TESTO 550s Instruction Manual
The Testo 550S is a digital manifold designed for servicing and maintaining refrigeration systems and heat pumps. It allows qualified technicians to measure and monitor pressure, temperature, and other critical parameters. The Testo 550S is compatible with most non-corrosive refrigerants, water and glycol but not compatible with refrigerants containing ammonia. It features two valve positioners for controlling refrigerant flow, a clear display for monitoring readings, and Bluetooth connectivity for data transfer and analysis.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
testo 550s / testo 557s digital manifold Instruction manual Contents Contents 1 2 3 4 5 6 6.1 About this document ...........................................................................5 Safety and disposal..............................................................................5 Product-specific approvals .................................................................5 Product-specific information...............................................................6 Use.........................................................................................................7 Product description .............................................................................8 Overview of the testo 550s ................................................................................. 8 6.2 Overview of the testo 557s ................................................................................. 9 6.3 Overview of main menu .................................................................................... 10 6.4 Control keys ..................................................................................................... 11 7 First steps ...........................................................................................11 7.1 Inserting (rechargeable) batteries ..................................................................... 11 7.2 Switching the instrument on and off .................................................................. 12 7.3 Setup wizard..................................................................................................... 12 8 Using the product...............................................................................13 8.1 Preparing for measurement .............................................................................. 13 8.1.1 Operating the valve positioners ................................................................................ 13 8.1.2 Automatic mode .......................................................................................................14 8.2 Measuring mode ............................................................................................... 14 8.2.1 Refrigeration ............................................................................................................14 8.2.2 Evacuation ...............................................................................................................19 8.2.3 Pressure Leak Test ..................................................................................................22 8.2.4 Target Superheat .....................................................................................................25 8.2.5 Compressor Test (DLT)............................................................................................29 8.2.6 Delta T .....................................................................................................................31 8.3 Bluetooth .......................................................................................................... 33 8.3.1 Probes compatible with the instrument ..................................................................... 33 8.3.2 Establishing a connection ......................................................................................... 34 8.3.3 Switching on/off........................................................................................................34 8.3.3.1 Switching on ............................................................................................................................35 8.3.3.2 Switching off ............................................................................................................................35 8.3.3.3 Manual probe selection ...........................................................................................................36 Contents 8.4 Settings ............................................................................................................ 36 8.4.1 Backlight duration .................................................................................................... 37 8.4.2 Backlight brightness................................................................................................. 38 8.4.3 Auto Off ................................................................................................................... 39 8.4.4 Auto Tfac (Temperature compensation factor) ......................................................... 40 8.4.5 Units ........................................................................................................................ 41 8.4.6 Language ................................................................................................................ 42 8.4.7 Setup Wizard ........................................................................................................... 43 8.4.8 Restore factory settings ........................................................................................... 44 8.4.9 Device Info .............................................................................................................. 45 9 Smart App........................................................................................... 46 9.1 App – user interface ......................................................................................... 46 9.2 Main menu ....................................................................................................... 47 9.3 Measurement menu.......................................................................................... 48 9.3.1 Basic view ............................................................................................................... 49 9.3.1.1 Graphic view ........................................................................................................................... 49 9.3.1.2 Table view ............................................................................................................................... 50 9.3.2 Refrigeration ............................................................................................................ 50 9.3.3 Target superheat ..................................................................................................... 53 9.3.4 System Leak Test .................................................................................................... 55 9.3.5 Evacuation .............................................................................................................. 57 9.4 Customer.......................................................................................................... 58 9.4.1 Creating and editing a customer .............................................................................. 58 9.4.2 Creating and editing measuring sites ....................................................................... 59 9.5 Memory ............................................................................................................ 61 9.5.1 Searching for and deleting measurement results ...................................................... 61 9.6 Sensors ............................................................................................................ 62 9.6.1 Information .............................................................................................................. 62 9.6.2 Settings ................................................................................................................... 63 9.7 Settings ............................................................................................................ 63 9.7.1 Language ................................................................................................................ 63 9.7.2 Measurement settings ............................................................................................. 64 9.7.3 Company details ...................................................................................................... 64 9.7.4 Privacy settings ....................................................................................................... 64 9.8 Help and Information ........................................................................................ 65 9.8.1 Instrument information ............................................................................................. 65 9.8.2 Tutorial .................................................................................................................... 65 9.8.3 Exclusion of liability ................................................................................................. 66 Contents 9.9 testo DataControl archiving software ................................................................ 66 9.9.1 System requirements ...............................................................................................66 9.9.1.1 Operating system ....................................................................................................................66 9.9.1.2 PC............................................................................................................................................67 9.9.2 Procedure ................................................................................................................67 10 Maintenance........................................................................................69 10.1 Calibration ........................................................................................................ 69 10.2 Cleaning the instrument .................................................................................... 69 10.3 Keeping connections clean ............................................................................... 69 10.4 Removing oil residues ...................................................................................... 69 10.5 Ensuring measuring accuracy ........................................................................... 69 10.6 Changing batteries/rechargeable batteries........................................................ 70 11 12 Technical data ....................................................................................70 Tips and assistance ...........................................................................73 12.1 Questions and answers .................................................................................... 73 12.2 Error Codes ...................................................................................................... 73 12.2.1 Main screen .............................................................................................................73 12.2.2 Status view ..............................................................................................................74 12.3 Accessories and spare parts ............................................................................. 74 13 Support ...............................................................................................74 1 About this document 1 About this document • The instruction manual is an integral part of the instrument. • Pay particular attention to the safety instructions and warning advice in order to prevent injury and damage to the product. • Please read this instruction manual through carefully and familiarize yourself with the product before putting it to use. • Familiarity with a PC as well as the Microsoft® products is assumed in this documentation. Symbols and writing standards Display Explanation Note: basic or further information 1 2 … > Menu [OK] Warning advice, risk level according to the signal word: Warning! Serious physical injury may occur. Caution! Minor physical injury or damage to the equipment may occur. > Take the specified precautionary measures. Action: several steps, the sequence must be followed Result of an action Requirement Action Elements of the instrument, the instrument display or the program interface. Control keys of the instrument or buttons of the program interface. 2 Safety and disposal Please observe the Testo information document (enclosed with the product). 3 Product-specific approvals Please find the current country approvals in the Approvals and Certifications document. 5 4 Product-specific information 4 Product-specific information • The measuring instrument being dropped or any other comparable mechanical stress may cause breakage of the pipe pieces in the refrigerant hoses. The valve positioners may also suffer damage, causing further damage inside the measuring instrument that is not necessarily visible externally. Therefore, always replace the refrigerant hoses with new ones after the measuring instrument is dropped or after any comparable mechanical stress. For your own safety, you should return the measuring instrument to the Testo Customer Service for technical inspection. • Electrostatic charging may destroy the instrument. Integrate all the components (system, manifold's valve block, refrigerant bottle, etc.) into the potential bonding (earthing). Please see the safety instructions for the system and the refrigerant used. • Refrigerant gases can harm the environment. Please note the applicable environmental regulations. • Use with A2L refrigerants Testo measuring instruments (as of July 2020) can be used in compliance with the prescribed laws, standards, directives and safety regulations for refrigeration systems and refrigerants as well as regulations of the manufacturers of refrigerants of safety group A2L as per ISO 817. Regional standardization and interpretation must always be observed. For example, DIN EN 378-Part 1-4 applies to the scope of the EN standards. During maintenance work, the employer must ensure that a hazardous explosive atmosphere is prevented (see also: TRBS1112, TRBS2152 VDMA 24020-3). A hazardous and potentially explosive atmosphere must be anticipated during maintenance and repair work on refrigeration systems with flammable refrigerants (e.g. those of category A2L and A3). Maintenance, repairs, removal of refrigerants and commissioning of systems may only be carried out by qualified personnel. 6 5 Use 5 Use testo 550s and testo 557s are digital manifolds for maintenance and service work on refrigeration systems and heat pumps. They may only be used by qualified authorized personnel. The functions of the instruments testo 550s and testo 557s mean they can replace mechanical manifolds, thermometers and pressure/temperature charts. Pressures and temperatures can be applied, adapted, tested and monitored. The instruments testo 550s and testo 557s are compatible with most noncorrosive refrigerants, water and glycol. The instruments testo 550s and testo 557s are not compatible with refrigerants containing ammonia. The products must not be used in potentially explosive atmospheres! 7 6 Product description 6 Product description 6.1 Overview of the testo 550s Mini-DIN probe socket for NTC 1 temperature probe, with socket cover 2 Foldable suspension device (rear). Rear • Battery compartment • 3 Display. Instrument status icons 4 5 Control keys 6 Sight glass for refrigerant flow 7 2 x valve positioner 8 3 x connections 7/16" UNF, brass 9 Left/right: Low/high pressure for refrigerant hoses with quick 8 Mini USB port for firmware update It is not possible to charge rechargeable batteries in the instrument. 3 x hose bracket for refrigerant hoses 6 Product description release screw fitting; passage can be closed via valve positioner Centre: e.g. for refrigerant bottles, with sealing cap, refrigerant hoses with quickrelease screw fitting; passage can be closed via valve positioner. 6.2 Overview of the testo 557s Mini-DIN probe socket for NTC 1 temperature probe, with socket cover 2 Foldable suspension device (rear). Rear • Battery compartment • Mini USB port for firmware update 3 Display. Instrument status icons 4 5 Control keys 6 Sight glass for refrigerant flow It is not possible to charge rechargeable batteries in the instrument. 9 6 Product description 7 4 x valve positioner 8 4 x hose bracket for refrigerant hoses Connection 7/16" UNF, brass. High pressure, for refrigerant 9 hoses with quick-release screw fitting, passage can be closed via valve positioner. 10 Connection 5/8" UNF, brass, for vacuum pump Connection 7/16" UNF, brass, 11 e.g. for refrigerant cylinders, with sealing cap Connection 7/16" UNF, brass. Low pressure, for refrigerant 12 hoses with quick-release screw fitting, passage can be closed via valve positioner. 6.3 Overview of main menu Measuring mode Bluetooth® Settings 10 Refrigeration Evacuation System Leak Test Target Superheat Compressor Test (DLT) Delta T Connection to the testo Smart App or Smart Probes Backlight duration Backlight brightness Auto Off Auto Tfac (Temperature compensation factor) Units Language Setup Wizard Restore factory settings Instrument information 7 First steps Control keys 6.4 Symbol Meaning • Open menu • Confirm input • Switch on the display illumination: Press and hold the key for >2s • Switch off the display illumination: Press and hold the key for >2s Change/navigate the display screen. / • Switches to the measurement view • Back to the menu • Switch off the instrument: Press and hold the key for >2s 7 First steps 7.1 Inserting (rechargeable) batteries 1 Unfold the suspension hook and open the battery compartment (clip lock). 2 Insert the batteries (scope of delivery) or rechargeable batteries (4 x 1.5V, type AA / Mignon / LR6) into the battery compartment. Observe the polarity! 3 Close the battery compartment. insertion of the batteries, the instrument switches on After automatically and goes into the settings menu. When not in use for a long period: Take out the (rechargeable) batteries. 11 7 First steps Switching the instrument on and off 7.2 Current status Action Instrument off Press Function Instrument is switched on. When the measuring instrument is started for the first time, the setup wizard guides you through the following setting parameters step by step: - Language - testo Smart App Instrument on Press and hold down (> 2 s) Instrument is switched off. The instrument setup that is implemented can be adapted at any time in the Settings menu. 7.3 Setup wizard When the testo 550s / testo 557s is started up for the first time and after the factory settings have been reset, the setup wizard is activated and guides you step-by-step through the following setup parameters. The instrument setup that is implemented can be adapted at any time in the Settings menu. Language selection and QR code Instrument is switched on and the initialization phase has been completed. 1 Select language: Press [▲] / [▼] and [Menu/Enter] to confirm. Selecting the language activates the appropriate presetting of the units of measurement 12 8 Using the product 2 Take a photo of the QR code of the testo Smart App and press [Menu/Enter] to confirm. - The measurement menu is displayed. 8 Using the product 8.1 Preparing for measurement 8.1.1 Operating the valve positioners With respect to the refrigerant path, the digital manifold behaves just like a conventional four-way manifold (only applies to the testo 557s): The passages are opened by opening the valves. The applied pressure is measured with the valves closed and the valves opened. - Open the valve: Turn valve positioner anticlockwise. - Close the valve: Turn valve positioner clockwise. WARNING Valve positioner tightened too tightly. • Damage to the PTFE seal (1). • Mechanical deformation of the valve piston (2) leading to the PTFE seal (1) falling out. Damage to the thread of the threaded spindle (3) and the valve screw (4). Broken valve knob (5). • Only tighten the valve positioner hand-tight. Do not use any tools to tighten the valve positioners. 13 8 Using the product 8.1.2 Automatic mode The manifold automatically detects the pressure difference between the lowpressure and high-pressure sides. If the measured pressure on the low pressure side is 1 bar higher than on the high pressure side, a dialogue appears and the display can be changed accordingly. If “yes” is selected, the low pressure moves from left to right and the high pressure moves from right to left. This mode is particularly suitable for air conditioning systems that provide cooling and heating. 8.2 Measuring mode WARNING Risk of injury caused by refrigerant that is under high pressure, hot, cold, or toxic! > Wear protective goggles and safety gloves. > Before applying pressure to the measuring instrument: Always fasten the measuring instrument on the suspension hook to prevent it from falling (danger of breakage). > Before each measurement, check the refrigerant hoses are intact and connected properly. Do not use any tools to connect the hoses; only tighten hoses hand-tight (max. torque 5.0 Nm/3.7 ft*lb). > Comply with the permissible measuring range (-1 to 60 bar/-14.7 to 870 psi). Pay particular attention to this in systems with R744 refrigerant, since these are frequently operated at higher pressures! 8.2.1 Refrigeration The Refrigeration application is used to determine the following system measuring values: • High pressure • Low pressure • Refrigerant evaporating temperature • Refrigerant condensation temperature • Temperature of suction line • Temperature of liquid line 14 8 Using the product • Superheating • Subcooling An NTC temperature probe (accessory) must be connected for measuring the pipe temperature and for automatic calculation of superheating and subcooling. These can be fixed cable temperature probes or Testo Smart Probes (e.g. testo 115i). Before each measurement, check that the refrigerant hoses are in flawless condition. Before each measurement, zero the pressure sensors. All connections must be pressureless (ambient pressure). Press the [▲] (P=O) key for 2 seconds to zero the sensors. The instrument is switched on and the measurement menu is displayed. All connections must be pressureless (ambient pressure). 1 Press [Menu/Enter] to confirm. - The main menu is displayed. 2 Press [Menu/Enter] to confirm. 15 8 Using the product 3 Select Refrigeration and press [Menu/Enter] to confirm. - The measurement menu is displayed. 4 Connect the refrigerant hoses. 4.1 Close the valve positioners. 4.2 Connect the refrigerant hoses for low-pressure side (blue) and highpressure side (red) to the measuring instrument. 4.3 Connect the refrigerant hoses to the system. 5 Connect testo 115i or fixed cable probes. 16 8 Using the product 6 Set refrigerant. 6.1 Press the key [▼] (Rxx) (refrigerant number according to ISO 817). - The refrigerant menu opens and the current refrigerant is highlighted. 6.2 Setting the refrigerant: Press [▲] or [▼] to select the refrigerant and press [Menu/Enter] to confirm. You have the option of setting up favourite refrigerants on your instrument and in the app. These then appear at the beginning of the refrigerant list. To do this, the app must be connected to the instrument via Bluetooth. In the refrigerant list (App), you can now choose the refrigerant as a favourite by clicking on the star. The new favourite refrigerant will now be synchronized to the testo 550s or testo 557s. Note: During synchronization, the refrigerant list/selection on the instrument must remain closed. - The newly set refrigerant is displayed in the refrigerant list. 7 Press the [▲] (P=O) key for 2 seconds to zero the sensors. - Zeroing takes place. 17 8 Using the product 8 Pressurize the measuring instrument. - The measurement starts automatically. - Measurement results are displayed: • Low/high pressure • Condensation and evaporation temperature • Suction and liquid line temperature • Superheating and subcooling With zeotropic refrigerants, the evaporation temperature to/Ev is displayed after complete evaporation/the condensation temperature tc/Co is displayed after complete condensation. The measured temperature must be assigned to the superheating or subcooling side (toh <--> tcu). Dependent on this assignment, the display will show toh/T1 resp. Δtoh/SH or tcu/T2 resp. Δtcu/SC, depending on the selected display. Reading and display illumination flash: • 1 bar/14.5 psi before reaching critical refrigerant pressure • When max. permissible pressure of 60 bar/870 psi is exceeded. All values can be saved and sent in the app. The data can also be transferred between the app and the testo DataControl software. 18 8 Using the product 8.2.2 Evacuation Via the Evacuation application, foreign gases and moisture can be removed from the refrigeration circuit. The testo 552i is recommended for carrying out the measurement. The measurement is also possible without the testo 552i, with testo 550s/testo 557s. However, this is not advisable due to insufficient accuracy. The instrument is switched on and the measurement menu is displayed. ® Bluetooth is enabled. Hoses are connected. 1 Press [Menu/Enter]. 2 Press [▲] / [▼] to select Measuring Mode and press [Menu/Enter] to confirm. - The Measuring Mode menu is displayed. 19 8 Using the product 3 Press [▲] / [▼] to select Evacuation and press [Menu/Enter] to confirm. - The Configure Target Lines menu is displayed. 4 Adjust the Target Line value 4.1 Press the [▲] key and in the Evacuation Target field, select Manual Input. 4.2 Press [Menu/Enter] to confirm. - The field is activated. 4.3 Press [▲] / [▼] to set the value. 4.4 Press [Menu/Enter] to confirm. 5 Adjust the Maximum Decay Target value. 5.1 Press the [▼] key and in the Maximum Decay Target field, select Manual Input. 20 8 Using the product 5.2 Press [Menu/Enter] to confirm. - The field is activated. 5.3 Press [▲] / [▼] to set the value. 5.4 Press [Menu/Enter] to confirm. 6 Confirm the entries in steps 4 and 5: Press [▼] to select OK and press [Menu/Enter] to confirm. - A connection is established with available Bluetooth® probes. - testo 552i is switched on and connected automatically. - The Evacuation measurement menu is displayed. 21 8 Using the product 7 Start measurement: Press the [▼] (Start) key. - Once the measuring range 0 to 20,000 microns / 0 to 26.66 mbar is reached, the current vacuum value is shown on the instrument display. The instrument also displays the current ambient temperature, the water evaporation temperature which corresponds to the vacuum reading, and the delta between these two temperatures. 8 End measurement: Press the [▼] (Stop) key. - The measurement result is displayed. Press the [▲] New key to reset the determined values. If necessary, a test can also be started again. 9 Press [Menu/Enter] to return to the main menu. 8.2.3 Pressure Leak Test The temperature compensated leakage test can be used to check the leak tightness of systems. For this purpose both the system pressure and the ambient temperature are measured over a defined period of time. To this end, a temperature probe that measures the ambient temperature or a Smart Probe for measuring the air temperature can be connected. As a result, information is provided about the temperaturecompensated differential pressure and about the temperature at the beginning/end of the test. Due to the temperature compensation, the actual pressure drop is displayed as delta P. If no temperature probe is connected, you may also perform the tightness test without temperature compensation. Surface temperature probes (e.g. testo 115i) can also be used for the temperature-compensated leakage testing, but must not be used to measure surface temperature. They must be positioned as far as possible to measure the air temperature. If a surface probe is used, in the Settings menu of the testo 550s / testo 557s, the Auto Tfac (Temperature compensation factor) must be switched off, see section 8.3.4. 22 8 Using the product The testo 550s or testo 557s manifold is used to carry out the measurement. The instrument is switched on and the measurement menu is displayed. Hoses are connected. 1 Press [Menu/Enter]. 2 Press [▲] / [▼] to select Measuring Mode and press [Menu/Enter] to confirm. - The Measuring Mode menu is displayed. 3 Press [▲] / [▼] to select Pressure Leak Test and press [Menu/Enter] to confirm. 23 8 Using the product - For the temperature-compensated leak test, a connection is established with available Bluetooth® probes. If cable probes are connected to the instrument, they are prioritized for the compensation. Please note that only air probes are ideal for temperaturecompensated leakage testing. - testo 905i / testo 605i is switched on and automatically connected. Other temperature probes compatible with testo 550s / testo 557s can be connected. - The Pressure Leak Test menu is displayed. T Comp is shown on the display if a compatible probe is connected via Bluetooth® or cable. The temperature compensation is used for the measurement result. 4 Press the [▼] (Start) key. - The leakage test is carried out. 5 Press the [▼] (Stop) key. - The leakage test is terminated. - The measurement result is displayed. Press the [▲] New key to reset the determined values. If necessary, a test can also be started again. The measurement result can be displayed graphically on the manifold as well as in the app. 6 Press [Menu/Enter] to return to the main menu. 24 8 Using the product Target Superheat 8.2.4 This feature makes it possible to connect the testo 550s and testo 557s manifolds to two additional testo 605i Smart Probes in order to calculate the target superheat. This application can only be used for split air conditioning systems/heat pumps with a fixed expansion valve. The two connected testo 605i Smart Probes determine the ODDB and RAWB values. The target superheat value appears on the display as a result. The following are used for the measurement: • testo 115i (clamp thermometer) or • fixed cable probes • testo 605i in practice. Alternatively, the values can be configured manually. Before each measurement, check that the refrigerant hoses are in flawless condition. Before each measurement, zero the pressure sensors. The instrument is switched on and the measurement menu is displayed. All connections must be pressureless (ambient pressure). ® Bluetooth is enabled. 1 Press [Menu/Enter]. 2 Press [▲] / [▼] to select Measuring Mode and press [Menu/Enter] to confirm. 25 8 Using the product - The Measuring Mode menu is displayed. 3 Press [▲] / [▼] to select Target Superheat and press [Menu/Enter] to confirm. - The Configure Target Superheat menu is displayed. The values can either be configured manually via Manual Input or recorded by the testo 605i via Smart Probe. When a Smart Probe is selected, available testo 650i instruments are displayed for the connection. 4 Adjust values for Outdoor Dry Bulb Temp. 4.1 Press the [▲] key and in the Outdoor Dry Bulb Temp. field, select Manual Input. 4.2 Press [Menu/Enter] to confirm. - The field is activated. 4.3 Press [▲] / [▼] to set the value. 4.4 Press [Menu/Enter] to confirm. 5 Adjust the Return Air Wet Bulb Temp. value 26 8 Using the product 5.1 Press the [▲] / [▼] key and in the Return Air Wet Bulb Temp. field, select Manual Input. 5.2 Press [Menu/Enter] to confirm. - The field is activated. 5.3 Press [▲] / [▼] to set the value. 5.4 Press [Menu/Enter] to confirm. 6 Confirm the entries in steps 4 and 5: Press [▼] to select Okay and press [Menu/Enter] to confirm. - The Target Superheat measurement menu is displayed. 7 Connect the refrigerant hoses. 7.1 Close the valve positioners. 27 8 Using the product 7.2 Connect the refrigerant hoses for low-pressure side (blue) and highpressure side (red) to the measuring instrument. 7.3 Connect the refrigerant hoses to the system. 8 Connect testo 115i/fixed cable probes. 9 Set refrigerant. 9.1 Press the key [▼] (Rxx) (refrigerant number according to ISO 817). - The refrigerant menu opens and the current refrigerant is highlighted. 9.2 Setting the refrigerant: Press [▲] or [▼] to select the refrigerant and press [Menu/Enter] to confirm. - The newly set refrigerant is displayed in the refrigerant list. 10 Press the [▲] (P=O) key for 2 seconds to zero the sensors. - Zeroing takes place. 11 Pressurize the measuring instrument. - The measurement starts automatically. - Measurement results are displayed: • Low/high pressure 28 • Condensation and evaporation temperature • Suction and liquid line temperature • Superheating and subcooling • Target superheat TSH 8 Using the product 8.2.5 Compressor Test (DLT) For this mode, 3 temperature probes are used. In addition to the conventional temperature sensors for superheating and subcooling, an additional temperature probe must be connected via Bluetooth. The testo 115i (clamp thermometer) or fixed cable probes are used to carry out the measurement. Before each measurement, check that the refrigerant hoses are in flawless condition. Before each measurement, zero the pressure sensors. The instrument is switched on and the measurement menu is displayed. 1 Press [Menu/Enter]. 2 Press [▲] / [▼] to select Measuring Mode and press [Menu/Enter] to confirm. - The Measuring Mode menu is displayed. 29 8 Using the product 3 Press [▲] / [▼] to select Compressor Test (DLT) and press [Menu/Enter] to confirm. measurement menu is The displayed. The DLT temperature is shown on the display. 4 Connect the refrigerant hoses. 4.1 Close the valve positioners. 4.2 Connect the refrigerant hoses for low-pressure side (blue) and highpressure side (red) to the measuring instrument. 4.3 Connect the refrigerant hoses to the system. 5 Connect 2 x testo 115i or 2 x fixed cable probes and third temperature probe to the compressor outlet. 6 Set refrigerant. 6.1 Press the key [▼] (Rxx) (refrigerant number according to ISO 817). - The refrigerant menu opens and the current refrigerant is highlighted. 30 8 Using the product 6.2 Setting the refrigerant: Press [▲] or [▼] to select the refrigerant and press [Menu/Enter] to confirm. - The newly set refrigerant is displayed in the refrigerant list. 7 Press the [▲] (P=O) key for 2 seconds to carry out zeroing. - Zeroing takes place. 8 Pressurize the measuring instrument. - The measurement starts automatically. - The measurement result is displayed. 9 Press [Menu/Enter] to return to the main menu. 8.2.6 Delta T Temperature 1 and temperature 2 are measured. The difference is shown on the display as the delta temperature. Two testo 115i (clamp thermometers) or fixed cable probes are used to carry out the measurement. The instrument is switched on and the measurement menu is displayed. The steps described in the Preparing for measurement section have been followed/carried out. testo 115i is switched on. 1 Place the testo 115i at the measuring points. 2 Press [Menu/Enter]. 31 8 Using the product 3 Press [▲] / [▼] to select Measuring Mode and press [Menu/Enter] to confirm. - The Measuring Mode menu is displayed. 4 Press [▲] / [▼] to select Delta T and press [Menu/Enter] to confirm. - The measurement result is displayed. 5 Press [Menu/Enter] to return to the main menu. 32 8 Using the product 8.3 Bluetooth testo 550s / testo 557s have the option of establishing a Bluetooth® connection with wireless probes as well as a connection to the testo Smart App at the same time. If the testo 550s or testo 557s is used with Smart Probes, they must be at least 20 cm apart. 8.3.1 Probes compatible with the instrument Smart Probes Order no. Designation 0560 2115 02 testo 115i – clamp thermometer with smartphone operation testo 605i – thermohygrometer with smartphone operation testo 552i – vacuum Smart Probe testo 905i – temperature probe with smartphone operation 0560 2605 02 0564 2552 01 0560 1905 NTC probes Order no. Designation 0613 1712 0613 5505 Robust air temperature probe (NTC) Clamp probe (NTC) for temperature measurements on pipes (Ø 6-35 mm), 1.5 m fixed cable Clamp probe (NTC) for temperature measurements on pipes (Ø 6-35 mm), 5 m fixed cable 2 x clamp probes (NTC) for temperature measurements on pipes (Ø 6-35 mm), 1.5 m fixed cable Temperature probe with Velcro (NTC) Pipe wrap probe (NTC), measuring range: 50 to +120 °C Waterproof surface temperature probe (NTC) for flat surfaces, measuring range: -50 to +150 °C 0613 5506 0613 5507 0613 4611 0613 5605 0613 1912 33 8 Using the product 8.3.2 Establishing a connection To establish a connection via Bluetooth®, you need a tablet or smartphone with the Testo Smart App installed on it. You can get the App for iOS instruments in the App Store or for Android instruments in the Play Store. Compatibility: Requires iOS 12.0 or later/Android 6.0 or later, requires Bluetooth® 4.0. Once the connection between the app and the Testo manifold has been successfully established, the app is in second screen mode. This is indicated by a yellow frame in the app. This means that all measurement data from the manifold is being mirrored on the app. The measurement can now be controlled from both instruments. It is possible to carry out the following actions: • Start measurement • Stop measurement • Reset measurement • Configure measurement • Select refrigerant 8.3.3 Switching on/off The instrument is switched on and the measurement menu is displayed. 1 Press [Menu/Enter]. 2 Press [▲] / [▼] to select Bluetooth: and press [Menu/Enter] to confirm. - The Bluetooth menu is displayed. 34 8 Using the product 8.3.3.1 Switching on The Bluetooth menu is selected. 1 [Menu/Enter] - In the On/Off switch icon, displayed. is 2 Enable Bluetooth®: Press [▼] to activate the [Completed] button and press [Menu/Enter] to confirm. - When the Bluetooth® icon is shown on the display, Bluetooth is switched on. - Bluetooth® automatically searches for and connects available probes. - After opening the App, the instrument is automatically connected if it is within range. The instrument does not have to be connected to the smartphone/tablet via settings beforehand. 8.3.3.2 Switching off The Bluetooth® menu is activated. 1 [Menu/Enter] - In the On/Off switch icon, displayed. is 3 Disable Bluetooth®: Press [▼] to activate the [Competed] button and press [Menu/Enter] to confirm. - When the Bluetooth® icon is not shown on the display, Bluetooth® is switched off. 35 8 Using the product Manual probe selection 8.3.3.3 If this menu is activated, it appears before a measurement. The Bluetooth® menu is activated (in the On/Off switch icon, displayed. is 1 Press [▼] to select Manual probe selection. Enable function: Via [Menu/Enter], set the switch to [ON]. An info window with the available probes appears before each measurement to be carried out. The information must be confirmed by pressing [Menu/Enter]/[Okay]. Disable function: Via [Menu/Enter], set the switch to [OFF]. If the advanced Bluetooth® settings are switched off, the instrument automatically connects to the first compatible Smart Probe. 2 Press [▼] to click on the [Completed] button and press [Menu/Enter] to confirm. In the Bluetooth® menu, you will obtain further information. Display Explanation flashes is permanently displayed is not displayed 8.4 There is no Bluetooth® connection, or a potential connection is being searched for. There is a Bluetooth® connection, the number of connected Bluetooth® probes is displayed next to it. Bluetooth® is disabled. Settings The instrument is switched on and the measurement menu is displayed. 1 Press [Menu/Enter]. 36 8 Using the product 2 Select Settings: Press [▼] and then [Menu/Enter] to confirm. - The Settings menu is displayed. Available settings: • Backlight duration 8.4.1 • Backlight brightness • Auto Off • Auto Tfac (Temperature compensation factor) • Units • Language • Setup Wizard • Restore factory settings • Instrument information Backlight duration Set the backlight duration for the display. The Settings menu is activated. 1 Press [▲] / [▼] to select Backlight duration and press [Menu/Enter] to confirm. - Menu properties are displayed. 37 8 Using the product 2 Press [▲] / [▼] to select the backlight duration and press [Menu/Enter] to confirm. 3 Press [ESC]: 1x main menu view, 2 x measurement menu view. 8.4.2 Backlight brightness Set the brightness for the display. The Settings menu is activated. 1 Press [▲] / [▼] to select Backlight brightness and press [Menu/Enter] to confirm. - Menu properties are displayed. 2 Press [▲] / [▼] to select the brightness value (25%, 50%, 75%, 100%) and press [Menu/Enter] to confirm. 3 Press [ESC]: 1x main menu view, 2 x measurement menu view. 38 8 Using the product Auto Off 8.4.3 You can manage the energy consumption for your instrument yourself. The Settings menu is activated. 1 Press [▲] / [▼] to select [Auto OFF] and press [Menu/Enter] to confirm. - Menu properties are displayed. 2 Select using [▲] / [▼] • On: The instrument automatically switches off after 30 minutes of inactivity. The instrument switches off automatically if no pressure is measured and no key was pressed within 10 minutes. As long as pressure is present, the instrument remains on. • Off: Continuous operation 3 Press [Menu/Enter] to confirm selection. 4 Press [ESC]: 1x main menu view, 2 x measurement menu view Unsaved readings are lost when the measuring instrument is switched off. 39 8 Using the product 8.4.4 Auto Tfac (Temperature compensation factor) A surface compensation factor has been set in the measuring instrument to reduce the measuring errors in the main field of applications. This reduces measuring errors when using surface temperature probes. Surface temperature probe An NTC temperature probe (accessory) must be connected for measuring the pipe temperature and for automatic calculation of superheating and subcooling. The Settings menu is activated. 1 Select Auto Tfac and press [Menu/Enter] to confirm. - Menu properties are displayed. 2 Press [▲] / [▼] to select On/Off and press [Menu/Enter] to confirm. Press [▲] / [▼] to select the question mark icon and [Menu/Enter] to open. You will obtain further information on temperature compensation. 3 Press [ESC]: 1x main menu view, 2 x measurement menu view 40 8 Using the product 8.4.5 Units The Settings menu is activated. 1 Press [▲] / [▼] to select [Units] and press [Menu/Enter] to confirm. - Menu properties are displayed. Adjustable units Measurement parameter Unit Description Temperature °C, °F Set unit for temperature. Pressure psi, kPa, MPa, bar Set unit for pressure. Prel, Pabs Depending on the chosen unit for pressure: Change between absolute and relative pressure displays. bar (Pabs) Set the current absolute pressure (current air pressure values for your region can be obtained, for example, from the local weather service or on the Internet). Pressure mode Absolute pressure Vacuum pressure Micron, mbar, Torr, mTorr inH2O, in Hg, hPa, Pa 41 8 Using the product 3 Press [ESC]: 1 x Unitsmenu, 2 x main menu view, 3 x measurement menu view. 8.4.6 Language The Settings menu is activated. 1 Press [▲] / [▼] to select [Language] and press [Menu/Enter] to confirm. - Menu properties are displayed. 2 Select language: Press [▲] / [▼] and [Menu/Enter] to confirm. Selecting the language activates the appropriate presetting of the units of measurement. 3 Press [ESC]: 1 x Units menu, 2 x main menu view, 3 x measurement menu view. 42 8 Using the product 8.4.7 Setup Wizard The Settings menu is activated. 1 Press [▲] / [▼] to select [Setup Wizard] and press [Menu/Enter] to confirm. - Language selection opens. 2 Press [▲] / [▼] to select the language. - The units for the respective country are set automatically. - Barcode is displayed and the app can be downloaded from the respective app store. 43 8 Using the product 8.4.8 Restore factory settings The instrument is reset to the factory settings. The Settings menu is activated. 1 Press [▲] / [▼] to select [Factory Reset] and press [Menu/Enter] to confirm. - Menu properties are displayed. 2 Start [Factory Reset]: Press [▲] / [▼] to select [Do a factory reset] and press [Menu/ESC] to confirm. Press [Back] to quit the process. - [Factory Reset] is carried out. 3 See Setup Wizard. 44 8 Using the product Device Info 8.4.9 The Settings menu is activated. 1 Press [▲] / [▼] to select [Device Info] and press [Menu/Enter] to confirm. - The Versions Info menu is displayed. The following information can be viewed: • Serial number • Firmware version • Refrigerant version • BLE (Bluetooth Low Energy) version 2 Press [ESC]: 1 x Units menu, 2 x main menu view, 3 x measurement menu view. 45 9 Smart App 9 Smart App 9.1 App – user interface 1 2 Open main menu Display of the measurement period 3 Display of calculated measurement results 4 Reading for each probe 5 Can be controlled with different function keys 6 Instrument status bar 7 Configuration 8 Edit reading display 46 9 Smart App Further symbols on the user interface (without numbering) One level back Exit view Share report Search Favourite Delete Further information Display report Multiple selection 9.2 Main menu The Main menu can be accessed via the icon at top left. To exit the main menu, select a menu or right-click on the guided menus. The last screen displayed is shown. Measure Customer Memory Sensors Settings Help and Information 47 9 Smart App Additional icons: One level back Deleting Exit view Further information Share measurement data/reports Display report Search Edit Favourite 9.3 Measurement menu The testo 550s / testo 557s has permanently installed measurement programs. These enable the user to carry out convenient configuration and implementation of specific measuring tasks. The testo 550s / testo 557s offers the following Measurement menus: Basic view Volume flow – duct Volume flow – outlet Differential temperature (ΔT) Differential pressure (ΔP) Refrigeration Target superheat Cooling and heating output Leakage test Evacuation 48 9 Smart App 9.3.1 Basic view In the Basic view application menu, the current measuring values can be read, recorded and saved. The Basic view is particularly suitable for fast, uncomplicated measurements without the specific requirements of a standardcompliant measurement. All Bluetooth® probes compatible with the testo Smart App are displayed in the Basic view angezeigt. In all application menus, apart from the volume flow measurement, there are three different screens for the measurement - Live (or also Basic view), Graphic and Table. 9.3.1.1 Graphic view In the Graphic view, the values for a maximum 4 channels can be displayed simultaneously in a chronological trend graph. All measured parameters can be displayed in the Graphic view via the channel selection (click on one of the four selection fields). Once a measurement parameter has been selected, the value is updated automatically. The Zoom touch function allows individual parts of the graphic to be viewed in more detail or time progressions to be displayed compactly. 1 Open the main menu 2 Change of display 3 Reading for selected channel 4 Measurement parameter and measurement unit 5 Graphic with selected channels and 4 Y-axes 6 Status bar 7 Open the configuration menu 8 Selection of other channels 9 Time axis 10 New/Start/Stop/Save button 49 9 Smart App 9.3.1.2 Table view 1 Open the main menu 2 Change of display 3 Column with date and time 4 Arrow keys to go directly to the end of the table 5 Status bar 6 Open the configuration menu 7 Probe ID measurement unit 8 Measuring values 9 New/Start/Stop/Save button 9.3.2 Refrigeration The Refrigeration application is used to determine the following system measuring values: • Low-pressure side: Evaporation pressure, refrigerant evaporation temperature to/Ev (T evap.) • Evaporation pressure: Measured temperature toh/T1 • Evaporation pressure: Superheating Δtoh/SH • High-pressure side: Condensation pressure, refrigerant condensation temperature tc/Co (T condens.) • Condensation pressure: Measured temperature tcu/T2 • Condensation pressure: Subcooling Δtcu/SC The testo 115i clamp thermometer is used for the measurement. An NTC temperature sensor (accessory) must be connected for measuring the pipe temperature and for automatic calculation of superheating and subcooling. Testo Smart Probes (e.g. testo 115i) can be used. 50 9 Smart App Before each measurement, check that the refrigerant hoses are in flawless condition. Before each measurement, zero the pressure sensors. All connections must be pressureless (ambient pressure). Press the button [▲] (P=O) for 2 seconds to zero the sensors. 1 Click on Measure. 2 Click on AC + Refrigeration. The Refrigeration measurement menu opens. 3 Click on . Configuration menu opens. 4 Make the required settings. 5 Click on Apply Configuration. 51 9 Smart App 6 Set refrigerant. You have the option of setting up favourite refrigerants in the App. These then appear at the beginning of the refrigerant list. To do this, click on the asterisk next to the refrigerant in the refrigerant list (App). The newly set refrigerant is displayed in the refrigerant list. 7 Click on Start. The measurement starts. currently being measured are Values displayed. Measured values can be saved or a new measurement can be started. With zeotropic refrigerants, the evaporation temperature to/Ev is displayed after complete evaporation/the condensation temperature tc/Co is displayed after complete condensation. The measured temperature must be assigned to the superheating or subcooling side (toh <--> tcu). Dependent on this assignment, the display will show toh/T1 resp. Δtoh/SH or tcu/T2 resp. Δtcu/SC, depending on the selected display. 52 9 Smart App Reading and display illumination flash: • 1 bar/14.5 psi before reaching critical refrigerant pressure • When max. permissible pressure of 60 bar(870 psi is exceeded. Target superheat 9.3.3 This feature allows the testo 550s / testo 557s manifold to calculate the target superheat in conjunction with the App and additional testo 605i Smart Probes. This application can only be used for split air conditioning systems/heat pumps with a fixed expansion valve. The two connected testo 605i Smart Probes determine the ODDB and RAWB values. As a result, the target superheat appears in the App. The following are used for the measurement: • testo 115i (clamp thermometer) • testo 605i Before each measurement, check that the refrigerant hoses are in flawless condition. Before each measurement, zero the pressure sensors. 1 Click on Measure. 2 Click on Target superheat. The Target superheat measurement menu opens. 3 Click on . Configuration menu opens. 53 9 Smart App 4 Make the required settings. 5 Click on Apply Configuration. 6 Set refrigerant. The newly set refrigerant is displayed in the refrigerant list. 54 9 Smart App 7 Click on Start. The measurement starts. Values currently being measured are displayed. Measured values can be saved or a new measurement can be started. 9.3.4 System Leak Test The temperature compensated leakage test can be used to check the leak tightness of systems. For this purpose both the system pressure and the ambient temperature are measured over a defined period of time. To this end, a temperature probe may be connected to measure the ambient temperature (recommendation: Deactivate the surface compensation factor and use an NTC air probe or Bluetooth® temperature Smart Probes) or Smart Probe for air temperature measurement. This provides information about the temperaturecompensated differential pressure and the temperature at the beginning/end of the test. Due to the temperature compensation, the actual pressure drop is displayed as delta P. If no temperature probe is connected, you may also perform the tightness test without temperature compensation. Surface temperature probes (e.g. testo 115i) can also be used for the temperature-compensated leakage test, but must not be used for measuring surface temperature. They must be positioned as far as possible to measure the air temperature. The 550i, 550s or 557s manifold is used to perform the measurement. 1 Click on Measure. 2 Click on Leakage test. The Leakage test measurement menu opens. 3 Click on . Configuration menu opens. 55 9 Smart App 4 Make the required settings. 5 Click on Apply Configuration. 6 Click on Start. The measurement starts. currently being measured are Values displayed. values are saved. The values can be exported or a report Measured can be created. 56 9 Smart App 9.3.5 Evacuation With the Evacuation application, foreign gases and moisture can be removed from the refrigeration circuit. 1 Click on Measure. 2 Click on Evacuation. The Evacuation measurement menu opens. 3 Click on . Configuration menu opens. 4 Make the required settings. 5 Click on Apply Configuration. 6 Click on Start. The measurement starts. 57 9 Smart App currently being measured are Values displayed. Measured values can be saved or a new measurement can be started. Customer 9.4 In the Customer menu, all customer and measuring site information can be created, edited and deleted. Fields marked with * are mandatory. Without any information in this field, no customers or measuring sites can be stored. 9.4.1 Creating and editing a customer 1 Click on . Main menu opens 2 Click on Customer. The Customer menu opens. 58 9 Smart App 3 Click on + New Customer. A new customer can be created. 4 Store all relevant customer data. 5 Click on Save. The new customer was saved. 9.4.2 Creating and editing measuring sites 1 Click on . Main menu opens 2 Click on Customer. The Customer menu opens. 3 Click on + New Customer. 4 Click on the right tab Measuring point. 59 9 Smart App 5 Click on + New Measuring Point. A new measuring site can be created. 6 Store all relevant measuring site information. 7 Click on right tab Parameters. 8 Select further parameters. For the duct, outlet or duct with k-factor measuring sites, further parameter settings can be implemented. 9 Click on Save. The new measuring site has been saved. 60 9 Smart App Memory 9.5 In the Memory menu, you can call up all the measurements stored with the testo 550s / testo 557s, analyze them in detail and also create and save csv data and PDF reports. When clicking on a measurement, an overview of the measurement results is displayed. 9.5.1 Searching for and deleting measurement results In the Memory menu, all stored measurements are sorted by date and time. The Memory menu is open. 1 Click on . Search field with measurements opens. 2 Enter the customer name or measuring site or date/time in the search field. The result is displayed. Deleting 1 Click on . A check box is displayed in front of each measurement. 2 Click on the required measurement. A tick is displayed in the respective box. 3 Click on . Information window is displayed. 4 Acknowledge the information. Selected measurements were deleted. 61 9 Smart App Sensors 9.6 All sensors used with the App can be found in the Sensors menu. There, you can view general information about currently connected probes as well as recently connected probes. 9.6.1 Information Information is stored for each probe. The App is connected to testo testo 550s / testo 557s. 1 Click on . Main menu opens. 2 Click on Sensors. The Sensors menu opens. 3 Click on one of the displayed probes. is displayed about the model, order number, serial number Information and firmware version. 62 9 Smart App 9.6.2 Settings Settings can also be made for each probe. The probe is connected to the App. 1 Click on . Main menu opens. 2 Click on Sensors. The Sensors menu opens. 3 Click on one of the displayed probes. 4 Click on the Settings tab. 5 Click on one of the displayed probes. Settings appear that can be changed if necessary. 9.7 Settings 9.7.1 Language 1 Click on Settings. The Settings menu opens. 2 Click on Language. A window with different languages opens. 3 Click on the required language. The required language is set. 63 9 Smart App 9.7.2 1 Measurement settings Click on Settings. The Settings menu opens. 2 Click on Measurement settings. A window with different basic settings for measurement opens. 3 Click on the required settings and change if necessary. The required measurement settings are set. 4 9.7.3 1 Exit Measurement settings. Company details Click on Settings. The Settings menu opens. 2 Click on Company details. A window with company details opens. 3 Click on the required data and enter if necessary. The required company details are set. 4 9.7.4 1 Exit Company details. Privacy settings Click on Settings. The Settings menu opens. 2 Click on Privacy settings. A window with privacy settings opens. 64 9 Smart App 3 Activate or deactivate the required settings. The required settings are set. 4 Exit Privacy settings. Help and Information 9.8 Under Help and Information, you will find information about the testo 550i, and the tutorial can be called up and implemented. This also where legal information can be found. 9.8.1 1 Instrument information Click on Help and Information. The Help and Information menu opens. 2 Click on Instrument information. current App version, Google Analytics instance ID, refrigerant The version and update are displayed for the connected instrument. Automatic updates for instruments can be enabled or disabled. > Use the slider to activate or deactivate Update for connected instruments. 9.8.2 1 Tutorial Click on Help and Information. The Help and Information menu opens. 2 Click on Tutorial. tutorial shows you the most important steps prior to The commissioning. 65 9 Smart App 9.8.3 Exclusion of liability 1 Click on Help and Information. The Help and Information menu opens. 2 Click on Exclusion of liability. data protection information and licence usage information is The displayed. testo DataControl archiving software 9.9 The free testo DataControl measurement data management and analysis software enhances the functionality of the testo Smart App measuring instrument with lots of useful functions: • Manage and archive customer data and measuring site information • Read out, evaluate and archive measurement data • Presenting readings in graphic form • Create professional measurement reports from the existing measurement data • Conveniently add pictures and comments to measurement reports • Data import from and data export to the measuring instrument System requirements 9.9.1 Administrator rights are required for installation. 9.9.1.1 Operating system The software can be run on the following operating systems: • Windows® 7 • Windows® 8 • Windows® 10 66 9 Smart App 9.9.1.2 PC The computer must meet the requirements of the operating system in each case. The following requirements must also be met: • Interface USB 2 or higher • DualCore processor with minimum 1 GHz • Minimum 2 GB RAM • Minimum 5 GB available hard disk space • Screen with a resolution of at least 800 x 600 pixels 9.9.2 Procedure To transfer the data from the App to testo DataControl, both instruments must be in the same network. For example: A notebook with installed testo DataControl and a smartphone with installed testo Smart App are connected to the same WLAN. 1 Open testo Smart App on the smartphone or tablet. 2 Open the testo DataControl archiving software on the PC. 3 Click on Select instrument. An overview with available instruments opens. 67 9 Smart App 4 Select instrument. A safety notice is displayed. 5 Click on Transfer data to DataControl and delete from instrument. Data has been successfully transferred. 68 10 Maintenance 10 Maintenance 10.1 Calibration The testo 550s / testo 557s is supplied with a factory calibration certificate as standard. Recalibration once every 12 months is recommended in many applications. This can be carried out by Testo Industrial Services (TIS) or other certified service providers. Please contact Testo for further information. 10.2 Cleaning the instrument Do not use any aggressive cleaning agents or solvents! Mild household cleaning agents and soap suds may be used. > If the housing of the instrument is dirty, clean it with a damp cloth. 10.3 Keeping connections clean > Keep screw connections clean and free of grease and other deposits; clean with a damp cloth as required. 10.4 Removing oil residues > Carefully blow out oil residues in the valve block using compressed air. 10.5 Ensuring measuring accuracy Testo Customer Service will be happy to help you as required. > Check the instrument regularly for leaks. Keep to the permissible pressure range! > Calibrate the instrument regularly (recommendation: once a year). 69 11 Technical data 10.6 Changing batteries/rechargeable batteries The instrument is switched off. 1 Fold out the suspension hook, release the clip and remove the battery compartment lid. 2 Remove the spent (rechargeable) batteries and insert new ones (4 x AA / Mignon / LR6) into the battery compartment. Observe the polarity! 3 Attach and close the battery compartment lid (the clip must click into place). 4 Switch the instrument on. 11 Technical data Feature Value Measurement parameters Sensor Pressure: kPa/MPa/bar/psi Temperature: °C/°F/K Vacuum: hPa / mbar/ Torr / mTorr / inH2O / micron / inHg / Pa testo 550s testo 557s Measuring cycle Connections: 3 Connections: 4 Valves: 3 Valves: 4 Pressure: 2 x pressure sensor Temperature: 2 x NTC Vacuum: via external probe Up to 4 Smart Probes via Bluetooth® connection 0.5 s Interfaces 70 Pressure ports: 3 x 7/16" UNF, 1 x 5/8" UNF NTC measurement External vacuum probe 11 Technical data Feature Value Measuring ranges HP/LP pressure measuring range: -100 to 6000 kPa/-0.1 to 6 Mpa/-1 to 60 bar (rel)/-14.7 to 870 psi Temperature measuring range: -50 to +150 °C / -58 to 302 °F Temperature measuring range of testo 115i: -40 to +150 °C / -40 to 302 °F Vacuum measuring range: 0 to 20,000 microns 65 bar; 6500 kPa; 6.5 MPa; 940 psi Overload Resolution Accuracy (nominal temperature 22 °C/71.6 °F) Measurable media Ambient conditions Housing IP class Resolution pressure: 0.01 bar/0.1 psi/1 kPa/0.001 Mpa Resolution temperature: 0.1 °C / 0.1 °F / 0.1 K Vacuum resolution: 1 micron (from 0 to 1000 microns) 10 microns (from 1000 to 2000 microns) 100 microns (from 2000 to 5000 microns) 500 microns (from 5000 to 10,000 microns) 5000 microns (from 10,000 to 20,000 microns) Pressure: ±0.5% of full scale value (±1 digit) Temperature (-50 to 150 °C): ±0.5 °C (±1 digit), ±0.9 °F (±1 digit), testo 115i temperature: ±2.3 °F (-4 to 185 °F) / ±1.3 °C (-20 to +85 °C), Vacuum: ±(10 microns + 10% of m.v.) (100 to 1000 microns) Measurable media: all media that are stored in the testo 557. Not measurable: ammonia (R717) and other refrigerants which contain ammonia. Operating temperature: -20 to 50 °C / -4 to 122 °F -10 to 50 °C / 14 to 122 °F (vacuum) Storage temperature: -20 to 60 °C / -4 to 140 °F Humidity application range: 10 to 90 %RH Material: ABS/PA/TPE Dimensions: approx. 235 x 121 x 80 mm Weight: 930 g (without batteries) 54 71 11 Technical data Feature Value Power supply Current source: Rechargeable batteries / batteries 4 x 1.5 V type AA / Mignon / LR6 Battery life: > 250 h (display illumination off, Bluetooth off, vacuum probe not connected) > 100 h (display illumination on, Bluetooth on, vacuum probe connected) 30 min, if enabled Auto off Display Directives, standards and tests type: Illuminated LCD Response time: 0.5 s EU Directive: 2014/30/EU You can find the EU declaration of conformity on the Testo website, www.testo.com, under the productspecific downloads. Available refrigerants Feature Value No. of refrigerants Selectable refrigerants in the instrument ~ 90 R114 R12 R123 R1233zd R1234yf R1234ze R124 R125 R13 R134a R22 R23 R290 R32 R401A R401B R402A R402B R404A R407A R11 72 R407C R407F R407H R408A R409A R410A R414B R416A R420A R421A R421B R422B R422C R422D R424A R427A R434A R437A R438A R442A R227 R444B R448A R449A R450A R452A R452B R453a R454A R454B R454C R455A R458A R500 R502 R503 R507 R513A R600a R718 (H2O) R744 (CO2) R417A 12 Tips and assistance Feature Value FX80 I12A R1150 R1270 R13B1 R14 R142B R152a R161 R170 R236fa R245fa R401C R406A R407B R407D R41 R411A R412A R413A R417B R417C R422A R426A R508A R508B R600 RIS89 SP22 12 Tips and assistance 12.1 Questions and answers Question Possible causes/solution flashes Batteries are almost empty. > Change the batteries. The device switches itself off. Residual capacity of batteries too low. > Change the batteries. Below range lights up instead of the measurement parameter display The value is below the permissible measuring range. > Keep within the permissible measuring range. Above range lights up instead of the measurement parameter display The value is above the permissible measuring range. > Keep within the permissible measuring range. 12.2 Error Codes 12.2.1 Main screen Code Possible cause / solution E 11 E 12 E 13 E 14 E 15 Remove the batteries and reinsert them in the instrument. If the error persists, please contact our service department. 73 13 Support 12.2.2 Status view Code Possible cause / solution E 30 An old version is still running on the testo 550s / testo 557s. Update the instrument. If the error persists, please contact our service department. testo 550s / testo 557s is still using the old version refrigerants. If you wish to use the new refrigerants, please update the refrigerant list again. If the error persists, please contact our service department. Remove the batteries and reinsert them in the instrument. If the error persists, please contact our service department. E 31 E 32 12.3 Accessories and spare parts Description Order no. Clamp probe for temperature measurement on pipes (1.5 m) 0613 5505 Clamp probe for temperature measurement on pipes (5 m) 0613 5506 2 x clamp temperature probes kit (NTC) for digital manifolds 0613 5507 Pipe wrap probe with Velcro tape for pipe diameters of up to max. 75 mm, Tmax +75 °C, NTC 0613 4611 Water tight NTC surface probe 0613 1912 Precise, robust NTC air probe 0613 1712 Spare valve set 0554 5570 Magnetic strap 0564 1001 External vacuum probe 0564 2552 For a complete list of all accessories and spare parts, please refer to the product catalogues and brochures or visit our website www.testo.com 13 Support You can find up-to-date information on products, downloads and links to contact addresses for support queries on the Testo website at: www.testo.com. If you have any questions please contact your local dealer or the Testo Customer Service. You can find contact details on the back of this document or online at www.testo.com/service-contact. 74 Testo SE & Co. KGaA Celsiusstr. 2 79822 Titisee-Neustadt Germany Phone: +49 (0)7653 681-0 E-mail: [email protected] www.testo.com 0970 5572 en 01
advertisement
Key Features
- Digital Manifold
- Refrigeration & Heat Pump Service
- Pressure & Temperature Measurement
- Two Valve Positioners
- Bluetooth Connectivity
- Data Logging & Analysis
- Compatible with Most Refrigerants
- User-friendly Interface