Scinomix TS-1078 Quick Start Manual
Scinomix TS-1078 is a replacement part for the swing arm on Sci-Print MP2 or MP2+ instruments. This manual provides step-by-step instructions for replacing the swing arm, ensuring proper installation and functionality. The procedure involves removing the old arm, installing the new arm at a 90-degree angle, and securing it with screws. It also includes instructions for reconnecting airlines and testing the instrument.
advertisement
▼
Scroll to page 2
of
5
advertisement
Key Features
- Swing arm replacement
- Sci-Print MP2/MP2+
- Step-by-step instructions
- Visual aids
- Troubleshooting guide
Frequently Answers and Questions
What tools are required to replace the swing arm?
You will need a 2mm Allen wrench, a 3mm Allen wrench, and a 2.5mm Allen wrench.
How do I remove the old swing arm?
Remove the four screws for the swing arm tower housing, unplug the flex cable, remove the flex cable/airline bracket, and then remove the top and middle shoulder screws. Finally, remove the broken arm.
What is the correct angle for installing the new swing arm?
Install the new swing arm at a 90-degree angle so the small dowel pin fits into the swing slot.