AquaLabo TOXmini Operating Manual
AquaLabo TOXmini is a portable luminometer designed for measuring bio- and chemiluminescence reactions. It's ideal for research, diagnostics, and hygiene monitoring. The TOXmini stores up to 2000 measurements with date and time stamps, allowing you to analyze results and transfer them to a PC. You can also define custom measurement protocols for different applications.
advertisement
Assistant Bot
Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.
▼
Scroll to page 2
of
46
Operating Manual Version: V 1.1 Date : 2020 microLAN b.v. Biesbosweg 2 NL-5145 PZ Waalwijk www.microlan.nl www.toxcontrol.com Contents TOXmini Contents Symbol Key ........................................................................................... 4 Safety Instructions ................................................................................ 5 1. Overview .......................................................................................................... 7 2. Instrument Set-Up ........................................................................................... 12 2.1. Control Panel .......................................................................................... 13 2.2 Measuring Chamber ................................................................................ 14 2.3 Connections ........................................................................................... 15 2.4 Battery Compartment .............................................................................. 16 2.5 Power Supply .......................................................................................... 17 2.6 Software .................................................................................................. 17 3 Getting started ................................................................................................. 18 4 Software Set-up and Operation ....................................................................... 20 4.1 Set-up ...................................................................................................... 20 4.2 Operation ................................................................................................ 21 4.2.1 Display mode .................................................................................. 21 4.2.2 Input field ........................................................................................ 21 4.2.3 Selection Fields (Text) .................................................................... 22 4.2.4 Option Fields................................................................................... 23 4.2.5 The Keys ......................................................................................... 24 5 Setting the Parameter ...................................................................................... 25 5.1 Password ................................................................................................. 26 5.2 User Input (Measurement Parameters) ................................................... 27 5.3 Date/Time ............................................................................................... 30 5.4 Language................................................................................................. 31 5.5 The RS232 Port....................................................................................... 32 5.6 The LCD Display .................................................................................... 34 6 Measuring ....................................................................................................... 35 6.1 Possible Adjustments to the Start Display .............................................. 35 6.2 Measurement Procedure......................................................................... 36 7 Processing the Results .................................................................................... 38 7.1 Readouts on the Display ......................................................................... 39 7.2 Transferring Data to a PC ....................................................................... 40 7.4 Deleting Results ...................................................................................... 42 8 Maintenance .................................................................................................... 44 9 Trouble Shooting ............................................................................................ 45 10 Technical Data ............................................................................................ 46 Symbol Key TOXmini Symbol Key ! 4 This symbol indicates a dangerous situation. Failing to heed this warning poses the threat of injuries or equipment damage! + This character indicates important information. This character is used to indicate processing steps. This character marks enumeration. [] Messages on the TOXmini's display are written in bold type and set in brackets [ ], for example [Measurement Start]. <> Function keys for the TOXmini are written in bold type and placed in arrow keys <ENT> or <ON>. Safety Instructions TOXmini Safety Instructions Prescribed Usage Instructions for Use of Applications The TOXmini is a small portable luminometer. It is designed for bio- and chemiluminescence measurements using reagents that generate a relatively constant light output. Typical applications are for example: research, clinical diagnostics and hygiene monitoring. For Your Safety The TOXmini has been designed and manufactured in accordance with state-of-the-art technology and regulations governing technical safety. The manufacturer has taken all steps to guarantee safe operation. The user must ensure that the instrument is set up and installed so as not to compromise its safe usage. Furthermore it is the operator’s responsibility to adhere to any local laws and regulations applicable to the installation and use of this instrument. Instruments are factory tested and delivered in a condition safe for operation. This manual contains instructions and warnings that must be followed in order to enable safe operation of the instrument! To disconnect the instrument from power pull out the power supply DC- connector or pull out the wall power supply from main power. Use only a approved power supply with a main voltage of 100V to 240V and a secondary voltage of 6VDC / 2,1A. The negative pole is on the inside of the plug. The main supply voltage fluctuations must not exceed +10% of the nominal voltage (230VAC). The instrument is designed for indoor use only. ! 5 Safety Instructions TOXmini The following safety instructions are to be observed before starting the instrument and during its operation: The instrument may only be put into operation by authorized persons and used by properly-trained personnel. All users of the instrument must first read the instruction manual. Users should only carry out maintenance and repair procedures found in the manual. During these procedures, no equipment other than that mentioned in the manual should be used. All maintenance and repair work not covered in the instruction manual is to be carried out only by service technicians authorized by microLAN. Make sure that the power supply provided is suitable for your line voltage. Always keep the sample holder clean! Avoid electrostatic charges! microLAN accepts no liability (also in cases of third-party involvement) for damages caused through improper use of the instrument. For the safety of the user and to ensure good working condition, the proper steps for starting, testing and maintaining the instrument are to be followed. 6 1.Overview TOXmini 1. Overview The TOXmini is a small portable luminometer which is suitable for measuring all types of ”glow” luminescence, i.e., bio- and chemiluminescent reactions that emit a constant light output over a given period of time. The TOXmini is used in research (molecular biology, medicine, biochemistry and pharmacy), diagnostics and hygiene monitoring. Common applications are: Reporter gene assays ATP measurements, including hygiene monitoring to detect microbial contamination NAD(P)H- and enzyme measurements Luminescence immunoassays and DNA probe assays The TOXmini is designed for rapid standard measurements of single samples. Measurement results are given in RLUs (Relative Light Units). These results can be classified in 3 categories by user-definition of two threshold values. The advantages of the TOXmini in a single glance: High accuracy of measurement with easy readout Simple operation of standard applications Storage of different measurement protocols Automatic storing of the last 2000 measurements, comprising the date, time, site identification number and test number Display of stored results Transfer of (selected) results to a PC Deletion of measured results Operation using mains supply or rechargeable batteries 7 1.Overview TOXmini ... and operation is so simple The Display After the TOXmini is switched on with the <ON> key, the Start display appears in the Display mode. The desired line can be selected using the arrow keys , and the <ENT> key activates the selected field (see figure 1.1): If you wish to select the Options field (for example, [Measurement Start]), measuring starts; the Input field (Number), the Input mode is activated and the numerator can be modified; the Selection field (for example, [Assay]), the Selection mode is switched on and selections can be made from the available options. Selections or input are confirmed by pressing <ENT>, which causes the return to the display mode. Pressing <ESC> in the Display mode leads back to the menu in the preceding level, for example, from the Start menu to the Main menu (see figure 1.2). Measurement Start Assay: ATP Location: 1 Test: 1 Insert sample Figure 1.1: Start display 8 Options field Selection field Input field Information line 1.Overview TOXmini The Key Functions (also see Chapter 4.2) Switches on the instrument. Press “ON” again the display illumination can be switched on or off. Switches off the instrument. Activates the selected field: Menu field, Input or Selection field. Input and Selection mode: Confirms the entry or selection. Display Mode: Move one level higher in the menu hierarchy. Input and Selection mode: Return to Display Mode without transferring changes. Display mode: Select a line. Selection mode: Page through the selection list. Input mode: Modify number values. Input mode: Move the cursor. Selection mode: Select Yes/No. 9 1.Overview TOXmini Overview of the menus Figure 1.2: The menu structure The user is directly in de main menu after switching on the instrument. There the users has a choice between the <Measure>, <Results> and <Settings> options: 10 <Measure> This option is called up if measurements should be carried out. <Results> The data administration is accessed via this option (call up or delete measurement results, data transfer to the computer). <Settings> The submenu for the setting of the instrument and test parameters is accessed by calling up this option. 1.Overview TOXmini How do I proceed? Step Chapter Page 1. Getting started 3 18 2. Define measurement protocol 5.2 27 3. Select measurement protocol 6.1 35 4. Measure 6.2 36 Where in the manual is the explanation found? Function Chapter Page Start display 6.1 35 Start measuring 6.2 36 Site identification 6.1 35 Test identification (in process) 6.1 35 Selecting a measurement protocol 6.1 35 Changing a measurement protocol 5.2 27 Saving measured results 6.2 36 Displaying measured results 7.1 39 Transferring measured results 5.5, 7.2 32, 40 11 2.Instrument Set-Up TOXmini 2. Instrument Set-Up Function, quality and reliability were the values most important during development. On the front panel of the TOXmini you have access to the measuring chamber containing the sample holder and photomultiplier (see Chapter 2.2) and the control panel (see Chapter 2.1). The connector RS232 for data transfer and connector for the accompanying power supply are located on the rear panel of the TOXmini (see Chapter 2.3). The battery compartment is placed on the bottom . The housing is protected against beating rain according to IP54 and can be operated in temperatures up to 30° C. Figure 2.1: Instrument set-up 12 2.Instrument Set-Up TOXmini 2.1. Control Panel The control panel contains 6 function keys that operate the software and 2 keys that control the power switch, as well as the graphic display. The display's background illumination is automatically switched to the energy-saving mode when the key pad has not been used for several seconds. It will automatically be switched back on with the next use of the key pad. Contrast and brightness of the display can be software adjusted (see Chapter 5.6). Figure 2.2: Control panel 13 2.Instrument Set-Up 2.2 TOXmini Measuring Chamber The measuring chamber contains both the sample holder and the photomultiplier. It can be opened using the black button. Only in its closed state the measuring chamber is light-tight, which is a prerequisite for luminescence detection. Individual test tubes are inserted into the sample holder for measurement. Tube formats range from 12 x 47 mm to 15 x 75 mm. Use the accompanying adapter for optimal results when measuring the smaller sized tube formats (article no. 04TM32737 Adapter vials, 15mm x 42 mm, 10 pieces). Black button to open lid Sample holder Photomultiplier Figure 2.3: Profile of the measuring chamber containing the sample holder and photomultiplier 14 2.Instrument Set-Up TOXmini To open the sample holder: Press the black button. To close: The cover should be pressed down until it clicks into place. Only then is the chamber completely closed to light and measurement process enabled. The Photomultiplier (Detector) The photomultiplier (detector) is a single photon counter that operates in a wavelength range of 380 – 630 nm. The linear range covers 6 decades. The measured counts are directly displayed as RLUs. 2.3 Connections There are two external connections. Both are located on the rear panel of the TOXmini and can be protected with the sliding panel: RS232 port for transferring data to a personal computer (serial port). A standard zero-modem cable is needed. Female connector to connect the accompanying power supply. The negative pole is on the inside of the plug. RS 232 power Figure 2.4: Connections (rear panel of the instrument) 15 2.Instrument Set-Up 2.4 TOXmini Battery Compartment Remove test tubes from the measuring chamber before turning the instrument over! Danger of damage from leaking fluids! The battery compartment is located on the instrument's underside. The compartment is covered by a panel, which is fastened by a screw. The three batteries must be inserted up as shown in figure 2.5. A warning message is displayed 15 minutes before the batteries become completely discharged. This is indicated by the message: Battery empty! The instrument then shuts down. For battery loading connect the instrument to the mains supply (see chapter 2.3). Loading only takes place if the instrument is switched on. Loading time max. 4-5 hours. (see chapter 2.5) Figure 2.5: Instrument underside with battery compartment Use only the recommended rechargeable battery! Ni/MH Akku Ansmann 1.2 V/4500 mAh microLAN accepts no liability for damages caused through use of other batteries! 16 2.Instrument Set-Up TOXmini 2.5 Power Supply The TOXmini operates with an input voltage of 6 V. The instrument can be run by means of either the power supply or the rechargeable batteries. When the instrument is connected to the power supply and t it is switched on, the batteries will be charged. Simultaneously, the TOXmini can be operated. The built-in real-time clock is maintained by means of a lithium battery (durability – min. 5 years – lasts usually approx. 10 yrs). Battery type Operating time without mains supply Charging time while operating Ni/MH 4500mAh 6h 4 – 5 h. Frequent partial depletion and recharging of batteries, together with aging, lowers their capacity. In order to reduce this, batteries should be completely discharged and recharged at regular intervals. A warning message is displayed 15 minutes before the batteries become completely discharged. This is indicated by the message: Battery empty! The instrument then shuts down. Use only the recommended rechargeable batteries! 2.6 Software The user software is set up in a series of menus accessed and controlled by using 6 function keys. The software, parameters, and up to 2000 measurements are stored in the non-volatile memory. The software will recognize overload conditions, that could be caused by excessive light emissions from a sample and display the warning message: Overload. For software operation, see Chapter 4. 17 3.Getting started TOXmini 3 Getting started Unpacking o Unpack the instrument and power supply and check the instrument for damage. Immediately report any problems to the distributor or manufacturer. Connecting and Switching On o o o o Slide the protective panel on the rear of the instrument to the open position to access the connections (see figure 3.1). Connect the power supply to a suitable mains outlet and plug the in line socket to the power inlet. The internal batteries will start to be charged if the instrument is switched on. Switch on the instrument by pressing the <ON> key. A few seconds after being switched on, the display will first show the instrument model number together with the software version, followed by the Start display (see Figure 3.2). Measurements are started here. 18 RS232 Power Figure 3.1: TOXmini rear panel 3.Getting started TOXmini Switching on the instrument in battery mode: If the TOXmini is not connected to the power supply, simply press the <ON> key. Measurement Start MAesss sa uny:g S ta r t ATP cation: A T P1 ALsosay: OTe rt:st: 11 T es t: 1 Insert sample P ro b e ein fü h ren Figure 3.2: Start display 19 4. Software Set-up and Operation TOXmini 4 Software Set-up and Operation 4.1 Set-up The TOXmini's software is set up for easy measurement of standard applications. After switching on the instrument the measurement menu Measurement Start appears on the display. A measurement can be started immediately by pressing the <ENT> key. In order to define the measurement parameters or the data output, the user simply changes to the Main menu, where the other menus can be selected by pressing the <ESC> key: Assay: Location Test: Measurement menu Assay: Location Test: ATP 1 1 ENT ESC ENT LB 9509 Main menu Results V 1.0 Measurement Results Settings ENT Password Customer Settings Date/Time Language RS232 System Figure 4.1: Software set-up 20 RLU: Result: insert sample ENT ATP 1 1 Screen PC Delete 111 pass 4. Software Set-up and TOXmini Operation 4.2 Operation The instrument is operated by use of 8 function keys. The readout follows on a graphic display. Results are given in Relative Light Units (RLUs), which means an integrated value over the measurement period. This RLU value is assembled by the measurement of RLU values in time intervals of 5 msec. The result is classified as pass, warn or fail according to the thresholds values set previously. Additionally, the classification is indicated by 3 LEDs: green = pass yellow = warn red (blinking) = fail red (blinking) = The measurement has been interrupted due to the opening of the sample holder. red (blinking) = Overload 4.2.1 Display mode When the TOXmini is switched on, the readout is automatically in the Display mode. The cursor may be moved from one line to the next by using the arrow key. This causes either the entire line or a field to be marked. In the Start display the following fields are used for operation (see Figure 4.3): Input field: to input values, Selection field: to select text, and Options field: to select an option or a menu. 4.2.2 Input field When you mark an Input field (text or figure values) using the cursor and press the <ENT> key, the Input mode will be activated, enabling the displayed value to be changed by means of the 4 arrow keys. The Input mode is identified by a blinking cursor. The blinking cursor marks the points at which changes may be made. The cursor is moved to the next point by pressing the key. 21 4. Software Set-up and Operation TOXmini The key enables the value at the selected point to be raised or lowered using a scale of 0–9. Pressing the <ENT> key a second time confirms the value and exits the input mode. Example: Changing test number 111 to 121 Measurement Start Assay: ATP Location: 1 Test: 111 ENT insert sample Measurement Start Assay: ATP Location: 1 Test: 0111 Measurement Start Assay: ATP Location: 1 Test: 0111 insert sample insert sample Measurement Start Assay: ATP Location: 1 Test: 121 Measurement Start Assay: ATP Location: 1 Test: 0121 ENT insert sample insert sample Figure 4.2: Changing a value ENT o Activate the Input mode with . The last digit of the number will be marked by the blinking cursor. o Press the be marked. o Press the o key once. The second-to-the-last digit will key once. The value will increase by 1. ENT Press the key in order to confirm the entry and exit the Input mode. 4.2.3 Selection Fields (Text) When you mark a Selection field using the cursor and press the <ENT> key, the Selection mode will be activated, and you will be able to page through the selection list using the arrow key. The Selection mode can also be identified by a blinking cursor. Pressing the <ENT> key a second time confirms the entry. The program returns to the Display mode. 22 4. Software Set-up and TOXmini Operation 4.2.4 Option Fields A selected menu or a selected option can be identified by the extended cursor block covering the entire line. If a menu or option has been marked with the cursor, pressing the <ENT> key activates the selected menu. The program enters the selected menu or activates the option. Examples (see Figure 4.1): When you select the [Results] menu while in the Main menu and subsequently press the <ENT> key, the [Results] sub-menu and its options will be displayed. If while in the [Results] sub-menu you select the [Time/ Date] option, the entry fields for date and time input will be displayed. Measurement Start Assay: ATP Mo ec sa s utio n gn:S ta r t 1 L Ae ss st: ay : A1 T1 P T 01 O rt: T es t: Options field Selection field Input field 1 1 insert sample Information line P ro b e ein fü h ren Figure 4.3: Start display 23 4. Software Set-up and Operation 4.2.5 TOXmini The Keys Switches on the instrument. Press “ON” again the display illumination can be switched on or off. Switches off the instrument. Activates the selected field: Menu field, Input or Selection field. Input and Selection mode: Confirms the entry or selection. Display Mode: Move one level higher in the menu hierarchy. Input and Selection mode: Return to Display Mode without transferring changes. Display mode: Select a line. Selection mode: Page through the selection list. Input mode: Modify number values. Input mode: Move the cursor. Selection mode: Select Yes/No. 24 5.Setting the Parameter TOXmini 5 Setting the Parameter In the [Settings] menu, the userspecific instrument and measurement parameters are defined. These parameters should be set following the first start-up of the instrument. The procedure for doing so is described in the next pages. In the Main menu, select [Settings] and press the <ENT> key. This causes the options for the parameter to be displayed. Measurement Start Assay: ATP Location: 1 Test: 1 insert sample ESC LB 9509 V 1.0 Measurement Results Settings LB 9509 V 1.0 Measurement Results Settings ENT Settings Password Customer Settings Date/Time Language RS232 System Figure 5.1: Selecting the menu [Settings] 25 5.Setting the Parameter 5.1 TOXmini Password If you want only authorized persons to change parameter settings and/or delete results, it is possible to define a password (max. 4 characters). Upon delivery, the TOXmini's password has been defined as ”0000”. If this preset password is not changed, each password query can be bypassed by using the <ENT> key. Changing the Password o In the [Settings] sub-menu, choose the [Password] option. o Press the <ENT> key. Settings Password Customer Settings Date/Time Language RS232 System ENT o Enter the old password. o Confirm the entry with the <ENT> key. The cursor moves to the [New Password] line. The Input mode is automatically activated. Enter the new password. Confirm the entry with the <ENT> key. The new password will be accepted, and the program returns to the [Settings] submenu. Password o o Old Password: 0000 New Password: 0 Figure 5.2: Password entry The sequence described above can be interrupted at any time by pressing <ESC>. The old password is still valid, and the program returns to the [Settings] sub-menu. 26 5.Setting the Parameter TOXmini 5.2 User Input (Measurement Parameters) The measurement parameters are defined in the [User Input] sub-menu. A total of up to 6 different programs containing measurement parameters, respectively protocols can be saved for various measurement requirements. The desired program can then be selected with the [Assay] query in the Start display prior to measurement start. A protocol record contains all of the parameters that define a measurement: the measuring time and the threshold settings. Defining the Measurement Parameters o In the [Settings] sub-menu, select the [Customer Settings] option. Settings Password Customer Settings Date/Time Language RS232 System ENT o Press the <ENT> key. Password o Enter the password and press the <ENT> key. Password: o (In the event that you have kept the password ”0000”, it will suffice to press the <ENT> key.) o Select the program for which you wish to change the parameter (for example, [ATP]). o Press the <ENT> key. The next screen will be displayed with the [Error] parameter selected. ENT Customer Settings Programs ATP Protocol 2 Protocol 3 Protocol 4 Protocol 5 Protocol 6 ENT Assay Figure 5.3: User Input 000 0 : ATP Fail : 10000 Pass : 5000 MeasTime : 10 27 5.Setting the Parameter o o o o o o o o o o 28 TOXmini Press the <ENT> key in order to activate the Input mode and enter the RLU value for the [limit fail] which will serve as the upper/lower threshold value to classify results as fail. Confirm the entry with the <ENT> key. Move the cursor to the [O.K.] parameter and press the <ENT> key. Enter the RLU value for the [limit pass] which will serve as the lower/upper threshold of the range that will be used to classify results as pass (Acceptance range). Be aware that results that fall between the two thresholds will be classified as warn (= Warning range). Confirm the entry with the <ENT> key. Move the cursor to the [Measuring Time] parameter and press the <ENT> key. Enter the desired measuring time in seconds and confirm the entry with <ENT>. When all individual parameters have been defined and confirmed with the <ENT> key, press the <ESC> key in order to return to the [Settings] sub-menu. If you wish, you can select a different program (2-6) and define its measurement parameters. Proceed as described for [ATP]. 5.Setting the Parameter TOXmini In accordance with the defined threshold values, the results will be displayed and classified: If [limit pass] is smaller than [limit fail], RLU values, which are smaller than [limit pass], are classified as pass. If [limit pass] is higher than [limit fail], RLU values, which are higher than [limit pass], are classified as pass. [limit pass] < [limit fail]: Measurement Class Results (RLU) Meaning LED RLU limit pass Acceptance range green limit pass < RLU < warn limit fail Warning range yellow RLU limit fail Error range red blinking [limit pass] > [limit fail]: Measurement Class Results (RLU) Meaning LED RLU limit pass Acceptance range green limit pass > RLU > warn limit fail Warning range yellow RLU limit fail Error range red blinking pass fail pass fail 29 5.Setting the Parameter 5.3 TOXmini Date/Time The date and time for the internal real-time clock can be entered or changed. All results will be saved with corresponding time and date entries. These also serve as selection criteria for the readout on the display, the transfer of information to a PC, and the deletion of results. In addition, the format for the date and time can be changed. Procedure o In the [Settings] sub-menu, select the [Date/ Time] option. o Press the <ENT> key. o o o o o o 30 Settings Password Einstellungen Customer Settings Passwort Date/Time Benutzereingaben Language Datum/Uhrzeit RS232 Sprache System RS232 LCD-Anzeige ENT ENT Mark the [Date] option, press the Date/Time set: <ENT> key, and enter the Eingabe Datum/Zeit: desired date as corresponds to Date : 12.01.1998 the preset format. Confirm the Time : :10:43 am Datum 12.01.1998 Format :: 10:43 am entry with the <ENT> key. Zeit Date : :Europe Format Mark the [Time] option, press the Time : Europa : 12h am/pm Datum <ENT> key, and enter the Zeit : 12h am/pm desired time as corresponds to Figure 5.4: the preset format. Confirm the Date and time entry entry with the <ENT> key. To change the Date and/or Time format, you should proceed as follows: Select option [Format]/[Date] and press the <ENT> key. Select the date format [USA] (= 04/24/1998) or [Europe] (= 24.04.1998) using the arrow keys. The current date entry will automatically change to the selected format. Confirm the selection with the <ENT> key. Select the [Format]/[Time] option and press the <ENT> key. Using the arrow keys, select the time format [12h am/pm] or [24h]. The current date entry will automatically change to the selected format. Confirm the selection with the <ENT> key. With the <ESC> key, return to the [Settings] sub-menu. 5.Setting the Parameter TOXmini 5.4 Language The TOXmini's operation language is selectable. Procedure o In the [Settings] sub-menu, select the [Language] option. o Press the <ENT> key. o Select the desired language and press the <ENT> key in order to confirm the selection. This causes the program to return to the [Settings] submenu. Settings Password Customer Settings Date/Time Language RS232 System ENT Language: Englisch German Figure 5.5: Selecting the language 31 5.Setting the Parameter 5.5 TOXmini The RS232 Port The RS232 port on the rear of the TOXmini enables the transfer of saved data (up to max. 2000 measurements) to a PC. This requires a standard zero-modem cable. The transmission parameters necessary for your computer's configuration are to be set in the [RS232] menu. In general it is possible to use the pre-set parameters. If the terminal program’s settings do not match the TOXmini's transmission parameters, adapt the program’s settings to the TOXmini’s requirements. Otherwise no data transfer is possible. Changing the Transmission Parameters o In the [Settings] sub-menu, select the [RS232] option. o Press the <ENT> key. o The transmission parameters for the RS232 port will be displayed. Modify these to correspond with those of your PC. Activate the desired parameter with the <ENT> key, select the corresponding value, and confirm the selection with the <ENT> key. Settings Password Customer Settings Date/Time Language RS232 System ENT [Baud Rate] Selection: [300], [600], [1200], [2400], [4800], [9600], [19200] and [38400] Baud. [Data Bits] Selection: [7] and [8]. [Parity] Selection: [None], [Odd], [Even]. [Stop Bits] Selection: [1] or [2] [Flowcontrol] 32 RS232: Baudrate Databits Parity : 19200 : 7 : None Flowcontrol: XON/XOFF Figure 5.6:Defining the transmission parameters 5.Setting the Parameter TOXmini Selection: o [XON/XOFF] = [Rts/Cts] = [None] = Pressing the <ESC> key, returns you to the [Settings] sub-menu. In order to transfer the selected data to your PC, first select the [Results] sub-menu in the Main menu and then the [PC] option. From this menu it is possible to select the data to be transferred according to date, time, and site (see Chapter 7.2). Please note: Your PC must have a terminal program in order to be able to process the transferred data. 33 5.Setting the Parameter 5.6 TOXmini The LCD Display Using this sub-menu, it is possible to modify brightness and contrast on the LCD display. Procedure o In the [Settings] sub-menu, select the [System] option. o Press the <ENT> key. o Select the [Brightness] System: parameter and press the <ENT> key. Using the arrow Brightness : 3 keys, the brightness Contrast : 7 may be set between levels 1– 4. The display illumination is changed simultaneously to Figure 5.7: this entry. Setting display brightness Once you have adjusted the and contrast brightness to the desired level, press the <ENT> key again. Select the [Contrast] parameter and press the <ENT> key. Using the arrow keys, the contrast level may be adjusted between levels 1–10. Once you have adjusted the contrast as desired, press the <ENT> key again. By pressing the <ESC> key, you return to the [Settings] sub-menu. o o o o 34 Settings Password Customer Settings Date/Time Language RS232 System ENT 6. Measuring TOXmini 6 Measuring To start a measurement in the Measure menu choose the [Measurement Start] option and press <ENT>. Parameters for the measurement can be selected and/or set in the Measure menu. This is achieved by moving the cursor to the appropriate line using the arrow keys and activating the Input and/or the Selection mode with the <ENT> key. Changes are made by using the 4 arrow keys. 6.1 Possible Adjustments to the Start Display [Assay] The desired measurement protocols can be selected from 6 variations. These are prepared in the [Settings]/ [User Input] menu. [Location] An identification number for the location (site) can be entered. [Test] An identification number for the measurement can be entered. This number will be automatically raised by 1 following each measurement, but can be changed as the user desires. Measurement Start Assay: ATP Location: 1 Test: 1 insert sample Measurement Start Assay: Protocol 1 Location: 1 Test: 1 insert sample : Measurement Start Protocol 6 Assay: Location: 1 Test: 1 insert sample Figure 6.1: Selecting the desired measurement protocol 35 6. Measuring 6.2 o o o o o o o 36 TOXmini Measurement Procedure Prepare the sample according Measurement Start Assay: ATP to the reagent manufacturer’s Location: 1 instructions. Test: 1 Select the desired insert sample measurement protocol and if necessary, enter a location, as ENT well as a test identification number. Assay: ATP Open the cover of the Location: 1 Test: 1 measuring chamber and insert Time: 6,5,4... the test tube into the sample holder. Close the measuring chamber. ENT The cap must click into place. The measuring chamber will Assay: ATP then be light tight and the Location: 1 Test: 1 measurement can be started. RLU: 111 At the [Measurement Start] Result: pass menu option, press the <ENT> key. The measurement will starts. The remaining measuring time Figure 6.2: in seconds is displayed while Measurement procedure the measurement is being taken. On completion of the measurement the RLU value is displayed. The measurement will also be classified according to the threshold value settings (pass = O.K., fail = Error range or warn = Warning range) and indicated by the illumination of the appropriate LED. 6. Measuring TOXmini Automatic Storing of Results The results are saved sequentially. If the memory is full (after the 2000th measurement), the oldest measurement data will be lost. Stopping a Measurement o A measurement in progress can be stopped by pressing the <ESC> key. A safety query appears asking if you are certain that you would like to abort the measurement. o Using the arrow keys , select [YES] or [NO] and confirm your answer using the <ENT> key. The measurement will be aborted following selection of the [YES] key, and the program returns to the Start display. Following selection of the [NO] key, the measurement continues. Returning to the Start Display and Restarting a Measurement o To return to the Start display, press the <ENT> key following the completion of the measurement. This causes a return to the Start display and automatically raises the test number by 1. You may immediately start taking a new measurement. 37 7. Processing the Results TOXmini 7 Processing the Results There are several options available in the [Results] menu for processing the stored results: The results, selected using Time and Site, may be transferred to a PC. The results, selected using Time and Site, may be shown on the LCD display. Selected results can be deleted. Choosing the [Results] menu o o When the readout shows the Start display, press the <ESC> key. Measurement Start Assay: ATP Location: 1 Test: 1 insert sample The Main menu appears. ESC LB 9509 o Select the [Results] menu. o Press the <ENT> key. o The options for processing the results will be displayed. V 1.0 Measurement Results Settings ENT Results Screen PC delete Figure 7.1: [Results] menu 38 7. Processing the Results TOXmini 7.1 Readouts on the Display The results can be selected using the time of measurement and the site identification number and subsequently shown one after the other on the LCD display. Procedure o o o o o o Select the [Screen] option. Press the <ENT> key. The [Display by starting at] display is shown with the selection criteria. Automatically the actual date, time and the site “0” are presented. If nothing is changed, the oldest result is shown. Enter [Date] and [Time] as the starting points. This causes all measurements taken between the entered point in time and the current date to be selected for readout. As a further criterion for selection, enter the [Site] identification number for the measurements that should be displayed. If you would like to display the measurements independently from the site identification, enter ”0”. Select [Start Search] with the cursor and press the <ENT> key. Results Screen PC delete ENT Display by Starting at: Date :12/28/1997 Time : 06:20 pm Location : 1 Start search ENT Assay: ATP Date: 12/28/1997 Time: 06:20 pm Location: 1 RLU: Result: 111 pass Figure 7.2: Displaying results o This causes the first (= earliest in terms of time) result corresponding to the selection criteria to be shown. o o Page further with the key. Pressing the <ESC> key returns you to the [Results] menu. 39 7. Processing the Results 7.2 TOXmini Transferring Data to a PC The measured results can be selected using the time of measurement and the site identification number and subsequently transferred to a PC. Procedure: o o o o o o o o o 40 Connect the RS232 port on the Results rear side of the instrument with the Screen serial port of your PC and start PC your terminal program (see delete Chap. 2.3). On the TOXmini’s display, select the [PC] option. ENT Press the <ENT> key. Display by The [Display by Starting Point at] Starting at: display is shown with the selection criteria. Automatically the actual Date : 12/28/1997 Time : 06:20 pm date, time and the site “0” are Location : 1 presented. If nothing is changed, start search the oldest result is transferred. Enter [Date] and [Time] as the ENT starting point. This causes all measurements taken between the Transfer of data entered point in time and the current date to be selected for transfer. As a further criterion for selection, enter the [Site] identification number for the measurements that should be transferred. If you would Figure 7.3: like to display the measurements Transferring results independently from the site to a PC identification, enter ”0”. Select [Start Search] with the cursor and press the <ENT> key. This causes all results corresponding to the time and site criteria to be transferred to the PC. The message [Transfer of data] will appear on the display. Pressing the <ESC> key returns you to the [Results] sub-menu. 7. Processing the Results TOXmini Format of Data Transfer a) Data transfer to PC Data is transferred in the following order: Date Time Program Limit pass Limit fail Measuring time Location identification number Test identification number Result (= integrated RLUs) or ”Overload” <CR> (= carriage return) <LF> (= line feed) If an overload occurs during a measurement, “Overload” is transferred instead of a result (RLU value). Each value is separated by using <TAB>. Example: 03/23/98<TAB>08:32:58<TAB>ATP<TAB>10000<TAB>5000 <TAB>10<TAB>1<TAB>3<TAB>1556<TAB><CR><LF> b) On-line data transfer to PC Here, data are transferred in the same format and sequence as in a) however, additionally 100 single RLU values are transferred, which are positioned in front of the integrated RLU value. Also the single RLU values are separated by using <TAB>. If the measurement is aborted, the values are transferred up until stopping the measurement and ”CANCEL” will be transferred. 41 7. Processing the Results 7.4 TOXmini Deleting Results The results to be deleted are selected using the time of measurement as well as the site identification number and are shown before deletion on the display. Results Screen PC delete ENT Procedure o Select the [Delete] option. o Press the <ENT> key and enter the password. Confirm the entry with the <ENT> key. o The [Display by starting at] display is shown with the selection criteria. Automatically the actual date, time and the site “0” are presented. If nothing is changed, the oldest result is deleted. o Enter [Date] and [Time] as the starting points. This causes all measurements taken between the entered point in time and the current date to be selected for deletion. o As a further criterion for selection, enter the [Site] identification number for the measurements that should be displayed. If you would like to delete the measurements independently from the site identification, enter ”0” o Select [Start Search] with the cursor and press the <ENT> key. Password Password: 0000 ENT Display by Starting at: ESC Date : 12/28/1997 Time : 06:20 pm Location : 1 start search ENT Assay: ATP Date: 12/28/1997 Time: 06:20 pm Location: 1 RLU: 111 Result: pass ENT ESC Assay: ATP Date: 12/28/1997 Time: 06:20 pm Location: 1 RLU: 111 Result: pass Delete result ? ENT ? Delete Result? Yes No Figure 7.4: Deleting results 42 7. Processing the Results TOXmini o This causes the first measured result corresponding to the entered point in time and the defined site identification to be displayed. o Page further with the key until you reach the readout of the result that you would like to display. o Press the <ENT> key to delete the displayed measurement. The safety query [Delete measurement?] will appear. Pressing the <ENT> key again will cause the displayed measurement to be deleted and the next measurement to be displayed. The deletion is aborted with the <ESC> key, and you can continue paging through the list. In this way, you can delete measurements at your discretion. Pressing the <ESC> key returns you to the [Results] sub-menu. o o o o 43 8. Maintenance TOXmini 8 Maintenance The TOXmini is extremely maintenance-friendly and requires almost no (or very little) maintenance . Please observe the following: 1. Sample Holder The sample holder must always be kept clean and free from dust, in order not to spoil measuring results! Fluids in the measuring chamber could damage the instrument! In the event that fluid enters the test chamber, immediately remove the test tube and turn the instrument upside down to remove the fluid before it enter the measuring chamber! Clean the sample holder. If fluid enters the measuring chamber, inform service personnel immediately! 2. Caring for the Rechargeable Batteries To prevent the capacities of the batteries from diminishing, they should be completely discharged on a regular basis (every 2–3 months) and then recharged. Discharging duration: ca. 6 h. Recharging duration: ca. 4–5 h. In order to discharge the batteries, detach the TOXmini from the power supply and leave it on for a given period of time, for example, overnight. To recharge the batteries, reattach the instrument to the power supply and switch it on. 44 9. Trouble Shooting TOXmini 9 Trouble Shooting Error Cause Solution Display - 10 HV part could be wet Turn off instrument and let it dry. Switch on again. If error message appears again, contact service. other error messages contact service 45 10. Technical Data TOXmini 10 Technical Data Detector Sensitivity Dynamic range Tube format Language Display Interface Software Measurement time Data storage PC-software Power supply Installation category Current consumption Batteries Humidity Altitude Temperature Recommended operation temperature Pollution degree Dimensions (WxDxH) Weight 46 Ultra fast single photon counter with spectral sensitivity range of 380-630 nm 1 fmol ATP using the ATP Bioluminescence Assay Kit CLSII (Boehringer Mannheim, FRG) more than 6 decades 12 mm diameter, 47-75 mm height; 15 mm diameter, 47-75 mm height English or German (software selectable) Graphic display with backlit LED, 128 x 64 dots RS232 C interface for data transfer to a PC built-in microprocessor, storage of 6 different measuring protocols selectable in steps of 1 s between 1 and 999 s storage of the last 2000 results data capture software program WinTerm for easy data transfer and evaluation Prim.: 230V / 50Hz; 115V / 60Hz Sec.: 6V DC, 2,1A (Negative pole is on the inside of the plug!) II max. 15 VA 3 rechargeable Ni/MH Akku Baby 1.2 V / 4500 mAh 80% at 31°C degreasing linearly to 50% at 40°C no condensation <2000m 5°C to 40°C +15°C to +30°C 2 150 mm x 280 mm x 170 mm approx. 2 kg net, incl. batteries 10. Technical Data TOXmini RS232 interface: pin configuration Pin1 Pin2 Pin3 Pin4 Pin5 Pin6 Pin7 Pin8 Pin9 I O O I O nc RxD TxD nc. GND nc. RTS CTS nc. Ordering Information TOXmini portable luminometer Disposable Glass Cuvettes 200 Adapter vials, 15x42 mm,10 pack Rechargeable batteries, 3 pack PC connecting cable Sturdy Metal Transport case PC Terminal software WinTerm Test Light source Ordering number 01TM32526-10 04AZF686004-200 01TM32737 06TM40650 07TM20246 01TM32700 07TM29890 01TM11070 47
advertisement
Key Features
- Portable design
- High accuracy
- Simple operation
- Stores 2000 measurements
- Data transfer to PC
- Custom measurement protocols
Frequently Answers and Questions
How do I change the measurement parameters?
You can change the parameters in the 'Customer Settings' menu within the 'Settings' sub-menu. You'll need to enter the password first.
What are the classification categories for results?
Results are classified as 'pass', 'warn', or 'fail' based on the thresholds you define. The classification is indicated by LEDs: green for pass, yellow for warn, and red (blinking) for fail.
How do I transfer data to a PC?
Use the RS232 port on the back of the TOXmini and a standard zero-modem cable to connect to your computer.