Gruetzner LUB-M Operating Instructions Manual

Gruetzner LUB-M Operating Instructions Manual

Gruetzner LUB-M is a compact lubrication system designed for industrial applications. It handles both oil and grease, up to NLGI class 2, and features a 1500 cm³ lubricant reservoir in a disposable cartridge. This system comes with up to 5 pump units, each offering 2 outlets for precise dispensing. The LUB-M can be configured for either time-controlled or pulse-controlled lubrication intervals, catering to a range of industrial lubrication needs.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Gruetzner LUB-M Lubrication System Operating Instructions Manual | Manualzz
Operating Instructions
Lubricus Lubrication System
LUB - M
(24 VDC)
Operating Instructions: LUB‐ M
Table of Contents
1) General Information
1.1 Warning
1.2 Scope of supply
1.3 Overview – the Lubricus M Lubrication System
1.4 Symbols and notes
1.5 Intended use
1.6 Liability and extent of guarantee
4
4
4
4
5
5
6
2) Safety Information
2.1 General safety information
2.2 Transport and storage
2.3 Installation
2.4 Maintenance
7
7
8
8
8
3) Lubricus M ‐ General information
3.1 Definition of terms
3.2 Display & Actuation Pad
3.3 Identification pump units and outlets
3.4 Grease cartridge Lubricus M
9
9
9
10
10
4) Function/Principle
4.1 Basic operation
4.2 Lubricus M operation menu (overview)
4.3 Start: control panel
4.4 Default indication control panel (time control)
4.5 Quick Check – back pressure control
11
11
12
13
13
14
5) Pro Program – time controlled lubrication intervals
5.1 Enter PIN
5.2 Input pause time and amount of lubricant
5.3 ESC – Exit the menu
5.4 CLr – delete error messages
5.5 FIL – first start‐up, venting the pump
5.6 Change PIN
5.7 Change Feedback (Pump function control)
5.8 Change operating mode: Time control / pulse control
15
15
16
18
18
19
20
20
21
6) Pro Program – pulse controlled lubrication intervals
6.1 Default indication control panel (pulse control)
6.2 Enter PIN
6.3 CLr – delete error messages
6.4 FIL – first start‐up, venting the pump
6.5 Change PIN
6.6 Change Feedback (Pump function control)
6.7 Change operating mode: Time control / pulse control
6.8 ESC – Exit the menu
22
22
23
23
24
25
25
26
26
7) Pulse signals to activate the outlets
7.1 Identification pump units and outlets
7.2 Pulse signals pump unit 1
7.3 Pulse signals pump unit 2
7.4 Pulse signals pump unit 3
7.5 Pulse signals pump unit 4
7.6 Pulse signals pump unit 5
7.7 pulse control ‐ acceptable Design
27
27
28
28
29
29
30
31
Operating Instructions: LUB‐M
Pulse control
2
Time control
Contents
2
8) Communication interface (M12x1 connector) Lubricus M
32
9) Errors, Messages
9.1 Error message E1: Level indicator
9.2 Error message E2: Red locking pin is missing
9.3 Error message E3: Pump unit works too slowly
9.4 Error message E4: Internal electrical fault
9.5 Error message E7: Back pressure too high/over‐current
33
33
33
34
34
35
10) Master PIN
35
11) Service Lubricus M
11.1 Readout of the device memory
11.2 Replacement of grease cartridge
11.3 Filling with oil
36
36
37
40
12) Mounting Lubricus M
41
13) Technical data
41
14) Disposal
41
Operating Instructions: LUB‐M
3
1) General information
Thank you very much for choosing the Lubricus M Lubrication System.
Please ensure you familiarise yourself with the safety instructions before using the pump and accessories supplied.
This manual contains important safety information.





The Lubricus Lubrication System is a very compact lubrication pump for oil and grease up to NLGI class 2.
The Lubricus type LUB-M is made for a 24 VDC power supply and requires a mains connection.
The lubricant reservoir (1500 cm³) is in a rigid, disposable cartridge.
Supply pressure is <70 bar.
Depending on the model, the Lubricus Lubrication System has 2, 4, 6, 8 or 10 outlets.
1.1 Warning
The Lubricus Lubrication System is designed for use in normal industrial environments, but not for use in or on motor vehicles.
Use only original Lubricus spare parts and accessories (cartridges, fittings, connectors, tubes and battery packs).
1.2 Scope of supply
The scope of supply may vary from country to country.
Standard supply includes:



the Lubricus M Lubrication System
tube connector attached to the outlet(s) for medium-pressure PA tube 6 x 4 (Ø 6mm outer and Ø 4mm inner diameter)
Operating Instructions
1.3 Overview – the Lubricus M Lubrication System
Installing the Lubricus Lubrication System is simple. These instructions should help you to install your pump swiftly and easily, and to
quickly learn the basics.
Actuation magnet
Housing/covering
Actuation pad
Outlets
Communication interface
(Output M12x1)
Please note!
Unused outlets should not be closed!
Operating Instructions: LUB‐M
4
1.4 Symbols & Notes
Please pay attention, not only to these safety instructions, but also to the special safety instructions on other pages.
This symbol warns of electrical voltage.
This general danger symbol denotes safety instructions, the disregarding of which could lead to bodily harm.
This heading is used if incorrect compliance or non-compliance with the Operating Instructions or specified work procedures
etc. may result in damage of the system.
Attention
Please note!
This term will be used to draw your attention to special information.
Instructions attached directly to the machine must be strictly followed and kept in clearly legible condition!
1.5 Intended Use
Attention
The Lubricus Lubrication System is only approved for industrial use. The Lubricus Lubrication System may only be used in
accordance with the Technical Data (see chapter “Technical Data”).
Unauthorised structural changes to the Lubricus Lubrication System are not permitted.
We do not assume liability for damages to persons or machines resulting from such.
Additional instructions for use in conformity with the intended use:
 pay attention to all information in the Operating Instructions;
 carry out all maintenance work;
 use only Gruetzner GmbH lubrication tubes and connectors;
 follow all appropriate regulations for work safety and accident prevention during the entire lifespan of the Lubricus Lubrication
System;
 only qualified and authorized personnel should carry out work of any kind on the Lubricus Lubrication System.
Attention
Other uses or uses beyond those described above shall be deemed as not in conformity with the intended use.
Operating Instructions: LUB‐M
5
1.6 Liability and extent of guarantee
Proper functioning of the Lubricus Lubrication System can only be achieved through use of recommended lubricants from the original
Lubricus range of accessories, and by observation of all installation, operation and maintenance instructions.
Gruetzner GmbH excludes all liability if these instructions are not observed.
Gruetzner GmbH grants guarantees concerning operating safety, reliability and performance only under the following conditions:
 assembly, installation, maintenance and repairs are carried out by qualified staff only;
 hazardous hot or cold machine parts must be shielded to prevent skin contact;
 the Lubricus Lubrication System must be used according to the Technical Data and instructions;
 the limit values stated in the Technical Data are under no circumstances to be exceeded;
 retrofitting and repair work on the Lubricus Lubrication System may only be carried out by the manufacturer;
 the lubrication system must be protected against damp and wet.
If a lubricant that has not yet been tested for compatibility with the Lubricus Lubrication System be supplied or requested for use in the
Lubricus Lubrication System by the purchaser/customer, the purchaser/customer undertakes the risk and responsibility.
This product is subject to strict production controls and fulfils our company’s factory specifications. However, guarantee of testing of each
individual case cannot be given due to the numerous factors involved.
We therefore strongly recommend that test runs be carried out.
All liability is excluded.
The Lubricus Lubrication System lubricant cartridge has been constructed and produced for one-time use only. Multiple use of the
lubricant cartridge is not envisaged. After being used once, the lubricant cartridge is to be completely replaced; refilling is not permitted.
Noncompliance will lead to nullification of the guarantee.
For guarantee conditions see the sales and delivery conditions of Gruetzner GmbH
Operating Instructions: LUB‐M
6
2) Safety Information
Basic information, which must be followed during assembly, operation and maintenance, is listed as follows. These Operating Instructions
must be read by all installation and operation staff before assembly and operation, and must be permanently available on site.
All personnel involved in installation, maintenance and operation of the system are to read these instructions carefully before beginning
assembly and operation.
2.1 General safety information
2.1.1 Qualification and training of personnel
All operation, maintenance, service and installation personnel must be appropriately qualified for this work. Responsibility
and supervision of the personnel must be clearly defined by the end user. If staff do not have the necessary knowledge,
they must be trained and instructed. The end user must ensure that staff have completely understood the contents of the
Operating Instructions.
2.1.2 Danger of non-compliance with the safety instructions
Non-compliance with the safety information can be dangerous for people, environment and machines. Noncompliance with the safety information can mean the loss of any or all damage claims. In certain cases, non-compliance
can, for example, lead to the following hazards:
 malfunction of important system functions.
 malfunction of prescribed methods of maintenance and servicing.
 danger to people due to electrical, mechanical and chemical effects.
 danger to the environment due to leakage of dangerous substances.
2.1.3 Safety information for end user / operating staff
 Moving, rotating, hot or cold machine parts are to be securely shielded on site to prevent accidental contact. The
shielding on moving or rotating parts is not to be removed.
 Leakages of dangerous materials are to be removed in such a way as not to endanger people or the environment.
 Legal regulations are to be complied with.
 Electrical hazards are to be eliminated.
2.1.4. Safety information for maintenance, inspection and assembly work
All maintenance, inspection and installation work may only be carried out by trained specialists who have been
informed appropriately by carefully studying the Operating Instructions.
All work must only be carried out when the machine is witched off and while wearing appropriate protective clothing.
Always comply with the procedures for switching off as described in the Operating Instructions.
All safety and protective equipment must be replaced immediately after completing work.
Environmentally hazardous substances must be disposed of in accordance with local regulations.
Secure the system during maintenance and repair work against intentional or unintentional operation.
Dispose of lubricants in accordance with the safety data sheets of the lubricant manufacturer.
2.1.5 Unauthorised modification and manufacture of spare parts
Modification and alteration of the system is not permitted. Original spare parts and accessories authorised by the
manufacturer are for safety purposes. Using other parts can result in loss of liability for claims resulting out of this.
Gruetzner GmbH does not guarantee and is not liable for components retrofitted by the end user.
2.1.6 Prohibited methods of operation
Operational security of the Lubricus Lubrication System is only guaranteed if it is operated in accordance with the Operating
Instructions. The limit values listed in the Technical Data are under no circumstances to be exceeded.
Operating Instructions: LUB‐M
7
2.1.7 General hazard warning
All components of the system are designed in accordance with the prevailing regulations of the construction of technical
machines regarding operational safety and accident prevention. Nevertheless, operation can lead to dangers for the user,
third parties or other technical facilities. Therefore, the Lubricus Lubrication System may only be used for its intended
purpose in technically fault-free condition. This may only be carried out under compliance with the safety regulations
and the Operating Instructions.
Therefore, please regularly check the Lubricus Lubrication System and its attachments for possible damage or leaks.
2.2 Transport and storage
Use suitable lifting gear for transport. Do not throw or expose the Lubricus Lubrication System to heavy buffeting. Store the Lubricus
Lubrication System in a cool and dry place to avoid corrosion of the system’s individual parts.
Comply with valid safety and accident prevention instructions when transporting.
Wear suitable protection clothing if necessary!
2.3 Installation
The following conditions have to be satisfied during the installation of this Lubricus Lubrication System so that it can be
assembled to a complete machine without affecting the health and safety of humans.
To prevent the formation of condensation, the housing of the Lubricus Lubrication System should not
be exposed to direct sunlight or radiant heat.
The Lubricus Lubrication System must be protected against damp and wet. We strongly recommend
that the lubrication system be enclosed in a housing.
Lubricus Lubrication Systems should never be exposed to outside weather conditions. If using
outdoors, ensure the system has suitable protection (e.g. housing) and never use this product in or
under water.
2.3.1 Electrical connection
 Only a qualified electrician should connect the power supply!
 Connection of electrical components is to be carried out professionally.
 Compare the voltage details with the existing power supply voltage!
2.4 Maintenance




Disconnect the system from the mains before starting with maintenance or repair work.
Maintenance and repair work may only be done with the system completely switched off.
Check the surface temperature of the Lubricus Lubrication System to avoid burns.
Always wear heat-resistance gloves!
Ensure the system cannot be restarted during maintenance and repair work!
Operating Instructions: LUB‐M
8
3) Lubricus M – General Information
Lubricus M
3.1 Definition of terms
Lubricus M is designed for applications with minimal amounts of lubricant:
o Depending on the configuration Lubricus M is available with up to 5 pump units (P1, P2, P3, P4, P5), each with 2 outlets.
o 1 pump unit is the housing for 2 piston pumps. 1 pump unit is dispensing equal quantities from both outlets.
o Minimum quantity lubrication means the controlled and efficient supply of a lubrication point with selected, high performance
lubricants
o Run = back pressure control – preventive check on a lubrication point: The measurement of the back pressure is done during the
pumping process and the result will be shown on the display. The displayed value is a first orientation over the pressure range.
3.2 Actuation Pad
Pump unit P1 … P5:
Error messages (ALARM):
Operating Instructions: LUB‐M
display green LED
display red LED
9
3.3 Identification pump units and outlets
Each active pump body (P1, P2, P3, P4, P5) is
indicated by a green LED on the control panel.
Each active outlet is displayed during a function
with a number (1 or 2) on the control panel.
3.4 Grease cartridge
The lubricant reservoir (1500 cm³) is in a disposable grease cartridge.
Operating Instructions: LUB‐M
10
4) Function/Principle
After starting, the integrated pump transports the lubricant to the outlets according to the settings that have been made.
The integrated electronics control and monitor the amount of lubricant and the time intervals between lubrications.
The Lubricus M Type uses a four-pin socket to connect a M12X1 plug allowing communication with the machine controller. An external 24
VDC power supply is used to switch ON and OFF (voltage 24 VDC +/- 5%), supplied to PIN 1. When voltage is supplied the lubrication
pump is in operating condition. If there are no errors (Pump OK) the input voltage will be transferred to the output signal (PIN 4) indicating
all is OK. Once the voltage supply to PIN 1 is stopped (switched OFF), the pump rests and saves the current operating conditions in
memory. When the pump is turned back on the saved operating conditions will be continued. PIN 4 relays the operating condition.
4.1 Basic operation
All changes to the settings are made with the actuation magnet (ventilation lock). Open the ventilation lock (turn clockwise to OPEN) and
remove it from the cover. The tip of the ventilation lock is your programming key. If there are no desired changes to the settings, please set
the actuation magnet back at its place and close the ventilation lock (turn counter clockwise to CLOSE).
Programming Lubricus M
 SELECT: Change settings - touch the operating pad with the actuation magnet
 ENTER: Confirmation of all changes - touch the actuation pad actuation magnet. The display flashes 2 times for
confirmation
 TIMEOUT: If no interaction takes place for a period, the device automatically returns to the previous display mode.
Operating Instructions: LUB‐M
11
4.2 Lubricus M operation menu (overview)
Two modes are possible: „time control“ and „pulse control“
Operating Instructions: LUB‐M
12
4.3 Start: control panel.
Apply operating voltage
Softwareversion n01 … n99
Operating mode
Operating mode
Time control
Pulse Control
4.4 Default indication control panel (time control)
 SELECT: Change settings - touch the operating pad with the actuation magnet
 ENTER: Confirmation of all changes - touch the actuation pad actuation magnet.
 If no interaction takes place for a period, the device automatically returns to the previous display mode.
Basic menu in time mode:
On:
run:
Pro:
ESC:
Operating mode – time control
Run = back pressure control – preventive check on a lubrication point: The measurement of the back pressure is done during the
dispensing pumping process and the result will be shown on the display. The displayed value is a first orientation over the
pressure range.
Other functions, PIN code protected
Key to leave and return to the main menu
Operating Instructions: LUB‐M
13
4.5 Quick Check – back pressure control
For test and sample purposes the installed lubrication pump can easily be used to dispense small amounts of lubricant. For this purpose,
the available pressure between the lubrication point and pump is measured. The value displayed is the back pressure in bar.
 SELECT: Change settings
 ENTER: Confirmation of all changes
 If no interaction takes place for a period, the device automatically
returns to the previous display mode
 Selection of pump body by SELECT
 Display shows selected pump body (green LED)
 Confirmation by ENTER
 Result is shown on display
(multiple repetitions possible by ENTER)
Remark: outlets of one pump unit are dispensing
lubricant alternately
Select other pump unit or timeout:
leave the menu
Operating Instructions: LUB‐M
14
5) Pro Program – time controlled lubrication intervals
The „pro program“ of the time mode contains the following editable options:
Time control
Enter PIN
Change
operating
mode
(time/pulse)
Input delivery
rates and
pasue times
Change
feedback
ESC: exit the
menu
CLr: delete
error messages
Change PIN
FIL: Deaerate
the pump
5.1 Enter PIN: Access to Pro Program is protected (PIN)
Selection of other menu items with PIN entry possible.
 SELECT: Change settings
 ENTER: Confirmation of all changes
If no interaction takes place for a period,
the device automatically returns to the
previous display mode
The settings are performed with „ENTER“ and „SELECT“.
The factory setting is: 000 (Master PIN siehe Service)
After entering the the PIN the lubrication intervals can be programmed.
Operating Instructions: LUB‐M
15
5.2 Input pause time and amount of lubricant
The following settings can be made for each pump body.
Time control
H: TIME (pause time): time between the lubrication intervals in hours (1 to 240 h are possible)
Attention: If H: TIME = 0 it is not possible to input cycles.
C: CYCLE (number of cycles) = amount of lubricant per interval (1 to 30 possible, 1 cycle = 0,15 cm³)
Pump units can be turned off by setting C: CYCLE = 0
Factory setting for every pump unit: H: TIME = 4h, C: CYCLE = 1
step-by-step approach:
 SELECT: Change settings
 ENTER: Confirmation of all changes
If no interaction takes place for a period,
the device automatically returns to the
previous display mode
Select pump body (P1-P5) via SELECT
indicated by a green LED on the control panel.
 Confirmation with ENTER
Input pause time (H: TIME) via SELECT
max. 240 h possible
 Confirmation with ENTER
Input number of cycles (C: Cycle) via SELECT,
max. 30 possible
 Confirmation with ENTER
Select other pump unit or timeout: exit the menu.
Important: Changes of the parameters are only applied by confirming.
Operating Instructions: LUB‐M
16
Attention
Lubricus M is designed for minimum quantity lubrication.
Warning: The ALARM LED flashes on the control panel and the display shows "InF" (for information) for 10s.
During this 10s, the operation of Lubricus M is blocked.
.
Acceptable Design:
The specified lubricant quantity refers to the expected life of a piston pump or a pump unit. A service for the verification of the
performance of Lubricus M is recommended. Service interval is 65,000 cycles per pump unit. 65,000 cycles correspond to 10.000
cm³ lubricant. See service: Readout of the device memory.
If a higher demand for lubricant at one lubricating point is required, the outlets of different pump units can be combined
externally.
Please note!
unused outlets must remain open to avoid damage to the pump!
Examples:
1 outlet supplying 1 lubrication point:
Allowed amount of lubricant for 1 lubrication point =
max. 5,000 cm³ at 50 bar back pressure
2 outlets supplying 1 lubrication point:
Allowed amount of lubricant for 1 lubrication point =
max. 10,000 cm³ at 50 bar back pressure
4 outlets supplying 1 lubrication point:
Allowed amount of lubricant for 1 lubrication point =
max. 20,000 cm³ at 50 bar back pressure
Operating Instructions: LUB‐M
17
Time control
The microprocessor calculates from the inputs of pause time and number of cycles a comparison value.
If this value is too high, a reduction in the lifetime of Lubricus M is accepted consciously.
5.3 ESC – Exit the menu
Confirm „ESC“ with "ENTER" => exit to main menu.
Time control
5.4 CLr – delete error messages and delete filling cycles
With Confirmation of „CLr“ error messages will be deleted.
In „FIL“-Modus the filling operation will be stopped with „CLr“.
Operating Instructions: LUB‐M
18
5.5 FIL – first start-up, venting the pump und filling of lubrication lines
Attention: Lubricus M lubrication pumps must be filled at initial operation. Each pump unit must be filled separately.
The operation is completed when lubricant is visible at the outlet.
Attention!
Time control
First start-up:
Prefill housing inlet with 50 ml of suitable lubricant (same grease
type as content of cartridge)
Attention: Low level indicator in the inlet must not be damaged.
Then place grease cartridge and start FIL function.
 Select „FIL“
 Confirm with ENTER
 Select pump body
 ENTER
Delete with CLr (Clear)
FIL means that the outlet of each pump unit is 15 times active.
Total time of FIL / pump unit: about 5 minutes
This function can also be used to pre-fill tubes.
15 cycles with 0,15cm³ = about. 2 cm³
Delete with CLr (Clear)
Select other pump units with:
 Pro
 PIN
 FIL
or timeout => exit the menu
Operating Instructions: LUB‐M
19
5.6 Change PIN , factory setting PIN: 000 (Master PIN see Service)
In the menu point „PIN“ you can change the PIN for the access of the PRO-Menues.
Example: Change PIN to 222
Time control
The new PIN is saved.
5.7 Change Feedback (Pump function control)
Feedback = Pump function control: During the function of the pump (about 7 s) the output signal at PIN 4 is switching from High (+
20...30 VDC) to Low (0 V). By counting the signals of the pump function control there is a possibility to estimate the delivered quantity of
lubricant and the remaining quantity in the lubricant reservoir (delivery rate per cycle =0,15 cm³).
Factory setting:
Alternative:
F 1 = Feedback “on”
F 0 = Feedback “off”
Operating Instructions: LUB‐M
20
5.8 Change operating mode: Time control / pulse control
Basic setting:
Time control Pu0 = Time control „on“, pulse control „off“
ON appears on the control panel and activated pump units flashing in sequence (indicated by a green LED).
Alternative:
Pulse Control Pu1 = Pulse control „on“, time control „off“
PUL appears on the control panel and installed pump units flashing in sequence (indicated by a green LED).
Changing from „time control“ to „Pulse control“:
 Select Pu0 Function
 Confirm with ENTER
 Change Pu0 to Pu1 via SELECT
 ENTER (Display is blinking 2 times)
Lubricus M ist changed to pulse control.
„PUL“ is shown on the Display.
Example Pulse control (5 installed pump units)
Operating Instructions: LUB‐M
21
Time control
Factory setting: Lubricus M works in the „time modus“.
6. pulse controlled lubrication intervals
6.1 Default indication control panel (pulse control)
Basic menu in control mode:
Pulse control
PUL:
Pro:
ESC:
pulse control
Other functions, PIN code protected
exit and return to the main menu
The „pro program“ of the pulse mode contains the following editable options:
Enter PIN
ESC: exit the menu
CLr: delete error
messages
Change operating
mode (time/pulse
FIL: Deaerate the
pump
Change Feedback
Operating Instructions: LUB‐M
Change PIN
22
6.2 Enter PIN: Access to Pro Program is protected (PIN)
Selection of other menu items with PIN entry possible.
If no interaction takes place for a period,
the device automatically returns to the
previous display mode
The settings are performed with „ENTER“ and „SELECT“.
The factory setting is: 000 (Master PIN siehe Service)
6.3 CLr – delete error messages and delete filling cycles
With Confirmation of „CLr“ error messages will be deleted.
In „FIL“-Modus the filling operation will be stopped with „CLr“.
Operating Instructions: LUB‐M
23
Pulse control
 SELECT: Change settings
 ENTER: Confirmation of all changes
6.4 FIL – first start-up, venting the pump und filling of lubrication lines
Attention: Lubricus M lubrication pumps must be filled at initial operation.
Each pump unit must be filled separately. The operation is completed when lubricant is visible at the outlet.
Pulse control
First start-up:
Attention!
Prefill housing inlet with 50 ml of suitable lubricant (same grease
type as content of cartridge)
Attention: Low level indicator in the inlet must not be damaged.
Then place grease cartridge and start FIL function.
 Select „FIL“
 Confirm with ENTER
 Select pump body
 ENTER
Delete with CLr (Clear)
FIL means that the outlet of each pump unit is 15 times active.
Total time of FIL / pump unit: about 5 minutes
This function can also be used to pre-fill tubes.
15 cycles with 0,15cm³ = about 2 cm³
Delete with CLr (Clear)
Select other pump units with:
 Pro
 PIN
 FIL
or timeout => exit the menu
Operating Instructions: LUB‐M
24
6.5 Change PIN , factory setting PIN: 000 (Master PIN siehe Service)
In the menu point „PIN“ you can change the PIN for the access of the PRO-Menues.
Pulse control
Example: Change PIN to 222
The new PIN is saved.
6.6 5.7 Change Feedback (Pump function control)
Feedback = Pump function control: During the function of the pump (about 7 s) the output signal at PIN 4 is switching from High (+
20...30 VDC) to Low (0 V). By counting the signals of the pump function control there is a possibility to estimate the delivered quantity of
lubricant and the remaining quantity in the lubricant reservoir (delivery rate per cycle =0,15 cm³).
Factory setting:
Alternative:
F 1 = Feedback “on”
F 0 = Feedback “off”
Operating Instructions: LUB‐M
25
6.7 Change operating mode: Time control / pulse control
Factory setting: Lubricus M works in the „time modus“.
Time control Pu0 = Time control „on“, pulse control „off“
ON appears on the control panel and activated pump units flashing in sequence (indicated by a green LED).
Alternative:
Pulse Control Pu1 = Pulse control „on“, time control „off“
PUL appears on the control panel and installed pump units flashing in sequence (indicated by a green LED).
Changing from „pulse control“ to „time control“:
 Select Pu1 Function
 Confirm with ENTER
 Change Pu1 to Pu0 via SELECT
 ENTER (Display is blinking 2 times)
Lubricus M ist changed to time control.
„ON“ is shown on the Display.
Example time control (5 installed pump units)
6.8 ESC – Exit the menu
Confirm „ESC“ with "ENTER" => exit to main menu.
 SELECT: Change settings
 ENTER: Confirmation of all changes
If no interaction takes place for a period,
the device automatically returns to the
previous display mode
Operating Instructions: LUB‐M
26
Pulse control
Basic setting:
7) Pulse signals to activate the outlets
Indication in seconds, accuracy ± 0,2 s, time between two signals > 20 s
7.1 Identification pump units and outlets
 green LED shows active pump
Operating Instructions: LUB‐M
27
Pulse control
Remark: During receiving a correct pulse signal at PIN 2, the display “PUL” flashes for confirmation. The green LED’s (indicating the
pump units) are not now active.
7.2 Pulse signals pump unit 1
7.2.1 Pulse signals pump unit 1 – outlet 1 or outlet 2
Delivery rate per pulse signal: 0,15cm³, indication: 2s, start of delivery rate at outlet 1 or outlet 2.
Outlets will be served alternately.
Pulse control
7.2.2. Pulse signals pump unit 1 – outlet 1 or outlet 2
Delivery rate per pulse signal: 0,15cm³, indication: 2s, time between two signals > 20 s, start of delivery rate at outlet 1 or outlet 2.
outlets will be served alternately.
7.3 Pulse signals pump unit 2
7.3.1 Pulse signals pump unit 2 - outlet 1 or outlet 2
Delivery rate per pulse signal: 0,15cm³, indication: 4s, start of delivery rate at outlet 1 or outlet 2.
Outlets will be served alternately.
7.3.2 Pulse signals pump unit 2 - outlet 1 and outlet 2
Delivery rate per pulse signal: 0,15cm³, indication: 4s, time between two signals > 20 s, start of delivery rate at outlet 1 or outlet 2.
Outlets will be served alternately
Operating Instructions: LUB‐M
28
7.4 Pulse signals pump unit 3
7.4.1 Pulse signals pump unit 3 – outlet 1 or outlet 2
Delivery rate per pulse signal: 0,15cm³, indication: 6s, start of delivery rate at outlet 1 or outlet 2.
Outlets will be served alternately.
Pulse control
7.4.2 Pulse signals pump unit 3 – outlet 1 and outlet 2
Delivery rate per pulse signal: 0,15cm³, indication: 6s, time between two signals > 20 s, start of delivery rate at outlet 1 or outlet 2.
Outlets will be served alternately.
7.5 Pulse signals pump unit 4
7.5.1 Pulse signals pump unit 4 – outlet 1 or outlet 2
Delivery rate per pulse signal: 0,15cm³, indication: 8s, time between two signals > 20 s, start of delivery rate at outlet 1 or outlet 2.
Outlets will be served alternately.
7.5.2 Pulse signals pump unit 4 – outlet 1 and outlet 2
Delivery rate per pulse signal: 0,15cm³, indication: 8s, time between two signals > 20 s, start of delivery rate at outlet 1 or outlet 2.
Outlets will be served alternately.
Operating Instructions: LUB‐M
29
7.6 Pulse signals pump unit 5
7.6.1 Pulse signals pump unit 5 – outlet 1 or outlet 2
Delivery rate per pulse signal: 0,15cm³, indication: 10s, time between two signals > 20 s, start of delivery rate at outlet 1 or outlet 2.
Outlets will be served alternately.
Pulse control
7.6.2 Pulse signals pump unit 5 – outlet 1 and outlet 2
Delivery rate per pulse signal: 0,15cm³, indication: 10s, time between two signals > 20 s, start of delivery rate at outlet 1 or outlet
Outlets will be served alternately.
Remark: A new pulse signal can be started sooner if the feedback signal (F 1 = feedback „on”) is analyzed.
Requirement: After reaching normal pump running there is a clear High signal for at least 5s.
Operating Instructions: LUB‐M
30
7.7 pulse control - acceptable Design
The specified lubricant quantity refers to the expected life of a piston pump or a pump unit. A service for the verification of the
performance of Lubricus M is recommended. Service interval is 65,000 cycles per pump unit. 65,000 cycles correspond to 10.000
cm³ lubricant. See service: Readout of the device memory.
If a higher demand for lubricant at one lubricating point is required, the outlets of different pump units can be combined
externally.
Please note!
Unused outlets must remain open to avoid damage to the pump!
Examples:
1 outlet supplying 1 lubrication point:
Allowed amount of lubricant for 1 lubrication point =
max. 5,000 cm³ at 50 bar back pressure
2 outlets supplying 1 lubrication point:
Allowed amount of lubricant for 1 lubrication point =
max. 10,000 cm³ at 50 bar back pressure
4 outlets supplying 1 lubrication point:
Allowed amount of lubricant for 1 lubrication point =
max. 20,000 cm³ at 50 bar back pressure
Operating Instructions: LUB‐M
31
8) Communication interface (M12x1 connector) Lubricus M
Connector PIN assignment M 12 x 1
PIN 1:
PIN 2:
PIN 3:
PIN 4:
Input voltage + 24 VDC (+/- 5%), stabilized operating voltage 24
VDC, color brown
Input pulse signal to activate pump unit, color white
Ground (GND), color blue
Output Signal, color black
Details (all data refer to 24 VDC):
Peak current Imax (during pump operation), approx. 350 mA, typically < 200
mA, Standby current (standby mode) < 50 mA, typically 20 mA
The outlet signal (PIN4) can be charged with max. 100mA
The peak current increases by the sampled output current: for example
350 mA + 100 mA = 450 mA
Output signal (PIN 4): High (24VDC) = OK, LOW (0 V) = Error
Warning: The exit is not short-circuit proof.
A time delay fuse 1A is normally required.
Operating Instructions: LUB‐M
32
9) Errors, Messages
Malfunctions will be recognized by the Lubricus electronics and shown on the display.
General:
Display on the control panel + LED
Mode time control:
Details appear on the control panel and activated pump units flashing in sequence
(indicated by a green LED).
Mode pulse control:
Details appear on the control panel and installed pump units flashing in sequence
(indicated by a green LED).
9.1 Error message E1: Level indicator (output signal PIN 4 = Low (0 V))
Display:
ALARM LED, Display E1
Cause:
Lubricant cartridge is empty or missing. Pumping is stopped!
Remedy:
Replace lubricant cartridge
9.2 Error message E2: Red locking pin is missing (output signal PIN 4 = Low (0 V))
Display:
ALARM LED, Display E2
Cause:
Red locking pin is missing. Pumping of all pump bodies is stopped
Remedy:
Put in the locking pin at the bottom of the housing. The error message is cleared automatically.
Pump will continue to run as per active program.
Operating Instructions: LUB‐M
33
9.3 Error message E3: Pump unit works too slowly (output signal PIN 4 = Low (0 V))
Display:
ALARM (red LED) is flashing, Display: Number of pump unit + E3 = 2E3
Cause:
Low-voltage, pump unit does not work in specified time. Pumping is stopped!
Remedy:
Eliminate cause, delete error in the program Pro with Clr. Pump will continue to run.
Example error message E3 time control (5 activated pump units)
Example error message E3 pulse control (5 activated pump units)
9.4 Error message E4: Internal electrical fault (output signal PIN 4 = Low (0 V))
Display:
ALARM (red LED) is flashing, Display: Number of pump unit + E4 = 2E4
Cause:
Internal electrical fault. Pumping is stopped!
Remedy:
Delete error in the program Pro with Clr. Pump will continue to run. After another error message E4 send the
pump for repair.
Example error message E4 time control (5 activated pump units)
Example error message E4 pulse control (5 activated pump units)
Operating Instructions: LUB‐M
34
9.5 Error message E7: Back pressure too high/over-current (output signal PIN 4 = Low (0 V))
Display:
ALARM (red LED) is flashing, Display: Number of pump unit + E7 = 2E7
Cause:
Back pressure was measured 3 x too high. The lubrication point could be clogged, the tube length could be
too long or the grease is too stiff or has hardened. Pumping is stopped!
Remedy:
Delete error in the program Pro with Clr. Pump will continue to run. After another error message E4 send the
pump for repair.
Example error message E7 time control (5 activated pump units)
Example error message E7 pulse control (5 activated pump units)
10) Master PIN: 485
The Master PIN – 4 – 8 – 5 - can be entered to access the Pro program.
Operating Instructions: LUB‐M
35
11) Service: Lubricus M
Lubricus M ist is designed for minimum quantity lubrication.
A service for the verification of the performance of Lubricus M recommended. Service interval is 65,000 cycles per pump unit.
65,000 cycles correspond to 10.000 cm³ lubricant.
11.1 Readout of the device memory
The number of cycles can be determined by reading the device memory:
Touch for 5s the operating pad (ENTER) with the programming key. Then remove the programming key. The display shows in sequence
the number of cycles for every installed pump unit. The cycles are counted to 65,535.
.
P1 = 17.853 cycles
Thereafter an overflow occurs. Number 1 appears in the display. The cycles are counted again, starting with number 1. 65.535 must be
added to the displayed number. A service of the pump unit is strongly recommended in order to ensure the performance of Lubricus M.
The only other service required is the change of grease cartridge.
Operating Instructions: LUB‐M
36
11.2 Service: Lubricus M – replacement of grease cartridge
1. Observe the safety instructions
2. Remove red locking pin (bottom of the housing)
3. Remove transparent cover:
Attention: The internal components of the transparent cover are spring-loaded.
Operating Instructions: LUB‐M
37
4. Remove the green actuation magnet and lock press plate with red locking pin.
Attention: The internal components of the transparent cover are spring-loaded.
5. Remove grease cartridge
Operating Instructions: LUB‐M
38
6. Remove protection from the new cartridge
7. Place the cartridge on the inlet, install the prepared transparent cover
8. Complete the installation of the housing by tightening the screws to a torque of 5 Nm, remove the locking pin, reinstall actuation magnet.
Operating Instructions: LUB‐M
39
9. Reinstall locking pin: Attention: Pump will not function until the locking pin is installed!
10. The error message is cleared automatically
11. Fill if required (see section FIL)
12. Pump will continue to run as per active program
11.3 Service: Lubricus M – filling with oil
Storage/volume Lubricus M (with oil): 2000 cm³ oil reservoir
1. Observe the safety instructions
2. Remove the programming key (turn clock wise to OPEN)
3. Fill Oil
4. Reinstall programming key (turn counter clockwise to CLOSE)
5. The error message is cleared automatically
6. Fill if required (see section FIL)
7. Pump will continue to run as per active program
Operating Instructions: LUB‐M
40
12) Mounting Lubricus M
2 screws (not supplied) Ø 10mm (for example, M10x80 or longer) are required to install your Lubricus M.
13) Technical data Lubricus M
Lubrication volume:
Lubricant type:
Functional principle:
Dose volume per stroke:
Number of outlets:
Connection:
Operating pressure:
Operating voltage:
Operating temperature:
Dimensions, max., W x H x D
Weight without lubricant:
Integrated controller:
Pressure monitoring:
Level monitoring:
Connector:
Protection class:
Progressive distributor control:
Material housing:
1500cm³ grease cartridge
grease up to NLGI 2
Piston pump
0,15 cm³
max. 10
high pressure tube Ø 6mm
max. 70 bar, 1.000 psi
24 VDC
-20°C up to +70ºC
Width x height x depth, 160 x 255 x 165 mm (without tube fitting)
approx. 4000g
Microelectronic
Integrated, electronic
Integrated, dry reed contact
M12x1, 4-pole
IP 54
Suitable
Aluminium
14) Disposal
Please note!
When changing lubricants, the waste disposal instructions of the lubricant manufacturer must be observed!
Regional applicable laws and regulations are to be observed when disposing of the Lubricus Lubrication System.
See your applicable national/regional disposal guidelines for more information.
Emptied cartridges contain lubrication residue!
Please dispose of together with other oleaginous waste!
Gruetzner GmbH
Kohlenhofstr. 60
90443 Nuremberg, Germany
info@G‐LUBE.com
www.lubricus.en
Tel.: +49‐911‐277 399 0
Fax: +49‐911‐277 399 99
Operating Instructions: LUB‐M
41

advertisement

Key Features

  • Compact design
  • Disposable cartridge
  • Up to 5 pump units
  • 2 outlets per unit
  • Time/Pulse control
  • Back pressure control
  • Error messaging
  • Communication interface

Frequently Answers and Questions

How do I change the lubrication interval on the LUB-M?
You can adjust the lubrication intervals using the actuation magnet. The device allows for both time-controlled and pulse-controlled intervals. Refer to the Pro Program section of the manual for detailed instructions.
What are the different types of lubricants that can be used with the LUB-M?
The LUB-M is compatible with both oil and grease, specifically up to NLGI class 2. Use only original Lubricus spare parts and accessories.
How do I know the LUB-M is working properly?
The LUB-M has a built-in back pressure control system that monitors the pressure between the pump and the lubrication point. This helps to ensure that the system is working correctly and that the lubrication points are receiving the correct amount of lubricant.

Related manuals

Download PDF

advertisement