Breeze | All-in-one Hardware System | Release & Installation Notice

BreezeSuite™ 6.4.1
Service Pack 6
Release & Installation Notice
This software is a Service Pack for BreezeSuite version 6.4.1 and is intended to be installed on any computer
that has BreezeSuite 6.4.1 already installed. This Service Pack replaces and includes all updates in Service
Packs 1-5. BreezeSuite 6.4.1 Service Pack 6 can be installed on workstations, but not on BreezeSuite 6.4.1
Servers.
Because some systems cannot be upgraded to the BreezeSuite 7 Series, Service Pack 6 has the ability to
automatically export patient data to a shared folder so that BreezeSuite 7 can import that data into a central
database. See Page 3 of these instructions for information on running both BreezeSuite 6.4.1 with Service
Pack 6 and BreezeSuite 7 Series.
Minimum System Requirements
 BreezeSuite 6.4.1
Warning:
If installing on versions previous to Service Pack 4, installing Service Pack 6 will over-write the
Microsoft Access reports (mainly used in Physician Review) with the default reports. If you have
made changes to the Access reports these will be lost; however, if your computer has Microsoft
Access installed you can recover your changes by running “Convert MS Access Reports” from
Start >Programs >MedGraphics. Service Pack 6 will not make any changes to your “Quick Print”
reports.
For additional information or help, contact MEDGRAPHICS Technical Support (1-800-333-4137)
or your local distributor before installing Service Pack 6.
INSTALLATION
1.
2.
3.
4.
5.
Close all applications.
Insert CD into CD Drive.
Wait for the installation to start automatically (~15 seconds) or run the BreezeSuite641SP6.exe file on
the CD.
After the software unzips the files, select Next on the Welcome screen.
After the files have been copied select Finish and remove the CD.
LANGUAGE TRANSLATIONS
Service Pack 6 has been translated into five languages: French, Italian, Spanish, Portuguese & Russian
1.
2.
3.
4.
5.
6.
Follow the Installation Instructions above before proceeding
With the SP6 CD in the CD Drive, browse to the Translations folder on the CD.
Double click on the appropriate language executable file.
At the “Location To Save Files” screen select Next.
At the “Overwrite Protection” screen select Yes To All.
After the files have been extracted and copied select Finish.
MGC P/N: 147781-003, Rev. A
MedGraphics® is a registered trademark of Medical Graphics Corporation.
BreezeSuite
is a trademark of Medical Graphics Corporation.
Page 1 of 6
Form 441 Rev C
Highlighted changes from BreezeSuite 6.4.1 Service Packs 1-4
























Added enhancements for WFA (Waveform) communications.
Corrected issue of needing Pulmonary Consult installed for Electronic Signature.
Corrected issue of negative A-a gradient.
Corrected issue of manually entering blood gas values without hemoglobin.
Enhanced the PDF report from Physician Review so that a multiple page report will be exported as a
single PDF file.
Corrected shortcut for USB 1.0 & 2.0 driver
Enhanced the Shapetable flow program for the Platinum Elite™ Series.
Added the ability to change the Y-Axis variable in the O2 Kinetics Graph
Added a 100 ms Interpolation Averaging Method to the Gas Exchange Test.
Enhanced the ability of the Gas Exchange software to continue updating if the GX data exceeds the
threshold limits.
Added entry to specify the number of decimal places to display the Percent Predicted/Percent Change
values in the Test Tab.
Corrected issue so that a value will be displayed in red with an asterisk when Spanish language is
selected in regional settings and if the data value displayed is less than 1.0, such as 0.5, the value will
be displayed as 0,5
Added ability to convert data from a NDD database or NDD export files via the "Convert Visit" menu in
DBTools
Added ability to enter a patients’ DOB which would make them less than 1 year old.
Corrected issue so that the treadmill stops when a protocol calls for two minutes of recovery and then
the script calls for the treadmill to stop
Corrected issue of the Disability Graph not printing the second page if trace goes beyond 10 seconds
Corrected issue for Wasserman treadmill obese male predicted values
Added a prompt “confirm: Go to Exercise?” when selecting the Exercise phase
Corrected issue so that the VO2000™ continues to collect appropriate data when the respiratory rate
exceeds 44 breaths per minute
Added support for the Ultima™ Tab 006 system
Added support for the CPFS/D™ USB Tab 003 system
Added “Gas Autocal” in the FRC shapetable program
Corrected shutter issue for standard Elite Plethysmograph in TGV and DLco testing
Corrected issue for standard Elite Plethysmograph where exhaled volume was not collected
Highlighted changes for BreezeSuite 6.4.1 Service Pack 5








Added support for BreezeSuite 6.4.1 SP5 and BreezeSuite 7.0 & above to share data
Corrected issue of shutter not closing in Static Compliance
Corrected issue of predicteds not calculating when using author Wang 1993
Corrected issue of fat calories or carbohydrates reported as negative numbers
Added the ability to export patient files directly from BreezeSuite
Corrected communication error with Trackmaster 225, 300 & 425 treadmills
Resized the Physician Review Station spreadsheet to see all columns without scrolling
Added information pertaining to NHANES III into the HELP file
Highlighted change for BreezeSuite 6.4.1 Service Pack 6
 Corrected the possibility of negative or non-physiologic estimated and measured Vd/Vt by modifying
equation for PECO2 using “True CO2” (VCO2 STPD/VD STPD)
MGC P/N: 147781-003, Rev. A
MedGraphics® is a registered trademark of Medical Graphics Corporation.
BreezeSuite
is a trademark of Medical Graphics Corporation.
Page 2 of 6
Form 441 Rev C
BreezeSuite 6.4.1 with Service Pack 6 and BreezeSuite 7 Series
Service Pack 6 for BreezeSuite 6.4.1 allows you to share patient data with BreezeSuite 7 Series. This may be
necessary because some older equipment is no longer supported by BreezeSuite 7.0 or greater. In these cases
the older equipment must stay on BreezeSuite 6.4.1 while the newer systems must be on BreezeSuite 7.0 or
greater.
Warning:
If you are using either BreezeConnect™ or BreezeLink™ to share data with you facilities
EMR (Electronic Medical Record), please contact the MEDGRAPHICS Network Specialist
before proceeding.
Service Pack 6 allows you to export patient data to a shared folder. BreezeSuite 7 Series can then import that
data into a central database. All data, regardless of testing system, can reside in the BreezeSuite 7 Series
database.
NOTE: When sharing a common folder, BreezeSuite 6.4.1 cannot automatically import data from a BreezeSuite
7 Series database. The data flows only in one direction: from BreezeSuite 6.4.1 to BreezeSuite 7. (see
diagram)
Auto export of patient
files whenever a visit
is saved with changes
Shared Folder on Server
for BreezeSuite 6.4.1
patient files
BreezeSuite 6.4.1 with
Service Pack 6 System
Auto import of 6.4.1
patient files. Auto
delete from shared
folder after import
BreezeSuite 7.0 or
greater System
NOTE: Testing and work stations that have BreezeSuite 7.0 or greater will be able to see all patients,
regardless of which system they were tested on. Work stations with BreezeSuite 6.4.1 will be
able to see only those patients tested on a system with BreezeSuite 6.4.1.
The following procedure presumes that there is a BreezeSuite 6.4.1 multi-user network already installed
and that some of the existing equipment cannot be upgraded to the BreezeSuite 7 Series software.
Assistance from your SQL Database Administrator is strongly recommended.
MGC P/N: 147781-003, Rev. A
MedGraphics® is a registered trademark of Medical Graphics Corporation.
BreezeSuite
is a trademark of Medical Graphics Corporation.
Page 3 of 6
Form 441 Rev C
Procedure
1. Backup all existing databases before proceeding.
2. Install the new BreezeSuite 7 Series multi-user software on the SQL Server and then on all systems that will
be running BreezeSuite 7.0 or greater. For installation instructions, refer to the instructions on the installation
CD or contact Technical Support at Medical Graphics.
NOTE: For SQL Server installation, it is important to not upgrade the BreezeSuite 6.4.1 central server database
if some systems will remain on BreezeSuite 6.4.1. When upgrading a workstation from 6.4.1 to 7.0 or greater,
ensure that you point to the BreezeSuite 7 Series Server Database.
3. Upgrade all a appropriate BreezeSuite 6.4.1 workstations to the BreezeSuite 7 Series software.
4. Upgrade all Physician Review Stations to the BreezeSuite 7 Series and point to the BreezeSuite 7 database.
If you do not upgrade a Physician Review Station and the review station remains on BreezeSuite 6.4.1, they will
not be able to review all patients.
5. Create a common folder on a central server to which both software versions have access. (remember the
name and location of the common folder)
6. On each BreezeSuite 6.4.1 workstation go to Tools>Options>Export/Import Tab. In the Auto Export section
place a checkmark (click on the box) for COW. In the Export Directory browse to the common folder where the
patient files will be exported. In the example below, the patient files will be exported to the X:\BreezeSuite
Patient Files folder.
MGC P/N: 147781-003, Rev. A
MedGraphics® is a registered trademark of Medical Graphics Corporation.
BreezeSuite
is a trademark of Medical Graphics Corporation.
Page 4 of 6
Form 441 Rev C
7. On each BreezeSuite 7 Series workstation go to Tools>Options>Export/Import Tab. In the Import section
place a checkmark (click on the box) for Auto Import and Refresh Import. In the Import Directory browse to the
common folder where the patient files will be imported. In the example below, the patient files will be imported
from the X:\BreezeSuite Patient Files folder.
8. Once this is set up, every time you close a patient in BreezeSuite 6.4.1 that has changes, the patient file will
be automatically exported to the common folder. Every time you open BreezeSuite 7.0 or greater, or refresh the
patient list, the software will automatically import any patient files in the common folder. Once they are
imported, they will automatically be deleted from the common folder.
NOTE: This system will be a dual multi-user setup. One multi-user setup will be BreezeSuite 6.4.1, while the
other will be BreezeSuite 7.0 or greater, connect through a common folder. See example below.
MGC P/N: 147781-003, Rev. A
MedGraphics® is a registered trademark of Medical Graphics Corporation.
BreezeSuite
is a trademark of Medical Graphics Corporation.
Page 5 of 6
Form 441 Rev C
BreezeSuite 6.4.1
Central Database
BreezeSuite 6.4.1
Workstation
BreezeSuite 6.4.1
Workstation
BreezeSuite 6.4.1
Workstation
Auto export of patient
files whenever a visit
is saved with changes
Shared Folder on Server
for BreezeSuite 6.4.1
patient files
BreezeSuite 7 Series
Workstation
BreezeSuite 7 Series
Workstation
Auto import of 6.4.1
patient files. Auto
delete from shared
folder after import
BreezeSuite 7 Series
Workstation
BreezeSuite 7
Series Central
Database
MGC P/N: 147781-003, Rev. A
MedGraphics® is a registered trademark of Medical Graphics Corporation.
BreezeSuite
is a trademark of Medical Graphics Corporation.
Page 6 of 6
Form 441 Rev C