GMI PS200 series User Manual

GMI PS200 series User Manual

The GMI PS200 series is a powerful and versatile gas detection instrument designed for a range of applications, including marine environments. Its intuitive interface allows for easy operation, and it features advanced features such as automatic calibration and intelligent alarms. This manual provides comprehensive instructions for using and maintaining the PS200 series device.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

GMI PS200 Series User Manual - Online & Download | Manualzz
!
Quick Operation Guide /
Prise en main rapide
Switch On / Mise en marche
2s
Backlight On / Rétro-éclairage de l’écran
Min
Max
1s
20s
Manual Datalog / Sauvegarde manuelle
Reset / Remise à zéro
Pump / M/A pompe
Self Test / Auto Test
1s
1s
Alarms / Alarmes gaz
Alarm Acknowledge / Acquittement
1s
Calibration Overdue / Date de calibrage dépassée
Switch Off / Mise à l’arrêt
&
3s


Flow Fault / Défaut débit
Charging / Mise en charge
4h
1s
Reset / Acquittement
Zero Fault / Défaut de la mise à zéro
Gas Measurement Instruments Ltd
www.gmiuk.com
Part Number: 64191 Issue 4
MARINE 4 QUICK START, CALIBRATION AND TROUBLESHOOTING GUIDE
IMPORTANT
MARINE 4 QUICK START
CALIBRATION AND
TROUBLESHOOTING GUIDE
READ BEFORE USE
MARINE 4 QUICK START, CALIBRATION AND TROUBLESHOOTING GUIDE
MARINE 4 QUICK START, CALIBRATION AND TROUBLESHOOTING GUIDE
Calibration using ABC station
Equipment required:
1. ABC Station and power adapter
2. USB stick supplied with ABC station – complete with software as
supplied
3. Test gas and tubing with connectors
4. Demand flow regulator
5. Marine 4
1. Connect demand flow regulator and tubing to gas bottle
3. Connect Air / Exhaust Tubing to ABC Station (if required)
If the ABC Station is located in a poorly ventilated area, it is
recommended that tubing is connected and routed to the nearest
window for the supply of fresh air/exhaust ventilation.
4. Connect power to ABC station
5. Allow 30 seconds for the unit to warm up, ensure that the light
sequence Orange‐Orange‐Red is seen during power‐up after which all
LEDs will light up green and when completed only the power light will
be lit
5 seconds
25 seconds
2. Connect tubing into the back of the ABC station
6. Put memory stick supplied into
PC/laptop and select “Open
Folder to View Files” (note actual
display may vary with different
operating systems)
MARINE 4 QUICK START, CALIBRATION AND TROUBLESHOOTING GUIDE
7. Select below from the folder
11. Put USB into the ABC station
The image shown will load.
8. Choose unit calibration
12. The LED’s will flash while the settings are uploaded. Wait for light
sequence to complete
9. Compare gas types and
ranges with gas bottle
and ensure they are
the same, make
changes if needed.
10. Ensure the USB stick is
still connected to
laptop/PC and press generate settings file button.
This will save them to the USB stick
13. Wait for the lights to STOP flashing and then REMOVE THE USB STICK
MARINE 4 QUICK START, CALIBRATION AND TROUBLESHOOTING GUIDE
14. Open the front cover of the ABC unit by pressing the latch
16. Close the front door until it ‘clicks’ in to place
15. Insert the unit to be tested into the ABC station connecting the gas
supply nozzle at the base of the unit. The marine 4 can be switched
on or off.
17. The process will start automatically when the door is closed.
While testing is in progress the lights will cycle through a sequence of
orange flashing lights – first will be the test in progress light as shown
below.
18. The marine 4 will also start up and go through a sequence of tests with
gas being delivered for 60 seconds
19. Audible alarms will also be present as the unit is tested and alarm set
points are reached.
MARINE 4 QUICK START, CALIBRATION AND TROUBLESHOOTING GUIDE
20. The unit will then indicate pass or fail by either a green or red led
23. Wait for light sequence to complete
PASS
FAIL
21. Remove the unit from the ABC station. If it has passed it can be put
into operation. If failed please refer to the troubleshooting guide.
22. When units have been calibrated successfully and removed from the
ABC station, Insert the USB stick to download the results from the ABC
internal memory.
24. Remove USB from the ABC
25. Put memory stick supplied into PC/laptop and select “Open Folder to
View Files” (note actual display may vary with different operating
systems)
MARINE 4 QUICK START, CALIBRATION AND TROUBLESHOOTING GUIDE
26. Select below from the folder
30. Select the test result that you wish to produce the certificate for from
the list of test results that are displayed in the new window
27. Click the button to generate the
calibration certificate
28. The below box will appear to
input specific information if
required
31. Click OK to open produced certificate
29. Click create certificate
MARINE 4 QUICK START, CALIBRATION AND TROUBLESHOOTING GUIDE
32. Example calibration certificate
33. Certificates that have been created can then be viewed in the
Generated certificates folder
MARINE 4 QUICK START, CALIBRATION AND TROUBLESHOOTING GUIDE
Screen Display
showing…
Audible
alarm status
Sounds once
every 2
seconds
LED status
Probable Cause
Solution
If not resolved
Red LED’s
flashing
Battery requires charging. There is less
than 30 minutes normal operation left
before the unit turns off
Place the unit on charge, it will take a
maximum of 4 hours to recharge fully
If battery will not charge
up please contact
service@martek‐
marine.com
Note: Gas alarms continue to operate
after the ‘LOW
BATTERY’ warning appears.
LOW BATTERY warning
Sounds
Red LED’s
continuously stay on
Battery requires charging. There is less
than 3 minutes normal operation left
before the unit turns off.
Place the unit on charge, it will take a
maximum of 4 hours to recharge fully
If battery will not charge
up please contact
service@martek‐
marine.com
NO ALARMS
NO LIGHTS
Battery has been allowed to become fully
depleted and requires charging
Place the unit on charge – allow up to 3
hours for charging symbol to appear – leave
on charge until fully charged. This may take
up to 8 hours.
If after leaving on
charge for 8+ hours the
unit will not turn on
please contact
service@martek‐
marine.com
Sounds once
every 2
seconds
Red LED’s
flashing
Sensor requires replacement or an
electrical fault exists.
Calibrate the unit (NOT bump test). To
confirm the fault.
If your unit is under
warranty please contact
service@martek‐
marine.com
BAT FAULT warning
NO DISPLAY
UNIT WILL NOT
TURN ON
If problem persists after calibration, replace
sensor.
SENSOR FAULT warning
To purchase a new
sensor please contact
Fastcalgas@martek‐
marine.com
MARINE 4 QUICK START, CALIBRATION AND TROUBLESHOOTING GUIDE
Screen Display
showing…
Audible
alarm status
Sounds once
every 2
seconds
LED status
Probable Cause
Solution
If not resolved
Red LED’s
flashing
Unit has been turned on in the presence
of gas
Leave the instrument ON for 30 minutes in
fresh air.
Then switch off and back on again still in
fresh air.
If the fault persists
contact
service@martek‐
marine.com
Or
The cell has been over exposed to gas
ALTERNATING WITH A GAS
READING HIGHER THAN 0%
AS SHOWN BELOW
NOTE: The instrument can still be used to
detect and alarm on the other sensor(s)
fitted. Press the
button to continue.
The faulty sensor will continue to display a
flashing spanner symbol to warn the user
that this sensor is NOT working correctly
Sounds once
every 2
seconds
Red LED’s
flashing
Cell is unable to zero.
Apply test gas for two minutes to allow the
display to return to zero then switch
instrument OFF and ON again.
If the fault persists
contact
service@martek‐
marine.com
Flow is restricted due to a blockage in the
sample tube, sample filter or probe.
Check for and clear any blockages found in
sample tube, filter or probe. Restart pump
by pressing
button
Connect a maximum of 30m sample line
Restart pump by pressing the
button
If no blockages found
but alarm move onto
the below step
If tubing is correct
length and no blockages
found but fault remains
contact
service@martek‐
marine.com
ALTERNATING WITH A GAS
READING OF 0%
AS SHOWN BELOW
Sounds
Red LED’s
continuously stay on
Or sample tube is in excess of 30m
LOW FLOW alarm
MARINE 4 QUICK START, CALIBRATION AND TROUBLESHOOTING GUIDE
Screen Display
showing…
Audible
alarm status
Sounds once
every 2
seconds
LED status
Probable Cause
Solution
If not resolved
Red LED’s
stay on
Flow is restricted due to a blockage in the
sample tube, sample filter or probe.
Check for and clear any blockages found in
sample tube, filter or probe. Restart pump
by pressing
button
If no blockages found
but the fault remains
contact
service@martek‐
marine.com
Sounds
Red LED’s
continuously stay on
Calibration of the unit is overdue. The
instrument will operate using its previous
calibration settings; however, as the
sensors response may have diminished,
the instrument should be recalibrated and
tested.
FLOW FAULT alarm
CALIBRATION OVERDUE
warning
None
CALIBRATION EXPIRED
warning
Red LED’s
will flash
every 30
seconds
Calibration of the unit is overdue
Press the
button once to
acknowledge that calibration due date is
overdue, cancel the audible / visual alarm,
and continue to the next display.
A ‘CAL EXPIRED’ warning, red backlighting
and red LED’s will flash every 30 seconds.
Calibrate the unit at the earliest opportunity
Calibrate the unit either manually or using
the auto calibration station
If warning remains after
calibration please
contact
service@martek‐
marine.com
If warning remains after
calibrations please
contact
service@martek‐
marine.com
MARINE 4 QUICK START, CALIBRATION AND TROUBLESHOOTING GUIDE
Screen Display
showing…
Audible
alarm status
None
LED status
Probable Cause
Solution
The battery has been allowed to
completely deplete and has reset the
internal clock
SWITCH THE INSTRUMENT ON
Wait until the following screen appears
Push the buttons in this sequence, with a
slight pause between each press.
DATE AND TIME
HAS RESET TO
2000
The countdown timer (top right of the
screen) will now alternate between a
number and the letter M. Once the
countdown has finished the screen will
display:
Normal
Calibrate
Alarms
Setup
Default
Clock
Scroll down (left button) until clock is
highlighted and enter (right button)
If not resolved

advertisement

Key Features

  • Gas detection
  • Marine application
  • Automatic calibration
  • Intelligent alarms
  • Intuitive interface
  • Durable design
  • Data logging
  • Customizable settings

Frequently Answers and Questions

How do I calibrate the PS200 series device?
The PS200 series can be calibrated either manually or using an automatic calibration station. Refer to the manual for detailed instructions.
What types of gas can the PS200 series detect?
The PS200 series can detect a range of gases, including methane, propane, and carbon monoxide. Check the specifications in the manual for a complete list.
What should I do if the PS200 series displays a 'LOW BATTERY' warning?
Plug the device into a charger and allow it to charge fully. Refer to the manual for the estimated charging time.

Related manuals

Download PDF

advertisement