- Industrial & lab equipment
- Measuring, testing & control
- Radiation detectors
- Golden Engineering
- XRS-3
- User manual
- 16 Pages
Golden engineering XRS-3 Operator's Manual
The Golden Engineering XRS-3 is a compact, portable X-ray generator designed for industrial applications. It operates on a removable battery pack and features a pulsed X-ray system capable of penetrating up to 1 inch of steel. With a maximum output of 270kVp, the XRS-3 provides clear images for various inspection needs.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
▼
Scroll to page 2
of
16
OPERATOR'S MANUAL MARCH 2018 - Golden Engineering �-Portable X-ray Technology TABLE OF CONTENTS PAGE 1.0 INTRODUCTION 2 2.0 WARNINGS 2 DUTY CYCLE 2 STORAGE 2 3.0 PHYSICAL DESCRIPTION 3 HIGH VOLTAGE PULSER/TUBEHEAD 3 BASE 3 BATTERY PACK 3 BATTERY CHARGER 3 3.5 CONTROL MODULE 4 CABLE CONNECTOR DIAGRAM 5 4.0 DESCRIPTION OF OPERATION 6 BLOCK DIAGRAM 6 5.0 OPERATING INSTRUCTIONS 7 OPERATING PRECAUTIONS 8 EXCLUSION ZONE 8 PULSE SETTINGS 8 6.0 NAVIGATING THE MENU 9 7.0 MAINTENANCE 11 DOSE MEASUREMENT 11 TUBE REPLACEMENT 11 8.0 TROUBLE-SHOOTING 11 9.0 INSTRUCTIONS FOR SERVICE 12 REMOVING BOARDS 12 BOARD INSTALLATION 12 HEAD REPLACEMENT 12 BATTERY DISPOSAL 13 10.0 WARRANTY INSTRUCTIONS TO RETURN FOR SERVICE AND MAINTENANCE 11.0 SPECIFICATIONS 13 14 PHYSICAL DIMENSIONS 14 X-RAY OUTPUT 14 ELECTRICAL & THERMAL CHARACTERISTICS 14 12.0 SPARE PARTS 1 13 14 1.0 INTRODUCTION The XRS3 produces high levels of radiation and must be operated by quali ed personnel who have read the WARNINGS and OPERATING INSTRUCTIONS sections of the manual before operating the device. The XRS3 is a small, lightweight X-ray generator that operates on its own removable battery pack. The XRS3 is a pulsed X-ray device that produces X-ray pulses of very short duration (25 nanoseconds). It produces a relatively low dose rate comparable to a 0.25 ma constant potential machine. The energy produced by the XRS3 is up to 270kVp, which makes it possible to radiograph up to one (1) inch (2.54 cm) of steel. XRS3 standard accessories are two keys, two battery packs, and one battery charger. Remote cable and carrying case are also common accessories. 2.0 WARNINGS The XRS3 is a pulsed X-ray generator that emits hazardous ionizing radiation when pulsing. The XRS3 should only be operated by authorized personnel who are properly trained to safely operate the generator. The XRS3 must be registered with proper authorities prior to use and should not be used to intentionally expose humans. Develop and closely follow a safe operating system for using the XRS3. The safe operating system must ensure that no one is exposed to radiation above the permissible limits which are 2 mR (0.02 mSv) per hour for a member of the public. The safe operating system must ensure the XRS3 is used within federal and state guidelines. All operators and users of the XRS3 X-ray machine must wear a personal radiation monitoring device, such as a TLD (thermoluminescent dosimeter), lm badge, and/or a pocket dosimeter consistent with the appropriate federal, territorial or provincial standards (note: an electronic dosimeter will not detect the XRS3 radiation pulses). Due to the short pulse width of the XRS3, survey meters of the Geiger-Mueller and scintillator type do not accurately detect the radiation emitted from the x-ray source. Survey meters should be of the ionization type and should be used in the integration mode. Survey meters must not be used in the rate mode because the XRS3 does not produce constant radiation. The XRS3 produces very high rates of radiation for very short periods of time resulting in either unrealistically high readings or no readings for a survey meter in rate mode. The XRS3 has no explosion proof rating and should not be used in an explosive atmosphere. The Spark Gap is vented to the air and could be a source of ignition. DUTY CYCLE WARNING. The XRS3 is a light duty machine that is not made to pulse continuously. The maximum duty cycle for the XRS3 is 200 pulses every four minutes (3000 pulses per hour). Exceeding the duty cycle will shorten the life of the tube and head. In temperatures above 90 °F (32.22 °C) or continual use, rest 30 seconds after every 50 pulses and 4 minutes after every 200 pulses. STORAGE & USE: Store and use the XRS3 in upright position. Upside down storage or use may cause premature failure including no output dose. 2 3.0 PHYSICAL DESCRIPTION HANDLE CONTROL MODULE KEY BATTERY BEAM ANGLE LABEL BASE COLLIMATOR HEAD RADIATION WARNING LABEL PICATINNY RAIL Figure 1: XRS3 X-ray Unit HIGH VOLTAGE PULSER/TUBEHEAD. The main body of the XRS3 is the tube head which contains the tube cavity, cold cathode type X-ray tube, spark gap, high voltage capacitor, and transformer. The standard collimator located on the front of the head limits the X-ray beam to 40 degrees. Special order collimators up to 85 degrees are available. TRIPOD MOUNT RELEASE TAB BASE. The base of the XRS3 contains an identi cation label and a ¼-20 brass insert compatible with standard camera tripods. The base also accommodates a quick release external tripod mount. The identi cation label located on the bottom of the generator lists the manufacturer’s name and address, model number, serial number, weight, volt, amp, and production date. BATTERY PACK. The standard battery pack is a DeWalt® 20V 2 amp hour Li Ion battery (DCB203). BATTERY CHARGER: The standard battery charger is the DeWalt® DCB107 110V charger or DCB105 220V charger. (Note DeWalt part numbers may change). Battery charge time is less than one hour. See battery charger manual for additional instructions and warnings. PICATINNY RAIL: The XRS3 has a 21 mm picatinny rail located on each side of the housing. 3 TRIPOD MOUNT Figure 2: Base 3.5 CONTROL MODULE MODEL IDENTIFICATION POWER ON LIGHT LCD X-RAY PULSING LIGHTS ENTER / EMERGENCY STOP DELAY SWITCH KEY DIRECTIONAL BUTTONS Figure 3: Control Module POWER ON LIGHT: Illuminates when battery voltage is applied to control module. RED X-RAY PULSING LIGHT: Blinks after time delay button or remote cable button is pressed to warn that the XRS3 is going to pulse. The light stays on continuously while the XRS3 is pulsing. This is a failsafe warning light. If the light does not work the X-ray unit will not pulse. See settings menu for fail safe override in emergency situations. LIQUID CRYSTAL DISPLAY (LCD): 80 digit LCD. Home displays number of pulses selected and delay time before the rst pulse. Home will also indicate if multiple pulse trains have been entered. Arrows allow operator to scroll through various menu options. DIRECTIONAL BUTTONS: Used to maneuver through the menu. DELAY SWITCH: Left and Right arrow buttons pressed simultaneously Initiates the delay mode. EMERGENCY STOP/ENTER SWITCH: Stops the unit before it begins pulsing or stops the unit in the middle of a pulse train. Also used as the enter button to select desired option. 4 XRS3 REAR VIEW/CABLE CONNECTOR CABLE CONNECTOR: Lemo “K” series ve pin connector located on the back of the control module beneath the battery receives the remote cable or imaging system cable. 1 PIN # 1 2 3 4 5 DESCRIPTION +5 VOLTS 100 ma MAXIMUM REMOTE SWITCH REMOTE SWITCH - NO DELAY X-RAY ON SIGNAL COMMON 0 VOLTS 2 5 3 4 REMOTE CABLE CONNECTOR TEMPERATURE SENSOR NEGATIVE - POSITIVE + REMOTE CABLE CONNECTOR BATTERY TERMINAL BLOCK REMOTE CONNECTOR: LEMO EEG.0K.305.CLN MATING CABLE PLUG: LEMO FGG.305.CYCC50Z PIN 2 PIN 5 Remote switch inputs are activated when grounded. 5 4.0 DESCRIPTION OF OPERATION The block diagram below illustrates how the XRS3 functions. The following sequence of events takes place each time the XRS3 is red: 1. 2. 3. 4. 5. 6. 7. User initiates operation of the machine. The control section sends a signal to the converter section to begin oscillating. Once oscillating, the converter section changes the 20 volts DC to 22Khz AC. The transformer charges the High Voltage Capacitor to about 9000 volts. The spark gap arcs after the High Voltage Capacitor reaches proper voltage. The pulse detector signals the control block that the unit has pulsed. As the High Voltage Switch is closed, a high voltage transient of approximately 270,000 volts and 25 nanoseconds in duration is applied across the x-ray tube generating x-rays. The closing of the High Voltage Switch produces an audible pulsing sound. The XRS3 cannot produce x-rays without the pulsing sound so it serves as an additional warning the XRS3 is functioning. This unit generates X-rays through high voltage bombardment of a tungsten target. The XRS3 does not contain radioactive materials. All the high voltage is contained within the aluminum canister and as long as the canister is not punctured the operator is not exposed to dangerous voltages. BLOCK DIAGRAM X-RAY TUBE SPARK GAP HIGH VOLTAGE CAPACITOR PULSE DETECTOR CONVERTER USER INTERFACE CONTROLLER BATTERY 6 5.0 OPERATING INSTRUCTIONS The following are basic operation instructions to take an X-ray image using the XRS3. Certain applications may require modi cations to these basic procedures. 7 *XRS3 should be positioned directly in front of the object to be X-rayed and the imager placed directly behind the object to be X-rayed. Imager should be placed as close to the object as possible. Distance between XRS3 and imager is usually 24 to 72 inches (30 to 180 cm). During operation XR200 should be stabilized on a at surface, a tripod, or a custom xture suitable for holding the 11 pound (5 Kg) XRS3. XRS3 OBJECT IMAGER 24 to 72 inches (30 to 180 cm) OPERATING PRECAUTIONS: The operator should always stand at least 10 feet (3m) behind the X-ray unit and clear all personnel at least 10 feet (3m) behind the unit or at least 100 feet (30m) from the front of the unit before pulsing. The exclusion zone (below) should be a controlled area free of all personnel while X-ray pulses. 25’ (7.6 m) 10’ (3m) 100’ (30m) X-ray Beam 10’ (3m) 10’ (3m) 25’ (7.6 m) Figure 4: Exclusion Zone SUGGESTED PULSE SETTINGS The chart below lists approximate pulses necessary to penetrate various materials. Settings vary greatly depending on imaging system used. Refer to imaging system instructions for more information. MATERIAL CARDBOARD / LIGHT WOOD / PLASTIC LIGHT METAL STEEL 1/4” STEEL ½” STEEL 1” PULSE SETTING (24” BETWEEN X-RAY & IMAGER) 2-5 10 25 50 99 8 The following is true when using lm or digital systems that generate a positive image. If the radiograph is too dark, the lm is underexposed. If the radiograph is too light the lm is overexposed. Underexposure can be corrected by increasing the number of pulses and/or decreasing the distance between the imaging medium and the XRS3. Overexposure can be corrected by reducing the number of pulses and/or increasing the distance between the imaging medium and XRS3. Underexposed Correct exposures (pulse setting) Overexposed Correct exposures (pulse setting) 6.0 NAVIGATING THE MENU LCD ERROR MESSAGE Low Battery 9 Battery voltage is too low to pulse the XRS3. Charge the battery. PWM not running Could occur at the start of a pulse train or during a pulse train. Power unit off and back on. If error message continues there is a problem with electronics. Slow Pulsing More than one second between pulses in a pulse train. Either low battery, problem with transformer, or electronics. Charge battery or replace oscillator board. No Feedback Current not detected. No current ow going to the head. Occurs at the beginning of pulse train. Check 20 amp fuse on the oscillator board. If fuse is ok problem could be in head or electronics. Replace board or send back for repair. Cycle Reached Displayed if duty cycle is reached (200 pulses in 4 minutes). >HOME PULSES: _ _ _ _ DELAY: _ _ _ _ MENU PRESETS HOME SET PULSE SET DELAY TRAINS >MENU PRESETS LIFE PC RESET PC SETTINGS HOME >SET PULSE SET DELAY TRAINS - MENU PRESETS LIFE PC RESET PC SETTINGS HOME SET PULSE >SET DELAY TRAINS - MENU HOME SET PULSE SET DELAY >TRAINS - MENU HOME SET PULSE SET DELAY TRAINS - MENU HOME SET PULSE SET DELAY TRAINS - MENU HOME SET PULSE SET DELAY TRAINS - MENU PRESETS LIFE PC RESET PC SETTINGS PRESETS LIFE PC RESET PC SETTINGS SET PULSE COUNT _ _ _ PULSES BACK SAVE SET DELAY TIME _ _ _ SECONDS BACK SAVE SET PULSE TRAINS _ _ _ PULSE TRAINS: _ _ _ SECONDS BETWEEN: BACK PRESETS >LIFE PC RESET PC SETTINGS SAVE LIFETIME PULSE COUNT _ _ __ _ _ PULSES PRESETS LIFE PC >RESET PC SETTINGS #1: _ _ _ _ #2: _ _ _ _ BACK RESETTABLE COUNTERS PRESETS LIFE PC RESET PC >SETTINGS >BACKLIGHT X-RAY INFO BACK >RESET RESET SETTINGS BACKLIGHT FEEDBACK ON SETTINGS BACKLIGHT FEEDBACK INDIVIDUAL PULSE X-RAY ON / X-RAY OFF >FEEDBACK BACK OFF BACK SETTINGS BACKLIGHT >E-RAY INFO BACK X-RAY INFO FEEDBACK SOFTWARE _ _ _ SERIAL # _ _ _ BACK HEAD # _ _ _ To Enter/Save Presets 1. Enter desired settings (Pulse, Delay Time, Pulse Trains) from main menu as shown above 2. Go to Preset Menu 3. Select Preset # 4. Select SAVE HOME > PRESET #1 PRESET #2 PRESET #3 MENU - PRESETS PRESET #1 PULSES: _ _ _ _ TRAINS: _ _ _ _ BACK RECALL PRESET #1 DELAY: _ _ _ _ BTWN: _ _ _ _ >SAVE PULSES: _ 50 TRAINS: 05 BACK RECALL DELAY: 15 BTWN: 45 >SAVE To Recall Presets 1. Go to Preset Menu 2. Select Preset # 3. Select Recall HOME > PRESET #1 PRESET #2 PRESET #3 MENU - PRESETS PRESET #1 PULSES: _ 50 TRAINS: 05 BACK >RECALL DELAY: 15 BTWN: 45 SAVE 10 7.0 MAINTENANCE X-RAY DOSE MEASUREMENT Using a dosimeter, the average X-ray dose for new tube can be established. With the dosimeter located 1 foot from the front of the case and in line with the center of the beam angle label, the reading for 10 pulses should be at least 20 mR to 36 mR. The leakage sheet illustrates the X-ray dose and maximum allowable radiation leakage levels for each X-ray unit. A completed copy of this form accompanies each X-ray. TUBE REPLACEMENT: If you have a tube replacement kit refer to instructional disk included with the kit. If you do not have a kit the unit must be sent back to Golden Engineering or an Authorized Distributor for tube replacement. Tube life is approximately 100,000 pulses. Under normal conditions the tube’s output will decrease slowly with use. If the tube is broken or the glass cracks the tube output will cease immediately. 8.0 TROUBLESHOOTING SYMPTOM No “power on” light TEST -Check battery voltage -Check battery connection Power on lights, X-ray pulsing light does not illuminate, X-ray does not pulse ACTION - Replace or charge battery - Make sure battery is securely attached and battery clips are not bent or broken. - Go to settings menu failsafe disable - To x light replace processor board X-ray pulses, but no image or black image. -Test X-ray output. -Replace the tube. Unit stops pulsing in the middle of a pulse train and LCD displays 00. -Check the battery voltage. -Check 15 amp fuse. -Check feedback line connection. - Charge battery if necessary. - Replace the fuse if blown. - Make sure the screw holding wire to the oscillator board. Unit makes loud popping noise while pulsing. - Stop immediately and return for repair. Oil leaking from unit. Return for repair. 11 9.0 INSTRUCTIONS FOR SERVICE INSTRUCTIONS TO REMOVE THE BOARDS AND THE HEAD. 1. Use T10 Torx driver to remove 6 screws holing control module in place. 2. Remove the control module. 3. Remove cable connecting the boards. 4. Control Module Removed 1 2 3 4 Remove keyswitch connector Remove 3 screws holding processor board to top Remove processor board. Disconnect ribbon cable. Remove the 8 screws that hold the housing together Remove half the housing. Remove the head. Loosen the four screws that hold the feedback wires in place. 12 Remove feedback wires. Remove the screws that connect the oscillator board to the head. Remove the oscillator board. INSTRUCTIONS FOR BATTERY DISPOSAL Follow all federal, state, and local laws for disposal of lithium-ion batteries. Batteries may be returned to Golden Engineering. RETURNING UNIT FOR SERVICE AND MAINTENANCE Complete the support form at http://www.goldenengineering.com/home/support and include a copy of the printed form with the repair. If you do not have internet access prior to sending generator, include a letter containing a brief description of the problem, contact name, phone number, and return address. Remove battery before shipping the unit. Accessories are not required. Be sure the unit is securely packaged for shipment and wrapped in plastic bag if there is an oil leak. Ship to address: Golden Engineering, Inc., 6364 Means Road, Centerville, In 47330 USA 10.0 WARRANTY Golden Engineering, Inc. warrants XRS3 X-ray unit made and sold by it or its authorized representatives to be free of defects in materials and workmanship for a period of twelve (12) months from the date of shipment to the end user. Warranty does not cover maintenance required due to life. To make a claim under this limited warranty, customer must ship the entire unit (or the component believed to be defective) to Golden Engineering, post-paid. Golden Engineering assumes no liability for units or components shipped until they are actually in the custody of Golden Engineering, Inc. Provided Golden Engineering, in its sole discretion, is satis ed that the failure is not the result of excessive use, abuse, misuse, accident, modi cation or improper disassembly or repair, Golden Engineering will provide parts and labor required for the repair. Golden Engineering reserves the right to use reconditioned and remanufactured components that meet original speci cations. The unit or component will be shipped to customer at customer’s expense. THIS EXPRESS LIMITED WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES AND GUARANTEES, EITHER EXPRESS OR IMPLIED OR CREATED BY OPERATION OF LAW. XRS3 Manufacturer European Representative Golden Engineering, Inc. Certi cation Experts Europe 6364 Means Road, Box 185 Nieuwstad 100 Centerville, IN 47330 USA 1381 CE Weesp Phone: 1-765-855-3493 The Netherlands Fax: Web: 1-765-855-3492 Serial Number:___________________ 13 www.goldenengineering.com Delivery Date:____________________ 11.0 SPECIFICATIONS PHYSICAL DIMENSIONS INCLUDING BATTERY PACK LENGTH 14.20 inches (36.07 cm) WIDTH 4.26 inches (10.82 cm) HEIGHT 7.44 inches (18.90 cm) WEIGHT 11.8 pounds (5.4 Kg) with battery X-RAY OUTPUT X-ray dose per pulse 2 mR/pulse min,(12 inches in front of unit) Pulses per battery charge 4000 Pulses per second 15 (Nominal) Expected tube life (glass tube) 100,000 pulses X-ray source size 1/8 in. (3mm) Maximum Photon Energy 270 kVp X-ray pulse width 25 nanoseconds ELECTRICAL AND THERMAL CHARACTERISTICS Battery voltage 18 volts - 20 volts Battery type Li Ion Battery recharge time One Hour Current draw 20 amps @ 18/20 volt battery Current ow 0.25 mA Temperature range -10 to 120 degrees F (-23 to 50 degrees C) Airborne Noise Emissions 80 dB at 10 cm Maximum duty cycle 200 pulses every 4 minutes (3000 pulses per hour) Warm-up None required 12.0 SPARE PARTS AND ACCESSORIES FOR THE XRS3 IITEM PART NUMBER Thumbwheel Key 2002000 Flat key 5951020 DeWalt® Battery 20V (2 Amp Hour / 40 Watt Hours) DCB105 1800106 DeWalt® Battery Charger (110V) DCB107 1800152 DeWalt® Battery Charger (220V) DCB105 or DCB115 1800164 Remote Cable 1809022 Tripod Mount 4000352 Handle 4000153 Carrying case (holds X-ray, 2 batteries, charger, cable) 1701656 14
advertisement
Key Features
- Portable
- Pulsed X-ray
- Up to 270kVp Output
- Penetrates Steel
- Removable Battery
- Adjustable Pulse Settings
- Safety Features
- Menu-Driven Control
Frequently Answers and Questions
What is the maximum duty cycle of the XRS-3?
The maximum duty cycle is 200 pulses every four minutes, or 3000 pulses per hour.
How do I adjust the pulse settings on the XRS-3?
Use the directional buttons and the Enter/Emergency Stop switch to navigate the menu and select your desired pulse count, delay time, and pulse trains.
What types of materials can the XRS-3 penetrate?
The XRS-3 can penetrate up to 1 inch of steel. It can also be used to inspect cardboard, wood, plastic, and lighter metals.