Microscan EZ Match Imager is a versatile imaging system designed for tube/cap detection and symbol decoding applications. With its high-resolution camera and powerful image processing algorithms, the EZ Match Imager can accurately and reliably identify objects even in challenging conditions. The imager features a user-friendly interface and can be easily configured using the EZ Match Imager Software. The imager is also equipped with a variety of I/O options, making it easy to integrate into existing systems.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

EZ Match B&W or Color Quick Start Guide | Manualzz

Quick Start Guide

EZ Match Imager

Step 1 — Check Hardware

1

Step 2 — Connect the System

Connecting by RS-232 / RS-422 / RS-485

• Connect the imager (1) to the IB-131/IC-332 interface (2) and (3).

• Connect the host cable (6) to the host and to the host port on the IB-131 (3).

• Connect the object detector (5) to the IB-131 (3).

• Connect the power supply (4) to the IB-131 (3).

• Apply power to the imager.

2

Scanner

3

6

1

4 5

Hardware Required

Caution: Be sure that all cables are connected BEFORE applying power to the system. Always power down BEFORE disconnecting any cables.

Item

1

2

3

4

5

6

Description

EZ Match Imager (B&W or Color)

IC-332 Adapter

IB-131 Interface Box

Power Supply

Object Detector (optional)

Host Cable

Part Number

FIS-0004-100XG

FIS-0001-0035G

99-400018-01

97-100004-15 (90-264 VAC, 24 VDC)

99-000017-01

61-300026-01

P/N 83-210051 Rev B

Step 3 — Position the Imager

• Place the imager in an area with as little ambient light as possible.

• Place a black surface behind the area in which test tubes, caps, or symbols will be inspected by the imager.

The black background provides the contrast necessary for the imager to detect the presence or absence of objects in its field of view.

• Position the imager at a focal distance between 2 and 6 inches from a sample tube, cap, or symbol.

• Position the imager relative to the object to avoid the glare of direct

(specular) reflection. Important: Avoid excessive skew or pitch. Maximum skew is ±30°; maximum pitch is ±30°.

6

4

2

10 ml

8

Place a black background behind the area where tubes, caps, or symbols will be inspected by the imager.

This provides the contrast necessary for the imager to detect the presence or absence of objects in the field of view.

EZ Match Imager and Object in Field of View

• Objects can be rotated at any angle; however, for best results, objects should be aligned with the FOV (field of view).

Note: For accuracy of testing and performance, Microscan recommends using a mounting arm adapter kit. Contact your Microscan sales manager for details about mounting arm adapter kits and other accessories.

Copyright ©20

15 Microscan Systems, Inc.

Step 4 — Install ESP

ESP Software can be found on the Microscan Tools CD that is packaged with the EZ Match.

1. Follow the prompts to install ESP from the CD.

2. Click on the ESP icon to run the program.

Note: ESP can also be installed from the Download Center at

www.microscan.com.

ESP System Requirements

• 166 MHz Pentium processor (recommended)

• Windows Vista, XP, or 2000 operating system

• Internet Explorer 5.0 or higher

• 64 MB minimum RAM

• 40 MB minimum disk space

Refer to the EZ Match Imager

User Manual for detailed information about using ESP to configure the EZ Match.

6

2

Scanner

3

4 5

Hardware Configuration

Caution: Be sure that all cables are connected BEFORE applying power to the system. Always power down BEFORE disconnecting any cables.

Step 5 — Select Model

The following menu will appear when you start ESP:

Important: An EZ Match

Imager must be connected to the host computer for its corresponding icon (shown at left) to appear on the model menu.

The icon will always be present in subsequent ESP sessions, even when the EZ

Match unit is not physically connected.

1. Click the button showing the EZ Match Imager.

2. Click OK.

Note: You can also double-click the EZ Match button to make your selection.

3. Click Yes when this dialog appears:

Note: If you need to select another model later, click the Switch

Model button near the top of the screen or use Model > New

Model in the menu toolbar.

Step 6 — Select Protocol and

Connect

• Choose the connection protocol you are using and click Connect.

Step 7 — Locate the Object

Locate the Object by ESP

• In ESP’s EZ Mode, click the Locate button to enable the blue target pattern.

Step 8 — Calibrate

EZ Match settings can be adjusted automatically for optimum tube/cap detection and symbol decoding performance by either the EZ Button or by ESP.

During the calibration routine, the imager will flash its amber Read Rate percent LEDs and red illumination LEDs while searching camera settings and determining the best configuration for detecting tubes or caps, or for decoding symbol data. Upon successful completion of this routine, a green

LED pattern will flash brightly and illuminate the object. If unsuccessful, the imager will emit 5 short beeps and stop searching.

• Once you select your communications mode, follow the simple prompts to establish your connection.

• When you are connected, you will see the green connection indicator in the status bar at the bottom right of your screen.

RS-232:

USB:

The object in the field of view will appear in the video view beneath the

Locate and Calibrate buttons, and you will see the blue target pattern projected from the front of the imager.

• Center the target pattern on the object.

At 2 to 3 inches, the pattern resembles an X . At 3 to 6 inches, the pattern resembles a V .

Important: The entire object of interest (tube, cap, or symbol) should fall within the field of view (FOV) of the imager. The field of view is what appears in ESP’s Locate/Calibrate window in EZ Mode.

• Click the Stop button to end the Locate function.

Locate the Object by EZ Button

If you are not connected to a host computer, the EZ Button allows you to locate an object in the imager’s field of view.

• Hold down the EZ Button for about one second and release when you hear one short beep. The amber 20% LED will illuminate, and you will see the blue target pattern projected from the front of the imager.

• Center the target pattern on the object.

Note: To end all EZ Button functions, press the EZ Button once and quickly release.

Calibrate by EZ Button

1. Hold down the EZ Button for about two seconds and release when you hear two short beeps. The 20% and 40% LEDs will illuminate.

2. The imager will search camera settings to determine the best configuration for detecting tubes or caps, or for decoding symbol data.

Note: To end all EZ Button functions, press the EZ Button once and quickly release.

Calibrate by ESP

1. Click the Calibrate button.

2. The imager will search camera settings to determine the best configuration for detecting tubes or caps, or for decoding symbol data.

A successful calibration will display a green frame around the object, and the following message will appear: “Uploading all reader parameters.”

After a moment the tube/cap data and/or symbol data will be presented in the field below the image display window.

Calibrate by Serial Command

Send <@CAL> from a terminal program to begin calibration.

Step 9 — Test Read Rate

Read Rate indicates the number of tube/cap detections or successful decodes per second achieved by the imager.

Test Read Rate by EZ Button

1. To start the Read Rate test, hold down the EZ Button about three seconds until you hear three short beeps. The

20%, 40%, and 60% LEDs will illuminate.

While the object is being inspected, the

Read Rate LEDs will indicate the read rate percentage on the back of the unit.

2. To end the Read Rate test, press the

EZ Button and quickly release.

Test Read Rate by ESP

1. Click the Test button to start the Read Rate test and Stop to end it.

If a tube or cap has been successfully detected, or a symbol has been successfully decoded, the object’s data and related features will be presented in the field below the image display window. Also, while the object is being inspected, the Read Rate LEDs will indicate the Read

Rate percentage on the back of the unit.

2. To end the test, click the Stop button.

Note: Read Rate can also be tested using the Read Rate interface in

Utilities.

Test Read Rate by Serial Command

You can also start a test with the <C> or <Cp> command and end it with the

<J> command.

Copyright ©20 15 Microscan Systems, Inc.

Step 10 — Configure the Imager

To make setup changes to the EZ Match Imager, click the App Mode button.

The following modes are accessible by clicking the buttons in the first row of

App Mode icons:

• Click the EZ Mode button to return to EZ Mode.

• Click the Autoconnect button to establish communication.

• Click the Send/Recv button to send or receive commands.

• Click the Switch Model button to open the model menu, or to return to a previous model.

• Click the EZ Match button to Test the imager and to configure Multi-Operation settings, tube/cap Library settings, and tube/cap detection Messages.

• Click the Parameters button to show the tabbed tree controls for Communication,

Read Cycle, Symbologies, I/O Parameters, Symbol Quality, Matchcode, and

Diagnostics.

• Click the Setup button to access a Camera Setup tree control and Video view, Evaluate image captures, Calibrate the imager, set the Window of

Interest, fine-tune capture settings and processing settings in the IP Database, set up output filters and parse symbol data in Ordered Output and Output

Format, and control multiple read cycle functions in Dynamic Setup.

• Click the Terminal button to display tube or cap detection data or decoded symbol data, and to send serial commands to the imager using text or macros.

• Click the Utilities button to test Read Rate, request or clear Counters, enable or disable the imager or send output pulses in Device Control, determine the

Differences from Default in the current settings, add or remove master symbol data in Master Database, and verify or update the imager’s Firmware.

For further details, see ESP Help in the dropdown Help menu.

Step 11 — Save Configuration

To make changes to a configuration setting:

1. Left click on the

+ to expand the desired tree.

2. Double click on the desired parameter and click once in the selection box to view options.

3. Place your cursor in the selection box, scroll down to the setting you want to change, and click once on the setting.

4. Left click on the open screen to complete your selection.

5. Right click on the open screen and select Save to

Reader to implement the command in the imager.

Saving Options

Send, No Save. Changes will be lost when power is re-applied to the imager.

Send and Save. This activates all changes in current memory and saves to the imager for power-on.

advertisement

Key Features

  • Connects to host via RS-232, RS-422, RS-485, or USB
  • Detects tubes, caps, or symbols
  • Configurable via ESP software
  • Read Rate indication
  • Object detection
  • Symbol decoding

Frequently Answers and Questions

How do I calibrate the Microscan EZ Match Imager?
The Microscan EZ Match Imager can be calibrated through the EZ Button or through the ESP software.
What are the hardware requirements for the Microscan EZ Match Imager?
The Microscan EZ Match Imager requires an EZ Match Imager (B&W or Color), IC-332 Adapter, IB-131 Interface Box, Power Supply, Object Detector (optional), and Host Cable.
How do I connect the Microscan EZ Match Imager to a host computer?
Connect the imager (1) to the IB-131/IC-332 interface (2) and (3). Connect the host cable (6) to the host and to the host port on the IB-131 (3). Connect the object detector (5) to the IB-131 (3). Connect the power supply (4) to the IB-131 (3).
How do I install the ESP software?
The ESP Software can be found on the Microscan Tools CD that is packaged with the EZ Match Imager or can be downloaded from the Microscan website.
How do I select the Microscan EZ Match Imager model in the ESP software?
In the ESP software, click the button showing the EZ Match Imager and then click OK.
How do I connect to the Microscan EZ Match Imager using ESP?
Once the Microscan EZ Match Imager is selected in the ESP software, choose the connection protocol and click Connect. Follow the prompts to establish the connection.
How do I locate an object using the Microscan EZ Match Imager?
To locate an object using the Microscan EZ Match Imager, click the Locate button in ESP's EZ Mode or hold down the EZ Button for about one second.
How do I calibrate the Microscan EZ Match Imager using the EZ Button?
To calibrate the Microscan EZ Match Imager using the EZ Button, hold down the EZ Button for about two seconds and release when you hear two short beeps.