
replaces the report objects in the report template with data to generate a report. See Report
Objects for more information.
Report objects include the following:
• Text Objects
• Table Objects
• Graphic Objects
• Formatting Objects
For a descriptive listing of XReport objects, see Appendix B, “Types of Report Objects.”
Text Objects
Text objects contain static text labels and associated customized data fields. To customize the
specific fields to display, right-click the object and choose Properties from the shortcut menu.
The Configuration dialog box opens. When the software combines a template with data to
generate a report, it replaces the text object fields with textual information (static text). See
Text Objects for more information.
Thermo Scientific
XReport User Guide
5
1
Introduction to XReport
Report Templates Overview
The following Xcalibur text objects are available in XReport:
• Annotation
• Component Ident/Detect/Cal Settings
• Component ISTD Settings Summary
• Component Settings Summary
• Component Sys Suit/Flags Settings
• Instrument Method
• Processing Method – General
• Processing Method – Qual
• Quan ISTD Peak Summary
• Quan Peak Summary
• Report Info
• Sample Header
Table Objects
When the Xcalibur data system combines a template with data, it replaces table objects with
tabulated information (creates a populated data table). Configure or change the properties of a
table object to customize its output.
6
XReport User Guide
Thermo Scientific
1
Introduction to XReport
Report Templates Overview
XReport provides the following table objects:
• Avalon Qual Events Table
• Proc. Method Programs Table
• Avalon Quan Events Table
• Proc. Method Sample Reports Settings
Table
• Calibration File Table
• Component Cal Level Table
• Component QC Level Table
• Component Spectrum Table
• Dilution Factor Table
• Electronic Signature
• Ion Ratio Confirmation Results Table
• IRC Settings Table
• IRC Summary Table
• Lib. Search Graph Table
• Proc. Method Summary Reports
Settings Table
• Qual Peak Table
• Qual Summary Table
• Quan Peak Table
• Quan Summary Table
• Sample Table (Quan Results)
• Sequence Table
• Spectrum Candidate Results Table
• Spectrum List Table
• Lib. Search Results Table
Graphic Objects
When XReport combines a report template with data, it replaces graphic objects with
graphical information. Configure or change the properties of a graphic object to customize its
output. See Graphic and Formatting Objects for more information.
The following graphic objects are available in XReport:
• Bitmap
• Chromatogram
• Component Cal (Calibration) Curve
• Spectrum
Thermo Scientific
XReport User Guide
7
1
Introduction to XReport
Report Templates Overview
Formatting Objects
These objects, which only format the report, are available in XReport:
• Columns
• Page Break
Data Source Files
Data source files are files that contain acquired data. For example, there are five types of
Xcalibur data source files:
• Processing method (PMD file)
• Raw data file (RAW file)
• Result (RST file)
• Sequence (SLD file)
• Calibration (XCAL file)
Report
Generate a report (or resolved document) using a report template. In an Xcalibur report, the
software combines the report objects in the report template with Xcalibur data. Save and
display reports in any of these formats:
• Microsoft Word document (DOC file)
• Text (TXT file)
• Microsoft Excel workbook (XLS file)
• Rich text (RTF file)
• PDF (PDF file)
• Web page (HTM or HTML file)
8
XReport User Guide
Thermo Scientific
1 Introduction to XReport
Configuring XReport Objects Overview
Configuring XReport Objects Overview
Use the Template pane of the XReport window to configure objects by right-clicking them
and choosing Properties from the shortcut menu. Set parameters in the configuration dialog
box that opens.
You can configure nearly all items in XReport, including the following:
• Repeating sections
• Tables
• Chromatograms
• Spectra
• Component calibration curves
• Text objects
For specific information on configuring each object, see Chapter 4, “Configuring Objects and
Sections.”
Configuring a Repeating Section Overview
Use the Repeating Section Properties dialog box to specify the following:
• Whether to exclude Internal Standards or Target Compounds from a Quan repeating
section
• The order in which peaks and components are repeated
• Any limit on number of peaks to be reported
• How the individual peaks or components are split across pages in the final report
For information about the parameters in the Repeating Section Properties dialog box, see
Repeating Section Properties Dialog Box.
Configuring a Table Overview
Use the Table Properties wizard to do the following:
• Specify the columns to be included in the table
• Define and customize table columns
• Specify an order for the table
For information about the parameters in the Table Properties wizard, see Table Properties
Wizard.
Thermo Scientific
XReport User Guide
9
1
Introduction to XReport
Configuring XReport Objects Overview
Using the Chromatogram Wizard Overview
Use the Chromatogram wizard to do the following:
• Select chromatogram type and peak type
• Use a demonstration (demo) file
• Select plot details
• Select smoothing parameters
• Use MS Quan traces
• Configure the Chromatogram display, style, labeling, axes, and normalization methods
For information about the parameters in the Chromatogram wizard, see Chromatogram
Properties Dialog Box and Wizard.
Configuring a Component Calibration Curve Overview
Use the Component Calibration Curve wizard to specify the name of the component plotted
for a nonrepeating Component Cal Curve report object in a report.
For information about the parameters in the Component Calibration Curve dialog box, see
Component Calibration Curve Dialog Box.
Using the Data Source Browser Overview
Use the Data Source Browser to select data files that can be used to generate reports from
within XReport.
For information about the parameters in the Data Sources dialog box, see Data Sources
Dialog Box.
Sample XReport Templates Overview
An XReport template is an XReport document (XRT file) with XReport objects embedded
within it. You do not have to create personal XReport templates. XReport provides
customized report templates for you to use with your data. For full details of these report
templates, see Appendix A, “Sample XReport Template Files.”
10
XReport User Guide
Thermo Scientific
1
Introduction to XReport
Using XReport
Using XReport
To use the XReport application, follow these topics:
• Starting XReport
• Selecting a Report Template
• Creating and Generating Reports with XReport
• Exiting XReport
Starting XReport
 To start XReport
1. From the data system computer, double-click the XReport icon,
, on the desktop,
or choose All Programs (or Programs) > Thermo Xcalibur > XReport from the
Windows Start menu.
When more than one application is installed that uses XReport for reporting (for
example, both the Xcalibur data system and a compatible version of LCquan™), the
Template Configuration dialog box opens (Figure 2).
Figure 2.
Thermo Scientific
Template Configuration dialog box
XReport User Guide
11
1
Introduction to XReport
Using XReport
2. Select the application for the report template and click OK.
XReport opens and creates a new template (Figure 3).
Figure 3.
XReport window, showing a new template
Selecting a Report Template
 To select a report template
1. In the Reports view, choose File > Open. A browse dialog box opens.
2. To select the required report template, click the template name. If the name does not
appear, browse to the correct folder.
The template name appears in the File name box.
3. To view the report template and close the dialog box, click Open.
12
XReport User Guide
Thermo Scientific
1
Introduction to XReport
Using XReport
Creating and Generating Reports with XReport
There are two distinct stages in the process of producing a customized Xcalibur report using
XReport:
• Creating a Report Template
• Generating a Report Using a Report Template
These stages are discussed in more detail in these chapters: Creating a Template and
Generating a Report. The Xcalibur data system provides sample XReport templates for you to
use. These are XRT files with XReport objects embedded within them and are located in the
Xcalibur\templates folder. For details on these report templates, see Appendix A, “Sample
XReport Template Files.”
Creating a Report Template
Use XReport to create a customized report template (.xrt file) that defines how the final report
looks and defines the specific information in the report.
 To create a report template
1. Click the XReport icon,
, on the desktop to start XReport.
A new XReport document appears in the report window (see Figure 3 on page 12).
2. Add sections and report objects to the report template by dragging items from the
Available Sections and Available Objects panes to the report preview pane.
3. Configure the sections and report objects included in the report template to adjust their
properties and customize their output.
Right-click objects and sections to configure and choose Properties from the shortcut
menu. Set parameters in the configuration dialog boxes. Use the menus, toolbars, or
mouse to align, resize, and reposition objects.
4. Save the report template.
For more detailed information on how to create and configure a report template, see
Chapter 2, “Creating a Template,” and Chapter 4, “Configuring Objects and Sections.”
Thermo Scientific
XReport User Guide
13
1
Introduction to XReport
Using XReport
Generating a Report Using a Report Template
After creating a report template, generate a report (or resolved document) by combining a
report template with one or more data source files (for example, an Xcalibur processing
method [PMD], raw data file [RAW], result file [RST], sequence file [SLD], and calibration
file [XCAL]).
For detailed information on how to generate a report from within the Xcalibur data system,
see Chapter 3, “Generating a Report.”
There are three ways to generate a report:
• From a report template open in XReport: Search for the required data source files and
combine the data with the template to generate a report. Preview, print, and save the
report as a text file (TXT), Word document file (DOC), Excel spreadsheet file (XLS), rich
text format file (RTF), portable document format file (PDF), or web page (HTM or
HTML).
• From a data processing application (for example, the Xcalibur data system): Associate the
data you produce in the application with a report template (for example, do this in Quan
Browser) and combine the data with the template to generate a single report.
• From a data processing application (for example, the Xcalibur data system): Insert the
name of the template to a processing method (such as Processing Setup) and generate
reports automatically when you batch process data (such as in Sequence Setup).
Exiting XReport
 To save the report template or report and XReport
Complete or cancel any active dialog boxes and choose File > Exit.
14
XReport User Guide
Thermo Scientific
2
Creating a Template
This chapter explains how to create a custom report template. When you create a custom
report template, you can save the template and use it to generate an Xcalibur report. For more
information about generating a report, see Chapter 3.
Contents
• Opening a New or an Existing XReport Template
• Adding, Ordering, and Configuring Sections
• Adding and Configuring Report Objects
• Previewing and Saving the Template
Thermo Scientific
XReport User Guide
15
2
Creating a Template
Opening a New or an Existing XReport Template
Opening a New or an Existing XReport Template
When you start XReport, XReport creates a new template that includes header and footer
sections (Figure 4). Either work with this new template, or choose File > Open to open an
existing template and select an XReport template (XRT) file.
Figure 4.
16
XReport User Guide
XReport window, showing a new template
Thermo Scientific
2 Creating a Template
Adding, Ordering, and Configuring Sections
Adding, Ordering, and Configuring Sections
To begin creating a template, select sections to include in the report template. Choose to
include Quan-based or Qual-based repeating sections or nonrepeating sections in your report
template. For a description of the different section types and how repeating sections work, see
Configuring a Repeating Section.
This section contains the following topics:
• Adding, Deleting, or Reordering Sections
• Resizing Sections
• Configuring a Qual or Quan Repeating Section
Adding, Deleting, or Reordering Sections
Follow these procedures.
 To add a section to the report
Drag it from the Sections pane (upper left) to the Template pane (center).
Add only one section to the template at a time.
 To delete a section from the report
1. Do one of the following:
• Click the section in the Report Template Outline pane (far right) and press
DELETE. The system prompts you to confirm the deletion. You cannot delete the
header or footer sections.
–or–
• When the Report Template Outline pane is not visible, choose View > Report
Template Outline or right-click a section and choose Delete Section from the
shortcut menu.
 To change the order of the sections in the template
Drag the sections in the Report Template Outline pane. Header and footer sections must
remain at the top and bottom of the list, respectively.
Thermo Scientific
XReport User Guide
17
2
Creating a Template
Adding, Ordering, and Configuring Sections
Resizing Sections
To resize a section, follow these procedures.
 To automatically remove the extra space at the top and bottom of a section
Select the section and choose Layout > Shrink Wrap.
 To automatically add a half page of extra space at the bottom of a section
Select the section and choose Layout > Expand.
 To manually add or remove extra space at the bottom of a section
Drag the bottom border of a section up or down.
Configuring a Qual or Quan Repeating Section
For more information about the Repeating Section Properties dialog box, see Repeating
Section Properties Dialog Box.
 To configure a Qual or Quan repeating section
1. Scroll to where you can see the QL or QN icon at the top left side of a section in the
Template pane.
2. Right-click the icon and choose Properties from the shortcut menu. The Repeating
Section Properties dialog box opens (Figure 5).
Figure 5.
Repeating Section Properties dialog box
3. Set the parameters and click OK.
18
XReport User Guide
Thermo Scientific
2 Creating a Template
Adding and Configuring Report Objects
Adding and Configuring Report Objects
After adding and configuring sections, select the report objects you want to include in a
specified section of the report template. Select from these options: text objects, table objects,
graphic objects, and text, depending on the selections you made in Configuring a Qual or
Quan Repeating Section.
XReport displays the objects that are available for the currently selected section in the Objects
pane (lower left). The contents of this pane change when you select different sections.
To add and configure report objects, follow these procedures:
• Adding, Deleting, Repositioning, and Resizing Objects
• Aligning Objects
• Centering Objects
• Using the Report Template Grid
• Using Columns
• Using Page Breaks
• Configuring Objects
• Adding User Text to a Text Object
• Adding a Formula to a Table
• Changing Font Settings
• Changing the Page Layout
Adding, Deleting, Repositioning, and Resizing Objects
You can add, delete, reposition, and resize objects in a report template.
 To add an object to the report
Drag an object from the Objects pane (lower left) to the currently selected section (in
blue) in the Template pane (center). Add only one object to the template at a time.
 To delete an object from the report
1. Select the object in the Template pane (center). Blue handlebars on each side and at each
corner become visible.
2. Press DELETE.
 To reposition an object
Drag the object to the preferred location.
Thermo Scientific
XReport User Guide
19
2
Creating a Template
Adding and Configuring Report Objects
 To resize an object
Drag the object by one of the corner handles (small blue squares) until the object is the
preferred size.
 To make two or more objects the same size
1. Select the first object to size.
The size of the first object selected determines the size of subsequent objects.
2. Hold down the SHIFT key and select other objects to resize.
3. Click the icon in the toolbar that corresponds to the dimension you want to resize:
Same Width
Same Height
Same Width and Height
 To resize the columns in a table object
1. Double-click the table.
Handles (small blue boxes) appear on the interior column borders.
2. Drag the handles to resize the columns.
Aligning Objects
 To align one or more objects
1. Select the first object to align. The position of the first object selected determines how
subsequent objects are aligned.
2. Hold down the SHIFT key and select each object to align.
3. Click the toolbar icon that corresponds to the edge to align.
Left Align icon
Right Align icon
Top Align icon
Bottom Align icon
Or click the Auto Align icon,
20
XReport User Guide
, to snap all objects to the nearest grid line.
Thermo Scientific
2 Creating a Template
Adding and Configuring Report Objects
4. Choose View > Report Template Grid to view the grid lines. To adjust the grid settings,
choose Layout > Preferences. The Preferences dialog box opens (Figure 6).
Centering Objects
 To center one or more objects
1. Select the first object to center. Then, hold down the SHIFT key and select each object to
center with respect to the first object.
2. To center the objects on the page, do the following:
• Click the Center Horizontally icon,
• Click the Center Vertically icon,
.
.
Using the Report Template Grid
The Template pane has a built-in grid.
 To view the grid lines
Choose View > Report Template Grid.
 To adjust the grid settings
1. Choose Layout > Preferences.
The Preferences dialog box opens (Figure 6). For more information about the Preferences
dialog box, see Preferences Dialog Box.
Figure 6.
Preferences dialog box
2. Set the spacing between grid lines and whether to snap objects to the grid as they are
added to the Template pane.
Thermo Scientific
XReport User Guide
21
2
Creating a Template
Adding and Configuring Report Objects
Using Columns
Add columns to arrange other reporting objects so that they appear in columns. To add a
column to the template, drag the column object from the Objects pane to the Template pane.
After adding a column object to the template, you can resize, reposition, and configure it.
 To configure a column object
1. Right-click the column object.
2. Choose Properties from the shortcut menu.
The Column Properties dialog box opens (Figure 7). To learn how to configure columns
and how columns work with different repeat types, see Configuring Columns.
Figure 7.
Column Properties dialog box
You can resize the entire column object (select it and drag the corner handles), but the
columns within the column object remain equal in size.
After you have finished configuring the columns, drag objects to the columns.
Using Page Breaks
If you do not manually insert page breaks, objects automatically shift to a new page when
there is not enough room on the current page.
 To preview automatic page breaks
1. Choose Layout > Preferences.
The Preferences dialog box opens (see Figure 6 on page 21).
2. Select the Show Page End check box.
22
XReport User Guide
Thermo Scientific
2 Creating a Template
Adding and Configuring Report Objects
 To add a page break to the template
1. Drag the Page Break object from the Objects pane to the Template pane.
2. After adding a page break to the template, drag the page break to reposition it.
3. Right-click the Page Break object and choose Properties from the shortcut menu.
The Page Setup dialog box opens. See Page Setup Dialog Box.
4. Choose the paper source and page orientation.
5. Click OK to return to the template pane.
The Page Break object displays the page orientation.
Configuring Objects
For information about the XReport objects, see Report Objects.
 To configure an object
1. Right-click the object.
2. Choose Properties from the shortcut menu.
The Properties dialog box opens.
For information about configuring each of the specific objects, see Chapter 4. For a
description of each object, including which sections each object can be used in, see
Appendix B, “Types of Report Objects.”
Adding User Text to a Text Object
To add additional customized text headings to text objects, use the <Blank Line > and <User
Text> Data Items. These items are available for all text objects except the Annotation Object
and the Instrument Method Object.
For more information, see Adding Custom Text Fields and Text Objects.
 To add custom text to a text object
1. Right-click the text object and choose Properties from the shortcut menu.
The Data Page of the Text Object dialog box opens.
2. To add a text item to the Selected list, do the following:
a. Select <User Text> in the Available Items list.
b. Click Add.
Thermo Scientific
XReport User Guide
23
2
Creating a Template
Adding and Configuring Report Objects
3. To position the <User Text> item where you want your text displayed, click Move Up or
Move Down.
4. In the Selected list, select the <User Text> item.
The User Text box appears.
5. In the User Text box, type the appropriate text.
6. Click Apply.
The text you entered replaces the <User Text> item in the Selected list.
 To edit a <User Text> item
1. In the Selected list, select the <User Text> item of interest.
2. In the User Text box, type the appropriate text.
3. Click Apply.
Adding a Formula to a Table
You can create custom formulas in XReport to perform mathematical operations on table data
and display the result in a table column.
 To add a formula to a table
1. In your XReport template, right-click a table object and choose Properties.
The Select Table Fields page opens (Figure 8). For information about the parameters on
this page, see Select Table Fields Page. The available parameters depend on the selected
table object. The screen captures in this procedure show the parameters for a Spectrum
List table object.
24
XReport User Guide
Thermo Scientific
2 Creating a Template
Adding and Configuring Report Objects
Figure 8.
Select Table Fields dialog box for a Quan Repeating – Spectrum List Table
2. Select one of the User Columns fields in the Available Columns list: [User Column 1],
[User Column 2], [User Column 3], [User Column 4], or [User Column 5].
For the five User Column fields available, you can create up to five separate formulas for a
single table.
3. Click Next to go to the Configure User Column Parameters Page (Figure 9).
For information about the parameters on this page, see Configure User Column
Parameters Page.
Figure 9.
Thermo Scientific
Configure User Column Parameters dialog box for the Spectrum List Table
XReport User Guide
25
2
Creating a Template
Adding and Configuring Report Objects
4. Type a title for this column in the User Column Title box.
5. Click Edit Formula.
The Edit Formula For Dialog Box opens (Figure 10). For information about the
parameters on this page, see Edit Formula For Dialog Box.
Figure 10. Edit Formula For dialog box
6. Enter a formula in the Formula box. Formulas can be entered in one of several ways:
• Type formulas directly in the box. Formulas obey the following syntax rules:
–
The formula must begin with an equals sign (=).
–
Each table column is represented by an alphabetic letter identifier (A, B, C, and
so on), displayed beneath the column title in the Table object.
• Type simple mathematical expressions (A+B, A–B, A*B, and A/B), as well as any of
the functions listed in the Functions list.
Mathematical formulas are computed on a row-by-row basis. That is, if you type
=A+B as a formula in a User Column, the results are computed as follows:
26
XReport User Guide
–
The value in the first row of column A is added to the value in the first row of
column B, and the result appears in the first row of the User Column.
–
The value in the second row of column A is added to the value in the second row
of column B, and the result appears in the second row of the User Column.
–
This process is continued until all rows of the table have been completed.
Thermo Scientific
2 Creating a Template
Adding and Configuring Report Objects
• Select a function from the Functions list and click Add Function. The Configure
Function Dialog Box opens. Select the columns to be used with the function. When
you configure your function and click OK, XReport enters the configured function
in the Formula box.
7. When you have entered your formula, click OK to close the Edit Formula For dialog box
and return to the Table Properties wizard. Continue through the remaining pages of the
wizard to finish configuring your table.
Changing Font Settings
In the XReport application, you can change font settings globally or locally for text and table
objects. When you change the font settings globally, each object added from that point
forward appears in the new fonts (objects that are already in the template are not updated).
When you change font settings locally, only the selected objects reflect the new font settings.
 To change the font settings for a single object
1. Do one of the following:
• Select the object and click
in the toolbar.
The Font dialog box opens (Figure 11).
Figure 11. Font dialog box for a table object
–or–
a. Right-click the object and choose Properties from the shortcut menu. For some
objects, the object type Properties dialog box opens. For table objects, a wizard opens.
b. Click the Font tab of the Properties dialog box or click Next in the table wizard until
a Fonts page opens.
2. Update the settings and click OK.
Thermo Scientific
XReport User Guide
27
2
Creating a Template
Adding and Configuring Report Objects
 To change the font settings for more than one text or table object at a time
1. Select the first object.
2. Hold down the SHIFT key while you select the other objects.
3. Click
in the toolbar.
The Font dialog box opens.
4. Update the settings and click OK.
 To globally change the font settings used for all new objects to add to the template
1. Click any section and confirm that no objects are selected.
2. Click
in the toolbar.
The Font dialog box opens.
3. Update the settings and click OK.
Changing the Page Layout
You can change the page layout for your XReport templates by using the Page Setup dialog
box.
 To change the page layout
1. Do one of the following:
• Choose File > Page Setup.
–or–
• Right-click a Page Break object in the template pane and choose Properties.
The Page Setup dialog box opens (Figure 12).
28
XReport User Guide
Thermo Scientific
2 Creating a Template
Adding and Configuring Report Objects
Figure 12. Page Setup dialog box
2. Change any of the following: header and footer section heights, portrait or landscape page
orientation, or margins.
3. Click OK to change the page layout.
Thermo Scientific
XReport User Guide
29
2
Creating a Template
Previewing and Saving the Template
Previewing and Saving the Template
Before you use a template to print reports, preview the template with the example data
provided with the Xcalibur data system or with a representative data file.
 To preview the appearance of the template
1. Choose Report > Simulate Report. The Preview window opens (Figure 13). The Preview
window shows the report with the default data.
Figure 13. Preview window
2. When the report is correct, click Close to close the preview window.
3. To save the template, choose File > Save. To save changes to a locked template choose
File > Save As and rename the file.
XReport uses the steroid example data that is provided with the Xcalibur data system as
the default data set. It consists of the following files:
• steroid.pmd
• steroids02.raw
• steroids02.rst
• steroid.sld
When your template is designed for a very different type of data set than the default data,
you might not see the results you expect when you preview the template. In that case,
choose appropriate data sources for the template and create the report as described in
Creating a Report Within XReport.
Use XReport to lock a template so that it can never be modified and saved with the same
filename. To modify a locked template, save it with a different filename. Choose File >
Open to open a locked template the same way you open any other template.
30
XReport User Guide
Thermo Scientific
2 Creating a Template
Previewing and Saving the Template
 To lock the template
1. Choose File > Save As.
The Save As dialog box opens (Figure 14).
Figure 14. Save As dialog box for XReport Templates
2. Give the report template an appropriate name and save it in the Xcalibur\templates folder
with the templates that ship with the Xcalibur data system.
3. Select the Lock Template check box and click Save.
Thermo Scientific
XReport User Guide
31
3
Generating a Report
To generate an Xcalibur report (resolved document), combine a report template with one or
more Xcalibur data source files (processing methods [PMD], raw data files [RAW], result files
[RST] sequence files [SLD], and calibration files [XCAL]).
There are two ways to generate reports:
• Use the XReport Data Source Browser to automatically search for the required data
source files and combine the data with a template to generate a report. This method
generates a single report.
• Associate the data you produced in the Xcalibur data system with a report template and
combine the data with the template to generate a report. This method generates multiple
reports from the data system.
Contents
• Creating a Report Within XReport
• Creating a Report from the Xcalibur Data System
Creating a Report Within XReport
To create a report within XReport, use these procedures:
• Choosing Data Sources for a Template
• Generating, Saving, and Printing a Report
Choosing Data Sources for a Template
The process of generating a report from an XReport template involves using the Data Sources
Dialog Box to search for the required data source files. The data source files required to
generate a report depend on the content (sections and report objects) of the report template.
To learn which data files are needed for each type of report object, see Appendix B, “Types of
Report Objects.”
Thermo Scientific
XReport User Guide
33
3
Generating a Report
Creating a Report Within XReport
 To resolve a template with data
1. To open the template (XRT file), choose File > Open in XReport.
The XReport Templates dialog box opens.
2. Select the template and click Open.
3. Choose Report > Data Sources.
The Data Sources dialog box opens (Figure 15).
Figure 15. Data Sources dialog box
4. In each box, enter or search for a data file.
If you specify a result file (RST) that references a raw data file (RAW) and that raw data
file does not exist or is in the wrong folder, XReport sends an error message. Make sure
the raw data file is in the correct folder.
5. Click OK.
Generating, Saving, and Printing a Report
 To create a report using selected data sources
1. Select the appropriate data sources as described in Choosing Data Sources for a Template.
2. Choose Report > Resolve Report in XReport.
The Preview window opens (Figure 16). The Preview window shows a report combined
with selected data sources.
34
XReport User Guide
Thermo Scientific
3 Generating a Report
Creating a Report from the Xcalibur Data System
Figure 16. Preview window
3. Click Save to save the report in the following file formats: Microsoft Word (DOC),
Microsoft Excel (XLS), web page (HTM or HTML), rich text (RTF), PDF (PDF), or text
(TXT).
4. To print the report, click Print.
5. To return to the XReport window, click Close.
Creating a Report from the Xcalibur Data System
When generating a report, the data system combines the data that you produced in the data
system with a report template. You can generate multiple reports from the data system. To
proceed, you must have a basic understanding of the Xcalibur data system. Refer to the
Xcalibur Getting Started Guide for more information.
Generate a report from the following areas of the application:
• See Generating a Report From Sequence Setup when running a sample or sequence and
batch processing.
• See Generating a Report From Quan Browser when using the Reports dialog box.
In addition, when generating reports from the Xcalibur data system, create reports containing
electronic signature information.
Always save and close the report template in XReport before using it to generate a report from
the Xcalibur data system.
Thermo Scientific
XReport User Guide
35
3
Generating a Report
Creating a Report from the Xcalibur Data System
Sample Reports and Summary Reports
To set up reporting options in the data system in Processing Setup or Quan Browser, select
two types of reports:
• Sample reports to print reports that give information about individual samples
• Summary reports to print reports that give summary information about an entire
sequence or set of samples
The example template files included with XReport indicate the specific templates to use to
generate sample reports or summary reports. For descriptions of sample files, see Sample
XReport Template Files.
Generating a Report From Sequence Setup
To generate reports from the Sequence Setup window, create a processing method that
includes the necessary report options on the Reports page of the method.
Use the Sequence Setup window to generate reports when running a sample or sequence or
batch processing a sequence.
See these topics for more information:
• Setting Report Options in a Processing Method
• Generating Reports when Running a Sequence
• Generating Reports when Batch Processing a Sequence
Setting Report Options in a Processing Method
To automated reporting, set up the appropriate report templates in your processing methods.
 To set up reports in a processing method
1. Open the Processing Setup window and set up a processing method. For more
information, refer to the Xcalibur Data Acquisition and Processing User Guide.
2. Choose View > Reports.
The Reports view opens.
3. To add one or more reports to either the sample or summary report tables, do the
following:
The sample table lists the reports to be produced for processed samples in a sequence. The
Summary table lists the reports to be produced for sequences or brackets.
a. Click the Enable column to display a check box and select the check box to enable
the report.
b. To create a sample table, select the Std, QC, Unk, and Other check boxes.
36
XReport User Guide
Thermo Scientific
3 Generating a Report
Creating a Report from the Xcalibur Data System
c. Select the output for the report in the Save As column. Choose from None (this
prints the report, but does not save it), Text, Doc, HTML, PDF, RTF, or XLS (select
one of these options to save the report in the selected format).
The Xcalibur data system saves reports created from the processing method in the
same directory as the source data files and names files. The reports use this format:
data file name template name.xxx
where
data file name is the name of the data file from which the report was generated
template name is the name of the template used to generate the report
xxx is the suffix [for example, (.doc) or (.pdf )] indicating the file type
When more than one report is generated from the same data file and template,
XReport adds a date stamp before the suffix.
d. Double-click the Report Name column to select an XReport template (XRT file) to
use for this report.
4. Save the processing method.
Generating Reports when Running a Sequence
You must create a sequence to run either a single sample or multiple samples from the
Sequence Setup window. When you click the Run Sample icon in the Sequence Editor
toolbar, the data system automatically selects the first row of the sequence. When you click the
Run Sequence icon, the data system automatically selects the entire sequence if you have not
already selected a subset of sequence rows.
 To generate reports when running a sequence
1. Set up the report options in a processing method as described in Setting Report Options
in a Processing Method.
2. In the Xcalibur Sequence Setup window, open the sequence that you want to run.
3. Specify the processing method that includes all of the report options in the Proc Meth
column.
Thermo Scientific
XReport User Guide
37
3
Generating a Report
Creating a Report from the Xcalibur Data System
4. To run a subset of the sequence, select the sequence rows that you want to run.
5. Choose Actions > Run This Sequence or Actions > Run This Sample.
The Run Sequence dialog box opens (Figure 17). The Run Rows box lists the row
selection.
Figure 17. Run Sequence dialog box
6. Select the Reports check box in the Processing Actions area.
7. Set all of the other parameters.
8. Click OK.
The data system submits the selected sample or samples to the acquisition queue.
38
XReport User Guide
Thermo Scientific
3 Generating a Report
Creating a Report from the Xcalibur Data System
Generating Reports when Batch Processing a Sequence
You can set up a batch processing job to print reports.
 To generate reports when batch processing a sequence
1. Set report parameters in the processing method as described in Setting Report Options in
a Processing Method.
2. In the Xcalibur Sequence Setup window, open the sequence that you want to batch
reprocess.
3. Enter the processing method that includes all of the report options in the Proc Meth
column.
4. Select the rows that you want to process.
5. Choose Actions > Batch Reprocess.
The Batch Reprocess Setup dialog box opens (Figure 18). The Process Rows box lists the
selected sequence rows.
Figure 18. Batch Reprocess Setup dialog box
6. If necessary, change the rows to be processed by entering the first and last row to be
processed in the Process Rows box. The format is either nRow for one sample or First
Row#–Last Row# for multiple samples.
7. Select Quan or Qual processing actions and select from active options.
Thermo Scientific
XReport User Guide
39
3
Generating a Report
Creating a Report from the Xcalibur Data System
8. Select from these XReport processing options:
• To create reports while batch processing, select the Reports check box.
• To print reports for the samples displayed in the Process Rows box, select the Print
Sample Reports option. You defined the Sample report parameters in the Reports
view of the Processing Setup window.
• Select the Print Summary Reports option to print reports for the samples that are
displayed in the Process Rows box. You defined the summary report parameters in the
Reports view of the Processing Setup window.
• Select programs, spreadsheet, and advanced options.
9. Click OK.
The data system initiates batch processing of the selected samples.
Generating a Report From Quan Browser
You can print reports for a sample set from the Quan Browser window.
 To generate a report from the Quan Browser window
1. Open the Quan Browser window.
The Open dialog box opens.
2. To select a sequence, select an SLD file with samples to use in the reports. Then, click
Open.
3. Choose View > Reports Dialog.
The Reports dialog box opens (Figure 19).
40
XReport User Guide
Thermo Scientific
3 Generating a Report
Creating a Report from the Xcalibur Data System
Figure 19. Reports dialog box in the Quan Browser window
4. To make a row of the Sample Reports area active, do the following:
a. Click the Enable box in the first row and click the Enable box again.
A check box appears in the box.
b. Click the check box.
When you click the check box again, the word Yes appears in the Enable box to
indicate that the entire row of options is selected.
5. To select the report type to print for the sample, do the following:
a. Click one of the following boxes: Stds, QCs, Unks, or Other.
A check box control appears.
b. Click the check box.
When you click the check box again, the word Yes appears in the Enable box to
indicate that reports are to be printed for all selected samples of this sample type.
6. Repeat step 4 and step 5 to print reports for needed sample types.
7. To display the Open Report Template dialog box, double-click the Report Template
Name box. Select the previously prepared report template document and click Open.
The full path and the file name of the report template (XRT) document appear in the
Report Template Name box.
Thermo Scientific
XReport User Guide
41
3
Generating a Report
Creating a Report from the Xcalibur Data System
8. To print reports for each report template document, repeat steps 4 through 7 for the next
row.
9. To make all of the sample reports you have selected for printing available, click the
Include Sample Reports check box at the bottom of the Reports dialog box. Click the
Include Summary Report check box to make summary reports available.
10. To select the samples for the report, do the following:
a. Click Select Samples to open the Select Report Samples dialog box.
The samples available from the sequence file opened in step 2 appear in the Sample
Choices box (Figure 20).
Figure 20. Select Report Samples dialog box
b. Select the samples:
• For a group of contiguous samples, click the first of the sequence, scroll until you
see the last sample in the sequence and, while holding down the SHIFT key, click
the last sample in the sequence. The data system highlights the selected sequence.
• For a group of non-contiguous samples, hold down the CTRL key while you
click each sample to select or clear it. The data system highlights the selected
samples.
c. To copy the selected samples to the Selected Samples area, click Add.
d. To return to the Reports dialog box, click OK.
42
XReport User Guide
Thermo Scientific
3 Generating a Report
Creating a Report from the Xcalibur Data System
11. To print reports, click Print Reports.
Prior to printing reports, the data system does the following:
• Displays the Printer Activity icon at the bottom of the Quan Browser window.
• Combines the data in the sample file with the selected Report Template document.
• Adds the report printing tasks to the processing queue and displays the Printing
message box.
After printing the reports, the data system returns you to the Quan Browser window.
The application saves reports created from the processing method in the same directory as the
source data files and names files. The reports use this format:
data file name template name.xxx
where
data file name is the name of the data file from which the report was generated,
template name is the name of the template used to generate the report, and
xxx is the suffix [for example, (.doc) or (.pdf )] indicating the file type.
When more than one report is generated from the same data file and template, XReport adds
a date stamp before the suffix.
Creating a Signed Report
When generating reports from Quan Browser or by batch processing in Sequence Setup, you
can create signed reports containing electronic signatures.
 To create a signed report
1. Create a report template containing the Electronic Signature table object or choose one of
the preconfigured templates supplied with XReport.
All of the preconfigured templates contain the Electronic Signature table object in the
footer.
2. Confirm that the Xcalibur Authorization Manager settings are set to require electronic
signatures at the report generating step.
See Configuring Authorization Manager Settings for Signed Reports for that process.
3. Generate the report.
Thermo Scientific
XReport User Guide
43
3
Generating a Report
Creating a Report from the Xcalibur Data System
Configuring Authorization Manager Settings for Signed Reports
To produce signed reports with electronic signatures, configure the Xcalibur Authorization
Manager Settings to require signatures. The Xcalibur data system administrators must
perform this configuration before setting permissions in either Quan Browser or Sequence
Setup.
 To set permissions for producing signed reports from Sequence Setup
1. Choose Start > All Programs > Thermo Foundation x.x > Authorization Manager.
The Authorization Manager window opens (Figure 21).
2. Expand the Home Page entry. In the directory tree, expand the Sequence Operations
entry in the permission level pane.
3. Select Batch Reprocess from the list of operations below Sequence Operations.
4. Select Signature List in the Permission Level area.
5. In the Signature List Groups area, select a group from the Available groups list and click
>> to move it to the Signatures Req’d list. The members of this group must type their
login names and passwords to electronically sign the report when it is generated. Repeat
this step for each group that you want to add to the Signatures Req’d list.
Note If you do not have any groups available or if the groups do not contain the
correct members, refer to the Foundation Administrator Guide or the Foundation Help
for information on how to create and modify groups.
6. To require a signature from the current user when the report is generated, select the
Current User Must Sign check box.
At this point, the settings should be similar to the settings in Figure 21.
44
XReport User Guide
Thermo Scientific
3 Generating a Report
Creating a Report from the Xcalibur Data System
Figure 21. Sequence Setup settings for producing signed reports
 To set permissions for producing signed reports from Quan Browser
1. Choose Start > All Programs > Thermo Foundation x.x > Authorization Manager.
The Authorization Manager window opens (Figure 22).
2. In the tree directory, expand the Quan Browser entry and expand the Print entry.
3. Select Print Reports.
4. In the Permission Level area, select Signature List.
Thermo Scientific
XReport User Guide
45
3
Generating a Report
Creating a Report from the Xcalibur Data System
5. In the Signature List Groups area, select a group from the Available groups list and click
>> to move it to the Signatures Req’d list. The members of this group are required to type
their log on names and passwords to electronically sign the report when it is generated.
Repeat this step for each group that you want to add to the Signatures Req’d list.
Note If you do not have any groups available or if the groups do not contain the
correct members, refer to the Foundation Administrator Guide for information on how
to create and modify groups.
6. To require a signature from the current user when the report is generated, select the
Current User Must Sign check box.
At this point, the settings should be similar to the settings below in Figure 22.
Figure 22. Quan Browser settings for producing signed reports
7. Click OK to save the settings and close the Authorization Manager dialog box.
46
XReport User Guide
Thermo Scientific
3 Generating a Report
Creating a Report from the Xcalibur Data System
Generating Signed Reports
Once you have a template containing an Electronic Signatures object, generate the report as
described in Generating Reports when Batch Processing a Sequence or in Generating a Report
From Quan Browser. The Xcalibur data system prompts you to type signature information
when the report is generated. Each user whose signature is required must type their log on
name and password to sign the report.
Figure 23 displays an example of a signature dialog box.
Figure 23. Example of a signature dialog box
Thermo Scientific
XReport User Guide
47
4
Configuring Objects and Sections
You can configure the following objects and sections in XReport. When you configure an
object, you can change or edit only the individual properties of that specific object. When you
add another object of the same type to the report, XReport inserts it with the default settings.
Contents
• Configuring a Repeating Section
• Configuring a Text Object
• Configuring an Annotation Object
• Configuring a Table Object
• Configuring a Bitmap Object
• Configuring a Chromatogram Object
• Configuring a Component Calibration Curve
• Configuring a Spectrum Object
• Configuring Columns
Configuring a Repeating Section
There are two different types of repeating sections in XReport:
Thermo Scientific
Qual repeating
section
Displays details related to a single Qual peak in some data. The
section is repeated for all Qual peaks found.
Quan repeating
section
Displays details related to a single Quan peak in some data or a
component configured in a processing method. The section is
repeated for all Quan peaks or components.
XReport User Guide
49
4
Configuring Objects and Sections
Configuring a Repeating Section
In the Repeating Section Properties dialog box, specify these parameters:
• Whether to exclude internal standards or target compounds from a Quan repeating
section
• The repeat order for peaks and components
• Any limit on the number of peaks to be reported
• How the individual peaks or components are split across pages in the final report
 To configure a Qual or Quan repeating section
1. Scroll to where you can see the QL or QN icon at the top left side of the template pane
and right-click the icon.
2. Choose Properties from the shortcut menu.
The Repeating Section Properties dialog box opens (Figure 24). This dialog box has three
pages: Exclude (available for a Quan repeating section only), Ordering, and Page Breaks.
Figure 24. Exclude page in the Repeating Section Properties dialog box
3. When configuring a Quan repeating section, use the Exclude page to exclude specific
elements from the process and discard any components that have not been found.
a. From the Exclude Component Type list, define the exclusions for the Quan repeating
section by selecting one of the following:
• None (no exclusions)
• Internal Standards (excludes internal standard compounds from the report)
• Target Compounds (excludes target compounds from the report)
The default selection is None.
50
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Repeating Section
b. To discard any components that have not been found, select the Show Only Found
Components check box.
For more information about the parameters on the Exclude page, see Exclude Page.
4. Click the Ordering tab to display the Ordering page (Figure 25). Use this page to change
the default ordering of processed results, defining the order of peaks and components in
either a Qual or Quan repeating section.
Figure 25. Ordering page in the Repeating Section Properties dialog box
5. To specify the order for the repeating section in the Repeating Section Ordering area,
select one of the following from the Order By list:
• RT (retention time)
• Component Name (Quan only)
• Peak Height Intensity
• Peak Area Intensity
• Peak Response Intensity (Quan only)
• Component Concentration (Quan only)
The default selection is RT.
6. To define the direction for the ordering in the Ordering Direction area, select one of these
options:
• Ascending: Sort data in ascending order
• Descending: Sort data in descending order
Thermo Scientific
XReport User Guide
51
4
Configuring Objects and Sections
Configuring a Repeating Section
7. To set a maximum number of peaks or components to order (up to 999) in the Repeating
Records Maximum area, select the Limit Number of Reported Peaks or Components
check box and select a value from the active box.
For more information about the parameters on the Ordering page, see Ordering Page.
8. Click the Page Breaks tab to display the Page Breaks page. Use this page to define the
layout for the individual peaks that appear in the Qual or Quan repeating section in the
final report (Figure 26).
Figure 26. Page Breaks page in the Repeating Section Properties dialog box
Select the option either to print each peak on its own pages or to avoid page breaks
between the individual peaks.
For more information about the parameters on the Page Breaks page, see Page Breaks
Page.
9. When you have made selections in the Repeating Sections Properties dialog box, click
OK.
10. Choose File > Save to save any changes.
52
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Text Object
Configuring a Text Object
Use the Text Object dialog box to configure a text object within the report.
To configure a text object, right-click the object and choose Properties. A dialog box opens
(Figure 27), named for the type of object you chose.
For a description of each text object, including which section each object is in, see
Appendix B, “Types of Report Objects.”
Figure 27. Data page in the Sample Header dialog box
Use the Text Object dialog box for these tasks:
• Configuring Text Fields
• Adding Custom Text Fields
• Configuring Text Attributes
• Setting Text Fonts
Thermo Scientific
XReport User Guide
53
4
Configuring Objects and Sections
Configuring a Text Object
Configuring Text Fields
Use the Data page of the Text Object dialog box to set up the text fields.
 To configure the text fields
1. In the template pane, open the Text Object dialog box by right-clicking a text object and
selecting Properties from the shortcut menu.
The Text Object dialog box opens to the Data page.
2. To include a custom title for the object, do the following:
a. Select the Include the Title on the Resolved Report check box.
The Title box appears.
b. Type a name in the Title box.
3. Select the items to include from the Available Items list.
• Select an item and click Add.
• Double-click an item.
• To move all available fields, click Add All.
The fields you have chosen appear in the selected list.
4. Select the fields in the Selected list that you do not want to include and click Remove.
5. To change the order of the selected items in the Selected list box, select an item and click
Move Up or Move Down.
6. To edit a data label, select it in the Selected list and edit its text in the Data label box.
Headings cannot be edited. However, use User Text items to add custom Headings to a
text object. See the next topic, Adding Custom Text Fields.
7. Click OK to save your changes and close the dialog box.
54
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Text Object
Adding Custom Text Fields
Add a custom text field to the text object to use as a heading or annotation. For more
information about the parameters on the Data page, see Data Page.
 To add a custom text field
1. In the template pane, right-click a text object and choose Properties from the shortcut
menu.
The Text Object dialog box opens to the Data page.
2. On the Data page of the Text Object dialog box, select <User Text> in the Available Items
list and click Add (Figure 28).
Figure 28. Adding a custom text field
3. Select <User Text> in the Selected list.
4. Type the text in the User Text box and click Apply (Figure 29).
Thermo Scientific
XReport User Guide
55
4
Configuring Objects and Sections
Configuring a Text Object
Figure 29. Customizing a text field
5. Click OK to save your changes.
Configuring Text Attributes
To configure text attributes, use the Attributes page of the Text Object dialog box (Figure 30).
For information about the parameters on the Attributes page, see Attributes Page.
Figure 30. Attributes page in the Sample Header (text object) dialog box
56
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Text Object
 To configure attributes
1. On the template pane, right-click a text object and choose Properties from the shortcut
menu.
The Text Object dialog box opens
2. Click the Attributes tab.
3. To format the text in columns, select the number of columns in the Number of Columns
area.
4. To include data labels in the object in the Horizontal Spacing area, select the Include
Data Label check box.
5. To add extra space between data labels and data, select the Include Spacing Between a
Label and Its Data check box.
6. Select a line spacing option in the Vertical Spacing area.
7. Click OK to save your changes.
Setting Text Fonts
Use the Fonts page of the Text Object dialog box to set the text fonts. You can select a font type
for the title, heading, data label, and data. For more information about the parameters on the
Fonts page for a text object, see Font Page.
 To open the Fonts page
1. On the template pane, right-click a text object and choose Properties from the shortcut
menu.
The Text Object dialog box opens.
2. Click the Font tab.
The Font page opens (Figure 31).
Thermo Scientific
XReport User Guide
57
4
Configuring Objects and Sections
Configuring a Text Object
Figure 31. Font page for a text object
 To set the font for a specific text item
1. Click a text type button.
The Font dialog box opens (Figure 32).
Figure 32. Font dialog box
2. Make the font selections and click OK.
Note The Heading Font selection is also used for User Text. See Adding Custom Text
Fields.
58
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring an Annotation Object
Configuring an Annotation Object
You can include an annotation object in any section type.
 To configure the Annotation object
1. On the template pane, right-click the annotation object in the template and choose
Properties from the shortcut menu.
The Annotation Properties dialog box opens (Figure 33).
Figure 33. Data page in the Annotation Properties dialog box
2. On the Data page, click the box on the Annotation Properties dialog box to get a cursor.
Delete the words Enter your text here and type annotation text.
3. On the Attributes page, select the number of columns to display annotation text in the
Number of Columns area (Figure 34).
For information about the parameters on the Annotation Properties – Attributes page, see
Attributes Page.
Figure 34. Attributes page in the Annotation Properties dialog box
4. In the Horizontal Alignment area, select an alignment option for the annotation text:
Left, Center, or Right.
Thermo Scientific
XReport User Guide
59
4
Configuring Objects and Sections
Configuring an Annotation Object
5. In the Vertical Spacing area, select a spacing option for the annotation text line: Single,
1.5 lines, or Double.
6. On the Font page, configure the font by setting the font parameters (Figure 35).
Figure 35. Font page in the Annotation Properties dialog box
7. Click OK to save your changes.
60
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Table Object
Configuring a Table Object
Use the Table Object wizard to configure a table object.
 To configure a table object
On the template pane, right-click a table object and choose Properties from the shortcut
menu.
The Table Object wizard opens (Figure 36), named for the type of table you selected. For a
description of each table, including the sections where a table can appear, see Appendix B,
“Types of Report Objects.”
Figure 36. Select Table Fields page of the Avalon Qual Events Table object wizard
Using the Table Object wizard, configure a table object included in the report template by
changing its properties. Use the wizard for the following:
• Choosing Table Columns
• Configuring Custom (User) Columns
• Specifying Table Sort Order
• Setting Table Fonts
• Setting Additional Table Properties
Thermo Scientific
XReport User Guide
61
4
Configuring Objects and Sections
Configuring a Table Object
Choosing Table Columns
To select the columns to include in the table, use the Select Table Fields page of the Table
Object wizard. The columns available for a table appear in alphabetical order, but can be
added to a report template in any order. Your column options to select depend on the type of
table object you are configuring.
For more information about the parameters on this page, see Select Table Fields Page.
 To select and arrange table columns
1. On the template pane, right-click a table object and choose Properties from the shortcut
menu.
The Table Object – Select Table Fields page opens.
2. To include a custom title for the table, type the title in the Title box and select the
Include the Title on the Resolved Report check box.
3. To select the columns to include in the table from the Available Columns list, do one of
the following:
• Select an item and click Add.
• Double-click an item.
• To move all available fields, click Add All.
The fields you have chosen appear in the selected list.
Note All table objects have user columns available for selection. Use the user
columns to define formulas for calculating table data and to display the results in the
table. For more information on user columns, see Configuring Custom (User)
Columns.
4. Type the column heading text in the Column Header box.
5. To arrange the order of the selected columns in the Selected Columns list, click Move Up
or Move Down.
6. Click Next to go to the next page of the wizard.
62
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Table Object
Configuring Custom (User) Columns
The Configure User Column Parameters page of the Table Object wizard appears only if you
chose to include custom User Columns in the table (on the Select Table Fields page). Use this
page to define parameters for each User Column in the table.
For more information about the parameters on this page, see Configure User Column
Parameters Page.
You can use the User Columns to define mathematical formulas for calculating table data.
Formulas obey the following syntax rules:
• The formula must begin with an equals sign (=).
• Each table column is represented by an alphabetic letter identifier (A, B, C, and so on)
that appears beneath the column title in the Table object.
• The mathematical expressions are kept simple, for instance, type A+B, A–B, A*B, or A/B.
This is true as well for any of the functions listed in the Functions list of the Edit Formula
dialog box. Formulas are computed on a row-by-row basis. That is, if you type =A+B as a
formula in a User Column, the results are computed as follows:
a. The Xcalibur data system adds the value in the first row of column A to the value in
the first row of column B and displays the result in the first row of the User Column.
b. The data system adds the value in the second row of column A to the value in the
second row of column B and displays the result in the second row of the User
Column.
c. The data system continues this process until all rows of the table have been
completed.
Figure 37. Configure User Column Parameters page
The data system displays the User Columns you have selected for the table in the User
Columns list.
Thermo Scientific
XReport User Guide
63
4
Configuring Objects and Sections
Configuring a Table Object
 To define each user column
1. On the template pane, right-click a table object and choose Properties from the shortcut
menu.
The Table Object – Select Table Fields page opens.
2. Add a User Column to the table and then click Next.
The Configure User Column Parameters page appears.
3. Select a user column in the User Columns list.
4. Type a title for the user column in the User Column Title box.
5. Click Edit Formula in the User Column Formula area.
The Edit Formula dialog box opens (Figure 38). For more information about the creating
formulas, see Edit Formula For Dialog Box and Adding a Formula to a Table.
Figure 38. Edit Formula For dialog box
6. For the selected column, edit a formula for the user column and configure valid fields and
associated functions. Below the functions list is a brief description of each function at the
bottom of the dialog box.
7. To enter a formula in the Formula box, select a function from the Functions list and click
Add Function or type directly in the box.
When you click Add Function, a configuration dialog box opens for that function. For
example, for the Average function, choose the column to average.
8. To return to the Configure User Column Parameters page, click OK.
64
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Table Object
9. Click Next to go to the next page of the wizard.
Specifying Table Sort Order
Use the Set Table Sort Filter page of the Table Object wizard to define a method for sorting
table data when you generate a report (Figure 39).
Figure 39. Set Table Sort Filter page
XReport displays the columns selected for the table in a list on the left side of the dialog box.
Set the order for the table using any of the selected columns. Use up to three columns for
ordering the table: The first order takes precedence over the second order, which takes
precedence over the third order.
 To define a method for ordering data in the table
1. On the template pane, right-click a table object and choose Properties from the shortcut
menu.
The Table Object dialog box opens.
2. Click Next until the Set Table Sort Filter page opens.
3. To select a variable to provide the first level of sorting for the table, double-click the item
in the column list to move it to the First Order box.
4. To define how to sort the First Order column data in the Current Sorting Configuration
area, select one of the following options:
• Ascending: Sort data in ascending order.
• Descending: Sort data in descending order.
Thermo Scientific
XReport User Guide
65
4
Configuring Objects and Sections
Configuring a Table Object
5. To define the second and the third variable used to order table columns, repeat the first
two steps.
6. To move to the Font page, click Next.
Setting Table Fonts
Use the final page of the Table Object wizard, the Set Table Font page, to set the font type for
elements in the table (Figure 40).
Figure 40. Set Table Font page
 To define the font for a specific element
1. On the template pane, right-click a table object and choose Properties from the shortcut
menu.
The Table Object dialog box opens.
2. Click Next until the Set Table Fonts page opens.
3. Click the appropriate button.
The Font dialog box opens.
4. Select font options and click OK.
Repeat step 3 and step 4 as needed for other table elements.
66
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Table Object
Setting Additional Table Properties
After completing the Table Object wizard, additional Properties dialog boxes appear for the
following table objects:
• Library Search Graphics Table Properties
• Library Search Results Table Properties
• Sample Table Properties
• Spectrum List Table Properties
• Electronic Signature Table Properties
Library Search Graphics Table Properties
Use the Library Search Graphics Table Properties dialog box to set the number of hits to be
displayed. For more information about the parameters in this dialog box, see Library Search
Graphics Table Properties Dialog Box.
 To open the Library Search Graphics Table Properties dialog box
1. On the template pane, right-click the Library Search Graphic Table object and choose
Properties from the shortcut menu.
The Library Search Graphics Table – Select Table Fields page opens.
2. Make changes as necessary.
3. Click Finish.
The Library Search Graphics Table Properties dialog box opens (Figure 41).
Figure 41. Library Search Graphics Table Properties dialog box
Thermo Scientific
XReport User Guide
67
4
Configuring Objects and Sections
Configuring a Table Object
 To set the parameters in the Library Search Graphics Table Properties dialog box
1. Type the total number of search results to display in the table in the Max # of hits to
display box.
2. To display column labels, select the Display Column Labels check box.
3. To show the peak spectrum, select the Show Peak Spectrum check box.
4. Click OK.
5. Choose File > Save to save any changes made.
Library Search Results Table Properties
Use the Library Search Results Table Properties dialog box to specify the maximum number of
hits to be displayed. For more information about the parameters in this dialog box, see Library
Search Results Table Properties Dialog Box.
 To open the Library Search Results Table Properties dialog box
1. On the template pane, right-click the Library Search Results Table object and choose
Properties from the shortcut menu.
The Library Search Results Table wizard opens to the Select Table Fields page.
2. Make the necessary changes.
3. Click Finish.
The Library Search Results Table Properties dialog box opens (Figure 42).
Figure 42. Library Search Results Table Properties dialog box
 To set the parameters in the Library Search Results Table Properties dialog box
1. Type the total number of search results to display in the table in the Max # of hits to
display box.
2. Click OK.
68
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Table Object
Sample Table Properties
When you click Finish on the last page of the Sample Table object wizard (in a Quan
Repeating section), the Sample Table (Quan Results) Properties dialog box opens (Figure 43).
For more information about the parameters in the Sample Table Properties dialog box, see
Sample Table Properties Dialog Box.
Figure 43. Sample Table Properties dialog box
 To set parameters in the Sample Table Properties dialog box
1. In the template pane, open the Sample Table (Quan Results) Properties dialog box as
follows:
a. In a Quan Repeating section, right-click a Sample Table object and choose
Properties.
The Sample Table wizard opens.
b. Complete the wizard and click Finish.
2. To include standards, QCs, blanks, or unknowns in the Sample table, select the
corresponding check boxes.
3. Click OK.
Thermo Scientific
XReport User Guide
69
4
Configuring Objects and Sections
Configuring a Table Object
Spectrum List Table Properties
When you click Finish on the last page of the Spectrum List Table object wizard (in a Quan or
Qual Repeating section), the Spectrum List Table Properties dialog box opens (Figure 44).
Figure 44. Options page in the Spectrum List Table Properties dialog box
This dialog box has four pages:
• Options Page
• Plot Type Page
• Enhancement Page
• Display Page
Options Page
For information about the parameters on the Spectrum List Table Properties – Options page,
see Options Page for a Spectrum Object.
 To set the parameters on the Options page
1. Open the Options page of the Spectrum List Table Properties dialog box as follows:
a. Right-click a Spectrum List Table object and choose Properties.
b. Complete the wizard and click Finish.
c. Click the Options tab.
2. Select or clear the Use Processing Method Properties for Plot Type and Enhancement
Info check box (Figure 44).
When you select this check box, the application hides the parameters on the Plot Type
and Enhancement pages.
70
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Table Object
3. To use the ISTD peak when repeating on a component, select the corresponding check
box. You must select the Use Processing Method Properties for Plot Type check box to
activate the ISTD peak option. (This option only applies when this object is used in a
Quan repeating section. It is not available when the Spectrum List Table is used in a Qual
Repeating Section.)
4. Click OK.
Plot Type Page
For information about the parameters on the Spectrum List Table Properties – Plot Type
page, see Plot Type Page for a Spectrum Object.
 To set parameters on the Plot Type page
1. Open the Plot Type page of the Spectrum List Table Properties dialog box as follows:
a. On the template pane, right-click a Spectrum List Table object and choose
Properties.
b. Complete the wizard and click Finish.
The Spectrum List Table Properties dialog box opens to the Options page.
c. Clear the Use Processing Method Properties for Plot Type and Enhancement Info
check box.
d. Click the Plot Type tab.
2. Type the mass range to display in the Mass range box. Separate the values by a dash:
lowmass–highmass. Type * to display the entire range (Figure 45).
Figure 45. Plot Type page in the Spectrum List Table Properties dialog box
3. Select the type of detector that you are using from the Detector list: MS or PDA.
4. Type the time range to display in minutes in the Time box. Type the values separated by a
dash: firsttime–secondtime. Type * to display the entire range.
Thermo Scientific
XReport User Guide
71
4
Configuring Objects and Sections
Configuring a Table Object
5. Enter the scan filter to use with the data in the Scan Filter list.
6. Type or search in the Demo File box for the name of the file you to use as a
demonstration file for the settings in the template.
7. Click OK.
Enhancement Page
For information about the parameters on the Spectrum List Table Properties – Enhancement
page, see Enhancement Page for a Spectrum List Table.
 To set parameters on the Enhancement page
1. Open the Plot Type page of the Spectrum List Table Properties dialog box as follows:
a. On the template pane, right-click a Spectrum List Table object and choose
Properties.
b. Complete the wizard and click Finish.
The Spectrum List Table Properties dialog box opens to the Options page.
c. Clear the Use Processing Method Properties for Plot Type and Enhancement Info
check box.
d. Click the Enhancement tab.
2. To add smoothing to the data, select the Enable check box in the Smoothing area
(Figure 46).
Figure 46. Enhancement page in the Spectrum List Table Properties dialog box
3. To select a type of smoothing, select Boxcar or Gaussian from the Type list.
4. Select the number of smoothing points (odd numbers from 3 to 15) to be used in the
Points list.
5. To refine the data, select the Enable check box in the Refine area.
72
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Table Object
6. To define an area for drawing a peak, type the number of seconds on either side of the
specified point. A reasonable initial value for the window is the peak width in seconds.
7. Type the cutoff point for peaks in the Noise Threshold box. To show all peaks in the
spectrum, type 0. Return to this box to eliminate more noise. Slowly increase the value
until the noise peaks are eliminated. Use this setting to eliminate peaks generated from
baseline noise.
8. To turn on a first baseline region that is used for assessing background subtraction, select
one or both of the Time Range check boxes. Type the time range to use in the box.
The Background Subtraction area shows details of background subtraction if applied to
the spectrum. The background contribution is determined by averaging the scans from
one or two baseline regions.
9. Click OK.
Display Page
For information about the parameters on the Spectrum List Table Properties – Display page,
see Display Page for a Spectrum List Table.
 To set parameters on the Display page
1. Open the Plot Type page of the Spectrum List Table Properties dialog box as follows:
a. On the template pane, right-click a Spectrum List Table object and choose
Properties.
b. Complete the wizard and click Finish.
The Spectrum List Table Properties dialog box opens to the Options page.
c. Click the Display tab.
2. To display all peaks in the spectrum, select the All Peaks check box or, to limit the display
to the strongest peaks, type a value in the Top box (Figure 47).
Figure 47. Display page in the Spectrum List Table Properties dialog box
Thermo Scientific
XReport User Guide
73
4
Configuring Objects and Sections
Configuring a Table Object
3. To order the data by mass or by intensity, select the appropriate option in the Order By
area.
4. Type a value representing the relative abundance range of mass peaks to include in the
table in the Normalization area.
5. Click OK.
Electronic Signature Table Properties
Use the Electronic Signature table object to specify whether the Created By signature block
appears before or after the Signed By signature block. For more information about the
parameters in the Electronic Signature Properties dialog box, see Electronic Signature
Properties Dialog Box.
 To configure the Electronic Signature table object
1. On the template pane, right-click the Electronic Signature table object and choose
Properties from the shortcut menu.
The Electronic Signature Properties dialog box opens (Figure 48).
Figure 48. Electronic Signature Properties dialog box
2. To display the Created By signature block before the Signed By signature block, select the
Created By First option in the Signature Information Order area or, to display the Signed
By block before the Created By block, select the Signed By First option.
3. Click OK.
74
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Bitmap Object
Configuring a Bitmap Object
For information about the parameters on the Bitmap file page of the Bitmap Properties dialog
box, see Bitmap Properties Dialog Box.
 To configure the bitmap object
1. On the template pane, right-click the bitmap object and choose Properties from the
shortcut menu.
The Bitmap Properties dialog box opens (Figure 49).
Figure 49. Bitmap Properties dialog box
2. Enter the name of the image file. Choose one of the following image file formats: BMP,
DIB, GIF. JPG, or ICO.
3. To set the size of the bitmap image to the size of the original file, click Reset to Original
Size.
4. Click OK.
Thermo Scientific
XReport User Guide
75
4
Configuring Objects and Sections
Configuring a Chromatogram Object
Configuring a Chromatogram Object
For information about the parameters in the Chromatogram Properties dialog box, see
Chromatogram Properties Dialog Box and Wizard.
 To configure a chromatogram report object
1. On the template pane, right-click the object and choose Properties from the shortcut
menu.
The Chromatogram Properties dialog box opens (Figure 50).
Figure 50. Chromatogram Properties dialog box
2. Click Configure.
The Chromatogram wizard opens.
The Chromatogram wizard guides you through the process of changing the properties of
a chromatogram report object contained in the report template.
The XReport Chromatogram wizard displays different pages and contains different
options depending on the type of section containing the chromatogram and the selections
for the Chromatogram Type and Peak Type properties.
76
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Chromatogram Object
Table 1 lists the wizard pages and their order. Each wizard page is described in a topic below.
See the appropriate topic to learn how to complete each wizard page.
Table 1. Possible Chromatogram wizard pages, given the section type, in order of appearance
Page in the XReport Chromatogram wizard
Nonrepeating section
(Chromatogram Type)
Repeating section
Raw
Quan
Quan
Qual
Use Container Supplied Quan Peak Page
Qual
X
Use Container Supplied Qual Peak Page
X
a
Xb
Select Chromatogram Type Page
X
X
X
X
Select Demo File Page
X
X
X
X
X
Xa
Xb
Select the Quan Peak Page
X
Xc
Select Component
Select the Qual Peak Page
X
Select Mass Spectrum Quan Traces to Display Page
X
Configure How the Peak Window Will Look Page
X
X
Xa
Xb
Xa
Xb
X
X
d
Xe
Select Plot Details Page
X
X
Select Enhance Details Page
X
Xd
Xe
Select Chromatogram Style Page
X
X
X
X
X
Select the Peak Labeling Page
X
X
X
X
X
Select How the Chromatogram Axes Will Be Configured Page
X
X
X
X
X
Configure Normalization for the Chromatogram Page
X
X
X
X
X
a
This page is only displayed when you select the Explicitly Specify the Chromatogram’s Behavior check box on the Use Container Supplied Quan
peak page.
b
This page is only displayed when you select the Explicitly Specify the Chromatogram’s Behavior check box on the Use Container Supplied Qual
peak page.
c
This page is only displayed when you select the Explicitly Specify the Chromatogram’s Behavior check box on the Use Container Supplied Quan
peak page and the Quan Results Based Chromatogram option on the Select Chromatogram Type page.
d
This page is only displayed when you select the Explicitly Specify the Chromatogram’s Behavior check box on the Use Container Supplied Quan
peak page and the Raw File Based Chromatogram option on the Select Chromatogram Type page.
e
This page is only displayed when you select the Explicitly Specify the Chromatogram’s Behavior check box on the Use Container Supplied Qual
peak page and the Raw File Based Chromatogram option on the Select Chromatogram Type page.
Note When you select the All Qual Peaks option for a Qual Result Based
Chromatogram, the wizard page order is the same as for the Qual Result Based
Chromatogram Type without the Select Qual Peak page.
Thermo Scientific
XReport User Guide
77
4
Configuring Objects and Sections
Configuring a Chromatogram Object
When the chromatogram report object to be configured is contained in a Quan or Qual
repeating section, the default Chromatogram Type is Quan or Qual Result Based with no
explicit Peak Type set. The next page of the wizard (Use Container Supplied Quan (or Qual)
asks you to confirm this.
Use Container Supplied Quan Peak Page
Use the Use Container Supplied Quan Peak page of the Chromatogram wizard to display the
details of the internal standard associated with the implicit Quan peak or ignore the implicit
Quan peak and explicitly configure the chromatogram type.
For more information about the parameters on this page, see Use Container Supplied Quan
Peak. For more information about the Chromatogram wizard, see Configuring a
Chromatogram Object.
 To explicitly configure a chromatogram type in a Quan repeating section
1. In a Quan repeating section, do the following:
a. Right-click a Chromatogram object and choose Properties.
b. Click Configure.
The Use Container Supplied Quan Peak page of the Chromatogram wizard opens.
By default, the Use Quan Peak’s Associated Internal Standard and Explicitly Specify
check boxes are clear.
2. To display the details of the internal standard associated with the implicit Quan peak,
select the Use Quan Peak’s Associated Internal Standard check box in the Implicit
Quan Peak Options area (Figure 51).
Figure 51. Use Container Supplied Quan Peak page
78
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Chromatogram Object
3. To ignore the implicit Quan peak supplied by the container and explicitly configure the
chromatogram type, select the Explicitly Specify the Chromatogram’s Behaviour check
box in the Explicit Chromatogram Type area.
4. Click Next to go to the next page of the wizard.
Use Container Supplied Qual Peak Page
Use the Use Container Supplied Qual Peak page of the Chromatogram wizard to ignore the
implicit Qual peak and explicitly configure the chromatogram type.
For information about the parameters on the Use Container Supplied Qual Peak page, see Use
Container Supplied Qual Peak. For more information about the Chromatogram wizard, see
Configuring a Chromatogram Object.
 To explicitly configure a chromatogram type in a Qual repeating section
1. In a Qual repeating section, do the following:
a. Right-click a Chromatogram object and choose Properties.
b. Click Configure.
The Use Container Supplied Qual Peak page of the Chromatogram wizard opens. By
default, the Explicitly Specify the Chromatogram’s Behaviour check box is clear.
2. To ignore the implicit Qual peak supplied by the container and explicitly configure the
chromatogram type, select the Explicitly Specify the Chromatogram’s Behaviour check
box in the Explicit Chromatogram Type area (Figure 52).
Figure 52. Use Container Supplied Qual Peak page
3. Click Next to go to the next page of the wizard.
One of these pages opens:
• The Select Chromatogram Type page opens if you selected the Explicitly Specify the
Chromatogram’s Behaviour check box.
• The Select Demo File page opens if you left the Explicitly Specify the
Chromatogram’s Behaviour check box clear.
Thermo Scientific
XReport User Guide
79
4
Configuring Objects and Sections
Configuring a Chromatogram Object
Select Chromatogram Type Page
Use the Select Chromatogram Type page of the Chromatogram wizard to set the
chromatogram type for the chromatogram report object being configured in the report
template.
For information about the parameters on the Select Chromatogram Type page, see Select
Chromatogram Type. For more information about the Chromatogram wizard, see
Configuring a Chromatogram Object.
 To open the Select Chromatogram Type page
1. Right-click a Chromatogram object and choose Properties.
2. Click Configure.
In a non-repeating section, the Select Chromatogram Type page opens. In a Qual or
Quan repeating section, the Use Container Supplied Qual or Quan Peak page of the
Chromatogram wizard opens.
3. In the Use Container Supplied Qual or Quan Peak page, select the Explicitly Specify the
Chromatogram’s Behaviour check box. Then click Next.
The Select Chromatogram Type page opens.
4. Define the chromatogram type in the Chromatogram Type area (Figure 53) by selecting
one of the following options:
• Raw file based chromatogram
• Quan results based chromatogram
• Qual results based chromatogram
Figure 53. Select Chromatogram Type page
When you select the Qual Results Based Chromatogram option, the Show All Qual Peaks
in Result File check box becomes available.
80
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Chromatogram Object
5. If you select a Qual results based chromatogram in step 4, select the Show All Qual
Peaks in Result File check box to display all Qual peaks in a resolved Xcalibur report.
This is a Peak Type setting.
6. Click Next to go to the next page of the wizard, the Select Demo File Page.
Select Demo File Page
Use the Select Demo File page of the Chromatogram wizard to associate a demonstration file
(RAW or RST file) with a chromatogram report object. A demo file shows how the
chromatogram report object appears in the report template and in wizard page previews with
its current settings when applied to real Xcalibur data.
For information about the parameters on the Select Demo File page of the Chromatogram
wizard, see Select Demo File.
 To open the Select Demo File page of the Chromatogram wizard
1. In a Qual, Quan, or non-repeating section, right-click a Chromatogram object and
choose Properties.
2. Click Configure.
• In a non-repeating section, the Select Chromatogram Type page opens. Go to step 4.
• In a Qual or Quan section, the Use Container Supplied Qual or Quan Peak page of
the Chromatogram wizard opens.
3. On the Use Container Supplied Qual or Quan page, do one of the following:
a. Select the Explicitly Specify the Chromatogram’s Behaviour check box. Click
Next.
The Select Chromatogram Type page opens.
b. Go to step 4.
–or–
• Leave the Explicitly Specify the Chromatogram’s Behaviour check box clear. Click
Next.
The Select Demo File page opens.
4. For an explicit chromatogram type, select one of the options in the Chromatogram Type
area and click Next.
The Select Demo File page opens.
Thermo Scientific
XReport User Guide
81
4
Configuring Objects and Sections
Configuring a Chromatogram Object
The appearance of the Select Demo File page depends on the chromatogram type and the
peak type settings chosen in the Chromatogram Type page of the wizard.
• Raw File Based Chromatogram
• Quan Result Based Chromatogram
• Qual Result Based Chromatogram – Explicit Peak
• Qual Result Based Chromatogram – All Available Peaks
Raw File Based Chromatogram
When configuring a chromatogram report object based on a raw data file, use this procedure
to select a file.
 To select a demo file for a raw file-based Chromatogram
1. To associate a demo file with the chromatogram report object, select the Use Demo File
check box on the Select Demo File page (Figure 54).
Figure 54. Select a raw data file based chromatogram
2. To search for the required demo file (RAW file), click the browse button in the Demo File
area.
The Select Raw Data File dialog box opens.
3. Select a demo file and click Open.
4. Click Next to go to the next page of the wizard.
82
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Chromatogram Object
Quan Result Based Chromatogram
When configuring a chromatogram report object based on Quan results, use this procedure to
select a file.
For information about the parameters on the Select Demo File page of the Chromatogram
wizard, see Select Demo File.
 To select a demo file for a Quan-based chromatogram
1. To associate a demo file with the chromatogram report object, select the Use Demo File
check box on the Select Demo File page (Figure 55).
Figure 55. Selecting a Quan result based chromatogram
2. To search for the required demo file (RAW file), click the browse button in the Demo File
area.
The Select Raw Data File dialog box opens.
3. Select a demo file and click Open.
4. To specify the peak to demo, click the browse button in the Demo Peak area.
The Select a Peak dialog box opens.
5. Select a peak from the list located in the result file.
6. Click OK to apply the selection and return to the wizard.
7. Click Next to go to the next page of the wizard.
Thermo Scientific
XReport User Guide
83
4
Configuring Objects and Sections
Configuring a Chromatogram Object
Qual Result Based Chromatogram – Explicit Peak
When configuring a chromatogram report object based on Qual Results with an explicit Peak
Type required, use this procedure to select a file.
For information about the parameters on the Select Demo File page of the Chromatogram
wizard, see Select Demo File.
 To select a demo file for a Qual-based chromatogram
1. Select the Use Demo File check box on the Select Demo File page to associate a demo file
with the chromatogram report object (Figure 56).
Figure 56. Selecting an explicit peak in a Qual result based chromatogram
2. To search for the required demo file (RAW file), click the browse button in the Demo File
area.
The Select Raw Data File dialog box opens.
3. Select a demo file and click Open.
4. To specify the peak to demo, click the browse button in the Demo Peak area.
The Select a Peak dialog box opens.
5. Select a peak from the list located in the result file.
6. Click OK to apply the selection and return to the wizard.
7. Click Next to go to the next page of the wizard.
84
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Chromatogram Object
Qual Result Based Chromatogram – All Available Peaks
When configuring a chromatogram report object that is Qual result based with Show all Qual
peaks set (peak type setting), use this procedure to select a file.
For information about the parameters on the Select Demo File page of the Chromatogram
wizard, see Select Demo File.
 To select a demo file for all available peaks
1. To associate a demo file with the chromatogram report object, select the Use Demo File
check box on the Select Demo File page.
2. To search for the required demo file (RAW file), click the browse button in the Demo File
area.
The Select Raw Data File dialog box opens.
3. Select a demo file and click Open.
4. Click Next to go to the next page of the wizard.
Select the Quan Peak Page
Use the Select The Quan Peak page (Figure 57) of the Chromatogram wizard to explicitly
specify the Quan peak (component) for a Quan results based chromatogram report object.
Figure 57. Select the Quan Peak page
For more information about the parameters on this page, see Select the Quan Peak. For more
information about the Chromatogram wizard, see Configuring a Chromatogram Object.
 To specify a Quan peak
1. Open the Select the Quan Peak page of the Chromatogram wizard as follows:
Thermo Scientific
XReport User Guide
85
4
Configuring Objects and Sections
Configuring a Chromatogram Object
a. In a Qual, Quan, or non-repeating section, right-click a Chromatogram object and
choose Properties.
b. Click Configure.
• In a non-repeating section, the Select Chromatogram Type page opens. Go to
step 1e.
• In a Qual or Quan section, the Use Container Supplied Qual or Quan Peak page
of the Chromatogram wizard opens.
c. On the Use Container Supplied Qual or Quan Peak page, select the Explicitly
Specify the Chromatogram’s Behaviour check box.
d. Click Next.
The Select Chromatogram Type page opens.
e. Select the Quan Results Based Chromatogram Option and click Next.
The Select Demo File page opens.
f.
Follow the instructions in Quan Result Based Chromatogram. Then click Next.
2. To specify the Quan peak, do one of the following:
• Type the text name of the component in the Quan Peak box.
–or–
• Select the Quan peak from the component list of a processing method as follows:
i.
Click the browse button.
The Select Processing File dialog box opens.
ii. Select a processing method file.
The Select Component dialog box opens. For more information about this
dialog box, see Select Component.
iii. Select a component. The list includes the components defined in the selected
processing method.
iv. Click OK to return to the wizard. The Next button is active only after you have
entered a component.
3. To use the associated internal standard of the Quan peak instead of the Quan peak, select
the Use Quan Peak’s Associated Internal Standard check box.
Note Select a valid peak for XReport to resolve the report correctly. If the file used to
resolve the report does not contain the specified peak, the Xcalibur data system
removes the chromatogram from the resolved report.
4. Click Next to go to the next page of the wizard, the Select Mass Spectrum Quan Traces
To Display page.
86
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Chromatogram Object
Select the Qual Peak Page
Use the Select the Qual Peak page of the Chromatogram wizard to explicitly specify the Qual
peak for a Qual Results Based chromatogram report object.
 To specify a Qual peak
1. Open the Select the Qual Peak page of the Chromatogram wizard as follows:
a. In a Qual, Quan, or non-repeating section, right-click a Chromatogram object and
choose Properties.
b. Click Configure.
• In a non-repeating section, the Select Chromatogram Type page opens. Go to
step 1e.
• In a Qual or Quan section, the Use Container Supplied Qual or Quan Peak page
of the Chromatogram wizard opens.
c. On the Use Container Supplied Qual or Quan Peak page, select the Explicitly
Specify the Chromatogram’s Behaviour check box.
d. Click Next.
The Select Chromatogram Type page opens.
e. Select the Qual Results Based Chromatogram Option, leave the Show All Qual
Peaks in Result File clear, and click Next.
The Select Demo File page opens.
f.
Follow the instructions in Qual Result Based Chromatogram – Explicit Peak. Then
click Next.
The Select The Qual Peak page of the Chromatogram wizard opens.
2. To specify the Qual peak in the Qual Peak To Display area, type the appropriate retention
time in the Retention time box (Figure 58).
The Next button is active only after you have selected a component in the Retention time
box. Set the retention time to any time within the start and end peak boundaries.
Note Select a valid retention time for XReport to resolve the report correctly. If the
file used to resolve the report does not contain the specified retention time, the
chromatogram is removed from the resolved report.
Thermo Scientific
XReport User Guide
87
4
Configuring Objects and Sections
Configuring a Chromatogram Object
Figure 58. Select the Qual Peak page
3. Click Next to go to the next page of the wizard, the Configure How The Peak Window
Will Look page.
Select Mass Spectrum Quan Traces to Display Page
Use the Select Mass Spectrum Quan Traces to Display page of the Chromatogram wizard to
configure the Quan Traces property for MS data. (This page is only applicable to MS data and
is inactive for other detector types.)
When a chromatogram report object displays a Quan component, there are potentially six
traces associated with the component, as follows:
• Quan Mass (QM) Trace (the actual component peak)
• Up to five Confirmation Ions
Note The Ion Ratio Confirmation settings on this page apply only to GC-MS data.
Select any of these possible six traces to be displayed in the Chromatogram or display all
available traces of the component.
For parameter information, see Select Mass Spectrum Quan Traces to Display.
88
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Chromatogram Object
 To specify a Quan trace
1. Select one of the following from the Quan Traces list (Figure 59):
• Quan Mass (QM Trace)
• First Qualifier Ion
• Second Qualifier Ion
• Third Qualifier Ion
• Fourth Qualifier Ion
• Fifth Qualifier Ion
• All Quan Traces
Figure 59. Select Mass Spectrum Quan Traces To Display page
2. To normalize all the traces the same way, select the Common Normalization for All
Traces check box.
3. Click Next to go to the next page of the wizard.
Thermo Scientific
XReport User Guide
89
4
Configuring Objects and Sections
Configuring a Chromatogram Object
Configure How the Peak Window Will Look Page
Use the Configure How the Peak Window Will Look page of the Chromatogram wizard to
control the display of the baselines (Quan and Qual peaks) and the peak window width
(active for Qual peak and Quan peak in some cases) when a peak from a result file is set to
display in a Chromatogram.
 To open the Configure How the Peak Window Will Look page
1. Right-click a Chromatogram object and choose Properties.
2. Click Configure.
Depending on the section type, one of these pages opens.
• In a Quan repeating section, the Use Container Supplied Quan Peak page opens. Go
to step 3.
• In a Qual repeating section, the Use Container Supplied Qual Peak page opens. Go
to step 3.
• In a non-repeating section, the Select Chromatogram Type page opens. Go to step 4.
3. Make the appropriate selection and click Next.
If you selected the Explicitly Specify the Chromatogram’s Behaviour check box, the Select
Chromatogram Type page opens. Otherwise, the Select Demo File page opens.
4. Depending on whether the Select Demo File page or the Select Chromatogram Type page
opens, do one the following:
If the Select Demo File page opens, do the following:
a. Follow the instructions in Select Demo File Page. Then click Next.
The Select Mass Spectrum Quan Traces To Display page opens.
b. Select the appropriate trace. Then click Next.
The Configure How the Peak Window Will Look page opens.
If the Select Chromatogram Type page opens, select either the Quan or Qual Results
Based Chromatogram option:
a. Select the Quan Results Based Chromatogram option. Then click Next.
The Select Demo File page opens.
b. Follow the instructions in Select Demo File Page. Then click Next.
The Select The Quan Peak page opens.
c. Follow the instructions in Select the Quan Peak Page. Then click Next.
The Select Mass Spectrum Quan Traces To Display page opens.
90
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Chromatogram Object
d. Follow the instructions in Select Mass Spectrum Quan Traces to Display Page. Then
click Next.
The Configure How The Peak Window Will Look page opens.
–or–
a. Select the Qual Results Based Chromatogram option. Then click Next.
The Select Demo File page opens.
b. Follow the instructions in Select Demo File Page. Then click Next.
If you left the Show All Qual Peaks in Result File check box clear, the Select The
Qual Peak page opens. Otherwise, the Configure How the Peak Window Will Look
page opens.
c. Follow the instructions in Select the Qual Peak Page. Then click Next.
The Configure How The Peak Window Will Look page opens.
 To specify the width of the peak window and whether the peak window shows an
integrated peak
1. Select the Show Baselines check box to display the baselines for a Chromatogram peak
(Quan or Qual) (Figure 60). When you clear this check box, the XReport application
does not display the baselines, peak tracing, or full peak annotation.
Figure 60. Configure How the Peak Window Will Look page
2. To manually override the default value of the peak window width (the default is peak
width plus 1 min), select the Set Custom Window Width check box. Use the Custom
width edit box to type a value for the window width (min). The default value for this field
is 0.75 min.
The Set Custom Window Width check box is unavailable if you selected the Show All
Qual Peaks in Result File check box in the Select Chromatogram Type page.
3. Click Next to go to the next page of the wizard.
Thermo Scientific
XReport User Guide
91
4
Configuring Objects and Sections
Configuring a Chromatogram Object
Select Plot Details Page
Use the Select Plot Details page of the Chromatogram wizard to configure how the plot is
drawn when raw data is being displayed in a chromatogram report object.
For more information about the parameters on this page of the wizard, see Select Plot Details.
 To configure a plot on a raw file-based chromatogram
1. Open the Select Plot Details page as follows:
a. Right-click a Chromatogram object and select Properties.
b. Click Configure.
• In a Qual repeating section, the Use Container Supplied Qual Peak page opens.
• In a Quan repeating section, the Use Container Supplied Quan Peak page opens.
• In a non-repeating section, the Select Chromatogram Type page opens.
c. In a Qual or Quan repeating section, do the following:
i.
Select the Explicitly Specify the Chromatogram’s Behaviour check box on the
Use Container Supplied Qual Peak page or the Use Container Supplied Quan
Peak page, respectively. Then click Next.
The Select Demo File page opens.
ii. Follow the instructions in Select Demo File Page. Then click Next.
The Select Chromatogram Type page opens.
d. For a Qual, Quan, or non-repeating section, select the Raw File Based
Chromatogram option on the Select Chromatogram Type page.
The Select Plot Details page opens.
2. Select the type of detector used to generate the raw data file from the Detector list in the
Plot Type area (Figure 61). The type of detector determines the available traces.
92
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Chromatogram Object
Figure 61. Select Plot Details page
3. Type in the offset of the start of a chromatogram trace in the Delay box. This box is
available only when you select a non-MS detector type, such as UV or PDA.
4. From the Filter list, select the processing applied to a subset of the scans in a raw data file.
This option is available only when you select MS detectors. The list displays scan filter
options stored in the selected demo RAW file. If no demo file is selected, type a valid filter
string.
5. From the Trace lists, select the following:
a. From the first list, select a basic chromatogram type, for example, TIC.
b. From the second list, select a logical operator: + or –. Selecting an operator makes the
third list active.
c. From the third list, select a second chromatogram type to add to, or subtract from,
the first trace (for example, Mass Range). The list features valid plot types.
The list of valid traces depends on the detector used to generate the data as follows.
Thermo Scientific
Detection
Valid traces
MS
Mass Range, TIC, and Base Peak
Analog
Analog 1–4
A/D card
A/D Card Ch 1–4
PDA
Wavelength Range, Total Scan, and Spectrum Maximum
UV
Channel A–D
XReport User Guide
93
4
Configuring Objects and Sections
Configuring a Chromatogram Object
6. Depending on the type of chromatogram you select, do the following:
a. If you choose MS detector types, specify the mass or mass range for Mass Range or
Base Peak Traces in the Mass list as follows:
If you choose Base Peak ± Mass Range or Mass Range ± Mass Range plot
combinations in step 4 above, the identification page displays a Mass (m/z) field for
each trace type.
b. If you choose other detector types, use this field to specify the wavelength or
wavelength range for the chromatogram. If you use a plot combination such as
Wavelength Range + Wavelength Range, an additional Wavelength (nm) box appears
for you to specify the second wavelength range.
The valid range depends on the configured detector. The format is
Low Mass/Wavelength–High Mass/Wavelength. Separate entries by a dash with no
spaces. For example, for the range m/z 123 through 456, type: 123–456.
7. Specify the lower and upper time limits for the plot (in minutes to 1 decimal place) in the
Time Range box. Separate entries by a dash with no spaces.
Type * (the default setting) to display the complete time range for the plot.
8. Select the Fixed Scale check box to set the y-axis maximum to a specific value.
Selecting the Fixed Scale check box activates the Normalize To area on the final page of
the wizard and the box where you type the maximum y-axis value on the current page.
9. Select the Autofilter check box to repopulate a chromatogram view with the following:
• A plot showing the chromatogram without any scan filters
• Plots for each scan filter applied to the chromatogram
10. Click Next to go to the next page of the wizard.
94
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Chromatogram Object
Select Enhance Details Page
Use the Select Enhance Details page of the Chromatogram wizard to enhance a
chromatogram plot by applying smoothing (Figure 62).
For more information about the parameters on this wizard page, see Select Enhance Details.
Figure 62. Select Enhance Details page
 To enhance a chromatogram
1. To open the Select Enhance Details page, open the Select Plot Details page as described in
step 1 of To configure a plot on a raw file-based chromatogram. Then click Next.
2. To apply smoothing to the chromatogram, select the Enable Smoothing check box.
3. To select a type of smoothing, select Boxcar or Gaussian from the Type list in the
Smoothing Details area.
4. Select the number of smoothing points (odd numbers from 3 to 15) to use in the Points
list.
5. Click Next to go to the next page of the wizard.
Select Chromatogram Style Page
Use the Select Chromatogram Style page of the Chromatogram wizard to determine the
appearance of the chromatogram. For more information about the parameters on this page of
the wizard, see Select Chromatogram Style.
This page of the Chromatogram wizard is available for all of the report sections and for all of
the chromatogram types whether implicit or explicit.
Thermo Scientific
XReport User Guide
95
4
Configuring Objects and Sections
Configuring a Chromatogram Object
 To define styles for a chromatogram
1. To define the plotting style, select one of the following options in the Plotting area
(Figure 63):
• Point To Point: Display point-to-point peak profiles.
• Stick: Display the chromatogram as vertical lines.
Figure 63. Select Chromatogram Style page
2. To define the plotting arrangement, select one of the following options in the
Arrangement area:
• Stack (2D): Stack plots vertically (same x axis) with no overlap (own y axis).
• Overlay (3D): Overlay plots vertically (same x axis). Use the 3D arrangement when
you selected Multiple Quan Traces on the Select Mass Spectrum Quan Traces page of
the wizard.
3. When you select an Overlay (3D) arrangement, set the following in the active 3D area:
a. To set the Elevation angle (from 0 to 60 degrees) for the 3D plot style, adjust the
Elevation slider bar. Either drag the Elevation slider bar or click the Elevation slider
left or right arrows until you reach the desired angle.
b. To set the skew angle for the 3D plot style, adjust the Skew slider bar. Either drag the
Skew slider or click the Skew slider left or right arrow until you reach the desired
angle (from 0 to 45 degrees).
c. To add a backdrop to 3D plots, select the Draw Backdrop check box.
4. Click Next to go to the next page of the wizard.
96
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Chromatogram Object
Select the Peak Labeling Page
Use the Select The Peak Labeling page of the Chromatogram wizard to choose the type and
style of peak labels. From this page, click Advanced to open a dialog box where you set
advanced labeling options.
This page of the Chromatogram wizard is available for all of the report sections and for all of
the chromatogram types whether implicit or explicit.
To label peaks in reports, see these topics:
• Basic Peak Labeling Options
• Advanced Labeling Options for a Quan Peak
• Advanced Labeling Options for a Qual Peak
Basic Peak Labeling Options
To set up the basic peak labeling options, follow this procedure.
 To select peak labeling options
1. To label peaks with the time in minutes in the chromatogram plot, select the Retention
Time check box in the Label With area (Figure 64). The data system displays RT before
the label.
Figure 64. Select the Peak Labeling page
2. To label peaks with the number of mass scans at peak maximum, select the Scan Number
check box. The data system displays S# before the label.
3. To label the base peak in the mass spectrum of the chromatogram peak, select the Base
Peak check box. The data system displays BP before the label.
Thermo Scientific
XReport User Guide
97
4
Configuring Objects and Sections
Configuring a Chromatogram Object
4. To label peaks with the signal-to-noise ratio at the peak maximum, select the
Signal-to-Noise check box. The data system displays SN before the label.
5. To label peaks with flags providing supplemental information about the peak data, select
the Flags check box. When a peak is saturated, the data system displays an S above the
peak.
6. To label peaks with the integrated area of the peaks, select the Area check box. The data
system displays AA before the label when detected automatically or MA when detected
manually.
7. To label peaks with the apex height, select the Height check box. The data system
displays AH before the label when detected automatically or MH when detected
manually.
8. To offset a label from its normal position to avoid conflict with another label, select the
Offset check box in the Label Styles area. Type the number for the offset (for the number
of characters) in the Size box.
9. To use vertical, rather than horizontal, labels, select the Rotated check box.
10. To place a rectangular outline around each peak label, select the Boxed check box.
11. To limit the labeling of peaks to those exceeding the specified percentage of the base peak,
set the Label threshold box to a percent value between 100 and 0.
12. To open the Advanced Quan Peak Annotation Options dialog box or the Advanced Qual
Peak Annotations dialog box, click Advanced.
• If you are configuring a Quan peak, see the next topic Advanced Labeling Options
for a Quan Peak.
• If you are configuring a Qual peak, see Advanced Labeling Options for a Qual Peak.
13. Click Next to go to the next page of the wizard, the Select How the Chromatogram Axes
Will Be Configured page.
Advanced Labeling Options for a Quan Peak
If are configuring a Quan peak in any of the report sections, the Advanced Quan Peak
Annotation Options dialog box opens when you click Advanced on the Select The Peak
Labeling page of the Chromatogram wizard. When you set Advanced Labeling options, the
data system labels only the Quan peak.
For more information about the parameters in this dialog box, see Advanced Quan Peak
Annotation Options. For information about the Select Peak Labeling page of the
Chromatogram wizard, see Select the Peak Labeling Page.
98
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Chromatogram Object
 To set advanced Quan peak annotation options
1. To activate advanced Quan peak options, select the Use Advanced Quan Peak
Annotation check box (Figure 65).
Figure 65. Advanced Quan Peak Annotation Options dialog box
2. To label the Quan peak with the component name, select the Component Name check
box in the Custom Peak Labels area.
3. To label the Quan peak with the component key, select the Component Key check box.
4. To specify a custom label, select the Custom Text Label check box. Type a custom label
for the Quan peak in the active box.
5. To label the Quan peak with the retention time, select the Retention Time check box.
If you selected Retention Time check box in the Label With area on the Select The Peak
Labeling page of the Chromatogram wizard, this check box is already selected. Clearing
the Retention Time check box in the Advanced Quan Peak Annotation Options dialog
box also clears the Retention Time check box on the Select The Peak Labeling page.
6. In the RT Labeling area, select the unit for the retention time from the Units list:
Minutes or Seconds.
7. Select the number of decimal places to be displayed in the retention time label from the
Decimals list: 1, 2, or 3.
8. Click OK to save the changes and close the dialog box.
9. To continue configuring the properties of the Chromatogram object, click Next on the
Select the Peak Labeling page of the Chromatogram wizard.
Thermo Scientific
XReport User Guide
99
4
Configuring Objects and Sections
Configuring a Chromatogram Object
Advanced Labeling Options for a Qual Peak
If are configuring a Qual peak in any of the report sections, the Advanced Qual Peak
Annotation Options dialog box opens when you click Advanced on the Select The Peak
Labeling page of the Chromatogram wizard. When you set Advanced Labeling options, the
data system labels only the Qual peak.
For more information about the parameters in this dialog box, see Advanced Qual Peak
Annotation Options. For information about the Select Peak Labeling page of the
Chromatogram wizard, see Select the Peak Labeling Page.
 To set advanced Qual peak annotation options
1. To activate the advanced Qual peak options, select the Use Advanced Qual Peak
Annotation check box (Figure 66).
Figure 66. Advanced Qual Peak Annotation Options dialog box
2. To specify a custom label, select the Custom Text Label check box in the active Custom
Peak Labels area. Type a custom label for the Qual peak in the box.
3. To label the Qual peak with the retention time, select the Retention Time check box.
If you selected Retention Time check box in the Label With area on the Select The Peak
Labeling page of the Chromatogram wizard, this check box is already selected. Clearing
the Retention Time check box in the Advanced Qual Peak Annotation Options dialog
box also clears the Retention Time check box on the Select The Peak Labeling page.
4. In the RT Labeling area, select the type of unit for the retention time from the Units list.
5. Select the number of decimal places to be displayed in the retention time label from the
Decimals list.
6. Click OK to save the changes and close the Advanced Qual Peak Annotation Options
dialog box.
100
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Chromatogram Object
7. To continue configuring the properties of the Chromatogram object, click Next on the
Select the Peak Labeling page of the Chromatogram wizard.
Select How the Chromatogram Axes Will Be Configured Page
Use the Axes page of the Chromatogram wizard to set chromatogram axes label and display
options. For more information about the parameters on this page, see Select How the
Chromatogram Axes Will Be Configured.
This page of the Chromatogram wizard is available for all of the report sections and for all of
the chromatogram types whether implicit or explicit. For information about the
Chromatogram wizard, see Configuring a Chromatogram Object.
 To specify the appearance for chromatogram axes
1. To set the name of the x axis in the X Axis area, type it in the Name box (Figure 67).
Figure 67. Select How the Chromatogram Axes Will Be Configured page
2. To offset the axis label from the chromatogram, select the Offset check box.
3. To split the chromatograms into two or more separate graphs with equal time ranges,
select the Split Time Range check box.
4. Type the number of split time range graphs displayed for each chromatogram in the
Divisions box. This box is active only when you select the Split Time Range check box.
5. In the Y Axis area, select either the From Detector option in the Label area to use the
detector-specific label for the y axis (stored in the RAW file) or the Custom option to use
your own label.
6. To specify the setting for the y-axis units, select either Absolute or Relative in the Units
area.
Thermo Scientific
XReport User Guide
101
4
Configuring Objects and Sections
Configuring a Chromatogram Object
7. If you have chosen to use a Custom label for the y axis, type a y-axis name in the Name
box.
8. To offset the axis label from the chromatogram, select the Offset check box.
9. To go to the next page of the wizard, click Next.
Configure Normalization for the Chromatogram Page
Use the Normalization page of the Chromatogram wizard to select the normalization (y-axis
scaling) method used for the chromatogram. For more information about the parameters on
this page, see Configure Normalization For The Chromatogram.
This page of the Chromatogram wizard is available for all of the report sections and for all of
the chromatogram types whether implicit or explicit. For information about the
Chromatogram wizard, see Configuring a Chromatogram Object.
 To specify normalization options
1. Define the normalizing method in the Normalize Method area (Figure 68) by selecting
one of the following options:
• Auto Zero: Optimize the y axis automatically for each chromatogram.
• Intensity Range (%): Set the range manually. Type the minimum and maximum
intensity required for the y axis. The valid range is –200.00 to +200.00%.
Figure 68. Configure Normalization for the Chromatogram page
102
XReport User Guide
Thermo Scientific
4 Configuring Objects and Sections
Configuring a Component Calibration Curve
2. Define a peak to use for normalization in the Normalize Each Plot area by selecting one
of the following options:
• Largest Peak in Subsection: Normalize each split time range to the largest peak in
the division.
• Largest Peak in Selected Time Range: Normalize the spectrum to the largest peak in
the displayed time range.
• Largest Peak in All Times: Normalize the spectrum to the largest peak in the entire
chromatogram.
The Normalize Each Plot area is active when you select a raw file-based chromatogram and
you have selected the Fixed Scale check box on the Plot Type Page (see Plot Type Page).
Finishing the Chromatogram Wizard
To change selections in the Chromatogram wizard, click Back to go back through the
previous pages. When the page is done, click Finish.
For more information about the Chromatogram wizard, see Configuring a Chromatogram
Object.
Configuring a Component Calibration Curve
To configure a Component Calibration Curve object, right-click it and choose Properties
from the shortcut menu. The Component Calibration Curve Properties dialog box opens
(Figure 69).
Figure 69. Component Calibration Curve Properties dialog box
If your report has a nonrepeating Component Cal Curve report object, use the Component
Calibration Curve wizard to specify the name of the component.
 To specify the name of the component
1. To search for the required processing method file (PMD), click the browse button. A
dialog box opens. By default, the wizard searches in the Xcalibur\methods folder.
Thermo Scientific
XReport User Guide
103
4
Configuring Objects and Sections
Configuring a Spectrum Object
2. To open a file, select it and click Open.
The Select Component dialog box opens (Figure 70).
Figure 70. Select Component dialog box
3. Select the required component from the list available for the chosen processing method
and click OK. The Component Name page opens, displaying the specified name of the
component in the box.
4. Click OK.
Configuring a Spectrum Object
Use the Spectrum Properties dialog box to configure a Spectrum object in an XReport
template. Use the XReport Spectrum Properties pages to do the following:
• Configure Options
• Configure Plot Types
• Enhance Spectra
• Configure Spectrum Style
• Configure Spectrum Peak Labels
• Configure Axes
• Select Normalization Options
• Specify a Peak
To configure the Spectrum report object, make selections on each page based on these
corresponding procedures.
For more information about the Spectrum Properties dialog box, see Spectrum Properties
Dialog Box.
 To open the Spectrum Properties dialog box
Right-click a Spectrum object and choose Properties from the shortcut menu.
104
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Spectrum Object
Configure Options
To configure Spectrum options, use the Options page of the Spectrum Properties dialog box.
For information about the parameters on this page, see Options Page for a Spectrum Object.
 To configure Spectrum options
1. In the template pane, right-click a Spectrum object and choose Properties.
The Spectrum Properties dialog box opens to the Options page (Figure 71).
Figure 71. Options page
2. To use processing method properties for Plot type and Enhance information, select the
Use Processing Method Properties for Plot Type and Enhance Info check box.
When you select this check box, the options on the Plot type and Enhance pages are not
available.
3. To use the Internal Standard peak when repeating on a component, select the Use ISTD
Peak When Repeating on Component check box. This option is only available when
you select the Use processing method properties for plot type and enhance info check
box.
Thermo Scientific
XReport User Guide
105
4
Configuring Objects and Sections
Configuring a Spectrum Object
Configure Plot Types
When raw data is being displayed in a Spectrum report object, use the Plot Type page of the
Spectrum Properties dialog box to configure how the plot is drawn. For more information
about the parameters on this page, see Plot Type Page for a Spectrum Object.
If you selected the Use processing method properties for plot type and enhance info check box
on the Options page, this page is blank.
 To configure plot types
1. In the template pane, right-click a Spectrum object and choose Properties.
The Spectrum Properties dialog box opens.
2. Click the Plot Types tab.
3. Specify the mass or mass range for MS detector types in the Mass Range box (Figure 72).
Figure 72. Plot type page
For PDA detector types, use this parameter to specify the wavelength or wavelength range
(nm). Type the lower and upper limits for the plot. Separate entries by a dash (no spaces).
To use the entire mass/wavelength range, type an asterisk:*.
4. Select the type of detector used to generate the raw data file from the Detector list. Valid
types are MS and PDA. The type of detector determines the available Plot types.
5. Select the time of the mass scan in the host chromatogram in the Time box. Type a new
value or range as required.
6. From the Scan Filter list, select a filter or type a new one in the box using the scan filter
format.
7. Click the browse button in the Demo File box and select the file. This box shows the path
and filename of the demo file (RAW) used to generate the selected plot.
106
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Spectrum Object
Enhance Spectra
Use the Enhance page of the Spectrum Properties dialog box to apply smoothing (the refine
algorithm) to the spectrum and set background subtraction time ranges. For information
about the parameters on this page, see Enhance Page for a Spectrum Object.
If you selected the Use processing method properties for plot type and enhance info check box
on the Options page, this page is blank.
 To define spectrum enhancement options
1. To turn on spectrum smoothing, select the Enable check box in the Smoothing area
(Figure 73).
Figure 73. Enhance page
2. To select a type of smoothing, select Boxcar or Gaussian from the Type list.
3. Type the number of points for spectrum smoothing in the Points list. Type an odd
number from 3 (minimum smoothing) to 15 (maximum smoothing).
4. To turn Refine spectrum enhancement on or off, select the Enable check box in the
Refine area.
5. To apply the refining algorithm and specify a length of time for mass chromatograms,
type a value in the Window box indicating the number of seconds on either side of the
specified point. A reasonable initial value is the peak width in seconds.
6. To eliminate peaks generated from baseline noise, type 0 (this shows all peaks in the
spectrum) in the Noise threshold box. After you close this box to display the results, you
can open the box again to increase the value until noise peaks are eliminated.
Thermo Scientific
XReport User Guide
107
4
Configuring Objects and Sections
Configuring a Spectrum Object
7. To specify the first baseline region used for assessing the background, select the Time
Range 1 check box in the Background Subtraction area. Type the time range in the box.
The Background Subtraction area shows details of background subtraction when this
value has been applied to the spectrum. The background contribution is determined by
averaging the scans from one or two baseline regions.
8. To specify the second baseline region used for assessing the background, select the Time
Range 2 check box. Type the time range in the box.
Configure Spectrum Style
To set parameters that determine the appearance of the spectrum, use the Style page of the
Spectrum Properties dialog box (Figure 74). For information about the parameters on this
page, see “Style Page for a Spectrum Object.
Figure 74. Style page
 To set plotting options
To define the plotting style, select one of the following options in the Plotting area:
• Automatic: Choose the graphic style based on the data acquisition method used for
the active spectrum.
• Point to Point: Display a point-to-point peak profile.
• Stick: Display spectral mass peaks as vertical lines.
• Shade: Display the spectrum as a shaded representation of intensity in each amu
band for the active spectrum.
108
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Spectrum Object
Configure Spectrum Peak Labels
To choose the type and style of peak labels, use the Labels page of the Spectrum Properties
dialog box. For information about the parameters on this page, see Labels Page for a Spectrum
Object.
 To configure spectrum peak labels
1. To label spectrum peaks with m/z value, select the Mass check box (Figure 75) in the
Label With area.
Figure 75. Labels page
2. To label spectrum peaks with flags, select the Flags check box. Flags provide supplemental
information about the peak data. For example, when a peak is saturated, the Xcalibur
data system displays an S above the peak.
3. To specify that other spectrum peaks are labeled relative to that value, type a value in the
Relative to box.
4. Type the number of decimal places to be used in the labels in the Decimals box.
5. To move a label from its normal position to avoid conflict with another label, select the
Offset check box in the Label Styles area. Choose the number for the offset (for the
number of characters) in the Size box.
6. To use vertical, rather than horizontal, labels, select the Rotated check box.
7. To place a rectangular outline around each peak label, select the Boxed check box. This
option does not apply to flags.
8. To limit the labeling of peaks to those exceeding the specified percentage of the base peak,
type a percent in the Label Threshold% box.
Thermo Scientific
XReport User Guide
109
4
Configuring Objects and Sections
Configuring a Spectrum Object
Configure Axes
To set spectrum axis labels and display options, use the Axis page of the Spectrum Properties
dialog box. For information about the parameters on this page, see Axis Page for a Spectrum
Object.
 To configure axis options
1. Type the x-axis name in the Name box in the X Axis area (Figure 76).
Figure 76. Axis page in the Spectrum Properties dialog box
2. To offset the axis label from the spectrum, select the Offset check box.
3. To split the spectra into two or more separate graphs with equal ranges, select the Split
Mass Range check box. Type the number of split range graphs displayed for each
spectrum in the Divisions box.
4. Select one of the following options in the Y Axis area under Label:
• From Detector: Use the detector-specific label for the y axis (stored in the RAW file).
• Custom: Use your own label.
5. To specify the setting for the y-axis units, select the Absolute or the Relative option.
6. Type a y-axis name in the Name box.
7. To move the axis label away from the spectrum, select the Offset check box.
110
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring a Spectrum Object
Select Normalization Options
Use the Normalization page of the Spectrum Properties dialog box to select normalization
options for spectrum plots. For information about the parameters on this page, see
Normalization Page for a Spectrum Object.
 To select normalization options
1. To define the normalization method, select one of the following options in the Normalize
Method area (Figure 77).
• Auto Zero: Automatically optimize the y axis.
• Intensity Range (%): Set the y-axis range. Use the box to type the minimum and
maximum intensity. The valid range is –200.00 to +200.00%.
Figure 77. Normalization page
2. To define the peak that the spectrum is normalized to, select one of the following options
in the Normalize To area:
• Largest Peak in Subsection: Normalize each split mass range to the largest peak in
the division.
• Largest Peak in Mass Range: Normalize the spectrum to the largest peak in the
displayed mass range.
• Largest Peak in Scan: Normalize the spectrum to the largest peak in the entire
spectrum.
Thermo Scientific
XReport User Guide
111
4
Configuring Objects and Sections
Configuring a Spectrum Object
3. To define how multiple scans are normalized, select one of the following options in the
Normalize Multiple Scans area:
• Individually: Normalize each mass plot individually.
• All the Same: Normalize all mass plots equally.
Specify a Peak
To explicitly specify the peak to display in the Spectrum report object, use the Peak page of
the Spectrum Properties dialog box. For information about the parameters on this page, see
Peak Page for a Spectrum Object.
 To specify a peak
1. To specify a specific type of peak to appear in the spectrum, select the Enable Explicit
Peak Type check box (Figure 78).
Figure 78. Peak page
2. To specify the peak type to appear in the Spectrum, do one of the following:
• To specify a Quan peak, select the Quan Peak option.
• To specify a Qual peak, select the Qual Peak option.
3. Depending on the peak type, do one of the following:
• For a Quan peak, type a name for the component in the Peak Identification area.
• For a Qual peak, type the retention time (any time between the start time and end
time of the peak) or click the browse button.
The Select a Peak dialog box opens.
4. Select a peak and click OK to close the dialog box.
112
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring Columns
Configuring Columns
Use the Column Properties dialog box to configure a column object.
For information about the parameters in this dialog box, see Column Properties Dialog Box.
For information about the column objects in the XReport application, see Columns Object.
After you have configured a column object, drag objects to the columns.
 To configure a column object
1. To configure a column object, right-click it and choose Properties from the shortcut
menu.
The Column Properties dialog box opens (Figure 79).
Figure 79. Column Properties dialog box
2. Select the number of columns to use in the Number of Columns area.
3. If the column object is in a nonrepeating section, configure the options in the Repeat type
area. Select Sample Repeat or Summary Repeat by clicking the appropriate option. For
an example of using these options, see Set Sample Repeat Options and Set Summary
Repeat Options.
The Repeat type options are not available in Quan repeating or Qual repeating sections.
4. If you selected Summary repeat, select whether you would like to repeat from Left to
Right or from Top to Bottom by clicking the appropriate option.
5. To start each sample on a new row, select the New Sample New Row check box.
Thermo Scientific
XReport User Guide
113
4
Configuring Objects and Sections
Configuring Columns
Set Sample Repeat Options
Use the Sample repeat option to present information relating to a raw data file of a single
sample and the results created when processing it. Use this repeat type as a layout tool to
indicate where objects are placed on the page of the resolved report.
Note The Sample repeat and Summary repeat options are applicable only in
nonrepeating report sections. These options are not available in other types of report
sections.
For example, when you place a 2-column object in a nonrepeating section, add a Quan-based
chromatogram to the left column and combine or resolve the template using data of 1 sample
with 4 components, the report appears as in Figure 80.
Figure 80. 2-column object behavior in a nonrepeating section, with a chromatogram object
Resolve
Chrom
Chrom 1
Chrom 2
Chrom 3
Chrom 4
Sample 1
In another example, place a 2-column object in a nonrepeating section and add a
chromatogram and a text object to the left column and a spectrum to the right column.
Combine (resolve) the template using data of 1 sample with 4 components. All objects in the
column are repeated for a single sample, and the report appears as in Figure 81.
Figure 81. 2-column object behavior in a nonrepeating section, with chromatogram and spectrum
objects
Chrom
Text
Spec
Resolve
Chrom 1
Spec 1
Text
Chrom 2
Spec 2
Text
Sample 1
Chrom 3
Spec 3
Text
Chrom 4
Spec 4
Text
114
XReport User Guide
Thermo Scientific
4
Configuring Objects and Sections
Configuring Columns
Set Summary Repeat Options
Use the Summary repeat option to present information relating to all of the samples in an
Xcalibur sequence and the results created when processing them.
Note The Sample repeat and Summary repeat options are applicable only in
nonrepeating report sections. These options are not available in other types of report
sections.
For example, when you place a 2-column object in a nonrepeating section, add a
chromatogram (in either column), and combine (resolve) the template using data of 2 samples
with 4 components each, the report appears as in Figure 82.
Figure 82. 2-column object behavior in a nonrepeating section, with a chromatogram, Left to right
option
Resolve
Chrom
Left to
right
Chrom 1
Chrom 2
Chrom 3
Chrom 4
Chrom 1
Chrom 2
Chrom 3
Chrom
Chrom 34
Sample 1
Sample 2
If you choose instead the Top To Bottom option, the report appears as in Figure 83.
Figure 83. 2-column object behavior in a nonrepeating section, with a chromatogram, Top to
bottom option
Resolve
Chrom
Top to
bottom
Chrom 1
Chrom 3
Chrom 2
Chrom 4
Chrom 1
Chrom 3
Chrom 2
Chrom 4
Sample 1
Sample 2
If you place an object in the right column and leave the left column empty, XReport combines
the data as if the object is placed in the left column, as illustrated above.
If you place an object in each column, the Xcalibur data system ignores the Top to bottom
and Left to right options. The report appears as in Figure 84.
Thermo Scientific
XReport User Guide
115
4
Configuring Objects and Sections
Configuring Columns
Figure 84. 2-column object behavior in a nonrepeating section, with chromatogram and spectrum
objects
Resolve
Chrom
Text
Spec
Chrom 1
Spec 1
Chrom 2
Spec 2
Chrom 3
Spec 3
Chrom 4
Spec 4
Chrom 1
Spec 1
Chrom 2
Spec 2
Chrom 3
Spec 3
Chrom 4
Spec 4
Sample 1
Sample 2
Text
116
XReport User Guide
Thermo Scientific
A
Sample XReport Template Files
The Xcalibur data system provides several sample XReport templates for you to use when
creating your own reports. This appendix provides details of these report templates stored at
the following location:
Xcalibur\templates
Table 2 describes the standard report templates.
Table 2. Sample XReport template files (Sheet 1 of 2)
Report template file name
Data source files
Description/Use
CalibrationFile.xrt
raw data file (RAW)
calibration file (XCAL)
Generates sample reports with
calibration data.
CompCalReport.xrt
result file (RST)
raw data file (RAW)
processing method (PMD)
sequence file (SLD)
Generates summary reports showing
processing method details, calibration
curve, and sample table (Quan
results). Can be used from Batch
Processing and Quan Browser in the
Xcalibur data system.
result file (RST)
raw data file (RAW)
Generates Qual sample reports that
display Qual peaks and library search
details. Use from Batch Processing in
the Xcalibur data system.
IonRatioConfirmation.xrt
result file (RST)
raw data file (RAW)
processing method (PMD)
Generates a Quan sample report
containing non-graphical ion ratio
confirmation results.
IonRatioConfirmationGraphical.xrt
result file (RST)
raw data file (RAW)
processing method (PMD)
Generates Quan sample reports
containing graphical ion ratio
confirmation results. These report
templates take advantage of the
explicit confirmation ion settings
available when using the XReport
Chromatogram wizard.
CompCalReport_Avalon.xrt
CompCalReport_Genesis.xrt
CompCalReport_ICIS.xrt
CustLibrSearRept.xrt
LibrarySearchReport.xrt
IonRatioConfirmationGraphical
Simple.xrt
IonRatioConfirmationGraphical
Stacked.xrt
Thermo Scientific
XReport User Guide
117
A
Sample XReport Template Files
Table 2. Sample XReport template files (Sheet 2 of 2)
Report template file name
Data source files
Description/Use
PeakIntegration.xrt
result file (RST)
raw data file (RAW)
processing method (PMD)
Generates a Quan sample report
containing peak integration data
generated using Genesis.
ProcessingMethod.xrt
processing method (PMD)
Generates processing method sample
reports. Use from Processing Method
Setup in the Xcalibur data system.
QualPeakReport.xrt
result file (RST)
raw data file (RAW)
Generates a simple Qual sample
report. Use for data from Batch
Processing in the Xcalibur data
system.
QuanPeakResults_ESTD.xrt
result file (RST)
raw data file (RAW)
processing method (PMD)
Generates Quan sample reports. Can
be used from Batch Processing or
Quan Browser in the Xcalibur data
system.
QuantifySampleReport.xrt
result file (RST)
raw file (RAW)
processing method (PMD)
sequence file (SLD)
Generates a Quan summary report
containing chromatograms and a
Quan summary table.
SequenceReport.xrt
sequence file (SLD)
Generates a summary report showing
the sequence being processed. Use for
batch processing in the Xcalibur data
system. Refer to the Xcalibur Data
Acquisition and Processing User Guide.
SpectrumCandidates.xrt
result file (RST)
raw data file (RAW)
Generates a Quan summary report
(only applicable to GC/MS). Use
from batch processing or Quan
Browser in the Xcalibur data system.
ProcessingMethod_Avalon.xrt
ProcessingMethod_Genesis.xrt
ProcessingMethod_ICIS.xrt
QuanPeakResults_ISTD.xrt
QuanResults.xrt
QuanSimple.xrt
118
XReport User Guide
Thermo Scientific
B
Types of Report Objects
The following tables list the report objects you can place onto a report template: text objects,
table objects, graphic objects, and formatting objects. The tables briefly describe an object and
indicate in which sections the object is available.
Contents
• Text Object Descriptions
• Table Object Descriptions
• Graphic Object Descriptions
• Formatting Object Descriptions
Text Object Descriptions
Table 3 describes the text objects.
Table 3. Text objects (Sheet 1 of 3)
Text object
Data source file
Description
Section placement
Annotation
none
Displays static text. Use this object for
adding headings or comments.
Any
Component Ident/Detect/Cal
Settings
PMD
Displays the peak detection, calibration, Quan repeating
and integration parameters for a single
component.
Component ISTD Settings
Summary
PMD
Displays a summary of the configuration Quan repeating
details for each component’s ISTD
configuration in the Quantitation
section of a processing method.
Component Settings Summary
PMD
Displays a short list of peak detection
and integration parameters for a
specified component.
Thermo Scientific
Quan repeating
XReport User Guide
119
B
Types of Report Objects
Text Object Descriptions
Table 3. Text objects (Sheet 2 of 3)
Text object
Data source file
Description
Component Sys Suit/Flags
Settings
PMD
Displays Calibration, Quantitation, and Quan repeating
Detection flags and system suitability
parameters to specify automatic
chromatographic checks that assign a
pass/fail qualification to a target peak.
Instrument Method
RAW
Displays general instrument method
parameters. This text object does not
have any editable properties.
Nonrepeating
Processing Method – General
PMD
Displays general processing method
parameters, including date created, date
modified, baseline definition,
chromatography type (LC/GC).
Nonrepeating
Processing Method – Qual
PMD
Displays Qual processing method
parameters, including the following:
Nonrepeating
•
•
•
•
•
•
Quan ISTD Peak Summary
RST or PMD
Section placement
Identification
Spectrum Enhancement
Detection Options
Advanced Detection Options
Library Search Constraints
Library Search Options
Displays summary parameters for a
Quan component’s associated ISTD
peak. For example:
Quan repeating
• ISTD NameBase Line
• Actual RTSignal To Noise
• Response Expected RT
Quan Peak Summary
RST or PMD
Displays Component Peak summary
parameters, for example:
•
•
•
•
•
120
XReport User Guide
Quan repeating,
header or footer
Actual RT% Diff
Calculated AmountArea Ratio
Response Area ISTD
Base Line Specified Amount
Signal To Noise
Thermo Scientific
B
Types of Report Objects
Table Object Descriptions
Table 3. Text objects (Sheet 3 of 3)
Text object
Data source file
Description
Section placement
Report Info
none
Displays general information about the
report itself.
Header or footer
Sample Header
RAW or SLD
Displays general information about a
sample, for example:
Nonrepeating,
header or footer
•
•
•
•
•
Sample Name
Operator
Sample Type
Run Time
Acquisition Date
Table Object Descriptions
Table 4 describes the table objects.
Table 4. Table objects (Sheet 1 of 3)
Table object
Data source file
Description
Avalon Qual Events Table
PMD
Displays details of the Avalon peak
Nonrepeating
detection algorithm events configured in
the Qual section of a processing method.
Avalon Quan Events Table
PMD
Displays details of the Avalon peak
detection algorithm events configured
for a component in the Quan section of
a processing method.
Quan repeating
Calibration File Table
XCAL, SLD,
RAW
Displays component names, response
factors, and %RSD of standards.
Nonrepeating
Component Cal Level Table
PMD
Shows the calibration levels, amounts,
and acceptance test values defined for a
specific component.
Quan repeating, or
Nonrepeating
Component QC Level Table
PMD
Shows the QC levels, amounts, and
percent test values defined for a specific
component.
Quan repeating
Component Spectrum Table
PMD
Displays m/z and intensity data for up to Quan repeating
fifty spectrum peaks. The data system
uses this data to identify components
when the Spectrum Peak Identification
Method is specified in the processing
method. This object is used only for GC
chromatography.
Thermo Scientific
Section placement
XReport User Guide
121
B
Types of Report Objects
Table Object Descriptions
Table 4. Table objects (Sheet 2 of 3)
Table object
Data source file
Description
Dilution Factor Table
RST
Displays the names of calibration levels Nonrepeating
and dilution factors for components that
are not internal standards.
Electronic Signature
none
Displays up to five electronic signatures
in table format. This object is not
editable in XReport.
Any
Ion Ratio Confirmation Results
Table
RST
Displays the following data about each
qualifier ion, based on search criteria
specified in the processing method:
Quan repeating
•
•
•
•
Section placement
Mass
Coelution test pass/fail
Ion Ratio Test Target (%)
Ion Ratio Test Window (%)
IRC Settings Table
PMD
Displays up to five qualifier ions within Quan repeating
tolerance limits to confirm the detection
of a target analyte. The ions are specified
in the processing method.
IRC Summary Table
RST
Displays all of the information for the
confirmation ion for each component.
Most useful when the Name column is
the first column.
Nonrepeating
Lib. Search Graph Table
RST, RAW
Displays the graphical results from a
library search on a Qual peak (for
example, Hit Spectrum, Delta, and
Compound Structure).
Qual repeating
Lib. Search Results Table
RST, RAW
Displays the results of a Library Search
on a Qual peak (for example, Name,
Probability, RSI, SI, CAS#, Area).
Qual repeating
Proc. Method Programs Table
PMD
Displays a list of programs to be run by
the Xcalibur data system after the
sample has been analyzed and the data
has been processed. The data system
runs the programs in the order listed.
Nonrepeating
Proc. Method Sample Reports
Table
PMD
Displays Sample Report information for Nonrepeating
each sample in a sequence, including
Template name and the Report file
name. Specify Sample Type in the
processing method.
122
XReport User Guide
Thermo Scientific
B
Types of Report Objects
Table Object Descriptions
Table 4. Table objects (Sheet 3 of 3)
Table object
Data source file
Description
Section placement
Proc. Method Summary Reports
Table
PMD
Displays Summary Reports to be issued
after processing the samples in a
bracketed or non-bracketed sequence.
Nonrepeating
Qual Peak Table
RST
Displays information specific to
chromatogram peaks detected by
qualitative analysis for a single sample.
Nonrepeating
Qual Summary Table
SLD, RST
Displays information specific to
chromatogram peaks detected by
qualitative analysis for all samples in a
sequence.
Nonrepeating
Quan Peak Table
RST, PMD
Displays information specific to
Nonrepeating
component peaks detected by
quantitative analysis for a single sample.
Quan Summary Table
SLD, RST,
PMD
Displays information specific to
component peaks detected by
quantitative analysis for all samples in a
sequence.
Sample Table (Quan Results)
SLD
Displays system suitability and flag
Quan repeating
processing results for each component in
a sample, and repeats this information
for every sample in a sequence.
Sequence Table
SLD
Displays the contents of the sequence
file.
Nonrepeating
Spectrum Candidate Results
Table
RST
Displays the results of spectrum
candidate processing for GC/MS data
on a per component basis. Parameters
include the following:
Nonrepeating
•
•
•
•
Spectrum List Table
Thermo Scientific
RST, RAW
Nonrepeating
Forward fitStart RT
Reverse fitEnd RT
MatchArea
Found RTHeight
Displays the text equivalent of the
Qual or Quan
Spectrum graphic object. The Spectrum repeating
List Table contains a header and lists
values for the m/z, intensity, and relative
intensity.
XReport User Guide
123
B
Types of Report Objects
Graphic Object Descriptions
Graphic Object Descriptions
Table 5 describes the graphic objects.
Table 5. Graphic objects
Graphic object
Data source file
Bitmap
BMP, DIB, GIF, Displays a graphical image, for example, Any
JPG, or ICO
a company logo.
Chromatogram
RAW, RST
Displays a graphical trace that can show Qual or Quan repeating
either raw data or processed data for any and nonrepeating
acquisition device supported by the data sections
system (MS, Analog, A/D Card, PDA,
UV).
Component Cal Curve
RST, PMD
Shows the calibration curve used to
quantify a specified component. The
curve is specified in the processing
method as response versus amount of
standard.
In a Quan repeating
section, the Component
Cal Curve object repeats
on the component peaks.
Displays a graphical spectrum that can
show either raw data or processed data
for any acquisition device supported by
the data system (mass spectrometer,
PDA detector).
Qual or Quan repeating
and nonrepeating
sections
Spectrum
RAW, RST
Description
Section placement
In a Nonrepeating
section, the Component
Cal Curve object uses the
specified component.
Formatting Object Descriptions
Table 6 describes the formatting objects.
Table 6. Formatting objects
Graphic object
Data source file
Description
Section placement
Columns
none
Specify the layout and repeating
properties of graphic and text objects.
Any
Page Break
none
Insert a manual page break in the report. Qual or Quan repeating
and nonrepeating
sections
124
XReport User Guide
Thermo Scientific
C
Sample Report
This topic appendix contains a sample Xcalibur report that is generated when you use
PeakIntegration.xrt (located in the Xcalibur\templates folder) to generate a Quan sample
report that contains certain peak integration data. The sample report uses the steroids02.rst
data file from the Xcalibur\examples\data folder.
Contents
• Peak Integration Report, Page 1
• Peak Integration Report, Page 2
• Peak Integration Report, Page 3
• Peak Integration Report, Page 4
Thermo Scientific
XReport User Guide
125
C
Sample Report
Peak Integration Report, Page 1
Peak Integration Report, Page 1
126
XReport User Guide
Thermo Scientific
C Sample Report
Peak Integration Report, Page 2
Peak Integration Report, Page 2
Thermo Scientific
XReport User Guide
127
C
Sample Report
Peak Integration Report, Page 3
Peak Integration Report, Page 3
128
XReport User Guide
Thermo Scientific
C Sample Report
Peak Integration Report, Page 4
Peak Integration Report, Page 4
Thermo Scientific
XReport User Guide
129
D
XReport Application Features
This appendix provides reference material about the XReport application, including the
XReport window and dialog boxes.
Contents
• Reports View
• XReport Dialog Boxes
• Report Sections
• Report Objects
• Report Templates
Reports View
Use the XReport application or the Reports view of the Processing Setup window to specify
how the Xcalibur data system produces reports for samples and sequences. The data system
provides several standard report formats. You can export results in a number of file formats
including spreadsheet documents (XLS) and web pages (HTM or HTML).
For more information about the Reports view of the Processing Setup window, refer to the
Xcalibur Data Acquisition and Processing User Guide.
Thermo Scientific
XReport User Guide
131
D
XReport Application Features
Reports View
XReport Toolbar Buttons
Table 7 describes the XReport toolbar icons and buttons.
Table 7. Toolbar buttons (Sheet 1 of 2)
Button
Description
New
Create a new report template for the program you
select.
Open
Find and open XRT files.
Save
Save the current template as an XRT file. This option
is not available if you are using a locked template. In
this case, choose File > Save As and save the template
with a different name.
Cut
Remove the currently selected objects and place them
on the clipboard.
Copy
Copy the currently selected objects to the clipboard.
Paste
Insert the contents of the clipboard in the currently
selected section.
Align Left
Align the left edges of the currently selected objects to
the left edge of the first selected object.
To select more than one object, click and drag a
selection box around the objects or hold down the
SHIFT key and click each object.
Align Right
Align the right edges of the currently selected objects
to the right edge of the first selected object.
To select more than one object, click and drag a
selection box around the objects or hold down the
SHIFT key and click each object.
Align Top
Align the top edges of the currently selected objects
to the top edge of the first selected object.
To select more than one object, click and drag a
selection box around the objects or hold down the
SHIFT key and click each object.
132
XReport User Guide
Thermo Scientific
D
XReport Application Features
Reports View
Table 7. Toolbar buttons (Sheet 2 of 2)
Button
Description
Align Bottom
Align the bottom edges of the currently selected
objects to the bottom edge of the first selected object.
To select more than one object, click and drag a
selection box around the objects or hold down the
SHIFT key and click each object.
Auto Align
Snap all objects (selected or not) in the current
section to the nearest upper and right grid lines.
Center Horizontally
Center the current object on the page so that it is an
equal distance from the left and right sides of the
page.
Center Vertically
Center the current object on the page so that it is an
equal distance from the top and bottom sides of the
page.
Make Same Height
Enlarge all selected objects (except for the first
selected) to match the height of the first selected
object.
Make Same Width
Enlarge all selected objects (except for the first
selected) to match the width of the first selected
object.
Make Same Height and Enlarge all selected objects (except for the first
Width
selected) to match the height and the width of the
first selected object.
Shrink Wrap
Shrink the template workspace to remove all unused
workspace at the bottom of the section.
Expand
Enlarge the template workspace to the full height of
the section size.
Zoom
Thermo Scientific
Increase or decrease the magnification of the template
display.
Font
Change the font that is used in the currently selected
object or change the global font settings if no object
is selected.
Resolve Report
Combine your report template with your data and
view, save, and print your report.
Help
Open Xcalibur Help.
XReport User Guide
133
D
XReport Application Features
Reports View
XReport Menus
The XReport window has the following menus:
• XReport File Menu
• XReport Edit Menu
• XReport View Menu
• XReport Layout Menu
• XReport Report Menu
• XReport Help Menu
XReport File Menu
Table 8 describes the File menu commands for the XReport application.
Table 8. File menu commands
134
XReport User Guide
Command
Description
New
Create a new template.
Open
Open a previously saved template.
Save
Save the current template.
Note This option is not available if you are using a locked
template. In this case, choose File > Save As and give the template
a different name.
Save As
Save the current template with a different file name.
Page Setup
Define page layout and margin settings.
Recently Used Files
View the most recently opened template files. To open the
template, select a filename in the list.
Close
Close the current template.
Exit
Close XReport.
Thermo Scientific
D
XReport Application Features
Reports View
XReport Edit Menu
Table 9 describes the Edit menu commands for the XReport application.
Table 9. Edit menu commands
Command
Description
Cut
Remove the currently selected objects and place them on the
clipboard.
Copy
Copy the currently selected objects to the clipboard.
Paste
Insert the contents of the clipboard in the currently selected
section.
Clear
Delete the currently selected objects.
Select All
Select all of the objects in the currently selected section.
Edit Object
Change the current object. For example, to enable resizing of
columns in the table object.
Properties
Configure the current object. For example, for a bitmap object,
you can select the source file.
Delete Section
Delete the currently selected section. This option is not available
when objects are selected.
XReport View Menu
Table 10 describes the View menu commands for the XReport application.
Table 10. View menu commands
Thermo Scientific
Command
Description
Report Template Grid
View or hide the grid in the current section. To change your grid
settings, choose Layout > Preferences.
Report Template
Outline
View or hide the Report Template Outline pane. This pane
displays an overview of all sections and objects in your report
template and opens at the right side of the XReport window.
Toolbars
View or hide the toolbars.
Status Bar
View or hide the status bar at the bottom of the XReport window.
XReport User Guide
135
D
XReport Application Features
Reports View
XReport Layout Menu
Table 11 describes the Layout menu commands for the XReport application.
Table 11. Layout menu commands (Sheet 1 of 2)
Command
Description
Align >
Left
Align the left edges of the currently selected objects to the left edge
of the first selected object.
To select more than one object, click and drag a selection box
around the objects or hold down the SHIFT key and click each
object.
Align >
Right
Align the right edges of the currently selected objects to the right
edge of the first selected object.
To select more than one object, click and drag a selection box
around the objects or hold down the SHIFT key and click each
object.
Align >
Top
Align the top edges of the currently selected objects to the top
edge of the first selected object.
To select more than one object, click and drag a selection box
around the objects or hold down the SHIFT key and click each
object.
Align >
Bottom
Align the bottom edges of the currently selected objects to the
bottom edge of the first selected object.
To select more than one object, click and drag a selection box
around the objects or hold down the SHIFT key and click each
object.
136
XReport User Guide
Align >
Auto
Snap all objects (selected or not) in the current section to the
nearest upper and right grid lines.
Make Same Size >
Width
Enlarge all selected objects (except for the first selected) to match
the width of the first selected object.
Make Same Size >
Height
Enlarge all selected objects (except for the first selected) to match
the height of the first selected object.
Make Same Size >
Both
Enlarge all selected objects (except for the first selected) to match
the height and the width of the first selected object.
Center In Page >
Vertical
Center the current object on the page so that it is an equal
distance from the left and right sides of the page.
Center In Page >
Horizontal
Center the current object on the page so that it is an equal
distance from the top and bottom sides of the page.
Shrink Wrap
Shrink the current section to fit all of the objects that it contains.
Thermo Scientific
D
XReport Application Features
Reports View
Table 11. Layout menu commands (Sheet 2 of 2)
Command
Description
Expand
Enlarges the current section so you can add additional objects.
Preferences
Change grid settings, enable actions, and view or hide template
configuration dialog box.
XReport Report Menu
Table 12 describes the Report menu commands for the XReport application.
Table 12. Report menu commands
Command
Description
Data Sources
Specify the data files for the report.
Simulate Report
View your template before adding real data.
Resolve Report
Display the report with data filled in. Select data in the Data
Sources dialog box.
Print Report
Print the current report.
XReport Help Menu
Table 13 describes the Help menu commands for the XReport application
Table 13. Help menu commands
Command
Description
Xcalibur Help
View Xcalibur Help.
Glossary
View the glossary.
How to Use Help
View How to Use Help.
About XReport
Open the About XReport dialog box.
The About XReport dialog box displays the version number of
XReport, along with copyright information.
Thermo Scientific
XReport User Guide
137
D
XReport Application Features
XReport Dialog Boxes
XReport Dialog Boxes
You can access the following dialog boxes from the XReport application:
• Repeating Section Properties Dialog Box
• Chromatogram Properties Dialog Box and Wizard
• Spectrum Properties Dialog Box
• Table Properties Wizard
• Edit Formula For Dialog Box
• Configure Function Dialog Box
• Spectrum List Table Properties Dialog Box
• Average Scan Filter Selection Dialog Box
• Library Search Graphics Table Properties Dialog Box
• Library Search Results Table Properties Dialog Box
• Sample Table Properties Dialog Box
• Text Object Properties Dialog Box
• Bitmap Properties Dialog Box
• Column Properties Dialog Box
• Component Calibration Curve Dialog Box
• Data Sources Dialog Box
• Electronic Signature Properties Dialog Box
• Font Dialog Box
• Preferences Dialog Box
• Preview Window
• Save As Dialog Box
• Template Configuration Dialog Box
• Page Setup Dialog Box
138
XReport User Guide
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Repeating Section Properties Dialog Box
Use the Repeating Section Properties dialog box to customize the following settings for your
repeating section:
• Including or exclude internal standards or target compounds from a Quan repeating
section
• Specifying the order in which peaks and components are repeated
• Defining any limit on the number of peaks reported
• Specifying page breaks in the final report
The Repeating Section dialog box has these pages:
• Exclude Page
• Ordering Page
• Page Breaks Page
Exclude Page
Use the Exclude page of the Repeating Section Properties dialog box to indicate components
to exclude from a Quan repeating section, and whether to discard any components that have
not been found.
For more information, see Configuring a Repeating Section.
Table 14 describes the parameters on the Repeating Section Properties – Exclude page.
Table 14. Exclude page parameters
Parameter
Description
Repeating Section Excludes
Exclude Component
Type
Exclude specific compounds from the Quan configurations:
• None
• Internal Standards
• Target Compounds
Show Only Found
Components
Thermo Scientific
Discard any components that have not been found.
XReport User Guide
139
D
XReport Application Features
XReport Dialog Boxes
Ordering Page
Set parameters on the Ordering page of the Repeating Section Properties dialog box to
indicate the order for peaks or components in the repeating section.
Table 15 describes the parameters on the Repeating Section Properties – Ordering page.
Table 15. Ordering page parameters
Parameter
Description
Repeating Section Ordering
Order By
Choose a value that determines the order for the peaks or
components in the section:
• RT
• Component Name
• Peak Height Intensity
• Peak Area Intensity
• Peak Response Intensity
• Component Concentration
Ordering Direction
Ascending or
Descending
Choose an order for the peaks or components in the section by the
property in the Order By list from lowest to highest (ascending) or
highest to lowest (descending).
Repeating Records Maximum
Limit number of
reported peaks or
components to
140
XReport User Guide
Limit the total number of peaks or components that are included
in the repeating section.
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Page Breaks Page
Set parameters on the Page Breaks page of the Repeating Section Properties dialog box to
indicate where to apply page breaks in your report.
Table 16 describes the parameters on the Repeating Section Properties – Page Breaks page.
Table 16. Page Breaks page parameters
Parameter
Description
Each peak contained on its
own page(s)
Insert a page break before and after each peak in the repeating
section.
No page breaks between the Remove page breaks between each peak in the repeating
individual peaks
section.
Page break before the
repeating section
Insert a page break before the repeating section.
Page break after the
repeating section
Insert a page break after the repeating section.
Chromatogram Properties Dialog Box and Wizard
The Chromatogram Properties dialog box displays a preview of the current chromatogram
settings.
 To change these settings
Click Configure.
The Chromatogram Properties wizard opens.
Use the Chromatogram Properties wizard to configure any of these settings for the
Chromatogram report object that is contained in your report template:
• Chromatogram and peak type
• Use of a demo file
• Plot details
• Smoothing parameters
• Use of MS Quan traces
• Chromatogram display, style, labeling, axes, and normalization methods
This Chromatogram Properties wizard displays different pages and contains different options,
depending on the type of section containing the chromatogram report object and the
selections for the chromatogram type and peak type properties.
Thermo Scientific
XReport User Guide
141
D
XReport Application Features
XReport Dialog Boxes
The Chromatogram Properties wizard can contain the following pages:
• Advanced Qual Peak Annotation Options
• Advanced Quan Peak Annotation Options
• Configure the Look of the Peak Window
• Configure Normalization For The Chromatogram
• Select A Peak Dialog Box
• Select Chromatogram Style
• Select Chromatogram Type
• Select Component
• Select Demo File
• Select Enhance Details
• Select How the Chromatogram Axes Will Be Configured
• Select Mass Spectrum Quan Traces to Display
• Select Plot Details
• Select the Peak Labeling
• Select the Qual Peak
• Select the Quan Peak
• Use Container Supplied Qual Peak
• Use Container Supplied Quan Peak
Advanced Qual Peak Annotation Options
Set advanced Qual peak annotation parameter on the Advanced Qual Peak Annotation
Options page of the Chromatogram Properties wizard. When you set advanced labeling
options, only the Qual peak is labeled.
Note If you are in a Quan or nonrepeating section, choose a Qual results-based
chromatogram.
142
XReport User Guide
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Table 17 describes the parameters on the Chromatogram Properties – Advanced Qual Peak
Annotation Options page.
Table 17. Advanced Qual Peak Annotation Options page parameters
Parameter
Description
Use Advanced Qual Peak
Annotation
Turn on the advanced Qual peak options.
Custom Peak Labels
Custom Text Label
Label the Qual peak with a custom label.
Retention Time
Label the Qual peak with the retention time.
RT Labeling
Units
Define the units for the retention time.
Decimals
Specify the number of decimal places to be displayed in the
retention time label.
Advanced Quan Peak Annotation Options
Set advanced Quan peak annotation parameters on the Advanced Quan Peak Annotation
Options page of the Chromatogram Properties wizard. When you set advanced labeling
options, only the Quan peak is labeled.
Note If you are in a Qual or nonrepeating section, choose a Quan results-based
chromatogram.
Table 18 describes the parameters on the Chromatogram Properties – Advanced Quan Peak
Annotation Options page.
Table 18. Advanced Quan Peak Annotation Options page parameters (Sheet 1 of 2)
Parameter
Description
Use Advanced Quan
Peak Annotation
Turn on the advanced Quan peak options.
Custom Peak Labels
Component Name
Label the Quan peak with the component name.
Component Key
Label the Quan peak with the component key.
Custom Text Label
Label the Quan peak with a custom text label.
Retention Time
Label the Quan peak with the retention time.
RT Labeling
Units
Thermo Scientific
Specify the units for the retention time.
XReport User Guide
143
D
XReport Application Features
XReport Dialog Boxes
Table 18. Advanced Quan Peak Annotation Options page parameters (Sheet 2 of 2)
Parameter
Description
Decimals
Define the number of decimal places to be displayed in the
retention time label.
Overlaid QI Identifier
Show QI Peak Label
Label overlaid qualifier ion peaks with a label identifying the
qualifier ion. This option is only available when you select one of
these options:
• The All Quan Traces option on the Select Mass Spectrum
Quan Traces to Display page.
• The Overlay (3D) option on the Select Chromatogram Style
page.
Configure the Look of the Peak Window
Set parameters on the Configure How The Peak Window Will Look page of the
Chromatogram Properties wizard to control the display of baselines (Quan and Qual peaks)
and the peak window width (Qual peak and in some cases Quan peaks) when a peak from a
result file is set to display in a chromatogram.
Table 19 describes the parameters on the Chromatogram Properties – Configure How The
Peak Window Will Look page.
Table 19. Configure How the Peak Window Will Look page parameters
144
XReport User Guide
Parameter
Description
Show Baselines
View the baselines for a chromatogram peak (Quan or Qual). If
you do not select this option, the Xcalibur data system hides the
baselines, peak tracing, and full peak annotation.
Set Custom Window
Width
Override the default value of the peak window width (the default
is peak width plus one min) manually. This option is always
available for Qual peaks and is sometimes available for Quan
peaks.
Custom Width (min)
Define the Custom window width. The default value is 0.75 min.
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Configure Normalization For The Chromatogram
Set parameters on the Configure Normalization For The Chromatogram page of the
Chromatogram Properties wizard to select the normalization (y-axis scaling) method used for
the chromatogram.
Table 20 describes the parameters on the Chromatogram Properties – Configure
Normalization For The Chromatogram page.
Table 20. Configure Normalization for the Chromatogram page parameters
Parameter
Description
Normalize Method
Auto Zero
Optimize the y axis automatically for each chromatogram.
Intensity Range
Set the range manually. Type the minimum and maximum
intensity required for the y axis. The value can be from –200.0%
to +200.0%.
Normalize Each Plot To
Largest Peak In
Subsection
Normalize each split time range to the largest peak in the division.
Largest Peak In
Selected Time Range
Normalize the chromatogram to the largest peak in the displayed
time range.
Largest Peak In All
Times
Normalize the chromatogram to the largest peak in the entire
chromatogram.
Select A Peak Dialog Box
Use the Select A Peak dialog box to select a peak in a demo file that opens in the report
preview.
Table 21 describes the parameters in the Select A Peak dialog box.
Table 21. Select a Peak dialog box parameters
Thermo Scientific
Parameter
Description
Result File
Select the name of the result file to be the demo file.
Peak Type
Define the peak type (Quan or Qual) in the result file.
Peak List
View the list of peaks that are in the result file.
XReport User Guide
145
D
XReport Application Features
XReport Dialog Boxes
Select Chromatogram Style
Use the Select Chromatogram Style page of the Chromatogram Properties wizard to
determine the appearance of the chromatogram.
Table 22 describes the parameters on the Chromatogram Properties – Select Chromatogram
Style page.
Table 22. Select Chromatogram Style page parameters
Parameter
Description
Plotting
Point-to-Point
(see Figure 63)
Display point-to-point peak profiles.
Stick
Display the chromatogram as vertical lines.
Arrangement
Stack (2D)
Stack plots vertically (same x axis) with no overlap (one y axis for
each plot).
Overlay (3D)
Move plots vertically (same x axis) with an optional horizontal
skew (y-axis time offset). When selected, the 3D display options
are turned on. Use the 3D arrangement if you selected Multiple
Quan Traces on the Select Mass Spectrum Quan Traces to Display
page.
3D
146
XReport User Guide
Elevation
Set the elevation angle (amount of overlay) from 0 to 60 degrees
for the 3D plot style.
Skew
Set the skew angle (time offset) from 0 to 45 degrees for the 3D
plot style.
Draw Backdrop
Add a drawn perspective backdrop to 3D plots.
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Select Chromatogram Type
Set the chromatogram type for the chromatogram report object being configured in your
report template on the Select Chromatogram Type page of the Chromatogram Properties
wizard.
Table 23 describes the parameters on the Chromatogram Properties – Select Chromatogram
Type page.
Table 23. Select Chromatogram Type page parameters
Parameter
Description
Chromatogram Type
Raw File Based
Chromatogram
Indicate the chromatogram is based on a raw data file (RAW).
Quan Results Based
Chromatogram
Indicate the chromatogram is based on Quan results.
Qual Results Based
Chromatogram
Indicate the chromatogram is based on Qual results.
Show All Qual Peaks In Display all Qual peaks in the finished report.
Result File
Select Component
Set parameters in the Select Component dialog box of the Chromatogram wizard to select a
component to display when configuring a chromatogram.
For information about selecting the component for a Quan peak, where the chromatogram
type is a Quan results based chromatogram, see Select the Quan Peak Page.
Table 24 describes the parameters in the Select Component dialog box of the Chromatogram
wizard.
Table 24. Select Component page parameters
Thermo Scientific
Parameter
Description
Processing
View the name of the currently selected processing method
(PMD) file or select a different processing method file.
Component List
View a list of components found in the processing method file.
Select a component from this list.
XReport User Guide
147
D
XReport Application Features
XReport Dialog Boxes
Select Demo File
Set parameters on the Select Demo File page of the Chromatogram wizard to associate a demo
file (RAW or RST) with a chromatogram report object. Use the demo file to see how the
chromatogram report object appears with its current settings when applied to real data.
The appearance of the Select Demo File page depends on the chromatogram type and the
peak type settings chosen on the Chromatogram Type page.
Table 25 describes the parameters on the Chromatogram wizard – Select Demo File page.
Table 25. Select Demo File page parameters
Parameter
Description
Use Demo File
Use Demo File
Connect a demo file with the chromatogram report object.
Demo File
Specify the path to the demo file.
Demo Peak
Quan Component
Specify the Quan peak to use in the demo file. Click the Browse
button. The Select A Peak Dialog Box opens.
This box appears only if you are configuring a Quan based
chromatogram.
Qual Peak RT
Specify the Qual peak to use in the demo file. Click the Browse
button. The Select A Peak Dialog Box opens.
This box appears only if you are configuring a Qual based
chromatogram, and not showing all Qual peaks. Select these
options on the Select Chromatogram Type page.
148
XReport User Guide
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Select Enhance Details
Set parameters on the Select Enhance Details page of the Chromatogram Properties wizard to
enhance a chromatogram plot by applying smoothing.
Table 26 describes the parameters on the Chromatogram Properties – Select Enhance Details
page.
Table 26. Select Enhance Details page parameters
Parameter
Description
Enable Smoothing
Apply smoothing to the chromatogram.
Smoothing Details
Type
Define the type of smoothing algorithm applied to the
chromatogram, either Boxcar or Gaussian.
Points
Specify the number of smoothing points to use in the smoothing
algorithm.
Select How the Chromatogram Axes Will Be Configured
Set chromatogram axes labels and display parameters on the Select How The Chromatogram
Axes Will Be Configured page of the Chromatogram Properties wizard.
Table 27 describes the parameters on the Chromatogram Properties – Select How The
Chromatogram Axes Will Be Configured page.
Table 27. Select How the Chromatogram Axes Will Be Configured page parameters (Sheet 1 of 2)
Parameter
Description
X axis
Name
View or change the name of the x axis.
Offset
Move the x-axis label away from the chromatogram.
Split Time Range
Split the chromatogram into two or more separate graphs with
equal time ranges.
Divisions
Define the number of split time range graphs to be displayed for
each chromatogram.
Y axis
Label
From Detector
Use the detector-specific label for the y axis (stored in the RAW
file).
Custom
Use your own label for the y axis.
Units
Absolute
Thermo Scientific
Use absolute units for the y axis.
XReport User Guide
149
D
XReport Application Features
XReport Dialog Boxes
Table 27. Select How the Chromatogram Axes Will Be Configured page parameters (Sheet 2 of 2)
Parameter
Description
Relative
Use relative units for the y axis.
Name
Define the custom name of the y axis. This box is available only
when you click the Custom option in the Label area.
Offset
Move the y axis label away from the chromatogram.
Select Mass Spectrum Quan Traces to Display
Set parameters on the Select Mass Spectrum Quan Traces To Display page of the
Chromatogram Properties wizard to configure the Quan Traces property for MS data.
Options on this page only apply to MS data and are ignored for other detector types. When a
chromatogram report object displays a Quan component, there are potentially six traces
associated with the component: the Quan Mass (QM) trace (the actual component peak) and
up to five qualifier ions.
Select any of these possible six traces to be displayed in the chromatogram or display all of the
available traces of the component.
Table 28 describes the parameters on the Chromatogram Properties – Select Mass Spectrum
Quan Traces To Display page.
Table 28. Select Mass Spectrum Quan Traces to Display page parameters
Parameter
Description
Quan Traces
Quan Traces
Specify the Quan traces to display.
Selections:
Common Normalization
For All Traces
Normalize all of the traces the same way.
This check box is available only when you select All Quan
Traces from the Quan Traces list.
Select Plot Details
Set parameters on the Select Plot Details page of the Chromatogram Properties wizard to
configure how the plot is drawn when you configure a raw file-based chromatogram.
 To configure a plot
1. If you are in a Qual or Quan repeating section, the Use Container Supplied Qual or
Quan Peak page opens. Select the Explicitly Specify the Chromatogram’s Behavior
check box and click Next.
150
XReport User Guide
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
2. On the Select Chromatogram Type page, select a raw file based chromatogram and follow
the pages of the wizard until the Select Plot Details page opens.
Table 29 describes the parameters on the Chromatogram Properties – Select Plot Details page.
Table 29. Select Plot Details page parameters (Sheet 1 of 2)
Parameter
Description
Plot Type
Detector
View the type of detector used to generate the raw data file.
Delay (min)
Specify the delay for the start of the chromatogram trace.
Filter
View the processing that is applied to a subset of the scans in the
RAW file. The list displays scan filter options that are stored in the
selected demo RAW file. If no demo file is selected, type a valid
filter string.
This option is only available when an MS detector is selected.
Trace
The first list indicates the basic chromatogram type, for example,
TIC.
The second list indicates a logical operator: + or –.
The third list indicates a second chromatogram type to add to, or
subtract from, the first trace. For example, Mass Range.
The list of traces depends on the detector used to generate the data
as follows:
• Detector Type
• Valid Traces
• MS
• Mass Range, TIC, Neutral Fragment, and Base Peak
• Analog
• Analog 1–4
• A/D Card
• A/D Card Ch. 1–4
• PDA
• Wavelength Range, Total Scan, and Spectrum Maximum
• UV
• Channel A–D
Thermo Scientific
XReport User Guide
151
D
XReport Application Features
XReport Dialog Boxes
Table 29. Select Plot Details page parameters (Sheet 2 of 2)
Parameter
Description
Mass (m/z)
Define the display for the mass or mass range for Mass Range or
Base Peak traces for MS detector types as follows:
If you choose Base Peak ± Mass Range or Mass Range ± Mass
Range plot combinations, the identification page displays a Mass
(m/z) field for each trace type.
If you use other detector types, use this field to specify the
wavelength or wavelength range for the chromatogram. If you use
a plot combination such as
Wavelength Range + Wavelength Range, an additional Wavelength
(nm) box opens.
The valid range depends on the configured detector. The format is
low mass /wavelength–high mass /wavelength. Separate entries with a
dash, no spaces. For example, for the range m/z 123 through 456,
type 123–456.
Time Range (min)
Specify the upper and lower time limits for the plot in minutes, to
one decimal place. Separate entries by a dash, with no spaces.
The default setting is “*”, which means to display the entire time
range for the plot.
Fixed Scale
Set the y-axis maximum to a specific value. Clear this check box to
activate the Normalize To area on the Configure Normalization
For The Chromatogram page.
Autofilter
Repopulate a chromatogram view with
• A plot showing the chromatogram without any scan filters
• Plots for each scan filter applied to the chromatogram
152
XReport User Guide
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Select the Peak Labeling
Set parameters on the Select the Peak Labeling page of the Chromatogram Properties wizard
to choose the type and style of peak labels.
Table 30 describes the parameters on the Chromatogram Properties – Select the Peak Labeling
page.
Table 30. Select the Peak Labeling page parameters
Parameter
Description
Label With
Retention Time
Label the peaks with the time in minutes, prefixed by RT.
Scan Number
Label the peaks with number of mass scans at peak maximum,
prefixed by S#.
Base Peak
Label the base peak in the mass spectrum of the chromatogram
peak, prefixed by BP.
Signal To Noise
Label the peaks with the signal-to-noise ratio at the peak
maximum, prefixed by SN.
Flags
Label the peaks with flags providing supplemental information
about the peak data. For example, if a peak is saturated, the data
system displays an S above the peak.
Area
Label the peaks with the integrated area of the peaks, prefixed by
AA if detected automatically or MA if detected manually.
Height
Label the peaks with the apex height, prefixed by AH if detected
automatically or MH if detected manually.
Label Styles
Thermo Scientific
Offset
Move a label away from its normal position to avoid conflict with
another label.
Size
Specify the offset size, measured in number of characters.
Rotated
Use vertical rather than horizontal mass labels. Flags are always
displayed unrotated.
Boxed
Place a rectangular outline around each peak mass label. Flags are
not boxed.
Label Threshold
Limit the labeling of peaks to those exceeding the specified
percentage of the base peak.
Advanced
Set advanced labeling options. Only the Quan or Qual peak is
labeled if you click this option.
XReport User Guide
153
D
XReport Application Features
XReport Dialog Boxes
Select the Qual Peak
Set the parameter on the Select the Qual Peak page to explicitly specify the Qual peak for a
Qual results-based chromatogram report object.
Table 31 describes the parameters on the Chromatogram Properties – Select the Qual Peak
page.
Table 31. Select the Qual Peak page parameters
Parameter
Description
Qual Peak To Display
Retention Time (min)
The retention time of the Qual peak. The retention time can be
any time between the start and end peak boundaries.
You must select a valid retention time to combine template and
data correctly. If the file used to resolve the report does not
contain the specified retention time, the Xcalibur data system
removes the chromatogram from the report.
Select the Quan Peak
Set parameters on the Select the Quan Peak page of the Chromatogram Properties wizard to
explicitly specify the Quan peak (component) for a Quan results-based chromatogram report
object.
Table 32 describes the parameters on the Chromatogram Properties – Select the Quan Peak
page.
Table 32. Select the Quan Peak page parameters
Parameter
Description
Quan Peak
Quan Peak
Set the name of the Quan peak. To select a processing method
(PMD) file and select a component, click the Browse button.
The Select Component dialog box opens.
Use Quan Peak’s Associated Use the associated internal standard of the Quan peak instead
Internal Standard
of the Quan peak.
You must select a valid peak for XReport to resolve the report
correctly. If the file used to resolve the report does not contain
the specified peak, the chromatogram is removed from the
resolved report.
154
XReport User Guide
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Use Container Supplied Qual Peak
Set the parameter on the Use Container Supplied Qual Peak page of the Chromatogram
Properties wizard to ignore the implicit Qual peak and explicitly configure the chromatogram
type.
Table 33 describes the parameters on the Chromatogram Properties – Use Container Supplied
Qual Peak page.
Table 33. Use Container Supplied Qual Peak page parameters
Parameter
Description
Explicit Chromatogram Type
Explicitly Specify The
Chromatogram’s Behaviour
Ignore the implicit Quan peak supplied by the container
and explicitly configure the chromatogram type.
Use Container Supplied Quan Peak
Set parameters on the Use Container Supplied Quan Peak page of the Chromatogram
Properties wizard to either display the details of the internal standard associated with the
implicit Quan peak or ignore the implicit Quan peak and explicitly configure the
chromatogram type.
Table 34 describes the parameters on the Chromatogram Properties – Use Container Supplied
Quan Peak page.
Table 34. Use Container Supplied Quan Peak page parameters
Parameter
Description
Implicit Quan Peak Options
Use Quan Peak’s Associated
Internal Standard
View the details of the internal standard associated with
the implicit Quan peak.
Explicit Chromatogram Type
Explicitly Specify The
Chromatogram’s Behaviour
Thermo Scientific
Ignore the implicit Quan peak supplied by the container
and explicitly configure the chromatogram type.
XReport User Guide
155
D
XReport Application Features
XReport Dialog Boxes
Spectrum Properties Dialog Box
Use the Spectrum Properties dialog box to configure the settings for your spectrum report
object.
Spectrum Properties dialog box pages:
• Axis Page for a Spectrum Object
• Enhance Page for a Spectrum Object
• Labels Page for a Spectrum Object
• Normalization Page for a Spectrum Object
• Options Page for a Spectrum Object
• Peak Page for a Spectrum Object
• Plot Type Page for a Spectrum Object
• Style Page for a Spectrum Object
Axis Page for a Spectrum Object
Set spectrum axis labels and display options on the Axis page of the Spectrum Properties
dialog box.
Table 35 describes the parameters on the Spectrum Properties – Axis page.
Table 35. Axis page parameters (Sheet 1 of 2)
Parameter
Description
X axis
Name
View or change the name for the x axis.
Offset
Move the axis label away from the spectrum.
Split Mass Range
Split the spectrum into two or more separate graphs with equal
ranges.
Divisions
View the number of split range graphs for each spectrum.
Y axis
Label
156
XReport User Guide
From Detector
Use the detector-specific label for the y axis (stored in the raw data
file).
Custom
Use your own label for the y axis.
Name
Specify the name for the y axis if you selected Custom label above.
Offset
Move the axis label away from the spectrum.
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Table 35. Axis page parameters (Sheet 2 of 2)
Parameter
Description
Units
Absolute
Use absolute units on the y axis.
Relative
Use relative units on the y axis.
Enhance Page for a Spectrum Object
Specify spectrum enhancement details on the Enhance page of the Spectrum Properties dialog
box.
Note Clear the Use Processing Method Properties For Plot Type And Enhancement Info
check box on the Options Page for a Spectrum List Table to modify the Enhance page
parameters.
Table 36 describes the parameters on the Spectrum Properties – Enhance page.
Table 36. Enhance page parameters (Sheet 1 of 2)
Parameter
Description
Smoothing
Enable
Turn on spectrum smoothing.
Type
Specify the type of smoothing algorithm to apply to the spectrum.
Choose from Boxcar or Gaussian.
Points
Specify the number of points for spectrum smoothing. This must
be an odd number from 3 (minimum smoothing) to 15
(maximum smoothing).
Refine
Thermo Scientific
Enable
Turn on refine spectrum enhancement.
Window
Define an area for drawing a peak by specifying the number of
seconds on either side of the specified point. A reasonable initial
value is the peak width in seconds.
Noise Threshold
Specify the cutoff point for peaks. Type 0 to show all peaks in the
spectrum and slowly increase the value until the noise peaks are
eliminated. Use this setting to eliminate peaks generated from
baseline noise.
XReport User Guide
157
D
XReport Application Features
XReport Dialog Boxes
Table 36. Enhance page parameters (Sheet 2 of 2)
Parameter
Description
Background Subtraction
This area shows details of background subtraction if it was applied to the spectrum. The
background contribution is determined by averaging the scans from one or two baseline
regions.
Time Range 1
Add a first baseline region that is used for assessing the
background. When you click this option, the Xcalibur data system
activates a box. Type the time range to use in the box.
Time Range 2
Add a second baseline region that is used for assessing the
background.
Labels Page for a Spectrum Object
Choose the type and style of peak labels on the Labels page of the Spectrum Properties dialog
box.
Table 37 describes the parameters on the Spectrum Properties – Labels page.
Table 37. Labels page parameters (Sheet 1 of 2)
Parameter
Description
Label With
Mass or Wavelength
Label the peaks with m/z values (if MS is selected as the detector
on the Plot Type Page for a Spectrum Object) or wavelength
values (if PDA is selected as the detector on the Plot Type Page for
a Spectrum Object).
Flags
Label spectrum peaks with flags. Flags provide supplemental
information about peak data. For example, if a peak is saturated,
the Xcalibur data system displays an S above the peak.
Relative To
Move the m/z labels for the peaks away from the peaks by the
specified amount.
Decimals
Specify the number of decimal places to be used in mass labels.
Label Styles
158
XReport User Guide
Offset
Move a label from its normal position to avoid conflicts with other
labels.
Size
Specify the offset size to use (in number of characters).
Rotated
Use vertical rather than horizontal mass labels. Flags are always
displayed unrotated.
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Table 37. Labels page parameters (Sheet 2 of 2)
Parameter
Description
Boxed
Place a rectangular outline around each peak mass label. Flags are
not boxed.
Label Threshold
Specify a peak height level, in percent of the base peak. Above that
level, XReport labels a peak.
Normalization Page for a Spectrum Object
Select normalization options for spectrum plots on the Normalization page of the Spectrum
Properties dialog box.
Table 38 describes the parameters on the Spectrum Properties – Normalization page.
Table 38. Normalization page parameters
Parameter
Description
Normalize Method
Auto Zero
Optimize the y axis automatically.
Intensity Range
Set the y-axis range. Use the box to type the minimum and
maximum intensity. The valid range is from –200.00 to
200.00%.
Normalize To
Largest Peak In Subsection
Normalize each split mass range to the largest peak in the
division.
Largest Peak In Range
Normalize the spectrum to the largest peak in the displayed
mass range.
Largest Peak In Scan
Normalize the spectrum to the largest peak in the entire
spectrum.
Normalize Multiple Scans
Thermo Scientific
Individually
Normalize each mass plot individually.
All The Same
Normalize all mass plots equally.
XReport User Guide
159
D
XReport Application Features
XReport Dialog Boxes
Options Page for a Spectrum Object
Set spectrum parameters on the Options page of the Spectrum Properties dialog box.
Table 39 describes the parameters on the Spectrum Properties – Options page.
Table 39. Options page parameters
Parameter
Description
Use Processing Method
Properties For Plot Type
and Enhance Info
Use properties from the selected processing method to set plot
type and enhance information for the spectrum.
Use ISTD Peak When
Repeating On Component
Use the selected ISTD (internal standard) peak when
repeating on a component. Select the Use Processing
Method Properties For Plot Type and Enhance Info option
to modify this parameter.
Peak Page for a Spectrum Object
Set parameters on the Peak page of the Spectrum Properties dialog box to explicitly specify the
peak to display in the spectrum report object.
Table 40 describes the parameters on the Spectrum Properties – Peak page.
Table 40. Peak page parameters
Parameter
Description
You can explicitly specify the peak that you want to appear in a spectrum control
Enable Explicit Peak
Type
Activate controls so you can specify the type of peak to appear in
the spectrum.
Peak Type
Quan Peak
Specify a Quan peak.
Qual Peak
Specify a Qual peak.
Peak Identification
Component Name
160
XReport User Guide
Define the name of the Quan peak or the retention time of the
Qual peak. If you are entering a Qual peak, you can type any time
between the start time and end time of the peak.
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Plot Type Page for a Spectrum Object
Configure how the spectrum is drawn on the Plot Type page of the Spectrum Properties
dialog box.
Note The parameters in the Plot Type page are enabled only when you clear the Use
Processing Method Properties For Plot Type And Enhancement Info check box on the
Options Page for a Spectrum List Table.
Table 41 describes the parameters on the Spectrum Properties – Plot Type page.
Table 41. Plot Type page parameters
Parameter
Description
Mass Range or
Wavelength Range
Specify the mass or mass range for MS detector types. For PDA
detector types, use this box to specify the wavelength or
wavelength range (nm).
Type the upper and lower limits for the plot. Separate entries by a
dash, with no spaces.
To use the entire mass or wavelength range, type an asterisk (*).
Detector
View the type of detector that was used to generate the raw data
file. Valid types are MS and PDA.
The type of detector determines the available plot types.
Thermo Scientific
Time
View or change the time of the mass scan in the host
chromatogram. Type a new value or range as required.
Scan Filter
Specify the scan filter to apply to your data. Select a filter from the
list or type a new filter using the scan filter format.
Demo File
Specify the path and filename of the demo file (raw data file) that
is used to generate a preview of the selected plot.
XReport User Guide
161
D
XReport Application Features
XReport Dialog Boxes
Style Page for a Spectrum Object
On the Style page of the Spectrum Properties dialog box, set parameters that determine the
appearance of the spectrum.
Table 42 describes the parameters on the Spectrum Properties – Style page.
Table 42. Style page parameters
Parameter
Description
Plotting
Automatic
Choose the graphic style based upon the data acquisition method
that was used for the active spectrum.
Point To Point
Display point-to-point peak profiles.
Stick
Display the spectral mass peaks as vertical lines.
Shade
Display the spectrum as a shaded representation of intensity in
each amu band for the active spectrum.
Table Properties Wizard
Use the table properties (Table Name Properties) wizard to customize the following settings for
your table:
• Which columns to include in the table
• Custom table column settings
• Table sort order
The Table Properties wizard has these pages:
• Configure User Column Parameters Page
• Select Table Fields Page
• Set Table Sort Filter Page
• Set Table Font Page
162
XReport User Guide
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Configure User Column Parameters Page
Configure any custom (User) columns that you selected to include in your table on the
Configure User Column Parameters page of the table properties wizard. Complete the
parameters on this page for each custom column that you selected.
For more information, see Configuring Custom (User) Columns.
Table 43 describes the parameters on the Configure User Column Parameters page of the
table properties wizard.
Table 43. Configure User Column Parameters page parameters
Parameter
Description
User Columns
Define the custom columns to include in your table. Select the
column to customize.
User Column Title
User Column Title
Provide a descriptive title for the column.
User Column Formula
User Column Formula
View the formula for the current column (read-only box). To edit
the formula, click Edit Formula.
Edit Formula
Change your formula.
Select Table Fields Page
Choose the columns for the table on the Select Table Fields page of the table properties
wizard. The columns available for a table are given in alphabetical order, but can be added to a
report template in any order. The columns that you can select depend on the type of table
object that you are configuring.
Table 44 describes the parameters on the Select Table Fields page of the table properties
wizard.
Table 44. Select Table Fields page parameters (Sheet 1 of 2)
Parameter
Description
Title
View the custom title of the selected object.
Include The Title In
The Resolved Report
Include the custom title in the final report.
Available Columns
View columns that are available for the selected table.
A user column is a column that you can configure to display what
you want on a later page of the wizard.
Thermo Scientific
XReport User Guide
163
D
XReport Application Features
XReport Dialog Boxes
Table 44. Select Table Fields page parameters (Sheet 2 of 2)
Parameter
Description
Add
Add one or more columns to the table. Select one or more
columns from the Available Columns list and click Add to add
them to the table.
Add All
Add all of the available columns to the table.
Remove
Remove one or more columns from the table. Select one or more
columns from the Selected list and click Remove to remove them
from the table.
Remove All
Remove all of the columns from the selected list.
Selected
View the columns that are currently selected to be included in the
table. The columns are listed in the order that they appear in the
table.
Column Header
View the heading for the currently selected column.
Decimal Setting
View or change the number of digits displayed after the decimal
point for values displayed in this column. This parameter is only
available for some table fields.
Move Up
Move a column up in the Selected list. Select a column in the
Selected list and click Move Up to move the column up in the list.
This action moves the column to the left in the finished table.
Move Down
Move a column down in the Selected list. Select a column in the
Selected list and click Move Down to move the column down in
the list. This action moves the column to the right in the finished
table.
Set Table Sort Filter Page
Set parameters on the Set Table Sort Filter page of the table properties wizard to define the
order for table data when you generate a report using the chosen table object in the report
template.
Note This page is not available for the Library Search Graphics Results table.
164
XReport User Guide
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Table 45 describes the parameters on the Select Table Sort Filter page of the table properties
wizard.
Table 45. Set Table Sort Filter page parameters
Parameter
Description
Column List
View the list of columns that are currently selected for the table.
Current Sorting Configuration
First Order column
Ascending or
Descending option
Second Order column
Ascending or
Descending option
Select the first column that the Xcalibur data system uses to sort
the table.
Select to sort in ascending or descending order.
Select the second column that the data system uses to sort the
table. For instance, use this option in columns that have identical
entries in the First Order column.
Select to sort in ascending or descending order.
Third Order column
Ascending or
Descending option
Select the third column that the data system uses to sort the table.
For instance, use this option in columns that have identical entries
in the First Order and Second Order columns.
Select to sort in ascending or descending order.
Buttons
Add
Add the selected column to the next available sort order.
Remove
Remove the last selected sort order.
Set Table Font Page
Configure the formatting of the currently selected table object on the Font page of the table
properties wizard.
Table 46 describes the parameters on the Select Table Font page of the table properties wizard.
Table 46. Set Table Font page parameters
Thermo Scientific
Parameter
Description
Title Font,
Heading Font,
Cell Font
Set and view the font settings for all of the title, headings, and cells
in the table object. To configure your font, click the button. The
Font Dialog Box opens.
XReport User Guide
165
D
XReport Application Features
XReport Dialog Boxes
Edit Formula For Dialog Box
Edit a formula for use in a custom column of your table in the Edit Formula For dialog box.
For more information on formula syntax, see Adding a Formula to a Table.
Table 47 describes the parameters in the Edit Formula For dialog box.
Table 47. Edit Formula For dialog box parameters (Sheet 1 of 2)
Parameter
Description
Formula
View the current formula. Type in this box directly or add functions from
the list.
Functions
A list of available functions. A brief description of each of the functions appears at the
bottom of the dialog box when each function is selected.
166
XReport User Guide
Function
Syntax
Description
ABS
ABS(x)
Return the positive value of a number of a
formula regardless of its actual positive or
negative value.
AND
AND(x,y)
Return the value 1 if the logical expressions x
and y are both true, or the value 0 if either
expression is false.
AVERAGE
AVERAGE(x1, x2,...,xn)
Return the average of the list of values.
COUNT
COUNT(list)
Return the count of the number of items in a
list.
FALSE
=FALSE
Return 0 (zero).
INT
INT(x)
Return the number to the left of the decimal
place in the value or formula x.
MAX
MAX(x1, x2,...,xn)
Return the largest value in a list.
MIN
MIN(x1, x2,...,xn)
Return the smallest value in a list.
MOD
MOD(x,y)
Return the remainder from dividing the value
x by the value y a number of times in integers.
NOT
NOT(x)
Return the value 0 (zero, false) if the logical
expression x is true, or the value 1 (true) of the
expression x is false.
OR
OR(x,y)
Return the value 1 (true) if either or both
logical expressions x and y are true, or the
value 0 (zero, false) if both expressions are
false.
PRODUCT
PRODUCT(x1, x2,...,xn)
Return the result of multiplying a list of
values.
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Table 47. Edit Formula For dialog box parameters (Sheet 2 of 2)
Parameter
Description
ROUND
ROUND(x,y)
Return the value of x rounded to the specified
number of decimal places y; x can be either a
number or a result of a formula.
SIGN
SIGN(x)
Return the value 1 if x is a positive value, or
the value –1 if x is a negative value.
SUM
SUM(x1, x2,...,xn)
Return the sum of a list of values or formulas.
TRUE
=TRUE
Return 1.
Add
Function
button
Configure the currently selected function and add it to the current formula.
Configure Function Dialog Box
Configure the settings for a function in a formula for use in a custom column of your table in
the Configure Function dialog box.
The Configure Function dialog box varies, depending on which function you are configuring.
FALSE and TRUE: No configuration options.
ABS, AVERAGE, INT, MAX, MIN, PRODUCT, ROUND, SIGN, SUM: Select one or
more columns to perform the function on. These expressions all have the same options, but
ROUND has the additional option to select the number of decimal places.
AND, NOT, and OR: enter one (NOT) or two (AND and OR) expressions to be evaluated.
These expressions all have the same options.
COUNT: Enter a list.
MOD: Select either a column or a value for each of the first and second parameters.
Thermo Scientific
XReport User Guide
167
D
XReport Application Features
XReport Dialog Boxes
Spectrum List Table Properties Dialog Box
Use the Spectrum List Table Properties dialog box to configure settings that apply to the
Spectrum List table.
The Spectrum List Table Properties dialog box has the following pages:
• Display Page for a Spectrum List Table
• Enhancement Page for a Spectrum List Table
• Options Page for a Spectrum List Table
• Plot Type Page for a Spectrum List Table
Display Page for a Spectrum List Table
Configure display parameters for the Spectrum List table on the Display page of the Spectrum
List Table Properties dialog box. For more information, see Display Page.
Table 48 describes the parameters in the Spectrum List Table Properties – Display page.
Table 48. Display page parameters
Parameter
Description
Display
All Peaks
Display all peaks in the Spectrum List table.
Top
Limit the number of peaks in the table to a specified number.
Order By
Mass/Intensity
or
Toggle between ordering the data in the Spectrum List table by
mass (or wavelength if the PDA Detector is selected) or intensity.
Wavelength/Intensity
Normalization
Intensity Range
Specify the relative abundance range of mass peaks that are
included in the table. The valid range is –200.000 to 200.000%.
To change the range of relative abundances, type the minimum
and maximum relative abundance to display in the Intensity
Range box, separated by a dash. For example, to display all peaks
in a mass spectrum with relative abundances ranging from 50 to
100%, type 50.000–100.000. XReport excludes spectrum peaks
with relative abundances that range from 0.000 to 49.999% from
the table.
168
XReport User Guide
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Enhancement Page for a Spectrum List Table
Configure spectrum enhancement parameters for the Spectrum List table on the
Enhancement page of the Spectrum List Table Properties dialog box. For more information,
see Enhancement Page.
Note Clear the Use Processing Method Properties for Plot Type and Enhancement
Info check box on the Options Page for a Spectrum List Table to modify parameters on
the Enhancement page.
Table 49 describes the parameters in the Spectrum List Table Properties – Enhancement page.
Table 49. Enhancement page parameters
Parameter
Description
Smoothing
Enable
Turn on spectrum smoothing.
Type
Specify the type of smoothing algorithm to apply to the spectrum.
Choose from Boxcar or Gaussian.
Points
Specify the number of points for spectrum smoothing. This must
be an odd number from 3 (minimum smoothing) to 15
(maximum smoothing).
Refine
Enable
Turn on refine spectrum enhancement.
Window
Define an area for drawing a peak by specifying the number of
seconds on either side of the specified point. A reasonable initial
value is the peak width in seconds.
Noise Threshold
Specify the cutoff point for peaks. Type 0 to show all peaks in the
spectrum and slowly increase the value until the noise peaks are
eliminated. Use this setting to eliminate peaks generated from
baseline noise.
Background Subtraction
This area shows details of background subtraction if it was applied to the spectrum. The
background contribution is determined by averaging the scans from one or two baseline
regions.
Thermo Scientific
Time Range 1
Activate a first baseline region that is used for assessing the
background. Clicking this option activates a box. Type the time
range to use in the box.
Time Range 2
Activate a second baseline region that is used for assessing the
background.
XReport User Guide
169
D
XReport Application Features
XReport Dialog Boxes
Options Page for a Spectrum List Table
Configure general parameters relating to the Spectrum List table on the Options page of the
Spectrum List Table Properties dialog box. For more information, see Options Page.
Table 50 describes the parameters in the Spectrum List Table Properties – Options page.
Table 50. Options page parameters
170
XReport User Guide
Parameter
Description
Use Processing Method
Properties For Plot Type
And Enhancement Info
Use the information in the processing method when
determining the plot type and other enhancement details.
Use ISTD Peak When
Repeating On Component
Include internal standard peaks in the spectrum list. This
parameter is only applicable to the Quan nonrepeating and is
not available when this object is used in a Qual nonrepeating.
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Plot Type Page for a Spectrum List Table
Configure plot information for the Spectrum List table on the Plot Type page of the Spectrum
List Table Properties dialog box. For more information, see Plot Type Page.
Note Clear the Use Processing Method Properties for Plot Type and Enhancement
Info check box on the Options Page for a Spectrum List Table to modify the parameters
in the Plot Type page.
Table 51 describes the parameters in the Spectrum List Table Properties – Plot Type page.
Table 51. Plot Type page parameters
Parameter
Description
Mass/
Wavelength
For Detector = MS:
View or change the current mass range of the active spectrum. To
change the mass range, type the first mass and last mass of the scan
in the Mass Range box. The format is First Mass–Last Mass. For
example to display mass/charge 100 through 200, type 100–200.
For Detector = PDA:
View or change the current wavelength range in nanometers. To
change the wavelength range, input the short wavelength and long
wavelength of the scan in the Wavelength Range box. The format
is Short Wavelength–Long Wavelength. For example, to display a
wavelength range of 195 through 795 nanometers, type 195–795.
Detector
View or change the currently selected detector data type:
• MS
• PDA
Thermo Scientific
Time
View or change the time range (in minutes) for the active
spectrum. The valid range is 0.00 to 200.00 minutes. To select a
time range, type the lower and upper time limits in minutes,
separated by a dash (no spaces) in the Time box. For example, to
select a time range from 0.10 to 9.10 minutes, type: 0.10–9.10.
Scan Filter
View or change the scan filter used for the active spectrum. Select
a filter from the list or use the scan filter format to type a scan
filter in the Filter box. This box is available only for MS detectors.
Demo File
View the filename of the demo file that is used in the template.
XReport User Guide
171
D
XReport Application Features
XReport Dialog Boxes
Average Scan Filter Selection Dialog Box
Configure the average scan filter that is used for the Spectrum List table in the Average Scan
Filter Selection dialog box.
Display the Plot Type page, and type an asterisk (*) in the Time box.
Table 52 describes the parameter in the Average Scan Filter Selection dialog box.
Table 52. Average Scan Filter Selection dialog box parameters
Parameter
Description
Scan Filter
View a list of all scan filters in the time range. If there is more than
one scan type in the selected time range, choose one scan type to
average.
Library Search Graphics Table Properties Dialog Box
Configure additional parameters that apply to the Library Search Graphics Table in the
Library Search Graphics Table Properties dialog box.
Table 53 describes the parameters in the Library Search Graphics Table Properties dialog box.
Table 53. Library Search Graphics Table Properties dialog box parameters
Parameter
Description
Max # Of Hits To
Display
Specify the maximum number of matches from the library search
to display in the Library Search Graphics table.
Display Column Labels Display labels at the top of the columns in the table.
Show Peak Spectrum
Display the peak spectrum in the table. This check box is available
only when you have selected to include the Hit Spectrum column
in the Table Properties dialog box.
Library Search Results Table Properties Dialog Box
Configure an additional parameter that applies to the Library Search Results Table in the
Library Search Results Table Properties dialog box.
Table 54 describes the parameter in the Library Search Results Table Properties dialog box.
Table 54. Library Search Results Table Properties dialog box parameters
Parameter
Description
General
Max # Of Hits To
Display
172
XReport User Guide
Specify the maximum number of matches from the library search
to display in the Library Search Results table.
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Sample Table Properties Dialog Box
Configure additional settings that apply to a Sample table in the Sample Table Properties
dialog box.
Table 55 describes the parameters in the Sample Table Properties dialog box.
Table 55. Sample Table Properties dialog box parameters
Parameter
Description
Include Standards
Include data from standards in the Sample table.
Include QCs
Include data from QCs in the Sample table.
Include Blanks
Include data from blank samples in the Sample table.
Include Unknowns
Include data from unknown samples in the Sample table.
Text Object Properties Dialog Box
Use the Text Object Properties dialog box to configure the settings of the currently selected
text object.
This dialog box has the following pages:
• Data Page
• Attributes Page
• Attributes Page for the Annotation Text Object
• Font Page
Thermo Scientific
XReport User Guide
173
D
XReport Application Features
XReport Dialog Boxes
Data Page
Indicate what text appears in the text object on the Data page.
The Data page of the Text Object Properties dialog box changes depending on the text object
you select.
To create an Annotation object, you can type the annotation text to add to your report. You
can either type the text directly in the box or copy and paste it in. Paste text in the box by
either right-clicking in the box and choosing Paste from the shortcut menu, or by pressing
CTRL+V. Add line breaks by pressing the ENTER key while the cursor is still in the box.
Table 56 describes the parameters on the Text Properties – Data page.
Table 56. Data page parameters (Sheet 1 of 2)
Parameter
Description
Title
The custom title of the selected object.
Include The Title In
The Resolved Report
Include the custom title in the final report.
Available Items
This box lists the items that are available to add to the text object. To add an item to the
Selected list, select it to highlight it and click Add.
The Available Items list varies, depending on which text object is selected:
• Component ISTD Settings Summary Object
• Component Settings Summary Object
• Component Sys Suit/Flags Settings Object
• Instrument Method Object
• Processing Method – General Object
• Processing Method – Qual Object
• Quan ISTD Peak Summary Object
• Quan Peak Summary Object
• Report Info Object
• Sample Header Object
Add
Add the currently highlighted item in the Available Items list to
the Selected list.
Add All
Add all of the items in the Available Items list to the Selected list.
Remove
Remove the currently highlighted item from the Selected list.
Remove All
Remove all of the items from the Selected list.
Selected
This box lists the items that are already selected to be displayed in the text object.
Move Up, Move Down Rearrange the order of the items in the Selected list by moving the
currently highlighted item up or down in the list.
174
XReport User Guide
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Table 56. Data page parameters (Sheet 2 of 2)
Parameter
Description
Data Label
Show the data label or heading text for the currently highlighted
data label or heading in the Selected list. Edit the data labels or
headings by typing in this box.
Decimal Setting
Specify the number of digits displayed after the decimal point for
values displayed with the selected data label. This parameter is
only available for some data labels.
Attributes Page
Format the currently selected text object on the Attributes page.
Table 57 describes the parameters on the Text Properties – Attributes page.
Table 57. Attributes page parameters
Parameter
Description
Number of Columns
1, 2, or 3
Display the data in 1, 2, or 3 columns.
Horizontal Spacing
Include Data Label
Show or hide the data labels adjacent to the data.
Include Spacing
Between Data Label
and Its Data
Add space between data labels and the data so that all data values
are aligned.
Vertical Spacing
Single, 1.5 Lines, or
Double
Thermo Scientific
Set the amount of space between each line in the text object.
XReport User Guide
175
D
XReport Application Features
XReport Dialog Boxes
Attributes Page for the Annotation Text Object
Format the currently selected Annotation text object on the Attributes page (Figure 85).
Figure 85. Attributes page of the Annotation Properties dialog box
Table 58 describes the parameters on the Annotation Properties – Attributes page.
Table 58. Annotation Properties – Attributes page parameters
Parameter
Description
Number of Columns
1, 2, or 3
Display the data in 1, 2, or 3 columns.
Horizontal Alignment
Left, Center, or Right
Aligns the text to the left, center, or right.
Vertical Spacing
Single, 1.5 Lines, or
Double
176
XReport User Guide
Set the amount of space between each line in the text object.
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Font Page
Format the currently selected text object on the Font page.
The parameters in the Font page depend on whether you are configuring an annotation or
other text object.
The Annotation Font page has the standard Microsoft Windows font chooser. Here you can
choose font, font size, color, emphasis, and so on.
Table 59 describes the parameters for the Font page.
Table 59. Font page parameters
Parameter
Description
Title Font,
Heading Font,
Data Label Font,
Data Font
Set and view the font settings for all of the title, headings, data
labels, and data in the text object.
To configure your font, click
. The Font Dialog Box opens.
Bitmap Properties Dialog Box
Specify the filename of the currently selected bitmap in the Bitmap Properties dialog box. For
more information, see Configuring a Bitmap Object.
Table 60 describes the parameters for the Bitmap Properties dialog box.
Table 60. Bitmap Properties dialog box parameters
Parameter
Description
Bitmap File
Specify the filename for the currently selected bitmap. Type the
full path or click the Browse button to browse for the file.
Reset to Original Size
Reset the size of the image in the template to the original size in
the file.
Column Properties Dialog Box
Set parameters in the Column Properties dialog box to specify the number of columns for the
column object and whether to repeat by sample or summary.
For more information about using the Columns object, see Using Columns.
Thermo Scientific
XReport User Guide
177
D
XReport Application Features
XReport Dialog Boxes
Table 61 describes the parameters for the Column Properties dialog box.
Table 61. Column Properties dialog box parameters (Sheet 1 of 2)
Parameter
Description
Number of Columns
2, 3, 4, or
5
Specify the number of columns in the column object.
Repeat Type
Sample
repeat
View information relating to a particular raw data file of a sample and the results created when
processing it. Use this repeat type as a layout tool to indicate where objects are placed on the page of the
resolved report.
Example 1: When you place a 2-column object in a Quan repeating section, add a chromatogram to the
right column, and resolve the template given data of 1 sample with 4 components each, the report
appears as follows:
Resolve
Chrom 1
Chrom
Chrom 2
Sample 1
Chrom 3
Chrom 4
Example 2: Place a 2-column object in a Quan repeating section, add a chromatogram to the left
column, a spectrum to the right column, and a text object below the column. Then, combine the data
with the template given data of 1 sample with 4 components. All objects in the section are repeated by
sample, and the report appears as follows:
Resolve
Chrom
Text
Spec
Chrom 1
Spec 1
Text
Chrom 2
Spec 2
Text
Sample 1
Chrom 3
Spec 3
Text
Chrom 4
Spec 4
Text
178
XReport User Guide
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Table 61. Column Properties dialog box parameters (Sheet 2 of 2)
Parameter
Description
Summary
repeat
View information relating to all of the samples in an Xcalibur sequence and the results created when
processing them. This setting is ignored when the column object is not located in a nonrepeating section.
Example 3: If you place a 2-column object in a nonrepeating section, add a chromatogram (in either
column), and resolve the template given data of 2 samples with 4 components each, the report appears as
follows:
Resolve
Chrom
Left to
right
Chrom 1
Chrom 2
Chrom 3
Chrom 4
Chrom 1
Chrom 2
Chrom 3
Chrom 4
Sample 1
Sample 2
If you choose instead the Top To Bottom option, the report appears as follows:
Resolve
Chrom
Top to
bottom
Chrom 1
Chrom 3
Chrom 2
Chrom 4
Chrom 1
Chrom 3
Chrom 2
Chrom 4
Sample 1
Sample 2
If you place an object in the right column, and leave the left column empty, the data is resolved as if the
object is placed in the left column, as illustrated above.
If you place an object in each column, the Xcalibur data system ignores the Top To Bottom and Left To
Right options. The report appears as follows:
Resolve
Chrom
Text
Spec
Chrom 1
Spec 1
Chrom 2
Spec 2
Chrom 3
Spec 3
Chrom 4
Spec 4
Chrom 1
Spec 1
Chrom 2
Spec 2
Chrom 3
Spec 3
Chrom 4
Spec 4
Sample 1
Sample 2
Text
If you click the New Sample New Row option, data for a new sample always starts on a new row.
Thermo Scientific
XReport User Guide
179
D
XReport Application Features
XReport Dialog Boxes
Component Calibration Curve Dialog Box
Set parameters in the Component Calibration Curve dialog box to specify the component
name that a nonrepeating Component Calibration Curve report object is plotted for in your
report.
Table 62 describes the parameters for the Component Calibration Curve dialog box.
Table 62. Component Calibration Curve dialog box parameters
Parameter
Description
Component Name
Specify the name of the component to display in your component
calibration curve.
Click the Browse button to open a dialog box and select a
processing method file. After you select a processing method file,
XReport provides a list of the components to choose from in that
file. After you select a component, XReport returns to the
Component Calibration Curve Dialog Box and adds the selected
component to the Component Name box.
Data Sources Dialog Box
Select data files to apply to the current report in the Data Sources dialog box.
Table 63 describes the parameters for the Data Sources dialog box.
Table 63. Data Sources dialog box parameters
180
XReport User Guide
Parameter
Description
Calibration File
Select the calibration (XCAL) file to apply to the current
report.
Processing Method File
Select the processing method (PMD) file to apply to the
current report.
Raw Data File
Select the raw (RAW) file to apply to the current report.
Result File
Select the result (RST) file to apply to the current report.
Sequence List File
Select the sequence list (SLD) file to apply to the current
report.
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Electronic Signature Properties Dialog Box
Configure the order of information displayed in the Electronic Signature Table Object object
in the Electronic Signature Properties dialog box.
Table 64 describes the parameters for the Electronic Signature Properties dialog box.
Table 64. Electronic Signature Properties dialog box parameters
Parameter
Description
Signature Information Order
Created By First
Display the Created By signature block first. This block
displays the electronic signature information for the person
who created the report.
Signed By First
Display the Signed By signature block first. This block
displays the electronic signature information for the people
who signed the report.
Font Dialog Box
Use the Font dialog box to select the global fonts for all objects in your report or customized
fonts for the selected objects in the report.
This dialog box has three pages that are the same as in the individual Font Properties pages
that you access when you configure an individual object:
• Annotation (see Font Page)
• Text (see Font Page)
• Tables (see Table Properties Wizard)
Preferences Dialog Box
Set grid and configuration options for your report in the Preferences dialog box.
Table 65 describes the parameters for the Preferences dialog box.
Table 65. Preferences dialog box parameters (Sheet 1 of 2)
Parameter
Description
Grid Spacing
Thermo Scientific
Width
Set the width between grid lines. (1 grid line unit = 0.1 inches)
Height
Set the height between grid lines. (1 grid line unit = 0.1 inches)
XReport User Guide
181
D
XReport Application Features
XReport Dialog Boxes
Table 65. Preferences dialog box parameters (Sheet 2 of 2)
Parameter
Description
Actions
Snap To Grid
Snap objects to the grid. For Snap To Grid to work, the grid must
be displayed (choose View > Report Template Grid).
Show Page End
Display a dashed blue line that indicates the location for page
breaks.
Show Template
Configuration
Show the Template Configuration Dialog Box when you open
XReport or choose File > New. Use the Template Configuration
dialog box to select a type of report to create (for example, the
Xcalibur data system). This option is available only when two or
more types of report templates are installed on the system (for
example, both the Xcalibur data system and the LCquan™
application.)
Preview Window
Preview and adjust the appearance of your report using the parameters in the Preview window.
In previewing the appearance of your report, adjust your template before you combine it with
actual data.
 To open the preview window
Choose Report > Simulate Report.
XReport displays a preview of your template filled with sample data.
 To view the report’s appearance with actual data
Choose Report > Resolve Report
The application shows a preview of the report combined with your actual data.
182
XReport User Guide
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
Table 66 describes the parameters for the Preview window.
Table 66. Preview window parameters
Parameter
Description
Print
Print your report preview. This option is available only for
resolved reports, not for simulated reports.
Save
Save your report in the following formats (available only for
resolved reports, not simulated reports):
• Microsoft word document (DOC file)
• Microsoft Excel workbook (XLS file)
• Web page (HTM or HTML file)
• Plain text (TXT file)
• Rich text format (RTF file)
• Adobe™ PDF (PDF file)
Next Page/Prev Page
Scroll the preview to the next or previous page in the report.
First Page/Last Page
Scroll the preview to the first or last page in the report.
Page X of Y
View X, the current page, and Y, the total number of pages in the
report. Type in a new value for X to scroll to that page.
Two Page
View the preview two pages at a time.
Zoom In/Zoom Out
Increase or decrease the size of the report in the preview window.
Close
Close the preview window and go to XReport.
Help
View XReport Help.
Save As Dialog Box
Use the Save As dialog box to save your report template.
The Save As dialog box is the standard Microsoft Save dialog box, with the addition of a Lock
Template check box. If you select this check box, the template cannot be saved with the same
name again. It can, however, be opened, edited, and resolved with data. To save an edited
version, save it as a different filename.
The title bar of XReport indicates whether the displayed template is locked or unlocked. For a
locked template, Locked appears in the title bar. For an unlocked template, nothing appears in
the title bar.
There is no audit trail when a new template is created or when an existing template is
changed.
Thermo Scientific
XReport User Guide
183
D
XReport Application Features
XReport Dialog Boxes
Template Configuration Dialog Box
Select the type of report to create in the Template Configuration dialog box. This dialog box
is only displayed if you have more than one type of report installed (for example, Xcalibur and
LCquan).
Table 67 describes the parameters for the Template Configuration dialog box.
Table 67. Template Configuration dialog box parameters
Parameter
Description
Report Types
View a list of available report types. Select the program for which
you would like to create a report.
Do Not Show This
Dialog Again
Hide this dialog box when you open XReport or choose File >
New again. To redisplay this dialog box, choose Layout >
Preferences and select the Show Template Configuration check
box.
Page Setup Dialog Box
Use the Page Setup dialog box to set up the page layout for reports printed from the Qual
Browser window, Sequence Setup view, or the XReport application.
The page setup settings for the Qual Browser window, Sequence Setup view, and XReport
application are independent of each other.
 To set up the page layout for a report template
1. Do one of the following:
• From the menu bar of the Qual Browser window, Sequence Setup view, or XReport
application, choose File > Page Setup.
–or–
• In the XReport application, set up the page breaks for the report template as follows:
i.
Drag the Page Break object to the template pane.
ii. Right-click the Page Break object and choose Properties from the shortcut
menu.
The Page Setup dialog box opens.
2. Set up the following properties:
a. In the Paper area, do the following:
• Select the paper size.
• Select the paper source.
184
XReport User Guide
Thermo Scientific
D
XReport Application Features
XReport Dialog Boxes
b. In the Orientation area, select the Portrait or Landscape option.
c. In the Margins area, type the appropriate margins in the Left, Right, Top, and
Bottom boxes.
The range of valid values depends on the paper size and page orientation.
d. (XReport only) In the Heights area, do the following:
• Type the appropriate value for the placement of the header in the Header box.
• Type the appropriate value for the placement of the footer in the Footer box.
As you change the properties, the graphic at the top of the dialog box displays the effect of
the changes.
3. Click OK to accept the changes and return to the template pane.
Table 68 describes the parameters in the Page Setup dialog box.
Table 68. Page Setup dialog box parameters
Parameter
Description
Paper
Size
Specifies the paper size for printing.
Default selection: Letter
Source
Specifies the paper source. Some printers have multiple paper
trays.
Default selection: Automatically Select
Orientation
Portrait
Selecting this option prints pages vertically.
Landscape
Selecting this option prints pages horizontally.
Margins (inches)
Left, Right, Top, and
Bottom
Adjusts the space between the edge of the document and the
corresponding edge of the page.
Heights (inches) XReport application only
Header
Specifies the distance of the header from the top of the page.
Footer
Specifies the distance of the footer from the bottom of the page.
Buttons
Thermo Scientific
OK
Accepts the new settings and closes the dialog box.
Cancel
Cancels the new settings and closes the dialog box.
XReport User Guide
185
D
XReport Application Features
Report Sections
Report Sections
You can divide your report template into sections and add report objects to those sections. To
add sections into a report template, drag them from the Sections pane.
The types of sections used in XReport are as follows:
• Nonrepeating Section
• Report Header and Footer
• Qual Repeating Section
• Quan Repeating Section
A report object that is placed in a repeating section can behave differently from the same
report object placed in a nonrepeating section. See Report Objects for information on how a
report object functions when it is placed inside or outside a repeating section.
Nonrepeating Section
A nonrepeating section is any portion of a report template that is not classified as a repeating
section. A nonrepeating section usually displays static details associated with either data or
methods that apply once and globally within the system.
A nonrepeating section can have these objects:
• Formatting Objects
–
Page Break Object
–
Columns Object
• Graphic Objects
186
XReport User Guide
–
Electronic Signature Table Object
–
Bitmap Object
–
Chromatogram Object
–
Component Cal Curve Object
–
Spectrum Object
Thermo Scientific
D
XReport Application Features
Report Sections
• Table Objects
–
Avalon Qual Events Table Object
–
Calibration File Table Object
–
Component Cal Level Table Object
–
Dilution Factor Table Object
–
IRC Summary Table Object
–
Processing Method Programs Table Object
–
Processing Method Sample Reports Settings Table Object
–
Processing Method Summary Reports Settings Table Object
–
Qual Peak Table Object
–
Quan Peak Table Object
–
Sequence Table Object
–
Spectrum Candidate Results Table Object
• Text Objects
–
Annotation Object
–
Instrument Method Object
–
Processing Method – General Object
–
Processing Method – Qual Object
–
Sample Table Object
Report Header and Footer
The report header and footer sections contain data to be repeated at the top and bottom of
each page in your report, for example, the report date, page number, and a company logo.
A Report Header or Footer section can have these objects:
• Formatting Objects
–
Columns Object
• Graphic Objects
Thermo Scientific
–
Bitmap Object
–
Electronic Signature Table Object
–
Table Objects
XReport User Guide
187
D
XReport Application Features
Report Sections
• Text Objects
–
Annotation Object
–
Report Info Object
–
Sample Header Object
Repeating Sections
A repeating section is a section of a report template that is automatically repeated for every
component or peak in a resolved document. You can change the properties of a repeating
section to customize its behavior. XReport can have two types of repeating sections:
• Qual Repeating Section
• Quan Repeating Section
Qual Repeating Section
A Qual Repeating section displays details relating to a single Qual peak within some data.
This section is then repeated for all Qual peaks found.
A Qual Repeating section can have these objects:
• Formatting Objects
–
Page Break Object
–
Columns Object
• Graphic Objects
–
Electronic Signature Table Object
–
Bitmap Object
–
Chromatogram Object
–
Spectrum Object
• Table Objects
–
Library Search Graphics Table Object
–
Library Search Results Table Object
–
Spectrum List Table Object
• Text Objects
–
188
XReport User Guide
Annotation Object
Thermo Scientific
D
XReport Application Features
Report Sections
Quan Repeating Section
A Quan repeating section is also called a Component section. Its purpose is to display details
relating to a single Quan peak in some data or a component configured in a processing
method. This section is then repeated for all Quan peaks or components.
A Quan repeating section can have the following objects:
• Formatting Objects
–
Page Break Object
–
Columns Object
• Graphic Objects
–
Electronic Signature Table Object
–
Bitmap Object
–
Chromatogram Object
–
Component Cal Curve Object
–
Spectrum Object
• Table Objects
–
Avalon Quan Events Table Object
–
Component Cal Level Table Object
–
IRC Settings Table Object
–
Component QC Level Table Object
–
Component Spectrum Table Object
–
IRC Settings Table Object
–
Spectrum List Table Object
–
Sample Table Object
• Text Objects
Thermo Scientific
–
Annotation Object
–
Component IdentDetectCal Settings Object
–
Component ISTD Settings Summary Object
–
Component Settings Summary Object
–
Component Sys Suit/Flags Settings Object
–
Quan ISTD Peak Summary Object
–
Quan Peak Summary Object
XReport User Guide
189
D
XReport Application Features
Report Objects
Report Objects
Report objects are predefined chunks of data that you can drag to the XReport template from
the Objects pane. When you combine a report template with Xcalibur data source files,
XReport replaces the report objects in the template with data to generate an Xcalibur report.
Report objects include the following:
• Graphic and Formatting Objects
• Table Objects
• Text Objects
Graphic and Formatting Objects
Graphic objects display graphics. Graphic objects can be configured and customized. The
graphic object is replaced with a graphic when a report template is resolved with Xcalibur
data. Formatting objects help to configure the layout of the report.
The following graphic and formatting objects are available in XReport:
• Chromatogram Object
• Spectrum Object
• Component Cal Curve Object
• Bitmap Object
• Columns Object
• Page Break Object
Chromatogram Object
The Chromatogram object displays a graphical trace that can show either raw data or
processed data for any acquisition device that is supported by the Xcalibur data system (MS,
Analog, A/D card, PDA, UV).
Required data file: RAW or RST
Section placement: Qual, Quan, and nonrepeating sections
190
XReport User Guide
Thermo Scientific
D
XReport Application Features
Report Objects
Spectrum Object
The Spectrum object displays a graphical spectrum that can show either raw data or processed
data for any acquisition device supported by the Xcalibur data system (MS, PDA).
Required data file: RAW or RST
Section placement: Qual, Quan, and nonrepeating sections
Component Cal Curve Object
The Component Calibration Curve object displays the calibration curve that is used to
quantify a specified component. The curve is specified in the processing method as the
response versus the amount of standard.
Required data file: PMD or RST
Section placement: Quan or nonrepeating sections. For the nonrepeating section, specify
which component to display in the Component Calibration Curve Dialog Box.
Bitmap Object
The Bitmap object displays a graphical image, for example, a company logo.
Required data file: BMP, DIB, GIF, JPG, or ICO
Section placement: possible in all sections
Columns Object
Use the Columns object to manipulate the layout and repeating properties of graphic and text
objects. Format the columns in the Column Properties Dialog Box.
Required data file: none
Section placement: possible in all sections.
Page Break Object
Use the Page Break object to insert a manual page break in your report.
Required data file: None
Section placement: Possible in Qual repeating, Quan repeating, and nonrepeating sections.
Thermo Scientific
XReport User Guide
191
D
XReport Application Features
Report Objects
Table Objects
Table objects contain tabular information. The columns of table objects can be added,
deleted, and ordered. You can also create custom columns. The columns in the table fill with
data when a report template is resolved with Xcalibur data.
These table objects are available in XReport:
• Avalon Qual Events Table Object
• Avalon Quan Events Table Object
• Calibration File Table Object
• Component Cal Level Table Object
• Component QC Level Table Object
• Component Spectrum Table Object
• Dilution Factor Table Object
• Electronic Signature Table Object
• Ion Ratio Confirmation Results Table Object
• IRC Settings Table Object
• IRC Summary Table Object
• Library Search Graphics Table Object
• Library Search Results Table Object
• Processing Method Programs Table Object
• Processing Method Sample Reports Settings Table Object
• Processing Method Summary Reports Settings Table Object
• Qual Peak Table Object
• Quan Peak Table Object
• Qual Summary Table Object
• Quan Summary Table Object
• Sample Table Object
• Sequence Table Object
• Spectrum Candidate Results Table Object
• Spectrum List Table Object
192
XReport User Guide
Thermo Scientific
D
XReport Application Features
Report Objects
Avalon Qual Events Table Object
The Avalon Qual Events table displays details of the Avalon peak detection algorithm events
configured in the Qual section of a processing method. This table provides information for
each peak with each sample. There is a separate row for each sample peak combination. The
rows are grouped so that all peaks in a sample are shown together.
Required data file: PMD
Section placement: Nonrepeating section
This table has the following possible fields:
• Time (ms)
• Kind
• Event OPCode
• Event
• Value
• Value 2
• User Columns 1 to 5
Avalon Quan Events Table Object
The Avalon Quan Events table displays details of the Avalon peak detection algorithm events
configured for a component in the Quan section of a processing method. This table provides
information for each peak with each sample. There is a separate row for each sample peak
combination. The rows are grouped so that all peaks in a sample are shown together.
Required data file: processing method (PMD)
Section placement: Nonrepeating section
This table has the following possible fields:
• Time
• Kind
• Event OPCode
• Event
• Value
• Value2
• User Columns 1 to 5
Thermo Scientific
XReport User Guide
193
D
XReport Application Features
Report Objects
Calibration File Table Object
The Calibration File table displays calibration levels, raw data file names, and acquisition
dates of standards. This table provides information for each peak with each sample.
Required data file: calibration file (XCAL), sequence file (SLD), or raw data file (RAW)
Section placement: Nonrepeating section
This table has the following possible fields:
• Level
• Response Factor
• Component Name
• Average Response Factor
• %RSD
• User Columns 1 to 5
Component Cal Level Table Object
The Component Cal Level table displays the calibration levels, amounts, and acceptance test
values defined for a specific component.
Required data file: processing method (PMD)
Section placement: Nonrepeating section
This table has the following possible fields:
• Component Name
• Cal Level
• Amount
• User Columns 1 to 5
194
XReport User Guide
Thermo Scientific
D
XReport Application Features
Report Objects
Component QC Level Table Object
The Component QC Level table displays the QC levels, amounts, and percent test values
defined for a specific component.
Required data file: processing method (PMD)
Section placement: Nonrepeating section
This table has the following possible fields:
• QC Level
• Amount
• Test (%)
• Component Name
• User Columns 1 to 5
Component Spectrum Table Object
The Component Spectrum table displays m/z and intensity data for up to fifty spectrum
peaks. The Xcalibur data system uses this data to identify components when the Spectrum
Peak Identification Method is specified in the processing method. This table provides
information for each peak with each sample. There is a separate row for each sample peak
combination. The rows are grouped so that all peaks in a sample are shown together.
Note This method is only used for GC chromatography.
Required data file: processing method (PMD)
Section placement: Nonrepeating section
This table has the following possible fields:
• #
• m/z
• Intensity (%)
• User Columns 1 to 5
Thermo Scientific
XReport User Guide
195
D
XReport Application Features
Report Objects
Dilution Factor Table Object
The Dilution Factor table displays the names of calibration levels and dilution factors for
components other than internal standards.
Required data file: result file (RST)
Section placement: Nonrepeating section
This table has the following possible fields:
• Calibration Level
• Dilution
• User Columns 1 to 5
Electronic Signature Table Object
The Electronic Signature table object displays up to six electronic signatures in table format.
Required data file: none
Section placement: possible in all sections.
Note The data system generates Electronic Signature information at the time the report is
created, and the signature information can only be included in reports generated from
either the Xcalibur Sequence Setup or Quan Browser. Reports generated from within
XReport do not display signature information.
The Electronic Signature table object displays two blocks of signatures. The Created By block
displays the signature of the person who created the report. The Signed By block displays the
signatures of the people who entered electronic signature information at the time the report
was created. This object displays the electronic signature data in the following columns:
• Logon Name
• Full Name
• Date and Time
• Comment
196
XReport User Guide
Thermo Scientific
D
XReport Application Features
Report Objects
Ion Ratio Confirmation Results Table Object
The Ion Ratio Confirmation Results table displays data about each qualifier ion, based on
search criteria specified in the processing method.
Required data file: result file (RST)
Section placement: Nonrepeating section
This table has the following possible fields:
• Qualifier Ion #
• Mass
• Qualifier Ion Found
• Coelution Passed
• Ion Ratio Passed
• Absolute Window (%)
• Actual Ion Ratio (%)
• Target (%)
• Target Range (%)
• User Columns 1 to 5
IRC Settings Table Object
The IRC Settings Table displays up to five qualifier ions within tolerance limits to confirm the
detection of a target analyte. The ions are specified in the processing method.
Required data file: processing method (PMD)
Section placement: Nonrepeating section
This table has the following possible fields:
• Qualifier Ion #
• m/z
• Target Ratio (%)
• Window (±%)
• User Columns 1 to 5
Thermo Scientific
XReport User Guide
197
D
XReport Application Features
Report Objects
IRC Summary Table Object
The IRC Summary table displays all of the information for the qualifier ion for each
component. Most useful when the Name column is the first column.
Required data file: result file (RST)
Section placement: Nonrepeating section
This table has the following possible fields:
• Component Name
• RT
• Mass
• Response
• Calculated Amount
• Actual Ion Ratio (%)
• Ion Ratio Passed
• Target Range (%)
• Qualifier Ion #
• User Columns 1 to 5
Library Search Graphics Table Object
The Library Search Graphics table displays the graphical results from a library search on a
Qual peak.
Required data file: result file (RST) and raw data file (RAW)
Section placement: Nonrepeating section
This table has the following possible fields:
• Hit Spectrum
• Delta
• Compound Structure
198
XReport User Guide
Thermo Scientific
D
XReport Application Features
Report Objects
Library Search Results Table Object
The Library Search Results table displays the results of a library search on a Qual peak.
Required data file: result file (RST) or raw data file (RAW)
Section placement: Non-repeating section
This table has the following possible fields:
• SI
• RSI
• Compound Name
• Cas #
• Probability
• Area
• Area (%)
• Library
• Molecular Formula
• Molecular Weight
• RT
• Scan #
• User Columns 1 to 5
Thermo Scientific
XReport User Guide
199
D
XReport Application Features
Report Objects
Processing Method Programs Table Object
The Processing Method Programs table displays a list of programs to be run by the data
system after the sample has been analyzed and the data has been processed. The application
runs programs in the order listed.
Required data file: processing method (PMD)
Section placement: Nonrepeating section
This table has the following possible fields:
• Num
• Enable
• Std
• QC
• Unk
• Action
• Other
• Parameters
• Program or Macro Name
• Sync
• User Columns 1 to 5
200
XReport User Guide
Thermo Scientific
D
XReport Application Features
Report Objects
Processing Method Sample Reports Settings Table Object
The Processing Method Sample Reports Settings table displays sample report information for
each sample in a sequence, including the template name and the report file name. Specify the
sample type in the processing method.
Required data file: processing method (PMD)
Section placement: Nonrepeating section
This table has the following possible fields:
• Num
• Enable
• Std
• QC
• Unk
• Other
• Report Template Name
• Save As
• User Columns 1 to 5
Processing Method Summary Reports Settings Table Object
The Processing Method Summary Reports Settings table displays summary reports to be
issued after processing the samples in a bracketed or non-bracketed sequence.
Required data file: processing method (PMD)
Section placement: Nonrepeating section
This table has the following possible fields:
• Num
• Enable
• Save As
• Report Template Name
• User Column 1 to 5
Thermo Scientific
XReport User Guide
201
D
XReport Application Features
Report Objects
Qual Peak Table Object
The Qual Peak table displays information specific to chromatogram peaks detected by
qualitative analysis for a single sample.
Note To display information on chromatogram peaks for an entire sequence, use the Qual
Summary Table Object.
Required data file: result file (RST)
Section placement: Nonrepeating section
This table has the following possible fields:
• RT
• Peak Area
• Peak Height
• BL
• S/N
• Area (%)
• Peak Width
• Saturated
• User Column 1 to 5
202
XReport User Guide
Thermo Scientific
D
XReport Application Features
Report Objects
Quan Peak Table Object
The Quan Peak table displays information specific to component peaks detected by
quantitative analysis for a single sample.
Note To display component peak information for all samples in a sequence, use the Quan
Summary Table Object object.
Required data file: processing method (PMD), result file (RST)
Section placement: Nonrepeating section
The Quan Peak table has the following possible fields:
• Component Name
• ISTD Area
• RT
• ISTD Height
• Calculated Amount
• ISTD Name
• Response
• ISTD Response
• BL
• Keys
• %Diff
• Peak Purity
• Area
• Peak Status
• Area (%)
• Response Ratio
• Area Ratio
• S/N
• Calibration Flags
• Sample Type
• Component Found
• Saturated
• Component Type
• Specified Amount
• Data Flags
• System Suitability
• Expected RT
• Units
• Height
• Width
• Height Ratio
• User Columns 1 to 5
• Ion Ratio Status
Thermo Scientific
XReport User Guide
203
D
XReport Application Features
Report Objects
Qual Summary Table Object
The Qual Summary table displays information specific to chromatogram peaks detected by
qualitative analysis for all samples in a sequence.
Note To display peak information for a single sample only, use the Qual Peak Table
Object object.
Required data file: sequence file (SLD) or result file (RST)
Section placement: Nonrepeating section
The Qual Summary table has the following possible fields:
• RT
• Peak Area
• Peak Height
• BL
• Data File Name
• Sample ID
• S/N
• Area (%)
• Peak Width
• Saturated
• User Column 1 to 5
204
XReport User Guide
Thermo Scientific
D
XReport Application Features
Report Objects
Quan Summary Table Object
The Quan Summary table displays information specific to component peaks detected by
quantitative analysis for all samples in a sequence.
Note To display component peak information for a single sample only, use the Quan Peak
Table Object object.
Required data file: sequence file (SLD), processing method (PMD), result file (RST)
Section placement: Nonrepeating section
The Quan Summary table has the following possible fields:
• Component Name
• ISTD Area
• RT
• ISTD Height
• Calculated Amount
• ISTD Name
• Response
• ISTD Response
• BL
• Keys
• %Diff
• Peak Purity
• Area
• Peak Status
• Area (%)
• Response Ratio
• Area Ratio
• S/N
• Calibration Flags
• Sample ID
• Component Found
• Sample Type
• Component Type
• Saturated
• Data File Name
• Specified Amount
• Data Flags
• System Suitability
• Expected RT
• Units
• Height
• Width
• Height Ratio
• User Columns 1 to 5
• Ion Ratio Status
Thermo Scientific
XReport User Guide
205
D
XReport Application Features
Report Objects
Sample Table Object
The Sample table displays system suitability and flag processing results for each component in
a sample, and repeats this information for every sample in a sequence. This table provides
information for each peak with each sample. There is a separate row for each sample peak
combination. The rows are grouped so that all peaks in a sample are shown together.
Required data file: sequence file (SLD)
Section placement: Nonrepeating section
The Sample table has the following possible fields:
• Sample ID
• Data File Name
• Area
• Area Ratio
• ISTD Area
• Specified Amount
• %Diff
• %RSD
• Actual RT
• Calculated Amount
• Calibration Flags
• Data Flags
• Excluded
• Height
• ISTD Height
• Height Ratio
• QC Failed
• System Suitability
• User Columns 1 to 5
206
XReport User Guide
Thermo Scientific
D
XReport Application Features
Report Objects
Sequence Table Object
The Sequence table displays the contents of the sequence file. This table provides information
for each peak with each sample.
Required data file: sequence file (SLD
Section placement: Nonrepeating section
The Sequence table has the following possible fields:
• Sample Type
• File Name
• Sample ID
• Path
• Inst Method
• Cal File
• Dil Factor
• Inj Vol
• ISTD Corr Amt
• Level
• Proc Method
• Sample Vol
• Sample Wt
• Position
• User Labels 1 to 5
• User Columns 1 to 5
Thermo Scientific
XReport User Guide
207
D
XReport Application Features
Report Objects
Spectrum Candidate Results Table Object
The Spectrum Candidate Results table displays the results of spectrum candidate processing
for GC/MS data on a per component basis.
Required data file: result file (RST)
Section placement: Nonrepeating section
The Spectrum Candidate Results table has the following possible fields:
• Component Name
• Cand. #
• Forward Fit
• Reverse Fit
• Match
• Area
• End RT
• Found RT
• Height
• Start RT
• User Columns 1 to 5
Spectrum List Table Object
The Spectrum List table displays the text equivalent of the spectrum graphic object. This table
provides information for each peak with each sample. There is a separate row for each sample
peak combination. The rows are grouped so that all peaks in a sample are shown together.
Required data file: result file (RST) or raw data file (RAW)
Section placement: Nonrepeating section
The Spectrum List table has the following possible fields:
• m/z
• Intensity
• Relative
• Wavelength
• User Columns 1 to 5
208
XReport User Guide
Thermo Scientific
D
XReport Application Features
Report Objects
Text Objects
Text objects contain text labels and associated text fields. These fields can be deleted,
cut/copied, and pasted anywhere in the report template (for example, a header). The text
object fields are replaced with textual information when a report template is resolved with
Xcalibur data.
These text objects are available in XReport:
• Annotation Object
• Component IdentDetectCal Settings Object
• Component ISTD Settings Summary Object
• Component Settings Summary Object
• Component Sys Suit/Flags Settings Object
• Instrument Method Object
• Processing Method – General Object
• Processing Method – Qual Object
• Quan ISTD Peak Summary Object
• Quan Peak Summary Object
• Report Info Object
• Sample Header Object
Annotation Object
Use the Annotation object to insert static text into your report. To enter the text, right-click
the object and choose Properties from the shortcut menu. The Annotation Properties dialog
box opens. Type your text in the box.
Required data file: None
Section placement: Possible in all sections.
Component IdentDetectCal Settings Object
The Component IdentDetectCal Settings object shows the peak detection, calibration, and
integration parameters for a single component.
Required data file: processing method (PMD)
Section placement: Quan repeating section, repeats on component peak
For this object’s possible fields and headings, see Table 69.
Thermo Scientific
XReport User Guide
209
D
XReport Application Features
Report Objects
Component ISTD Settings Summary Object
The Component ISTD Settings object displays a summary of the configuration details for
each component’s ISTD configuration in the Quantitation section of a processing method.
Required data file: processing method (PMD)
Section placement: Quan repeating section, repeats on component peak
Table 69 lists the this object’s possible fields.
Table 69. Component ISTD Settings object fields
• Component Name
• Expected Retention
Time
• Detector Type
• Filter
• Trace
• Mass Range
Genesis Peak
Integration/Detection
• Retention Time
Window
• Genesis Smoothing
Points
• View Width
• Valley Detection
Enabled
• Wavelength Range 1
• Retention Time
Reference
• Wavelength Range 2
• Component Type
• ISTD Name
• Peak Detection
Algorithm
• Genesis Constrain Peak
Width
• Genesis Identify By
• Genesis Minimum
Peak Height
ICIS Peak
Integration/Detection
• ICIS Smoothing Points
• ICIS Constrain Peak
Width
• ICIS Identify By
• ICIC Minimum Peak
Height
Avalon Peak
Integration/Detection
• Avalon Smoothing
Points
Other
• <Blank Line > and
<User Text> Data
Items
Component Settings Summary Object
The Component Settings Summary object displays a short list of peak detection and
integration parameters for a specified component.
Required data file: processing method (PMD)
Section placement: Quan repeating section, repeats on component peak
Table 70 lists this object’s possible fields and headings.
210
XReport User Guide
Thermo Scientific
D
XReport Application Features
Report Objects
Table 70. Component Settings Summary object fields/headings
• Component Name
• Detector Type
• Filter
• Trace
• Mass Range
• Expected Retention
Time
Genesis Peak
Integration/Detection
• Retention Time
Window
• Genesis Smoothing
Points
• View Width
• Valley Detection
Enabled
• Wavelength Range 1
• Retention Time
Reference
• Wavelength Range 2
• Component Type
• ISTD
• Standard Dilution –
Base Amount
• Peak Detection
Algorithm
• Genesis Constrain Peak
Width
• Genesis Identify By
• Genesis Minimum
Peak Height
ICIS Peak
Integration/Detection
• ICIS Smoothing Points
• ICIS Constrain Peak
Width
• ICIS Identify By
• ICIC Minimum Peak
Height
Avalon Peak
Integration/Detection
• Avalon Smoothing
Points
Other
• <Blank Line > and
<User Text> Data
Items
Thermo Scientific
XReport User Guide
211
D
XReport Application Features
Report Objects
Component Sys Suit/Flags Settings Object
The Component Sys Suit/Flags Settings object displays calibration, quantitation, and
detection flags along with system suitability parameters, which allow automatic
chromatographic checks that assign a pass/fail qualification to a target peak.
Required data file: processing method (PMD
Section placement: Quan nonrepeating, repeats on component peak
Table 71 lists this object’s possible fields and headings.
Table 71. Component Sys Suit/Flags Settings object fields/headings
System Suitability
Detect Tailing
• Component Name
• Tail Peak Height
• Tail Failure Threshold
Resolution Parameters
• Resolution Threshold
Detect Column Overload
• Over Peak Height
Symmetry Parameters
• Symmetry Peak Height
• Symmetry Threshold
• Over Failure Threshold
Detect Baseline Clipping
Quantitation Flags
• Num. of Peak Widths
for Noise Detection
• Detection Limit
Detect Minimum
Signal-to-Noise Ratio
• Signal-to-Noise Ratio
Calibration and
Quantitation Flags
Calibration Flag
Peak Classification
Parameters
• R-Squared
Detect Peak Width
• Quantitation Limit
• Linearity Limit
• Carry Over Limit
Detection Flags
• Area Threshold
• Height Threshold
Other
• <Blank Line > and
<User Text> Data
Items
• Width Peak Height
• Min. Peak Width
• Max. Peak Width
Instrument Method Object
The Instrument Method object displays general instrument method parameters.
Required data file: raw data file (RAW)
Section placement: Nonrepeating section
This object displays instrument method information for all the instruments associated with a
RAW file. This object uses the default data, attribute, and font settings from Qual Browser,
and has no configurable properties.
212
XReport User Guide
Thermo Scientific
D
XReport Application Features
Report Objects
Processing Method – General Object
The Processing Method – General object displays general processing method parameters.
Required data file: processing method (PMD)
Section placement: Nonrepeating section
Table 72 lists this object’s possible fields and headings.
Table 72. Processing Method – General object fields/headings
Identification
• Created By
• When Created
• Last Modification
• Modified By
• Summary
Baseline
• Baseline and Noise
Window
• Baseline Noise
Tolerance
Options
• Chromatography By
• Calibration By
Other
• <Blank Line > and
<User Text> Data
Items
• Min. Number of Scans
in Baseline
• Void Time
Thermo Scientific
XReport User Guide
213
D
XReport Application Features
Report Objects
Processing Method – Qual Object
The Processing Method – Qual object displays Qual processing method parameters.
Required data file: processing method (PMD)
Section placement: Nonrepeating section
Table 73 lists this object’s possible fields and headings.
Table 73. Processing Method – Qual object fields/headings (Sheet 1 of 2)
Identification
•
•
•
•
•
•
•
•
•
•
•
Detector Type
Detector Delay
Filter
1st Trace Type
Operator
2nd Trace Type
Mass Range 1
Wavelength Range 1
Mass Range 2
Wavelength Range 2
Range
ICIS Advanced
Parameters
•
•
•
•
•
Detection Options
• Peak Detection
Algorithm
Genesis Peak Integration
Avalon Peak Integration
• Avalon Smoothing
Points
Limit Peaks
• Select Top Peaks
• Genesis Smoothing
• Select By
Points
• Number To Select
• S/N Threshold
• Relative Peak Height
• Valley Detection
Threshold
• Expected Peak Width
• Genesis Constrain Peak • Percent of Highest
Peak
Width
• Genesis Peak Height
• Genesis Tailing Factor Advanced Detection
Options
Spectrum Enhancement
Spectrum Enhancement
Noise Method
Usage
Minimum Peak Width
Multiplet Resolution
Area Tail Extension
Area Scan Window
Refine
• Window Size
• Noise Threshold
Genesis Peak Detection
• Genesis Identify By
• Minimum Masses
Required
• Minimum Percent of
Masses Found
• Minimum Peak Height
Genesis Peak Edge
Detection
• Peak S/N Cutoff
• Report Noise As
Combine
Background Subtraction
Left Region
• Left Region Width
• Region End
Peak Top Region
• Width
214
XReport User Guide
Thermo Scientific
D
XReport Application Features
Report Objects
Table 73. Processing Method – Qual object fields/headings (Sheet 2 of 2)
Options
• Maximum Number of
Hits
• Reverse Search
• Search with MW
Append to User Library
•
•
•
•
•
Library Search
Constraints
• Molecular Weight
Range
• Other Databases
• Name Fragment
• Element Constraints
Enabled
• Elements in
Compound
• Element Infiltration
• Mass Spectral Peak
Constraints Enabled
• Mass Spectral Peak
Measurement
Availability
User Library
Match Factor
Reverse Match Factor
Probability
Search Libraries
ICIS Peak Integration
Genesis Apex Detection
•
•
•
•
•
ICIS Smoothing Points
• Window Size
Baseline Window
• Filter Width
Area Noise Factor
Genesis Valley Detection
Peak Noise Factor
ICIS Constrain Peak
• Rise %
Width
• Valley S/N
• ICIS Peak Height
• ICIS Tailing Factor
Genesis Background
Subtraction
• Recomputation
Interval
• Number of Scans in
Background
Thermo Scientific
Background Subtraction
Right Region
• Right Region Width
• Region Start
Threshold
• Cutoff Threshold
Library Search Options
Search Type
Peak Purity Options
• Scan Threshold
• Peak Coverage
• Limit Scan Wavelength
Range
Other
• <Blank Line > and
<User Text> Data
Items
• Identity
• Similarity
XReport User Guide
215
D
XReport Application Features
Report Objects
Quan ISTD Peak Summary Object
The Quan ISTD Peak Summary object displays summary parameters for a Quan
component's associated ISTD peak.
Required data file: processing method (PMD) or result file (RST)
Section placement: Quan nonrepeating, repeats on component peak
Table 74 lists this object’s possible fields.
Table 74. Quan ISTD Peak Summary object fields
• Component Name
• Height
• Signal-to-Noise
• System Suitability
• ISTD Name
• Width
• Saturated
• Data Flags
• Actual RT
• Baseline Flags
• Expected RT
• Response
• Area
• Peak Status
• <Blank Line > and
<User Text> Data
Items
Quan Peak Summary Object
The Quan Peak Summary object displays component peak summary parameters.
Required data file: processing method (PMD) or result file (RST)
Section placement: Quan nonrepeating, repeats on component peak
Table 75 lists this object’s possible fields.
Table 75. Quan Peak Summary object fields
• Component Name
• Area
• Peak Purity
• Component Type
• Actual RT
• Area Ratio
• Baseline Flags
• Peak Status
• Specified Amount
• Height
• Signal-to-Noise
• System Suitability
• Calculated Amount
• Height Ratio
• Saturated
• Data Flags
• %Diff
• Width
• Expected RT
• Calibration Flags
• Response
• Ion Ratio Status
• Sample Type
• <Blank Line > and
<User Text> Data
Items
• Response Ratio
216
XReport User Guide
Thermo Scientific
D
XReport Application Features
Report Objects
Report Info Object
The Report Info object displays general information about the report itself.
Required data file: None
Section placement: Report header and footer
Table 76 lists this object’s possible fields.
Table 76. Report Info object fields
• Page [a] of [b]
• Layered App Version
• Date and Time
• <Blank Line > and <User Text> Data Items
• Template File
• Layered App Name
Sample Header Object
The Sample Header object displays general information about a sample.
Required data file: raw data file (RAW) or sequence file (SLD)
Section placement: Report header or footer or Non-repeating section
Table 77 lists this object’s possible field.
Table 77. Sample Header object fields
• Data File
• Run Time
• Sample Weight
• Revision
• Original Data Path
• Comments
• ISTD Amount
• Barcode
• Current Data Path
• Vial
• Calibration Level
• Barcode Status
• Sample Type
• Injection Volume
• Dilution Factor
• User Labels 1 to 5
• Sample ID
• Scans
• Instrument Method
• Sample Name
• Low Mass
• Processing Method
• <Blank Line > and
<User Text> Data
Items
• Operator
• High Mass
• Acquisition Date
• Sample Volume
• Current Processing
Method
• Calibration File
Thermo Scientific
XReport User Guide
217
D
XReport Application Features
Report Objects
System Suitability Flags
The Quan Summary Table Object, Quan ISTD Peak Summary Object, Sample Table Object,
and Quan Peak Table Object objects have an available System Suitability field, which displays
the results of any system suitability tests carried out on the specified peak. Each test is
represented by a flag letter. XReport displays the flags in categories according to whether the
peak passed, failed, or was not tested.
Table 78 describes the System Suitability flags.
Table 78. System Suitability flags
218
XReport User Guide
Flag
Description
S
Symmetry: A peak passes this test if its calculated symmetry is greater than the
symmetry threshold specified in the System Suitability page of Quan View in
Processing Setup.
R
Resolution: A peak passes this test if its calculated resolution is greater than the
resolution threshold specified in the System Suitability page of Quan View in
Processing Setup.
W
Peak Width: A peak passes this test if its width falls between the minimum and
maximum values specified in the System Suitability page of Quan View in
Processing Setup.
T
Tailing: A peak passes this test if its calculated tailing falls below the threshold
specified in the System Suitability page of Quan View in Processing Setup.
O
Column Overload: A peak passes this test if its calculated overloading falls
below the threshold specified in the System Suitability page of Quan View in
Processing Setup.
B
Baseline Clipping: A peak passes this test if baseline clipping is not detected.
N
Signal to Noise Ratio: A peak passes this test if its signal-to-noise ratio falls
above the threshold specified in the System Suitability page of Quan View in
Processing Setup.
C
Concave: A peak passes this test if it does not exhibit a concave depression due
to noise.
D
Saturation: A peak passes this test if the detector did not saturate during data
acquisition.
Thermo Scientific
D
XReport Application Features
Report Objects
<Blank Line > and <User Text> Data Items
All text objects except for the Annotation object and Instrument Method object have two
special data items, <Blank Line> and <Text>. These data items appear in the Available Items
list of the Data Page.
When you insert the <Blank Line> data item in the Selected list, XReport inserts a blank line
at that spot. You can use this to control the spacing between items in a text object. Unlike
ordinary data items, a <Blank Line> can appear in the Selected list more than once.
The <User Text> item allows you to insert additional custom text into the text object. The
<User Text> item can be added to the Selected list more than once, and each instance can be
configured to display different custom text. Custom text is displayed using the font settings
for Headings.
Baseline Flags
The Quan Peak Summary Object, Quan ISTD Peak Summary Object, Quan Peak Table
Object, Qual Peak Table Object, Qual Summary Table Object, and Quan Summary Table
Object objects have an available Baseline flags field that displays how the data system detected
the left and right peak edges of the current peak. The flags are displayed in pairs, with the first
letter of the pair indicating the detection method for the left edge of the peak, and the second
letter indicating the detection method for the right edge of the peak.
Table 79 describes the baseline flags.
Table 79. Baseline flags
Thermo Scientific
Flag
Description
B
Baseline: The edge of the peak is at baseline level.
V
Valley: The edge of the peak is in a peak valley.
M
Manual: The edge of the peak has been adjusted manually.
S
Stripe: The edge of the peak reached the Constrain Peak Height Percent
specified in the method.
T
Tail: The edge of the peak reached the Constrain Peak Height Tailing Factor
limit before the Height Percent.
–
Tilt: An error occurred before the edge of the peak could be determined.
?
Unknown: An unknown error occurred.
XReport User Guide
219
D
XReport Application Features
Report Objects
Data Flags
XReport can display Data Flag letters in the Quan Summary Table Object, Quan ISTD Peak
Summary Object, Quan Peak Table Object, and Quan Summary Table Object objects to
indicate when a peak has exceeded threshold area or height values specified in Quan Browser.
These are the data flags that can appear.
Table 80 describes the data flags.
Table 80. Data flags
Flag
Description
A
Indicates that the peak area is outside the specified tolerance.
H
Indicates that the height is outside the specified tolerance.
Calibration Flags
XReport can display Calibration Flag letters in the Quan Peak Summary Object, Quan
Summary Table Object, and Quan Peak Table Object objects to indicate when certain peak
characteristics exceed values specified in Quan Browser. These are the calibration flags that
can appear.
Table 81 describes the calibration flags.
Table 81. Calibration flags
220
XReport User Guide
Flag
Description
E
Detection limit: Indicates that the detection threshold for the concentration of
the quantified peak is outside the specified tolerance.
L
Carry Over limit: Indicates that the carryover threshold for the concentration of
the quantified peak is outside the specified tolerance.
Q
R-squared: Indicates that the coefficient of determination is outside the specified
tolerance.
U
Quantitation limit: Indicates that the quantitation threshold for the
concentration of the quantified peak is outside the specified tolerance.
Y
Linearity limit: Indicates that the linearity threshold for the concentration of the
quantified peak is outside the specified tolerance.
Thermo Scientific
D
XReport Application Features
Report Templates
Report Templates
An XReport template is an XReport document (XRT file) that defines the content, layout,
and style of an Xcalibur report. You can include sections and report objects in the report
template. An Xcalibur report is generated when a report template is resolved (combined) with
Xcalibur data source files).
XReport comes with several example Xcalibur report templates for you to use. These
templates, described below, are located in the Xcalibur\Templates folder.
Table 82 describes the standard XReport templates.
Table 82. XReport templates (Sheet 1 of 2)
Report names
Description
Files needed
CalibrationFile
For reporting calibration data.
RAW and XCAL
CompCalReport:
_Avalon,_Genesis,_ICIS
Generates summary reports showing processing method
RST, RAW, PMD,
details, calibration curve, and sample table (Quan results). Use and SLD
this report with results from Xcalibur batch processing and
Quan Browser.
CustLibrSearRept
LibrarySearchReport
Generates Qual sample reports that display Qual peaks and
RST and RAW
library search details. Use this report with results from Xcalibur
batch processing.
IonRatio
Confirmation
Generates a Quan sample report containing non-graphical ion RST, RAW, and
ratio confirmation results.
PMD
IonRatio
Confirmation:
Generates Quan sample reports containing graphical ion ratio RST, RAW, and
confirmation results. These report templates take advantage of PMD
the explicit qualifier ion settings available when using the
Chromatogram Properties Dialog Box and Wizard.
Graphical, GraphicalSimple,
GraphicalStacked
Peak Integration
Generates a Quan sample report containing peak integration
data generated using the Genesis peak integration algorithm.
RST, RAW, and
PMD
ProcessingMethod:
_Avalon, _Genesis, _ICIS
Generates processing method reports. Use this report with
results from Xcalibur Processing Method Setup.
PMD
QualPeak Report
Generates a simple Qual sample report. Use this report with
results from Xcalibur batch processing.
RST and RAW
QuanPeakResults:
_ISTD, _ESTD
QuanResults
QuanSimple
Generates Quan sample reports. Use this report with results
from Xcalibur batch processing or Quan Browser.
RST, RAW, and
PMD
QuantifySampleReport
Generates Quan sample reports containing chromatograms
and a Quan summary table.
SLD, PMD, RST,
and RAW
Thermo Scientific
XReport User Guide
221
D
XReport Application Features
Report Templates
Table 82. XReport templates (Sheet 2 of 2)
Report names
Description
Files needed
Sequence
Report
Generates a summary report showing the sequence being
processed. Use this report with results from Xcalibur batch
processing.
SLD
Spectrum
Candidates
Generates a Quan sample report (only applicable to GC/MS).
Use this report with results from Xcalibur batch processing or
Quan Browser.
RST and RAW
222
XReport User Guide
Thermo Scientific
I
Index
Numerics
C
2-column objects
in a nonrepeating section, with chromatogram and
spectrum 114
in a nonrepeating section, with chromatogram and
spectrum objects 116
in nonrepeating section with chromatogram 114
Left to right option 115
Top to bottom option 115
Calibration File table 194
centering objects 21
choosing
data sources 33
table columns 62
chromatogram object 190
Chromatogram Properties dialog box 76, 141
Chromatogram Properties wizard
Configure How the Peak Window Will Look page 144
Configure Normalization For The Chromatogram
page 145
Select Chromatogram Style page 146
Select Chromatogram Type page 147
Select Component page 147
Select Demo File page 148
Select Enhance Details page 149
Select How The Chromatogram Axes Will Be Configured
page 149
Select Mass Spectrum Quan Traces to Display
page 150
Select Plot Details page 150
Select the Peak Labeling page 153
Select the Qual Peak page 154
Select the Quan Peak page 154
Use Container Supplied Qual Peak page 155
Use Container Supplied Quan Peak page 155
chromatograms, configuring 76
column objects
configuring 113
introduction 191
repeating behavior 114
resizing in a table 20
summary repeating 114–115
using 22
Column Properties dialog box
adding columns 22
parameters 177
combined (resolved) documents 8
A
adding
custom text fields 55
objects 19
sections 17
Advanced Qual Peak Annotation Options page 142
Advanced Qual Peak Annotation Options, setting 100
Advanced Quan Peak Annotation Options page 143
aligning objects 20
annotation object 209
Annotation Properties dialog box
Attributes page 59
configuring an object 59
Font page 60
Avalon Qual Events table, XReport 193
Avalon Quan Events table, XReport 193
Average Scan Filter Selection dialog box 172
B
Batch Reprocess dialog box 39
batch reprocessing, generating reports while 39
bitmap object
configuring 75
details 191
Bitmap Properties dialog box 177
blank line, adding to a report 219
Thermo Scientific
XReport User Guide
223
Index: D
Component Cal Curve object 191
Component Cal Level table 194
Component Calibration Curve Configuration wizard 10
Component Calibration Curve object, configuring 103
Component IdentDetect Cal Settings object 209
Component Ion Ratio table 197
Component ISTD Settings Summary 210
Component QC Level table 195
Component Settings Summary 210
Component Spectrum table 195
Component Sys Suit/Flags Settings 212
Configure Component Calibration Curve dialog box 180
Configure Function dialog box 167
Configure Repeating Section dialog box
Exclude page 139
Ordering page 140
Page Breaks page 141
reference 139
configuring
annotation objects 59
bitmap objects 75
chromatograms 76
Component Calibration Curve objects 103
repeating sections 18
spectra 105
table objects 61
text attributes 56
XReport Objects 9
contacting us ix
creating
reports from Xcalibur 35
reports within XReport 33
templates 13
custom
columns, configuring 63
text fields, adding 55
text fields, editing 56
D
Data Source Browser 10
data source files 8
Data Sources dialog box
reference 180
selecting data sources 34
default data set 30
deleting
objects 19
sections 17
Dilution Factor table 196
224
XReport User Guide
E
Edit Formula For dialog box
configuring custom columns 64
parameters 166
electronic signatures
creating a report with 43
table object 196
exiting XReport 14
F
Font dialog box 58
fonts
settings 27, 57
table 66
footer sections 4
formatting objects 8
G
generating reports
using a report template 14
when running a sample or sequence 37
graphic objects 7
grid, report template 21
H
header sections 4
I
instrument method 212
Ion Ratio Confirmation Result table 197
IRC Settings table 197
IRC Summary table 198
L
Layout menu 136
Library Search Graphic Table Properties dialog box
parameters 172
setting properties 68
Library Search Graphics table 198
Library Search Results table 199
Library Search Results Table Properties dialog box
parameters 172
setting properties 68
locking a template 30
N
nonrepeating sections 5, 186
Thermo Scientific
Index: O
O
objects, XReport
adding 19
aligning 20
annotation 59
centering 21
configuring 23
configuring bitmap 75
deleting 19
formatting 8
graphic 7
repositioning 19
resizing 19
table 6, 61, 121
text 5, 119
text objects 209
types of objects 5
P
page breaks 22, 191
page layout 28
Page Setup dialog box 29, 184
Preferences dialog box, reference 181
Preview window 30, 35, 182
previewing a template 30
processing methods
fields 213
general 213
Programs table 200
Qual 214
Sample Reports Settings table 201
sample table 201
Summary reports settings table 201
summary table 201
Q
Qual Peak table 202
Qual repeating section 188
Qual Summary table 204
Quan Browser window - Reports dialog box 41
Quan Browser, generating a report from 40
Quan ISTD Peak summary 216
Quan Peak summary 216
Quan Peak table 203
Quan repeating section 189
Quan Summary table 205
R
reordering sections 17
repeating objects in columns 114
Thermo Scientific
Repeating Section Configuration dialog box 9
Repeating Section Properties dialog box
configuring a repeating section 18
Exclude page 50
Ordering page 51
Page Breaks page 52
repeating sections 5
report info 217
report template grid 21
Report Template Outline pane 3
report templates 221
reports
adding to processing methods 36
creating from Xcalibur 35
creating within XReport 33
generating 14
generating from Quan Browser 40
generating from Sequence Setup 36
generating when batch reprocessing a sequence 39
generating when running a sample or sequence 37
resolved (combined) documents 8
Reports view 131
repositioning objects 19
resizing objects 19
resizing sections 18
Run Sequence dialog box 38
S
sample header object 217
sample reports 36
sample table
fields 206
properties 69
Sample Table Properties dialog box 69
Save As dialog box 31, 183
saving, XReport templates 30
sections, XReport
adding 17
deleting 17
footer 4
header 4
nonrepeating 5, 186
Qual repeating 188
Quan repeating 189
reordering 17
repeating 5
repeating, configuring 18, 49
reports 186
resizing 18
types of sections 4
Select A Peak dialog box 145
Select Component dialog box 104
XReport User Guide
225
Index: T
Select Report Samples dialog box 42
Sequence Setup, generating a report from 36
sequence table, XReport object 207
setting table fonts 66
setting table properties 67
signatures, electronic 43
sorting tables 65
spectra, configuring report object 105
Spectrum Candidate Results table 208
Spectrum Configuration dialog box
Axis page 156
Enhance page 157
Labels page 158
Normalization page 159
Options page 160
Peak page 160
Plot Type page 161
Style page 162
Spectrum List Table Properties dialog box
Display page 73, 168
Enhancement page 169
Options page 70, 170
Plot Type page 209
reference 168
spectrum object 191
Spectrum Properties dialog box
Enhance page 107
Labels page 109
Normalization page 111
Options page 105
Plot Type page 106
Style page 108
starting XReport 11
summary reports 36
Select Table Fields page
configuring a table object 61
parameter descriptions 163
Set Table Sort Filter page 164
table properties, setting 67
table sort order, specifying 65
Template Configuration dialog box 184
templates
choosing data sources 33
creating 13, 15
definition 4
locking 30
overview 4
prebuilt 10
previewing 30
saving 30
text attributes, configuring 56
text fields
adding custom 55
configuring 54
Text Object Properties dialog box
adding custom text fields 55
Attributes page 175
Attributes page for annotations 176
Data page 53, 174
Font page 58, 177
reference 173
text objects 5
T
XReport Spectrum Control Properties 104
XReport window
features 2
new template 12
Report Template Outline pane 3
table columns, choosing 62
Table Configuration wizard 9
table objects 6
table properties
library search graphics 67
library search results 68
sample table 69
spectrum list 70
Table Properties dialog box
Configure User Column Parameters page 63
Set Table Sort Filter page 65
Table Properties wizard
Configure User Column Parameters page 163
Font page 165
reference 162
226
XReport User Guide
U
user columns, configuring 63
user text, adding 219
X
Thermo Scientific